ds dusthunter c200 en 8013191
DESCRIPTION
Stack Dust Control Sick MaihakTRANSCRIPT
-
DATA SHEETContinuous measurement of dust with very low to very high concentrations
Intended purpose The DUSTHUNTER C200 serves for con-tinuous measurement of very low to very high dust concentrations in indus-trial plants as well as for monitoring limit values as required by the applica-ble directives and regulations.
ModelsThe combination of scattered light mea-suring technology and transmission measurement principle makes the DUSTHUNTER C200 ideal for efficient use in a wide spectrum of applications.
The DUSTHUNTER C200 consists of
DUSTHUNTER C200Transmissiometer and Scattered Lightthe components: Sender/receiver unit DHC-T Reflector/Scattered light receiver
DHC-R Flanges with tube MCU control unit (with/without
purge air supply) External purge air unit (option) Connection cable Purge air hose DN25 (for MCU-P
control unit with integrated purge air supply)
certified
-
2Overview of DUSTHUNTER C200 device componentsDUSTHUNTER C200 | SICK
Fig. 1 Layout of the DUSTHUNTER components onsite at the measuring location with internal duct pressure < 2 hPa
Fig. 2 Layout of the DUSTHUNTER components onsite at the measuring location with internal duct pressure < 30 hPa
Connection cable includedOnsite wiring
MCU-P control unit (integrated purge air)
Sender/receiver unit
Purge air hose DN 25 mm
Power supply:85 ... 250 V AC, 47 ... 63 Hz
Reflector/scattered light receiver
Internal duct pressure
-
3Fig. 3 Layout of the DUSTHUNTER components onsite at the measuring location with remote MCU < 1,000 mReflector/scattered DUSTHUNTER C200 | SICK
Plug connector for connection to onsite cable
Fig. 4 Pin assignment
For easier opening, insert the plug in the socket on the sender/receiver unit.Recommendation: For example, cable type UNITRONIC Li2YCYv(TP) 2 x 2 x 0.5 mm2 with reinforced outer sheath.
POWER
FAILURE
MAINTENANCEREQUEST
Status: Measuring
Status Ack Menu
No. Purge air hose Length Part No. Remark(4) Purge air hose DN 40 mm 5 m or
10 m
Sold by the meter
5034683
Included in scope of delivery of external purge air unitOrder separately
(5) Purge air reduction 40/25 7047538 Order separately
No. Signal cable for connec-tion Length Part No. Remark
(1) MCU control unit sender/receiver unit (SR)
Onsite
(2) SR-unit reflector 5 m10 m20 m
204541620454172048674
Order separately
(3) Plug connector for onsite cable
7045569 Order separately
Sender/receiver unit
(3)
light receiver
(2)
(5)
Max. 1,000 m
(1)
MCU-N con-trol unit (with-out purge air)
Power supply:85 ... 250 V AC, 47 ... 63 Hz
Optional modules (I/O, Interface)
External purge air unit
(4)
Purge air hoseDN 40 mm
Purge air reducer
Onsite cable
ClosedClosed
Open Open A
A
A -A
+24 V1
2
3 RS485 B
4 RS485 A
5
6gnd
7Shield
-
4Project planning information
Task Requirements Work stepDUSTHUNTER C200 | SICK
Determine the measuring and installation location for the device components
Inlet and outlet paths accord-ing to DIN EN 13284-1 (or com-parable regulation) inlet at least 5x hydraulic diameter dh, outlet at least 3x dh; distance to stack opening at least 5x dh
For round and square ducts: dh = duct diameter
Follow specifications for new equipment
Select best possible location for existing equipment
For too short inlet/outlet paths: Inlet path > outlet path
For rectangular ducts:dh = 4x cross-section divided by circumference
Uniform flow distribution Representative dust distri-
bution
Whenever possible, no deflec-tions, cross-section variations, feed and drain lines, flaps or fittings in the area of the inlet and outlet paths
If conditions cannot be ensured, define flow profile according to DIN EN 13284-1 and select best possible location
Assembly position for sender/receiver unit and reflector/scattered light receiver
Do not fit vertically on horizon-tal or slanted ducts; max. measuring axis angle to horizontal 45
Select best possible location
Accessibility, accident preven-tion
The device components must be easily and safely accessible
Provide platforms or pedestals as required
Installation free of vibrations Acceleration < 1 g Eliminate/reduce vibrations throughsuitable measures
Ambient conditions Limit values according to Tech-nical Data
If necessary: Provide weatherproof covers/sun
protection Enclose or lag device components
Select the purge air sup-ply
Sufficient primary purge air pressure depending on internal duct pressure
Up to +2 hPa, control unit with integrated purge air supply
Above +2 hPa to +30 hPa, optional external purge air unit
Select supply type
Clean intake air Whenever possible, low amount of dust, no oil, mois-ture or corrosive gases
Select best possible location for air intakeDetermine required purge air hose length
Select device components
Measuring path, duct wall thickness with insulation
Sender/receiver unit, reflector/scattered light receiver, flange with tube
Select components according to Configuration Tables (see Operat-ing Instructions);
if necessary, plan additional mea-sures to fit the flange with tube (see Operating Instructions)
Internal duct pressure Type of purge air supplyFitting locations Cable and purge air hose
lengthsPlan calibra-tionopenings
Access Easy and safe Provide platforms or pedestals as
Distances to measuring level No mutual interference between calibration probe and measuring system
Plan sufficient distance between measuring and calibration level (approx. 500 mm)
Plan power supply
Operating voltage, power requirements
According to Technical Data Plan adequate cable cross-sections and fuses
-
5Preparations for installation
Mounting suggestion for the DUSTHUNTER system componentsDUSTHUNTER C200 | SICK
Fig. 5 Example: Mounting options on a vertical duct/stack with external purge air unit
Active measuring path
Flange flange distance:0.5 3 m
Flange with tube
Reflector/ scattered light receiver for 0.5 - 3 m
Sender/receiver unit
76
470 mm
Min
. 700
mm
Approx. 600mm
470
mm
Attachment (e.g. 4 steel pipes, 50 x 5 mm) for external purge air unit
25 25
Sender/receiver unit
Active measuring path
Flange flange distance:2.5 8 m
Reflector/scattered light receiver for 2.5 - 8 m
Min
. 700
mm
76
15
9
25 40
470 mm
Approx. 600mm
470
mm
Attachment (e.g. 4 steel pipes, 50 x 5 mm) for external purge air unit
-
6Standard connectionDUSTHUNTER C200 | SICK
Fig. 6 Standard connection of DUSTHUNTER C200
Oper
a-tio
n/m
al-
Mai
nte-
Chec
k cy
cle
Lim
it va
lue
War
ning
*Onsite Connection cable (SICK cable or onsite cable)
Onsi
te
MCU processor board
-
7 Technical Data DUSTHUNTER C200 Transmission and Scattered LightDUSTHUNTER C200 | SICK
Measuring ParametersMeasured variable Transmission, opacity, extinction, scattered light intensity, dust concentrationMeasuring range (freely adjustable)
Min. Max.
Transmission measure-ment
100 ... 90% 100 ... 0%
Opacity 0 ... 10% 0 ... 100%
Extinction 0 ... 0.045 0 ... 2
Dust concentration Scattered light measurement0 ... 5 mg/m 0 ... 200 mg/mTransmission measurement (depending on active measuring path and particle characte-ristics; see special graphics)0 ... 200 mg/m 0 10,000 mg/m
Measurement uncertainty 2 %Damping time 1 ... 600 s; freely selectable
Measuring ConditionsActive measuring path1)
1) Upper limit only with distortion-free fitting
0.5 ...3; 2.5...8 mSwivel error 2)
2) With swivel angle 0.3; total swivel range 1
0.2% transmissionGas temperature (above dew point)
25 ... 300C Higher on request
Sample gas pressure 50 hPa ... +2 hPa50 hPa ... +30 hPa
MCU-P control unitOptional external purge air unit
Ambient temperature 40 ... +60 C
40 ... +45 C
Sender/receiver unit, reflector/scattered light receiver, MCU-N control unitMCU-P control unit, intake temperature for purge air
Function CheckAutomatic self-test Linearity, drift, aging, contamination
Contamination limit value: from 30% warning; from 40% malfunctionManual linearity check Using a reference filter
Output Signals Analog output 3 outputs 0/2/4 ... 22 mA, max. load 750 ; resolution 10 bits; electrically isolated,
further analog outputs by using I/O modules (option)Relay output 5 potential-free outputs (changeover contacts) for status signals; load 120 V AC, 1 A (30
V DC, 2 A); further relay outputs by using I/O modules (option)
Input SignalsAnalog input 2 inputs 0...20 mA (standard; without electric isolation) Digital input 4 inputs to connect potential-free contacts (e.g. for external maintenance switch, trigge-
ring check cycles)
Communication InterfacesUSB 1.1, RS232 (on termi-nals)
For measured value inquiries and software updates per PC/laptop using the operating program
RS485 To connect the sender/receiver unitOptional Interface module To communicate with the Host PC, optional for PROFIBUS, Ethernet
-
8Energy SupplyDUSTHUNTER C200 | SICK
Control unit Power supply:Power consumption:
85...250 V AC, 47...63 Hz; opt. 24 V DC 2 V Max. 15 W without purge air supplyMax. 70 W with purge air supply
Sender/receiver unit Power supply:Power consumption:
24 V from control unitMax. 4 W
Optional external purge air unit(with blower 2BH13)
Power supply:
Rated current:Motor rating:
200 ... 240 V/345 ... 415 V at 50 Hz;220 ... 275 V/380 ... 480 V at 60 Hz2.6 A/Y 1.5 A0.37 kW at 50 Hz; 0.45 kW at 60 Hz
Dimensions (W x H x D)Sender/receiver unit 420 mm x 250 mm x 212 mm Reflector/scattered light receiver
395 mm x 250 mm x 212 mm420 mm x 268 mm x 265 mm
DHC-R0DHC-R1
Control unit 300 mm x 400 mm x 220 mm200 mm x 200 mm x 130 mm
MCU-PMCU-N
Optional external purge air unit
550 mm x 550 mm x 258 mm; with weatherproof cover 605 mm x 550 mm x 350 mm
WeightSender/receiver unit 6.5 kgReflector/scattered light receiver
6.5 kg8.0 kg
DHC-R0DHC-R1
Control unit 13.5 kg 3.7 kg
MCU-PMCU-N
Optional external purge air unit
14 kg
Miscellaneous Degree of protection IP 66
IP 54
Sender/receiver unit, reflector/scattered light receiver, control unitOptional external purge air unit
Connection cable length 5 m, 10 m Other lengths on requestPurge air hose length 5 m, 10 m Other lengths on requestSender LED White light, wavelength between 450 nm and 700 nmLaser Degree of protection 2; capacity < 1 mW; wavelength between 640 nm and 660 nmPurge air feed volume Max. 20 m/h
Max. 63 m/hMCU-P control unitOptional external purge air unit
For more information on the DUSTHUNTER series and its components, please see the following documents: Operating Instructions DUSTHUNTER T, Part No. 8012428 Operating Instructions DUSTHUNTER SF, Part No. 8012424 Operating Instructions DUSTHUNTER SP, Part No. 8012426 Operating Instructions DUSTHUNTER SB, Part No. 8012422 Operating Instructions DUSTHUNTER C, Part No. 8011952 Data Sheet SLV4 Purge Air Unit, Part No.8008088
-
9DimensionsAll measures are specified in mm.DUSTHUNTER C200 | SICK
Fig. 7 DUSTHUNTER C200 sender/receiver unit
406
241
205
140
104 2
12r = 280 212
-
10
Fig. 8 Reflector/scattered light receiver DHC-R0 for active measuring paths of 0.5 3 mDUSTHUNTER C200 | SICK
Fig. 9 Reflector/scattered light receiver DHC-R1 for active measuring paths of 2.5 8 m
347 70
241
205
140
104 2
12
r = 240 220
352 191
308
140
212
r = 240223
104208
265
212
-
11
Fig. 10 MCU-P control unit
260DUSTHUNTER C200 | SICK
Fig. 11 MCU-N control unit without integrated purge air supply
Meas40
0
300 220
420
455
Meas
300
210
160
320
340
120
-
8013
192/
2009
-04
Prin
ted
in G
erm
any
(200
9-04
) S
ubje
ct to
mod
ifica
tions Fig. 12 Flange with tubeSICK MAIHAK GmbH I GermanyNimburger Str. 11 I 79276 Reute I www.sick.comPhone +49 7641 469-0 I Fax +49 7641 469-1149 I [email protected]
Fig. 13 Weatherproof cover for sender/receiver unit
D1
LD2D3
45 45
G
Dimen-sion
Flange with tube k100 k225
D1 76 159D2 100 225D3 130 265L 130, 240, 500 240, 500G M10 M12
Name Part No. Usage onFlange type k100Flange with tube, Di = 70.2 Length 130 mm, St37 2017845
DHC-T, DHC-R0
Flange with tube, Di = 70.2 Length 240 mm, St37 2017847Flange with tube, Di = 70.2 Length 500 mm, St37 2017849Flange with tube, Di = 70.2 Length 130 mm, 1.4571 2017846Flange with tube, Di = 70.2 Length 240 mm, 1.4571 2017848Flange with tube, Di = 70.2, length 500 mm, 1.4571 2017850Flange type k225Flange with tube, Di = 152 Length 240 mm, St37 2045415 DHC-R1Flange with tube, Di = 152 Length 500 mm, St37 2045418Flange with tube, Di = 152 Length 240 mm, 1.4571 2045419Flange with tube, Di = 152 Length 500 mm, 1.4571 2045420
492 365
360
239
Intended purposeOverview of DUSTHUNTER C200 device componentsPlug connector for connection to onsite cable
Project planning informationPreparations for installationStandard connection
Dimensions
/ColorImageDict > /JPEG2000ColorACSImageDict > /JPEG2000ColorImageDict > /AntiAliasGrayImages false /CropGrayImages true /GrayImageMinResolution 300 /GrayImageMinResolutionPolicy /OK /DownsampleGrayImages true /GrayImageDownsampleType /Bicubic /GrayImageResolution 300 /GrayImageDepth -1 /GrayImageMinDownsampleDepth 2 /GrayImageDownsampleThreshold 1.50000 /EncodeGrayImages true /GrayImageFilter /DCTEncode /AutoFilterGrayImages true /GrayImageAutoFilterStrategy /JPEG /GrayACSImageDict > /GrayImageDict > /JPEG2000GrayACSImageDict > /JPEG2000GrayImageDict > /AntiAliasMonoImages false /CropMonoImages true /MonoImageMinResolution 1200 /MonoImageMinResolutionPolicy /OK /DownsampleMonoImages true /MonoImageDownsampleType /Bicubic /MonoImageResolution 1200 /MonoImageDepth -1 /MonoImageDownsampleThreshold 1.50000 /EncodeMonoImages true /MonoImageFilter /CCITTFaxEncode /MonoImageDict > /AllowPSXObjects false /CheckCompliance [ /None ] /PDFX1aCheck false /PDFX3Check false /PDFXCompliantPDFOnly false /PDFXNoTrimBoxError true /PDFXTrimBoxToMediaBoxOffset [ 0.00000 0.00000 0.00000 0.00000 ] /PDFXSetBleedBoxToMediaBox true /PDFXBleedBoxToTrimBoxOffset [ 0.00000 0.00000 0.00000 0.00000 ] /PDFXOutputIntentProfile () /PDFXOutputConditionIdentifier () /PDFXOutputCondition () /PDFXRegistryName () /PDFXTrapped /False
/Description > /Namespace [ (Adobe) (Common) (1.0) ] /OtherNamespaces [ > /FormElements false /GenerateStructure true /IncludeBookmarks false /IncludeHyperlinks false /IncludeInteractive false /IncludeLayers false /IncludeProfiles true /MultimediaHandling /UseObjectSettings /Namespace [ (Adobe) (CreativeSuite) (2.0) ] /PDFXOutputIntentProfileSelector /NA /PreserveEditing true /UntaggedCMYKHandling /LeaveUntagged /UntaggedRGBHandling /LeaveUntagged /UseDocumentBleed false >> ]>> setdistillerparams> setpagedevice