QUAL
-QA-
TEM
P-02
20_r
ev15
Part Number: RAN429G
Date: 22/9/14ECR: 5567
ABX Pentra 60C+ / Pentra ES60Kit Version 3.0.1 - Eproms V3.0.0-1 V3.0.0-2
Tn
TN
Kit Version 3.0.1 - Eproms V3.0.0-1 V3.0.0-2 TN: RAN429G
Objective Software version update for P60C+/PES60 with or without takasago valves
Intervention
informationIntervention priority For information - As needed - Mandatory on next visit - Immediate and mandatory
Tracking number TN-RAN429G Intervention estimated duration: 30 min
User process impact YES NO Potential patient result impact YES NO
User training needed YES NO User documentation update YES NO
Applied in factory
on instrument serial #P60C+: PCP13129, 13152PES60: PES12976, 12980, 12985, 12990, 12994, 13038
since instrument serial #P60C+: PCP13156PES60: PES13075
Note history Note index Reason for change
C to D Release of version V2.8.1 (ECR4284)
D to E Release of version V3.0.0 (ECR5180)
E to F Warning regarding the installation on Windows 7 for the Chinese language pack (ECR5437)
F to G V3.0.0 to V3.0.1 (ECR5567)
Software
version
Note index Previous New Mini. requirements
A to B
V2.5.0 PES60V2.4.1 P60C+
V2.7.0EPROM V2.3.0P60C+ version > v2.2.9
B to C canceled canceled
C to D V2.7.0 V2.8.1 EPROM V2.3.3 + Takasago Valves
D to E V2.8.1 V3.0.0 EPROM V3.0.0
E to F V3.0.0 V3.0.0 EPROM V3.0.0
F to G V3.0.0 V3.0.1 EPROM V3.0.0
New parts Parts designation Previous New Reason for change
KIT V3.0.1 P60C+/ES60
XEC130BS XEC130CS Software release
Motherboard P60 OT/C+/ES TAKA
- XAA722AS Software release
Motherboard P60 OT/C+/ES NON TAKA
XAA472DS XAA472ES Software release
New
adjustments
Adjustments Previous New Reason for change
2/8
Kit Version 3.0.1 - Eproms V3.0.0-1 V3.0.0-2 TN: RAN429G
New
documents
Documentation Previous New Reason for change
Documentation CD-ROM
PES60 : RAX071EP60C+ : RAX059F
PES60 : RAX071FP60C+ : RAX059G
V3.0.1 software release
Additionnal
requirements
Designation Part number Link
Part details kit or Assy DesignationComponent p/n
Component designation Qtyavailable for order
XEC130CS KIT V3.0.1 P60C+/ES60 Yes
EPROM V3.0.0-1 1 No
EPROM V3.0.0-2 VTK 1 No
NAJ128BP60 RANGE WORKSTATION V3.0.1 CD-ROM
1 No
RAN429GV3.0.1 PES60/P60C+ technical note
1 No
RAX071F Documentation CD-ROM PES60 1 Yes
RAX059G Documentation CD-ROM P60C+ 1 Yes
3/8
Kit Version 3.0.1 - Eproms V3.0.0-1 V3.0.0-2 TN: RAN429G
1. Modifications and evolutionsThis new software version operates under Windows XP and Windows 7. It no longer operates under Windows NT.
1.1. Russian interface
The ABX Pentra 60C+ and the Pentra ES60 interface is now available in Russian.
1.2. Correction of the PCT Unit
The unit for PCT has been changed for SI and mmol/L. It has been changed from “10-2 /L” to “10-2 L/L”.
1.3. Optical Gain Tolerance Change
Following recent findings, the optical gain frame has been increased. It still allows a minimum tension of 5.5 V, but itcan now allow a maximum tension of 7.5 V.Refer to RAN589A for more information on this adjustment.
2. EPROM installation■ Switch the instrument off and remove the left-hand side panel.
■ Using an antistatic mat, replace the instrument EPROM (refer to the diagram on the following page).Replace the EPROM by the correct one, depending on whether the instrument is equipped with Takasago valves ornot. The EPROM V3.0.0-1 must be used for instruments not equipped with Takasago valves, and the EPROM V3.0.0-2 VTK must be used for instruments equipped with Takasago valves.
■ Re-install the left-hand side panel.
Without Takasago valves With Takasago valves
4/8
Kit Version 3.0.1 - Eproms V3.0.0-1 V3.0.0-2 TN: RAN429G
Without Takasago valves With Takasago valvesV3.0.0-1 V3.0.0-2 - VTK
U43 EVEN U42 ODD
5/8
Kit Version 3.0.1 - Eproms V3.0.0-1 V3.0.0-2 TN: RAN429G
3. V3.0.1 Software installation
■ Create a new worklist before installing the new version in order to have an archive file.
■ Switch the instrument and the workstation off.
■ Switch the instrument and the workstation on and log in as the administrator.
■ Insert the CD-ROM NAJ128B in the computer CD-ROM drive.
■ Start the Windows Explorer application and click the CD-ROM drive device letter.
■ Run "setup.exe" from the CD-ROM root. The following window is displayed:
■ Click "Yes" to continue the installation. The following window is displayed:
■ Click "Next".
■ Select the correct instrument from the scrolling list.
Version V3.0.1 must be installed over version 2.2.9 (or higher) except for the computers Fujitsu Siemens E620 and E5615, which operate under Windows XP. For Fujitsu Siemens E620 and E5615, version 3.0.1 can be installed directly or over version 2.2.9 (or higher).Refer to the latest version of RAN056 to check the compatibility between the software and the computer.
The installation program installs all the files needed to carry out the implementation of the software in the case of a first installation (case of the installation on the Fujitsu Siemens E620 and E5615). In other cases it updates the software but keeps the customer settings and data.
The installation of software version V3.0.1 on Windows 7 is not compatible and not functional for the Chinese language pack.EPSON 6100, EPSON EPL-59000 and Lexmark E210 printers are not compatible with Windows 7.
6/8
Kit Version 3.0.1 - Eproms V3.0.0-1 V3.0.0-2 TN: RAN429G
■ Click "Next".A status bar is displayed during the copy:
◆ The instrument proceeds to the software version installation. At the end of the installation the following window is displayed:
◆ Click "Finish" to restart the computer.◆ Open a session with the "ABX" login.◆ Check the instrument and user setup.◆ Control the correct operation of the instrument.