See discussions, stats, and author profiles for this publication at: https://www.researchgate.net/publication/7612501 Evaluation of a clinical aberrometer for lower-order accuracy and repeatability, higher-order repeatability, and instrument myopia Article in Optometry - Journal of the American Optometric Association · September 2005 DOI: 10.1016/j.optm.2005.07.006 · Source: PubMed CITATIONS 39 READS 88 2 authors, including: Corina Van de Pol Marshall B. Ketchum University 37 PUBLICATIONS 319 CITATIONS SEE PROFILE All in-text references underlined in blue are linked to publications on ResearchGate, letting you access and read them immediately. Available from: Corina Van de Pol Retrieved on: 11 August 2016

Upload: others

Post on 01-Aug-2021

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Evaluation of a clinical aberrometer for lower-order accuracy and … · 2019. 11. 18. · The goal ofaberrometry is to measure the wavefront formed by the eye’s op-tics; therefore,aberrometers

Seediscussions,stats,andauthorprofilesforthispublicationat:https://www.researchgate.net/publication/7612501

Evaluationofaclinicalaberrometerforlower-orderaccuracyandrepeatability,higher-orderrepeatability,andinstrumentmyopia

ArticleinOptometry-JournaloftheAmericanOptometricAssociation·September2005

DOI:10.1016/j.optm.2005.07.006·Source:PubMed

CITATIONS

39

READS

88

2authors,including:

CorinaVandePol

MarshallB.KetchumUniversity

37PUBLICATIONS319CITATIONS

SEEPROFILE

Allin-textreferencesunderlinedinbluearelinkedtopublicationsonResearchGate,

lettingyouaccessandreadthemimmediately.

Availablefrom:CorinaVandePol

Retrievedon:11August2016

Page 2: Evaluation of a clinical aberrometer for lower-order accuracy and … · 2019. 11. 18. · The goal ofaberrometry is to measure the wavefront formed by the eye’s op-tics; therefore,aberrometers

EfraC

a

F

B

M

R

C

K

Sro7

CLINICAL RESEARCH

V

valuation of a clinical aberrometeror lower-order accuracy andepeatability, higher-order repeatability,nd instrument myopia

ol Thomas O. Salmon, O.D., Ph.D.,a,b Ltc Corina van de Pol, O.D., Ph.D.b

College of Optometry, Northeastern State University, Tahlequah, Oklahoma; and bUS Army Aeromedical Research Laboratory,ort Rucker, Alabama

ackground: Refractive surgery has stimulated the develop-ment of aberrometers, which are instruments that measurehigher-order aberrations. The purpose of this study was totest one clinical aberrometer, the Complete Ophthalmic Anal-ysis System (COAS), for its accuracy, repeatability, and

BackgroundContext of the studyUntil recently, the clinical measurement and correction of

rwhhtwotowpokmsSrcfsputa

AArkhtrd

instrument myopia for measuring sphere and astigmatismand its repeatability for measuring higher-order aberrations.

ethods: Aberrations of 56 normal eyes (28 subjects) weremeasured with and without cycloplegia using a COAS, aconventional autorefractor and by subjective refraction.We evaluated lower-order accuracy (sphere and astigma-tism) of the COAS and autorefractor by comparing thatdata with that of subjective refraction. We also testedCOAS lower- and higher-order repeatability for 5 mea-surements taken in less than 1 minute. We evaluatedinstrument myopia by comparing cycloplegic and noncy-cloplegic measurements of the same eye. Data wereanalyzed for a 5.0-mm-diameter pupil.

esults: Mean COAS spherical error was between – 0.1 and�0.4 diopters (D), depending on cycloplegia and the kindof sphere power computation selected. Cylinder powererrors were less than 0.1 D. COAS repeatability coeffi-cients were better than 0.25 D, and instrument myopiawas less than 0.4 D. These were comparable with thoseof autorefraction. Higher-order repeatability was sufficientto allow reliable measurement of normal third-order ab-errations and spherical aberration.

onclusions: Accuracy, repeatability, and instrument myopiaof the COAS are similar to those of a conventionalautorefractor. Accuracy and repeatability are also similarto those of subjective refraction. Like an autorefractor, theCOAS provides instantaneous, objective measurements ofsphere and astigmatism, but it also measures higher-orderaberrations. We found that it is capable of reliablymeasuring problematic higher-order aberrations and istherefore a valuable asset for modern clinical eye care.

ey Words: Aberrations, aberrometer, aberrometry, wavefront,refractive errors, accuracy, repeatability, instrument myopia

almon TO and van de Pol C. Evaluation of a clinical aber-ometer for lower-order accuracy and repeatability, higher-rder repeatability, and instrument myopia. Optometry 2005;

a6:461-472.

OLUME 76 / NUMBER 8 / AUGUST 2005

efractive errors was limited to sphere and astigmatism,hereas more complicated refractive errors, known as

igher-order aberrations, were ignored. For most patients,igher-order aberrations had little effect on vision, andhey were too difficult to measure or correct. This changedith the advent of refractive surgery, which was designed

riginally to correct sphere and, more recently, astigma-ism, but in many cases inadvertently created large higher-rder aberrations that caused poor vision.1,2 These un-anted aberrations were troublesome particularly foratients with small treatment zones, higher prescriptions,r large pupils.3–9 The new wavefront-guided laser in situeratomileusis (LASIK) procedures are designed to mini-ize higher-order aberrations. The popularity of refractive

urgery raised a number of concerns for Army aviation.hould the Army accept pilot candidates who have hadefractive surgery? Even if sphere and astigmatism wereorrected, could higher-order aberrations affect flight per-ormance? Should the Army establish new visual or opticaltandards for pilots who have had refractive surgery? Asart of a larger study to answer these questions, we eval-ated the instrument we would use to measure aberra-ions: the Complete Ophthalmic Analysis System (COAS)berrometer.

berrometersberrometers are instruments that measure optical aber-

ations. In the 1970s, Roland Shack developed what is nownown as the Shack-Hartmann (SH) aberrometer,10 whichas been used by astronomers to dramatically improveelescope images.11 Junzhong Liang first applied SH aber-ometry to the human eye in about 1990,12 and within aecade vision scientists around the world had adopted this

s the preferred technique for measuring ocular aberra-

461

OPTOMETRY

Page 3: Evaluation of a clinical aberrometer for lower-order accuracy and … · 2019. 11. 18. · The goal ofaberrometry is to measure the wavefront formed by the eye’s op-tics; therefore,aberrometers

tboueisralc

SArtroBeapt

wwiaeeFtt(mtc

Mpemctt1Occaadtrtmoesnwbc

STetHlrtapcm

F

CLINICAL RESEARCH

4

O

ions.13–26 In 2001, WaveFront Science, Inc, (Al-uquerque, NM) introduced the first commercialphthalmic SH aberrometer, the COAS (see Fig-re 1). Other technologies for measuring theye’s higher-order aberrations, include Tschern-ng and Howland aberroscopy, laser ray tracing,patially resolved refractometry, and slit-scanetinoscopy.27 SH-type aberrometers, however,re the most popular because of their high-reso-ution measurements, known laboratory accura-y,28 and mechanical simplicity.

pecifying ocular aberrationsberrometers measure aberrations of the eye by

ecording the path taken by light rays passinghrough the pupil.29 In SH aberrometry, light iseflected off the retina and measured as it passesut through the eye’s optics (see Figure 2A and). A camera captures the light emitted from theye and records an image of the pupil filled with pattern of dots (see Figure 2C). By analyzing theosition of each dot, the instrument computes

igure 1 The COAS manufactured by Wavefront Sciences, Inc., one ofthe first commercial ophthalmic aberrometers.

he path of each ray and reconstructs the optical e

62

PTOMETRY

avefront formed by the eye. By analyzing theavefront’s shape, all of its refractive errors,

ncluding sphere, astigmatism, and higher-orderberrations can be completely identified. In anye with perfect optics, the wavefront of lightmitted from the eye would be perfectly flat (seeigure 2A), but if the eye contains any aberra-ions, they will distort the wavefront accordingo the type and magnitude of aberrations presentsee Figure 2B). The goal of aberrometry is to

easure the wavefront formed by the eye’s op-ics; therefore, aberrometers are sometimesalled wavefront sensors.

ost eyes produce a wavefront that has a com-lex, irregular shape formed by the combinedffects of multiple aberrations, but by mathe-atical analysis, aberrations which are present

an be determined. This can be done by fittinghe wavefront to an algebraic equation such ashe polynomials developed by Frits Zernike in934.30 Zernike polynomials, as defined by theptical Society of America31,32 and the Ameri-

an National Standards Institute,33,34 are the ac-epted system for recording ocular higher-orderberrations. A Zernike polynomial can containn infinite number of terms, each representing aiscrete aberration, but a limited number oferms is usually sufficient to represent the aber-ations in an eye.13,14,35 The individual aberra-ions are sometimes referred to as Zernikeodes, and they are organized into a hierarchy

f Zernike orders.31 This allows each mode (ab-rration) to be numerically identified by a sub-cript, which tells its order, and a positive oregative superscript, which designates the modeithin that order. For example, the mode la-eled Z3

-1 is the third-order aberration, verticaloma.

tudy objectiveshe principles of Shack-Hartmann (SH) ab-rrometry are well established, and labora-ory devices have proven reliable.15,19,20,28

owever, relatively little work has been pub-ished on the accuracy of the new clinical aber-ometers.26,28,36 Our objectives were to evaluatehe COAS in terms of its (1) accuracy, (2) repeat-bility, and (3) instrument myopia for a normalopulation of Army flight school applicants. Ac-uracy describes how correctly an instrumenteasures what it is supposed to measure. We

valuated accuracy for lower-order aberrations

VOLUME 76 / NUMBER 8 / AUGUST 2005

Page 4: Evaluation of a clinical aberrometer for lower-order accuracy and … · 2019. 11. 18. · The goal ofaberrometry is to measure the wavefront formed by the eye’s op-tics; therefore,aberrometers

oCRt(

rcat

F

CLINICAL RESEARCH

V

nly (sphere and astigmatism), by comparingOAS measurements to subjective refraction.epeatability describes measurement consis-

ency. We assessed repeatability for both lower-

igure 2 Shack-Hartmann aberrometers measure optical quality of the eyfrom the eye. In an eye with perfect optics, a flat wavefront eThe raw data image recorded by a Shack-Hartmann aberromecan reconstruct the wavefront emitted from the eye.

sphere and astigmatism) and higher-order aber- l

OLUME 76 / NUMBER 8 / AUGUST 2005

ations by making multiple measurements andomputing statistics related to the variance. Welso measured instrument myopia, which is theendency of the eye to overaccommodate when

rojecting a point onto the retina and measuring the wavefront of light emitteds (A). Aberrations such as simple myopia (B) bend and distort the wavefront.an array of spots in the pupil (C). By analyzing the position of the spots, we

e by pmergeter is

ooking into tabletop instruments.

463

OPTOMETRY

Page 5: Evaluation of a clinical aberrometer for lower-order accuracy and … · 2019. 11. 18. · The goal ofaberrometry is to measure the wavefront formed by the eye’s op-tics; therefore,aberrometers

MSWcaIhA

IcriewtrDti

PTpsrLdtd

Taewapsms

AItaumatuepsd

Tf

CLINICAL RESEARCH

4

O

ethodsubjectse recruited 28 volunteers from among pilot

andidates who were undergoing a physical ex-mination for entry into Army aviation school.nclusion criteria included the vision and ocularealth requirements for flight school stated inrmy Regulation AR 40-501:

● Spherical refractive error between –0.75 di-opters (D) and �3.00 D

● Astigmatism � 0.75 D● Uncorrected visual acuity of 20/50 or better

in each eye● Best corrected visual acuity of 20/20 or bet-

ter in each eye● No evidence of ocular disease

n addition, subjects must not have worn rigidontact lenses within 6 months, never have hadefractive surgery, and have no medical contra-ndications to the use of cycloplegic or anestheticye drops. One female and 27 male subjects,ith a mean age of 24.7 � 3.3 years, volun-

eered. The mean noncycloplegic refractive er-or was �0.30 � 0.41 D sphere and –0.21 � 0.26

cylinder. Cycloplegia shifted the mean subjec-ive refraction �0.25 D sphere and 0.09 D cyl-nder.

rocedureshe investigators briefed subjects on the study’surpose, duration, procedures and their rights asubjects. Subjects provided informed consent asequired by the US Army Aeromedical Researchaboratory’s Human Use Committee. The rawata for our study were provided by autorefrac-ion, subjective refraction, and aberrometryone in the following order:Noncycloplegic refractions● Auto-refraction (3 measurements with the

Nidek ARK-700A in � 1 minute)● Subjective refraction (1 measurement)● Aberrometry (5 measurements with the

COAS in � 1 minute)Cycloplegic refractions (2 drops of 1%

cyclopentolate)● Aberrometry (5 measurements in � 1

minute)● Subjective refraction (1 measurement)● Auto-refraction (3 measurements in � 1

minute)

64

PTOMETRY

he COAS aberrometer was configured to reportberrations for a 5-mm-diameter pupil up to theighth Zernike order (44 modes). Measurementsere made with dim room illumination, and

fter each measurement, sphere, cylinder, axis,upil diameter, and Zernike coefficients wereaved in a database for subsequent analysis. Weeasured right and left eyes, but analyzed them

eparately.

ccuracydeally, accuracy testing requires that we knowhe true value to be measured. We had no way tobsolutely know each eye’s aberrations, but wesed the subjective refraction as our best esti-ate of the lower-order aberrations, sphere, and

stigmatism. Because we had no way to knowhe higher-order aberrations, we could not eval-ate higher-order accuracy, although we didvaluate higher-order repeatability. Our ap-roach to testing lower-order accuracy was toee how much the COAS sphere and astigmatismiffered from the subjective refraction.

he following steps summarize data processingor the accuracy assessment.26

● Convert each subjective and COAS refrac-tion (sphere, cylinder, axis) to a power vec-tor. This was necessary because astigmaticdata with dissimilar axes cannot be addedtogether directly. Power vectors can beadded directly, which makes it possible tocompute differences, means, and variances.The sphere (S), cylinder (C), and axis (�) maybe converted to power vector [ J45, M, J180]according to Equations 1-3. The J180 termused in this paper is the same as the J0 termin Thibos’ paper on power vectors.37

J45 � (�C ⁄ 2)sin(2�) (1)

M � S � C ⁄ 2 (2)

J180 � (�C ⁄ 2)cos(2�) (3)

● Compute the mean COAS power vector (vector C)for each eye by averaging the J45, M, and J180 for thefive COAS refractions.

● Find the COAS refraction error (vector E) for eacheye by taking the difference between the meanCOAS power vector (vector C) and the subjectiverefraction (vector S), according to Equation 4.

E� �C� �S� (4)

VOLUME 76 / NUMBER 8 / AUGUST 2005

Page 6: Evaluation of a clinical aberrometer for lower-order accuracy and … · 2019. 11. 18. · The goal ofaberrometry is to measure the wavefront formed by the eye’s op-tics; therefore,aberrometers

CsmWrb

BboSamjtsttttaa

R(Be

tldoaeAstd

RAbZmtofaacyceFcevdC

CLINICAL RESEARCH

V

● Compute the overall mean COAS error by averag-ing vector E across 28 subjects.

● To simplify interpretation, we converted the meanCOAS error vector to the equivalent sphere, cylin-der and axis using Equations 5 through 7.

C � �2�J452 � J180

2 (5)

S � M � C ⁄ 2 (6)

� � [tan�1(J45 ⁄ J180)] ⁄ 2 (7)

omputation of the axis required an additionaltep to ensure that it conformed to standardinus-cylinder notation (0 � � � 180 degrees).e corrected the axis value based on the initial

esult for � (Equation 7) and the logical testselow.IF J180� 0, axis � � � 90IF J180 � 0 AND IF J45 � 0, axis � 135IF J180 � 0 AND IF J45 � 0, axis � 45IF J180 � 0 AND IF J45 � 0, axis � � � 180IF J180 � 0 AND IF J45 � 0, axis � �● We also summarized the COAS error in

vector E by combining the 3 numbers J45,M, and J180 into a single magnitude value(m) according to Equation 8. This is similarto the clinical practice of simplifyingspherocylindrical errors with the sphericalequivalent. We did this for each eye, andfound the mean across 28 eyes.

m � �J452 � M2 � J180

2 (8)

y default, the COAS computes sphere powerased on the value of mode Z2

0 (defocus), but itffers an alternate computation referred to as theeidel sphere. It includes mode Z4

0 (sphericalberration) in the calculation and may betteratch the way a human eye responds to a sub-

ective refraction. A complete description of howhe COAS computes both the default and Seidelpheres may be found in another article.26 Weested COAS accuracy for both the default andhe Seidel sphere. For comparison, we alsoested accuracy of the autorefractor followinghe procedures outlined above. Accuracy wasnalyzed separately for right and left eyes withnd without cycloplegia.

epeatability for lower-order aberrationssphere and astigmatism)ecause 5 COAS measurements were made of

ach eye by the same investigator, we were able m

OLUME 76 / NUMBER 8 / AUGUST 2005

o evaluate repeatability, which we did for bothower- (sphere and astigmatism) and higher-or-er aberrations. Repeatability testing for lower-rder aberrations made use of both the defaultnd Seidel spheres and evaluated right and leftyes, with and without cycloplegia separately.utorefractor repeatability was computed in the

ame manner, except that 3 measurements wereaken of each eye. We processed lower-orderata for each eye as follows:

● Convert each of the 5 COAS refractions(sphere, cylinder, axis) to a power vector(Equations 1-3).

● Compute the mean COAS refraction as themean of the 5 original power vectors.

● Subtract the mean from each of the 5 origi-nal power vectors. This gave 5 differencevectors.

● Compute the magnitude (Equation 8) ofeach difference vector and the mean ofthese 5 magnitudes. This gave the meandeviation, in diopters, for each eye.

● Square and sum the mean deviations for 28eyes and divide by 28 to obtain the RMS(root mean squared) deviation.

● Compute a repeatability coefficient, definedas the RMS deviation multiplied by 1.96.This follows the method developed byBland and Altman38 and used in other clin-ical studies39–41 to evaluate repeatability ofdiagnostic instruments.

epeatability for higher-order aberrationsberrometers record higher-order aberrationsy using a numerical coefficient for each of theernike modes. Each Zernike coefficient tells theagnitude (positive or negative) of the aberra-

ion. For normal eyes, the most significant higher-rder aberrations are contained in the third andourth Zernike orders. Aberrations in the fifthnd higher orders are normally much smallernd have little effect on vision.24 For the sake ofompleteness, we evaluated aberrations well be-ond this—through the eighth order, which in-ludes 39 higher-order modes. We measuredach eye’s aberrations with the COAS 5 times.rom the 5 measurements for each mode, weomputed a standard deviation (SD), standardrror (SE � SD⁄�5) and 95% confidence inter-al (CI � SE ● 2.78), then averaged the confi-ence intervals across 28 eyes. The mean 95%Is for these aberrations were interpreted as a

easure of instrument noise and used as our

465

OPTOMETRY

Page 7: Evaluation of a clinical aberrometer for lower-order accuracy and … · 2019. 11. 18. · The goal ofaberrometry is to measure the wavefront formed by the eye’s op-tics; therefore,aberrometers

sorc

IWa

Ietct

Twdmca

RAAiTmtsleSaweDSsisdT

FtrdFamDDtF

CLINICAL RESEARCH

4

O

ummary statistic for higher-order repeatabilityf the COAS. This process was applied sepa-ately to right and left eyes, with and withoutycloplegia.

nstrument myopiae computed instrument myopia for the COAS

nd autorefractor by the following steps26:● Cycloplegia may have induced a slight

change in the true refraction, so for each eyewe computed the change (vector �) as thedifference between the subjective noncy-cloplegic (vector Sm) and subjective cyclo-plegic (vector Sc) power vectors (Equation9).

�� �S�m � S� c (9)

nstrument myopia (vector I in Equation 10) forach eye was defined as the difference betweenhe COAS noncycloplegic (vector Cm) and COASycloplegic (vector Cc) power vectors minus therue change, vector �.

I��C� m � C� c��� (10)

he mean of all COAS instrument myopia valuesas computed for right and left eyes for theefault and Seidel spheres. Mean instrumentyopia power vectors were converted to sphere,

ylinder, and axis and the mean spherical equiv-

Table 1. COAS Error for measuring sphere and cylindcycloplegia, using the default and Seidel sp

Eye/condition Sphere typeSphere e

(D)

O.D./no cyclo Default �0.10 �O.S./no cyclo Default �0.14 �O.D./no cyclo Seidel �0.08 �O.S./no cyclo Seidel �0.08 �O.D./cyclo Default �0.14 �O.S./cyclo Default �0.11 �O.D./cyclo Seidel �0.44 �O.S./cyclo Seidel �0.41 �

Note: A negative sphere error indicates that the instrument overestimatedmean error in terms of power vector magnitude (bold). Analysis pupil diam

lent. g

66

PTOMETRY

esultsccuracyccuracy for measuring sphere and astigmatism

s summarized in Table 1 for the COAS and inable 2 for the autorefractor. The error vectoragnitudes (Table 1, right column, bold) provide

he simplest way to describe COAS accuracy forphere and astigmatism. All rows showed errorsess than 0.5 D, which is similar to the range ofrror seen with the autorefractor (see Table 2).mallest COAS error was found with cycloplegiand the default sphere (about 0.3 D). Otherwise,ithout cycloplegia, power vector error using

ither the default or Seidel sphere was about 0.4. For all conditions except one (cycloplegia,eidel sphere), the mean error for measuringphere (see Table 1, column 3) only was approx-mately �0.1 D. The Seidel sphere shifted thepherical error about �0.25 D for a 5.0-mm-iameter pupil. For all rows, cylinder error (seeable 1, column 4) was less than �0.1 D.

igure 3 shows the magnitude of COAS vec-or errors for each eye across a range ofefractive errors (right, left, default, and Sei-el spheres) when no cycloplegia was used.or both the default sphere (large symbols)nd Seidel sphere (small symbols), approxi-ately 80% of the errors were less than 0.6. In a few cases, magnitudes exceeded 1.0. The Seidel sphere resulted in fewer ex-

reme errors, so that none exceeded 0.9 D.igure 4 shows a similar plot when cyclople-

r right (O.D.) and left (O.S.) eyes, without and with(mean ± 1 SD).

r Cylinder error(D)

Vector magnitude(D)

�0.09 � 0.27 0.43 � 0.25�0.07 � 0.34 0.43 � 0.30�0.09 � 0.27 0.38 � 0.22�0.07 � 0.34 0.43 � 0.24�0.08 � 0.31 0.29 � 0.14�0.09 � 0.32 0.29 � 0.20�0.08 � 0.30 0.45 � 0.24�0.09 � 0.32 0.41 � 0.22

ia; a positive error indicates hyperopic error. The last column shows theas 5.0 mm.

er fohere

rro

0.600.640.550.640.420.460.420.40

myopeter w

ia was used. The distribution of errors was

VOLUME 76 / NUMBER 8 / AUGUST 2005

Page 8: Evaluation of a clinical aberrometer for lower-order accuracy and … · 2019. 11. 18. · The goal ofaberrometry is to measure the wavefront formed by the eye’s op-tics; therefore,aberrometers

cmtdFamcCsfrm

RTccIast3tFepemsmTi(ctd�

svmcp

IItmsifpt

DAotitdatmaCmrCd

AW

CLINICAL RESEARCH

V

oncentrated closer to zero, with approxi-ately 90% of the errors less than 0.6 D. In

his case, fewer outliers were seen when theefault sphere was used. For comparison,igure 5 shows a similar analysis for theutorefractor. The distribution of errors isarginally better than the COAS without

ycloplegia and marginally worse than theOAS with cycloplegia. Cycloplegia did notignificantly improve the error distributionor the autorefractor. About 80% of the er-ors were less than 0.6 D. The autorefractoreasured across a 3.5-mm-diameter pupil.

epeatabilityable 3 shows mean COAS repeatability coeffi-ients for right and left eyes, with and withoutycloplegia, using the default and Seidel sphere.n all cases it was less than 0.25 D. COAS repeat-bility was marginally better with the defaultphere and cycloplegia. For comparison, the au-orefractor’s repeatability coefficients (see Table, right column) were about 0.1 D smaller thanhat of the COAS.igure 6 shows COAS repeatability for high-r-order aberrations. Data points indicate re-eatability in terms of the mean 95% CIs forach eye/condition for each mode. Approxi-ately 90% of the points fall within the

haded region, which generalizes mode-by-ode repeatability based on these results.hat is, COAS repeatability for each mode is

ndicated by the height of the shaded regionsin micrometers), which declines in each suc-essive order. The respective values for thehird, fourth, fifth, sixth, and sevenths or-ers are 0.035, 0.025, 0.02, 0.015 and 0.010

Table 2. Mean autorefractor (Nidek ARK-700A) erroeyes, with and without cycloplegia

Eye/conditionSphere error

(D)

O.D./no cyclo �0.29O.S./no cyclo �0.28O.D./cyclo �0.46O.S./cyclo �0.43

Note: A negative sphere error indicates the instrument overestimated myomean error in terms of power vector magnitude (bold). The autorefractor m

m. Eighth-order values (not shown) were b

OLUME 76 / NUMBER 8 / AUGUST 2005

imilar to those in the seventh order. Thesealues can be interpreted as our estimate ofeasurement noise; that is, variability

aused by the instrument or measurementrocedure.

nstrument myopianstrument myopia had little effect on astigma-ism (� 0.1 D change in each case), so we sum-arized instrument myopia in terms of the

pherical equivalent power (see Table 4). COASnstrument myopia was smaller when the de-ault sphere was used, about –0.25 D. For com-arison, instrument myopia with the autorefrac-or was about –0.2 D.

iscussionberrometers, such as the COAS, provide thenly means for measuring higher-order aberra-ions in a clinical setting and the COAS was thenstrument chosen by our laboratory to studyhe ocular aberrations of Army pilots. Beforeoctors can depend on the data provided byberrometers, they need to know how reliablehese instruments are. Because aberrometerseasure lower-order aberrations as well, they can

lso function as autorefractors. We assessedOAS accuracy, repeatability, and instrumentyopia for lower-order aberrations and only

epeatability for higher-order aberrations. AllOAS analyses were done for a 5.0-mm-iameter pupil.

ccuracye expressed accuracy for measuring the com-

sphere and cylinder for right (O.D.) and left (O.S.)

Cylinder error(D)

Vectormagnitude (D)

�0.15 0.40�0.12 0.41�0.09 0.47�0.05 0.47

positive error indicates hyperopic error. The last column shows thered across a 3.5-mm diameter pupil.

r for

pia; aeasu

ined lower-order aberrations of sphere and

467

OPTOMETRY

Page 9: Evaluation of a clinical aberrometer for lower-order accuracy and … · 2019. 11. 18. · The goal ofaberrometry is to measure the wavefront formed by the eye’s op-tics; therefore,aberrometers

avsTtnsaCwseF

0aepCstrAe

TaftoSftsjWbctdbaip

F

F

F

CLINICAL RESEARCH

4

O

stigmatism by the magnitude of the mean errorector. COAS accuracy was best with the defaultphere and cycloplegia—0.3 D by this statistic.he autorefractor’s accuracy was 0.4 D. To put

his into perspective, a 0.3-D error vector mag-itude is equivalent to a 1/8-D error in both thephere and cylinder combined with a 12-degreexis error. Thus, we found that, on average, theOAS was capable of the same level of accuracye can expect for a good subjective refraction. In

ome cases, however, COAS and autorefractorrror vector magnitudes exceeded 1.0 D (see

igure 3 Magnitude of COAS vector errors across a range of refractiveerrors without cycloplegia. All units are in diopters. Approxi-mately 80% of the errors, using both the default (largesymbols) and Seidel spheres (small symbols) are within0.60 D.

igure 4 Magnitude of COAS vector errors across a range of refractiveerrors with cycloplegia. All units are in D. Approximately 90%of the errors, using both the default sphere (large symbols) andSeidel sphere (small symbols) are within 0.60 D.

igures 3 through 5), which is equivalent to a l

68

PTOMETRY

.5-D error in both the sphere and cylinder with30-degree axis error. Our results with human

yes were only marginally worse than the re-orted accuracy of the COAS with model eyes.heng et al28 reported mean errors of �0.1 Dphere, �0.1D cylinder and �2° axis (equivalento a 0.16-D vector error) across a broad range ofefractive errors (�4.00 to �3.00) on model eyes.ccuracy declined slightly for greater refractiverrors in that study.

here is still debate among vision scientistsbout how to best estimate the subjective sphererom aberrometer data. The COAS default set-ing computes the sphere directly from the sec-nd-order aberration Z2

0 (defocus), whereas theeidel sphere option takes into account theourth-order aberration Z4

0 (spherical aberra-ion).26 Some scientists believe that the Seidelphere should give a better estimate of the sub-ective sphere, especially with large pupils.

hen no cycloplegia was used, we did not findetter accuracy with the Seidel sphere. Withycloplegia it was marginally worse than withhe default sphere. It is possible that our pupiliameters (5.0 mm) were not large enough toenefit from the Seidel computation, because, innother study, we found slightly better accuracyn larger pupils with the Seidel sphere.26 It ap-ears, therefore, that users should normally

igure 5 Magnitude of autorefractor vector errors across a range ofrefractive errors for the right and left eyes without (circles)and with (diamonds) cycloplegia. All units are in diopters.Without cycloplegia, 84% of the errors were less than 0.60 D.With cycloplegia, about 80% of the errors were less than0.60 D.

eave the default sphere setting in place and

VOLUME 76 / NUMBER 8 / AUGUST 2005

Page 10: Evaluation of a clinical aberrometer for lower-order accuracy and … · 2019. 11. 18. · The goal ofaberrometry is to measure the wavefront formed by the eye’s op-tics; therefore,aberrometers

rm

Aweasfbm((

Tl

RRpi(eps

Fof ins

CLINICAL RESEARCH

V

eserve the Seidel option for widely dilated (�6m) pupils.

s mentioned in the Methods section, weere not able to evaluate accuracy for high-

r-order aberrations in this study. Cheng etl28 were able to evaluate COAS accuracy forome higher-order modes using model eyes,or which the higher-order aberrations coulde computed by ray-tracing. They foundean errors of less than 0.01 �m for Z4

0

spherical aberration), less than 0.03 for Z31

Table 3. Repeatability coefficients for the COAS for d

Eye/conditionCOAS default

sphere (D)

O.D./ no cyclo 0.20O.S./ no cyclo 0.18O.D./ cyclo 0.12O.S./ cyclo 0.18

Note: Analysis pupil diameters were 5.0 mm for the COAS and 3.5 mm fo

igure 6 Repeatability of the COAS for measuring higher-order aberratileft (O.S.) eyes, without and with cycloplegia. Each data poin90% of the data points and indicates a generalized estimate

coma) and �0.3 �m for Z42 (5.0-mm pupil). c

OLUME 76 / NUMBER 8 / AUGUST 2005

hese correspond to the respective equiva-ent diopter24 values of 0.1, 0.3, and 0.3 D.

epeatabilityepeatability refers to the variability of re-eated measurements. The COAS repeatabil-ty coefficients for lower-order aberrations�0.25 D) were similar to what we wouldxpect from subjective refraction. For com-arison, the following 5 refractions of theame eye would have a repeatability coeffi-

lt and Seidel spheres and for autorefraction

COAS Seidelsphere (D)

Autorefractor(D)

0.24 0.120.23 0.200.18 0.080.22 0.09

autorefractor.

xpressed as 95% CI for 5 readings taken within 1 minute for right (O.D.) andsents the mean for 28 eyes (27 for O.S. cyclo). The shaded region containstrument noise for each mode. Pupil diameter was 5.0 mm.

efau

r the

ons, et repre

ient of 0.20 D:

469

OPTOMETRY

Page 11: Evaluation of a clinical aberrometer for lower-order accuracy and … · 2019. 11. 18. · The goal ofaberrometry is to measure the wavefront formed by the eye’s op-tics; therefore,aberrometers

Btrs

Wewp60f0TwctmcpcrNp

TotIvtp

scw

CIvmm5Tmpateeadfatpmtppsewo

Ioa

CLINICAL RESEARCH

4

O

● �1.00 – 1.25 � 180● �1.00 – 0.75 � 180● �1.00 – 1.00 � 170● �1.25 – 1.00 � 180● �0.75 – 1.00 � 180

ecause COAS repeatability was marginally bet-er with the default sphere (mean 0.17 D), weecommend using it rather than the Seidelphere for most cases.

e also evaluated COAS repeatability for high-r-order aberrations. Higher-order repeatabilityas generally the same with or without cyclo-legia. As shown by the shaded region in Figure, it was approximately 0.035 �m (equivalent to.04 D) for third-order modes, 0.025 �m (0.03 D)or fourth-order modes, and declined to less than.02 �m (0.02 D) for the fifth order and above.hese values are important to keep in mindhen interpreting aberrometry, because Zernike

oefficients less than the noise level are essen-ially immeasurable. For example, if the instru-ent reports –0.02 �m of mode Z3

-1 (verticaloma) aberrations, it cannot be assumed that theatient has any of this aberration, because itould just be caused by instrument noise. Theseesults apply for a pupil diameter of 5.0 mm.oise would probably increase with larger pu-ils and decrease with smaller pupils.

he variability that we measured for higher-rder aberrations was only slightly worse thanhat reported in another study by Cheng et al.36

n that study, they estimated that most of theariability of COAS was attributable to fluctua-ions in accommodation, the tear film, or eye

Table 4. Instrument myopia, expressed as the changeequivalent power, in diopters (mean ± 1 SD

Instrument EyeSpheremethod

COAS O.D. DefaultCOAS O.S. DefaultCOAS O.D. SeidelCOAS O.S. SeidelAutorefractor O.D. NAAutorefractor O.S. NA

Note: A negative value indicates that the instrument overestimated myop

osition rather than to the instrument itself. The t

70

PTOMETRY

COAS may be subject to axial,transverse, or angular position-ing errors, but they showedthat, within the range of mis-alignments expected for normalclinical use, these caused no sig-nificant error.28,36

Instrument myopiaThe COAS showed marginallymore instrument myopia thanthe autorefractor (mean, 0.28versus 0.19 D). We previouslyfound almost no instrumentmyopia with the COAS, but allof those subjects were myo-pic.26 The slightly greater in-

trument myopia in this study may have beenaused by the inclusion of young hyperopes,ho tend to habitually overaccommodate.

onclusionsn normal eyes, higher-order aberrations areery small and have little effect on vision. Theagnitudes of higher-order aberrations in nor-al eyes (averaged across several studies), for

.0- and 6.0-mm pupils, are shown in Figure 7.42

he figure also shows the noise range we deter-ined for a 5.0-mm pupil, for which the most

rominent aberrations—all third-order modesnd fourth-order mode Z4

0 (spherical aberra-ion)—are measurable by the COAS because theyxceed measurement noise (shaded zone). How-ver, for a 5.0-mm pupil, the other fourth-ordernd fifth-order aberrations are so small that theyo not exceed the noise limits and would, there-ore, be difficult to measure. To detect the subtleberrations in an eye with good optics, we,herefore, recommend measuring with as large aupil as possible. Clinicians, however, are pri-arily interested in measuring abnormal aberra-

ions, which would be larger than the valueslotted in Figure 7. Based on this analysis, allroblematic aberrations should be easily mea-urable with the COAS. This is particularly rel-vant to refractive surgery patients or othershose subnormal vision may be caused by poor

ptics.

n addition to its capacity to measure higher-rder aberrations, the COAS can serve as anutorefractor by measuring sphere and astigma-

spherical

strumentyopia

0.24 � 0.420.24 � 0.350.36 � 0.430.29 � 0.410.19 � 0.330.19 � 0.37

in)

Inm

������

ia.

ism. Like a conventional autorefractor, it occa-

VOLUME 76 / NUMBER 8 / AUGUST 2005

Page 12: Evaluation of a clinical aberrometer for lower-order accuracy and … · 2019. 11. 18. · The goal ofaberrometry is to measure the wavefront formed by the eye’s op-tics; therefore,aberrometers

semtrvaamutaal

DTraUoif

AWOa

R

1

1

1

1

F over a

CLINICAL RESEARCH

V

ionally has larger-than-average measurementrrors, so when accuracy is critical, we recom-end comparing COAS refractions to that ob-

ained by a careful subjective refraction. Weecommend using the default sphere, except forery large pupils. Cycloplegia slightly improvesccuracy and repeatability for measuring spherend astigmatism, but it would not be helpful ineasuring the higher-order aberrations of a nat-

ral eye, because cycloplegia itself can changehose aberrations.43 We conclude that the COASberrometer provides clinicians with a fast, reli-ble, easy-to-use method to objectively measureower- and higher-order aberrations of eyes.

isclaimerhe opinions, interpretations, conclusions, andecommendations in this article are those of theuthors and are not necessarily endorsed by theS Army and/or Department of Defense. Nonef the authors of this article have any financialnterest in WaveFront Sciences, Inc., the manu-acturer of the COAS.

cknowledgementse are grateful for the valuable assistance of Major Linda Knapp Glisson,

.D., US Army, for helping acquire data; Jon B. Sawyer for data acquisitionnd analysis; and Leonita M. Newman for help in preparing the manuscript.

eferences1. Maguire L. Keratorefractive surgery, success and the

igure 7 Magnitudes of higher-order aberrations expected for normal eyefrom 5 studies that surveyed the aberrations of normal eyes

public health. Am J Ophthalmol 1994;117:394-8.

OLUME 76 / NUMBER 8 / AUGUST 2005

2. Howland H. The history and methods of ophthalmicwavefront sensing. J Refract Surg 2000;16:S552-3.

3. Boxer Wachler BS. Effect of pupil size on visual func-tion under monocular and binocular conditions inLASIK and non-LASIK patients. J Cataract Refract Surg2003;29:275-8.

4. Boxer Wachler BS, Huynh VN, El-Shiaty AF, et al. Evaluationof corneal functional optical zone after laser in situ keratom-ileusis. J Cataract Refract Surg 2002;28:948-53.

5. Casson EJ, Jackson WB, Mintsioulis G, et al. Visualperformance under dilated conditions following Exci-mer photorefractive keratectomy. In: Vision Science andits Applications. Santa Fe, NM: 1996;222-5.

6. Haw WW, Manche EE. Effect of preoperative pupilmeasurements on glare, halos, and visual function afterphotoastigmatic refractive keratectomy. J Cataract Re-fract Surg 2001;27:907-16.

7. Lee YC, Hu FR, Wang IJ. Quality of vision after laser insitu keratomileusis. Influence of dioptic correction andpupil size on visual function. J Cataract Refract Surg2003;29:769-77.

8. Martínez CE, Applegate RA, Klyce SD, et al. Effect ofpupillary dilation on corneal optical aberrations after pho-torefractive keratectomy. Arch Ophthalmol 1998;116:1053-62.

9. Boxer Wachler BS, Durrie DS, Assil KK, Krueger RR.Role of clearance and treatment zones in contrast sen-sitivity: significance in refractive surgery. J CataractRefract Surg 1999;25:16-23.

0. Platt B, Shack R. History and principles of Shack-Hart-mann wavefront sensing. J Ref Surg 2001;17:S573-7.

1. Fugate RQ, Wild WJ. Untwinkling the stars—part I. Sky& telescope 1994;87:24-31 (May).

2. Liang J, Grimm B, Goelz S, et al. Objective measure-ment of wave aberrations of the human eye with the useof a Hartmann-Shack wave-front sensor. J Opt Soc Am AOpt Image Sci Vis1994;11:1949-57.

3. Liang J, Williams DR. Aberrations and retinal imagequality of the normal human eye. J Opt Soc Am A Opt

.0-mm and 6.0-mm-diameter pupils. Data were obtained by computing meansrange of refractive errors.

s for 5

Image Sci Vis 1997;14:2873-83.

471

OPTOMETRY

Page 13: Evaluation of a clinical aberrometer for lower-order accuracy and … · 2019. 11. 18. · The goal ofaberrometry is to measure the wavefront formed by the eye’s op-tics; therefore,aberrometers

1

1

1

1

1

1

2

2

2

2

2

2

2

2

2

2

3

3

3

3

3

3

3

3

3

3

4

4

4

4

CLINICAL RESEARCH

4

O

4. Liang J, Williams DR, Miller DT. Supernormal vision andhigh-resolution retinal imaging through adaptive optics. JOpt Soc Am A Opt Image Sci Vis 1997;14:2884-92.

5. Salmon TO, Thibos L, Bradley A. Comparison of theeye’s wave-front aberration measured psychophysicallyand with the Shack-Hartmann wave-front sensor. J OptSoc Am A Opt Image Sci Vis 1998;15:2457-65.

6. Thibos L, Hong X. Clinical applications of the Shack-Hartmann aberrometer. Optom Vis Sci 1999;76:817-25.

7. Hamam H. A quick method for analyzing Hartman-Shack patterns: application to refractive surgery. J RefSurg 2000;16:S636-42.

8. Miller D. Retinal imaging and vision at the frontiers ofadaptive optics. Physics Today 2000;53:31-6.

9. Moreno-Barriuso E, Navarro R. Laser ray tracing versusHartmann-Shack sensor for measuring optical aberra-tions in the human eye. J Opt Soc Am Opt Image Sci Vis2000;17:974-84.

0. Prieto P, Vargas-Martin F, Goelz S, et al. Analysis of theperformance of the Hartmann-Shack sensor in the humaneye. J Opt Soc Am A Opt Image Sci Vis 2000;17:1388-98.

1. Hofer H, Artal P, Singer B, et al. Dynamics of the eye’swave aberration. J Opt Soc Am A Opt Image Sci Vis2001;18:497-506.

2. Porter J, Guirao A, Cox I, et al. Monochromatic aberra-tions of the human eye in a large population. J Opt SocAm A Opt Image Sci Vis 2001;18:1793-1803.

3. Marcos S, Diaz-Santana L, Llorente L, et al. Ocularaberrations with ray tracing and Shack-Hartmannwave-front sensors: Does polarization play a role? J OptSoc Am A Opt Image Sci Vis 2002;19:1063-72.

4. Thibos L, Hong X, Bradley A, et al. Statistical variationof aberration structure and image quality in a normalpopulation of healthy eyes. J Opt Soc Am A Opt Image SciVis 2002;19:2329-48.

5. Yoon GY, Williams DR. Visual performance after correct-ing the monochromatic and chromatic aberrations of theeye. J Opt Soc Am A Opt Image Sci Vis 2002;19:266-75.

6. Salmon TO, West RW, Gasser W, et al. Measurement ofrefractive errors in young myopes using the COAS Shack-Hartmann aberrometer. Optom Vis Sci 2003;80:6-14.

7. Krueger R, Applegate RA, MacRae S. Wavefront Cus-tomized visual correction—the quest for super vision II,2nd ed. Thorofare, NJ: Slack, Inc, 2004.

8. Cheng X, Himebaugh N, Kollbaum P, et al. Validationof a clinical Shack-Hartmann aberrometer. Optom VisSci 2003;80:587-95.

9. Salmon TO, West RW, Optical wavefront sensing of thehuman eye. In: Pandalai S, ed. Recent research develop-ments in optics. Trivandrum, Kerala, India: ResearchSignpost, 2002:183-214.

0. deCarvalho LAV, de Castro JC. Preliminary results ofan instrument for measuring the optical aberrations of

the human eye. Braz J Phys 2003;33:148-57.

72

PTOMETRY

1. Atchison D, Scott D, Cox M. Mathematical treatment ofocular aberrations: a user’s guide. In: Lakshminaray-anan V, ed. Vision science and its applications. Santa Fe,NM:Optical Soc Am, 2000:110-30.

2. Thibos L, Applegate R, Schwiegerling J, et al. Standardsfor reporting the optical aberrations of the eye. In:Lakshminarayanan V, ed. Vision science and its applica-tions. February 11-14, Santa Fe, NM [conference book-let]: Optical Soc Am, 2000:232-44.

3. Atchison D. Recent advances in representation ofmonochromatic aberrations of human eyes. Clin ExpOptom 2004;87:138-48.

4. American National Standard for Ophthalmics. Methodsfor reporting optical aberrations of eyes. ANSI Z80.28-2004; 2004.

5. Salmon TO. Corneal contribution to the wavefront aberra-tion of the eye [PhD thesis]. Bloomington, IN: IndianaUniversity, 1999.

6. Cheng X, Himebaugh N, Kollbaum P, et al. Test-retestreliability of clinical Shack-Hartmann measurements.Invest Ophthalmol Vis Sci 2004;45:351-60.

7. Thibos LN, Wheeler W, Horner D. Power vectors: anapplication of Fourier analysis to the description andstatistical analysis of refractive error. Optom Vis Sci1997;74:367-75.

8. Bland J, Altman D. Statistical methods for assessingagreement between two methods of clinical measure-ment. Lancet 1986;1:307-10.

9. Zadnik K, Mutti DO, Adams AJ. The repeatability ofmeasurement of the ocular components. Invest Ophthal-mol Vis Sci 1992;33:2325-33.

0. Rosenfield M, Chiu NN. Repeatability of subjective andobjective refraction. Optom Vis Sci 1995;72:577-9.

1. Walline JJ, Kinney KA, Zadnik K, et al. Repeatabilityand validity of astigmatism measurements. J RefractSurg 1999;15:23-31.

2. Salmon TO, van de Pol C. Zernike coefficient norms—comparison of studies. In: American Academy of Optom-etry Annual Meeting. Tampa, FL, 2004.

3. Carkeet A, Velaedan S, Tan YK, et al. Higher orderocular aberrations after cycloplegic and non-cycloplegicpupil dilation. J Refract Surg 2003;19:316-22.

Corresponding author:Thomas O. Salmon, O.D. PhD.

College of Optometry, Northeastern StateUniversity

1001 N. Grand AvenueTahlequah, OK 74464-7017

[email protected]

VOLUME 76 / NUMBER 8 / AUGUST 2005