extending mirqc’s dynamic range: amplifying the view of limiting rna samples - download the...
TRANSCRIPT
Scientific article
Abstract: The original microRNA quality control (miRQC) study provided an in-depth analysis of
commercially available microRNA (miRNA) quantification platforms. Specifically, twelve different
microarray, real-time PCR and small RNA sequencing platforms were assessed for reproducibility,
sensitivity, accuracy, specificity and concordance of differential expression using a variety of
sample types. Overall, each platform exhibited specific strengths and weaknesses, leading to the
final suggestion that a platform should be chosen on the basis of the experimental setting and the
specific research questions. With this suggestion in mind, and the fact that liquid miRNA biopsies
are an area of intense interest, we sought to expand the original miRQC study. For our “miRQC
extension,” we benchmarked the QIAGEN miScript® PCR System with and without preamplification,
and included a specific focus on routinely used biofluids. Concurrently, we benchmarked the
miScript PCR System against another SYBR® Green miRNA detection platform. Overall, QIAGEN
miScript demonstrated strong reproducibility and accuracy as well as superior detection rate
and sensitivity in biofluids. Collectively, QIAGEN miScript provides the leading solution for novel
miRNA discoveries.
Brian Dugan and Jonathan M. Shaffer
QIAGEN Sciences Inc., Frederick, MD, USA
Extending miRQC’s dynamic range: amplifying the view of limiting RNA samples
Contents
Introduction __________________________________________ 1
Materials and methods ________________________________ 2
Samples ________________________________________ 2
miRNA quantification _____________________________ 2
Data pre-processing ______________________________ 3
Results and discussion _________________________________ 3
miScript PCR System: reproducibility and accuracy ____ 3
Comparison of QIAGEN and Exiqon platforms _______ 4
Reproducibility, accuracy and titration response ______ 5
Detection rate and sensitivity _______________________ 6
miRNA detection in biofluids _______________________ 7
Detection of miRNA from equal sample inputs ________ 7
Conclusions __________________________________________ 9
References __________________________________________ 12
Introduction
miRNAs are approximately 21 nucleotide small noncoding
RNAs that play a prominent role in virtually all normal and
disease biological processes. Originally discovered only 20 years
ago, miRNA changes have been correlated with gene expression
changes in development, differentiation, signal transduction,
infection, aging and disease. Specifically in cancer, miRNA
have been associated with cell proliferation, resistance to
apoptosis, invasiveness and differentiation. Interestingly, miRNAs
exhibit stable expression in circulation and continually growing
evidence associates circulating miRNAs with normal and disease
biology. As a result, miRNAs are currently being explored
as minimally invasive biomarkers for disease detection and
monitoring using serum, plasma and cerebrospinal fluid (CSF),
as well as virtually all biofluids (1).
For circulating miRNA expression analysis, where miRNAs are
often expressed at very low levels and cDNAs are analyzed
www.qiagen.com2 QIAGEN
for the quantification of miRNA including reproducibility,
sensitivity, accuracy, specificity and concordance of differential
expression. In the study, 12 different microarray, real-time PCR
and small RNA sequencing platforms were benchmarked using
20 standardized positive and negative control samples,
including human universal reference RNA, human brain RNA and
titrations thereof, human serum samples and synthetic spikes from
miRNA family members with varying homology. Overall, each
platform exhibited various strengths. Specifically, QIAGEN’s
miScript PCR System showed high levels of reproducibility,
accuracy and sensitivity, which are key features important to
every miRNA quantification study. However, the conclusions
were drawn using only 372 of miScript’s industry leading 2400
primer assays, and miScript’s preamplification platform was not
benchmarked. As a result, Biogazelle and QIAGEN teamed
together to extend the miRQC study to the entire portfolio of
miScript human assays, benchmark the miScript preamplification
workflow and specifically focus on the quantification of miRNA
from biofluids. Concurrently, the QIAGEN miScript PCR System
was compared with the other main SYBR Green platform; the
Exiqon miRCURY LNA™ microRNA qPCR System.
Materials and methods
Samples
Four reference total RNA samples from the original miRQC
study were tested, including miRQC A (Agilent Technologies:
Universal Human miRNA Reference RNA), miRQC B (Agilent
Technologies: human brain RNA), miRQC C (75% miRQC A +
25% miRQC B) and miRQC D (25% miRQC A + 75% miRQC B).
Total RNA was isolated from serum (BioreclamationIVT) and CSF
(generously provided by Professor James Dear, The University
of Edinburgh) samples using the miRNeasy Serum/Plasma Kit
(QIAGEN).
miRNA quantification
All experiments were performed by Biogazelle (Ghent, Belgium).
Three platforms were used for miRNA quantification. First, the
QIAGEN miScript PCR System without preamplification was
used in conjunction with the miScript miRBase Profiler (miRNome)
Arrays (2402 total assays). Second, the miScript PCR System
with preamplification was used in conjunction with the miScript
miRBase Profiler miRNome Arrays. Third, the Exiqon miRCURY
across hundreds of qPCR wells, special considerations need to
be taken, paramount being that miRNA molecules must be at
a high enough copy per qPCR reaction to reduce the effect of
sampling ambiguity (i.e., technical variability). According to
Ståhlberg and Kubista, the minimal number of target molecules
per qPCR well must be 35 to reduce sampling ambiguity (2).
Consider the following two scenarios for quantification of
a miRNA present at 100 copies. In scenario 1, which does
not include preamplification, the 100-copy miRNA is reverse-
transcribed, and the cDNA is applied to a plate of 384 miRNA
assays. This results in the 100-copy miRNA being present
at 0.26 copies per qPCR well, which is markedly below the
recommended minimum copy number for efficient, reproducible
quantification. In scenario 2, which includes 12 cycles of
intermediate amplification after cDNA synthesis, the 100-copy
target would be present at more than 500 copies per qPCR
well. This is well within the recommended range for efficient,
reproducible quantification, and enables efficient, reproducible
quantification. If preamplification is not included in biofluid
analyses, particularly for CSF, expression will only be observed
for the most robustly expressed miRNAs, leaving you with a
truncated view of the full expression profile of your sample;
ultimately, you will be missing data.
The QIAGEN miScript PCR System provides various platforms
and optimized workflows to quantify any miRNA from any
sample type, whether it is derived from cells, fresh or frozen
tissue, FFPE tissue or biofluids. Specifically for low RNA content
samples, such as biofluids, preamplification options are
available that enable robust quantification from the lowest RNA
amounts, including cell-free nucleic acid, miRNA encapsulated
in exosomes and even single cells. Each of the miScript platforms
can be paired with bench-verified individual miScript Primer
Assays or pre-formatted miScript miRNA PCR Arrays to enable
quantification of over 2400 annotated human miRNAs. In
addition to human, bench-verified assays are available for
common model organisms, including but not limited to mouse,
rat, dog and cow. Additionally, custom assays can be derived
for any miRNA of interest, including novel miRNAs identified
during small RNA sequencing. The miScript System offers the
most comprehensive sample, species and assay solutions for
quantitation studies.
The original microRNA quality control (miRQC) study represented
the first comprehensive assessment of the factors most important
Scientific article www.qiagen.com 3
LNA microRNA qPCR System was used in conjunction with
the Exiqon microRNA Ready-to-Use PCR, Human panel I+II,
V4.M. The QIAGEN miScript PCR System includes up to 2402
miRNA assays across seven 384-well PCR arrays, while the
Exiqon system consists of up to 752 assay targets across two
384-well plates. Both platforms utilize SYBR Green as well as
a universal reverse-transcription reaction. All experiments were
done in technical duplicate and performed according to the
manufacturer’s specific instructions unless indicated otherwise.
Data pre-processing
Raw CT: values were processed as described in the miRQC
study. Briefly, a detection cutoff was defined based on the
fraction of single positives (i.e., miRNAs only expressed in
one of two replicates) of replicate expression profiles. The CT:
cutoff was chosen as such that it eliminates 95% of the single
positives. For the QIAGEN platforms, two different CT: cutoffs
were defined, one for the workflow without pre-amplification
(QNP) and one for the workflow with pre-amplification (QP).
The CT: cutoff for the workflow without pre-amplification was
based on replicate CT: values from all four miRQC samples. This
resulted in a CT: cutoff of 31.86 cycles. A similar CT: cutoff was
obtained when considering replicate serum samples profiled
without pre-amplification (CT = 31.35) suggesting that the
algorithm applied to determine the CT: cutoff is robust and
sample-independent. For the workflow with pre-amplification,
replicate serum samples were considered, resulting in a CT: cutoff
of 30.05. For the Exiqon platform, a CT: cutoff of 34.92 was
obtained when considering replicate miRQC samples. These
cutoffs were applied throughout the entire data analysis
workflow. Data were normalized using the global mean as
described in the original miRQC study. For details about the
performance parameters and how they were calculated, refer
to the original miRQC study (3).
Results and discussion
The original miRQC study tested multiple methodologies
for miRNA quantification, including qPCR and small RNA
sequencing. Each platform exhibited specific strengths and
weaknesses, leading to the final suggestion that a platform
should be chosen on the basis of the experimental setting and
the specific research questions. With this suggestion in mind,
and the fact that liquid miRNA biopsies are an area of intense
interest, we sought to expand the original miRQC study. For our
“miRQC extension,” we benchmarked the QIAGEN miScript
PCR System and the QIAGEN miScript PCR System with
preamplification. In addition, we incorporated routine biofluids
such as serum and CSF in the testing. Concurrently, we
benchmarked the miScript PCR System against the Exiqon
miRCURY LNA microRNA qPCR System. Overall, this miRQC
extension was intended to measure key metrics paramount for
successful miRNA quantification studies of the most relevant
sample types including reproducibility, accuracy, detection rate
and sensitivity in biofluids. The size of the QIAGEN miRNome
and the Exiqon miRNome vary substantially, with QIAGEN
offering over 2400 assays in preformatted arrays, while Exiqon
only offers 752. With over 1700 more miRNA targets, there is
greater opportunity for novel discovery outside the original 750
miRNA set. Due to Exiqon offering only 752 assays, a common
set of 746 miRNAs was selected for the direct comparison
experiments.
miScript PCR System: reproducibility and accuracy
A high-performance miRNA quantification platform should
exhibit strong reproducibility and accuracy. In practice, this
enables high confidence that “observed results” are in fact “real
results.” To determine reproducibility, miRQC sample C (75%
miRQC A + 25% miRQC B) and miRQC D (25% miRQC A + 75%
miRQC B) were quantified and analyzed for the replicate
expression difference (log2) using the miScript PCR System
without preamplification and the full 2402 assay miRNome.
Figure 1, panel A shows a replicate expression correlation plot.
Both replicate 1 and replicate 2 show high levels of concordance
and extremely low variability sample to sample. Panel B depicts
the cumulative distribution of replicate expression differences.
This is measuring the deviation in the observed replicate CT values
from replicate 1 to replicate 2. The goal is to limit expression
differences between replicates, and the lower the value the
better. In the plot, the area left of the curve is measured and
correlated to an average fold change. Here the observed ALC
between both replicates was 0.346, or an average of only
1.27-fold replicate expression difference. Collectively, this
figure demonstrates that the miScript PCR System, across the
entire portfolio of miScript miRNA assays, enables highly
reproducible results.
www.qiagen.com4 QIAGEN
Accuracy of the miScript PCR System without preamplification
was evaluated using miRNAs from the total 2402 assay list that
were only expressed in either miRQC A or miRQC B (as these
should theoretically have a 3-fold difference in expression
between miRQC C and D). Figure 2 shows the fold change
Replicate 222
–6
–6 22Replicate 1 (log2)
Replicate expression difference (log2)
Cumulative fraction (%)1.0
0.8
0.6
0.4
0.2
0
0 0.80.60.40.2 1.0
ALC = 0.346
Figure 1. Assessment of miScript PCR System reproducibility. A Correlation of miRNA expression between miRQC samples C and D and their replicates. B ALC-value (Area Left of the Curve) of 0.346 (corresponding to an average 1.27-fold replicate expression difference) was obtained.
Figure 2. Assessment of miScript PCR System accuracy. Fold-change (miRQC C/D or miRQC D/C) are shown for miRNAs only expressed in miRQC A or miRQC B. The dashed line indicates the theoretical 3-fold fold change, and the black line indicates the median fold-change.
MAQC C/D (D/C)8
6
4
2
0quartile 1-4 quartile 2-4
1st quartileExpression
2nd quartile 3rd quartile 4th quartile
between C and D when considering all miRQC A/B-specific
miRNAs (quartile 1–4) or only those miRQC A/B-specific miRNAs
with expression values in the upper 3 quartiles (quartile 2–4).
The theoretical fold change (3-fold) is indicated by a dashed
line, and the median fold-change is indicated by the black
line. Whether quartile 1–4 or only quartile 2–4 was assessed,
the median fold difference was close to the theoretical 3-fold
change demonstrating high accuracy of the miScript PCR System.
Comparison of QIAGEN and Exiqon platforms
Accurately quantifying miRNA is a challenge, but there are a
number of key aspects that differentiate SYBR Green and probe-
based systems. First, both the QIAGEN and Exiqon systems
employ universal reverse-transcription. By using universal reverse-
transcription, all miRNA can be converted to cDNA. In addition,
as referenced in the miRQC study, probe-based qPCR, Taqman,
showed the lowest reproducibility when compared to SYBR Green
miRNA qPCR. In addition, SYBR green based technologies showed
superior reproducible detection of miRNA’s from miRQC samples.
To benchmark the QIAGEN miScript PCR System, another SYBR
Green platform, the Exiqon miRCURY LNA qPCR System was
chosen. Both the QIAGEN and Exiqon systems are universal
reverse-transcription platforms and use SYBR Green. The goal
was to determine specific differences in performance of the
technologies. QIAGEN currently offers more than 3x the number
of assays for miRNA profiling, so only the overlapping targets
were compared for this discussion.
In order to directly compare the platforms, a common set of
748 assays was selected, and reproducibility, accuracy and
titration response were directly compared. Further, the ability to
reproducibly detect miRNAs from serum and CSF were also
compared. Finally, since the standard miScript PCR System
(without preamplification) and the Exiqon platform suggest
different starting amounts of RNA, each platform was run with
each suggested starting amount, 20 ng per 384-well plate
(recommended by Exiqon for quantifiable RNA from cells and
tissues) and 500 ng per 384-well plate (recommended by
QIAGEN for quantifiable RNA from cells and tissues). With these
varying input recommendations, there is supposition that the
starting input directly affects the ability to detect miRNA in samples.
With this last test, we are testing sensitivity and debunking
the theory of whether or not one platform offers improved
sensitivity versus the other when preamplification is not included.
Scientific article www.qiagen.com 5
Reproducibility, accuracy and titration response
In Figure 3, the reproducibility of the QIAGEN platform without
preamplification is compared to the Exiqon platform using
miRQC samples A–B. Panels 3A and 3B show the replicate
correlation plots for miScript and Exiqon, respectively. For each
platform, replicates 1 and 2 show high levels of concordance
and extremely low variability sample to sample. Panels 3C and
3D depict the cumulative distribution of replicate expression
differences, and measures the deviation in the observed
replicate CT values from replicate 1 to replicate 2. In the plot,
the area left of the curve is measured and correlated to an
average fold change. Here the observed ALC between both
replicates was very good for both platforms (< 0.4) with
miScript exhibiting a slight edge in reproducibility.
Figure 3. Assessment of miScript PCR System reproducibility. Correlation of miRNA expression between miRQC samples A-D for the QIAGEN A and Exiqon B platforms. C and D ALC-values, which are measurements for reproducibility, were determined for both the QIAGEN and Exiqon platforms.
Replicate 2
QIAGEN
22
–6
–6 22Replicate 1 (log2)
Replicate 2
Exiqon
12
–8
–8 12Replicate 1 (log2)
Replicate expression difference (log2)
Cumulative fraction (%)1.0
0.8
0.6
0.4
0.2
0
0 0.80.60.40.2 1.0
ALC0 0.80.60.40.2 1.0
Figure 4. Assessment of QIAGEN and Exiqon platform accuracy using a common set of assays. Fold-change (miRQC C/D or miRQC D/C) are shown for miRNAs only expressed in miRQC A or miRQC B.The dashed line indicates the theoretical 3-fold fold change, and the black line indicates the median fold-change. In addition, the media deviation from the expected C/D or D/C ratio is shown.
MAQC C/D (D/C)8
6
4
2
0quartile 1-4
QIAGEN Exiqon
median deviation1.1 1.41.31.2 1.5
MAQC C/D (D/C)8
6
4
2
0quartile 2-4
QIAGEN Exiqon
median deviation1.1 1.41.31.2 1.5
1st quartileExpression
2nd quartile 3rd quartile 4th quartile
www.qiagen.com6 QIAGEN
Accuracy of the QIAGEN and Exiqon platforms were evaluated
using miRNAs that were only expressed in either miRQC A or
miRQCB (as these should theoretically have a 3-fold difference
in expression between miRQC C and D). Figure 4 illustrates the
fold change between C and D when considering all miRQC
A/B-specific miRNAs (quartile 1–4) or only those miRQC A/B-
specific miRNAs with expression values in the upper 3 quartiles
(quartile 2–4). The theoretical fold change (3-fold) is indicated
by a dashed line, and the median fold-change is indicated by
the black line. Whether quartile 1–4 or only quartile 2–4 was
assessed, the median fold difference was close to the theoretical
3-fold change demonstrating comparable accuracy of both
QIAGEN and Exiqon platforms.
Titration response is the ability of a miRNA quantification
platform to call results “correctly.” Figure 5 summarizes the results
of titration response testing for both the QIAGEN and the Exiqon
platforms. Assessments of titration response rely on expression
patterns of miRNAs in all four miRQC samples A–D, and the
theory is tied to the ratio of the miRQC C (75% A + 25% B) and
miRQC D (25% A + 75% B) samples. For example, if “miRNA 1”
has greater expression in miRQC A than miRQC B, then
AmiRNA 1 > CmiRNA 1 > DmiRNA 1 > BmiRNA 1. Titration
response is the ability to detect the samples in the correct order,
and the fraction of “titrating miRNAs” can be determined.
Panels 5A and 5B summarize the fraction of titrating miRNAs
of the QIAGEN and Exiqon platforms, respectively, across bins
of increasing expression difference between samples miRQC A
and miRQC B (x-axis). The titration response was then plotted
as a function of the cumulative bin size (Panel 5C), and the area
under the curve (AUC: Panel 5D) was determined. An AUC = 1
would indicate that all miRNAs are titrating. Both QIAGEN
and Exiqon demonstrate high titration responses with both have
AUC values close to 0.8.
Detection rate and sensitivity
In order to discover novel miRNA signatures from low RNA
content samples, such as biofluids, a platform should have
high detection sensitivity. Detection rate and sensitivity can
be assessed by measuring the number of double positives as
well as single positives during expression profiling of replicate
samples. The higher the number of double positives and the
lower the number of single positives results in a higher detection
rate sensitivity. For this analysis, miRQC samples A–D were
included. In Figure 6, miScript displayed a substantially higher
number of double positives (50%: 600 to 400) and a lower
percentage of single positives (6% vs 10%) when compared to
Exiqon. Together, these results demonstrate a superior detection
rate and enhanced sensitivity for QIAGEN platform compared
to the Exiqon platform.
Figure 5. Assessment of QIAGEN and Exiqon titration response. For the A QIAGEN and B Exiqon platforms, the fraction of titrating miRNAs is plotted across bins of increasing expression difference between miRQCA and miRQCB. C The titration response for the miRNAs is plotted. D The AUC values, which reflect titration response, demonstrate the titration response is virtually identical for both platforms (QIAGENt: ~0.788 & Exiqon: ~0.792)
0.75 0.76 0.77 0.78 0.79 0.80AUC
QIAGENFraction titrading miRNAs 1
0
–3 30Expression difference (MAQCA-MAQCB)
Expression difference (MAQCA-MAQCB)
ExiqonFraction titrading miRNAs 1
0
–3 30
FC ranked miRNAs as fraction of total
Fraction titrading miRNAs 1
0
0 10
0.75 0.76 0.77 0.78 0.79 0.80AUC
QIAGENFraction titrading miRNAs 1
0
–3 30Expression difference (MAQCA-MAQCB)
Expression difference (MAQCA-MAQCB)
ExiqonFraction titrading miRNAs 1
0
–3 30
FC ranked miRNAs as fraction of total
Fraction titrading miRNAs 1
0
0 10
Scientific article www.qiagen.com 7
miRNA detection in biofluids
Thus far, the head-to-head analyses suggest a competitive
advantage for QIAGEN with respect to detection rate and
sensitivity, both of which are critical for biofluid detection.
However, these tests were performed on non-limiting amounts
of RNA, and may not reflect the realities of detection from low
RNA content samples, such as serum and CSF. This is becoming
increasingly important, as many applications are currently
focused on the ability to detect biomarkers, including miRNA,
through the use of low sample input or less invasive techniques,
which is driving an evolution in research testing. One of these
approaches is liquid biopsy, or the ability to detect analytes
through blood-based testing. To test these real world type
samples, both platforms were compared in their ability to detect
the conserved set of miRNA targets across platforms.
For this experiment, the ability to reproducibly detect miRNA
from two CSF samples as well as two serum samples was
compared (Figure 7). In a comparison of two CSF samples, CSF
A and CSF B there was a significant difference in the number
of miRNA reproducibly detected when comparing platforms. In
CSF A, for example, Exiqon detected 36 miRNA. From the
same sample, which was isolated using the same technique, the
QIAGEN miScript system was able to detect 150 miRNA –
almost 5x greater miRNA than Exiqon. The same was true for
CSF B sample as Exiqon detected 24 miRNA while QIAGEN
again detected 5x the amount of miRNA at 127. Serum samples
were tested under similar conditions, an A and B replicate were
tested across both platforms. Here again, the detection sensitivity
was measured using the number of like miRNA that were able
to be detected from the same samples. QIAGEN miScript
detected miRNA at 2x greater sensitivity, in Serum A sample
247 miRNA were detected for the miScript system while 90
were detected using Exiqon miRCURY. Serum B sample yielded
a similar result, as QIAGEN miScript detected 287 compared
to MiRCURY’s 98.
While many aspects of both the QIAGEN and Exiqon per-
formance were comparable for reference samples, such as
reproducibility, accuracy and titration response, there was a
clear difference in limited sample performance. Detection of
nucleic acids in biofluids can be highly variable. As a result,
sensitivity in these fluids is a key to detecting and quantifying
these miRNA.
Detection of miRNA from equal sample inputs
Sample input amount/concentration may be a key factor in
detection sensitivity. Theoretically, the lower the minimum input
level, the assay may be assumed to exhibit greater sensitivity.
As sample input concentrations decrease, the number of available
copies decreases. This holds true in comparing the difference
between a 500 ng input and a 20 ng input amount.
This is an important feature in the comparison of miRNA
quantification systems, as lower sample input levels can be
Figure 7. Assessment of QIAGEN and Exiqon detection from common biofluids. For CSF samples A and Serum samples B, the number of reproducibly detected miRNAs were determined for replicate samples.
Figure 6. Assessment of QIAGEN and Exiqon detection rate and sensitivity. Across the replicates of all four miRQC samples (A-D), the number of double positives A and percentage of single positives B was determined.
Reproducibly detected miRNAs
CSF
Serum
150125100
755025
0CSF A CSF B CSF A CSF B
QIAGEN Exiqon
Reproducibly detected miRNAs 300250200150100
500
Serum A Serum B Serum A Serum B
QIAGEN Exiqon
0 200 400 600 800
Single positives (%)
Double positives (%)
0 2 4 6 8 10 12 14 16
www.qiagen.com8 QIAGEN
perceived to offer greater sensitivity when compared to kits with
higher recommended levels. To test this theory, QIAGEN again
employed Biogazelle to compare the detection of a like set of
miRNA, 750 targets, with two distinct sample input levels. Both
platforms were tested across the miRQC A reference samples.
This time each platform was tested at the QIAGEN protocol
input, 500 ng, as well as the Exiqon protocol input, 20 ng.
Both arrays were then performed with qPCR and analysis being
conducted by Biogazelle. By looking at both sets of array
performance, QIAGEN shows greater detection sensitivity for
the reference sample at both input levels (Figure 8). For 500 ng
of input, QIAGEN detects roughly 10% more miRNA 555 to
517. In demonstration of the flexibility of the miScript platform,
at the Exiqon recommended levels of 20 ng, the miScript system
demonstrated a 25% greater detection, 424 to 344 observed
miRNA.
While the performance of both systems shows that increased
sample input will yield increased miRNA detection, the question
remains – does the lower input amount lead to greater sensitivity?
By comparing the recommended input levels of both platforms
against each other, again miScript considerably out performs
Exiqon MiRCURY. When looking at how many miRNA are
detected from the reference sample, miScript shows an almost
2 to 1 advantage.
Number of miRNA 600500400300200100
0QIAGEN InputLevel (500 ng)
Exiqon InputLevel (20 ng)
QIAGEN Exiqon
Number of miRNA 600500400300200100
0miRNA detected by both recommended protocols
QIAGEN Exiqon
Figure 8. Assessment of miRNA detection from equal sample input. A The number of miRNA detected for both the recommended QIAGEN RNA Input (500 ng per 384-well plate) and Exiqon RNA Input (20 ng per 384-well plate) were determined. B Number of miRNAs detected for the QIAGEN and Exiqon platforms when their respective RNA amount input requirement is followed.
Scientific article www.qiagen.com 9
Conclusion
The purpose of this study was to extend the miRQC study of
the QIAGEN miScript PCR System, including a comparison to
a similar SYBR Green platform. In doing this, both platforms’
ability to reproducibly detect and quantify miRNA was
benchmarked on multiple sample types, at multiple input levels
and compared across a number of key performance metrics.
The data demonstrates that the QIAGEN platform has superior
performance, particular when compared to the Exiqon platform
during detection rate, sensitivity and biofluid detection, with
respect to reproducibility, accuracy, detection of miRNA in
body fluids and detection sensitivity at miRNA detected per
nanogram of input.
Limiting samples including the liquid biopsy samples are the
focus of many studies looking to identify potential biomarkers
or utilize miRNA as possible diagnostic or prognostic markers.
By looking at serum and CSF, these limiting inputs were tested to
demonstrate real world application performance. Here again,
QIAGEN miScript showed superior detection sensitivity in CSF
where 5x miRNA were detected on average versus MiRCURY
as well in serum, where miScript detected 3x the number of
miRNA targets.
Finally, discussions of sensitivity may often include the input
levels into quantification experiments. By comparing two
different sample input levels, 500 ng and 20 ng, both platforms’
performance were assessed. Here again, miScript showed
greater sensitivity at both the 500 ng and 20 ng levels.
QIAGEN miScript miRNA PCR System offers superior sensitivity,
reproducibility and accuracy in miRNA quantification from
samples as low as a single cell to any tissue type. In addition,
miScript offers the most comprehensive target list with the
largest miRNome, most species (animals and plants) and the
largest collection of cataloged pathway- and disease-focused
PCR arrays available. QIAGEN miScript is the most flexible,
most accurate and most sensitive miRNA quantification platform
available today.
www.qiagen.com10 QIAGEN
Scientific article www.qiagen.com 11
For up-to-date licensing information and product-specific disclaimers, see the respective QIAGEN kit handbook or user
manual. QIAGEN kit handbooks and user manuals are available at www.qiagen.com or can be requested from QIAGEN
Technical Services or your local distributor.
For more information on QIAGEN’s miRNA portfolio, visit www.qiagen.com/miScript.
Trademarks: QIAGEN®, Sample to Insight®, miScript® (QIAGEN Group), SYBR® (Life Technologies Corporation), Exiqon®, miRCURY®, LNA™ (Exiqon).
© 2016 QIAGEN, all rights reserved. PROM-9221-001
Ordering www.qiagen.com/contact Technical Support support.qiagen.com Website www.qiagen.com
Ordering Information
Product Contents Cat. no.
miRNeasy Serum Plasma Isolate total RNA from serum, plasma and Biofluids 217184
miRNeasy Mini Kit Isolate high quality total RNA from animal tissues and cells 217004
miScript miRNome PCR Array Profile all available miRNA from human, mouse or rat – V16 331222
miScript miRNA miRBase Profiler Profile all the most up to date human miRNome, 2402 Targets Available in 384-well format
331223
miScript PreAMP Primer Mix Reagents and specific assays for amplification of miRNA targets for low sample input
331241
miScript PreAMP PCR Kit 331452
miScript Single Cell qPCR Kit Universal preamplification of all miRNA from as low as a single cell. Can also be used for ultra-low sample amounts greater than an individual cell
331053
1099072 01/2016
References
1. Weber, J.A. et al. (2010) The microRNA spectrum in 12 body fluids. Clinical Chemistry 56,1733.
2. Ståhlberg, A. and Kubista, M. (2014) The workflow of single-cell expression profiling using quantitative real-time PCR. Expert. Rev. Mol Diagn. 14, 323.
3. Mestdagh, P. et al. (2014) Evaluation of quantitative miRNA expression platforms in the microRNA quality control (miRQC) study. Nature Methods 11, 971.