experiment 7 - determination of caffeine and vitamin b6 in energy drinks using hplc
DESCRIPTION
hplc experimentTRANSCRIPT
*Based on the procedure outlined in:
Leacock, R. E.; Stankus, J. J.; Davis, J. M. Simultaneous Determination of Caffeine and Vitamin B6 in
Energy Drinks by High-Performance Liquid Chromatography (HPLC). J. Chem. Educ. 2010, 88, 232-234.
Chemistry 315
Experiment 7:
Determination of Caffeine and Vitamin B6 in Energy Drinks using
High-Performance Liquid Chromatography*
In this experiment you will calculate the amount of caffeine and vitamin B6 (pyridoxine
hydrochloride) in an energy drink using the standard addition method by high
performance liquid chromatography (HPLC).
Required Reading
1. B. L. Karger, “HPLC: Early and Recent Perspectives”, J. Chem. Educ. 1997, 74, 45.
DOI: 10.1021/ed074p45
2. Chapter 26, “An Introduction to Chromatographic Separations”, Skoog, et al.
3. Chapter 28, “High Performance Liquid Chromatography”, Skoog, et al.
4. Section 1D-3, “Standard Addition Methods”, Skoog et al.
or
5. Section 5.1, “Fundamentals of Chromatographic Separations”, pp. 159-170, Kellner
et al.
6. Section 5.3, “Liquid Chromatography”, pp. 185-208, Kellner, et al.
7. L. R. Snyder, “Modern Practice of Liquid Chromatography, Before and After 1971”,
J. Chem. Educ. 1997, 74, 37.
8. Section 12.2.7, pp. 743-744 (standard addition method), Kellner et al.
2
Introduction
Energy drink consumption has increased exponentially since the beverages were
first introduced in the 1960s. Compared to a traditional cup of coffee, which contains
between 77 and 150 mg of caffeine, energy drinks can have as much as 505 mg of
caffeine per can. The United States Food and Drug Administration (FDA) set a limit to
the amount of caffeine a standard soda can contain of 200 parts per million (ppm) for a
standard soda.1,2
In addition to providing the standard nutritional information on the label
of a food or beverage product, it is important to be able to quantify the amounts of toxic
substances, such as caffeine3 present in consumer products prior to marketing. High-
performance liquid chromatography (HPLC) will be used to simultaneously determine
the amount of caffeine and vitamin B6 (pyridoxine hydrochloride) present in an energy
drink and compare the results to the amounts printed on the label.
Basic Chromatographic Concepts
Chromatography allows for the separation of analytes from a complex sample
matrix based upon their adsorption affinities to a support. There are many different
methods of performing chromatographic separations, but all depend upon the partition of
an analyte between the mobile phase and stationary phase. The classification of
chromatographic techniques is often divided into two major groups: gas chromatography
and liquid chromatography.
For this experiment, the main focus will be on liquid chromatography, but the
same general principles apply to gas chromatography. In standard column
chromatography, a liquid solution is poured through a column containing silica gel. The
solvent acts as the mobile phase and carries the analytes of interest down the column due
to gravity while the silica gel acts as the stationary phase. Because silica gel contains
alcohol groups, polar components have higher affinities for adsorbing to the stationary
phase and elute from the column later than those components that are nonpolar. The time
required for a component to elute is termed the retention time. Thin layer
chromatography follows the same principles as column chromatography, but the mobile
phase is drawn up the side of the film by capillary action, rather than downward due to
gravity. The use of a polar stationary phase is known as normal-phase chromatography,
but reverse-phase chromatography, containing a non-polar stationary phase, is much
more common. Size exclusion and ion exchange supports can also be used to separate
chemical compounds in column chromatography.
The ability for a given column to separate peaks is known as the column
efficiency and can be quantified as the height equivalent of a theoretical plate (HETP).
This is often related to the length of the column according to the equation:
where N is the number of theoretical plates, L is the length of the column in cm, and H is
the height of the theoretical plates in cm. This provides a useful number for comparing
different columns when considering which one to use for a given separation.
Experimentally, this can be calculated according to the following equation:
( ⁄
)
3
where tR is retention time in seconds and W1/2 is the width of the peak at half of the height
of the peak in cm.
An additional parameter useful for characterizing separations is resolution, or the
distance between two peaks relative to the widths of the peaks. This may be calculated as
[( ) ( ) ]
where A and B are two different components, tR is their retention times in s, and W is the
width of the peaks in cm. Components that elute at very similar times or that have overly
broad peaks will exhibit very poor resolution. Optimization experiments are often
performed to increase the resolution of a separation. Such experiments include changes in
solvent viscosity, changes to flow rate, and even changes in temperature over the duration
of the elution.
Chromatography, in itself, is not a detection technique. It is often coupled to some
type of detector, such as a mass spectrometer or, in the case of this experiment, a UV-
visible spectrometer.
High-Performance Liquid Chromatography (HPLC)
Due to the dependence on gravity for flow, column chromatography requires a
significant amount of time to complete a separation and can exhibit poor resolution
between components as the effects of longitudinal diffusion become more prominent. In
high-performance liquid chromatography, we take advantage of smaller stationary phase
particle sizes which provide more surface area for components to adsorb. A high-pressure
pump is required to draw the eluting solvent through the smaller pores, however,
increasing the cost of the instrumentation. The expense one must consider is offset by the
vast information one is able to obtain from HPLC, though, since it has the ability to
resolve different components from a very complex sample matrix, such as an energy
drink which contains high fructose corn syrup, food coloring, and various other additives.
The HPLC column material in this experiment is nonpolar (reverse-phase) porous
silica with covalently bonded organic side-chains (C18). Differences in binding affinity
of the vitamin B6 and caffeine to the C18 groups located on the column allow for higher
resolution between the two components in this experiment. For this experiment, an
isocratic elution will be performed using a mobile phase that consists of a 1:1
water:methanol + 1% acetic acid solution. The HPLC system in the lab is set up so that a
UV-visible detector, set at 292 nm, monitors the elution of the analytes.
Standard Addition
Because the sample matrix of the energy drink is so complex, the use of internal
standards cannot be used since there is no guarantee that certain peaks will not overlap
with each other. We will therefore explore the standard addition method in order to
determine the concentration of caffeine and vitamin B6 in the energy drink. In the
method of standard additions, one spikes the unknown sample with known concentrations
of a standard solution. By plotting the added concentration versus peak area or peak
height, one can extrapolate the best-fit line determined through linear regression analysis
to the x-axis. The absolute value of the x-intercept is the actual concentration of the
unknown.
4
Example:
References
(1) Nutrition, C. for F. S. and A. FDA Basics - Why isn’t the amount of caffeine a product contains required on a food label?
(2) Reissig, C. J.; Strain, E. C.; Griffiths, R. R. Caffeinated Energy Drinks -- A Growing Problem. Drug Alcohol Depend 2009, 99, 1-10.
(3) Peters, J. M. Factors Affecting Caffeine Toxicity. The Journal of Clinical Pharmacology 1967, 7, 131 -141.
y = 1000x + 20000 R² = 1
0
10000
20000
30000
40000
50000
60000
70000
-30 -20 -10 0 10 20 30 40 50
Pe
ak A
rea
Standard Concentration (ppm)
Unknown Conc = 20 ppm
5
Prelab assignment
1. Look up the MSDS for caffeine, pyridoxine hydrochloride, and methanol. What are
the hazards associated these chemicals? What actions should be taken should these
come into contact with your body (e.g. your skin/eyes)?
2. Calculate the required amounts in grams of caffeine and pyridoxine hydrochloride to
create the stock solutions. What are the concentrations in mg/L and molarity?
3. Complete the following table according to the experimental procedures:
Sample Volume
Caffeine
Stock
(mL)
Volume
Pyridoxine
HCL
Stock
(mL)
Total
Volume
(mL)
Standard
Caffeine
Concentration
(ppm)
Standard
Pyridoxine
HCl
Concentration
(ppm)
1
2
3
4
5
4. Draw a block (functional) diagram of the HPLC apparatus.
5. Draw the structure of the exterior of a silica gel particle for the column used in this
procedure.
6. Explain three reasons for band broadening in a chromatogram. Be sure to discuss the
van Deemter equation.
7. Which of the compounds (caffeine or pyridoxine HCl) do you expect to elute first?
Why?
6
Materials
Equipment:
Column – Agilent ZORBAX Eclipse XDB-C18
Guard Column – Agilent ZORBAX Eclipse XDB-C18
Pump – PerkinElmer Flexar Isocratic LC Pump
Detector –PerkinElmer Flexar UV-Vis LC Detector
7 × 100 mL volumetric flasks (for sample solutions)
1 × 5 mL volumetric pipet
1 × 1 mL volumetric pipets
1 × 2 mL volumetric pipets
1 × 3 mL volumetric pipets
1 × 4 mL volumetric pipets
2 × 100 mL Erlenmeyer flasks (for mixing standards)
1 stirring rod (for quantitative transfer between flasks)
1 × 50 mL beaker (for pipet rinse)
1 × 50 µL blunt-tipped syringe
Chemicals:
Caffeine
Pyridoxine hydrochloride
Methanol (in 20 mL scintillation vial for column rinse)
Distilled water
Miscellaneous:
Energy Drink
Nitrile gloves
Pasteur pipettes/bulbs (for topping off volumetric flasks)
7
Procedure
Special Notes:
Make sure to clean ALL glassware between each sample injection or dilution.
Everything except methanol may go down the drain. There is a waste container
provided for methanol waste.
Turn on the instrument and start the LC Pump when you first enter to allow for
proper equilibration
Standard Addition Preparation:
1. Prepare two stock solutions, one of approximately 650 ppm caffeine and the other of
approximately 500 ppm pyridoxine hydrochloride, in 100 mL volumetric flasks. The
weighed amounts do not have to be exact since you can calculate the actual
concentration of your final dilutions using the mass you record. Bring to volume with
de-ionized water. Ensure adequate mixing by inverting the volumetric flasks after
capping the top with the provided ground glass stoppers.
2. Pipet 5 mL of the energy drink to each of five 100 mL volumetric flasks. Pipet the
appropriate amount of stock solution to each flask according to the table below:
Sample Caffeine (mL) Pyridoxine HCL (mL)
1 0 0
2 1 1
3 2 2
4 3 3
5 4 4
Bring to volume with de-ionized water. Ensure adequate mixing by inverting the
volumetric flasks after capping the top with the provided ground glass stoppers.
Instrument Setup
1. Turn on the UV-Vis detector and the LC Pump by clicking the buttons on the front of
the instrument (the detector needs to be held down until it turns on). Open the
Chromera Manager software. Double-click “Honeybadger” under “labuser” in the
Configuration window. Wait for Chromera to load.
2. Under Control Mode on the left of the screen, make sure “Sequence” is selected.
8
3. Go to File > Open Sequence. Click the plus sign next to Sequence Group in the
window that opens. Select “Experiment 7” and then click “Open”.
4. Verify the Method and Sequence being used:
a. Click on Method on the bottom-left of the screen. Go to File > Open Method.
Click the plus sign next to Method Group in the window that opens. Select
“Experiment 7” and then click “Open”. Print the Method using the default
options.
b. Click on Sequence on the bottom-left of the screen. Go to File > Open
Sequence. Click the plus sign next to Sequence Group in the window that
opens. Select “Experiment 7” and then click “Open”. Print the sequence using
the default options.
c. Compare the edited dates on the printed sheets to those taped to the desk. If
they do not match, inform your TA. DO NOT continue the experiment
until these dates match.
5. Press the Play button ( ) under Sequence and wait for the Instrument window to
open. If the pump status is listed as “Shut down”, click “Start LC Pump” in the
Control Panel on the right of the screen. Wait for the pump to equilibrate if necessary.
Once you get the “Waiting for manual injection” message and you have verified that
the pump pressure is stable around 2600-3000 psi, you are ready to continue. You
will not have to pause or stop the sequence at any point unless you make a mistake.
6. Load the blunt-tipped syringe with the caffeine stock solution (slightly more than 20
μL) and remove any air bubbles by dipping the syringe into the sample solution and
quickly pumping the syringe by moving the plunger up and down until all bubbles are
removed.
7. Make sure that the valve on the left side of the detector is in the 8 o’clock LOAD
position and insert the syringe into the small, white circular opening on the front of
the valve. Push the needle in until the glass part of the syringe touches the white
opening. There may be some resistance toward the end.
8. Push in the plunger on the syringe and remove the syringe from the valve. This
LOADS the sample into the valve’s sample loop.
9
9. Quickly turn the valve clockwise toward the INJECT position. This injects the sample
into the HPLC column, zeroes the detector, and starts the sequence on the computer.
The sequence will take 10 minutes to complete. When the run is finished and the
chromatogram resets, turn the valve toward the 8 o’clock LOAD position.
Loop injector valve switching patterns. The load position allows the loop to be filled with sample. The
inject position injects the contents of the loop onto the column.
10. The pump may need to equilibrate between runs. Check the status window on the
bottom right of the screen to ensure that the Sequence Status is “Running”, the
Detector Status is “Ready”, and the Pump Status is “Ready” before continuing.
11. Repeat steps 6-10 with the Vitamin B6 stock solution, the standard addition samples,
and methanol rinse according to the order below. For the methanol steps, only inject
20 µL of methanol (provided in the scintillation vial) in order to clean the column.
1. Caffeine 7. Sample 3
2. Vitamin B6 8. Methanol
3. Sample 1 9. Sample 4
4. Methanol 10. Methanol
5. Sample 2 11. Sample 5
6. Methanol 12. Methanol
Make sure to rinse the syringe with the sample to be injected between each run to
prevent contamination.
The order is very important because the sequence automatically names each
sample. You can check which sample the program is waiting for by simply
looking at the list on the screen. If you make an error during an injection, allow
10
the run to finish. Then, stop the sequence by pressing the Stop button ( ). Select
the sample you would like to repeat by clicking on the row number to the left of
the sample (it is recommended to select the Column Rinse step right before the
erroneous sample to clean the column first). Once selected, click the Start Row
button ( ). All of the samples above the selected one should turn grey. Then,
click on the Play button ( ) and continue with the sample you selected.
12. The reports for your runs have automatically been saved in the “HPLC Reports”
folder on the desktop. Ensure that they have been saved and that each file contains a
summary of the data collected as well as an image of the chromatogram. You may
print the PDFs or save the Excel files to a flash drive to work with later. You must
attach each of your standard addition sample chromatograms and the final
column rinse chromatogram to the back of your report.
13. Click “Stop LC Pump” in the Control Panel and exit Chromera. Right click on the
Chromera icon on the bottom right of the screen, and close the Chromera Manager.
Turn off the pump module and the detector by holding down the power buttons on the
front of the instruments.
14. Clean your work station. Only solutions containing energy drink, caffeine, and
pyridoxine hydrochloride may go down the drain. Methanol must go in the waste
container. Rinse all glassware with distilled water.
For your Laboratory Report:
Results:
a) Identify the retention times of caffeine and pyridoxine hydrochloride.
b) Prepare plots of the areas for the caffeine peaks and the peak heights of the
caffeine peaks versus their concentrations in each sample.
c) Prepare a plot of the peak areas of the pyridoxine hydrochloride peaks versus their
concentrations in each sample.
d) Perform a linear least squares analysis on the data to find the equations of the best
fit lines. Also, include the R2 correlation values.
e) Find the concentrations of caffeine and pyridoxine hydrochloride in the energy
drink in ppm, mg/L, and molarity.
11
f) Compare your findings to the reported values on the can. (WolframAlpha
[http://www.wolframalpha.com] is able to calculate percent daily values on the
can to a useful number)
g) Calculate the number and height of the theoretical plates using your data and the
column information provided on Compass for caffeine.
h) Using the vitamin B6 peak and the closest neighboring peak in Sample 5,
determine the resolution between the peaks. Where tR is the retention time and W
is the width of the peak at the base, where B is the second species eluted. Use the
ruler to determine the width of the peaks if necessary.
Questions:
1) Why are peaks visible when a component elutes, rather than just vertical lines?
How does the use of peak heights instead of peak areas for the determination of
the caffeine concentration affect your results? Does this introduce any errors? If
so, what were they?
2) Calculate the error between your findings and the concentrations listed on the can.
Discuss possible reasons for any differences. Is HPLC a suitable platform for this
type of measurement?
3) What are the implications of using a UV-visible detector to monitor the elution of
analytes? In other words, what are some errors that may be introduced into the
measurement? What are some benefits?
4) Additional Chemical Engineering Question: How might you improve this
experiment to obtain higher resolution between peaks? Be sure to discuss such
factors as column construction and mobile phase composition.