MVCBCT at UCSF A Team Creation
by Michèle AubinIGRT Specialist
Head & Neck Alignment
![Page 2: MVCBCT at UCSF A Team Creation by Michèle Aubin IGRT Specialist Head & Neck Alignment](https://reader034.vdocuments.us/reader034/viewer/2022051819/551654725503469d698b4cdf/html5/thumbnails/2.jpg)
H&N Patient
Here is a typical Coherence screen after the 3D image acquisition and the automatic registration are done for a head & neck patient.
Here is a typical Coherence screen after the 3D image acquisition and the automatic registration are done for a head & neck patient.
![Page 3: MVCBCT at UCSF A Team Creation by Michèle Aubin IGRT Specialist Head & Neck Alignment](https://reader034.vdocuments.us/reader034/viewer/2022051819/551654725503469d698b4cdf/html5/thumbnails/3.jpg)
- Click the ‘Fusion Bending’ tool, and the Blending Dialog will appear.
- Click the ‘Fusion Bending’ tool, and the Blending Dialog will appear.
![Page 4: MVCBCT at UCSF A Team Creation by Michèle Aubin IGRT Specialist Head & Neck Alignment](https://reader034.vdocuments.us/reader034/viewer/2022051819/551654725503469d698b4cdf/html5/thumbnails/4.jpg)
- In the Blending Dialog window, slide to the left the blending tool (Cone Beam (CB) Image),
- Select Invert Grayscale in the Lookup Table and, adjust the Window-Values tool.
- In the Blending Dialog window, slide to the left the blending tool (Cone Beam (CB) Image),
- Select Invert Grayscale in the Lookup Table and, adjust the Window-Values tool.
![Page 5: MVCBCT at UCSF A Team Creation by Michèle Aubin IGRT Specialist Head & Neck Alignment](https://reader034.vdocuments.us/reader034/viewer/2022051819/551654725503469d698b4cdf/html5/thumbnails/5.jpg)
- In the Blending Dialog window, slide to the right the blending tool (CT Reference Image),
- In the Blending Dialog window, slide to the right the blending tool (CT Reference Image),
![Page 6: MVCBCT at UCSF A Team Creation by Michèle Aubin IGRT Specialist Head & Neck Alignment](https://reader034.vdocuments.us/reader034/viewer/2022051819/551654725503469d698b4cdf/html5/thumbnails/6.jpg)
- Select Grayscale in the Lookup Table and, adjust the Window-Values tool.
- Select Grayscale in the Lookup Table and, adjust the Window-Values tool.
![Page 7: MVCBCT at UCSF A Team Creation by Michèle Aubin IGRT Specialist Head & Neck Alignment](https://reader034.vdocuments.us/reader034/viewer/2022051819/551654725503469d698b4cdf/html5/thumbnails/7.jpg)
- Slide the blending tool to 1/3.
The CT and the CB appear together overlapping each other.
- Slide the blending tool to 1/3.
The CT and the CB appear together overlapping each other.
![Page 8: MVCBCT at UCSF A Team Creation by Michèle Aubin IGRT Specialist Head & Neck Alignment](https://reader034.vdocuments.us/reader034/viewer/2022051819/551654725503469d698b4cdf/html5/thumbnails/8.jpg)
-Adjust the image with the Zoom tool.
-Adjust the image with the Zoom tool.
![Page 9: MVCBCT at UCSF A Team Creation by Michèle Aubin IGRT Specialist Head & Neck Alignment](https://reader034.vdocuments.us/reader034/viewer/2022051819/551654725503469d698b4cdf/html5/thumbnails/9.jpg)
- Click on the Manual Registration tool and adjust the CT over the CB in the three views (transversal, coronal and sagital).
- Click on the Manual Registration tool and adjust the CT over the CB in the three views (transversal, coronal and sagital).
![Page 10: MVCBCT at UCSF A Team Creation by Michèle Aubin IGRT Specialist Head & Neck Alignment](https://reader034.vdocuments.us/reader034/viewer/2022051819/551654725503469d698b4cdf/html5/thumbnails/10.jpg)
- When the adjustment is done, click the Accept button.
- When the adjustment is done, click the Accept button.
![Page 11: MVCBCT at UCSF A Team Creation by Michèle Aubin IGRT Specialist Head & Neck Alignment](https://reader034.vdocuments.us/reader034/viewer/2022051819/551654725503469d698b4cdf/html5/thumbnails/11.jpg)
In the right bottom window are showed the table shifts: -On the left are the initial table coordinates.-On the middle are the new table coordinates (absolute).-On the right are the relative table moves.
In the right bottom window are showed the table shifts: -On the left are the initial table coordinates.-On the middle are the new table coordinates (absolute).-On the right are the relative table moves.