![Page 1: Identifying Image Artifacts, Their Causes, and How to Fix Themamos3.aapm.org/abstracts/pdf/127-35528-418554-126827.pdf · 8/1/2017 2 Speakers Robert Pooley –MRI Beth Schueler -](https://reader036.vdocuments.us/reader036/viewer/2022062920/5f01fb807e708231d401fe72/html5/thumbnails/1.jpg)
8/1/2017
1
AAPM Annual
Meeting 20171
SAM Practical Medical Physics
TU-B-201-0
Abstract
This course will teach the participant to
identify common artifacts found clinically in
MR, DR, CT, PET, to determine the causes of
artifacts, and to make recommendations for
how to resolve artifacts.
2
Learning Objectives
1. Identify common artifacts found clinically
2. Determine causes of various clinical
artifacts
3. Describe how to resolve various clinical
artifacts
3
![Page 2: Identifying Image Artifacts, Their Causes, and How to Fix Themamos3.aapm.org/abstracts/pdf/127-35528-418554-126827.pdf · 8/1/2017 2 Speakers Robert Pooley –MRI Beth Schueler -](https://reader036.vdocuments.us/reader036/viewer/2022062920/5f01fb807e708231d401fe72/html5/thumbnails/2.jpg)
8/1/2017
2
Speakers
Robert Pooley – MRI
Beth Schueler - Digital Radiography
Jim Kofler - CT
Brad Kemp - PET
4
Robert A. Pooley, Ph.D.
Mayo Clinic Florida
AAPM Annual
Meeting 20175
Outline: Artifact Identification, Cause,
How to Fix
Motion / ghosting
Aliasing / wrap
Radiofrequency interference
Metal
Corduroy
![Page 3: Identifying Image Artifacts, Their Causes, and How to Fix Themamos3.aapm.org/abstracts/pdf/127-35528-418554-126827.pdf · 8/1/2017 2 Speakers Robert Pooley –MRI Beth Schueler -](https://reader036.vdocuments.us/reader036/viewer/2022062920/5f01fb807e708231d401fe72/html5/thumbnails/3.jpg)
8/1/2017
3
Motion / Ghosting
Identification: May appear as repetition of “ghosts” across
image, may be distinct or blurred depending on type of motion
Typically occurs in phase encoding direction
Causes:
Patient motion, physiologic / involuntary, voluntary; equipment vibration / instability
Image space k-spaceFT
K-space trajectory
12
3456789
Kx
Ky
![Page 4: Identifying Image Artifacts, Their Causes, and How to Fix Themamos3.aapm.org/abstracts/pdf/127-35528-418554-126827.pdf · 8/1/2017 2 Speakers Robert Pooley –MRI Beth Schueler -](https://reader036.vdocuments.us/reader036/viewer/2022062920/5f01fb807e708231d401fe72/html5/thumbnails/4.jpg)
8/1/2017
4
![Page 5: Identifying Image Artifacts, Their Causes, and How to Fix Themamos3.aapm.org/abstracts/pdf/127-35528-418554-126827.pdf · 8/1/2017 2 Speakers Robert Pooley –MRI Beth Schueler -](https://reader036.vdocuments.us/reader036/viewer/2022062920/5f01fb807e708231d401fe72/html5/thumbnails/5.jpg)
8/1/2017
5
Propeller / Blade
Reduce sensitivity to motion
May be used for uncooperative patients
Pipe, Magn. Reson. Med. 1999; 42:963-969
K-space
Propeller
![Page 6: Identifying Image Artifacts, Their Causes, and How to Fix Themamos3.aapm.org/abstracts/pdf/127-35528-418554-126827.pdf · 8/1/2017 2 Speakers Robert Pooley –MRI Beth Schueler -](https://reader036.vdocuments.us/reader036/viewer/2022062920/5f01fb807e708231d401fe72/html5/thumbnails/6.jpg)
8/1/2017
6
Propeller / Blade
Image from Siemens Applications guide
Motion / ghosting How to fix:
Remind patient to hold still, make patient more comfortable, decrease scan time, consider sedation
Use respiratory triggering / navigator pulses, motion correction, breath-hold options when available
Implement Propeller / Blade sequence options
Address equipment problems, test with phantom, test with cold-head off
Swap phase and frequency if artifact obscures pathology
![Page 7: Identifying Image Artifacts, Their Causes, and How to Fix Themamos3.aapm.org/abstracts/pdf/127-35528-418554-126827.pdf · 8/1/2017 2 Speakers Robert Pooley –MRI Beth Schueler -](https://reader036.vdocuments.us/reader036/viewer/2022062920/5f01fb807e708231d401fe72/html5/thumbnails/7.jpg)
8/1/2017
7
Aliasing / wrap
Identification:
Patient anatomy appears in incorrect
locations
Causes:
Insufficient sampling
Small FOV with anatomy (signal) outside
FOV
![Page 8: Identifying Image Artifacts, Their Causes, and How to Fix Themamos3.aapm.org/abstracts/pdf/127-35528-418554-126827.pdf · 8/1/2017 2 Speakers Robert Pooley –MRI Beth Schueler -](https://reader036.vdocuments.us/reader036/viewer/2022062920/5f01fb807e708231d401fe72/html5/thumbnails/8.jpg)
8/1/2017
8
![Page 9: Identifying Image Artifacts, Their Causes, and How to Fix Themamos3.aapm.org/abstracts/pdf/127-35528-418554-126827.pdf · 8/1/2017 2 Speakers Robert Pooley –MRI Beth Schueler -](https://reader036.vdocuments.us/reader036/viewer/2022062920/5f01fb807e708231d401fe72/html5/thumbnails/9.jpg)
8/1/2017
9
Aliasing / wrap
How to fix:
Increase FOV
Increase over-sampling in phase encoding
and / or slice directions (for 3D)
Use saturation bands for anatomy outside
FOV
Turn off coils outside imaging volume
RF Interference / Zippers
Identification:
Typically appears as single or multiple lines
(“zipper”) in the phase encoding direction
Causes:
Unexpected radiofrequency signal from
equipment inside room, or outside room with
poor RF shielding
![Page 10: Identifying Image Artifacts, Their Causes, and How to Fix Themamos3.aapm.org/abstracts/pdf/127-35528-418554-126827.pdf · 8/1/2017 2 Speakers Robert Pooley –MRI Beth Schueler -](https://reader036.vdocuments.us/reader036/viewer/2022062920/5f01fb807e708231d401fe72/html5/thumbnails/10.jpg)
8/1/2017
10
![Page 11: Identifying Image Artifacts, Their Causes, and How to Fix Themamos3.aapm.org/abstracts/pdf/127-35528-418554-126827.pdf · 8/1/2017 2 Speakers Robert Pooley –MRI Beth Schueler -](https://reader036.vdocuments.us/reader036/viewer/2022062920/5f01fb807e708231d401fe72/html5/thumbnails/11.jpg)
8/1/2017
11
RF Interference / Zippers
How to fix:
Identify source, replace electronic
components generating unwanted RF signal
Check integrity of RF shielding, clean RF
door threshold and RF fingers / plates
Metal
Identification:
Signal void, often with adjacent very bright
signal
Geometric distortion
Causes:
Magnetic susceptibility, induced eddy
currents, spin dephasing
Metal
![Page 12: Identifying Image Artifacts, Their Causes, and How to Fix Themamos3.aapm.org/abstracts/pdf/127-35528-418554-126827.pdf · 8/1/2017 2 Speakers Robert Pooley –MRI Beth Schueler -](https://reader036.vdocuments.us/reader036/viewer/2022062920/5f01fb807e708231d401fe72/html5/thumbnails/12.jpg)
8/1/2017
12
![Page 13: Identifying Image Artifacts, Their Causes, and How to Fix Themamos3.aapm.org/abstracts/pdf/127-35528-418554-126827.pdf · 8/1/2017 2 Speakers Robert Pooley –MRI Beth Schueler -](https://reader036.vdocuments.us/reader036/viewer/2022062920/5f01fb807e708231d401fe72/html5/thumbnails/13.jpg)
8/1/2017
13
![Page 14: Identifying Image Artifacts, Their Causes, and How to Fix Themamos3.aapm.org/abstracts/pdf/127-35528-418554-126827.pdf · 8/1/2017 2 Speakers Robert Pooley –MRI Beth Schueler -](https://reader036.vdocuments.us/reader036/viewer/2022062920/5f01fb807e708231d401fe72/html5/thumbnails/14.jpg)
8/1/2017
14
![Page 15: Identifying Image Artifacts, Their Causes, and How to Fix Themamos3.aapm.org/abstracts/pdf/127-35528-418554-126827.pdf · 8/1/2017 2 Speakers Robert Pooley –MRI Beth Schueler -](https://reader036.vdocuments.us/reader036/viewer/2022062920/5f01fb807e708231d401fe72/html5/thumbnails/15.jpg)
8/1/2017
15
Metal
How to fix / reduce: Properly screen patient, remove any metal that
can be removed
Turbo spin echo and spin echo sequences reduce artifact compared to gradient echo
Decrease TE and echo spacing; increase bandwidth and resolution in frequency encoding direction
Swap phase and frequency to modify shape
Corduroy
Identification: Pattern of regularly spaced lines extending
across image
Can occur at different spatial frequencies and different angles
Multiple sets can combine to appear as a cross-hatch pattern
Causes: Spike(s) in k-space
Image space k-spaceFT
![Page 16: Identifying Image Artifacts, Their Causes, and How to Fix Themamos3.aapm.org/abstracts/pdf/127-35528-418554-126827.pdf · 8/1/2017 2 Speakers Robert Pooley –MRI Beth Schueler -](https://reader036.vdocuments.us/reader036/viewer/2022062920/5f01fb807e708231d401fe72/html5/thumbnails/16.jpg)
8/1/2017
16
Image space k-spaceFT
Image space k-spaceFT
![Page 17: Identifying Image Artifacts, Their Causes, and How to Fix Themamos3.aapm.org/abstracts/pdf/127-35528-418554-126827.pdf · 8/1/2017 2 Speakers Robert Pooley –MRI Beth Schueler -](https://reader036.vdocuments.us/reader036/viewer/2022062920/5f01fb807e708231d401fe72/html5/thumbnails/17.jpg)
8/1/2017
17
Corduroy
How to fix:
Confirm room humidity in specification
Perform scanner spike check
Attempt to localize source if in room
Call Service
![Page 18: Identifying Image Artifacts, Their Causes, and How to Fix Themamos3.aapm.org/abstracts/pdf/127-35528-418554-126827.pdf · 8/1/2017 2 Speakers Robert Pooley –MRI Beth Schueler -](https://reader036.vdocuments.us/reader036/viewer/2022062920/5f01fb807e708231d401fe72/html5/thumbnails/18.jpg)
8/1/2017
18
Bonus
Artifact:
Reddit: “Camera Malfunction”
Courtesy Catherine Sturchio
Suggested Reading
DW McRobbie, EA Moore, MJ Graves, MR
Prince, “MRI: From Picture to Proton,” 2nd
edition, Cambridge University Press
Dietrich, “Artifacts in 3-Tesla MRI: Physical
background and reduction strategies” EJR
2008 65(1) 29-35
Thank You
Mayo Clinic Florida
![Page 19: Identifying Image Artifacts, Their Causes, and How to Fix Themamos3.aapm.org/abstracts/pdf/127-35528-418554-126827.pdf · 8/1/2017 2 Speakers Robert Pooley –MRI Beth Schueler -](https://reader036.vdocuments.us/reader036/viewer/2022062920/5f01fb807e708231d401fe72/html5/thumbnails/19.jpg)
8/1/2017
19
Flow
Identification:
Ghosting, appears as repetition (phase
encoding direction) of blood
Causes:
Velocity induced phase effects
![Page 20: Identifying Image Artifacts, Their Causes, and How to Fix Themamos3.aapm.org/abstracts/pdf/127-35528-418554-126827.pdf · 8/1/2017 2 Speakers Robert Pooley –MRI Beth Schueler -](https://reader036.vdocuments.us/reader036/viewer/2022062920/5f01fb807e708231d401fe72/html5/thumbnails/20.jpg)
8/1/2017
20
![Page 21: Identifying Image Artifacts, Their Causes, and How to Fix Themamos3.aapm.org/abstracts/pdf/127-35528-418554-126827.pdf · 8/1/2017 2 Speakers Robert Pooley –MRI Beth Schueler -](https://reader036.vdocuments.us/reader036/viewer/2022062920/5f01fb807e708231d401fe72/html5/thumbnails/21.jpg)
8/1/2017
21
Flow
How to fix:
Gradient moment nulling / flow
compensation
Spatial saturation bands
Swap phase and frequency if artifact mimics
pathology