ezmm

32

Upload: rakshit-srinivas

Post on 21-Nov-2015

5 views

Category:

Documents


0 download

DESCRIPTION

.

TRANSCRIPT

  • Manual for K-Notes

    Why K-Notes?

    Towards the end of preparation, a student has lost the time to revise all the chapters from his /

    her class notes / standard text books. This is the reason why K-Notes is specifically intended for

    Quick Revision and should not be considered as comprehensive study material.

    What are K-Notes?

    A 40 page or less notebook for each subject which contains all concepts covered in GATE

    Curriculum in a concise manner to aid a student in final stages of his/her preparation. It is highly

    useful for both the students as well as working professionals who are preparing for GATE as it

    comes handy while traveling long distances.

    When do I start using K-Notes?

    It is highly recommended to use K-Notes in the last 2 months before GATE Exam

    (November end onwards).

    How do I use K-Notes?

    Once you finish the entire K-Notes for a particular subject, you should practice the respective

    Subject Test / Mixed Question Bag containing questions from all the Chapters to make best use

    of it.

  • Measurement & Instrumentation

    State characteristics of measuring system

    1) Accuracy

    Degree of closeness in which a measured value approaches a true value of a quality under

    measurement .

    When accuracy is measured in terms of error :

    Guaranteed accuracy error (GAE) is measured with respect to full scale deflation.

    Limiting error (in terms of measured value)

    GAE * FullscaledeflationLE

    Measuredvalue

    2) Precision

    Degree of closeness with which reading in produced again & again for same value of input

    quantity.

    3) Sensitivity

    Change the output quantity per unit change in input quantity.

    o

    i

    qS

    q

    4) Resolution

    Smallest change in input which can be measured by an instrument

    5) Threshold

    Minimum input required to get measurable output by an instrument

    6) Zero Drift

    Entire calibration shifts gradually due to permanent set

  • 7) Span Drift

    If there is proportional change in indication all along upward scale is called span drift.

    8) Dead zone & Dead time

    The range of input for which there is no output this portion is called Dead zone.

    To respond the pointer takes a minimum time is called dead time.

    TYPES OF ERROR

    a) Gross Error : Error due to human negligency, i.e. due to loose connection, reading the value

    etc.

    b) Systematic error : Errors are common for all observers like instrumental errors,

    environmental errors and observational errors.

    c) Random errors : Error due to unidentified causes & may be positive or negative.

    Absolute Errors :

    m rA A A

    mA Measured value

    rA True value

    Relative Errors :

    r =

    A

    T

    AbsoluteErrors

    Truevalue A

    mm rT T

    r

    AA A A 1

    1

  • Composite Error :

    i) Sum of quantities

    1 2

    X X X

    x x1 x2

    ii) Difference of quantities

    1 2

    X X X

    x x1 x2 So for sum & difference absolute errors are added.

    iii) Multiplication of quantities

    1 2 3

    X X X X

    31 2

    1 2 3

    XX XX

    X X X X

    iv) Division of quantities

    1

    2

    XX

    X

    1 2

    1 2

    X XX

    X X X

    So, for multiplication & division, fractional or relative errors are added.

    If m m

    1 2

    p

    3

    X XX

    X

    31 2

    1 2 3

    XX XXm n p

    X X X X

    Precision Index

    Indicates the precision for a distribution

    1h

    2

  • Probable Error

    r = 0.6745

    0.4769r

    h

    Standard deviation of combination of quantities

    2 2 2

    1 2n1 2

    n

    2 2 2x x x x

    X X X......

    X X X

    Probable Error

    2 2 2

    1 2 nn1 2

    22 2x x x x

    X X Xr r r ...... r

    X X X

    Electro-Mechanical Instruments

    1) Permanent magnet moving Coil (PMMC)

    Deflecting Torque

    Td = nIAB

    Where n = no. of turns

    I = current flowing in coil

    A = Area of coil

    B = magnetic flux density

    Deflection G

    Ik

    G = NBA & K = Spring constant

    Eddy current damping & spring control torque in used.

    For pure AC signal, the pointer vibrates around zero position.

    It is used to measured DC or average quantity.

    It can directly read only up to 50mV or 100mA.

  • Enhancement of PMMC

    i) Ammeter

    For using PMMC as an ammeter with wide range, we connect a small shunt resistance in

    parallel to meter.

    m

    Im

    I multiplication factor

    Basically, m is ratio of final range (as an ammeter) to initial range of instrument.

    m

    sh

    RR

    m 1

    ; mR = meter resistance

    ii) Voltmeter

    A series multiples resistance of high magnitude is connected in series with the meter.

    M = multiplication factor

    m

    Vm

    V

    s mR R m 1

    Sensitivity of voltmeter

    s mv

    fsd

    R R1S / V

    I V

    Application of PMMC

    1) Half wave rectifier meter

    mavg

    II I

  • RMS

    s m f

    2V

    R R R

    RMS

    avg

    s m f

    0.45VI

    R R R

    ; For Ac input

    For DC input

    DCavg

    s m f

    VI

    R R R

    avg AC DCavgI 0.45 I (Assuming DC RMSV V )

    AC DC(Sensitivity) 0.45(Sensitivity)

    2) Full wave rectifier meter

    RMS

    ACf

    avgs m

    2 2VI

    R R 2R

    RMS

    s m f

    0.9V

    R R 2R

    DCavg DC

    s m f

    VI

    R R 2R

    avg DCACavgI 0.9 I (Assuring RMS DCV V )

    AC DCSensitivity 0.9 Sensitivity

    2) Moving iron meter

    Deflecting torque, 2d1 dL

    T I2 d

    I = current flowing throw the meter

    L = Inductance

    = deflection

    Under steady state

  • 21 dL

    K I2 d

    2 I

    MI meter measures both ac & dc quantities. In case of AC, It measures RMS value.

    1T 2

    2

    RMS

    0

    1I i t dt

    T

    If 0 1 2i t I I sinwt I sin2wt .......

    2 2 2RMS 0 1 21

    I I I I .......2

    Air friction Damping is used

    Condition for linearity

    dLcons tant

    d

    MI meter cannot be used beyond 125Hz, as then eddy current error is constant.

    3) Elector dynamometer

    Deflecting Torque, d 1 2dM

    T i id

    For DC, 1 2i i I

    2ddM

    T Id

    2 I

    For AC, 1 m1i I sin t

    2 m2i I sin t

    1 2d avg

    dMT I I cos

    d

    Where m11

    II

    2 & 2

    I2I

    2

  • Applications of dynamometer

    1) Ammeter

    Fixed coils are connected in series.

    1 2I I I

    0 (Angel between 1 2I & I )

    2d

    dMT I

    d

    At balance, c dT T

    2dM

    K Id

    2 I

    It reads both AC & DC & for AC it reads RMS.

    2) Voltmeter

    sR Series multiplier resistance

    2 1

    s

    VI I

    R , 0

    cos 1

    2

    d 2

    s

    V dMT

    dR

    At balance, cdT T

    2

    2s

    V dMK

    dR

    2 V

    It reads both AC & DC & for AC it reads RMS.

  • 3) Wattmeter

    Fixed coils carry same current as load & as called as current coils.

    Moving coil is connected across voltage and thus current voltage, a high non-inductive

    load is connected in series with MC to limit the current.

    d 1 2

    dMT I I cos

    d

    avg

    s s

    PV dM dMI cosR d R d

    At balance, d

    k T

    avg P

    Symbol :

    Two wattmeter method

    1 RY R RY RW V I cos V & I

    L LV I cos 30

    2 BY B BY BW V I cos V & I

    L LV I cos 30

    Where LV is line to line voltage

  • LI is line current

    These expression remain same for -connected load.

    1 23P W W

    L L3V I cos

    2 13Q 3 W W

    L L3V I sin

    2 1

    1 2

    3

    3

    Q 3 W Wtan

    P W W

    2 11

    1 2

    3 W Wtan

    W W

    for lag load

    2 11

    1 2

    3 W Wtan

    W W

    for lead load

    = Remember, In our case 1

    W is wattmeter connected to R-phase and 2

    W is wattmeter

    connected to B-phase.

    = If one of the wattmeter indicates negative sign, then pf < 0.5

    Errors in wattmeter

    a) Due to potential coil connection

    2

    T

    cLr

    I r% * 100

    P

    LI = load current

    Cr = CC Resistance

    TP = True Power

    2

    T

    rs

    V% *100

    R P

    V = voltage across PC

  • sR = Series multiplier resistance

    TP = True Power

    b) Due to self inductance of PC

    If PC has finite inductance

    p p sZ R R jwLp

    p sR R p sZ R jwLp

    r% tan tan *100

    = load pf angle

    p1

    s

    Ltan

    R

    4) Energy meter

    Energy = Power * Time

    T

    VIcos tW * kwhr

    1000 3600

    TW = True energy

    It is based on principle of induction.

    It is an integrating type instrument.

    mt

    W VIsin * kwhr3600

    Where mW = measured Energy

    = angle between potential coil voltage & flux produced by it.

    = load pf angle

    Error = m TW W

    Energy constant = No.ofRe voluations N

    kwhr P.t

    Measured Energy = mTotalno.ofrevolutions

    KW

  • True Energy =

    TVIcos t

    * kw.hr1000 3600

    W

    Error = m TrT

    W W% *100

    W

    Creeping Error in energy meter

    If friction is over compensated by placing shading loop nearer to PC, then disc starts rotating

    slow with only PC excited without connecting any load is creeping.

    Otherwise if over voltage is applied on pressure coil then also creeping may happen due to

    stray magnetic fields.

    To remove creeping holes are kept on either side of disc diametrically opposite & the torque

    experienced by both holes is opposite & they stop creeping.

    % creeping error = TotalNo.ofRew / kwhr due to creeping

    *100TotalNo.ofRew / kwhr due to load

    Thermal Instruments

    These instruments work on the principle of heating and are called as Thermal Instruments.

    These are used for high frequency measurements.

    They can measure both AC & DC.

    In case of AC, they measure RMS value.

    Electrostatic voltmeter

    Deflecting torque, 2d

    1 dcT V

    2 d

    At Balance,

    cdT T

    21 dc

    V k2 d

    2 V

    Condition for linearity

  • dc

    constantd

    For increasing the range, we connect another capacitor in series

    To increase the range from mV to V

    m

    s

    CC

    m 1 ;

    m

    Vm

    V

    Potentiometer / Null Detector

    wI = working current

    B

    wh

    VI

    R l.r _____________(1)

    Switch at (A)

    If gI 0

    s w 1V I l r

    sw1

    VI

    l r _____________(2)

    Switch at (B)

    2x wV I l r

    xw

    2

    VI

    l r ________(3)

    s x

    1 2

    V V

    l r l r

    2x s1

    lV V

    l

  • r = resistance of slide wire (/ m)

    l = Total length of slide wire (m)

    1l = length at which standard cell ( sV ) is balanced

    2l = length at which test voltage ( xV ) is balanced

    Measuring a low resistance

    R

    s

    VR S

    V

    Instrument Transformer

    Current transformer

    Equivalent circuit

    Turns Ratio = Nominal Ratio 2

    1

    Nn

    N

    1 l s

    l s

    X Xtan

    R R

    R = Actual Ratio

    s

    I cos I sinn

    I

  • Errors in current transformer

    1) Ratio Error :

    Current ratio p

    s

    I

    I is not equal to turns ratio due to no-load component of current.

    r

    K R% *100

    R

    K = n = Nominal Ratio

    R = Actual Ratio

    2) Phase Angel Ratio :

    Ideally, Phase difference between p sI & I should be 0180 but due to no-load component of

    current, it deviates from that value.

    Phase angle error =

    s

    I cos I sin 180*

    nI degrees

    Phase angle between primary & secondary currents

    = 180 degrees

    Potential Transformer

    Equivalent circuit

    Turns Ratio = n = 2

    1

    N

    N

    Actual Transformation Ratio = R = P

    S

    V

    V

    SP P P P

    S

    I1R n R cos X sin I R I X

    V n , Where

    1 XtanR

    Phases angle error

    SP P P P

    s

    IX cos R sin I X I R

    nnV

  • AC & DC Bridges

    AC Bridges

    Balance condition : DI 0

    41 2 3Z Z Z Z

    41 2 3Z Z Z Z

    41 2 3

    432 3

    1 24

    Z ZZ

    Z

    Quality Factor & dissipation factor

    Quality Factor (Q) Dissipation Factor

    (D)

    1 wLQ

    R

    RD

    wL

    2 RQ

    wL

    wLD

    R

    3 1Q

    wCR

    D =wcR

    4 Q = wcR 1D

    wCR

    Measurement of Inductance

    (i) Maxwells Inductance Bridge

    Here, we try to measure 1

    R & 1

    L

  • 2 31

    4

    R RR

    R

    2 31

    4

    L LL

    R

    (ii) Maxwells Inductance Capacitance Bridge

    2 31

    4

    R RR

    R

    41 2 3L R R C

    This bridge is only suitable for coils where 1 < Q < 10

    Q = Quality Factor

    (iii) Hays Bridge

    Used for coils having high Q value

    2 24 42 3

    1 2

    R R R CR

    11

    Q

    42 31 2

    R R CL

    11

    Q

    4 4

    1Q

    R C

    (iv) Andersons Bridge

    This Bridge is used for low Q coils.

    2 31 1

    4

    R RR r

    R

    3 41 2 2 34

    CRL R R r R R

    R

  • (v) Owens Bridge

    431

    2

    R CR

    C

    2 41 3L R R C

    Measurement Of Capacitance

    De-Sautys Bridge

    31 2 2 14

    Rr R r R

    R

    42

    31

    RC C

    R

    D = dissipation factor

    = 1 1

    C r

    1r = internal resistance of

    1C

    Schering Bridge

    431

    2

    R CR

    C

    4 21

    3

    R CC

    R

    dissipation factor = D = 4 4C R

    Measurement of frequency

    Wien Bridge Oscillator

    Balancing Condition

    3 1 2

    4 2 1

    R R C

    R R C

    Frequency of Osculation

    1 2 1 2

    1f

    2 R R C C

  • Measurement of Resistance

    Classification of Resistance

    1) Low Resistance : R 1 , Motor and Generator

    2) Medium Resistance : 1 < R < 100k , Electronic equipment

    3) High Resistance : R > 100 k, winding insulation of electrical motor

    DC Bridges

    Medium Resistance Measurement

    1. Wheatstone Bridge

    Finding Theremin Equivalent

    thggth

    VI

    R R

    Th

    P RV V

    P Q R S

    Th

    PQ RSR

    P Q R S

    For Balance Condition

    gI 0

    Th

    V 0

    PS = RQ

  • Sensitivities

    1) Current sensitivity , i

    g

    SI

    mm/mA

    = deflection of Galvan meter in mm

    2) Voltage sensitivity,

    Th

    SV

    mm/V

    3) Bridge Sensitivity , B

    SR /R

    mm

    ThB

    vV SSR / R

    B

    vV.SSSR 2

    S R

    For Maximum Sensitivity

    SRS R = 1

    B, maxvV.SS

    4

    2. Carey foster slide wire Bridge

    r = slide wire resistance in m

    .

    for case (1).

    At balance

    1

    1

    R rP

    Q S L r

    .(1)

    For case (2)

    R & S is reversed

    2

    2

    S rP

    Q R L r

    ..(2)

    From (1) & (2)

    1 2

    1 2

    R r S r

    S L r R L r

  • 3. Voltmeter Ammeter Method

    a) Ammeter near the load

    vm X AA

    VR R R

    I

    vV = voltage across voltmeter

    A

    I = Ammeter current

    X

    R = Test resistance, A

    R = Animator resistance

    % error = m T A

    xT

    R R R100 100%

    R R

    b) Voltmeter near the load

    X

    v Xm

    vA

    VVR

    I I I

    vXm

    X v vX

    X X

    R R1R

    I I R R

    V V

    % error = m X

    X

    R R100%

    R

    If a vXR R R , voltmeter is connected near the load

    a vXR R R , ammeter is connected near the load

    4. Ohmmeter

    a) Series Type

    when XR 0

    m FSDI I = Full scale deflection

    when XR

    mI 0 = zero deflection

    for Half scale deflection

    msh

    seX hmsh

    R .RR R R

    R R

  • b) Shunt Type

    S

    R = current limiting resistor

    If X

    R 0

    mI 0 = zero deflection

    If xR

    m FSDI I = Full scale deflection

    For Half scale Deflection

    m S

    x hm S

    R RR R

    R R

    Measurement of Low Resistance

    Kelvins Double Bridge Method

    Unknown resistance

    qr pP P

    R SQ p q r Q q

    P, Q = outer ratio arms

    p, q = inner ratio arms

    S = standard resistance

    r = lead resistance

    R = Test resistance

    High Resistance Measurement

    Loss of charge Method

    tRc

    eCV t V

    10

    C

    0.4343tR

    VC log

    V

    t = time in (seconds)

    V = source voltage

    CV = Capacitor voltage

  • Cathode Ray Oscilloscope (CRO)

    The velocity of e is charged by charging the pre-accelerating & accelerating anode

    potential

    KE =PE

    2

    a1

    mv qV2

    a2qV

    m

    Deflection sensitivity

    D = deflection height on screen

    d = distance between plates

    d = length of vertical deflecting plates

    L = distance between centre of plate & screen

    aV = anode potential

    yV = Vertical plate Potential

    yd

    a

    L VVD

    mm2dV

    deflection sensitivity

    d

    y a

    LD VSmmV 2dV

    Lissajous Pattern

    If both horizontal & vertical deflection plates of CRT is applied with the sinusoidal signal,

    the wave form pattern appearing on screen is called Lissajous Pattern.

  • Case 1: Both signals have same frequency

    x m xV V sin w t

    y m yV V sin w t

    x y mV V V

    x yw w w

    = variable

    S.No Lissayous Pattern

    1

    0 or 360

    2

    0 90

    Or

    270 360

    3

    90 or 270

    4

    90 270

    Or

    180 270

    5

    180

  • Finding

    1) Lissajous Pattern in Ist & IIIrd Quadrant

    1 11 1

    2 2

    X Ysin sin

    X Y

    for clockwise orientation phase difference = (180 )

    for anti-clockwise orientation

    2) Lissahous Pattern in IInd & IVth Quadrant

    2

    1 1X

    180 sinX

    2

    1 1Y

    180 sinY

    for clockwise orientation

    for anti-clockwise orientation = 360

    Case 2

    x yw w

    x m xV V sinw t

    y m yV V sinw t

    y y

    x x

    w f Number of horizental tangencies

    w f Number of vertical tangencies

    y

    x

    f 42

    f 2

  • Digital Meters

    Type of converter Maximum Conversion Time

    1) Dual slope ADC n 12 Clocks

    2) Successive Approximation Register (ADC) n Clocks

    3) Counter ADC n2 Clocks

    4) Flash ADC 1 Clock

    Dual slope A/D Converter

    aV = analog input

    R

    V = Reference input

    Ra 2 11

    VV T T

    T

    1n

    CLKT 2 T

    Maximum conversion time = n 1

    CLK2 T

    Successive Approximation Register

    Suppose = REF aV 1 V

    and aV = 12V

    3D 2D 1D 0D

    10 5 2.5 1.25

    1

    T 1 0 0 0 10V < 12V

    2T 1 1 0 0 15V > 12 V

    3T 1 0 1 0 12.5 > 12 V

    4T 1 0 0 1 11.25 < 12 V

    In first clock cycle, MSB is set to get voltage corresponding to the digital o/p

    If 0

    V < aV , then in next cycle next bit is set else,

    If 0

    V > aV , MSB is reset & next bit is set

    We continue the same process till we reach LSB.

  • Specifications of Digital Voltmeter

    1) Resolution

    The smallest value of input that can be measured by digital meter is called resolution.

    n

    1R

    10

    n = No. of full Digits (0, 1,.., 9)

    2) Sensitivity

    S = Resolution x Range

    3) Over Ranging

    The extra 12

    digit is called over-ranging

    If n = 3, we can measure from 0 999

    Resolution , 3

    1R 0.001

    10

    if 1n 32

    digit, 12

    digit can be 0 & 1.

    we can measure from 0 1999

    Resolution, 1

    R 0.0052000

    if 34

    digit is there than MSB can be 0 3.

    4) Total Error

    Error = (% error in reading) x reading + (NO. of counts) Full Scale

    Range of meter

    Q meter / Voltage Magnifier

    If works on the principal of series resonance.

    At series resonance

    L CX X

    V

    IR

    CCV IX

    LCX X

    V VR R

    C

    V = V. Q

    CV Q

  • Practical Q-meter

    Also includes series resistance of source (oscillator)

    True T

    wLQ

    R

    Measured Q,

    Tm

    sh shsh

    QwL wLQ

    R RR RR 1 1

    R R

    msh

    T

    RQ Q 1

    R

    Measurement of unknown capacitance

    Test capacitance is connected at 43T & T .

    Circuit is resonated at C = 1

    C

    fr= T1

    1

    2 2 C C (1)

    TC = Test Capacitance

    TC is removed & circuit is resonated at C = 2C

    fr =

    2

    1

    2 LC (2)

    from (1) & (2)

    T 2 1C C C

    Measurement OF self Capacitance

    Resonance is achieved at C = 1

    C

    1

    1 d

    1fr

    2 L C C

  • At C = 2

    C , resonance is achieved at 2

    fr

    2

    2 d

    1fr

    2 L C C

    = n fr1,

    21 2

    d 2

    C n CC

    n 1

  • EMI.pdfEMI_final.pdfBack Page.pdf