emittance measuremet on the ceb6 electron gun for the

20
Emittance Measuremet on the CeB6 Electron Gun for the SPring-8 Compact SASE Source 26th International Free Electron Laser Conference (FEL2004) September 2, 2004 Trieste, Italy T. Shintake, H. Baba, T. Inagaki, K. Onoe, T. Tanaka SPring-8 / RIKEN Harima Institute H. Matsumoto High Energy Accelerator Research Organization (KEK) K. Togawa SPring-8 / RIKEN Harima Institute

Upload: others

Post on 22-Jan-2022

2 views

Category:

Documents


0 download

TRANSCRIPT

THBOC03_talk.pdfEmittance Measuremet on the CeB6 Electron Gun for the SPring-8 Compact SASE Source
26th International Free Electron Laser Conference (FEL2004)
September 2, 2004 Trieste, Italy
T. Shintake, H. Baba, T. Inagaki, K. Onoe, T. Tanaka SPring-8 / RIKEN Harima Institute
H. Matsumoto High Energy Accelerator Research Organization (KEK)
K. Togawa SPring-8 / RIKEN Harima Institute
1:20
C o
lli m
a to
Traveling Wave
1428 MHz
Traveling Wave
2 e
Long Lifetime
>10,000 hours
Properties Very flat surface
(surface roughness <1 µm)
Low workfunction (~2.4 eV)
Long lifetime (>10,000 hours)
Rapid recovery from contamination
Thermal Emittance
Cathode Radius r=1.5 mm Temperature T=1450 (1723 K)
Emission Current Density Richardson-Dashman's Formula (Ideal Case)
)kBT/exp(T4.120J 2 φ'−=
CeB6 Cathode Assembly (New Model) SCSS
Cathode Assembly Heated Cathode in Stem
CeB6 Single Crystal 3mm
Cathode Temperature ~1500 C
Graphite Heater
The gun voltage=500 kV Temperature was measured at the graphite sleeve by a radiation monitor.
500 kV Electron Gun
Pulse Transformer
M ag
ne tic
Le ns
S te
er in
g C
oi l
G at
e V
al ve
C T
M on
ito r
M ov
ab le
S lit
M ov
ab le
S lit
M ov
ab le
S lit
M ov
ab le
S lit
B ea
m D
um p
HV Bushing
C-band klystron modulator is used as a HV pulsed power supply.
C-band Klystron Modulator SCSS
Same model of the C-band klystron modulator is used for the 500 kV electron gun.
3 µs
517 kV
Circuit Diagram
Gun Pulse Waveform
PFN : 13 section (for 500 kV Gun) PFN Impedance : ~4 Charging Voltage : 50 kV max Max. Repetition Rate : 60 pps
500 kV Electron Gun Chamber SCSS
CeramicsCorona Dome Magnetic Sheld
Beam Waveform I-V Curve
We operate the gun in temperature limited region to reduce emittance growth due to space charge.
Beam Current (CT Monitor)
Gun Voltage (CVD Monitor)
Space Charge Limit Schottky Effect Droop of CT Monotor
~1 ns part will cut out from the flattop by a beam deflector, and be used for the SCSS accelerator.
Emittance Measurement
x222 xxxx ′⋅−′βγ=
∫ ∫ i ∫ ∫ i=
2 2
dxdx'x'x dxdx'x'xxx
Electron Beam Beamlet Sub-beamlet
Emittance Monitor SCSS
Beamlet Profiles (400 keV, 0.9 A)
Beamlet spread due to space charge is about 15% at 50 µm width.
Beamlet Width (FWHM)
Beamlet spreads due to not only the thermal momentum but also the space charge force.
Opening width of the upstream slit must be narrow enough that the space charge effect can be neligible.
Emittance of 500 keV Beam SCSS
Beam Profile Phase Space Profile
Requirement : <1π mm.mrad Experiment : 1.1π mm.mrad
Emittance (εn,rms)Beam Energy : 500 keV Peak Current : 1.0 A Pulse Width : 3 µs
φ 5.6 mm (FWHM)
φ 5.6 mm (FWHM)
Linear Space Charge
x' Phase Space Profile
Electrons diffuse to the outside by thermal motion. Non-linear space charge force breaks emittance.
cut
Cathode
Anode
2/r 'xx'xx
= 0.6 π mm.mrad (expected)
Summary SCSS
1) We have succeeded in producing a 500 keV, 1 A beam from the CeB6 gun.
2) Measured emittance was ~1π mm mrad.