fh-s•thread:m6 •rechargeable battery pack integrated, standard, operating time up to 12 h...
TRANSCRIPT
Measuring system
Measuring range [Max]: 2 N
Overload protection (of [Max]): 150 %
Precision (of [Max]): 0,005
Readout [d]: 0,001 N
Test frequency: 2000 Hz
Display
Display digit height: 1 cm
Functions
Data hold function: yes
Function to set limits: yes
Obtaining average: yes
Peak function: yes
Statistical function: yes
Environmental conditions
Maximum operating temperature: 40 OC
Minimum ambient temperature: 10 OC
Power supply
Charging time: 4 h
Input voltage: 100 V - 240 V
n/a: CH EURO
Power supply
Operating time (Backlight off): 12 h
Service
DAkkS Calibration certificate 963-261
(Pressure):
DAkkS Calibration certificate 963-161
(Tensile):
DAkkS Calibration certificate 963-361
(Tensile/Pressure):
ISO calibration certificate: 961-361
ISO Calibration certificate (Pressure): 961-261
ISO Calibration certificate (Tensile): 961-161
ISO Calibration certificate (Ten- 961-361
sile/Pressure):
Category
Brand:
Category: Force gauges
Product Group: Electronic force-measuring
device
Packaging & shipping
Delivery: 24 h
Dimensions packaging (WxDxH): 312 x 170 x 60 mm
Gross weight: 1,300 kg
Net weight: 640 g
Shipping method: Parcel service
UN number: 3496
Construction
Dimensions of display device
(WxDxH):
230 x 66 x 36 mm
Overall dimensions mounted
(WxDxH):
230 x 66 x 36 mm
n/a: Mains adapter external
1
X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com
FH-S X-SENSORS GmbH
Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung
01
OPTION
Model
Measuring range
[Max] N
Readout
[d]
N
Option ISO Calibration Certificate Tension Compression Tension/Compression
FH 2. 2 0,001 961-161 961-261 961-361
FH 5. 5 0,001 961-161 961-261 961-361
FH 10. 10 0,005 961-161 961-261 961-361
FH 20. 20 0,01 961-161 961-261 961-361
FH 50. 50 0,01 961-161 961-261 961-361
FH 100. 100 0,05 961-161 961-261 961-361
FH 200. 200 0,1 961-161 961-261 961-361
FH 500. 500 0,1 961-161 961-261 961-361
Digital force gauge FH-S
Universal digital force gauges for tension and compression tests
with integrated measuring cell and RS-232 data interface
Features
•Turnable display with backlight
• Can be mounted on all test stands
•Digital force gauge with internal sensor
•Data interface RS-232, included
• Standard attachments: as shown below,
extension rod: 90 mm
• Delivered in a hard carrying case
•Selectable measuring units: N, lb, kg •Peak-Hold function to capture peaks (measure-
ment result will be “frozen” for a short time) or
Track function mode for a continuous measure-
ment indication (period of time approx. 10 s)
•Function to set limits, programming of Max./
Min., in pull and push direction, with output
of acousitc and optical signal. Ideal mode for
efficient and accurate testing of standard parts
•Auto-Power-Off
STANDARD
•Internal memory for up to 10 measurements
•Mini Statistics Kit: calculates the average
result from up to ten stored single results,
min., max., n
Technical data
•High resolution: up to 10,000 points
(total measuring range)
•Internal measuring frequency: 2000 Hz
•Precision: 0,5 % of [Max]
•Overload protection: 150 % of [Max]
•Dimensions WxDxH 66x36x230 mm
•Thread: M6 •Rechargeable battery pack integrated,
standard, operating time up to 12 h without
backlight, charging time approx. 4 h
•Net weight approx. 0,64 kg
Accessories
•Relais module, serves to amplify the output
signal of the dynamometer to control direct
actions, AFH-02
•Force-time data transfer software for
graphical representation on the PC and data
transfer to Microsoft Excel,
AFH FAST •Force-displacement data transfer software
with graphic display of the measurement
process, AFH FD
•Thermal printer, YKB-01N
• Standard attachments, AC 43
•Further accessory see www. .eu and
page 26 et seqq.
X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com
FH-S X-SENSORS GmbH
Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung
Adjusting program (CAL):
For quick setting of the balance’s accuracy.
External adjusting weight required.
Calibration block:
standard for adjusting or correcting
the measuring device.
Peak hold function:
capturing a peak value within a
measuring process.
Scan mode:
continuous capture and display
of measurements.
Push and Pull:
the measuring device can capture tension
and compression forces.
Length measurement:
captures the geometric dimensions of
a test object or the movement during
a test process.
Focus function:
increases the measuring accuracy of a
device within a defined measuring range.
Internal memory:
to save measurements
in the device memory.
Data interface RS-232:
bidirectional, for connection
of printer and PC.
Data interface USB:
To connect the balance to a printer,
PC or other peripheral devices.
Data interface Infrared:
To transfer data from the balance to a
printer, PC or other peripheral devices.
Control outputs
(optocoupler, digital I/O):
to connect relays, signal lamps, valves, etc.
Analogue interface:
to connect a suitable peripheral device for
analogue processing of the measurements.
Statistics:
using the saved values, the device
calculates statistical data, such as
average value, standard deviation etc.
PC Software:
to transfer the measurements from
the device to a PC.
Printer:
a printer can be connected to the
device to print out the measurements.
GLP/ISO record keeping:
of measurements with date, time and
serial number. Only with printers.
Measuring units:
Weighing units can be switched to
e.g. non-metric at the touch of a key.
Please refer to website for more details.
Measuring with tolerance range:
Upper and lower limiting can
be programmed individually,
e.g. for sorting and dosing.
ZERO:
Resets the display to “0”.
Battery operation:
Ready for battery operation.
The battery type is specified
for each device.
Rechargeable battery pack:
rechargeable set.
Mains adapter:
230V/50Hz in standard version for EU.
On request GB, AUS or USA version
available.
Power supply:
Integrated, 230V/50Hz in EU.
More standards e.g. GB, AUS
or USA on request.
Motorised drive:
The mechanical movement is carried out
by a motorised drive.
Fast-Move:
the total length of travel can be covered
by a single lever movement.
ISO Calibration:
The time required for ISO calibration
is shown in days in the pictogram.
Package shipment:
The time required for internal
shipping preparations is shown
in days in the pictogram.
Pallet shipment:
The time required for internal
shipping preparations is shown
in days in the pictogram.
Warranty:
The warranty period is shown
in the pictogram.
Pictograms:
X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com
FH-S X-SENSORS GmbH
Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung
GmbH • Ziegelei 1 • 72336 Balingen • Germany • Telephone + 49-7433-9933-0 • Fax + 49-7433-9933-149 • www. -sohn.com • info@ .eu AFH_02-BA-e-1410
User's manual
AFH-02
AFH-02 relay module
Machine control by means of control outputs
Features
•In order to control the machine operations with the use of a dynamometer it is necessary to process the dynamometer control signals in terms of the machine control system.
•As the output signals of FH dynamometer are too low for direct control of the
operations, a relay module is used for their processing which allows for
operation with constant/variable voltage from 20 to 35 V.
•The dynamometers of FH series of company make it possible after setting
the values in the device to control various outputs. It is possible to control 3
outputs. 1. the measurement result is lower than the lower limit value (LoLt option in FH dynamometer menu), 2. the measurement result is higher than the upper limit (HILt option in FH dynamometer menu), 3. the measurement result exceeds Stop value (StoP option in FH dynamometer menu).
• The system of the relay module connections allows setting 4 control options: - value below the lower limit,
- value is OK (between the lower and upper limit), - value above the upper limit, - exceeding of the Stop value.
•The relay module is intended to be fixed on 35/15 mm rails which conform
to DIN 50022 standard as well as for universal assembly in the control cabinet.
•The relay module is equipped in 4 relays, each with one potential-free
change-over.
•The dynamometer is connected to the relay module with the use of SUB-D plug-in socket .
• The scope of delivery : a relay module, a connection cable ca. 1.5m long.
Assignment of terminals
Break contact Common potential- free contact
Make contact
Lower limit value
Ö1 C1 S1
OK Ö2 C2 S2
Upper limit value
Ö3 C3 S3
Stop Ö4 C4 S4
Technical data
Input voltage Relay connection voltage Dimensions
from 20 to 35 V, constant/variable voltage 250 V/10 A 90 × 77 × 41 mm
Compatibility with the equipment software of FS dynamometer from S 5.28 version with full range of functions, below with limited operation (no useful 'Stop' signal).
X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com
FH-S X-SENSORS GmbH
Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung
Thank you for buying a force gauge. We hope you are
pleased with your high quality force gauge with its big
functional range. If you have any queries, wishes or helpful
suggestions, do not hesitate to call our service number.
„Sensor Inside“ means the measuring cell is inside the
housing.
1. Included in delivery
- FH incl. rechargeable battery
- Carrying Case
- Charger
- Standard Attachments as shown below,
5 pcs M3 x 8 screws
Important annotation:
By pressing the RESET key (on the right side of housing,
see illustration on the left), individual settings and
memorised values can be re-set or erased, in example for
a new start of the instrument after an operating error.
Data in mm
2. Working Conditions
10°C to 30°C / 15% up to 80% humidity
3. Electrical Power Supply
Either by rechargeable battery or current power supply
Current power supply:
are charged
- Connection by power adapter
- Rechargeable batteries
simultaneously
Rechargeable battery pack for mobile applications:
- Type: Ni 8.4V / 600 mAh
- Charging time: approx. 1 hour
Instruction Manual
FH
Sensor
Inside
Adaptersockel
Re Set Taste
Set
LCD
Grenzwert Anzeige
RS 232
Kraftaufnehmer
Power plug
Reset button
Se
Display
Pilot lamp
RS-232 port
Force Sensor
FH_SI-BA-e-1317 1
X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com
FH-S X-SENSORS GmbH
Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung
Instruction Manual
FH
4. Technical Data
- Accuracy: ± 0,5 % of Capacity (measuring range)
- Data Sampling Rate: 2.000 Hz
- Weight: 640 g
5. Operation
Display
(1) Measuring Result
(2) Measuring Units
(3) Activation of PRINT Function
(4) Indication of power charging status
(5) PEAK or AUTO-PEAK Mode
(6) Average value of stored peak values
(7) Force direction
(8) Occupancy of storing spaces
(9) AVERAGE- or Saving Mode
Operation keys
ON / OFF:
ON / OFF key
(For ON, press 1 sec.)
UNIT:
- Press shortly:
- Press for 2 sec.:
Select unit:
N, kg or lb
Display return
ZERO:
Three functions:
- Zeros the measuring result (Tara function)
- Cleans the peak value (in Peak mode)
- Saves a setting (in SET mode)
SET:
- 1 x Press: Upper Limit [HidT / Hi Lt]. To
change press: ▲ or ▼ (see section
High/Low Limit function)
- 1 x Press: Lower Limit [LodT]. To change
press ▲ or ▼ (see section High/Low Limit
function)
- 1 x Press: Minimum limit to save peak values
in the instrument’s memory. [LE.SET / Lo
Pe]. Please see section „Min Limit Peak
Save“. To change press: ▲ or
▼ (Only active in „Peak-Mode“) (see
section min. limit peak save)
- 1 x Press: Sets the Auto-STOP force value
that stops the movement of a test stand.
Adjustment of the force value. To change
press: ▲ or ▼ (only active in „Battery-
Mode“)
1 x Press: Auto-OFF Function. Turns the
instrument off after a here defined time
period in sec. [P.OFF]. To change press: ▲
or ▼ (Only active in „Battery-Mode“)
- 1 x Press: Peak-Freeze-Time [PE.2E / A.PE /
HoldT]: Time period (in sec.) in which a
peak value is being shown in the display in
sec. To change press: ▲ or ▼
- 1 x Press: Data Transfer [rS232] to PC (PC) or
to the printer (Print) or (in version U 5.1) to
the test stand
- PC: Pressing SET once: Saves the Settings
- PRINT: Pressing SET twice: sending data to the printer
- STAND: Pressing SET three times: sending a signal to
the Test Stand to stop its movement (in version U 5.1)
BACK LIGHT:
Hi Lt
Lo dT
Lo Pe
P.OFF
HoldT
rS232
StoP
FH_SI-BA-e-1317 2
X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com
FH-S X-SENSORS GmbH
Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung
- Track mode (continuous measurement)
- Peak mode (capture of maximum values)
- Auto-Peak mode, same as Peak-mode, only
without the „Min limit peak save“ function
MEMORY:
Saves the peak values to calculate the average
value (please see section “Memory mode”
DELETE Function
Deletes stored peak values (only in „Memory
mode“ active)
PRINT:
Sends the stored peak values to a PC or Printer
(please see section „Data Output”)
High / Low limit function
LEDs to display OK / NOT OK Tests
▼
●
▲
Lower than lower limit
Lights if the STOP value is reached
Higher than highest limit
This function allows efficient testing of OK / NOT OK’S
measurements of similar or identical testing objects.
A lower and an upper limit value can be defined. The
instrument compares the individual measuring results with
these limit values and shows the OK or NOT OK result by
green or red light diode and by sound.
To set these limit values, please see the SET Menu in
section „Operating keys“.
Measurement (Track Modus)
Display (1) shows the continuous force in a defined
direction (6)
To zero the display, press:
Instruction Manual
FH
PEAK: Peak-Hold Function (Peak Mode)
Please press:
Three functions available:
Auto-Peak-Hold-Function (Auto-Peak Mode)
Please press:
Min. limit peak function to activate storage of
measurements
This function allows to eliminate unwanted „Pre-Peak
values“ that are lower than the main peak value (Fp). The
“Min limit peak save” value (Fo) takes care, that these
“Pre-Peak values” are not saved.
The „Min limit peak save“ function is only possible in Peak-
Mode.
To set this Min limit value, please see the SET Menu in
section „Operating keys“.
Memory mode and average value (from up to 10 peak
values)
Saving peak values in the instrument
Activating the „AUTO PEAK Function“ by PEAK key
Deactivating the „Average Function“ by MEMORY key
Now, all peak values are stored automatically in the
instrument
To browse through the stored values, please use the ▲
or ▼ keys. (The values will be shown in the upper
display segment)
By pressing the MEMORY key, the average value of the
stored peak values can be shown (in the upper display
segment)
To delete every stored value, press the▼-key in the
AVERAGE-Mode
FH_SI-BA-e-1317 3
X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com
FH-S X-SENSORS GmbH
Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung
Instruction Manual
FH
6. Configuration of RS 232
SUB-D 9pm
FH_SI-BA-e-1317 4
Pin Signal Illustration
2 TxD Output signal
3 RxD Input signal
5 GND Ground
6 +1.6 to + 2 V Over upper limit
7 +1.6 to + 2 V Lower than lower limit
8 +1.6 to + 2 V OK
6.1 Output Protocol
RS-232 Parameter
Baud rate: 9600
Data-Bit: 8
Parity: none
Stop-Bit: 1
The measured value is requested by the PC by the ASCII
Sign “9”.
The measured value that comes from the instrument has
this format:
e.g. 0011.70 means -11,70 Newton, if
Newton is the selected unit
||-----|
| | > the other 6 places describe the
measured value as ASCII-Signs
| > the first place describes the
direction of the force (0 = minus = Pressure;
1 = plus = Tension)
or: 1021.15 means +21,15 N (Tension)
| > the first place describes the
direction of the force (0 = minus = Pressure;
1 = plus = Tension)
or: 1021.15 means +21,15 N (Tension)
7. Warning
Intended use
The instrument you have acquired serves to determine the
measuring value of the material to be measured. It is
intended to be used as a “non-automatic“ instrument, i.e.
the material to be measured is manually and carefully
attached at the instrument. The measuring value can be
read off after a stable measurement value has been
obtained.
Inappropriate use
Do not use the instrument for medical measurements. In
the event that small quantities are removed or added to the
material to be measured, incorrect measuring results can
be displayed due to the “stability compensation“ in the
instrument. (Example: Slow draining off of liquid from a
container suspended from the instrument). Do not attach a
continuous load. This can damage the measuring unit as
well as the parts, relevant to safety.
Prevent jolts, torsion and oscillation (e.g. by appending
slopingly) of all kinds. Be sure to prevent overloading the
instrument in excess of the stated maximum load (max.),
minus any tare weight that may possibly exist. This could
damage the instrument (risk of breakage).
Important:
•Always make sure that there are no people or materials
below the load that could be injured or damaged!
•The instrument is not suitable for measuring people. Do
not use as baby scales!
•The instrument does not comply with the medical product
law (MPG).
Never operate the instrument in hazardous locations. The
series design is not explosion-proof. Structural alterations
may not be made to the instrument. This can lead to
incorrect measuring results, faults concerning safety
regulations as well as to destruction of the instrument. The
instrument may only be used in compliance with the
described guidelines. Varying areas of application/ planned
use must be approved by in writing.
Guarantee
The guarantee is not valid in following cases:
•Non-observation of our guidelines in the operating
instructions
• Use outside the described applications
• Alteration to or opening of the device
• Mechanical damage and damage caused by media,
liquids
• Natural wear and tear
• Inappropriate erection or electric installation
• Overloading of the measuring equipment
X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com
FH-S X-SENSORS GmbH
Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung
Instruction Manual
FH
Monitoring the test substances
The metrology features of the instrument and any possible
available adjusting weight must be checked at regular
intervals within the scope of quality assurance.
For this purpose, the answerable user must define a
suitable interval as well as the nature and scope of this
check. Information is available on the home page
(www.SensorShop.com) with regard to the monitoring of
instrument test substances and the test weights required
for this. Test weights and instruments can be adjusted
quickly and at a reasonable price in ’s accredited DKD
calibration laboratory (return to national normal).
Fundamental safety information
Do not use the hanging instrument to transport loads.
Prevent jolts, torsion and oscillation (e.g. by appending
slopingly) of all kinds.
Never use the hanging instrument over the maximum
permitted weight (!! Danger of breaking!!).
Always make sure that there are no living beings or
materials below the load that could be injured or damaged.
The hanging electronic instruments from the instrument
are only suitable for hand-held use or use in a test stand.
They are not suitable for hanging from a mechanical hook,
e.g. a crane hook.
Observe the information in the operating instructions
Please read the operating instructions carefully before
erecting
and commissioning, even if you already have experience
with instruments.
Staff training
The device may only be operated and looked after by
trained members of staff.
8. CE Declaration of Conformity
FH_SI-BA-e-1317 5
X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com
FH-S X-SENSORS GmbH
Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung
Instruction Manual
FH
FH Adjustment Procedure
FH_SI-BA-e-1317 6
1 Start the
instrument
Press ON/OFF The green
light is on
2 Enter the
adjustment
procedure
Directly after pressing
ON/OFF, press PEAK
and PRINT together for
more times and very
short intervals until the
red light is on.
The lower
red light is
on
3 Instrument
Type
Press SET
Directly after pressing
PEAK + PRINT
XXX N is
shown
a (Back to the
normal
mode ???)
(If you are in the
operation mode again,
the instrument has to be
shut off. Start again with
1. Eventually pressing
the buttons faster)
4 Select the
instrument
type
Press ▼▲ The
correspondi
ng value to
the instr.
appears in
the display
5 Save
instrument
type
Press SET
6 Enter the
adjustment
sequence
Press MEMORY The upper
red light is
on
7 Choose the
adjustment
weight
Press UNIT and
with▼▲ specify the
calibration weight in
Newton (X kg * 9,81)
The weight
is written in
the display
8 Save the
adjustment
weight
Press SET and UNIT
simultaneously
9 Attach the
weight to the
instrument
(by hanging)
Press ZERO The
instrument
changes to
operation
mode.
Adjustment
procedure is
finished
X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com
FH-S X-SENSORS GmbH
Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung
FH Schnittstellenbeschreibung
17.05.2016 FH-SCHNITTSTELLEN BESCHREIBUNG DP
Senden vom PC Gerät
ASCⅡ Dezimal Hex
2 50 32 Anzeige auf „Null“ stellen
3 51 33 Einheit umstellen auf "KN"
4 52 34 Einheit umstellen auf "TF"
5 53 35 Einheit umstellen auf "KLBF"
6 54 36 Schaltet Gerät in den Track Modus
7 55 37 Schaltet Gerät in den Peak Modus
9 57 39 Gerät sendet aktuellen wert an den PC
Sendeformat
1. bei der Einheit kN sieht das Daten Byte folgender maßen aus:
bit 1 bit 2 bit 3 bit 4 bit 5 bit 6 bit 7
Kraft-
richtung
data 1
data 2
Dezimal
punkt
data 3
data 4
data 5
Bit 1 Kraftrichtung
1: Kraftrichtung “Zug”
0: Kraftrichtung “Druck”
Wenn das Gerät 0,005 kN in “Zugrichtung misst sieht
das Datenbyte folgender maßen aus:
bit 1 bit 2 bit 3 bit 4 bit 5 bit 6 bit 7
1 0 0 . 0 0 5
Wenn das Gerät 0,005 kN in “Druckrichtung misst sieht das Datenbyte folgender maßen
aus:
bit 1 bit 2 bit 3 bit 4 bit 5 bit 6 bit 7
0 0 0 . 0 0 5
- Baudrate: 9600 - Daten-Bit: 8 - Parität: keine - Stop-Bit: 1
Beispiel:
Der Messwert wird durch das ASCII-Zeichen „9“ angefordert.
Der zurückgelieferte Messwert sieht folgendermaßen aus:
bedeutet -11,70 Newton, wenn Newton als Einheit eingestellt ist z.B. 0011.70
||-----|
| | >
| >
die restlichen 6 Stellen beschreiben den Messwert als ASCII-Zeichenkette
erstes Zeichen beschreibt das Vorzeichen (0 = minus = Druck; 1 = plus = Zug)
oder: 1021.15 bedeutet +21,15 N (Zugkraft)
X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com
FH-S X-SENSORS GmbH
Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung
FH Data Connection protocol
17.05.2016 FH-SCHNITTSTELLEN BESCHREIBUNG DP
Senden vom PC Device
ASCⅡ decimal hex
3 50 32 Set the unit to “zero”
3 51 33 make the unit as "KN"
4 52 34 make the unit as "TF"
5 53 35 make the unit as "KLBF"
6 54 36 make the gauge on the tracking mode
7 55 37 make the gauge on the peak mode
9 57 39 send the value data to pc
format of sending
1. when the unit is " KN", the format sending as:
bit 1 bit 2 bit 3 bit 4 bit 5 bit 6 bit 7
bit of
force
direction
data 1
data 2
point
data 3
data 4
data 5
bit of force direction
1: force direction is pull 0:
force direction is press
if the pull force is 0.005KN, and the unit is "KN" the
format of sending is:
bit 1 bit 2 bit 3 bit 4 bit 5 bit 6 bit 7
1 0 0 . 0 0 5
if the pressure is 0.005KN, and the unit is "KN"
the format of sending is:
bit 1 bit 2 bit 3 bit 4 bit 5 bit 6 bit 7
0 0 0 . 0 0 5
- Baudrate: 9600 - Data-Bit: 8 - Parity: none - Stop-Bit: 1
Example: The measured value requested through sending ASCII-sign „9“ from the PC to the force gauge
The requested data looks like:
z.B. 0011.70 means -11.70 N if the force gauge is set to unit Newton
||-----|
| | > the other 6 digits are the measured value in ASCII
| > the first sign in the byte is the pre sign (0 = minus = comprehension;
1 = plus = Tension)
or: 1021.15 means +21,15 N
X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com
FH-S X-SENSORS GmbH
Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung
Kommandos zum steuern der Prüfstände
17.05.2016 FH-SCHNITTSTELLEN BESCHREIBUNG DP
Commands to manage the test stands
TVM-N Startet Bewegung “AUF” ASCII 7c
TVM-N Startet Bewegung “AB” ASCII 7d
TVM-N STOP ASCII 7e
TVM-N Start movement UP ASCII 7c
TVM-N Start movement DOWN ASCII 7d
TVM-N STOP ASCII 7e