Presentation Contents
Feature Overview
Creating an Encoded C-scan from a Sectorial Scan
Creating an Encoded C-scan from a Linear Scan
Creating a Timed C-scan from a Linear Scan
Reviewing Acquired C-scan Data
-
8/19/2019 Encoded C Scan
3/101
Presentation Contents Demonstration Procedures
Attac!ing t!e "ini-#!eel Encoder
Encoder Cali$ration
Ti%s for #eld &ns%ection
Demonstration Procedure' (ero Degree Flaw &ns%ection
-
8/19/2019 Encoded C Scan
4/101
-
8/19/2019 Encoded C Scan
5/101
O%tion Overview
T!e encoded C-Scan software o%tion is an added cost o%tion to t!estandard EPOC) *+++i instrument,
&t is t!e same %art num$er and %rice as t!e %reviousl released manualLinear scan and C-scan software, P./ EP*+++&-ECSCA/ 012*3+*3*4
#!en t!is software o%tion is %urc!ased it can $e remotel activated ont!e EPOC) *+++i via a software code %rovided,
T!is o%tion adds' Encoded and timed C-Scan Linear scan ca%a$ilit 0L-scan or E-scan for electronic scan4 &ncreases t!e %!ased arra configuration from t!e standard *5'*5 to *5'53,
-
8/19/2019 Encoded C Scan
6/101
O%tion Overview
T!e Encoded C-scan o%tion allows ou to create %!ased arra C-scans from $ot! linear and sector scan data, T!is is an added feature from t!e %revious o%tion w!ic! onl allowed C-
scan creation from linear scans,
C-scan creation from S-scans ma6es t!is software o%tion es%eciall!el%ful in weld ins%ection as ou can view gated data over a swee% ofangles at multi%le %ro$e %ositions,
-
8/19/2019 Encoded C Scan
7/101
Required Equi%ment
Encoded C-Scan Software O%tion"ini-#!eel Encoder Ot!er encoders can also $e used
Encoder Ada%tor Ca$le7L*5-A*+P or 7L53-A*8 Pro$e 0or equivalent4 An %!ased arra %ro$e t!at attac!es to an encoder and !as *5 elements
for S-Scans and u% to 53 elements for Linear Scans can $e used
Encoder Com%ati$le #edges S%ecial %!ased arra wedges are required for use wit! encoded scans wit!
t!e A*+9 A*+P and A*8 st le %ro$es, T!ese wedges !ave t!readed insertsfor connecting to encoders,
-
8/19/2019 Encoded C Scan
8/101
Creating an Encoded C-scanfrom a Sectorial Scan (S-scan)
-
8/19/2019 Encoded C Scan
9/101
Creating a C-scan from an S-scan
To create a C-scan from a sector scan9 $egin wit! a fullcali$rated S-scan
Completed
ero and !aincali"rations
-
8/19/2019 Encoded C Scan
10/101
De%t! :ates
:ates can $e dis%la ed in True De%t! mode on t!e S-scan allowingou to collect C-scan measurements from a constant de%t! area in t!e
%art,
To set u% True De%t! gates'*, Select !ate Setup # Setup8, Set !ates $%pe to &ept'
:ate t %e is a glo$al settingt!at is a%%lied to all activegates
:ates will onl dis%la as!ori;ontal lines on t!e S-scanonl 9 A-scan gates will remainin sound%at! mode
-
8/19/2019 Encoded C Scan
11/101
-
8/19/2019 Encoded C Scan
12/101
Attac!ing t!e Encoder
For encoded C-scans9 a %ositional encoder is used to trac6distance traveled along t!e scan a>is, Encoder-com%ati$le wedges and an EPOC) *+++i ada%tor ca$le are
required for %ro%er setu%,
For detailed information on setting u% t!e mini-w!eel encoder9%lease use t!e following lin6' Attac!ing t!e "ini-#!eel Encoder
-
8/19/2019 Encoded C Scan
13/101
Cali$rating t!e Encoder
Once t!e encoder !as $een attac!ed to t!e %ro$e9 it must$e cali$rated to ensure %ro%er distance values arecollected during t!e scan,
For instructions on !ow to cali$rate t!e encoder9 %leaseuse t!e following lin6' Encoder Cali$ration
-
8/19/2019 Encoded C Scan
14/101
C-scan Setu%
/ow ou are read to set u% our C-scan using t!efollowing ste%s'*, Turn O t!e o%tion in t!e C Scan # C Scan menu,
Once t!e C-scan is turned on9 additional C-scan %arameters willdis%la in t!e menu,
-
8/19/2019 Encoded C Scan
15/101
-
8/19/2019 Encoded C Scan
16/101
C-scan Setu% 0cont,4
@, Set t!e desired Start and End Co-ordinates for t!e scan2, Select w!et!er to Save 3scan
images, &f :ate 8 is $eing used9 select t!e
3mplitude &ecimation factor For encoded C-scans9 gate *am%litude com%ression isautomaticall set to ma>imum andde%t! com%ression for gates * and8 is set to minimum,
*+, T!e current Scan File Si4e $ased
on t!e selected %arameters isdis%la ed, T!is %arameter cannote>ceed t!e +a5 File Si4e ofa%%ro>imatel @+"$,
**, Press t!e 6ESC3PE7 6e to return tot!e live screen,
-
8/19/2019 Encoded C Scan
17/101
Creating a C-scan
To acquire data for our C-scan*, Set t!e C Scan # +ode to Start8, "ove our encoder and %ro$e along our test %iece at t!e a%%ro%riate
scan rate so t!at ou do not receive an LOS measurement
-
8/19/2019 Encoded C Scan
18/101
Additional /otes
Encoded C-scans can also $e created from a ;ero degree sector scan
T!ese t %es of encoded C-scans can $e ver !el%ful in weldins%ection a%%lications, For additional information9 refer to ourTi%s for #eld &ns%ection ,
-
8/19/2019 Encoded C Scan
19/101
eturn to Presentation Contents
eviewing 3c8uired C-Scan &ata
-
8/19/2019 Encoded C Scan
20/101
Creating an Encoded C-scanfrom a 9inear Scan (9-scan)
-
8/19/2019 Encoded C Scan
21/101
Creating a C-scan from a Linear Scan
To create a C-scan from a linear scan9 $egin wit! a fullcali$rated linear scan
Completedero and !ain
cali"rations
-
8/19/2019 Encoded C Scan
22/101
:ate Ad=ustment
For ;ero degree C-scans9Sound%at! gates can $e usedBou s!ould set our gate%osition so t!at ou areca%turing t!e desired distance
wit!in t!e %art for our C-scan, Ad=ust t!e !1Start and
!1 idt' %arameters to ca%turet!e desired distance wit!in t!etest %iece
A second gate can also $e!el%ful for $ac6wall monitoring
Ad=ust t!e !:Start and!: idt' %arameters tosurround t!e $ac6wall ec!o
-
8/19/2019 Encoded C Scan
23/101
Attac!ing t!e Encoder
For encoded C-scans9 a %ositional encoder is used to trac6distance traveled along t!e scan a>is, Encoder-com%ati$le wedges and an EPOC) *+++i ada%tor ca$le are
required for %ro%er setu%,
For detailed information on setting u% t!e mini-w!eel encoder9%lease use t!e following lin6' Attac!ing t!e "ini-#!eel Encoder
-
8/19/2019 Encoded C Scan
24/101
Cali$rating t!e Encoder
Once t!e encoder !as $een attac!ed to t!e %ro$e9 it must$e cali$rated to ensure %ro%er distance values arecollected during t!e scan,
For instructions on !ow to cali$rate t!e encoder9 %leaseuse t!e following lin6' Encoder Cali$ration
-
8/19/2019 Encoded C Scan
25/101
C-scan Setu%
/ow ou are read to set u% our C-scan using t!efollowing ste%s'*, Turn O t!e o%tion in t!e C Scan # C Scan menu,
Once t!e C-scan is turned on9 additional C-scan %arameters willdis%la in t!e menu,
-
8/19/2019 Encoded C Scan
26/101
C-scan Setu% 0cont,4
8, Select C Scan # Setup toenter t!e setu% screenimumscanning s%eed,
2 Encoder esolution willautomaticall %o%ulate fromt!e encoder cali$ration
-
8/19/2019 Encoded C Scan
27/101
C-scan Setu% 0cont,4
@, Set t!e desired Start and End Co-ordinates for t!e scan2, Select w!et!er to Save 3scan
images, &f :ate 8 is $eing used9 select t!e
3mplitude &ecimation factor
For encoded C-scans9 gate *am%litude com%ression isautomaticall set to ma>imum andde%t! com%ression for gates * and8 is set to minimum,
*+, T!e current Scan File Si4e $ased
on t!e selected %arameters isdis%la ed, T!is %arameter cannote>ceed t!e +a5 File Si4e ofa%%ro>imatel @+"$,
**, Press t!e 6ESC3PE7 6e to return tot!e live screen,
-
8/19/2019 Encoded C Scan
28/101
Creating a C-scan
To acquire data for our C-scan*, Set t!e C Scan # +ode to Start8, "ove our encoder and %ro$e along our test %iece at t!e
a%%ro%riate scan rate so t!at ou do not receive an LOSmeasurement
-
8/19/2019 Encoded C Scan
29/101
eturn to Presentation Contents
eviewing 3c8uired C-Scan &ata
&emonstration Procedure* ero &egree Flaw .nspection
-
8/19/2019 Encoded C Scan
30/101
Creating a $imed C-scanfrom a 9inear Scan (9-scan)
-
8/19/2019 Encoded C Scan
31/101
Creating a C-scan from a Linear Scan
To create a C-scan from a linear scan9 $egin wit! a fullcali$rated linear scan
Completedero and !ain
cali"rations
-
8/19/2019 Encoded C Scan
32/101
:ate Ad=ustment
For ;ero degree C-scans9Sound%at! gates can $e usedBou s!ould set our gate%osition so t!at ou areca%turing t!e desired distance
wit!in t!e %art for our C-scan, Ad=ust t!e !1Start and
!1 idt' %arameters to ca%turet!e desired distance wit!in t!etest %iece
A second gate can also $e!el%ful for $ac6wall monitoring
Ad=ust t!e !:Start and!: idt' %arameters tosurround t!e $ac6wall ec!o
-
8/19/2019 Encoded C Scan
33/101
C-scan Setu%
/ow ou are read to set u% our C-scan using t!efollowing ste%s'*, Turn O t!e o%tion in t!e C Scan # C Scan menu,
Once t!e C-scan is turned on9 additional C-scan %arameters willdis%la in t!e menu,
-
8/19/2019 Encoded C Scan
34/101
C-scan Setu% 0cont,4
8, Select C Scan # Setup to entert!e setu% screenceed t!e +a5 File Si4e ofa%%ro>imatel @+"$,
@, Press t!e 6ESC3PE7 6e to returnto t!e live screen,
-
8/19/2019 Encoded C Scan
35/101
Creating a C-scan
To acquire data for our C-scan*, Set t!e C Scan # +ode to Start8, Once started9 t!e scan will automaticall move at a fi>ed rate along t!e
timed a>is
-
8/19/2019 Encoded C Scan
36/101
eturn to Presentation Contents
eviewing 3c8uired C-Scan &ata
&emonstration Procedure* ero &egree Flaw .nspection
-
8/19/2019 Encoded C Scan
37/101
eviewing 3c8uired C-scan &ata
-
8/19/2019 Encoded C Scan
38/101
Com%leting a C-scan
Once ou !ave acquired all t!e data from a %articularsection or !ave finis!ed our entire scan9 ou can c!ooseto Stop or Free4e t!e C-scan, Sto%%ing a scan will com%lete t!e scan, /ew C-scans will $e
started from t!e start coordinate or ;ero %oint de%ending on t!escan mode,
Free;ing a scan will $ring ou to Review mode, #!en t!e scan isunfro;en ou can continue t!e scan from t!e %oint w!ere t!e scanwas fro;en,
vs
C Scan # +ode6F EE E7
-
8/19/2019 Encoded C Scan
39/101
Start.Sto% Function
T!e Start;Stop function is used to end our C-scan For convenience9 t!e 6CHEC
-
8/19/2019 Encoded C Scan
40/101
Free;e Function
#!en ou !ave finis!ed ourscan section9 or to %auseand review our acquireddata9 %ress t!e 6F EE E7 6e to enter Review "ode ,
An F icon will dis%la to t!eleft of t!e scan to indicatet!at ou are reviewing afro;en scan,
T!e Start.Sto% +odecontrols will $e disa$ledw!en review mode is active,
-
8/19/2019 Encoded C Scan
41/101
Free;e Function 0cont,4
Entering Review "ode allows use of t!e followingfunctions.features9 w!ic! are e>%lained furt!er in t!issection'*, &mage cursors 0D,T,9 Angle9 &nde>48, iewing source S-scan or E-scan images
-
8/19/2019 Encoded C Scan
42/101
-
8/19/2019 Encoded C Scan
43/101
&mage Cursors Distance Traveled 0D,T,4 Cursor
For encoded C-scans9 t!e D,T,cursor will select a s%ecificencoder %osition along t!e scana>is in inc!es or mm,For timed C-scans9 t!is cursor will
corres%ond to a s%ecific screen%i>el along t!e scan a>is,
To ad=ust t!e D,T, cursor select P3 Cursors # & $
Encoded C-scanfrom S-scan
Encoded C-scanfrom linear scan
$imed C-scanfrom linear scan
-
8/19/2019 Encoded C Scan
44/101
&mage Cursors Angle.&nde> Cursor
For C-scans created fromS-scans9 t!e Angle cursoris used to select a s%ecificfocal law wit!in t!e selected
angle range of t!e sectorscan,For C-scans created fromlinear scans9 t!e &nde>cursor selects a %articulara%erture grou% of elements
along t!e lengt! of t!e%ro$e,
To ad=ust t!e Angle.&nde> cursor select %resst!e 63 !9E7 6e
Encoded C-scan
from linear scan
Encoded C-scanfrom S-scan
-
8/19/2019 Encoded C Scan
45/101
Screen iews
&n Review mode9 ou can view t!e S-scan or linear scan from w!ic! a C-scan%oint was generated, /OTE' Bou will not $e a$le to ad=ust gate %osition or range on t!e stored C-scan or
linear scans, Angle.&nde> can $e ad=usted,
After selecting t!e s%ecific C-scan %oint using t!e image cursors9 sim%l ad=ustt!e C Scan # Screen view to dis%la t!e sector or linear scan for t!e selected
C-scan %oint 0D,T,9 &nde>.Angle4,
-
8/19/2019 Encoded C Scan
46/101
Si;ing Cursors
P!ased arra si;ing cursors are also availa$le for $asic flaw si;ing ont!e C-scan image, ?ased on our screen view9 t!e and B cursors will $e corres%ond to focal
law 0angle4 or %osition along t!e %ro$e 0inde>4 and distance traveled 0scana>is4,
To turn t!e si;ing cursors O 9 select P3 Cursors # Cursors , T!e and B cursors can t!en $e ad=usted to ta6e measurements in $ot!
directions along t!e scan, &f ou are selected in 3uto mode9 t!e relevant cursor measurements will $e
dis%la ed in t!e measurement $o>es a$ove t!e scan,
-
8/19/2019 Encoded C Scan
47/101
Si;ing Cursors 0cont,4
Encoded C-scans created from S-scans %rovide information over aswee% of angles and s!ow flaw %osition along t!e encoded scan a>is,
Encoded C-scans created from linear scans %rovide an electronic scanof inde>es t!at %ro%erl scale defects along t!e lengt! of t!e %ro$e and
s!ow flaw %osition along t!e encoded scan a>is,
Timed C-scans do not %rovide %osition information along t!e scan a>isas t!is timed a>is is not to scale,
-
8/19/2019 Encoded C Scan
48/101
Si;ing Cursors 0cont,4
G cursors si;e in scan direction
GB cursors si;e along %ro$e lengt! using inde>es
Encoded C-scan from linear scanwit' si4ing cursors
$imed C-scan from linear scan wit'si4ing cursors
G cursors not used since scan direction is timedand not to scale
GB cursors si;e along %ro$e lengt! using inde>es
-
8/19/2019 Encoded C Scan
49/101
Si;ing Cursors S!ading&n order to more clearl view t!e area $eing si;ed9 ou can c!oose tos!ade t!e area surrounded $ t!e and B si;ing cursors,To s!ade t!e si;ing area9 select &ispla% Setup # Contrast , As t!e contrast is increased9 t!e si;ing area will $ecome dar6er s!aded in $lue /OTE' T!is contrast feature also controls t!e contrast of t!e weld overla if t!is
feature is $eing used,
-
8/19/2019 Encoded C Scan
50/101
:ate Source DataDe%t! and am%litude data for gates * and 8 are collected w!en a C-scan is
created, After acquiring a C-scan9 t!e user can ad=ust $ot! C Scan # Source %arametersto view ot!er relevant C-scan data,
!ate 1
!ate :
3mplitude
&ept'
-
8/19/2019 Encoded C Scan
51/101
Com%ressed iewSince encoded C-scans allow for data collection for lengt!s longer t!an t!escreen widt!9 ou can com%ress a C-scan to view t!e entire scan on screen allat once,To turn com%ressed view O 9 select C Scan # Compressed
-
8/19/2019 Encoded C Scan
52/101
Color Palettes Additional color %alettes are availa$le for %!ased arra images to %rovidedifferent color scales for use wit! varied a%%lications,To c!ange t!e color %alette Select P3 &ispla% # Scan Palette 0&ept' Palette for de%t! C-Scans4 Or select &ispla% Setup # Color Setup ,
T!e user can also manuall modif t!e color scale of an of t!e PA color
%alettes 0 &ispla% Setup # Color Setup 4 O$E* Color Palette ad=ustment will not affect an of t!e raw data dis%la ed on t!e
A-Scan and will onl c!ange t!e color scale on t!e PA scan image,
S-scan wit' 3mplitude Palette S-scan wit' Corrosion Palette
-
8/19/2019 Encoded C Scan
53/101
Color Palettes Color Ad=ustment' Am%litudeEac! of t!e color %alettes !as a %reset color scale t!at is a%%lied over t!e fullrange of signal am%litudes from +-**+H FS),
T!e user can ad=ust t!e ma>imum and minimum am%litude color values usingt!e %arameters $elow under Am%litude Palette Control settings'
Palette Start* sets t!e u%%er colort!res!old for signals a$ove t!e setam%litude
Palette End* sets t!e lower colort!res!old for signals $elow t!e setam%litude
T!e new ad=usted color scale will$e dis%la ed to t!e left of t!esettings in t!e Color Setup menu,
-
8/19/2019 Encoded C Scan
54/101
PA Color Palette Ad=ustment- Am%litude
GPalette Start ad=usted from 0>to :0>
G 3ll ec'oes "elow :0>FSH aredispla%ed in w'ite on t'e S-
Scan
&efault settings for 3mplitudeScan Palette
3d=usted S-Scan using modified3mplitude Scan Palette*
-
8/19/2019 Encoded C Scan
55/101
Color Palettes Color Ad=ustment' De%t!For De%t! C-Scans9 t!e color %alette scale is set in one of two wa s de%endingon !ow t!e Source %arameter is set'
!ate* T!e color scale is set wit! t!e gatestart and end %oints as t!e minimum andma>imum values of t!e scale wit! t!ecolors distri$uted evenl over t!e gatewidt!,
Custom* T!e user manuall sets t!estart and end %oints of t!e color scale,T!e colors are t!en distri$uted evenl$etween t!e start and end %oints,
I O$E* T!e Reference Cursors on t!e A-Scan can $e !el%ful in determining startand end %oints,
T!e user can also select !ow man color levels to use tos%an across t!e start and end %oints using &ept'Palette 9evels
-
8/19/2019 Encoded C Scan
56/101
Color Palettes Color Ad=ustment' De%t!
&efault Palette Source set to !ate*
G Four palette levels distri"uted overgate widt'
1?
-
8/19/2019 Encoded C Scan
57/101
Color Palettes Color Ad=ustment' De%t!
1?
G !ate 1 points ad=usted to capture onl%t'e area of part containing flaws
G Four palette levels now distri"uted overflaw area onl%
G Colors can also "e ad=usted to includeentire part t'ic,ness wit'outinterference from gated initial pulse or"ac,wall ec'o
Palette Source c'anged to Custom*
-
8/19/2019 Encoded C Scan
58/101
PA Color Palette Ad=ustment- De%t!
.nitial timed C-Scan using
Corrosion Scan Palette and!ate Source
3d=usted timed C-Scan usingCustom &ept' Scan Palette Colorgradient restructured around 'ole
dept' rat'er t'an gate position
-
8/19/2019 Encoded C Scan
59/101
Saving a C-scan
T!e com%lete C-scan and com%ressed A-scans can also $e stored int!e instrument datalogger for later review,To save a C-scan' Create and O%en a file of t!e desired t %e #it! t!e C-scan on screen9 %ress t!e 6S3@E7 6e
T!e C-scan can $e saved eit!er w!ile fro;en in Review "ode or w!enin Sto%%ed mode,
-
8/19/2019 Encoded C Scan
60/101
Recalling a C-scan
After recalling a saved C-scan from t!e Reviewdatalogger menu9 %ress t!e6ESC3PE7 6e to e>it Recallmode,To enter Review mode9 %resst!e 6F EE E7 6e , All t!efeatures availa$le w!enreviewing a fro;en C-scanwill now $e availa$le,
-
8/19/2019 Encoded C Scan
61/101
Recalling a C-scan
&f ou !ave recalled a C-scan from an incremental 0&/C4file9 ou will need to %ress t!e 6ESC3PE7 6e after t!escan is recalled to return to t!e live screen, T!e on t!e rig!t side of t!e waveform indicates t!at ou are in
Recall mode,
-
8/19/2019 Encoded C Scan
62/101
eturn to Presentation Contents
$ips for eld .nspection
&emonstration Procedure* ero &egree Flaw .nspection
-
8/19/2019 Encoded C Scan
63/101
-
8/19/2019 Encoded C Scan
64/101
#eld Sam%le An weld sam%le wit! 6nown flaws
similar to t!e one s!own a$ovewill wor6 for t!is demonstration
Cou%lant D-*8 :el T %e cou%lant
Transducer and #edge 7L*5- ,5 *+-A*+P-P-8,7-O" *5 element PA %ro$e SA*+-/77S wedge
Encoder and Ada%tor ca$le E/C*-8,7-DE "ini-#!eel Encoder CA?L-*++*5-+++2 EP*+++i
Ada%tor Ca$le
Required Equi%ment
-
8/19/2019 Encoded C Scan
65/101
Ti%s for #eld &ns%ectionFor demonstration of t!e encodedC-scan feature on a weld9 use aweld sam%le wit! 6nown flaws T!is can $e done on an weld
sam%le9 w!ic! will !ave differentflaws t!an t!e one used !ere
Encoded C-scans add anadditional dimension to standardS-scans $ ca%turing data acrossa swee% of angles at multi%le%ro$e %ositions along t!e welda>is,
A0 to B0&egree 3ngle
Sweep
E n c o
d e
d S c a n
3 5
i s
-
8/19/2019 Encoded C Scan
66/101
Locating #eld :eometr?egin wit! a full cali$rated S-scan9 and face t!e %ro$e toward a side of t!eweld %late and determine t!e %osition of %ro$e to see first and second legsignals from to% and $ottom of t!e %late,T!en %osition t!e %ro$e at t!is same offset distance to t!e weld to viewreflections from t!e weld root and weld crown,
eld!eometr%
-
8/19/2019 Encoded C Scan
67/101
#eld OverlaT!e #eld Overla feature canalso $e used to assist inlocating weld geometr on
our sector scan, ? in%utting our distance
awa from t!e centerline oft!e weld9 ou can %osition a$asic image of t!e weld over
our S-scan to aid inidentif ing weld ec!oes
eldCenterline
Pro"e FrontCursor
-
8/19/2019 Encoded C Scan
68/101
De%t! :ate Ad=ustment?ased on t!e s%ecific flawsin our %articular weldsam%le9 ou can set t!ede%t! gates to ca%ture a%articular flaw in a C-scan,De%t! gates are es%eciall
!el%ful in weld ins%ectiona%%lications $ecause oucan ca%ture an entire leg ofa %art in a single scan, 1sing leg indicators will
allow ou to %osition t!e
gate to create a com%leteC-scan for a given leg,
e sure to set %our rangeso t'at %ou are capturing
dept' data at all focal lawsin a given leg
&ept' gates set to capturesecond leg data
-
8/19/2019 Encoded C Scan
69/101
-
8/19/2019 Encoded C Scan
70/101
-
8/19/2019 Encoded C Scan
71/101
Angled Linear Scans
Angled linear scans can $e $eneficial w!en loo6ing for %orosit in welds
Crac,
Porosit%
-
8/19/2019 Encoded C Scan
72/101
-
8/19/2019 Encoded C Scan
73/101
&emonstration Procedure*ero &egree Flaw .nspection
-
8/19/2019 Encoded C Scan
74/101
Test ?loc6 EP*+++-PA?LOCJ-* P!ased
Arra Aluminum Demo ?loc6 Cou%lant D-*8 :el T %e cou%lant
Transducer and #edge 7L53-
-
8/19/2019 Encoded C Scan
75/101
*, ?egin wit! a full cali$rated linear scan
C-scan Setu%
-
8/19/2019 Encoded C Scan
76/101
8, Set ange to *,+++
-
8/19/2019 Encoded C Scan
77/101
7, Attac! 7L53-A*8 %ro$e to encoder 5, Perform encoder cali$ration@, Turn C-scan o%tion On 0C Scan # C Scan 42, Set Source 0P
-
8/19/2019 Encoded C Scan
78/101
C-scan Setu% 0cont,4*+, Select C Scan # Setup to enter t!e setu% screen**, Set t!e Encoder +ode to i &irectional *8, Set t!e &irection to 9eft to ig't*
-
8/19/2019 Encoded C Scan
79/101
-
8/19/2019 Encoded C Scan
80/101
(ero Degree C-scan Overview
Stored 3-scan forselected C-scan point
3mplitude C-scanof D F Hs
9inear Scan 35is(lengt' along pro"e)
Encoded Scan 35is (distance traveled "% encoder)
-
8/19/2019 Encoded C Scan
81/101
1sing Si;ing Cursors on t!e C-scanEncoded ;ero degree C-scans %rovide a to%-down %lanar view of t!e test %iecewit! %ro%erl scaled data collected along $ot! a>es of t!e scan,Si;ing cursors %ositioned along t!e lengt! of t!e %ro$e and t!e encoded scana>is can $e used to find t!e a%%ro>imate dimensions of a given flaw,
1se t!e P3 Cursors grou% menuto ad=ust and B cursor %ositions
6P17 toggles t!e and B cursorsOn and Off
P89 P
-
8/19/2019 Encoded C Scan
82/101
1sing Si;ing Cursors on t!e C-scan 0cont,4
To find t!e idt' or 9engt' oft!e F?)'*, Position t!e * or B* cursor to
intersect t!e ellow am%litudearea to t!e left of t!e F?)indication 0-5d? %oint ofimage4
8, Position t!e 8 or B8 cursor tointersect t!e ellow am%litudearea to t!e rig!t of t!e$ac6wall indication 0-5d? %ointof image4
-
8/19/2019 Encoded C Scan
83/101
Reviewing Saved Linear Scan Data A linear scan image will $e saved for eac! %oint on t!e C-scan,1se t!e D,T, 0 P3 Cursors # & $ 4 and &nde> 063 !9E7 4 cursors tolocate t!e %ea6 A-scan ec!o from t!e large F?) on t!e C-scan,
& $ Cursor
.nde5 Cursor
-
8/19/2019 Encoded C Scan
84/101
Reviewing Saved Linear Scan Data 0cont,4Switc! t!e C Scan # Screen to view t!e linear scan t!at was used tocreate t!at %articular C-scan %oint,
ac,walleflection
9arge F H
ac,wall&ropout
-
8/19/2019 Encoded C Scan
85/101
1sing Si;ing Cursors on t!e Linear ScanT!e linear scan can $e used to find t!e true de%t! of t!e selected F?),To find t!e &ept' of t!e F?)'
*, Position t!e B* cursor tointersect t!e %ea6am%litude of t!e F?)indication
8, Position t!e B8 cursor tointersect t!e %ea6am%litude of t!e $ac6wallindication
-
8/19/2019 Encoded C Scan
86/101
eturn to Presentation Contents
-
8/19/2019 Encoded C Scan
87/101
-
8/19/2019 Encoded C Scan
88/101
Required Equi%mentT!e following slides detail !ow to %ro%erl attac! t!e "ini-#!eel encoder to
our 7L*5-A*+P and 7L53-A*8 %ro$es for encoded C-scans,Please note t!at t!ere are man wa s to attac! t!ese %arts de%ending ona%%lication and user %reference,
"ini-#!eel Encoder 012@2+* @4
)e>agonal 6e screwdriver9 mounting $rac6et %in 0included wit! encoder4Flat!ead screwdriver 7L*5-A*+P 012
-
8/19/2019 Encoded C Scan
89/101
Encoder Attac!ment 7L*5-A*+P
*, ?egin wit! our SA*+-/77S 0orSA*+-+L4 wedge attac!ed to our%ro$e using cou%lant
8, /ote t!e location of t!e t!readed!ole on t!e side of t!e wedge
-
8/19/2019 Encoded C Scan
90/101
Encoder Attac!ment 7L*5-A*+P 0cont,47, Screw %art of t!readed end of
t!e mounting %in into t!et!readed !ole on t!e wedge sot!at t!e wedge is facing in t!esame direction as t!e encoderw!eel
5, 1se t!e flat!ead screwdriver toscrew t!e mounting %in all t!ewa into t!e wedge until it istig!t Do not overtig!ten or ou can
damage t!e t!readsK
-
8/19/2019 Encoded C Scan
91/101
-
8/19/2019 Encoded C Scan
92/101
-
8/19/2019 Encoded C Scan
93/101
Encoder Attac!ment 7L53-A*8 0cont,47, Screw %art of t!readed end of
t!e mounting %in into t!et!readed !ole on t!e wedge sot!at t!e wedge is facing in t!esame direction as t!e encoderw!eel
5, 1se t!e flat!ead screwdriver toscrew t!e mounting %in all t!ewa into t!e wedge until it istig!t Do not overtig!ten or ou can
damage t!e t!readsK
-
8/19/2019 Encoded C Scan
94/101
-
8/19/2019 Encoded C Scan
95/101
eturn to Presentation Contents
eturn to Encoder Cali"ration (from S-scan)
eturn to Encoder Cali"ration (from 9inear scan)
-
8/19/2019 Encoded C Scan
96/101
Encoder Cali"ration
-
8/19/2019 Encoded C Scan
97/101
Required Equi%ment?efore creating an encoded C-scan9 ou must cali$rate t!e encoderto determine t!e correct encoder resolution to ensure accuratedistance traveled data,Required Equi%ment'
Encoder P!ased arra %ro$e and wedge Ruler9 ideall at least 5 in, 0*7+ mm4
-
8/19/2019 Encoded C Scan
98/101
Cali$ration Setu%
Gsing a greater cali"rationdistance will increase encoderresolution accurac% as error isaveraged over a larger distance
Once t!e encoder !as $een %ro%erl attac!ed to our %ro$e'*, Enter t!e Encoder # Cali"ration menu8, Press 6PD7to eset t!e encoder %ulses and D,T, values
-
8/19/2019 Encoded C Scan
99/101
Encoder Cali$ration3, Position t!e front of t!e %ro$e
at t!e ;ero mar6 0start %oint4on our ruler
7, Press 6P17 Cal Start to $egint!e cali$ration
5, "ove t!e %ro$e until t!e frontis at t!e end %oint 05 in,4 on
our ruler @, Press 6P17 End Cal to
com%lete t!e cali$ration
-
8/19/2019 Encoded C Scan
100/101
Encoder Cali$ration 0cont,4 After t!e cali$ration !as $een com%lete9 t!e newl calculated EncoderResolution will dis%la ,
T!e resolution for t!e "ini-#!eel Encoder s!ould $e a%%ro>imatel @5 %ulses.inc!0< %ulses.mm4,
Press t!e 6ESC7 6e to return to t!e live screen
-
8/19/2019 Encoded C Scan
101/101
eturn to Presentation Contents
eturn to C-scan Setup (from S-scan)
eturn to C-scan Setup (from 9inear scan)