electrode_transducers.pptx

Upload: prasanth-chitturi

Post on 14-Apr-2018

213 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/28/2019 Electrode_transducers.pptx

    1/141

    Electrode Skin Interface

    Apart from the electrode electrolyte interface, there isalso the skin interface.

    when an electrode is placed on the skin surface, thereis some electrical resistance at the electrode-skininterface.

    The skin consists of three layers, Epidermis, dermis

    and subcutaneous layer.

  • 7/28/2019 Electrode_transducers.pptx

    2/141

    Stratum corneum(we see dead cells)

    Stratum granulosum

    Stratum (basale) germinativum(where new skin cells form)

    Deep layers of skin consist ofvascular and nervous

    components, as well as sweat glands, sweat ducts and hairfollicles. With the exception of sweat glands, no particular characteristics

    affecting the electrode performance.

  • 7/28/2019 Electrode_transducers.pptx

    3/141

  • 7/28/2019 Electrode_transducers.pptx

    4/141

  • 7/28/2019 Electrode_transducers.pptx

    5/141

    Magnified section of skin, showing the various

    layers

  • 7/28/2019 Electrode_transducers.pptx

    6/141

    Epidermis

    We are most interested at epidermis, as that is themain contact with the electrode

    Since the skins natural resistance is high comparedto the resistance of the fluids, the selected skin site isto be well prepared by cleaning withAlcohol orAcetoneand by applying a commercially available

    conducting jelly (electrode paste).

    This ensures a low value of electrode-skin interfaceresistance.

  • 7/28/2019 Electrode_transducers.pptx

    7/141

    Surface electrodes usually have resistances of 2000 to10000 ohms depending on their size, whereas, smallneedle electrodes have a much higher resistance.

    A simple series equivalent circuit of an electrode-electrolyte interface is shown in figure.

  • 7/28/2019 Electrode_transducers.pptx

    8/141

    Here, Ehc is the half-cell potential; Rd and Cd represent the impedance associatedwith electrode-electrolyte interface; and RS is the total resistance in the circuit due toresistance in electrolyte and electrode lead wire.

    However, when an electrode makes contact with the skinvia an electrolyte paste, the equivalent circuit modified asshown.

    Now, Rs becomes the effective resistance of the paste between theelectrode and the skin.

    The epidermis of the skin may be considered as a semipermeable membrane and the potential difference across itis represented by Ese.

  • 7/28/2019 Electrode_transducers.pptx

    9/141

  • 7/28/2019 Electrode_transducers.pptx

    10/141

    The epidermic layer has also an electrical impedance,which is represented by the parallel circuit Ce,Re.

    The dermis and subcutaneous layer under it behave ingeneral as pure resistance Ru, as shown in figure.

    Thus, it can be seen that to obtain a more stable

    electrode, the effect of the epidermis (stratumcorneum) has to be reduced.

  • 7/28/2019 Electrode_transducers.pptx

    11/141

    This is achieved by many ways: by rubbing with a padsoaked in acetone or by puncturing the epidermis withdental burrs.

    All these methods tend to short out Ese, Ce and Re,thus improving the stability of the signal.

    Psychogenic electrodermal responses or the galvanicskin reflex (GSR), is the contribution of the sweatglands and sweat ducts.

  • 7/28/2019 Electrode_transducers.pptx

    12/141

    The fluid secreted by sweat glands contains Na+, K+,and Cl ions, the concentrations of which differ fromthose in the extracellular fluid.

    Thus there is a potential difference between the lumenof the sweat duct and the dermis and subcutaneous layers.

  • 7/28/2019 Electrode_transducers.pptx

    13/141

    There also is a parallel RpCp combination in serieswith this potential that represents the wall of thesweat gland and duct, as shown by the broken lines.

    These components are often neglected when weconsider biopotential electrodes unless the electrodes

    are used to measure the electrodermal response orGSR.

  • 7/28/2019 Electrode_transducers.pptx

    14/141

  • 7/28/2019 Electrode_transducers.pptx

    15/141

    Motion artifact

    If a pair of electrodes is in an electrolyte and onemoves while the other remains stationary, a potentialdifference appears between the two electrodes duringthis movement.

    This potential is known as motion artifact and can be

    a serious cause of interference in the measurement ofbiopotentials.

  • 7/28/2019 Electrode_transducers.pptx

    16/141

    Because motion artifact results primarily frommechanical disturbances of the distribution of chargeat the electrodeelectrolyte interface, it is reasonableto expect that motion artifact is minimal fornonpolarizable electrodes.

    This artifact can be significantly reduced when thestratum corneum is removed bymechanical abrasion

    with a fine abrasive paper.

  • 7/28/2019 Electrode_transducers.pptx

    17/141

  • 7/28/2019 Electrode_transducers.pptx

    18/141

    Remembering the dynamic nature of the epidermis,note also that the stratum corneum can regenerateitself in as short a time as 24 h, thereby renewing thesource of motion artifact.

  • 7/28/2019 Electrode_transducers.pptx

    19/141

    Stretching the skin changes this skin potential by 5 to10 mV, and this change appears as motion artifact.

    Ten 0.5 mm skin punctures through the barrier layershort-circuits the skin potential and reduces thestretch artifact to less than 0.2 mV.

  • 7/28/2019 Electrode_transducers.pptx

    20/141

  • 7/28/2019 Electrode_transducers.pptx

    21/141

    Invasive and Non Invasive

    Sensors that are used to measure electrical, chemical,physical activities from human body.

  • 7/28/2019 Electrode_transducers.pptx

    22/141

    Invasive

    Ionization (radiation) X-ray, UV, -ray

    Contact with bloodIntrusion into the body

    Minimally invasive

    Contact with blood

    Intrusion into the body

    Non-invasive

    Surface or remote diagnosis / therapy

  • 7/28/2019 Electrode_transducers.pptx

    23/141

    Types of electrodes

    A wide variety of electrodes can be used to measurebioelectric events, but nearly all can be classified asbelonging to one of three basic types:

    1) Surface Electrodes: Used to measure ECG,EEG andEMG potentials on the surface of the skin.

    2) Needle Electrodes: Used to penetrate the skin to

    record EEG potentials from a local region of the brain,or EMG potentials from a specific group of muscles.

    3) Micro Electrodes: Used to measure bioelectricpotentials near a single cell.

  • 7/28/2019 Electrode_transducers.pptx

    24/141

    METAL-PLATE ELECTRODES: SURFACE

    In its simplest form, it consists of a metallic conductorin contact with the skin.

    An electrolyte soaked pad or gel is used to establishand maintain the contact.

    Metal-plate electrode used for application to limbs isshown.

    It consists of a flat metal plate that has been bent into

    a cylindrical segment. A terminal is placed on its outside surface near one

    end; this terminal is used to attach the lead wire to theelectrocardiograph.

  • 7/28/2019 Electrode_transducers.pptx

    25/141

    The electrode is traditionally made ofGerman silver (anickelsilver alloy).

    Before it is attached to the body with a rubber strap ortape, its concave surface is covered with electrolyte gel.

    A limb electrode

  • 7/28/2019 Electrode_transducers.pptx

    26/141

    Metal-disk electrode

    The structure shown can be used as a chest electrodefor recording the ECG or in cardiac monitoring forlong-term recordings.

    In these applications the electrode is often fabricatedfrom a disk ofAg that may have an electrolyticallydeposited layer ofAgCl on its contacting surface. It iscoated with electrolyte gel and then pressed against

    the patients chest wall.

    Metal-disk electrode

  • 7/28/2019 Electrode_transducers.pptx

    27/141

    In recording EMGs, investigators use stainless steel,platinum, or gold-plated disks to minimize the chancethat the electrode will enter into chemical reactionswith perspiration or the gel.

    Electrodes used in monitoring EMGs or EEGs aregenerallysmaller in diameter than those used inrecording ECGs.

  • 7/28/2019 Electrode_transducers.pptx

    28/141

    Disk-shaped electrodes such as these have also beenfabricated from metal foils (primarily silver foil) and

    are applied as single-use disposable electrodes.

    The thin, f lexible foil allows the electrode to conformto the shape of the body surface.

    In choosing suitable cardiac electrodes for patient-monitoring applications, physicians are more andmore turning to pregelled, disposable electrodes with

    the adhesive alreadyin place.

    These devices are ready to be applied to the patientand are disposed after use.

  • 7/28/2019 Electrode_transducers.pptx

    29/141

    Disposable foam-pad electrodes

    It consists of a relativelylarge disk of plastic foam materialwith a silver plated disk on one side attached to a silver-plated snap similar to that used on clothing in the center ofthe other side.

    The silver-plated disk serves as the electrode and may becoated with an AgCl layer.

    A layer of electrolyte gel covers the disk.

    The electrode side of the foam is covered with an adhesivematerial that is compatible with the skin.

  • 7/28/2019 Electrode_transducers.pptx

    30/141

    A protective cover or strip of release paper is placedover this side of the electrode and foam, and thecomplete electrode is packaged in a foil envelope sothat the water component of the gel will not evaporateaway.

  • 7/28/2019 Electrode_transducers.pptx

    31/141

    SUCTION ELECTRODES

    Amodification of the metal-plate electrode thatrequires no straps or adhesives for holding it in place isthe suction electrode.

    Such electrodes are frequently used inelectrocardiography as the precordial (chest) leads,because they can be placed at particular locations andused to take a recording.

  • 7/28/2019 Electrode_transducers.pptx

    32/141

    They consist of a hollow metallic cylindrical electrodethat makes contact with the skin at its base.

    An appropriate terminal for the lead wire is attached

    to the metal cylinder, and a rubber suction bulb fitsover its other base.

    Electrolyte gel is placed over the contacting surface ofthe electrode, the bulb is squeezed, and the electrodeis then placed on the chest wall.

  • 7/28/2019 Electrode_transducers.pptx

    33/141

    Used as a precordial electrode

  • 7/28/2019 Electrode_transducers.pptx

    34/141

    The bulb is released and applies suction against theskin, holding the electrode assembly in place.

    This electrode can be used only for short periods oftime; the suction and the pressure of the contactsurface against the skin can cause irritation.

  • 7/28/2019 Electrode_transducers.pptx

    35/141

    Although the electrode itself is quite large, that theactual contacting area is relatively small.

    This electrode thus tends to have a higher sourceimpedance than the relatively large-surface-area metalplate electrodes used for ECG limb electrodes.

  • 7/28/2019 Electrode_transducers.pptx

    36/141

    FLOATING ELECTRODES

    We noted that one source ofmotion artifact inbiopotential electrodes is the disturbance of thedouble layer of charge at the electrodeelectrolyte

    interface. The use of nonpolarizable electrodes, such as the

    Ag/AgCl electrode, can greatly diminish this artifact.

    But it still can be present, and efforts to stabilize the

    interface mechanically can reduce it further. Floating electrodes offer a suitable technique to do so.

  • 7/28/2019 Electrode_transducers.pptx

    37/141

    Figure (a) depicts a floating electrode known as a top-hat electrode; its internal structure is illustrated incross section in Figure(b).

    The principal feature of the electrode is that the actualelectrode element or metal diskis recessed in a cavityso that it does not come in contactwith the skin itself.

    Instead, the element is surrounded by electrolyte gel inthe cavity.

  • 7/28/2019 Electrode_transducers.pptx

    38/141

  • 7/28/2019 Electrode_transducers.pptx

    39/141

    The cavity and hence the gel does not movewithrespect to the metal disk, so it does not produce anymechanical disturbance of the double layer of charge.

    The electrode element can be a disk made of a metalsuch as silver coatedwithAgCl or sinteredAg/AgCl

    pellet instead of a metal disk. These electrodes are found to be quite stable and are reusable afterappropriate cleaning between uses.

  • 7/28/2019 Electrode_transducers.pptx

    40/141

    A single-use, disposable modification of the floating

    electrode is shown in cross section in Figure (c).

    It has one added componenta disk of thin, open-cellfoam saturated with electrolyte gel.

    The other surface of the foam that is placed against theskin is able to move with the skin, therebydiminishing

    the motion artifact that sometimes results fromdifferential movement between the skin and theelectrolyte gel.

  • 7/28/2019 Electrode_transducers.pptx

    41/141

  • 7/28/2019 Electrode_transducers.pptx

    42/141

  • 7/28/2019 Electrode_transducers.pptx

    43/141

    The below figure shows one technique employed to provide

    flexible electrodes.

    Acarbon-filled silicone rubber compound in the form of athin strip or disk is used as the active element of an

    electrode.

    The carbon particles in the silicone make it an electricconductor.

    Apin connector is pushed into the lead connector hole,and the electrode is used in the same way as a similar typeof metal-plate electrode.

  • 7/28/2019 Electrode_transducers.pptx

    44/141

    Carbon-filled silicone rubber electrode

  • 7/28/2019 Electrode_transducers.pptx

    45/141

    Flexible thin-film electrode

    The basic electrode consists of a 13 mm-thick Mylarfilm on which an Ag and AgCl film have beendeposited, as shown in Figure.

  • 7/28/2019 Electrode_transducers.pptx

    46/141

    The flexible lead wire is attached to the Mylarsubstrate by means of a conducting adhesive, and asilver film approximately 1 micrometer thick is deposited

    over this and the Mylar.

    An AgCl layer is then grown on the surface of the silver

    film via the electrolytic process.

  • 7/28/2019 Electrode_transducers.pptx

    47/141

    Advantage

    In addition to the advantage of being flexible andconforming to the shape of the newborns chest, theseelectrodes have a layer ofsilver thin enough to be

    essentiallyx-ray transparent, so theyneed not beremoved when chest x rays of the infant are taken.

    Consequently, the infants skin is also protected fromthe irritation caused by removing and reapplying theadhesive tape that holds the electrode in place.

  • 7/28/2019 Electrode_transducers.pptx

    48/141

    INTERNAL ELECTRODES:NEEDLE

    Electrodes can also be usedwithin the bodyto detectbio potentials.

    They can take the form ofpercutaneous electrodes, inwhich the electrode itself or the lead wire crosses theskin, or they may be entirelyinternal electrodes, in

    which the connection is to an implanted electroniccircuit such as a radio telemetry transmitter.

  • 7/28/2019 Electrode_transducers.pptx

    49/141

    No electrolyte gel is required to maintain this

    interface, because extracellular fluid is present.

    The basic needle electrode consists of a solid needle,usually made ofstainless steel, with a sharp point.

    The shank of the needle is insulated with a coatingsuch as an insulating varnish; only the tip is left

    exposed. A lead wire is attached to the other end of the needle,

    and the joint is encapsulated in a plastic hub to protectit.

  • 7/28/2019 Electrode_transducers.pptx

    50/141

    This electrode, frequently used in electromyography, is

    shown in Figure(a).When it is placed in a particular muscle, it obtains an

    EMG from that muscle acutely and can then beremoved.

    A shielded percutaneous electrode can be fabricated inthe form shown in Figure(b).

    It consists of a small-gage hypodermic needle that hasbeen modified by running an insulated fine wire downthe center of its lumen and filling the remainder of thelumen with an insulating material such as an epoxyresin.

  • 7/28/2019 Electrode_transducers.pptx

    51/141

  • 7/28/2019 Electrode_transducers.pptx

    52/141

    The needle itself is connected to ground through theshield of a coaxial cable, thereby extending the coaxialstructure to its very tip.

    Multiple electrodes in a single needle can be formed asshown in Figure(c).

    Here two wires are placed within the lumen of theneedle and can be connected differentiallyso as to besensitive to electrical activity only in the immediatevicinity of the electrode tip.

  • 7/28/2019 Electrode_transducers.pptx

    53/141

  • 7/28/2019 Electrode_transducers.pptx

    54/141

    Percutaneous wire electrodes

    The needle electrodesjust described are principally foracute measurements, because their stiffness and sizemake them uncomfortable for long term implantation.

    When chronic recordings are required, percutaneouswire electrodes are more suitable.

  • 7/28/2019 Electrode_transducers.pptx

    55/141

    A fine wireoften made ofstainless steel ranging indiameter from 25 to 125 micromis insulated with aninsulating varnish to within a few millimeters of thetip.

    This noninsulated tip is bent back on itself to form aJ-shaped structure.

    The tip is introduced into the lumen of the needle, asshown in Figure(d).

  • 7/28/2019 Electrode_transducers.pptx

    56/141

  • 7/28/2019 Electrode_transducers.pptx

    57/141

    The needle is inserted through the skin into themuscle at the desired location, to the desired depth.

    It is then slowly withdrawn, leaving the electrode inplace, as shown in Figure(e).

    Note that the bent-over portion of wire serves as a barb holding thewire in place in the muscle.

    To remove the wire, the technician applies a mild uniform force tostraighten out the barb and pulls it out through the wires track.

  • 7/28/2019 Electrode_transducers.pptx

    58/141

    Realizing that wire electrodes chronically implanted inactive muscles undergo a great amount of f lexing as

    the muscle moves (which can cause the wire to slip as it passesthrough the skin and increase the irritation and risk ofinfection at this

    point, or even cause the wire to break), they developed thehelical electrode and lead wire shown in Figure(f).

  • 7/28/2019 Electrode_transducers.pptx

    59/141

  • 7/28/2019 Electrode_transducers.pptx

    60/141

    It, too, is made from a very fine insulated wire coiledinto a tight helix of approximately150 microm

    diameter that is placed in the lumen of the insertingneedle.

    The uninsulated barb protrudes from the tip of theneedle and is bent back along the needle before

    insertion. It holds the wire in place in the tissue when the needle

    is removed from the muscle.

    Of course, the external end of the electrode now passes through theneedle and the needle must be removedor at least protectedbeforethe electrode is connected to the recording apparatus.

  • 7/28/2019 Electrode_transducers.pptx

    61/141

    Intracutaneous needles

    Another group of percutaneous electrodes are thoseused for monitoring fetal heartbeats.

    (a) Suction electrode. (b) Cross-sectional viewof suction electrode in place, showing penetration ofprobe through epidermis.(c) Helical electrode that is attached to fetal skin by

    corkscrew-type action.

  • 7/28/2019 Electrode_transducers.pptx

    62/141

    Implantable electrodes

    Often when implantable wireless transmission is used,we want to implant electrodes within the bodyand notpenetrate the skin with any wires.

    In this case the radio transmitter is implanted in thebody.

    The simplest electrode for this application is shown inFigure(a).

  • 7/28/2019 Electrode_transducers.pptx

    63/141

    Wire-loop electrode

  • 7/28/2019 Electrode_transducers.pptx

    64/141

    Insulated multistranded stainless steel or platinumwire suitable for implantation has one end stripped sothat an eyelet can be formed from the strands of wire.

    The eyelet can be sutured to the point in the body atwhich electric contact is to be established.

    Platinum -sphere cortical-surface potential

  • 7/28/2019 Electrode_transducers.pptx

    65/141

    Platinum sphere cortical surface potential

    electrode

    Figure(b) shows another example of an implantableelectrode for obtaining cortical-surface potentials fromthe brain applied this electrode for the radiotelemetry

    of subdural EEGs. .

  • 7/28/2019 Electrode_transducers.pptx

    66/141

    The electrode consists of a 2 mm-diameter metallicsphere located at the tip of the cylindrical Tefloninsulator through which the electrode lead wirepasses.

    The calvarium is exposed through an incision in thescalp, and a burr hole is drilled.

    A small slit is made in the exposed dura, and the silversphere is introduced through this opening so that itrests on the surface of the cerebral cortex.

    The assembly is then cemented in place onto thecalvarium by means of a dental acrylic material.

  • 7/28/2019 Electrode_transducers.pptx

    67/141

    Multielement depth electrode

    Deep cortical potentials can be recorded from multiplepoints using the technique as shown in Figure(c).

  • 7/28/2019 Electrode_transducers.pptx

    68/141

    This kind of electrode consists of a cluster of fineinsulated wires held together by avarnish binder.

    Each wire has been cut transverselyto expose anuninsulated cross section that serves as the activeelectrode surface.

    By staggering the ends of the wires as shown, we canproduce electrodes located at known differences indepth in an array.

    The other ends of the electrodes can be attached to appropriateimplantable electronic devices or to a connector cemented on the skullto allow connection to an external recording apparatus.

  • 7/28/2019 Electrode_transducers.pptx

    69/141

    ELECTRODE ARRAYS

    Although implantable electrode arrays can befabricated one at a time using clusters of fine insulatedwires, this technique is both time-consuming andexpensive.

    Furthermore, when such clusters are madeindividually, each one will be somewhat different fromthe other.

    A way to minimize these problems is to utilize microfabrication technology to fabricate identical two- andthree dimensional electrode arrays.

  • 7/28/2019 Electrode_transducers.pptx

    70/141

    One-dimensional plunge electrode array

  • 7/28/2019 Electrode_transducers.pptx

    71/141

    Two -dimensional array

  • 7/28/2019 Electrode_transducers.pptx

    72/141

    Three -dimensional array

  • 7/28/2019 Electrode_transducers.pptx

    73/141

    MICROELECTRODES

    To be able to measure potential differences across thecell membrane we must have an electrode within thecell.

    Such electrodes must be small with respect to the celldimensions to avoid causing serious cellular injury andtherebychanging the cells behavior.

  • 7/28/2019 Electrode_transducers.pptx

    74/141

    In addition to being small, the electrode used formeasuring intracellular potential must also be strongso that it canpenetrate the cell membrane and remain

    mechanically stable.

  • 7/28/2019 Electrode_transducers.pptx

    75/141

    Electrodes that meet these requirements are known as

    microelectrodes.

    They have tip diameters ranging from approximately

    0.05 to 10 mirom.

    Microelectrodes can be formed from solid-metalneedles, from metal contained within or on the surface

    of a glass needle, or from a glass micropipette having alumen filled with an electrolytic solution.

  • 7/28/2019 Electrode_transducers.pptx

    76/141

    METAL MICROELECTRODES

    A fine needle of a strong metal that is insulatedwithan appropriate insulator up to its tip.

    The structure of a metal microelectrode for intracellular recordings.

  • 7/28/2019 Electrode_transducers.pptx

    77/141

    The metal needle is prepared byelectrolytic etching,using an electrochemical cell in which the metalneedle is the anode.

    The electric current etches the needle as it is slowlywithdrawn from the electrolyte solution.

    Veryfine tips can be formed in this way, but a greatdeal ofpatience and practice are required to gain theskill to make them.

  • 7/28/2019 Electrode_transducers.pptx

    78/141

    Suitable strong metals for these microelectrodes are

    stainless steel, platinumiridium alloy, and tungsten. The compound tungsten carbide is also used because

    of its great strength.

    The microelectrode and supporting shaft are usuallyinsulated by a film of some polymeric material orvarnish.

    Only the extreme tip of the electrode remains uninsulated.

  • 7/28/2019 Electrode_transducers.pptx

    79/141

    SUPPORTED-METAL MICROELECTRODES

    The properties oftwo different materials are used toadvantage in supported metal microelectrodes.

    A strong insulating material that can be drawn to afine point makes up the basic support, and a metalwith good electrical conductivityconstitutes thecontacting portion of the electrode.

  • 7/28/2019 Electrode_transducers.pptx

    80/141

    Metal filled glass micropipette

    The classic example of this form is a glass tube drawnto a micropipette structure with its lumen filled withan appropriate metal.

  • 7/28/2019 Electrode_transducers.pptx

    81/141

    Preparation

    This is prepared by first filling a glass tube with ametal that has a melting point near the softening point of theglass.

    The tube can then beheated to the softening point

    and pulledto form a narrow constriction.

    When it is broken at the constriction, two micropipettes filled with metal are formed.

    In this type of structure, the glass not only provides the mechanicalsupport but also serves as the insulation.

    The active tip is the onlymetallic area exposed in crosssection where the pipette was broken away.

  • 7/28/2019 Electrode_transducers.pptx

    82/141

    (a) Section of fine-bore glass capillary.(b) Capillary narrowed through heating and stretching.(c) Final structure of glass-pipette microelectrode.

  • 7/28/2019 Electrode_transducers.pptx

    83/141

    .

    Glass micropipette, coated with metal film

    The figure shows the cross section of the tip of adeposited-metal-film microelectrode.

    A solid glass rod or glass tube is drawn to form the

    micropipette.

    A metal film is deposited uniformly on this surface to

    a thickness of the order of tenths of a micrometer.A polymeric insulation is then coated over this, leaving

    just the tip, with the metal film exposed.

  • 7/28/2019 Electrode_transducers.pptx

    84/141

    Glass micropipette, coated with metal film

  • 7/28/2019 Electrode_transducers.pptx

    85/141

    MICROPIPETTE ELECTRODES

    A glass micropipette electrode filled with an electrolyticsolution.

    Electrolyte solution that is frequently3M KCl.

    A cap containing a metal electrode is then sealed to the pipette.

    The metal electrode contacts the electrolyte within the pipette.

    The electrode is frequently a silver wire prepared with anelectrolytic AgCl surface.

    Platinum or stainless steelwires are also occasionally used.

    MICROELECTRODES BASED ON MICROELECTRONIC

  • 7/28/2019 Electrode_transducers.pptx

    86/141

    MICROELECTRODES BASED ON MICROELECTRONIC

    TECHNOLOGY

    The basic structure consists of narrow gold stripsdeposited on a silicon substrate the surface of whichhas been first insulated by growing an SiO2 film.

    The gold strips are then further insulated bydepositing SiO2 over their surface.

    The silicon substrate is next etched to a thin, narrow

    structure that is just wide enough to accommodate thegold strips in the region of the tip.

  • 7/28/2019 Electrode_transducers.pptx

    87/141

    The silicon substrate is etched a millimeter or two backfrom the tip so that only the gold strips and their SiO2

    insulation remain.

    The insulation is etched awayfrom the verytip of thegold strips to expose the contacting surface of theelectrodes.

  • 7/28/2019 Electrode_transducers.pptx

    88/141

    Beam-lead multiple electrode

  • 7/28/2019 Electrode_transducers.pptx

    89/141

    Multielectrode silicon probe

  • 7/28/2019 Electrode_transducers.pptx

    90/141

    Multiple-chamber electrode

  • 7/28/2019 Electrode_transducers.pptx

    91/141

    Peripheral-nerve electrode

  • 7/28/2019 Electrode_transducers.pptx

    92/141

    ELECTRICAL PROPERTIES OF MICROELECTRODES

    We must derive an electrical equivalent circuit fromphysical considerations.

    The microelectrode contributes a series resistance Rsthat is due to the resistance of the metal itself.

    A major contributor to this resistance is the metal inthe shank and tip portion of the microelectrode,because the ratio of length to cross-sectional area ismuch higher in this portion than it is for the shaft.

  • 7/28/2019 Electrode_transducers.pptx

    93/141

    The metal is coated with an insulating material over allbut its most distal tip, so a capacitance is set upbetween the metal and the extracellular fluid.

    This is a distributed capacitance Cd that we canrepresent in lumped form by separating the shank andtip from the shaft.

    In the shank region, we can consider the microelectrode to be a coaxialcylinder capacitor; the capacitance per unit length (F/m) is given by

    Electrode with tip placed within a cell, showing origin

  • 7/28/2019 Electrode_transducers.pptx

    94/141

    p p , g g

    of distributed capacitance

  • 7/28/2019 Electrode_transducers.pptx

    95/141

  • 7/28/2019 Electrode_transducers.pptx

    96/141

    Here the ratio of diameters would be practically unity, so we cansimplify the calculation by unwrapping the circumferential surface ofthe shaft and considering the system to be a parallel-plate capacitor ofarea equal to the circumferential surface area and of thickness equal to

    t, the thickness of the insulation layer.

    The capacitance per unit length (F/m) is given by

    Note that this capacitance comes from only that portion of the electrode shaftthat is submerged in the extracellular fluid. Often only the shank issubmerged, so Cd2 is zero.

  • 7/28/2019 Electrode_transducers.pptx

    97/141

    Equivalent circuit

  • 7/28/2019 Electrode_transducers.pptx

    98/141

  • 7/28/2019 Electrode_transducers.pptx

    99/141

    The tip of the microelectrode is within a cell, so there is a seriesresistance Ri, associated with the electrolytewithin the cell membraneand another series resistance Re due to the extracellular fluid.

    The cell membrane itself can be modeled simply as a variable potentialEmp, but in more detailed analyses an equivalent circuit of greatercomplexity is required.

    Some of the distributed capacitance of the shank,Cd1, is between themicroelectrode and the extracellular fluid, as shown in the equivalentcircuit, whereas the remainder of it is between the microelectrode andthe intracellular fluid.

  • 7/28/2019 Electrode_transducers.pptx

    100/141

    Simplified equivalent circuit

  • 7/28/2019 Electrode_transducers.pptx

    101/141

    The above circuit neglects the impedance of the reference electrodeand the series-resistance contribution from the intracellular andextracellular fluid and lumps all the distributed capacitance together.

    Under circumstances in which the input impedance of the amplifier

    connected to this electrode is not sufficiently large, we see that thiscircuit can behave as a high-pass filter and significant waveformdistortion can result.

    The effective impedance of metal microelectrodes is frequency

    dependent and can be of the order of10 to 100 M.

  • 7/28/2019 Electrode_transducers.pptx

    102/141

    We can, however, lower this impedance byincreasingthe effective surface area of the tip of themicroelectrode through the application ofplatinum

    black.

    At lower frequencies, the impedance can be reducedby applying anAg/AgCl surface to the electrode tip.

    Gl i i tt i l t d

  • 7/28/2019 Electrode_transducers.pptx

    103/141

    Glass micropipette microelectrode

    The internal electrode in the micropipette gives themetal electrolyte interface components Rma, Cma,and Ema.

    In series with this is a resistive element Rt

    corresponding to the resistance of the electrolyte inthe shank and tip region of the microelectrode.

    Connected to this is the distributed capacitance Cdcorresponding to the capacitance across the glass inthis region.

    The distributed capacitance due to the shaft region has been neglected,because the glass wall of the electrode is much thicker in this region andthe capacitive contribution is quite small.

    Electrode with its tip placed within a cell, showing the origin of distributed

    capacitance

  • 7/28/2019 Electrode_transducers.pptx

    104/141

    capacitance

  • 7/28/2019 Electrode_transducers.pptx

    105/141

    There are two potentials associated with the tip of the micropipettemicroelectrode.

    The liquid-junction potential Ej corresponds to the

    liquid junction set up between the electrolyte in themicropipette andthe intracellular fluid.

    In addition, a potential known as the tip potential Et

    arises because the thin glass wall surrounding the tipregion of the micropipette behaves like a glassmembrane and has an associated membrane potential.

  • 7/28/2019 Electrode_transducers.pptx

    106/141

    The equivalent circuit also includes resistancescorresponding to the intracellular Ri and extracellularRe fluids.

    These are coupled to the microelectrode through thedistributive capacitance Cd, as is the case for the metalmicroelectrode.

    The equivalent circuit for the reference electroderemains unchanged from that shown for the previouselectrode.

    E i l i i

  • 7/28/2019 Electrode_transducers.pptx

    107/141

    Equivalent circuit

  • 7/28/2019 Electrode_transducers.pptx

    108/141

    Unlike the metal microelectrode, the micropipettesmajor impedance contribution is resistive.

    This can be illustrated by approximating the

    equivalent circuit.

  • 7/28/2019 Electrode_transducers.pptx

    109/141

    Here the overall series resistance of the electrode is

    lumped together as Rt.

    This resistance generally ranges in value from 1 to 100M.

    The total distributed capacitance is lumped togetherto form Ct, which can be on the order of tens of picofarads.

    All the associated dc potentials are lumped together inthe source Em, which is given by

  • 7/28/2019 Electrode_transducers.pptx

    110/141

    Note that the micropipette-type microelectrode behaves asa low-pass filter.

    The high series resistance and distributed capacitancecause the electrode output to respond slowly to rapidchanges in cell-membrane potential.

    To reducethis problem, positive-feedback, negative-capacitance amplifiers are used to reduce the effectivevalue of Ct.

    TRANSDUCER

  • 7/28/2019 Electrode_transducers.pptx

    111/141

    TRANSDUCER

    Generally, transducer is required to convertphysiological variables into electrical signals which areeasier to be processed.

    The relationship between input and output variablecan be linear, logarithmic or square.

    The transducer can be active or passive dependingupon conversion of non electrical variable intoelectrical signal.

  • 7/28/2019 Electrode_transducers.pptx

    112/141

    The active transducer directlyconverts input variableinto electrical signals while passive transducermodifies either excitation voltages or modulates the

    carrier signals.

    The passive transducers are externally poweredwhileactive transducers are self generating and require no

    external power.

    ACTIVE TRANSDUCERS

  • 7/28/2019 Electrode_transducers.pptx

    113/141

    PASSIVE TRANSDUCERS

  • 7/28/2019 Electrode_transducers.pptx

    114/141

    PASSIVE TRANSDUCERS

    There are only three passive circuit elements that canbe used to change voltage at the output of the circuitaccording to the physical variable :

    (1) resistors

    (2) capacitors and

    (3) inductors.

    The passive transducer is part of a circuit normally anarrangement similar to awheatstone bridgewhich ispowered by an ac or dc excitation.

  • 7/28/2019 Electrode_transducers.pptx

    115/141

    INDUCTIVE PASSIVE TRANSDUCERS

    Variable induction :

    The property of inductance is varied in the circuit tochange the output voltage in accordance with the

    input variable.

    The inductance L = n2G

    (n = number of turns in coil, G = form factor of coil and = permeability of core material inside the coil).

    Induction Displacement Transducer

  • 7/28/2019 Electrode_transducers.pptx

    116/141

    Induction Displacement Transducer

  • 7/28/2019 Electrode_transducers.pptx

    117/141

    Variable reluctance :

    In this, core remains stationaryinside the coil but the

    air gap in the magnetic path of the core is varied tochange the net permeability, thereby varying theoutput signal as per the input variable (displacement).

    i bl l d

  • 7/28/2019 Electrode_transducers.pptx

    118/141

    Variable Reluctance Transducer

    Linear Variable differential transformer

  • 7/28/2019 Electrode_transducers.pptx

    119/141

    Linear Variable differential transformer

    Applications

  • 7/28/2019 Electrode_transducers.pptx

    120/141

    Applications

    1. LVDT is used as a catheter-tip blood pressuretransducer.

    In this service, the core of the LVDT is affixed to asmall, circular, elastic diaphragm exposed to bloodpressure.

    The mass of the core and the diaphragm are very smalland the system has a high stiffness.

  • 7/28/2019 Electrode_transducers.pptx

    121/141

    2. The movement of a bourdon tube and, thereby, thepressure inside the tube can be measured byconnecting the core of the LVDT to the tip of the

    bourdon tube.

    3. LVDT is used in Ballistocardiography for picking upthe movement of the ballistocardiograph platform (on

    which a subject lies supine) due to the pumping actionof the heart.

  • 7/28/2019 Electrode_transducers.pptx

    122/141

    4.Vibrations of several body segments can bemonitored by using LVDT to study the effects ofvibrations on the human body placed on a vibration

    platform.

    PASSIVE CAPACITANCE TRANSDUCERS

  • 7/28/2019 Electrode_transducers.pptx

    123/141

    PASSIVE CAPACITANCE TRANSDUCERS

    Variable capacitance :

    The capacitance (C) of a capacitor having two parallel

    plates of area A which are separated by a distance d is:

    C = o r A/d

    (0 = dielectric constant of free space and r = relative

    dielectric constant).

    Variation of Capacitance with Displacement

  • 7/28/2019 Electrode_transducers.pptx

    124/141

    Variation of Capacitance with Displacement

    Variation of Capacitance with Displacement

  • 7/28/2019 Electrode_transducers.pptx

    125/141

    Variation of Capacitance with Displacement

    Variation of Capacitance with Angular Displacement

  • 7/28/2019 Electrode_transducers.pptx

    126/141

    Variation of Capacitance with Angular Displacement

    Applications

  • 7/28/2019 Electrode_transducers.pptx

    127/141

    Applications

    1. The capacitance method has been applied to themeasurement of physiological events, particularlyblood pressure.

    An elastic element exposed to blood pressureconstitutes one plate of the capacitor, the other plate isnearby and fixed.

    To obtain a rapid response time, the elastic element ismade as small and as stiff as possible.

  • 7/28/2019 Electrode_transducers.pptx

    128/141

    2. The capacitance microphone is used for thedetection ofheart sounds.

    3. A miniature capacitance microphone is used as a

    high fidelitypulse pick-up. 4.An unusual application is that the dielectric property

    of the living tissues itself as part of the capacitor.

    This principle is used by placing electrodes on thechest and back and correlating the output of this, torecordvolume and cardiac output for each heart beat.

    TEMPERATURE MEASUREMENT

  • 7/28/2019 Electrode_transducers.pptx

    129/141

    TEMPERATURE MEASUREMENT

    It has been seen that a person in shock has reducedblood pressure in circulating system which results intolow body temperature.

    Infection and illness are usually reflected by a highbody temperature.

    Special heated incubators are used for maintaining thebody temperature of infants.

    The temperature of the joint of an arthritic patient isclosely linked with the amount of local inflammation.

  • 7/28/2019 Electrode_transducers.pptx

    130/141

    The temperature can be measured by

    (1) thermocouples

    (2) thermistor and

    (3) radiation and fiber optic detectors

  • 7/28/2019 Electrode_transducers.pptx

    131/141

    Resistance Temperature Detector (RTD)

  • 7/28/2019 Electrode_transducers.pptx

    132/141

    Electrical resistance of an electrical conductor is afunction of temperature

  • 7/28/2019 Electrode_transducers.pptx

    133/141

  • 7/28/2019 Electrode_transducers.pptx

    134/141

    Thin Film Gold Temperature Sensor

  • 7/28/2019 Electrode_transducers.pptx

    135/141

    Thin-Film Gold Temperature Sensor

    Thermistors

  • 7/28/2019 Electrode_transducers.pptx

    136/141

    The resistance of thermistor can be given as

    Thermistors can be formed into disks, beads, rods orany desired shapes.

    Thermistor probes are available with resistance from afew hundred ohms to several megohms.

    Most thermistor thermometers use the principle ofwheatstone bridge to obtain a voltage output whichvaries as per input temperature.

  • 7/28/2019 Electrode_transducers.pptx

    137/141

  • 7/28/2019 Electrode_transducers.pptx

    138/141

  • 7/28/2019 Electrode_transducers.pptx

    139/141

  • 7/28/2019 Electrode_transducers.pptx

    140/141

    The thermograph is an infrared thermometer

    incorporated into a scanner which can be used to scanentire surface of body or some part of body like atelevision camera.

    The infrared energy detected in scanning is used tomodulate the intensity of a light beam so that to getthe image on the photographic film in which thebrightness depends on the detected infrared radiation.

    The image is called a thermogram.

    Applications

  • 7/28/2019 Electrode_transducers.pptx

    141/141

    pp

    1. They are used for continuous measurement ofskinand body core temperatures.

    2. For the measurement ofblood f low.A heatedthermistor is mounted on the tip of a catheter or

    hypodermic needle which is inserted in to the bloodvessel. (change in R)

    3. The use of a thermistor for the respiration ratemeasurement is of special interest.

    Sufficient amount of current passed through thethermistor to raise its quiescent temperature to approx