EPH AROMATIC HYDROCARBONS
Trilogy 255 nm EPH Calibration
Sitelab’s PAH calibration kit is used on the UVF-3100’s EDRO
optics for applications testing EPH Aromatics or “Total PAHs.”
This test matches well to laboratory GC/FID methods for
Extractable Petroleum Hydrocarbons. This test is mostly used
by Sitelab for TPH fingerprinting applications. Trilogy’s higher
readings are due to subtle differences in the emission optics.
PPM
RFU
Curve r2 = 0.997
UVF-3100
RFU
0
35
67
309
544
746
Trilogy
RFU
15
82
140
535
996
1,342
Calibration
Standard
Solvent Blank
0.05 ppm
0.1 ppm
0.5 ppm
1.0 ppm
1.5 ppmUVF-3100 (ppm)
Trilogy (ppm)
PAH
CAL-060
Instead of DRO and GRO, some states
require laboratories to perform EPH and
VPH analysis. These methods separate the
aliphatics and aromatics into fractions. The
aromatics have lower regulatory cleanup
limits. Fluorescence is not sensititive to
aliphatic hydrocarbons.
C9-C18
Aliphatics
C19-C36
Aliphatics
C11-C22
Aromatics
UVF-3100 EPH
Aromatics
7,700
11,000
14,000
3,700
9,300
20,000
4,800
14,000
21,000
5,250
14,400
23,600
0
500
1000
1500
0 0.5 1 1.5 2
TRILOGY ALMOST 2X HIGHER,
BUT RESULTS ARE PROPORTIONATE
MassDEP EPH Results by GC/FID
Concentrations in ppm (mg/KG)FACTORY SITE
GOOD,
ACCURATE
254 LAMP 255 LED
RESPONSE AT 100 PPM EPH EPH
Heat Transfer Fluid 35 55
Heavier Crude Oil 39 64
Lighter Crude Oil 8 14
Old Home Heating Oil 20 40
Motor Oil 9.5 20
Diesel Fuel 3.5 7
Vacuum Pump Oil 2.7 4
Jet Fuel 1.5 3
Gasoline, 93 Octane 0.56 0.88
Gasoline, 87 Octane 0.23 0.34
97
SOIL SAMPLE IN PPM
No. 6 Fuel Oil, Factory Site 60 97
SOILS FROM
NAPL PLUME
Samples shown here not available for Trilogy study.
UVF-TRILOGY
EDRO Module
© SITELAB CORPORATION 2018 TRILOGY EVALUATION STUDY PAGE 6 OF 8