Model 115 Portable FID 1
PID Analyzers LLC 2012
Introduction
The Model 115 is a small hand held analyzer
Smallest amp Lightest FID on the market today- analyzer (2) hydride tank (22)- most are 11-12
Safe low pressure H2 storage as a metal hydride (gt 60 hours of operation)
Bluetooth enabled
H2 flow regulator is stable to +- 1-2 through the pressure range-keeping the operating T and sensitivity constant
2
PID Analyzers LLC 2012
The Model 115 is Easy to use
ThermoFoxboro- TVA 1000-FID 27 step startup procedure
PID-115- 8 steps with Calibration Turn on H2 for 30 sec by opening the H2 tank valve Turn on analyzer-Press the On button Press red button (piezo starter) to ignite Check T reading should be increasinf if the flame is litb (takes about
20 min to reach operating T) Press cal button Cal with zero gas (1 min) Cal with Cal gas (1 min) Ready to operate amp calibrated
3
PID Analyzers LLC 2012
Description Photo
The fill system consists of a customer H2 (99999) Tank with a regulator capable of setting to 250 PSI and a fluidics cable (supplied) with a quick disconnect connector at one end and a frac14rdquo connector at the other end (H2 tank)
Easy to fill Cylinder 4
PID Analyzers LLC 2012
Model 115 FID H2 Tank
Model 115 FID amp Tank 5
PID Analyzers LLC 2012
Detail of Model 115 Analyzer 6
PID Analyzers LLC 2012
FID and Belt Mounted Tank FID Belt Clip
Accessories for the 115 7
PID Analyzers LLC 2012
Shoulder strap for the 115 is also available
8
PID Analyzers LLC 2012
Tank Pressure vs H2 Flow
PID Analyzers LLC 2012
9
0
5
10
15
20
25
0 50 100 150 200 250
H2
Flo
w
Tank Pressure (PSI)
Effect of flow on Tank Presure
Linearity of Model 115
0
500
1000
1500
2000
2500
3000
3500
4000
4500
0 500 1000 1500 2000 2500 3000 3500 4000 4500
FID
Re
ad
ing
115
ppm Methane
FID Linearity
10
PID Analyzers LLC 2012
Schematic of FID
PID Analyzers LLC 2012
11
1 Determine when to put on an SCBA amp protective clothing 2 Find buried barrels of haz waste amp check drums for VOC
waste 3 Check soil or water headspace for contamination 4 Leak Detection- liquid fuels natural gas 5 Emergency response 6 EPA Method 21 ndashleak detection
httpwwwhnucom
What can I do with a portable FID analyzer
115 Specifications
PID Analyzers LLC 2012
13
Range- 0-50000 ppm 0-5000 ppm Detection limit- 01 ppm Response time (sec)- 3 sec Weight Analyzer- 2 Weight Tank- 22 H2 (liters) whenfull 70 Battery life- 12 hours Battery Capacity- 0-100 (7 V to 5 VDC) Dimensions- Analyer- 115rdquo x 3rdquo x 2375rdquo Tank- 85rdquo x 275rdquo x 25rdquo Analog output- 0-1 VDC ADC- 16 bit Precision- +- 5 Operating Range- 0-45oC Humidity range - 0-98 RH (non condensing_ Flame Temp Readout- 0-250oC Cal Gas 0-1000 or 0-10000 ppm CH4
- Model 115 Portable FID
- Introduction
- The Model 115 is Easy to use
- Easy to fill Cylinder
- Model 115 FID amp Tank
- Detail of Model 115 Analyzer
- Accessories for the 115
- Slide Number 8
- Tank Pressure vs H2 Flow
- Linearity of Model 115
- Schematic of FID
- 115 Specifications
-
Introduction
The Model 115 is a small hand held analyzer
Smallest amp Lightest FID on the market today- analyzer (2) hydride tank (22)- most are 11-12
Safe low pressure H2 storage as a metal hydride (gt 60 hours of operation)
Bluetooth enabled
H2 flow regulator is stable to +- 1-2 through the pressure range-keeping the operating T and sensitivity constant
2
PID Analyzers LLC 2012
The Model 115 is Easy to use
ThermoFoxboro- TVA 1000-FID 27 step startup procedure
PID-115- 8 steps with Calibration Turn on H2 for 30 sec by opening the H2 tank valve Turn on analyzer-Press the On button Press red button (piezo starter) to ignite Check T reading should be increasinf if the flame is litb (takes about
20 min to reach operating T) Press cal button Cal with zero gas (1 min) Cal with Cal gas (1 min) Ready to operate amp calibrated
3
PID Analyzers LLC 2012
Description Photo
The fill system consists of a customer H2 (99999) Tank with a regulator capable of setting to 250 PSI and a fluidics cable (supplied) with a quick disconnect connector at one end and a frac14rdquo connector at the other end (H2 tank)
Easy to fill Cylinder 4
PID Analyzers LLC 2012
Model 115 FID H2 Tank
Model 115 FID amp Tank 5
PID Analyzers LLC 2012
Detail of Model 115 Analyzer 6
PID Analyzers LLC 2012
FID and Belt Mounted Tank FID Belt Clip
Accessories for the 115 7
PID Analyzers LLC 2012
Shoulder strap for the 115 is also available
8
PID Analyzers LLC 2012
Tank Pressure vs H2 Flow
PID Analyzers LLC 2012
9
0
5
10
15
20
25
0 50 100 150 200 250
H2
Flo
w
Tank Pressure (PSI)
Effect of flow on Tank Presure
Linearity of Model 115
0
500
1000
1500
2000
2500
3000
3500
4000
4500
0 500 1000 1500 2000 2500 3000 3500 4000 4500
FID
Re
ad
ing
115
ppm Methane
FID Linearity
10
PID Analyzers LLC 2012
Schematic of FID
PID Analyzers LLC 2012
11
1 Determine when to put on an SCBA amp protective clothing 2 Find buried barrels of haz waste amp check drums for VOC
waste 3 Check soil or water headspace for contamination 4 Leak Detection- liquid fuels natural gas 5 Emergency response 6 EPA Method 21 ndashleak detection
httpwwwhnucom
What can I do with a portable FID analyzer
115 Specifications
PID Analyzers LLC 2012
13
Range- 0-50000 ppm 0-5000 ppm Detection limit- 01 ppm Response time (sec)- 3 sec Weight Analyzer- 2 Weight Tank- 22 H2 (liters) whenfull 70 Battery life- 12 hours Battery Capacity- 0-100 (7 V to 5 VDC) Dimensions- Analyer- 115rdquo x 3rdquo x 2375rdquo Tank- 85rdquo x 275rdquo x 25rdquo Analog output- 0-1 VDC ADC- 16 bit Precision- +- 5 Operating Range- 0-45oC Humidity range - 0-98 RH (non condensing_ Flame Temp Readout- 0-250oC Cal Gas 0-1000 or 0-10000 ppm CH4
- Model 115 Portable FID
- Introduction
- The Model 115 is Easy to use
- Easy to fill Cylinder
- Model 115 FID amp Tank
- Detail of Model 115 Analyzer
- Accessories for the 115
- Slide Number 8
- Tank Pressure vs H2 Flow
- Linearity of Model 115
- Schematic of FID
- 115 Specifications
-
The Model 115 is Easy to use
ThermoFoxboro- TVA 1000-FID 27 step startup procedure
PID-115- 8 steps with Calibration Turn on H2 for 30 sec by opening the H2 tank valve Turn on analyzer-Press the On button Press red button (piezo starter) to ignite Check T reading should be increasinf if the flame is litb (takes about
20 min to reach operating T) Press cal button Cal with zero gas (1 min) Cal with Cal gas (1 min) Ready to operate amp calibrated
3
PID Analyzers LLC 2012
Description Photo
The fill system consists of a customer H2 (99999) Tank with a regulator capable of setting to 250 PSI and a fluidics cable (supplied) with a quick disconnect connector at one end and a frac14rdquo connector at the other end (H2 tank)
Easy to fill Cylinder 4
PID Analyzers LLC 2012
Model 115 FID H2 Tank
Model 115 FID amp Tank 5
PID Analyzers LLC 2012
Detail of Model 115 Analyzer 6
PID Analyzers LLC 2012
FID and Belt Mounted Tank FID Belt Clip
Accessories for the 115 7
PID Analyzers LLC 2012
Shoulder strap for the 115 is also available
8
PID Analyzers LLC 2012
Tank Pressure vs H2 Flow
PID Analyzers LLC 2012
9
0
5
10
15
20
25
0 50 100 150 200 250
H2
Flo
w
Tank Pressure (PSI)
Effect of flow on Tank Presure
Linearity of Model 115
0
500
1000
1500
2000
2500
3000
3500
4000
4500
0 500 1000 1500 2000 2500 3000 3500 4000 4500
FID
Re
ad
ing
115
ppm Methane
FID Linearity
10
PID Analyzers LLC 2012
Schematic of FID
PID Analyzers LLC 2012
11
1 Determine when to put on an SCBA amp protective clothing 2 Find buried barrels of haz waste amp check drums for VOC
waste 3 Check soil or water headspace for contamination 4 Leak Detection- liquid fuels natural gas 5 Emergency response 6 EPA Method 21 ndashleak detection
httpwwwhnucom
What can I do with a portable FID analyzer
115 Specifications
PID Analyzers LLC 2012
13
Range- 0-50000 ppm 0-5000 ppm Detection limit- 01 ppm Response time (sec)- 3 sec Weight Analyzer- 2 Weight Tank- 22 H2 (liters) whenfull 70 Battery life- 12 hours Battery Capacity- 0-100 (7 V to 5 VDC) Dimensions- Analyer- 115rdquo x 3rdquo x 2375rdquo Tank- 85rdquo x 275rdquo x 25rdquo Analog output- 0-1 VDC ADC- 16 bit Precision- +- 5 Operating Range- 0-45oC Humidity range - 0-98 RH (non condensing_ Flame Temp Readout- 0-250oC Cal Gas 0-1000 or 0-10000 ppm CH4
- Model 115 Portable FID
- Introduction
- The Model 115 is Easy to use
- Easy to fill Cylinder
- Model 115 FID amp Tank
- Detail of Model 115 Analyzer
- Accessories for the 115
- Slide Number 8
- Tank Pressure vs H2 Flow
- Linearity of Model 115
- Schematic of FID
- 115 Specifications
-
Description Photo
The fill system consists of a customer H2 (99999) Tank with a regulator capable of setting to 250 PSI and a fluidics cable (supplied) with a quick disconnect connector at one end and a frac14rdquo connector at the other end (H2 tank)
Easy to fill Cylinder 4
PID Analyzers LLC 2012
Model 115 FID H2 Tank
Model 115 FID amp Tank 5
PID Analyzers LLC 2012
Detail of Model 115 Analyzer 6
PID Analyzers LLC 2012
FID and Belt Mounted Tank FID Belt Clip
Accessories for the 115 7
PID Analyzers LLC 2012
Shoulder strap for the 115 is also available
8
PID Analyzers LLC 2012
Tank Pressure vs H2 Flow
PID Analyzers LLC 2012
9
0
5
10
15
20
25
0 50 100 150 200 250
H2
Flo
w
Tank Pressure (PSI)
Effect of flow on Tank Presure
Linearity of Model 115
0
500
1000
1500
2000
2500
3000
3500
4000
4500
0 500 1000 1500 2000 2500 3000 3500 4000 4500
FID
Re
ad
ing
115
ppm Methane
FID Linearity
10
PID Analyzers LLC 2012
Schematic of FID
PID Analyzers LLC 2012
11
1 Determine when to put on an SCBA amp protective clothing 2 Find buried barrels of haz waste amp check drums for VOC
waste 3 Check soil or water headspace for contamination 4 Leak Detection- liquid fuels natural gas 5 Emergency response 6 EPA Method 21 ndashleak detection
httpwwwhnucom
What can I do with a portable FID analyzer
115 Specifications
PID Analyzers LLC 2012
13
Range- 0-50000 ppm 0-5000 ppm Detection limit- 01 ppm Response time (sec)- 3 sec Weight Analyzer- 2 Weight Tank- 22 H2 (liters) whenfull 70 Battery life- 12 hours Battery Capacity- 0-100 (7 V to 5 VDC) Dimensions- Analyer- 115rdquo x 3rdquo x 2375rdquo Tank- 85rdquo x 275rdquo x 25rdquo Analog output- 0-1 VDC ADC- 16 bit Precision- +- 5 Operating Range- 0-45oC Humidity range - 0-98 RH (non condensing_ Flame Temp Readout- 0-250oC Cal Gas 0-1000 or 0-10000 ppm CH4
- Model 115 Portable FID
- Introduction
- The Model 115 is Easy to use
- Easy to fill Cylinder
- Model 115 FID amp Tank
- Detail of Model 115 Analyzer
- Accessories for the 115
- Slide Number 8
- Tank Pressure vs H2 Flow
- Linearity of Model 115
- Schematic of FID
- 115 Specifications
-
Model 115 FID H2 Tank
Model 115 FID amp Tank 5
PID Analyzers LLC 2012
Detail of Model 115 Analyzer 6
PID Analyzers LLC 2012
FID and Belt Mounted Tank FID Belt Clip
Accessories for the 115 7
PID Analyzers LLC 2012
Shoulder strap for the 115 is also available
8
PID Analyzers LLC 2012
Tank Pressure vs H2 Flow
PID Analyzers LLC 2012
9
0
5
10
15
20
25
0 50 100 150 200 250
H2
Flo
w
Tank Pressure (PSI)
Effect of flow on Tank Presure
Linearity of Model 115
0
500
1000
1500
2000
2500
3000
3500
4000
4500
0 500 1000 1500 2000 2500 3000 3500 4000 4500
FID
Re
ad
ing
115
ppm Methane
FID Linearity
10
PID Analyzers LLC 2012
Schematic of FID
PID Analyzers LLC 2012
11
1 Determine when to put on an SCBA amp protective clothing 2 Find buried barrels of haz waste amp check drums for VOC
waste 3 Check soil or water headspace for contamination 4 Leak Detection- liquid fuels natural gas 5 Emergency response 6 EPA Method 21 ndashleak detection
httpwwwhnucom
What can I do with a portable FID analyzer
115 Specifications
PID Analyzers LLC 2012
13
Range- 0-50000 ppm 0-5000 ppm Detection limit- 01 ppm Response time (sec)- 3 sec Weight Analyzer- 2 Weight Tank- 22 H2 (liters) whenfull 70 Battery life- 12 hours Battery Capacity- 0-100 (7 V to 5 VDC) Dimensions- Analyer- 115rdquo x 3rdquo x 2375rdquo Tank- 85rdquo x 275rdquo x 25rdquo Analog output- 0-1 VDC ADC- 16 bit Precision- +- 5 Operating Range- 0-45oC Humidity range - 0-98 RH (non condensing_ Flame Temp Readout- 0-250oC Cal Gas 0-1000 or 0-10000 ppm CH4
- Model 115 Portable FID
- Introduction
- The Model 115 is Easy to use
- Easy to fill Cylinder
- Model 115 FID amp Tank
- Detail of Model 115 Analyzer
- Accessories for the 115
- Slide Number 8
- Tank Pressure vs H2 Flow
- Linearity of Model 115
- Schematic of FID
- 115 Specifications
-
Detail of Model 115 Analyzer 6
PID Analyzers LLC 2012
FID and Belt Mounted Tank FID Belt Clip
Accessories for the 115 7
PID Analyzers LLC 2012
Shoulder strap for the 115 is also available
8
PID Analyzers LLC 2012
Tank Pressure vs H2 Flow
PID Analyzers LLC 2012
9
0
5
10
15
20
25
0 50 100 150 200 250
H2
Flo
w
Tank Pressure (PSI)
Effect of flow on Tank Presure
Linearity of Model 115
0
500
1000
1500
2000
2500
3000
3500
4000
4500
0 500 1000 1500 2000 2500 3000 3500 4000 4500
FID
Re
ad
ing
115
ppm Methane
FID Linearity
10
PID Analyzers LLC 2012
Schematic of FID
PID Analyzers LLC 2012
11
1 Determine when to put on an SCBA amp protective clothing 2 Find buried barrels of haz waste amp check drums for VOC
waste 3 Check soil or water headspace for contamination 4 Leak Detection- liquid fuels natural gas 5 Emergency response 6 EPA Method 21 ndashleak detection
httpwwwhnucom
What can I do with a portable FID analyzer
115 Specifications
PID Analyzers LLC 2012
13
Range- 0-50000 ppm 0-5000 ppm Detection limit- 01 ppm Response time (sec)- 3 sec Weight Analyzer- 2 Weight Tank- 22 H2 (liters) whenfull 70 Battery life- 12 hours Battery Capacity- 0-100 (7 V to 5 VDC) Dimensions- Analyer- 115rdquo x 3rdquo x 2375rdquo Tank- 85rdquo x 275rdquo x 25rdquo Analog output- 0-1 VDC ADC- 16 bit Precision- +- 5 Operating Range- 0-45oC Humidity range - 0-98 RH (non condensing_ Flame Temp Readout- 0-250oC Cal Gas 0-1000 or 0-10000 ppm CH4
- Model 115 Portable FID
- Introduction
- The Model 115 is Easy to use
- Easy to fill Cylinder
- Model 115 FID amp Tank
- Detail of Model 115 Analyzer
- Accessories for the 115
- Slide Number 8
- Tank Pressure vs H2 Flow
- Linearity of Model 115
- Schematic of FID
- 115 Specifications
-
FID and Belt Mounted Tank FID Belt Clip
Accessories for the 115 7
PID Analyzers LLC 2012
Shoulder strap for the 115 is also available
8
PID Analyzers LLC 2012
Tank Pressure vs H2 Flow
PID Analyzers LLC 2012
9
0
5
10
15
20
25
0 50 100 150 200 250
H2
Flo
w
Tank Pressure (PSI)
Effect of flow on Tank Presure
Linearity of Model 115
0
500
1000
1500
2000
2500
3000
3500
4000
4500
0 500 1000 1500 2000 2500 3000 3500 4000 4500
FID
Re
ad
ing
115
ppm Methane
FID Linearity
10
PID Analyzers LLC 2012
Schematic of FID
PID Analyzers LLC 2012
11
1 Determine when to put on an SCBA amp protective clothing 2 Find buried barrels of haz waste amp check drums for VOC
waste 3 Check soil or water headspace for contamination 4 Leak Detection- liquid fuels natural gas 5 Emergency response 6 EPA Method 21 ndashleak detection
httpwwwhnucom
What can I do with a portable FID analyzer
115 Specifications
PID Analyzers LLC 2012
13
Range- 0-50000 ppm 0-5000 ppm Detection limit- 01 ppm Response time (sec)- 3 sec Weight Analyzer- 2 Weight Tank- 22 H2 (liters) whenfull 70 Battery life- 12 hours Battery Capacity- 0-100 (7 V to 5 VDC) Dimensions- Analyer- 115rdquo x 3rdquo x 2375rdquo Tank- 85rdquo x 275rdquo x 25rdquo Analog output- 0-1 VDC ADC- 16 bit Precision- +- 5 Operating Range- 0-45oC Humidity range - 0-98 RH (non condensing_ Flame Temp Readout- 0-250oC Cal Gas 0-1000 or 0-10000 ppm CH4
- Model 115 Portable FID
- Introduction
- The Model 115 is Easy to use
- Easy to fill Cylinder
- Model 115 FID amp Tank
- Detail of Model 115 Analyzer
- Accessories for the 115
- Slide Number 8
- Tank Pressure vs H2 Flow
- Linearity of Model 115
- Schematic of FID
- 115 Specifications
-
8
PID Analyzers LLC 2012
Tank Pressure vs H2 Flow
PID Analyzers LLC 2012
9
0
5
10
15
20
25
0 50 100 150 200 250
H2
Flo
w
Tank Pressure (PSI)
Effect of flow on Tank Presure
Linearity of Model 115
0
500
1000
1500
2000
2500
3000
3500
4000
4500
0 500 1000 1500 2000 2500 3000 3500 4000 4500
FID
Re
ad
ing
115
ppm Methane
FID Linearity
10
PID Analyzers LLC 2012
Schematic of FID
PID Analyzers LLC 2012
11
1 Determine when to put on an SCBA amp protective clothing 2 Find buried barrels of haz waste amp check drums for VOC
waste 3 Check soil or water headspace for contamination 4 Leak Detection- liquid fuels natural gas 5 Emergency response 6 EPA Method 21 ndashleak detection
httpwwwhnucom
What can I do with a portable FID analyzer
115 Specifications
PID Analyzers LLC 2012
13
Range- 0-50000 ppm 0-5000 ppm Detection limit- 01 ppm Response time (sec)- 3 sec Weight Analyzer- 2 Weight Tank- 22 H2 (liters) whenfull 70 Battery life- 12 hours Battery Capacity- 0-100 (7 V to 5 VDC) Dimensions- Analyer- 115rdquo x 3rdquo x 2375rdquo Tank- 85rdquo x 275rdquo x 25rdquo Analog output- 0-1 VDC ADC- 16 bit Precision- +- 5 Operating Range- 0-45oC Humidity range - 0-98 RH (non condensing_ Flame Temp Readout- 0-250oC Cal Gas 0-1000 or 0-10000 ppm CH4
- Model 115 Portable FID
- Introduction
- The Model 115 is Easy to use
- Easy to fill Cylinder
- Model 115 FID amp Tank
- Detail of Model 115 Analyzer
- Accessories for the 115
- Slide Number 8
- Tank Pressure vs H2 Flow
- Linearity of Model 115
- Schematic of FID
- 115 Specifications
-
Tank Pressure vs H2 Flow
PID Analyzers LLC 2012
9
0
5
10
15
20
25
0 50 100 150 200 250
H2
Flo
w
Tank Pressure (PSI)
Effect of flow on Tank Presure
Linearity of Model 115
0
500
1000
1500
2000
2500
3000
3500
4000
4500
0 500 1000 1500 2000 2500 3000 3500 4000 4500
FID
Re
ad
ing
115
ppm Methane
FID Linearity
10
PID Analyzers LLC 2012
Schematic of FID
PID Analyzers LLC 2012
11
1 Determine when to put on an SCBA amp protective clothing 2 Find buried barrels of haz waste amp check drums for VOC
waste 3 Check soil or water headspace for contamination 4 Leak Detection- liquid fuels natural gas 5 Emergency response 6 EPA Method 21 ndashleak detection
httpwwwhnucom
What can I do with a portable FID analyzer
115 Specifications
PID Analyzers LLC 2012
13
Range- 0-50000 ppm 0-5000 ppm Detection limit- 01 ppm Response time (sec)- 3 sec Weight Analyzer- 2 Weight Tank- 22 H2 (liters) whenfull 70 Battery life- 12 hours Battery Capacity- 0-100 (7 V to 5 VDC) Dimensions- Analyer- 115rdquo x 3rdquo x 2375rdquo Tank- 85rdquo x 275rdquo x 25rdquo Analog output- 0-1 VDC ADC- 16 bit Precision- +- 5 Operating Range- 0-45oC Humidity range - 0-98 RH (non condensing_ Flame Temp Readout- 0-250oC Cal Gas 0-1000 or 0-10000 ppm CH4
- Model 115 Portable FID
- Introduction
- The Model 115 is Easy to use
- Easy to fill Cylinder
- Model 115 FID amp Tank
- Detail of Model 115 Analyzer
- Accessories for the 115
- Slide Number 8
- Tank Pressure vs H2 Flow
- Linearity of Model 115
- Schematic of FID
- 115 Specifications
-
Linearity of Model 115
0
500
1000
1500
2000
2500
3000
3500
4000
4500
0 500 1000 1500 2000 2500 3000 3500 4000 4500
FID
Re
ad
ing
115
ppm Methane
FID Linearity
10
PID Analyzers LLC 2012
Schematic of FID
PID Analyzers LLC 2012
11
1 Determine when to put on an SCBA amp protective clothing 2 Find buried barrels of haz waste amp check drums for VOC
waste 3 Check soil or water headspace for contamination 4 Leak Detection- liquid fuels natural gas 5 Emergency response 6 EPA Method 21 ndashleak detection
httpwwwhnucom
What can I do with a portable FID analyzer
115 Specifications
PID Analyzers LLC 2012
13
Range- 0-50000 ppm 0-5000 ppm Detection limit- 01 ppm Response time (sec)- 3 sec Weight Analyzer- 2 Weight Tank- 22 H2 (liters) whenfull 70 Battery life- 12 hours Battery Capacity- 0-100 (7 V to 5 VDC) Dimensions- Analyer- 115rdquo x 3rdquo x 2375rdquo Tank- 85rdquo x 275rdquo x 25rdquo Analog output- 0-1 VDC ADC- 16 bit Precision- +- 5 Operating Range- 0-45oC Humidity range - 0-98 RH (non condensing_ Flame Temp Readout- 0-250oC Cal Gas 0-1000 or 0-10000 ppm CH4
- Model 115 Portable FID
- Introduction
- The Model 115 is Easy to use
- Easy to fill Cylinder
- Model 115 FID amp Tank
- Detail of Model 115 Analyzer
- Accessories for the 115
- Slide Number 8
- Tank Pressure vs H2 Flow
- Linearity of Model 115
- Schematic of FID
- 115 Specifications
-
Schematic of FID
PID Analyzers LLC 2012
11
1 Determine when to put on an SCBA amp protective clothing 2 Find buried barrels of haz waste amp check drums for VOC
waste 3 Check soil or water headspace for contamination 4 Leak Detection- liquid fuels natural gas 5 Emergency response 6 EPA Method 21 ndashleak detection
httpwwwhnucom
What can I do with a portable FID analyzer
115 Specifications
PID Analyzers LLC 2012
13
Range- 0-50000 ppm 0-5000 ppm Detection limit- 01 ppm Response time (sec)- 3 sec Weight Analyzer- 2 Weight Tank- 22 H2 (liters) whenfull 70 Battery life- 12 hours Battery Capacity- 0-100 (7 V to 5 VDC) Dimensions- Analyer- 115rdquo x 3rdquo x 2375rdquo Tank- 85rdquo x 275rdquo x 25rdquo Analog output- 0-1 VDC ADC- 16 bit Precision- +- 5 Operating Range- 0-45oC Humidity range - 0-98 RH (non condensing_ Flame Temp Readout- 0-250oC Cal Gas 0-1000 or 0-10000 ppm CH4
- Model 115 Portable FID
- Introduction
- The Model 115 is Easy to use
- Easy to fill Cylinder
- Model 115 FID amp Tank
- Detail of Model 115 Analyzer
- Accessories for the 115
- Slide Number 8
- Tank Pressure vs H2 Flow
- Linearity of Model 115
- Schematic of FID
- 115 Specifications
-
1 Determine when to put on an SCBA amp protective clothing 2 Find buried barrels of haz waste amp check drums for VOC
waste 3 Check soil or water headspace for contamination 4 Leak Detection- liquid fuels natural gas 5 Emergency response 6 EPA Method 21 ndashleak detection
httpwwwhnucom
What can I do with a portable FID analyzer
115 Specifications
PID Analyzers LLC 2012
13
Range- 0-50000 ppm 0-5000 ppm Detection limit- 01 ppm Response time (sec)- 3 sec Weight Analyzer- 2 Weight Tank- 22 H2 (liters) whenfull 70 Battery life- 12 hours Battery Capacity- 0-100 (7 V to 5 VDC) Dimensions- Analyer- 115rdquo x 3rdquo x 2375rdquo Tank- 85rdquo x 275rdquo x 25rdquo Analog output- 0-1 VDC ADC- 16 bit Precision- +- 5 Operating Range- 0-45oC Humidity range - 0-98 RH (non condensing_ Flame Temp Readout- 0-250oC Cal Gas 0-1000 or 0-10000 ppm CH4
- Model 115 Portable FID
- Introduction
- The Model 115 is Easy to use
- Easy to fill Cylinder
- Model 115 FID amp Tank
- Detail of Model 115 Analyzer
- Accessories for the 115
- Slide Number 8
- Tank Pressure vs H2 Flow
- Linearity of Model 115
- Schematic of FID
- 115 Specifications
-
115 Specifications
PID Analyzers LLC 2012
13
Range- 0-50000 ppm 0-5000 ppm Detection limit- 01 ppm Response time (sec)- 3 sec Weight Analyzer- 2 Weight Tank- 22 H2 (liters) whenfull 70 Battery life- 12 hours Battery Capacity- 0-100 (7 V to 5 VDC) Dimensions- Analyer- 115rdquo x 3rdquo x 2375rdquo Tank- 85rdquo x 275rdquo x 25rdquo Analog output- 0-1 VDC ADC- 16 bit Precision- +- 5 Operating Range- 0-45oC Humidity range - 0-98 RH (non condensing_ Flame Temp Readout- 0-250oC Cal Gas 0-1000 or 0-10000 ppm CH4
- Model 115 Portable FID
- Introduction
- The Model 115 is Easy to use
- Easy to fill Cylinder
- Model 115 FID amp Tank
- Detail of Model 115 Analyzer
- Accessories for the 115
- Slide Number 8
- Tank Pressure vs H2 Flow
- Linearity of Model 115
- Schematic of FID
- 115 Specifications
-