Product Training
FreeStyle Optium Neo H Blood Glucose and Ketone Monitoring System
Healthcare professional portrayal.
![Page 2: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/2.jpg)
Glucose Ketones
TheFreeStyleNeoHcanperformbothbloodglucoseandbloodKetonetests.Ensuretofamiliariseyourselfwithbothmodules.Clickonthemodulebelow.
Patient testing
Quality Control
Maintenance
Patient testing
Quality Control
Maintenance
![Page 3: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/3.jpg)
AnIntroduc+ontoFreeStyleOp+umNeoH
![Page 4: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/4.jpg)
TheFreeStyleOpAumNeoHSystemataGlance
High contrast display Displays meter home screen, test results
and other important information
Up button • Use to mark a venous blood test
• Adjust settings
Test strip port Where the test strip is inserted
Logbook
Down button • Used to mark a control solution test
• Adjust settings
Power button Press to:
• Turns meter on and off
• Returns to the home screen
• Saves settings
USB port Insert a data cable for uploading test results to a computer
![Page 5: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/5.jpg)
TheFreeStyleOpAumNeoHBloodGlucoseTestStrip
TheFreeStyleOpAumNeoHmeterisdesignedtobeusedonlywithFreeStyleOpAumNeoHBloodGlucoseTestStrips.• 0.6μLsamplesize• 5-secondresultcountdown• Assayrange1.1-27.8mmol/L• Haematocritrange15-65%• ToporApsampleapplicaAon• VisualconfirmaAonofsufficientsample• Filltriggerensuressufficientsample• SampleTypes–Capillary,venous,arterial&neonatal• Eachpackcontains100individuallyfoilwrappedteststrips,calibraAonbarandteststripinsert
![Page 6: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/6.jpg)
MetercalibraAon
• ThecalibraAonprocedureprogrammesthemeterwiththelotnumber,expirydateandteststriptechnology
• Alwayscalibratethemeterwitheachnewpackofteststrips–failuretocalibrateproperlywillcauseincorrectresults
• Withthelotnumberfacingtowardyou,insertthecontactbarsofthecalibratorintothemeter.ThemeterturnsonautomaAcally
• LOTandthelotnumberappearinthedisplaywindow• Checkthatthelotnumberonthemeterdisplaywindowmatches
withthenumberontheteststripcalibratorandthelastfivedigitsonboththeteststripfoilpacketandteststripinsert
• RetainthecalibraAonbarfortheduraAonofthatpack• CalibraAoniscomplete• Onlyusethecalibratorthatispackagedinthepackofteststrips
youareusing
![Page 7: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/7.jpg)
PerformingaPa+entTestThe FreeStyle Optium Neo H system enables easy and accurate bedside testing of blood glucose levels
} 5 second test time
} Small 0.6 μL sample size
} Calibrator provides expiry check and
calibrates the meter
} Control solution mark
Blood glucose testing
![Page 8: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/8.jpg)
PaAentTesAng1. Preparethesamplesite:
i. Beforeyouobtainablooddrop,makesurethesamplesiteisclean,dryandwarm
ii. Hangthearmdownbeforelancingthesamplesite,tohelpaidbloodflow
2. Opentheindividuallyfoilwrappedteststripatthenotchandteardown
3. InserttheteststripintothemeterunAlitstops.Thiswillturnthemeteron.
4. CheckthattheLOTnumberofteststripyouareusingmatchestheLOTnumbershownonthemeterdisplay
5. The blinks,indicaAngthemeterisreadyforyoutoapplyasampletotheteststripi. EnsureyoufolloworganisaAonalpolicyforsamplecollecAon
![Page 9: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/9.jpg)
PaAenttesAng
6. IftesAngglucoseonvenousbloodpressthetomarkthetestunAlappearsindicaAngthatthemeterisnowreadyforyoutoapplyvenousbloodtotheteststrip7. Applyasamplewiththemeterinahorizontal(flat) posiAontoavoiditenteringtheteststripport8. Ifyouareapplyingthesamplefromasyringe,applyaway
fromthedirecAonoftheteststripport
![Page 10: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/10.jpg)
PaAenttesAng
9. HoldsampletotheteststripunAl3shortlinesappearonthescreen
10. Theresultswillbedisplayedafera5-secondcountdown
![Page 11: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/11.jpg)
PaAenttesAng
Whentheresultappearsonthemeterscreenthetestiscomplete.Theresultswillbestoredinthemeter’smemory.Pressandholdthebuhontoturnoffthemeter.
![Page 12: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/12.jpg)
SummaryforbloodglucosetesAng
1. Insertsteststrip,blinks
2.Applyblood3.Sampledetected,5secondscountdownbegins
4.Resultisdisplayed
![Page 13: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/13.jpg)
Outofrangeindicators
Highbloodglucoseresults• AresultfallingoutsidetherangesetbyyourorganisaAonmay
requiresomeintervenAonfromtheoperator,suchasrepeaAngthetest– Themeterwilldisplayasolidyellowarrow
• Forguidanceonoutofrangeresults,pleasefollowyourorganisaAonalpolicy
Lowbloodglucoseresults• AresultfallingoutsidetherangesetbyyourorganisaAonwill
requiresomeintervenAonfromtheoperator,suchasrepeaAngthetest– Themeterwilldisplayasolidredarrow
• Forguidanceonoutofrangeresults,pleasefollowyourorganisaAonalpolicy
![Page 14: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/14.jpg)
DisplayMessages
Result is higher than or equal to 13.3 mmol/L
Glucose result is above 27.8 mmol/L
Result is less than 1.1 mmol/L
![Page 15: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/15.jpg)
Results&Memory
Thetestiscompletewhentheresultappearsonthemeterscreen.• Theresultisstoredinthemeter’slogbook• Themeterlogbookcanstoreupto1,000events–includingbloodglucose,bloodketone,andcontrolsoluAonresults,andothermeterinformaAon• ThelogbookcanbeaccessedbygoingtotheHomeScreenandselecAng• TorecallpaAentresults:WiththeFreeStyleOpAumNeoHturnedon:
1. Pressthe toviewlogbookevents2. Press or toscrollthroughlogbookevents
While on the Home screen, press to view logbook events
![Page 16: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/16.jpg)
Logbook:ViewLogbookEvents
Press to scroll through to view logbook events
Blood Glucose Result
Glucose Control Solution Result
Blood Ketone Result Venous Blood Result
Ketone Control Solution Result
Control solution test results will alternate between ‘CtL’ and result
![Page 17: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/17.jpg)
Logbook:ViewBloodGlucoseAverages
Average
Number of tests In this average
1. While in the logbook, press any time to view the 7-day average
2. Press or to scroll through 7-, 14- and 30-day averages. Press to return to logbook events.
![Page 18: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/18.jpg)
ErrorMessages
Message What it means
E1 Temperature is too hot or too cold for the meter to work correctly
E2/E9 Meter error
E3 i. Blood drop is too small
ii. Incorrect test procedure
iii. There may be a problem with the test strip
E4 i. The blood glucose level may be too high to be read by the system
ii. There may be a problem with the test strip
E5 i. Blood was applied to the test strip too soon
E6 i. Calibration error
ii. Test strip error
E7 i. Test strip may be damaged, used or the meter does not recognise it
![Page 19: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/19.jpg)
PerformingaQualityControlTest
![Page 20: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/20.jpg)
PerformingaQualityControlTest
• IfQCtesAngisNOTperformed,misleadingresultsmayleadtoinappropriatetreatment
• AlwaysfollowyourorganisaAonalpolicytoavoidpumngpaAentsatrisk• TheFreeStyleOpAumNeoHmetermaypromptwhenthisisdue• OnlyuseMediSenseGlucoseandKetoneControlSoluAonsandkeepthemeter
flattopreventexcesssoluAonrunningintothemeter• QCSoluAonisstablefor90daysfollowingopeningorunAltheexpirydate
printedonthesoluAonbohles,whicheverissooner.DiscardanddoNOTusecontrolsoluAonpastitsexpirydate
• CheckresultsagainsttheteststripinsertinsidethepackofteststripsbeingusedasQCrangeswillvarybetweenteststripLOTs
• ItistheresponsibilityofalluserstoensurethemeterhasbeentestedbeforepaAentuse
![Page 21: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/21.jpg)
PerformingQualityControl(QC)Tests1. CheckthattheLOTnumberprintedontheteststripandinstrucAonsforuse
match2. Opentheindividualfoilteststripatthenotchandteardown3. Inserttheteststripintothemeterwhichwillturnthemeteronturnon4. WhentheblinkspressthebuhonunAlthisiconappears5. ThemeterisnowreadyforyoutoapplythelowlevelcontrolsoluAontothetest
stripi. InvertthebohletomixthesoluAonii. WipetheQCnozzlewithcleangauzebeforeandafereachtestiii. Replacethecapsecurelyonthebohleimmediatelyaferuse
7. ApplyadropofthecontrolsoluAonwiththemeterinahorizontalposiAontoavoiditenteringtheteststripport
![Page 22: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/22.jpg)
PerformingQualityControl(QC)Tests8. ApplyadropofthecontrolsoluAontotheteststripunAl3shortlinesappearon
thescreen9. Theresultswillbedisplayedafera5-secondcountdown
i. ComparethecontrolsoluAonresultstotherangeprintedonthebloodglucoseteststripinsert; theresultsshouldfallwithinthisrangei. TheresultswillbestoredinmemoryasacontrolsoluAonresultii. RecordtheresultintheQCrecordbook
10. Ifresultsareout-of-rangei. Repeatthetestii. Iftherepeatedresultisout-of-rangefolloworganisaAonalpolicyforout-of-rangeresults
11. Removetheteststrip,repeattheprocedureforhighlevelcontrolsoluAon12. RecordHIandLOQCtestresultintheQCrecordbook
![Page 23: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/23.jpg)
QualityControlTesAngSummary
1. Insertteststrip,blinks.
2.PressandholdtheunAlappears.
3.ThemeterisnowreadyfortheoperatortoapplycontrolsoluAontotheteststrip.
4.ApplycontrolsoluAon.5.Sampledetected,countdown
begins.5.Resultisdisplayed.
![Page 24: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/24.jpg)
Maintenance
![Page 25: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/25.jpg)
Maintenanceofthemeter
1. GentlycleantheexterioroftheFreeStyleOpAumNeoHmeterwithadampclothi. AcceptablecleaningsoluAonsincludemilddetergentandwater,70%isopropylalcohol
oramixtureof1parthouseholdbleachand9partswaterii. Donotimmersethemeterinwaterorotherliquids.Avoidgemngdust,blood,control
soluAonandanyothersubstancesintheteststripport,USBportandbaherycompartment
iii. Allowthemetertodry
2. Donottrytocleantheteststripport.3. Donotpourliquidintotheteststripportorontothebuhons.4. Donotimmersethemeterinwateroranyotherliquid.
![Page 26: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/26.jpg)
Maintenanceofthemeter1. Replacecoincellbaherieswhen‘LowBaheryPower’messageisdisplayed
1. Themeterrequires2replaceableCR2032coincellbaheries.2. Tochangethebaheriesturnthemeteroverandslidethebaherydoor
(locatedonthesideofthedevice)open,asshown.2. PerformQCtestasperorganisaAonalpolicy3. RefertotroubleshooAngguideinFreeStyleOpAumNeoHoperator’smanualfor
errormessages4. Whennotinuse,alwaysstoretheFreeStyleOpAumNeoHmeterasper
organisaAonalpolicy
![Page 27: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/27.jpg)
Bloodβ-KetoneTes+ngwithFreeStyleOp+umNeoH
The FreeStyle Optium Neo H system enables easy and accurate bedside testing of blood ketone levels
} 10 second test time
} 1.5 μL sample size
} Purple calibrator calibrates the meter
} Control solution mark
Blood ketone testing
![Page 28: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/28.jpg)
AnIntroduc+ontoFreeStyleOp+umNeoH
![Page 29: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/29.jpg)
TheFreeStyleOpAumNeoHSystemataGlance
High contrast display Displays meter home screen, test results
and other important information
Up button • Use to mark a venous blood test
• Adjust settings
Test strip port Where the test strip is inserted
Down button • Used to mark a control solution test
• Adjust settings
Power button Press to:
• Turns meter on and off
• Returns to the home screen
• Saves settings
USB port Insert a data cable for uploading test results to a computer
![Page 30: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/30.jpg)
FreeStyle Optium H Blood β-Ketone Test Strips
TheFreeStyleOpAumHBloodβ-KetoneTestStrip:• 1.5μLsamplesize• 10-secondresultcountdown• AssayRange0.0-8.0mmol/L• HaematocritRange30-60%• ToporApsampleapplicaAon• VisualconfirmaAonofsufficientsample• Filltriggerensuressufficientsample• SampleType–capillary(finger)andvenouswholebloodsamples• Eachpackcontains10individuallyfoilwrappedteststrips,calibraAonbarandteststripinsert
![Page 31: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/31.jpg)
β-Ketone Benefits
Clinicalaccuracy• Measuresβ-hydroxybutyrate(themostsignificantketonebody)• AquanAtaAvevs.qualitaAvemeasure• ResultsunaffectedbymanycommonmedicaAons
Pa+entconvenience/compliance• Bloodvs.urinesampling• Foilwrappedteststrips• Testintegratedwithinbloodglucosemeters
![Page 32: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/32.jpg)
MetercalibraAon
• ThecalibraAonprocedureprogrammesthemeterwiththelotnumber,expirydateandteststriptechnology
• Alwayscalibratethemeterwitheachnewpackofteststrips–failuretocalibrateproperlywillcauseincorrectresults
• Withthelotnumberfacingtowardyou,insertthecontactbarsofthecalibratorintothemeter.ThemeterturnsonautomaAcally
• LOTandthelotnumberappearinthedisplaywindow• Checkthatthelotnumberonthemeterdisplaywindowmatches
withthenumberontheteststripcalibratorandthelastfivedigitsonboththeteststripfoilpacketandteststripinsert
• RetainthecalibraAonbarfortheduraAonofthatpack• CalibraAoniscomplete• Onlyusethecalibratorthatispackagedinthepackofteststrips
youareusing
![Page 33: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/33.jpg)
PerformingaPa+entTest
![Page 34: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/34.jpg)
PaAentTesAng1. Preparethesamplesite:
i. Beforeyouobtainablooddrop,makesurethesamplesiteisclean,dryandwarm
ii. Hangthearmdownbeforelancingthesamplesite,tohelpaidbloodflow
2. Opentheindividuallyfoilwrappedteststripatthenotchandteardown
3. InserttheteststripintothemeterunAlitstops.Thiswillturnthemeteron.
4. CheckthattheLOTnumberofteststripyouareusingmatchestheLOTnumbershownonthemeterdisplay
5. The blinks,indicaAngthemeterisreadyforyoutoapplyasampletotheteststripi. EnsureyoufolloworganisaAonalpolicyforsamplecollecAon
![Page 35: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/35.jpg)
PaAenttesAng
6. IftesAngketonesonvenousbloodpressthetomarkthetestunAlappearsindicaAngthatthemeterisnowreadyforyoutoapplyvenousbloodtotheteststrip7. Applyasamplewiththemeterinahorizontal(flat) posiAontoavoiditenteringtheteststripport8. Ifyouareapplyingthesamplefromasyringe,applyaway
fromthedirecAonoftheteststripport
![Page 36: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/36.jpg)
PaAenttesAng
9. HoldsampletotheteststripunAl3shortlinesappearonthescreen
10. Theresultswillbedisplayedafera10-secondcountdown
![Page 37: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/37.jpg)
PaAenttesAng
Whentheresultappearsonthemeterscreenthetestiscomplete.Theresultswillbestoredinthemeter’smemory.Pressandholdthebuhontoturnoffthemeter.
![Page 38: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/38.jpg)
DisplayMessages
Message Explanation
Result above 8.0 mmol/L
Thisappearswhenaresultishigherthan8.0mmol/LThissuggestsveryhighbloodketonelevelsorthattheremaybeanerrorwiththeteststripRepeatthetestwithanewteststripThisrequiresimmediateacAonasdefinedbyyourfacility’spoliciesandprocedures
![Page 39: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/39.jpg)
1. Insert ketone test strip, blinks and KET appears
2. Apply blood
3. Sample detected, 10 seconds countdown begins
4. Result is displayed
Blood Ketone Testing summary
![Page 40: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/40.jpg)
ErrorMessages
Message What it means
E1 Temperature is too hot or too cold for the meter to work correctly
E2/E9 Meter error
E3 i. Blood drop is too small
ii. Incorrect test procedure
iii. There may be a problem with the test strip
E4 i. The blood glucose level may be too high to be read by the system
ii. There may be a problem with the test strip
E5 i. Blood was applied to the test strip too soon
E6 i. Calibration error
ii. Test strip error
E7 i. Test strip may be damaged, used or the meter does not recognise it
![Page 41: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/41.jpg)
PerformingaQualityControlTest
![Page 42: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/42.jpg)
PerformingaQualityControlTest
• IfQCtesAngisNOTperformed,misleadingresultsmayleadtoinappropriatetreatment
• AlwaysfollowyourorganisaAonalpolicytoavoidpumngpaAentsatrisk• TheFreeStyleOpAumNeoHmetermaypromptwhenthisisdue• OnlyuseMediSenseGlucoseandKetoneControlSoluAonsandkeepthemeter
flattopreventexcesssoluAonrunningintothemeter• QCSoluAonisstablefor90daysfollowingopeningorunAltheexpirydate
printedonthesoluAonbohles,whicheverissooner.DiscardanddoNOTusecontrolsoluAonpastitsexpirydate
• CheckresultsagainsttheteststripinsertinsidethepackofteststripsbeingusedasQCrangeswillvarybetweenteststripLOTs
• ItistheresponsibilityofalluserstoensurethemeterhasbeentestedbeforepaAentuse
![Page 43: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/43.jpg)
PerformingQualityControl(QC)Tests1. CheckthattheLOTnumberprintedontheteststripandinstrucAonsforusematch2. Opentheindividualfoilteststripatthenotchandteardown3. Inserttheteststripintothemeterwhichwillturnthemeteronturnon4. WhentheblinkspressthebuhonunAlthisiconappears5. ThemeterisnowreadyforyoutoapplythelowlevelcontrolsoluAontothetest
stripi. InvertthebohletomixthesoluAonii. WipetheQCnozzlewithcleangauzebeforeandafereachtestiii. Replacethecapsecurelyonthebohleimmediatelyaferuse
7. ApplyadropofthecontrolsoluAonwiththemeterinahorizontalposiAontoavoiditenteringtheteststripport*KETwillappearonthescreenasapurpleketoneteststriphasbeeninserted
![Page 44: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/44.jpg)
PerformingQualityControl(QC)Tests8. ApplyadropofthecontrolsoluAontotheteststripunAl3shortlinesappearon
thescreen9. Theresultswillbedisplayedafera10-secondcountdown
i. ComparethecontrolsoluAonresultstotherangeprintedonthebloodglucoseteststripinsert; theresultsshouldfallwithinthisrangei. TheresultswillbestoredinmemoryasacontrolsoluAonresultii. RecordtheresultintheQCrecordbook
10. Ifresultsareout-of-rangei. Repeatthetestii. Iftherepeatedresultisout-of-rangefolloworganisaAonalpolicyforout-of-rangeresults
11. Removetheteststrip,repeattheprocedureforhighlevelcontrolsoluAon12. RecordHIandLOQCtestresultintheQCrecordbook
![Page 45: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/45.jpg)
QualityControlTesAngSummary
1. Insertteststrip,blinks.
2.PressandholdtheunAlappears.
3.ThemeterisnowreadyfortheoperatortoapplycontrolsoluAontotheteststrip.
4.ApplycontrolsoluAon.5.Sampledetected,countdown
begins.5.Resultisdisplayed.
![Page 46: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/46.jpg)
Maintenance
![Page 47: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/47.jpg)
Maintenanceofthemeter
1. GentlycleantheexterioroftheFreeStyleOpAumNeoHmeterwithadampclothi. AcceptablecleaningsoluAonsincludemilddetergentandwater,70%isopropylalcohol
oramixtureof1parthouseholdbleachand9partswaterii. Donotimmersethemeterinwaterorotherliquids.Avoidgemngdust,blood,control
soluAonandanyothersubstancesintheteststripport,USBportandbaherycompartment
iii. Allowthemetertodry
2. Donottrytocleantheteststripport.3. Donotpourliquidintotheteststripportorontothebuhons.4. Donotimmersethemeterinwateroranyotherliquid.
![Page 48: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/48.jpg)
Maintenanceofthemeter1. Replacecoincellbaherieswhen‘LowBaheryPower’messageisdisplayed
1. Themeterrequires2replaceableCR2032coincellbaheries.2. Tochangethebaheriesturnthemeteroverandslidethebaherydoor
(locatedonthesideofthedevice)open,asshown.2. PerformQCtestasperorganisaAonalpolicy3. RefertotroubleshooAngguideinFreeStyleOpAumNeoHoperator’smanualfor
errormessages4. Whennotinuse,alwaysstoretheFreeStyleOpAumNeoHmeterasper
organisaAonalpolicy
![Page 49: FreeStyle Optium Neo H · 2016-05-10 · The FreeStyle Opum Neo H System at a Glance High contrast display Displays meter home screen, test results and other important information](https://reader030.vdocuments.us/reader030/viewer/2022040506/5e3dbbe0e5a82c516226d82a/html5/thumbnails/49.jpg)
CongratulaAonsoncompleAngthismodule
FreeStyle blood glucose meters assist in the monitoring of blood glucose (and ketone) levels for people diagnosed with diabetes mellitus. People diagnosed with diabetes with elevated blood ketone levels should seek medical advice. Information contained herein for distribution outside of the U.S. Only. FreeStyle and related brand marks are trademarks of Abbott Diabetes Care Inc. in various jurisdictions. All other trademarks are the property of their respective owners. Abbott Diabetes Care Australia, 666 Doncaster Road, Doncaster VIC 3180. For further information call Customer Service on 1800 801 478 or visit www.myfreestyle.com.au MSE160420044321