Download - AMTI Force Plate System
AMTI Force Plate AMTI Force Plate SystemSystem
AMTI Force Plate & AmplifierAMTI Force Plate & Amplifier
AMTI AmplifierAMTI Amplifier
Zero Reset Button. Press before each trial.
Incorrect Amplifier Settings Clips Incorrect Amplifier Settings Clips ForceForce
0
200
400
600
800
1000
1200
1400
1600
0 50 100 150 200 250 300 350
Time (ms)
Ve
rtic
al F
orc
e (
N)
0
200
400
600
800
1000
1200
1400
0 50 100 150 200 250 300 350
Time (ms)
Ve
rtic
al F
orc
e (
N)
Notice the SQUARE tops on the force curve on the right. The force values are being clipped. Reduce the Fz AMTI Amplifier Gain from 4000 to 2000, or from 2000 to 1000 to eliminate clipping of force.
Over amplification causes clipping of force signals.
Software Amplifier Settings Must Software Amplifier Settings Must Match AMTI Amplifier Gain Settings.Match AMTI Amplifier Gain Settings.
These settings (4000, 2000, 1000) must match.