VAPORIZERS
Dr. deepak kumar Resident in Anaesthesia deptt
S.P.Medical College & A.G.Hospitals, Bikaner
![Page 2: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/2.jpg)
What is a Vaporizer?
A VAPORIZER IS AN INSTRUMENT DESIGNED TO FACILITATE THE CHANGE OF A LIQUID ANAESTHETIC AGENT INTO A VAPOR
AND ADD A CONTROLLED AMOUNT OF THIS VAPOR TO THE
FGF.
![Page 3: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/3.jpg)
Function of Vaporizers
• Produce vaporisation of volatile agent• Mix vapour with fresh gas flow• Control the mixture despite variables
►To deliver safe and accurate concentration of inhalational agents to the patient.
![Page 4: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/4.jpg)
Applied Physics
• Factors affecting vaporization of a liquid
1.Temperature
2.Volatility
3.Surface area
4.Removal of vapor from the vicinity of the liquid.
![Page 5: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/5.jpg)
Applied Physics • Vapor• Gas • Critical temperature• Vapor pressure• Boiling point• Partial pressure • Heat of vaporization• Specific heat• Thermal conductivity• Thermal capacity
![Page 6: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/6.jpg)
Vapor
• Vapor is the gaseous phase of a substance which is normally a liquid at room temp. and atm. pressure
Gas
• Gas is a substance which exists only in the gaseous state at room temp. and atm.
![Page 7: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/7.jpg)
Critical temperature
• It is the temperature, above which no amount of pressure will convert a gas to a liquid.
02=-118.4°C; N2O= 36.5°C; CO2= 31°C.
• The pressure required to liquefy a gas at its critical temp. is the CRITICAL PRESSURE .
![Page 8: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/8.jpg)
Vapor pressure
• When enclosed in a container molecules of a volatile liquid break away to form vapor.
• Vapor pressure is the pressure with which they bombard the walls of the container.
• Vapor pressure depends only on temp and nature of liquid.
• Vapor pressure of an agent determines how much of vapour will be formed from 1 ml of the liquid.
• Since diff anaesthetic agents have diff vapour pressure so need separate vaporizers.
![Page 9: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/9.jpg)
Boiling point
• Boiling point of a liquid is that temp at which the vapor pressure is equal to the atmospheric pressure.
• Lower the atmospheric pressure, lower the boiling point.
![Page 10: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/10.jpg)
Partial pressure
• The part of total pressure due to any one gas in the mixture
• Depends on – Nature of liquid and Temp.• Clinical significance-
Patient uptake and anaesthetic depth are directly related to partial pressure.
![Page 11: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/11.jpg)
• HEAT OF VAPORIZATION-
It is the number of calories needed to convert 1gm or 1ml of liquid to vapor.
• SPECIFIC HEAT-
Quantity of heat required to raise the temp of 1gm or 1ml of substance by 1*c .
![Page 12: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/12.jpg)
• THERMAL CONDUCTIVITY-
A measure of speed with which heat flow through a substance .
• THERMAL CAPACITY-
Amount of heat stored in vaporizer body
= specific heat x Mass
![Page 13: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/13.jpg)
CLINICAL SIGNIFICANCE-• liquid with low specific heat vaporize easier.• Higher the thermal conductivity, better a
substance conduct heat• Vaporizers construct of material with high
thermal conductivity and specific heat e.g.- copper, bronze to minimise temp changes when in use.
![Page 14: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/14.jpg)
Related physics
THERMOSTABILISATION Methods :-
Vaporizer constructed of metal with high thermal conductivity.
Heavy metal parts act as heat reservoir. Wicks to be in contact with the metal part so that
heat loss due to vaporization is quickly replaced. Immerse vaporizer chamber in a large mass of water.
![Page 15: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/15.jpg)
Related physicsTHERMOCOMPENSATION-• Some means to maintain the vaporizer
output constant despite any temp changes
Methods -
1.Alteration in splitting ratio(automatic compensation) e.g.- bimetallic strip in Tec vaporizers, ether filled bellows in penlon vaporizers, EMO(Epstein Macintosh Oxford)
![Page 16: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/16.jpg)
Related physics
2.Computer control-electronic vaporizers.
3.Manually adjust flow- measured flow , drager vapor.
4.Supplied heat- Tec 6(electrically heated).
![Page 17: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/17.jpg)
Properties Of Common Anaesthetic Agents
Agent Boiling Point at 100 kPa(degree Celsius)
Vapor Pressure(at 20 deg C)
MAC value(vol% ) in 100% O2
1) Ether 34.5 440 19
2) Halothane 50.2 243 0.75
3) Enflurane 56.5 175 1.68
4) Isoflurane 48.5 238 1.15
5) Desflurane 22.8 669 6.4
6) Sevoflurane 58.6 157 2.0
7) Trilene 87.5 57 0.17
![Page 18: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/18.jpg)
Terminology for vaporizers
PLENUM- FGF is pushed in to the vaporizer(high resistance).
DRAWOVER- Gas is pulled in to the vaporizer by the patients own inspiratory effort(low resistance) e.g.- Goldman,EMO,OMV
INHALER- A drawover vaporizer in which the carrier is air.
![Page 19: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/19.jpg)
SPLITTING RATIO• The ratio of the bypass gas to the gas going to
the vaporizing chamber is called the splitting ratio
• It depends on-
1.The resistance of the two pathways, which inturn depends on the variable orifice of the inlet/outlet.
2.Temp of the liquid/carrier gas
3. Flow rate of gases.
![Page 20: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/20.jpg)
CLASSIFICATION
• Dorsch & Dorsch classification (1979)
1) Regulation of output concentration
2) Method of vaporization
3) Location of the vaporizer
4) Temperature compensation
5) Specific anaesthetic agent
![Page 21: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/21.jpg)
CLASSIFICATION
• Dorsch & Dorsch classification
1) Regulation of output concentration
a) Variable Bypass type
b) Measured flow type
2) Method of vaporization
3) Location of the vaporizer
4) Temperature compensation
5) Specific anaesthetic agent
![Page 22: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/22.jpg)
Flowmeter
Patient
VARIABLE BYPASS
![Page 23: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/23.jpg)
Flowmeter
Patient
![Page 24: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/24.jpg)
Flowmeter
Patient
![Page 25: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/25.jpg)
MEASURED FLOW VAPORIZER
• Uses a measured flow of carrier gas to pick up agent
• Consist of-
1.Vaporizer
2.flowmeter assembly
3.Vaporizer circuit control valve
![Page 26: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/26.jpg)
Cont...
• Operator has to set the flow to the vaporizer and bypass with separate flowmeters
• This means that respective flows have to be calculated for a given temp and vapour output.
![Page 27: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/27.jpg)
Measured Flow Vaporizer
V V
C
L
![Page 28: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/28.jpg)
CLASSIFICATION
• Dorsch & Dorsch classification
1) Regulation of output concentration
2) Method of vaporization
a) Flow over
b) Bubble through
3) Location of the vaporizer
4) Temperature compensation
5) Specific anaesthetic agent
![Page 29: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/29.jpg)
Flow over
• A stream of gas passes over liquid surface • Efficiency depend on-
1. Gas-liquid interface e.g.- baffles, spiral tracks, wicks.
2. Velocity of carrier gas flow
3. Height of gas
![Page 30: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/30.jpg)
Bubble through
• Bubble the gas through the liquid • efficiency depends on-
1. size of bubbles
2. depth of the liquid
3. velocity of carrier gas
E.g.- sintered diffuser
-cowl in Boyle’s bottle
![Page 31: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/31.jpg)
BUBBLE THROUGH VAPORIZER
![Page 32: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/32.jpg)
CLASSIFICATION
• Dorsch & Dorsch classification
1) Regulation of output concentration
2) Method of vaporization
3) Location of the vaporizer
a) VOC(Vaporizer outside the circuit)
b) VIC (vaporizer inside the circuit)
4) Temperature compensation
5) Specific anaesthetic agent
![Page 33: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/33.jpg)
In system vaporizer(VIC)
• Should have standard male and female 22 mm fittings or standard screw threaded fittings,inlet and outlet ports should have arrows.
• Should have low resistence.• Vaporiser concentration and output vary• Easy to wash and clean• Drawover vaporizers eg- goldman,EMO,OMV
![Page 34: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/34.jpg)
OUT OF SYSTEM VAPORIZERS(VOC)
• Usually on back bar – b/w flowmeter and FGF outlet
• Most variable bypass vaporizers and all measured flow vaporizers are VOC type.
![Page 35: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/35.jpg)
CLASSIFICATION
• Dorsch & Dorsch classification
1) Regulation of output concentration
2) Method of vaporization
3) Location of the vaporizer
4) Temperature compensation
a) Non compensated
b) Compensated
5) Specific anaesthetic agent
![Page 36: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/36.jpg)
EFFECTS OF TEMPERATURE
![Page 37: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/37.jpg)
GIVING HEAT
![Page 38: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/38.jpg)
![Page 39: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/39.jpg)
![Page 40: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/40.jpg)
TEMP COMPENSATION
• To maintain a constant output from the vaporizer,mechanisms to compensate for the fluctuations in temp are to be employed.
• Method –
1. automatic- alteration in the splitting ratio(bimetallic strip in tec vaporizer,ether filled bellows in penlon vaporizers,EMO
2.Supplied heat- tec 6(electrically heated)
3.Computer control- electronic vaporizers
![Page 41: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/41.jpg)
TEMPERATURE COMPENSATED
![Page 42: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/42.jpg)
TEMPERATURE COMPENSATED
![Page 43: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/43.jpg)
BIMETALLIC STRIP
![Page 44: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/44.jpg)
BIMETALLIC STRIP
![Page 45: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/45.jpg)
TEMPERATURE COMPENSATED
![Page 46: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/46.jpg)
CLASSIFICATION
• Dorsch & Dorsch classification
1) Regulation of output concentration
2) Method of vaporization
3) Location of the vaporizer
4) Temperature compensation
5) Specific anaesthetic agent
a) Agent specific
b) Multiple agent
![Page 47: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/47.jpg)
CLASSIFICATION
• Gray & Nunn Classification (1971)
1) Plenum vaporizers
2) Draw over vaporizers
3) Vaporizers for use inside a circle anaesthetic system
![Page 48: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/48.jpg)
CLASSIFICATION
• Gray & Nunn Classification
1) Plenum vaporizers
Fresh gas flow is forced into a chamber
Unidirectional gas flow
Relatively high resistance to gas flow
e.g. Boyle vaporizer, copper kettle,
Tec series vaporizers
![Page 49: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/49.jpg)
CLASSIFICATION
• Gray & Nunn Classification
2) Draw over vaporizers/ Inhalers: Air or anaesthetic gases drawn over or through the vaporizer either by pt’s own respiratory efforts or by a self inflating bag or manual bellows.
Very low resistance to gas flowe.g. EMO vaporizer, Oxford
miniature inhaler, emotril, tecota vaporizers.
![Page 50: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/50.jpg)
CLASSIFICATION
• Gray & Nunn Classification
3) Vaporizers for use inside a circle anaesthetic system
e.g.- Goldman vaporizer
Rawbotham vaporizer
![Page 51: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/51.jpg)
NEWER CLASSIFICATION.
METHOD OF REGULATING OUTPUT CONCENTRATION(1) Conc. Calibrated(2) Measured flow
METHOD OF VAPORISATION(1) Flow over(2) Bubble through(3) Injection
TEMP. COMPENSATION(1)Thermocompensation(2) Supplied heat
![Page 52: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/52.jpg)
Factors affecting vaporizer performance
• Flow rate• Temperature• Barometric pressure• Intermittent back pressure• Gas direction• Liquid levels• Anaesthetic agents
• Carrier gas composition (N2O causes transient drop)
![Page 53: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/53.jpg)
Factors affecting vaporizer performance
• Flow rate At high flows,the vaporizer delivers less
anaesthetic concentration than is set on the dial
problem is solved by increase the surface area of contact b/w the fresh gas and anesthetic agent
Eg- WICKS
![Page 54: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/54.jpg)
TEMPRATURE• As anaesthetic molecule escape temp
decreases more difficult for the remaining molecules to escape less vaporisation.
• Thus at lower temp there is less vaporization.
• Problem solving- Giving heat Giving flow
![Page 55: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/55.jpg)
Giving heat
1.Vaporizer material
good conductor(high thermal conductivity)
act as a heat reservoir(high specific heat)
2.Supplied heat (tec-6 electrically heated)
![Page 56: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/56.jpg)
Giving flow
• Giving flow (thermocompensation) by alteration in splitting ratio
1. Mechanically
2. Compuer controlled
![Page 57: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/57.jpg)
Factors affecting vaporizer performance
• Intermittent back pressurePumping effect(hill and lowe effect)Pressurizing effect(cole effect)
![Page 58: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/58.jpg)
Pumping effect(Hill and lowe effect)
• The effect of changing pressure during IPPV increasing the output of the vaporizer is called the “pumping effect”
• During positive pressure ventilation, pressure transmitted back results in compression of gas.
• Since the vaporizing chamber volume is much larger than the ‘by pass’ channel volume, more fresh gas gets compressed in to it.
![Page 59: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/59.jpg)
Cont…
• This extra fresh gas that enters the vaporizing chamber collects anesthetic vapour
• Some of the rapidly expanding gas (containing vapour) enter the inlet of the vaporizer and cross over into the ‘bypass’ channel.
• The addition of the ‘bypass’ vapour to the vaporising chamber raises the final concentration of anesthetic delivered.
![Page 60: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/60.jpg)
Pumping Effect
![Page 61: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/61.jpg)
Pumping Effect
![Page 62: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/62.jpg)
Pumping Effect
![Page 63: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/63.jpg)
Pumping Effect
![Page 64: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/64.jpg)
Pumping Effect
![Page 65: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/65.jpg)
Pumping Effect
![Page 66: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/66.jpg)
Modifications to reduce the ‘pumping effect’
• By pass channel made larger• Inlet tube made longer• Exclude wicks from inlet• Increased resistance – a high internal
resistance to “resist” changes to flow.• Check valve at outlet of the vaporizer -Fluotec
2 • Check valve upstream to junction with the
oxygen flush.• Pressure relief valve.
![Page 67: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/67.jpg)
Large bypass channel
![Page 68: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/68.jpg)
Long inlet tube
![Page 69: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/69.jpg)
Increased resistance
![Page 70: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/70.jpg)
One way valve (open)
![Page 71: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/71.jpg)
One way valve (closed)
![Page 72: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/72.jpg)
Pressurising effect
• k/a ‘cole effect’ – decreased output of anesthetic agent.
• An inreased in pressure causes an increased pressure inside the vaporizer.The carrier gas is compressed but the vapour pressure of the volatile anesthetic is unaffected.Net result decreased conentration of anesthetic delivered.
![Page 73: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/73.jpg)
Pressurizing Effect
![Page 74: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/74.jpg)
Carrier gas composition
• Most vaporisers are calibrated using 100% o2• Composition of carrier gas affects output in
many(vaporiser aberrance)• If add N2O- decreased output(25% less with
100% N2O)due to solubility of N2O in agent .As N2O dissolves in liquid anaesthetic.Flow of gases exiting vaporiser decreases.
• Once saturated with N2O output gradually increases but is less than before(10% less with 100% N2O.
![Page 75: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/75.jpg)
EFFECT OF BAROMETRIC PRESSURE
• Vaporizers are calibrated at std. atmospheric pressure(at sea level)
• Low boiling point,high SVP anaesthetic volatile agents- more susceptible to influence by barometric pressure.
• Vapour pressure of agent is independent of barometric pressure.
• Effect of low atm. Pressure- deliver HIGHER concentration if measured in volume% but deliver same partial pressure
![Page 76: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/76.jpg)
Cont….
• Clinically effect unchanged
• Effect of high atm.pressure-
increased density of gas increased resistance through vapourising chamber decreased vapour output both in partial pressure and volume%
![Page 77: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/77.jpg)
SPECIFIC VAPORIZERS
![Page 78: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/78.jpg)
Early methods• OPEN DROP METHOD:-
Inhalational anaesthesia by vaporization of a liquid anaesthetic placed drop by drop on a gauze/ mask covering the mouth and nose.
Devices- Schimmelbusch mask
other modification- Yankauers, Bellamy gardner
![Page 79: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/79.jpg)
CONT..
• SEMI OPEN-
a frame added to “keep the ether in” in an enclosed area- permitted some degree of rebreathing.
e.g.- Ogston’s inhaler
Flagg’s can
![Page 80: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/80.jpg)
Early devices- open drop method(schimmelbusch mask,
yankaeur mask and bellamy gardner wire mask)
![Page 81: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/81.jpg)
Semi open drop method
Ogston mask with schimmelbusch frame
![Page 82: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/82.jpg)
MORTONS ETHER INHALER
![Page 83: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/83.jpg)
OPEN ETHER ADMINISTRATION
• TECHNIQUE- Volatile anaesthetic dripped on to a gauze:1. ETHER- 16 layers gauze over mask2. CHLOROFORM/ETHYL CHLORIDE- 12 layers of gauze/1
layer lint(chloroform dropped over half the area)- gradually increased no. Of drops/min- During inspiration air passed through the gauze and
vaporizes the liquid anaesthetic into high concentration.
![Page 84: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/84.jpg)
Open drop ether
• Bellamy gardner dropper
- amber coloured
- control on pouring
- capacity – 90 ml ether
![Page 85: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/85.jpg)
INDUCTION• WITH ETHER • RATE OF DROPS 1ST min = 12 drops = 1 % 2nd min = 25 drops = 3 % 3rd min = 50 drops = 6 % 4th min = 100 drops = 10-12 %
• ETHYL CHLORIDE - 3 to 5 ml - 3 to 5 %• Rate of drops 1st min = 30 drops 2nd min = 60 drops 3rd min = 90 drops
![Page 86: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/86.jpg)
MAINTENANCE• Conc.for maint. with ether is 6 -8 %• Heat loss = 200-300 cal/min• Temp. above and below mask = 2-3 degrees <
room temp.• Temp. at mask = 0 – 1 degrees C
• Gas comp.under mask 0% ether = 80% N2 + 20% O2 5% ether = 76% N2 + 18% O2 10% ether = 72% N2 + 16% O2
![Page 87: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/87.jpg)
Advantage of open drop
• Easy to administer• Low dead space 40-60 ml• Low resistence• Wide margin of safety • Relatively cheap
![Page 88: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/88.jpg)
Disadvantage of open drop• Significant rebreathing• Hypoxic mixtures may occur• Poor control of inspired gas concentration• Inability to assist or control ventilation• No conservation of heat or humidity• Difficult airway management especially during
head and neck procedure• Pollution of the operating room• Hazardous especially with flammable agents.• Skin burn.• Eye injury
![Page 89: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/89.jpg)
In system vaporizers(VIC)
• There are two ways that gas flows through a vaporizer
- push through
-Drawover
![Page 90: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/90.jpg)
Drawover anaesthesia
• Drawover system :
- provide anaesthesia without a supply of compressed gases.
-Atmospheric air – main carrier gas
-drawn by the patient’s inspiratory efforts
- volatile agent (ether or halothane) added to vaporizer
- Inhaled by the patient via a non-rebreathing valve.
![Page 91: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/91.jpg)
Drawover anaesthesia
• The component of a drawover circuit
![Page 92: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/92.jpg)
EMO Vaporizer Epstein, Macintosh, Oxford (EMO) introduced
in 1952 Classification-• Variable bypass/conc. calibrated• Flow over with wicks • Temperature compensated by supplied heat and
altered flow. • Agent specific-used for ether, halothane, chloroform
and trilene.• Can be a part of a drawover system or used as a
plenum vaporizer• Low resistance(<1.25cm water at 40 Lpm flow)
![Page 93: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/93.jpg)
EMO VAPORIZER
![Page 94: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/94.jpg)
• Wt- 6.5 kg ; ht 24cm ;dia 23cm
• TRANSIT position- seals ether chamber
• CONTROL lever-upto 20%
• INLET/OUTLET – R to L
• TAP for filling /draining water chamber at bottom
• Outlet(male) inlet(female)
• 1250ml water filled in water chamber
![Page 95: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/95.jpg)
A- Inlet B- Outlet C- Water compartment D- Ether E- Vaporizing chamber F- Thermo compensating
valve G- Off/on valve H- Mixing chamber I- Water drain
![Page 96: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/96.jpg)
Ctd• FILLER-depress to fill (control lever at 0-not
transit- for air to escape) springs back automatically.
• LEVEL INDICATOR- moves only after 150 ml ; add 300ml for full (fill with control at 0 –not at ‘in transit’.
• TEMP.INDICATOR-rod with black & red bands and metal top
20-25 degrees-black line with metal top >32 degrees – red band- temp above
working range
![Page 97: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/97.jpg)
EMO(ctd)• Thermocompensation mechanism at outlet of
v.c.– metal bellows with liquid Ether[ether capsule] &
connected to plunger– temp. range; 15-29 degree Celsius
• Water jacket serves as heat reservoirChecks(1) check level indicator-> put “in transit”-> invert –
indicator should fall to full.(2) close outlet- connect OIB to inlet-> put “in transit”->
press bellows-> open filler –no air should escape.(3) release filler-set at 10% -rpt above(4) attach bellows to outlet->block inlet –> set at 2 %
-suck air –> should hear a hissing if safety release valve is working
![Page 98: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/98.jpg)
EMO(ctd)• Care-Mark I--empty Al water jacket every 3 months, Mark II &
III- yearly water check EVALUATION 1. Calibration of EMO is accurate only for intermittent gas
flows; maintains output at 5-13L/min flows. Highest conc. delivered 16%. If use as plenum i.e. blow air into it –increase output
2. Climate; Cool-add antifreeze (2% glycol) Warm- cool by allowing agent to vaporise -refrigerate -air will deposit water in cooler vc 3. Splashing during transit if in ON position. 4. Sticking of rotor-PTFE coating in Mk4 (Stetson) 5. Advantage- compact, low cost, portable, useful in mass
casualties, no effect of altitude, easy maintenance, no need for sterilisation
![Page 99: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/99.jpg)
OXFORD INFLATING BELLOWS
• Self inflating bellows used with spontaneous/controlled ventillation
• Bellows sit vertically,internal volume maintained by a spring; 6 bellows- 150 ml each
• 2 unidirectional flap valves• Magnet to inactivate distal unidirectional
valve.
![Page 100: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/100.jpg)
Oxford inflating bellows
![Page 101: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/101.jpg)
Spontaneous ventilation
![Page 102: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/102.jpg)
Assisted ventilation
![Page 103: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/103.jpg)
Assisted ventilation
![Page 104: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/104.jpg)
Oxford miniature vaporizer(OMV)
Introduced by Epstein, Macintosh and Mendelssohn in 1941.
• Conc. Calibrated• Draw over vaporizer (low resistance)• Flow over with wicks• Temperature regulated by means of calcium
chloride( supplied heat)• Outside the circuit, can be used as plenum
vaporizer
![Page 105: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/105.jpg)
OXFORD MINIATURE VAPORISER(OMV)
• Simple portable inhaler• for less volatile agents – halo, trilene, chloroform• Fairly accurate over a short period of time• 13.5 cm high,1060 gms with full water jacket.• Control lever, alternative scales for halo(0-4%),tri(0-1.5%),
methoxy (0-0.6%)• water jacket at base with 25% glycol• Body stainless steel/wicks of stainless steel gauze• Plugged into outlet of EMO-performance unaffected by IPPV- can
place on pt side of bellows• Highest conc delivered 3.5% hal
![Page 106: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/106.jpg)
OMV (ctd)
• Special filler with 2 springs
light pressure-air relief
more pressure-opens filler • Funnel around filler has capacity of 10ml, covers 1/8th of level
indicator. A second 10ml can be added• cleaning-drain by tipping after pressing filler lever, wash out
with alcohol or Ether.• If used with EMO flow is R to L • Another version for use with continuous flow machine
then flow is L to R• direction of gas flow marked with an arrow
Disadvantage of OMV-only 20 ml ; cannot mount on backbar
![Page 107: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/107.jpg)
Goldman Vaporizer
• Classification
- concentration calibrated
- flow over without wick
- no temp compensation
- multiple agents – halothane,trilene
- in or out of system
![Page 108: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/108.jpg)
Goldman vaporizer
• Small glass bowl• Capacity 20 ml• Bowl attached to a head, which divide gas b/w
bypass and vaporizing chamber• Control lever at top; max conc. Delivered at 3rd
mark of 2.21%• Young modification- added a wick(increase to
4%)• Halls modification – 2 in series
![Page 109: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/109.jpg)
GOLDMAN VAPORIZER
MARK I MARK II MARK III
1. Self locks Click stops No lockingin off position in each setting
2. DIVISIONS Off-1-2-3-ON Off-1-2- ONOff -1-2-3-ON
3. Max conc Max conc Max concdelivered at delivered at delivered at3 position. 3 position. On.
![Page 110: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/110.jpg)
ROWBOTHAM VAPORISER
• Has a wire gauze wick
• 2marks to fill till
• Top mark and blue mark
• Max. at full on 3.10%
![Page 111: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/111.jpg)
ADVANTAGES
• Portable• Easy to operate• Low resistance-used as VIC• Calibrated at high flows of 30 Lpm so safely
use with O2 flush• Small, inexpensive• Safe- cannot deliver high conc.’s
![Page 112: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/112.jpg)
DISADVANTAGES• No temperature compensation- Level of
halothane kept at full mark• Tilting - pouring of liquid in respiratory
tract• Back pressure or pumping effect• Small capacity vaporising chamber - so
delivers low halothane concentration.• agitation/splashing -5%
![Page 113: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/113.jpg)
Boyle’s bottle vaporizer
• Classification• 1.Variable bypass• 2.Flow over or bubble through• 3.Not temperature
compensated• 4. Agent specific
(ether,halothane,trilene)• 5.Outside the circuit
![Page 114: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/114.jpg)
![Page 115: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/115.jpg)
Boyle’s bottle
![Page 116: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/116.jpg)
![Page 117: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/117.jpg)
Boyle Bottles• Ether Bottle • Larger vc-300 ml filled fully• U tube & hood of Cu• Has 4 lines between off & on-begins to operate
at 2nd mk• Trilene bottle -100ml for ½ inch liquid depth• Chrome plated U tube& hood; cowl adjusted by
stainless steel plunger• Delivers 0.5-2 %
![Page 118: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/118.jpg)
Boyle Bottles
• Halothane bottle• Uses only control tap –no plunger/hood• Control lever marked 1-10 (8%)
starts at 3 , at 4 about 1%• Inlet tube plugged at end; hole on side 1
cm above
![Page 119: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/119.jpg)
BOYLES BOTTLE• FACTORS AFFECTING OUTPUT
1. Temp. of liquid
2. Plunger level
3. Control lever position
4. Level of liquid
5. Eccentricity of hood
6. Agitation of vaporiser as during pouring of liquid in bottle(>5%x 15 secs)
![Page 120: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/120.jpg)
BOYLES BOTTLECARE & CLEANING• Empty after use/allow to dry• special grease for free rotation of drum• Plunger loose- tighten the gland nut• replace packing in gland nut- cotton, neoprene,
nylon• bottle may chip off leading to leakage• bottle washer may get damaged• pressure build up in unused ether bottle• static charges on cork-chain
![Page 121: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/121.jpg)
Copper Kettle vaporizer Described by Lucein Morris in 1952.
Classification-
• Measured flow
• Bubble through
• Temperature compensated by supplied heat and manual flow alteration.
• Multiple agent ( chloroform, ether,halothane)
• Outside the circuit
• 2 models 160ml/400ml
![Page 122: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/122.jpg)
Copper kettle vaporizer
• Constructed of copper
-High heat capacity
- high thermal conductivity
-High degree of accuracy
![Page 123: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/123.jpg)
Copper kettle vaporizer
![Page 124: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/124.jpg)
• A- Filling funnel• B- Drain• C- Inlet • D- Sintered bronze
porex disc• E- Outlet
![Page 125: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/125.jpg)
Obstetric inhalers• Emotril• Cyprane
• Provide TV 250-1000 ml • work over RR’s of 12-30/min• resistance of breathing to be
<1.25 cm H2O at 30 LPM.
![Page 126: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/126.jpg)
TEC VAPORIZERS
• CLASSIFICATION (TEC 1 to 5)1.Variable bypass2.Flow over with wick3.Out of system4.Temp. Compensated by automatic flow
alteration5.Conc. Calibrated6.Agent specific
![Page 127: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/127.jpg)
TEC 2
• Used only for Halothane & methoxyflurane• Capacity 150 ml• Calibrated upto 4% (in increments of 0.5% )•Temp. compensation by bimetallic strip• Filling tap at side, draining at bottom •Level indicator on side•Conc. Dial in front, attached with a spindle
![Page 128: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/128.jpg)
TEC 2• Not accurate below 4L/min
• <2L/min flow and <2% dial setting->delivers less
• >2% dial setting -> deliver more
• With N2O it gives greater output at lower setting and less output at higher setting
• Prone to pressure changes:
pumping effect at low flows
pressurizing effect at high flows .
![Page 129: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/129.jpg)
• Care and cleaning
Drain halothane every 2 week and discard as THYMOL accumulates sticking of spindle and bimetallic strip.
HAZARDS-
1. Tipping
2. Agitation high output
3. Reverse flow:->back pressure changes
4. Sticking of control dial due to thymol
5. Between off and 0.5% some output can occur varying with FGF.
6. Small leak in off position.
![Page 130: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/130.jpg)
TEC 3
Construction:-1.Conc. Control dial is on top
2.Calibrated from off to 5% in 0.5% gradations
3.Locking lever to be depressed before dial can be turned
4.Screw cap filler with drain at bottom
5.Optional pin safety system for filling
6.Sight window for liquid level on left
![Page 131: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/131.jpg)
TEC 3
•Used for Halothane, enflurane, isoflurane & sevoflurane
•Capacity : 135ml (with dry wicks)
100ml ( with wet wicks)
•A bimetallic strip increases flow through the bypass chamber when temperature increases.
•Negligible back pressure changes
•No non return valve
•Less affected by fresh gas flow and composition of carrier gas
•Control knob less likely to stick
![Page 132: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/132.jpg)
Cont...Internal structure:• Completely redesigned• Has 2 sections- lower vaporizing chamber and upper
duct and valve system• 2 bypass channels- one direct gas stream over
bimetallic strip• Bimetallic strip at inlet of 2nd bypass• Gas exits VC by way of the control channel and joins
gas coming from the bypass• Bypass is located concentrically within the vaporizing
chamber.
![Page 133: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/133.jpg)
![Page 134: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/134.jpg)
TEC 3• Accur. falls off at high
flow rates & dial settings
• All are accurate with low dial settings
![Page 135: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/135.jpg)
TEC 3EVALUATION • Sudden increase or decrease in carrier gas flow,
intermittent back pressure and upstream O2 flush has negligible effect on vapor output.
• N2O has got little effect on output• Performance in 0-0.5 % range governed mainly
by conc. dial & less by FGF.• Upto 90° tipping has no effect
HAZARDS: • FAULTY LOCKING LEVER.
• TIPPING TO 180 DEGREE INCREASES CONCENTRATION DELIVERED TO > 12%.
• LEAKS SMALL AMOUNT OF VAPOUR IN OFF POSITION.
• REVERSE FLOW INCREASES OUTPUT.
![Page 136: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/136.jpg)
TEC 4Used for halothane, enflurane and isoflurane Entire new look-> Select-a-Tec manifoldCapacity 135ml (with dry wicks)
100 ml ( with wet wicks)Graduation from 0-5% ( in 0.25% increments from 0 -1%, and 0.5% increments thereafter)Depress release button on left of control dial to turn on the vaporizerLocking lever on rear- Vaporizer can be turned on only if locked on manifold.Two filling mechanism
1. screw cap with drain plug
2.keyed filling device
![Page 137: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/137.jpg)
Improvement over tec 3• Output unaffected by back pressure changes under
clinical conditions• Unaffected by tipping even upto 180°
Limitations-
1. Excess pressure(>400 mmHg) cause decrease in output.
2.Not so accurate at low flow rates, low dial settings and larger pressure fluctuations.
3.Overfilling possible.
4.Use of N2O decreases output.
5. Difficulty in operation one handed.
![Page 138: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/138.jpg)
TEC 4
![Page 139: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/139.jpg)
TEC 5Used for Halothane, isoflurane , enflurane & sevofluraneCapacity 300ml ( with dry wicks)
225ml (with wet wicks)Graduation 0-5% ( 0.2% increments from 0-1% and 0.5% increments thereafter)
Features-
1.Top control dial
2.Locking lever
3. Release button at rear of dial
4.Sight glass – bottom right
5.Keyed filling device:-FILLING DRAINING PORT
-LOCKING LEVER TO SECURE FILLER BLOCK
-SMALL LEVER AT BASE ALOWS LIQUID TO BE ADDED OR DRAINED
![Page 140: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/140.jpg)
Tec 5Features :• Internal baffle system• VC lies within the
bypass, which lies along side of the vaporizer.
• Bimetallic strip at the base in bypass.
• Before reaching VC –helical IPPV assembly--spiral wick.
![Page 141: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/141.jpg)
TEC 5• Introduced in 1989 responding to criticism of the Tec 4.
Improved features of TEC 5:-
1. Bypass chamber at the base; an improved bimetallic strip.
2. Improved safety interlock.
3. Agent capacity increased from 125 ml to 300 ml.
4. Helical IPPV assembly to minimize effects of PPV.
5. One handed dial control and more obvious “off” position.
6. Service interval now three years.
7. Improved characteristics with tubular woven cotton wicks
8. Accurate with gas flow 5Lpm, dial settings < 3 %.
9. Greatest accuracy between 15-35 degree celcius
![Page 142: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/142.jpg)
Hazards of TEC 5
• More prone to pumping effect then Tec 4.• Large liquid loss if filling port is opened.• Overfilling – bottle adaptor loose, vaporizer
on• Reverse flow increases output.• Carrier gas composition affects output.
![Page 143: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/143.jpg)
TEC 6• Classification –1.Conc. Calibrated
2.Injection vaporizer
3.Thermocompensated by supplied heat or electrically heated.
4.Electromechanically controlled dual circuit.
5.Gas-vapour blender.
6.pressurized
7.Single agent – Desflurane
![Page 144: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/144.jpg)
• Used only for desflurane.• Capacity 390ml • Graduation 1-18% ( 1% increments from 1-
10 and 2% increments thereafter)• Maximum permissible flow 20L/min• Since VC is sealed from atm. special filler
system is required.• Various LEDs in front pannel:-
Amber : without alarm- warm up
Amber : with alarm- agent level below 50 ml
Green : operational
Red :no output due to:--1. Low agent level <20ml
2. Power failure
3. Malfunction
4. Tilted vaporizer beyond 20 degree
![Page 145: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/145.jpg)
TEC 6•Desflurane heated to 39 deg celcius in a sealed chamber, adjusted by H.•VP 1300mmHg in sump.•Carrier gas flow restricted by O, so that pressure is ~ to flow.•Pressure sensed by P , which readjusts R1 so that desflurane flow is ~ to FGF. •Control dial adjusts R2, and thus the output conc.
• H-heater• O-fixed orifice• P-differential pressure
transducer• R1-adjusted by P• R2-adjusted by control
dial
![Page 146: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/146.jpg)
EVALUATION:-• Output almost linear at 3%,5%, 7%.• Slightly low output at <5l/min• Ideal temp 18-30 deg celcius• Tilting resistant • Pumping effect insignificant• Carrier gas effect minimal.
![Page 147: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/147.jpg)
TEC 7•Similar to TEC 5
•Used for isoflurane, enflurane , sevoflurane
•Capacity 300ml (with dry wicks)
225ml (with wet wicks)
•Graduation 0-5% ( 0.2% increments from 0-1 and 0.5% increments thereafter)
•Available with 3 filling devices– funnel filler, Quikfil, Easyfil.
•New ergonomics and design.
•Soldered sump assembly eliminating seals
•Improved sight glass design
•Clear agent color identification
![Page 148: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/148.jpg)
Tec 7
![Page 149: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/149.jpg)
Characteristics of Drager Vaporizers
MOST ACCURATE
CLASSIFICATION-• Conc. calibrated• Flow over• Temperature compensated• Agent specific• Out of circuit
![Page 150: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/150.jpg)
Drager vapourizers
![Page 151: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/151.jpg)
Drager Vapor 19.1
Used for Halothane, isoflurane, enflurane and sevoflurane
Capacity 200ml
Calibrated from 0-5% (o.2% increments in between 0-1 and 1% thereafter)
![Page 152: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/152.jpg)
DRAGER VAPOUR 2000
![Page 153: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/153.jpg)
Penlon sigma delta vaporizer
• Specifications-• wt :5 kg approx• Capacity volume : at MAX mark 250 ml• Flow range: operating flow range .2 to 15 lt/min
SAFETY FEATURE-
1. Keyed filler
2. low filling port
3. Interlocks
4. Secured vaporizers
![Page 154: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/154.jpg)
Cont..
![Page 155: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/155.jpg)
ALADIN CASSETTE VAPORIZER
Classification-1.Conc. Calibrated
2.Flow over
3.Automatic thermocompensation
4. Agent specific
Features-
1.Cassettes containing liquid anaesthetic inserted into a port
2.Agent recognized and dispensed into the stream of FGF
3.Tipping resistant and maintenance free
4.Power battery backup and adequate O2 pressure mandatory
5.Fixed output irrespective of fresh gas mixture
6.Extremely light and can be removed with one hand.
.
![Page 156: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/156.jpg)
Use with datex-Engstrom AS/3 ADU
![Page 157: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/157.jpg)
SIEMENSClassification• Conc calibrated
• Injection
• No thermocompensation
• A caliberated throttlevalve is opened and closed by user and thus regulate the pressure exerted by FGF on surface of liquid anaesthetic agent
• This pressure tends to force liquid to atomize at the injector nozzle
• The liquid droplets vaporize in the flowing FGF.
![Page 158: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/158.jpg)
Agent specific filling system
• A vaporizer designed for a single agent be fitted with a permanently attached agent specific device to prevent accidental filling with wrong agent.
• Reduce air pollution
TYPES-
1.Keyed filling system
2.Screw capped filling system
3.Pin safety system
![Page 159: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/159.jpg)
TEC 4 (KEY FILLER)
![Page 160: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/160.jpg)
TEC 4 (KEY FILLER)
![Page 161: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/161.jpg)
TEC 4 (FUNNEL FILLER)
![Page 162: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/162.jpg)
TEC 4 (FUNNEL FILLER)
![Page 163: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/163.jpg)
![Page 164: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/164.jpg)
![Page 165: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/165.jpg)
![Page 166: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/166.jpg)
![Page 167: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/167.jpg)
KEY FILLERS
![Page 168: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/168.jpg)
KEY FILLER
![Page 169: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/169.jpg)
KEY FILLER
![Page 170: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/170.jpg)
Easy filling system
![Page 171: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/171.jpg)
Quick fill system
![Page 172: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/172.jpg)
Vaporizer mounting systemIf >1 vaporizer can be switched on at a time:-• The patient exposed to a overdose of
anaesthetic agent• The downstream vaporizer is contaminated.
TYPES-
1.Select a tec back bar- a switch on the back bar may be used to direct gas flow through only one vaporizer at a time.
2.A mechanical locking system
3.A mechanical interconnector
![Page 173: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/173.jpg)
Selectatec system• Pair of port valves for
each vaporizer• Vaporizer is mounted
and locked on back bar• When ON 2 plungers
open the valve ports & activate extension rods that prevent other vaporizer.
![Page 174: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/174.jpg)
Vaporizer Mounting Systems
• Only one vaporizer can be turned on
• Gas enters only the “on” vaporizer
• Leak of trace gas is minimized.
![Page 175: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/175.jpg)
BACK BAR DEVICES
• Ohmeda selectatec –has pins in manifold linked to control dial
• If one on –extend to prevent other
![Page 176: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/176.jpg)
DRAGER LOCK
• For Drager 19.2 has rotating bar on manifold with teeth that fit into a cut out on the control dial
![Page 177: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/177.jpg)
Order of Vap. Less potent – upstream More potent – downstream If equipotent: low VP – upstream High VP – downstream
ALSO , If explosive – downstream Trilene – downstream Easy to clean - downstream
![Page 178: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/178.jpg)
ORDER OF VAPORISERS
UP STREAM DOWN STREAM
SEVOFLURANE ENFLURANE ISOFLURANE HALOTHANE DESFLURANE
VP-157 175 238 243 669
![Page 179: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/179.jpg)
Hazards of contemporary vaporizers
Incorrect agent/ Misfilling
Tipping
Overfilling
Reversed flow
Simultaneous inhaled agent administration
Leaks
Electronic failure
![Page 180: Vaporizers 2015 spmc](https://reader034.vdocuments.us/reader034/viewer/2022052214/55a61b3a1a28abd1098b47a1/html5/thumbnails/180.jpg)
IDEAL VAPORIZER
• Deliver fixed desired conc. • Independent of-
temp
flow rate
carrier gas alteration• No effect of back pressure• easy to maintain and clean• Agent specific