Full Terms & Conditions of access and use can be found at https://www.tandfonline.com/action/journalInformation?journalCode=iedd20 Expert Opinion on Drug Delivery ISSN: (Print) (Online) Journal homepage: https://www.tandfonline.com/loi/iedd20 User experience for manual injection of 2 mL viscous solutions is enhanced by a new prefillable syringe with a stacked 8 mm ultra-thin wall needle Aurélie Pager , Anne Combedazou , Karen Guerrero , Tzvetelina Tzvetkova- Chevolleau , Didier Morel , Cécile Frolet & Stanislav Glezer To cite this article: Aurélie Pager , Anne Combedazou , Karen Guerrero , Tzvetelina Tzvetkova- Chevolleau , Didier Morel , Cécile Frolet & Stanislav Glezer (2020): User experience for manual injection of 2 mL viscous solutions is enhanced by a new prefillable syringe with a stacked 8 mm ultra-thin wall needle, Expert Opinion on Drug Delivery, DOI: 10.1080/17425247.2020.1796630 To link to this article: https://doi.org/10.1080/17425247.2020.1796630 © 2020 Becton Dickinson France SAS. Published by Informa UK Limited, trading as Taylor & Francis Group View supplementary material Accepted author version posted online: 23 Jul 2020. Published online: 09 Aug 2020. Submit your article to this journal Article views: 18 View related articles View Crossmark data

Upload: others

Post on 05-Sep-2021

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Drug Delivery Systems, Pharmaceutical Systems - BD - User ......Published by Informa UK Limited, trading as Taylor & Francis Group View supplementary material Accepted author version

Full Terms & Conditions of access and use can be found athttps://www.tandfonline.com/action/journalInformation?journalCode=iedd20

Expert Opinion on Drug Delivery

ISSN: (Print) (Online) Journal homepage: https://www.tandfonline.com/loi/iedd20

User experience for manual injection of 2 mLviscous solutions is enhanced by a new prefillablesyringe with a stacked 8 mm ultra-thin wall needle

Aurélie Pager , Anne Combedazou , Karen Guerrero , Tzvetelina Tzvetkova-Chevolleau , Didier Morel , Cécile Frolet & Stanislav Glezer

To cite this article: Aurélie Pager , Anne Combedazou , Karen Guerrero , Tzvetelina Tzvetkova-Chevolleau , Didier Morel , Cécile Frolet & Stanislav Glezer (2020): User experience for manualinjection of 2 mL viscous solutions is enhanced by a new prefillable syringe with a stacked 8 mmultra-thin wall needle, Expert Opinion on Drug Delivery, DOI: 10.1080/17425247.2020.1796630

To link to this article: https://doi.org/10.1080/17425247.2020.1796630

© 2020 Becton Dickinson France SAS.Published by Informa UK Limited, trading asTaylor & Francis Group

View supplementary material

Accepted author version posted online: 23Jul 2020.Published online: 09 Aug 2020.

Submit your article to this journal

Article views: 18 View related articles

View Crossmark data

Page 2: Drug Delivery Systems, Pharmaceutical Systems - BD - User ......Published by Informa UK Limited, trading as Taylor & Francis Group View supplementary material Accepted author version

ORIGINAL RESEARCH

User experience for manual injection of 2 mL viscous solutions is enhanced by a new prefillable syringe with a stacked 8 mm ultra-thin wall needleAurélie Pagera, Anne Combedazou a, Karen Guerreroa, Tzvetelina Tzvetkova-Chevolleaua, Didier Morelb, Cécile Froleta and Stanislav Glezerc

aBD Medical - Pharmaceutical Systems, Becton Dickinson and Company, Le Pont De Claix, France; bGlobal Clinical Development, Becton Dickinson and Company, Le Pont De Claix, France; cDiabetes Care, Becton Dickinson and Company, Franklin Lakes, NJ, USA

ABSTRACTObjectives: User experience was compared between a new pre-fillable 2.25 mL glass syringe equipped with an ultra-thin-wall (UTW) 8 mm stacked needle and a marketed BD Neopak™ syringe equipped with a special-thin-wall (STW) 12.7 mm stacked needle.Methods: Participants simulated subcutaneous injections with both syringes alone (formative Human Factors study) and in combination with a needlestick-prevention device (validation Human Factors study).Results: Usability results of both studies showed higher success rates for delivering the full dose of 2 mL viscous solution (30 cP) with the 8mmUTW syringe than with the 12.7mmSTW one (63% vs. 42% in the formative study). The use of the 8mmUTW syringe demonstrated also better ease of use and acceptance results and 72% of formative study participants preferred this new syringe over the current one when delivering the viscous solution. Using a shorter needle also showed a benefit in decreasing the injection-related anxiety. Besides, in the case of a non-recommended injection technique, the calculated risk of accidental intramuscular injection is reduced by 2 to 13 times with the 8mmUTW syringe.Conclusion: Altogether, the results obtained demonstrated an improvement of the user experience with this new syringe compared to the current one in the manual delivery of 2 mL viscous solutions.

ARTICLE HISTORYReceived 26 June 2020 Accepted 10 July 2020

KEYWORDSHuman factors; usability; user experience; viscous drug delivery; manual subcutaneous injection; prefillable syringe; 8 mm needle; ultra-thin wall cannula; needlestick prevention device; chronic care

1. Introduction

Subcutaneous injection is one of the most common routes of administration used for administration of therapeutic proteins, peptides, and other biologics used in chronic diseases (growth hormone, insulin, heparin, interferons, monoclonal antibo-dies), mainly due to low oral bioavailability, gastric degrada-tion of oral drugs, local toxicity at the site of absorption, suboptimal pharmacokinetics associated with gastrointestinal absorption or lack of efficient gastrointestinal transport mechanisms for these molecules [1,2]. Moreover, the benefits of home administration by patients themselves or lay care-givers and the increasing interest in new biotherapeutics, such as monoclonal antibody-based therapies, has challenged the limits of volume and viscosity of drug formulations [1,3,4]. Although subcutaneous administration improves patient com-pliance and quality of life by enabling at-home injections, as compared to intravenous or intramuscular route, the asso-ciated anxiety, pain, discomfort, and the frequency of injec-tions can play a significant role in patient non-adherence with chronic subcutaneous treatments [5–8]. In multiple sclerosis patients, anxiety related to injections was often associated with poor adherence to treatment, and the anxiety perceived by some patients with diabetes, for example, resulted in fewer

than recommended daily insulin injections performed [9,10]. In order to address these challenges, through reducing the required frequency of subcutaneous injections and increasing injection comfort for novel biotherapeutics, drug delivery sys-tems should be designed to enable the administration of large-volume injections and/or high-viscosity drugs [11,12].

High viscosity and large-volume solutions raise a new tech-nical challenge to keep on allowing manual drug administra-tion for the proportion of users who want to control their injection compared to injections with auto-injectors. These solutions must be designed without affecting the user’s experience in terms of pain, anxiety, ease of use, and accep-tance, while several factors can influence the perception of pain, such as injection volume, drug viscosity, or needle inser-tion [6,13–20].

For subcutaneous insulin therapy, short needles had been introduced to increase the certainty of delivering into the targeted subcutaneous tissue. Outside diabetes, despite the risk of injecting into muscular tissue, the most commonly used needle length for subcutaneous chronic drug delivery is the one-half inch (12.7 mm) [21]. In this context, the accidental administration of insulin into muscular tissue can lead to faster insulin absorption and hypoglycemia episodes [22–25]. Relative to intramuscular, subcutaneous injections also

CONTACT Anne Combedazou [email protected] BD Medical - Pharmaceutical Systems, Becton Dickinson and Company, Le Pont De Claix, France

Supplemental data for this article can be accessed here.

EXPERT OPINION ON DRUG DELIVERYhttps://doi.org/10.1080/17425247.2020.1796630

© 2020 Becton Dickinson France SAS. Published by Informa UK Limited, trading as Taylor & Francis Group This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivatives License (http://creativecommons.org/licenses/by-nc-nd/4.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the original work is properly cited, and is not altered, transformed, or built upon in any way.

Page 3: Drug Delivery Systems, Pharmaceutical Systems - BD - User ......Published by Informa UK Limited, trading as Taylor & Francis Group View supplementary material Accepted author version

improve the efficacy of Interferon-beta-1a treatment in multi-ple sclerosis [26,27]. Moreover, several studies demonstrated an improved patient satisfaction and preference for the sub-cutaneous route of administration [28,29]. Therefore, the use of a 12.7 mm needle length is not necessary and may be associated with unwarranted risk of accidental intramuscular injection, as demonstrated in diabetes care, especially when self-injecting patients do not apply the recommended subcu-taneous injection technique. Injection at an angle of 90° or 45° with a skinfold is recommended for subcutaneous administra-tions with this needle length [30]. In 2010, Frid, A., et al., studied subcutaneous injection practices in 4300 patients using insulin. Despite growing availability of injection techni-que training, they observed that 43% of these patients released the skinfold too soon and 17% were using an incor-rect technique for lifting a skinfold [31,32]. While these errors could result in accidental intramuscular injections, the use of a needle shorter than 12.7 mm may help mitigate this risk. In 2010, Gibney, M., et al., measured the skin thickness and subcutaneous fat depth in the thigh, arm, abdomen, and buttock in adult patients with diabetes [33]. Based on these measurements, they estimated the location of the drug depot (intradermal, intramuscular, or subcutaneous) depending on the needle length. Assuming a needle insertion at 90° without lifting a skinfold, the risk of intramuscular administration is reduced with an 8 mm needle (15%) compared to a 12.7 mm needle (45%). In 2014, Hirsch, L., et al., confirmed that the estimated probability of unintended intramuscular administra-tion is significantly reduced with 8 mm needles compared to 12.7 mm needles [34]. The same trend is also observed in children and adolescents with diabetes, for whom the use of 8 mm needles is estimated to significantly reduce the occur-rence of intramuscular injections by about 50% relative to the use of 12.7 mm needles [35]. Accidental intramuscular admin-istration might be responsible for potential side effects such as painful administration, hematoma formation, or risk of ana-phylaxis across the range of chronic injectable drugs [36–38]. However, the risk-benefit ratio is likely to be less favorable if patients are exposed to the risk of intradermal administration. In contrast to subcutaneous tissue, the dermis is composed of many specialized immune cells (i.e. dermal dendritic cells), which may induce an undesirable immune response against protein therapeutics intended to treat diseases such as Crohn’s disease or rheumatoid arthritis [39].

Therefore, to support the delivery of higher drug viscosities and volumes into the targeted subcutaneous tissue, we devel-oped a pre-fillable glass syringe featuring a stacked 8 mm needle with an ultra-thin wall (UTW) cannula technology, BD Neopak™ XtraFlow™ 2.25 mL, hereafter referred to ‘8mmUTW’ (Figure 1(a)). In comparison to the currently marketed BD Neopak™ 2.25 mL pre-fillable glass syringe equipped with a 12.7 mm special thin wall (STW) needle, hereafter named as 12.7mmSTW, the 8mmUTW offers an optimized flow capa-city due to shorter needle length and larger inner diameter of the cannula (Figure 1(a)). The UTW technology makes the use of smaller gauge (larger outer diameter) needles unnecessary, which can positively impact the pain perception when

enabling the injection of viscous solutions [40,41]. Simulated- use studies with the 8mmUTW syringe alone or combined with an anti-needlestick device have been conducted to assess their safety and efficacy and identify the potential use-related risks. A formative study was undertaken to evaluate the usabil-ity, the ease of pushing the plunger rod, and the acceptance of the force required for pushing the plunger rod with the 8mmUTW syringe versus the 12.7mmSTW syringe. The BD UltraSafe Plus™ Passive 2.25 mL Needle Guard, intended to offer needlestick-prevention (X225PLUS), is compatible with the 8mmUTW syringe (Figure 1(b)). Among the objectives of the validation study, some were to compare the usability, the ease of use, the acceptance, and the adoption of the X225PLUS–8mmUTW versus the X225PLUS–12.7mmSTW com-bination product.

2. Participants and methods

2.1. Products description

Based on the marketed standard BD Neopak™ 2.25 mL glass syringe equipped with a 12.7 mm 27 G needle with Special Thin Wall (STW), hereafter referred to 12.7mmSTW syringe, a new syringe, BD Neopak™ XtraFlow™ 2.25 mL, equipped with a shorter 8 mm needle and an Ultra-Thin Wall (UTW) technology is currently under development, hereafter referred to 8mmUTW syringe (Figure 1(a)). This 8 mm needle offers identical outer diameter (same 27 G) but larger inner diameter and is intended to help the delivery of higher viscosities and/or volumes without degrading the user experience. In addition to the different needle technologies, 2 rigid needle shields (RNS), RNS-8 mm and RNS-12.7 mm, were made of the same elastomer (BD260). The plunger rod, the stopper, and the glass barrel were the same for the 2 tested syringes (Figure 1(a)).

As the standard 12.7mmSTW, the 8mmUTW syringe can be assembled with a non-sterile and disposable device with a passive activation of a needle covering safety system named BD UltraSafe Plus™ Passive needle guard 2.25 mL (hereafter abbreviated ‘X225PLUS’), intended to reduce the risk of acciden-tal needle stick injuries after injection. In this study, participants used X225PLUS devices equipped or not with an add-on finger flange providing a larger flange for a better grip when pushing the plunger rod (Figure 1(b)).

All the syringes were filled with polyethylene glycol 400 EMPROVE® ESSENTIAL Ph Eur, JP (Merck), diluted to obtain 10 cP or 30 cP solutions. Both solutions behave as Newtonian fluids. The viscosity is not affected by the shear rate of the needle.

2.2. Simulated studies participants

Based on FDA Guidance (2016), ‘Applying Human Factors and Usability Engineering to Medical Devices’ and AAMI/ANSI-HE -75:2009 Part 9.3.2/Table 9.2., a minimum of 5 or 15 partici-pants per user group were recruited for the formative and the validation studies, respectively. Representative end-users over

2 A. PAGER ET AL.

Page 4: Drug Delivery Systems, Pharmaceutical Systems - BD - User ......Published by Informa UK Limited, trading as Taylor & Francis Group View supplementary material Accepted author version

18 years old were recruited either in the UK or in the US for the formative study and representative end-users over 21 years old were recruited in the US for the validation study:

● Naïve patients, diagnosed with at least one chronic dis-ease, did not have any experience in injecting them-selves or others with syringes, pens, or autoinjectors in the last 10 years.

● Experienced patients diagnosed with at least one chronic disease. In the formative study, these participants had experienced a minimum of 12 self-injections with pre-filled syringes for at least 6 months in the past 2 years (only one participant used prefilled syringes with a safety add-on). Experienced participants involved in the valida-tion study had at least 1 year of experience giving self- injections using syringes with active or passive safety systems (a maximum of 10% experienced patients with only self-injecting experience using pen for diabetes

treatments and a maximum of 20% of them having only self-injecting experience using an autoinjector).

● Hand-impaired patients, diagnosed with Rheumatoid Arthritis or Multiple Sclerosis affecting their fingers and/ or hands, characterized by a Cochin score between 10 and 40 at the recruiting stage for the formative study and on the day of testing for the validation study [42,43].

● 16 Health care workers (HCW) with at least 2 years of working experience and 16 Caregivers (CGV) with at least 6 months of experience and a minimum of 12 injections provided to another individual participated in the valida-tion study, as part of the intended users’ population. Less than 10% and 20% of the caregivers had experience only using reusable pens or autoinjectors, respectively.

All participants signed an informed consent form before parti-cipating in the described studies. Both studies got IRB approval for testing conducted in the US.

Figure 1. (a). BD Neopak™ 2.25 mL glass pre-fillable syringe with special thin-wall 12.7 mm (left) and ultra thin-wall 8 mm (right) needles. (b). BD UltraSafe Plus™ 2.25 mL Passive needle guard combined with BD Neopak™ 2.25 mL glass pre-fillable syringes with special thin-wall 12.7 mm needle with (left panel) and without the Add-on finger flange (right panel).

EXPERT OPINION ON DRUG DELIVERY 3

Page 5: Drug Delivery Systems, Pharmaceutical Systems - BD - User ......Published by Informa UK Limited, trading as Taylor & Francis Group View supplementary material Accepted author version

2.3. Study design of the formative study

The human factors formative study was a multicenter study conducted between August and September 2019 in the US (Boston, MA) and the UK (London). The primary objective was to assess the usability of the 8mmUTW syringe in comparison to the marketed 12.7mmSTW syringe, filled with 2 mL of 10 cP or 30 cP solutions. Secondary objectives were to evaluate the ease of pushing the plunger rod and the acceptance of the force required to push the plunger rod when delivering the solution, as well as the preference between both syringes. This formative testing was performed to confirm that the 8mmUTW syringe is suitable for its intended use without revealing new use-related hazards, use errors, or even without increasing the occurrence of any known risks in comparison to the marketed 12.7mmSTW syringe.

The 39 participants enrolled were not trained on how to use the syringes before the session, but the instructions for use (IFU) of both syringes were available to read during the entire simulated testing. Each participant was asked to per-form 4 simulated injections with the 2 types of syringe filled with 2 different viscous solutions, according to an assigned randomization order. The participants tested both syringes by simulating subcutaneous injections into a foam pad attached to a belt that they placed on their body (thigh or abdomen). Acceptance and preference were assessed after each two simulations with a same viscosity, while the root cause inter-view was performed once the 4 simulations done.

2.3.1. Task assessmentThe usability of the syringes was evaluated for the following critical and essential tasks: ‘Remove the needle cover,’ ‘Perform the injection’, ‘Deliver the full dose’ and ‘Disposal of the used syringe.’ As per FDA guidance, critical tasks would or could cause serious harm to the patient or user in case of use error or failure and essential tasks are the ones necessary for suc-cessful use of the product and associated with high risk but resulting in severity level below than critical tasks. Correct use was defined as achieving all assessed tasks in compliance with the IFU, including success and success with operational diffi-culties (SOD). The moderator recorded success with and with-out operational difficulties and use errors. Usability data are described as categorical data (N), and proportion frequen-cies (%).

2.3.2. Anxiety assessmentBefore starting the simulated injections, naïve patients were asked to evaluate both uncapped 8mmUTW and 12.7mmSTW syringes, provided in front of them, in terms of anxiety if they were going to self-inject a treatment. Subjective patients’ perception and feeling of anxiety was measured using a visual analogue scale (VAS) ranging from ‘no anxiety’ (0%) to ‘worst anxiety’ (100%). Scores were measured in millimeters (120 mm-VAS) and transformed in percentage.

Causes of potential anxiety and feedbacks on the shorter needle perception were recorded for the experienced popula-tion after all simulations.

2.3.3. Ease of pushing the plunger rod and acceptance of the force required to push the plunger rodThe ease of pushing the plunger rod and the acceptance of the force required to push it were recorded through a VAS scale after having tested each syringe for a same viscosity (two simulations). This approach allowed the participants to com-pare the two tested syringes for the same viscosity and marked each syringe on the same VAS scale. Subjective patients’ perception of ease and acceptance was measured using a 120 mm-VAS ranging from ‘very easy/very acceptable’ (0%) to ‘very difficult/very unacceptable’ (100%).

2.3.4. PreferenceThe moderator asked the participants which syringe they have preferred to use between the 8mmUTW and the 12.7mmSTW and the reason(s) why, after each set of 2 simulations for a same viscosity.

2.4. Study design of the validation study

The human factors validation study was a multi-center study conducted between December 2019 and January 2020 in the US (Newark, NJ; Philadelphia PA; Baltimore, MD and Boston, MA). This study aimed to assess the safe and effective use of the X225PLUS product. Additional objectives were to compare the usability performance of the X225PLUS-12.7mmSTW and X225PLUS-8mmUTW products when filled with 2 mL of a 30 cP viscous solution, as well as the ease and the acceptance of pushing the plunger rod when delivering the full dose was assessed for both product configurations.

No training was provided to the 63 participants enrolled, but all of them received the IFU and were free to read it. Each participant was asked to perform two simulated injections with the X225PLUS-12.7mmSTW and then two simulated injec-tions with the X225PLUS-8mmUTW products. All the partici-pants simulated the subcutaneous injections into a foam pad placed on the thigh or on the abdomen for patients or on a manikin for health care workers and caregivers. The injection technique is limited at a 90° angle with the X225PLUS- 8mmUTW product instead of a 45° insertion angle to prevent the risk of intradermal administration. Debriefings on the root cause, ease, and acceptance of pushing the plunger rod as well as willingness to adopt both products were performed after each simulation.

2.4.1. Task assessmentThe primary endpoint was the number and percentage of participants that correctly used the X225PLUS–8mmUTW in comparison to the X225PLUS-12.7mmSTW. For both products, 5 critical tasks were evaluated: ‘Remove the needle cover,’ ‘Perform the injection,’ ‘Deliver the full dose,’ ‘Activate the guard,’ and ‘Disposal of the used product.’ For each simula-tion, the moderator assessed all tasks and recorded the results based according to the IFU of the products, without giving any feedback or indication of performance to the participant. The moderator recorded any use errors, SOD as well as

4 A. PAGER ET AL.

Page 6: Drug Delivery Systems, Pharmaceutical Systems - BD - User ......Published by Informa UK Limited, trading as Taylor & Francis Group View supplementary material Accepted author version

success. For each use error or difficulty encountered by parti-cipants, the root cause was investigated. Usability data are described as categorical data (N), and proportion frequencies (%). Both performance-based and subjective data were ana-lyzed to ensure that no new risks were identified.

2.4.2. Ease of pushing the plunger rod and acceptance of the force required to push the plunger rodSubjective responses were assessed through a questionnaire using a 6-points Likert scale. The percentage of answers with a score ≥5 per question was calculated, representing only strong positive feedbacks.

Likert scale was defined as following: ‘1: very difficult/fully unacceptable’; ‘2: difficult/unacceptable’; ‘3: somewhat diffi-cult/somewhat unacceptable’; ‘4: somewhat easy/somewhat acceptable’; ‘5: easy/acceptable’ and ‘6: very easy/very acceptable.’

2.4.3. Willingness to adopt the productUser’s willingness to adopt each combination if prescribed was assessed using close-ended question. The number and percen-tage of positive answers were calculated.

2.5. Mathematical model of intramuscular and intradermal risks

A mathematical model was developed with the R statistical software (version 3.6.3) based on published skin thickness data at 4 common injection sites for adults with diabetes (abdo-men, arm, buttock, and thigh) [33]. This model estimates intradermal (ID) and intramuscular (IM) injection risks using several needle penetration depths (NPD) for two needle lengths (12.7mmSTW and 8mmUTW) at two injection angles (90° and 45°). These data were stratified by factor (gender, Body Mass Index, injection site). Tissue ID thickness, IM depth, and NPD values were generated using a random number generator function from the respective distributions, each defined by their mean and standard deviation. Logarithm transform was used for IM depth to correct for non-normality of the observed distribution.

Monte Carlo simulations were run for a sample size of 500 for each combination of tissue thickness and

NPD. For each pair of ID depth and NPD and IM depth and NPD, the ID and IM risks were calculated as the proportion of simulated samples where NPD < ID depth and NPD > IM depth, respectively.

Percentages of intramuscular and intradermal risks and ratios were computed between the two syringes of interest (at 45° and 90° injection angles).

3. Results

3.1. Participants demographics

The demographic characteristics of all the participants who took part in the studies are summarized in Table 1.

In the formative study 39 participants, Experienced patients (EXP), Naïve patients (NVP) and Hand-impaired patients (IMP) with a Cochin score between 1 and 86 were included in analyzes.

In the validation study, evaluating the use of the 8mmUTW syringe with the needlestick-prevention device X225PLUS, 63 representative end-users were included (Health Care Workers (HCW), Caregivers (CGV), Experienced patients (EXP), and Naïve patients (NVP)). Among the patient participants, 15 were having hand impairment (IMP) characterized by a Cochin Score in the range 12–38.

3.2. Formative study 8mmUTW vs. 12.7mmSTW syringes – Anxiety

Having the choice between 12.7mmSTW and 8mmUTW syr-inges, naïve patients indicated to be less anxious by using the syringe with the shorter needle and indicated that they would favor it for self-injection. Indeed, the median VAS score was lower for the 8mmUTW (34%) syringe than for the 12.7mmSTW syringe (53%) (Figure 2). Among the 23 experi-enced participants (with or without hand-impairment), 43% reported to be anxious before to self-inject their treatments because they either anticipated the potential pain or side

Table 1. Demographic characteristics of the participants.

HF formative #1 (8mmUTW vs. 12.7mmSTW) HF summative #2 (X225PLUS–8mmUTW vs. X225PLUS-12.7mmSTW)

EXP NVP IMP EXP NVP IMP* HCW CGV

Total N 13 16 10 15 16 15 16 16Age (years),

[min–max]45 [26–66]

55 [45–73]

58 [49–74]

55 [37–68]

59 [38–81]

57 [45–72]

49 [28–63]

50 [33–68]

Gender female N (%) 10 (77%) 5 (31%) 6 (60%) 10 (59%) 12 (71%) 11 (73%) 14 (88%) 11 (69%)Dominant hand

Right 12 (92%) 15 (94%) 8 (80%) 13 (81%) 13 (81%) 10 (67%) 15 (92%) 14 (87%)Left 1 (8%) 1 (6%) 0 (0%) 2 (13%) 3 (19%) 4 (27%) 1 (6%) 2 (13%)Ambidextrous 0 (0%) 0 (0%) 2 (10%) 1 (6%) 0 (0%) 1(7%) 0 (0%) 0 (0%)

Hand impairmentCochin score > 40 NA NA 4 (40%) 9 9 0 NA NA10 ≤ Cochin

score ≤40NA NA 4 (40%) 7 (47%) 8 (50%) 15 (100%) NA NA

Cochin score < 10 NA NA 2 (20%) 0 0 0 NA NA

*Hand-impaired patients (IMP) = subgroup of Experienced (EXP, N = 7) and Naïve patients (NVP, N = 8). HCW: Health Care Workers; CGV: Caregivers. NA: Not Applicable.

EXPERT OPINION ON DRUG DELIVERY 5

Page 7: Drug Delivery Systems, Pharmaceutical Systems - BD - User ......Published by Informa UK Limited, trading as Taylor & Francis Group View supplementary material Accepted author version

effects, or they were afraid by long needles) and 78% thought a shorter needles could be an advantage (more comfortable, less painful, and intimidating).

3.3. Intradermal and intramuscular injection risks

The risks of injecting in a wrong tissue, either intradermal or intramuscular ones, with the 8mmUTW versus the 12.7mmSTW syringes, were estimated based on Gibney M., et al. data (Table 2) [33]. Regardless of the injection sites, the

use of shorter needles (8 mm) compared to conventional ones would not increase the risk of intradermal injections when inserting the needle at 45° or 90° without raising a skinfold. The estimated risk of accidental injection into the muscle according to the 4 considered injection sites (thigh, abdomen, arm, and buttocks), in case of unsuitable injection technique used (injection at 90° without/or with incorrect skinfold) is reduced with the 8mmUTW by 59% to 84%. Considering the injection at 45° without skinfold, the risk of intramuscular delivery is also reduced (from 76% to 92%) for the 8mmUTW syringe versus the 12.7mmSTW syringe (Table 2).

3.4. Formative study 8mmUTW vs. 12.7mmSTW syringes – usability

The overall usability results obtained by the participants were similar to the 8mmUTW and the 12.7mmSTW syringes, filled with 2 mL of 10 cP or 30 cP solutions (Figure 3(a)).

The 8mmUTW syringe improved the removal of the needle cover compared to the 12.7mmSTW syringe (Figure 3(b) and 3(c). While no success with operational difficulty was observed to remove the RNS-8mm, the success rate was increased by 18% and 17%, at 10 cP and 30 cP, respectively, against the removal of the RNS-12.7mm (Figure 3(b) and 3(c)). The same trend was observed for all user groups (Figure S1A). The success rate of removing the needle cover was increased with the use of the 8mmUTW syringe relative to the 12.7mmSTW syringe for the experienced and naïve patients, regardless of the viscosity. No improvement was observed for the hand-impaired patients, but the removal of the needle cover was correctly achieved by almost all these participants.

Similar rates of use errors were observed when performing the task ‘Perform injection’ for the two types of syringes at 10

Figure 2. Boxplot distributions of the VAS rating of the anxiety perceived by naive self-injecting patients during the formative study. Horizontal lines indicate median values, boxes indicate interquartile range, and whiskers indicate minimum and maximum values. VAS scale ranging from 0% = no anxiety at all to 100% = worst anxiety ever.

Table 2. Estimated risks of intradermal (ID) and intramuscular (IM) injections with the 8mmUTW and the 12.7mmSTW syringes per site of injection and angle of insertion. The percentage of reduction in IM and ID risks and the risk ratio comparing the 8mmUTW and 12.7mmSTW syringes were calculated. N = 500 sample size. Percentages and risk ratio values are rounded up.

Site Risk

Insertion angle

(degrees)

12.7mm STW (%)

8mm UTW (%)

Percentage of risk

reduction between

12.7mmSTW and

8mmUTW

Risk ratio (12.7mm STW/

8mmUTW)

Thigh ID 90 0.0% 0.0% NA NAAbdomen ID 90 0.0% 0.0% NA NAArm ID 90 0.0% 0.0% NA NAButtocks ID 90 0.0% 0.0% NA NAThigh IM 90 62.4% 25.3% 59.4% 2.5Abdomen IM 90 37.1% 8.4% 77.4% 4.4Arm IM 90 54.8% 17.2% 68.6% 3.2Buttocks IM 90 26.0% 4.2% 83.9% 6.2Thigh ID 45 0.0% 0.0% NA NAAbdomen ID 45 0.0% 0.0% NA NAArm ID 45 0.0% 0.0% NA NAButtocks ID 45 0.0% 0.0% NA NAThigh IM 45 33.8% 8.3% 75.5% 4.1Abdomen IM 45 13.2% 1.6% 87.9% 8.3Arm IM 45 24.8% 4.2% 83.2% 6.0Buttocks IM 45 7.2% 0.6% 92.0% 12.6

6 A. PAGER ET AL.

Page 8: Drug Delivery Systems, Pharmaceutical Systems - BD - User ......Published by Informa UK Limited, trading as Taylor & Francis Group View supplementary material Accepted author version

cP and 30 cP (Figure 3(b) and 3(c)). Overall syringe configura-tions, impaired and naïve patients experienced 61% and 56% of use errors, respectively, while only 13% were recorded for experienced users (Figure S1B).

Regarding the task ‘Deliver the full dose,’ participants experienced more operational difficulties when using both syringes, 8mmUTW, and 12.7mmSTW, filled with the 30 cP solution compared to those filled with the 10 cP solution.

Figure 3. Usability of 8mmUTW and 12.7mmSTW syringes during the formative study. Overall task completion rates (a) and per task assessed (b). Success, use error and success with operational difficulty (SOD) rates are indicated for syringes filled with 2 mL of 10 cP solution. N number of simulated injections; %. (c). Success, use error and success with operational difficulty (SOD) rates are indicated for syringes filled with 2 mL of 30 cP solution. N number of simulated injections; %. * 71% (for 12.7mmSTW) and 83% (for 8mmUTW) of the use errors were caused by syringe priming. (d) Rates of success, use error, and artifact in regard to the injection technique used by patients. N number of simulated injections; %.

EXPERT OPINION ON DRUG DELIVERY 7

Page 9: Drug Delivery Systems, Pharmaceutical Systems - BD - User ......Published by Informa UK Limited, trading as Taylor & Francis Group View supplementary material Accepted author version

Indeed, more participants were using two hands instead of one to deliver the full dose. Interestingly, while the success rate to deliver the full dose decreased for the high viscos-ity, the syringe configuration also affects the performance of participants for this task. When participants used the 8mmUTW syringe, they experienced more successes (42% to 63%) and fewer operational difficulties (39% to 21%) on delivering the full dose at 30 cP than when using the 12.7 mm syringe (Figure 3(c)). The participants with prior practice in self-injecting experienced more successes to deliver the full dose when using the 8mmUTW syringe relative to the 12.7mmSTW syringe filled with the 30 cP solution (75% vs. 37.5%) (Figure S1C). Regarding both syr-inges filled with a 30 cP solution, a greater proportion of the hand-impaired patient population correctly achieved the full dose delivery when using the 8mmUTW syringe rather than using the 12.7mmSTW syringe (50% vs. 13%) (Figure S1C). Moreover, hand-impaired patients experi-enced more successes with operational difficulties with the 12.7mmSTW syringe relative to the 8mmUTW syringe (50% vs. 25%) (Figure S1C). The increase in the number of use errors recorded for impaired and naïve patients was caused by using a wrong injection technique. Injections at a 45° or 90° angle into a skinfold, or at a 45° without a skinfold were considered acceptable. However, 35% of the participants, including experienced ones, simulated the injection at a 90° angle without lifting a skinfold, 1% inserted the needle in the pad with an angle of less than 45° with both 8mmUTW and 12.7mmSTW syringes and 9% used inappropriate injection techniques because of the simulated environment but would correctly inject them-selves in standard conditions (Figure 3(d)).

Comparable usability results were recorded for the two syringes, 8mmUTW and 12.7mmSTW, and the two tested visc-osities, 10 cP and 30 cP, concerning the disposal of the used syringe task (Figure 3(b) and 3(c)). Overall, the proportion of use errors for naïve patients completing this task was twice higher than for others user groups (Figure S1D). 62% of these use errors were due to participants that attempted to recap the needle after completing the simulated injection. As the moderator intervened to stop the recapping to avoid needle stick injury after the first simulation, less use errors were recorded for the following simulations, regardless of the syr-inge type.

Overall, the usability of the 8mmUTW syringe is at least equivalent to the marketed 12.7mmSTW syringe, thus there is no difference in risk profile. Regardless of the viscosity, parti-cipants experienced less operational difficulties to remove the needle cover with the 8mmUTW syringe. Nonetheless, the use of 8mmUTW syringe filled with a high viscous solution of 30 cP improves the full dose delivery and reduces the operational difficulties encountered by all representative patient groups relative to the 12.7mmSTW syringe, in particular for patients with prior self-injecting experience or those suffering from hand-impairment.

3.5. Formative study 8mmUTW vs. 12.7mmSTW syringes – ease of use, acceptance and preference

As expected, the viscosity directly influenced both ease of pushing and acceptance of the force required to push the plunger rod for all participants.

Across all participants, the ease of pushing the plunger rod and the acceptance of the force required for it were similarly rated for both syringes, the 8mmUTW, and the 12.7mmSTW, filled with 10 cP solution (Figure 4(a) and 4(b) respectively). These two para-meters were rated better at 10 cP than at 30 cP for all participants (Figure 4(a) and 4(b)). According to fewer operational difficulties experienced by participants to deliver the full dose at 30 cP with the 8mmUTW syringe compared to the 12.7mmSTW syringe (Figure 3(c)), participants perceived the pushing of the plunger rod easier with the 8mmUTW syringe than the 12.7mmSTW syr-inge (overall median VAS score of 43% vs. 61%; Figure 4(a)). Consistently, the force required to push the plunger rod was more acceptable for the 8mmUTW syringe versus the 12.7mmSTW syringe (overall median VAS score of 42% vs. 61%; Figure 4(b)). Thus, at 30 cP, the 8mmUTW syringe provides an improved ease of pushing the plunger rod and a better accep-tance of the force required to push it, especially for naïve and hand-impaired patients. Likewise, 72% of the participants expressed a preference for the 8mmUTW syringe filled at 30cP because of the reduced injection force (57%), the shorter needle (36%) or the easier cap removal (7%). Regardless of the assessed viscosities, the 8mmUTW was preferred by 64% of the participants. At 10 cP the preference toward the 8mmUTW versus the 12.7mmSTW syringe, for the same reasons as mentioned above, was formulated to a lesser extend (56% versus 21%) (Table 3).

3.6. Validation study needlestick-prevention device (X225PLUS) equipped with the 8mmUTW syringe – usability

The usability of the needlestick-prevention device X225PLUS was assessed when equipped with the 8mmUTW syringe or with the 12.7mmSTW syringe, both filled with a high viscous solution (30 cP).

Overall, the usability results were similar between the X225PLUS device equipped with the 8mmUTW syringe or with the 12.7mmSTW syringe (Figure 5(a)).

Regardless of the user groups, the number of successes and use errors were not significantly different for participants using the X225PLUS-8mmUTW and the X225PLUS- 12.7mmSTW products for the task ‘Remove needle cover’ (98% vs. 94% of successes, respectively).

39% and 41% of use errors were recorded for the task ‘Perform the injection’ with the X225PLUS-8mmUTW and the X225PLUS-12.7mmSTW products, respectively. 39% of the recorded use errors were due to the unappropriated injection technique employed (insertion at 90° without skinfold). 19% of the naïve and 27% of the experienced patients did not per-form a skinfold while injecting at a 90° angle.

8 A. PAGER ET AL.

Page 10: Drug Delivery Systems, Pharmaceutical Systems - BD - User ......Published by Informa UK Limited, trading as Taylor & Francis Group View supplementary material Accepted author version

In comparison to the X225PLUS-12.7mmSTW product, the use of the X225PLUS-8mmUTW product improves the success of the ‘Deliver the full dose’ task (Figure 5(b)). No use error was recorded across all user groups with the X225PLUS-8mmUTW product against 4% of use errors with the X225PLUS- 12.7mmSTW product. Moreover, the operational difficulties encountered to deliver the full dose was reduced by 22% when participants were using the X225PLUS-8mmUTW pro-duct instead of the X225PLUS-12.7mmSTW product. Besides some operational difficulties, users were able to deliver 30 cP

solution and activate the needlestick-prevention product. Indeed, the use of the 8mmUTW does not impede the activa-tion of the needle guard, as 100% of the needle guards were correctly activated with both X225PLUS-8mmUTW and X225PLUS-12.7mmSTW products (Figure 5(b)).

No significant difference was observed for the task ‘Disposal of the used syringe,’ for which the participants experienced 92% of successes with the X225PLUS-8mmUTW product and 90% with the X225PLUS-12.7mmSTW product (Figure 5(b)).

Figure 4. Boxplot distribution of the VAS scores of the ease of pushing the plunger rod (a) and of the acceptance of the force required to push the plunger rod (b) for all participants of the formative study. Horizontal lines indicate median values, boxes indicate interquartile range, and whiskers indicate minimum and maximum values. VAS scales ranging from 0 = very easy/very acceptable to 100% = very difficult/very unacceptable.

Table 3. Willingness to adopt the X225 PLUS–8mmUTW and the X225 PLUS–12.7mmSTW products across all participants during the validation study. N number of answers; (%).* The total N number for the naïve patients is reduced when assessing the adoption of the X225PLUS–8mmUTW because some participants did not complete some of the simulations.** The Hand-impaired group is composed of Naïve and Experienced patients.

Syringe preference at 10 cP Syringe preference at 30 cP Overall syringe preference

User group 8mmUTW 12.7mmSTW No preference 8mmUTW 12.7mmSTW No preference 8mmUTW 12.7mmSTW No preference

Experienced patients (N = 13)

7 (54%) 4 (31%) 2 (15%) 12 (92%) 0 (0%) 1 (8%) 19 (73%) 4 (15%) 3 (12%)

Naïve patients (N = 16)

8 (50%) 4 (25%) 4 (25%) 11 (69%) 1 (6%) 4 (25%) 19 (59%) 5 (16%) 8 (25%)

Hand-impaired patients (N = 10)

7 (70%) 0 (0%) 3 (30%) 3 (40%) 4 (40%) 3 (30%) 10 (50%) 4 (20%) 6 (30%)

All participants (N = 39)

22 (56%) 8 (21%) 9 (23%) 28 (72%) 3 (8%) 8 (20%) 50 (64%) 11 (14%) 17 (22%)

EXPERT OPINION ON DRUG DELIVERY 9

Page 11: Drug Delivery Systems, Pharmaceutical Systems - BD - User ......Published by Informa UK Limited, trading as Taylor & Francis Group View supplementary material Accepted author version

3.7. Validation study needlestick-prevention device (X225PLUS) equipped with the 8mmUTW syringe – ease of use and acceptance

The participants found easier to push the plunger rod with the X225PLUS-8mmUTW product versus the X225PLUS- 12.7mmSTW product (41% vs. 15% of easy and very easy rating answers) (Figure 5(chealth care workers (59% vs. 13%) (Figure 5(c)). Similarly, all participants better accepted the force required to push the plunger rod with the X225PLUS- 8mmUTW product than with the X225PLUS-12.7mmSTW pro-duct (53% vs. 20% of acceptable and very acceptable rating answers) (Figure 5(d)).

Therefore, the use of the X225PLUS–8mmUTW product improves the successful delivery of the full dose of high viscous solution (30 cP), which is supported by better perceptions of the ease of pushing the plunger rod and the acceptance of the force required to push the plunger rod. Additionally, the increased proportion of the partici-pants willing to adopt the X225PLUS–8mmUTW product (86%) relative to the X225PLUS-12.7mmSTW product

(61%) confirmed the perceived ease of use and acceptance (Table 4).

Figure 5. Usability of the safety X225 PLUS device equipped either with the 8mmUTW or the 12.7mmSTW syringes during the validation study. Overall task completion rates (a) and per task assessed (b) for X225 PLUS–8mmUTW and X225 PLUS–12.7mmSTW devices. Rates of success, use error and success with operational difficulty (SOD) are indicated. N number of simulated injections; %. * 20% (for X225-12.7mmSTW) and 28% (for X225PLUS-8mmUTW) of the use errors were caused by syringe priming. Ease of pushing the plunger rod (c) and acceptance of the force required to push the plunger rod (d) for X225 PLUS–8mmUTW and X225 PLUS–12.7mmSTW products. N represents the number of answers rated at 5 (easy/acceptable) and 6 (very easy/very acceptable); (%).

Table 4. Willingness to adopt the X225 PLUS–8mmUTW and the X225 PLUS– 12.7mmSTW products across all participants during the validation study. N number of answers(%).

Willingness to adopt X225 PLUS – 8mmUTW

product at 30 cP

Willingness to adopt X225 PLUS –

12.7mmSTW product at 30 cP

User group YES NO YES NO

Health Care Workers (N = 16)

30 (94%) 2 (6%) 25 (78%) 7 (19%)

Caregivers (N = 16)

25 (81%) 6 (19%) 13 (41%) 19 (59%)

Experienced patients (N = 15)

23 (77%) 7 (23%) 18 (60%) 12 (40%)

Naïve patients (N = 15 or 16)*

28 (93%) 2 (7%) 21 (66%) 11 (34%)

Hand-impaired patients ** (N = 15)

24 (87%) 4 (13%) 14 (47%) 16 (53%)

All participants (N = 63)

106 (86%) 17 (14%) 77 (61%) 49 (39%)

*The total N number for the naïve patients is reduced when assessing the adoption of the X225PLUS–8mmUTW because some participants did not complete some of the simulations

**The Hand-impaired group is composed of Naïve and Experienced patients.

10 A. PAGER ET AL.

Page 12: Drug Delivery Systems, Pharmaceutical Systems - BD - User ......Published by Informa UK Limited, trading as Taylor & Francis Group View supplementary material Accepted author version

4. Discussion

The two human factors studies, presented in this manuscript, aimed to evaluate the usability of a new pre-fillable glass syringe equipped with a short 8mm and ultra thin-wall needle (BD Neopak™ XtraFlow™ 2.25 mL, 8mmUTW) in comparison to a marketed pre-fillable syringe (BD Neopak™ 2.25 mL, 12.7mmSTW) alone or combined with a needlestick- prevention device (BD UltraSafe Plus™ 2.25 mL Passive needle guard, X225PLUS) in order to improve the subcutaneous administration of high volume (2 mL) and high viscous solu-tions (30 cP). These human factors studies were conducted per Food and Drug Administration (FDA) guidance and the International Electrotechnical Commission (IEC) 62366–1, and the spectrum of intended user groups were evaluated.

The full delivery of the 30cP viscous solution is improved with the 8mmUTW syringe versus the 12.7mmSTW syringe, resulting in fewer operational difficulties and more successes for all participants. The same trend was observed for the 8mmUTW assembled with the X225PLUS needlestick- prevention device, more successes were recorded to deliver the full dose in comparison to the X225PLUS–12.7mmSTW product. However, the reduced rate of use errors recorded with the X225PLUS–8mmUTW was most probably attributed to the gained experience during previous simulations done with X225PLUS–12.7mmSTW products. Indeed, after 5 simula-tions with the X225PLUS-12.7mmSTW product, to assess the usability of this product, the users were able to learn how to use this needlestick-prevention device before simulating the injections with the X225PLUS-8mmUTW product. Although this bias must be considered in the interpretation, the user experience is improved, with an easier delivery and a more acceptable force required to push on the plunger rod with the X225PLUS-8mmUTW than the X225PLUS-12.7mm product. A significant difference between the rate of use errors during the validation (0% to 4% of use errors) and the formative (16% to 18% of use errors) studies was observed. During the for-mative study, the use errors caused by a priming prior the simulations representing 83% and 71% of the use errors for the 8mmUTW and the 12.7mmSTW were recorded during the assessment of the task ‘Deliver the full dose’ as the users failed to deliver the full dose due to the expelled volume expelled before the simulation creating a potential risk of underdosing. These use errors were recorded during the evaluation of the task ‘Perform injection’ in the validation study as occurring before the needle insertion. Despite the difference in terms of task assessment between the two studies, these use errors were not related to the performance of the tested products but were due to an inappropriate use by the participants.

The task ‘Perform injection’ was considered slightly differ-ently between the two studies. It was considered as essential in the formative study and as critical in the validation study. Usually, the instructions in terms of injection technique are précised into the final combination product and therefore are under pharmaceutical companies’ responsibility. However, during the validation study, the injection technique for the X225PLUS-8mmUTW product was restricted (insertion at a 90°

angle within a skinfold) and communicated through the IFU to the participants due to the encumbrance of the X225PLUS product related to the shorter needle length, 8 mm, which may increase the risk of intradermal administration if inserted at a 45° angle. Therefore, to assess this mitigation measure the used injection techniques were recorded and the task ‘Perform injection’ was defined as critical during the validation study. Nevertheless, this task was assessed in terms of usability and root cause evaluation during both studies, regardless of its classification.

In terms of safety, regardless of the syringe used, the for-mative study pointed a hazardous behavior. Indeed, some users attempted to recap the needle once the simulations performed. Although the moderator stopped patients to pre-vent the recapping of the needle and thus needlestick injury, among them 62.5% were experienced patients in using pre-filled syringes (with or without hand impairment) and 37.5% were naïve. Participants located in Boston (US) explained their action because they did not have access to a sharp container at home unlike most of the participants located in London (UK). Thus, the availability of a sharp containers might be critical to change users’ habit toward needle recapping and prevent against needlestick injuries.

Besides the safety and efficacy of drug delivery devices, patients’ perception should also be considered when select-ing a drug delivery solution to improve injection experience and potentially their treatment adherence [44–46]. The shorter needle combined with an optimized inner diameter (UTW technology) positively impacts the patients’ perception of the force required to deliver the full dose when using a shorter UTW needle during simulated subcutaneous injec-tions especially for high viscosity solution (30 cP) in both studies. The ability to hold a device may be strongly affected in case of hand and/or fingers deformities and pain observed for some patients suffering from rheumatoid arthritis [45]. Our studies revealed that hand-impaired patients also perceived the 8mmUTW syringe alone and combined with the X225PLUS product as easier to push and better accepted the force required to push on the plunger rod than with the 12.7mmSTW syringe. During the validation study, two participants did not perform any simu-lation with the X225PLUS-8mmUTW product, as they felt some discomfort after the simulations done with the X225PLUS-12.7mmSTW product. The users’ feedbacks on the ease of pushing the plunger rod and the acceptance of the force required to push the plunger rod may support the interest in using the X225PLUS-8mmUTW product to protect users from fatigue due to the injection of high volumes and/ or high viscous solutions. Overall, participants expressed a preference in terms of ease of use and acceptance for the 8mmUTW syringe relative to the 12.7mmSTW syringe due to the perceived reduction of the force required to deliver the solution.

Moreover, in the formative study, the use of an incorrect injection technique observed supports the benefit of using 8 mm needles versus 12.7 mm needles, in terms of chance to reach the correct tissue. Indeed, 35% of participants,

EXPERT OPINION ON DRUG DELIVERY 11

Page 13: Drug Delivery Systems, Pharmaceutical Systems - BD - User ......Published by Informa UK Limited, trading as Taylor & Francis Group View supplementary material Accepted author version

including experienced ones, simulated the injection at 90° angle without lifted a skinfold. Based on the described math-ematical model the unintended intramuscular injections by reducing the needle length from 12.7 mm to 8 mm when using a non-recommended injection technique, insertion at a 90° angle without raising a skinfold was estimated to be reduced by 59% and 77% in the thigh or abdomen, respectively.

In the formative study, the proportion of participants employ-ing an incorrect injection technique might be overestimated for naïve patients as they were not were aware of the subcutaneous injection technique in contrast with experienced patients, includ-ing those with hand-impairment. Moreover, although it is impor-tant to note that our data were obtained from a simulated study and with a small sample size and might not reflect real injections, the analysis of a survey addressed to adults with diabetes, using 8 mm or longer needles, showed that only 64% of patients lifted a skinfold, among them 75% did it properly, and less than 50% of them released correctly the skinfold [47]. Therefore, preliminary results obtained in the formative study, as well as published data, highlight that patients’ population often does not know or use the appropriate injection techniques to deliver their treatment in the targeted subcutaneous tissue. In real conditions, these use errors would have increased the risk of injecting in the intramuscular tissue, more with the 12.7mmSTW syringe than with the 8mmUTW syringe, potentially affecting the drug absorption, espe-cially, in case of using an inappropriate injection technique at 90° without raising a skinfold [7,33–35]. In addition to reducing acci-dental intramuscular administration, the use of the 8mmUTW syringe does not increase the risk of targeting the dermis instead of the subcutaneous tissue. A previous study estimated that the injection at a 90° angle without lifting a skinfold with short needles (4 mm to 8 mm) did not result in accidental intradermal injection, based on skin thickness measurements of the abdomen and the upper arm in adults with diabetes [48]. The data presented in this article confirm that the 8 mm needle is not more likely to induce dermis administration than the 12.7 mm needle when injecting at a 90° as well as at a 45° angle without raising a skinfold regardless of the injection sites (abdomen, arm, thigh, and buttocks).

Limitations of these studies are inherent to the nature of simulated use studies, meaning that participants may perform differently their injections when observed during a simulation study than in their actual use. All subjective feedbacks were recorded according to participants’ point of view including feelings, perceptions, and concerns at the time of the simu-lated testing. Therefore, clinical studies of the full combination product are required to confirm the results obtained in those human factors studies and go further by evaluating pain and depot localization of the drugs.

5. Conclusion

The new 8mmUTW syringe combined or not to a needlestick- prevention device (X225PLUS) improves the manual delivery of high volumes (2 mL) and viscous solutions (up to 30 cP) for the targeted end-users, especially for patients with prior experience in self-injecting chronic treatments, as well as for patients

suffering from hand impairment. The 8mmUTW syringe presents a better usability profile for manual subcutaneous injections in comparison to the 12.7mmSTW syringe by increasing the rate of success and decreasing the occurrence of operational difficulties. Many users perceived the 8mmUTW syringe alone or assembled with the X225PLUS device as easier to push on the plunger rod to deliver the solution and they better accepted the force required to do so in comparison to the 12.7mmSTW syringe. Consistently, most of the users expressed a preference toward the 8mmUTW syringe and were willing to adopt it in combination with the X225PLUS device against the 12.7mmSTW syringe. Our formative study highlighted the use of inappropriate injection technique at a 90° angle without raising a skinfold in 35% of the simulated injections, even for self-injecting experienced patients. In doing so, the needle length reduction could increase the chances of targeting the right tissue (subcutaneous) by 2 to 6 times, thus ensuring an optimal efficacy of chronic injectable therapies while avoiding potential adverse effects for the patients. The 8mmUTW syringe provides an acceptable and easy to use alternative to the current 12.7 mm needle syringes for the delivery of viscous solutions. Additionally, the shorter needle length tends to reduce the injection-related anxiety felt by patients. Altogether, these results demonstrate that the user experience is improved with the BD Neopak™ XtraFlow™ syringe (8mmUTW) which might contribute to enhance treatment compliance in chronic care.

Author contributionsA Pager, K Guerrero, T Tzvetkova-Chevolleau contributed to the concep-tion, design and analysis of the human factors testing. D Morel developed the mathematical model data analysis methods. A Combedazou wrote the paper with the inputs from all authors. All authors reviewed the manu-script. C Frolet and S Glezer supervised the project.

Funding

This work was sponsored by BD Medical - Pharmaceutical Systems.

Declaration of interestAll authors are employees of BD Medical Pharmaceutical Systems. The authors have no other relevant affiliations or financial involvement with any organiza-tion or entity with a financial interest in or financial conflict with the subject matter or materials discussed in the manuscript apart from those disclosed.

Reviewer disclosuresPeer reviewers on this manuscript have received an honorarium from Expert Opinion on Drug Delivery for their review work, but have no other relevant financial relationships to disclose.

ORCIDAnne Combedazou http://orcid.org/0000-0002-0046-5057

References

Papers of special note have been highlighted as either of interest (•) or of considerable interest (••) to readers.

12 A. PAGER ET AL.

Page 14: Drug Delivery Systems, Pharmaceutical Systems - BD - User ......Published by Informa UK Limited, trading as Taylor & Francis Group View supplementary material Accepted author version

1. Viola M, Sequeira J, Seiça R, et al. Subcutaneous delivery of mono-clonal antibodies: how do we get there? J Control Release. 2018;286:301–314.

2. Usach I, Martinez R, Festini T, et al. Subcutaneous injection of drugs: literature review of factors influencing pain sensation at the injection site. Adv Ther. 2019;36(11):2986–2996.

3. Collins DS, Kourtis LC, Thyagarajapuram NR, et al. Optimizing the bioavailability of subcutaneously administered biotherapeutics through mechanochemical drivers. Pharm Res. 2017;34 (10):2000–2011.

4. Zharkov A, Barton B, Heinzmann D, et al., Development pathways for subcutaneous formulations of biologics versus biosimilar development. Expert Review of Precision Medicine and Drug Development, 2019: p. 1–8.

5. Bolge SC, Goren A, Tandon N. Reasons for discontinuation of sub-cutaneous biologic therapy in the treatment of rheumatoid arthritis: a patient perspective. Patient Prefer Adherence. 2015;9:121–131.

6. Zijlstra E, Jahnke J, Fischer A, et al. Impact of injection speed, volume, and site on pain sensation. J Diabetes Sci Technol. 2018;12(1):163–168.

7. Jin JF, Zhu -L-L, Chen M, et al. The optimal choice of medication administration route regarding intravenous, intramuscular, and subcutaneous injection. Patient Prefer Adherence. 2015;9:923–942.

8. Schweighofer CD, Wendtner CM. First-line treatment of chronic lymphocytic leukemia: role of alemtuzumab. Onco Targets Ther. 2010;3:53–67.

9. Turner AP, Williams RM, Sloan AP, et al., Injection anxiety remains a long-term barrier to medication adherence in multiple sclerosis. Rehabil Psychol. 2009;54(1): 116–121.

• The study demonstrates that anxiety has been associated with poor treatment adherence in multiple sclerosis.

10. Zambanini A, Newson RB, Maisey M, et al. Injection related anxiety in insulin-treated diabetes. Diabetes Res Clin Pract. 1999;46 (3):239–246.

11. Garidel P, Kuhn AB, Schäfer LV, et al. High-concentration protein formulations: how high is high? Eur J Pharm Biopharm. 2017;119:353–360.

12. McLennan DN, Porter CJ, Charman SA. Subcutaneous drug delivery and the role of the lymphatics. Drug Discov Today Technol. 2005;2 (1):89–96.

13. Bittner B, Richter W, Schmidt J. Subcutaneous administration of biotherapeutics: an overview of current challenges and opportunities. BioDrugs. 2018;32(5):425–440.

• This review presents the new challenges for upcoming thera-pies intended to be subcutaneously and self-administered especially in rheumatoid arthritis, primary immunodeficien-cies, and multiple sclerosis

14. Wago KJ, Skarsvåg TI, Lundbom JS, et al. The importance of needle gauge for pain during injection of lidocaine. J Plast Surg Hand Surg. 2016;50(2):115–118.

15. Jorgensen JT, Rømsing J, Rasmussen M, et al. Pain assessment of subcutaneous injections. Ann Pharmacother. 1996;30(7–8):729–732.

16. Mathaes R, Koulov A, Joerg S, et al. Subcutaneous injection volume of biopharmaceuticals-pushing the boundaries. J Pharm Sci. 2016;105(8):2255–2259.

17. Berteau C, Filipe-Santos O, Wang T, et al. Evaluation of the impact of viscosity, injection volume, and injection flow rate on subcuta-neous injection tolerance. Med Devices (Auckl). 2015;8:473–484.

•• This study demonstrates that injections of large volumes up to 3mL and viscous solutions up to 15-20 cP are feasibles and well tolerated without pain.

18. Alam M, Geisler A, Sadhwani D, et al. Effect of needle size on pain perception in patients treated with botulinum toxin type a injections: a randomized clinical trial. JAMA Dermatol. 2015;151 (11):1194–1199.

19. Heise T, Nosek L, Dellweg S, et al. Impact of injection speed and volume on perceived pain during subcutaneous injections into the

abdomen and thigh: a single-centre, randomized controlled trial. Diabetes Obes Metab. 2014;16(10):971–976.

20. Juul KA, Bengtsson H, Eyving B, et al. Influence of hypodermic needle dimensions on subcutaneous injection delivery–a pig study of injection deposition evaluated by CT scanning, histology, and backflow. Skin Res Technol. 2012;18(4):447–455.

21. Watt RP, Khatri H, Dibble ARG. Injectability as a function of viscosity and dosing materials for subcutaneous administration. Int J Pharm. 2019;554:376–386.

22. Karges B, Boehm BO, Karges W. Early hypoglycaemia after acciden-tal intramuscular injection of insulin glargine. Diabet Med. 2005;22 (10):1444–1445.

23. Vaag A, Handberg A, Lauritzen M, et al. Variation in absorption of NPH insulin due to intramuscular injection. Diabetes Care. 1990;13 (1):74–76.

24. Vaag A, Pedersen KD, Lauritzen M, et al. Intramuscular versus subcutaneous injection of unmodified insulin: consequences for blood glucose control in patients with type 1 diabetes mellitus. Diabet Med. 1990;7(4):335–342.

25. Frid A, Ostman J, Linde B. Hypoglycemia risk during exercise after intramuscular injection of insulin in thigh in IDDM. Diabetes Care. 1990;13(5):473–477.

26. Guo S, Bozkaya D, Ward A, et al. Treating relapsing multiple sclero-sis with subcutaneous versus intramuscular interferon-beta-1a: modelling the clinical and economic implications. Pharmacoeconomics. 2009;27(1):39–53.

27. Calabrese M, Bernardi V, Atzori M, et al. Effect of disease-modifying drugs on cortical lesions and atrophy in relapsing-remitting multi-ple sclerosis. Mult Scler. 2012;18(4):418–424.

28. Arthur V, Jubb R, Homer D. A study of parenteral use of metho-trexate in rheumatic conditions. J Clin Nurs. 2002;11(2):256–263.

29. Klein CG, Cicinnati V, Schmidt H, et al. Compliance and tolerability of subcutaneous hepatitis B immunoglobulin self-administration in liver transplant patients: a prospective, observational, multicenter study. Ann Transplant. 2013;18:677–684.

30. Frid AH, Kreugel G, Grassi G, et al. New insulin delivery recommendations. Mayo Clin Proc. 2016;91(9):1231–1255.

31. Frid A, Hirsch L, Gaspar R, et al. New injection recommendations for patients with diabetes. Diabetes Metab. 2010;36(Suppl 2):S3–18.

32. Strauss K, Gols HD, Hannet I, et al. A pan-European epidemiologic study of insulin injection technique in patients with diabetes. Pract Diabetes Int. 2002;19(3):71–76.

33. Gibney MA, Arce CH, Byron KJ, et al., Skin and subcutaneous adipose layer thickness in adults with diabetes at sites used for insulin injections: implications for needle length recommendations. Curr Med Res Opin. 2010;26(6): 1519–1530.

•• This study provides robust data for skin thickness and subcu-taneous adipose layer thickness in order to improve the nee-dle legntnh choice.

34. Hirsch L, Byron K, Gibney M. Intramuscular risk at insulin injection sites–measurement of the distance from skin to muscle and ratio-nale for shorter-length needles for subcutaneous insulin therapy. Diabetes Technol Ther. 2014;16(12):867–873.

35. Tubiana-Rufi N, Belarbi N, Du Pasquier-Fediaevsky L, et al., Short needles (8 mm) reduce the risk of intramuscular injections in children with type 1 diabetes. Diabetes Care. 1999;22(10): 1621–1625.

•• This study shows that 8mm needles significantly reduce the risk of accidental intramuscular insulin administration in dia-betic children and adolescents.

36. Tsapatsaris NP. Low-dose heparin. A cause of hematoma of rectus abdominis. Arch Intern Med. 1991;151(3):597–599.

37. Thomas BL. Heparin-induced hematomas. Arch Intern Med. 1992;152(1):204.

38. Kim L, Nevis I, Potts R, et al. Patients on subcutaneous allergen immunotherapy are at risk of intramuscular injections. Allergy Asthma Clin Immunol. 2014;10(1):22.

EXPERT OPINION ON DRUG DELIVERY 13

Page 15: Drug Delivery Systems, Pharmaceutical Systems - BD - User ......Published by Informa UK Limited, trading as Taylor & Francis Group View supplementary material Accepted author version

39. Baker MP, Reynolds HM, Lumicisi B, et al. Immunogenicity of pro-tein therapeutics: the key causes, consequences and challenges. Self Nonself. 2010;1(4):314–322.

40. Arendt-Nielsen L, Egekvist H, Bjerring P. Pain following controlled cutaneous insertion of needles with different diameters. Somatosens Mot Res. 2006;23(1–2):37–43.

41. Kreugel G, Beijer H, Kerstens MN, et al. Influence of needle size for subcutaneous insulin administration on metabolic control and patient acceptance. Eur Diabetes Nurs. 2007;4(2):51–55.

42. Hudry C, Lebrun A, Moura B, et al. Evaluation of usability and acceptance of a new autoinjector intended for methotrexate sub-cutaneous self-administration in the management of rheumatoid arthritis. Rheumatol Ther. 2017;4(1):183–194.

43. Duruoz MT, Poiraudeau S, Fermanian J, et al. Development and validation of a rheumatoid hand functional disability scale that assesses functional handicap. J Rheumatol. 1996;23(7):1167–1172.

44. Koren MJ, Lundqvist P, Bolognese M, et al. Anti-PCSK9 monother-apy for hypercholesterolemia: the MENDEL-2 randomized,

controlled phase III clinical trial of evolocumab. J Am Coll Cardiol. 2014;63(23):2531–2540.

45. Keininger D, Coteur G. Assessment of self-injection experience in patients with rheumatoid arthritis: psychometric validation of the Self-Injection Assessment Questionnaire (SIAQ). Health Qual Life Outcomes. 2011;9(1):2.

46. Schiff M, Saunderson S, Mountian I, et al. Chronic disease and self-injection: ethnographic investigations into the patient experi-ence during treatment. Rheumatol Ther. 2017;4(2):445–463.

47. Frid AH, Hirsch LJ, Menchior AR, et al., Worldwide injection techni-que questionnaire study: population parameters and injection practices. Mayo Clin Proc. 2016;91(9): 1212–1223.

•• This large survey focuses on patient characteristics and prac-tical aspects of injection technique.

48. Sim KH, Hwang MS, Kim SY, et al. The appropriateness of the length of insulin needles based on determination of skin and subcutaneous fat thickness in the abdomen and upper arm in patients with type 2 diabetes. Diabetes Metab J. 2014;38(2):120–133.

14 A. PAGER ET AL.