Four Channels Data Acquisition System for Silicon Photomultipliers
Mateusz Baszczyk, Piotr Dorosz, Sebastian Głąb,Wojciech Kucewicz, Łukasz Mik, Maria Sapor
Department of Electronics,AGH – University of Science and Technology,Al. Mickiewicza 30, 30-059 Krakow, Polande-mail [email protected]@agh.edu.pl
![Page 2: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/2.jpg)
Agenda
• Fluorescence;• Silicon Photomultiplier description;• Description of previous measurement system;• Four channel data acquisition system;• Results;• Summary;• Future work.
2
![Page 3: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/3.jpg)
Why do we measure fluorescence light intensity?
• Dye-labelled cells can be count by fluorescence light intensity measurement;
• At low concentrations the fluorescence intensity is generally proportional to the concentrations of fluorophore.
3
![Page 4: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/4.jpg)
Fluorescein and Bio-Tek filters spectra
fluorescein resorufin
4
![Page 5: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/5.jpg)
Optical block
Blue LED
S500LLB4G-H+
485/20nm optical filter
1.5 ml cuvette with fluorescein solution
530/25nm optical filter
5mm diameter optical channel
5mm diameter optical collimator
and fiber cable with 940um core
5
![Page 6: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/6.jpg)
Silicon Photomultiplier
The idea appears in the late 90-ties:•„Photon Detection with High Gain Avalanche Photodiode Arrays” S. Vasile, P. Gothoskar, R. Farrell, D. Sdrulla•„Silicon avalanche photodiodes on the base of metal-resistor-semiconductor (MRS) structures” V. Saveliev, V. Golovin 6
![Page 7: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/7.jpg)
Structure
7
![Page 8: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/8.jpg)
Current - Voltage Characteristic
8
![Page 9: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/9.jpg)
Photodiode Operation Mode
1 photon generates 1 electron-hole pair9
![Page 10: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/10.jpg)
Avalanche Operation Mode
1 photon generates 10-200 electron-hole pairs10
![Page 11: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/11.jpg)
Geiger Operation Mode
1 photon generates 105 - 106 electron-hole pairs11
![Page 12: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/12.jpg)
Typical SiPM Parameters
sensL s1020 Hamamatsu S10362-11-025C
Spectral Range (λ) 400-1100 nm 320-900 nm
Peak Spectral Response (λ) 490 nm 440 nm
Breakdown Voltage (Vbr) 28.6 V ---
Operating Voltage (V) 30.6 V 70.82 V
Microcell Gain 106 2.75 · 105
SPM Pixel Active Area Φ 1 mm Φ 1 mm
Number of Pixels 400 1600
Dark Rate (MHz) 0.42 0.3
12
![Page 13: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/13.jpg)
Previous Measurement System For Fluorescence
Picosecond 2600C
generator
Cuvette holderwith
specimen
GPAamplifier
QDC device
PC with Labview
application
excitationfilter
emissionfilter
Agilent33250
generator
trigger signal gate pulses
13
![Page 14: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/14.jpg)
First Measurement System For Fluorescence
14
![Page 15: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/15.jpg)
Acquisition System
15
![Page 16: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/16.jpg)
Acquisition System
16
![Page 17: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/17.jpg)
Front-end ASIC
17
![Page 18: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/18.jpg)
Front-end ASIC
18
![Page 19: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/19.jpg)
Front-end ASIC
19
![Page 20: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/20.jpg)
Linearity – high gain
20
![Page 21: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/21.jpg)
Linearity – low gain
21
![Page 22: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/22.jpg)
Measurement results
22
![Page 23: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/23.jpg)
Measurement results
23
![Page 24: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/24.jpg)
Measurement results
......),,,(22
22
21
21
22
21111
mxmx
eaeamaxf 24
![Page 25: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/25.jpg)
Measurement results
25
![Page 26: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/26.jpg)
Measurement results
26
![Page 27: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/27.jpg)
Measurement results
27
![Page 28: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/28.jpg)
Light Intensity
28
![Page 29: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/29.jpg)
Light Intensity
29
![Page 30: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/30.jpg)
Light Intensity
30
![Page 31: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/31.jpg)
Without Coincidence
31
![Page 32: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/32.jpg)
Coincidence
32
![Page 33: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/33.jpg)
Results of fluorescence light measurements
976.6 [pg/ml]61.05 [pg/ml]
y = -572500 x² + 223.6 x + 6458000 – fluoresceiny = -373800 x² + 170.3 x + 14030000 – resorufin
33
![Page 34: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/34.jpg)
Influence of Temperature and Bias
34
![Page 35: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/35.jpg)
Dark Current Rate
35
![Page 36: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/36.jpg)
Influence of Temperature and Bias
36
![Page 37: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/37.jpg)
Conclusions
• We have designed, built and optimized readout system for detection of single photons;
• We have tested readout circuit with light and confirmed that it is linear;
• We can distinguish a dozen or so photons and eliminate thermal generation from measurement data;
• System with silicon photomultiplier has a good sensitivity and can be used to measure fluorescence light intensity for fluorescein concentration from 1ug/ml to about 60pg/ml and resorufin at the level of 1ng/ml up to 1ug/ml.
37
![Page 38: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/38.jpg)
Future work
• Stabilization system compensating temperature influence on SiPM gain by varying supply voltage;
• Measurements of fluorescent dyes in microbioreactors.
scanning probe
micro-bioreactor
38
![Page 39: Four Channels Data Acquisition System for Silicon Photomultipliers](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813480550346895d9b5e49/html5/thumbnails/39.jpg)
Thank you for your attention
39