filter integrity testers for the pharmaceutical industry integrity testers for the pharmaceutical...
TRANSCRIPT
Filter Integrity Testers for the Pharmaceutical Industry
Inline Filter Integrity Tests on Filters on huge Vessels, Freeze Dryers, Autoclaves, Filling -and Pressure Lines
1
[email protected] www.campilo.net
Filter Integrity Testers for the Pharmaceutical Industry
The portable Filter Integrity Tester "it-01"
The Filter Integrity Tester " it-01" can be used for integrity tests on membrane filters. In addition to integrity tests it is also possible to check unknown volumes (e.g. the upstream volume of a filter
housing).The Instrument can also be used as a Calibration Pressure Gauge (Range 0-4 bar, 0-6 or 0-8bar)
Membrane Filter Integrity Tests :� Diffusion Test (Forward-Flow)� Pressure Drop Test� Diffusion and Bubble Point Test� Diff.Tests with automatic filter wetting� Water Intrusion Test (WIT)� Water Intrusion Test on huge housings (150L)� WIT Tests with automatic filter drying� WIT Tests with automatic housing cooling� Leakage Tests on sterile bags and containers
Volume Measurement:� Volumes Checks from 0.1 to 32L
"it-01" is small and portable and can be easily used in inaccessable places e.g. on huge vessels and in autoclave rooms.
Electrical Power: Selectable Battery or Mains driven.
For the testing of small filters "it-01" has a special holder installed on the backside, e.g for small capsules, 50 mm Disposables, mini cartridges and injection filters.
A pneumatic driven valve can be directly controlled by the instrument: e.g for the automatic wetting procedure, filter drying or filling of huge housings.
2
[email protected] www.campilo.net
Filter Integrity Testers for the Pharmaceutical Industry
The Printer is waterproof inside the case behind the cover (No thermo printer!).
A short Operation Instructionis shown on the main plate inside the instrument.
Up to 100 filter test programs can be stored in the instrument`s memory and an unlimited amount on a PC.A software (Windows Xp) for the central administration of filter test programs is available.
FDA Standard 21 CFR 11: A software version for this standard is available.
Up to 40 filter test records can be stored in the instrument (only with CFR-version).
PC-program for the Administration of Filter Test Programs(System Software: Windows>Xp.)
With the software "it-trans" a lot of filter test programs can be created and administrated on a central place.
If there are different instruments are installed in the company , the QA Administrator can create for every intrument the necessary filter test programs. The test programs are loaded in via interface cable. This high securety level guarantes, that the test programs can not be changed and the application is correct.
3
[email protected] www.campilo.net
Filter Integrity Testers for the Pharmaceutical Industry
Explosion proof Filter Integrity Tester "it-01-Ex"
The Filter Integrity Tester " it-01-Ex" is specially designed for the application in areas where an explosive atmosphere is possible (Ex-Zone 1).In addition to filter integrity tests it is also possible to check unknown volumes (e.g. the upstream volume of a filter housing).
The Instrument can also be used as an Calibration Pressure Gauge (Range 0-4 bar)
Membrane Filter Integrity Tests :
� Diffusion Test (Forward-Flow)� Pressure Drop Test� Diffusion and Bubble Point Test� Water Intrusion Test (WIT)
Volume Measurement:� Volumes Checks from 0.1 to 32L
The explosion proof Integrity Tester "it-01-Ex" can be used in Ex-Zone 1 –T4.
"it-01-Ex" is built small and compact, the cabinet is made of stainless steel (1.4301 electro polished).
The "it-01-Ex" is in accordance to the ATEX guidelines. It will be delivered with German TÜV and British BASEFA Certificates.
Electrical Power: "it-01-Ex" needs for operation an ex-proofed mains connector 110 or 220V /AC
Line Pressure: 6 bar air or nitrogen . A separate source for nitrogen protective gas is not needed.
4
[email protected] www.campilo.net
Filter Integrity Testers for the Pharmaceutical Industry
The 2-Channel Filter Integrity Test Station "ita-02-TS" for Membrane Filters
The 2-Channel Filter Test Station carries out the filter integrity tests for two different filters simultanously. All usual tests like Diffusion Test , Bubble Point Test or Water Intrusion Test (WIT) can be used. The liquid filters are automatically wetted during the test procedure. The test records can be collected on a PC or printed out in place. Pressure range 200-6000 mbarg
The 2-channels on the test station "ita-02-TS work independant and simulatanouslye.g At the same time aOn Channel 1 --> Diffusion-Bubblepoint Test On Channel 2 --> Water Intrusion Test
Automatic Filter Wetting ProcedureThe wetting of filter membranes can be optimized (important at PES membranes). The wetting procedure is part of the integrity test.
The Bubble Point Tester "it-BP" for the QA- Laboratory
The Bubble Point Tester "it-BP" can be used for Bubble Point measurements on membrane filters. Caused by the huge measuring range 10 mbar -7000 mbarg also wiremesh filters, diaphragma or ceramic filters can be tested. The instrument is used for the quality comparison of different filter materials.
2 Versions are available:
The low pressure version"it-BP-LP" Measuring Range 10 -1000 mbarFor wiremash filters, diaphragma filters and ceramic filters
The high pressure version "it-BP-HP" Measuring Range 200-7500 mbarFor membrane filters and ceramic filters
5
www.campilo.net
[email protected] www.campilo.net
Filter Integrity Testers for the Pharmaceutical Industry
Calibration, Maintainance and Qualification :
Once a year the sensor and the pressure control system have to be checked and calibrated.
For maintainance and calibration the instrument normally has to be sent back to the manufacturer, a special transportation box is part of the delivery. For countries in the EU or Switzerland the maintainance work can also be carried out in place. All necessary measuring devices are DKD (Deutscher-Kalibrier-Dienst) certified. A continious controlled schedule for calibration of all measuring devices is guaranteed.
Qualification (IQ, OQ,PQ):
If an integrity Tester should be qualified, our qualification service will carry out dthe different parts according the the customers requirements. The qualification documents are created in English or German
6
[email protected] www.campilo.net
Filter Integrity Testers for the Pharmaceutical Industry
Accessories: The Filter Testbox "itb-08"
Up to 8 small disposable filters or capsules can be tested simultaneously in one test run (even injection filters). The filter holders are designed in such a way that most of the standard filters can be easily fixed . For filters with other dimensions, special adapters are available.
The kind of filter testing: All filters are tested in parallel. When using the Bubble Point Test, the filter with the lowest Bubble Point Value determines the test result. For separate tests the filter holder on the instrument can be used.
The complete test run as well as the wetting procedure of the filters is fully automatic.
Installation:
The Filter Test Box "itb-08" is connected on one side to the pressure vessel with the wetting medium and on the other side to the integrity tester "it-01" (see picture ).After teh test is started, the wetting fluid flows for a selected time through the filters. After that time the test run starts and the result is printed out after approx. 5 min.
Remarks:If alcohol is used as weeting medium, and the box is sealed, the alcohol steam cannot leak into the room atmosphere.
7
[email protected] www.campilo.net
Filter Integrity Testers for the Pharmaceutical Industry
Accessories: . The Low Pressure Testbox „itb-LP“ for Leakage Tests on Sterile bags and Containers
Description: The low pressure test box „itb-LP“ is an accessory for the filter integrity tester „it-01“ Controlled by the integrity tester documented leakage tests on strile package bags and containers can be carried out. The leakage test can be done in a test pressure range between 50 and 500 mbar.
How to operate the testbox "itb-LP“ ?The inlet side of the testbox is linked to a line pressure source (1-3,5bar).The filling valve is taken away from the integrity tester. The low pressure testbox is plugged into the same connector On the outlet side of the testbox (Triclamp TC 1,5“) a PTFE sterile filter capsule is mounted. On the sterile side of the capsule the sterile bag is fixed. The test program for the specific leakage test is started on the integrity tester. When the test is over, the sterile bag is emptied by the use of an inbuilt vakuum pump.
Specifications:
Cabinet Stainless Steel 1.4301
Cabinet Dimensions 200x300x80 mm
Weight 2,5Kg
Pressure Line Connector.
Male Stäubli RBE 03 (1.4404)
Press. SensorConnector
Male Stäubli RBE 03 (1.4404)
Press. Outlet Triclamp TC 50,5mm
Line Pressure: 0,5- 3 bar
Test Pressure 50-500 mbar
Protection Classe IP 65
8
[email protected] www.campilo.net
[email protected] www.campilo.net
Filter Integrity Testers for the Pharmaceutical Industry
Specifications:
Housing material Stainless steel 1.4301
Housing size (without handhold, and holders) 200x300x155
Weight: 8,6 kg
Protection classe: IP 55
Power Supply: Battery 7,2 V/ 3,8Ah (oder Mains)
Keyboard: Infrared,alpha numeric (water protected)
Software protection: Stored in flash EPROMs
Matrix printer: Normal paper, 24 characters/ line
Amount of storable test programs: 100
Kind of storage Flash EPROMs
Pneumatic connectors: Stäubli RBE 03 / stainless steel 1.4404
Internal pressure sensor : 0-6000 mbar; rel. Class l 0.5
Barometric pressure sensor: 800-1090 mbar Class 0,5
External pressure sensor: 0-6000 mbar, rel. Class.0,15
Main CPU -board: HC-12 Motorola
Program language: C
Display: 4 lines illuminated - 20 characters/line
Data transfer: RS 232 (19.200 bd)
Languages: German, English, French, Italian, Spanish, Danish,
Measuring units:
Pressure : mbar
Water flow : µL/ t (test time)
Diffusion: mL/min (1000 mbar /20°C)
Volume: mL
Possible integrity tests: Water Intrusion Test (WIT)Diffusion Test, Pressure Drop TestBubble Point Test , Diff.+Bubblepoint TestVolume Measurement
Cleaning of pneumatic parts: Steaming of the external sensor and Filling valve
Flushing of the internal valves (80°C)
Measuring Accuracy
WIT: 4% of actual value >= 0.1mL/minmax.resolution 0,008mL
Diffusion: 5% of actual value >= 5mL/minmax.resolution 0,1mL
Bubble Point 1% F.S.
Environment temperature: 5- 40°C
Storage temperature: 2- 50°C
Rel. huminity: 0-80%
9
[email protected] [email protected] www.campilo.net
Filter Integrity Testers for the Pharmaceutical Industry
Installation: Water Intrusion Test (WIT)
10
Gas-Fi l ter
STOP
START
1 2 3
4 5 6
7 8 9
0FUNCTI ON ENTER
OFF ON VOLUME
i t -01
Bat t .
6L Press.Vessel
Sterile Area
Pressure L ine 6-8bar
Fi l l ing
Valve
STOP
START
1 2 3
4 5 6
7 8 9
0FUNCTION ENTER
OFF ON VOLUME
i t -01
Bat t .
Installation: Water Intrusion Test (WIT)
Pressure Line
6-8bar
STOP
START
1 23
4 5 6
7 8 9
0FUNCTION ENTER
OFF ON VOLUME
i t -01
Filling Valve
Pressure Line6-8bar
Installation: Water Intrusion Test (WIT)with Cooling
Press. vessel
(i t-006L)
Press.
Sensor Press. Sensor
Press.
Sensor
OverflowValve
Filling Valve
In let
Gas -Fi l ter
Vent Fi l ter
on a T ank
Inlet
Installation:"Remote Controlled"Water Intrusion Test
[email protected] www.campilo.net
Filter Integrity Testers for the Pharmaceutical Industry
Installation:: Diffusion- Bubble Point Test and Volume Mesurement
11
STOP
START
1 2 3
4 5 6
7 8 9
0FUNCTI ON ENTER
OFF ON VOLUME
i t - 01
Bat t .
Fi l l ing Valve
Liquid Fi l ter
Pressure
Sensor
In let
Pressure L ine
6-8bar
Installation:Diffusion TestBubblepoint Test STOP
START
1 2 3
4 5 6
7 89
0FUNCTI ON ENTER
OFF ON VOLUME
i t -01
Inlet
Ref erence
Volume
--> it-350
--> it-800
Pressure
Line 6-8bar
Filling Valv e
Liquid Filter
Volume Measurement
STOP
START
1 2 3
4 5 6
7 8 9
0FUNCTI ON ENTER
OFF ON VOLUME
i t -01
Bat t .
Press. Sensor with Valve Group
Inlet
10L Pressure Vessel for the wetting medium
(it-010L)
Press. Regulator
0.5-->0.8 bar)
Vent-Nozzle
Pressure L ine
6-8bar
Pneumatic Valv e for filter wetting
Installation:Diffusion TestBubblepoint Testwith Filter Wetting
Filter Integrity Testers for the Pharmaceutical Industry
2-Channel Filter Integrity Test Station for wetting and testing of Membrane Filters
12
[email protected] www.campilo.net