electrosurgical life-test fixture team e.l.f. design review mechanical engineers mary hamann brad...
TRANSCRIPT
![Page 1: Electrosurgical Life-test Fixture Team E.L.F. Design Review Mechanical Engineers Mary Hamann Brad Watson Naomi Sanders Electrical Engineers Tony Giedl](https://reader036.vdocuments.us/reader036/viewer/2022062421/56649da25503460f94a8f196/html5/thumbnails/1.jpg)
Electrosurgical Life-test Fixture
Team E.L.F. Design Review
Mechanical Engineers
Mary Hamann
Brad Watson
Naomi Sanders
Electrical Engineers
Tony Giedl
Jim Michalk
![Page 2: Electrosurgical Life-test Fixture Team E.L.F. Design Review Mechanical Engineers Mary Hamann Brad Watson Naomi Sanders Electrical Engineers Tony Giedl](https://reader036.vdocuments.us/reader036/viewer/2022062421/56649da25503460f94a8f196/html5/thumbnails/2.jpg)
Design Review Outline
•Background•Problem Definition•Project Needs•Flow Diagram•Components
FrameWand MountPressure ControlAngle AdjustmentTissue Holder
•Integrated Design
![Page 3: Electrosurgical Life-test Fixture Team E.L.F. Design Review Mechanical Engineers Mary Hamann Brad Watson Naomi Sanders Electrical Engineers Tony Giedl](https://reader036.vdocuments.us/reader036/viewer/2022062421/56649da25503460f94a8f196/html5/thumbnails/3.jpg)
Background
ArthroCare manufactures a device used in surgery for the removal of tissue. This tool is more precise than previous methods of tissue removal and minimizes damage to healthy tissue. Currently, technicians have to hand-test a certain number of Arthro-Wands out of every new batch that is manufactured to ensure the life of each device. This is a very time and labor intensive task. ArthroCare would like an apparatus that is capable of mechanically testing their Arthro-Wands. This will cut down on labor costs of the technicians and allow more wands to be manufactured and tested.
![Page 4: Electrosurgical Life-test Fixture Team E.L.F. Design Review Mechanical Engineers Mary Hamann Brad Watson Naomi Sanders Electrical Engineers Tony Giedl](https://reader036.vdocuments.us/reader036/viewer/2022062421/56649da25503460f94a8f196/html5/thumbnails/4.jpg)
Problem Definition
The purpose of this project is to develop a fixture that will automatically test Arthro-Wands. The device has to mimic the condition that the wand will observe during an actual surgery in order to ensure the reliability of the product.
![Page 5: Electrosurgical Life-test Fixture Team E.L.F. Design Review Mechanical Engineers Mary Hamann Brad Watson Naomi Sanders Electrical Engineers Tony Giedl](https://reader036.vdocuments.us/reader036/viewer/2022062421/56649da25503460f94a8f196/html5/thumbnails/5.jpg)
Project Needs
•Downsize the footprint of the apparatus (roughly 1/4
original size)
•Easy reproducibility
•Short set-up time
•System completes one lifetime without traveling over used
tissue
•Capable of accommodating a variety of handle sizes
•Adjustable to a variety of angles and positions
![Page 6: Electrosurgical Life-test Fixture Team E.L.F. Design Review Mechanical Engineers Mary Hamann Brad Watson Naomi Sanders Electrical Engineers Tony Giedl](https://reader036.vdocuments.us/reader036/viewer/2022062421/56649da25503460f94a8f196/html5/thumbnails/6.jpg)
Flow Diagram
Setup
Tissue Fixture Wand
Input
Time Speed
Run Fixture
Wand Oscillation Increment
Control Motion
![Page 7: Electrosurgical Life-test Fixture Team E.L.F. Design Review Mechanical Engineers Mary Hamann Brad Watson Naomi Sanders Electrical Engineers Tony Giedl](https://reader036.vdocuments.us/reader036/viewer/2022062421/56649da25503460f94a8f196/html5/thumbnails/7.jpg)
Frame Designs
Concept #1 – X-Y Gantry Assembly
Concept #2 – Last Years Design
Concept #3 - “Toy Grabber”
Guide Rods
Lead Screws
Sliders
Wand Mount
![Page 8: Electrosurgical Life-test Fixture Team E.L.F. Design Review Mechanical Engineers Mary Hamann Brad Watson Naomi Sanders Electrical Engineers Tony Giedl](https://reader036.vdocuments.us/reader036/viewer/2022062421/56649da25503460f94a8f196/html5/thumbnails/8.jpg)
Frame Concept #1X-Y Gantry Assembly
linear motion video
Advantages:
•Smaller Footprint than last years, due to the stationary tray
•Could Eliminate Scotch Yoke
•Stationary Tray and one moving assembly
Disadvantages:
•Vertical Lead Screw located to one side creates a torque on the top lead screw assembly
•Frequent direction change on the stepper motors
•Indexing of the wand is coupled with oscillatory motion
This Assembly has both the horizontal and vertical linear motion located on the top of the frame. The sinusoidal motion can be created by either an improved scotch yoke or the vertical linear motion lead screw.
![Page 9: Electrosurgical Life-test Fixture Team E.L.F. Design Review Mechanical Engineers Mary Hamann Brad Watson Naomi Sanders Electrical Engineers Tony Giedl](https://reader036.vdocuments.us/reader036/viewer/2022062421/56649da25503460f94a8f196/html5/thumbnails/9.jpg)
Frame Concept #2Last Years Design
Advantages:
•Horizontal and vertical motion are de-coupled
•No moment created from off-centered lead screw
•Slides and Frame material all stock
•Could Eliminate the Scotch Yoke
Disadvantages:
•Larger Footprint due to moving tray
•Frequent direction change of the stepper motors
•Both the tray and the tissue mount observe movement
Concept #2 is similar to last years design, with the horizontal-motion on the top and the vertical motion on the bottom. The sinusoidal motion can be created by either an improved scotch yoke or the vertical motion of the tray.
![Page 10: Electrosurgical Life-test Fixture Team E.L.F. Design Review Mechanical Engineers Mary Hamann Brad Watson Naomi Sanders Electrical Engineers Tony Giedl](https://reader036.vdocuments.us/reader036/viewer/2022062421/56649da25503460f94a8f196/html5/thumbnails/10.jpg)
Frame Concept #3 Toy Grabber
Advantages:
•Decreased moment on the horizontal lead screw assembly
•Smaller footprint in the vertical direction due to stationary tray
•Eliminates Scotch Yoke
Disadvantage:
•Less Space to adjust the position and angle of the wand
•Larger Footprint in the horizontal direction due to the vertical motion lead screw and guide rods
•Frequent direction change of the stepper motors
This design has the horizontal and vertical motion on top, with both lead screw assemblies having linear motion along the top rails of the frame. In this design the lead screws connect at the location of the wand mount platform.
![Page 11: Electrosurgical Life-test Fixture Team E.L.F. Design Review Mechanical Engineers Mary Hamann Brad Watson Naomi Sanders Electrical Engineers Tony Giedl](https://reader036.vdocuments.us/reader036/viewer/2022062421/56649da25503460f94a8f196/html5/thumbnails/11.jpg)
Wand Mount Designs
Concept #3 – “Christmas Tree Mount”
Concept #2 – Vice Clamp
Concept #1 – Clamp
![Page 12: Electrosurgical Life-test Fixture Team E.L.F. Design Review Mechanical Engineers Mary Hamann Brad Watson Naomi Sanders Electrical Engineers Tony Giedl](https://reader036.vdocuments.us/reader036/viewer/2022062421/56649da25503460f94a8f196/html5/thumbnails/12.jpg)
Wand Mount Concept #1Clamp
Advantages:
•Securely hold a variety of wand sizes and shapes
•Pre-manufactured options
•Clamp handle easily attaches to the wand mount platform
•Prevents motion of the wand during testing
•Quick set-up time
Disadvantages:
•May not accommodate all sizes of wand handles
This design is a spring loaded clamp that will hold the wand securely in place and will accommodate a variety of sizes
![Page 13: Electrosurgical Life-test Fixture Team E.L.F. Design Review Mechanical Engineers Mary Hamann Brad Watson Naomi Sanders Electrical Engineers Tony Giedl](https://reader036.vdocuments.us/reader036/viewer/2022062421/56649da25503460f94a8f196/html5/thumbnails/13.jpg)
Wand Mount Concept #2Vice Clamp
Advantages:
•Securely holds a variety of wand sizes and shapes
•Accommodates all sizes of wand handles
•Quick set-up time
Disadvantages:
•Doesn’t easily attach to the wand platform
•May allow movement of the wand during testing
This design is a contoured vice clamp that can be adjusted with a screw to fit wand handles of various sizes
![Page 14: Electrosurgical Life-test Fixture Team E.L.F. Design Review Mechanical Engineers Mary Hamann Brad Watson Naomi Sanders Electrical Engineers Tony Giedl](https://reader036.vdocuments.us/reader036/viewer/2022062421/56649da25503460f94a8f196/html5/thumbnails/14.jpg)
Wand Mount Concept #3Christmas Tree Mount
Advantages:
•Highly adjustable for a variety of wand shapes and sizes
•Easily Manufactured
Disadvantages:
•Difficult to align in the holder
•May allow motion of the wand during testing
•Increased set-up time due to the multiple screw adjustment
•Does not easily attach to the wand mount platform
This is a mounting device with three equidistant screws that can be adjusted to accommodate any size of wand.
![Page 15: Electrosurgical Life-test Fixture Team E.L.F. Design Review Mechanical Engineers Mary Hamann Brad Watson Naomi Sanders Electrical Engineers Tony Giedl](https://reader036.vdocuments.us/reader036/viewer/2022062421/56649da25503460f94a8f196/html5/thumbnails/15.jpg)
Pressure Control Designs
Design Concept #1-
Spring Load Control
Design concept #2 – Gravitational Load Control
![Page 16: Electrosurgical Life-test Fixture Team E.L.F. Design Review Mechanical Engineers Mary Hamann Brad Watson Naomi Sanders Electrical Engineers Tony Giedl](https://reader036.vdocuments.us/reader036/viewer/2022062421/56649da25503460f94a8f196/html5/thumbnails/16.jpg)
Pressure Control Concept #1Spring Load Control
Advantages:
•Works at all wand angles
•Design fully contained – no separate parts or weights
Disadvantages:
•Difficult to set desired pressure
•Contains more components
•Pressure not always constant
This design uses springs to create a constant pressure on the wand and allows vertical motion of the wand with the change in pressure of the test subject
![Page 17: Electrosurgical Life-test Fixture Team E.L.F. Design Review Mechanical Engineers Mary Hamann Brad Watson Naomi Sanders Electrical Engineers Tony Giedl](https://reader036.vdocuments.us/reader036/viewer/2022062421/56649da25503460f94a8f196/html5/thumbnails/17.jpg)
Pressure Control Concept #2Gravitational Load Control
Advantages:
•Works at all wand angles
•Pressure always constant
•Easy set-up
Disadvantages:
•Requires additional parts -need separate weights to apply the constant load
This design simply uses gravity to maintain a constant force on the surface of the chicken. The assembly has a set weight applied and it can slide freely up and down with the contours of the chicken.
![Page 18: Electrosurgical Life-test Fixture Team E.L.F. Design Review Mechanical Engineers Mary Hamann Brad Watson Naomi Sanders Electrical Engineers Tony Giedl](https://reader036.vdocuments.us/reader036/viewer/2022062421/56649da25503460f94a8f196/html5/thumbnails/18.jpg)
Angle Adjustment Designs
Concept #3 - Slotted Runner
Concept #1 – Screw adjustment
Concept #2 – Screw
Clamped Adjustor
![Page 19: Electrosurgical Life-test Fixture Team E.L.F. Design Review Mechanical Engineers Mary Hamann Brad Watson Naomi Sanders Electrical Engineers Tony Giedl](https://reader036.vdocuments.us/reader036/viewer/2022062421/56649da25503460f94a8f196/html5/thumbnails/19.jpg)
Angle Adjustment Concept #1Screw Adjustment
Advantages:
•Quick set-up time
•Fine adjustment capabilities
•Easily connected to the wand platform
Disadvantages:
•Wand tip is only centered at a vertical angle
A threaded rod is used to attach to the wand holder to the pressure control device. The wand holder can be adjusted to any angle and bolted in place.
![Page 20: Electrosurgical Life-test Fixture Team E.L.F. Design Review Mechanical Engineers Mary Hamann Brad Watson Naomi Sanders Electrical Engineers Tony Giedl](https://reader036.vdocuments.us/reader036/viewer/2022062421/56649da25503460f94a8f196/html5/thumbnails/20.jpg)
Angle Adjustment Concept #2
Advantages:
•Fine adjustment capabilities
•Easily connected to the wand platform
Disadvantages:
•More complicated set-up
•Wand tip is only centered at a vertical angle
A rod extending from the wand holder is clamped into the pressure control device. The wand holder can be adjusted to any angle and clamped in place.
![Page 21: Electrosurgical Life-test Fixture Team E.L.F. Design Review Mechanical Engineers Mary Hamann Brad Watson Naomi Sanders Electrical Engineers Tony Giedl](https://reader036.vdocuments.us/reader036/viewer/2022062421/56649da25503460f94a8f196/html5/thumbnails/21.jpg)
Angle Adjustment Concept #3
Advantages:
•Quick and easy adjustment
•Wand tip is always centered
Disadvantages:
•Height differential is limited
•Does not connect easily to the wand platform
This is a slotted semi-circular device to which the wand can be attached to in a variety of angles and heights
![Page 22: Electrosurgical Life-test Fixture Team E.L.F. Design Review Mechanical Engineers Mary Hamann Brad Watson Naomi Sanders Electrical Engineers Tony Giedl](https://reader036.vdocuments.us/reader036/viewer/2022062421/56649da25503460f94a8f196/html5/thumbnails/22.jpg)
Tissue Holder Design
The design for the tissue holder is essentially the same design as last year’s. A frame is placed over the chicken and bolted down to hold it still during the testing period.
![Page 23: Electrosurgical Life-test Fixture Team E.L.F. Design Review Mechanical Engineers Mary Hamann Brad Watson Naomi Sanders Electrical Engineers Tony Giedl](https://reader036.vdocuments.us/reader036/viewer/2022062421/56649da25503460f94a8f196/html5/thumbnails/23.jpg)
Integrated Design
![Page 24: Electrosurgical Life-test Fixture Team E.L.F. Design Review Mechanical Engineers Mary Hamann Brad Watson Naomi Sanders Electrical Engineers Tony Giedl](https://reader036.vdocuments.us/reader036/viewer/2022062421/56649da25503460f94a8f196/html5/thumbnails/24.jpg)
Assembled Design Components
Slider to allow for large vertical adjustment
Tray sits on platform
and indexes as well as oscillates the chicken
Mass applies constant load on chicken
Wand angle set by the wand holder
Wand mount slides with the contour of the chicken