FDR PresentationFDR Presentation
Personal Health Monitoring Personal Health Monitoring DeviceDevice
Bassam Noaman
Dina El-Eissa
Mentor: Prof Zhenyu Guo
Instructor: Prof J Silverman
4/12/00
![Page 2: FDR Presentation Personal Health Monitoring Device Bassam Noaman Dina El-Eissa Mentor: Prof Zhenyu Guo Instructor: Prof J Silverman 4/12/00](https://reader030.vdocuments.us/reader030/viewer/2022033107/56649dba5503460f94aaaf9c/html5/thumbnails/2.jpg)
OverviewOverview
Sensors ECG Blood PressureTemperature
Signal ConditioningFilters, Amplifiers,Logic
PIC Microcontroller program interface
Palm pilotPC
![Page 3: FDR Presentation Personal Health Monitoring Device Bassam Noaman Dina El-Eissa Mentor: Prof Zhenyu Guo Instructor: Prof J Silverman 4/12/00](https://reader030.vdocuments.us/reader030/viewer/2022033107/56649dba5503460f94aaaf9c/html5/thumbnails/3.jpg)
Overall Block DiagramOverall Block Diagram
![Page 4: FDR Presentation Personal Health Monitoring Device Bassam Noaman Dina El-Eissa Mentor: Prof Zhenyu Guo Instructor: Prof J Silverman 4/12/00](https://reader030.vdocuments.us/reader030/viewer/2022033107/56649dba5503460f94aaaf9c/html5/thumbnails/4.jpg)
SpecificationSpecification Power Supply: 9.0 VDC Battery
Operating time 2.8 hours(from Standard 9.0 VDC Duracell battery) Inputs
Body Temperature SensorDetect Temperature in degree CelsiusRange between 25 C – 50.5 C
Blood Pressure Range between 0 – 5 Psi(259mmHg)Accuracy 0.125 @ 25 CManual cuff inflationMaximum systolic pressure 250 mmHg
ECGBandwidth 0.3Hz – 150 HzDetect Heart Rate in bpmOne lead ECG
![Page 5: FDR Presentation Personal Health Monitoring Device Bassam Noaman Dina El-Eissa Mentor: Prof Zhenyu Guo Instructor: Prof J Silverman 4/12/00](https://reader030.vdocuments.us/reader030/viewer/2022033107/56649dba5503460f94aaaf9c/html5/thumbnails/5.jpg)
SpecificationSpecification
• Monitor Performance Specification Palm Pilot screen record and plays the patient’s data file
on the screen PC screen displays the patient data file
Recorded body temperature Systolic and diastolic blood pressure ECG waveform Heart rate Email this data file
Interfaces Cables to interface the human body to the device Serial interface from the device to the Palm Serial interface from the palm to PC
![Page 6: FDR Presentation Personal Health Monitoring Device Bassam Noaman Dina El-Eissa Mentor: Prof Zhenyu Guo Instructor: Prof J Silverman 4/12/00](https://reader030.vdocuments.us/reader030/viewer/2022033107/56649dba5503460f94aaaf9c/html5/thumbnails/6.jpg)
SpecificationSpecification• Indicators
LED power statusLED blood pressure visual detector
Memory Storage1 MB of Palm is allocated for health monitoring 55.6 minutes of display range
![Page 7: FDR Presentation Personal Health Monitoring Device Bassam Noaman Dina El-Eissa Mentor: Prof Zhenyu Guo Instructor: Prof J Silverman 4/12/00](https://reader030.vdocuments.us/reader030/viewer/2022033107/56649dba5503460f94aaaf9c/html5/thumbnails/7.jpg)
Block DiagramBlock Diagram Bassam
Knob
Dina
Bassam
Hi Pass Filter
Instrumentation Amplifier
Notch Filter
Low Pass Filter
2.5 VDC offset
Amplifier
ECG Sensor Blood Pressure Sensor Capacitive
Microphone
Comparison Circuit
Current Temperature Sensor
Amplifier
Super Diode
PIC16C72 8-bit Microcontroller
8-bit multichunnel A/D Converter
Palm Pilot
Band Pass Amplifier
LED
PC
MMV
SAmplifier
![Page 8: FDR Presentation Personal Health Monitoring Device Bassam Noaman Dina El-Eissa Mentor: Prof Zhenyu Guo Instructor: Prof J Silverman 4/12/00](https://reader030.vdocuments.us/reader030/viewer/2022033107/56649dba5503460f94aaaf9c/html5/thumbnails/8.jpg)
Temperature signal conditioner simulationTemperature signal conditioner simulation
R4
1k
R6
1k
U2
LM741
3
2
74
6
1
5+
-
V+V-
OUT
OS1
OS2
V1-270mVdcV R1
10k
298.0uA0
D1
D1N4002
R5
100k
0V
R7
100k
0
0
0
I1298uAdc
R2
1k
0
V3-5Vdc
V2
5Vdc
0
2.980V
R3
1k
0
U1
LM741
3
2
74
6
1
5+
-
V+V-
OUT
OS1
OS2
![Page 9: FDR Presentation Personal Health Monitoring Device Bassam Noaman Dina El-Eissa Mentor: Prof Zhenyu Guo Instructor: Prof J Silverman 4/12/00](https://reader030.vdocuments.us/reader030/viewer/2022033107/56649dba5503460f94aaaf9c/html5/thumbnails/9.jpg)
Output voltage of the body Temperature sensorOutput voltage of the body Temperature sensor
![Page 10: FDR Presentation Personal Health Monitoring Device Bassam Noaman Dina El-Eissa Mentor: Prof Zhenyu Guo Instructor: Prof J Silverman 4/12/00](https://reader030.vdocuments.us/reader030/viewer/2022033107/56649dba5503460f94aaaf9c/html5/thumbnails/10.jpg)
Output voltage of the signal conditioner of the Output voltage of the signal conditioner of the body temperature sensorbody temperature sensor
![Page 11: FDR Presentation Personal Health Monitoring Device Bassam Noaman Dina El-Eissa Mentor: Prof Zhenyu Guo Instructor: Prof J Silverman 4/12/00](https://reader030.vdocuments.us/reader030/viewer/2022033107/56649dba5503460f94aaaf9c/html5/thumbnails/11.jpg)
Blood pressure measurementBlood pressure measurement
Blood pressure:
is defined as the intra-arterial pressure generated by
the pulsatic flow of blood through an artery. The blood pressure is indicated from two numbers.
Systolic:represents the pressure while the heart is beating.
maximum value for blood pressure
Diastolic: represents the pressure when the heart is resting between beats.
![Page 12: FDR Presentation Personal Health Monitoring Device Bassam Noaman Dina El-Eissa Mentor: Prof Zhenyu Guo Instructor: Prof J Silverman 4/12/00](https://reader030.vdocuments.us/reader030/viewer/2022033107/56649dba5503460f94aaaf9c/html5/thumbnails/12.jpg)
Flowchart for blood pressure measurementFlowchart for blood pressure measurement
![Page 13: FDR Presentation Personal Health Monitoring Device Bassam Noaman Dina El-Eissa Mentor: Prof Zhenyu Guo Instructor: Prof J Silverman 4/12/00](https://reader030.vdocuments.us/reader030/viewer/2022033107/56649dba5503460f94aaaf9c/html5/thumbnails/13.jpg)
Signal Conditioner of Capacitive Signal Conditioner of Capacitive MicrophoneMicrophone
+5
out1
+4
out2 LED OUTPUT
Microphone
JK FF
-2
R2
10k
R5
1k
+5
out4
K
+VCC
LED OUTPUT
-5
-VCC
OPA4137
R4
10k
-4
-1 +2C1
1n
C2
1n
+1
R3
1k
R6 1k
R1
1k
J
PUSH BUTTON
![Page 14: FDR Presentation Personal Health Monitoring Device Bassam Noaman Dina El-Eissa Mentor: Prof Zhenyu Guo Instructor: Prof J Silverman 4/12/00](https://reader030.vdocuments.us/reader030/viewer/2022033107/56649dba5503460f94aaaf9c/html5/thumbnails/14.jpg)
ECG Signal ConditionerECG Signal Conditioner
R7
1k
C2
1n
out2
R1
680
+4
C4
1u
R2
1k
AD620
LL
-4
C6.47u
-1
R9
1k
R10
40k
+2
+5
out4
out
RG 5k
+1
C1
1u 3.3M
1k
-2
RA
-5R82.7k
+5
R4
1k
+VCC
-VCC
LA
OPA4137
-5R3
1k
C5
1u
out1
R6 2.7k
+2.5
![Page 15: FDR Presentation Personal Health Monitoring Device Bassam Noaman Dina El-Eissa Mentor: Prof Zhenyu Guo Instructor: Prof J Silverman 4/12/00](https://reader030.vdocuments.us/reader030/viewer/2022033107/56649dba5503460f94aaaf9c/html5/thumbnails/15.jpg)
Simulation of the frequency Simulation of the frequency domain of ECGdomain of ECG
![Page 16: FDR Presentation Personal Health Monitoring Device Bassam Noaman Dina El-Eissa Mentor: Prof Zhenyu Guo Instructor: Prof J Silverman 4/12/00](https://reader030.vdocuments.us/reader030/viewer/2022033107/56649dba5503460f94aaaf9c/html5/thumbnails/16.jpg)
Microcontroller circuitMicrocontroller circuit
8
1 14
+5+5
3
32
C51u
C21u
24
+5
temp
11
C6
33n
6
+5
43
C7
33n
C4
1u
C1 1u
123
BP
1
40
10
PIC 16F874
13
ECG
MAX232
2
C3
1u
31
![Page 17: FDR Presentation Personal Health Monitoring Device Bassam Noaman Dina El-Eissa Mentor: Prof Zhenyu Guo Instructor: Prof J Silverman 4/12/00](https://reader030.vdocuments.us/reader030/viewer/2022033107/56649dba5503460f94aaaf9c/html5/thumbnails/17.jpg)
Software design block diagramSoftware design block diagram
![Page 18: FDR Presentation Personal Health Monitoring Device Bassam Noaman Dina El-Eissa Mentor: Prof Zhenyu Guo Instructor: Prof J Silverman 4/12/00](https://reader030.vdocuments.us/reader030/viewer/2022033107/56649dba5503460f94aaaf9c/html5/thumbnails/18.jpg)
Actual Physical LayoutActual Physical Layout
![Page 19: FDR Presentation Personal Health Monitoring Device Bassam Noaman Dina El-Eissa Mentor: Prof Zhenyu Guo Instructor: Prof J Silverman 4/12/00](https://reader030.vdocuments.us/reader030/viewer/2022033107/56649dba5503460f94aaaf9c/html5/thumbnails/19.jpg)
Physical layout of the program Physical layout of the program interfaceinterface
![Page 20: FDR Presentation Personal Health Monitoring Device Bassam Noaman Dina El-Eissa Mentor: Prof Zhenyu Guo Instructor: Prof J Silverman 4/12/00](https://reader030.vdocuments.us/reader030/viewer/2022033107/56649dba5503460f94aaaf9c/html5/thumbnails/20.jpg)
DemonstrationDemonstration
Measure current temperature and blood pressure and display it on the palm.
Display temperature,blood pressure and ECG wave in Hyper terminal, then take the data and save it in a file. Finally display it on java applet.
![Page 21: FDR Presentation Personal Health Monitoring Device Bassam Noaman Dina El-Eissa Mentor: Prof Zhenyu Guo Instructor: Prof J Silverman 4/12/00](https://reader030.vdocuments.us/reader030/viewer/2022033107/56649dba5503460f94aaaf9c/html5/thumbnails/21.jpg)
ConclusionConclusion
The Palm pilot is not only an organizer, but also a device to monitor data(Blood pressure, Temperature)