BA084D/06/en/12.0350103643
Valid as of software version:V 1.01.01 (amplifier)
PROline Prowirl 72Vortex Flow Measuring System
Operating Instructions
Brief operating instructions PROline Prowirl 72
2 Endress+Hauser
Brief operating instructions
These brief operating instructions explain how to commission your measuring device quickly and easily:
Safety instructions Page 7
Installation Page 11
Wiring Page 21
Display and operating elements Page 27
Commissioning with "QUICK SETUP" Page 40
You can commission your measuring device quickly and easily using the special Quick Setup menu. It allows you to configure important basic functions via the local display, for example display language, measured variables, engineering units, signal type etc.
Customer-specific configuration /Description of device functions
Page 75 ff.
Complex measurement tasks require the configuration of additional functions which you can individually select, set and adapt to your process conditions using the function matrix. The function matrix of the measuring device and all the functions are described in detail in the "Description of device functions" section.
PROline Prowirl 72 QUICK SETUP for quick commissioning
Endress+Hauser 3
QUICK SETUP for quick commissioning
Note!The display returns to the QUICK SETUP COMMISSIONING cell if you press the ESC key combination during interrogation.
Only the output (current output or pulse/status output) not yet configured in the current Quick Setup is offered for selection after the first cycle.
The YES option appears as long as a free output is still available. "NO" is the only option displayed when no further outputs are available.
When "YES" is selected, the flow is assigned to line 1 of the local display and the totalizer to line 2.
++ +E EEsc
E- +
Yes
Yes
No
No
Measuring Unit Type
Configuration another output ?
Automatically configuration display ?
Automaticallyparameterization
of the display
Language
Application
Quick Setup
HOME-POSITION
Quick SetupCommission
Corrected Volume FlowVolume Flow Calculated Mass Flow
QuitCurrentoutput
Pulse /Status output
Selection output type
UnitFlow
UnitFlow
UnitTotalizer
UnitFlow
UnitTotalizer
UnitTotalizer
UnitDensity
UnitDensity
OperatingDensity
OperatingDensity
ReferenceDensity
OperationMode
Pulse Status
Assignstatus
Pulsevalue
Currentspan
Onvalue
Pulsewidth
Value20 mA
Offvalue
Outputsignal
Timeconstant
Timeconstant
Failsafemode
Failsafemode
Vortex Freq. PFM
F06-72xxxxxx-19-xx-xx-en-000
QUICK SETUP for quick commissioning PROline Prowirl 72
4 Endress+Hauser
PROline Prowirl 72 Contents
Endress+Hauser 5
Contents
1 Safety instructions . . . . . . . . . . . . . . . . . 71.1 Designated use . . . . . . . . . . . . . . . . . . . . . . . . 71.2 Installation, commissioning and operation . . . 71.3 Operational safety . . . . . . . . . . . . . . . . . . . . . . 71.4 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81.5 Notes on safety conventions and icons . . . . . . 8
2 Identification . . . . . . . . . . . . . . . . . . . . . . 92.1 Device designation . . . . . . . . . . . . . . . . . . . . . 9
2.1.1 Nameplate on the transmitter . . . . . . . 92.1.2 Nameplate of the sensor, remote
version . . . . . . . . . . . . . . . . . . . . . . . 102.2 CE mark, declaration of conformity . . . . . . . . 102.3 Registered trademarks . . . . . . . . . . . . . . . . . 10
3 Installation . . . . . . . . . . . . . . . . . . . . . . . . 113.1 Incoming acceptance, transport, storage . . . 11
3.1.1 Incoming acceptance . . . . . . . . . . . 113.1.2 Transport . . . . . . . . . . . . . . . . . . . . . 113.1.3 Storage . . . . . . . . . . . . . . . . . . . . . . . 11
3.2 Installation conditions . . . . . . . . . . . . . . . . . . 123.2.1 Dimensions . . . . . . . . . . . . . . . . . . . 123.2.2 Installation location . . . . . . . . . . . . . 123.2.3 Orientation . . . . . . . . . . . . . . . . . . . . 123.2.4 Heat insulation . . . . . . . . . . . . . . . . . 133.2.5 Inlet and outlet run . . . . . . . . . . . . . . 143.2.6 Vibrations . . . . . . . . . . . . . . . . . . . . . 153.2.7 Limiting flow . . . . . . . . . . . . . . . . . . . 15
3.3 Installation instructions . . . . . . . . . . . . . . . . . 163.3.1 Mounting the sensor . . . . . . . . . . . . 163.3.2 Rotating the transmitter housing . . . 173.3.3 Mounting the transmitter (remote
version) . . . . . . . . . . . . . . . . . . . . . . . 183.3.4 Rotating the local display . . . . . . . . . 19
3.4 Post-installation check . . . . . . . . . . . . . . . . . . 19
4 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 214.1 Connecting the remote version . . . . . . . . . . . 21
4.1.1 Connecting the sensor . . . . . . . . . . . 214.1.2 Cable specifications . . . . . . . . . . . . . 22
4. 2 Connecting the measuring unit . . . . . . . . . . . 224.2.1 Connecting the transmitter . . . . . . . . 224.2.2 Terminal assignment . . . . . . . . . . . . 244.2.3 HART connection . . . . . . . . . . . . . . . 25
4.3 Degree of protection . . . . . . . . . . . . . . . . . . . 264.4 Post-connection check . . . . . . . . . . . . . . . . . 26
5 Operation . . . . . . . . . . . . . . . . . . . . . . . . . 275.1 Display and operating elements . . . . . . . . . . 275.2 The function matrix: layout and use . . . . . . . . 28
5.2.1 General notes . . . . . . . . . . . . . . . . . . 295.2.2 Enabling the programming mode . . 295.2.3 Disabling the programming mode . . 29
5.3 Error message display . . . . . . . . . . . . . . . . . . 305.4 Communication (HART) . . . . . . . . . . . . . . . . . 31
5.4.1 Operating options . . . . . . . . . . . . . . . 315.4.2 Device variables and process
variables . . . . . . . . . . . . . . . . . . . . . . 325.4.3 Universal / common practice
HART commands . . . . . . . . . . . . . . . 325.4.4 Device status / error messages . . . . . 375.4.5 Switching HART write protection
on/off . . . . . . . . . . . . . . . . . . . . . . . . . 38
6 Commissioning . . . . . . . . . . . . . . . . . . . 396.1 Function check . . . . . . . . . . . . . . . . . . . . . . . . 396.2 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . 39
6.2.1 Switching on the measuring device . 396.2.2 Commissioning Quick Setup. . . . . . 40
7 Maintenance . . . . . . . . . . . . . . . . . . . . . . 42
8 Accessories . . . . . . . . . . . . . . . . . . . . . . . 43
9 Trouble-shooting . . . . . . . . . . . . . . . . . 459.1 Trouble-shooting instructions . . . . . . . . . . . . . 459.2 System error messages . . . . . . . . . . . . . . . . . 469.3 Process errors without messages . . . . . . . . . . 489.4 Response of outputs to errors . . . . . . . . . . . . 509.5 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . 519. 6 Installing and removing electronics boards . . 52
9.6.1 Non-Ex, Ex-i version . . . . . . . . . . . . . 529.6.2 Ex-d version . . . . . . . . . . . . . . . . . . . 54
9.7 Software history . . . . . . . . . . . . . . . . . . . . . . . 56
10 Technical data . . . . . . . . . . . . . . . . . . . . 5710.1 Technical data at a glance . . . . . . . . . . . . . . . 57
10.1.1 Application . . . . . . . . . . . . . . . . . . . . 5710.1.2 Function and system design . . . . . . . 5710.1.3 Input . . . . . . . . . . . . . . . . . . . . . . . . . 5710.1.4 Output . . . . . . . . . . . . . . . . . . . . . . . . 5810.1.5 Power supply . . . . . . . . . . . . . . . . . . 5910.1.6 Performance characteristics . . . . . . . 6010.1.7 Mechanical construction . . . . . . . . . . 6210.1.8 Human interface . . . . . . . . . . . . . . . . 6310.1.9 Certificates and approvals . . . . . . . . 6310.1.10 Accessories . . . . . . . . . . . . . . . . . . . 6410.1.11 Documentation . . . . . . . . . . . . . . . . . 64
10.2 Dimensions of transmitter, remote version . . . . . . . . . . . . . . . . . . . . . . . . 64
10.3 Dimensions of Prowirl 72 W . . . . . . . . . . . . . . 6510.4 Dimensions of Prowirl 72 F . . . . . . . . . . . . . . . 6610.5 Dimensions of Prowirl 72 F, Dualsens version 7010.6 Dimensions of flow conditioner . . . . . . . . . . . . 73
Contents PROline Prowirl 72
6 Endress+Hauser
11 Description of device functions . . . 7511.1 Illustration of the function matrix . . . . . . . . . . 7511.2 Description of functions . . . . . . . . . . . . . . . . 76
11.2.1 Group MEASURED VALUES . . . . . . 7611.2.2 Group SYSTEM UNITS . . . . . . . . . . . 7711.2.3 Group QUICK SETUP . . . . . . . . . . . . 8111.2.4 Group OPERATION . . . . . . . . . . . . . 8211.2.5 Group USER INTERFACE . . . . . . . . 8411.2.6 Group TOTALIZER . . . . . . . . . . . . . . 8611.2.7 Group CURRENT OUTPUT . . . . . . . 8811.2.8 Group PULSE/STATUS OUTPUT . . . 9011.2.9 Information on the response
of the status output . . . . . . . . . . . . . . 9711.2.10 Group COMMUNICATION . . . . . . . . 9811.2.11 Group PROCESS PARAMETER . . . . 9911.2.12 Group SYSTEM PARAMETER . . . . 10311.2.13 Group SENSOR DATA . . . . . . . . . . 10411.2.14 Group SUPERVISION . . . . . . . . . . . 10611.2.15 Group SIMULATION SYSTEM . . . . 10811.2.16 Group SENSOR VERSION . . . . . . 10911.2.17 Group AMPLIFIER VERSION . . . . . 109
11.3 Factory settings . . . . . . . . . . . .