Validación de un sistema de sensores inerciales para medición de cinemática
articular: Estudio de cinemática cervical.
Poster · September 2018
DOI: 10.13140/RG.2.2.18521.36965
CITATIONS
0READS
16
6 authors, including:
Some of the authors of this publication are also working on these related projects:
R&D Corporación de Rehabilitación Club de Leones Cruz del Sur View project
Patricio Eduardo Barria Aburto
Rehabilitation Center Club de Leones Cruz del Sur, Chile
5 PUBLICATIONS 0 CITATIONS
SEE PROFILE
All content following this page was uploaded by Patricio Eduardo Barria Aburto on 11 September 2018.
The user has requested enhancement of the downloaded file.
Validación de un sistema de sensores inerciales para medición de cinemática articular: Estudio
de cinemática cervical.Terreros Ghisoni, Andrés (1); Barría Aburto, Patricio (1,2); Raya López, Rafael (3); Moris Rebolledo, André (1); Andrade Gallardo, Asterio (1); Azorín Poveda, José M. (2)
1 Corporación de Rehabilitación Club de Leones Cruz del Sur, Eduardo Frei 02281, Punta Arenas, Chile.2 Universidad Miguel Hernández de Elche, Av. de la Universidad S/N, 03202 Elche, España
3 Universidad CEU San Pablo, Madrid, España.
La goniometría es la medición de los ángulos creados por la intersección de
los ejes longitudinales de los huesos en las articulaciones. Dentro de sus usos
destaca su aplicación para describir las alteraciones de la alineación del sistema
osteoarticular. La fotogrametría es considerada el gold standard para la medición
del movimiento, sin embargo, su uso es complejo y tiene un alto costo. Una
alternativa a esos sistemas son los sensores inerciales, que utilizan distintas
tecnologías para registrar el movimiento, sin las restricciones de los sistemas
más elaborados.
OBJETIVO
Determinar la confiabilidad de los sensores inerciales en la medición de la
cinemática cervical, en comparación con la fotogrametría.
Se midió el movimiento activo en 8 personas sin patología mediante
fotogrametría y sensores inerciales simultáneamente. El procedimiento fue
desarrollado por profesionales capacitados en cada sistema. Se obtuvo
autorización del comité de revisión institucional y consentimiento informado
previo al estudio.
MATERIAL Y MÉTODOS
Se observó una excelente correlación (ICC=0.936) entre ambos instrumentos
de medición al analizar todos los datos de cinemática cervical. Individualmente,
el plano sagital presenta los mejores resultados de correlación (ICC=0.846),
seguido por el plano transverso (ICC=0.514) y finalmente el plano frontal
(ICC=0.095).
RESULTADOS
Los resultados demuestran que los sensores inerciales representan una
alternativa viable para la evaluación de la cinemática articular. Se encontró una
excelente correlación global con el gold standard.
Si bien el plano frontal muestra una pobre correlación entre ambos métodos,
se debe considerar que la diferencia de rango articular registrada es menor a 5°,
por lo tanto, no presenta diferencias estadísticamente significativas.
Futuros estudios deberían mejorar el procedimiento de evaluación con
sensores inerciales para aumentar la confiabilidad de la medición en los planos
individualmente.
CONCLUSIONES
INTRODUCCIÓN
7. Tf ÉC, Sc M, Rubio-romero JA, Sc MDM. DE PRUEBAS DIAGNóSTICAS Statistical methods for evaluating diagnostic test agreement and reproducibility. 2010;61(3):247–55. 8. Fletcher J, Bandy W. Intrarater Reliability of CROM Measurement of Cervical Spine Active Range of Motion in Persons With and Without Neck Pain. J Orthop Sport Phys Ther. 2008;38(10):640–5. 9. Li X, Gonzalez C G-CJ. Fiabilidad y validez de la medida de la movilidad cervical en pacientes con espondiloartritis axial utilizando un sensor inercial. Rehabilitación. 2017;51(1):17–21. 10. Theobald PS, Jones MD, Williams JM. Do inertial sensors represent a viable method to reliably measure cervical spine range of motion ? Man Ther [Internet]. Elsevier Ltd; 2012;17(1):92–6.11. Duc C, Salvia P, Lubansu A, Feipel V, Aminian K. Medical Engineering & Physics. A wearable inertial system to assess the cervical spine mobility : Comparison with an optoelectronic-based motion capture evaluation. Med Eng Phys [Internet]. Institute of Physics and Engineering in Medicine; 2014;36(1):49–56.
Unidad de Investigación y Desarrollo - Corporación de Rehabilitación Club de Leones Cruz del Sur - Av Eduardo Frei 02281, Punta Arenas Chile - Contacto: [email protected]
SENSOR INERCIAL
ENLAZA (Werium AS®)
Cinemática del Tronco
PROTOCOLO DE MEDICIÓN
REGISTRO DE CINEMÁTICA ARTICULAR CON SISTEMAS DE
SENSORES INERCIALES Y FOTOGRAMETRÍA
TODOS LOS DATOS
Coeficiente de correlación intraclase
Correlaciónintraclase
Intervalo de confianza95% Prueba F con valor verdadero 0
Límiteinferior
Límitesuperior Valor gl1 gl2 Sig.
Medidas individuales
,879 ,834 ,912 16,136 143 143 ,000
Medidaspromedio
,936 ,909 ,954 16,136 143 143 ,000
DATOS PLANO SAGITAL
Coeficiente de correlación intraclase
Correlaciónintraclase
Intervalo de confianza95% Prueba F con valor verdadero 0
Límiteinferior
Límitesuperior Valor gl1 gl2 Sig.
Medidas individuales
,733 ,569 ,841 6,496 47 47 ,000
Medidaspromedio
,846 ,725 ,914 6,496 47 47 ,000
DATOS PLANO FRONTAL
Coeficiente de correlación intraclase
Correlaciónintraclase
Intervalo de confianza95% Prueba F con valor verdadero 0
Límiteinferior
Límitesuperior Valor gl1 gl2 Sig.
Medidas individuales
,050 -,235 ,327 1,105 47 47 ,366
Medidaspromedio
,095 -,614 ,493 1,105 47 47 ,366
PLANO TRANSVERSO
Coeficiente de correlación intraclase
Correlaciónintraclase
Intervalo de confianza95% Prueba F con valor verdadero 0
Límiteinferior
Límite superior Valor gl1 gl2 Sig.
Medidasindividuales
,346 ,071 ,572 2,058 47 47 ,007
Medidaspromedio
,514 ,133 ,728 2,058 47 47 ,007
1. Taboadela CH. Goniometría: una herramienta para la evaluación de las incapacidades laborales. 1a. ART A, editor. Buenos Aires; 2007. 2. Prushansky T, Dvir Z. La prueba de la movilidad cervical : metodología e implicaciones clínicas. 2008;3(3):108–14. 3. Access O. Reliability of the universal goniometer for assessing active cervical range of motion in asymptomatic healthy persons. 2016.4. Williams JM, Cuesta-vargas AI, Gala A. The use of inertial sensors system for human motion analysis. 2011; 5. Raya R, Rocon E, Gallego J, Ceres R, Pons J. A Robust Kalman Algorithm to Facilitate Human-Computer Interaction for People with Cerebral Palsy, Using a New Interface Based on Inertial Sensors. Sensors (Basel). 2012;12(3):3049–67. 6. Hallgren KA. NIH Public Access. 2012;8(1):23–34.
Fiabilidad IntraclaseDetermina la variación en los datos
obtenidos por dos o más evaluadores, en el mismo grupo de sujetos..
Excelente: ICC > 0.75
Adecuada: ICC 0.40 a < 0.74
Pobre: ICC < 0.40