fag detector iii - quick start manual
TRANSCRIPT
8/4/2019 FAG Detector III - Quick Start Manual
http://slidepdf.com/reader/full/fag-detector-iii-quick-start-manual 1/8
8/4/2019 FAG Detector III - Quick Start Manual
http://slidepdf.com/reader/full/fag-detector-iii-quick-start-manual 2/8
Screw the magnet onto the sensor.Then fit the coiled cable to the
sensor and rotate the union nut untilit is tight. Now connect the red endof the helix cable to the red BNC jackon the device. Fit the rechargeablebattery to the device.
Switch on the FAG Detector III(press “ “ for 2 seconds).Using the cursor keys will alwaysbring you to the required menu item.If the language is set to German,change this to English as follows:
- Select “System-Menü”
- Press “ “
- Select “Sprache ändern“
- Press “ “
- Select “English“
- Press “ “
- Then press “ “ in order tomove up to the next menu
1. Setup the device
Page 1
magnet sensor
8/4/2019 FAG Detector III - Quick Start Manual
http://slidepdf.com/reader/full/fag-detector-iii-quick-start-manual 3/8
2. Select the machine
Ensure that this is a new machine or a machine suitable for continuousoperation.
3. Select the measurement position
Irrespective of the machine type, the
measuring position should be in theload zone wherever possible. If thisis not known, a measuring pointshould be measured in a horizontal,vertical and axial direction.
The measurement position must:- be made of ferrous material in
order to ensure adhesion of the
sensor/magnet- be marked such that the sensor
can always be fitted at the sameposition for all measurements inthe follow-on period
- have good metallic contact withthe bearing and shaft (no looseplates, covers etc.)
- be clean and free of grease- be flat- be larger than the diameter of the
sensor - be entered in the measurement
table using an unambiguous name
First make the required number of copies of the measurement table
template (see page 7).Use one measurement table for eachmachine component.
Ensure that vibration measurementis carried out in a safe manner in
each case!
Page 2
8/4/2019 FAG Detector III - Quick Start Manual
http://slidepdf.com/reader/full/fag-detector-iii-quick-start-manual 4/8
4. Define the machine class (ISO 10816)
Figure 1: Example measurement table
Page 3
Each machine class is allocated different alarm thresholds for vibration.Assign your machine to an ISO class (see Figure 2) and enter thecorresponding prealarm and main alarm thresholds in the measurement table.
Rough power classification
< 15 kW
15-75 kW resp.15-300 kW
300 kW-10 MW
< 10 MW
Class
I
II
III
IV
E xact description according to ISO 10816
Individual parts of engines and machines, integrallyconnected to the complete machine in its normaloperating condition. (Electric drive motors of up to 15 kWare typical examples of machines in this category.)
Medium-sized machines (typically electrical motors with15 kW to 75 kW output) without special foundations,
ridgly mounted engines or machines (up to 300 kW) onspecial foundations.
Large prime-movers and other large machines withroating masses mounted on rigid and heavy foundationswhich are relatively stiff in the direction of vibrationmeasurements.
Large prime-movers and other large machines withrotating masses mounted on foundations which arerelatively soft in the direction of vibration measurements
(for example, turbogenerator sets and gas turbines withoutputs greater than 10 MW).
The standard ISO 10816 defines various machine classes since the vibrationbehaviour of large machines is different from that of small machines.
8/4/2019 FAG Detector III - Quick Start Manual
http://slidepdf.com/reader/full/fag-detector-iii-quick-start-manual 5/8
5. Carry out measurements
Place the sensor/magnet carefully on the measurement position (with one sidefirst, then the other).
Page 4
Figure 2: Alarm thresholds according to ISO 10816
8/4/2019 FAG Detector III - Quick Start Manual
http://slidepdf.com/reader/full/fag-detector-iii-quick-start-manual 6/8
1. Switch on the FAG Detector III(press “ ” for 2 seconds)
2. Select “Single measurements”
3. Press “ ”
4. Select “ISO 10816” and press “ ”
5. Select the machine class requiredfor measurement (Class 1-4) andpress “ ”
6. Enter the value displayed in thecorrect column in the measurementtable
7. Move the sensor to the nextmeasurement position
8. Press “ ” again to start a newmeasurement on the same machineclass
After the final measurement, switch the
device off.
Page 5
Measurement will start. Ensure that neither the machine nor themeasurement system are subjected to any disruptions during
measurement. The measurement result will be shown in the display.
8/4/2019 FAG Detector III - Quick Start Manual
http://slidepdf.com/reader/full/fag-detector-iii-quick-start-manual 7/8
6. Evaluate the measurement values
The first 10 measurement values in a column in the measurement table(measured over a period of 2-5 months) must be taken as the normal varianceof machine vibration. If significantly increasing deviations from this “normalcondition” occur in the follow-on period, this indicates a change in themachine.
If you require further information or have any queries, please contact us asfollows:
SupportPhone: +49 (0)2407-9149-99E-mail: [email protected]
SalesPhone: +49 (0)2407-9149-66E-mail: [email protected]
Internet
www.FAG-DetectorIII.comwww.fis-services.com
Page 6
8/4/2019 FAG Detector III - Quick Start Manual
http://slidepdf.com/reader/full/fag-detector-iii-quick-start-manual 8/8
F A G I n d
u s t r i a l S e r v i c e s • K a i s e r s t r a
s s e 1 0 0 • 5 2 1 3 4 H e r z o g e n r a
t h ( G e r m a n y )
P h o n e : + 4 9 ( 0 ) 2
4 0 7 - 9 1 4 9 - 9 9 • F a x : + 4 9 ( 0 ) 2 4 0 7 - 9 1 4 9 - 5 9 • E - m a i l : s u p p
o r t @ f i s - s e r v i c e s . c o m
M e a s u r e
m e n t t a b l e
P a g e 7