BEIR VII-Key Results
• The report supports the LNT model.
• The radiations risk values are “similar” to previous reports and other groups like UNSCEAR and ICRP.
• The uncertainties have been reduced for lower dose.
![Page 3: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/3.jpg)
Radiation Dose from Medical Imaging
“The Historical View”• The radiation dose from medical imaging
was low and a small fraction of the dose from natural sources.
• Imaging was viewed as being done only when clearly needed so it was well justified.
• The data on radiation effects was seen as uncertain at doses that would overlap the range of common medical procedures.
![Page 4: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/4.jpg)
DoseDose
Ris
kR
isk
![Page 5: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/5.jpg)
DoseDose
Ris
kR
isk
MedicalImaging
![Page 6: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/6.jpg)
DoseDose
Ris
kR
isk
MedicalImaging
![Page 7: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/7.jpg)
DoseDose
Ris
kR
isk
MedicalImaging
Recently - Doses from Imaging Have Increased
![Page 8: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/8.jpg)
DoseDose
Ris
kR
isk
MedicalImaging
BEIR VII - The Degree of Uncertainty Has Decreased
![Page 9: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/9.jpg)
BEIR VII Suggests
• We need to pay more attention to the justification of medical imaging.
• Practitioners should be well informed about The medical aspects of the procedure so that it
can be justified, The radiation risks from the procedure, and The medical physics aspects of the procedure
so it can be optimized.
![Page 10: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/10.jpg)
BEIR VII Suggests
• We need to pay more attention to ALARA to insure that properly justified procedures are properly done Optimization of the dose-image quality
tradeoff Robust quality assurance techniques to
insure high quality images
![Page 11: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/11.jpg)
Recent DevelopmentsIn Imaging
• Great increases in the number of imaging studies
• Increased use of higher dose techniques Computed Tomography Fluoroscopy
![Page 12: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/12.jpg)
-20%
0%
20%
40%
60%
80%
100%
120%
140%
Conventional Mammography CT Nuclear Imaging
Change in the Number of Imaging ProceduresChange in the Number of Imaging Procedures
1993-19991993-1999
Maitino et al Radiology 2003;227:113-117
![Page 13: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/13.jpg)
0
5
10
15
20
25
2003 2004
Rat
e p
er 1
000
sub
scri
ber
s
Radiologists
Non-Radiologists
Levin JACR 2006;3:90-95
+13%
+29%
Rate Changes 2003-2004
![Page 14: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/14.jpg)
Dose and Collective Dose
• CT accounts for 15% of procedures
• CT accounts for 75% of the total effective dose
Wiest et al Sem in US, CT & MRI 2002, 23;402-410
![Page 15: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/15.jpg)
US Exposures
Medical
15%
Consume
r
2%
Natural
83%
Other
0.28%
NCRP 93
![Page 16: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/16.jpg)
17%
34%
48%
1%
Natural
Radon
Medical
Consumer
Huda-Frey Estimate2.8 mSv
![Page 17: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/17.jpg)
Levin JACR 2006;3:90-95
0% 10% 20% 30% 40% 50% 60% 70% 80% 90%
Podiatry
Chiropractic
Family Practise
Pulmonary Medicine
General Practise
Orthopedic Surgery
Radiology
Un
ac
ce
pta
ble
Im
ag
es
Unacceptable Image Quality: Pennsylvania Blue Shield Study
How Well is Imaging Done?How Well is Imaging Done?
![Page 18: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/18.jpg)
Are Practitioners Are Practitioners Knowledgeable About Knowledgeable About Radiation Dose and Radiation Dose and
Dose Effects?Dose Effects?
![Page 19: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/19.jpg)
0%
10%
20%
30%
40%
50%
60%
CR>CT CT x1-x10
CR
CT x10-
x100 CR
CT x100-
x250 CR
CT > 500
CR
Patients
Radio logis ts
ER Phys ic ians
![Page 20: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/20.jpg)
Changes in the Medical Imaging Environment
• The uncertainties about radiation effects have decreased.
• The number of imaging procedures have increased.
• The increase has been highest in the high dose studies (CT).
• It is not obvious that all studies are well justified.
![Page 21: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/21.jpg)
How Are These Issues Addressed?
• Federal Law & Regulation
• Federal Payment Structure
• State Law & Regulation
• Professional Societies
• Business and Insurance
• Public
![Page 22: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/22.jpg)
Professional Societies
• The AAPM and ACR have undertaken a number of initiatives to address these problems Accreditation programs Task Group Reports
• CTDI Values• DR Exposure Index
![Page 23: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/23.jpg)
Professional Society –Regulatory Partnership
• Cooperation between the regulatory community and professional societies can work to reduce the effects of these many problems.
![Page 24: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/24.jpg)
Some Suggested Some Suggested ApproachesApproaches
&&Some Current ActivitiesSome Current Activities
![Page 25: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/25.jpg)
Practitioners
• Practitioners who use fluoroscopy or direct CT examinations should be able to demonstrate knowledge of: Radiation effects Radiation risks Basic physics of the modality Image optimization
![Page 26: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/26.jpg)
Practitioners
• Regulation like that used to determine status as an authorized user in Part 35 would be prudent and relevant.
![Page 27: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/27.jpg)
Medical Physicists
• Medical physicists play a key roles in: Radiation protection Image optimization Definition and supervision of quality
assurance
• Many state regulations are based on obsolete roles for diagnostic medical physicists.
![Page 28: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/28.jpg)
Medical Physicists
• Limitations in existing State regulations mean that individuals who act as medical physicists can often lack the knowledge to Properly optimize image quality Design and specify proper quality
assurance programs
![Page 29: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/29.jpg)
Medical Physicists
• Existing State regulations have created a group of “diagnostic physicists” who can only deal with prescriptive regulations
• But cannot Optimize image quality Provide effective quality assurance
programs
![Page 30: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/30.jpg)
Medical Physicists
• State Regulations should require that medical physicists should be able to demonstrate competence.
• The AAPM definition of a qualified medical physicist should form the basis of standards which would be used to define the competence of a medical physicist.
![Page 31: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/31.jpg)
Medical Physicists
• New regulations should concentrate on establishing the competence of medical physicists and then allow medical physicists to exercise their professional judgment in many areas
• Prescriptive regulations cannot keep up with the rapid changes in medical equipment so professional judgment is required
![Page 32: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/32.jpg)
Equipment Regulations
• Difficult because of rapid changes in equipment
• Also regulations have to encompass both new and old equipment
![Page 33: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/33.jpg)
Current Regulation
• Stresses testing of details of equipment performance There is frequently little or no evidence
that specific items in these regulations protect the public in an effective way
They may create a false sense of security
![Page 34: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/34.jpg)
False Sense of Security
• Equipment that meets regulations but produces poor images
• Equipment that meets regulations but is used poorly
• Reliance on CTDI values that significantly underestimate actual patient doses significantly distorts the situation
![Page 35: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/35.jpg)
Issues with Current Regulations
CT as an Example
• Dose testing is highly prescriptive but does not set standards
• Image quality does not set standards
![Page 36: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/36.jpg)
CT as an Example
• Dose Measurement in SSR Use phantom that underestimates the
dose for most most body studies Require measurements for each head,
body or whole body scan Do not set limitations for the measured
values
![Page 37: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/37.jpg)
CT Scanning
• Current SSR require image quality testing.
• However there is no minimum performance that is required.
• There is no definition of the phantom that is used.
![Page 38: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/38.jpg)
How Might Regulation Be Improved?
(CT as an Example)• Dose testing should relate to national
and international standards for reference doses.
• Testing should be done to insure an acceptable level of image quality.
• The image quality should be determined by reference to a national standard.
![Page 39: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/39.jpg)
European Union Guidelines
Examination CTDIw mGy DLP
mGy-cmRoutine Head 60 1,050
Face & Sinus 35 360
Routine Chest 30 650
High Res Chest
35 280
Routine Abdomen
35 780
European Guidelines on quality criteria for CT (EUR 16262EN)
![Page 40: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/40.jpg)
ACR CT Accreditation
Examination CTDIw mGyRoutine Head 60
Adult Abdomen
35
Pediatric Abdomen
25
![Page 41: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/41.jpg)
I Negus – Berreford Hospital, Plymouth, GD
![Page 42: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/42.jpg)
Improved RegulationImage Quality Testing
• National Standard for Performance
• National Phantom
• This is commonly done for mammography and should be adopted for CT.
![Page 43: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/43.jpg)
Image Quality Testing
• ACR accreditation testing
• National standard for performance
![Page 44: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/44.jpg)
![Page 45: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/45.jpg)
Quality Assurance• Regulations cannot keep up with the
rapid changes in imaging equipment.
• Qualified medical physicists should have some latitude in specifying quality assurance programs and image quality as long as the programs relate to a national standard.
![Page 46: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/46.jpg)
Dose Tracking Trends
• The NEXT program has long been the standard for determining radiation dose trends.
• NEXT Data has been of great value in determining trends in use and dose.
![Page 47: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/47.jpg)
NEXT Limitations
• Equipment has become more complex so it is difficult to make meaningful measurement.
• The variety of procedures changes rapidly so NEXT data tends to “lag” current practice.
![Page 48: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/48.jpg)
Electronic Monitoring
• Modern equipment has the ability to collect dose data automatically.
• The medical and regulator communities should support efforts to collect and analyze this data in an automatic fashion.
![Page 49: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/49.jpg)
Summary
• BEIR VII suggests we should be more concerned about high dose medical exposure.
• Recent data suggests that not all medical imaging is well justified or well done.
![Page 50: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/50.jpg)
Summary
• Professional Societies (AAPM, ACR, etc) are aware of these problems
• The Regulatory Community (CRCPD) is also aware of these problems
• A strong partnership between these communities can address these problems in effective ways
![Page 51: BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina](https://reader030.vdocuments.us/reader030/viewer/2022032709/56649ec15503460f94bcc73a/html5/thumbnails/51.jpg)
Summary
• Regulatory Philosophies should change: Insure that medical practitioners and
medical physicists are well qualified and stay well qualified
Regulations should become more flexible in detail
Regulations should require testing to a national standard