![Page 1: Michael Wright Bioanalytical Services LGC Fordham, UK · Science for a safer world Case studies: Troubleshooting failed system suitability tests Michael Wright Bioanalytical Services](https://reader034.vdocuments.us/reader034/viewer/2022042206/5ea8e3ff577ae3285923d5d3/html5/thumbnails/1.jpg)
Sciencefor a safer world
Case studies: Troubleshooting failed system suitability tests
Michael Wright
Bioanalytical Services
LGC
Fordham, UK
![Page 2: Michael Wright Bioanalytical Services LGC Fordham, UK · Science for a safer world Case studies: Troubleshooting failed system suitability tests Michael Wright Bioanalytical Services](https://reader034.vdocuments.us/reader034/viewer/2022042206/5ea8e3ff577ae3285923d5d3/html5/thumbnails/2.jpg)
13/10/2016 2
Case Study No.1
Expected for SST:
VMA
RT - 1.08 (0.98-1.18)
S:N - 80:1 (>20:1)
5HIAA
RT - 2.3 (2.15-2.45)
S:N - 100:1 (>20:1)
HVA
RT - 2.7 (2.6-2.9)
S:N – 30:1 (>15:1)
Pressure = 3300-3600 psi
NB: Uses different MRM periods
VMA
5HIAA
HVA
![Page 3: Michael Wright Bioanalytical Services LGC Fordham, UK · Science for a safer world Case studies: Troubleshooting failed system suitability tests Michael Wright Bioanalytical Services](https://reader034.vdocuments.us/reader034/viewer/2022042206/5ea8e3ff577ae3285923d5d3/html5/thumbnails/3.jpg)
1) Pressure = 2694 psi (lower
limit 3300)
2) New column….
Case Study No.1
“Where did everybody go?”
![Page 4: Michael Wright Bioanalytical Services LGC Fordham, UK · Science for a safer world Case studies: Troubleshooting failed system suitability tests Michael Wright Bioanalytical Services](https://reader034.vdocuments.us/reader034/viewer/2022042206/5ea8e3ff577ae3285923d5d3/html5/thumbnails/4.jpg)
Step 1 – start simple
• Correct method being run? Source?
• Is the LC connected to the MS?
YES
Want to see if the peaks have
moved outside of their MRM
periods
Moved the time periods back and…
Hello Peaks!
VMA
RT - 1.67 (0.98-1.18)
S:N - 40:1 (>20:1)
5HIAA
RT – 3.13 (2.15-2.45)
S:N - 50:1 (>20:1)
HVA
RT – 3.5 (2.6-2.9)
S:N – 11:1 (>15:1)
![Page 5: Michael Wright Bioanalytical Services LGC Fordham, UK · Science for a safer world Case studies: Troubleshooting failed system suitability tests Michael Wright Bioanalytical Services](https://reader034.vdocuments.us/reader034/viewer/2022042206/5ea8e3ff577ae3285923d5d3/html5/thumbnails/5.jpg)
13/10/2016 5
Step 2 – look for general causes of low back pressure
• Leak somewhere? Nothing detected
• Correct mobile phases? Made fresh and the same RT
• Lines not purged properly or leaking purge valve? No
• Fluctuating Back pressure? No
Step 1 cont….
Is to do with the new column? No
Struggling to find the cause
Pressure = 2406 psi (even lower)
Peaks eluting even later – VMA peak
now being cut off
![Page 6: Michael Wright Bioanalytical Services LGC Fordham, UK · Science for a safer world Case studies: Troubleshooting failed system suitability tests Michael Wright Bioanalytical Services](https://reader034.vdocuments.us/reader034/viewer/2022042206/5ea8e3ff577ae3285923d5d3/html5/thumbnails/6.jpg)
Same instrument – Clozapine run was submitted
Clozapine SST looked fine
![Page 7: Michael Wright Bioanalytical Services LGC Fordham, UK · Science for a safer world Case studies: Troubleshooting failed system suitability tests Michael Wright Bioanalytical Services](https://reader034.vdocuments.us/reader034/viewer/2022042206/5ea8e3ff577ae3285923d5d3/html5/thumbnails/7.jpg)
13/10/2016 7
Leak here
Leak worse when the pressure
was highest
Analyst had checked the front of
the valve for a leak (with a finger)
not the back
Learning points
• Now check the front and back
of the valve for leaks
• Now check for leaks using blue
tissue paper
![Page 8: Michael Wright Bioanalytical Services LGC Fordham, UK · Science for a safer world Case studies: Troubleshooting failed system suitability tests Michael Wright Bioanalytical Services](https://reader034.vdocuments.us/reader034/viewer/2022042206/5ea8e3ff577ae3285923d5d3/html5/thumbnails/8.jpg)
13/10/2016 8
Case Study No.2
SST – 25OH VitD
Retention Time Shift
Retention
time shift just
within the
criteria
![Page 9: Michael Wright Bioanalytical Services LGC Fordham, UK · Science for a safer world Case studies: Troubleshooting failed system suitability tests Michael Wright Bioanalytical Services](https://reader034.vdocuments.us/reader034/viewer/2022042206/5ea8e3ff577ae3285923d5d3/html5/thumbnails/9.jpg)
25OH Vitamin D batch submitted
Retention
time at start
of run
Retention
time at end
of run
The 2nd plate of samples had failed
![Page 10: Michael Wright Bioanalytical Services LGC Fordham, UK · Science for a safer world Case studies: Troubleshooting failed system suitability tests Michael Wright Bioanalytical Services](https://reader034.vdocuments.us/reader034/viewer/2022042206/5ea8e3ff577ae3285923d5d3/html5/thumbnails/10.jpg)
The following day the routine maintenance was performed
“Replace fluid in pump seal wash bottles (10% Ethanol)”
![Page 11: Michael Wright Bioanalytical Services LGC Fordham, UK · Science for a safer world Case studies: Troubleshooting failed system suitability tests Michael Wright Bioanalytical Services](https://reader034.vdocuments.us/reader034/viewer/2022042206/5ea8e3ff577ae3285923d5d3/html5/thumbnails/11.jpg)
![Page 12: Michael Wright Bioanalytical Services LGC Fordham, UK · Science for a safer world Case studies: Troubleshooting failed system suitability tests Michael Wright Bioanalytical Services](https://reader034.vdocuments.us/reader034/viewer/2022042206/5ea8e3ff577ae3285923d5d3/html5/thumbnails/12.jpg)
Pump D seal
New Seal
![Page 13: Michael Wright Bioanalytical Services LGC Fordham, UK · Science for a safer world Case studies: Troubleshooting failed system suitability tests Michael Wright Bioanalytical Services](https://reader034.vdocuments.us/reader034/viewer/2022042206/5ea8e3ff577ae3285923d5d3/html5/thumbnails/13.jpg)
Learning points
• Our criteria for retention time wasn’t tight enough
to spot this
• We weren’t changing pump seals frequently
enough
Case Study No.2
![Page 14: Michael Wright Bioanalytical Services LGC Fordham, UK · Science for a safer world Case studies: Troubleshooting failed system suitability tests Michael Wright Bioanalytical Services](https://reader034.vdocuments.us/reader034/viewer/2022042206/5ea8e3ff577ae3285923d5d3/html5/thumbnails/14.jpg)
Case Study No.3…. 1 week later
![Page 15: Michael Wright Bioanalytical Services LGC Fordham, UK · Science for a safer world Case studies: Troubleshooting failed system suitability tests Michael Wright Bioanalytical Services](https://reader034.vdocuments.us/reader034/viewer/2022042206/5ea8e3ff577ae3285923d5d3/html5/thumbnails/15.jpg)
Unacceptable drop in
intensity & run failed
Started easy
Leak somewhere after the column? No
Gas OK? Yes
Vacuum OK? Yes
Correct ion probe? Yes APCI
Capillary extended in probe? Yes
Probe Blocked? No
APCI Corona needle positioned correctly? Yes
MS fouled or out of calibration? Ahhh
![Page 16: Michael Wright Bioanalytical Services LGC Fordham, UK · Science for a safer world Case studies: Troubleshooting failed system suitability tests Michael Wright Bioanalytical Services](https://reader034.vdocuments.us/reader034/viewer/2022042206/5ea8e3ff577ae3285923d5d3/html5/thumbnails/16.jpg)
A fibre in the
skimmer plate!
![Page 17: Michael Wright Bioanalytical Services LGC Fordham, UK · Science for a safer world Case studies: Troubleshooting failed system suitability tests Michael Wright Bioanalytical Services](https://reader034.vdocuments.us/reader034/viewer/2022042206/5ea8e3ff577ae3285923d5d3/html5/thumbnails/17.jpg)
Following removal
of the offending
fibre…
Learning points
• Our criteria for intensity wasn’t tight enough
• We needed to change the way in which we performed
a weekly source clean on the Mass Spec
![Page 18: Michael Wright Bioanalytical Services LGC Fordham, UK · Science for a safer world Case studies: Troubleshooting failed system suitability tests Michael Wright Bioanalytical Services](https://reader034.vdocuments.us/reader034/viewer/2022042206/5ea8e3ff577ae3285923d5d3/html5/thumbnails/18.jpg)
XIC of +MRM (4 pairs): 383.200/257.300 Da ID: 25OHD3 1 from Sample 224 (2016_03_30 TNHS) of SST.wiff (Heated Nebulizer) Max. 3850.0 cps.
0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0Time, min
0
200
400
600
800
1000
1200
1400
1600
1800
2000
2200
2400
2600
2800
3000
3200
3400
3600
3800
Inte
ns
ity
, c
ps
4.74
4.98
5.29
Case Study No.4
XIC of +MRM (4 pairs): 383.200/257.300 Da ID: 25OHD3 1 from Sample 208 (2016_03_07 TN) of SST.wiff (Heated Nebulizer) Max. 4385.0 cps.
0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0Time, min
0
200
400
600
800
1000
1200
1400
1600
1800
2000
2200
2400
2600
2800
3000
3200
3400
3600
3800
4000
4200
4385
Inte
ns
ity
, c
ps
4.95
5.20
XIC of +MRM (4 pairs): 383.200/257.300 Da ID: 25OHD3 1 from Sample 207 (2016_03_07 TN) of SST.wiff (Heated Nebulizer) Max. 1.0e4 cps.
0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0Time, min
0.00
500.00
1000.00
1500.00
2000.00
2500.00
3000.00
3500.00
4000.00
4500.00
5000.00
5500.00
6000.00
6500.00
7000.00
7500.00
8000.00
8500.00
9000.00
9500.00
1.00e4
Inte
ns
ity
, c
ps
5.34
Supposed to
look like this
Now….
Purged mobile
phases…
XIC of +MRM (4 pairs): 383.200/257.300 Da ID: 25OHD3 1 from Sample 208 (2016_03_07 TN) of SST.wiff (Heated Nebulizer) Max. 4385.0 cps.
0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0Time, min
0
200
400
600
800
1000
1200
1400
1600
1800
2000
2200
2400
2600
2800
3000
3200
3400
3600
3800
4000
4200
4385
Inte
ns
ity
, c
ps
4.95
5.20
Fresh mobile
phases….Checked the
obvious (e.g.
leaks)
![Page 19: Michael Wright Bioanalytical Services LGC Fordham, UK · Science for a safer world Case studies: Troubleshooting failed system suitability tests Michael Wright Bioanalytical Services](https://reader034.vdocuments.us/reader034/viewer/2022042206/5ea8e3ff577ae3285923d5d3/html5/thumbnails/19.jpg)
13/10/2016 19
Back pressure not
fluctuating and purge
valves not leaking
Back pressure looks
normal but tested flow
rate anyway – OK
Peaks are late but
drifting backward and
forward so unlikely to
be column at fault
XIC of +MRM (4 pairs): 383.200/257.300 Da ID: 25OHD3 1 from Sample 208 (2016_03_07 TN) of SST.wiff (Heated Nebulizer) Max. 4385.0 cps.
0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0Time, min
0
200
400
600
800
1000
1200
1400
1600
1800
2000
2200
2400
2600
2800
3000
3200
3400
3600
3800
4000
4200
4385
Inte
ns
ity
, c
ps
4.95
5.20
XIC of +MRM (4 pairs): 383.200/257.300 Da ID: 25OHD3 1 from Sample 205 (2016_03_07 TN) of SST.wiff (Heated Nebulizer) Max. 1.3e4 cps.
0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0Time, min
0.00
500.00
1000.00
1500.00
2000.00
2500.00
3000.00
3500.00
4000.00
4500.00
5000.00
5500.00
6000.00
6500.00
7000.00
7500.00
8000.00
8500.00
9000.00
9500.00
1.00e4
1.05e4
1.10e4
1.15e4
1.20e4
1.25e4
Inte
ns
ity
, c
ps
5.37
XIC of +MRM (4 pairs): 383.200/257.300 Da ID: 25OHD3 1 from Sample 206 (2016_03_07 TN) of SST.wiff (Heated Nebulizer) Max. 170.0 cps.
0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0Time, min
0
10
20
30
40
50
60
70
80
90
100
110
120
130
140
150
160
170
180
190
200
210
220
230
240
250
260
Inte
ns
ity
, c
ps
XIC of +MRM (4 pairs): 383.200/257.300 Da ID: 25OHD3 1 from Sample 208 (2016_03_07 TN) of SST.wiff (Heated Nebulizer) Max. 4385.0 cps.
0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0Time, min
0
200
400
600
800
1000
1200
1400
1600
1800
2000
2200
2400
2600
2800
3000
3200
3400
3600
3800
4000
4200
4385
Inte
ns
ity
, c
ps
4.95
5.20
![Page 20: Michael Wright Bioanalytical Services LGC Fordham, UK · Science for a safer world Case studies: Troubleshooting failed system suitability tests Michael Wright Bioanalytical Services](https://reader034.vdocuments.us/reader034/viewer/2022042206/5ea8e3ff577ae3285923d5d3/html5/thumbnails/20.jpg)
13/10/2016 20
Auto-sampler
Pump(s) 1 Pump(s) 2
Column
OvenMass
Spectrometer
2 position divert valve
On-line SPE cartridge
Analytical column
WASTE
![Page 21: Michael Wright Bioanalytical Services LGC Fordham, UK · Science for a safer world Case studies: Troubleshooting failed system suitability tests Michael Wright Bioanalytical Services](https://reader034.vdocuments.us/reader034/viewer/2022042206/5ea8e3ff577ae3285923d5d3/html5/thumbnails/21.jpg)
13/10/2016 21
Wanted to rule out issues with the
pumps and gradient mixers so
swapped pumps A/B with pumps
C/D
Wanted to rule out any issues with
the auto-sampler so bypassed it
with a manual injector
1st dimension of on-line SPE
seems to work fine in isolation….
![Page 22: Michael Wright Bioanalytical Services LGC Fordham, UK · Science for a safer world Case studies: Troubleshooting failed system suitability tests Michael Wright Bioanalytical Services](https://reader034.vdocuments.us/reader034/viewer/2022042206/5ea8e3ff577ae3285923d5d3/html5/thumbnails/22.jpg)
12
3
4 5
6
![Page 23: Michael Wright Bioanalytical Services LGC Fordham, UK · Science for a safer world Case studies: Troubleshooting failed system suitability tests Michael Wright Bioanalytical Services](https://reader034.vdocuments.us/reader034/viewer/2022042206/5ea8e3ff577ae3285923d5d3/html5/thumbnails/23.jpg)
12
3
4 5
6
12
3
4 5
6
A B
![Page 24: Michael Wright Bioanalytical Services LGC Fordham, UK · Science for a safer world Case studies: Troubleshooting failed system suitability tests Michael Wright Bioanalytical Services](https://reader034.vdocuments.us/reader034/viewer/2022042206/5ea8e3ff577ae3285923d5d3/html5/thumbnails/24.jpg)
Rotor seal in a switching valve
![Page 25: Michael Wright Bioanalytical Services LGC Fordham, UK · Science for a safer world Case studies: Troubleshooting failed system suitability tests Michael Wright Bioanalytical Services](https://reader034.vdocuments.us/reader034/viewer/2022042206/5ea8e3ff577ae3285923d5d3/html5/thumbnails/25.jpg)
13/10/2016 25
The scratched rotor seal was possibly causing liquid to leak to
varying degrees. It wasn’t leaking during the start of the run
(when BP was measured)
The instrument was recently brought into a higher throughput
environment but the maintenance schedule hadn’t been
changed
Learning points
• We introduced more regular servicing for the switching
valve.
• Highlighted the importance of having an system
capable of generating a pressure trace.
Case Study No.4
![Page 26: Michael Wright Bioanalytical Services LGC Fordham, UK · Science for a safer world Case studies: Troubleshooting failed system suitability tests Michael Wright Bioanalytical Services](https://reader034.vdocuments.us/reader034/viewer/2022042206/5ea8e3ff577ae3285923d5d3/html5/thumbnails/26.jpg)
Conclusions
Running and documenting SST can:
• Stream-line troubleshooting
• Guide to creating a more robust LC-MS service
• Help plan preventative maintenance
• Highlight areas where your systems could be
improved
13/10/2016 26
![Page 27: Michael Wright Bioanalytical Services LGC Fordham, UK · Science for a safer world Case studies: Troubleshooting failed system suitability tests Michael Wright Bioanalytical Services](https://reader034.vdocuments.us/reader034/viewer/2022042206/5ea8e3ff577ae3285923d5d3/html5/thumbnails/27.jpg)