Perchlorate
April 14, 2004
An Integrated Approach to Ensure Reliable Perchlorate Analytical Results
![Page 2: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/2.jpg)
Agenda
Perchlorate Chemistry and Sources Current Methods for Testing Ion Chromatography LC/MS/MS An Integrated Approach A Success Story Summary and Conclusions
![Page 3: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/3.jpg)
Perchlorate Ion
Perchlorate is a compound containing one chlorine atom and four oxygen atoms.
Perchlorate is an ion, meaning it carries an electrical charge, as indicated by the negative sign.
![Page 4: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/4.jpg)
Natural Sources of Perchlorate
Chilean fertilizer deposits New Mexican potash Canadian potash Californian hanksite Bolivian playa crusts
![Page 5: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/5.jpg)
Anthropogenic Sources of Perchlorate
Rocket Fuel Fireworks High explosives Flares Herbicides Automobile airbags Tracer munitions Detergents ?
![Page 6: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/6.jpg)
How is perchlorate measured?
EPA Method 314.0Ion chromatography/conductivity detector
Enhanced (Modified) 314.0 Suppressed Conductivity Detector
SW-846 Method 9058 (draft)Ion Chromatography/conductivity detector
![Page 7: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/7.jpg)
How is perchlorate measured? Other analytical methods that can detect perchlorate:
IC “Improved” Method 314.1 approved use of 2nd column confirmation and pre-concentration of samples.
IC/MS (ion chromatography/mass spectrometry) Proposed Method 330.0
LC/MS (liquid chromatography/mass spectrometry)
LC/MS/MS (liquid chromatography/mass spectrometry/mass spectrometry) Method 8321 Modified, Proposed Method 331.0
![Page 8: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/8.jpg)
Ion Chromatography
![Page 9: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/9.jpg)
Ion Chromatography An aqueous sample, comprised of
components A and B, is injected into the ion chromatograph.
The sample enters the analytical column which is packed with a material that can adsorb the components of the sample.
Some components are more strongly adsorbed.
Diagram from Fundamentals of Analytical Chemistry, Skoog, 1988.
![Page 10: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/10.jpg)
Ion Chromatography
Fresh solvent is continuously injected into the instrument.
The solvent carries the sample components, A and B, through the column.
Component B is more strongly adsorbed onto the column and is moving thought the column at a slower rate.
The boundary between A and B is indistinct at this point.
Diagram from Fundamentals of Analytical Chemistry, Skoog, 1988.
Detector
![Page 11: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/11.jpg)
Ion Chromatography
Component B is almost completely separated from component A.
The boundary between the two components is more distinct.
Diagram from Fundamentals of Analytical Chemistry, Skoog, 1988.
Detector
![Page 12: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/12.jpg)
Ion Chromatography
Component B is completely separated from component A.
Component A exits the column through the detector at time t3.
The detector records component A as a peak (pictured below the column) in the sample chromatogram (graph).
Diagram from Fundamentals of Analytical Chemistry, Skoog, 1988.
Detector
![Page 13: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/13.jpg)
Ion Chromatography Component B exits the
column through the detector at time t4.
The detector records component B as a peak (pictured below the column) in the sample chromatogram.
Diagram from Fundamentals of Analytical Chemistry, Skoog, 1988.
Detector
Solvent
![Page 14: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/14.jpg)
Diagram from Fundamentals of Analytical Chemistry, Skoog, 1988.
![Page 15: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/15.jpg)
Identification of Perchlorate by Ion Chromatography
11.2.4 The width of the retention time window used to make identifications should be based upon measurements of actual retention time variations of standards measured over several days. Three times the standard deviation of retention time may be used as a suggested window size but the retention time window should not extend beyond ± 5% of the retention time for perchlorate. The experience of the analyst should weigh heavily in the interpretation of these chromatograms.
EPA Method 314.0
![Page 16: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/16.jpg)
IC Perchlorate Standard
![Page 17: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/17.jpg)
High Conductivity
![Page 18: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/18.jpg)
Carry Over
![Page 19: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/19.jpg)
Matrix Spike Confirmation
![Page 20: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/20.jpg)
Matrix Spike Non- Confirmation
![Page 21: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/21.jpg)
Matrix Spike Non- Confirmation
![Page 22: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/22.jpg)
Liquid Chromatography/MassSpectrometry/Mass
Spectrometry
![Page 23: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/23.jpg)
LC/MS/MS
Leading edge technology
DoD Recommended
Pending Regulatory Approval
![Page 24: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/24.jpg)
![Page 25: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/25.jpg)
API MS/MS
![Page 26: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/26.jpg)
Bowling Balls and Feathers: a simplification of mass spec separations
![Page 27: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/27.jpg)
Identification By LC/MS/MS Liquid chromatography separates perchlorate
from other sample components in the same manner that ion chromatography does
Mass spectrometers (MS) are used as detectors in place of the conductivity detector used in IC.
Unlike conductivity detectors, an MS is a specific detector
![Page 28: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/28.jpg)
Detection By LC/MS/MS The MS counts the number of ions that have a specific molecular weight.
Perchlorate has a characteristic molecular weight (MW). The first MS counts the number of ions that have the MW of perchlorate:
1 Chlorine (35) + 4 Oxygens (16) = 99 The second MS removes one of the oxygen atoms and then counts the
number of “daughter” ions that have that MW: 1 Chlorine (35) + 3 Oxygens (16) = 83
Chlorine has a less abundant isotope (MW 37), so ions with MW 101 are also counted by the first MS and ions with MW 85 are counted by the second MS.
The retention time of the peak for MW 99 must be the same as the retention time of the peak for MW 83 for perchlorate to be confirmed. If the retention times are different, then the two molecular weight were detected from two different components and perchlorate is not confirmed.
![Page 29: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/29.jpg)
Non-detect Analysis by LC/MS/MS Analysis by LC/MS/MS Reporting limit is 2 g/L.
Bottom chromatogram shows the counts of ions with molecular weight 83 (1 chlorine + 3 oxygens).
Top chromatogram shows the counts of ions with molecular weight (the isotope of chlorine).
Perchlorate is not detected.
![Page 30: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/30.jpg)
Comparison of LC/MS/MSNon-Detect Detect
![Page 31: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/31.jpg)
Reporting limit 2 g/L.
Bottom left shows MW 83 (1 chlorine + 3 oxygens).
Top left shows MW 85 (the isotope of chlorine).
Right chromatogram show the 18O internal standard (recovery 88%).
Perchlorate is detected.
Detect Analysis by LC/MS/MS
![Page 32: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/32.jpg)
Perchlorate Decision Tree
![Page 33: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/33.jpg)
Success Using the Perchlorate Decision Tree
Approximately 1000 samples were collected and analyzed for perchlorate from August 2003 through January 2004.
89 samples (about 9% of total) were pretreated and reanalyzed.
39 samples (about 3% of total) were reanalyzed by LC/MS/MS - 24 detects and 15 nondetects.
10 perchlorate detects (about 0.4% of total) were confirmed by LC/MS/MS – 58% false positive rate and a 0% false negative rate.
![Page 34: Click master to edit styles](https://reader034.vdocuments.us/reader034/viewer/2022051404/586bacac1a28ab3f0a8ba448/html5/thumbnails/34.jpg)
Summary Why use 314.0?
Cost CapacityAvailabilityPromulgated Methods
When to use LC/MS/MS?Definitive DataCritical SamplesSmall Projects