electrical impedance tomography: image algorithms and applications

36
A.Adler, Oct 28, 2002 Electrical Impedance Tomography 1 Electrical Impedance Tomography: Image Algorithms and Applications Andy Adler Assistant Professor, SITE, U.Ottawa

Upload: gkcalat

Post on 04-Dec-2015

226 views

Category:

Documents


10 download

DESCRIPTION

Electrical ImpedanceTomography:Image Algorithms andApplications

TRANSCRIPT

Page 1: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 1

Electrical Impedance Tomography:

Image Algorithms and Applications

Andy Adler

Assistant Professor, SITE, U.Ottawa

Page 2: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 2

Outline

• Electrical Impedance Tomography• Physics• Image Reconstruction• Applications• Future Work

Page 3: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 3

Electrical Impedance Tomography

• Relatively new medical imaging technique (early 1990’s)

• Body Surface Electrodes apply current patterns and measure the resulting voltages

• Distribution of conductivity is calculated

Page 4: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 4

EIT: Block Diagram

+

+

CurrentSource

Amplifiers

Controler

Dataacquisition

Medium

+

Imaging SystemΩ

Page 5: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 5

EIT: Applications

• EIT can image physiological processes involving movement of conductive fluids and gasses

• Lungs• Heart / perfusion• GI tract• Brain• Breast

Page 6: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 6

EIT: Advantages

EIT is a relatively low resolution imaging modality, but

• Non-invasive• Non-cumbersome• Suitable for monitoring• Underlying technology is low cost

Page 7: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 7

Application: Breathing

Chest images of tidal breathing in normal

Page 8: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 8

Application: Heart Beat

EIT signal in ROI around heart and ECG

Page 9: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 9

Non-invasive

0.1

1

10

100

30 100 300 1K 3K 10K 30K

Frequency (Hz)

Current(mA)

Percentile 99Percentile 50Percentile 1EIT system

Thresholds for cutaneous perception of electric current vs. frequency and EIT system

Page 10: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 10

EIT: Physics

• Within medium there is E and J.

dtd

od

c

EJ

EJ

εε

σ

=

=

( )EJ ojωεεσ −=

Page 11: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 11

EIT: Physics

In the absence of magnetic fields

No charge build up in conductive medium

We have

V−∇=E

0J =−=•∇t∂

∂ρ

( ) Ω=∇−•∇ in 0Vj oωεεσ

Page 12: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 12

Image Reconstruction:Static Imaging

Static imaging reconstructs the absolute conductivity from measurements.

Algorithms:• Iterative (Newton-Raphson)• Layer Stripping

Page 13: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 13

CurrentInjection

Patient

andEIT data

measurement

Real Data

Finite Element Model

EIT datasimulation Simulation Data

Does simulationapproximate

real data

Finishedyes

Updateconductivitydistribution

no

Block Diagram of Iterative Algorithm

Page 14: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 14

Static Imaging Difficulties

• Extremely sensitive to uncertainties in electrode position

• Ill-conditioned problem• Numerical instability

Page 15: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 15

Dynamic Imaging

• Calculate change in conductivity distribution from change in measurements

• Inverse problem linearized• Much reduced sensitivity to electrode

and hardware errors.• Very suitable for physiological imaging:

lung, heart, GI

Page 16: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 16

Dynamic imaging techniques

• Backprojection: Voltages are “projected” along the equipotential lines.

Equipotential Region

Current Injection I+_I

M+M_

Page 17: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 17

Inverse Techniques

• We can pose dynamic imaging as linear inverse, using a sensitivity matrix

j

jhhj δ

δσσ )()( +−=

zzz

σ∆= Hz

Page 18: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 18

Inverse Techniques

• Classic least-squares inverse

( ) zHHHx

Hxztt 1

ˆ−

=

=

Page 19: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 19

Matrix Techniques

However, problem is:• ill-conditioned: measurements depend

much more on data near electrodes than in centre

• ill-formed: more unknowns than measurements

Page 20: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 20

Regularized Imaging

Given Linear Model:

Maximum A Posteriori (MAP) estimate is:

( ) ( )∞−−−−− ++=

+=

xRRHRHRHx

nHxz

11111ˆ xnt

xnt z µµ

Page 21: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 21

Regularized Imaging• Parameters W, Q, , represent a priori

statistical knowledge of problem

[ ]( ) ( )[ ] [ ]

[ ]

==

−==

=

∞∞∞∞

22

21

00

σσ

nnR

xxxxx-xx-xR

x x

n

x

t

ttt

E

EE

E

∞x

Page 22: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 22

Advantages of Regularization

• Stabilizes ill-conditioned inverse• Introduction of a priori information• Control of resolution-noise performance

trade-off• MAP inverse justifies the formulation in

terms of Bayesian statistics

Page 23: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 23

Regularized Inverse

Parameters:• W: models measurement noise• Q: penalizes image features which are

greater than data supports• : represents the background

conductivity distribution (heart,lungs,etc)• : “hyper-parameter” amount of

regularization

∞x

Page 24: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 24

Noise – Resolution Tradeoff

F: Meas: No Noise Reconst: NF= 2.0G: Meas: -3dB SNR Reconst: NF= 2.0

Page 25: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 25

Noise – Resolution Tradeoff

F: Meas: No Noise Reconst: NF= 0.4G: Meas: -3dB SNR Reconst: NF= 0.4

Page 26: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 26

Electrode Movement

• Electrodes on chest move during breathing

• Figure shows rough movement pattern of rib cage

A B

Page 27: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 27

Electrode Movement

• FEM of electrical and mechanical properties of thorax to simulate

• Signal from expansion shown to contribute 10-20% of conductivity signal

Page 28: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 28

Applications: Lung Function and Disease

Monitoring of lung function is important in ICU. Issues are:

• Distribution of ventilation• Ventilation – perfusion match• Lung Edema• Flow limitation (obstructive lung

disease)

Page 29: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 29

Lung function tests

• Patients with obstructive lung disease (emphysema, bronchitis) tend to dynamically hyperinflate: lower inspiration than expiration.

• Simulated with PEEP (positive end expiratory pressure)

Page 30: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 30

Lung Function: Protocol

• Initially, inflations to 3 kPa• dog expires to the applied PEEP• ventilation stopped, and dog expires against PEEP for 15 s

Page 31: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 31

Lung Function

A: image of section thorax due to ∆VL 800 mlB: image due to a ∆VL 400 where left main

stem bronchus was plugged.

Page 32: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 32

• Volume estimates by EIT, Pao and Pes, after step volume increases of 100ml, 500ml, 900ml.

• Note that EIT signal does not display overshoot

Page 33: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 33

Pulmonary Edema

• Pulmonary Edema (lungs filled with fluid) plagues ICU patients for heart disease, accident victims

• Typical monitoring techniques are invasive (Swan-Ganz catheter)

• We looked at EIT ability to monitor level of Edema.

Page 34: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 34

Pulmonary Edema: Results

• Change in lung liquid volume by EIT vs liquid volume instilled

Page 35: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 35

Imaging

• Algorithms to use data from internal electrodes

• How to manage electrode artefacts• Arbitrary electrode placements• Image interpretation and error bounds

given artefacts and noise

Page 36: Electrical Impedance Tomography: Image Algorithms and Applications

A.Adler, Oct 28, 2002 Electrical Impedance Tomography 36

Applications of interest

• Longer term monitoring in ICU• Small animal lung function• Contrast agents• Inhomogeneous ventilation on smaller

scale