1.1 Range of molar absorptivity
Electronic excitation of outer valence (i.e. bonding) electron
How probable for this electronic excitation? (allowed transition, or forbidden)
judged by the range of = 104 -105 L mol-1 cm-1, strong absorption < 103 L mol-1 cm-1, low intensity
*MhM excitation
![Page 2: Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species](https://reader035.vdocuments.us/reader035/viewer/2022062502/56814c1c550346895db91d13/html5/thumbnails/2.jpg)
1.2 Which electron get excited?1.2.1 Organic molecules
, (bonding) and n (non-bonding) orbitals*, * (anti-bonding) orbitals
* E large ( < 150 nm, out of range) = 10 -10,000 Lmol-1cm-1
n * E smaller ( = 150 - 250 nm) = 200-2000 Lmol-1cm-1
* n * E smallest ( = 200 - 700 nm) = 10-10,000 Lmol-1cm-1
Ideal for UV-Vis spectrometry of organic chromophore
![Page 3: Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species](https://reader035.vdocuments.us/reader035/viewer/2022062502/56814c1c550346895db91d13/html5/thumbnails/3.jpg)
![Page 4: Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species](https://reader035.vdocuments.us/reader035/viewer/2022062502/56814c1c550346895db91d13/html5/thumbnails/4.jpg)
1.2.2 Inorganic moleculesMost transition metal ions are colored (absorption in Vis) due to d d electronic transition
Fig. 14-4 (p.371)Fig. 14-3 (p.370)
![Page 5: Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species](https://reader035.vdocuments.us/reader035/viewer/2022062502/56814c1c550346895db91d13/html5/thumbnails/5.jpg)
1.2.3 Charge-transfer absorption
A: electron donor, metal ionsD: electron acceptor, ligand > 10,000
DAhAD excitation
Fig. 14-5 (p.371)
![Page 6: Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species](https://reader035.vdocuments.us/reader035/viewer/2022062502/56814c1c550346895db91d13/html5/thumbnails/6.jpg)
2.1Approximation of T and A
A = -log T = log (P0/P) = ··b·c: molar absorptivity at one particular
wavelength (L·mol-1cm-1)b: path length of absorption (cm)c: molar concentration (mol·L-1)
Fig. 6-25 (p.158)
Fig. 13-1 (p.337)
Light loss due to reflection (17.3%), scattering, …
![Page 7: Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species](https://reader035.vdocuments.us/reader035/viewer/2022062502/56814c1c550346895db91d13/html5/thumbnails/7.jpg)
PP
PPA
PP
PPT
solution
solvent
solvent
solution
0
0
loglog
![Page 8: Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species](https://reader035.vdocuments.us/reader035/viewer/2022062502/56814c1c550346895db91d13/html5/thumbnails/8.jpg)
2.2 Application of Beer’s law to mixtures
Absorbance is additive
Atotal = A1 + A2 + … = 1bc1 + 2bc2
For a 2-component mixture, we measure the absorption at two different wavelength, respectively
A1 = 1,1·b·c1 + 2,1·b·c2
A2 = 1,2·b·c1 + 2,2·b·c2
![Page 9: Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species](https://reader035.vdocuments.us/reader035/viewer/2022062502/56814c1c550346895db91d13/html5/thumbnails/9.jpg)
2.3 Limitations of Beer’s law
2.3.1 Real deviations
At low concentration
A = -log T = log (P0/P) = ··b·c
At c > 0.01 M solute-solute interaction, hydrogen-bond, …can alter the electronic absorption at a given wavelength
dilute the solution
![Page 10: Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species](https://reader035.vdocuments.us/reader035/viewer/2022062502/56814c1c550346895db91d13/html5/thumbnails/10.jpg)
2.3.2 Chemical effects analyte associates, dissociates or reacts with a solvent to
give molecule with different
Example: acid-base equilibrium of an indicator
430 570 HIn 6.30 x102 7.12x103 (measured in HCl solution)In- 2.06 x 1049.61 x102 (measured in NaOH solution)
What’s the absorbance of unbuffered solution at c = 2 x 10-5M?
073.0][][
236.0][][
1088.0][
1012.1][
570,570,570
430,430,430
5
5
HInbInbA
HInbInbA
MHIn
MIn
HInIn
HInIn
5 a 102.1K InHHIn
][][
][][
1042.1][
]][[ 5
IncHIn
InHHInInHKa
![Page 11: Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species](https://reader035.vdocuments.us/reader035/viewer/2022062502/56814c1c550346895db91d13/html5/thumbnails/11.jpg)
Fig. 13-3 (p.340)=[HIn] + [In-]
InHHIn
![Page 12: Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species](https://reader035.vdocuments.us/reader035/viewer/2022062502/56814c1c550346895db91d13/html5/thumbnails/12.jpg)
2.3.3 Instrumental deviations due to polychromatic radiation
Beer’s law applies for monochromatic absorption only.
If a band of radiation consisting of two wavelength ’, and ”Assuming Beer’s law applies to each wavelength
bc"bcm
bc"
bc
PPPP
PPPP
A
PP"
PP
bcPP
A
"0
''0
"0
'0
"0
'0
"0
''0
'0
1010log
"'log
absorbance Measured
10"h wavelengtsecond For the
10'
''
log'
length first waveFor
Fig. 7-11 (p.176)
![Page 13: Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species](https://reader035.vdocuments.us/reader035/viewer/2022062502/56814c1c550346895db91d13/html5/thumbnails/13.jpg)
2.3.3 Instrumental deviations due to polychromatic radiation
bc"bcm PPPP
PPPPA "
0'
0
"0
'0
"0
'0
1010log
"'log
absorbance Measured
Fig. 13-4 (p.341)
′″
′″
′″ Non-linear calibration curve
![Page 14: Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species](https://reader035.vdocuments.us/reader035/viewer/2022062502/56814c1c550346895db91d13/html5/thumbnails/14.jpg)
How to avoid :
Select a wavelength band near its maximum absorption where the absorptivity changes little with wavelength
Fig. 13-5 (p.341)
![Page 15: Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species](https://reader035.vdocuments.us/reader035/viewer/2022062502/56814c1c550346895db91d13/html5/thumbnails/15.jpg)
2.3.4 Other physical effects
stray light – the scattering, reflection radiation from the instrument, outside the nominal wavelength band chosen for measurement
mismatched cell for the sample and the blank
![Page 16: Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species](https://reader035.vdocuments.us/reader035/viewer/2022062502/56814c1c550346895db91d13/html5/thumbnails/16.jpg)
3.1 Standard deviation of c
TTs
css
sTcbTT
c
Tb
c
Tc
TT
cc
Tc
log434.0
)(
434.0
log1
2
2
222
![Page 17: Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species](https://reader035.vdocuments.us/reader035/viewer/2022062502/56814c1c550346895db91d13/html5/thumbnails/17.jpg)
3.2 Sources of instrumental noise Case I Limited readout resolution (31/2-digit displays 0.1% uncertainty from 0%T -100% T) Thermal noise in thermal detector, etc (particularly for IR and neat IR spectrophotometer)
Case IIShot noise in photon detector (random emission of photon from the light source orrandom emission of electrons from the cathode in a detector)
Case III Flicker noise,
Fail to position sample and blank cells reproducibly in replicate measurements (as a result, different sections of cell window are exposed to radiation, and reflection and scattering losses change)
1ksT
TTksT 22
TksT 3
![Page 18: Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species](https://reader035.vdocuments.us/reader035/viewer/2022062502/56814c1c550346895db91d13/html5/thumbnails/18.jpg)
3.2 Sources of instrumental noise
Fig. 13-3 (p.344)
![Page 19: Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species](https://reader035.vdocuments.us/reader035/viewer/2022062502/56814c1c550346895db91d13/html5/thumbnails/19.jpg)
4.1 Designsa. Single beam
b. Double-beam-in-space
c. Double-beam-in-time
Advantage of double beam configuration
• Compensate for fluctuation in the radiant output, drift in transducer, etc.
• Continuous recording of spectra
Fig. 13-13 (p.352)
![Page 20: Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species](https://reader035.vdocuments.us/reader035/viewer/2022062502/56814c1c550346895db91d13/html5/thumbnails/20.jpg)
Shimadzu UV-2450 Spectrophotometer
![Page 21: Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species](https://reader035.vdocuments.us/reader035/viewer/2022062502/56814c1c550346895db91d13/html5/thumbnails/21.jpg)
Wavelength Range
190 to 900nm (performance guaranteed range). Extendable to 1,100nm through the use of the optional photomultiplier. (The measurable range maybe restricted in the shorter wavelength side depending on the type of photomultipler used.)
Monochromator System
UV-2450: Single monochromator with a high-performance blazed holographic grating in the aberration corrected Czerny-Turner mounting.
Resolution 0.1nmSpectral Bandwidth 0.1, 0.2, 0.5, 1, 2 and 5nm
Wavelength Accuracy ±0.3nm
Wavelength Repeatability ±0.01nm
Wavelength Scanning Speed FAST, MEDIUM, SLOW, and SUPER SLOW
Light Source 50W halogen lamp (2,000 hours of life) and D2lamp (500 hours of life)Light Source lamp switching Selectable between 282nm and 393nm
Stray Light UV-2450: Less than 0.015% at 220nm and 340nmDetector Photomultiplier R-928Photometric System Double beam, direct ratio system with dynode feedback
Photometric Mode Absorbance (Abs.), transmittance (%), reflectance (%) and energy (E).
Photometric Range
Abosrbance: -4~5 Abs. (0.001 Abs. increments)Transmittance: 0~999.9% (0.01 increments)Reflectance: 0~999.9% (0.01% increments)
Photometric Accuracy
±0.002Abs(0~0.5Abs), ±0.004Abs(0.5~1Abs),±0.3T (0~100%T) (all determined with NIST 930D standard filter)
Photometric Repeatability 0.001Abs (0~0.5Abs), ±0.1%T
Baseline Correction Selectable with storage in firmware
Baseline Flatness Within ±0.001Abs. (excluding noise, 2nm slit width and SLOW wavelength scanning speed)
Drift Less than 0.0004 Abs. per hour (after 2 hours warm-up)Dimensions 570 (W) x 660 (D) x 275 (H) (mm)Weight 36kgPower Supply AC 100V/120V/220V/240V, 50/60Hz 250VA (swithc-selectable)
![Page 22: Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species](https://reader035.vdocuments.us/reader035/viewer/2022062502/56814c1c550346895db91d13/html5/thumbnails/22.jpg)
Shimadzu UV-2450 Spectrophotometer
![Page 23: Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species](https://reader035.vdocuments.us/reader035/viewer/2022062502/56814c1c550346895db91d13/html5/thumbnails/23.jpg)
d. Effects of monchromator exit slit width on spectraNarrow exit slit width improves the spectrum resolutionbut it also significantly reduce the radiant power
Trade-off between resolution and S/N ratio
![Page 24: Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species](https://reader035.vdocuments.us/reader035/viewer/2022062502/56814c1c550346895db91d13/html5/thumbnails/24.jpg)
e. Multichannel spectrometerNo monochromator, but disperses transmitted light and measures “all wavelength at once”
No Scanning-simple and fastMore expensiveLimited resolution
Fig. 13-15 (p.353)
![Page 25: Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species](https://reader035.vdocuments.us/reader035/viewer/2022062502/56814c1c550346895db91d13/html5/thumbnails/25.jpg)
5.1Quanlitative spectra
Solvent effects on the UV-Vis spectra
Polar solvents “blur” vibrational features
Polar solvents shift absorption maximan * blue shift * red shift -
UV-Vis not reliable for qualitative but excellent for quantitative analysis
![Page 26: Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species](https://reader035.vdocuments.us/reader035/viewer/2022062502/56814c1c550346895db91d13/html5/thumbnails/26.jpg)
5.2 Quantitative analysis- Determining the relationship between A and c
External StandardsStandard-Addition
![Page 27: Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species](https://reader035.vdocuments.us/reader035/viewer/2022062502/56814c1c550346895db91d13/html5/thumbnails/27.jpg)
Fig. 14-14 (p.382)
5.3 Spectrophotometric kinetics
![Page 28: Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species](https://reader035.vdocuments.us/reader035/viewer/2022062502/56814c1c550346895db91d13/html5/thumbnails/28.jpg)
Fig. 14-16 (p.384)