What are the dose quantities in CT ?
IAEA/RCA Kampala
Kampala
![Page 2: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/2.jpg)
12345678
12
3
45
6
7
8
CT Principles
Kampala
![Page 3: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/3.jpg)
Kampala
![Page 4: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/4.jpg)
Reminder : CT terminology • Axial scanning
– 1 rotation of beam followed by stepwise couch movement
– Increment – distance moved by couch• Helical/spiral scanning
– Continuous table movement & beam rotation– Pitch – table movement relative to nominal
slice width
Kampala
![Page 5: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/5.jpg)
CT Slice
Kampala
![Page 6: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/6.jpg)
CT Slice
Kampala
![Page 7: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/7.jpg)
CT Slice
Kampala
![Page 8: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/8.jpg)
Dose Profile
Kampala
![Page 9: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/9.jpg)
Dose profile for a series of slices
Kampala
![Page 10: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/10.jpg)
Dose Profiles in Phantoms / PatientsEffects of scatter
Broadened profilePeak kerma dependent on beam widthLong scatter tails for wide beams
Kampala
![Page 11: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/11.jpg)
Which Dosemeter ?
Kampala
![Page 12: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/12.jpg)
Integrated Profile
Kampala
![Page 13: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/13.jpg)
Integrated Profile
Kampala
![Page 14: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/14.jpg)
CTDose Index
• Measure of dose for a single axial slice (mGy)– no account of Helical/MDCT gaps or overlaps
• CTDI = area under dose profile (mGy.mm) nominal slice width, T (mm)
Kampala
kerma
T
CTDI(mGy
)
Z-axis (mm)
CTDI = height of rectangle if area
under profile was all contained
within nominal slice width
![Page 15: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/15.jpg)
CTDI100 – What is it ?• Analagous to
Multiple Scan Average Dose.
• Originally defined for single slice scanners.
Kampala
![Page 16: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/16.jpg)
• In air : CTDI100,Air
• In air : CTDI100,Air
• Sometimes normalise to 1 mAs:
• nCTDI100,Air
• Just like in radiography!
Kampala
![Page 17: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/17.jpg)
In Perspex / PMMA
Kampala
![Page 18: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/18.jpg)
CT dosimetry in PerspexThe computed tomography air kerma index is also measured inside PMMA head and body phantoms
The quantity CTDI100,c is measured at the centre of the standard CT dosimetry phantom and CTDI100,p is the average of values measured at four positions around the periphery of the same phantom.
Kampala
![Page 19: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/19.jpg)
CTDI100
• CTDI can be measured in air or in phantoms
• CTDI in air can be used as a measure of tube output
• For dosimetry purposes CTDI is measured in perspex phantoms• 16cm diameter ‘head’• 32cm diameter ‘body’
Kampala
![Page 20: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/20.jpg)
Patient Size.• CTDI is a measure of scanner output.• CTDI does take patient size into account• CTDI is NOT patient dose• Important for paediatrics, small adults
and possibly bariatrics.• You can correct to patient size
– To be mentioned later.
Kampala
![Page 21: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/21.jpg)
Phantoms for CT dosimetry
• Cylindrical PMMA phantoms with holes for pencil chamber
• 32 cm body phantom• 16 cm head phantom
Kampala
![Page 22: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/22.jpg)
Centre Periphery – 1cm depth (mean of 4 positions)
C
P1
P2
P3
P4
32 cm
‘Head’ ‘Body’
16 cm
140 mm
CTDI Phantoms
Kampala
![Page 23: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/23.jpg)
40
‘Head’
‘Body’
40
40 40
40
20
20 20
20
10
Periphery : centre 1:1 Periphery : centre 2:1
Typical CTDI100 ValuesCTDI100,p values quoted in mGy
Kampala
![Page 24: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/24.jpg)
CTDIW• Weighted average CTDI: represents the
average dose for contiguous irradiation • CTDIw = 1/3 CTDIcentre + 2/3 CTDIperiphery • nCTDIw = CTDIw/mAs
CTDIC CTDIP
Kampala
![Page 25: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/25.jpg)
CTDIvol
Pitch = 1 CTDIvol = CTDIw
Pitch = 2 CTDIvol = CTDIw
2
Pitch = 0.5 CTDIvol = 2 x CTDIw
Kampala
![Page 26: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/26.jpg)
CTDIvol
• CTDIW is the average dose in the scan plane
• Volume CTDI considers non-contiguous scanning
• The CTDIvol can be considered to be the average CTDIW in the scan volume
• CTDIvol = CTDIW Pitch
Kampala
![Page 27: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/27.jpg)
Dose Length Product
• CTDIvol is a measure of absorbed dose, energy absorbed per unit mass
• Radiation risk is based on total energy absorbed
• Estimate using dose length product (DLP)• DLP = CTDIvol x scan length• Units are mGy cm
Kampala
![Page 28: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/28.jpg)
Dose Quantities in CT
Kampala
![Page 29: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/29.jpg)
Dose Summary Sheet - Commercial CT Scanner
Huda W , Mettler F A Radiology 2011;258:236-242
Kampala
![Page 30: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/30.jpg)
CTDI on the Console
• The CTDI value displayed will depend on the phantom that it is based on.
• Body : 32 cm; Head 16 cm.• CTDI for 16cm reference twice that of
CTDI for 32cm reference.• Paediatric protocol MAY reference to 16
cm phantom• Implementation manufacturer
dependent.Kampala
![Page 31: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/31.jpg)
CTDI MEASUREMENT IN PRACTICE
Kampala
![Page 32: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/32.jpg)
Equipment for CT measurements
• CT ionization chamber and electrometer calibrated for CT beam qualities
• Chamber support stand• CT head & body phantoms
Kampala
![Page 33: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/33.jpg)
Methodology for CT measurements in air
• Position chamber in stand so beyond couch
• Adjust couch position so chamber along axis of scanner and beam centred at central point of chamber
• Record dose measurements for single axial scan at selected parameters (no couch movement)
• Record all data
Kampala
![Page 34: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala](https://reader036.vdocuments.us/reader036/viewer/2022082215/5a4d1b6a7f8b9ab0599b2b31/html5/thumbnails/34.jpg)
Methodology for CT measurements in phantoms
• Position phantom on couch (use head rest if appropriate)
• Position chamber in central position in phantom• Adjust couch position so chamber centred as
before• Record dose measurements for single axial
scan at selected parameters (no couch movement)
• Repeat for peripheral positions (north, south, east & west)
Kampala