electron spectroscopy for surface analysis
DESCRIPTION
Electron Spectroscopy for Surface Analysis. Introduction XPS Background XPS Instrument How Does XPS Technology Work? Auger Electron Cylindrical Mirror Analyzer (CMA) Equation KE versus BE Spectrum Background Identification of XPS Peaks X-rays vs. e - Beam XPS Technology. Outline. - PowerPoint PPT PresentationTRANSCRIPT
Electron Spectroscopy for Surface Analysis
1
Outline Introduction XPS Background XPS Instrument How Does XPS Technology Work? Auger Electron Cylindrical Mirror Analyzer (CMA) Equation KE versus BE Spectrum Background Identification of XPS Peaks X-rays vs. e- Beam XPS Technology
2
Introduction
Electron spectroscopy (ES) is a technique that uses characteristic electrons emitted from solid for elemental analysis, not for imaging as in electron microscopy;
The characteristic electrons (either Auger electrons or photoelectrons) exhibit characteristics energy levels, revealing the nature of chemical elements in specimen being examined;
3
Auger or photoelectrons can only escape from the uppermost atomic layers of solid (a depth of 10 nm or less) because their energies are relative low (generally 20-2000 eV);
While the characteristic X-rays can escape from a much greater depth (several micrometers from the surface);
Thus, ES is a technique for surface analysis; There two types of ES: Auger electron
spectroscopy (AES) and X-ray photoelectron spectroscopy (XPS);
Auger electrons and photoelectrons are different in their physical origins, but both types of electrons carry similar information about chemical elements in material surfaces.
4
X-ray Photoelectron SpectroscopySmall Area Detection
X-ray BeamX-ray Beam
X-ray penetration X-ray penetration depth ~1depth ~1m.m.Electrons can be Electrons can be excited in this excited in this entire volume.entire volume.
X-ray excitation area ~1x1 cmX-ray excitation area ~1x1 cm22. Electrons . Electrons are emitted from this entire areaare emitted from this entire area
Electrons are extracted Electrons are extracted only from a narrow solid only from a narrow solid angle.angle.
1 mm1 mm22
10 nm10 nm
5
XPS Background XPS technique is based on Einstein’s idea about the
photoelectric effect, developed around 1905
The concept of photons was used to describe the ejection of electrons from a surface when photons were impinged upon it
During the mid 1960’s Dr. Siegbahn and his research
group developed the XPS technique.
In 1981, Dr. Siegbahn was awarded the Nobel Prize in Physics for the development of the XPS technique
6
X-Rays Irradiate the sample surface, hitting the core electrons (e-) of the
atoms.
The X-Rays penetrate the sample to a depth on the order of a micrometer.
Useful e- signal is obtained only from a depth of around 10 to 100 Å on the surface.
The X-Ray source produces photons with certain energies: MgK photon with an energy of 1253.6 eV AlK photon with an energy of 1486.6 eV
Normally, the sample will be radiated with photons of a single energy (MgK or AlK). This is known as a monoenergetic X-Ray beam.
7
Why the Core Electrons? An electron near the Fermi level is far from the nucleus,
moving in different directions all over the place, and will not carry information about any single atom. Fermi level is the highest energy level occupied by an
electron in a neutral solid at absolute 0 temperature. Electron binding energy (BE) is calculated with respect to the
Fermi level.
The core e-s are local close to the nucleus and have binding energies characteristic of their particular element.
The core e-s have a higher probability of matching the energies of AlK and MgK.
Core e-
Valence e-
Atom
8
Binding Energy (BE)
These electrons are attracted to the
proton with certain binding energy x
This is the point with 0 energy of attraction
between the electron and the nucleus. At this point the electron is free from
the atom.
The Binding Energy (BE) is characteristic of the core electrons for each element. The BE is determined by the attraction of the electrons to the nucleus. If an electron with energy x is pulled away from the nucleus, the attraction between the electron and the nucleus decreases and the BE decreases. Eventually, there will be a point when the electron will be free of the nucleus.
0
x
p+
B.E.
9
Energy Levels
Vacumm Level
Fermi Level
Lowest state of energy
BE
Ø, which is the work function
At absolute 0 Kelvin the electrons fill from the lowest energy states up. When the electrons occupy up to this level the neutral solid is in its “ground state.”
10
X-ray Photoelectron Spectrometer
11
XPS Instrument
XPS is also known as ESCA (Electron Spectroscopy for Chemical Analysis).
The technique is widely used because it is very simple to use and the data is easily analyzed.
XPS works by irradiating atoms of a surface of any solid material with X-Ray photons, causing the ejection of electrons.
University of Texas at El Paso, Physics DepartmentFront view of the Phi 560 XPS/AES/SIMS UHV System
12
XPS Instrument
The XPS is controlled by using a computer system.
The computer system will control the X-Ray type and prepare the instrument for analysis.
University of Texas at El Paso, Physics DepartmentFront view of the Phi 560 XPS/AES/SIMS UHV System and
the computer system that controls the XPS. 13
XPS Instrument
The instrument uses different pump systems to reach the goal of an Ultra High Vacuum (UHV) environment.
The Ultra High Vacuum environment will prevent contamination of the surface and aid an accurate analysis of the sample.University of Texas at El Paso, Physics Department
Side view of the Phi 560 XPS/AES/SIMS UHV System
14
XPS Instrument
X-Ray Source
Ion Source
SIMS Analyzer
Sample introductionChamber
15
Sample Introduction Chamber
The sample will be introduced through a chamber that is in contact with the outside environment
It will be closed and pumped to low vacuum.
After the first chamber is at low vacuum the sample will be introduced into the second chamber in which a UHV environment exists.
First ChamberSecond Chamber UHV
16
Diagram of the Side View of XPS System
X-Ray source
Ion source
Axial Electron Gun
Detector
CMA
sample
SIMS Analyzer
Sample introduction Chamber
Sample Holder
Ion PumpRoughing Pump Slits
17
How Does XPS Technology Work?
A monoenergetic x-ray beam emits photoelectrons from the from the surface of the sample.
The X-Rays either of two energies:
Al Ka (1486.6eV) Mg Ka (1253.6 eV)
The x-ray photons The penetration about a micrometer of the sample
The XPS spectrum contains information only about the top 10 - 100 Ǻ of the sample.
Ultrahigh vacuum environment to eliminate excessive surface contamination.
Cylindrical Mirror Analyzer (CMA) measures the KE of emitted e-s.
The spectrum plotted by the computer from the analyzer signal.
The binding energies can be determined from the peak positions and the elements present in the sample identified.
18
Why Does XPS Need UHV? Contamination of surface
XPS is a surface sensitive technique.• Contaminates will produce an XPS signal and lead to incorrect
analysis of the surface of composition.
The pressure of the vacuum system is < 10-9 Torr
Removing contamination To remove the contamination the sample surface is bombarded
with argon ions (Ar+ = 3KeV). heat and oxygen can be used to remove hydrocarbons
The XPS technique could cause damage to the surface, but it is negligible.
19
Remove adsorbed gases from Remove adsorbed gases from the sample.the sample.
Eliminate adsorption of Eliminate adsorption of contaminants on the sample. contaminants on the sample.
Prevent arcing and high voltage Prevent arcing and high voltage breakdown.breakdown.
Increase the mean free path for Increase the mean free path for electrons, ions and photons.electrons, ions and photons.
Degree of VacuumDegree of Vacuum1010
1010
1010
1010
1010
22
-1-1
-4-4
-8-8
-11-11
Low VacuumLow Vacuum
Medium VacuumMedium Vacuum
High VacuumHigh Vacuum
Ultra-High VacuumUltra-High Vacuum
PressurePressureTorrTorr
Why UHV for Surface Analysis?
20
XPS spectral lines are XPS spectral lines are identified by the shell from identified by the shell from which the electron was which the electron was ejected (1s, 2s, 2p, etc.).ejected (1s, 2s, 2p, etc.).
The ejected photoelectron has The ejected photoelectron has kinetic energy:kinetic energy:
KE=hv-BE-KE=hv-BE- Following this process, the Following this process, the
atom will release energy by atom will release energy by the emission of an Auger the emission of an Auger Electron.Electron.
Conduction BandConduction Band
Valence BandValence Band
L2,L3L2,L3
L1L1
KK
FermiFermiLevelLevel
Free Free Electron Electron LevelLevel
Incident X-rayIncident X-rayEjected PhotoelectronEjected Photoelectron
1s1s
2s2s
2p2p
The Photoelectric Process
21
X-Rays on the Surface
Atoms layers
e- top layer
e- lower layer with collisions
e- lower layer but no collisions
X-Rays
Outer surface
Inner surface
22
X-Rays on the Surface The X-Rays will penetrate to the core e- of the atoms in the
sample.
Some e-s are going to be released without any problem giving the Kinetic Energies (KE) characteristic of their elements.
Other e-s will come from inner layers and collide with other e-s of upper layers These e- will be lower in lower energy. They will contribute to the noise signal of the spectrum.
23
X-Rays and the ElectronsX-RayElectron without collision
Electron with collision
The noise signal comes from the electrons that collide with other electrons of different layers. The collisions cause a decrease in energy of the electron and it no longer will contribute to the characteristic energy of the element.
24
What e-s can the Cylindrical Mirror Analyzer Detect?
The CMA not only can detect electrons from the irradiation of X-Rays, it can also detect electrons from irradiation by the e- gun.
The e- gun it is located inside the CMA while the X-Ray source is located on top of the instrument.
The only electrons normally used in a spectrum from irradiation by the e- gun are known as Auger e-s. Auger electrons are also produced by X-ray irradiation.
25
X-Rays and Auger Electrons
When the core electron leaves a vacancy an electron of higher energy will move down to occupy the vacancy while releasing energy by: photons Auger electrons
Each Auger electron carries a characteristic energy that can be measured.
26
Two Ways to Produce Auger Electrons
1. The X-Ray source can irradiate and remove the e- from the core level causing the e- to leave the atom
2. A higher level e- will occupy the vacancy. 3. The energy released is given to a third higher
level e-. 4. This is the Auger electron that leaves the atom.
The axial e- gun can irradiate and remove the core e- by collision. Once the core vacancy is created, the Auger electron process occurs the same way.
27
L electron falls to fill core level L electron falls to fill core level vacancy (step 1).vacancy (step 1).
KLL Auger electron emitted to KLL Auger electron emitted to conserve energy released in conserve energy released in step 1.step 1.
The kinetic energy of the The kinetic energy of the emitted Auger electron is: emitted Auger electron is:
KE=E(K)-E(L2)-E(L3).KE=E(K)-E(L2)-E(L3).
Conduction BandConduction Band
Valence BandValence Band
L2,L3L2,L3
L1L1
KK
FermiFermiLevelLevel
Free Free Electron Electron LevelLevel
Emitted Auger ElectronEmitted Auger Electron
1s1s
2s2s
2p2p
Auger Relation of Core Hole
28
Auger Electron Spectroscopy (AES)
Atom layers
e- released fromthe top layer
Outer surface
Inner surfaceElectron beam from the e- gun
29
Cylindrical Mirror Analyzer (CMA)
The electrons ejected will pass through a device called a CMA.
The CMA has two concentric metal cylinders at different voltages.
One of the metal cylinders will have a positive voltage and the other will have a 0 voltage. This will create an electric field between the two cylinders.
The voltages on the CMA for XPS and Auger e-s are different.
30
Cylindrical Mirror Analyzer (CMA)
When the e-s pass through the metal cylinders, they will collide with one of the cylinders or they will just pass through. If the e-’s velocity is too high it will collide with the
outer cylinder If is going too slow then will collide with the inner
cylinder. Only the e- with the right velocity will go through the
cylinders to reach the detector.
With a change in cylinder voltage the acceptable kinetic energy will change and then you can count how many e-s have that KE to reach the detector.
31
Cylindrical Mirror Analyzer (CMA)
Slit
Detector
Electron Pathway through the CMA
0 V
+V
0 V 0 V
0 V
+V
+V
+V
X-RaysSource
SampleHolder
32
KEKE = hv - = hv - BEBE - - specspec
Where: Where: BEBE= Electron Binding Energy= Electron Binding Energy
KEKE= Electron Kinetic Energy= Electron Kinetic Energy
specspec= Spectrometer Work Function= Spectrometer Work Function
Photoelectron line energies: Photoelectron line energies: DependentDependent on photon energy.on photon energy.
Auger electron line energies: Auger electron line energies: Not DependentNot Dependent on photon energy.on photon energy.
If XPS spectra were presented on a kinetic energy If XPS spectra were presented on a kinetic energy scale, one would need to know the X-ray source energy used to scale, one would need to know the X-ray source energy used to collect the data in order to compare the chemical states in the collect the data in order to compare the chemical states in the sample with data collected using another source.sample with data collected using another source.
XPS Energy Scale- Kinetic energy
Equation
33
XPS Energy Scale- Binding energy
BEBE = hv - = hv - KEKE - - specspec
Where: Where: BEBE= Electron Binding Energy= Electron Binding Energy
KEKE= Electron Kinetic Energy= Electron Kinetic Energy
specspec= Spectrometer Work Function= Spectrometer Work Function
Photoelectron line energies: Photoelectron line energies: Not DependentNot Dependent on photon on photon energy.energy.
Auger electron line energies: Auger electron line energies: DependentDependent on photon on photon energy.energy.
The binding energy scale was derived to make uniform The binding energy scale was derived to make uniform comparisons of chemical states straight forward.comparisons of chemical states straight forward.
34
KE versus BE
E E E
KE can be plotted depending on BE
Each peak represents the amount of e-s at a certain energy that is characteristic of some element.
1000 eV 0 eV
BE increase from right to left
KE increase from left to right Binding energy
# o
f ele
ctr
on
s
(eV)
35
Interpreting XPS Spectrum: Background The X-Ray will hit the e-s
in the bulk (inner e- layers) of the sample
e- will collide with other e- from top layers, decreasing its energy to contribute to the noise, at lower kinetic energy than the peak .
The background noise increases with BE because the SUM of all noise is taken from the beginning of the analysis. Binding energy
# o
f ele
ctr
on
sN1
N2
N3
N4
Ntot= N1 + N2 + N3 + N4
N = noise
36
XPS Spectrum The XPS peaks are sharp.
In a XPS graph it is possible to see Auger electron peaks.
The Auger peaks are usually wider peaks in a XPS spectrum.
Aluminum foil is used as an example on the next slide.
37
XPS Spectrum
O 1s
O becauseof Mg source
CAl
Al
O 2s
O Auger
Sample and graphic provided by William Durrer, Ph.D.Department of Physics at the Univertsity of Texas at El Paso
38
Auger Spectrum
Characteristic of Auger graphsThe graph goes up as KE increases.
Sample and graphic provided by William Durrer, Ph.D.Department of Physics at the Univertsity of Texas at El Paso
39
Identification of XPS Peaks
The plot has characteristic peaks for each element found in the surface of the sample.
There are tables with the KE and BE already assigned to each element.
After the spectrum is plotted you can look for the designated value of the peak energy from the graph and find the element present on the surface.
40
X-rays vs. e- Beam
X-Rays Hit all sample area simultaneously
permitting data acquisition that will give an idea of the average composition of the whole surface.
Electron Beam It can be focused on a particular area
of the sample to determine the composition of selected areas of the sample surface.
41
XPS Technology Consider as non-
destructive because it produces soft
x-rays to induce photoelectron emission from the sample surface
Provide information about surface layers or thin film structures
Applications in the industry: Polymer surface Catalyst Corrosion Adhesion Semiconductors Dielectric materials Electronics packaging Magnetic media Thin film coatings
42
Applications of X-ray Photoelectron Spectroscopy (XPS)
43
XPS Analysis of Pigment from Mummy Artwork
150 145 140 135 130
Binding Energy (eV)Binding Energy (eV)
PbO2
Pb3O4
500 400 300 200 100 0Binding Energy (eV)Binding Energy (eV)
O
Pb Pb
Pb
N
Ca
C
Na
Cl
XPS analysis showed XPS analysis showed that the pigment used that the pigment used on the mummy on the mummy wrapping was Pbwrapping was Pb33OO44
rather than Ferather than Fe22OO33
Egyptian Mummy Egyptian Mummy 2nd Century AD2nd Century ADWorld Heritage MuseumWorld Heritage MuseumUniversity of IllinoisUniversity of Illinois
44
Analysis of Carbon Fiber- Polymer Composite Material by XPS
Woven carbon Woven carbon fiber compositefiber composite
XPS analysis identifies the functional XPS analysis identifies the functional groups present on composite surface. groups present on composite surface. Chemical nature of fiber-polymer Chemical nature of fiber-polymer interface will influence its properties.interface will influence its properties.
-C-C--C-C-
-C-O-C-O
-C=O-C=O
-300 -295 -290 -285 -280
B inding energy (eV )
N(E
)/E
45
Analysis of Materials for Solar Energy Collection by XPS Depth Profiling- The amorphous-SiC/SnO2 Interface
The profile indicates a reduction of the SnOThe profile indicates a reduction of the SnO22
occurred at the interface during deposition. occurred at the interface during deposition. Such a reduction would effect the collector’s Such a reduction would effect the collector’s efficiency.efficiency.
Photo-voltaic CollectorPhoto-voltaic Collector
Conductive Oxide- SnOConductive Oxide- SnO22
p-type a-SiCp-type a-SiC
a-Sia-Si
Solar EnergySolar Energy
SnOSnO22
SnSn
Depth500 496 492 488 484 480
Binding Energy, eV
Data courtesy A. Nurrudin and J. Abelson, University of Illinois46
Angle-resolved XPS
=15° = 90°
More Surface More Surface SensitiveSensitive
Less Surface Less Surface SensitiveSensitive
Information depth = dsinInformation depth = dsind = Escape depth ~ 3 d = Escape depth ~ 3 = Emission angle relative to surface= Emission angle relative to surface==Inelastic Mean Free PathInelastic Mean Free Path
47
Angle-resolved XPS Analysis of Self-Assembling Monolayers
Angle Resolved XPS Can Angle Resolved XPS Can DetermineDetermineOver-layer ThicknessOver-layer ThicknessOver-layer CoverageOver-layer Coverage
Data courtesy L. Ge, R. Haasch and A. Gewirth, University of IllinoisData courtesy L. Ge, R. Haasch and A. Gewirth, University of Illinois
0 2 0 4 0 6 0 8 0 1 0 00 . 1
0 . 2
0 . 3
0 . 4
0 . 5
0 . 6
C (W )
C (A u)
A u
S iW O12 40 d
E lectron E m iss ion A ng le , degrees
Expt. Data
M odel
48