User manual
User manual• Viofor JPS Classic• Viofor JPS Standard• Magnetic & Light Therapy
Med & Life 2005
PART A
Prior to use of Viofor JPS System apparatus, please read the present manual. The User Manual for Viofor JPS System Apparatus is not meant as the main source of medical information. Any information necessary for optimal use of the apparatus can be found in professional publications and at Symposiums organized by Med & Life. It is also advisable to seek medical consultations available on Med & Life Medical Infoline, share information and experience with other MD’s and to visit the site: www.medandlife.com
Consult instructions for use
NOTICE: please read the enclosed documentation!
Do not use during pregnancy (see: contraindications)!
Do not use if wearing electronic implants (see: contraindications)!
Beam of intensive red or/and infrared radiation! Protect your eyes!Do not look at the light!Direct eye contact with the applicator can cause risk as in case of class 3A laser.
Fragments marked in italics present viofortherapy theory elements and are mainly for doctor’s use. You do not need to know these parts of the manual in order to be able to operate the apparatus.
Application of Viofor JPS System
3
User Manual for Viofor JPS System has been designed for a wide range of private users as well as health care professionals, i.e. medical doctors of various specializations (including dentists and veterinarians) as well as physiotherapists and rehabilitation special-ists, who use or are interested in application of variable magnetic fields in medicine.
In recent years, medical world has been intensively seeking new solutions in diagnos-tics and therapy, especially of noninvasive character, as well as treatment methods not sufficiently implemented until now.
Dynamic development observed in math-ematics, physics and technology (especially electronics) has resulted in a new quality
of physical medicine. Thanks to modern technologies applied in medical devices, the application and effectiveness of variable magnetic fields in medicine has also beenexpanded.
Viofor JPS System (further referred to as Viofor JPS) is a set consisting of: a control-ler with basic and/or additional accessories. Unique design solutions enable a wide ap-plication of therapeutic properties of variable magnetic fields. Viofor JPS provides a selec-tion of physical parameters of variable mag-netic fields used in:1. magnetostimulation in JPS System 2. magnetostimulation in JPS system paired
with light energy (optional)
Application of Viofor JPS System
1 CHARACTERISTICS OF VIOFOR JPS
1.1 Usage
Viofor JPS System is used for the following types of therapy:- Magnetostimulation of the whole body
or its parts- Magnetostimulation paired
with light energy of selected body parts
Viofor JPS System can be used:- in medical and physiotherapy clinics- in dentistry- in rehabilitation- in sport medicine- in veterinary medicine- at home as recommended by a physician- at home prophylaxis
2.1 Magnetostimulation2.1.1 Characteristics of magnetostimulation
Magnetic fields in magnetostimulationgenerated by Viofor JPS system are impulse fields of complex impulse shapes and signalstructure of multi-peak frequency spectrum. Basic frequencies are within 180 – 195 Hz, frequencies of impulse packets are within 12,5 to 29 Hz, packet groups within 2,8 – 7,6 Hz,
and series 0,08 – 0,3 Hz, whereas median induction of generated fields are significantlylower than those used in magnetotherapy.
Viofor JPS generates variable magnetic fields, which cause changes in the organismsimilar to those caused by physical activity (physical strain, physical activity); therefore they support biological regeneration. Both mental and physical regeneration is tightly connected with exercise. Variable magnetic field plays a similar beneficial role. Magneto-
2 TYPES OF THERAPY
INTRODUCTION
Application of Viofor JPS System
4
stimulation stimulates in the organism electro-motor forces of such properties, which enable homeostasis, and in case of its disturbance stimulate return to a normal state.
Biological effect of variable magnetic fieldgenerated by Viofor JPS is the result of:1. electrodynamic effect of such field on ion
currents in the organism2. magnetomechanic effect of the magnetic
field on particles with uncompensatedmagnetic spin
3. ionic cyclotron resonance of cations and anions in the body fluids.Body fluids i.e. blood, the lymph and inter-
cellular fluid constitute so-called second modeconductors of the ionic character of electrical conductivity.
The essence of the Viofor JPS system is the optimal choice of two different electrical (and thus magnetic) signals as well as their mutual combination, in order to achieve the aforementioned three fundamental mecha-nisms of variable magnetic field absorptionby the living matter. ELF (Extremely Low Fre-quency) electromagnetic radiation may affect an organism in various ways. In magnetostim-ulation, due to extremely low value of magnetic induction, the most important are bio-electric, biochemical and bioenergetic effect.1. The bioelectric effect causes normaliza-
tion in the potential of the cell membrane. In pathological cases, the potential drops due to penetration through the cell mem-brane to the cell interior of positive ions, e.g. Na+. In order to reverse the process, the cell needs energy, which can be ob-tained from ATP hydrolysis.
2. The biochemical effect is based on an increased enzymatic activity as well as reduced processes related with ATP.
3. The bioenergetic effect is a factor that stimulates the nutrition and growth of cells and regulates intercellular processes lead-ing to regeneration of the organism. The main biological effects of magneto-
stimulation mechanisms are actions:- regenerative- vasodilatatory- angiogenetic
- stabilizing cell membranes- anti-inflammatory- relaxative- anti-spastic- analgesic
Analgesic action takes place mainly due to an increase in the secretion of endo-genous opiates from the ß-endorphin group, a substance responsible for the increase in pain threshold. The effect of analgesic action occurs not only during exposure to magnetic fields, but is also observed after exposure isdiscontinued. This determines the biological hysteresis of magnetic field effect. Preciselysuch a mechanism of analgesic action on the internal opioid system confirms an an-tagonist of opiates blocking this effect (Naloxone).
The basis for regeneration effect is mainly the intensification of the processesof oxygen utilization and tissue respira-tion as a result of intensified diffusion aswell as oxygen capture by hemoglobin and cytochrome. The increased capture of oxy-gen is associated with a stimulation of the processes of tissue respiration, DNA syn-thesis and acceleration of the mitotic cycle. Thanks to this, magnetic fields treat trophiclesions of the shank, accelerate the heal-ing of ulcerous niches, improve the cellular metabolism of the central nervous system and accelerate regeneration of damaged peripheral nerves.
Very weak currents generated due to in-teraction on piezoelectric substances e.g. collagen, stimulate the activity of osteoblasts. It is aided by intensification of vasodilatatoryeffects. It has also been observed that a mag-netic field inhibits demineralization processesof the bones and, at the same time, increases their mineralization – which is measured by densitometry.
Anti-inflammatory effect is connected with the stimulation of formation of c-AMP and E prostaglandin. Prostaglandin influ-ences the accumulation of c-AMP, which decreases the secretion of inflammationmediators from neutrophiles, basophiles, mastocytes and lymphocytes.
Application of Viofor JPS System
5
In case of the nervous system, magnetic fields also improve the interneuron conductivityand modulation of neuron activity as well as the pineal gland’s twenty-four hour rhythms of mela-tonin secretion, which is called” the scavenger of free radicals” responsible, among others, for the aging process.
The variable magnetic fields generatedby the Viofor JPS clearly modify the ampli-tude value of alpha and theta wave rhythms in proper directions, characteristic for the physiological state of relaxation - objectively measurable and indicated by value of the de-mobilization of the autonomous system. The results of psychophysical testing, as a part of psychotherapy of emotionally immature sub-jects point to magnetostimulation as a method of choice for these people all the more due to the fact that this is a non-invasive method ap-plied equally in medicine as in psychology.
2.1.2 Indications for magnetostimulationIndications of magnetostimulation as
a supplementary therapy (therapy of choice) include:
I. Analgesic action1. Osseous-joint system:
- degeneration of the spine and osseous-joint system of the upper and lower limbs
- rheumatoid inflammation of the joints- stiffening inflammation (rigor)
of the spinal joints- joint injuries
2. soft tissue:- extrajoint rheumatism - so-called
fibromyalgia- traumas of soft tissue- conditions after injury (cutting) of tissue
continuity (including post-surgical)3. migraines
II. Regeneration action1. osseous system:
- osteoporosis, particularly post-meno-pausal (concerns prophylaxis as well as therapy)
- Sudeck syndrome
- conditions after bone fracture and breakage
- conditions after operations on the osseous system
- conditions after operations in which anastomosis has been carried out with screws or nails
- pseudo joints2. soft tissue:
- conditions after injury with or without damage to skin continuity
- burns- bedsores
III. Action improving peripheral circulation1. improper local blood flow in the limbs:
- in diabetic angiopathy- in atherosclerotic angiopathy - in crural ulceration
2. accelerated absorption of haematomas3. decreased lymphatic edema/swelling
IV. Anti-spastic action- conditions after apoplexy- degenerative diseases of the nervous
system, accompanied by spasticity
V. Relaxative action- neuroses- sleep disorders- prophylactic calming action,
to counteract everyday stress- stress in the common meaning of the
word, and particularly after long-term psychic/mental tension
- concentration disorders
VI. Action decreasing oxygen deficit througheffect on the processes of:- aerobic tissue respiration- anaerobic tissue respiration
VII. Magnetostimulation in dentistry:- Dental treatment, e.g.: biological pulp
treatment, pulpitis treatment, complica-tions related to pulp gangrene or para-dontitis treatment, periodontitis, compli-cations following endodontic treatment
- Paradontitis treatment
Application of Viofor JPS System
6
(anti-inflammatory treatment, fightingpain, tissue regeneration, treatment of post-surgical scarring)
- Diseases of mucous membrane of oral cavity, diseases of small salivary glands, gingivitis
- After surgery (24 hours), e.g.: after tooth extraction, after gouging following apicectomy, in implant therapy,
- Complications after dental surgery (e.g. infected dry socket)
- Complications after anesthesia, nerve damage
- Orthodontic treatment (during treatment with permanent devices)
- Prosthetic treatment (after prosthesis application)
2.1.3 Contraindications to magnetostimulation
Contraindications apply to magnetostim-ulation as a method, and not the Viofor JPS apparatus. Emitted fields are a contraindication in case of an existing disease – whereas they are not a disease-developing factor. The diseases mentioned in contraindications are caused by different factors. At present, there is no information in literature regarding contraindications to magnetostimulation. However, in view of the comparatively short period of existence of magnetostimulation in rela-tion to magnetotherapy, the contraindica-tions should be the same as in case of magnetotherapy, i.e.:- pregnancy- active neoplastic disease- active tuberculosis of the lungs- bleeding from the alimentary tract- serious infections of a viral, bacterial
and mycotic origin- the presence of electronic implants- conditions after organ transplants
Pregnancy is a state in which the dynam-ics of forming new tissues is the greatest. As a rule, apart from necessary measures, everything else should be avoided during preg-nancy. Until now, the influence of magneticfields on the fetus has been unknown. Due to
lack of information on the effect of magnetic fields on the fetus, the application of variablemagnetic fields in pregnancy should be con-sidered inadvisable.
Neoplastic disease in the majority of cases has been of a vague etiology, and the natural history of neoplastic diseases is also known only in some cases. Experimental data regarding the effect of magnetic fields on neo-plastic cell cultures and animals with an active neoplastic disease are contradictory. In accor-dance with the principle primum non nocere, variable magnetic fields should not be appliedin a neoplastic disease, although many data indicate a positive effect of magnetic fields onneoplasms.
Active tuberculosis of the lungs is a contraindication to the application of mag-netic fields, due to their interactive effect ontissue. The influence of magnetic fields onthe utilization of oxygen, as well as on the blood flow, is beneficial. Koch’s bacilli areaerobes, therefore, based on knowledge of the biological effects of variable magnetic fields, in case of TBC, they should not beapplied.
Bleeding from the alimentary tract should be divided into massive hemorrhag-ing – from the varices of the gullet and peptic ulcers – and bleeding accompanying gastritis type diseases. In case of intensive bleeding, the therapy of choice is an endo-scopic or surgical intervention. In case of bleeding observed with gastritis erosiva and gastritis haemorrhagica, caution is also rec-ommended in the application of magnetic fields. In case of gastric ulcer the relaxativeand anti-spastic action of variable magnetic fields could be considered advisable, nev-ertheless the decision of their application in this disease should be consulted with a doctor, since it is frequently accompanied by gastritis
Serious infections of a viral, bacterial and mycotic origin are a contraindication to the application of magnetic fields. Eachtype of these diseases should be treated with antiviral, antibiotic and antimycotic phar-maceuticals.
Application of Viofor JPS System
7
Electronic implants applied in mod-ern medicine, in most cases have good screening from external interference. From physical point of view, the fields applied inmagnetic field therapy, particularly in mag-netostimulation, have little influence on theoperation of electronic implants. The person ordering the therapy rarely has knowledge of the electronic implant’s screening qual-ity that is why it is safer for the patient to refrain from exposure to variable magnetic fields.
2.1.4 Side effects, which may occur due to magnetostimulation
Until now, no serious side effects of mag-netostimulation have been reported. The fol-lowing effects might occur temporarily: - tingling sensation - numbing sensation- warmth sensation- sleep disorders, particularly in elderly
people- irritation- concentration disorders- temporary intensification of pain.
Tingling, cramped and warmth sensa-tion may occur particularly in the first threeor four days of exposure. This applies in par-ticular to post-traumatic areas of the osseous system.
Sleep disorders are observed only in ap-prox. 3% of elderly patients. This could be easily eliminated by exposure to increasing level of induction.
Irritation and/or concentration dis-orders concern particularly women during menopause as well as people with increased tension of the neurovegetative system. The above symptoms are usually observed in the first days after exposure. The frequencyof their occurrence may decrease through the application of an increasing value of in-tensity. It is also advisable to inform patients about possibility of side effects. The above symptoms affect approx. 4% of women and 1% of men.
Temporary intensification of pain usually occurs during the first period of ex-
posure and affects people with significantdegenerative changes of the osseous-joint system.
Summing up, having taken into consider-ation the above observations, it should be stated that the application of variable mag-netic fields is safe and the progress of ex-perimental and clinical knowledge will in all likelihood result in reduction of the number of contraindications.
2.2 Magnetostimulation paired with light energy
2.2.1 Characteristics of magnetostimulation paired with light energy
The method of magnetostimulation paired with light energy utilizes the magnetic fieldeffect described in paragraph 2.1 of the Manual.
Light energy has a mainly topical effect on tissues. Its body penetration depends on the light wavelength. Tissue reaction depends on energy absorption in each tis-sue layer. The absorption is mainly affected by: tissue thickness, blood supply and flow,water content and presence of pigment. Infrared reactions start at the level of cell membrane, and reactions to red light – in mitochondria.
At the tissue level, mechanism of biologi-cal effect of variable magnetic fields and lightenergy are of similar nature. When applying magnetostimulation together with light, a syn-ergy effect could be expected, depending on individual characteristics.
Main biological effects of magnetostimu-lation paired with light energy should be as follows: - improved oxygen assimilation by its car-
riers, which improves ATP synthesis in oxidation-reduction systems aerobic and anaerobic respiration path; increased ATP concentration is directly related to the increased activity of ATP-dependent enzymes, responsible, among oth-ers, for protein and collagen synthesis,
Application of Viofor JPS System
8
which better stimulates tissue regenera-tion process.
- Strong stimulation of DNA synthesis and cell proliferation
- Vasodilatation effect due to direct relax-ation of vessel smooth muscular coat, related with accelerated processes of angiogenesis and tissue perfusion, which improves regeneration
- Direct effect on liquid crystal mem-brane structure and on modification ofmembrane enzyme activity related with increased ATP-dependent ion pumps through increased ATP synthesis in mi-tochondria; each of the above processes leads to changed membrane permeability, thus to change in electrolyte and water distribution between cells or their organelle and surrounding areas; both processes have a strong anti-inflammatory and anti-swelling effect
- Restored ATP-dependent membrane pump activity, which has been lost due to disease-related changes – sodium-po-tassium and calcium pumps in neurons, which leads to membrane hyperpolariza-tion and stimulus blocking, even those of high amplitude; as a result, pain stimuli conduction in afferent fibers is decreased,therefore a clear analgesic effect can be observed
- Changed activity of serotoninergic synapses and stimulation of beta-endor-phin release and stimulation of pituitary gland-adrenal gland axis accompanied by increased release of clycocorticoids adds to a strong analgesic effect
- Regulation of immune system and rheo-logical blood properties
2.2.2 Indications for magnetostimulation with light energy
I. Analgesic effect 1. osseous-joint system:
- degeneration of the spine and osse-ous-joint system of the upper and lower limbs
- joint strain and injuries- rheumatoid inflammation of the joints- stiffening inflammation (rigor) of the spi-
nal joints2. soft tissue:
- extrajoint rheumatism – so-called fibromyalgia
- traumas of soft tissue- zoster- neuralgia
II. Regenerative effect- skin and soft tissue inflammation- skin and soft tissue post-inflammation
condition- psoriasis- burns- bedsores- chronic and subacute arthritis- initial treatment preceding massage,
kinesitherapy and iontophoresis
III. Improvement of peripheral circulation1. improper local blood flow in the limbs:
- in diabetic angiopathy- in atherosclerotic angiopathy- in crural ulceration
2. accelerated absorption of haematomas4. decreased lymphatic edema
IV. Anti-spastic effect- degenerative diseases of the nervous
system, accompanied by spasticity
V. Decreased oxygen deficit and debt through affecting the processes of:- aerobic tissue respiration- anaerobic tissue respiration
2.2.3 Contraindications to magnetostimulation and light energy treatment
Although there is no proof for negative effect of magnetostimulation on each meta-bolic function, contraindications should be the same as in case of variable magnetic field (see paragraph 2.1.3.). Special attention should be paid to local neoplastic
Application of Viofor JPS System
9
changes and changes that are difficult to identify, predisposition to bleeding as well as acute inflammation of skin and soft tissue in the area exposed to therapy. Patients who are administered drugs, which cause photosensitization or show symptoms of photosensitization should not be treated – it is easy to verify by checking patient’s history.
2.2.4 Side effects which may occur due to magnetostimulation with light energy treatment
No significant side effects should be ex-pected. The following effects might occur temporarily:- tingling sensation,- numbing sensation,- warmth sensation,- skin erythema in the area of treatment.
3 THERAPY SELECTION
Viofor JPS apparatus is a unique and original design that provides the user with the selection of one of three options of variable magnetic field therapy:1. magnetostimulation in JPS System2. magnetostimulation in JPS System paired
with light energy
In order to establish the application param-eters and therapy options for a given patient, the following factors should be considered:- age- general health- course of the main illness- co-existing illnesses- administered pharmaceuticals - the number of treatments and their duration - other types of physical therapy.
3.1 Magnetostimulation1. Therapy should be conducted upon
a diagnosis and doctor’s consultation2. Magnetostimulation, like other forms of
physical therapy, usually does not bring immediate effects. They depend on the nature of the illness, its length and coexist-ing illnesses. The individual reaction to this type of treatment is also important.
3. Magnetostimulation complements primary treatment. Changes in treatment (par-ticularly pharmacological) should be intro-duced only after doctor’s consultation.
4. When lying down on a large applicator (mat, frame bed), the head should always be positioned near the cable outlet.
5. Due to the possibility of sleep disorders, par-ticularly in elderly patients, treatment should be avoided late in the evening. It particularly concerns treatment of the head and neck area with the use of a pad, elliptic and point applicator as well as ring applicators.
6. Therapy should be preceded by applica-tion of dEF1 or dEF2 program (see below), since it allows to determine individual sen-sibility to magnetic field therapy. The listsof programs are available for Classic version
Prophylaxis – based on default parameters preprogrammed by manufacturer (dEF1 and dEF2 lists are used only for magnetostimulation)
Application of Viofor JPS System
10
only (displayed by remote control). Stan-dard version users can set the below-listed parameters (M and I) manually – except for program (P), which is a default program and equivalent of P2 program in Classic version.
7. Each applicator can be plugged in any socket of Viofor JPS Classic controller. If a default program has been selected, i.e. dEF1 and dEF2, applicators assigned to these programs should be plugged in ac-cordance with markings on the controller and tables indicated in item 6.
8. Magnetostimulation with the use of a large applicator (cot, mat, short mat) can be ap-plied in intervals of minimum six hours, two or three times within 24 hours.
9. In case of long-term therapy in chronic ill-nesses, a week interval is recommended after several weeks of application.
10. During therapy it is not possible to interfere with the application’s parameters – all keys, apart from START/STOP and SOUND, are inactive at this time.
11. Possible subjective side effects are usually of temporary character.
12. Usage of an adapter enables to use two ap-plicators at the same time without possibility of applying different application parameters.
A table containing applicator combinations is on page 11. The applicators, when used simultaneously on a patient, intensify the effect of variable magnetic field, and, there-fore, accelerate the therapeutic effect. Two elliptic applicators of the same color cannot be used on one patient in respect to the same body part. It is, however, possible to use any two applicators on two patients. Other combinations require consultations with the distributor.
3.1.1 Application parameters components
3.1.1.1 Programs (P)1. P11) – JPS system with one impulse type
of simplified structure2. P2 – JPS system with two impulse types 3. P31) – JPS system with two impulse types
of different impulse order and structure as compared with P2
3.1.1.2 Application modes (M)1. M1 – application with a constant intensity;
the selected field intensity is constantthroughout the whole time of application; recommended in persons of good health
2. M2 – application with increasing intensity; the degree of field intensity increases every 10 or 12 seconds from 0,5 to the se-lected intensity, cyclically during the time of application; recommended in persons of lowered general condition
1) for version Classic
Application of Viofor JPS System
11
3. M3 – application with an increasing-de-creasing intensity; the degree of field inten-sity increases in 2 min. from 0,5 to the se-lected value, during the next 8 min. remains at the selected level, during the last 2 min. decreases to a level of 0,5; apply from an intensity of 3 or higher.
3.1.1.3 Magnetic field intensity ( I )Intensity from 0,5 to 12 corresponds
with proportional values of magnetic field emitted by the applicators. It is recom-mended to start therapy from magnetic field intensity set at 1 and successively set to higher levels. The speed of setting to higher levels depends on the age and gen-eral health of the user. For persons of high sensitivity to magnetic fields it is recom-mended to start application from a level of
0,5. For older people and children, chang-es may be introduced every 5-7 days, for others every 1-2 days.
Proposed magnetic field intensity for adults: - level 0,5 do 121) (102)), depending on age
and individual sensitivity to magnetic fields.Proposed intensity for children:
- 0-2 years level 0,5-1- 2-5 years level 1-2- 5-11 years level 1-4- 11-15 years level 1-6
3.1.1.4 Selection of application parameters
Application parameters are selected through a combination of: program (P)1), mode (M) and intensity (I)1. P11) + M1 – application with constant in-
tensity (8 min.)2. P11) + M2 – application with increasing
intensity (12 min.)3. P11) + M3 – application with increasing-de-
creasing intensity (12 min.)
Applicator combinations
1) for version Classic2) for versioni Standard
Application of Viofor JPS System
12
4. P21) 2) + M1 i P31) + M1 – JPS system with constant intensity (10 min.)
5. P21) 2) + M2 i P31) + M2 – JPS system with increasing intensity (12 min.)
6. P21) 2) + M3 i P3 + M3 – JPS system with increasing-decreasing intensity (12 min.)The recommended modes of applica-
tion are based on wide clinical experience of the teams of practicing physicians as well as multicenter research that evaluate the ef-fectiveness of magnetostimulation with the Viofor JPS apparatus.
Each of the application types can be com-bined with selected field intensity according tothe above-mentioned instructions.
In order to establish the application param-eters for a given patient, the following factors should be considered:- age- general fitness- course of the main illness- co-existing illnesses- pharmaceuticals administered- time of the day, when therapy
will be carried out- duration of magnetostimulation- other types of physical therapy.
3.1.2 Using more than one applicatorAdditional accessories (adapter, extra ap-
plicators) enable optional usage of several applicators in one or more patients in order to increase the therapy effectiveness or to use the apparatus more economically.
Example:– patient with rheumatoid inflammation of
joints and significant degenerative chang-es of hip joints – simultaneous application of a mat (whole body) and a pad (hips)
– several large applicators used for several athletes, e.g. after training.Two-color coding of applicators is used to
identify the direction of magnetic field polesgenerated by an applicator. In case of the same impulse direction, each element of the applicator when, for example, facing white side up, will have the same magnetic field
pole direction. It is important in case of using of two applicators in one patient.
NOTICE!
Never use two applicators with the same color sides facing each other.
When a patient’s body or a part of the body is placed between two applicators generating heterogeneous magnetic field, the effect isclose to the homogenous magnetic field.
1 m connection between two elements of the flat applicator enables simultaneous ex-posure of two body parts.
Two or three large applicators can be used at the same time (connected with one or two adapters) as well as different combinations of applicators (e.g. large and point applicator, large and small, large and point). Point applicators work with large or small applicators facing the white side up. In case of using applicators in two or three patients at the same time, therapy parameters are identical for all patients. There is no possibility of setting separate parameters for each applicator when used simultaneously.
3.2 Magnetostimulation with light energy
CAUTION!
STRONG RED AND/OR INFRARED RADIATION!
PROTECT YOUR EYES!DO NOT LOOK AT THE LIGHT!
1. Magnetostimulation paired with light en-ergy is used locally during the series of repeated treatments with the use of one of three magnetic-light applicators.
2. Each applicator generates a heteroge-neous variable magnetic field in the JPSsystem. The light area of 48 diodes is lo-cated in within the coil and equals 19,6 cm2.
3. Light energy penetrating the tissue de-pends on the radiation properties (wave-lengths, angle of incidence), and on tissue
1) for version Classic2) for version Standard
Application of Viofor JPS System
13
properties (rate of hydration, perfusion, pigment content).
4. Tissue penetration will be the highest in case of infrared magnetic-light applica-tor (48 diodes, wavelength 855 ± 5 nm and one diode, wavelength 630 ± 5 mn), intermediate in case of combined mag-netic-light applicator (24 diodes, wave-length 855 ± 5 nm, 24 diodes, wavelength 630 ± 5 nm), and the lowest in case of red magnetic-light applicator (47 diodes, wavelength 630 ± 5 nm).
5. Energy power of each light type will in-crease to intensities set on the controller. Between 0.5 – 6 it increases proportionally to intensity, and between 6 – 12 – it re-mains on the same level.
6. Treatment with magnetic-light applicators should be started from low induction val-ues of variable magnetic field, due to theexpected synergy of magnetostimulation and light energy.
7. If the magnetostimulation has started with the use of elliptic applicator, then it is rec-ommended to lower down the induction of magnetic-light applicator by 50% as com-pared to the previous one in respect to the same body area.
8. Due to the applicator contact with the skin, it is recommended to wash and dry the skin area to be exposed.
9. When using magnetic-light applicators, never direct the light in the eye area!
3.2.1 Application parameter components
3.2.1.1 Programs ( P )When using magnetic-light applicators, all 3
programs are recommended, i.e. P11) P22) and P31). Energy light parameters are indicated in technical data tables, pp. 32-34.
3.2.1.2 Application modes ( M )Viofor JPS apparatus enables three
different application modes (M) for vari-able magnetic field described in paragraph3.1.1.2.
When using magnetic-light applicators, M1 mode is recommended. It is the applica-
tion with constant intensity, where selected intensity of variable magnetic field does notchange during the application.
3.2.1.3 Magnetic field intensity ( I )
Intensity from 0,5 to 121) (102)) corresponds with proportionate values of variable magnetic field emitted by ellipticapplicators. The highest mean impulse value of variable heterogeneous magnetic field (in microteslas) on the surface of mag-netic-light applicators is 8.9 x 1 at P2M1 (intensity level selected on the Viofor JPS controller).
It is recommended to start the therapy from the lowest intensity and increase it gradually.
3.2.1.4 Selection of application parametersApplication parameters are selected
through a combination of one of three pro-grams (P11), P22) or P31)), application mode M1 (recommended) and magnetic field inten-sity (I).
Treatment duration is set automatically depending on P and M combination.
When the induction in Viofor JPS con-troller increases from 0.5 to 6, light energy density increases as well, whereas above these values, i.e. from 6 to 12, only variable magnetic field induction will increase. Due tothe expected synergy of both types of radia-tion, the therapy should be started from low induction values of variable magnetic field,and, therefore, from low light energy density level.
Selection of one of three magnetic-light applicators should depend on the type of light required for a given form of therapy.
1. Red magnetic-light applicator R (wavelength 630 ± 5 nm) is used mainly for skin disease treatment, such as unhealing wounds and ulcerations (including bedsores), common acne and zoster. It is recommended in dermatology and cosmetic surgery.
1) for version Classic2) for version Standard
Application of Viofor JPS System
14
Application of Viofor JPS System
2. Infrared magnetic-light applicator IR (wavelength 855 ± 5 nm) in motor sys-tem diseases (arthrosis; pain syndromes in discopathy; syndromes due to strain of muscles and periarticular soft tissue; inflammation of tendons, fascia, tendon sheaths and joint bursa; periarticular in-flammation; bone fractures and delayed union of fractured bones) as well as in neurological diseases (neuralgia of pe-ripheral nerves – intercostal neuralgia, trigeminal neuralgia, post-zoster neural-gia, sciatica; diabetic neuropathy).
3. Combined magnetic-light applicator R+IR (wavelength 630/855 ± 5 nm) in chronic skin inflammations; keloid, psoriasis, psori-atic joint inflammation, ankylosing spondy-litis, rheumatoid inflammation of the joints, post-thrombotic syndrome.
Energy doses as well as type of light and the number of treatments should be selected individually; their parameters should be simi-lar to those used in laser biostimulation, de-pending on experience and preferences of the therapist.
14
Usage of Viofor JPS System
15
PART B
Usage of Viofor JPS System
16
Usage of Viofor JPS System
4 VIOFOR JPS CLASSIC
4.1 General InformationThe front panel of Viofor JPS Classic contains: membrane keyboard control unit, applica-
tion time display and LED diode indicating selected settings and passage of current through an applicator. These elements are in the main control panel of the apparatus.
4.2 Viofor JPS Classic back panel
Drawing 1: Viofor JPS Classic back panel
1 – power switch2 – voltage switch (default setting 200/230V)3 – fuse base with a 0,25 A (250 mA) delay
switch4 – power cable
5 – socket A36 – socket A27 – socket A18 – service socket
– DO NOT USE!
4.3 Viofor JPS Classic front panel
Drawing 2: Viofor JPS Classic front panel
Usage of Viofor JPS System
17
1 – START/STOP button – starts or stops an application
2. Light-emitting diode / LED (yellow), which flashes when the apparatus is operating
3. Button to increase the magnetic field in-tensity (>)
4. Button to decrease the magnetic field in-tensity (<)
5. A series of light-emitting diodes / LED (7 green, 3 yellow, 3 orange), which signal the set and used magnetic field intensity
6. SOUND button – switches on / off beeper signaling of apparatus operation
7. Light-emitting diode / LED (green), which lights up whenever the beeper is switched on
8. ON button, which activates the apparatus from STAND-BY mode and restores indi-cator lighting
9. Light-emitting diode / LED (green), which signals switching on of the apparatus found in STAND-BY mode, when all indi-cators are unlit
10. Four-digit display, which indicates applica-tion time (before pushing START/STOP button), or remaining application time
as well as programmed and programmable lists.
11. LED diodes (green) signal current polariz ation of the field impulses
12. A key, for applicator socket selection13. Light-emitting diode / LED, which signals
selection of A1 socket – mat: lights green if the applicator circuit is closed, red – if there is a break in the applicator’s circuit
14. Light-emitting diode / LED, which signals selection of the A2 socket – pad, color lighting as in point 13
15. Light-emitting diode / LED, which signals selection of the A3 socket, color lighting as in point 13
16. M button (MODE) – mode of magnetic field application
17. Light-emitting diodes / LED (green), which signal the chosen application mode: M1, M2 or M3
18. P button (program) – selection of the ap-propriate application program
19. Orange light-emitting diodes / LED, which signal the selected program: P1, P2 or P3
20. Infrared receiver window (for remote control)
4.4 Remote control unit
Remote control enables fast programming and setting.
4.4.1 Remote control details1– START/STOP button switches
the application on and off2. ON button - activation
of the apparatus from STAND-BY mode and restoration of indicator lighting
3. A1, A2, A3 applicator selector buttons
4. M1, M2, M3 application mode selector buttons
5. P1, P2, P3 program selector buttons
6. 0,5 to 12 intensity selection buttons7. B, C, D, E active set application
parameter selector buttons
8. SEL button selects successive application parameter sets LI1, LI2, dEF1, dEF2
Drawing 3: Remote control unit
Usage of Viofor JPS System
18
9. OK button confirms set selection (pre-ceded by the SEL button) or monitoring of the active set on the apparatus front panel display (not preceded by the SEL button)
10. SOUND button - switches the beeper on and off, to indicate whether the unit is in operation
11. CODE button – in standard version used as one of the elements enabling counter reading; in special version conditions the switching on of the apparatus
4.4.2 Battery installation1. Remove the battery cover on the back of
the remote control unit in the direction indi-cated by an arrow
2. Place in the correct position (in accor-dance with the diagram inside the contain-er) two AAA alkaline batteries (or equiva-lent)
3. Replace the battery cover ADDITIONAL INFORMATION
- use only alkaline AAA batteries - batteries will last depending on remote
control usage, however they should be changed at least once a year
- if the remote control unit is not used for a long time, remove the batteries
- batteries should be removed during transportation
- do not mix old batteries with new or batteries of different types
- do not connect the batteries to an elec-tric current, take them apart, heat or put into fire.
Change of batteries is not covered by the warranty and is within user’s discretion.
4.5 Apparatus installation
Load batteries in the control unit. Set the control unit on a stable base
(e.g. on a table) and connect the applicators. Plug in the cable. If a mat is used, it should be laid on a flat, hard surface at a minimumdistance of 50 cm from metal objects (e.g. radiators).
For best effects, select places free from the interference of other magnetic fields.
4.6 Apparatus operational use
1. Press the power switch (1; see fig. 1,p. 16) on the back panel of the appara-tus. The parameters will be set after a short test:
Applicator A1 – mat (mat socket) Program P1Mode M1 (application with constant intensity)Intensity – 1Time – 8:00 min2. Press the A button (on the control unit) or
A1, A2, A3 (on the remote control) to select the applicator – the green diode correspond-ing to this applicator will light up. If the LED lights red, this signals a circuit break or too large resistance in the applicator circuit (e.g. a loose pin).
3. Press the P button to select the appro-priate program (P1, P2, P3). An orange LED will show the program currently se-lected.
4. Press the M button to choose the appropri-ate mode of application (M1, M2, M3). A green LED will show the mode currently selected.
5. Press the „>”(3) or „<”(4) buttons on the ap-paratus front panel or ”0,5” to ”12” buttons on the remote control unit to increase or decrease the intensity to the selected level. A series of lit LED shows the selected level of intensity. CAUTION! Each time the intensity button on the front panel is pressed, the intensity level changes by one point - it is signaled by a beep. With a change of program (P) and application mode (M), the application time shown on the display also changes automatically (8 min, 10 min, 12 min).
6. After setting the appropriate application parameters press the START/STOP but-ton – this is accompanied by a prolonged beep. The clock is then activated to count down the application time in sec-onds from the set value to zero. During the application, a yellow LED pulsates next to the START/STOP button. The
Usage of Viofor JPS System
19
application stops automatically after the set time has elapsed and is signaled by 3 beeps.
7. When START/STOP button is pressed all buttons apart from START/STOP and SOUND will become inactive – stop re-sponding.
CAUTION! The START/STOP button will not work if the circuit of the chosen applica-tor is broken (the applicator’s LED will light red). If a break in the circuit occurs during the application, it will be interrupted, the display will show “Err” and then the initial time of the application.
8. Depending on requirements, press the SOUND button to switch on or off the sound signaling operation of the apparatus.
9. If the apparatus is inactive for longer than 2 minutes, it switches to as STAND-BY mode. The green LED then lights up. To re-activate the apparatus press ON button.
4.7 Programming of application parameters
Programming applies to sets LI1 and LI2. Sets dEF1 and dEF2 (see item 6 p. 9) are permanently programmed by the manufacturer.
To program application parameters follow the instructions:1. press the power switch (1; see fig. 1, p. 16)
on the back panel of the control unit or, if it is in STAND-BY mode, activate it with the ON button.
2. set the required applicator (A)3. select the application parameters i.e. pro-
gram (P), mode (M) and intensity4. press the SEL button on the remote control
unit to retrieve set LI1 or LI25. press the ON button on the front panel of
the control unit – the message SAVE will appear on the display
6. on the remote control unit press one of the B, C, D or E buttons, to store the applica-tion parameters
7. a prolonged beep informs us that the application parameters have been stored.
By repeating the procedure described in points 2-6, 8 different sets of application parameters can be programmed (4 for set LI1 and 4 for set LI2). The parameters will be stored in the apparatus’s memory until re-placed by new ones.
4.8 Retrieving preprogrammed application parameters:1. Press the ON button to activate the ap-
paratus2. Press the OK button on the remote con-
trol unit to check which set is active – if the required set is displayed, proceed to the action described in pt.5, and then onto pt. 6; if not:
3. Press the SEL button on the remote control unit repeatedly until the required set is displayed
4. Press the OK button on the remote con-trol unit to confirm the set selection
5. Press one of the B, C, D or E buttons on the remote control unit to retrieve selected preprogrammed application parameters
6. Press the START/ STOP button to start the application.
4.9 PLACEBO mode
When the apparatus is off, press and hold the intensity ”>” (3) button and press the power switch. After the message ”8:00 min” appears on the display release the ”>” (3) button. The apparatus will simulate all func-tions, but impulses will be sent to a subsid-iary circuit. All applicators will remain inactive at this time.
4.10 Counter
When ON-CODE-ON buttons on the remote unit are pressed in sequence within 1 second, the counter will appear on the display. Each digit is separated by a pe-riod. The treatment is measured if longer than 30 seconds. The user cannot reset the counter.
Usage of Viofor JPS System
20
5 VIOFOR JPS STANDARD5.1 General Information
The front panel of Viofor JPS Standard contains: membrane keyboard control unit, application time display and LED diode in-
dicating selected settings and passage of current through an applicator. These ele-ments are in the main control panel of the apparatus.
1 – power switch2 – voltage switch (default setting 200/230V)3 – fuse base with a 0,25 A (250 mA)
delay switch
4 – power cable5 – applicator socket6 – service socket
– DO NOT USE!
5.2 Viofor JPS Standard back panel
Drawing 4: Viofor JPS Standard back panel
Drawing 5: Viofor JPS Standard front panel
5.3 Viofor JPS Standard front panel
1 – START/STOP button – starts or stops an application
2. Light-emitting diode / LED (yellow), which flashes when the apparatus is operating
3. Button to increase the magnetic field intensity (>)
4. Button to decrease the magnetic field intensity (<)
Usage of Viofor JPS System
21
5. A series of light-emitting diodes / LED (7 green, 3 yellow, 3 orange), which signal the set and used magnetic field intensity
6. SOUND button – switches on / off beeper signaling of apparatus operation
7. Light-emitting diode / LED (yellow), which lights up whenever the beeper is switched on
8. ON button, which activates the apparatus from STAND-BY mode and restores indi-cator lighting
9. Light-emitting diode / LED (green), which signals switching on of the apparatus found in STAND-BY mode, when all indi-cators are unlit
10. Four-digit display, which indicates applica-tion time (before pushing START/STOP but-ton), or remaining application time as well as programmed and programmable lists.
11. LED diodes (green) signal current polar-ization of the field impulses
12. A LED diode, lights green if the applicator circuit is closed, red – if there is a break in the applicator’s circuit
13. M button (MODE) – mode of magnetic field application
14. Light-emitting diodes / LED (green), which signal the chosen application mode: M1, M2 or M3
5.4 Apparatus installation
Set the control unit on a stable base (e.g. on a table) and connect the applicators. Plug in the cable.
If a mat is used, it should be laid on a flat,hard surface at a minimum distance of 50 cm from metal objects (e.g. radiators).
For best effects, select places free from the interference of other magnetic fields.
5.5 Apparatus operational use
1. Press the power switch (1; see fig. 1, p. 16)on the back panel of the apparatus. The pa-rameters will be set after a short test:Mode M1 (application with constant intensity)Intensity – 1Time – 10:00 min
If an applicator is plugged in, a green di-ode will light up. If the diode turns red, this signals a circuit break or too large resistance in the applicator circuit (e.g. a loose pin).2. Press the M button to choose the appropriate
mode of application (M1, M2, M3). A green LED will show the mode currently selected.
3. Press the „>”(3) or „<”(4) buttons on the apparatus front panel or ”0,5” to ”12” but-tons on the remote control unit to increase or decrease the intensity to the selected level. A series of lit LED shows the select-ed level of intensity.
CAUTION! Each time the intensity button on the front panel is pressed, the intensity level changes by one point - it is signaled by a beep. With a change of program (P) and application mode (M), the application time shown on the display also changes automati-cally (8 min, 10 min, 12 min).
4. After setting the appropriate application parameters press the START/STOP button – this is accompanied by a prolonged beep. The clock is then activated to count down the application time in seconds from the set value to zero. During the application, a yellow LED pulsates next to the START/STOP but-ton. The application stops automatically after the set time has elapsed and is signaled by 3 beeps.
5. When START/STOP button is pressed all buttons apart from START/STOP and SOUND will become inactive – stop responding.
CAUTION! The START/STOP button will not work if the circuit of the chosen applicator is broken (the applicator’s LED will light red). If a break in the circuit occurs during the ap-plication, it will be interrupted, the display will show “Err” and then the initial time of the application.
6. Depending on the needs the button SOUND can switch on / off beeper signaling of appa-ratus operation.
7. After 2 minutes of idleness the apparatus switches in the STAND-BY state, which is ac-companied by turning off whole signalization. Then the green diode lights up. Reactivation of the apparatus can be done by pressing the ON button.
Usage of Viofor JPS System
22
6 MAGNETIC & LIGHT THERAPY6.1 General Information
Magnetic & Light Therapy set is used for magnetostimulation paired with light energy treatment. Viofor JPS controller generates control signal for all applicators manufactured by the company. ZAMS 01/2 power supply adjusts the output signal in the controller to the input signal in magnetic-light applicator. Magnetic-light applicator gener-ates heterogeneous variable magnetic fieldof a specific signal structure, defined as JPSsystem, and emits light impulses of selected parameters synchronized with magnetic fieldimpulses.
6.2 ZAMS 01/2 power supply unitZAMS 01/2 power supply unit enables
the users of Viofor JPS Classic or Viofor JPS Standard to apply magnetic and light therapy. It adjusts the output controller signal to the input signal of the magnetic-light applicator and syn-chronizes with light impulses.
The power supply unit is equipped with 6-pin socket to connect the applicator and the cable with a large jack, connected to a socket in Viofor JPS System controller. Large jack has a blue label wit the power supply unit symbol. A diode (ON) in the middle of the front panel lights in blue when the power supply unit is plugged in.
6.2.1 ZAMS 01/2 front panel
Drawing 6. ZAMS 01/2 front panel
1 – signal diode2 – cable with
a large jack3 – 6 –pin socket
6.2.2 ZAMS 01/2 back panel
1 – power on2 – fuse base with a 100 mA delay switch3 – power cable4 – power voltage switch5 – power unit serial number
Drawing 7. ZAMS 01/2 back panel
Usage of Viofor JPS System
23
6.3 Magnetic-light applicators
The applicator casing contains a coil for magnetic field therapy and LED diodes setfor light therapy. On the top of the casing there is a triangular warning label, signal diode in color depending on the applicator type (see Table below) and company logo. On the bottom there is a transparent glass, through which the light is emitted.
Each applicator is equipped with a cable with a 6-pin jack. The cable is labeled with different color labels indicating:- applicator serial number
(from the jack’s side),- applicator type identifier
(from the casing side – R, IR, RIR)Applicator is attached to the patient with
a regulated belt with a buckle.
6.4 Magnetic & Light Therapy set usage1. Make sure that the power voltage switches
(on the back panel of power supply unit
and controller) are set at the proper volt-age. If not – switch them accordingly and exchange the fuses according to the de-scription on the back panels of the power supply unit and controller under the fuse socket (electric cable should be discon-nected).
2. Connect ZAMS 01/2 power supply unit to the Viofor JPS controller socket with the large jack. Connect the plug of mag-netic-light applicator to the 6-pin power supply unit socket and turn it right to block it.
3. Turn the power on in controller and power supply unit.
4. Place the applicator over the treated area. Set the controller. Press „START”. NOTICE:Due to direct contact of the ap-
plicator with the patient’s skin it is recom-mended to clean the applicator with a disin-fectant (Notice, do not allow the applicator to get wet!).
In case of therapy of open wounds and bacterial or mycotic skin deformations, do not touch the applicator to the skin. Keep it 0,5-1 cm above the skin.
Drawing 8: LED diode layout in applicators: a) red light R, b) infrared light IR, c) mixed light RIR (infrared applicator is equipped with one red LED diode that signals emission of invisible infrared radiation IR during treatment)
a) b) c)
Usage of Viofor JPS System
24
8 CLEANING, MAINTENANCE, REPAIRS8.1 General information- Clean the controller with the plastic clean-
er, do not use water- Applicators can be cleaned with a cloth
with a mild detergent and, if necessary, with disinfectant
- Wash ZAMS 01/2 only with detergent for plastics
- Device does not require any maintenance- Repairs can be done only by authorized
service of Med & Life, as described in the warranty certificate. Any unauthorized re-pairs and/or breaking of guarantee seals will result in warranty termination!
- When changing the fuse, unplug the con-troller and power supply unit
- In case of broken fuse in controller or ZAMS, exchange it (when the controller and power supply unit are plugged off!). Make sure to use the proper fuse, i.e. according to the label at the fuse socket.
- Do not short-circuit 6-pin sockets in ZAMS 01/2!
- In case of damaged power cable or plug, do not use the set and have it serviced.
- Magnetic-light applicators are not resistant to high temperatures. Do not use auto-clave for disinfection!
- If disposal is necessary, have it done by a specialized company
7 SAFE USAGE OF VIOFOR JPS 1. The apparatus transported or stored in
temperatures below the temperature in the room, where it will be installed, it can be plugged in after at least 4 hours from deliv-ery. The equipment cannot be operated in humid places
2. When installing Viofor JPS, check if the voltage is set correctly.
NOTICE! Devices distributed in Europe are set by default to 203 V. When switch-ing the voltage to 115 V (when using in other countries), fuse should be ex-changed as indicated on a label.
3. Install and connect the apparatus accord-ing to the instruction contained in the pres-ent manual.
4. Do not plug the device if the cable or plug is damaged. Have the device serviced
5. Protect the device out of reach of small children and mentally disabled persons.
6. Keep away from pets.7. If the device is flooded or gets wet, do not
plug it in – have it serviced immediately.8. Before plugging the device in, put away
magnetic cards and electronic watches, since they might be damaged. It applies both to the patients and the operators.
9. Turn off all other electronic devices (e.g. hearing aids, pain stimulators, etc.)
10. For hygienic purposes, it is recommended to cover the applicators with disposable foil or paper covers.
11. When the device is working:- The personnel must keep at least
1 meter away from an active applicatorCaution! When using magnetic-light applica-tors, do not direct the light in the eye area!- The device can be plugged in and out
only when the power switch is off.- Plug the applicator in and out only
when the power is off.12. After work:
- Do not roll the mat, but fold it twice (see drawing)
- Do not bend the pad- Before cleaning the device, read appro-
priate section of the instruction - Do not open the zipper around the mat.
13. It is recommended to have the device checked every 1 to 2 years (service is free of charge, transportation costs are covered by the client).
14. The device transported to the Med & Life service should be well protected against damages (e.g. in original package)
15. Use only AAA alkaline batteries for the remote control.
Usage of Viofor JPS System
25
8.2 Service pack
8.2.1 Fuse exchange1) Unplug the device from the wall socket.2) Remove the fuse handle and the fuse
with a proper screwdriver. The fuse handle is removed by one-quarter turn counter-clockwise.
3) Remove the fuse from the handle. Insert a new fuse in accordance with the instruction on the back panel.
4) Place the handle in the socket with a proper screwdriver and make one-quarter turn clockwise.
5) Plug the device in the wall socket and turn on the power. If the device is not operating, it should be serviced.
NOTICE: do not use fuses with parameters other than indicated.
8.2.2 Exchange of applicator plug and adapter1) Cut the damaged plug from the applicator
cable.2) Unscrew the new plug casing. Install
the new plug on the applicator cable.3) Remove external insulation from applicator
cable – approx. 1.5 cm.4) Remove approx. 0.5 cm of insulation
from the brown (or black) wire.5) Remove approx. 1.5 cm of insulation
from the blue (or white) wire.6) Cover the uncovered wire ends
with soldering agent.7) Solder the brown (or black) wire
to the middle plug electrode.8) Solder the blue (or white) wire
to the side plug electrode.9) Screw the plug casing together.10) Test the applicator with MFI indicator.
If the applicator is not working, it should be serviced.
9 WARRANTY TERMS
Warranty cover free repair or exchange of the device or its elements in case of manu-facture and material defects. Exchange will apply to the damaged element only.
The warranty does not cover:• mechanical damages
- due to device being dropped or hit - abrasions- scratches- dents- tears- discoloration due to usage of improper
detergents and cleaning agents- detachment of magnetic-light
applicator glass due to improper cleaning method
- torn or broken power supply cable, ap-plicator cable or plug
- improper disinfection method (e.g. in autoclave)
• regular wear and tear of applicators • applicator deformation• effects of atmospheric discharge
• pollution• flooding• damages due to
- insufficient knowledge of the user - incorrect installation or usage of the
device- exploiting of the device in conditions
others then specified by the manufacturer- incorrect parameters of power supply- incorrect setting of the voltage switch- incorrect fuse
• broken warranty seals• damages during the transport of the
device to Med & Life authorized service
In case of repair or structural changes by unauthorized persons, Med & Life shall have the right to revoke warranty obligations.
The manufacturer shall not be held responsi-ble for any bodily damages or injuries due to in-correct usage of the device, its usage for purpo-ses other than those indicated in the instruction manual, or any changes introduced to the devi-ce without prior approval of the manufacturer.
Technical data and accessories
26
PART C
Technical data and accessories
27
Technical data and accessories10 TECHNICAL DATA
10.1 Controllers
Controller Microprocessor
Trade name Viofor JPS System Classic/Standard
Voltage rating and frequency 220/230 V – 110/115 V (±10%, 50-60 Hz)
Power input in ready mode <8 VA
Maximum power input <15 VA
Output voltage in controller socket Constant voltage, max 20 V
Fuse in primary circuit T 0,25 A / with delay (for 110/115 V / T 0,5A)
Working mode Continuous running
Storage and transport temperature From -15 to +70oC
Relative working humidity Max 80%
Admissible storage and transportation humidity Max 90%
Output controller socket Jack
Safety class II
Protection type IP40 (water unprotected)
Protection B (from electric shock)
Magnetic induction intensity regulation in Classic controller Gradual 0,5; 1; 2.................; 12
Magnetic induction intensity regulation in Standard controller Gradual 0,5; 1; 2.................; 10
Application time 81), 10, 12 min
Automatic field direction change Depends on program and application
Continuation on the next page
1) for Classic controller
Technical data and accessories
28
Amplitude of current running through applicator coils for intensity 0,5 0,05 ÷ 0,1 A
Amplitude of current running through applicator coils for intensity 12 (Classic) 1,5 ÷ 2,0 A
Amplitude of current running through applicator coils for intensity 10 (Standard) 1,4 ÷ 1,7 A
Continuation from page 27
10.1.1 Compatibility All Viofor JPS controllers
are compatible with all Viofor JPS applicators and can work with magnetic-light applicators through ZAMS 01/2 power supply unit. Viofor JPS Classic also works with infrared remote control unit.
10.1.2 CasingA special ABS casing has been used,
of maximal dimensions: width 330mmdepth 280mmheight 280mm
10.1.3 Mass Mass of the device without applicators - < 4kg
10.2 Power supply unit ZAMS 01/2
Power 230V/115V (±10% -10%) – T100/200 mA
Input signal from Viofor JPS controller
Impulse output signal synchronized with controller signal: 21 V ± 0,3 V
Max. power input 12 VA ± 2 VA
Working temperature 10°C – 40°C
Relative working humidity max 80 %
Storage and transport temperature -15°C - +70°C
Admissible humidity of storage and transport max 90%
Safety class II
Type of protection IP40
Protection B (from electric shock)
Working mode Continuous running
Technical data and accessories
29
10.3 Magnetic-light applicatorsPower and input signal from ZAMS 01/2 power supply unit
Output signal Light impulses synchronized with magnetic field impulses: - applicator light parameters: see table p. 32-34 - magnetic field parameters: peak value 100 x I[µT], mean value: 8,9 x I [µT] where: I – intensity set on the controller
Max power input 0,63 A
Working temperature 10°C – 40°C
Relative working humidity max 80 %
Storage and transport temperature -15°C - +70°C
Admissible humidity of storage and transport max 90%
Safety class for light energy 3A
Protection B (from electric shock)
Working mode Continuous running
11.1 Applicators for magnetostimulation
11.1.1 Large applicator - matUsed for the whole body treatment. Contains 3 pairs of coils placed along the mat and generating heterogeneous variable
magnetic field of different intensity: coil located in the feet area are the strongest, and in the shoulder area – the weakest. Head should always be located near the output cable. Mat is a portable applicator, in two colors, can be folded in 3. Size 180 x 65, folded 60 x 65 cm. Installation: place the applicator on a hard
surface, away from any household appliances and TV or radio sets. Plug the applicator to any controller socket*. Lay down on the mat, with your head near the output cable. Set the desired application parameters and press START/STOP button.
11.1.2 Large applicator – short matUsed for the whole body treatment. Generates het-erogeneous variable mag-netic field. Used in cases, when it is recommended to
use two mats on one patient or if a regular mat is not handy (e.g. after surgery). Two colors. Size 157 x 65 cm, side cable output.
11 APPLICATORS
* In case of using default application parameters, mat, pad and point applicators should be connected in accordance with symbols located over the controller sockets.
Technical data and accessories
30
Installation: plug the applicator into any con-troller socket*. Cover the patient with the mat. Set the desired application parameters and press START/STOP button.
11.1.3 Large applicator - frame bedMetal frame bed, generates heterogeneous variable magnetic field. Used instead of large applicator – mat in doctor’s office.
Equipped with applicator roll guides for the large ring applicator. Size: length 200 x width 60 x height 55 cm.Installation: place the applicator away from household appliances and TV or radio sets. Plug the applicator into any controller socket*. Set the desired application param-eters and press START/STOP button.
11.1.4 Small applicator - padUsed for local treatment, can be placed over or un-der selected body part. Contains one pair of coils generating heterogeneous
variable magnetic field. The strongest effect of magnetic field is located near each coil. Do not bend the pad. Two color. Size 50 x 30 cm.Installation: Plug the applicator into any controller socket*. Place the applicator on or under the treated body part. Set the desired application parameters and press START/STOP button.
11.1.5 Point applicator (S and Z)For local application. Generates heterogeneous variable magnetic field of point-like characteristics. Used in dentistry, on selected acupuncture points and in cases when a narrow beam of variable magnetic field of significant induction
is required. Can be used together with two-color applicator facing white side up.Installation: Plug the applicator into any controller socket*. Point the plastic applicator end to the place exposed to treatment. Set the desired application parameters and press START/STOP button.
11.1.6 Elliptic applicator (light and green)For local application, fastened with the belt on limbs or with the tape in other places of application. Generates heterogeneous variable magnetic field, whose area of activity is within approx. 6 cm circle. Used mostly for treatment of small and medium joints,
in dentistry and laryngology. White and greed applicators, when used separately, work identically. If two elliptic applicators are used on the same spot, e.g. wrist in the same patient, only different color applicators can be connected.Installation: plug the applicator into any controller socket*. Place the applicator on the treated area, fasten with the strap, with an adhesive tape or hold it in one position. Set the desired application parameters and press START/STOP button.
11.1.7 Ring applicator – small, 30 cm diameter
For local application on limbs. Generates homogenous variable magnetic field of line of force parallel to the coil axis, therefore parallel
to patient’s body, which increases application effectiveness. Equipped with two cable sockets, marked H and L. When applicator is connected to socket H, the induction of magnetic field generated by the applicator is three times stronger than in socket L.
* In case of using default application parameters, mat, pad and point applicators should be connected in accordance with symbols located over the controller sockets.
Technical data and accessories
31
Installation: select therapy type: H or L and plug in the cable in the selected applicator socket, in accordance with symbols on the cable. Plug the other end to any controller socket. Place the limb inside the applicator (applicator should be place exactly above the treated area) on a flat base. Set the desired parameters and press START/STOP button.
11.1.8 Ring applicator – large, 65 cm diameter
Used together with the large applicator – frame bed. Equipped with rolls that can be moved in the frame bed guides. In the area of applica-tion, generates homogenous variable magnetic field, of line of force parallel to the coil axis, therefore parallel to patient’s body,
which increases application effectiveness. Used as local applicator for selected body parts.Installation: Plug properly (see assembly instruction) ring applicator installed on the guide rails of the frame-bed applicator to any controller socket. Lie down on the applicator and set the ring applicator. Set the desired application parameters and press START/STOP button. If frame-bed applicator and large ring applicators have to be used simultaneously, they should be connected through an adapter.
11.2 Magnetic-light applicators
Three magnetic-light applicators for lo-cal use (red (R), infrared (IR) and combined red/infrared (R+IR)). Generate heteroge-neous variable magnetic field of physical parameters of magnetostimulation in JPS System, whose area of activity is within ap-prox. 6 cm circle. The applicator coil con-tains:- Red applicator R - 48 red LED diodes
- Infrared applicator IR - 47 infrared LED diodes, one LED red diode
- Combined applicator R+IR - 24 infrared and 24 red LED diodesLight area of magnetic-light applicators
is 19,6 cm2. LED diode terminals touch polycarbonate plate 0,75 mm thick. Plate is place directly on patient’s body. Applicators are fastened with the belt on limbs and with tape on other parts of the body. Synergy of both types of electromagnetic radiation is expected.
11.2.1 Red magnetic-light applicator (R) (wavelength 630 ±5 nm – shal-low penetration) is used mainly for skin disease treatment, such as unhealing wounds and ulcerations (including bedsores),
common acne and zoster. It is recommended in dermatology and cosmetic surgery.
11.2.2 Infrared magnetic-light applicator (IR)(wavelength 855 ±5 nm – deep penetration)in motor system diseases (arthrosis; pain syndromes in discopathy; syndromes
due to strain of muscles and periarticular soft tissue; inflammation of tendons, fascia, tendon sheaths and joint bursa; periarticular inflammation; bone fractures and delayed union of fractured bones) as well as in neurological diseases (neuralgia of peripheral nerves – intercos-tal neuralgia, trigeminal neuralgia, post-zoster neuralgia, sciatica; diabetic neuropathy).
11.2.3 Mixed magnetic-light applicator (RIR)
(wavelength 630/855 ±5 nm – medium penetration) in chronic skin inflammations; keloid, psoriasis, psoriatic joint inflammation, ankylosing
spondylitis, rheumatoid inflammation of the joints, post-thrombotic syndrome.
Technical data and accessories
32
11.3 Maximal induction of magnetic field generated by Viofor JPS System coils
B[µT] B[µT]Type of applicator Mean value Impulse peak value Comments (at setting M1 P2) Large applicatorMat/frame bed 2.5 x I 28 x I Heterogeneous field*
Small applicator - pad 5 x I 56 x I Heterogeneous field
Point applicator S 10 x I 112 x I Heterogeneous field
Point applicator Z 100 x I 1120 x I Heterogeneous field
Elliptic applicator 8.9 x I 100 x I Heterogeneous field(light and green)
Small ring applicator 0.9 x I 10 x I for L Homogenous field**Ø 30 cm 2.7 x I 30 x I for H Large ring applicator 0.9 x I 10 x I Homogenous fieldØ 65 cm
Magnetic-light 8.9 x I 100 x I Heterogeneous fieldapplicator
Accuracy ±10%.I – is the intensity set on Viofor JPS System controller* Heterogeneous field measured on the applicator surface in the area of maximal device readings** Homogenous field** measured in the geometric center of the applicator. Homogeneity with accuracy ± 5 % obtainedin approx. 70 % applicator internal volume.
11.4 Radiation value parameters of magnetic-light applicators connected to Viofor JPS controllerLight area 19,6 cm2.
In red applicators R: 48 red LED diodes, ø5 mm, 2Θ1/2 = 15°
In infrared applicators IR: 47 infrared LED diodes, ø5 mm, 2Θ1/2 = 30°; 1 red diode, ø5 mm, 2Θ1/2 = 15°
In mixed applicators (RIR): 24 red LED diodes, ø5 mm, 2Θ1/2 = 15° and 24 infrared LED diodes, ø5 mm, 2Θ1/2 = 30°
Red and infrared light parameters constitute 50 % of below listed parameter values
Technical data and accessories
33
RED APPLICATOR R
Value Unit 0,5 1 2 3 4 5 6 ÷ 12
Impulse power mW 210 210 210 210 210 210 210
Mean power for P11) mW 10,9 25,2 30,2 31,5 32,3 32,8 33,6
Mean power for P2 mW 9,7 30,7 35,4 36,7 37,5 38,0 39,0
Mean power for P31) mW 9,9 31,8 36,7 38,0 38,8 39,3 40,3
Impulse power density mW/cm2 10,7 10,7 10,7 10,7 10,7 10,7 10,7
K1 impulse energy mJ 0,27 0,36 0,76 0,79 0,81 0,82 0,84
K2 impulse energy mJ 0,10 0,71 0,78 0,79 0,81 0,82 0,84
K1 impulse energy density µJ/ cm2 13,9 32,1 38,6 40,2 41,2 41,8 42,9
K2 impulse energy density µJ/ cm2 5,4 36,4 40,0 40,7 41,2 41,8 42,9
Energy provided during treatment for P11) J 5,2 12,1 14,5 15,1 15,5 15,7 16,1
Energy provided during treatment for P2 J 5,8 18,4 21,3 22,0 22,5 22,8 23,4
Energy provided during treatment for P31) J 5,9 19,1 22,0 22,8 23,3 23,6 24,2
Energy density for P11) J/cm2 0,27 0,62 0,74 0,77 0,79 0,80 0,82
Energy density for P2 J/cm2 0,30 0,94 1,09 1,12 1,14 1,17 1,19
Energy density for P31) J/cm2 0,30 0,98 1,12 1,17 1,19 1,20 1,23
1) for Viofor JPS Classic controller
The table shows light parameters of the red light applicator depending on the intensity set on the Viofor JPS controller.
INFRARED APPLICATOR IR
Value Unit 0,5 1 2 3 4 5 6 ÷ 12
Impulse power mW 1440,00 1440,00 1440,00 1440,00 1440,00 1440,00 1440,00
Mean power for P11) mW 74.88 172.80 207.36 216.00 221.76 224.64 230.40
Continuation on the next page
Technical data and accessories
34
Mean power for P2 mW 66.59 210.59 243.07 251.83 257.24 260.58 267.26
Mean power for P31) mW 67.74 218.42 251.60 260.58 266.11 269.57 276.48
Impulse power density mW/cm2 73.47 73.47 73.47 73.47 73.47 73.47 73.47
K1 impulse energy mJ 1.87 4.32 5.18 5.40 5.54 5.62 5.76
K2 impulse energy mJ 0.72 4.90 5.33 5.47 5.54 5.62 5.76
K1 impulse energy density µJ/ cm2 95.51 220.41 264.49 275.51 282.86 286.53 293.88
K2 impulse energy density µJ/ cm2 36.73 249.80 271.84 279.18 282.86 286.53 293.88
Energy provided during treatment for P11) J 35.94 82.94 99.53 103.68 106.44 107.83 110.59
Energy provided during treatment for P2 J 39.95 126.35 145.84 151.10 154.34 156.35 160.36
Energy provided during treatment for P31) J 40.64 131.05 150.96 156.35 159.67 161.74 165.89
Energy density for P11) J/cm2 1.83 4.23 5.08 5.29 5.43 5.50 5.64
Energy density for P2 J/cm2 2.04 6.45 7.44 7.71 7.87 7.98 8.18
Energy density for P31) J/cm2 2.07 6.69 7.70 7.98 8.15 8.25 8.46
1) for Viofor JPS Classic controller
The table shows light parameters of infrared light applicator depending on the intensity set on the Viofor JPS controller.
Wavelengths of emitted light (λP) are as follows:• 630±5 nm for red light,• 855±5 nm for infrared light.
CAUTION!
STRONG RED AND/OR INFRARED RADIATION!PROTECT YOUR EYES!DO NOT LOOK AT THE LIGHT!
Continuation from page 33
35
TABLE OF CONTENTS1 CHARACTERISTICS OF VIOFOR JPS ................................................................................ 3
1.1 Usage ............................................................................................................................ 32 TYPES OF THERAPY .......................................................................................................... 3
2.1 Magnetostimulation ....................................................................................................... 32.1.1 Characteristics of magnetostimulation .................................................................. 32.1.2 Indications for magnetostimulation ........................................................................ 52.1.3 Contraindications to magnetostimulation .............................................................. 62.1.4 Side effects, which may occur due to magnetostimulation .................................... 7
2.2 Magnetostimulation paired with light energy ................................................................. 72.2.1 Characteristics of magnetostimulation paired with light energy ............................ 72.2.2 Indications for magnetostimulation with light energy ............................................. 82.2.3 Contraindications to magnetostimulation and light energy treatment ................... 92.2.4 Side effects which may occur due to magnetostimulation
with light energy treatment .................................................................................... 93 THERAPY SELECTION ........................................................................................................ 9
3.1 Magnetostimulation ...................................................................................................... 93.1.1 Application parameters components ................................................................... 10
3.1.1.1 Programs (P) ............................................................................................... 103.1.1.2 Application modes (M) ..................................................................................113.1.1.3 Magnetic field intensity ( I ) ..........................................................................113.1.1.4 Selection of application parameters ............................................................ 12
3.1.2 Using more than one applicator .......................................................................... 123.2 Magnetostimulation with light energy ......................................................................... 13
3.2.1 Application parameters components ................................................................... 133.2.1.1 Programs ( P ) ............................................................................................. 133.2.1.2 Application modes ( M ) ............................................................................... 133.2.1.3 Magnetic field intensity ( I ) ......................................................................... 143.2.1.4 Selection of application parameters ............................................................ 14
4 VIOFOR JPS CLASSIC ....................................................................................................... 164.1 General information .................................................................................................... 164.2 Viofor JPS Classic back panel ................................................................................... 164.3 Viofor JPS Classic front panel .................................................................................... 164.4 Remote control unit .................................................................................................... 17
4.4.1 Remote control details ........................................................................................ 174.4.2 Battery installation ............................................................................................... 18
4.5 Apparatus installation ................................................................................................. 184.6 Apparatus operational use .......................................................................................... 184.7 Programming of application parameters .................................................................... 194.8 Retrieving preprogrammed application parameters ................................................... 194.9 PLACEBO mode ........................................................................................................ 194.10 Counter ..................................................................................................................... 20
5 VIOFOR JPS STANDARD ................................................................................................... 20 5.1 General information ................................................................................................... 20 5.2 Viofor JPS Standard back panel ............................................................................... 20
Table of contents
35
36
5.3 Viofor JPS Standard front panel ................................................................................ 205.4 Apparatus installation ................................................................................................ 215.5 Apparatus operational use ........................................................................................ 21
6 MAGNETIC & LIGHT THERAPY ......................................................................................... 226.1 General information .................................................................................................... 226.2 ZAMS 01/2 power supply unit .................................................................................... 22
6.2.1 ZAMS 01/2 front panel ........................................................................................ 226.2.2 ZAMS 01/2 back panel ........................................................................................ 22
6.3 Magnetic-light applicators ........................................................................................... 236.4 Magnetic & Light Therapy set usage .......................................................................... 23
7 SAFE USAGE OF VIOFOR JPS ........................................................................................ 248 CLEANING, MAINTENANCE, REPAIRS ............................................................................ 24
8.1 General information ....................................................................................................... 248.2 Service pack .................................................................................................................. 25
8.2.1 Fuse exchange ...................................................................................................... 258.2.2 Exchange of applicator plug and adapter .............................................................. 25
9 WARRANTY TERMS .......................................................................................................... 2510 TECHNICAL DATA ............................................................................................................ 27
10.1 Controllers ................................................................................................................ 2710.1.1 Compatibility ...................................................................................................... 2810.1.2 Casing ............................................................................................................... 2810.1.3 Mass .................................................................................................................. 28
10.2 Power supply unit ZAMS 01/2 .................................................................................. 2810.3 Magnetic-light applicators ......................................................................................... 29
11 APPLICATORS .................................................................................................................. 2911.1 Applicators for magnetostimulation .............................................................................. 29
11.1.1 Large applicator - mat ........................................................................................ 2911.1.2 Large applicator – short mat .............................................................................. 2911.1.3 Large applicator - frame bed ............................................................................. 3011.1.4 Small applicator - pad ........................................................................................ 3011.1.5 Point applicator (S and Z ) ................................................................................. 3011.1.6 Elliptic applicator (white and green ) .................................................................. 3011.1.7 Ring applicator – small, 30 cm diameter ........................................................... 3011.1.8 Ring applicator – large, 65 cm diameter ............................................................ 31
11.2 Magnetic-light applicators ......................................................................................... 3111.2.1 Red magnetic-light applicator (R) ...................................................................... 3111.2.2 Infrared magnetic-light applicator (IR) ............................................................... 3111.2.3 Combined magnetic-light applicator (RIR) ......................................................... 31
11.3 Maximal induction of magnetic field generated by Viofor JPS System coils ............. 3211.4 Radiation value parameters of magnetic-light applicators connected to Viofor JPS controller ............................................................................ 32
AAVI-EN-I-04/2005
Table of contents
Technical data and accessories
37
Viofor JPS has won many prestigious national and international awards, such as:- award of “Polish Product of the Future 2000”- gold medal at the Health Spa and Rehabilitation Fair “Zdroj 2001”- awards at 50th World Exhibition of Development, Research and New Technologies
“Brussels Eureka” in Brussels:• gold medal and distinction• special award of the Italian Delegation • special award of Croatian Society of Inventors INOVA
- gold medal at „Concours Lepine” Fairs 2002 Paris- gold medal at IENA 2002 Fairs in Nuremberg- gold medal at the International Fair “Invest Technology 2002” in Warsaw- metal at Intarg Fair 2002 in Katowiceand many others.
Awards and honors
38
Med & Life, the manufacturer of Viofor JPS system has a certified Quality System according to ISO norms: ISO 9001:2000 and ISO 13485:2003.
Med & Life is also a recipient of many prestigious awards: - Leader of Entrepreneurship 2000- Cup and Diploma of the Minister of Education
and President of KBN for innovative activity- Leader of Innovation 2002- Award at “Huzar Europy 2003” competition.
For more information on the manufacturer see: www.medandlife.com
Awards and honors
39
Additional informations
EC CONFORMITY DECLARATIONMed & Life herewith declares, that:Systems for therapy and rehabilitation• Viofor JPS System Standard with magnetic field, with or without
light source (medical product class IIa - Annex IX, Rule 9)• Viofor JPS System Classic with magnetic field, with or without
light source (medical product class IIa - Annex IX, Rule 9)• Viofor JPS System Clinic with magnetic field, with or without
light source (medical product class IIa - Annex IX, Rule 9)• Viofor JPS System Delux with magnetic field, with or without
light source (medical product class IIa - Annex IX, Rule 9) conform the essential requirements and the guidelines of the Council Directive
93/42/EEC - Medical Devices and with harmonized standards:– EN IEC 60601-1 „Medical electrical equipment” with the supplements
and linked standards.– PN-EN 60825:2000/A2:2002 (U) Safety of laser products. Equipment classification,
requirements and user`s guide.– is a subject of the procedure mentioned in the Annex II of the Directive 93/42/EECMed & Life has implemented the Quality Management System according to ISO 9001:2000/ISO 13485:2003 under the surveillance of the Notified Body nr 0120, SGS United Kingdom Ltd Systems & Services Certification.
40