Overview Natural resource scarcity is creating higher demand for faster
and more reliable analyzers. A new microchip GC technology
that includes an injection system, column and detector in a
unique compact cartridge was developed by Thermo Scientific
for rapid analysis of natural gas composition. This micro GC
increases analytical productivity by accurately analyzing the
calorific value of natural gas, providing reliable results in
seconds. It also reduces maintenance costs, consumes less gas
and can be used both inside laboratories and on-line.
Exchangeable column cartridges can be easily installed for
flexible configuration and ease of use.
A temperature control allows the micro GC columns to be
programmed with a speed of 240°C per minute for the analysis
of a wider range of components. An integrated stream selector
for auto calibration provides on-line, reliable analytical results.
The micro GC has dedicated instrument control and data
handling software to run on a system, generating rapid results
from the time the micro GC is installed. Reporting is available in
compliance with current ISO, ASTM and GPA methods.
Natural Gas Analysis in less than 1 Minute by Micro GC with
Temperature Programing – According to ASTM D1945 Chris van Tilburga , Cristiane de Oliveira Silvab
aThermo Fisher Scientific, Enschede, The Netherlands bNova Analítica, São Paulo, SP, Br
Natural Gas Analysis
Final Considerations Unique features of the C2V Micro GC for High Speed on-line Gas Analysis:
• Replaceable Cartridge:
Highest level of hardware integration
Simple application change
Easy exchange by operator, no external service
• Flexible Configuration:
Modular 1-4 Channels
Add channels, split systems into individual mGCs
No pre-investment in future hardware
• Automatic Stream Selection:
Calibration gas and sample
• T-ramp programming.
C2V-200 Micro GC from Thermo Scientific
System Design
FIGURE 6. Natural gas analysis up to C3
FIGURE 7. Natural gas analysis C3-C6 on PDMS
FIGURE 8. Natural gas analysis C3-C10 on PDMS (T-ramp)
Temperature Ramp:
50-180 °C in 25 s
Channel Interfacing
Replaceable Cartridge
Contains: MEMS injector,
Column and microTCD
Main Components
FIGURE 5. Natural gas response factor analysis
Instrument Control Software
FIGURE 1. System Set Up
FIGURE 2. Automation Analysis Monitor
FIGURE 3. Automation Result Reporting
FIGURE 4. Automation Trend Reporting