  • DATA SHEETContinuous measurement of dust with very low to very high concentrations

    Intended purpose The DUSTHUNTER C200 serves for con-tinuous measurement of very low to very high dust concentrations in indus-trial plants as well as for monitoring limit values as required by the applica-ble directives and regulations.

    ModelsThe combination of scattered light mea-suring technology and transmission measurement principle makes the DUSTHUNTER C200 ideal for efficient use in a wide spectrum of applications.

    The DUSTHUNTER C200 consists of

    DUSTHUNTER C200Transmissiometer and Scattered Lightthe components: Sender/receiver unit DHC-T Reflector/Scattered light receiver

    DHC-R Flanges with tube MCU control unit (with/without

    purge air supply) External purge air unit (option) Connection cable Purge air hose DN25 (for MCU-P

    control unit with integrated purge air supply)

    certified

  • 2Overview of DUSTHUNTER C200 device componentsDUSTHUNTER C200 | SICK

    Fig. 1 Layout of the DUSTHUNTER components onsite at the measuring location with internal duct pressure < 2 hPa

    Fig. 2 Layout of the DUSTHUNTER components onsite at the measuring location with internal duct pressure < 30 hPa

    Connection cable includedOnsite wiring

    MCU-P control unit (integrated purge air)

    Sender/receiver unit

    Purge air hose DN 25 mm

    Power supply:85 ... 250 V AC, 47 ... 63 Hz

    Reflector/scattered light receiver

    Internal duct pressure

  • 3Fig. 3 Layout of the DUSTHUNTER components onsite at the measuring location with remote MCU < 1,000 mReflector/scattered DUSTHUNTER C200 | SICK

    Plug connector for connection to onsite cable

    Fig. 4 Pin assignment

    For easier opening, insert the plug in the socket on the sender/receiver unit.Recommendation: For example, cable type UNITRONIC Li2YCYv(TP) 2 x 2 x 0.5 mm2 with reinforced outer sheath.

    POWER

    FAILURE

    MAINTENANCEREQUEST

    Status: Measuring

    Status Ack Menu

    No. Purge air hose Length Part No. Remark(4) Purge air hose DN 40 mm 5 m or

    10 m

    Sold by the meter

    5034683

    Included in scope of delivery of external purge air unitOrder separately

    (5) Purge air reduction 40/25 7047538 Order separately

    No. Signal cable for connec-tion Length Part No. Remark

    (1) MCU control unit sender/receiver unit (SR)

    Onsite

    (2) SR-unit reflector 5 m10 m20 m

    204541620454172048674

    Order separately

    (3) Plug connector for onsite cable

    7045569 Order separately

    Sender/receiver unit

    (3)

    light receiver

    (2)

    (5)

    Max. 1,000 m

    (1)

    MCU-N con-trol unit (with-out purge air)

    Power supply:85 ... 250 V AC, 47 ... 63 Hz

    Optional modules (I/O, Interface)

    External purge air unit

    (4)

    Purge air hoseDN 40 mm

    Purge air reducer

    Onsite cable

    ClosedClosed

    Open Open A

    A

    A -A

    +24 V1

    2

    3 RS485 B

    4 RS485 A

    5

    6gnd

    7Shield

  • 4Project planning information

    Task Requirements Work stepDUSTHUNTER C200 | SICK

    Determine the measuring and installation location for the device components

    Inlet and outlet paths accord-ing to DIN EN 13284-1 (or com-parable regulation) inlet at least 5x hydraulic diameter dh, outlet at least 3x dh; distance to stack opening at least 5x dh

    For round and square ducts: dh = duct diameter

    Follow specifications for new equipment

    Select best possible location for existing equipment

    For too short inlet/outlet paths: Inlet path > outlet path

    For rectangular ducts:dh = 4x cross-section divided by circumference

    Uniform flow distribution Representative dust distri-

    bution

    Whenever possible, no deflec-tions, cross-section variations, feed and drain lines, flaps or fittings in the area of the inlet and outlet paths

    If conditions cannot be ensured, define flow profile according to DIN EN 13284-1 and select best possible location

    Assembly position for sender/receiver unit and reflector/scattered light receiver

    Do not fit vertically on horizon-tal or slanted ducts; max. measuring axis angle to horizontal 45

    Select best possible location

    Accessibility, accident preven-tion

    The device components must be easily and safely accessible

    Provide platforms or pedestals as required

    Installation free of vibrations Acceleration < 1 g Eliminate/reduce vibrations throughsuitable measures

    Ambient conditions Limit values according to Tech-nical Data

    If necessary: Provide weatherproof covers/sun

    protection Enclose or lag device components

    Select the purge air sup-ply

    Sufficient primary purge air pressure depending on internal duct pressure

    Up to +2 hPa, control unit with integrated purge air supply

    Above +2 hPa to +30 hPa, optional external purge air unit

    Select supply type

    Clean intake air Whenever possible, low amount of dust, no oil, mois-ture or corrosive gases

    Select best possible location for air intakeDetermine required purge air hose length

    Select device components

    Measuring path, duct wall thickness with insulation

    Sender/receiver unit, reflector/scattered light receiver, flange with tube

    Select components according to Configuration Tables (see Operat-ing Instructions);

    if necessary, plan additional mea-sures to fit the flange with tube (see Operating Instructions)

    Internal duct pressure Type of purge air supplyFitting locations Cable and purge air hose

    lengthsPlan calibra-tionopenings

    Access Easy and safe Provide platforms or pedestals as

    Distances to measuring level No mutual interference between calibration probe and measuring system

    Plan sufficient distance between measuring and calibration level (approx. 500 mm)

    Plan power supply

    Operating voltage, power requirements

    According to Technical Data Plan adequate cable cross-sections and fuses

  • 5Preparations for installation

    Mounting suggestion for the DUSTHUNTER system componentsDUSTHUNTER C200 | SICK

    Fig. 5 Example: Mounting options on a vertical duct/stack with external purge air unit

    Active measuring path

    Flange flange distance:0.5 3 m

    Flange with tube

    Reflector/ scattered light receiver for 0.5 - 3 m

    Sender/receiver unit

    76

    470 mm

    Min

    . 700

    mm

    Approx. 600mm

    470

    mm

    Attachment (e.g. 4 steel pipes, 50 x 5 mm) for external purge air unit

    25 25

    Sender/receiver unit

    Active measuring path

    Flange flange distance:2.5 8 m

    Reflector/scattered light receiver for 2.5 - 8 m

    Min

    . 700

    mm

    76

    15

    9

    25 40

    470 mm

    Approx. 600mm

    470

    mm

    Attachment (e.g. 4 steel pipes, 50 x 5 mm) for external purge air unit

  • 6Standard connectionDUSTHUNTER C200 | SICK

    Fig. 6 Standard connection of DUSTHUNTER C200

    Oper

    a-tio

    n/m

    al-

    Mai

    nte-

    Chec

    k cy

    cle

    Lim

    it va

    lue

    War

    ning

    *Onsite Connection cable (SICK cable or onsite cable)

    Onsi

    te

    MCU processor board

  • 7 Technical Data DUSTHUNTER C200 Transmission and Scattered LightDUSTHUNTER C200 | SICK

    Measuring ParametersMeasured variable Transmission, opacity, extinction, scattered light intensity, dust concentrationMeasuring range (freely adjustable)

    Min. Max.

    Transmission measure-ment

    100 ... 90% 100 ... 0%

    Opacity 0 ... 10% 0 ... 100%

    Extinction 0 ... 0.045 0 ... 2

    Dust concentration Scattered light measurement0 ... 5 mg/m 0 ... 200 mg/mTransmission measurement (depending on active measuring path and particle characte-ristics; see special graphics)0 ... 200 mg/m 0 10,000 mg/m

    Measurement uncertainty 2 %Damping time 1 ... 600 s; freely selectable

    Measuring ConditionsActive measuring path1)

    1) Upper limit only with distortion-free fitting

    0.5 ...3; 2.5...8 mSwivel error 2)

    2) With swivel angle 0.3; total swivel range 1

    0.2% transmissionGas temperature (above dew point)

    25 ... 300C Higher on request

    Sample gas pressure 50 hPa ... +2 hPa50 hPa ... +30 hPa

    MCU-P control unitOptional external purge air unit

    Ambient temperature 40 ... +60 C

    40 ... +45 C

    Sender/receiver unit, reflector/scattered light receiver, MCU-N control unitMCU-P control unit, intake temperature for purge air

    Function CheckAutomatic self-test Linearity, drift, aging, contamination

    Contamination limit value: from 30% warning; from 40% malfunctionManual linearity check Using a reference filter

    Output Signals Analog output 3 outputs 0/2/4 ... 22 mA, max. load 750 ; resolution 10 bits; electrically isolated,

    further analog outputs by using I/O modules (option)Relay output 5 potential-free outputs (changeover contacts) for status signals; load 120 V AC, 1 A (30

    V DC, 2 A); further relay outputs by using I/O modules (option)

    Input SignalsAnalog input 2 inputs 0...20 mA (standard; without electric isolation) Digital input 4 inputs to connect potential-free contacts (e.g. for external maintenance switch, trigge-

    ring check cycles)

    Communication InterfacesUSB 1.1, RS232 (on termi-nals)

    For measured value inquiries and software updates per PC/laptop using the operating program

    RS485 To connect the sender/receiver unitOptional Interface module To communicate with the Host PC, optional for PROFIBUS, Ethernet

  • 8Energy SupplyDUSTHUNTER C200 | SICK

    Control unit Power supply:Power consumption:

    85...250 V AC, 47...63 Hz; opt. 24 V DC 2 V Max. 15 W without purge air supplyMax. 70 W with purge air supply

    Sender/receiver unit Power supply:Power consumption:

    24 V from control unitMax. 4 W

    Optional external purge air unit(with blower 2BH13)

    Power supply:

    Rated current:Motor rating:

    200 ... 240 V/345 ... 415 V at 50 Hz;220 ... 275 V/380 ... 480 V at 60 Hz2.6 A/Y 1.5 A0.37 kW at 50 Hz; 0.45 kW at 60 Hz

    Dimensions (W x H x D)Sender/receiver unit 420 mm x 250 mm x 212 mm Reflector/scattered light receiver

    395 mm x 250 mm x 212 mm420 mm x 268 mm x 265 mm

    DHC-R0DHC-R1

    Control unit 300 mm x 400 mm x 220 mm200 mm x 200 mm x 130 mm

    MCU-PMCU-N

    Optional external purge air unit

    550 mm x 550 mm x 258 mm; with weatherproof cover 605 mm x 550 mm x 350 mm

    WeightSender/receiver unit 6.5 kgReflector/scattered light receiver

    6.5 kg8.0 kg

    DHC-R0DHC-R1

    Control unit 13.5 kg 3.7 kg

    MCU-PMCU-N

    Optional external purge air unit

    14 kg

    Miscellaneous Degree of protection IP 66

    IP 54

    Sender/receiver unit, reflector/scattered light receiver, control unitOptional external purge air unit

    Connection cable length 5 m, 10 m Other lengths on requestPurge air hose length 5 m, 10 m Other lengths on requestSender LED White light, wavelength between 450 nm and 700 nmLaser Degree of protection 2; capacity < 1 mW; wavelength between 640 nm and 660 nmPurge air feed volume Max. 20 m/h

    Max. 63 m/hMCU-P control unitOptional external purge air unit

    For more information on the DUSTHUNTER series and its components, please see the following documents: Operating Instructions DUSTHUNTER T, Part No. 8012428 Operating Instructions DUSTHUNTER SF, Part No. 8012424 Operating Instructions DUSTHUNTER SP, Part No. 8012426 Operating Instructions DUSTHUNTER SB, Part No. 8012422 Operating Instructions DUSTHUNTER C, Part No. 8011952 Data Sheet SLV4 Purge Air Unit, Part No.8008088

  • 9DimensionsAll measures are specified in mm.DUSTHUNTER C200 | SICK

    Fig. 7 DUSTHUNTER C200 sender/receiver unit

    406

    241

    205

    140

    104 2

    12r = 280 212

  • 10

    Fig. 8 Reflector/scattered light receiver DHC-R0 for active measuring paths of 0.5 3 mDUSTHUNTER C200 | SICK

    Fig. 9 Reflector/scattered light receiver DHC-R1 for active measuring paths of 2.5 8 m

    347 70

    241

    205

    140

    104 2

    12

    r = 240 220

    352 191

    308

    140

    212

    r = 240223

    104208

    265

    212

  • 11

    Fig. 10 MCU-P control unit

    260DUSTHUNTER C200 | SICK

    Fig. 11 MCU-N control unit without integrated purge air supply

    Meas40

    0

    300 220

    420

    455

    Meas

    300

    210

    160

    320

    340

    120

  • 8013

    192/

    2009

    -04

    Prin

    ted

    in G

    erm

    any

    (200

    9-04

    ) S

    ubje

    ct to

    mod

    ifica

    tions Fig. 12 Flange with tubeSICK MAIHAK GmbH I GermanyNimburger Str. 11 I 79276 Reute I www.sick.comPhone +49 7641 469-0 I Fax +49 7641 469-1149 I [email protected]

    Fig. 13 Weatherproof cover for sender/receiver unit

    D1

    LD2D3

    45 45

    G

    Dimen-sion

    Flange with tube k100 k225

    D1 76 159D2 100 225D3 130 265L 130, 240, 500 240, 500G M10 M12

    Name Part No. Usage onFlange type k100Flange with tube, Di = 70.2 Length 130 mm, St37 2017845

    DHC-T, DHC-R0

    Flange with tube, Di = 70.2 Length 240 mm, St37 2017847Flange with tube, Di = 70.2 Length 500 mm, St37 2017849Flange with tube, Di = 70.2 Length 130 mm, 1.4571 2017846Flange with tube, Di = 70.2 Length 240 mm, 1.4571 2017848Flange with tube, Di = 70.2, length 500 mm, 1.4571 2017850Flange type k225Flange with tube, Di = 152 Length 240 mm, St37 2045415 DHC-R1Flange with tube, Di = 152 Length 500 mm, St37 2045418Flange with tube, Di = 152 Length 240 mm, 1.4571 2045419Flange with tube, Di = 152 Length 500 mm, 1.4571 2045420

    492 365

    360

    239

    Intended purposeOverview of DUSTHUNTER C200 device componentsPlug connector for connection to onsite cable

    Project planning informationPreparations for installationStandard connection

    Dimensions

    /ColorImageDict > /JPEG2000ColorACSImageDict > /JPEG2000ColorImageDict > /AntiAliasGrayImages false /CropGrayImages true /GrayImageMinResolution 300 /GrayImageMinResolutionPolicy /OK /DownsampleGrayImages true /GrayImageDownsampleType /Bicubic /GrayImageResolution 300 /GrayImageDepth -1 /GrayImageMinDownsampleDepth 2 /GrayImageDownsampleThreshold 1.50000 /EncodeGrayImages true /GrayImageFilter /DCTEncode /AutoFilterGrayImages true /GrayImageAutoFilterStrategy /JPEG /GrayACSImageDict > /GrayImageDict > /JPEG2000GrayACSImageDict > /JPEG2000GrayImageDict > /AntiAliasMonoImages false /CropMonoImages true /MonoImageMinResolution 1200 /MonoImageMinResolutionPolicy /OK /DownsampleMonoImages true /MonoImageDownsampleType /Bicubic /MonoImageResolution 1200 /MonoImageDepth -1 /MonoImageDownsampleThreshold 1.50000 /EncodeMonoImages true /MonoImageFilter /CCITTFaxEncode /MonoImageDict > /AllowPSXObjects false /CheckCompliance [ /None ] /PDFX1aCheck false /PDFX3Check false /PDFXCompliantPDFOnly false /PDFXNoTrimBoxError true /PDFXTrimBoxToMediaBoxOffset [ 0.00000 0.00000 0.00000 0.00000 ] /PDFXSetBleedBoxToMediaBox true /PDFXBleedBoxToTrimBoxOffset [ 0.00000 0.00000 0.00000 0.00000 ] /PDFXOutputIntentProfile () /PDFXOutputConditionIdentifier () /PDFXOutputCondition () /PDFXRegistryName () /PDFXTrapped /False

    /Description > /Namespace [ (Adobe) (Common) (1.0) ] /OtherNamespaces [ > /FormElements false /GenerateStructure true /IncludeBookmarks false /IncludeHyperlinks false /IncludeInteractive false /IncludeLayers false /IncludeProfiles true /MultimediaHandling /UseObjectSettings /Namespace [ (Adobe) (CreativeSuite) (2.0) ] /PDFXOutputIntentProfileSelector /NA /PreserveEditing true /UntaggedCMYKHandling /LeaveUntagged /UntaggedRGBHandling /LeaveUntagged /UseDocumentBleed false >> ]>> setdistillerparams> setpagedevice