REFRACTION USING A
PHOROPTER
![Page 2: Refraction using a phoropter](https://reader035.vdocuments.us/reader035/viewer/2022081505/587205b51a28ab176b8b60bb/html5/thumbnails/2.jpg)
A phoropter is a common name for an ophthalmic testing device, also called a refractor. It is commonly used by eye care professionals during an eye examination, and contains different lenses used for refraction, to measure an individual's refractive error and determine his or her eyeglass prescription.
![Page 3: Refraction using a phoropter](https://reader035.vdocuments.us/reader035/viewer/2022081505/587205b51a28ab176b8b60bb/html5/thumbnails/3.jpg)
PARTS OF A PHOROPTER
![Page 4: Refraction using a phoropter](https://reader035.vdocuments.us/reader035/viewer/2022081505/587205b51a28ab176b8b60bb/html5/thumbnails/4.jpg)
PRELIMINARY STEPS OF REFRACTION:1. Place the patient in a comfortable sitting position
behind the phoropter.2. Adjust the IPD to centre the patients eyes on the
lenses3. Level the instrument to perfectly horizontal
position.4. Adjust the back vertex distance to 12mm.5. Starting point of refraction is determined by
objective refraction in any of these ways:a. Retinoscopyb. Auto refractometer readingsc. Glasses power
![Page 5: Refraction using a phoropter](https://reader035.vdocuments.us/reader035/viewer/2022081505/587205b51a28ab176b8b60bb/html5/thumbnails/5.jpg)
STEPS OF SUBJECTIVE REFRACTION PROPER
1. To test baseline visual acuity.2. Establishing spherical power.3. Cylindrical axis refinement.4. Cylindrical power refinement.5. Spherical power refinement.6. Binocular balancing.
![Page 6: Refraction using a phoropter](https://reader035.vdocuments.us/reader035/viewer/2022081505/587205b51a28ab176b8b60bb/html5/thumbnails/6.jpg)
1. To test baseline visual acuity.
Test the visual acuity of each eye separately using Snellen's chart.
![Page 7: Refraction using a phoropter](https://reader035.vdocuments.us/reader035/viewer/2022081505/587205b51a28ab176b8b60bb/html5/thumbnails/7.jpg)
2. Establishing spherical power.Start from one line above the visual acuity line.
Occlude one eye.“Fog” the eye with a plus lens.Start decreasing the power by 0.25D
each time and ask the patient which is better.
Stop when the patient no longer feels any improvement.
![Page 8: Refraction using a phoropter](https://reader035.vdocuments.us/reader035/viewer/2022081505/587205b51a28ab176b8b60bb/html5/thumbnails/8.jpg)
Use the least minus or the most plus power.(If we do not give the patient the least minus power, we leave the patient to accommodate which can cause asthenopia. Hence it is said that treat minus power like money. Give it to the patient only in return of something, ie. Better vision.)
![Page 9: Refraction using a phoropter](https://reader035.vdocuments.us/reader035/viewer/2022081505/587205b51a28ab176b8b60bb/html5/thumbnails/9.jpg)
FOGGING:Fogging means to make the eye artificially myopic.Because in a state off hyperopia, the patient tends to accommodate.Hence we put enough plus lenses in front of the eye such that accommodation is relaxed and the refractive error is stabilised.
![Page 10: Refraction using a phoropter](https://reader035.vdocuments.us/reader035/viewer/2022081505/587205b51a28ab176b8b60bb/html5/thumbnails/10.jpg)
3. Cylindrical axis refinement.Two methods:
1. Jackson cross cylinder (JCC)
2. Astigmatic fan test.
![Page 11: Refraction using a phoropter](https://reader035.vdocuments.us/reader035/viewer/2022081505/587205b51a28ab176b8b60bb/html5/thumbnails/11.jpg)
Jackson’s cross cylinder:Used to refine the axis obtained
by objective refraction (OR).Place the JCC over the cylindrical
lens. The axes should be at 45 and 135 degrees (the axis line on the JCC).
After setting the axis of cylinder as per the axis obtained in OR, flip the JCC and ask the patient which position is better?
![Page 12: Refraction using a phoropter](https://reader035.vdocuments.us/reader035/viewer/2022081505/587205b51a28ab176b8b60bb/html5/thumbnails/12.jpg)
In minus cylinder,chase the red line.
In plus cylinderchase the white line.
Continue this for every 15 degrees in one direction till direction changes.And then fine tune it for 5 degrees in the other direction.
![Page 13: Refraction using a phoropter](https://reader035.vdocuments.us/reader035/viewer/2022081505/587205b51a28ab176b8b60bb/html5/thumbnails/13.jpg)
Astigmatic clock dial:Fog the eye.Ask the patient to look at the astigmatic dial and
identify the darkest and sharpest line.Add minus cylinders perpendicular to this axis
progressively till all lines appear equally blurred (since eye is fogged).
This helps to find the axis as well as power.
![Page 14: Refraction using a phoropter](https://reader035.vdocuments.us/reader035/viewer/2022081505/587205b51a28ab176b8b60bb/html5/thumbnails/14.jpg)
4. Cylindrical power refinement.After finding the axis, place the JCC in the power axes.
On flipping the JCC ask the patient which is better?
If white- add plus power.If red- add minus power.
Continue till both appear equally blurred.For every 1D cylindrical power added or
reduced 0.5D of spherical should be added or reduced respectively.
![Page 15: Refraction using a phoropter](https://reader035.vdocuments.us/reader035/viewer/2022081505/587205b51a28ab176b8b60bb/html5/thumbnails/15.jpg)
5. Spherical power refinement:
After the astigmatism axis and power is refined the sphere power is refined.
This is done once again by fogging method.
This is then confirmed by:Duochrome test.Pinhole test.
![Page 16: Refraction using a phoropter](https://reader035.vdocuments.us/reader035/viewer/2022081505/587205b51a28ab176b8b60bb/html5/thumbnails/16.jpg)
DUOCHROME TEST::It is based on principle of chromatic
aberration.Due to difference in wavelengths, in
emmetropes yellow light focusses on the retina while red behind and green in front.
Hence to an emmetrope both the sides appear equally clear.
If red is clearer- under corrected myopia or over corrected hyperopia.
![Page 17: Refraction using a phoropter](https://reader035.vdocuments.us/reader035/viewer/2022081505/587205b51a28ab176b8b60bb/html5/thumbnails/17.jpg)
![Page 18: Refraction using a phoropter](https://reader035.vdocuments.us/reader035/viewer/2022081505/587205b51a28ab176b8b60bb/html5/thumbnails/18.jpg)
PINHOLE TESTING:If improvement
occurs using the pinhole, the prescription is incorrect.
![Page 19: Refraction using a phoropter](https://reader035.vdocuments.us/reader035/viewer/2022081505/587205b51a28ab176b8b60bb/html5/thumbnails/19.jpg)
6. BINOCULAR BALANCINGTo equalize the accommodative
effort in both the eyes.Fogging and alternate cover test
Duochrome test with fogging
Prism dissociation test.
![Page 20: Refraction using a phoropter](https://reader035.vdocuments.us/reader035/viewer/2022081505/587205b51a28ab176b8b60bb/html5/thumbnails/20.jpg)
Fogging and alternate cover test:
With the best accepted lenses in the frame both eyes are fogged with a 1D lens.
A rapid alternate cover test is performed.
Patient is asked to tell which image is clearer.
If the eyes are in balance the patient will report equal blur.
![Page 21: Refraction using a phoropter](https://reader035.vdocuments.us/reader035/viewer/2022081505/587205b51a28ab176b8b60bb/html5/thumbnails/21.jpg)
Prism dissociation test:After fogging, a 3 or 4 PD prism is kept base down in front of right eye and base up in front of left eye.
A difference between the clarity of the upper and lower lines is used to correct the error.
![Page 22: Refraction using a phoropter](https://reader035.vdocuments.us/reader035/viewer/2022081505/587205b51a28ab176b8b60bb/html5/thumbnails/22.jpg)
THANK YOU