Calibration of Concave Calibration of Concave 106106Ru Applicators at NISTRu Applicators at NIST
Christopher G. SoaresIonizing Radiation Division
National Institute of Standards and Technology
Physics LaboratoryPhysics Laboratory
UNITED STATES DEPARTMENT OF COMMERCE
National Institute of Standards and TechnologyGaithersburg, MD 20899-0001
UNITED STATES DEPARTMENT OF COMMERCE
National Institute of Standards and TechnologyGaithersburg, MD 20899-0001
![Page 2: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/2.jpg)
HistoryHistory
1937 Fialla builds first extrapolation chamber
1947 Beta-ray applicators in use
1952 Amersham makes first metal foil sources
1953 Loevinger benchmark paper on extrapolation chambers
1976 Establishment of NBS calibration service by Pruitt & Loevinger
1988 Publication of serious discrepancy between NBS and Amersham
1990 Reestablishment of revised NIST calibration service by Soares
1996 Establishment of the UW calibration service by DeWerd
1997 ICRU Report Committee for medical betas formed
1998 First international dosimetry intercomparison for applicators
200? Establishment of national standards at NPL and PTB
![Page 3: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/3.jpg)
Extrapolation Chamber Measurement GeometriesExtrapolation Chamber Measurement Geometries
30 mm diametercollecting electrode
10 mm diametersource
300 mm longair path
1 mm sourcein TE plastic block
1 mm diameter collecting electrodein contact with block surface
Protection Level
10 mm diametersource
4 mm diameter collecting electrode in contactor at 1 mm in tissue-equivalent plastic
Ocular Therapy
Brachytherapy
![Page 4: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/4.jpg)
Extrapolation Chamber MeasurementsExtrapolation Chamber Measurementsof Beta-Particle Sourcesof Beta-Particle Sources
Source Geometry Application 1 sigma
near point 10s cm in air protection +2%
planar mm in tissue ocular therapy +6%
seed/line mm in tissue brachytherapy +10%
![Page 5: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/5.jpg)
Water-equivalent plasticHigh-voltageelectrode/window
Ionization
Collectingelectrode
Insulatinggap
Air gap=0.40 mm
Extrapolation Chamber SchematicExtrapolation Chamber Schematic
Electrometer
79.54 pA
Air gap, mm
![Page 6: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/6.jpg)
Water-equivalent plasticHigh-voltageelectrode/window
Collectingelectrode
Insulatinggap
Air gap=0.35 mm
Extrapolation Chamber SchematicExtrapolation Chamber Schematic
Electrometer
69.73 pA
Ionization
Air gap, mm
![Page 7: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/7.jpg)
Water-equivalent plasticHigh-voltageelectrode/window
Collectingelectrode
Insulatinggap
Air gap=0.30 mm
Extrapolation Chamber SchematicExtrapolation Chamber Schematic
Electrometer
59.82 pA
Ionization
Air gap, mm
![Page 8: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/8.jpg)
Water-equivalent plasticHigh-voltageelectrode/window
Collectingelectrode
Insulatinggap
Air gap=0.25 mm
Extrapolation Chamber SchematicExtrapolation Chamber Schematic
Electrometer
49.85 pA
Ionization
Air gap, mm
![Page 9: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/9.jpg)
Water-equivalent plasticHigh-voltageelectrode/window
Collectingelectrode
Insulatinggap
Air gap=0.20 mm
Extrapolation Chamber SchematicExtrapolation Chamber Schematic
Electrometer
39.89 pA
Ionization
Air gap, mm
![Page 10: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/10.jpg)
Water-equivalent plasticHigh-voltageelectrode/window
Collectingelectrode
Insulatinggap
Air gap=0.15 mm
Extrapolation Chamber SchematicExtrapolation Chamber Schematic
Electrometer
29.92 pA
Ionization
Air gap, mm
![Page 11: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/11.jpg)
Water-equivalent plasticHigh-voltageelectrode/window
Collectingelectrode
Insulatinggap
Air gap=0.10 mm
Extrapolation Chamber SchematicExtrapolation Chamber Schematic
Electrometer
19.95 pA
Ionization
Air gap, mm
![Page 12: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/12.jpg)
Water-equivalent plasticHigh-voltageelectrode/window
Collectingelectrode
Insulatinggap
Air gap=0.05 mm
Extrapolation Chamber SchematicExtrapolation Chamber Schematic
Electrometer
9.99 pA
Ionization
Air gap, mm
![Page 13: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/13.jpg)
NIST Medical Extrapolation ChamberNIST Medical Extrapolation ChamberCollecting ElectrodesCollecting Electrodes
4 mm0.6 mm
1.2 mm
Aeff=14.68 mm2 Aeff=0.648 mm2
![Page 14: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/14.jpg)
![Page 15: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/15.jpg)
[ ]0 D(z0) = . (W/e) Sair
medium
0 A k’
dd
k I()
Extrapolation Chamber Dose EquationExtrapolation Chamber Dose Equation
D(z0) absorbed dose rate in medium at depth z0 (W/e) average energy to create an ion pair (33.97 J/C)
Sair stopping power of medium relative to air (1.12)
0 air density at reference conditions (1.197 kg/m3)
A area of the collecting electrode (14.68 mm2)
I () net current at air gap [ ]0 slope of current vs air gap function at zero air gap
k’ product of corrections which are independent of air gap
k product of corrections which vary with air gap
medium
.
![Page 16: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/16.jpg)
Determination of Net CurrentDetermination of Net Current
C the external feedback capacitance
U1,2 initial and final voltages on the feedback capacitor
t integration time
I+,- = C(U2-U1)/t
I = (I+ + |I-|)/2
![Page 17: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/17.jpg)
Corrections to Measured CurrentCorrections to Measured Currentfor Near Geometry Measurementsfor Near Geometry Measurements
Constant during the extrapolation curve measurement:
k’ba difference in electron backscatter between collector and tissue
Varying during the extrapolation curve measurement:
kTp correction to reference conditions of temperature and pressure
kre correction for recombination losses
kdi correction for radiation field divergence
![Page 18: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/18.jpg)
Near Geometry Divergence EffectNear Geometry Divergence Effect
Less side losses with decreasing air gap
![Page 19: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/19.jpg)
Calibration Chain at NISTCalibration Chain at NIST
Extrapolation Extrapolation Chamber withChamber with
4 mm Electrode4 mm Electrode
Extrapolation Extrapolation Chamber withChamber with
1 mm Electrode1 mm Electrode
RadiochromicRadiochromicFilmFilm
Source to beSource to beCalibratedCalibrated
Planar Planar 9090Sr/YSr/YReferenceReference
SourceSource
ReferenceReferenceLine SourceLine Source
Well-TypeWell-TypeIonizationIonizationChamberChamber
![Page 20: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/20.jpg)
ICRU Report Committee:ICRU Report Committee:
Dosimetry of Beta Rays andDosimetry of Beta Rays andLow Energy Photons forLow Energy Photons for
Brachytherapy with Sealed SourcesBrachytherapy with Sealed Sources
Committee Charter:Collection of existing data
New intercomparison of dosimetryof ophthalmic applicators
Commission Sponsors:R. CaswellA. Wambersie
Committee:W.G. CrossH. Järvinen (Chairman)C.G. SoaresS. VynckierK. Weaver
Consultant:D. Flühs
![Page 21: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/21.jpg)
National Institute of Standards & TechnologyUSA Chris Soares
National Physical LaboratoryUK Tudor Williams
Radiation and Nuclear Safety AuthorityFinland Hannu Järvinen
Catholic University of Louvain, St.-Luc HospitalBelgium Stefaan Vynckier
Algemeen Ziekenhuis MiddelheimBelgium Bob Schaeken
University Hospital of EssenGermany Dirk Flühs
Participants in the Measurement ComparisonParticipants in the Measurement Comparison
![Page 22: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/22.jpg)
Field ParameterizationField Parameterization
D(z,r) = D(z0,r0) [D(z,r0)/D(z0,r0)] {D(z,r)/D(z,r0)}
D(z0,r0) = reference absorbed dose z0 = 1 mm r0 = 0 mm
[D(z,r0)/D(z0,r0)] = relative central-axis depth dose
{D(z,r)/D(z,r0)} = relative off-axis dose
![Page 23: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/23.jpg)
Effective Meas. Phys.
Detector Inst. thickness Covering Diam. Diam. ApplicationExtrap. Chamb. NIST 0 0.0016 PET 0.91 500 1,2
4.32Extrap. Chamb. NPL 0 0.0016 PET 4 140 1GAF films NIST 0.007 0 >0.1 >1 1,2,3
0.0017 0GAF film STUK 0.007 0 3 >1 1TLDs UCL 0.3 0 5 5 1,2
1.0 0Alanine AZM 1.2 0 4.9 4.9 1,2Scintillator NIST 0.4 0.2 PE 1 6 1,2,3Scintillator Essen 1 0.02 Al 1 2.55 2,3Silicon diode STUK 0.05 0.375 PMMA 4 7 2Diamond STUK 0.3 0.65 PS 0.5 7.1 2Ion chamb. STUK 0 0.03 PET + 3 61 2 0.2 water . All dimensions in mm Applications: 1= Reference dose rate; 2=Relative central axis depth dose; 3=Relative off axis dose
Detectors Used for the MeasurementsDetectors Used for the Measurements
![Page 24: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/24.jpg)
Eye PhantomsEye Phantoms
![Page 25: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/25.jpg)
Planar Film Irradiation GeometryPlanar Film Irradiation Geometry
![Page 26: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/26.jpg)
Source GeometriesSource Geometries
90Sr planar 106Ru planar 106Ru concave
1 cm
![Page 27: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/27.jpg)
Tracerlab & ICN/Tracerlab New England Nuclear
Atlantic Research Corp/Atomchem
Manning Research
Technical Operations
3M
Isotope Products Lab/Nucl. Assocs.
Amersham International
Typical Source ProfilesTypical Source Profiles
![Page 28: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/28.jpg)
Peak Dose RatePeak Dose Rate 230 mGy/s230 mGy/s
Calibrated AverageCalibrated AverageDose Rate 170 mGy/sDose Rate 170 mGy/s
![Page 29: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/29.jpg)
9090Sr/Y Planar Source Field Profile at 1 mmSr/Y Planar Source Field Profile at 1 mm
![Page 30: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/30.jpg)
106106Ru/Rh Concave Source Field Profile at 5 mmRu/Rh Concave Source Field Profile at 5 mm
![Page 31: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/31.jpg)
Results of Reference Dose Rate DeterminationsResults of Reference Dose Rate Determinations
Extrap. Extrap. GAF GAF 0.3mm 0.3mm 0.4mm 1mm 1.2mm Chamb. Chamb. Film Film TLD diamond scin TLD alanineSource (NIST) (NPL) (NIST) (STUK) (UCL) (STUK) (NIST) (UCL) (AZM)
90Sr 0.237 0.260* 0.229 0.258 0.248 --- 0.278 0.308 0.272NEN Gy/s Gy/s Gy/s Gy/s Gy/s Gy/s Gy/s Gy/s0258
106Ru 1.61** 1.82** 1.62 1.63** 1.40 1.99 1.69 1.75 1.49BEBIG mGy/s mGy/s mGy/s mGy/s mGy/s mGy/s mGy/s mGy/s mGy/splanar
106Ru --- --- 2.48 2.10*** 2.55 3.17 --- 2.79 2.89BEBIG mGy/s mGy/s mGy/s mGy/s mGy/s mGy/s mGy/s mGy/s mGy/sconcave
*from a contact measurement with a factor of 0.573 applied to correct to 1 mm** from a contact measurement with a factor of 0.742 applied to correct to 1mm***from a measurement at 1.2 mm with a factor of 1.082 applied to correct to “
![Page 32: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/32.jpg)
0.00001
0.0001
0.001
0.01
0.1
1
10
0 200 400 600 800 1000 1200
Depth in material /mg cm-2
Re
lati
ve
do
se
ra
te
NIST RCF
0.3 mm TLD
1 mm TLD
Alanine
STUK diode
NIST EC
STUK IC
NIST scin
Essen scin
ACCEPTCode
Depth Dose Measurements of a Depth Dose Measurements of a 9090Sr Ophthalmic ApplicatorSr Ophthalmic Applicator
Gradient at 1 mm:6.5% per 0.1 mm
![Page 33: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/33.jpg)
0
0.25
0.5
0.75
1
1.25
1.5
1.75
2
0 200 400 600 800 1000 1200
Depth in material /mg cm-2
Rel
ativ
e d
ose
rat
e
NIST RCF
0.3 mm TLD
1 mm TLD
Alanine
STUK diode
NIST EC
STUK IC
NIST scin
Essen scin
ACCEPTCode
Depth Dose Measurements of a Depth Dose Measurements of a 9090Sr Ophthalmic ApplicatorSr Ophthalmic Applicator
Gradient at 1 mm:6.5% per 0.1 mm
![Page 34: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/34.jpg)
0.0001
0.001
0.01
0.1
1
10
0 400 800 1200 1600
Depth in material / mg cm-2
Re
lati
ve
do
se
ra
te
NIST RCF
1 mm TLD
STUKdiode
NIST scin
Essenscin
STUKdiamond
0.3 mmTLD
Alanine
ACCEPTCode
Depth Dose Measurements of a PlanarDepth Dose Measurements of a Planar106106Ru Ophthalmic ApplicatorRu Ophthalmic Applicator
Gradient at 1 mm:3% per 0.1 mm
![Page 35: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/35.jpg)
0
0.25
0.5
0.75
1
1.25
1.5
0 400 800 1200 1600
Depth in material / mg cm-2
Re
lati
ve
do
se
ra
te
NIST RCF
1 mm TLD
STUKdiode
NIST scin
Essenscin
STUKdiamond
0.3 mmTLD
Alanine
ACCEPTCode
Depth Dose Measurements of a PlanarDepth Dose Measurements of a Planar106106Ru Ophthalmic ApplicatorRu Ophthalmic Applicator
Gradient at 1 mm:3% per 0.1 mm
![Page 36: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/36.jpg)
0.0001
0.001
0.01
0.1
1
10
0 400 800 1200 1600
Depth in material / mg cm-2
Re
lati
ve
do
se
ra
te
NIST RCF
0.3 mmTLD
1 mm TLD
Alanine
STUKdiamond
STUKdiode
Essenscin
ACCEPTCode
Depth Dose Measurements of a Curved Depth Dose Measurements of a Curved 106106Ru Ophthalmic ApplicatorRu Ophthalmic Applicator
Gradient at 1 mm:2% per 0.1 mm
![Page 37: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/37.jpg)
0
0.25
0.5
0.75
1
1.25
0 400 800 1200 1600
Depth in material / mg cm-2
Re
lati
ve
do
se
ra
te
NIST RCF
0.3 mmTLD
1 mm TLD
Alanine
STUKdiamond
STUKdiode
Essenscin
ACCEPTCode
Depth Dose Measurements of a Curved Depth Dose Measurements of a Curved 106106Ru Ophthalmic ApplicatorRu Ophthalmic Applicator
Gradient at 1 mm:2% per 0.1 mm
![Page 38: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/38.jpg)
Spread in Measurement ResultsSpread in Measurement Results
Source Contact 1 mm 2 mm 3 mm 4 mm 5 mm 7 mm 10 mm 90Sr 6.2% 9.6% 5.3% 3.4% 4.8% 8.8% --- ---NEN (10) (8) (10) (10) (9) (9)0258
106Ru 5.5% 10.5% 4.0% 6.1% 6.4% 6.8% 10% 19%BEBIG (8) (9) (8) (8) (8) (8) (7) (7)planar
106Ru 8.2% 14% 4.8% 6.9% 7.3% 7.0% 9.4% 18% BEBIG (8) (6) (8) (8) (8) (8) (8) (8)concave
All percentages are single standard deviations. Yellow values are for reference absorbed dose rate; white values are for relative central-axis dose.
![Page 39: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/39.jpg)
Effective Point of MeasurementEffective Point of Measurement
t
z z+t
D(z)
The depth of an infinitely thin detector that gives thesame dose rate as that averaged over a detector of finite thickness, t
Davg(t,z) = D(z) dz
t
z+t
z
If the gradient is constant across the detector, thenthe effective point of measurement is in the center.Otherwise, calculate
and find the value of z which gives the value
Davg(t,z) in the function D(z).
![Page 40: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/40.jpg)
Measured Off-Axis Dose Function forMeasured Off-Axis Dose Function forthe the 9090Sr/Y Planar sourceSr/Y Planar source
Essen scintillator NIST radiochromic film NIST EGS4 calculation
![Page 41: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/41.jpg)
Essen scintillator NIST radiochromic film NIST EGS4 calculation
Measured Off-Axis Dose Function forMeasured Off-Axis Dose Function forthe the 106106Ru/Rh Planar sourceRu/Rh Planar source
![Page 42: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/42.jpg)
Essen scintillator NIST radiochromic film NIST EGS4 calculation
Measured Off-Axis Dose Function for the Measured Off-Axis Dose Function for the 106106Ru/Rh Concave sourceRu/Rh Concave source
![Page 43: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/43.jpg)
Applicator Types StudiedApplicator Types Studied
CCB
CCX
CIB
![Page 44: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/44.jpg)
CCB Applicator Dose Profile in Plane 2 mm from CenterCCB Applicator Dose Profile in Plane 2 mm from Center
![Page 45: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/45.jpg)
CIB Applicator Dose Profile in Plane 4 mm from CenterCIB Applicator Dose Profile in Plane 4 mm from Center
![Page 46: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/46.jpg)
results
![Page 47: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/47.jpg)
Distance off axis, mm
Depth, mm 0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5 9 9.5 10 10.5 11
0 1.29
0.5 1.13 1.15 1.17 1.19 1.21 1.23 1.25
1 0.99 0.99 1.00 1.00 1.02 1.06 1.10 1.15 1.19 1.19
1.5 0.87 0.87 0.88 0.90 0.92 0.94 0.97 1.01 1.05 1.12 1.20 1.31
2 0.77 0.78 0.80 0.83 0.86 0.90 0.94 0.98 1.03 1.09 1.14 1.19 1.25 1.31
2.5 0.69 0.70 0.70 0.71 0.71 0.72 0.73 0.75 0.77 0.79 0.82 0.86 0.91 0.98 1.06
3 0.53 0.54 0.56 0.58 0.60 0.62 0.65 0.69 0.73 0.77 0.82 0.88 0.94 1.00 1.06 1.12
4 0.42 0.43 0.43 0.44 0.45 0.47 0.48 0.50 0.53 0.56 0.59 0.63 0.68 0.73 0.79 0.85 0.91 0.97
5 0.30 0.31 0.32 0.33 0.34 0.35 0.37 0.38 0.40 0.43 0.45 0.48 0.51 0.55 0.59 0.64 0.69 0.75 0.82 0.89
7 0.16 0.16 0.16 0.17 0.17 0.18 0.18 0.19 0.19 0.20 0.20 0.20 0.19 0.19 0.18 0.16 0.14 0.12 0.099 0.075 0.055 0.044
10 0.031 0.031 0.032 0.033 0.034 0.035 0.036 0.037 0.038 0.039 0.039 0.038 0.037 0.036 0.034 0.032 0.029 0.025 0.022 0.018 0.014 0.010 0.005
Sample of Off Axis Dosimetry MeasurementsSample of Off Axis Dosimetry Measurementsfor CIB Source Along Non-Cutout Ordinal Radiifor CIB Source Along Non-Cutout Ordinal Radii
![Page 48: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/48.jpg)
0
0.2
0.4
0.6
0.8
1
1.2
1.4
1.6
1.8
0 1 2 3 4 5 6 7 8
1.25
1.5
1.75
2
2.25
2.5
2.75
3
0 1 2 3 4 5 6 7 8
Along cutout ordinal radius
Along non-cutout ordinal radii
Sample of Off Axis Dosimetry Measurements for CIB SourceSample of Off Axis Dosimetry Measurements for CIB Source
![Page 49: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/49.jpg)
0
5
10
15
20
25
30
35
40
45
0 2 4 6 8 10 12
Depth in solid water, mm
Sp
eci
fic d
ose
ra
te, m
Gy/
s/M
Bq
/mm
2
CCB 784
CCX 41
CIB 208
ICRU
An Attempt to Make a Depth Dose Curve Independent of Source TypeAn Attempt to Make a Depth Dose Curve Independent of Source Type
plot dose rate per unit activity density
![Page 50: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/50.jpg)
A Curved-electrode Extrapolation Chamber?A Curved-electrode Extrapolation Chamber?
use the applicator as the high-voltage electrode
Electrometer
9.99 pA
![Page 51: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/51.jpg)
Source MisalignmentSource Misalignment
![Page 52: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/52.jpg)
Mechanisms to Check Source UniformityMechanisms to Check Source Uniformity
![Page 53: Calibration of Concave 106 Ru Applicators at NIST Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology Physics](https://reader036.vdocuments.us/reader036/viewer/2022062719/56649ef35503460f94c0571e/html5/thumbnails/53.jpg)