Control of Bioaerosols
![Page 2: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/2.jpg)
Concentrations in Air
• Concentrations of indoor bacteria range from <100-1500 cfu/m3; avg 0-500– heated dwellings lower
• Avg concentrations of outdoor bacteria 200-1000
• Avg concentrations of indoor Fungi range from 10-500
• Avg concentrations of Outdoor Fungi range from 100-1000
![Page 3: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/3.jpg)
Air Quality Control Methods
SourceControl
VentilationPressureControl
AirTreatment
![Page 4: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/4.jpg)
![Page 5: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/5.jpg)
![Page 6: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/6.jpg)
![Page 7: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/7.jpg)
Air Filtration
• Part of most HVAC systems• Hospitals have very intricate systems• No universal standard method for evaluating
filter performance– Thus advertised efficiencies may be misleading
• Most difficult range of particles to filter 0.2 to 0.3 micron; generally particles categorized as > 0.5 or < 0.5 microns
• Peak Particle Removal with 12-15 ACH
![Page 8: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/8.jpg)
Types of Filters
• Fiberglass- typical household filter 5-7% effective
• Electronic- lose filtering 50% capacity in first week
• HEPA- High Efficiency Particulate Air, 99.97% of pollutants at 0.3 microns
• ULPA- Ultra Low Penetration Air, 99.9997% of particles as 0.1 micron
• Electrostatic- “static charged”, 95%
![Page 9: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/9.jpg)
![Page 10: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/10.jpg)
![Page 11: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/11.jpg)
![Page 12: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/12.jpg)
![Page 13: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/13.jpg)
![Page 14: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/14.jpg)
![Page 15: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/15.jpg)
![Page 16: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/16.jpg)
![Page 17: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/17.jpg)
![Page 18: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/18.jpg)
Other Air Treatments
• Ionizer- negative ion generation
• Ozone- good disinfectant, safety risk
• UV- good disinfectant, 253.7nm, long exposure times?
![Page 19: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/19.jpg)
![Page 20: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/20.jpg)
![Page 21: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/21.jpg)
Cundith et al.Console filtration
![Page 22: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/22.jpg)
Cundith et al., Meat Processing Plant
• Deployed EP filters with scanning UV
• Three Rooms (100% recirculation)– Processing- 10ºC, 110m3, 3 m/min– Chill Cooler- 0-2ºC, 130m3, 21 m/min– Aging Cooler- 0-2ºC, 360m3, 8m/min
![Page 23: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/23.jpg)
![Page 24: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/24.jpg)
![Page 25: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/25.jpg)
![Page 26: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/26.jpg)
![Page 27: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/27.jpg)
Kujundzic E, et al..•
Effects of ceiling-mounted HEPA-UV air filters on airborne bacteria concentrations in an indoor therapy pool building. J Air Waste Manag Assoc. 2005 Feb;55(2):210-8.
–
The purpose of this study was to assess the effectiveness of a new generation of high-volume, ceiling-mounted high-efficiency particulate air (HEPA)-ultraviolet (UV) air filters (HUVAFs) for their ability to remove or inactivate bacterial aerosol. In an environmentally controlled full-scale laboratory chamber (87 m3), and an indoor therapy pool building, the mitigation ability of air filters was assessed by comparing concentrations of total bacteria, culturable bacteria, and airborne endotoxin with and without the air filters operating under otherwise similar conditions. Controlled chamber tests with pure cultures of aerosolized Mycobacterium parafortuitum cells showed that the HUVAF unit tested provided an equivalent air-exchange rate of 11 hr(-1). Using this equivalent air-exchange rate as a design basis, three HUVAFs were installed in an indoor therapy pool building for bioaerosol mitigation, and their effectiveness was studied over a 2-year period. The HUVAFs reduced concentrations of culturable bacteria by 69 and 80% during monitoring periods executed in respective years. The HUVAFs reduced concentrations of total bacteria by 12 and 76% during the same monitoring period, respectively. Airborne endotoxin concentrations were not affected by the HUVAF operation.
![Page 28: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/28.jpg)
Laboratory Control
• BSCs
• Centrifuge containment
• HVAC
![Page 29: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/29.jpg)
Class I
• Personnel and Environmental Protection
• No Product Protection• Similar to fume hood
with HEPA exhaust• 75 LPM• Hard-ducted to BLDG
exhaust• 0.5 inches water
![Page 30: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/30.jpg)
Class I
• Some have integral blower
• Must be turned off in case of BLDG exhaust failure
• Good design to surround all positive pressure plenums with vacuum
![Page 31: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/31.jpg)
Class IIA/B3 (Now Class II A2)
• 30/70 exhaust/recirculation• HEPA filtered on exhaust
and recirculation• 75 LPM• Hard-ducting not
recommended • Not rated for volatiles or
toxics
![Page 32: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/32.jpg)
Class IIB1• Originated with NCI designed
cabinet• 70/30 exhaust/recirculation• HEPA filtered on exhaust,
recirculation and inflow• Personnel, environment and
product protection• 100 LPM• 70% downflow through rear grill;
30% front grill• Hard-ducted; preferably
dedicated
![Page 33: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/33.jpg)
Class IIB2
• Total exhaust • Personnel, environment and
product protection• HEPA filtered on inflow and
exhaust• 100 LPM• Hard-ducted;1.5 inches
water; must be interlocked; ideally dedicated
• With internal blower may not be hard-ducted
![Page 34: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/34.jpg)
Class III
• Designed for work with BSL 4 agents• Interchange box• Heavy duty rubber gloves• HEPA filtered on supply and exhaust (double or
incinerator)• Hard-ducted; dedicated exhaust 0.5 inches water
![Page 35: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/35.jpg)
BAG-IN BAG-OUT (BIBO)
• Additional layer of protection
• Between BSC and BLDG exhaust
• HEPA• Required for BSL3/BSL4
laboratories
![Page 36: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/36.jpg)
![Page 37: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/37.jpg)
Respirators/PAPRs
![Page 38: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/38.jpg)
Fit TestingBefore an employee uses any respirator with a negative or positive pressure tight-fitting facepiece, the employee must be fit tested with the same make, model, style, and size of respirator that will be used.
![Page 39: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/39.jpg)
Qualitative Fit Test (QLFT)
A pass/fail fit test to assess the adequacy of respirator fit that relies on the individual’s response to the test agent.
![Page 40: Control of Bioaerosols](https://reader035.vdocuments.us/reader035/viewer/2022062321/56813079550346895d965485/html5/thumbnails/40.jpg)
Quantitative Fit Test (QNFT)An assessment of the adequacy of respirator fit by numerically measuring the amount of leakage into the respirator.