BD FACSCount CD4 Reagents Features Provides absolute CD4 counts and CD4 percentage results  Uses unlysed whole blood  30-minute incubation time  Includes automated data analysis and built-in quality control  Runs on trusted BD FACSCount system New CD4 reagents on the proven BD FACSCount™ system provide absolute CD4 counts and percentage results in an easy-to-use, single-tube assay . The results you need – simply, reliably , and economically The BD FACSCount system is a turnkey, automated instrument and reagent system designed for CD4 counting in a wide range of laboratory environments. Setup and operator training are minimal. The software identies the lymphocyte populations of interest and calculates absolute CD4 counts (cells/µL) and CD4 percentages automatically, saving time, minimizing human errors, and reducing costs. More efcient workow BD FACSCount CD4 reagents are provided in one ready-to-use reagent tube. The assay uses unlysed whole blood, simplifying sample preparation. Incubation time is 30 minutes, speeding time to results. Fluorescence reference beads, included in the reagent tube, ensure accurate enumeration of lymphocytes of interest with no operator intervention. The software includes built-in quality control and a retrievable results le containing data from the last 100 runs, allowing access to previously run samples and transfer to a third-party computer. These features provide laboratories with a more efcient workow and greater condence in results. A familiar, proven system The BD FACSCount system has been an integral part of CD4 monitoring in global HIV/AIDS treatment programs for over a decade. Around the world, its simplicity of operation and robust performance in a wide range of laboratory settings make it a system of choice. With this quality of results, clinicians can depend on it for patient management. Global training and support from BD Biosciences BD Biosciences offers comprehensive training and service programs, carried out by experienced, knowledgeable BD support and eld service teams around the world. The ready-to-use BD FACSCount CD4 reagent tube contains every- thing required for absolute CD4 counts and CD4 percentages. US Orders: 877.232.8995 bdbiosciences.com “This new BD FACSCount system is a simple and reliable  system for enumeration of absolute and percentage CD4 +  T lymphocytes. Having one system giving both results should reduce the cost and thus increase access to CD4 testing . . . “ Kovit Pattanapanyasat, PhD. Visit bdbiosciences.com for more information.

Upload: adittia-heryanti

Post on 06-Jan-2016

237 views

Category:

Documents


0 download

DESCRIPTION

hjhjhjh

TRANSCRIPT

Page 1: Facscount Data Sheet

7/17/2019 Facscount Data Sheet

http://slidepdf.com/reader/full/facscount-data-sheet 1/2

BD FACSCount CD4 Reagents

Features

Provides absolute CD4 counts and CD4 percentage results

  Uses unlysed whole blood

  30-minute incubation time

  Includes automated data analysis and built-in qualitycontrol

  Runs on trusted BD FACSCount system

New CD4 reagents on the proven BD FACSCount™ systemprovide absolute CD4 counts and percentage results in an

easy-to-use, single-tube assay.The results you need – simply, reliably, and economicallyThe BD FACSCount system is a turnkey, automated instrumentand reagent system designed for CD4 counting in a wide rangeof laboratory environments. Setup and operator training areminimal. The software identifies the lymphocyte populations ofinterest and calculates absolute CD4 counts (cells/µL) and CD4percentages automatically, saving time, minimizing human errors,and reducing costs.

More efficient workflowBD FACSCount CD4 reagents are provided in one ready-to-usereagent tube. The assay uses unlysed whole blood, simplifyingsample preparation. Incubation time is 30 minutes, speedingtime to results. Fluorescence reference beads, included in thereagent tube, ensure accurate enumeration of lymphocytes ofinterest with no operator intervention. The software includesbuilt-in quality control and a retrievable results file containingdata from the last 100 runs, allowing access to previously runsamples and transfer to a third-party computer. These featuresprovide laboratories with a more efficient workflow and greaterconfidence in results.

A familiar, proven systemThe BD FACSCount system has been an integral part of CD4monitoring in global HIV/AIDS treatment programs for over adecade. Around the world, its simplicity of operation and robustperformance in a wide range of laboratory settings make ita system of choice. With this quality of results, clinicians can

depend on it for patient management.Global training and support from BD BiosciencesBD Biosciences offers comprehensive training and serviceprograms, carried out by experienced, knowledgeableBD support and field service teams around the world.

The ready-to-use BD FACSCount CD4 reagent tube contains every-thing required for absolute CD4 counts and CD4 percentages.

US Orders: 877.232.8995bdbiosciences.com

“This new BD FACSCount system is a simple and reliable system for enumeration of absolute and percentageCD4+ T lymphocytes. Having one system giving bothresults should reduce the cost and thus increase accessto CD4 testing . . . “ Kovit Pattanapanyasat, PhD.

Visit bdbiosciences.com for more information.

Page 2: Facscount Data Sheet

7/17/2019 Facscount Data Sheet

http://slidepdf.com/reader/full/facscount-data-sheet 2/2

Class I (1) laser product.For In Vitro Diagnostic Use.This product is CE IVD under the IVD directive.BD, BD Logo and all other trademarks are property of Becton, Dickinson and Company. © 2009 BD

23-9977-02

AccuracyCD4 absolute count and percentageresults were compared with results fromBD Tritest™ CD3 FITC/CD4 PE/CD45PerCP with BD Trucount™ tubes on the

BD FACSCalibur™ flow cytometer usingBD Multitest™ software.

PrecisionPrecision was estimated usingBD™ Multi-check low and normalcontrols. Two replicates of each controlwere analyzed in each run, and two runswere performed per day for a total of

21 days.

User-Reportable RangesAbsolute counts: 50 to 5,000 CD4 cells/µL

Percentages: 5% to 65%

StabilityStaining: Within 24 hours of draw

Analysis: Within 48 hours of staining

70

60

50

40

30

20

10

0

0 2010 30 40 50 60 70

BD Tritest CD4 percentage 

   B   D    F

   A   C   S   C  o  u  n   t   C   D   4  p  e  r  c  e  n   t  a  g  e

5000

4000

3000

2000

1000

0

0 1000 2000 3000 4000 5000

 BD Tritest CD4 absolute count

   B   D    F

   A   C   S   C  o  u  n   t   C   D   4  a   b  s  o   l  u   t  e  c  o  u  n   t

 

BD FACSCount CD4 Reagents

Parameters n R2 Slope Intercept Range

Absolute CD4 (cells/µL) 101 0.981 0.971 12.695 59 – 3,404

CD4 Percentage 99 0.99 0.999 -0.391 5.51 – 64.69

CD4 Absolute Counts CD4 Percentages

LowControl

NormalControl

LowControl

NormalControl

% CV SD

Within Device 4.82 4.28 0.38 1.28

Within Run 4.04 3.46 0.35 1.15

Regression plot of test versus predicate for CD4 absolute counts Regression plot of test versus predicate for CD4 percentages

Ordering Information

Product description QuantityReg

Status

Cat No.

BD FACSCount CD4 Reagents 50 Tests IVD 339010

BD FACSCount CD4 Software and User’s Guide Assembly 4 Disks — 339011

BD Biosciences

2350 Qume Drive

San Jose, CA 95131

US Orders: 877.232.8995

bdbiosciences.com