User’sManual ADMAG TI Series
AXG1A Magnetic Flowmeter Remote TransmitterBRAIN Communication Type
IM 01E22C02-01EN
IM 01E22C02-01EN
3rd Edition
1
IM 01E22C02-01EN
ADMAG TI SeriesAXG1A Magnetic Flowmeter Remote TransmitterBRAIN Communication Type
IM 01E22C02-01EN 3rd Edition
Contents1. Introduction ...................................................................................................5
1.1 For Safe Use of Product ...................................................................................... 81.2 Warranty .............................................................................................................. 10
2. Operation by Display unit ..........................................................................112.1 Basic Operating Procedures ............................................................................ 112.2 DisplayandBasicConfiguration ..................................................................... 12
2.2.1 DisplayConfiguration .......................................................................... 122.2.2 Operation Level ................................................................................... 14
2.3 Display Mode and Setting Mode ....................................................................... 142.4 Parameter Setting from Display Panel ............................................................ 17
2.4.1 SettingExampleofSelectTypeParameter(Flowrateunit) ............... 172.4.2 SettingExampleofNumericTypeParameter(Flowratespan) ......... 182.4.3 SettingExampleofAlphanumericTypeParameter(TagNo.) ............ 19
2.5 microSD Card Insertion/Removal .................................................................... 20
3. OperationwithBRAINConfigurationTool ............................................. 213.1 ConnectingBRAINConfigurationTool ........................................................... 213.2 BT200 Operating Procedures ........................................................................... 23
3.2.1 KeyLayoutandScreenDisplay .......................................................... 233.2.2 KeyDescriptions ................................................................................. 243.2.3 FunctionKeys ..................................................................................... 25
3.3 Calling Up Menu Addresses ............................................................................. 263.4 Parameter Update and Upload/ Download Function ..................................... 273.5 ParameterConfiguration .................................................................................. 28
3.5.1 SettingExampleofSelecttypeParameter(FlowRateUnit) .............. 283.5.2 SettingExampleofNumericTypeParameter(Flowratespan) ......... 293.5.3 SettingExampleofAlphanumerictypeParameter(TagNo.) ............. 29
4. Functions ................................................................................................... 304.1 Basic Settings .................................................................................................... 32
4.1.1 Overview ............................................................................................. 324.1.2 PVMappingofProcessValue ............................................................ 334.1.3 DisplayoftheProcessValue .............................................................. 334.1.4 UnitSetting .......................................................................................... 344.1.5 SpanSetting........................................................................................ 35
3rd Edition: Mar. 2020AllRightsReserved,Copyright©2019,YokogawaElectricCorporation
2
IM 01E22C02-01EN
4.1.6 DampingTimeConstantSetting ......................................................... 364.1.7 Low-cutFunctionSetting .................................................................... 374.1.8 MeterfactorSetting ............................................................................. 394.1.9 SensorTypeSetting ............................................................................ 404.1.10 Sensor’sNominalSizeSetting ........................................................... 414.1.11 DensitySetting .................................................................................... 414.1.12 TemperatureSetting ........................................................................... 424.1.13 SpanFlowVelocityDisplay ................................................................. 424.1.14 UserSpanFunction ............................................................................ 434.1.15 ZeroAdjustment .................................................................................. 45
4.2 Totalization Function ......................................................................................... 474.2.1 TotalizedValueandUnitSetting ......................................................... 474.2.2 TotalizedValueDisplayandTotalizationFunction .............................. 474.2.3 TotalizationSwitchFunction ............................................................... 494.2.4 TotalizerOperationatAlarmOccurrence ............................................ 504.2.5 Start/StopSettingforTotalizationFunction......................................... 504.2.6 TotalizationDirectionSetting .............................................................. 514.2.7 TotalizedValueReset/PresetFunction ............................................... 52
4.3 Pulse Output, Frequency Output, and Status Output .................................... 534.3.1 Pulse/StatusOutputModeSetting ..................................................... 534.3.2 PulseOutput/FrequencyOutputMapping .......................................... 544.3.3 AlarmOutputFunction ........................................................................ 544.3.4 PulseWidthSetting ............................................................................. 554.3.5 ActiveDirectionSetting ....................................................................... 564.3.6 ActivePulseSetting ............................................................................ 574.3.7 PulseRateSetting .............................................................................. 574.3.8 FrequencyOutputRangeSetting ....................................................... 594.3.9 StatusOutputFunctionSetting ........................................................... 604.3.10 PulseOptionSetting ........................................................................... 63
4.4 Status Input......................................................................................................... 644.4.1 ActiveDirectionSettingforStatusInput ............................................. 644.4.2 StatusInputFunctionSetting .............................................................. 65
4.5 Analog Input and Analog Output ..................................................................... 664.5.1 AnalogOutputSelect .......................................................................... 664.5.2 AnalogOutputHigh/LowLimitFunction ............................................. 674.5.3 AbsoluteRangeFunction ................................................................... 684.5.4 AlarmOutputFunction ........................................................................ 694.5.5 AnalogOutputPriority ......................................................................... 694.5.6 CurrentValueAdjustment ................................................................... 704.5.7 AnalogInputFunctionSetting ............................................................. 714.5.8 DisplayAnalogInput ........................................................................... 714.5.9 AnalogInputUnitSetting .................................................................... 71
3
IM 01E22C02-01EN
4.5.10 AnalogInputRangeSetting ................................................................ 724.5.11 AnalogInputHigh/LowLimitFunction ................................................ 72
4.6 Multi Range Function ........................................................................................ 734.6.1 MultiRangetype ................................................................................. 734.6.2 MultiRangeSetting ............................................................................. 734.6.3 MultiRangeOperation ........................................................................ 744.6.4 Forward/ReverseRange .................................................................... 764.6.5 CombinationofMultiRangeandForward/ReverseRange ............... 784.6.6 ExternalContactRange ...................................................................... 794.6.7 ExternalContactRange(withAnswer) ............................................... 804.6.8 CombinationofExternalContactRangeandForward/Reverse
Range .................................................................................................. 814.7 Auxiliary Calculation Function ......................................................................... 83
4.7.1 FluidFlowDirectionSetting ................................................................ 834.7.2 RateLimitFunctionSetting ................................................................. 834.7.3 PulsingFlowSupportFunctionSetting ............................................... 864.7.4 PowerFrequencySynchronizationSetting ........................................ 864.7.5 DensityCorrectionCalculation ........................................................... 874.7.6 CalorieCalculation .............................................................................. 894.7.7 LimitofAlarmOutputFunction............................................................ 90
4.8 Alarm ................................................................................................................... 914.8.1 ErrorsandCountermeasures ............................................................. 914.8.2 OperationattheTimeofError ............................................................ 974.8.3 AlarmDisplaySetting ........................................................................ 1014.8.4 AlarmHistoryFunction...................................................................... 1024.8.5 AlarmMaskFunction ........................................................................ 1034.8.6 BurnoutDirectionDisplay ................................................................. 1094.8.7 AlarmTerminalActiveDirectionSetting ............................................ 109
4.9 Display ............................................................................................................... 1104.9.1 LanguageSetting ...............................................................................1104.9.2 DisplayItemSetting ...........................................................................1114.9.3 Decimal-PointPositionSetting ..........................................................1124.9.4 DisplayLineCountandScrollSettings ..............................................1144.9.5 TrendGraphSetting ..........................................................................1164.9.6 OtherSettings ....................................................................................1184.9.7 microSDCardSetting ....................................................................... 120
4.10 Device Information ........................................................................................... 1224.10.1 OrderInformation .............................................................................. 1224.10.2 DeviceRevision ................................................................................ 1234.10.3 MemoFunction ................................................................................. 1234.10.4 DateandTimeInformation ............................................................... 1244.10.5 DisplayOperationTime .................................................................... 124
4
IM 01E22C02-01EN
4.10.6 ExplosionProtectionSetting ............................................................. 1254.11 Self-diagnostic Function ................................................................................. 126
4.11.1 TypesofDiagnosisFunctions ........................................................... 1264.11.2 AlarmHigh/LowLimitFunction ......................................................... 1264.11.3 ElectrodeAdhesionDetection .......................................................... 1284.11.4 SensorEmptyCheck ........................................................................ 1294.11.5 WiringConnectionDiagnosis ........................................................... 1304.11.6 Verification(DeviceHealthDiagnosis)Function .............................. 1314.11.7 ElectrodeInsulationDeteriorationDiagnosis ................................... 1354.11.8 FlowNoiseDiagnosis ....................................................................... 1364.11.9 LowConductivityDiagnosis .............................................................. 1374.11.10 CoilInsulationDiagnosis................................................................... 1384.11.11 OtherSetting ..................................................................................... 138
4.12 Test Mode .......................................................................................................... 1394.12.1 TestModeSetting ............................................................................. 1394.12.2 TestValueSetting ............................................................................. 1404.12.3 Test2Mode ....................................................................................... 1414.12.4 TestModeAutoReset ....................................................................... 141
4.13 Event Management Function .......................................................................... 1424.13.1 BackupFunction ............................................................................... 1424.13.2 Restore/DuplicateParameter ........................................................... 1454.13.3 BackupandRestoreParameters ..................................................... 1494.13.4 DataLoggingFunction ...................................................................... 153
4.14 Parameter Protection ...................................................................................... 1554.14.1 SoftwareWriteProtectionFunction .................................................. 1554.14.2 DisplayRestrictionFunction ............................................................. 156
5. Parameter Lists........................................................................................ 1576. Parameter Menu Tree .............................................................................. 191
6.1 Display Menu Tree ........................................................................................... 191
Revision Information ......................................................................................... 208
<1. Introduction> 5
IM 01E22C02-01EN
1. IntroductionThismanualexplainsbasicoperationsofADMAGTISeriesAXG1AMagneticFlowmeterRemoteTransmitterwithBRAINCommunicationProtocol.Foritemswhicharenotcoveredinthismanual,readtheapplicableuser’smanualslistedin“Table1.1ManualandGeneralSpecificationsList”intheADMAGTISeriesInstallationManual.ThesedocumentscanbedownloadedfromtheYOKOGAWAwebsite.Toensurethecorrectuseoftheproduct,readthesemanualsthoroughlyandfullyunderstandhowtooperatetheproductbeforeoperatingit.Toconfirmthemodelnameandspecificationsoftheproduct,refertothegeneralspecifications.Websiteaddress:http://www.yokogawa.com/fld/doc/
�� Precautions Related to the Protection, Safety, and Alteration of the Product
Thefollowingsafetysymbolmarksareusedinthismanualandtheproduct.
WARNING
AWARNINGsigndenotesahazard.Itcallsattentiontoaprocedure,practice,condition,orthelike,which,ifnotcorrectlyperformedoradheredto,couldresultininjuryordeathofpersonnel.
CAUTIONACAUTIONsigndenotesahazard.Itcallsattentiontoaprocedure,practice,condition,orthelike,which,ifnotcorrectlyperformedoradheredto,couldresultindamagetoordestructionofpartoralloftheproduct.
IMPORTANTAnIMPORTANTsigndenotesthatattentionisrequiredtoavoiddamagetotheproductorsystemfailure.
NOTEANOTEsigndenotesinformationnecessaryforanessentialunderstandingofoperationandfeatures.
<1. Introduction> 6
IM 01E22C02-01EN
Thefollowingsymbolsareusedintheproductandthemanualtoindicatetheaccompanyingsafetyprecautions:
Protectivegroundingterminal
Functionalgroundingterminal(Thisterminalshouldnotbeusedasaprotectivegroundingterminal.)
Alternatingcurrent
Directcurrent
CautionThissymbolindicatesthattheoperatormustrefertoanexplanationintheuser’smanualinordertoavoidtheriskofinjuryordeathofpersonnelordamagetotheproduct.
Fortheprotectionandsafeuseoftheproductandthesysteminwhichthisproductisincorporated,besuretofollowtheinstructionsandprecautionsonsafetythatarestatedinuser’smanualwheneveryouhandletheproduct.Takespecialnotethatifyouhandletheproductinamannerthatviolatestheseinstructions,theprotectionfunctionoftheproductmaybedamagedorimpaired.Insuchacase,YOKOGAWAdoesnotguaranteethequality,performance,function,orsafetyoftheproduct.
�� Regarding This User’s Manual• Thismanualshouldbeprovidedtotheenduser.
• Thecontentsofthismanualaresubjecttochangewithoutpriornotice.
• NopartofthismanualmaybereproducedinanyformwithoutYOKOGAWA’swrittenpermission.
• YOKOGAWAmakesnowarrantyofanykindwithregardtothismanual,including,butnotlimitedto,impliedwarrantyofmerchantabilityandfitnessforaparticularpurpose.
• Ifanyquestionsariseorerrorsarefound,orifanyinformationismissingfromthismanual,informthenearestYOKOGAWAsalesoffice.
• Thespecificationscoveredbythismanualarelimitedtothoseforthestandardtypeunderthespecifiedmodelnumberbreak-downanddonotcovercustom-madeproducts.
• Notethatchangesinthespecifications,construction,orcomponentpartsoftheproductmaynotimmediatelybereflectedinthismanualatthetimeofchange,providedthatpostponementofrevisionswillnotcausedifficultytotheuserfromafunctionalorperformancestandpoint.
• Thismanualisintendedforthefollowingpersonnel:
Engineersresponsiblefortheinstallationandwiringoftheproduct.
Personnelresponsibleforthenormaldailyoperationoftheproduct(operator).
• Toensurecorrectuse,readthismanualandtheapplicablemanualsthoroughlybeforestartingoperation.Readthegeneralspecificationsforspecificationsoftheproduct.
<1. Introduction> 7
IM 01E22C02-01EN
�� Trademarks• AllthebrandsornamesofYokogawaElectric’sproductsusedinthismanualareeither
trademarksorregisteredtrademarksofYokogawaElectricCorporation.
• Allothercompanyandproductnamesmentionedinthismanualaretradenames,trademarksorregisteredtrademarksoftheirrespectivecompanies.
• Inthismanual,trademarksandregisteredtrademarksarenotmarkedwith“™”or“®”.
<1. Introduction> 8
IM 01E22C02-01EN
1.1 For Safe Use of ProductFortheprotectionandsafeuseoftheproductandthesysteminwhichthisproductisincorporated,besuretofollowtheinstructionsandprecautionsonsafetythatarestatedinuser’smanualwheneveryouhandletheproduct.Takespecialnotethatifyouhandletheproductinamannerthatviolatestheseinstructions,theprotectionfunctionoftheproductmaybedamagedorimpaired.Insuchacase,YOKOGAWAshallnotbeliableforanyindirectorconsequentiallossincurredbyeitherusingornotbeingabletousetheproduct.
�� General
WARNING
• Donotopenthecoverinwetweatherorhumidenvironments.Whenthecoverisopen,thestatedenclosureprotectionisnotapplicable.
• Whenopeningthecover,waitformorethan20minutesafterturningoffthepower.Onlyanexpertengineerorskilledpersonnelispermittedtoopenthecover.
�� Operation
WARNING
Besuretoenablethewriteprotectfunctiontopreventparametersfrombeingoverwrittenafterfinishingparametersetting.Inrarecases,theIRswitchesmayrespondunexpectedlytowaterdropsorextraneoussubstancesstickingonthesurfaceofthedisplaypanelduetocharacteristicsoftheoperatingprinciples.Thepossibilityofmalfunctionarisesafterrainorcleaningneartheplacewheretheflowmeterisinstalled.Repeatedlyturningaflashlight,etc.onandoffinthedirectionoftheIRswitchmayalsobethecauseofmalfunction.Refertotheinstallationmanualforthehardwarewriteprotectfunction,andSection4.14forthesoftwarewriteprotectfunction.
�� Maintenance
WARNING
• Ifdirt,dust,orothersubstancesadherestotheglassofthedisplay,wipethemcleanwithasoftdrycloth.
• Maintenanceofthisproductshouldbeimplementedinamaintenanceserviceshopwherenecessitytoolsandenvironmentconditionareprovided.Therequiredenvironmentalconditionisthattheambienttemperatureshouldbe5to40°C(themaximumrelativehumidityis80%fortemperature5to31°C,anddecreasinglinearlyto50%relativehumidityat40°C).
<1. Introduction> 9
IM 01E22C02-01EN
�� microSD card
IMPORTANT• DonotstoreorusethemicroSDcardinplaceswithstaticelectricity,nearelectricallycharged
objects,orwhereelectricalnoiseispresent.DoingsocanresultinshockordamagetothemicroSDcard.
• DonotdisassembleormodifythemicroSDcard.• Donotphysicallyshock,bend,orpinchthemicroSDcard.• Duringreading/writingofdata,donotturnoffthepower,applyvibrationorshock,orpullout
thecard.Datacanbecorruptedorpermanentlylost.• UseonlythemicroSDcardsdesignatedbyYOKOGAWA.Theoperationofthedevicecannot
beguaranteedwhenothercardsareused.• WheninsertingthemicroSDcardintotheproduct,makesuretoorientthemicroSDcard
correctly(faceupordown)andinsertitsecurely.Ifnotinsertedcorrectly,themicroSDcardwillnotberecognizedbytheproduct.
• DonottouchthemicroSDcardwithwethands.• DonotusethemicroSDcardifitisdustyordirty.• ThemicroSDcardcomesformatted.IfyouwouldliketoformatthemicroSDcard,usethe
product’sFormatfunction.• YOKOGAWAprovidesnowarrantyfordamageto,orlossofdatarecordedonthemicroSD
card,regardlessofthecauseofsuchdamageorloss.Werecommendregularlymakingbackupcopiesofyourdata.
<1. Introduction> 10
IM 01E22C02-01EN
1.2 Warranty• Thewarrantyshallcovertheperioddescribedinthequotationpresentedtothepurchaserat
thetimeofpurchase.Problemsthatmayoccurduringthewarrantyperiodshallberepairedfreeofcharge.
• Incaseofproblems,thecustomershouldcontacttheYOKOGAWArepresentativefromwhichtheproductwaspurchasedorthenearestYOKOGAWAoffice.
• Ifaproblemariseswiththisproduct,pleaseinformYOKOGAWAofthenatureoftheproblemandthecircumstancesunderwhichtheproblemdeveloped,includingthemodelspecificationandserialnumber.Anydiagrams,dataandotherinformationyoucanincludeinyourcommunicationwillalsobehelpful.
• ResponsiblepartforrepaircostsoftheproblemsshallbedeterminedbyYOKOGAWAbasedonourinvestigation.
• Thepurchasershallbeartheresponsibilityforrepaircosts,evenduringthewarrantyperiod,ifthemalfunctionisdueto:
- Improperand/orinadequatemaintenancebythepurchaser.
- Failureordamageduetoimproperhandling,use,orstoragewhichdoesnotconformtodesignconditions.
- UseoftheproductinquestioninalocationnotconformingtothestandardsspecifiedbyYOKOGAWA,orproblemsduetoimpropermaintenanceoftheinstallationlocation.
- FailureordamageduetomodificationorrepairbyanypartyexceptYOKOGAWAoranapprovedrepresentativeofYOKOGAWA.
- Malfunctionordamagefromimproperrelocationoftheproductinquestionafterdelivery.
- Reasonofforcemajeuresuchasfires,earthquakes,storms/floods,thunder/lightning,orothernaturaldisasters,aswellasdisturbances,riots,warfare,orradioactivecontamination.
<2. Operation by Display unit> 11
IM 01E22C02-01EN
2. Operation by Display unit
2.1 Basic Operating ProceduresTheparametersettingscanbechangedbyusingthreeIR(infra-red)switches;[SET],[SHIFT]and[▼]switchesonthedisplay.ThethreeIRswitchesenabletheusertosetparametersfromtheoutsideoftheglassofthedisplaycover.ThissectionprovidesdescriptionsofbasicparameterconfigurationonthedisplayandoperationproceduresoftheIRswitches.ThisproductcanbealsooperatedbyusingthededicatedhandheldterminalortheFieldMate(VersatileDeviceManagementWizard).Fordetailsaboutthesettingprocedure,refertoChapter4.
WARNING
Besuretoenablethewriteprotectfunctiontopreventparametersfrombeingoverwrittenafterfinishingparametersetting.Inrarecases,theIRswitchesmayrespondunexpectedlytowaterdropsorextraneoussubstancesstickingonthesurfaceofthedisplaypanelduetocharacteristicsoftheoperatingprinciples.Thepossibilityofmalfunctionarisesafterrainorcleaningneartheplacewheretheflowmeterisinstalled.Turningonandofftheflashlight,etc.towardstheIRswitchmayalsobethecauseofthemalfunction.RefertotheInstallationManualforthehardwarewriteprotectfunction,andSection4.14forthesoftwarewriteprotectfunction.
IMPORTANT
Operatethedisplayunderconditionswheredirectsunlight,etc.donotshinetotheIRswitchesdirectlywhentheparametersettingoperationiscarriedout.
NOTE
• Alwayskeepthedisplaycoverclosedandoperatethesettingswitchesfromtheoutsideoftheglasswindow.
• Ifdirt,dust,orothersubstancesadherestotheglassofthedisplay,wipethemcleanwithasoftdrycloth.
• Operationwithdirtyglovesmaycauseaswitchresponseerror.
NOTE
Thelanguageonthedisplayissetto“English”asdefaultatfactoryshipment.SelecttheadequatelanguagereferringtoSubsection4.9.1.Themenupassofthedisplayonthismanualissetto“English”.
<2. Operation by Display unit> 12
IM 01E22C02-01EN
2.2 DisplayandBasicConfigurationThedisplayoftheAXG1Aremotetransmitterhasthevariousfunctionsbelow.
2.2.1 DisplayConfiguration
22.000mA0.0%
0.00000m/s
00 : 00VEL
FLP
A01SET SFT INC
Status icon and time
Data display(Abbreviation: see Table 2.2.1)Switch On/Off status(On: inverted)IR switches
F0201.ai
microSD card(Optional code: /MC)
(1) Basic operation of IR switchesThebasicoperationsofthedisplaypanelaredonebythethreeIRswitches;[SET],[SHIFT],and[▼].CombiningtwoIRswitchesprovidesadifferentfunction,andthefunctionoftheswitchisdisplayedonthedisplay.
IR switch (Note 1) Function of switch (Note 2) Function
[SET►] SET ▪Applyparameteranddata(Note3)▪Movetonextmenu
[SHIFT] SFT▪Set/resetMultipleselectableoptions (Selecttypeparameter)▪Movecursorright (Numerictypeparameter)
[▼] INC
▪Movecursordown (Selecttypeparameter)▪Incrementvalue (Numerictypeparameter)▪Changepositionofdecimalpoint (Numerictypeparameter)
[SHIFT]+[▼](=[▲]) DEC▪Movecursorup (Selecttypeparameter)▪Decrementvalue (Numerictypeparameter)
[SHIFT]+[SET►](=[ESC◄]) ESC ▪Cancel
▪Backtopreviousmenu
Note1:[A]+[B](=[C]):Thefunctionischangedtoswitch[C]whenswitch[B]ispressedwhilepressingswitch[A].Note2:[SET],[SFT],[INC],[DEC]and[ESC]indicatetheassignedfunctioninaccordancewithdisplaymodeatthattime.All
functionscannotbeusedsimultaneously.Note3:“Apply”and“Enter”areexecutedbypressing[SET]twice.Iftheexecutiondoesnotworkproperly,releaseyourfinger
fromthedisplayglasscompletelyafterinitiallypressing[SET],andthen,pressthatkeyagain.
<2. Operation by Display unit> 13
IM 01E22C02-01EN
(2) Status iconsIcon Description Icon Description
WriteprotectionfunctionDisable
WriteprotectionfunctionEnable
DeviceBusy DeviceFaultmicroSDcardReady
microSDcardAccessing
microSDcardDisabletoaccess
ParameterUploading
ParameterDownloading
TrendgraphfunctionRunning
SystemAlarmOccurs BRAINcommunication
ProcessAlarmOccurs
SettingAlarmOccurs
WarningOccurs
InformationOccurs
DisplayDampingValid
Operation level:Operator
Operation level:Maintenance
Operation level:Specialist
(3) Data indication partUptoeightprocessvaluescanbeselectedonthedisplay.Itispossibletodisplayuptofouritemsonthedisplayatthesametime,andtheremainingfouritemscanbedisplayedifscrolled.Table 2.2.1 Abbreviation table of process values to be indicated on the display
Abbreviation DescriptionFLP*1 Flowrate%PRV*1 ProcessvalueVEL*1 FlowvelocityVFL*1 VolumetricflowrateMFL*1 MassflowrateFLB Flowratein%bargraphCAL*1 CalorieTL1*1 Totalizedvalue1TL2*1 Totalizedvalue2TL3*1 Totalizedvalue3TAG TagNo.COM CommunicationprotocolADH Adhesiondiagnosticlevel(AlarmatLevel4)AO1*1 Analogoutputvalue1AO2*1 Analogoutputvalue2FNL Flowlevelofflownoisediagnosis(AlarmatLevel4)TC1 Total1countTC2 Total2countTC3 Total3count
*1: Atrendgraphcanbedisplayedonline.
NOTE
ThePRV(PVvalue)andFLP(flowrate%)operateinthesamewayasAnalogoutput1.IfthelowcutfunctionisenableforAnalogoutput1(G01:AO1LOWCUT/Lowcut),thePRVandFLPareaffectedbythelowcutfunction.Theirvalueafterlowcutaredisplayed.Ontheotherhand,theVEL(flowvelocity),VFL(volumeflow),MFL(massflow),andCAL(calorificvalue)arenotaffectedbythelowcutfunction.Theirvaluesaredisplayedastheyare.
<2. Operation by Display unit> 14
IM 01E22C02-01EN
2.2.2 Operation LevelWhensettingaparameterfromthedisplay,configurableparametersdifferdependingonthethreeoperationlevelsshowninTable2.2.2.Formaintenanceandspecialistlevels,thepasscodeneedstobeentered.Table 2.2.2 Parameter Setting from Display and Operation Level
Operation Level Parameter DescriptionReading Writing
Operator AllparametersPossibletosetadisplaylanguageonthedisplayandparametersrelatedtodisplayitems.
Nopasscoderequired
Maintenance AllparametersPossibletosetparametersallowedtosetwiththeoperatorlevelandparametersrelatedtozeroadjustment.
PasscoderequiredDefaultvalue:0000
Specialist Allparameters Possibletosetallwritableparameters
PasscoderequiredDefaultvalue:0000
Apasscodecanbeconfiguredonlyfromthedisplay.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Accesscfg►(seebelow)
Parameter Description
ChgmainteChangeofpasscode(maintenancecode)withmaintenanceoperationlevelauthority.Tochangethelevel,maintenanceorspecialistoperationlevelauthorityisnecessary.
Chgspecial Changeofpasscode(specialcode)withspecialistoperationlevelauthority.Tochangethelevel,specialistoperationlevelauthorityisnecessary.
2.3 Display Mode and Setting ModeThedevicerunsinDisplayModewhenthepowerison.Tocheckorchangeparameters,SettingModemustbeactivated.ThefollowingprocedureexplainshowtochangetoSettingModefromSettingMode.ForthefunctionsoftheIRswitches,refertoSubsection2.2.1.
[Procedure]1) Keeptouching[SET]switchforfewseconds.
F0513.ai
0.00000m/s0.00000l/min
00 : 00PRV
VEL
FLBSET SFT INC
2) Touch[SFT]+[INC]switches.
F0514.ai
Touch SFT + INC00 : 00
SET SFT INC
3) “No“isselected. Touch[INC]switchandselect“Yes”.
F0515.ai
Setting Mode
No Yes▲
▼
00 : 00
SET SFT INC
<2. Operation by Display unit> 15
IM 01E22C02-01EN
4) Touch[SET]switch.
F0516.ai
Setting Mode
Yes No▲
▼
00 : 00
SET SFT INC
5) “Yes”isblinking. Touch[SET]switchagain.
F0517.ai
00 : 00
SET SFT INC
Setting Mode Yes
6) ThescreenmovestothemenuofOperationLevel.
F0518.ai
Operation level Specialist Exit Operator
▲
▼
00 : 00
SET SFT INC
7) Selectanappropreateoperationlevelbymovingthecursorwith[INC]or[DEC]switch. Passcodeisnotnecessaryfor“Operator”.For“Maintenance”and“Specialist”,passcode
isnecessaryforeach.Forpasscodesetting,[SFT]isforpositionchange,and[INC]isfornumber,thentwice[SET]isforentrycompletion.
Thedefaultpasscodeatthefactoryshipmentissetto“0000”.
F0519.ai
Pass code **** 0000 0000 9999
00 : 00
SET SFT INC
8) WhentheOperationLevelisdetermined,thescreenmovesto“Devicesetup”astheSettingModewhereparameterscanbeconfigured.
9) Aftercompletingparametersetting,push[ESC]switch.ThescreenreturnstotheDisplayMode.
NOTE
FollowtheprocedurebellowwhenIndsoftrev(K52:INDBREV/Indsoftrev)isR2.01.02orearlier.
[Procedure]1) TurnonthepowerandwaitforseveralsecondstostartDisplayMode.
F0212.ai
0.00000m/s0.00000l/min
00 : 00PRV
VEL
FLBSET SFT INC
2) Pressandholdthe[SET]switchfortwoseconds.Thescreenwillswitchtothe“Operationlevel”menu.
F0213.ai
Operation level Maintenance Specialist Exit
▲
▼
00 : 00
SET SFT INC
<2. Operation by Display unit> 16
IM 01E22C02-01EN
NOTE
3) Selectanappropriateoperationlevelbymovingthecursorwiththe[INC]or[DEC]switch. The“Operator”doesnotrequireapasscode,butfor“Maintenance”and“Specialist”,the
passcodeisnecessary. Whensettingthepasscode,[SFT]isforpositionchange,[INC]isfornumberchange,and
pressing[SET]twiceisforentrycompletion. Thedefaultpasscodeissetto“0000”whenshippedfromthefactory.
F0214.ai
Pass code **** 0000 0000 9999
00 : 00
SET SFT INC
4) Whentheoperationlevelisdetermined,thescreenmovesto“Devicesetup”andswitchestoSettingMode.InSettingMode,parameterscanbeconfigured.
5) Aftercompletingtheparametersettings,pressthe[ESC]switch.ThescreenwillreturntoDisplay Mode.
IMPORTANT
Menu pathDisplay Devicesetup►Easysetupwizard
WhenparametersonthemenudescribedabovearesetintheWizardofEasysetup,besuretoexecute“Settingdownload”ineachmenuaftersettingtheparameters.If“Settingdownload”isnotexecuted,anychangedparameterswillnotbecorrectlyappliedtothedevice.
NOTE
Ifthereisnooperationfor10minutesinSettingMode,thescreengoesbacktoDisplayMode.
Parameter formThefollowingthreeparameterformsareavailable.
Type Example of display Description
SelecttypeUnit m3
l(liter) cm3▲
▼
00 : 00
SET SFT INC
Selectstheadequatetypeofdatafromamongthepre-determinedalternatives.
NumerictypeSpan +7200.00 +2200.00 +0.00001 +999999
00 : 00
SET SFT INC
Specifiesthedatawithacombinationofnumberandadecimalpointforeachdigit.
Alphanumerictype
T0500_03.ai
Tag No FT-1234 FT-1234
00 : 00
SET SFT INC
Thedataisconfiguredwiththecombinationofalphanumericcharacters.(TagNo.,Specialunit,etc.)
Thealphanumerictypedisplaysthefollowingalphanumericcharactersinorder.0123456789ABCDEFGHIJKLMNOPQRSTUVWXYZabcdefghijklmnopqrstuvwxyz!"#$%&'()*+,-./:;<=>?@[\]^_`{|}~"space"
<2. Operation by Display unit> 17
IM 01E22C02-01EN
2.4 Parameter Setting from Display PanelThissectionexplainshowtospecifytheparametersfromthedisplaypanel.RefertoSection2.3,andsettheoperationlevelto“Specialist”.AlsoselectaparametertobespecifiedintheSettingMode.
NOTE
Theflowrateunit,spanflow,andtagNo.areexplainedasanexample.Ifaparticularparametersisspecifiedatthetimeofordering,thisproductisshippedwiththeparameterspecified.Ifaparameterisnotspecifiedatthetimeofordering,thatparameterneedstobesetbythecustomer.
2.4.1 Setting Example of Select Type Parameter (Flow rate unit)
ThefollowingistheprocedureofchangingtheflowrateunitfortheSelecttypeparameter.Thephysicalunitandtimeunitneedtobespecifiedindividuallyfortheflowrateunit.Whensettingthevolumetricflowrateunittol/min,set“l(liter)”tothephysicalunitand“/min”tothetimeunit.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Provar►Volume►(seebelow)
Parameter DescriptionUnit Specifiesthephysicalunitofthevolumetricflowrate.Timeunit Specifiesthetimeunitofthevolumetricflowrate.
F0215.ai
Volume Damp pls/ttl Unit Time unit
▲
▼
00 : 00
SET SFT INC
Specifythephysicalunitforvolumetricflowrate.Movethecursorwith[INC]and[DEC]accordingtothemenupathdescribedabove,press[SET]todetermine,andselectthe“Unit”.
F0216.ai
Unit m3
l(liter) cm3▲
▼
00 : 00
SET SFT INC
Movethecursorwith[INC]and[DEC],press[SET]todetermine,andselectthe“l(liter)”.Whiletheselectedunitisblinking,press[SET]todeterminetheunit.Thescreenreturnstothesettingpageaftercompletingthesetting.
F0217.ai
Volume Unit Time unit Span
▼
00 : 00
SET SFT INC
▲
Specifythetimeunitforvolumetricflowrate.Movethecursorwith[INC]and[DEC]accordingtothemenupathabove,press[SET]todetermine,andselectthe“Timeunit”.
F0218.ai
Time unit /s /min /h
▼
00 : 00
SET SFT INC
▲
Movethecursorwith[INC]and[DEC],select“/min”,andthenpress[SET].Whiletheselectedunitisblinking,press[SET]todeterminetheunit.Thescreenreturnstothesettingpageaftercompletingthesetting.
F0219.ai
0.00000m/s0.00000l/min
00 : 00PRV
VEL
FLBSET SFT INC
Aftercompletingtheparametersetting,youcanreturnthescreentoDisplayModebypressing[ESC].
<2. Operation by Display unit> 18
IM 01E22C02-01EN
NOTE
Besuretosettheflowrateunitfirstwhenchangingtheflowrateunitandflowspanvalueatthesametime.Whentheflowrateunitischanged,thespanflowrateisconvertedinconjunctionwiththechangedunit.
2.4.2 Setting Example of Numeric Type Parameter (Flow rate span)
Thefollowingistheprocedureofchangingtheflowratespanasnumerictypeparameter.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Provar►Volume►Span
NOTE
Besuretosettheflowrateunitfirstwhenchangingtheflowrateunitandflowspanvalueatthesametime.Whentheflowrateunitischanged,thespanflowrateisconvertedinconjunctionwiththechangedunit.
F0220.ai
Volume Time unit /h Span +7200.00 Setting download
00 : 00
SET SFT INC
▲
▼
Specifytheflowratespanunit.Movethecursorwith[INC]and[DEC]accordingtothemenupathabove,press[SET]todetermine,andselectthe“Span”.
F0221.ai
Span +7200.00 +2200.00 +0.00001 +999999
00 : 00
SET SFT INC
Theswitch'sfunctionalityforsettingthespanflowrateisasbelow:Plus/minusandnumericchange:[INC]Movementofdigits:[SFT]Determinationofparameter:[SET]:Minimumvalue:Maximumvalue
Press[SET]todeterminetheparameterwhilethesettingvalueoftheflowratespanisblinking.Thescreenreturnstothesettingpageaftercompletingthesetting.
<2. Operation by Display unit> 19
IM 01E22C02-01EN
2.4.3 Setting Example of Alphanumeric Type Parameter (Tag No.)
ThefollowingistheprocedureofchangingtheTagNo.ofalphanumerictypeparameter.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Deviceinfo►Orderinfo►TagNo
F0222.ai
Device info Date/Time Order info Ver/Num info
00 : 00
SET SFT INC
▲
▼
►
►
►
SpecifytheTagNo..Movethecursorwith[INC]and[DEC]accordingtothemenupathabove,press[SET]todetermine,andselectthe“TagNo”.
F0223.ai
Tag No FT-1234 FT-1234
00 : 00
SET SFT INC
TheTagNo.canbesetwithuptoeightcharactersfromthedisplay.Theswitch'sfunctionalityforsettingtheTagNo.isasfollows:Numberandcharacterchange:[INC]Movementofdigits:[SFT]Determinationofparameter:[SET]Availablecharacters:ASCIIcharactersPress[SET]todeterminetheparameterwhilethevalueofTagNo.isblinking.Thescreenreturnstothesettingpageaftercompletingthesetting.
<2. Operation by Display unit> 20
IM 01E22C02-01EN
2.5 microSD Card Insertion/RemovalIftheoptionalcodeMCisselected,theparametersettingcanbestoredintothededicatedmicroSDcardinsertedintothedisplayunit.Thestoreddatacanberestoredtothedevice.Fordetailedonthefunctions,refertoChapter4.
WARNING
UseonlyamicroSDcarddesignatedbyYOKOGAWA.Theoperationofthedevicecannotbeguaranteedwhenothercardsareused.
(1) Insertion of microSD CardCarefullyinsertthededicatedmicroSDcardintotheslotonthedisplayuntiltheslotholdsthecard.(seeFigure2.5).
22.000mA0.0%
0.00000m/s
00 : 00VEL
FLP
AO1SET SFT INC
F0224.ai
microSD card
Figure 2.5 Insertion of microSD Card
(2) Removal of microSD CardRemovethemicroSDfromtheslotshowninFigure2.5.WhenremovingthemicroSDcard,select“Execute”fromthefollowingparameters.Menu path
Display Devicesetup►microSD►Unmount
IMPORTANT
IfthemicroSDcardisremovedwithoutexecuting“Unmount”onthesoftware,itmaycausethestoreddatatobeerasedortheproducttooperateabnormally.
<3.OperationwithBRAINConfigurationTool> 21
IM 01E22C02-01EN
3. OperationwithBRAINConfigurationTool
ThischapterdescribestheconnectionmethodofthisproductandBRAINconfigurationtool(BRAINTERMINAL(BT200)orFieldMate(VersatileDeviceManagementWizard)),andtheoperationmethodusingBT200.Readtheuser’smanualofBT200(IM01C00A11-01E)fordetailsabouttheBT200.Readtheuser’smanualofFieldMate(IM01R01A01-01E)fordetailsabouttheFieldMate.
NOTEInthismanual,theoperationwiththeconfigurationtoolisexplainedusingthescreenofBT200forwhichlanguageissettoEnglish.
3.1 ConnectingBRAINConfigurationToolInBRAINcommunication,thecommunicationsignalissuperimposedontothe4to20mADCanalogsignaltobetransmitted.TheBRAINconfigurationtoolcanbeconnectedtoanyrelayterminalintheinstrumentroomortransmissionloopforcommunication,providedthereistheminimumloadresistanceof250ΩbetweenthisproductandtheconnectionterminaloftheBRAINsettingtool.TheBRAINconfigurationtoolmustbeconnectedinparallelwiththisproductforcommunication,buttheconnectionisnotaffectedbythepolarity.SeeFigure3.1foraconnectionexample.
4 to 20 mA DCSignal Transmission Line DistributorTerminal
Board
BT200F0301.ai
Iout+Iout-
USBFieldMate
Modem
USB
PC/FieldMate
Magnetic FlowmeterTerminal Box
Relaying Terminals
Control Room
LoadResistance
250 to 450 Ω
Figure3.1 ConnectingtheBRAINConfigurationTool
IMPORTANTCommunicationsignalissuperimposedonoutputsignalatcommunication.Itisrecommendedtosetalow-passfilter(approximately0.1s)tothereceiverinordertoreducetheoutputeffectfromcommunicationsignal.Beforeonlinecommunication,confirmthatcommunicationsignaldoesnotgiveeffectonthehigh-levelsystem.
IMPORTANTRestrictionsexistwithregardtothedistanceoverwhichcommunicationispossible.Readthegeneralspecifications.
<3.OperationwithBRAINConfigurationTool> 22
IM 01E22C02-01EN
AfterconnectingtheBT200,checkthatitoperatesproperly.(1) WhentheBT200isturnedon,themessage“Pleasewait..”.isdisplayedontheBT200
screenforseveralseconds.Then,presstheEnterkey.
(2) Theinitialdatascreenisdisplayed.Pressthefunctionkey“F4”[OK]ortheEnterkey.
F0302.ai
PARAM01:MODEL 02:TAG NO 03:SELF CHECK GOOD
OK
Initial data screen
[01:MODEL]:Productmodelname[02:TAGNO]:TagNo.(ifspecifiedatordering)[03:SELFCHECK]:AbnormalitycheckforthisproductGOODorERROR
(3) Whenthemenuscreenisdisplayed,theparameterscanbedisplayedandchangedusingtheBT200.
IfthereisaprobleminthewiringsystemwiththeBT200,“communicationerror”isdisplayed.
F0303.ai
MENUA:DISPLAYB:EASY SETUPC:BASIC SETUPD:TOTAL SETE:PULSE OUTF:STATUS I/OHOME SET ADJ ESC
communication error
ESC
Menu screen Communication error (Faulty wiring)
<3.OperationwithBRAINConfigurationTool> 23
IM 01E22C02-01EN
3.2 BT200 Operating Procedures3.2.1 Key Layout and Screen Display
Alphanumeric keys• Enter numbers• Enter alphabetic
characters in combination with the shift keys.
F0304.ai
Power ON/OFF key
Shift keys
Screen Display(21 characters × 8 lines)
Movement keys• Move the cursor• Change pages
Function keysUsed to execute commands displayed at the bottom of the screen.(See Table 3.2.)
Enter key• Enter selected
items• Set up input
data on the connected instrument.
• Run BT200 functions.
F0305.ai
Menu(Max: 6 lines)
Menu screen
Screen title Function commands
Parameter(Max: 3 items)
Function commands
Parameter screen
MENUA:DISPLAYB:EASY SETUPC:BASIC SETUPD:TOTAL SETE:PULSE OUTF:STATUS I/OHOME SET ADJ ESC
PARAMA10:FLOW RATE(%) 0.0%A20:FLOW RATE 0.00000m3/hA21:FLW RATE(mA) 0.000mADATA DIAG PRNT ESC
<3.OperationwithBRAINConfigurationTool> 24
IM 01E22C02-01EN
3.2.2 Key DescriptionsUsethealphanumerickeysinconjunctionwiththeshiftkeystoenternumbers,symbols,andalphabeticcharacters.
F0306.ai
(1) Entering Numbers, Spaces and HyphenSimplypressthealphanumerickeystoinputkeys.
Entry Key-in sequence
–4
0.3
1 –9F0307.ai
(2) Entering Alphabetic CharactersPresseithertheleftorrightshiftkeyandthenanalphanumerickeytoenterthedesiredalphabeticcharacter.
Theshiftkeymustbepressedeachtimeanalphabeticcharacterisentered.
Entry
F0308.ai
(3) Entering Uppercase and LowercaseThecasetogglesbetweenuppercaseandlowercaseeachtime“F2”[CAPS]ispressed.
Entering lowercase CODE caps CLR ESC
Entering uppercase CODE CAPS CLR ESC
Entry
F0309.ai
<3.OperationwithBRAINConfigurationTool> 25
IM 01E22C02-01EN
(4) Entering SymbolsPressthefunctionkey“F1”[CODE]toselectthefollowingsymbols.Thefollowingsymbolswillappearinsequence,oneatatime,atthecursoreachtime“F1”[CODE]ispressed:/.−,+*)(’&%$#"!Toentercharactersnexttothesesymbols,pressthemovementkeytomovethecursor.
Entry Key-in Sequence
F0310.ai
3.2.3 Function KeysThefunctionsofthefunctionkeysdependonfunctioncommandsdisplayedonthescreen.
F0311.ai
Function commands
Function keys
MENUA:DISPLAYB:EASY SETUPC:BASIC SETUPD:TOTAL SETE:PULSE OUTF:STATUS I/OHOME SET ADJ ESC
Table 3.2 Function CommandCommand Meaning
ADJ DisplaystheADJmenuCAPS/caps ChangesthecasemodeCODE Selectssymbol/signCLR ErasesinputdataordeletesalldataDATA UpdatesparameterdataDEL DeletesonecharacterDIAG Callstheself-checkpanelESC ReturnstothemostrecentdisplayHOME DisplaysthemenupanelNO Resetting ReturntosettingscreenOK ProceedstothenextpanelPRAM EnterstheparameternumbersetupmodeSET DisplaystheSETmenuSLOT ReturnstotheslotselectionpanelUTIL Callstheutilitypanel*COPY Printsoutparametersondisplay*FEED Paperfeed*LIST Listsallparametersinthemenu*PON/POFF Automaticprintoutmodeonoroffforsettingchangedata*PRNT Changestotheprintmode*GO Startsprinting*STOP Cancelsprinting
*: AvailableonBT200-P00(withprinter).
<3.OperationwithBRAINConfigurationTool> 26
IM 01E22C02-01EN
3.3 Calling Up Menu AddressesCalling-upmenuonthescreenisasfollows.Pressmovementkeystomovethecursor.Byusingthemovementkeyleftandright,anitemcanbemovedbyonescreen.
"HOME" menu screen "SET" menu screen
Utility screen
Function screen
"ADJ" menu screen
[UTIL]
[SET]
[ESC]
[ADJ]
The utility screen contains the following items.1. ID settings for BT2002. Security code settings3. Selecting language of messages4. Display contrast settings for BT2005. Adjusting printout tone (BT200-P00 only)
F0312.ai
Startupscreen
Initial datascreen
Menuscreen
Parameterscreen
Setupscreen
MENUA:DISPLAYB:EASY SETUPC:BASIC SETUPD:TOTAL SETE:PULSE OUTF:STATUS I/OHOME SET ADJ ESC
MENUB:EASY SETUPC:BASIC SETUPD:TOTAL SETE:PULSE OUTF:STATUS I/OG:ANALOG I/OHOME SET ADJ ESC
MENUL:DIAGNOSISM:TESTN:EVENT MANAGEP:PROTECT
HOME SET ADJ ESC
PARAMA10:FLOW RATE(%) 0.0%A20:FLOW RATE 0.00000m3/hA21:FLW RATE(mA) 0.000mADATA DIAG PRNT ESC
PARAMB10:LANGUAGE EnglishB20:VOL DMP A/F 0.0sB21:VOL DMP P/T 0.0sDATA DIAG PRNT ESC
PARAML10:ADHESION CHK EnableL11:ADH LEVEL1 0.10MohmL12:ADH LEVEL2 0.50MohmDATA DIAG PRNT ESC
SETL10:ADHESION CHK Enable <Enable > <Disable >
ESC
FUNC1.MENU2.UPLOAD TO HHT3.DOWNLOAD TO INST4.PRINT ALL DATA
HOME SET ADJ
UTILITY1.ID2.SECURITY CODE3.LANGUAGE SELECT4.LCD CONTRAST5.PRINTER ADJUST
ESC
PARAM01:MODEL 02:TAG NO
03:SELF CHECK GOOD
OK
--WELCOME--BRAIN TERMINAL ID:
check connectionpush ENTER key
UTIL FEED
SETB10:LANGUAGE English <English > <French > <German > <Italian >
ESC
<3.OperationwithBRAINConfigurationTool> 27
IM 01E22C02-01EN
3.4 Parameter Update and Upload/ Download Function
(1) Updating ParametersThedisplayoftheparametersisupdatedautomaticallyeverysevenseconds.Forthefollowingparameters,refertothefollowingtable.
Item Name Item NameA10 FLOWRATE(%) I54 ALMRECORD3A20 FLOWRATE I55 ALMTIME3A21 FLWRATE(mA) I56 ALMRECORD4A30 TOTAL1 I57 ALMTIME4A31 TOTAL2 K02 OPERATETIMEA32 TOTAL3 K03 CURDATEA33 TOTAL1CNT K04 CURTIMEA34 TOTAL2CNT L16 ADHMEASVALA35 TOTAL3CNT L17 ADHSTATUSC40 VELOCITYCHK L25 FLNOISEVALF14 SI3STATE L26 FLNOISESTSG23 AIVALUE L31 CONDUCTIVITYH22 IEXPWRFREQ L40 VPEAKHOLDH23 MESPWRFREQ L41 IEXCOILRH37 MEASTEMP L42 ELECVOLAH38 CORRCTDENS L43 ELECVOLBH42 CALVALUE L44 EMPTYSTSI50 ALMRECORD1 N21 RESTORERSLTI51 ALMTIME1 P20 WRTPROTECTI52 ALMRECORD2 P23 SOFTSEALI53 ALMTIME2
(2) Upload/Download FunctionTheuploadfunctionisusedforcopyingtheparametersofthisproducttoBT200.ThedownloadfunctionisusedforsettingthecopiedparametersintheBT200intootherproduct.Forthefollowingparameters,refertothefollowingtable.
Item Name Item NameB10/J32 LANGUAGE B31/E14 P1RATEVALB20/C12 VOLDMPA/F B32/E17 F1AT0%B21/C13 VOLDMPP/T B33/E18 F1AT100%B22/C32 VOLFUNIT B40/J10 LINE1 SELB23/C35 TIMEUNIT B41/J11 LINE2 SELB24/C36 VOLFSPAN B42/J12 LINE3 SELB30/E13 P1RATEUNIT B50/C51 AUTOZEROEXE
<3.OperationwithBRAINConfigurationTool> 28
IM 01E22C02-01EN
3.5 ParameterConfigurationThissectiondescribesproceduresforsettingparametersusingtheBT200.
IMPORTANTDonotturnofftheproductrightafterchangingparameterwiththeBT200.Ifthisproductisturnedoffbefore30secondsaftersettingtheparameters,thesettingswillnotbestoredcorrectly.Pleasekeeptheproductturnedonforover30secondsaftersettingtheparameters.
IMPORTANTBesuretoenablethewriteprotectfunctiontopreventparametersfrombeingoverwrittenafterfinishingparametersetting.RefertotheInstallationManualforthehardwarewriteprotectfunction,andSection4.14forthesoftwarewriteprotectfunction.
NOTEBeforeupdatinganysetting,refertoChapter5andchecktheparametercontentyouwanttochange.
3.5.1 Setting Example of Select type Parameter (Flow Rate Unit)
With[B22:VOLFUNIT]oftheselecttypeparameter,changetheflowrateunitfromm3tol(liter).
(1) Movefrom[B:EASYSETUP]to[B22:VOLFUNIT].(2) Movetothesetupscreen,selecttheflowrateunit.Movethecursorto[l(liter)],andpress
theEnterkey. Whentheselectedunitblinks,presstheEnterkeytodetermine.
F0313.ai
SETB22:VOL F UNIT m3 <Kl(kiloliter) > <l(liter) > <cm3 > <kcf >
ESC
(3) Theflowrateunitisnowchangedto“l(liter)”.Pressthefunctionkey“F4”[OK],andthenthescreenisbacktotheparameterscreen.Pressthefunctionkey“F3”[NO],andthenthescreenisbacktothesetupscreen.
F0314.ai
SETB22:VOL F UNIT l(liter)
FEED NO OK
<3.OperationwithBRAINConfigurationTool> 29
IM 01E22C02-01EN
3.5.2 Setting Example of Numeric Type Parameter (Flow rate span)
With[B24:VOLFSPAN]ofthenumeric-typeparameter,changetheflowratespanfrom“100.000l/min”to“120.000l/min”.
(1) Movefrom[B:EASYSETUP]to[B24:VOLFSPAN].(2) Movetothesetupscreen,andentertheflowspanvalue.Enterthevalue“120.000”,and
presstheEnterkey.Whentheenteredvalueblinks,presstheEnterkeytodetermine.
F0315.ai
SETB24:VOL F SPAN 100.000l/min +120.000
DEL CLR ESC
(3) Theflowspanisnowchangedto“120.000l/min”.Pressthefunctionkey“F4”[OK],andthenthescreenisbacktotheparameterscreen.Pressthefunctionkey“F3”[NO],andthenthescreenisbacktothesetupscreen.
F0316.ai
SETB24:VOL F SPAN 120.000l/min
FEED NO OK
3.5.3 Setting Example of Alphanumeric type Parameter (Tag No.)
With[K01:TAGNO]ofthealphanumerictypeparameter,changeTagNo.from“F-123”to“F-456”.
(1) Moveto[K:DEVICEINFO]->[K01:TAGNO].(2) Movetothesetupscreen,andentertheTagNo.Enterthealphanumericcharacter“F-456”,
andpresstheEnterkey.Whenenteredthealphanumericcharacterblinks,presstheEnterkeytodetermine.
F0317.ai
SETK01:TAG NO F-123 F-456
CODE CAPS CLR ESC
(3) TheTagNo.isnowchangedto“F-456”.Pressthefunctionkey“F4”[OK],andthenthescreenisbacktotheparameterscreen.Pressthefunctionkey“F3”[NO],andthenthescreenisbacktothesetupscreen.
F0318.ai
SETK01:TAG NO F-456
FEED NO OK
<4. Functions> 30
IM 01E22C02-01EN
4. FunctionsThischapterdescribesthefunctionsoftheproduct.Followingisanoverviewofeachfunction.
�� Basic settingsThisproductcanmeasuretheprocessvaluesoftheflowvelocity,volumetricflowrate,massflowrate,calorie,andflownoiseatthesametime.Adampingtimeconstantcanbespecifiedforeachprocessvalue.Fordetailsabouthowtocheckmeasuredresultsandthesettingprocedure,refertoSection4.1.
�� Totalization functionThisproducthasthreetotalizersfortotalizingselectedprocessvalues.Inadditiontothetotalizedvaluedisplayfunction,theproducthasatotalizationcounterfunctiontoscaletotalizedvalueswiththeconversionfactorandtocountaspecificflowrate.Theproductalsohasatotalizationswitchfunctionthatcomparesaspecifiedtargetvaluewithatotalizedvaluetooutputtheresultintheformofthestatusoutput,andatotalizationpresetfunctionthattotalizationstartsfromavaluesetinadvance.Fordetailsaboutthetotalizationfunctionandsettingprocedures,refertoSection4.2.
�� Pulse output, Frequency output, and Status outputAmeasurementresultcanbeoutputaspulseoutput,frequencyoutput,orstatusoutput.Whenapulseoutputisused,thepulsewidthorpulseratecanbeselected.Whenafrequencyoutputisused,outputsat0%and100%forthespanoftheprocessvaluecanbespecified.Whenastatusoutputisused,theproductstatuscanbeoutputasastatusoutput.Forbothpulseoutputandfrequencyoutput,thelow-cutvaluescanbeset.Fordetailsabouteachoutputandthesettingprocedure,refertoSection4.3.
�� Status inputSincethisproducthasastatusinputterminal,thetotalizerpresetfunctionandzeroadjustmentfunction,etc.canbeusedinresponsetothestatusinputfromtheoutside.Fordetailsaboutthesettingprocedureofthesefunctions,refertoSection4.4.
�� Analog output/Analog inputUptotwoanalogoutputsareavailable.Thehigh/lowlimitfunction,forward/reverseflowratefunction(reverseflowrate:4to12mA,forwardflowrate:12to20mA),alarmoutputfunction,low-cutfunctionandotherfunctionsareavailable.Analoginputisusefulforinputtingfluidtemperatureviaanexternaltemperaturetransmitter.Fordetailsaboutsettingproceduresfortheanalogoutputandanaloginput,refertoSection4.5.
�� Multi range functionThemultirangefunctionmakesitpossibletomeasureaflowratewhileswitchingmultipleranges.Arangecanbeswitchedinresponsetoaflowrate,flowdirection,andstatusinput.Fordetailsaboutthemultirangefunction,refertoSection4.6.
�� Auxiliary calculation functionThisfunctionisusedtocalculatethetemperaturecorrectionfordensityorcaloriebyinputtingthetemperaturefromanexternaldeviceintheformofananaloginput.Theaccuracyofmassflowratemeasurementisimprovedbysettingthetemperaturecorrectionfordensity.Fordetailsabouttheauxiliarycalculationfunction,refertoSection4.7.
<4. Functions> 31
IM 01E22C02-01EN
�� AlarmAdetectederrorcanbenotifiedasanalarmorwarning.ThisfunctioncanshowtheerrorstatusbasedinaccordancewithNAMURNE107tosuitparametersettings.Itisalsopossibletorecordalarmsthatoccurredinthepastinthehistoryandmaskunnecessarysothattheyarehiddenonthedisplay.Fordetailsaboutcontentsandsettingprocedures,refertoSection4.8.
�� DisplayThisdisplaysupportsmultiplelanguagesforuseonthedisplay.Thisfunctioncanalsoshowthetimechangesoftheselectedparameterinatrendgraphonthedisplay.Fordetailsaboutdisplaysettings,refertoSection4.9.
NOTEThelanguageonthedisplayissetto“English”asdefaultatfactoryshipment.Ifnecessary,refertoSubsection4.9.1toselecttheadequatelanguage.Themenupassofthedisplayonthismanualissetto“English”.
�� Device informationWiththisfunction,theparametersspecifiedatthetimeoforder,modelcode,andsuffixcodeofthisproductcanbecheckedonthedisplay.Fordetailsabouthowtocheckthedeviceinformation,refertoSection4.10.
�� Self-diagnostic functionTheself-diagnosticfunctioncanbeusedtodiagnosefailuresoftheproductortheprocessstate.Forexample,thisfunctionisusefulfordiagnosingthehealthoftheproductbyusingtheelectrodeadhesiondetectingfunction,sensoremptycheckfunction,orverificationfunction.Fordetailsaboutthevariousself-diagnosticfunctions,refertoSection4.11.
�� Test modeWiththismode,theprocessvalueandthevalueoutputfromaconnectionterminalcanbearbitrarilyspecifiedtotestaresponsefromthedevice.Fordetailsaboutthetestmode,refertoSection4.12.
�� Backup, Restore, and Duplicate of ParameterThebackupfunctioncanstoresettingsoftheparametersintothebuilt-inmemoryinthedisplay.IftheoptionalcodeMCisselected,settingparametersarestoredinthemicroSDcardsuppliedwiththisproductinadditiontothebuilt-inmemoryinthedisplay(displayboard).Thebackupdatacanbeusedtorestoresettingsoftheproductforwhichdataisbackeduporduplicatesettingstoanotherproduct.Fordetailsaboutthebackup,restore,andduplicatefunctions,refertoSection4.13.
�� Software write protection functionAwriteprotectcanbechangedusingthehardwarewriteprotectionswitchandtheparametersettings(softwarewriteprotection).Fordetailsaboutthehardwarewriteprotectionswitch,refertotheInstallationManual.Fordetailsaboutthesoftwarewriteprotectionfunction,refertoSection4.14.
<4. Functions> 32
IM 01E22C02-01EN
4.1 Basic Settings4.1.1 Overview
Thisproductcanmeasureflowvelocity,volumetricflowrate,massflowrate,calorie,andflownoiseatthesametime.Ameasuredresultcanbeoutputasananalogoutput,frequencyoutput,pulseoutput,andstatusoutput.Refertothetablebelowforcodesandconnectionterminalsofcommunication/input-outputandinputandoutputforeachterminal.
Communicationand I/O Code Connection Terminal
BRAIN Iout1 +/- P/Sout1 +/- ALM +/- So11 +So12 +
Si11 +Si12 + MIO +/-
D0
Iout1(Active)
P/Sout1(Passive)
AlarmOutput(Passive)
Sout1,2(Passive)
Sin1,2(No-
voltage)
-
D2 Iin(Active)
D3 P/Sout2(Passive)
D4P/Sout2(Active,
withoutinternalresistor)
D5P/Sout2(Active,
withinternalresistor)
D6 Iout2(Active)
Iout1:Analogoutput1(BRAINcommunicationsignalsuperimposed)Iout2:Analogoutput2Iin:AnaloginputP/Sout1:Pulse/Statusoutput1P/Sout2:Pulse/Statusoutput2Sout1,2:Statusoutput1,2Sin1,2:Statusinput1,2ThepositionofCommunicationandI/Ocode:AXG1A-□□□□□□□■■□□
NOTEAvailablefunctionsvarydependingontheconnectionterminaltypespecifiedatthetimeoforder.Beforeusingeachfunctionofthisproduct,besuretochecktheterminaloftheproductwiththetableabove.
<4. Functions> 33
IM 01E22C02-01EN
4.1.2 PV Mapping of Process ValueTheflowvelocity,volumetricflowrate,massflowrate,andflownoisecanbemappedtotheprimaryvariable(PV).Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Provar►PVflowselectBRAIN C:BASICSETUP►C30:PVFLOWSEL
Selection DescriptionDisplay BRAINVelocity Velocity SpecifiestheflowvelocitytothePrimaryValue.
Volume Volume SpecifiesthevolumetricflowratetothePrimaryValue.
Mass Mass SpecifiesthemassflowratetothePrimaryValue.Diag Diag SpecifiestheflownoisetothePrimaryValue.
Settingexample:IfthevolumetricflowrateissettoPV,andthevolumetricflowratespanissetto100m3/hforuse,settheparametersasfollows. C30:PVFLOWSEL=“Volume” C32:VOLFUNIT=“m3” C35:TIMEUNIT=“/h” C37:VOLFSPAN=“100”
Settingexample:IfthemassflowrateissettoPV,withthemassflowratespanbeingsetto10,000kg/handthedensityto1000kg/m3foruse,settheparametersasfollows. H31:DENSITYUNIT=“kg/m3” H32:FIXEDDENS=“1000” C30:PVFLOWSEL=“Mass” C33:MASSFUNIT=“kg” C35:TIMEUNIT=“/h” C38:MASSFSPAN=“10000”
4.1.3 Display of the Process ValueTheflowvelocity,volumetricflowrate,massflowrate,totalizedvalue,calorie,andflownoisecanbecheckedwiththefollowingparameters.
�� Flow rate (PV)Menu path
Display Devicesetup►Processvariables►(seebelow)BRAIN A:DISPLAY►(seebelow)
Parameter DescriptionDisplay BRAIN
Flowrate(%) FLOWRATE(%) DisplaysthepercentagetotherangefortheprocessvaluesettothePrimaryValue.
Flowrate FLOWRATE DisplaystheprocessvaluesettothePrimaryValue.
<4. Functions> 34
IM 01E22C02-01EN
�� Flowvelocity,Volumetricflowrate,Massflowrate,Totalizedvalue,Calorie
Menu pathDisplay Devicesetup►Processvariables►(seebelow)BRAIN A:DISPLAY►(seebelow)
Parameter DescriptionDisplay BRAINVelocity (*) Displaystheflowvelocity.Volume (*) Displaysthevolumetricflowrate.Mass (*) Displaysthemassflowrate.Totalizer►Totalizer1 TOTAL1 Displaysthetotalizedvalueoftotalizer1.Totalizer►Totalizer2 TOTAL2 Displaysthetotalizedvalueoftotalizer2.Totalizer►Totalizer3 TOTAL3 Displaysthetotalizedvalueoftotalizer3.Calorie (*) Displaysthecalorie.
(*) OnlytheprocessvalueselectedwithC30:PVFLOWSELcanbecheckedfromBRAINcommunication.
�� Flow noiseMenu path
Display Devicesetup►Diag/Service►Diagnosis►Flownoise►Result►ValueBRAIN L:DIAGNOSIS►L25:FLNOISEVAL
Thisparameterdisplaystheflownoise.
4.1.4 Unit SettingUnitscanbespecifiedfortheflowvelocity,volumetricflowrate,massflowrate,andcalorie.Eachflowrateunitcanbespecifiedwiththephysicalunitandthetimeunit,individually.However,notethatthetimeunitiscommontoallflowrates.Also,theunitoftheflowvelocityisfixedto“cm/s”anditcannotbechanged.
Example)Tosetthevolumetricflowrateunitto“m3/h”Thevolumetricunitcanbesetto“m3/h”andthetimeunitto“/h”.Atthistime,thetimeunitforthemassflowrateandcaloriearealsosetto“/h”.
Thissettingcanbeconfiguredwiththefollowingparameters.
�� Physical unitMenu path
Display Devicesetup►Detailedsetup►Provar►(seebelow)BRAIN C:BASICSETUP►(seebelow)
Parameter DescriptionDisplay BRAINVelocity►Unit C31:VELOFUNIT Specifiesthephysicalunitoftheflowvelocity.
Volume►Unit C32:VOLFUNIT Specifiesthephysicalunitofthevolumetricflowrate.
Mass►Unit C33:MASSFUNIT Specifiesthephysicalunitofthemassflowrate.Calorie►Unit C34:CALUNIT Specifiesthephysicalunitofthecalorie.
<4. Functions> 35
IM 01E22C02-01EN
�� Time unitMenu path
Display Devicesetup►Detailedsetup►Provar►(seebelow)BRAIN C:BASICSETUP►(seebelow)
Parameter DescriptionDisplay BRAINVolume►Timeunit C35:TIMEUNIT Specifiesthetimeunitofthevolumetricflowrate.Mass►Timeunit C35:TIMEUNIT Specifiesthetimeunitofthemassflowrate.Calorie►Timeunit C35:TIMEUNIT Specifiesthetimeunitofthecalorie.
4.1.5 Span SettingAspancanbespecifiedfortheflowvelocity,volumetricflowrate,massflowrate,calorie,andflownoise.However,theunitforthespanisthesameastheunitspecifiedinSubsection4.1.4.Iftheunitischanged,thespanvalueisconvertedtomatchthechangedunit.Thissettingcanbeconfiguredwiththefollowingparameters.
�� Flowvelocity,Volumetricflowrate,Massflowrate,CalorieMenu path
Display Devicesetup►Detailedsetup►Provar►(seebelow)BRAIN C:BASICSETUP►(seebelow)
Parameter DescriptionDisplay BRAINVelocity►Span C36:VELOFSPAN Specifiesthespanoftheflowvelocity.Volume►Span C37:VOLFSPAN Specifiesthespanofvolumetricflowrate.Mass►Span C38:MASSFSPAN Specifiesthespanofthemassflowrate.Calorie►Span C39:CALFSPAN Specifiesthespanofthecalorie.
�� Flow noiseMenu path
Display Devicesetup►Diag/Service►Diagnosis►Flownoise►SpanBRAIN L:DISGNOSIS►L28:FLNOISESPN
Thisparameterspecifiesthespanoftheflownoise.
NOTEBesuretonotethefollowingpointswhenspecifyingtheflowratespan.• Foralinewithasignificantflowchange,settheflowratespantothemaximumflowrate.If
theflowrateexceedstheflowratespan,theerroroftheflowrate%increases.• Foralinewithastableflowrate,settheflowratespantoapproximately1.5to2.0timesthe
normalflowrate.• Settheflowrateforwhichtheflowvelocityiscomparabletotherangefrom0.3to10
m/s.Theflowvelocitycanbecheckedusingthesizingdatadescribedinthegeneralspecifications.Whentheflowvelocityischeckedusingtheparameter,itdisplaysthevalueobtainedbyconvertingthespecifiedflowratespantotheflowvelocity.
<4. Functions> 36
IM 01E22C02-01EN
NOTEBesuretosettheflowrateunitfirstwhenthespanvalueanditsflowrateunitarechangedatthesametime.
4.1.6 Damping Time Constant SettingThedampingtimeconstant(63.2%response)canbespecifiedfortheflowvelocity,volumetricflowrate,massflowrate,calorie,andflownoise.Toreduceanoutputfluctuationorchangetheresponsespeed,changethedampingtimeconstant(defaultvalueis3.0seconds).Itispossibletomeasureapulsingflowofupto1Hzwithanoutputdampingof0.1secondsforapistonpump,etc.Thedampingtimeconstantcanbesetfortheoutputofeachprocessvalue.Thedampingtimeconstantisacommonsettingfortheanalogoutput,frequencyoutput,pulseoutput,andtotalizedvalue.However,thedampingsettingfortheflownoiseiscommontoalloutputs.Thissettingcanbeconfiguredwiththefollowingparameters.
�� Analog output/Frequency outputMenu path
Display Devicesetup►Detailedsetup►Provar►(seebelow)BRAIN C:BASICSETUP►(seebelow)
Parameter DescriptionDisplay BRAIN
Velocity►DampAO/F C10:VELODMPA/F Specifiesthedampingtimeconstantoftheflowvelocity.
Volume►DampAO/F C12:VOLDMPA/F Specifiesthedampingtimeconstantofthevolumetricflowrate.
Mass►DampAO/F C14:MASSDMPA/F Specifiesthedampingtimeconstantofthemassflowrate.
Calorie►DampAO/F C16:CALDMPA/F Specifiesthedampingtimeconstantofthecalorie.
�� Pulse output/TotalizerMenu path
Display Devicesetup►Detailedsetup►Provar►(seebelow)BRAIN C:BASICSETUP►(seebelow)
Parameter DescriptionDisplay BRAIN
Velocity►Damppls/ttl C11:VELODMPP/T Specifiesthedampingtimeconstantoftheflowvelocity.
Volume►Damppls/ttl C13:VOLDMPP/T Specifiesthedampingtimeconstantofthevolumetricflowrate.
Mass►Damppls/ttl C15:MASSDMPP/T Specifiesthedampingtimeconstantofthemassflowrate.
Calorie►Damppls/ttl C17:CALDMPP/T Specifiesthedampingtimeconstantofthecalorie.
<4. Functions> 37
IM 01E22C02-01EN
�� Flow noiseMenu path
Display Devicesetup►Diag/Service►Diagnosis►Flownoise►DampBRAIN L:DIAGNOSIS►L27:FLNOISEDMP
Thisparameterspecifiesthedampingtimeconstantoftheflownoise.
NOTETheoutputfluctuationincreasesifthedampingtimeconstantissettoalowervalue.Setthedampingtimeconstantto5secondsorlongerforcontrolprocessingapplication.
4.1.7 Low-cut Function SettingAlow-cutvaluecanbespecifiedfortheanalogoutput,frequencyoutput,pulseoutput,andtotalizeroutputvalue.Ifthelow-cutfunctionisused,theflowratebelowsetvaluescanbestoppedfrombeingoutput.Thisfunctionhelpsreduceerroneousoutputwhentheflowis“0”.However,theunitofthelow-cutvalueisthesameastheunitspecifiedinSubsection4.1.4.Iftheunitischanged,thelow-cutvalueischangedtomatchthechangedunit.Thelow-cutvalueisnotsynchronizedwiththespanvalue.Ifthespanvalueischanged,besuretoresetthelow-cutvalue.Setthelow-cutvalueto“0”ifthelow-cutfunctionisnotused.Thissettingcanbeconfiguredwiththefollowingparameters.
�� Analog outputMenu path
Display Devicesetup►Detailedsetup►Analogout/in►(seebelow)BRAIN G:ANALOGI/O►(seebelow)
Parameter DescriptionDisplay BRAIN
AO1►Lowcut G01:AO1LOWCUTSpecifiesthelow-cutvalueofanalogoutput1.IfPVissetforthemeasurementscreenonthedisplay,aPVvaluewhichislow-cutwiththevaluesethereisdisplayed.
AO2►Lowcut G11:AO2LOWCUT Specifiesthelow-cutvalueofanalogoutput2.
�� Frequency output/Pulse outputMenu path
Display Devicesetup►Detailedsetup►Pulse/Statusout►(seebelow)BRAIN E:PULSEOUT►(seebelow)
Parameter DescriptionDisplay BRAIN
PO1/SO1►Lowcut E15:P1LOWCUT Specifiesthelow-cutvaluesoffrequencyoutput1andpulseoutput1.
PO2/SO2►Lowcut E37:P2LOWCUT Specifiesthelow-cutvaluesoffrequencyoutput2andpulseoutput2.
<4. Functions> 38
IM 01E22C02-01EN
�� TotalizerMenu path
Display Devicesetup►Detailedsetup►Totalizer►(seebelow)BRAIN D:TOTALSET►(seebelow)
Parameter DescriptionDisplay BRAINTotalizer1►Lowcut D12:TOT1LOWCUT Specifiesthelow-cutvalueoftotalizer1.Totalizer2►Lowcut D22:TOT2LOWCUT Specifiesthelow-cutvalueoftotalizer2.Totalizer3►Lowcut D32:TOT3LOWCUT Specifiesthelow-cutvalueoftotalizer3.
Theactualvalueforwhichthelow-cutiseffectivehas±0.5%ofthehysteresisfromthesetlow-cutvalue.Thehysteresiswidthonthenegativeside(goingintothelow-cutWhenflowrateisdecreasing)andthepositiveside(goingoutoflow-cut.Whenflowrateisincreasing)areasfollows:(1) Negativeside
=Low-cutsetvalue–(Minimumspansetwithmultirange×0.5%)
(2) Positiveside
=Low-cutsetvalue+(Minimumspansetwithmultirange×0.5%)
Example:Whensettingtospanofvolumetricflowrate=10.0m3/h, Low-cutvalue=1.0m3/h
(1) Negativeside
=1.0[m3/h]–(10.0[m3/h]×0.5[%])
=0.95[m3/h]
(2) Positiveside
=1.0[m3/h]+(10.0[m3/h]×0.5[%])
=1.05[m3/h]
Inthiscase,whentheflowrateisdecreasing,theoutputflowratevaluebecomes0.0[m3/h]withthelow-cutfunctioniftheactualflowrategoesbelow0.95[m3/h].Ontheotherhand,whentheflowrateisincreasing,theflowratevalueisoutputiftheactualflowrategoesover1.05[m3/h].
F0401.ai0
Output [m3/h]
1
Hysteresis1 %
Input [m3/h]1
(1)
(2)
1.050.95
<4. Functions> 39
IM 01E22C02-01EN
NOTENotethatthetotalizationmightbecountedduetotheinfluenceoftheoutputfluctuationnear0%outputifasmalllow-cutvalueisset.Inparticular,ifthevalueoftheflowratespan,dampingtimeconstant,orconductivityislow,thetotalizationiseasilycountedwhentheflowrateis“0”.Insuchacase,increasetheflowratespan,dampingtimeconstant,orlow-cutvalue.
NOTEWhenchangingtheoutputprocessvaluetobeoutput,itisnecessarytospecifythelow-cutvalueagain.
4.1.8 Meter factor SettingWhenorderingonlyAXG1A,orwhenchangingtheremotesensortocombinewith,themeterfactoroftheremotesensormustbespecified.Thissettingcanbeconfiguredwiththefollowingparameters.WhenthecustomerordersAXG1Aandtheremotesensoratthesametimeasacombination,themeterfactoroftheremotesensortocombinewithissettoAXG1A.So,thecustomerisnotrequiredtosetthemeterfactor.Menu path
Display Devicesetup►Detailedsetup►Sensor►(seebelow)BRAIN C:BASICSETUP►(seebelow)
Parameter DescriptionDisplay BRAINLowMF C20:LOWMF Specifiesthemeterfactorofthelow-frequencyside.
HighMF C21:HIGHMF Specifiesthemeterfactorofthehigh-frequencyside.
NOTEThemeterfactorisdescribedintheMETERFACTORcolumnontheremotesensornameplate.
Example:
F0445.ai
<4. Functions> 40
IM 01E22C02-01EN
4.1.9 Sensor Type SettingWhenonlyAXG1Aisordered,orwhentheremotesensortocombinewith,thetypeoftheremotesensormustbespecified.Thissettingcanbeconfiguredwiththefollowingparameters.WhenthecustomerordersAXG1Aandtheremotesensoratthesametimeasacombination,thetypeoftheremotesensortocombinewithissettoAXG1Aatthefactorybeforeshipment.So,thecustomerisnotrequiredtosetthesensortype.Menu path
Display Devicesetup►Detailedsetup►Sensor►FlowsensorselBRAIN C:BASICSETUP►C24:FLOWSNSOR
Fromthetablebelow,selectthesensortype.Selection DescriptionDisplay BRAIN
ADMAGAXG ADMAGAXG ConfiguretousetheADMAGAXGSeriesremotesensor.
ADMAGAXW ADMAGAXW ConfiguretousetheADMAGAXWSeriesremotesensor.
ADMAGAXF ADMAGAXF ConfiguretousetheADMAGAXFSeriesremotesensor.
ADMAG ADMAG ConfiguretousetheADMAGSeriesremotesensor.
ADMAGAE ADMAGAE ConfiguretousetheADMAGAESeriesremotesensor.
ADMAGSE ADMAGSE ConfiguretousetheADMAGSESeriesremotesensor.
YEWMAG YEWMAG ConfiguretousetheYEWMAGSeriesremotesensor.
Calibrator Calibrator ConfiguretousetheAM012(calibrator).Other1 Other1
Configuretouseanotherremotesensor.Other2 Other2Other3 Other3
IMPORTANTWhenchangingthecombinationofthesensors,themeterfactorneedstobere-adjustedbasedonactualflowcalibrationtosecuretheaccuracy.Whenusinganotherremotesensor,suchasothercompanies'products,contactaYokogawasalesofficeorrepresentative.
<4. Functions> 41
IM 01E22C02-01EN
4.1.10 Sensor’s Nominal Size SettingWhenonlyAXG1Aisordered,orwhenchangingtheremotesensortocombine,thenominalsizeoftheremotesensormustbespecified.Thissettingcanbeconfiguredwiththefollowingparameters.WhenthecustomerordersAXG1Aandtheremotesensoratthesametimeasacombination,thenominalsizeoftheremotesensortocombinewithissettoAXG1Aatthefactorybeforeshipment.So,thecustomerisnotrequiredtosetthenominalsize.Menu path
Display Devicesetup►Detailedsetup►Sensor►(seebelow)BRAIN C:BASICSETUP►(seebelow)
Parameter DescriptionDisplay BRAINNominalsizeunit C26:SIZEUNIT Specifiestheunitofthenominalsize.Nominalsize C27:NOMINALSIZE Specifiesnominalsize.
4.1.11 Density SettingThedensitymustbesetinordertomeasurethemassflowrate.Thedensitycanbeselectedfromthedensitycompensatedwithfixeddensityordensitycorrectedbytemperature.Fordetailsabouthowtomeasurethemassflowrateorcorrectthedensitybytemperature,refertoSubsection4.7.5.Ifdensityissetto“0”whilethemassflowrateismappedtoPV,asettingerrorwillresult.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Provar►Density►(seebelow)BRAIN H:AUXCALC►(seebelow)
Parameter DescriptionDisplay BRAIN
Valueselect H30:DENSITYSELSelectsoneofthetemperature-baseddensitycorrectionsfromthefollowing.• Fixedvalue :Usesthefixeddensity• Correctionvalue:Usesthecorrecteddensity
Unit H31:DENSITYUNIT Specifiestheunitofthedensity.Fixeddensity H32:FIXEDDENS Specifiesthevalueofthefixeddensity.
Std density H33:STDDENSITYSpecifiesthereferencedensityvalueofthestandardconditionwhenusingthedensitycompensatedwithtemperature.
Correctdensity H38:CORRCTDENS Displaysthecorrecteddensity.
<4. Functions> 42
IM 01E22C02-01EN
4.1.12 Temperature SettingThetemperaturesettingisrequiredwhencalculatingthecaloriebasedonthisvalue’sdifferencefromthetemperaturethatisinputviatheanaloginputprocess.Fordetailsaboutthecaloriecalculation,refertoSubsection4.7.6.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Provar►Temperature►(seebelow)BRAIN H:AUXCALC►(seebelow)
Parameter DescriptionDisplay BRAIN
Stdtemperature H34:STDTEMPSpecifiesthereferencestandardtemperatureforusingthetemperature-baseddensitycorrectionfunction.
Meastemperature H37:MEASTEMP Displaysthetemperaturethatisinputfromanaloginput.
Fixedtemperature H41:CALFIXTEMPSpecifiesthereferencetemperaturetouseforthecaloriecalculationwiththetemperaturedifferencefromthetemperaturewhichisinputfromanaloginput.
4.1.13 Span Flow Velocity DisplayThespanofthePV-mappedprocessvalueinSubsection4.1.2canbedisplayedintheflowvelocityunit.Thisinformationcanbeviewedwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Provar►VelocitycheckBRAIN C:BASICSETUP►C40:VELOCITYCHK
<4. Functions> 43
IM 01E22C02-01EN
4.1.14 User Span FunctionTheuserspanfunctioncandisplayaprocessvalueassignedtoanalogoutput1andanalogoutput2withthecustomer’sdesiredspecialunit.Ifthesameprocessvalueisassignedtobothanalogoutputs,thesettingforanalogoutput1takesprecedence.Thissettingcanbeconfiguredwiththefollowingparameters.
�� Using user spanMenu path
Display Devicesetup►Detailedsetup►UserSpan►(seebelow)BRAIN C:BASICSETUP►(seebelow)
Parameter DescriptionDisplay BRAINUserspanAO1►Select C41:USRSPNSEL1 Specifiestheuserspanfunctiontotheprocess
valueallocatedtoanalogoutput1.UserspanAO2►Select C44:USRSPNSEL2 Specifiestheuserspanfunctiontotheprocess
valueallocatedtoanalogoutput2.
From the table below, select the user span function.Selection DescriptionDisplay BRAIN
No No Notusinguserspanfunction.Yes Yes Usinguserspanfunction.
�� User unit settingMenu path
Display Devicesetup►Detailedsetup►UserSpan►(seebelow)BRAIN C:BASICSETUP►(seebelow)
Parameter DescriptionDisplay BRAIN
UserspanAO1►Unit C42:USRUNIT1 Specifiesthedesiredunitofupto8characterstotheprocessvalueassignedtoanalogoutput1.
UserspanAO2►Unit C45:USRUNIT2 Specifiesthedesiredunitofupto8characterstotheprocessvalueassignedtoanalogoutput2.
<4. Functions> 44
IM 01E22C02-01EN
�� User span settingMenu path
Display Devicesetup►Detailedsetup►UserSpan►(seebelow)BRAIN C:BASICSETUP►(seebelow)
Parameter DescriptionDisplay BRAIN
UserspanAO1►Span C43:USRSPAN1 Specifiesthedesiredspantotheprocessvalueassignedtoanalogoutput1.
UserspanAO2►Span C46:USRSPAN2 Specifiesthedesiredspantotheprocessvalueassignedtoanalogoutput2.
IMPORTANTTheoutputintheuserunitisdisplayedasfollows:Outputinuserunit=(Outputinaphysicalunitoftheprocessvalue/Spaninaphysicalunitoftheprocessvalue)×Userspan
Example:Tosetthespansettingforanalogoutput1to100dl/s,settheparametersasfollowsdueto100dl=10L.Displays100dl/swhenoutputting100%. PV=“Volumeflow” Spanofvolumetricflowrate=“10l/s” Usinguserspanforanalogoutput1=“Yes” Userunitforanalogoutput1=“dl/s” Userspanforanalogoutput1=“100”
<4. Functions> 45
IM 01E22C02-01EN
4.1.15 Zero AdjustmentThezeroadjustmentiscarriedouttosettheoutputforzeroflowvelocityto0%(4mA).Althoughtheadjustmenttozeroisperformedatthemanufacturingfactorypriortoshipment,thisproceduremustbecarriedoutonceagainfollowingtheinstallationofpipingtomatchthemagneticflowmetertoitsoperatingconditions.Thissubsectiondescribesthezeroadjustmentprocedureusingthedisplay.
IMPORTANT• Thezeroadjustmentshouldbecarriedoutbeforetheactualoperation.Notethatsetting
andupdatingotherparameterscannotbecarriedoutduringthezeroadjustment(forapproximately30seconds).
• Thezeroadjustmentshouldbeexecutedonlyafterthesensorisfilledwithmeasuringfluidandthefluidvelocityisreducedtozerobyclosingthevalve.
• Eachtimethemeasuringfluidischanged,besuretoperformthezeroadjustmentforthatchangedfluid.
�� Execution of zero adjustmentZeroadjustmentusingthedisplaycanbeexecutedwiththefollowingparameters.Menu path
Display Devicesetup►Diag/Service►Autozero►Execute►Execute
F0402.ai
Device setup Process variables Diag/Service Easy setup wizard
▲
▼
00 : 00
SET SFT INC
►
►
►
Select“Diag/Service”inaccordancewiththemenupathabove.
F0403.ai
Diag/Service Verification Autozero AO/AI trim
▲
▼
00 : 00
SET SFT INC
►
►
►
Select“Autozero”.
F0404.ai
Autozero Execute Result▲
▼
00 : 00
SET SFT INC►
Select“Execute”.
F0405.ai
Execute Not execute Execute Not execute▲
▼
00 : 00
SET SFT INC
Select“Execute”.
F0406.ai
Execute Execute
00 : 00
SET SFT INC
When“Execute”blinks,touch[SET]toexecute.
F0407.ai
Autozero functionexecuting...
00 : 00
00 : 08SET SFT INC
The time remaining until the end.
Thezeroadjustmentstarts,andtheprogressisdisplayedwiththeremainingtimeandabargraph.Waitforthezeroadjustmenttocomplete.
<4. Functions> 46
IM 01E22C02-01EN
F0408.ai
Autozero Execute Result▲
▼
00 : 00
SET SFT INC►
Afterthezeroadjustmentisfinished,thedisplayreturnsto“Autozero”menu.
ZeroadjustmentbyBRAINcommunicationcanbeexecutedwiththefollowingparameters.Menu path
BRAIN C:BASICSETUP►C51:AUTOZEROEXE
�� ConfirmationofzeroadjustmentresultThezeroadjustmentresultusingthedisplaycanbeconfirmedwiththefollowingparameters.Menu path
Display Devicesetup►Diag/Service►Autozero►Result►Zerovalue
F0409.ai
Result Zero value ▲
▼
00 : 00
SET SFT INC
Fortheresultofthezeroadjustment,select“Result”andthen“Zerovalue”.
F0410.ai
Zero value – 00.018cm/s – 00.018cm/s – 99.999 + 99.999
00 : 00
SET SFT INC
Theresultofzeroadjustmentisdisplayedasontheleftofthescreen.
NOTEWhenthezeroadjustmentresultexceedsthedefinedvalue,thewarning[092:AZwarn]isdisplayed.
ThezeroadjustmentresultusingBRAINcommunicationcanbeconfirmedwiththefollowingparameters.Menu path
BRAIN C:BASICSETUP►C52:ZERO
<4. Functions> 47
IM 01E22C02-01EN
4.2 Totalization Function4.2.1 Totalized Value and Unit Setting
Thisfunctioncantotalizethevolumetricflowrate,massflowrate,andcalorie.Thisproducthasthreetotalizersandcantotalize3processvaluesatthesametime.Totalizer1totalizesthevaluesintheunitofthePV-mappedprocessvalueinSubsection4.1.2.Totalizers2and3totalizetheprocessvalueoftheselectedunit.Thissettingcanbedisplayedandconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Totalizer►(seebelow)BRAIN D:TOTALSET►(seebelow)
Parameter DescriptionDisplay BRAINTotalizer1►Unit D10:TOT1UNIT Displaystheunitoftotalizer1.Totalizer2►Unit D20:TOT2UNIT SpecifiestheunitofTotalizer2.Totalizer3►Unit D30:TOT3UNIT SpecifiestheunitofTotalizer3.
NOTETheflowvelocityandflownoisecannotbetotalized.IfflowvelocityorflownoiseisselectedforthePrimaryValue,totalizer1totalizesthevolumetricflowrateandunitusing“m3”.
4.2.2 Totalized Value Display and Totalization FunctionThetotalizationresultcanbedisplayedwithatotalizedvalueoratotalizedvaluewhichisscaledwiththeconversionfactor(totalizedcountvalue).Whenthetotalizedvalueisscaledwiththeconversionfactor,aspecificflowrateistotalizedin1-countincrements,whichcanbeusedasatotalizercounter.However,notethattheconversionfactorcannotbesetto“0”.Ifthetotalizedvalueonthedisplayexceeds±99999999,themaximumvalueofthedisplayeddigits,thedisplayedvalueisresetto0.Inaddition,ifthevaluesofA30-A35exceedthelimitofthedisplayeddigits,theyareresetto0inthesamewayasthevalueonthedisplay.Thetotalizedvaluecanbedisplayedandconfiguredwiththefollowingparameters.
�� Displaying totalized valueMenu path
Display Devicesetup►Processvariables►Totalizer►(seebelow)BRAIN A:DISPLAY►(seebelow)
Parameter DescriptionDisplay BRAINTotalizer1 A30:TOTAL1 Displaysthetotalizedvalueoftotalizer1.Totalizer2 A31:TOTAL2 Displaysthetotalizedvalueoftotalizer2.Totalizer3 A32:TOTAL3 Displaysthetotalizedvalueoftotalizer3.
<4. Functions> 48
IM 01E22C02-01EN
NOTEIfMainsoftrev(K50:MAINBREV/Mainsoftrev)isR2.01.02orearlier,orIndsoftrev(K52:INDBREV/Indsoftrev)isR2.01.01orearlier,thetotalizedvalueonthedisplayisheldattheupperlimitifitexceeds±99999999,themaximumvalueofthedisplayeddigits.Inaddition,ifthevaluesofA30-A35exceedthelimitofthedisplayeddigits,theyareheldattheupperlimitinthesamewayasthevalueonthedisplay.Fordetailsabouthowtocheckthedevicerevision(MainsoftrevandIndsoftrev),refertoSubsection4.10.2.
�� Setting conversion factor for scalingMenu path
Display Devicesetup►Detailedsetup►Totalizer►(seebelow)BRAIN D:TOTALSET►(seebelow)
Parameter DescriptionDisplay BRAINTotalizer1►Convfactor D11:TOT1CONVFC Specifiestheconversionfactoroftotalizer1.Totalizer2►Convfactor D21:TOT2CONVFC Specifiestheconversionfactoroftotalizer2.Totalizer3►Convfactor D31:TOT3CONVFC Specifiestheconversionfactoroftotalizer3.
�� Display of the totalized count value that is scaled with the conversion factor
Menu pathDisplay Devicesetup►Processvariables►Totalizer►(seebelow)BRAIN A:DISPLAY►(seebelow)
Parameter DescriptionDisplay BRAINTotalizer1count A33:TOTAL1CNT Displaysthescaledtotalizedvalueoftotalizer1.Totalizer2count A34:TOTAL2CNT Displaysthescaledtotalizedvalueoftotalizer2.Totalizer3count A35:TOTAL3CNT Displaysthescaledtotalizedvalueoftotalizer3.
Example:Settheunitoftotalizer2to“m3”andtheconversionfactorto2.->Ifthetotalizedvalueoftotalizer2issetto“10.123m3”,thetotalizedvalueisscaledto“10.123÷2=5”.
NOTEIfMainsoftrev(K50:MAINBREV/Mainsoftrev)isR2.01.02orearlier,orIndsoftrev(K52:INDBREV/Indsoftrev)isR2.01.01orearlier,thetotalizedvalueonthedisplayisheldattheupperlimitifitexceeds±99999999,themaximumvalueofthedisplayeddigits.Inaddition,ifthevaluesofA30-A35exceedthelimitofthedisplayeddigits,theyareheldattheupperlimitinthesamewayasthevalueonthedisplay.Fordetailsabouthowtocheckthedevicerevision(MainsoftrevandIndsoftrev),refertoSubsection4.10.2.
<4. Functions> 49
IM 01E22C02-01EN
4.2.3 Totalization Switch FunctionWhenatargetvalue(threshold)tototalizeisspecified,itcanbeusedasatotalizationswitchfunction.Thetotalizationswitchfunctioncancomparethespecifiedtargetvaluewiththetotalizedvalueandoutputtheresultinstatusoutput.Thestatusoutputisactivewhilethetotalizedvalueexceedsthespecifiedtargetvalue.Evenifthetotalizedvalueexceedsthedisplayeddigitlimitandisresetto0underthatstate,thestatusoutputremainsactive.Fordetailsabouttheoutput,activedirection,andstatusoutputfunctionsettingforeachterminal,refertoSection4.3.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Totalizer►(seebelow)BRAIN D:TOTALSET►(seebelow)
Parameter DescriptionDisplay BRAINTotalizer1►Setpoint D18:TOT1SETPNT Specifiesthetargetvalueoftotalizer1.Totalizer2►Setpoint D28:TOT2SETPNT Specifiesthetargetvalueoftotalizer2.Totalizer3►Setpoint D38:TOT3SETPNT Specifiesthetargetvalueoftotalizer3.
Example: Setting procedure to use the totalization switch function with the pulse/status output 1.
Followthestepsbelowtosetthestatusoutputofpulse/statusoutput1to“Onactive”whenthetotalizedvalueoftotalizer1reachesthetargetvalue.(1) RefertoSubsection4.3.1tospecifytheoutputofpulse/statusoutput1to“Statusoutput”.
(2) RefertoSubsection4.3.5tospecifytheactivedirectionofpulse/statusoutput1to“Onactive”.
(3) RefertoSubsection4.3.9tospecifythestatusoutputfunctionto“Totallimit1”.
(4) Refertothissubsectiontospecifythetargetvalueoftotalizer1.
<4. Functions> 50
IM 01E22C02-01EN
4.2.4 Totalizer Operation at Alarm OccurrenceThetotalizeroperationcanbespecifiedtodealwithanalarmthataffectsthetotalizationfunction.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Totalizer►(seebelow)BRAIN D:TOTALSET►(seebelow)
Parameter DescriptionDisplay BRAIN
Totalizer1►Failureopts D13:TOT1FOPTS Specifiesthetotalizer1operationtobeperformedwhenanalarmoccurs.
Totalizer2►Failureopts D23:TOT2FOPTS Specifiesthetotalizer2operationtobeperformedwhenanalarmoccurs.
Totalizer3►Failureopts D33:TOT3FOPTS Specifiestotalizer3operationtobeperformedwhenanalarmoccurs.
Fromthetablebelow,selecttheoperationofthetotalizationfunction.Selection DescriptionDisplay BRAIN
Measuredvalue Measuredvalue Continuesthetotalizationfunctionwhileanalarmisoccurring.
Stop Stop Stopsthetotalizationfunctionwhileanalarmisoccurring.
Last valid Last validContinuesthetotalizationfunctionwiththelastvalidvaluerightbeforethealarmoccurredwhileanalarmisoccurring.
4.2.5 Start/Stop Setting for Totalization FunctionThestart/stopoperationofthetotalizationfunctioncanbeset.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Totalizer►(seebelow)BRAIN D:TOTALSET►(seebelow)
Parameter DescriptionDisplay BRAIN
Totalizer1►Start/Stop D15:TOT1EXEC SpecifiesStart/Stoptothetotalizationfunctionoftotalizer1.
Totalizer2►Start/Stop D25:TOT2EXEC SpecifiesStart/Stoptothetotalizationfunctionoftotalizer2.
Totalizer3►Start/Stop D35:TOT3EXEC SpecifiesStart/Stoptothetotalizationfunctionoftotalizer3.
NOTEThetotalizationfunctionissetto“Stop”atshipmentfromthemanufacturingfactory.Tostartthetotalizationfunction,besuretosetitto“Start”.
<4. Functions> 51
IM 01E22C02-01EN
4.2.6 Totalization Direction SettingThetotalizationdirectioncanbespecifiedwhenusingthetotalizationfunction.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Totalizer►(seebelow)BRAIN D:TOTALSET►(seebelow)
Parameter DescriptionDisplay BRAINTotalizer1►Options D14:TOT1OPTS Specifiesthetotalizationdirectionoftotalizer1.Totalizer2►Options D24:TOT2OPTS Specifiesthetotalizationdirectionoftotalizer2.Totalizer3►Options D34:TOT3OPTS Specifiesthetotalizationdirectionoftotalizer3.
Fromthetablebelow,selectthetotalizationdirection.Selection DescriptionDisplay BRAIN
Balanced Balanced Totalizesthedifferentialflowratebetweentheforwardandreversedirections.
Absolute Absolute Totalizestheabsolutevalueoftheflowrate.Only positive Only positive Totalizesonlytheflowrateintheforwarddirection.Onlynegative Onlynegative Totalizesonlytheflowrateinthereversedirection.
Hold Hold Stopstotalizationprocessing(holdsthecurrenttotalizedvalue).
<4. Functions> 52
IM 01E22C02-01EN
4.2.7 Totalized Value Reset/Preset FunctionThereset/presetfunctioncanbespecifiedforthetotalizedvalue.Iftheresetfunctionisused,thefunctionresetsthetotalizedvalueto“0”.Ifthepresetfunctionisused,itsetsthepresetvaluespecifiedinadvancetothetotalizedvalue.Thepresetfunctionisusedwhenstartingtocounttotalizationfromthespecifiedvalue.Thissettingcanbeconfiguredwiththefollowingparameters.
�� Use of the reset/preset functionMenu path
Display Devicesetup►Detailedsetup►Totalizer►(seebelow)BRAIN D:TOTALSET►(seebelow)
Parameter DescriptionDisplay BRAINTotalizer1►Reset/Preset D16:TOT1PRESET Usesthereset/presetfunctionoftotalizer1.Totalizer2►Reset/Preset D26:TOT2PRESET Usesthereset/presetfunctionoftotalizer2.Totalizer3►Reset/Preset D36:TOT3PRESET Usesthereset/presetfunctionoftotalizer3.
Fromthetablebelow,selectthereset/presetfunction.Selection DescriptionDisplay BRAIN
Notexecute Notexecute Doesnotusethereset/presetfunctionofthetotalizedvalue.
Reset Reset Usestheresetfunctionofthetotalizedvalue.Preset Preset Usesthepresetfunctionforthetotalizedvalue.
�� Preset value settingMenu path
Display Devicesetup►Detailedsetup►Totalizer►(seebelow)BRAIN D:TOTALSET►(seebelow)
Parameter DescriptionDisplay BRAINTotalizer1►Presetvalue D17:TOT1PREVAL Specifiesthepresetvalueoftotalizer1.Totalizer2►Presetvalue D27:TOT2PREVAL Specifiesthepresetvalueoftotalizer2.Totalizer3►Presetvalue D37:TOT3PREVAL Specifiesthepresetvalueoftotalizer3.
NOTETheparameterreturnsto“Notexecute”afterthereset/presetfunctionofthetotalizedvaluehasbeencompleted.
<4. Functions> 53
IM 01E22C02-01EN
4.3 Pulse Output, Frequency Output, and Status Output
4.3.1 Pulse/Status Output Mode SettingSelecttheoutputmodeofthepulse/status.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Pulse/Statusout►(seebelow)BRAIN E:PULSEOUT►(seebelow)
Parameter DescriptionDisplay BRAINPO1/SO1►Outputmode E10:P1OUTMODE Specifiestheoutputmodeofpulse/statusoutput1.
PO2/SO2►Outputmode E31:P2OUTMODE Specifiestheoutputmodeofpulse/statusoutput2.
Fromthetablebelow,selecttheoutputmode.Selection
DescriptionDisplay BRAIN
Nofunction Nofunction Doesnotusetheterminal.Fixedpulseoutput Fixedpulseoutput Specifiestheoutputtothefixedwidthpulseoutput.
Frequencyoutput Frequencyoutput Specifiestheoutputtothefrequencyoutput(Duty50%).
Statusoutput StatusoutputSpecifiestheoutputtothestatusoutput.Fordetailsaboutthestatusoutput,refertoSubsection4.3.9.
<4. Functions> 54
IM 01E22C02-01EN
4.3.2 Pulse Output/Frequency Output MappingWhenusingasapulseoutputorfrequencyoutput,theprocessvaluetobeoutputneedstobeselected.Forpulse/statusoutput1,aprocessvaluedoesnotneedtobesetbecausethePV-mappedprocessvalueinSubsection4.1.2isanoutputtarget.Forpulse/statusoutput2,theprocessvaluetobeoutputcanbeselectedamongtheflowvelocity,volumetricflowrate,massflowrate,calorie,andflownoise.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Pulse/Statusout►PO2/SO2►PulseselectBRAIN E:PULSESET►E32:P2SELECT
Thisparameterspecifiesaprocessvalueoutputfrompulse/statusoutput2.Fromthetablebelow,selecttheprocessvaluetobeoutput.
Selection DescriptionDisplay BRAINNoconnect Non-Connect Nooutput.Velocity Velocity Specifiestheflowvelocitytotheoutput.Volumeflow Volumeflow Specifiesthevolumetricflowratetotheoutput.Massflow Massflow Specifiesthemassflowratetotheoutput.Calorie Calorie Specifiesthecalorietotheoutput.Diag Diag Specifiestheflownoisetotheoutput.
4.3.3 Alarm Output FunctionItispossibletospecifytheoutputbehaviorforwhenanalarmoccurswhilebeingusedasapulseoutputorfrequencyoutput.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Pulse/Statusout►(seebelow)BRAIN E:PULSEOUT►(seebelow)
Parameter DescriptionDisplay BRAINPO1/SO1►Alarmout E16:P1ALMOUT Specifiesthealarmoutputfunctionofpulseoutput1
andfrequencyoutput1.PO2/SO2►Alarmout E33:P2ALMOUT Specifiesthealarmoutputfunctionofpulseoutput2
andfrequencyoutput2.
Fromthetablebelow,selecttheoutputoperationofthealarmoutputfunction.Selection DescriptionDisplay BRAIN
0 pps 0 pps Nooutput.
Measuredvalue Measuredvalue Outputsthefrequencyorpulsewiththemeasuredvalue.
Last valid Last valid Outputsthefrequencyorpulsewiththelastvalidvaluebeforeanalarmactivates.
Maxpps Maxpps Outputsthefrequencyorpulseat12500pps.
<4. Functions> 55
IM 01E22C02-01EN
4.3.4 Pulse Width SettingThepulsewidthcanbeselectedforuseasafixedwidthpulse.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Pulse/Statusout►(seebelow)BRAIN E:PULSEOUT►(seebelow)
Parameter DescriptionDisplay BRAINPO1/SO1►Fixwidth E12:P1WIDTH Specifiesthepulsewidthofpulseoutput1.
PO2/SO2►Fixwidth E34:P2WIDTH Specifiesthepulsewidthofpulseoutput2.
Fromthetablebelow,selectthepulsewidth.Pulse width [ms] Max. output frequency [Hz]
0.05 100000.1 50000.5 10001 50020 2533 1550 10100 5200 2.5330 1.5500 1.01000 0.52000 0.25Dutycycle50% ―
NOTEThemaximumpulsewidthwhen“Dutycycle50%”isselectedisupto300seconds(5minutes).Tooutputpulsewhoseperiodexceeds10minutes,thepulsewidthisfixedto300seconds.
<4. Functions> 56
IM 01E22C02-01EN
4.3.5 Active Direction SettingWhenthepulseoutputorstatusoutputisused,eitherONorOFFcanbesettoactiveforthepulsesignal.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Pulse/Statusout►(seebelow)
Parameter DescriptionDisplay BRAINPO1/SO1►Activemode
E:PULSEOUT►E11:P1ACTMODE
Specifiestheactivedirectionofthepulsesignalofpulse/statusoutput1.
PO2/SO2►Activemode
E:PULSEOUT►E33:P2ACTMODE
Specifiestheactivedirectionofthepulsesignalofpulse/statusoutput2.
SO11►Activemode F:STATUSI/O►F20:SO11ACTMODE
Specifiestheactivedirectionofthepulsesignalofstatusoutput11.
SO12►Activemode F:STATUSI/O►F30:SO12ACTMODE
Specifiestheactivedirectionofthepulsesignalofstatusoutput12.
Fromthetablebelow,selecttheactivedirectionofthepulsesignal.Selection DescriptionDisplay BRAIN
Onactive Onactive SetstoActivewhenthepulsesignalisseton.Offactive Offactive SetstoActivewhenthepulsesignalissetoff.
Example:Ifthefixedpulseoutputisspecifiedandthepulsewidthissetto“1ms”,itissettoactiveasshownbelow.
F0411.ai
Active:ON
Active:OFF
1 ms
<4. Functions> 57
IM 01E22C02-01EN
4.3.6 Active Pulse SettingWhencommunicationandI/OcodeD4isselected,thesensorwillbesetandshippedsothatthesensoroperateswithpulseoutput2specifiedasamechanicalcounter.Thissettingcanbechangedbythecustomer.Themechanicalcounterspecificationshavethefollowinglimitations.
Maximum pulse rate 2ppsPulse width 20ms,33ms,50ms,100ms
Active direction Onactive
RefertotheInstallationManualonhowtoconnectwiththemechanicalcounter.Menu path
Display Devicesetup►Detailedsetup►Pulse/Statusout►PO2/SO2►ActivepulseBRAIN E:PULSEOUT►E30:P2ACTPULSE
Selection DescriptionDisplay BRAINNormal Normal NormalpulseFormagneticcounter Formagneticcounter Mechanicalcounterspecificationpulse
4.3.7 Pulse Rate SettingWhenthepulseoutputisused,thepulseratevalueperpulseandpulseratescalingcanbespecified.TheunitofthepulserateisthesameastheunitofthePV-mappedprocessvalueinSubsection4.1.2.Thissettingcanbeconfiguredwiththefollowingparameters.Ifthepulseratescalingischanged,thepulseratevalueisalsoconvertedtomatchthechangedunit.
�� Pulse rate valueMenu path
Display Devicesetup►Detailedsetup►Pulse/Statusout►(seebelow)BRAIN E:PULSEOUT►(seebelow)
Parameter DescriptionDisplay BRAINPO1/SO1►Ratevalue E14:P1RATEVAL Specifiesthepulserateofpulseoutput1.
PO2/SO2►Ratevalue E36:P2RATEVAL Specifiesthepulserateofpulseoutput2.
<4. Functions> 58
IM 01E22C02-01EN
�� Scaling pulse rateMenu path
Display Devicesetup►Detailedsetup►Pulse/Statusout►(seebelow)BRAIN E:PULSEOUT►(seebelow)
Parameter DescriptionDisplay BRAINPO1/SO1►Rateunit E13:P1RATEUNIT Specifiesthescalingofthepulserateofpulse
output1.PO2/SO2►Rateunit E35:P2RATEUNIT Specifiesthescalingofthepulserateofpulse
output2.
Fromthetablebelow,selectthepulseratescaling.Selection Description
nunit/P 10-9×unitsperpulseuunit/P 10-6×unitsperpulsemunit/P 10-3×unitsperpulseUnit/P 1unitperpulsekunit/P 103×unitsperpulseMunit/P 106×unitsperpulsenP/unit 10-9×pulsesperunituP/unit 10-6×pulsesperunitmP/unit 10-3×pulsesperunitP/unit 1pulseperunitkP/unit 103×pulsesperunitMP/unit 106×pulsesperunit
Unit:Theunitoftheprocessvaluetobeoutputasapulseoutput
NOTEThemaximumpulserateandpulsewidthmustbespecifiedsothatthefollowingconditionsaresatisfied.Maximumpulseratevalue[pps] =Flowratespan[Unit/s]×Pulserate[P/Unit] ≤10[kpps] ≤1/(Pulsewidth×2)
Example:Whenthepulsewidthissetto“0.1ms”,themaximumpulseratevalueissetto“1/(0.0001×2)=5000[pps]”.Ifthespecifiedpulserateexceedsthisvalue,itcausesasettingerror,andanalarmis displayed.
<4. Functions> 59
IM 01E22C02-01EN
4.3.8 Frequency Output Range SettingWhenthefrequencyoutputisused,thefrequencyat0%and100%canbespecifiedforthespanoftheprocessvalue.Thefrequencyoutputrangecanbesetbyspecifyingthefrequency.TheoutputfrequencyistobesetforthespanofthePV-mappedprocessvalueinSubsection4.1.2.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Pulse/Statusout►(seebelow)BRAIN E:PULSEOUT►(seebelow)
Parameter DescriptionDisplay BRAINPO1/SO1►Frequencyat0% E17:F1AT0% Specifiesthefrequencywhentheprocessvalueto
beoutputwithfrequencyoutput1issetto0%.PO1/SO1►Frequencyat100% E18:F1AT100% Specifiesthefrequencywhentheprocessvalueto
beoutputwithfrequencyoutput1issetto100%.PO2/SO2►Frequencyat0% E39:F2AT0% Specifiesthefrequencywhentheprocessvalueto
beoutputwithfrequencyoutput2issetto0%.PO2/SO2►Frequencyat100% E40:F2AT100% Specifiesthefrequencywhentheprocessvalueto
beoutputwithfrequencyoutput2issetto100%.
<4. Functions> 60
IM 01E22C02-01EN
4.3.9 Status Output Function SettingTheproductstatecanbeoutputasacontactpointtousethestatusoutput.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Pulse/Statusout►(seebelow)
Parameter DescriptionDisplay BRAINPO1/SO1►SO1function
E:PULSEOUT►E20:SO1FUNCTION Specifiesstatusoutput1function.
PO2/SO2►SO2function
E:PULSEOUT►E41:SO2FUNCTION Specifiesstatusoutput2function.
SO11►Function F:STATUSI/O►F21:SO11FUNC Specifiesstatusoutput11function.
SO12►Function F:STATUSI/O►F31:SO12FUNC Specifiesstatusoutput12function.
Fromthetablebelow,selectthestatusoutputfunction.Selection DescriptionDisplay BRAIN
Nofunction Nofunction Thestatusoutputisnotavailablebecausethestatusoutputfunctionisnotenabled.
Warningoutput Warningoutput Thestatusoutputisactivatedwhenawarningoccurs.
Totallimit1 Totallimit1Thestatusoutputisactivatedwhenthetotalizedvalueoftotalizer1exceedsthespecifiedtargetvalue.Fordetailsaboutthetotalizationswitchfunction,refertoSubsection4.2.3.
Totallimit2 Totallimit2Thestatusoutputisactivatedwhenthetotalizedvalueoftotalizer2exceedsthespecifiedtargetvalue.Fordetailsaboutthetotalizationswitchfunction,refertoSubsection4.2.3.
Totallimit3 Totallimit3Thestatusoutputisactivatedwhenthetotalizedvalueoftotalizer3exceedsthespecifiedtargetvalue.Fordetailsaboutthetotalizationswitchfunction,refertoSubsection4.2.3.
H/Lalarm H/Lalarm
ThestatusoutputisactivatedwhenthePV-mappedprocessvalueisequaltoorlowerthanthespecifiedlowlimit(L)orequaltoorgreaterthanthespecifiedhighlimit(H).FordetailsaboutthePVmappingfortheprocessvalueandalarminformation,refertoSubsection4.1.2,Section4.8,andSubsection4.11.2.
HH/LLalarm HH/LLalarm
ThestatusoutputisactivewhilethePV-mappedprocessvalueisequaltoorlowerthanthespecifiedlow-lowlimit(LL)orequaltoorgreaterthanhigh-highlimit(HH).FordetailsaboutthePVmappingfortheprocessvalueandalarminformation,refertoSubsection4.1.2,Section4.8,andSubsection4.11.2.
Fwd/Revrange Fwd/Revrange
Usedforforward/reverserange.Aflowratecanbemeasuredbyswitchinganalogoutput1betweentheforwardrangeandreverserangeinresponsetoaflowdirection.Thestatusoutputisactivatedwhenthepresentrangeisthereverserange.Fordetails,refertoSubsection4.6.4.
Auto2range Auto2range
Usedforthemultirange.Aflowratecanbemeasuredwhileswitchinganalogoutput1betweenrange1andrange2inresponsetotheflowrate.Thestatusoutputisactivatedwhenthepresentrangeisrange2.Fordetails,refertoSubsection4.6.3.
<4. Functions> 61
IM 01E22C02-01EN
Selection DescriptionDisplay BRAIN
Auto3range Auto3range
Usedforthemultirange.Aflowratecanbemeasuredwhileswitchinganalogoutput1betweenrange1andrange3inresponsetotheflowrate.Displaysthepresentrangebycombiningthestateofstatusoutputs11and12.Canbespecifiedonlywithstatusoutputs11and12.Fordetails,refertoSubsection4.6.3.
Auto4range Auto4range
Usedforthemultirange.Aflowratecanbemeasuredwhileswitchinganalogoutput1betweenrange1andrange4inresponsetotheflowrate.Displaysthepresentrangebycombiningthestateofstatusoutputs11and12.Canbespecifiedonlywithstatusoutputs11and12.Fordetails,refertoSubsection4.6.3.
Ext2answer Ext2answerThestatusoutputisactivewhenthepresentrangeisrange2whileoperatingwithExt2rangesettings.Fordetails,refertoSubsection4.6.7.
Ext3answer Ext3answer
Displaysthepresentrangebycombiningthestateofstatusoutputs11and12whileoperatingwithExt3rangesettings.Canbespecifiedonlywithstatusoutputs11and12.Fordetails,refertoSubsection4.6.7.
Ext4answer Ext4answer
Displaysthepresentrangebycombiningthestateofstatusoutputs11and12whileoperatingwithExt4rangesettings.Canbespecifiedonlywithstatusoutputs11and12.Fordetails,refertoSubsection4.6.7.
Systemalarm Systemalarm ThestatusoutputisactivatedwhenSystemalarmoccurs.Canbespecifiedonlywithstatusoutput2.
Processalarm Processalarm ThestatusoutputisactivatedwhenProcessalarmoccurs.Canbespecifiedonlywithstatusoutput2.
Settingalarm Settingalarm ThestatusoutputisactivatedwhenSettingalarmoccurs.Canbespecifiedonlywithstatusoutput2.
Signaloverflow 50:Sigoverflow ThestatusoutputisactivatedwhenSignaloverflowoccurs.Canbespecifiedonlywithstatusoutput2.
Emptydetect 51:EmptydetectThestatusoutputisactivatedwhenEmptypipedetectionoccurs.Canbespecifiedonlywithstatusoutput2.
Adhoverlv4 53:Adhoverlv4ThestatusoutputisactivatedwhenAdhesionoverlevel4occurs.Canbespecifiedonlywithstatusoutput2.
Cablemiscon 85:CablemisconThestatusoutputisactivatedwhenCablemisconnectoccurs.Canbespecifiedonlywithstatusoutput2.
Coilinsulation 86:Coilinsulate ThestatusoutputisactivatedwhenCoilinsulationoccurs.Canbespecifiedonlywithstatusoutput2.
Adhesionlv3 87:Adhoverlv3 ThestatusoutputisactivatedwhenAdhesionlv3occurs.Canbespecifiedonlywithstatusoutput2.
LCwarn 88:LCwarnThestatusoutputisactivatedwhenLowconductivityWarningoccurs.Canbespecifiedonlywithstatusoutput2.
Insudetect 89:InsudetectThestatusoutputisactivatedwhenInsulationdetectionoccurs.Canbespecifiedonlywithstatusoutput2.
FLNoverlv3 90:FLNoverlv3ThestatusoutputisactivatedwhenFlownoiseoverlevel3occurs.Canbespecifiedonlywithstatusoutput2.
FLNoverlv4 91:FLNoverlv4ThestatusoutputisactivatedwhenFlownoiseoverlevel4occurs.Canbespecifiedonlywithstatusoutput2.
Verifwarn 93:VerifwarnThestatusoutputisactivatedwhenVerificationwarningoccurs.Canbespecifiedonlywithstatusoutput2.
<4. Functions> 62
IM 01E22C02-01EN
Selection DescriptionDisplay BRAIN
Factnoisewarn 94:FCnoisewarnThestatusoutputisactivatedwhenFactorynoisewarningoccurs.Canbespecifiedonlywithstatusoutput2.
Lowalarm LoalarmThestatusoutputisactivatedwhenthePV-mappedprocessvalueisequaltoorlessthanthelowlimit(L)andexceedsthelow-lowlimitvalue(LL).Canbespecifiedonlywithstatusoutput2.
LLalarm LoLoalarmThestatusoutputisactivatedwhenthePV-mappedprocessvalueisequaltoorlessthanthespecifiedlow-lowlimitvalue(LL).Canbespecifiedonlywithstatusoutput2.
Highalarm Hialarm
ThestatusoutputisactivatedwhenthePV-mappedprocessvalueisequaltoorgreaterthanthespecifiedhighlimit(H)andnomorethanthehigh-highlimit(HH).Canbespecifiedonlywithstatusoutput2.
HHalarm HiHialarmThestatusoutputisactivatedwhenthePV-mappedprocessvalueisequaltoorgreaterthanthespecifiedhigh-highlimit(HH).Canbespecifiedonlywithstatusoutput2.
<4. Functions> 63
IM 01E22C02-01EN
4.3.10 Pulse Option SettingAninstantaneousflowrateistotalizedforthefixedwidthpulseoutputorfrequencyoutputandisoutputasapulsewhenthetotalizedvalueexceeds1.Withthepulseoption,thetotalizationmethodcanbeselected.
Display Devicesetup►Detailedsetup►Pulse/Statusout►(seebelow)BRAIN E:PULSEOUT►(seebelow)
Parameter DescriptionDisplay BRAINPO1/SO1►Options E21:P1OPTS Specifiespulse1option.PO2/SO2►Options E42:P2OPTS Specifiespulse2option.
Selection DescriptionDisplay BRAIN
Balanced Balanced
Totalizesaflowrateintheforwarddirectionwithapositivevalueandaflowrateinthereversedirectionwithanegativevalue,andoutputsavalueasapulseiftheforwardandnegativetotalizedvaluesexceed1.
Absolute Absolute Totalizesanabsolutevalueoftheflowrateandoutputsitasapulseifthetotalizedvalueexceeds1.
Only positive Only positiveTotalizesonlyaflowrateintheforwarddirectionandoutputsitasapulseifthetotalizedvalueexceeds1.(Thenegativeflowrateisnottotalizedas0)
Onlynegative Onlynegative
Convertsanegativeflowratetoapositivevalueandtotalizesit.Outputsthevalueasapulseifthevalueexceeds1.(Thenegativeflowrateisnottotalizedintheforwarddirectionas0)
Use“Balanced”and“Absolute”inthefollowingcases.
Balanced
Whenraising0%pointwithfrequencyoutputSettingexample)Pulsemode=Freq,Freqat0%=100Hz,Freqat100%=500Hz,Pulseoption=BalancedSince100Hzisoutputat0%withtheabovesettings,aninstantaneousflowratecanbemeasuredupto-25%infrequencyoutput.(Possibletouseinthesamewayastheanalogoutput)
Absolute
Whenusingincombinationwithforward/reverserangeSettingexample)Pulsemode=Fixed,Pulseoption=Absolute,SOfunction=Fwd/RevrangePleasesetyourdesiredvalueforpulsewidth/rate/rateunit,andreversespan1.Withtheabovesettings,thesamepulseoutputcanbereproducedaswhentheforward/reverserangesettingismadewiththeoldmodelAXF.
<4. Functions> 64
IM 01E22C02-01EN
4.4 Status Input4.4.1 Active Direction Setting for Status Input
Whenthestatusinputisused,eitherONorOFFcanbesettoactiveforthestatusinput.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Statusin►(seebelow)BRAIN F:STATUSI/O►(seebelow)
Parameter DescriptionDisplay BRAINSI11►Activemode F22:SI11ACTMODE Specifiestheactivedirectionofstatusinput11.SI12►Activemode F32:SI12ACTMODE Specifiestheactivedirectionofstatusinput12.
Fromthetablebelow,selecttheactivedirectionofthestatusinput.Selection DescriptionDisplay BRAIN
Onactive Short(On)act SpecifiesthestatusinputtoActivewhenitisshorted-circuited.
Offactive Open(Off)act SpecifiesthestatusinputtoActivewhenitisopen.
<4. Functions> 65
IM 01E22C02-01EN
4.4.2 Status Input Function SettingWhenthestatusinputisused,thefunctionofthestatusinputcanbeset.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Statusin►(seebelow)BRAIN F:STATUSI/O►(seebelow)
Parameter DescriptionDisplay BRAINSI11►Function F23:SI11FUNC Specifiesstatusinput11function.SI12►Function F33:SI12FUNC Specifiesstatusinput12function.
Fromthetablebelow,selectthestatusinputfunction.Selection DescriptionDisplay BRAIN
Nofunction Nofunction Thestatusinputisnotavailablebecausethestatusinputfunctionisnotenabled.
0%signallock 0%SignalLock Analogoutput1isfixedat4mAwhenthestatusinputbecomesactive.
Extautozero Extautozero
Thezeroadjustmentfunctionisautomaticallyusedwhenthestatusinputisactive.Notethatzeroadjustmentcannotbestoppedevenifthestatusinputischangedwhilethezeroadjustmentfunctionisused.
Totalpreset1 Totalpreset1ThepresetvaluespecifiedinSubsection4.2.7issetasthevaluefortotalizer1andtotalizationstartsfromthevaluewhenthestatusinputisactive.
Totalpreset2 Totalpreset2ThepresetvaluespecifiedinSubsection4.2.7issetasthevaluefortotalizer2andtotalizationstartsfromthevaluewhenthestatusinputbecomesactive.
Totalpreset3 Totalpreset3ThepresetvaluespecifiedinSubsection4.2.7issetasthevaluefortotalizer3andtotalizationstartsfromthevaluewhenthestatusinputbecomesactive.
Ext2ranges Ext2ranges
Therangeofanalogoutput1isswitchedfromrange1torange2whilethestatusinputisactive.UsedforExternalContactRange.Fordetailsontheexternalcontactrange,refertoSubsection4.6.6.
Ext3ranges Ext3rangesUsedforExternalContactRange.Therangeofanalogoutput1isarbitrarilyswitchedfromrange1to3bycombiningthestateofstatusinput11and12.Fordetails,refertoSubsection4.6.6.
Ext4ranges Ext4rangesUsedforExternalContactRange.Therangeofanalogoutput1isarbitrarilyswitchedfromrange1to4bycombiningthestateofstatusinput11and12.Fordetails,refertoSubsection4.6.6.
0%signallock2 0%SignalLock2Aflowrateisfixedat0whenthestatusinputbecomesactive.Duetothis,theanalogoutput,pulseoutputandflowrateonthedisplaybecome0.
<4. Functions> 66
IM 01E22C02-01EN
4.5 Analog Input and Analog Output4.5.1 Analog Output Select
Theprocessvaluetobeassignedtoanalogoutput2isset.Inthecaseofanalogoutput1,theprocessvaluedoesnotneedtobesetbecausethePV-mappedprocessvalueinSubsection4.1.2isanoutputtarget.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Analogout/in►AO2►AO2selectBRAIN G:AnalogI/O►G10:AO2SELECT
Fromthetablebelow,selecttheprocessvaluefortheanalogoutput.Selection DescriptionDisplay BRAIN
Noconnect Non-Connect Nooutput.Velocity Velocity Specifiestheflowvelocitytotheoutput.Volumeflow Volumeflow Specifiesthevolumetricflowratetotheoutput.Massflow Massflow Specifiesthemassflowratetotheoutput.Calorie Calorie Specifiesthecalorietotheoutput.Diag Diag Specifiestheflownoisetotheoutput.
<4. Functions> 67
IM 01E22C02-01EN
4.5.2 Analog Output High/Low Limit FunctionThehighandlowlimitsoftheanalogoutputcanbesetforeachterminal.Thehighlimitcanbelimitedbetween4.0mAand21.6mA,andthelowlimittobetween2.4mAand20.0mA.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Analogout/in►(seebelow)BRAIN G:ANALOGI/O►(seebelow)
Parameter DescriptionDisplay BRAINAO1►Highlimit G02:AO1HILIMIT Specifiesthehighlimitofanalogoutput1.AO1►Lowlimit G03:AO1LOLIMIT Specifiesthelowlimitofanalogoutput1.AO2►Highlimit G12:AO2HILIMIT Specifiesthehighlimitofanalogoutput2.AO2►Lowlimit G13:AO2LOLIMIT Specifiesthelowlimitofanalogoutput2.
Example:
Ifthehighlimitissetto16mAandthelowlimitto8mA,theresultisasshownbelow.
F0412.ai
0
4
8
12
16
20
Time [sec]
Current [mA]
Low limit
High limitCurrent output before limit function applied.
<4. Functions> 68
IM 01E22C02-01EN
4.5.3 Absolute Range FunctionTheabsoluterangefunctionisavailableforusinganalogoutput1.Iftheabsoluterangefunctionisused,theforwardflowrateandreverseflowratetowardaspancanbeoutputtoanalogoutputatthesametime.With0%flowratesetat12mA,thereverseflowratecanbeoutputtoananalogoutputof4to12mA,andtheforwardflowratetoananalogoutputof12to20mA.Theabsoluterangefunctionandlow-cutfunctioncanbeusedatthesametime.Fordetailsaboutthelow-cutfunction,refertoSubsection4.1.7.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Analogout/in►AO1►RangemodeBRAIN G:ANALOGI/O►G05:AO1RNGMODE
Thisparametersetstheuseoftheabsoluterangefunctionforanalogoutput1.Fromthetablebelow,selecttheuseoftheabsoluterangefunction.
Selection DescriptionDisplay BRAINNormalrange Normalrange Doesnotusetheabsoluterangefunction.Absrange Absrange Usestheabsoluterangefunction.
Example:(1)Withabsoluterangeandnolowcut
F0413.ai-100
4
8
12
16
20
Flow rate [%]
Current [mA]
0 100
(2)Withabsoluterangeandlowcut
F0414.ai-100
4
8
12
16
20
Flow rate [%]
Current [mA]
0 100
Low cut
NOTETheabsoluterangefunctionandmultirangefunctioncannotbeusedatthesametime.
<4. Functions> 69
IM 01E22C02-01EN
4.5.4 Alarm Output FunctionBehavioroftheanalogoutputwhenatargetalarmoccurscanbesetwiththealarmoutputfunction.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Analogout/in►(seebelow)BRAIN G:ANALOGI/O►(seebelow)
Parameter DescriptionDisplay BRAINAO1►Alarmout G04:AO1ALMOUT Specifiesthealarmoutputfunctionofanalogoutput
1.AO2►Alarmout G14:AO2ALMOUT Specifiesthealarmoutputfunctionofanalogoutput
2.
Fromthetablebelow,selecttheuseofthealarmoutputfunction.Selection DescriptionDisplay BRAIN
<2.4mA <2.4mA Outputslowerthan2.4mA.3.8mA 3.8mA Outputs3.8mA.4mA 4mA Outputs4mA.20.5mA 20.5mA Outputs20.5mA.>21.6mA >21.6mA Outputshigherthan21.6mA.
Measuredvalue Measuredvalue Outputsmeasuredcurrent(thecurrentvalueisundefinedduetoanerror).
Hold Hold Outputscurrentthatisappliedwhenanalarmoccurs.
4.5.5 Analog Output PriorityThefunctionoftheprioritizedanalogoutputvariesdependingonthesettingsoftheproduct.
Priority level Output mode
High
Low
Testmodebit(Fordetails,refertoSection4.12.)StatusinputfunctionOutputof0%signallock(Fordetails,refertoSubsection4.4.2.)Alarmoutputfunction(Fordetails,refertoSubsection4.5.4.)
Verificationfunction,Outputduringofflinediagnosis(Fordetails,refertoSubsection4.11.6.)
Normaloutput
<4. Functions> 70
IM 01E22C02-01EN
4.5.6 Current Value AdjustmentAfterthesensorisinstalledtothecustomer’spiping,ananaloginputvalueoranalogoutputvaluecanbeadjustedinthecustomer’sinstallationenvironmentToadjustananalogoutputvalue,outputcurrentsof4mAand20mAfromthisproduct,connectanammeterforcalibration,andmeasurethecurrentvalue.Setthemeasuredcurrentvaluetotheparameterandadjusttheanalogoutputvalue.Toadjusttheanaloginputvalue,inputcurrentsof4mAand20mAfromanexternaldevice,anddisplaytheanaloginputvalueonthedisplayofthisproduct.Fordetailsaboutsettingsofthedisplayitem,refertoSubsection4.9.2.Setthedisplayedanalogvalue(adjustedvalue)totheparameterandadjusttheanaloginputvalue.Thissettingcanbeconfiguredwiththefollowingparameters.
�� Analog inputMenu path
Display Devicesetup►Diag/Service►AO/AItrim►AItrim►(seebelow)BRAIN G:ANALOGI/O►(seebelow)
Parameter DescriptionDisplay BRAIN
Trim4mA G51:AIat4mA Specifiestheadjustmentvaluewheninputting4mAofanaloginput.
Trim20mA G52:AIat20mA Specifiestheadjustmentvaluewheninputting20mAofanaloginput.
Trimclear G50:AITRIMCLR Clearstheadjustmentvalueforinputtinganaloginput.
�� Analog outputMenu path
Display Devicesetup►Diag/Service►AO/AItrim►AOtrim►(seebelow)BRAIN G:ANALOGI/O►(seebelow)
Parameter DescriptionDisplay BRAIN
AO1trim4mA G31:AO1at4mA Specifiestheadjustmentvalueofanalogoutput1(4mA).
AO1trim20mA G32:AO1at20mA Specifiestheadjustmentvalueofanalogoutput1(20mA).
AO1trimclear G30:AO1TRIMCLR Clearstheadjustmentvalueofanalogoutput1.
AO2trim4mA G41:AO2at4mA Specifiestheadjustmentvalueofanalogoutput2(4mA).
AO2trim20mA G42:AO2at20mA Specifiestheadjustmentvalueofanalogoutput2(20mA).
AO2trimclear G40:AO2TRIMCLR Clearstheadjustmentvalueofanalogoutput2.
IMPORTANTIftheinput/outputdoesnotagreewiththereadingsofthecurrentvaluewhenthecurrentvalueadjustmentfunctionisused,readjustthecurrentvalue.
<4. Functions> 71
IM 01E22C02-01EN
4.5.7 Analog Input Function SettingAprocessvaluefromanexternaldevicecanbeimportedtotheproductasananaloginput.Theanaloginputcanbeusedfortemperature-baseddensitycorrectioncalculationandcaloriecalculationbyusingtheanaloginputasatemperatureinput.Fordetailsaboutthetemperature-inputdensitycorrectioncalculationandcaloriecalculation,refertoSubsections4.7.5and4.7.6.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Analogout/in►AI►FunctionBRAIN G:ANALOGI/O►G20:AIFUNCTION
Thisparameterspecifiestheanaloginputfunction.Fromthetablebelow,selecttheanaloginputfunction.
Selection DescriptionDisplay BRAIN
Nofunction Nofunction Ananaloginputisnotavailablebecausetheanaloginputfunctionisnotenabled.
Monitoring MonitoringTheanaloginputvaluecanbemonitored.However,thisfunctiondoesnotgiveanyinfluencesoneitherthedensitycorrectioncalculationorthecaloriecalculation.
Difftemperature DifftemperatureTheanaloginputisusedasthetemperaturedifference.Caloriecanbecalculated.
Exttemperature Exttemperature Theanaloginputisusedasabsolutetemperature.Densitycorrectionandcaloriecanbecalculated.
4.5.8 Display Analog InputThecurrentvalueoftheanaloginputcanbecheckedwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Analogout/in►AI►ValueBRAIN G:ANALOGI/O►G23:AIVALUE
4.5.9 Analog Input Unit SettingAunitcanbesetforwhenananaloginputisusedasaninputfortemperature.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Analogout/in►AI►UnitBRAIN G:ANALOGI/O►G24:AIUNIT
<4. Functions> 72
IM 01E22C02-01EN
4.5.10 Analog Input Range SettingArangecanbesetfortheprocessvaluefortheanaloginputbyspecifying0%and100%oftheprocessvaluefortheanaloginput.TheunitfortherangeisthesameastheunitspecifiedinSubsection4.5.10.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Analogout/in►AI►(seebelow)BRAIN G:ANALOGI/O►(seebelow)
Parameter DescriptionDisplay BRAIN
URV G25:AIURV Specifiesthevaluethatisatthe100%pointoftheprocessvaluefortheanaloginput.
LRV G26:AILRV Specifiesthevaluethatisatthe0%pointoftheprocessvaluefortheanaloginput.
4.5.11 Analog Input High/Low Limit FunctionHighandlowlimitscanbesetfortheanaloginput.Thehighlimitcanbesettobetween4.0mAand21.6mA,andthelowlimittobetween2.4mAand20.0mA.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Analogout/in►AI►(seebelow)BRAIN G:ANALOGI/O►(seebelow)
Parameter DescriptionDisplay BRAIN
Highlimit G21:AIHILIMIT Specifiesthehighlimitvalueoftheanaloginputvalue.
Lowlimit G22:AILOLIMIT Specifiesthelowlimitvalueoftheanaloginputvalue.
<4. Functions> 73
IM 01E22C02-01EN
4.6 Multi Range Function4.6.1 Multi Range type
Themultirangefunctioncanbeusedforanalogoutput1.Themultirangefunctionshowsthepresentrangewiththestatusinputorstatusoutput.ThemultirangefunctionmakesitpossibletomeasureaspanforthePV-mappedprocessvalueinSubsection4.1.2whileswitchingmultipleranges.Themultirangefunctionhasthreeranges;themultirange,theforward/reverserange,andtheexternalcontactrange,andoneoftherangesshouldbeselected.Thecontentsofeachrangeareasfollows:
Multi range Thestateoftherangeisoutputinthestatusoutputbymeasuringaflowwhileswitchingseveralrangesinresponsetoflowrate.
Forward/Reverse range
Thestateoftherangeisoutputinthestatusoutputbymeasuringaflowwhileswitchingseveralrangesinresponsetotheflowdirection.
External Contact Range Flowismeasuredwhileswitchingseveralrangesbyusingthestatusinputs.
NOTE• Theabsoluterangefunctionandmultirangefunctioncannotbeusedatthesametime.• Themultirangefunctionisnotavailableforanalogoutput2.
4.6.2 Multi Range SettingWhenusingthemultirangefunction,eachrangeneedstobeconfigured.Forunitsetting,refertoSubsection4.1.4.Thissettingcanbeconfiguredwiththefollowingparameters.RefertoSubsection4.1.5.forthespansetting(forwardspan1)whenthemultirangefunctionisnotused.Menu path
Display Devicesetup►Detailedsetup►Multirange►(seebelow)BRAIN F:STATUSI/O►(seebelow)
Parameter DescriptionDisplay BRAINForwardspan2 F40:FWDSPAN2 Specifiesforwardspan2.Forwardspan3 F41:FWDSPAN3 Specifiesforwardspan3.Forwardspan4 F42:FWDSPAN4 Specifiesforwardspan4.Reversespan1 F43:REVSPAN1 Specifiesreversespan1.Reversespan2 F44:REVSPAN2 Specifiesreversespan2.Reversespan3 F45:REVSPAN3 Specifiesreversespan3.Reversespan4 F46:REVSPAN4 Specifiesreversespan4.
Thetypeoftherangeforthemultirangefunctioncanbesetwiththestatusoutputandstatusinputfunctions.Fordetails,refertoSubsections4.3.9.and4.4.2.Itisnecessarytosetaspanforeachrangeinresponsetotherangetypeofthemultirangefunction.Seteachspanvaluesoastosatisfytherelationoflowerrange≤upperrange.Forexample,iftheforward/reverserangeistobeusedincombinationwiththemultirange(range1and2),spansofforwardspan1,forwardspan2,reversespan1,andreversespan2needtobeset,andtherelationsofforwardspan1≤forwardspan2andreversespan1≤reversespan2shouldbesatisfied.Iftheseconditionsarenotmet,“Analogoutput1multirangeerror”occursandthemultirangefunctiondoesnotoperate.
<4. Functions> 74
IM 01E22C02-01EN
4.6.3 Multi Range OperationThemultirangeallowstworangestobeautomaticallyswitchedformeasurementinresponsetotheflowrate.Whenameasuredflowrateislargerthanthelowerrange,therangeisswitchedtothehigherrange.Whenameasuredflowrateissmallerthanthehigherrange,therangeisswitchedtothelowerrange.Therangebeingusedisshownwiththestateofthestatusoutput.Whenthemultirangeisusedinrange1and2,therelationoftherangeandstatusoutputisasfollows:
Range Stateofstatusoutput(auto2rangeindicated)Forward span 1 NotactiveForward span 2 Active
Whentherangeisswitchedfromthehigherrangetothelowerrange,ithashysteresis.Thehysteresissetsapercentageforthelowerrange.Thehysteresisformultirangeswitchingcanbesetwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Multirange►AutorangehystBRAIN F:STATUSI/O►F50:AUTORNGHYS
Whenmultirangeisusedinrange1and2,themultirangecanbesetwiththefollowingprocedures.(1) RefertoSubsection4.1.2tospecifythePVprocessvalue.
(2) RefertoSubsections4.1.4and4.1.5tospecifythespan(forwardspan1).
(3) Specifythemultirangeswitchinghysteresis.
(4) RefertoSubsection4.3.1tospecifytheoutputoftheterminaltothestatusoutput.
(5) RefertoSubsection4.3.9tospecifythestatusoutputfunctiontoauto2range.
(6) RefertoSubsection4.6.2tospecifyforwardspan2.
Specifyavaluegreaterthanforwardspan1.
Example:PV=Volumetricflowrate,Flowratespan(range1)=50m3/h,Range2=100m3/h,WhensettoHysteresiswidth=10%->Hysteresisvalue =50[m3/h]×10[%]
=5[m3/h]
(1) Whentherangeisswitchedfromlowerrange1tohigherrange2
F0415.ai
0
100
Status output
Output [%]
Flow rate [m3/h]
50 100
Forward range 1Forward range 2
<4. Functions> 75
IM 01E22C02-01EN
(2) Whentherangeisswitchedfromhigherrange2tolowerrange1(waitingforhysteresis)
F0416.ai
0
100
Status output
Output [%]
Flow rate [m3/h]
50 10045
Forward range 1Forward range 2
Ifstatusoutput11and12arecombined,aflowratecanbemeasuredbyautomaticallyswitchinganalogoutput1toranges1through3ortorange1through4inresponsetotheflowrate.Whenusingthemultirangeatranges1through3,setthefunctionsofstatusoutput11and12toauto3range.Also,setspanvaluesatforwardspans1through3.Whenusingthemultirangeatranges1through4,setthefunctionsofstatusoutput11and12toauto4range.Also,setspanvaluesatforwardspans1through4.Therelationbetweenarangeandthestatusoutputisasfollows:
Range Stateofstatusoutput11(Auto3(or4)range)
Stateofstatusoutput12(Auto3(or4)range)
Forward span 1 Notactive NotactiveForward span 2 Active NotactiveForward span 3 Notactive ActiveForward span 4* Active Active
*: Ifsettouseranges1through3,therangeisintransitionuptorange3.
NOTETochangethePVprocessvaluesetinSubsection4.1.2,itisnecessarytore-setthemultirange.
<4. Functions> 76
IM 01E22C02-01EN
4.6.4 Forward/Reverse RangeIntheforward/reverserange,thisproductmeasuresaflowbyautomaticallyswitchingtheforwarddirectionrange(forwardrange)andreversedirectionrange(reverserange)inresponsetotheflowdirection.Boththeforwardrangeandreverserangecanbesetfortherange,respectively.Therangebeingusedisshownwiththestateofthestatusoutput.Whentheforward/reverserangeisused,therelationoftherangeandstatusoutputisasfollows:
Flow direction Range Stateofthestatusoutput(Forward/Reverserange)
Forward Forwardspan1 NotactiveReverse Reversespan1 Active
Whentherangeisswitchedfromtheforwardrangetothereverserange,ithashysteresis.Thehysteresissetsapercentageforeithertheforwardrangeorthereverserange,whicheverissmaller.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Multirange►BidirectionhystBRAIN F:STATUSI/O►F51:BIDIRECHYS
Theforward/reverserangecanbeconfiguredwiththefollowingprocedure.(1) RefertoSubsection4.1.2tospecifythePVprocessvalue.
(2) RefertoSubsections4.1.4and4.1.5tospecifythespan(forwardspan1).
(3) Specifytheforward/reverserangeswitchinghysteresis.
(4) RefertoSubsection4.3.1tospecifytheoutputoftheterminaltothestatusoutput.
(5) RefertoSubsection4.3.9tospecifythestatusoutputfunctiontoFwd/Revrange.
(6) RefertoSubsection4.6.2tospecifyreversespan1.
<4. Functions> 77
IM 01E22C02-01EN
Example:PV=Volumetricflowrate,Flowratespan(forwardspan1)=50m3/h,Reversespan1=100m3/h,WhensettoHysteresis=5%->BecauseofForwardrange<ReverserangeHysteresisvalue =50[m3/h]×5[%]
=2.5[m3/h]
(1) Whentherangeisswitchedfromthereverserangetotheforwardrange(waitingforhysteresis)
F0417.ai
-100
100
Status output
Output [%]
Flow rate [m3/h]0
Reverse range 1Forwardrange 1
502.5
(2) Whentherangeisswitchedfromtheforwardrangetothereverserange(waitingforhysteresis)
F0418.ai
-100
100
Status output
Output [%]
Flow rate [m3/h]0
Reverse range 1Forward range 1
50-2.5
NOTETochangethePVprocessvaluesetinSubsection4.1.2,itisnecessarytore-settheforward/reverserange.
<4. Functions> 78
IM 01E22C02-01EN
4.6.5 Combination of Multi Range and Forward/Reverse Range
Themultirangecanbeusedincombinationwiththeforward/reverserangeforthisproduct.Whenthemultirange(range1to2)andtheforward/reverserangearecombined,therelationoftherangeandstatusoutputisasfollows:
Flow direction Range Stateofthestatusoutput(Forward/Reverserange)
Stateofthestatusoutput(Auto2range)
Forward Forward span 2 Notactive ActiveForward span 1 Notactive Notactive
Reverse Reverse span 1 Active NotactiveReverse span 2 Active Active
Themultirangeandtheforward/reverserangeeachhaveadifferenthysteresis.Forthehysteresesofthemultirangeandtheforward/reverserange,refertoSubsections4.6.3and4.6.4.
Thecombinationoftheforward/reverserangeandthemultirange(ranges1through2)canbesetusingthefollowingprocedures.(1) RefertoSubsection4.1.2tospecifythePVprocessvalue.
(2) RefertoSubsections4.1.4and4.1.5tospecifythespan(forwardspan1).
(3) RefertoSubsections4.6.3and4.6.4tospecifythehysteresesforthemultirangeandtheforward/reverserange.
(4) RefertoSubsection4.3.1tospecifytheoutputofeachterminaltothestatusoutput.
(5) RefertoSubsection4.3.9tospecifythestatusoutputfunctiontoauto2rangeandforward/reverserange.
(6) RefertoSubsection4.6.2tospecifyforwardspan2,reversespan1,andreversespan2.
Bycombiningstatusoutput11and12withanystatusoutput(exceptforstatusoutput11and12),aflowratecanbemeasuredbyautomaticallyswitchinganalogoutput1toforward/reverseranges1through3ortoforward/reverseranges1through4inresponsetoaflowrate.Whenusingthecombinationintheforward/reverseranges1through3,thefunctionsofstatusoutput11and12shouldbesettoauto3rangeandanystatusoutput(exceptfor11and12)shouldbesettotheforward/reverserange.Also,setthespanvaluesforuptoforwardspans1through3anduptoreversespans1through3.Whenusingthecombinationintheforward/reverseranges1through4,thefunctionsofstatusoutput11and12shouldbesettotheauto4rangeandanystatusoutput(exceptfor11and12)shouldbesettotheforward/reverserange.Also,setthespanvaluesforforwardspans1through4andreversespans1through4.Therelationbetweenarangeandthestatusoutputisasfollows:
Flow direction Range
Stateofthestatusoutput
(Forward/Reverserange)
Stateofstatusoutput11(Auto3(or4)range)
Stateofstatusoutput12(Auto3(or4)range)
Forward
Forward span 4* Notactive Active ActiveForward span 3 Notactive Notactive ActiveForward span 2 Notactive Active NotactiveForward span 1 Notactive Notactive Notactive
Reverse
Reverse span 1 Active Notactive NotactiveReverse span 2 Active Active NotactiveReverse span 3 Active Notactive ActiveReverse span 4* Active Active Active
*: Whenforward/reverseranges1through3aretobeused,therangeisintransitionuptoforwardspan3orreversespan 3.
<4. Functions> 79
IM 01E22C02-01EN
4.6.6 External Contact RangeTheexternalcontactrangecanbeusedwiththisproduct.Intheexternalcontactrange,aflowratecanbemeasuredbyarbitrarilyswitchingtworangesinresponsetostatusinput.Therelationbetweenarangeandthestatusinputisasfollows:
State of status input(Ext 2 ranges) Range
Not active Range1Active Range2
Whenusingtheexternalcontactrangeinranges1and2,itcanbesetwiththefollowingprocedures.(1) RefertoSubsection4.1.2tospecifythePVprocessvalue.
(2) RefertoSubsections4.1.4and4.1.5tospecifythespan(forwardspan1).
(3) RefertoSubsection4.4.2tospecifythefunctionofthestatusinputtoExt2ranges.
(4) RefertoSubsection4.6.2tospecifyforwardspan2.
Example:PV=Volumetricflowrate,Flowratespan(forwardspan1)=50m3/h,Whensettingtheforwardspan2=100m3/h
F0419.ai
Output [%]
Flow rate [m3/h]50 1000
100Forward range 2
Status input
Forward range 1
Ifstatusinput11and12arecombined,aflowratecanbemeasuredbyarbitrarilyswitchingtheanalogoutput1rangetoranges1through3ortoranges1through4.Whenusingtherangeinranges1through3,setthefunctionsofstatusinput11and12toExt3ranges.Also,setthespanvaluefromforwardspans1through3.Whenusingtherangeinranges1through4,setthefunctionsofstatusinput11and12toExt4ranges.Also,setthespanvaluefromforwardspans1through4.Therelationbetweenarangeandthestatusinputisasfollows:
State of status input 11(Ext 3 (or 4) range)
State of status input 12(Ext 3 (or 4) range) Range
Not active Not active Range1Active Not active Range2
Not active Active Range3Active Active Range4*
*: Forthesettingtouseranges1through3,therangeisintransitionuptorange3(therangeoperatesinrange3evenifthestatusinputissettorange4).
<4. Functions> 80
IM 01E22C02-01EN
4.6.7 External Contact Range (with Answer)Thisproductcandisplaythepresentrangeintheformofthestateofstatusoutputwhileoperatingintheexternalcontactrange.The relation between a range and the status output is as follows:
State of status input(Ext 2 ranges) Range Stateofthestatusoutput
(Ext2answer)Not active Range1 Notactive
Active Range2 Active
Theexternalcontactrangewithanswercanbeconfiguredusingthefollowingprocedure.(1) RefertoSubsection4.1.2tospecifythePVprocessvalue.
(2) RefertoSubsections4.1.4and4.1.5tospecifythespan(forwardspan1).
(3) RefertoSubsection4.4.2tospecifythefunctionofthestatusinputtoExt2ranges.
(4) RefertoSubsection4.3.9tospecifythefunctionofthestatusoutputtoExt2answer.
(5) RefertoSubsection4.6.2tospecifyforwardspan2.
Example:PV=Volumetricflowrate,Flowratespan(forwardspan1)=50m3/h,Whensettingtoforwardspan2=100m3/h
OutputForward range 1
Forward range 2
Flow rate[m3/h]
Status input
Status output
50 1000
100
F0446.ai
Whilearbitrarilyswitchingtheanalogoutput1rangefromranges1through3orfromranges1through4bycombiningstatusinput11and12withstatusoutput11and12,thepresentrangecanbedisplayedbycombiningstatusoutput11and12.Whenusingtherangeinranges1through3,statusinput11and12functionsshouldbesettotheExt3ranges,andthestatusoutput11and12functionsshouldbesettoExt3answerafterthespanvaluesforforwardspans1through3arealsoset.Whenusingtherangeinranges1through4,statusinput11and12functionsshouldbesettotheExt3ranges,andthestatusoutput11and12functionsshouldbesettotheExt4answerafterthespanvaluesforforwardspans1through4arealsoset.
<4. Functions> 81
IM 01E22C02-01EN
State of status input 11
(Ext 3 (or 4) range)
State of status input 12
(Ext 3 (or 4) range)Range
Stateofstatusoutput11
(WithExt3(or4)rangeanswer)
Stateofstatusoutput12
(WithExt3(or4)rangeanswer)
Not active Not active Range1 Notactive NotactiveActive Not active Range2 Active Notactive
Not active Active Range3 Notactive ActiveActive Active Range4* Active Active
*: Forthesettingtouseranges1through3,therangeisintransitionuptorange3(therangeoperatesinrange3evenifthestatusinputissettorange4).
4.6.8 Combination of External Contact Range and Forward/Reverse Range
Theexternalcontactrangecanbeusedincombinationwiththeforward/reverserangeforthisproduct.Therelationoftherangeandstatusoutputisasfollowswhentheforward/reverserangeandExternalContactRange(range1and2)arecombined.
Condition OperationFlow
directionState of status input
(Ext 2 ranges) Range Stateofthestatusoutput(Forward/Reverserange)
Stateofthestatusoutput(Ext2answer)
Forward Active Forwardspan2 Notactive ActiveNot active Forwardspan1 Notactive Notactive
Reverse Not active Reversespan1 Active NotactiveActive Reversespan2 Active Active
Thecombinationoftheforward/reverserangeandtheexternalcontactrange(range1and2)canbesetinthefollowingprocedures.(1) RefertoSubsection4.1.2tospecifythePVprocessvalue.
(2) RefertoSubsections4.1.4and4.1.5tospecifyspan(forwardspan1).
(3) RefertoSubsections4.6.3and4.6.4tospecifythehysteresesforthemultirangeandtheforward/reverserange.
(4) RefertoSubsection4.3.9tospecifythefunctionofthestatusoutputtotheforward/reverserange.
(5) RefertoSubsection4.4.2tospecifythefunctionofthestatusinputtoExt2ranges.
(6) RefertoSubsection4.6.2tospecifyforwardspan2,reversespan1,andreversespan2.
Whilearbitrarilyswitchingtheanalogoutput1rangefromranges1through3orfromranges1through4bycombiningstatusinput11and12withstatusoutput11and12,thepresentrangecanbedisplayedbycombiningstatusoutput11and12.Whenusingthecombinationinranges1through3,thestatusinput11and12functionsshouldbesettotheExt3ranges,thestatusoutput(exceptfor11and12)shouldbesettotheforward/reverserange,andthestatusoutput11and12functionsshouldbesettotheExt3answerafterthespanvaluesforforward1through3rangesandreverse1through3rangesarealsoset.Whenusingthecombinationinranges1through4,thestatusinput11and12functionsshouldbesettotheExt3ranges,thestatusoutput(exceptfor11and12)shouldbesettotheforward/reverserange,andthestatusoutput11and12functionsshouldbesettotheExt4answerafterthespanvaluesforforward1through4rangesandreverse1through4rangesarealsoset.
<4. Functions> 82
IM 01E22C02-01EN
Flow direction
State of status input 11
(Ext 3 (or 4) range)
State of status input 12
(Ext 3 (or 4) range)
Range
Stateofthestatusoutput(Forward/Reverserange)
Stateofstatusoutput11(WithExt3(or4)rangeanswer)
Stateofstatusoutput12(WithExt3(or4)rangeanswer)
Forward
Active Active Forwardspan4* Notactive Active ActiveNot active Active Forwardspan3 Notactive Notactive Active
Active Not active Forwardspan2 Notactive Active NotactiveNot active Not active Forwardspan1 Notactive Notactive Notactive
Reverse
Not active Not active Reversespan1 Active Notactive NotactiveActive Not active Reversespan2 Active Active Notactive
Not active Active Reversespan3 Active Notactive ActiveActive Active Reversespan4* Active Active Active
*: Forthesettingtouseranges1through3,therangeisintransitionuptorange3(therangeoperatesinrange3evenifthestatusinputissettorange4).
<4. Functions> 83
IM 01E22C02-01EN
4.7 Auxiliary Calculation Function4.7.1 Fluid Flow Direction Setting
Thearrowonthesurfaceofthesensorindicatesthefluidflowdirection.Uponshipmentfromthemanufacturingfactory,theflowrateismeasuredassumingthatthearrowdirectionisforward.Bychangingtheparametersettings,thisproductcanmeasuretheflowrate,assumingthatthereversedirectionisforwardagainstthearrowdirection.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►AUXcalculation►FlowdirectBRAIN H:AUXCALC►H10:FLOWDIRECT
Thisparameterspecifiesthefluidflowdirection.Fromthetablebelow,selectthefluidflowdirection.
Selection DescriptionDisplay BRAINForward Forward Thearrowdirectionofthesensorisforward.
Reverse Reverse Thereversedirectionofthearrowofthesensorisforward.
4.7.2 Rate Limit Function SettingIftheratelimitfunctionisused,itbecomespossibletoreducenoisesthatcannotallbeclearedonlybylengtheningthedampingtimeconstant.Whenastepsignalorasuddensignalduetoaslurryfluidisinput,thisfunctionjudgeswhetherthesignalisaflowratesignaloranoisesignal.Thisjudgmentismadebasedonthehigh/lowlimitvalue(ratelimitvalue)andtheratelimitfunctioncontinuationtime(deadtime),causingthenoisesignalovertheratelimitvaluetobecutoff.Theratelimitvalueisspecifiedwiththepercentage(%)forthespanofthePV-mappedprocessvalueinSubsection4.1.2.Thedeadtimeissetto“0”iftheratelimitfunctionisnotused.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►AUXcalculation►(seebelow)BRAIN H:AUXCALC►(seebelow)
Parameter DescriptionDisplay BRAINRatelimit H11:RATELIMIT Specifiestheratelimitvalue.Deadtime H12:DEADTIME Specifiesthedeadtime.
Noisefilter H13:NOISEFILTER Specifiesthenoisefilter(ratelimitvalueanddeadtime).*1
<4. Functions> 84
IM 01E22C02-01EN
*1: Fromthetablebelow,selectthenoisefilter(ratelimitvalueanddeadtime).Selection Rate limit value Dead timeDisplay BRAIN
Manual ManualThevalueisspecifiedintheparameter“Ratelimit”.
Thevalueisspecifiedintheparameter“Deadtime”.
Level 1 Level 1 0.5% 0.5sLevel 2 Level 2 1.0% 1.0sLevel 3 Level 3 5.0% 3.0s
NOTEIfeithertheratelimitvalueorthedeadtimeisspecified,thenoisefilterissetto“Manual”.
NOTEDeterminingtheratelimitvalueanddeadtime
T0
T0
2%
2%
The Rate limit value:Determine the level which should be cut the output fluctuation. For example, if its level is 2%, the noise of 2% or larger would be cut as shown in the following figure.
The Dead time (T0):Determine the value depending on the width of the output fluctuation. Choose the larger value when the noise which is over the dead time as shown in the following figure.
F0420.ai
NOTEFortheratelimitfunction,thedeadtimeissetto“0”uponshipmentfromthemanufacturingfactory.Besuretosetthedeadtimewhentheratelimitfunctionisused.
Signal processing for rate limit functionTheproductcalculatestosetthespecificratelimitvaluetotheprimarydelayresponsevalueofthepreviouslysampledflowratevalue.Iftheflowratevaluesampledatthistimeexceedstheratelimitvalueabove,itshighorlowlimitissettotheflowratevalueatthistime.Furthermore,ifthesamplingcountoccurswithinthedeadtimewhilethesignaloverthehigh/lowlimitisinthesamedirection,thissignalisjudgedtobeaflowratesignal.
<4. Functions> 85
IM 01E22C02-01EN
Example:(1) Wheninput=0to10%,Dampingtimeconstant=5seconds,Deadtime=3seconds,
andRatelimitvalue=1%,theoutputforthestepinputisobtainedasshownbelow.
F0421.ai
10%
1%
Step signal
Flow rate valueafter rate limitprocessing
63.2%
100 225Deat time: 3 s
(a) (b) (c) (d) Flow rate value after damping
Input: 0 to 10%Damping time constant: 5 sDead time: 3 sRate limit value: 1%
Number of signal samples
• Incondition(a),thesignalexceedstheratelimitvalueascomparedwiththepreviousvalue;therefore,theresponseissetto1%.
Theactualoutput,whichisdamped,isprocessedasindicatedbythesolidline.
• Then,theflowratevalueinthedeadtimeissetto“flowrateafterdampingcalculation+signalofratelimitvalue(1%)”.
• Theinputsignaldoesnotreturntotheratelimitvalueorlesswithinthedeadtime;therefore,itisjudgedtobeaflowratesignalatthetimeof(c).
• Theoutputsignalstartsfollowingthestepsignalalongthedampingcurve.
Thefigurebelowshowsanoutputexamplewhenaslurrynoisehasoccurred.(2) Wheninput=0to10%,Dampingtimeconstant=1second,Deadtime=1seconds,
andRatelimitvalue=1%,theoutputforaslurrynoiseisobtainedasshownbelow.
F0422.ai
+1%
-1%
Dead time: 1 s
Flow rate value after damping
Flow rate valueafter rate limitprocessing
Slurry noise
Time
Input: 0 to 10%Damping time constant: 1 sDead time: 1 sRate limit value: 1%
In the figure on the left, it is determined that the slurries noise signal is not a flow rate signal.
<4. Functions> 86
IM 01E22C02-01EN
4.7.3 Pulsing Flow Support Function SettingIfaplungerpumpisused,itmaycauseanerrorintheaverageoftheflowrateduetotheinfluenceofthepulsingflow.Ifthepulsingflowsupportfunctionisused,anerrorduetoapulsingflowcanbereducedbyfollowingaflowchangewhilecontrollingtheflowratecalculation.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►AUXcalculation►PulsingflowBRAIN H:AUXCALC►H14:PULSINGFLOW
Thisparameterspecifiestheuseofthepulsingflowsupportfunction.Fromthetablebelow,selecttheuseofthepulsingflowsupportfunction.
Selection DescriptionDisplay BRAINNo No Donotusethepulsingflowsupportfunction.Yes Yes Usethepulsingflowsupportfunction.
4.7.4 Power Frequency Synchronization SettingThisfunctioncanbespecifiedwhethertheexcitationfrequency(internalsignalprocessingfrequency)andpowerfrequencyaresynchronousorasynchronous.Whensettingtheexcitationfrequencyandpowerfrequencytoasynchronous,theexcitationfrequencyisdeterminedbythesetvalueofthepowerfrequency.Thepowerfrequencysynchronous/asynchronousmodeandthepowerfrequencycanbeconfiguredwiththefollowingparameters.
Menu pathDisplay Devicesetup►Detailedsetup►AUXcalculation►(seebelow)BRAIN H:AUXCALC►(seebelow)
Parameter DescriptionDisplay BRAIN
Powersync H20:POWERSYNCH Setstheexcitationfrequencyandpowerfrequencytosynchronous.*1
Setpowerfreq H21:SETPWRFREQ Specifiesthepowerfrequencywhentheexcitationfrequencyandpowerfrequencyareasynchronous.
*1: Setsthesynchronous/asynchronousmodeofpowerfrequencyfromthetablebelow.Selection DescriptionDisplay BRAIN
No No Setstheexcitationfrequencyandpowerfrequencyasynchronous.
Yes Yes Setstheexcitationfrequencyandpowerfrequencysynchronous.
IMPORTANTWhenusingtheDCpowerasthetransmitterpower,setthecommerciallyavailablepowerfrequencyoftheplacewherethetransmitterwillbeused.Inthiscase,set“Powersync”toOfffirstbeforesetting“Setpowerfreq”.
<4. Functions> 87
IM 01E22C02-01EN
Theexcitationfrequencyandpowerfrequencycanbecheckedwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►AUXcalculation►(seebelow)BRAIN H:AUXCALC►(seebelow)
Parameter DescriptionDisplay BRAIN
Iexpowerfrequency H22:IEXPWRFREQ Displaysthepowerfrequency(synchronouswiththeexcitationfrequency).
Measpowerfreq H23:MESPWRFREQ Displaysthemeasuredpowerfrequency.
4.7.5 Density Correction CalculationWhenthetemperaturefromanotherdevice(temperaturetransmitter,etc.)isusedasananaloginput,thedensitycorrectioncalculationcanbeperformedbasedontemperature.Themeasurementaccuracyforthemassflowrateisimprovedbyusingthisfunction.Transmitter
Iin-
Iin+24 V
0 V
Temperature transmitter, etc.
F0423.ai
Themassflowrateisobtainedusingthefollowingformula.Vm=Vf×ρ Vm:Massflowrate[kg/s] Vf:Volumetricflowrate[m3/s] ρ:Density[kg/m3]
Whenthedensitycorrectioncalculationisperformedbasedontemperature,thedensityisobtainedusingthefollowingformula.Thedensitycompensatingratesmustbespecifieddependingonapplicationsthatusethisproduct. ρr=ρn×{1+a1×(Tr-Tn)×10-2+a2×(Tr-Tn)2×10-6} ρr:Densitycorrectedbasedonthemeasuredtemperature[kg/m3] ρn:Standarddensity[kg/m3] Tr:Measuredtemperature[°C] Tn:Standardtemperature[°C] a1:Primarycompensatingrate a2:Secondarycompensatingrate
Thestandardtemperatureandthecompensatingratescanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Provar►Temperature►(seebelow)BRAIN H:AUXCALC►(seebelow)
Parameter DescriptionDisplay BRAINCoefA1 H35:TEMPCOEFA1 Specifiestheprimarycompensatingrate.CoefA2 H36:TEMPCOEFA2 Specifiesthesecondarycompensatingrate.
<4. Functions> 88
IM 01E22C02-01EN
Thedensitycorrectioncalculationcanbesetusingthefollowingprocedure.(1) RefertoSubsection4.11.1tospecifytheuseofthetemperaturecorrecteddensity.
(2) RefertoSubsection4.5.7tospecifythefunctionoftheanaloginputtotemperature.
(3) RefertoSubsection4.5.10tospecifytherangeoftheanaloginput.
(4) RefertoSubsections4.1.11and4.1.12tospecifythestandardtemperatureandstandarddensity.
(5) Specifytheprimarycompensatingrateandsecondarycompensatingrateofthecompensationformula.
NOTEThedensitycorrectioncalculationisperformedonlywhenitems(1)and(2)abovearespecified.Iftheyarenotspecified,thedensityisassignedtothefixedvaluethatisspecifiedinSubsection4.1.11.
Example: Ifawaterdensityiscorrectedbasedonthetemperatureinthefollowingconditions,theresultisobtainedasshownbelow.analoginputrange=0.0to100.0°C,Standarddensity=1000kg/m3,Standardtemperature=20°C,Primarycompensatingrate=-0.02,Secondarycompensatingrate=-3.9
F0424.aiTemperature [°C]
Den
sity
[kg/
m3 ]
950
960
970
980
990
1000
10 20 30 40 50 60 70 80 90 100
<4. Functions> 89
IM 01E22C02-01EN
4.7.6 Calorie CalculationWhentheabsolutetemperatureorthetemperaturedifferencefromanotherdevice(temperaturetransmitter,etc.)isusedastheanaloginput,thecaloriecalculationisexecutable.
Thecalorieisobtainedusingthefollowingformula.Q=c×Vm×∆t Q:Calorie[J/s] c:Specificheat[J/kg•K] Vm:Massflowrate[kg/s] ∆t:Temperaturedifference[K]
ThetemperaturedifferenceofΔtvariesdependingontheanaloginputfunctionspecifiedinSubsection4.5.8.If“Difftemperature”isselectedintheanaloginputfunction,thetemperatureinputintheformofanaloginputisused.If“Exttemperature”isselected,thedifferencebetweenthetemperatureusedastheanaloginputandthereferencetemperaturespecifiedinSubsection4.1.12isused.
Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Provar►Calorie►SpecificheatBRAIN H:AUXCALC►H40:SPECHEAT
Thisparameterspecifiesthespecificheat.
Thecaloriecalculationcanbeconfiguredusingthefollowingprocedure.
• When the absolute temperature is selected as the analog input:(1) RefertoSubsection4.5.7tospecifythefunctionoftheanaloginputto“Exttemperature”.
(2) RefertoSubsection4.5.10tospecifytherangeoftheanaloginput.
(3) RefertoSubsection4.1.12tospecifythereferencetemperature.
(4) Specifythespecificheat.
•Whenthetemperaturedifferenceisselectedastheanaloginput:(1) RefertoSubsection4.5.7tospecifythefunctionoftheanaloginputto“Difftemperature”.
(2) RefertoSubsection4.5.10tospecifytherangeoftheanaloginput.
(3) Specifythespecificheat.
<4. Functions> 90
IM 01E22C02-01EN
4.7.7 Limit of Alarm Output FunctionThealarmoutputfunctionofanalogoutput1canbelimited.Fordetailsaboutthealarmoutputfunction,refertoSubsection4.5.5.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►AUXcalculation►SetSILBRAIN H:AUXCALC►H50:SETSIL
Fromthetablebelow,selecttheuseofthealarmoutputfunction.Selection DescriptionDisplay BRAIN
No No Doesnotrestrictthealarmoutputfunctionofanalogoutput1.
Yes Yes Limitsthealarmoutputfunctionofanalogoutput1.
NOTEIfthisfunctionissetto“Yes”,thealarmoutputfunctionforanalogoutput1istobelimitedto“<2.4mA”and“>21.6mA”.Ifanotheroptionisset,thealarmoutputfunctionforanalogoutput1issetto“>21.6mA”.
<4. Functions> 91
IM 01E22C02-01EN
4.8 Alarm4.8.1 Errors and Countermeasures
ExplanationofNE107status:NE107 status Status of the device
F Failure Devicemalfunction,Partsmalfunction
C FunctionCheck Theoutputsignalistemporarilyinvalidforthelocaloperationormanualoperationinput.
S Outofspecification Thedeviceworksinoutofspecification.Theoutputsignalisuncertainfortheprocessortheambiance.
M Maintenancerequired Themaintenanceisrequiredinthenearfuture.
N NoEffect Stateotherthanmentionedabove
Thefollowingtableshowspossiblecountermeasures.
�� System alarmThedevicebreaksdownandcausesabnormalmeasurement.Devicereplacementisneeded.
NE107 Status
Error Message Error Description CountermeasureDisplay BRAIN
F 010:MainCPUFAIL 10:MainCPUFAIL CPU(Mainboard)failurewasdetected.
ContactYokogawaservicecenter.
F 011:RevcalcFAIL 11:RevcalFAIL Failureofreversecalculationwasdetected.
ContactYokogawaservicecenter.
F 012:MainEEPFAIL 12:MainEEPFAILFailureofEEPROM(Mainboard)wasdetected.
Turnonthepoweragainwithintherangeofthetemperaturespecifications.Ifthesituationdoesnotimprove,contactYokogawaservicecenter.
F 013:MainEEPdflt 13:MainEEPdfltEEPROM(Mainboard)wasresettodefaultvalues.
ContactYokogawaservicecenter.
F 014:SnsrbdFAIL 14:SnsrbdFAIL Failureofsensorboardwasdetected.
ContactYokogawaservicecenter.
F 015:SnsrcommERR 15:SnsrcommERR
Communicationerrorofthesensorwasdetected.
ContactYokogawaservicecenter.
F 016:AD1FAIL[Sig] 16:A/D1FAILFailureofA/Dtransmitter1(flowvelocitysignal)wasdetected.
ContactYokogawaservicecenter.
F 017:AD2FAIL[Excit] 17:A/D2FAILFailureofA/Dtransmitter2(Excitingcurrent)wasdetected.
ContactYokogawaservicecenter.
F 018:Coil open 18:Coil open Coilofsensorwasdisconnected.
Turnoffthepower,checkcoilofsensorandexcitationcable.
F 019:Coilshort 19:Coilshort Coilofsensorwasshorted.
ContactYokogawaservicecenter.
F 020:ExciterFAIL 20:ExciterFAIL Failureoftheexcitationcircuitwasdetected.
ContactYokogawaservicecenter.
F 021:PWM1stop
21-22:PWMstop
Errorofpulsewidthmodulation1wasdetected.
ContactYokogawaservicecenter.
F 022:PWM2stopErrorofpulsewidthmodulation2wasdetected.
ContactYokogawaservicecenter.
<4. Functions> 92
IM 01E22C02-01EN
NE107 Status
Error Message Error Description CountermeasureDisplay BRAIN
F 023:Optbdmismatch 23:Optmismatch Mismatchofoptionboardwasdetected.
ContactYokogawaservicecenter.
F 024:OptbdEEPFAIL 24:OptEEPFAILFailureofEEPROM(optionboard)wasdetected.
ContactYokogawaservicecenter.
F 025:OptbdA/DFAIL 25:OptA/DFAIL FailureofA/D(optionboard)wasdetected.
ContactYokogawaservicecenter.
F 026:OptbdSPIFAIL 26:OptSPIFAIL FailureofSPI(optionboard)wasdetected.
ContactYokogawaservicecenter.
F 027:RestoreFAIL 27:RestoreFAIL Restoreofparameterswasfailed.
Retrytherestorefunctionoftheparameter.
F 028:IndbdFAIL
28-32:IndctFAIL
Failureofdisplayboardwasdetected.
Checkthattheambienttemperatureofdisplayiswithintherange.Ifthesituationdoesnotimprove,contactYokogawaservicecenter.
F 029:IndbdEEPFAILFailureofEEPROM(displayboard)wasdetected.
Turnonthepoweragainwithintherangeofthetemperaturespecifications.Ifthesituationdoesnotimprove,contactYokogawaservicecenter.
F 030:LCDdrvFAIL Failureofdisplaydriverwasdetected.
ContactYokogawaservicecenter.
F 031:Indbdmismatch Mismatchofdisplayboardwasdetected.
ContactYokogawaservicecenter.
F 032:IndcommERRCommunicationerrorofdisplayboardwasdetected.
Checkthatthemainanddisplayboardsareconnected.
F 033:microSDFAIL 33:microSDFAIL FailureofmicroSDcardwasdetected.
ReplacethemicroSDcard.
<4. Functions> 93
IM 01E22C02-01EN
�� Process alarmThedeviceworksnormally,butsomeissueoftheprocesscausesabnormalmeasurement.Maintenanceworkisneeded.
NE107 Status
Error Message Error Description CountermeasureDisplay BRAIN
S 050:Signaloverflow 50:Sigoverflow Failureofinputsignalwasdetected.
Checkthesignalandgroundingcablesareconnected.
S 051:Emptydetect 51:EmptydetectInsideofsensorwasdetectedtobeempty.(Emptypipedetection)
Fillthesensorwithwater.
N 052:H/LHH/LLalm 52:H/LHH/LLalmAflowrateexceedsthehigh/lowlimitsorhigh-high/low-lowlimits.
Checktheparametersettingrelatedtoflowrateandalarmhigh/lowlimitfunction.
S 053:Adhoverlv4 53:Adhoverlv4
TheresistancevalueoftheelectrodesexceededLevel 4.(Adhesiondetectionofinsulationtotheelectrode)
Recommendcleaningtheelectrode.
�� Setting alarmThedeviceworksnormally,buttheparametersettingerroroccurs.Parametersettingisneeded.
NE107 Status
Error Message Error Description CountermeasureDisplay BRAIN
S 060:SpancfgERR 60:SpancfgERRSettingerrorofflowspanwasdetected.(fulfill“0.05m/s<spanflowrate<16m/s”)
Checktheparametersettingrelatedtospan.
S 061:PVFcfgERR 61:PVFcfgERRFlownoisewassettoPVwhenflownoisecouldnotbeused.
CheckorchangethePVFLOWSELparametersetting.
S 062:AO14-20lmt 62:AO14-20lmtSettingerrorofanalogoutput1wasdetected.(fulfill“LRV<HRV”)
Checktheparametersettingrelatedtoanalogoutput1.
S 063:AO24-20lmt 63:AO24-20lmtSettingerrorofanalogoutput2wasdetected.(fulfill“LRV<HRV”)
Checktheparametersettingrelatedtoanalogoutput2.
S 064:AO1mltrng 64:AO1mltrng
Settingerrorofthemultirangefunctionwasdetected.(Spanvaluesettingsofeachrangedonotsatisfytherelationshipoflowerrange≤higherrange,multirangesettingsforthestatusinputandstatusoutputareerroneous,orthemultirangeandtheabsoluterangeareselectedatthesametime)
Checktheparametersettingrelatedtothemultirange.
N 065:H/LcfgERR 65:AOH/Lcfg
Settingerrorofthealarmhigh/lowlimitfunctionwasdetected.(fulfill“HRV-LRV>Hysteresis”and“HHRV-LLRV>Hysteresis”)
Checktheparametersettingrelatedtothealarmhigh/lowlimitfunction.
<4. Functions> 94
IM 01E22C02-01EN
NE107 Status
Error Message Error Description CountermeasureDisplay BRAIN
S 066:DensitycfgERR 66:DenscfgERRThesettingerrorofdensityvaluewasdetectedwhenPVwassettomassflowrate.
Checktheparametersettingsrelatedtodensity.
S 067:Pls1cfgERR 67:Pls1cfgERR Settingerrorofpulseoutput1wasdetected.
Checktheparametersettingrelatedtopulseoutput1.
S 068:Pls2cfgERR 68:Pls2cfgERR Settingerrorofpulseoutput2wasdetected.
Checktheparametersettingrelatedtopulseoutput2.
C 069:Nomisizecfg 69:Nomisizecfg
Configurationerrorofnominalsizewasdetected.(fulfill“0.99mm<sensornominalsize<3000.10mm(0.01inch<sensornominalsize<120.10inch)”)
Checktheparametersettingsrelatedtonominalsize.
C 070:AdhcfgERR 70:AdhcfgERR
Settingerrorofelectrodeadhesiondetectionfunctionwasdetected.(fulfill“Level1<Level2<Level3<Level4”)
Checktheparametersettingsrelatedtoadhesiondetection.
C 071:FLNcfgERR 71:FLNcfgERR
Settingerrorofflownoiseverificationfunctionwasdetected.(fulfill“Level1<Level2<Level3<Level4”)
Changetheparametersettingsrelatedtoflownoise.
C 072:Lognotstart 72:Lognotstart Dataloggingfailedtostart.
InsertthemicroSDcard.
<4. Functions> 95
IM 01E22C02-01EN
�� WarningThedeviceworksnormallyandmeasurementisalsonormalbutwarningoccurs.
NE107 Status
Error Message Error Description CountermeasureDisplay BRAIN
S 080:AO1saturate
80-81:AOsaturt
SaturationofAnalogoutput1wasdetected.
Checktheparametersettingsrelatedtotheprocessvalueandanalogoutput1.
S 081:AO2saturate SaturationofAnalogoutput2wasdetected.
Checktheparametersettingsrelatedtotheprocessvalueandanalogoutput2.
S 082:Pls1saturate
82-83:Plssaturt
Saturationofpulseoutput1wasdetected.
Checktheparametersettingrelatedtoprocessvalueandpulseoutput1.
S 083:Pls2saturate Saturationofpulseoutput2wasdetected.
Checktheparametersettingrelatedtoprocessvalueandpulseoutput2.
S 084:AIsaturate 84:AIsaturate SaturationofAnaloginputwasdetected.
Checkanaloginputvalueandanaloginputparametersetting.
C 085:Cablemiscon 85:Cablemiscon Misconnectionofcablewasdetected.
Checkthesignalcableandexcitationcableconnection.
C 086:Coilinsulation 86:Coilinsulate Insulationdeteriorationofthecoilwasdetected.
ContactYokogawaservicecenter.
M 087:Adhesionlv3 87:Adhoverlv3
TheresistancevalueoftheelectrodesexceededLevel 3.(Adhesiondetectionofinsulationtotheelectrode)
Recommendcleaningtheelectrode.
N 088:LCwarn 88:LCwarn Decreaseofconductivitywasdetected.
Checkfluidconductivity.
M 089:Insudetect 89:Insudetect Insulationdeteriorationofelectrodewasdetected.
ContactYokogawaservicecenter.
N 090:FLNoverlv3 90:FLNoverlv3FlownoiseexceededLevel 3.(Detectionofflownoise)
Checkifthefluidhasaproblem(conductivityandbubble).
N 091:FLNoverlv4 91:FLNoverlv4FlownoiseexceededLevel 4.(Detectionofflownoise)
Checkifthefluidhasaproblem(conductivityandbubble).
C 092:AZwarn 92:AZwarn Resultofzeroadjustmentexceeded10cm/s.
Checkfluidisstoppedwhenexecutingzeroadjustment.
C 093:Verifwarn 93:Verifwarn Interruptionofverificationfunctionwasdetected.
ExecutetheVerificationfunctionagain.
C 094:Factnoisewarn 94:FCnoisewarn Thefluctuationofflowbecamelarger.
Checkifthereisaproblemwiththefluid.
C 095:Simulateactive 95:Simactive
Atestmodefortheflowvelocity,volumetricflowrate,massflowrate,calorie,analogoutput,pulseoutput,statusinput,orstatusoutputwasexecuted.
Releasesimulationortestmode.
<4. Functions> 96
IM 01E22C02-01EN
NE107 Status
Error Message Error Description CountermeasureDisplay BRAIN
S 096:AO1fix
96-97:AOfixed
Itwasdetectedthatanalogoutput1isfixed.
Checkwhetheranalogoutput1isinthetestmodeornot.
S 097:AO2fix Itwasdetectedthatanalogoutput2isfixed.
Checkwhetheranalogoutput2isinthetestmodeornot.
S 098:Pls1fix
98-99:Plsfixed
Itwasdetectedthatpulseoutput1isfixed.
Checkwhetherpulseoutput1isinthetestmodeornot.
S 099:Pls2fix Itwasdetectedthatpulseoutput2isfixed.
Checkwhetherpulseoutput2isinthetestmodeornot.
S 100:AIfix 100:AIfix Itwasdetectedthattheanaloginputisfixed.
Checkwhethertheanaloginputisinthetestmodeornot.
C 101:Paramrestorerun 101:PrmrestoreTherestorefunctionoftheparameterwasexecuted.
-
N 102:Disp over 102:Disp overThenumberofdigitsavailablefordisplayexceededthelimit.
Checktheparametersettingrelatedtodisplay.
N 103:SDsizewarn 103:SDsizewarnFreespaceofmicroSDcarddecreasedtolessthan10%.
IncreasethefreespaceofthemicroSDcard.
M 104:Bkupincmplt 104:Bkupincmplt Parameterbackupfailed.Retrythebackupfunctionoftheparameter.
S 105:SDmismatch 105:SDmismatch MismatchofmicroSDcardwasdetected.
ReplacethemicroSDcard.
M 106:SDremovalERR 106:SDremovERR
RemovalofmicroSDcardfailed.
AfterexecutingtheremovalofthemicroSDcardwiththeparameter,removethemicroSDcardfromtheproduct.
N 131:Transmismatch 131:Trnmismatch Mismatchofsensorandtransmitterwasdetected.
ContactYokogawaservicecenter.
�� InformationTheinstrumentworksnormallyandmeasurementisalsonormal.Thesemessagesarejustreferenceinformation.
NE107 Status
Error Message Error Description CountermeasureDisplay BRAIN
N 120:Watchdog (*) ErrorofWatchdogtimerwasdetected.
ContactYokogawaservicecenter.
N 121:Poweroff (*) Power-offwasdetected. -
N 122:InstpowerFAIL (*) Instantaneouspowerfailurewasdetected. -
N 123:Parambkuprun (*) Parameterbackupfunctionisrunning. -
N 124:Datalogrun (*) Data-loggingfunctionisrunning. -
(*) NotdisplayedforSelfcheckofBRAINcommunication.
<4. Functions> 97
IM 01E22C02-01EN
4.8.2 Operation at the Time of ErrorThefollowingtableshowsthebehaviorofoutputsandreadingsatthetimeoferror.
�� System alarm
NE107 Error Message
OutputTotalizer Process
Value DisplayAlarm Warning Current Pulse/Status
F 010:MainCPUFAIL Active Non-Active Burnout Stop Stop Stop Not
defined
F 011:RevcalcFAIL Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”Normal
Operation Alarm
F 012:MainEEPFAIL Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”Normal
Operation Alarm
F 013:MainEEPdflt Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”Normal
Operation Alarm
F 014:SnsrbdFAIL Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”HoldPriorsetting Alarm
F 015:SnsrcommERR Active Non-Active Burnout AlarmOut Based on
“FailOpts”HoldPriorsetting Alarm
F 016:AD1FAIL[Sig] Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”HoldPriorsetting Alarm
F 017:AD2FAIL[Excit] Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”HoldPriorsetting Alarm
F 018:Coil open Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”HoldPriorsetting Alarm
F 019:Coilshort Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”HoldPriorsetting Alarm
F 020:ExciterFAIL Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”Normal
Operation Alarm
F 021:PWM1stop Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”Normal
Operation Alarm
F 022:PWM2stop Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”Normal
Operation Alarm
F 023:Optbdmismatch Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”Normal
Operation Alarm
F 024:OptbdEEPFAIL Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”Normal
Operation Alarm
F 025:OptbdA/DFAIL Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”Normal
Operation Alarm
F 026:OptbdSPIFAIL Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”Normal
Operation Alarm
F 027:RestoreFAIL Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”Normal
Operation Alarm
F 028:IndbdFAIL Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”Normal
Operation Alarm
F 029:IndbdEEPFAIL Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”Normal
Operation Alarm
F 030:LCDdrvFAIL Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”Normal
Operation Alarm
F 031:Indbdmismatch Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”Normal
Operation Alarm
F 032:IndcommERR Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”Normal
Operation Alarm
F 033:microSDFAIL Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”Normal
Operation Alarm
<4. Functions> 98
IM 01E22C02-01EN
�� Process alarm
NE107 Error Message
OutputTotalizer Process
Value DisplayAlarm Warning Current Pulse/Status
S 050:Signaloverflow Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”HoldPriorsetting
Alarm
S 051:Emptydetect Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”HoldPriorsetting
Alarm
N 052:H/LHH/LLalm Active Non-Active Normal Normal Continue Normal
Operation Alarm
S 053:Adhoverlv4 Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”Normal
Operation Alarm
�� Setting alarm
NE107 Error Message
OutputTotalizer Process
Value DisplayAlarm Warning Current Pulse/Status
S 060:SpancfgERR Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”Normal
Operation Alarm
S 061:PVFcfgERR Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”Normal
Operation Alarm
S 062:AO14-20lmt Active Non-Active Normal Normal Continue Normal
Operation Alarm
S 063:AO24-20lmt Active Non-Active Normal Normal Continue Normal
Operation Alarm
N 065:H/LcfgERR Active Non-Active
AlarmOut Normal Based on
“FailOpts”Normal
Operation Alarm
S 066:DensitycfgERR Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”Normal
Operation Alarm
S 067:Pls1cfgERR Active Non-Active
AlarmOut AlarmOut Continue Normal
Operation Alarm
S 068:Pls2cfgERR Active Non-Active
AlarmOut AlarmOut Continue Normal
Operation Alarm
C 069:Nomisizecfg Active Non-Active
AlarmOut AlarmOut Based on
“FailOpts”Normal
Operation Alarm
C 070:AdhcfgERR Active Non-Active Normal Normal Continue Normal
Operation Alarm
C 071:FLNcfgERR Active Non-Active Normal Normal Continue Normal
Operation Alarm
C 072:Lognotstart Active Non-Active Normal Normal Continue Normal
Operation Alarm
�� Warning
NE107 Error Message
OutputTotalizer Process
Value DisplayAlarm Warning Current Pulse/Status
S 080:AO1saturate Non-Active Active Normal Normal Continue Normal
Operation Warning
S 081:AO2saturate Non-Active Active Normal Normal Continue Normal
Operation Warning
S 082:Pls1saturate Non-Active Active Normal Normal Continue Normal
Operation Warning
S 083:Pls2saturate Non-Active Active Normal Normal Continue Normal
Operation Warning
<4. Functions> 99
IM 01E22C02-01EN
NE107 Error Message
OutputTotalizer Process
Value DisplayAlarm Warning Current Pulse/Status
S 084:AIsaturate Non-Active Active Normal Normal Continue Normal
Operation Warning
C 085:Cablemiscon Non-Active Active Normal Normal Continue Normal
Operation Warning
C 086:Coilinsulation Non-Active Active Normal Normal Continue Normal
Operation Warning
M 087:Adhesionlv3 Non-Active Active Normal Normal Continue Normal
Operation Warning
N 088:LCwarn Non-Active Active Normal Normal Continue Normal
Operation Warning
M 089:Insudetect Non-Active Active Normal Normal Continue Normal
Operation Warning
N 090:FLNoverlv3 Non-Active Active Normal Normal Continue Normal
Operation Warning
N 091:FLNoverlv4 Non-Active Active Normal Normal Continue Normal
Operation Warning
C 092:AZwarn Non-Active Active Normal Normal Continue Normal
Operation Warning
C 093:Verifwarn Non-Active Active Normal Normal Continue Normal
Operation Warning
C 094:Factnoisewarn Non-Active Active Normal Normal Continue Normal
Operation Warning
C 095:Simulateactive Non-Active Active Normal Normal Continue Normal
Operation Warning
S 096:AO1fix Non-Active Active Normal Normal Continue Normal
Operation Warning
S 097:AO2fix Non-Active Active Normal Normal Continue Normal
Operation Warning
S 098:Pls1fix Non-Active Active Normal Normal Continue Normal
Operation Warning
S 099:Pls2fix Non-Active Active Normal Normal Continue Normal
Operation Warning
S 100:AIfix Non-Active Active Normal Normal Continue Normal
Operation Warning
C 101:Paramrestorerun Non-Active Active Normal Normal Continue Normal
Operation Execute
N 102:Disp over Non-Active Active Normal Normal Continue Normal
Operation Warning
N 103:SDsizewarn Non-Active Active Normal Normal Continue Normal
Operation Warning
M 104:Bkupincmplt Non-Active Active Normal Normal Continue Normal
Operation Warning
S 105:SDmismatch Non-Active Active Normal Normal Continue Normal
Operation Warning
M 106:SDremovalERR Non-Active Active Normal Normal Continue Normal
Operation Warning
N 131:Transmismatch Non-Active Active Normal Normal Continue Normal
Operation Normal
<4. Functions> 100
IM 01E22C02-01EN
�� Information
NE107 Error Message
OutputTotalizer Process
Value DisplayAlarm Warning Current Pulse/Status
N 120:Watchdog Non-Active Active Normal Normal Continue Normal
Operation Normal
N 121:Poweroff Non-Active Active Normal Normal Continue Normal
Operation Normal
N 122:InstpowerFAIL Non-Active Active Normal Normal Continue Normal
Operation Normal
N 123:Parambkuprun Non-Active Active Normal Normal Continue Normal
Operation Execute
N 124:Datalogrun Non-Active Active Normal Normal Continue Normal
Operation Icon
<4. Functions> 101
IM 01E22C02-01EN
4.8.3 Alarm Display Setting(1) Alarm displayIfanerroroccursonthisproduct,analarmappearsonthedisplay.Thealarmdisplaymodesareclassifiedintotwotypes:onemodetodisplayaprocessvalueandalarmname,andanothermodetodisplayanalarmnameandaction.Ifmultiplealarmsoccuronthisproduct,theywillbedisplayedinsequenceonthedisplay.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Displayset►Optionalconfig►AlarmdisplayBRAIN J:DISPLAYSET►J28:DISPALARM
Thisparametersetsthealarmdisplay.
Fromthetablebelow,selectthealarmdisplay.Selection DescriptionDisplay BRAIN
Normal Normal Displaystheprocessvalueandalarmname.
F0425.ai
01 : 00VEL
AO1FLP
SET SFT INC
0.00000m/s
mA4.000018: Coil open
0.0%
Detail Detail Displaysthealarmnameandaction.
F0426.ai
01 : 00
SET SFT INC
System Alarm
Cut the power andcheck coil & EX cable.
018: Coil open
(2) Alarm display based on NAMUR NE107AprefixcanbeassignedtothealarmnamebasedonNAMURNE107.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Displayset►Optionalconfig►NE107displayBRAIN J:DISPLAYSET►J27:DISPNE107
ThisparametersetsthealarmdisplaybasedonNAMURNE107.
Fromthetablebelow,selectthealarmdisplay.Selection DescriptionDisplay BRAIN
Normal Normal Setstothenormalalarmdisplay.NE107 NE107 SetsthealarmdisplaybasedonNAMURNE107.
<4. Functions> 102
IM 01E22C02-01EN
4.8.4 Alarm History FunctionThealarmhistoryfunctioncanrecordanalarmthatoccurredinthepastinthehistory.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Diag/Service►Sts/Selftest►Alarm►Alarmrecord►(seebelow)BRAIN I:ALARM►(seebelow)
Parameter DescriptionDisplay BRAINRecordalarm1 I50:ALMRECORD1 Displaysthenameofthefirstnewalarm.
Recordtime1 I51:ALMTIME1 Displaystheoperationtimewhenthefirstnewalarmoccurs.
Recordalarm2 I52:ALMRECORD2 Displaysthenameofthesecondnewalarm.
Recordtime2 I53:ALMTIME2 Displaystheoperationtimewhenthesecondnewalarmoccurs.
Recordalarm3 I54:ALMRECORD3 Displaysthenameofthethirdnewalarm.
Recordtime3 I55:ALMTIME3 Displaystheoperationtimewhenthethirdnewalarmoccurs.
Recordalarm4 I56:ALMRECORD4 Displaysthenameofthefourthnewalarm.
Recordtime4 I57:ALMTIME4 Displaystheoperationtimewhenthefourthnewalarmoccurs.
Theoperationtimewhenanalarmoccurredisdisplayedintheformatof“dddddDhh:mm”.“dddddD”indicatestheday,“hh”indicatesthehour,and“mm”indicatestheminute.
Example: “00031D12:34”isdisplayed.Thisexampleshowsthatanalarmhasoccurredwhentheproducthasbeenoperatedfor31days,12hours,and34minutes.
<4. Functions> 103
IM 01E22C02-01EN
4.8.5 Alarm Mask FunctionThealarmmaskfunctioncanbeconfiguredsothatasetalarmismasked,alarmnotificationsarehidden,andthealarmhistoryisnotrecorded.Themaskfunctioncanbesetforboththealarmnotificationandalarmrecord,respectively.Thissettingcanbeconfiguredwiththefollowingparameters.
�� AlarmnotificationmaskIfthealarmnotificationmaskfunctionisturned“On”,itdisablesalarmnotification.Settingexampleforalarm“50:Sigoverflow”: Todisablethealarmnotification,set“Sigoverflowon”toI32:OUTMSKSET2. Toenablethealarmnotification,set“Sigoverflowoff”toI32:OUTMSKSET2.Menu path
Display Devicesetup►Diag/Service►Sts/Selftest►Alarm►Alarmoutmask►(seebelow)BRAIN I:ALARM►(seebelow)
Parameter DescriptionDisplay BRAIN
Mask1-1 I30:OUTMSKSET1(forsetting)I31:OUTMSKSTS1(forchecking)
Specifiesthemaskfunctionforalarmnotification1-1.
Mask1-2 Specifiesthemaskfunctionforalarmnotification1-2.
Mask2-1 I32:OUTMSKSET2(forsetting)I33:OUTMSKSTS2(forchecking)
Specifiesthemaskfunctionforalarmnotification2-1.
Mask2-2 Specifiesthemaskfunctionforalarmnotification2-2.
Mask3-1 I34:OUTMSKSET3(forsetting)I35:OUTMSKSTS3(forchecking)
Specifiesthemaskfunctionforalarmnotification3-1.
Mask3-2 Specifiesthemaskfunctionforalarmnotification3-2.
�� Alarm record maskIfthealarmrecordmaskfunctionisturned“On”,itdisablesthealarmrecord.Settingexampleforalarm“51:Emptydetect”: Todisablethealarmrecord,set“Emptydetecton”toI42:RECMSKSET2. Toenablethealarmrecord,set“Emptydetectoff”toI42:RECMSKSET2.Menu path
Display Devicesetup►Diag/Service►Sts/Selftest►Alarm►Alarmrecordmask►(seebelow)BRAIN I:ALARM►(seebelow)
Parameter DescriptionDisplay BRAINMask1-1 I40:RECMSKSET1(for
setting)I41:RECMSKSTS1(forchecking)
Specifiesthemaskfunctionforalarmrecord1-1.
Mask1-2 Specifiesthemaskfunctionforalarmrecord1-2.
Mask2-1 I42:RECMSKSET2(forsetting)I43:RECMSKSTS2(forchecking)
Specifiesthemaskfunctionforalarmrecord2-1.
Mask2-2 Specifiesthemaskfunctionforalarmrecord2-2.
DuetothecharacteristicsofBRAINcommunication,parametersforsettingandcheckingareseparate.Usethesettingparameterwhencheckingthepresentmaskstateandthesettingparameterwhenchangingthemasksettings.
<4. Functions> 104
IM 01E22C02-01EN
NOTENotethatanalarmmaskedbythealarmnotificationmaskfunctionisnotrecordedinthealarmrecord.
Thealarmmaskfunctionsettingisasfollows:Parameter name Indicatesthenameofthemasksettingparameter.Alarm name Indicatesthealarmname.
Default value Indicatesthedefaultvalue(uponshipmentfromthemanufacturingfactory).(:Masked,-:Notmasked)
Attribute Indicateswhetherthemasksettingisenabledordisabled.(:Settingenabled,-:Settingdisabled)
�� AlarmnotificationmaskfunctionDisplay BRAIN Default
Value AttributeParameter Name Alarm Name Parameter Name Alarm Name- 010:MainCPUFAIL - 10:MainCPUFAIL - -- 011:RevcalcFAIL - 11:RevcalFAIL - -- 012:MainEEPFAIL - 12:MainEEPFAIL - -
Mask1-1 013:MainEEPdflt I30:OUTMSKSET1 13:MainEEPdflt -
- 014:SnsrbdFAIL - 14:SnsrbdFAIL - -- 015:SnsrcommERR - 15:SnsrcommERR - -- 016:AD1FAIL[Sig] - 16:A/D1FAIL - -- 017:AD2FAIL[Excit] - 17:A/D2FAIL - -- 018:Coil open - 18:Coil open - -- 019:Coilshort - 19:Coilshort - -- 020:ExciterFAIL - 20:ExciterFAIL - -
Mask1-2 021:PWM1stop I30:OUTMSKSET1 21-22:PWMstop
- Mask1-2 022:PWM2stop -
Mask1-2 023:Optbdmismatch I30:OUTMSKSET1 23:Optmismatch -
Mask1-2 024:OptbdEEPFAIL I30:OUTMSKSET1 24:OptEEPFAIL -
Mask1-2 025:OptbdA/DFAIL I30:OUTMSKSET1 25:OptA/DFAIL -
Mask1-2 026:OptbdSPIFAIL I30:OUTMSKSET1 26:OptSPIFAIL -
- 027:RestoreFAIL I30:OUTMSKSET1 27:RestoreFAIL - -
Mask1-2 028:IndbdFAIL
I30:OUTMSKSET1 28-32:IndctFAIL
Mask1-2 029:IndbdEEPFAIL - Mask1-2 030:LCDdrvFAIL - Mask1-2 031:Indbdmismatch - Mask1-2 032:IndcommERR -
Mask1-2 033:microSDFAIL I30:OUTMSKSET1 33:microSDFAIL -
Mask2-1 050:Signaloverflow I32:OUTMSKSET2 50:Sigoverflow -
Mask2-1 051:Emptydetect I32:OUTMSKSET2 51:Emptydetect -
Mask2-1 052:H/LHH/LLalm I32:OUTMSKSET2 52:H/LHH/LLalm
<4. Functions> 105
IM 01E22C02-01EN
Display BRAIN Default Value AttributeParameter Name Alarm Name Parameter Name Alarm Name
Mask2-1 053:Adhoverlv4 I32:OUTMSKSET2 53:Adhoverlv4
Mask2-1 060:SpancfgERR I32:OUTMSKSET2 60:SpancfgERR -
Mask2-1 061:PVFcfgERR I32:OUTMSKSET2 61:PVFcfgERR -
Mask2-1 062:AO14-20lmt I32:OUTMSKSET2 62:AO14-20lmt -
Mask2-1 063:AO24-20lmt I32:OUTMSKSET2 63:AO24-20lmt -
Mask2-1 064:AO1mltrng I32:OUTMSKSET2 64:AO1mltrng -
Mask2-1 065:H/LcfgERR I32:OUTMSKSET2 65:AOH/Lcfg -
Mask2-1 066:DensitycfgERR I32:OUTMSKSET2 66:DenscfgERR -
Mask2-2 067:Pls1cfgERR I32:OUTMSKSET2 67:Pls1cfgERR -
Mask2-2 068:Pls2cfgERR I32:OUTMSKSET2 68:Pls2cfgERR -
Mask2-2 069:Nomisizecfg I32:OUTMSKSET2 69:Nomisizecfg -
Mask2-2 070:AdhcfgERR I32:OUTMSKSET2 70:AdhcfgERR -
Mask2-2 071:FLNcfgERR I32:OUTMSKSET2 71:FLNcfgERR -
Mask2-2 072:Lognotstart I32:OUTMSKSET2 72:Lognotstart -
Mask2-2 080:AO1saturate I32:OUTMSKSET2 80-81:AOsaturt
- Mask2-2 081:AO2saturate Mask2-2 082:Pls1saturate I32:OUTMSK
SET2 82-83:Plssaturt Mask2-2 083:Pls2saturate
Mask2-2 084:AIsaturate I32:OUTMSKSET2 84:AIsaturate
Mask2-2 085:Cablemiscon I32:OUTMSKSET2 85:Cablemiscon -
Mask2-2 086:Coilinsulation I32:OUTMSKSET2 86:Coilinsulate
Mask2-2 131:Transmismatch I32:OUTMSKSET2 131:Trnmismatch -
Mask3-1 087:Adhesionlv3 I34:OUTMSKSET3 87:Adhoverlv3
Mask3-1 088:LCwarn I34:OUTMSKSET3 88:LCwarn
Mask3-1 089:Insudetect I34:OUTMSKSET3 89:Insudetect
Mask3-1 090:FLNoverlv3 I34:OUTMSKSET3 90:FLNoverlv3
Mask3-1 091:FLNoverlv4 I34:OUTMSKSET3 91:FLNoverlv4
Mask3-1 092:AZwarn I34:OUTMSKSET3 92:AZwarn
Mask3-1 093:Verifwarn I34:OUTMSKSET3 93:Verifwarn
Mask3-1 094:Factnoisewarn I34:OUTMSKSET3 94:FCnoisewarn
Mask3-1 095:Simulateactive I34:OUTMSKSET3 95:Simactive -
<4. Functions> 106
IM 01E22C02-01EN
Display BRAIN Default Value AttributeParameter Name Alarm Name Parameter Name Alarm Name
Mask3-1 096:AO1fix I34:OUTMSKSET3 96-97:AOfixed
-
Mask3-1 097:AO2fix -
Mask3-1 098:Pls1fix I34:OUTMSKSET3 98-99:Plsfixed
-
Mask3-1 099:Pls2fix -
Mask3-1 100:AIfix I34:OUTMSKSET3 100:AIfix -
Mask3-2 101:Paramrestorerun I34:OUTMSKSET3 101:Prmrestore
Mask3-2 102:Disp over I34:OUTMSKSET3 102:Disp over
Mask3-2 103:SDsizewarn I34:OUTMSKSET3 103:SDsizewarn
Mask3-2 104:Bkupincmplt I34:OUTMSKSET3 104:Bkupincmplt -
Mask3-2 105:SDmismatch I34:OUTMSKSET3 105:SDmismatch
Mask3-2 106:SDremovalERR I34:OUTMSKSET3 106:SDremovERR
Mask3-2 120:Watchdog*1 I34:OUTMSKSET3 120:Watchdog*1
Mask3-2 121:Poweroff*1 I34:OUTMSKSET3 121:Poweroff*1
Mask3-2 122:InstpowerFAIL*1 I34:OUTMSKSET3 122:InstPWFAIL*1
Mask3-2 123:Parambkuprun I34:OUTMSKSET3 123:bkuprun
Mask3-2 124:Datalogrun I34:OUTMSKSET3 124:Datalogrun
*1: Recordedinthealarmrecordregardlessofthesettingsofthealarmnotificationmaskfunction.
<4. Functions> 107
IM 01E22C02-01EN
�� Alarm record mask functionDisplay BRAIN Default
Value AttributeParameter Name Alarm Name Parameter Name Alarm Name- 010:MainCPUFAIL - 10:MainCPUFAIL - -- 011:RevcalcFAIL - 11:RevcalFAIL - -- 012:MainEEPFAIL - 12:MainEEPFAIL - -
Mask1-1 013:MainEEPdflt I40:RECMSKSET1 13:MainEEPdflt -
- 014:SnsrbdFAIL - 14:SnsrbdFAIL - -- 015:SnsrcommERR - 15:SnsrcommERR - -- 016:AD1FAIL[Sig] - 16:A/D1FAIL - -- 017:AD2FAIL[Excit] - 17:A/D2FAIL - -- 018:Coil open - 18:Coil open - -- 019:Coilshort - 19:Coilshort - -
Mask1-2 021:PWM1stop I40:RECMSKSET1 21-22:PWMstop
- Mask1-2 022:PWM2stop -
Mask1-2 023:Optbdmismatch I40:RECMSKSET1 23:Optmismatch -
Mask1-2 024:OptbdEEPFAIL I40:RECMSKSET1 24:OptEEPFAIL -
Mask1-2 025:OptbdA/DFAIL I40:RECMSKSET1 25:OptA/DFAIL -
Mask1-2 026:OptbdSPIFAIL I40:RECMSKSET1 26:OptSPIFAIL -
- 027:RestoreFAIL - 27:RestoreFAIL - -Mask1-2 028:IndbdFAIL
I40:RECMSKSET1 28-32:IndctFAIL
- Mask1-2 029:IndbdEEPFAIL - Mask1-2 030:LCDdrvFAIL - Mask1-2 031:Indbdmismatch - Mask1-2 032:IndcommERR -
Mask1-2 033:microSDFAIL I40:RECMSKSET1 33:microSDFAIL -
Mask2-1 050:Signaloverflow I42:RECMSKSET2 50:Sigoverflow -
Mask2-1 051:Emptydetect I42:RECMSKSET2 51:Emptydetect -
Mask2-1 052:H/LHH/LLalm I42:RECMSKSET2 52:H/LHH/LLalm -
Mask2-1 053:Adhoverlv4 I42:RECMSKSET2 53:Adhoverlv4 -
- 060:SpancfgERR - 60:SpancfgERR -- 061:PVFcfgERR - 61:PVFcfgERR -- 062:AO14-20lmt - 62:AO14-20lmt -- 063:AO24-20lmt - 63:AO24-20lmt -- 064:AO1mltrng - 64:AO1mltrng -- 065:H/LcfgERR - 65:AOH/Lcfg -- 066:DensitycfgERR - 66:DenscfgERR -- 067:Pls1cfgERR - 67:Pls1cfgERR -- 068:Pls2cfgERR - 68:Pls2cfgERR -- 069:Nomisizecfg - 69:Nomisizecfg -- 070:AdhcfgERR - 70:AdhcfgERR -- 071:FLNcfgERR - 71:FLNcfgERR -- 072:Lognotstart - 72:Lognotstart -- 080:AO1saturate
- 80-81:AOsaturt -- 081:AO2saturate -- 082:Pls1saturate
- 82-83:Plssaturt -
- 083:Pls2saturate -
<4. Functions> 108
IM 01E22C02-01EN
Display BRAIN Default Value AttributeParameter Name Alarm Name Parameter Name Alarm Name
- 084:AIsaturate - 84:AIsaturate -
Mask2-2 085:Cablemiscon I42:RECMSKSET2 85:Cablemiscon -
- 086:Coilinsulation - 86:Coilinsulate -- 131:Transmismatch - 131:Trnmismatch -- 087:Adhesionlv3 - 87:Adhoverlv3 -- 088:LCwarn - 88:LCwarn -- 089:Insudetect - 89:Insudetect -- 090:FLNoverlv3 - 90:FLNoverlv3 -- 091:FLNoverlv4 - 91:FLNoverlv4 -- 092:AZwarn - 92:AZwarn -- 093:Verifwarn - 93:Verifwarn -- 094:Factnoisewarn - 94:FCnoisewarn -- 095:Simulateactive - 95:Simactive -- 096:AO1fix
- 96-97:AOfixed -- 097:AO2fix -- 098:Pls1fix
- 98-99:Plsfixed -- 099:Pls2fix -- 100:AIfix - 100:AIfix -- 101:Paramrestorerun - 101:Prmrestore -- 102:Disp over - 102:Disp over -- 103:SDsizewarn - 103:SDsizewarn -- 104:Bkupincmplt - 104:Bkupincmplt -- 105:SDmismatch - 105:SDmismatch -- 106:SDremovalERR - 106:SDremovERR -- 120:Watchdog*1 - 120:Watchdog*1 - -- 121:Poweroff*1 - 121:Poweroff*1 - -- 122:InstpowerFAIL*1 - 122:InstPWFAIL*1 - -- 123:Parambkuprun - 123:bkuprun -- 124:Datalogrun - 124:Datalogrun -
*1: Recordedinthealarmrecordregardlessofthesettingsofthealarmnotificationmaskfunction.
<4. Functions> 109
IM 01E22C02-01EN
4.8.6 Burnout Direction DisplayTheburnoutfunctionsetsananalogoutputdirectionwhenaCPUfailureoccurs.Theanalogoutputdirectionatburnoutcanbecheckedwiththefollowingparameters.Menu path
Display Devicesetup►Diag/Service►H/Lalarmcfg►4-20burnoutBRAIN I:ALARM►I20:4-20BURNOUT
DisplaysburnoutdirectionSelection
DescriptionDisplay BRAIN
High High Outputs21.6mAormoreatburnout.Low Low Outputs2.4mAorlessatburnout.
NOTEIftheoptionalcodeC1orC2isselected,theburnoutdirectionissetto“High”.Otherwise,thedirectionissetto“Low”.Also,theburnoutdirectioncanbechangedwiththehardwareburnoutswitch.Fordetailsaboutthehardwareburnoutswitch,besuretorefertotheInstallationManual.
4.8.7 Alarm Terminal Active Direction SettingThealarmterminalisactivewhenasystemalarm,processalarm,and/orsettingalarmoccurs.TheONorOFFsettingofthealarmterminalcanbeactivated.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Diag/Service►Sts/Selftest►Alarm►Alarmoutput►ActivemodeBRAIN I:ALARM►I15:ALMOUTACT
Fromthetablebelow,selecttheactivedirection.Selection
DescriptionDisplay BRAIN
Onactive Onactive Specifiesactivewhenalarmterminalison.Offactive Offactive Specifiesactivewhenthealarmterminalisoff.
<4. Functions> 110
IM 01E22C02-01EN
4.9 Display4.9.1 Language Setting
Thelanguagetobeshownonthedisplaycanbeselectedfromninelanguages.Selectthedesireddisplaylanguagefromthelanguagesthatareincludedinthelanguagepackagespecifiedatthetimeoforder.Thissettingcanbeconfiguredwiththefollowingparameters.
�� Selecting languageMenu path
Display Devicesetup►LanguageBRAIN J:DISPLAYSET►J32:LANGUAGE
Thisparameterspecifiesthelanguagetobeusedonthedisplay.Fromthetablebelow,selectthelanguagetobeusedonthedisplay.
SelectionDescription
Display BRAINEnglish English
Package1:Japanese,English,French,German,Italian,Spanish,Portuguese,Russian
Package2:English,Chinese
French FrenchGerman GermanItalian ItalianSpanish SpanishPortuguese PortugueseRussian RussianChinese ChineseJapanese Japanese
�� Display of language packageMenu path
Display Devicesetup►Detailedsetup►Displayset►Optionalconfig►LanguagepackageBRAIN J:DISPLAYSET►J52:LANGPACKAGE
Thisparameterindicatesthelanguagepackageforthedisplay.
<4. Functions> 111
IM 01E22C02-01EN
4.9.2 Display Item SettingThisproductcanshowuptoeightitemsonthedisplaybyscrollingit.Eachdisplayitemistobesettotheeight-linemode.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Displayset►Lineselect►(seebelow)BRAIN J:DISPLAYSET►(seebelow)
Parameter DescriptionDisplay BRAINLine 1 J10:LINE1SEL Specifiesitem1tobeshownonthedisplay.Line 2 J11:LINE2SEL Specifiesitem2tobeshownonthedisplay.Line 3 J12:LINE3SEL Specifiesitem3tobeshownonthedisplay.Line 4 J13:LINE4SEL Specifiesitem4tobeshownonthedisplay.Line 5 J14:LINE5SEL Specifiesitem5tobeshownonthedisplay.Line 6 J15:LINE6SEL Specifiesitem6tobeshownonthedisplay.Line 7 J16:LINE7SEL Specifiesitem7tobeshownonthedisplay.Line 8 J17:LINE8SEL Specifiesitem8tobeshownonthedisplay.
Fromthetablebelow,selecttheitemstobeshownonthedisplay.Selection
DescriptionDisplay BRAIN
None None Doesnotdisplayitems(item1isnotselectable).
Flowrate(%) Flowrate(%) DisplaystheflowrateforthespanofthePV-mappedprocessvalueofSubsection4.1.2.
PV PV DisplaysthePV-mappedprocessvalueofSubsection4.1.2.
Velocity Velocity Displaystheflowvelocity.Volumeflow Volumeflow Displaysthevolumetricflowrate.Massflow Massflow Displaysthemassflowrate.
Flowrate(%Bar) Flowrate(%Bar)DisplaystheflowrateforthespanofthePV-mappedprocessvalueofSubsection4.1.2byusingabargraph.
Calorie Calorie Displaysthecalorie.Totalizer1*1 Totalizer1 Displaysthetotalizedvalueoftotalizer1.Totalizer2*1 Totalizer2 Displaysthetotalizedvalueoftotalizer2.Totalizer3*1 Totalizer3 Displaysthetotalizedvalueoftotalizer3.Tagnumber Tagnumber Displaysthetagnumber.Communprotocol CommunProtocol Displaysthecommunicationprotocol.
Adhesion Adhesion Displaystheadhesionleveloftheelectrodeadhesiondetectingfunction.
Analogout1 Analogout1 Displaysthecurrentvalueofanalogoutput1Analogout2 Analogout2 Displaysthecurrentvalueofanalogoutput2
Flownoiselevel Flownoiselevel Displaysthenoiseleveloftheflownoisediagnosisfunction.
Totalizer1count*1 Totalizer1count Displaysthecountvalueoftotalizer1.Totalizer2count*1 Totalizer2count Displaysthecountvalueoftotalizer2.Totalizer3count*1 Totalizer3count Displaysthecountvalueoftotalizer3.
*1: Theupper/lowerlimitsonthedisplayarelimitedbythenumberofdigitsthatcanbedisplayedaccordingtothesettingsofthedecimalpoint.Forexample,ifthedecimalpointpositionofTotalizer3issettodecimalpoint2,theupperandlowerlimitsofTotalizer3are+999999.99and-999999.99.Ifthetotalizedvalueexceedstheupperandlowerlimits,itisreset to 0.
Fordetailsaboutdecimal-pointpositionsetting,refertoSubsection4.9.3. Ifthetotalizedvalueisresetto0,thetotalizedvalueonthedisplaymaybedifferentfromthetotalizedvaluein
Subsection4.2.2.Thisisbecausethenumberofdigitsthatcanbedisplayedisdifferentforbothvalues.Whenre-totalizing,usethereset/presetfunctionforthetotalizedvalue.
Theupperandlowerlimitsofthetotalizedcountvalueare+99999999and-99999999.Ifthetotalizedcountvalueexceedstheseupperandlowerlimits,thecountvalueisresetto0.
<4. Functions> 112
IM 01E22C02-01EN
NOTEWhenMainsoftrev(K50:MAINBREV/Mainsoftrev)isR2.01.02orearlier,orIndsoftrev(K52:INDBREV/Indsoftrev)isR2.01.01orearlier,theupper/lowerlimitsonthedisplayarelimitedbythenumberofdigitsthatcanbedisplayedaccordingtothesettingsofthedecimalpoint.Forexample,ifthedecimalpointpositionofTotalizer3issettodecimalpoint2,theupperandlowerlimitsofTotalizer3are+999999.99and-999999.99.Fordetailsaboutdecimal-pointpositionsetting,refertoSubsection4.9.3.Theupperandlowerlimitsofthetotalizedcountvalueare+99999999and-99999999.Ifthetotalizedvalueandtotalizedcountvalueexceedtheseupperandlowerlimits,theyareheldattheupperlimitorlowerlimit.Fordetailsabouthowtocheckthedevicerevision(MainsoftrevandIndsoftrev),refertoSubsection4.10.2.
4.9.3 Decimal-Point Position SettingThenumberofdecimalplacescanbesettotheautomaticadjustmentorfixmodeforthePV-mappedtotalizedvalueorprocessvalueofSubsection4.1.2.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Displayset►Dispformat►(seebelow)BRAIN J:DISPLAYSET►(seebelow)
Parameter DescriptionDisplay BRAIN
FormatPV J20:FORMATFR Specifiesthedecimal-pointpositionfortheprocessvaluePV-mappedinSubsection4.1.2.
Formattotal1 J21:FORMATTOT1 Specifiesthedecimal-pointpositionforthetotalizedvalueoftotalizer1.
Formattotal2 J22:FORMATTOT2 Specifiesthedecimal-pointpositionforthetotalizedvalueoftotalizer2.
Formattotal3 J23:FORMATTOT3 Specifiesthedecimal-pointpositionforthetotalizedvalueoftotalizer3.
Fromthetablebelow,selectthepositionofthedecimalpoint.Processvariable
Selection DescriptionDisplay BRAIN
Auto Auto AutomaticallyadjuststhenumberofdigitsafterthedecimalpointwiththePV-mappedprocessvalue.*1
0digit 0digit Fixesthenumberofdecimalplacesto“0”.1digit 1digit Fixesthenumberofdecimalplacesto“1”.2digit 2digit Fixesthenumberofdecimalplacesto“2”.3digit 3digit Fixesthenumberofdecimalplacesto“3”.4digit 4digit Fixesthenumberofdecimalplacesto“4”.5digit 5digit Fixesthenumberofdecimalplacesto“5”.
Auto2 Auto2AutomaticallyadjuststhenumberofdigitsafterthedecimalpointwiththespanofthePV-mappedprocessvalue.*2
<4. Functions> 113
IM 01E22C02-01EN
TotalizedvalueSelection DescriptionDisplay BRAIN
Auto Auto Automaticallyadjuststhenumberofdigitsafterthedecimalpointwithatotalizedvalue.
0digit 0digit Fixesthenumberofdecimalplacesto“0”.1digit 1digit Fixesthenumberofdecimalplacesto“1”.2digit 2digit Fixesthenumberofdecimalplacesto“2”.3digit 3digit Fixesthenumberofdecimalplacesto“3”.4digit 4digit Fixesthenumberofdecimalplacesto“4”.5digit 5digit Fixesthenumberofdecimalplacesto“5”.6digit 6digit Fixesthenumberofdecimalplacesto“6”.7digit 7digit Fixesthenumberofdecimalplacesto“7”.
*1: When“Auto”isselected,thedisplayformatisautomaticallyswitchedinresponsetotheprocessvalueselectedforPV. Thetablebelowshowsdetails.
Judgment range Supported display format|processvalue|≥ 100000 0digit
100000 >|processvalue|≥ 10000.0 1digit10000.0 >|processvalue|≥ 1000.00 2digit1000.00 >|processvalue|≥ 100.000 3digit100.000 >|processvalue|≥ 10.0000 4digit10.0000 >|processvalue|≥ 0.00000 5digit
Sincethedisplayformatisswitchedinresponsetotheprocessvalues,thedisplayformatmaybeswitchedwhentheproductisusednearthethresholdlistedintheabovetable.
*2: When“Auto2”isselected,thedisplayformatisautomaticallyswitchedinresponsetothespanvalueoftheprocessselectedforPV.
Thetablebelowshowsdetails.Judgment range Supported display format
|spanvalue|> 900 0digit900 ≥|spanvalue|> 90 1digit90 ≥|spanvalue|> 9 2digit9 ≥|spanvalue|> 0 3digit
Sincethedisplayformatisswitchedinresponsetospanvaluesoftheprocess,thedisplayformatisnotchangedaslongasthespanisnotchanged.
<4. Functions> 114
IM 01E22C02-01EN
4.9.4 Display Line Count and Scroll SettingsThisproductcanshowuptoeightitemsonthedisplaybyscrolling,withfourlinesmax.shownatatime.ThescrollmethodcanbeselectedfromtheautomaticdisplayswitchingorthedisplayswitchingusingtheIRswitch.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Displayset►Optionalconfig►(seebelow)BRAIN J:DISPLAYSET►(seebelow)
Parameter DescriptionDisplay BRAIN
Linemode J25:DISPLINE Specifiesthenumberoflinestobeshownonthedisplay.*1
Scrollmode J29:DISPSCROLL Specifiesthedisplayscrollmethod.*2
*1: Fromthetablebelow,selectthenumberofdisplaylines. Thefontsizeisautomaticallyadjusteddependingonthenumberofthedisplayedline.
SelectionDescription
Display BRAIN1line(big) 1Line(Big) 1-linedisplaywithoutunit.
Thenumericvalueisdisplayedinalargefont.
1 line 1 Line 1-linedisplaywithunit.
2 line 2 Line 2-linedisplaywithunits
3 line 3 Line 3-linedisplaywithunits
<4. Functions> 115
IM 01E22C02-01EN
SelectionDescription
Display BRAIN4 line 4 Line 4-linedisplaywithunits.
*2: Fromthetablebelow,selectthescrollmethod.Selection DescriptionDisplay BRAIN
Off Off Doesnotscroll.Manual Manual SpecifiestothescrollusingtheIRswitch.
Auto(2s) Auto(2s) Specifiestotheautomaticscrollat2-secondintervals.
Auto(4s) Auto(4s) Specifiestotheautomaticscrollat4-secondintervals.
Auto(8s) Auto(8s) Specifiestotheautomaticscrollat8-secondintervals.
NOTEThedefaultvalueofthescrollmethod(J29:DISPSCROLL/Scrollmode)is“Off”.Ifthevalueofthescrollmethodissetto“Off”,the5thlineandbeyondcannotbecheckedonthedisplay. Todisplaythe5thlineandbeyond,setanoptionotherthan“Off”.
NOTEWhenMainsoftrev(K50:MAINBREV/Mainsoftrev)isR2.01.02orearlier,thedefaultvalueofthescrollmethod(J29:DISPSCROLL/Scrollmode)is“Manual”.Ifthevalueofthescrollmethodissetto“Manual”,thedisplaylineisscrolledeachtimetheinfraredswitch[INC]reacts.
<4. Functions> 116
IM 01E22C02-01EN
4.9.5 Trend Graph SettingThetrendgraphdisplayfunctiondisplaysthetimechangeoftheselecteditemasatrendgraph.Uptofouritemscanbeselectedforatrendgraph.Atrendgraphisscaledautomatically,andthetimeaxisflowsfromthelefttotheright.
01 : 00
SET SFT INC
01/01
F0432.ai
Process valueDisplay start time
Display end time
01:00
01/0101:01
PRV
(1) Trend graph display settingTheprocessvalueandlowlimitvalue/highlimitvaluethatcanbeshowninatrendgraphareasfollows:
Process value Display Low limit High limitFlowrate(%) FLP 0% 100%PV PRV 0 Spanvaluespecifiedin
Subsection4.1.5Flowvelocity VELVolumetricflowrate VFLMassflowrate MFLCalorie CALAnalogoutput1 AO1 4mA 20mAAnalogoutput2 AO2Totalizer1 TL1 Presetvaluespecifiedin
Subsection4.2.7TargetvalueofthetotalizerspecifiedinSubsection4.2.3Totalizer2 TL2
Totalizer3 TL3
Atrendgraphcanbesetwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Displayset►Optionalconfig►DisplaymodeBRAIN J:DISPLAYSET►J40:DISPMODE
Fromthetablebelow,selectthetrendgraphdisplay.Selection DescriptionDisplay BRAIN
Normal Normal Doesnotdisplayatrendgraph(normaldisplay).Trend Trend Displaysatrendgraph.
NOTEThetrendgraphdisplayfunctionisnotsynchronizedwiththemultirangeandtheforward/reverserange.
(2) Trend graph display item settingThissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Displayset►Trendselect►(seebelow)BRAIN J:DISPLAYSET►(seebelow)
Parameter DescriptionDisplay BRAINTrend1 J43:TREND1SEL Specifiesitem1tobeshowninatrendgraph.Trend2 J44:TREND2SEL Specifiesitem2tobeshowninatrendgraph.Trend3 J45:TREND3SEL Specifiesitem3tobeshowninatrendgraph.
<4. Functions> 117
IM 01E22C02-01EN
Trend4 J46:TREND4SEL Specifiesitem4tobeshowninatrendgraph.
Fromthetablebelow,selectthedisplayitemoftrendgraph.Selection DescriptionDisplay BRAIN
None None Doesnotsetanyitems(item1isnotselectable).
Flowrate(%) Flowrate(%) SpecifiestheflowrateforthespanofthePV-mappedprocessvalueofSubsection4.1.2.
PV PV SpecifiesthePV-mappedprocessvalueofSubsection4.1.2.
Velocity Velocity Specifiesthedisplayitemtotheflowvelocity.Volumeflow Volumeflow Specifiesthedisplayitemtothevolumetricflowrate.Massflow Massflow Specifiesthedisplayitemtothemassflowrate.Calorie Calorie Specifiesthedisplayitemtocalorie.Analogout1 Analogout1 Specifiestothelowlimitofanalogoutput1.Analogout2 Analogout2 Specifiestothelowlimitofanalogoutput2.Totalizer1 Totalizer1 Specifiesthetotalizedvalueoftotalizer1.Totalizer2 Totalizer2 Specifiesthetotalizedvalueoftotalizer2.Totalizer3 Totalizer3 Specifiesthetotalizedvalueoftotalizer3.
(3) Update period settingThissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Displayset►Optionalconfig►PeriodBRAIN J:DISPLAYSET►J26:DISPPERIOD
Fromthetablebelow,selectthetrendgraphupdateperiod.Selection DescriptionDisplay BRAIN
0.2 s 0.2s Setstheupdateperiodto0.2sec.0.4 s 0.4s Setstheupdateperiodto0.4sec.1.0 s 1.0s Setstheupdateperiodto1sec.2.0 s 2.0s Setstheupdateperiodto2sec.4.0 s 4.0s Setstheupdateperiodto4sec.8.0 s 8.0s Setstheupdateperiodto8sec.
(4) Trend graph high/low limit settingThissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Displayset►Optionalconfig►(seebelow)BRAIN J:DISPLAYSET►(seebelow)
Parameter DescriptionDisplay BRAINTrendofflnURV J42:OFFTRENDURV Specifiesthehighlimittodisplayinatrendgraph.TrendofflnLRV J41:OFFTRENDLRV Specifiesthelowlimittodisplayinatrendgraph.
<4. Functions> 118
IM 01E22C02-01EN
4.9.6 Other Settings(1) Display contrast settingThecontrastofthedisplaycanbeadjustedto11levels(+5to-5).Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Displayset►Optionalconfig►ContrastBRAIN J:DISPLAYSET►J24:DISPCONTR
+5 +4 +3 +2 +1 0 -1 -2 -3 -4 -5High Low
(2) Display damping time constant settingThisfunctioncanspecifythedampingtimeconstantforthedisplayindependentlyofthatspecifiedinSubsection4.1.6.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Displayset►Optionalconfig►DampBRAIN J:DISPLAYSET►J30:DISPDAMPING
NOTEThedampingtimeconstantofthedisplayiseffectiveonlyforthedisplay.RefertoSubsection4.1.6tospecifythedampingtimeconstantfortheoutputofthephysicalquantity.
(3) Date display format settingThedatedisplayformatcanbeset.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Displayset►Optionalconfig►FormatdateBRAIN J:DISPLAYSET►J31:FORMATDATE
Fromthetablebelow,specifythedatedisplayformat.Selection DescriptionDisplay BRAIN
MM/DD/YYYY MM/DD/YYYY Displaysthedateas“month/day/year”.DD/MM/YYYY DD/MM/YYYY Displaysthedateas“day/month/year”.YYYY/MM/DD YYYY/MM/DD Displaysthedateas“year/month/day”.
(4) Display black/white inverse settingTheblack/whiteinverseisavailableforthedisplay.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Displayset►Optionalconfig►InversionBRAIN J:DISPLAYSET►J47:DISPINVERSE
<4. Functions> 119
IM 01E22C02-01EN
Fromthetablebelow,selectthedisplayblack/whiteinversemode.Selection DescriptionDisplay BRAIN
Normal NormalDoesnotsetthedisplaytotheblack/whiteinversemode.(Charactercolor:Black,Backgroundcolor:White)
Invert Invert Setsthedisplaytotheblack/whiteinversemode.(Charactercolor:White,Backgroundcolor:Black)
(5) Display squawk setting (squawk)Thebacklightofthedisplaycanbeblinked(squawked)at4-secondintervalstoidentifyacommunicatingproductifanumberofthesamemodelisinstalled.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Diag/Service►Dispindicator►SquawkBRAIN J:DISPLAYSET►J51:DISPSQUAWK
Fromthetablebelow,selectwhetherornottoblinkthedisplay.Selection DescriptionDisplay BRAIN
Off Off Doesnotblinkthedisplay.On On Blinksthedisplay(continuously).Squawkonce SquawkOnce Blinksthedisplay(onlyonce).
(6) IR switch functionTheIRswitchfunctionisafunctiontodrivetheIRswitchfordisplayoperation.ForthebasicoperationsoftheIRswitch,refertoSubsection2.2.1.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
BRAIN J:DISPLAYSET►J54:IRSWOPERATE
Fromthetablebelow,selecthowtheIRswitchfunctionistobeused.Selection Description
Disable DisablestheIRswitchfunction.Enable EnablestheIRswitchfunction.
NOTEIftheIRswitchfunctionissetto“Disable”,theswitchfunctionisnotdisplayedonthedisplay,andtheIRswitchcannotbeoperated.
22.000mA0.0%
0.00000m/s
00 : 00VEL
FLP
AO1SET SFT INC
F0447.ai
If “Enable”
22.000mA0.0%
0.00000m/s
00 : 00VEL
FLP
AO1
F0448.ai
If “Disable”
<4. Functions> 120
IM 01E22C02-01EN
4.9.7 microSD Card SettingIftheoptionalcodeMCisselected,thebackupparametersandloggingdatacanbeusedbyinsertingthededicatedmicroSDcardintothedisplay.Fordetailsaboutbackingupparametersandloggingdata,refertoSection4.13.
(1) Removing the microSD card
IMPORTANTIfthemicroSDcardisremovedwithoutexecuting“Unmount”onthesoftware,itmaycausedatastoredonthemicroSDcardtobeerasedorthedevicetooperateabnormally.
ThemicroSDcardcanberemovedwiththefollowingparameters.Menu path
Display Devicesetup►microSD►Unmount
Fromthetablebelow,selectwhethertoremovethemicroSDcard.Selection Description
Cancel CancelstheremovalofthemicroSDcard.Execute MakesitpossibletoremovethemicroSDcardinsafety.
(2) microSD card format
IMPORTANTIftheformatfunctionofthisproductisnotusedtoformatamicroSDcard,itmaycauseadeviceoperationfailure.
<4. Functions> 121
IM 01E22C02-01EN
Formattingispossiblewiththefollowingparameters.Menu path
Display Devicesetup►microSD►Format
Fromthetablebelow,selectwhethertoformatthemicroSDcard.Selection Description
Cancel Cancelsformatting.Execute Executesformatting.
(3) Checking microSD contentsDataonthemicroSDcardcanbecheckedwiththefollowingparameters.Menu path
Display Devicesetup►microSD►Contents
(4) Checking the property of microSD CardThetotalspace,availablespaceandfilesystemofthemicroSDcardcanbecheckedwiththefollowingparameters.
�� Total spaceMenu path
Display Devicesetup►microSD►Property►Totalspace
�� Displaying available spaceMenu path
Display Devicesetup►microSD►Property►Availablespace
�� DisplayingfilesystemMenu path
Display Devicesetup►microSD►Property►Filesystem
<4. Functions> 122
IM 01E22C02-01EN
4.10 Device Information4.10.1 Order Information
Orderinformationcanbespecifiedforthisproduct.Ifaparticularparameterisspecifiedatthetimeoforder,thisproductisshippedwiththeparameterspecified.Ifaparameterisnotspecifiedatthetimeoforder,thatparameterneedstobesetbythecustomer.Themodelandsuffixcodeareindicatedinthefollowingformat.
Transmitter:AXG1A - □□□□□□□□□□□ /□ (1) (2) (3)
(1)Modelcode,(2)Suffixcode,(3)Optionalcode
Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Deviceinfo►Orderinfo►(seebelow)BRAIN K:DEVICEINFO►(seebelow)
Parameter DescriptionDisplay BRAINTagNo K01:TAGNO SpecifiestheTagNo.Max.16characters.
MScode►Modelcode K20:MODEL CODE Specifiesthemodelcodeoftheintegralflowmeterorremotetransmitter.
MScode►Suffixconfig1 K21:SUFFIXCONF1 Specifiesthesuffixcodeoftheintegralflowmeterorremotetransmitter.MScode►Suffixconfig2 K22:SUFFIXCONF2
MScode►Option1 K23:OPTION1Specifiestheoptionalcodeoftheintegralflowmeterorremotetransmitter.
MScode►Option2 K24:OPTION2MScode►Option3 K25:OPTION3MScode►Option4 K26:OPTION4
RSMScode►Modelcode K27:RSMDLCD Specifiesthemodelcodeoftheremotesensor.
RSMScode►Suffixconfig1 K28:RSSUFCONF1 Specifiesthesuffixcodeoftheremotesensor.RSMScode►Suffixconfig2 K29:RSSUFCONF2
RSMScode►Option1 K30:RSOPT1Specifiestheoptionalcodeoftheremotesensor.
RSMScode►Option2 K31:RSOPT2RSMScode►Option3 K32:RSOPT3RSMScode►Option4 K33:RSOPT4
Menu pathDisplay Devicesetup►Detailedsetup►Deviceinfo►Ver/Numinfo►(seebelow)BRAIN K:DEVICEINFO►(seebelow)
Parameter DescriptionDisplay BRAIN
TransserialNo K34:TRNSSRNO Displaystheserialnumber(deviceNo.)ofthetransmitter.
Sensor serial No K35:FTSRNO Displaystheserialnumber(deviceNo.)ofthesensor.
<4. Functions> 123
IM 01E22C02-01EN
NOTEWhenanyparametersrelatedtotheorderinformationabouttheproductarechanged,theorderinformationuponshipmentfromthemanufacturingfactorycannotbereferredto.Tostoretheorderinformationdefineduponshipmentfromthemanufacturingfactory,itisrecommendedtorefertoSection4.13tomakeabackup.
4.10.2 Device RevisionTherevisionofthesoftwareusedforthisproductcanbechecked.Thisinformationcanbeviewedwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Deviceinfo►Ver/Numinfo►(seebelow)BRAIN K:DEVICEINFO►(seebelow)
Parameter DescriptionDisplay BRAINTransmittertype K11:TRNSTYPE Displaysthetypeofthetransmitter.OptionboardID K12:OPTBOARDID Displaysthetypeoftheoptionboard.Mainsoftrev K50:MAINBREV Displaysthesoftwarerevisionofthemainboard.Snsrsoftrev K51:SENSORBREV Displaysthesoftwarerevisionofthesensorboard.Indsoftrev K52:INDBREV Displaysthesoftwarerevisionofthedisplayboard.
4.10.3 Memo FunctionThreeparameterscanbeusedforthememofunction.Thememofunctioncanbesettoupto16characters.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Deviceinfo►Memo►(seebelow)BRAIN K:DEVICEINFO►(seebelow)
Parameter DescriptionDisplay BRAINMemo1 K40:MEMO1 Specifiesmemo1.Memo2 K41:MEMO2 Specifiesmemo2.Memo3 K42:MEMO3 Specifiesmemo3.
<4. Functions> 124
IM 01E22C02-01EN
4.10.4 Date and Time InformationThepresentdateandtimecanbeset.ThedisplayformatfordateandtimeistheformatsetinSubsection4.9.6.Thissettingcanbedisplayedandconfiguredwiththefollowingparameters.
�� Displaying date and timeMenu path
Display Devicesetup►Detailedsetup►Deviceinfo►Date/Time►(seebelow)BRAIN K:DEVICEINFO►(seebelow)
Parameter DescriptionDisplay BRAINCurrentdate K03:CURDATE Displayscurrentdate.Currenttime K04:CURTIME Displayscurrenttime.
�� Setting date and timeMenu path
Display Devicesetup►Detailedsetup►Deviceinfo►Date/Time►(seebelow)BRAIN K:DEVICEINFO►(seebelow)
Parameter DescriptionDisplay BRAINSet date - Specifiesthecurrentdate.Settime - Specifiesthecurrenttime.- K05:SETCURDAY Setthecurrentday(1-31).- K06:SETCURMONTH Setthecurrentmonth(1-12).- K07:SETCURYEAR Setthecurrentyear(1900-2155).- K08:SETCURHR Setthecurrenttime(0-23).- K09:SETCURMIN Setthecurrentminute(0-59).- K10:SETCURSEC Setthecurrentsecond(0-59).
4.10.5 Display Operation TimeDisplaysoperationtime.Operationtimereferstotheoperationtimeuptothepresentfromthetimewhenthepowerwasturnedonforthefirsttime.However,timeisnotcountedwhenpowerisnotturnedon.Thisinformationcanbeviewedwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Deviceinfo►Date/Time►OperationtimeBRAIN K:DEVICEINFO►K02:OPERATETIME
Theoperationtimeisdisplayedintheformatof“dddddDhh:mm”.“dddddD”indicatestheday,“hh”indicatesthehour,and“mm”indicatestheminute.Example:“00031D12:34”Thisexampleshowsthattheproducthasbeenoperatedfor31days,12hours,and34minutes.
<4. Functions> 125
IM 01E22C02-01EN
4.10.6 Explosion Protection SettingWhethertheproductisastandardproductoranexplosionprotectionproductcanbeset.Whenusinganexplosionprotectionsensor,setthefollowingparameterto“Yes”.Menu path
Display Devicesetup►Detailedsetup►Deviceinfo►Orderinfo►ExplosionprotectionBRAIN K:DEVICEINFO►K17:EXPROTECT
Fromthetablebelow,selectexplosionprotectionsetting.Selection DescriptionDisplay BRAIN
No No Noexplosionprotectionsetting.Yes Yes Explosionprotection.
<4. Functions> 126
IM 01E22C02-01EN
4.11 Self-diagnostic Function4.11.1 Types of Diagnosis Functions
Thisproducthasaself-diagnosticfunctionfordiagnosingdevicefailuresorprocessstates.Diagnosticfunctionsofthisproductareasfollows:
Diagnosis function Description
Alarmhigh/lowlimitfunction Displaysanalarmwhenthespecifiedvalueisexceededandoutputsitasthestatusoutput.
Electrodeadhesiondetection Diagnosesanelectrodeadhesionfromtheresistancevalueoftheelectrode,anddisplaysawarningoralarmifadhesionisdetected.
Sensoremptycheck Checkswhetherthesensorisintheemptypipestate,anddisplaysanalarmiftheemptypipestateisdetected.
Misconnectiondetection Checkswhetherthetransmittersignallineandtheexcitationlinearecorrectlyconnected,anddisplaysawarningifamisconnectionisdetected.
Verification(devicehealthdiagnosis)function Diagnosesthehealthoftheproduct,anddisplaysitsresult.
Electrodeinsulationdeteriorationdiagnosis
Diagnosesdeteriorationofinsulationfromaresistancevalueoftheelectrode,anddisplaysawarningifitisdetected.
Flownoisediagnosis Measuresaflownoisedetectedinthesensor,anddisplaysawarningifthemeasuredvalueexceedsthespecifiedvalue.
Lowconductivitydiagnosis Obtainstheconductivityfromtheresistancevalueandsizeoftheelectrode,anddisplaysawarningiftheconductivityfallsbelowthespecifiedvalue.
Coilinsulationdiagnosis Diagnosesthecurrentvalueofexcitingcurrent,anddisplayswarningifinsulationisdetected.
4.11.2 Alarm High/Low Limit FunctionThehigh/lowlimitfunctionofthealarmdisplaysanalarmwhenthePV-mappedphysicalquantityofSubsection4.1.2exceedsthespecifiedvalue.Thisfunctionalsocanoutputanalarmoccurrenceasastatusoutput.Fordetailsaboutthestatusoutput,refertoSubsection4.3.9.Asavaluetojudgeanalarm,thefollowingfourvaluescanbeset:highlimit,high-highlimit,lowlimit,andlow-lowlimit.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Diag/Service►H/Lalarmcfg►(seebelow)BRAIN I:ALARM►(seebelow)
Parameter DescriptionDisplay BRAINHighalarm I10:HIGHALARM Specifiesthehighlimittojudgeanalarm.Lowalarm I11:LOWALARM Specifiesthelowlimittojudgeanalarm.HHalarm I12:HIHIALARM Specifiesthehigh-highlimittojudgeanalarm.LLalarm I13:LOLOALARM Specifiesthelow-lowlimittojudgeanalarm.
Whenthehighandlowlimitalarmsarereset,eachhasahysteresis.Thehysteresiswidthshouldbesetwiththepercentage(%)forthespanofthePV-mappedphysicalquantityofSubsection4.1.2.Thehysteresisineachcasecanbespecifiedwiththefollowingparameters.Menu path
Display Devicesetup►Diag/Service►H/Lalarmcfg►H/LalarmhystBRAIN I14:H/LALMHYS
<4. Functions> 127
IM 01E22C02-01EN
(1) Hysteresisvalue =Spanormax.rangexHysteresiswidth[%]
(2) Valuethatcausesahighorhigh-highlimitalarmtobereset
=Sethighlimitorhigh-highlimit-Hysteresisvalue
(3) Valuethatcausesaloworlow-lowlimitalarmtobereset
=Specifiedlowlimitorlow-lowlimitvalue+Hysteresisvalue
Example:Spanofvolumetricflowrate=300m3/h,High-highlimit=300m3/h,Low-lowlimit=25m3/h,Highlimitvalue=250m3/h,Lowlimitvalue=50m3/h,WhensettoHysteresiswidth=5%
(1) Hysteresisvalue=300[m3/h]×5[%] =15[m3/h]
(2-1) Valuethatcausesahigh-highlimitalarmtobereset =285[m3/h] =300[m3/h]–15[m3/h](2-2) Valuethatcausesahighlimitalarmtobereset =235[m3/h] =250[m3/h]–15[m3/h](3-1) Valuethatcausesalowlimitalarmtobereset =65[m3/h] =50[m3/h]+15[m3/h](3-2) Valuethatcausesalow-lowlimitalarmtobereset =40[m3/h] =25[m3/h]+15[m3/h]
F0433.ai
Time
Flow rate[m3/h]
300
250
5025
6540
285
235
Low LimitLow-Low Limit
High Limit
High-High Limit
Hysteresis
Status Output(High/Low limit)Status Output(High-High/Low-Low limit)
NOTEWhenthephysicalquantitytobeoutputischanged,specifythealarmjudgmentvalueagain.
<4. Functions> 128
IM 01E22C02-01EN
4.11.3 Electrode Adhesion DetectionTheelectrodeadhesiondetectingfunctiondiagnosesanelectrodeadhesionwiththeresistancevalueoftheelectrode,anddisplaysawarningoralarmifanadhesionisdetected.Theelectrodeadhesiondetectionisdisplayedonthedisplayinfourlevels:level1tolevel4.Avaluetojudgeeachlevelcanbespecifiedindividually.Displayawarningwhenadhesionexceedslevel3,andanalarmwhenadhesionexceedslevel4.Fordetailsaboutthealarmandwarning,refertoSubsection4.8.1.
F0434.ai
Alarm
Level 4
Warning
Level 3Level 2Level 1
Thissettingcanbeconfiguredwiththefollowingparameters.
Menu pathDisplay Devicesetup►Diag/Service►Diagnosis►Adhesion►(seebelow)BRAIN L:DIAGNOSIS►(seebelow)
Parameter DescriptionDisplay BRAIN
Function L10:ADHESIONCHK Specifiestheuseoftheelectrodeadhesiondetectingfunction.*1
Thresholdlevel1 L11:ADHLEVEL1 Specifiesthevaluetojudgelevel1.Thresholdlevel2 L12:ADHLEVEL2 Specifiesthevaluetojudgelevel2.Thresholdlevel3 L13:ADHLEVEL3 Specifiesthevaluetojudgelevel3.Thresholdlevel4 L14:ADHLEVEL4 Specifiesthevaluetojudgelevel4.Result►Value L16:ADHMEASVAL Displaystheresistancevalueoftheelectrode.Result►Status L17:ADHSTATUS Displaystheelectrodeadhesiondetectionlevel.
Checkcycle L18:ADHCHKCYC Specifiesthedataupdatecycleforelectrodeadhesiondetection.
*1: Fromthetablebelow,selecttheuseoftheelectrodeadhesiondetectionfunction.Selection DescriptionDisplay BRAIN
Disable Disable Doesnotusetheelectrodeadhesiondetectingfunction.
Enable Enable Usestheelectrodeadhesiondetectingfunction.
NOTEIftheelectrodeadhesiondetectingfunctionisnotused,theelectroderesistancevalueandlevelarecleared.
NOTEBesuretousetheelectrodeadhesiondetectingfunctionwhenthesensorisinthefullpipestate.Ifthesensorisintheemptypipestate,thefunctiontodetectadhesionmaynotrunnormally.
<4. Functions> 129
IM 01E22C02-01EN
NOTETheelectrodeadhesiondetectingfunctionrestrictsfluidconductivity.Therecommendedconductivitiesareasfollows.
Meter size Conductivity2.5to10mm 30μS/cmorlarger15to400mm(0.5to16in.) 10μS/cmorlarger500mm 20μS/cmorlarger
NOTEIfthedataupdatecyclefortheadhesiondetectionisshortened,theerroroftheelectroderesistancevalueincreases.Donotchangethedefaultvalueunlessespeciallyspecified.
4.11.4 Sensor Empty CheckThesensoremptycheckfunctioncheckswhetherthesensorisintheemptypipestate,anddisplaysanalarmiftheemptypipestateisdetected.Fordetailsaboutthealarmandwarning,refertoSubsection4.8.1.Thisinformationcanbeviewedwiththefollowingparameters.Menu path
Display Devicesetup►Diag/Service►Diagnosis►Emptycheck►(seebelow)BRAIN L:DIAGNOSIS►(seebelow)
Parameter DescriptionDisplay BRAIN
Emptystatus L44:EMPTYSTS Displaytheresultofthesensoremptycheckfunction.*1
ElectrodevoltageA L42:ELECVOLA DisplaysthevoltageofelectrodeA.ElectrodevoltageB L43:ELECVOLB DisplaysthevoltageofelectrodeB.
*1:Fromthetablebelow,checktheresultofthesensoremptycheckfunction.Selection DescriptionDisplay BRAIN
Full Full Indicatesthatthesensorisinthefullpipestate.Empty Empty Indicatesthatthesensorisintheemptypipestate.
IMPORTANT• Ifthesensorisintheemptypipestate,outputfluctuationoremptycheckalarmmayoccur.
Besuretousethemagneticflowmeterbeingfilledwithliquidfully.• Theemptycheckisdeterminedbymeasuringtheresistancebetweentheelectrodeandthe
ground.Forthatreason,theemptypipestatemaynotbedetectedduetothepipingcondition,electrodecondition,andenvironmentalnoise.Inparticular,notethattheemptycheckfunctionmaynotoperateproperlyforhigh-viscosityfluidsandadhesivefluids.
• Ittakes10to15minutestodiagnosetheemptypipestate.Otherprocessalarmsmayoccurbeforetheemptycheckalarmoccursafterthepipeisintheemptypipestate.
<4. Functions> 130
IM 01E22C02-01EN
4.11.5 Wiring Connection DiagnosisThewiringconnectiondiagnosticfunctiondiagnoseswhetherthesignallineandtheexcitationlinebetweentheremotetypesensorandtransmitterarecorrectlyconnected,anddisplaysawarningifamisconnectionisdetected.Forexample,iftherearetwosets(AandB)ofremotesensorsandremotetransmitters,thisfunctioncheckswhetherthesignallineoftransmitterAismisconnectedtothesignalterminalofsensorB,ortheexcitationlineoftransmitterAismisconnectedtotheexcitationterminalofsensor B.Ittakesapproximately10secondstodiagnosethewiringconnection.Fordetailsaboutthealarmandwarning,refertoSubsection4.8.1.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Diag/Service►Diagnosis►(seebelow)BRAIN L:DIAGNOSIS►(seebelow)
ParameterDescription
Display BRAIN
Diagnosticexecute L33:DIAGEXE Specifiestheuseofthewiringconnectiondiagnosticfunction.*1
Diagnosticoutput L49:DIAGOUTPUT Specifiestheoutputtousethewiringconnectiondiagnosticfunction.*2
*1: Fromthetablebelow,selecttheuseofthewiringconnectiondiagnosticfunction.Selection DescriptionDisplay BRAIN
Connectcheckexe ConnChkexe Startsthewiringconnectiondiagnosticfunction.
*2: Fromthetablebelow,selecttheoutputforwhichthewiringconnectiondiagnosticfunctionisexecuted.Selection Analog output Totalizer Pulse output Frequency
outputDisplay BRAIN
Zero Zero 0%output Input0(Outputfixed) 0 pps 0%output
Measuredvalue MeasuredvalueOutputof
calculatedvalue(Undefined)
Totalizationofcalculatedvalue(Undefined)
Outputofcalculatedvalue(Undefined)
Outputofcalculatedvalue(Undefined)
Hold HoldFixesthelastvalidvaluebeforethe
diagnosisstarts.
Totalizesthelastvalidvaluebeforethe
diagnosisstarts.
Fixesthelastvalidvaluebeforethe
diagnosisstarts.
Fixesthelastvalidvaluebeforethe
diagnosisstarts.
IMPORTANT• Beforeusingthewiringconnectiondiagnosticfunction,besuretodisconnectthisproduct
fromthecontrolloop.• Whenthewiringmisconnectiondiagnosticfunctionisused,theanalogoutputandpulse
outputarenotoutputcorrectly.
<4. Functions> 131
IM 01E22C02-01EN
NOTEWhenusingthewiringconnectiondiagnosticfunction,thefollowingconditionsmustbesatisfied.Iftheconditionsarenotsatisfied,themagneticflowmetermaynotcorrectlydiagnosethewiringconnectionstate.• Thesensorisinthefullpipestate.• Noinfluenceofnoise.• ThemagneticflowmetertransmitterandAM012(calibrator)arenotconnected.
4.11.6 Verification(DeviceHealthDiagnosis)FunctionTheverificationfunctiondiagnosesthehealthoftheproductanddisplaysthediagnosisresult.Ittakesabout15minutestocompletetheverificationfunction,andtheprogressoftheverificationfunctioncanbecheckedviathebargraphandtheremainingtimeonthedisplay.
10:00
Verification executionExcitation circuit...
01 : 00
SET SFT INC
Thisfunctioncaninspectconditionsofthemagneticcircuit,excitationcircuitandcalculationcircuit,andexecutesthedevicehealthdiagnosisoftheproductbasedontheirinternalalarmrecordalongwiththeirwiringmisconnection.Thisfunctionevaluatesthediagnosisresultbasedontheconditionsoftheproductandshows“Passed”ifnoproblemisfoundor“Failed”ifaproblemisfoundonthedisplay.Thefollowingexecutionresultsareshownonthedisplay.
• Examples of “Passed” resultsVFcheckresults PassedVFOperationtime 00001D10:01Magneticcircuitresult PassedExcitingcircuitresult PassedCalculationcircuitresult PassedDevicestatusresult PassedConnectionstatusresult Passed
• Examples of “Failed” resultsVFcheckresults FailedVFOperationtime 00001D10:01Magneticcircuitresult PassedExcitingcircuitresult PassedCalculationcircuitresult FailedDevicestatusresult PassedConnectionstatusresult Passed
ContactYokogawaservicecenterifthe“Failed”messageisdisplayedforitemsfrom“Magneticcircuitresult”to“Devicestatusresult”.If“Failed”isdisplayedfor“Connectionstatusresult”,checkforwiringmisconnectionbetweenthesensorandtransmitter(refertoSubsection4.11.5fordetail)anddamageonthecables.
Theverificationfunctioncanbeusedintwowaysdependingonthestateofthefluid;onestatewhereafluidisflowingandanotherstatewherenofluidisflowing.Thetwodiagnosisresults(currentandprevious)arestoredinthedevicememory,andtheycanbecheckedlater.
<4. Functions> 132
IM 01E22C02-01EN
Iftheverificationfunctionisused,thefollowingresultscanbedisplayed.Totaljudgmentresult CalculationcircuitdiagnosisresultOperationtimeofverificationfunction DevicealarmdiagnosisresultMagneticcircuitdiagnosisresult WiringconnectiondiagnosisresultExcitationcircuitdiagnosisresult
Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Diag/Service►Diagnosis►(seebelow)BRAIN L:DIAGNOSIS►(seebelow)
Parameter DescriptionDisplay BRAIN
Diagnosticoutput L49:DIAGOUTPUT Specifiestheoutputtoexecutetheverificationfunction.*1
Menu pathDisplay Devicesetup►Diag/Service►Verification►(seebelow)BRAIN L:DIAGNOSIS►(seebelow)
Parameter DescriptionDisplay BRAIN
Mode L50:VFMODE Specifiesthefluidstatewhenperformingtheverificationfunction.*2
Execute L51:VFEXE Specifiestheexecutionoftheverificationfunction.*3VFNo L52:VFNO Specifiesthediagnosisresultdisplaytime.*4
VFtargetselectL47:VFTGTSET(forsetting)L48:VFTGTSTS(forchecking)
Specifiesthetargetfordiagnosis.*5
Result - DisplaysthediagnosisresultofthetimesetbytheVFnumber.*6
*1: Fromthetablebelow,selecttheoutputtoexecutetheverificationfunction.Selection Analog output Totalizer Pulse output Frequency
outputDisplay BRAIN
Zero Zero 0%output Input0(Outputfixed) 0 pps 0%output
Measuredvalue MeasuredvalueOutputof
calculatedvalue(Undefined)
Totalizationofcalculatedvalue(Undefined)
Outputofcalculatedvalue(Undefined)
Outputofcalculatedvalue(Undefined)
Hold HoldFixesthelastvalidvaluebeforethe
diagnosisstarts.
Totalizesthelastvalidvaluebeforethe
diagnosisstarts.
Fixesthelastvalidvaluebeforethe
diagnosisstarts.
Fixesthelastvalidvaluebeforethe
diagnosisstarts.
*2: Fromthetablebelow,selectthefluidstatusrequiredtoexecutetheverificationfunction.Selection DescriptionDisplay BRAIN
Noflow Noflow Executestheverificationfunctionwiththefluidnotflowing.
Flow Flow Executestheverificationfunctionwiththefluidflowing.
<4. Functions> 133
IM 01E22C02-01EN
*3: Fromthetablebelow,selectwhetherornottoexecutetheverificationfunction.Selection DescriptionDisplay BRAIN
Notexecute Notexecute Doesnotexecutetheverificationfunction.Execute Execute Executestheverificationfunction.
*4: Fromthetablebelow,selecttheexecutionresultoftheverificationfunction.Selection DescriptionDisplay BRAIN
Factory Factory Displaystheresultobtaineduponshipmentfromthemanufacturingfactory.
Previous Previous Displaysthepreviousresult.Present Present Displaystheresultatthistime.
*5: Fromthetablebelow,selectthetargetfordiagnosis.Selection DescriptionDisplay BRAIN
Magneticcircuit - MagneticcircuitdiagnosisExcitecircuit - ExcitationcircuitdiagnosisCalccircuit - CalculationcircuitdiagnosisDevicestatus - DevicealarmdiagnosisConnectstatus - Wiringmisconnectioncheck- Allon Setalldiagnosticsresultsvalid.- Alloff Setalldiagnosticsresultsinvalid.- Magneticon Settheresultsofmagneticcircuitdiagnosisvalid.- Magneticoff Settheresultsofmagneticcircuitdiagnosisinvalid.- Excitationon Settheresultsofexcitationcircuitdiagnosisvalid.- Excitationoff Settheresultsofexcitationcircuitdiagnosisinvalid.- Calculationon Settheresultsofcalculationcircuitdiagnosisvalid.- Calculationoff Settheresultsofcalculationcircuitdiagnosisinvalid.- Devstatuson Settheresultsofdevicealarmdiagnosisvalid.- Devstatusoff Settheresultsofdevicealarmdiagnosisinvalid.- Connstatuson Settheresultsofmisconnectiondiagnosisvalid.- Connstatusoff Settheresultsofmisconnectiondiagnosisinvalid.
Inthedefaultstate,allofthediagnosisresultsarereflectedinthefinalresultsoftheverification.EvenifanitemissetonL47:VFTGTSETsoasnottoreflectthediagnosisresultstothefinalresultsoftheverification,executiontimefortheverificationdoesnotchangebecausethediagnosisitselfisexecuted.Inaddition,thediagnosisresultsselectedtoDisablewithL47:VFTGTSETaredisplayedas“Skip”.NotethatverificationitselfcannotbeexecutedifalldiagnosesareuncheckedonL47:VFTGTSET(Alloff).
*6: Fromthetablebelow,selecttheresult.Parameter DescriptionDisplay BRAIN
Failed/Passed L53:VFCHKRES Executionresult*7VFoperatetime L54:VFOPETIME OperationtimeatthestartofverificationMagneticcircuit L55:VFMAGRES Magneticcircuitdiagnosisresult*7Excitecircuit L56:VFEXCITRES Excitationcircuitdiagnosisresult*7Calccircuit L57:VFCALCRES Calculationcircuitdiagnosisresult*7Devicestatus L58:VFDEVRES Devicealarmdiagnosisresult*7Connectstatus L59:VFCONNRES Wiringmisconnectioncheckresult*7
<4. Functions> 134
IM 01E22C02-01EN
*7:Fromthetablebelow,selecttheresultoftheverificationfunction.Result
DescriptionDisplay BRAIN
Passed Passed Therearenoproblemsconcerningthediagnosisresult.
Failed Failed Thereisaproblemconcerningthediagnosisresult.Canceled Canceled Cancelsthediagnosis.
No data No Data Nodiagnosisresultdata(Theverificationfunctionisnotused.)
Unknown Unknown Cannotperformadiagnosis.Skip Skip Outofverificationtarget
IMPORTANT• Beforeusingtheverificationfunction,besuretodisconnectthisproductfromthecontrolloop.• Whentheelectrodeinsulationdeteriorationdiagnosisfunctionisused,theanalogoutputand
pulseoutputarenotoutputcorrectly.• Notethatparameterscannotbechangedwhiletheverificationfunctionisexecuted.
NOTE• Besuretousetheverificationfunctionwhenthesensorisinthefullpipestate.Ifthesensoris
intheemptypipestate,thefunctiontodetectadhesionmaynotrunnormally.• Whenusingtheverificationfunction,correctlyspecifythefluidstatuswiththeparameter.• Ifthereisaproblemwiththeverificationresult,refertotheMaintenanceManual.
<4. Functions> 135
IM 01E22C02-01EN
4.11.7 Electrode Insulation Deterioration DiagnosisTheelectrodeinsulationdeteriorationdiagnosisfunctiondiagnosesadeteriorationofinsulationwitharesistancevalueoftheelectrodeanddisplaysawarningifdeteriorationofinsulationisdetected.Ittakesapproximately5minutestocompletethisdiagnosis.Fordetailsaboutthealarmandwarning,refertoSubsection4.8.1.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Diag/Service►Diagnosis►(seebelow)BRAIN L:DIAGNOSIS►(seebelow)
Parameter DescriptionDisplay BRAIN
Diagnosticexecute L33:DIAGEXE Specifiestheexecutionoftheelectrodeinsulationdeteriorationdiagnosticfunction.*1
Diagnosticoutput L49:DIAGOUTPUT Specifiestheoutputtoexecutetheelectrodeinsulationdeteriorationdiagnosticfunction.*2
*1: Fromthetablebelow,selecttheuseoftheelectrodeinsulationdeteriorationdiagnosticfunction.Selection DescriptionDisplay BRAIN
Electrodeinsulexe Elecinsexe Startstheelectrodeinsulationdeteriorationdiagnosisfunction.
*2: Fromthetablebelow,selecttheoutputwhiletheelectrodeinsulationdeteriorationdiagnosticfunctionisexecuted.Selection Analog output Totalizer Pulse output Frequency
outputDisplay BRAIN
Zero Zero 0%output Input0(Outputfixed) 0 pps 0%output
Measuredvalue MeasuredvalueOutputof
calculatedvalue(Undefined)
Totalizationofcalculatedvalue(Undefined)
Outputofcalculatedvalue(Undefined)
Outputofcalculatedvalue(Undefined)
Hold HoldFixesthelastvalidvaluebeforethe
diagnosisstarts.
Totalizesthelastvalidvaluebeforethe
diagnosisstarts.
Fixesthelastvalidvaluebeforethe
diagnosisstarts.
Fixesthelastvalidvaluebeforethe
diagnosisstarts.
IMPORTANT• Beforeusingtheelectrodeinsulationdeteriorationdiagnosisfunction,besuretodisconnect
thisproductfromthecontrolloop.• Whentheelectrodeinsulationdeteriorationdiagnosisfunctionisused,theanalogoutputand
pulseoutputarenotoutputcorrectly.
NOTEBesuretousetheelectrodeinsulationdeteriorationdiagnosticfunctionwhenthesensorisinthefullpipestate.Ifthesensorisintheemptypipestate,theelectrodeinsulationdeteriorationdiagnosisfunctionmaynotrunnormally.
<4. Functions> 136
IM 01E22C02-01EN
4.11.8 Flow Noise DiagnosisThisfunctionmeasuresaflownoisedetectedinthesensor,anddisplaysawarningifitexceedsthespecifiedvalue.Theflownoisediagnosisisshownonthedisplayinfourlevels:level1tolevel4.Avaluetojudgeeachlevelcanbespecifiedindividually.Ifanoutputexceedsthelevel3orlevel4value,awarningmessageisdisplayed.Fordetailsaboutthealarmandwarning,refertoSubsection4.8.1.Fordetailsabouttheflownoisespananddampingtimeconstant,refertoSubsections4.1.5and4.1.6.
F0436.ai
Warning
Level 4
Warning
Level 3Level 2Level 1
Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Diag/Service►Diagnosis►Flownoise►(seebelow)BRAIN L:DIAGNOSIS►(seebelow)
Parameter DescriptionDisplay BRAIN
Function L20:FLNOISECHK Specifiestheuseoftheflownoisediagnosisfunction.*1
Thresholdlevel1 L21:FLNOISELV1 Specifiesthevaluetojudgelevel1.Thresholdlevel2 L22:FLNOISELV2 Specifiesthevaluetojudgelevel2.Thresholdlevel3 L23:FLNOISELV3 Specifiesthevaluetojudgelevel3.Thresholdlevel4 L24:FLNOISELV4 Specifiesthevaluetojudgelevel4.Result►Value L25:FLNOISEVAL Displaystheflownoisevalue.Result►Status L26:FLNOISESTS Displaystheflownoiselevel.
*1: Fromthetablebelow,selecttheuseoftheflownoisefunction.Selection DescriptionDisplay BRAIN
Disable Disable Doesnotusetheflownoisediagnosisfunction.Enable Enable Usestheflownoisediagnosisfunction.
NOTEIftheflownoisediagnosisfunctionissetto“Disable”,theflownoisevalueandlevelarecleared.
<4. Functions> 137
IM 01E22C02-01EN
4.11.9 Low Conductivity DiagnosisThelowconductivitydiagnosisfunctioncalculatesconductivityfromtheresistancevalueandsizeoftheelectrodeanddisplaysawarningiftheconductivityfallsbelowthespecifiedvalue.Fordetailsaboutthealarmandwarning,refertoSubsection4.8.1.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Diag/Service►Diagnosis►Conductivity►(seebelow)BRAIN L:DIAGNOSIS►(seebelow)
Parameter DescriptionDisplay BRAIN
Function L30:LOCONDFUNC Specifiestheuseofthelowconductivitydiagnosisfunction.*1
Result►Value L31:CONDUCTIVITY Displaysthecalculatedconductivity.
Lowlimit L32:CONDUCTLMT Specifiesthevalueusedtojudgethelowconductivity.
Menu pathDisplay Devicesetup►Detailedsetup►Deviceinfo►Orderinfo►(seebelow)BRAIN K:DEVICEINFO►(seebelow)
Parameter DescriptionDisplay BRAINElectrodesize K15:EL SIZE Displaystheelectrodesize.
*1: Fromthetablebelow,selecttheuseofthelowconductivitydiagnosisfunction.Selection DescriptionDisplay BRAIN
Disable Disable Doesnotusethelowconductivitydiagnosisfunction.
Enable Enable Usesthelowconductivitydiagnosisfunction.
NOTETheelectrodesizeisspecifieduponshipmentfromthemanufacturingfactory.Whenanyparametersrelatedtotheorderinformationabouttheinstrumentarechanged,theorderinformationuponshipmentfromthemanufacturingfactorycannotbereferredto.Tostoretheorderinformationdefineduponshipmentfromthemanufacturingfactory,itisrecommendedtorefertoSection4.13tomakeabackup.
NOTEBeforeusingthelowconductivitydiagnosisfunction,alwaysmakesurethatthesensorisinthefullpipestate.IfthesensorisintheemptypipestateorthetransmitterisconnectedwithAM012(calibrator),thefunctiontodiagnoseconductivitymaynotrunnormally.
NOTEIfthisfunctionissetto“Disable”,thelowconductivityvalueiscleared.
<4. Functions> 138
IM 01E22C02-01EN
4.11.10 Coil Insulation DiagnosisCoilInsulationDiagnosisisthefunctiontodisplayawarningwhentheinsulationisdeterioratedbydiagnosingthecurrentvalueoftheexcitingcurrent.Fordetailsaboutthealarmandwarning,refertoSubsection4.8.1.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Diag/Service►Diagnosis►(seebelow)BRAIN L:DIAGNOSIS►(seebelow)
Parameter DescriptionDisplay BRAIN
IEXcompare L38:IEXCOMP DisplaysthecurrentvalueofExcitingcurrenttojudgecoilinsulation.
Coilinsulthreshold L37:COILINSTH Specifiesthevaluetojudgecoilinsulation.
4.11.11 Other SettingInaddition,themaximumvoltagevaluesoftheflowratesignalandcoilresistancevaluearedisplayedasdiagnosticinformation.Thisinformationcanbeviewedwiththefollowingparameters.Menu path
Display Devicesetup►Diag/Service►Diagnosis►(seebelow)BRAIN L:DIAGNOSIS►(seebelow)
Parameter DescriptionDisplay BRAINVpeakhold L40:VPEAKHOLD Displaysthemaximumvoltageoftheflowsignal.IEXresistance L41:IEXCOILR Displaysthecoilresistancevalue.
<4. Functions> 139
IM 01E22C02-01EN
4.12 Test Mode4.12.1 Test Mode Setting
Ifthetestmodeisexecuted,aprocessvalueorthevaluetobeoutputfromaconnectionterminalcanbearbitrarilyset,andaresponsefromthedevicecanbetested.Awarningisdisplayedtoindicatethatthetestmodeisinusewhilethistestmodeisused.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Diag/Service►Test►(seebelow)BRAIN M:TEST►(seebelow)
Parameter DescriptionDisplay BRAINInputtest►Testmode M02:TESTMODE(for
setting)M03:TESTSTATUS(forchecking)
Selectsthetestitem(input).
Outputtest►Testmode Selectsthetestitem(output).
Fromthetablebelow,selecttheinput/outputandprocessvaluetotest.Selection DescriptionDisplay BRAIN
- Testalloff Ends all tests.Velocitytest Veloteston Startstestingtheflowvelocity.Volumetest VolFteston Startstestingthevolumetricflowrate.Mass test MassFteston Startstestingthemassflowrate.Calorie test Calorie test on Startstestingthecalorie.AO1test AO1teston Startstestinganalogoutput1.PO1test P1teston Startstestingpulseoutput/frequencyoutput1.SO1 test SO1 test on Startstestingstatusoutput1.AO2test AO2teston Startstestinganalogoutput2.PO2test P2teston Startstestingpulseoutput/frequencyoutput2.SO2 test SO2 test on Startstestingstatusoutput2.AItest AIteston Startstestingtheanaloginput.SO11 test SO11 test on Startstestingstatusoutput11.SI11 test SI11 test on Startstestingstatusinput11.SO12 test SO12 test on Startstestingstatusoutput12.SI12 test SI12 test on Startstestingstatusinput12.- Testallon Starts all tests.Alarmtest Almteston Startstestingalarmoutput.
Atestmodevalueistransmittedintheorderoftheflowvelocity,volumetricflowrate,massflowrate,andcalorie.Forexample,ifthevolumetricflowrateissetasatestvalue,themassflowrateandcaloriecanbeobtainedfromthetestvalue.Inthiscase,thetestvalueoftheflowvelocityisexecutedwiththenormalmeasurementvalue.
NOTETospecifyeachterminaloutput,refertoSubsection4.3.1.Ifthespecifiedoutputisdifferentfromtheteststartingoutput,thetestmodecannotbeused.Forexample,ifthepulse/statusoutputissetasafrequencyoutputtostartthestatusoutputtest,thetestcannotbestarted.
<4. Functions> 140
IM 01E22C02-01EN
4.12.2 Test Value SettingItisnecessarytosettestvaluesfortheinputandoutputandprocessvaluetotest.Thissettingcanbeconfiguredwiththefollowingparameters.However,theunitoftheprocessvaluetotestisthesameastheunitspecifiedinSubsection4.1.4.Iftheunitischanged,theprocessvalueisalsochangedtomatchtothechangedunit.
�� Input testMenu path
Display Devicesetup►Diag/Service►Test►Inputtest►(seebelow)BRAIN M:TEST►(seebelow)
Parameter DescriptionDisplay BRAINVelocity M10:VELOTVAL Specifiestheflowvelocity.Volume M11:VOLUMTVAL Specifiesthevolumetricflowrate.Mass M12:MASSTVAL Specifiesthemassflowrate.Calorie M13:CALOTVAL Specifiesthecalorie.AI M30:AITESTVAL Specifiestothecurrentvalueofanaloginput.SI11 M41:SI11TVAL Specifiesthestateofstatusinput11.SI12 M43:SI12TVAL Specifiesthestateofstatusinput12.
�� Output testMenu path
Display Devicesetup►Diag/Service►Test►Outputtest►(seebelow)BRAIN M:TEST►(seebelow)
Parameter DescriptionDisplay BRAINAO1 M14:AO1TESTVAL Specifiestothecurrentvalueofanalogoutput1.
PO1 M15:P1TESTVAL Specifiesthefrequencyofpulseoutput/frequencyoutput1.
SO1 M16:SO1TESTVAL Specifiesthestateofstatusoutput1.AO2 M20:AO2TESTVAL Specifiestothecurrentvalueofanalogoutput2.
PO2 M21:P2TESTVAL Specifiesthefrequencyofpulseoutput/frequencyoutput2.
SO2 M22:SO2TESTVAL Specifiesthestateofstatusoutput2.SO11 M40:SO11TVAL Specifiesthestateofstatusoutput11.SO12 M42:SO12TVAL Specifiesthestateofstatusoutput12.Alarm M44:ALMTESTVAL Specifiesthestateofalarmoutput.
<4. Functions> 141
IM 01E22C02-01EN
4.12.3 Test 2 ModeThetest2modeisthefunctiontocollectivelytestaprocessvalue,analogoutput,totalizedvalue,andpulse.Thetestvaluecanbesetintherangebetween-10.0%and110.0%.Inputtestvalues(%)aretestedwithascalevalueinresponsetothespanofeachprocessvalue.
�� Test 2 mode settingMenu path
Display Devicesetup►Diag/Service►Test►Test2modeBRAIN M:TEST►M50:TEST2MODE
Selectswhethertoenabletest2mode.Selection
Display BRAINNormal NormalTest Test
�� Test 2 mode value settingMenu path
Display Devicesetup►Diag/Service►Test►Test2valueBRAIN M:TEST►M51:TEST2OUT
Test2modevalueissetwiththepercentage(%)forthespan.
4.12.4 Test Mode Auto ResetWhenaspecifictimelapseswithnoparameterschangedafterthetestmodeisenabled,thetestmodewillberesetautomaticallyWhenanytestmodeparameterischanged,thetestmoderesettimeisextended.Thetimetoautomaticallyresetthetestmodecanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Diag/Service►Test►ReleasetimeBRAIN M:TEST►M01:TESTRELTIM
Fromthetablebelow,selectthetestmodeautoresettime.Selection DescriptionDisplay BRAIN
10min 10min Specifiestheresettimeto10min.30min 30min Specifiestheresettimeto30min.1h 1h Specifiestheresettimeto1hr.3h 3h Specifiestheresettimeto3hrs.12h 12h Specifiestheresettimeto12hrs.
<4. Functions> 142
IM 01E22C02-01EN
4.13 Event Management Function4.13.1 Backup Function
Thebackupfunctioncanbackupparametersettingsinthedisplay’sbuilt-inmemory(displayboard)ormicroSDcard(withtheoptionalcodeMCselected).Ifthebackupdataisrestored,theparametersettingscanbeduplicatedtoanotherdevice.Thedisplay’sbuilt-inmemorycanstorethedataforthreebackupsandthemicroSDcardcanstoreasmuchdataasthecapacityallows.Therearethreebackupmethodsavailable:backupfromthemainboardoftheproducttothememoryonthedisplayboard,backupfromthemainboardoftheproducttothemicroSDcard,andbackupfromthememoryonthedisplayboardtothemicroSDcard.Thefilename,backupname,anddatecanbespecifiedusingthebackupfunction.ThedatabackeduptothemicroSDcardisstoredinthe“YOKOGAWA”folderasa“.PAR”file.
Thissettingcanbeconfiguredwiththefollowingparameters.Fordetailsaboutthebackupparameters,refertoSubsection4.13.3.Menu path
Display Devicesetup►Diag/Service►Parambkup/restore►(seebelow)BRAIN N:EVENTMANAGE►(seebelow)
Parameter DescriptionDisplay BRAIN
Fbackupname N10:FBCKUPNAME Displaysthebackupnamedefinedatshipmentfromthemanufacturingplant.
Fbackupdate N11:FBCKUPDATE Displaysthebackupdateatshipmentfromthemanufacturingplant.
SDbackupname N12:SDBCKNAME SpecifiesthenameofthefiletobebackeduptothemicroSDcard.
Backupname1 N13:BCKNAME1 Specifiesbackupname1.Upto16characters
Backupdate1 N14:BCKDATE1 Specifiesdate1.
Backupname2 N15:BCKNAME2 Specifiesbackupname2.Upto16characters
Backupdate2 N16:BCKDATE2 Specifiesdate2.
Backupname3 N17:BCKNAME3 Specifiesbackupname3.Upto16characters
Backupdate3 N18:BCKDATE3 Specifiesdate3.Backupexecute N19:BACKUPEXEC Specifiestheuseofthebackupfunction.*1Backupresult N22:BACKUPRSLT Displaystheresultofthebackupfunction.*2
*1: Fromthetablebelow,selecthowthebackupfunctionistobeused.Selection DescriptionDisplay BRAIN
Notexecute NotExecute Doesnotbackupdata.
Storemainto1 Store Main to 1 Backsupparametersettingsfromthemainboardtomemory1inthedisplay’sbuilt-inmemory.
Storemainto2 Store Main to 2 Backsupparametersettingsfromthemainboardtomemory2inthedisplay’sbuilt-inmemory.
Storemainto3 Store Main to 3 Backsupparametersettingsfromthemainboardtomemory3inthedisplay’sbuilt-inmemory.
StoremaintoSD Store Main to SD BacksupparametersettingsfromthemainboardtothemicroSDcard.
StoreEEP1toSD StoreEEP1toSD Backsupparametersettingsfrommemory1inthedisplay’sbuilt-inmemorytothemicroSDcard.
StoreEEP2toSD StoreEEP2toSD Backsupparametersettingsfrommemory2inthedisplay’sbuilt-inmemorytothemicroSDcard.
StoreEEP3toSD StoreEEP3toSD Backsupparametersettingsfrommemory3inthedisplay’sbuilt-inmemorytothemicroSDcard.
<4. Functions> 143
IM 01E22C02-01EN
*2: Theresultofthebackupfunctionisdisplayedasshownbelow.Selection DescriptionDisplay BRAIN
Unexecuted Unexecuted Backupnotexecuted.Success Success Backupsuccessful.Failure Failure Backupfailed.Running Running Parameterbackuprunning.
Thetablebelowshowsalarmswhoseparameterscanbebackedupwhenanalarmoccurs.� :Executablewhenanalarmoccurs. ―:Notexecutablewhenanalarmoccurs.
Alarm name BackupDisplay BRAIN010:MainCPUFAIL 10:MainCPUFAIL ―011:RevcalcFAIL 11:RevcalFAIL ―012:MainEEPFAIL 12:MainEEPFAIL ―013:MainEEPdflt 13:MainEEPdflt ―014:SnsrbdFAIL 14:SnsrbdFAIL 015:SnsrcommERR 15:SnsrcommERR 016:AD1FAIL[Sig] 16:A/D1FAIL 017:AD2FAIL[Excit] 17:A/D2FAIL 018:Coil open 18:Coil open 019:Coilshort 19:Coilshort 020:ExciterFAIL 20:ExciterFAIL 021:PWM1stop 21-22:PWMstop ―022:PWM2stop ―023:Optbdmismatch 23:Optmismatch ―024:OptbdEEPFAIL 24:OptEEPFAIL ―025:OptbdA/DFAIL 25:OptA/DFAIL ―026:OptbdSPIFAIL 26:OptSPIFAIL ―027:RestoreFAIL 27:RestoreFAIL 028:IndbdFAIL
28-32:IndctFAIL
―029:IndbdEEPFAIL ―030:LCDdrvFAIL ―031:Indbdmismatch ―032:IndcommERR ―033:microSDFAIL 33:microSDFAIL ―050:Signaloverflow 50:Sigoverflow 051:Emptydetect 51:Emptydetect 052:H/LHH/LLalm 52:H/LHH/LLalm 053:Adhoverlv4 53:Adhoverlv4 060:SpancfgERR 60:SpancfgERR ―061:PVFcfgERR 61:PVFcfgERR ―062:AO14-20lmt 62:AO14-20lmt ―063:AO24-20lmt 63:AO24-20lmt ―064:AO1mltrng 64:AO1mltrng ―065:H/LcfgERR 65:AOH/Lcfg ―066:DensitycfgERR 66:DenscfgERR ―067:Pls1cfgERR 67:Pls1cfgERR ―068:Pls2cfgERR 68:Pls2cfgERR ―069:Nomisizecfg 69:Nomisizecfg ―070:AdhcfgERR 70:AdhcfgERR ―071:FLNcfgERR 71:FLNcfgERR ―072:Lognotstart 72:Lognotstart ―080:AO1saturate
80-81:AOsaturt 081:AO2saturate
<4. Functions> 144
IM 01E22C02-01EN
Alarm name BackupDisplay BRAIN082:Pls1saturate
82-83:Plssaturt 083:Pls2saturate 084:AIsaturate 84:AIsaturate 085:Cablemiscon 85:Cablemiscon 086:Coilinsulation 86:Coilinsulate 131:Transmismatch 131:Trnmismatch ―087:Adhesionlv3 87:Adhoverlv3 088:LCwarn 88:LCwarn 089:Insudetect 89:Insudetect 090:FLNoverlv3 90:FLNoverlv3 091:FLNoverlv4 91:FLNoverlv4 092:AZwarn 92:AZwarn 093:Verifwarn 93:Verifwarn 094:Factnoisewarn 94:FCnoisewarn 095:Simulateactive 95:Simactive 096:AO1fix
96-97:AOfixed 097:AO2fix 098:Pls1fix
98-99:Plsfixed 099:Pls2fix 100:AIfix 100:AIfix 101:Paramrestorerun 101:Prmrestore ―102:Disp over 102:Disp over ―103:SDsizewarn 103:SDsizewarn ―104:Bkupincmplt 104:Bkupincmplt 105:SDmismatch 105:SDmismatch ―106:SDremovalERR 106:SDremovERR ―120:Watchdog 120:Watchdog*1 121:Poweroff 121:Poweroff*1 122:InstpowerFAIL 122:InstPWFAIL*1 123:Parambkuprun 123:bkuprun ―124:Datalogrun 124:Datalogrun
IMPORTANT• Notethatparameterscannotbechangedwhenthebackupfunctionisexecuted.Ittakes
approximately20secondstocompletebackupprocessing.• TakecarenottoduplicatethefilenamewhenbackingupdatatothemicroSDcard.• NotethatifthemicroSDcardrunsoutoffreespace,thesubsequentdatawillnotbestored.
NOTEThebackupnameanddatedonotaffectthebackupfunction.Usethisasthememocolumnwhenperformingthebackupfunction.
<4. Functions> 145
IM 01E22C02-01EN
4.13.2 Restore/Duplicate ParameterIftherestorefunctionisused,itbecomespossibletorestoreparametersettings,whicharebackedupinthedisplay’sbuilt-inmemoryormicroSDcard(withtheoptionalcodeMCselected),totheproduct.Inaddition,itispossibletoreturnparameterstodefaultvaluesdefinedwhenshippedfromthefactoryandtoduplicatethebacked-upparameterstootherdevicebasedonthesamespecifications.
IMPORTANT• Fortheremotetype,itisnecessarytosetthedeviceinformationoftheremotesensortothe
parametersoftheremotetransmitter.• Beforeusingtherestorefunction,besuretoremovethisproductfrompipinglineorstopthe
line.• Afterexecutingtherestorefunction,checkthatbackupdataissetforthetargetparameter.
Therestoremethodcanbeselectedfromthefollowingfourtypes.DuplicateData Restoresthetargetparameter(excludingthefactoryadjustedvalueofthetransmitter)
fromtheproductforwhichtheparameterisbackedup.Restoreisalsoexecutableforotherproducts.
RestoreData Restoresthetargetparameter(includingthefactoryadjustedvalueofthetransmitter)fromtheproductforwhichtheparameterisbackedup.Restoreisnotexecutableforotherproducts.
CompulsionData Restoresthetargetparameter(specifiedatfactoryshipping)fromtheproductforwhichtheparameterisbackedup.Restoreisalsoexecutableforotherproducts.
Restorefactory Restoresalltheparameters(specifieduponshipmentfromthemanufacturingfactory).
Whenusingtherestorefunction,alwaysmakesurethatthebacked-updeviceinformationmatchesthedeviceinformationtoberestored.Thetablebelowshowstheconsistencyofthedeviceinformation. :ItemstobeMatched.
Device Information Duplicate Data Restore Data Compulsion DataOption Board ID -TransmitterSerialNo - -Sensor Serial No - -MainBoardSoftwareRev. -SensorBoardSoftwareRev. -DisplayBoardSoftwareRev. -Model(Note) -CommunicationandI/OCode -
NOTEDatacanonlybeduplicatedorrestoredbetweenproductsofthesamemodel.
<4. Functions> 146
IM 01E22C02-01EN
Thissettingcanbeconfiguredwiththefollowingparameters.Fordetailsabouttherestorableparameters,refertoSubsection4.13.3.Menu path
Display Devicesetup►Diag/Service►Parambkup/restore►(seebelow)BRAIN N:EVENTMANAGE►(seebelow)
Parameter DescriptionDisplay BRAINRestoreexecute N20:RESTOREEXEC Specifiestheexecutionoftherestorefunction.*1Restoreresult N21:RESTORERSLT Displaystheresultoftherestorefunction.*2
*1: Fromthetablebelow,selecttheexecutionoftherestorefunction.Selection DescriptionDisplay BRAIN
Notexecute NotExecute Does not restore data.
Duplicatedata1 DuplicateData1Restorestheparameter(DuplicateData)settingfrommemory1onthedisplayboardtothemainboardoftheproduct.
Duplicatedata2 DuplicateData2Restorestheparameter(DuplicateData)settingfrommemory2onthedisplayboardtothemainboardoftheproduct.
Duplicatedata3 DuplicateData3Restorestheparameter(DuplicateData)settingfrommemory3onthedisplayboardtothemainboardoftheproduct.
DuplicateSD DuplicateSDRestorestheparameter(DuplicateData)settingfromthemicroSDcardtothemainboardoftheproduct.
Restoredata1 RestoreData1Restorestheparameter(RestoreData)settingfrommemory1onthedisplayboardtothemainboardoftheproduct.
Restoredata2 RestoreData2Restorestheparameter(RestoreData)settingfrommemory2onthedisplayboardtothemainboardoftheproduct.
Restoredata3 RestoreData3Restorestheparameter(RestoreData)settingfrommemory3onthedisplayboardtothemainboardoftheproduct.
RestoreSD RestoreSD Restorestheparameter(RestoreData)settingfromthemicroSDcardtothemainboardoftheproduct.
Compulsiondata1 CompulsionData1Restorestheparameter(CompulsionData)settingfrommemory1onthedisplayboardtothemainboardoftheproduct.
Compulsiondata2 CompulsionData2Restorestheparameter(CompulsionData)settingfrommemory2onthedisplayboardtothemainboardoftheproduct.
Compulsiondata3 CompulsionData3Restorestheparameter(CompulsionData)settingfrommemory3onthedisplayboardtothemainboardoftheproduct.
CompulsionSD CompulsionSDRestorestheparameter(CompulsionData)settingfromthemicroSDcardtothemainboardoftheproduct.
Restorefactory RestoreFactory Restorestothestatusthatissetuponshipmentfromthemanufacturingfactory.
*2: Theresultoftherestorefunctionisdisplayedasshownbelow.Selection DescriptionDisplay BRAIN
Unexecuted Unexecuted HavenotrestoredSuccess Success Succeededinrestoration.Failure Failure Failedinrestoration.Running Running Restorerunning
<4. Functions> 147
IM 01E22C02-01EN
Thetablebelowshowsalarmswhoseparameterscanberestoredorduplicatedwhenanalarmoccurs.� :Executablewhenanalarmoccurs. ―:Notexecutablewhenalarmoccurs.
Alarm name RestoreDisplay BRAIN010:MainCPUFAIL 10:MainCPUFAIL ―011:RevcalcFAIL 11:RevcalFAIL ―012:MainEEPFAIL 12:MainEEPFAIL ―013:MainEEPdflt 13:MainEEPdflt ―014:SnsrbdFAIL 14:SnsrbdFAIL 015:SnsrcommERR 15:SnsrcommERR 016:AD1FAIL[Sig] 16:A/D1FAIL 017:AD2FAIL[Excit] 17:A/D2FAIL 018:Coil open 18:Coil open 019:Coilshort 19:Coilshort 020:ExciterFAIL 20:ExciterFAIL 021:PWM1stop 21-22:PWMstop ―022:PWM2stop ―023:Optbdmismatch 23:Optmismatch ―024:OptbdEEPFAIL 24:OptEEPFAIL ―025:OptbdA/DFAIL 25:OptA/DFAIL ―026:OptbdSPIFAIL 26:OptSPIFAIL ―027:RestoreFAIL 27:RestoreFAIL 028:IndbdFAIL
28-32:IndctFAIL
―029:IndbdEEPFAIL ―030:LCDdrvFAIL ―031:Indbdmismatch ―032:IndcommERR ―033:microSDFAIL 33:microSDFAIL ―050:Signaloverflow 50:Sigoverflow 051:Emptydetect 51:Emptydetect 052:H/LHH/LLalm 52:H/LHH/LLalm 053:Adhoverlv4 53:Adhoverlv4 060:SpancfgERR 60:SpancfgERR ―061:PVFcfgERR 61:PVFcfgERR ―062:AO14-20lmt 62:AO14-20lmt ―063:AO24-20lmt 63:AO24-20lmt ―064:AO1mltrng 64:AO1mltrng ―065:H/LcfgERR 65:AOH/Lcfg ―066:DensitycfgERR 66:DenscfgERR ―067:Pls1cfgERR 67:Pls1cfgERR ―068:Pls2cfgERR 68:Pls2cfgERR ―069:Nomisizecfg 69:Nomisizecfg ―070:AdhcfgERR 70:AdhcfgERR ―071:FLNcfgERR 71:FLNcfgERR ―072:Lognotstart 72:Lognotstart ―080:AO1saturate
80-81:AOsaturt 081:AO2saturate 082:Pls1saturate
82-83:Plssaturt 083:Pls2saturate 084:AIsaturate 84:AIsaturate 085:Cablemiscon 85:Cablemiscon 086:Coilinsulation 86:Coilinsulate 131:Transmismatch 131:Trnmismatch ―087:Adhesionlv3 87:Adhoverlv3 088:LCwarn 88:LCwarn
<4. Functions> 148
IM 01E22C02-01EN
Alarm name RestoreDisplay BRAIN089:Insudetect 89:Insudetect 090:FLNoverlv3 90:FLNoverlv3 091:FLNoverlv4 91:FLNoverlv4 092:AZwarn 92:AZwarn 093:Verifwarn 93:Verifwarn 094:Factnoisewarn 94:FCnoisewarn 095:Simulateactive 95:Simactive 096:AO1fix
96-97:AOfixed 097:AO2fix 098:Pls1fix
98-99:Plsfixed 099:Pls2fix 100:AIfix 100:AIfix 101:Paramrestorerun 101:Prmrestore ―102:Disp over 102:Disp over ―103:SDsizewarn 103:SDsizewarn ―104:Bkupincmplt 104:Bkupincmplt 105:SDmismatch 105:SDmismatch ―106:SDremovalERR 106:SDremovERR ―120:Watchdog 120:Watchdog*1 121:Poweroff 121:Poweroff*1 122:InstpowerFAIL 122:InstPWFAIL*1
123:Parambkuprun 123:bkuprun ―124:Datalogrun 124:Datalogrun
IMPORTANTWhenusingtherestorefunction,besuretoprepareabackupfileinthebuilt-inmemoryorthemicroSDcard.Notethattherestorefunctionisnotexecutableifabackupfileisnotprovided.
<4. Functions> 149
IM 01E22C02-01EN
4.13.3 Backup and Restore ParametersThelistbelowshowstheparametersthatcanbebackeduporrestored.However,thedatetoberestoredwith“CompulsionData”isthevalue(s)specifiedbythecustomeruponshipmentfromthemanufacturingfactory.(Defaultvaluewhennotdesignated)
Parameter Restore or Duplicate
Display BRAIN Duplicate Data
Restore Data
Compulsion Data
Restore Factory
DampAO/F C10:VELODMPA/F ―
Damppls/ttl C11:VELODMPP/T ―
DampAO/F C12:VOLDMPA/F ― Damppls/ttl C13:VOLDMPP/T ― DampAO/F C14:MASSDMPA/F ― Damppls/ttl C15:MASSDMPP/T ― DampAO/F C16:CALDMPA/F ― Damppls/ttl C17:CALDMPP/T ― LowMF C20:LOWMF HighMF C21:HIGHMF Flowsensorsel C24:FLOWSENSOR ― Nominalsizeunit C26:SIZEUNIT Nominalsize C27:NOMINALSIZE PVflowselect C30:PVFLOWSEL Unit C31:VELOFUNIT ― ― Unit C32:VOLFUNIT ― ― Unit C33:MASSFUNIT ― ― Timeunit C35:TIMEUNIT ― ― Span C36:VELOFSPAN Span C37:VOLFSPAN Span C38:MASSFSPAN Span C39:CALFSPAN ― Zerovalue C52:ZERO ― Convfactor D11:TOT1CONVFC ― ― Lowcut D12:TOT1LOWCUT Failureopts D13:TOT1FOPTS ― Options D14:TOT1OPTS ― Start/Stop D15:TOT1EXEC ― Reset/Preset D16:TOT1PRESET ― Presetvalue D17:TOT1PREVAL ― Set point D18:TOT1SETPNT ― Lowcut D22:TOT2LOWCUT ― Failureopts D23:TOT2FOPTS ― Options D24:TOT2OPTS ― Start/Stop D25:TOT2EXEC ― Reset/Preset D26:TOT2PRESET ― Presetvalue D27:TOT2PREVAL ― Set point D28:TOT2SETPNT ― Lowcut D32:TOT3LOWCUT ― Failureopts D33:TOT3FOPTS ― Options D34:TOT3OPTS ― Start/Stop D35:TOT3EXEC ― Reset/Preset D36:TOT3PRESET ―
Presetvalue D37:TOT3PREVAL ―
Set point D38:TOT3SETPNT ― Outputmode E10:P1OUTMODE Activemode E11:P1ACTMODE ― Fixwidth E12:P1WIDTH ―
<4. Functions> 150
IM 01E22C02-01EN
Parameter Restore or Duplicate
Display BRAIN Duplicate Data
Restore Data
Compulsion Data
Restore Factory
Rateunit E13:P1RATEUNIT ― ― Ratevalue E14:P1RATEVAL Lowcut E15:P1LOWCUT Alarmout E16:P1ALMOUT ― Frequencyat0% E17:F1AT0% Frequencyat100% E18:F1AT100% SO1function E20:SO1FUNCTION ― Options E21:P1OPTS ― Activepulse E30:P2ACTPULSE ― Outputmode E31:P2OUTMODE ― Pulseselect E32:P2SELECT ― Activemode E33:P2ACTMODE ― Fixwidth E34:P2WIDTH ― Ratevalue E36:P2RATEVAL ― Lowcut E37:P2LOWCUT ― Alarmout E38:P2ALMOUT ― Frequencyat0% E39:F2AT0% ― Frequencyat100% E40:F2AT100% ― SO2function E41:SO2FUNCTION ― Options E42:P2OPTS ― Activemode F20:SO11ACTMODE ― Function F21:SO11FUNC ― Activemode F30:SO12ACTMODE ― Function F31:SO12FUNC ― Activemode F22:SI11ACTMODE ― Function F23:SI11FUNC ― Activemode F32:SI12ACTMODE ― Function F33:SI12FUNC ― Forwardspan2 F40:FWDSPAN2 ― Forwardspan3 F41:FWDSPAN3 ― Forwardspan4 F42:FWDSPAN4 ― Reversespan1 F43:REVSPAN1 ― Reversespan2 F44:REVSPAN2 ― Reversespan3 F45:REVSPAN3 ― Reversespan4 F46:REVSPAN4 ― Autorangehyst F50:AUTORNGHYS ― Bidirectionhyst F51:BIDIRECHYS ― Lowcut G01:AO1LOWCUT ― Highlimit G02:AO1HILIMIT ― Lowlimit G03:AO1LOLIMIT ― Alarmout G04:AO1ALMOUT Rangemode G05:AO1RNGMODE ― AO2select G10:AO2SELECT ― Lowcut G11:AO2LOWCUT ― Highlimit G12:AO2HILIMIT ― Lowlimit G13:AO2LOLIMIT ― Alarmout G14:AO2ALMOUT ― Function G20:AIFUNCTION ― URV G25:AIURV ― LRV G26:AILRV ― AO1trim4mA G31:AO1at4mA ― AO1trim20mA G32:AO1at20mA ―
<4. Functions> 151
IM 01E22C02-01EN
Parameter Restore or Duplicate
Display BRAIN Duplicate Data
Restore Data
Compulsion Data
Restore Factory
AO2trim4mA G41:AO2at4mA ― AO2trim20mA G42:AO2at20mA ― Flowdirect H10:FLOWDIRECT ― Ratelimit H11:RATELIMIT ― Deadtime H12:DEADTIME ― Noisefilter H13:NOISEFILTER ― Pulsingflow H14:PULSINGFLOW ― Powersync H20:POWERSYNCH Setpowerfreq H21:SETPWRFREQ Valueselect H30:DENSITYSEL ― Fixeddensity H32:FIXEDDENS Std density H33:STDDENSITY ― Stdtemperature H34:STDTEMP ― CoefA1 H35:TEMPCOEFA1 ― CoefA2 H36:TEMPCOEFA2 ― Specificheat H40:SPECHEAT ― Fixedtemperature H41:CALFIXTEMP ― Highalarm I10:HIGHALARM ― Lowalarm I11:LOWALARM ― HHalarm I12:HIHIALARM ― LLalarm I13:LOLOALARM ― H/Lalarmhyst I14:H/LALMHYS ― Line 1 J10:LINE1SEL ― ― Line 2 J11:LINE2SEL ― ― Line 3 J12:LINE3SEL ― ― Line 4 J13:LINE4SEL ― ― Line 5 J14:LINE5SEL ― ― Line 6 J15:LINE6SEL ― ― Line 7 J16:LINE7SEL ― ― Line 8 J17:LINE8SEL ― ― FormatPV J20:FORMATFR ― ― Linemode J25:DISPLINE ― ― Disp install J53:DISPINSTALL ― ― - J54:IRSWOPERATE ― ― TagNo K01:TAGNO ― ― Electrodesize K15:EL SIZE ― Modelcode K20:MODEL CODE ― Suffixconfig1 K21:SUFFIXCONF1 ― Suffixconfig2 K22:SUFFIXCONF2 ― Option 1 K23:OPTION1 ― Option 2 K24:OPTION2 ― Option 3 K25:OPTION3 ― Option 4 K26:OPTION4 ― Modelcode K27:RSMDLCD ― Suffixconfig1 K28:RSSUFCONF1 ― Suffixconfig2 K29:RSSUFCONF2 ― Option 1 K30:RSOPT1 ― Option 2 K31:RSOPT2 ― Option 3 K32:RSOPT3 ― Option 4 K33:RSOPT4 ― TransserialNo K34:TRNSSRNO ― Sensor serial No K35:FTSRNO ―
<4. Functions> 152
IM 01E22C02-01EN
Parameter Restore or Duplicate
Display BRAIN Duplicate Data
Restore Data
Compulsion Data
Restore Factory
Function L10:ADHESIONCHK Thresholdlevel1 L11:ADHLEVEL1 ― Thresholdlevel2 L12:ADHLEVEL2 ― Thresholdlevel3 L13:ADHLEVEL3 ― Thresholdlevel4 L14:ADHLEVEL4 ― Checkcycle L18:ADHCHKCYC ― Function L20:FLNOISECHK ― Thresholdlevel1 L21:FLNOISELV1 ― Thresholdlevel2 L22:FLNOISELV2 ― Thresholdlevel3 L23:FLNOISELV3 ― Thresholdlevel4 L24:FLNOISELV4 ― Damp L27:FLNOISEDMP ― Span L28:FLNOISESPN Function L30:LOCONDFUNC ― Lowlimit L32:CONDUCTLMT ― Coilinsulthreshold L37:COILINSTH ― IEXcompare L38:IEXCOMP ― Diagnosticoutput L49:DIAGOUTPUT ― Mode L50:VFMODE ― VFNo L52:VFNO ― Releasetime M01:TESTRELTIM ―
<4. Functions> 153
IM 01E22C02-01EN
4.13.4 Data Logging FunctionWhentheoptionalcodeMC(microSDcard)isselected,thedataloggingfunctioncanstoreuptofourprocessvaluestothemicroSDcard.Tousethisfunction,itisnecessarytospecifyafilename,datastorageinterval,andendingtime.Storeddataissavedinthe“YOKOGAWA”folderasa“.TRD”file.Thissettingcanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Diag/Service►Datalog►(seebelow)BRAIN N:EVENTMANAGE►(seebelow)
Parameter DescriptionDisplay BRAINFilename N30:LOGGINGFILE Specifiesthenameofthefiletobestored.Intervaltime N31:LOGINTRTIM Specifiesthedatastorageinterval.*1Start date N32:LSTARTDATE Displaysthedatetostartthedataloggingfunction.Starttime N33:LSTARTTIME Displaysthetimetostartthedataloggingfunction.Endtime N34:LOGENDTIME Specifiesthetimetoendthedataloggingfunction.*2Log1 N35:LOG1SELECT Specifiesprocessvalue1tobestored.*3Log2 N36:LOG2SELECT Specifiesprocessvalue2tobestored.*3Log3 N37:LOG3SELECT Specifiesprocessvalue3tobestored.*3Log4 N38:LOG4SELECT Specifiesprocessvalue4tobestored.*3
Execute N39:LOGGINGEXEC Specifiestheexecutionofthedataloggingfunction.*4
*1: Fromthetablebelow,selectthedatastorageinterval.Selection DescriptionDisplay BRAIN
1 s 1sec Setsthestorageintervalto1sec.10 s 10sec Setsthestorageintervalto10sec.30 s 30sec Setsthestorageintervalto30sec.1min 1min Setsthestorageintervalto1min.5min 5min Setsthestorageintervalto5min.30min 30min Setsthestorageintervalto30min.1h 1h Setsthestorageintervalto1hour.
*2: Fromthetablebelow,selecttheendingtimeofthedataloggingfunction.Selection
DescriptionDisplay BRAIN
10min 10min Setstheendingtimeto10minuteslater.30min 30min Setstheendingtimeto30minuteslater.1h 1h Setstheendingtimeto1hourlater.3h 3h Setstheendingtimeto3hourslater.12h 12h Setstheendingtimeto12hourslater.24h 24h Setstheendingtimeto24hours(1day)later.72h 72h Setstheendingtimeto72hours(3days)later.240h 240h Setstheendingtimeto240hours(10days)later.720h 720h Setstheendingtimeto720hours(30days)later.1440h 1440h Setstheendingtimeto1440hours(60days)later.
<4. Functions> 154
IM 01E22C02-01EN
*3: Fromthetablebelow,selecttheprocessvaluetobestored.Selection DescriptionDisplay BRAIN
Velocity Velocity Storestheflowvelocity.Volumeflow Volumeflow Storesthevolumetricflowrate.Massflow Massflow Storesthemassflowrate.Calorie Calorie Storesthecalorie.
PV PV StoresthePV-mappedprocessvalueofSubsection4.1.2.
Adhesion Adhesion Storestheresistancevalueoftheelectrodeadhesiondetection.
Flownoise Flownoise Storestheflownoisevalue.ElectrodeA ElectrodeA StoresthevoltageofelectrodeA.ElectrodeB ElectrodeB StoresthevoltageofelectrodeB.Vpeak Vpeak Storesthepeakvalueoftheflowratesignal.
*4: Fromthetablebelow,selecttheuseofthedataloggingfunction.Selection DescriptionDisplay BRAIN
Notexecute Notexecute Doesnotexecutethedataloggingfunction.Execute Execute Executesthedataloggingfunction.
SettingdataisstoredonthemicroSDcardataspecifiedstorageintervalduringaperiodfromthestartofthedataloggingfunctiontotheend.Thefilestoredbythedataloggingfunctioncanbeopenedasatextfile.Example:Ifthestorageintervalissetto“1min”,dataisstoredasshownbelow.
F0437.ai
…………
…………
Date and time Process value 1
Process value 2
Process value 3
Process value 4
2017/01/0112:00:002017/01/0112:01:002017/01/0112:02:002017/01/0112:03:002017/01/0112:04:002017/01/0112:05:00
+9.9863E-01+9.9909E-01+9.9906E-01+9.9859E-01+9.9870E-01+9.9829E-01
+2.8235E+01+2.8248E+01+2.8248E+01+2.8234E+01+2.8237E+01+2.8226E+01
+1.4117E+04+1.4124E+04+1.4124E+04+1.4117E+04+1.4118E+04+1.4113E+04
+4.5600E-01+3.9717E-01+3.1753E-01+4.0430E-01+3.6609E-01+4.1892E-01
TheaccessongoingtothemicroSDcardcanbeconfirmedbytheicononthedisplay.Forexample,theiconsof“microSDcardready”and“AccesstomicroSDcard”aredisplayedalternatelyandrepeatedlybecausetheproductaccessesthemicroSDcardperiodicallyduringitsdatalogging.WhenthemicroSDcardcannotbeaccessedtoremove,theiconof“ErroraccessingmicroSDcard”isdisplayed.InsertthemicroSDcardagainwhenitisnecessarytousethisfunctionagain.Analarmof“microSDfailure”occurswhenthemicroSDcardhasanyproblems.
ReadyformicroSDcard AccessingmicroSDcard
DisabletoaccessmicroSDcard
IMPORTANT• Whenusingthedataloggingfunction,besuretosetthedateandtimeinformation.• Beforeusingthedataloggingfunction,makesurethatthemicroSDcardisprepared.Ittakes
approximately1minuteuntiltheproductisreadytostoredataonthemicroSDcardafterithasbeenturnedon.
• Notethatthedateandtimeinformationareresetifthisproductisturnedoff.• NotethatifthemicroSDcardrunsoutoffreespace,thesubsequentdatawillnotbestored.
<4. Functions> 155
IM 01E22C02-01EN
4.14 Parameter Protection4.14.1 Software Write Protection Function
Writeprotectcanbespecifiedbytwomethods;thehardwarewriteprotectswitchandsoftwarewriteprotect.Whenthewriteprotectissetusingoneofthesemethods,datacannotbewrittentoaparameter.Fordetailsaboutthehardwarewriteprotectionswitch,refertotheInstallationManual.Toenablethesoftwarewriteprotectionfunction,itisnecessarytospecifya“Newpassword”(eightalphanumericcharacters,excludinglowercase).Byinputtingthepasswordin“Enablewrite”,itbecomespossibletodisablethewriteprotectiononlyfor10minutes.Thewriteprotectionfunctioncanbecompletelydisabledbyenteringeightspacesin“Newpassword”whiletheprotectionisdisabledtemporarily.Thewriteprotectionfunctioncanbeconfiguredwiththefollowingparameters.Menu path
Display Devicesetup►Detailedsetup►Protection►(seebelow)BRAIN P:PROTECT►(seebelow)
Parameter DescriptionDisplay BRAINWriteprotectsts P20:WRTPROTECT Indicatestheuseofthewriteprotectionfunction.*1
Enablewrite P21:ENABLEWRITE Cancelsthewriteprotectionfunctiononlyfor10minutes.
Newpassword P22:NEWPASSWORD Specifiesanewpasswordforthewriteprotectionfunction.
*1: DisplayoftheusestatusofthewriteprotectionfunctionValue DescriptionDisplay BRAIN
No No Thewriteprotectionfunctionisnotinuse.(Parametercanbechanged)
Yes Yes Thewriteprotectionfunctionisbeingused.(Parametercannotbechanged)
NOTETheusagestatusofthewriteprotectfunctioncanbecheckedbytheicondisplayedontheparameterordisplay.Thefollowingiconsaredisplayed.
Icon DescriptionWriteprotectDisable(Parameterscanbechanged)WriteprotectEnable(Parametercannotbechanged)
NOTEWhenthewriteprotectionswitchontheamplifierboardisturnedon,noparameterscanbechanged.Tochangeaparameter,cancelthewriteprotectionfunctionwiththepre-specifiedpasswordandspecifyanewpassword.
<4. Functions> 156
IM 01E22C02-01EN
Ifthepasswordisforgotten,itispossibletotemporarilydisablethesoftwarewriteprotectionfunctionbyusingtheJokerpassword.WhenthewriteprotectionfunctionisdisabledusingtheJokerpassword,“Break”isdisplayedasthesoftwareseal(becausethewriteprotectionisnotdisabledwiththeregularroute).Tosetthesoftwaresealbackto“Keep”,itneedstosetthewriteprotectionagain.Whenthewriteprotectionistemporarilydisabledusingtheformalpassword,thesoftwaresealissetbackfrom“Break”to“Keep”.TheuseoftheJokerpasswordsealcanbecheckedbythefollowingparameter.Menu path
Display Devicesetup►Detailedsetup►Protection►SoftsealstatusBRAIN P:PROTECT►P23:SOFTSEAL
ChecktheuseoftheJokerpassworddependingonthefollowingparameters.Value DescriptionDisplay BRAIN
Keep Keep Normal
Break Break DisablethewriteprotectionfunctionusingtheJokerpassword.
NOTEIfitisnecessarytousetheJokerpassword,contactYokogawasalesofficeorrepresentative.
4.14.2 Display Restriction FunctionAccesstotheservicemodeislimited.Menu path
Display Devicesetup►Detailedsetup►Protection►Keycode
<5. Parameter Lists> 157
IM 01E22C02-01EN
5. Parameter ListsThischaptershowsparameterlistsofthedisplayandBRAINcommunication.
Eachparameterissetasspecifiedatthetimeofordering.Sinceotherparametershavedefaultvaluesset,besuretorefertothischapterwhenchangingthem.
IMPORTANT
Ifthisproductisturnedoffbefore30secondsaftersettingtheparameters,thesettingswillnotbesavedcorrectly.Pleasekeeptheproductturnedonforover30secondsaftersettingtheparameters.
NOTE
Toobtaincorrectflowsignals,itisnecessarytosetthenominalsize,flowratespanandmeterfactoroftheflowsensor.Thenominalsizeandmeterfactoroftheflowsensoraresetonthisproductwhenshippedfromthemanufacturingfactory,andacustomerisnotrequiredtosetthem.Ifaparticularparametersisspecifiedatthetimeoforder,thisproductisshippedwiththeparameterspecified.Ifaparameterisnotspecifiedatthetimeoforder,thatparameterneedstobesetbythecustomer.
NOTE
Theavailablefunctionsandparametersdisplayedvarydependingonatypeoftheconnectionterminalspecifiedatthetimeofordering.
:Parameterdisplayed
Function of terminal DescriptionCommunication and I/O
CodeD0 D2 D3 D4 D5 D6
Analogoutput1 Regardlessofthecodeselection,parametersrelatedtoanalogoutput1aredisplayed.
Analogoutput2Dependingonthecodeselection,parametersrelatedtoanalogoutput2maynotbedisplayed.No.7-6to7-10,7-21to7-23,13-9
- - - - -
Alarmoutput Regardlessofthecodeselection,parametersrelatedtothealarmoutputaredisplayed.
PulseStatusoutput1 Regardlessofthecodeselection,parametersrelatedtopulse/statusoutput1aredisplayed.
PulseStatusoutput2Dependingonthecodeselection,parametersrelatedtopulse/statusoutput2maynotbedisplayed.No.5-11to5-22,13-10to13-11
- - -
Statusoutput Dependingonthecodeselection,parametersrelatedtothestatusoutputmaynotbedisplayed.
StatusinputDependingonthecodeselection,parametersrelatedtothestatusinputmaynotbedisplayed.No.6-3to6-5,13-14
AnaloginputDependingonthecodeselection,parametersrelatedtotheanaloginputmaynotbedisplayed.No.7-11to7-17,7-24to7-26,8-13to8-21,13-5,13-12
- - - - -
<5. Parameter Lists> 158
IM 01E22C02-01EN
Theparameterlistconsistsofthefollowingitems.Parameter Name Indicatesaparametername.
R/W
Indicatesparameterswhichcanbedisplayedandset.R:Displayedonly,RW:Displayedandspecified.However,whenchangingparametersonthedisplay,aparameterwhichcanbesetdependson operation levels.Operator:Possibletosetadisplaylanguageonthedisplayandparametersrelatedtodisplayitems.Maintenance:Possibletosetparametersallowedtosetwiththeoperatorlevelandparametersrelatedtozeroadjustment.Specialist:Possibletosetallwritableparameters.
Data RangeIndicatesoptionsforselecttypedata.Indicatesthesettingrangeandthenumberofdecimalplacesfornumeric-typedata.Indicatesthelimitednumberofcharactersforalphanumerictypedata.
Default Value Indicatesthedefaultvalueuponshipmentfromthemanufacturingfactory.Unit Indicatestheunitofthedatarange.Reference Referencepageofparametercontents
NOTE
WhensettingnumericdataforBRAINcommunication,themaximumvaluewhichcanbesetis“32000”andtheminimumvalueis“-32000”regardlessofthedecimalpointposition.Pleasenotethatwhensettingthevaluewhosemaximumvalueexceeds"32000”,thevalueissetto“32000”,andwhensettingthevaluewhoseminimumvalueexceeds“-32000”,thevalueissetto“-32000”.Setdatainthefollowingrangebytakingintoaccountthedecimalpointposition.
Number of digits after the decimal point
Range
0 0 to 320001 0.0 to 3200.02 0.00 to 320.003 0.000 to 32.0004 0.0000 to 3.20005 0.00000 to 0.32000
ExampleWhensetting333.33tonumerictypedata,itcannotbesetsince"33333"ofthenumberstringexcludingthedecimalpointgoesover“32000”.Itneedstoset“333.3”bytakingintoaccountthedecimalpointposition.
<5. Parameter Lists> 159
IM 01E22C02-01EN
(1) Display parametersTheseparametersarerelatedtoflowandtotalizerdisplay.
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
1-1 A10:FLOWRATE(%) Flowrate(%) -999.999to+999.999Numberofdecimalplaces:3
-99999.9to+99999.9Numberofdecimalplaces:1
R - % P.33
1-2 A20:FLOWRATE Flowrate -999999to+999999Numberofdecimalplaces:0to5
-999999to+999999Numberofdecimalplaces:0 to 5
R - Flowvelocity:No.3-15 Volumetricflowrate: No.3-16/No.3-19Massflowrate:No.3-17/No.3-19Flownoisecm/s
P.33
1-3 A21:FLWRATE(mA) Analogout1 0.000to+99.999Numberofdecimalplaces:3
0.000to+99.999Numberofdecimalplaces:3
R - mA -
1-4 A30:TOTAL1 Totalizer1 -999999to+999999Numberofdecimalplaces:0to5
-99999999to+99999999Numberofdecimalplaces:0 to 5
R - No.4-1 P.47
1-5 A31:TOTAL2 Totalizer2 -999999to+999999Numberofdecimalplaces:0to5
-99999999to+99999999Numberofdecimalplaces:0 to 5
R - No.4-10 P.47
1-6 A32:TOTAL3 Totalizer3 -999999to+999999Numberofdecimalplaces:0to5
-99999999to+99999999Numberofdecimalplaces:0 to 5
R - No.4-19 P.47
1-7 A33:TOTAL1CNT Totalizer1count -INF*2to+INF*2
Numberofdecimalplaces:0-99999999to+99999999Numberofdecimalplaces:0
R - - P.48
1-8 A34:TOTAL2CNT Totalizer2count -INF*2to+INF*2
Numberofdecimalplaces:0-99999999to+99999999Numberofdecimalplaces:0
R - - P.48
1-9 A35:TOTAL3CNT Totalizer3count -INF*2to+INF*2
Numberofdecimalplaces:0-99999999to+99999999Numberofdecimalplaces:0
R - - P.48
*2:Therangeofsingleprecisionfloat(IEEE754).
<5. Parameter Lists> 160
IM 01E22C02-01EN
(2) EasyconfigurationparametersTheseparametersarefrequentlyusedgeneralfunctions.ParametersforNo.2-2to2-14canbesetbyusingtheEasysetupwizardfunction.
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
2-1 B10:LANGUAGE Language EnglishFrenchGermanItalianSpanishPortugueseRussianChineseJapanese
EnglishFrenchGermanItalianSpanishPortugueseRussianChineseJapanese
RW English - P.110
2-2 B20:VOLDMPA/F DampAO/F +0.1to+200.0Numberofdecimalplaces:1
+0.1to+200.0Numberofdecimalplaces:1
RW 3.0 s P.36
2-3 B21:VOLDMPP/T Damppls/ttl +0.1to+200.0Numberofdecimalplaces:1
+0.1to+200.0Numberofdecimalplaces:1
RW 3.0 s P.36
2-4 B22:VOLFUNIT Unit Ml(Megaliter)m3
kl(Kiloliter)l(liter)cm3
Ml(Megaliter)m³kl(kiloliter)l(liter)cm³
RW m³*1 - P.34
2-5 B23:TIMEUNIT Timeunit /s/min/h
/s/min/h
RW /h*1 - P.35
2-6 B24:VOLFSPAN Span +0.00001to+INF*2
Numberofdecimalplaces:0to5+0.00001to+999999Numberofdecimalplaces:0 to 5
RW 1.00 *1 No.3-16/No.3-19
P.35
2-7 B30:P1RATEUNIT P1unit nunit/Puunit/Pmunit/PUnit/Pkunit/PMunit/PnP/unituP/unitmP/unitP/unitkP/unitMP/unit
nunit/Puunit/Pmunit/PUnit/Pkunit/PMunit/PnP/unituP/unitmP/unitP/unitkP/unitMP/unit
RW Unit/P*1 - P.58
2-8 B31:P1RATEVAL P1val 0.00000to+INF*2
Numberofdecimalplaces:0to50.00000to+999999Numberofdecimalplaces:0 to 5
RW 0.0 No.5-4 P.57
2-9 B32:F1AT0% F1at0% 0 to 12500Numberofdecimalplaces:0
0 to 12500Numberofdecimalplaces:0
RW 0 Hz P.59
2-10 B33:F1AT100% F1at100% 0 to 12500Numberofdecimalplaces:0
0 to 12500Numberofdecimalplaces:0
RW 0 Hz P.59
2-11 B40:LINE1 SEL Line 1 Flowrate(%)PVVelocityVolumeflowMassflowFlowrate(%Bar)CalorieTotalizer1Totalizer2Totalizer3TagnumberCommunProtocolAdhesionAnalogout1Analogout2FlownoiselevelTotalizer1countTotalizer2countTotalizer3count
Flowrate(%)PVVelocityVolumeflowMassflowFlowrate(%Bar)CalorieTotalizer1Totalizer2Totalizer3TagnumberCommunprotocolAdhesionAnalogout1Analogout2FlownoiselevelTotalizer1countTotalizer2countTotalizer3count
RW PV - P.111
*1:Determinedbasedontheinformationaboutspecifieditemsatthetimeandorderingorthesensortobecombinedwith.*2:Therangeofsingleprecisionfloat(IEEE754).
<5. Parameter Lists> 161
IM 01E22C02-01EN
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
2-12 B41:LINE2 SEL Line 2 NoneFlowrate(%)PVVelocityVolumeflowMassflowFlowrate(%Bar)CalorieTotalizer1Totalizer2Totalizer3TagnumberCommunProtocolAdhesionAnalogout1Analogout2FlownoiselevelTotalizer1countTotalizer2countTotalizer3count
NoneFlowrate(%)PVVelocityVolumeflowMassflowFlowrate(%Bar)CalorieTotalizer1Totalizer2Totalizer3TagnumberCommunprotocolAdhesionAnalogout1Analogout2FlownoiselevelTotalizer1countTotalizer2countTotalizer3count
RW Flowrate(%) - P.111
2-13 B42:LINE3 SEL Line 3 RW Analogout1 - P.111
2-14 B50:AUTOZEROEXE Autozeroexe NotexecuteExecute
NotexecuteExecute
RW Notexecute - P.45
*1:Determinedbasedontheinformationaboutspecifieditemsatthetimeandorderingorthesensortobecombinedwith.*2:Therangeofsingleprecisionfloat(IEEE754).
<5. Parameter Lists> 162
IM 01E22C02-01EN
(3) BasicconfigurationparametersTheseparametersarerelatedtobasicsettings,suchasaflowsensor.
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
3-1 C10:VELODMPA/F DampAO/F +0.1to+200.0Numberofdecimalplaces:1
+0.1to+200.0Numberofdecimalplaces:1
RW 3.0 s P.36
3-2 C11:VELODMPP/T Damppls/ttl +0.1to+200.0Numberofdecimalplaces:1
+0.1to+200.0Numberofdecimalplaces:1
RW 3.0 s P.36
3-3 C12:VOLDMPA/F DampAO/F +0.1to+200.0Numberofdecimalplaces:1
+0.1to+200.0Numberofdecimalplaces:1
RW 3.0 s P.36
3-4 C13:VOLDMPP/T Damppls/ttl +0.1to+200.0Numberofdecimalplaces:1
+0.1to+200.0Numberofdecimalplaces:1
RW 3.0 s P.36
3-5 C14:MASSDMPA/F DampAO/F +0.1to+200.0Numberofdecimalplaces:1
+0.1to+200.0Numberofdecimalplaces:1
RW 3.0 s P.36
3-6 C15:MASSDMPP/T Damppls/ttl +0.1to+200.0Numberofdecimalplaces:1
+0.1to+200.0Numberofdecimalplaces:1
RW 3.0 s P.36
3-7 C16:CALDMPA/F DampAO/F +0.1to+200.0Numberofdecimalplaces:1
+0.1to+200.0Numberofdecimalplaces:1
RW 3.0 s P.36
3-8 C17:CALDMPP/T Damppls/ttl +0.1to+200.0Numberofdecimalplaces:1
+0.1to+200.0Numberofdecimalplaces:1
RW 3.0 s P.36
3-9 C20:LOWMF LowMF +0.0100to+3.0000Numberofdecimalplaces:4
+0.0100to+3.0000Numberofdecimalplaces:4
RW 1.0000 *1 - P.39
3-10 C21:HIGHMF HighMF +0.0100to+3.0000Numberofdecimalplaces:4
+0.0100to+3.0000Numberofdecimalplaces:4
RW 1.0000 *1 - P.39
3-11 C24:FLOWSENSOR Flowsensorsel ADMAGAXGADMAGAXWADMAGAXFADMAGADMAGAEADMAGSEYEWMAGCalibratorOther1Other2Other3
ADMAGAXGADMAGAXWADMAGAXFADMAGADMAGAEADMAGSEYEWMAGCalibratorOther1Other2Other3
RW ADMAGAXG*1
- P.40
3-12 C26:SIZEUNIT Nominalsizeunit mminch
mminch
RW mm - P.41
3-13 C27:NOMINALSIZE Nominalsize +0.01000to+3000.0Numberofdecimalplaces:0to5
+0.01000to+3000.0Numberofdecimalplaces:0 to 5
RW 100 *1 No.3-12 P.41
3-14 C30:PVFLOWSEL PVflowselect VelocityVolumeMassDiag
VelocityVolumeMassDiag
RW Volume*1 - P.33
3-15 C31:VELOFUNIT Unit m/sft/s
m/sft/s
RW m/s*1 - P.34
3-16 C32:VOLFUNIT Unit Ml(Megaliter)m3
kl(Kiloliter)l(liter)cm3
kcfcfmcfMgal(US)kgal(US)gal(US)mgal(US)kbbl(USOil)bbl(USOil)mbbl(USOil)ubbl(USOil)kbbl(USBeer)bbl(USBeer)mbbl(USBeer)ubbl(USBeer)
Ml(Megaliter)m³kl(kiloliter)l(liter)cm³kcfcfmcfMgal(US)kgal(US)gal(US)mgal(US)kbbl(USOil)bbl(USOil)mbbl(USOil)ubbl(USOil)kbbl(USBeer)bbl(USBeer)mbbl(USBeer)ubbl(USBeer)
RW m³*1 - P.34
3-17 C33:MASSFUNIT Unit tkggklblb
tkggklblb
RW kg*1 - P.34
3-18 C34:CALUNIT Unit MJkJJkcalcalBTU
MJkJJkcalcalBTU
RW J - P.34
*1:Determinedbasedontheinformationaboutspecifieditemsatthetimeoforderingandthesensortobecombinedwith.*2:Therangeofsingleprecisionfloat(IEEE754).
<5. Parameter Lists> 163
IM 01E22C02-01EN
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
3-19 C35:TIMEUNIT Timeunit /s/min/h/d
/s/min/h/d
RW /h*1 - P.35
3-20 C36:VELOFSPAN Span +0.00001to+INF*2
Numberofdecimalplaces:0to5+0.00001to+999999Numberofdecimalplaces:0 to 5
RW 1.00000 *1 No.3-15 P.35
3-21 C37:VOLFSPAN Span +0.00001to+INF*2
Numberofdecimalplaces:0to5+0.00001to+999999Numberofdecimalplaces:0 to 5
RW 1.00000 *1 No.3-16/No.3-19
P.35
3-22 C38:MASSFSPAN Span +0.00001to+INF*2
Numberofdecimalplaces:0to5+0.00001to+999999Numberofdecimalplaces:0 to 5
RW 1.00000 *1 No.3-17/No.3-19[kg/h]
P.35
3-23 C39:CALFSPAN Span +0.00000to+INF*2
Numberofdecimalplaces:0to5+0.00001to+999999Numberofdecimalplaces:0 to 5
RW 1.00000 No.3-18/No.3-19[J/h]
P.35
3-24 C40:VELOCITYCHK Velocitycheck -99.999to+99.999Numberofdecimalplaces:3
-99.999to+99.999Numberofdecimalplaces:3
R - m/s P.42
3-25 C41:USRSPNSEL1 Select NoYes
NoYes
RW No - P.43
3-26 C42:USRUNIT1 Unit ASCII8characters ASCII8characters RW AllSpace - P.433-27 C43:USRSPAN1 Span +0.00001to+INF*2
Numberofdecimalplaces:0to5+0.00001to+999999Numberofdecimalplaces:0 to 5
RW 100.000 - P.44
3-28 C44:USRSPNSEL2 Select NoYes
NoYes
RW No - P.43
3-29 C45:USRUNIT2 Unit ASCII8characters ASCII8characters RW AllSpace - P.433-30 C46:USRSPAN2 Span +0.00001to+INF*2
Numberofdecimalplaces:0to5+0.00001to+999999Numberofdecimalplaces:0 to 5
RW 100.000 - P.44
3-31 C51:AUTOZEROEXE Execute NotexecuteExecute
NotexecuteExecute
RW Notexecute - P.45
3-32 C52:ZERO Zerovalue -99.999to+99.999Numberofdecimalplaces:3
-99.999to+99.999Numberofdecimalplaces:3
RW 0.000 cm/s P.46
*1:Determinedbasedontheinformationaboutspecifieditemsatthetimeoforderingandthesensortobecombinedwith.*2:Therangeofsingleprecisionfloat(IEEE754).
<5. Parameter Lists> 164
IM 01E22C02-01EN
(4) TotalizationconfigurationparametersTheseparametersarerelatedtothetotalizationfunction.
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
4-1 D10:TOT1UNIT Unit Ml(Megaliter)m3
kl(Kiloliter)l(liter)cm3
tkggkcfcfmcfMgal(US)kgal(US)gal(US)mgal(US)kbbl(USOil)bbl(USOil)mbbl(USOil)ubbl(USOil)kbbl(USBeer)bbl(USBeer)mbbl(USBeer)ubbl(USBeer)klb(US)lb(US)
Ml(Megaliter)m³kl(kiloliter)l(liter)cm³tkggkcfcfmcfMgal(US)kgal(US)gal(US)mgal(US)kbbl(USOil)bbl(USOil)mbbl(USOil)ubbl(USOil)kbbl(USBeer)bbl(USBeer)mbbl(USBeer)ubbl(USBeer)klb(US)lb(US)
R - - P.47
4-2 D11:TOT1CONVFC Convfactor -INF*2to+INF*2
Unabletoinput0Numberofdecimalplaces:0to5
-999999to+999999Numberofdecimalplaces:0 to 5
RW 1.00000 No.4-1 P.48
4-3 D12:TOT1LOWCUT Lowcut 0.00000to+INF*2
Numberofdecimalplaces:0to50.00000to+999999Numberofdecimalplaces:0 to 5
RW 0.00000 *1 Volumetricflowrate: No.3-16/No.3-19Massflowrate:No.3-17/No.3-19Calorie: No.3-18/No.3-19
P.38
4-4 D13:TOT1FOPTS Failureopts MeasuredvalueStopLast valid
MeasuredvalueStopLast valid
RW Stop - P.50
4-5 D14:TOT1OPTS Options BalancedAbsoluteOnly positiveOnlynegativeHold
BalancedAbsoluteOnly positiveOnlynegativeHold
RW Only positive - P.51
4-6 D15:TOT1EXEC Start/Stop StopStart
StopStart
RW Stop - P.50
4-7 D16:TOT1PRESET Reset/Preset NotexecuteResetPreset
NotexecuteResetPreset
RW Notexecute - P.52
4-8 D17:TOT1PREVAL Presetvalue -INF*2to+INF*2
Numberofdecimalplaces:0to5-999999to+999999Numberofdecimalplaces:0 to 5
RW 0.00000 Volumetricflowrate: No.3-16Massflowrate:No.3-17Calorie: No.3-18
P.52
4-9 D18:TOT1SETPNT Set point 0.00000to+INF*2
Numberofdecimalplaces:0to50.00000to+999999Numberofdecimalplaces:0 to 5
RW 0.00000 Volumetricflowrate: No.3-16Massflowrate:No.3-17Calorie: No.3-18
P.49
*1:Determinedbasedontheinformationaboutspecifieditemsatthetimeandorderingorthesensortobecombinedwith.*2:Therangeofsingleprecisionfloat(IEEE754).*3:Displayedonlywhenanaloginputisvalid.
<5. Parameter Lists> 165
IM 01E22C02-01EN
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
4-10 D20:TOT2UNIT Unit Ml(Megaliter)m3
kl(Kiloliter)l(liter)cm3
tkggkcfcfmcfMgal(US)kgal(US)gal(US)mgal(US)kbbl(USOil)bbl(USOil)mbbl(USOil)ubbl(USOil)kbbl(USBeer)bbl(USBeer)mbbl(USBeer)ubbl(USBeer)klb(US)lb(US)MJ*3
kJ*3
J*3
kcal*3cal*3BTU*3
Ml(Megaliter)m³kl(kiloliter)l(liter)cm³tkggkcfcfmcfMgal(US)kgal(US)gal(US)mgal(US)kbbl(USOil)bbl(USOil)mbbl(USOil)ubbl(USOil)kbbl(USBeer)bbl(USBeer)mbbl(USBeer)ubbl(USBeer)klb(US)lb(US)MJ*3
kJ*3
J*3
kcal*3cal*3BTU*3
RW m³ - P.47
4-11 D21:TOT2CONVFC Convfactor -INF*2to+INF*2
Unabletoinput0Numberofdecimalplaces:0to5
-999999to+999999Numberofdecimalplaces:0 to 5
RW 1.00000 No.4-10 P.48
4-12 D22:TOT2LOWCUT Lowcut 0.00000to+INF*2
Numberofdecimalplaces:0to50.00000to+999999Numberofdecimalplaces:0 to 5
RW 0.00000 No.4-10/No.3-19
P.38
4-13 D23:TOT2FOPTS Failureopts MeasuredvalueStopLast valid
MeasuredvalueStopLast valid
RW Stop - P.50
4-14 D24:TOT2OPTS Options BalancedAbsoluteOnly positiveOnlynegativeHold
BalancedAbsoluteOnly positiveOnlynegativeHold
RW Onlynegative - P.51
4-15 D25:TOT2EXEC Start/Stop StopStart
StopStart
RW Stop - P.50
4-16 D26:TOT2PRESET Reset/Preset NotexecuteResetPreset
NotexecuteResetPreset
RW Notexecute - P.52
4-17 D27:TOT2PREVAL Presetvalue -INF*2to+INF*2
Numberofdecimalplaces:0to5-999999to+999999Numberofdecimalplaces:0 to 5
RW 0.00000 No.4-10 P.52
4-18 D28:TOT2SETPNT Set point 0.00000to+INF*2
Numberofdecimalplaces:0to50.00000to+999999Numberofdecimalplaces:0 to 5
RW 0.00000 No.4-10 P.49
*1:Determinedbasedontheinformationaboutspecifieditemsatthetimeandorderingorthesensortobecombinedwith.*2:Therangeofsingleprecisionfloat(IEEE754).*3:Displayedonlywhenanaloginputisvalid.
<5. Parameter Lists> 166
IM 01E22C02-01EN
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
4-19 D30:TOT3UNIT Unit Ml(Megaliter)m3
kl(Kiloliter)l(liter)cm3
tkggkcfcfmcfMgal(US)kgal(US)gal(US)mgal(US)kbbl(USOil)bbl(USOil)mbbl(USOil)ubbl(USOil)kbbl(USBeer)bbl(USBeer)mbbl(USBeer)ubbl(USBeer)klb(US)lb(US)MJ*3
kJ*3
J*3
kcal*3cal*3BTU*3
Ml(Megaliter)m³kl(kiloliter)l(liter)cm³tkggkcfcfmcfMgal(US)kgal(US)gal(US)mgal(US)kbbl(USOil)bbl(USOil)mbbl(USOil)ubbl(USOil)kbbl(USBeer)bbl(USBeer)mbbl(USBeer)ubbl(USBeer)klb(US)lb(US)MJ*3
kJ*3
J*3
kcal*3cal*3BTU*3
RW m³ - P.47
4-20 D31:TOT3CONVFC Convfactor -INF*2to+INF*2
Unabletoinput0Numberofdecimalplaces:0to5
-999999to+999999Numberofdecimalplaces:0 to 5
RW 1.00000 No.4-19 P.48
4-21 D32:TOT3LOWCUT Lowcut 0.00000to+INF*2
Numberofdecimalplaces:0to50.00000to+999999Numberofdecimalplaces:0 to 5
RW 0.00000 No.4-19/No.3-19
P.38
4-22 D33:TOT3FOPTS Failureopts MeasuredvalueStopLast valid
MeasuredvalueStopLast valid
RW Stop - P.50
4-23 D34:TOT3OPTS Options BalancedAbsoluteOnly positiveOnlynegativeHold
BalancedAbsoluteOnly positiveOnlynegativeHold
RW Balanced - P.51
4-24 D35:TOT3EXEC Start/Stop StopStart
StopStart
RW Stop - P.50
4-25 D36:TOT3PRESET Reset/Preset NotexecuteResetPreset
NotexecuteResetPreset
RW Notexecute - P.52
4-26 D37:TOT3PREVAL Presetvalue -INF*2to+INF*2
Numberofdecimalplaces:0to5-999999to+999999Numberofdecimalplaces:0 to 5
RW 0.00000 No.4-19 P.52
4-27 D38:TOT3SETPNT Set point 0.00000to+INF*2
Numberofdecimalplaces:0to50.00000to+999999Numberofdecimalplaces:0 to 5
RW 0.00000 No.4-19 P.49
*1:Determinedbasedontheinformationaboutspecifieditemsatthetimeandorderingorthesensortobecombinedwith.*2:Therangeofsingleprecisionfloat(IEEE754).*3:Displayedonlywhenanaloginputisvalid.
<5. Parameter Lists> 167
IM 01E22C02-01EN
(5) PulseoutputconfigurationparametersTheseparametersarerelatedtothepulse/statusoutput1/2.
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
5-1 E10:P1OUTMODE Outputmode NofunctionFixedpulseFrequencyoutputStatusoutput
NofunctionFixedpulseoutputFrequencyoutputStatusoutput
RW Fixedpulseoutput - P.53
5-2 E11:P1ACTMODE Activemode OnactiveOffactive
OnactiveOffactive
RW Onactive - P.56
5-3 E12:P1WIDTH Fixwidth 0.05ms0.1ms0.5ms1ms20ms33ms50ms100ms200ms330ms500ms1000ms2000msDutycycle50%
0.05ms0.1ms0.5ms1ms20ms33ms50ms100ms200ms330ms500ms1000ms2000msDutycycle50%
RW Dutycycle50% - P.55
5-4 E13:P1RATEUNIT Rateunit nunit/Puunit/Pmunit/PUnit/Pkunit/PMunit/PnP/unituP/unitmP/unitP/unitkP/unitMP/unit
nunit/Puunit/Pmunit/PUnit/Pkunit/PMunit/PnP/unituP/unitmP/unitP/unitkP/unitMP/unit
RW Unit/P*1 - P.58
5-5 E14:P1RATEVAL Ratevalue 0.00000to+INF*2
Numberofdecimalplaces:0to50.00000to+999999Numberofdecimalplaces:0 to 5
RW 0.0 *1 No.5-4 P.57
5-6 E15:P1LOWCUT Lowcut 0.00000to+INF*2
Numberofdecimalplaces:0to50.00000to+999999Numberofdecimalplaces:0 to 5
RW 0.0 *1 Flowvelocity:No.3-15Volumetricflowrate: No.3-16/No.3-19Massflowrate:No.3-17/No.3-19Flownoisecm/s
P.37
5-7 E16:P1ALMOUT Alarmout 0 ppsMeasuredvalueLast validMaxpps
0 ppsMeasuredvalueLast validMaxpps
RW 0 pps - P.54
5-8 E17:F1AT0% Frequencyat0% 0 to 12500Numberofdecimalplaces:0
0 to 12500Numberofdecimalplaces:0
RW 0 Hz P.59
5-9 E18:F1AT100% Frequencyat100% 0 to 12500Numberofdecimalplaces:0
0 to 12500Numberofdecimalplaces:0
RW 0 Hz P.59
5-10 E20:SO1FUNCTION SO1function NofunctionWarningoutputTotallimit1Totallimit2Totallimit3H/LalarmHH/LLalarmFwd/RevrangeAuto2rangeExt2answer
NofunctionWarningoutputTotallimit1Totallimit2Totallimit3H/LalarmHH/LLalarmFwd/RevrangeAuto2rangeExt2answer
RW Nofunction - P.60
5-11 E21:P1OPTS Options BalancedAbsoluteOnly positiveOnlynegative
BalancedAbsoluteOnly positiveOnlynegative
RW Only positive - P.63
5-12 E30:P2ACTPULSE Activepulse NormalFormagneticcounter
NormalFormagneticcounter
RW Normal - P.57
5-13 E31:P2OUTMODE Outputmode NofunctionFixedpulseFrequencyoutputStatusoutput
NofunctionFixedpulseoutputFrequencyoutputStatusoutput
RW Nofunction - P.53
*1:Determinedbasedontheinformationaboutspecifieditemsatthetimeoforderingandthesensortobecombinedwith.*2:Therangeofsingleprecisionfloat(IEEE754).
<5. Parameter Lists> 168
IM 01E22C02-01EN
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
5-14 E32:P2SELECT Pulseselect Non-ConnectVelocityVolumeflowMassflowCalorieDiag
NoconnectVelocityVolumeflowMassflowCalorieDiag
RW Noconnect - P.54
5-15 E33:P2ACTMODE Activemode OnactiveOffactive
OnactiveOffactive
RW Onactive - P.56
5-16 E34:P2WIDTH Fixwidth 0.05ms0.1ms0.5ms1ms20ms33ms50ms100ms200ms330ms500ms1000ms2000msDutycycle50%
0.05ms0.1ms0.5ms1ms20ms33ms50ms100ms200ms330ms500ms1000ms2000msDutycycle50%
RW Dutycycle50% - P.55
5-17 E35:P2RATEUNIT Rateunit nunit/Puunit/Pmunit/PUnit/Pkunit/PMunit/PnP/unituP/unitmP/unitP/unitkP/unitMP/unit
nunit/Puunit/Pmunit/PUnit/Pkunit/PMunit/PnP/unituP/unitmP/unitP/unitkP/unitMP/unit
RW Unit/P - P.58
5-18 E36:P2RATEVAL Ratevalue 0.00000to+INF*2
Numberofdecimalplaces:0to50.00000to+999999Numberofdecimalplaces:0 to 5
RW 0.0 No.5-16 P.57
5-19 E37:P2LOWCUT Lowcut 0.00000to+INF*2
Numberofdecimalplaces:0to50.00000to+999999Numberofdecimalplaces:0 to 5
RW 0.0 Flowvelocity:No.3-15Volumetricflowrate: No.3-16/No.3-19Massflowrate:No.3-17/No.3-19Flownoisecm/s
P.37
5-20 E38:P2ALMOUT Alarmout 0 ppsMeasuredvalueLast validMaxpps
0 ppsMeasuredvalueLast validMaxpps
RW 0 pps - P.54
5-21 E39:F2AT0% Frequencyat0% 0 to 12500Numberofdecimalplaces:0
0 to 12500Numberofdecimalplaces:0
RW 0 Hz P.59
5-22 E40:F2AT100% Frequencyat100% 0 to 12500Numberofdecimalplaces:0
0 to 12500Numberofdecimalplaces:0
RW 0 Hz P.59
*1:Determinedbasedontheinformationaboutspecifieditemsatthetimeoforderingandthesensortobecombinedwith.*2:Therangeofsingleprecisionfloat(IEEE754).
<5. Parameter Lists> 169
IM 01E22C02-01EN
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
5-23 E41:SO2FUNCTION SO2function NofunctionWarningoutputTotallimit1Totallimit2Totallimit3H/LalarmHH/LLalarmFwd/RevrangeAuto2rangeExt2answerSystemalarmProcessalarmSettingalarm50:Sigoverflow51:Emptydetect53:Adhoverlv485:Cablemiscon86:Coilinsulate87:Adhoverlv388:LCwarn89:Insudetect90:FLNoverlv391:FLNoverlv493:Verifwarn94:FCnoisewarnLoalarmLoLoalarmHialarmHiHialarm
NofunctionWarningoutputTotallimit1Totallimit2Totallimit3H/LalarmHH/LLalarmFwd/RevrangeAuto2rangeExt2answerSystemalarmProcessalarmSettingalarmSignaloverflowEmptydetectAdhoverlv4CablemisconCoilinsulationAdhesionlv3LCwarnInsudetectFLNoverlv3FLNoverlv4VerifwarnFactnoisewarnLowalarmLLalarmHighalarmHHalarm
RW Nofunction - P.60
5-24 E42:P2OPTS Options BalancedAbsoluteOnly positiveOnlynegative
BalancedAbsoluteOnly positiveOnlynegative
RW Only positive - P.63
*1:Determinedbasedontheinformationaboutspecifieditemsatthetimeoforderingandthesensortobecombinedwith.*2:Therangeofsingleprecisionfloat(IEEE754).
<5. Parameter Lists> 170
IM 01E22C02-01EN
(6) StatusinputandoutputconfigurationparametersTheseparametersarerelatedtothestatusinput,statusoutputandmultirange.
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
6-1 F20:SO11ACTMODE Activemode OnactiveOffactive
OnactiveOffactive
RW Onactive - P.56
6-2 F21:SO11FUNC Function NofunctionWarningoutputTotallimit1Totallimit2Totallimit3H/LalarmHH/LLalarmFwd/RevrangeAuto2rangeAuto3rangeAuto4rangeExt2answerExt3answerExt4answer
NofunctionWarningoutputTotallimit1Totallimit2Totallimit3H/LalarmHH/LLalarmFwd/RevrangeAuto2rangeAuto3rangeAuto4rangeExt2answerExt3answerExt4answer
RW Nofunction - P.60
6-3 F22:SI11ACTMODE Activemode Short(On)activeOpen(Off)active
Short(On)actOpen(Off)act
RW Short(On)act - P.64
6-4 F23:SI11FUNC Function Nofunction0%SignalLockExtautozeroTotalpreset1Totalpreset2Totalpreset3Ext2rangesExt3rangesExt4ranges0%SignalLock2
Nofunction0%signallockExtautozeroTotalpreset1Totalpreset2Totalpreset3Ext2rangesExt3rangesExt4ranges0%signallock2
RW - P.65
6-5 F24:SI11STATE State OpenShort
OpenClose
R - - -
6-6 F30:SO12ACTMODE Activemode OnactiveOffactive
OnactiveOffactive
RW Onactive - P.56
6-7 F31:SO12FUNC Function NofunctionWarningoutputTotallimit1Totallimit2Totallimit3H/LalarmHH/LLalarmFwd/RevrangeAuto2rangeAuto3rangeAuto4rangeExt2answerExt3answerExt4answer
NofunctionWarningoutputTotallimit1Totallimit2Totallimit3H/LalarmHH/LLalarmFwd/RevrangeAuto2rangeAuto3rangeAuto4rangeExt2answerExt3answerExt4answer
RW Nofunction - P.60
6-8 F32:SI12ACTMODE Activemode Short(On)activeOpen(Off)active
Short(On)actOpen(Off)act
RW Short(On)act - P.64
6-9 F33:SI12FUNC Function Nofunction0%SignalLockExtautozeroTotalpreset1Totalpreset2Totalpreset3Ext2rangesExt3rangesExt4ranges0%SignalLock2
Nofunction0%signallockExtautozeroTotalpreset1Totalpreset2Totalpreset3Ext2rangesExt3rangesExt4ranges0%signallock2
RW - P.65
6-10 F34:SI12STATE State OpenShort
OpenClose
R - - -
6-11 F40:FWDSPAN2 Forwardspan2 0.00001to+INF*2
Numberofdecimalplaces:0to50.00001to+999999Numberofdecimalplaces:0 to 5
RW 1.0 Flowvelocity:No.3-15Volumetricflowrate: No.3-16/No.3-19Massflowrate:No.3-17/No.3-19Flownoisecm/s
P.73
*2:Therangeofsingleprecisionfloat(IEEE754).
<5. Parameter Lists> 171
IM 01E22C02-01EN
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
6-12 F41:FWDSPAN3 Forwardspan3 0.00001to+INF*2
Numberofdecimalplaces:0to50.00001to+999999Numberofdecimalplaces:0 to 5
RW 1.0 Flowvelocity:No.3-15Volumetricflowrate: No.3-16/No.3-19Massflowrate:No.3-17/No.3-19Flownoisecm/s
P.73
6-13 F42:FWDSPAN4 Forwardspan4 0.00001to+INF*2
Numberofdecimalplaces:0to50.00001to+999999Numberofdecimalplaces:0 to 5
RW 1.0 Flowvelocity:No.3-15Volumetricflowrate: No.3-16/No.3-19Massflowrate:No.3-17/No.3-19Flownoisecm/s
P.73
6-14 F43:REVSPAN1 Reversespan1 0.00001to+INF*2
Numberofdecimalplaces:0to50.00001to+999999Numberofdecimalplaces:0 to 5
RW 1.0 Flowvelocity:No.3-15Volumetricflowrate: No.3-16/No.3-19Massflowrate:No.3-17/No.3-19Flownoisecm/s
P.73
6-15 F44:REVSPAN2 Reversespan2 0.00001to+INF*2
Numberofdecimalplaces:0to50.00001to+999999Numberofdecimalplaces:0 to 5
RW 1.0 Flowvelocity:No.3-15Volumetricflowrate: No.3-16/No.3-19Massflowrate:No.3-17/No.3-19Flownoisecm/s
P.73
6-16 F45:REVSPAN3 Reversespan3 0.00001to+INF*2
Numberofdecimalplaces:0to50.00001to+999999Numberofdecimalplaces:0 to 5
RW 1.0 Flowvelocity:No.3-15Volumetricflowrate: No.3-16/No.3-19Massflowrate:No.3-17/No.3-19Flownoisecm/s
P.73
6-17 F46:REVSPAN4 Reversespan4 0.00001to+INF*2
Numberofdecimalplaces:0to50.00001to+999999Numberofdecimalplaces:0 to 5
RW 1.0 Flowvelocity:No.3-15Volumetricflowrate: No.3-16/No.3-19Massflowrate:No.3-17/No.3-19Flownoisecm/s
P.73
6-18 F50:AUTORNGHYS Autorangehyst 0 to 15Numberofdecimalplaces:0
0 to 15Numberofdecimalplaces:0
RW 10 % P.74
6-19 F51:BIDIRECHYS Bidirectionhyst 0 to 8Numberofdecimalplaces:0
0 to 8Numberofdecimalplaces:0
RW 2 % P.76
*2:Therangeofsingleprecisionfloat(IEEE754).
<5. Parameter Lists> 172
IM 01E22C02-01EN
(7) Analogoutput/inputconfigurationparametersTheseparametersarerelatedtoanalogoutput1/2andanaloginput.
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
7-1 G01:AO1LOWCUT Lowcut 0.00000to+INF*2
Numberofdecimalplaces:0to50.00000to+999999Numberofdecimalplaces:0 to 5
RW 0.0 Flowvelocity:No.3-15Volumetricflowrate: No.3-16/No.3-19Massflowrate:No.3-17/No.3-19Flownoisecm/s
P.37
7-2 G02:AO1HILIMIT Highlimit +4.000to+21.600Numberofdecimalplaces:3
+4.000to+21.600Numberofdecimalplaces:3
RW 20.500 mA P.67
7-3 G03:AO1LOLIMIT Lowlimit +2.400to+20.000Numberofdecimalplaces:3
+2.400to+20.000Numberofdecimalplaces:3
RW 3.800 mA P.67
7-4 G04:AO1ALMOUT Alarmout <2.4mA3.8mA4mA20.5mA>21.6mAMeasuredvalueHold
<2.4mA3.8mA4mA20.5mA>21.6mAMeasuredvalueHold
RW >21.6mA - P.69
7-5 G05:AO1RNGMODE Rangemode NormalrangeAbsrange
NormalrangeAbsrange
RW Normalrange - P.68
7-6 G10:AO2SELECT AO2select Non-ConnectVelocityVolumeflowMassflowCalorieDiag
NoconnectVelocityVolumeflowMassflowCalorieDiag
RW Noconnect - P.66
7-7 G11:AO2LOWCUT Lowcut 0.00000to+INF*2
Numberofdecimalplaces:0to50.00000to+999999Numberofdecimalplaces:0 to 5
RW 0.0 No.7-6 P.37
7-8 G12:AO2HILIMIT Highlimit +4.000to+21.600Numberofdecimalplaces:3
+4.000to+21.600Numberofdecimalplaces:3
RW 20.500 mA P.67
7-9 G13:AO2LOLIMIT Lowlimit +2.400to+20.000Numberofdecimalplaces:3
+2.400to+20.000Numberofdecimalplaces:3
RW 3.800 mA P.67
7-10 G14:AO2ALMOUT Alarmout <2.4mA3.8mA4mA20.5mA>21.6mAMeasuredvalueHold
<2.4mA3.8mA4mA20.5mA>21.6mAMeasuredvalueHold
RW >21.6mA - P.69
7-11 G20:AIFUNCTION Function NofunctionMonitoringDifftemperatureExttemperature
NofunctionMonitoringDifftemperatureExttemperature
RW Nofunction - P.71
7-12 G21:AIHILIMIT Highlimit +4.000to+21.600Numberofdecimalplaces:3
+4.000to+21.600Numberofdecimalplaces:3
RW 20.500 mA P.72
7-13 G22:AILOLIMIT Lowlimit +2.400to+20.000Numberofdecimalplaces:3
+2.400to+20.000Numberofdecimalplaces:3
RW 3.800 mA P.72
7-14 G23:AIVALUE Value 0.000to+22.730Numberofdecimalplaces:3
0.000to+22.730Numberofdecimalplaces:3
R 0.000 mA P.71
7-15 G24:AIUNIT Unit degCdegFK
°C°FK
RW degC/°C - P.71
7-16 G25:AIURV URV -999.9to+999.9Numberofdecimalplaces:1
-999.9to+999.9Numberofdecimalplaces:1
RW 120.0 No.7-15 P.72
7-17 G26:AILRV LRV -999.9to+999.9Numberofdecimalplaces:1
-999.9to+999.9Numberofdecimalplaces:1
RW 0.0 No.7-15 P.72
7-18 G30:AO1TRIMCLR AO1trimclear NotexecuteExecute
NotexecuteExecute
RW Notexecute - P.70
7-19 G31:AO1at4mA AO1trim4mA 3.200 to 5.600Numberofdecimalplaces:3
3.200 to 5.600Numberofdecimalplaces:3
RW 4.000 mA P.70
7-20 G32:AO1at20mA AO1trim20mA 18.400 to 21.600Numberofdecimalplaces:3
18.400 to 21.600Numberofdecimalplaces:3
RW 20.000 mA P.70
7-21 G40:AO2TRIMCLR AO2trimclear NotexecuteExecute
NotexecuteExecute
RW Notexecute - P.70
7-22 G41:AO2at4mA AO2trim4mA 3.200 to 5.600Numberofdecimalplaces:3
3.200 to 5.600Numberofdecimalplaces:3
RW 4.000 mA P.70
7-23 G42:AO2at20mA AO2trim20mA 18.400 to 21.600Numberofdecimalplaces:3
18.400 to 21.600Numberofdecimalplaces:3
RW 20.000 mA P.70
*2:Therangeofsingleprecisionfloat(IEEE754).
<5. Parameter Lists> 173
IM 01E22C02-01EN
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
7-24 G50:AITRIMCLR Trimclear NotexecuteExecute
NotexecuteExecute
RW Notexecute - P.70
7-25 G51:AIat4mA Trim4mA 3.200 to 5.600Numberofdecimalplaces:3
3.200 to 5.600Numberofdecimalplaces:3
RW 4.000 mA P.70
7-26 G52:AIat20mA Trim20mA 18.400 to 21.600Numberofdecimalplaces:3
18.400 to 21.600Numberofdecimalplaces:3
RW 20.000 mA P.70
*2:Therangeofsingleprecisionfloat(IEEE754).
<5. Parameter Lists> 174
IM 01E22C02-01EN
(8) AuxiliaryfunctionconfigurationparametersTheseparametersarerelatedtotheflowdirection,ratelimit,lowcut,temperaturecorrectionfordensity,orcaloriecalculation.
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
8-1 H10:FLOWDIRECT Flowdirect ForwardReverse
ForwardReverse
RW Forward - P.83
8-2 H11:RATELIMIT Ratelimit 0.0 to 10.0Numberofdecimalplaces:1
0.0 to 10.0Numberofdecimalplaces:1
RW 5.0 % P.83
8-3 H12:DEADTIME Deadtime 0.0 to 15.0Numberofdecimalplaces:1
0.0 to 15.0Numberofdecimalplaces:1
RW 0.0 s P.83
8-4 H13:NOISEFILTER Noisefilter ManualLevel1Level2Level3
ManualLevel 1Level 2Level 3
RW Manual - P.83
8-5 H14:PULSINGFLOW Pulsingflow NoYes
NoYes
RW No - P.86
8-6 H20:POWERSYNCH Powersync NoYes
NoYes
RW Yes - P.86
8-7 H21:SETPWRFREQ Setpowerfreq 47.00 to 63.00Numberofdecimalplaces:2
47.00 to 63.00Numberofdecimalplaces:2
RW 50.00 Hz P.86
8-8 H22:IEXPWRFREQ Iexpowerfrequency 0.00 to 99.99Numberofdecimalplaces:2
0.00 to 99.99Numberofdecimalplaces:2
R - Hz P.86
8-9 H23:MESPWRFREQ Measpowerfreq 0.00 to 99.99Numberofdecimalplaces:2
0.00 to 99.99Numberofdecimalplaces:2
R - Hz P.86
8-10 H30:DENSITYSEL Valueselect FixedvalueCorrectionvalue
FixedvalueCorrectionvalue
RW Fixedvalue - P.41
8-11 H31:DENSITYUNIT Unit kg/m3
lb/gallb/cf
kg/m3
lb/gallb/cf
RW kg/m3 - P.41
8-12 H32:FIXEDDENS Fixeddensity 0.00000 to 999999Numberofdecimalplaces:0to5
0.00000 to 999999Numberofdecimalplaces:0 to 5
RW 0 *1 No.8-11 P.41
8-13 H33:STDDENSITY Std density 0.00000to+999999Numberofdecimalplaces:0to5
0.00000to+999999Numberofdecimalplaces:0 to 5
RW 0 No.8-11 P.41
8-14 H34:STDTEMP Stdtemperature -999.99to+999.99Numberofdecimalplaces:2
-999.99to+999.99Numberofdecimalplaces:2
RW 20.00 No.7-15 P.42
8-15 H35:TEMPCOEFA1 CoefA1 -INF*2to+INF*2
Numberofdecimalplaces:0to5-32000.0to+32000.0
-999999to+999999Numberofdecimalplaces:0 to 5-32000.0to+32000.0
RW 0.0 - P.87
8-16 H36:TEMPCOEFA2 CoefA2 -INF*2to+INF*2
Numberofdecimalplaces:0to5-32000.0to+32000.0
-999999to+999999Numberofdecimalplaces:0 to 5-32000.0to+32000.0
RW 0.0 - P.87
8-17 H37:MEASTEMP Meastemperature -999.99to+999.99Numberofdecimalplaces:2
-999.99to+999.99Numberofdecimalplaces:2
R - No.7-15 P.42
8-18 H38:CORRCTDENS Correctdensity 0.0 to 999999Numberofdecimalplaces:0to5
0.0 to 999999Numberofdecimalplaces:0 to 5
R - No.8-11 P.41
8-19 H40:SPECHEAT Specificheat 0.0to+99999.9Numberofdecimalplaces:1
0.0to+99999.9Numberofdecimalplaces:1
RW 4184.0 J/kg*K P.89
8-20 H41:CALFIXTEMP Fixedtemperature -999.99to+999.99Numberofdecimalplaces:2
-999.99to+999.99Numberofdecimalplaces:2
RW 20.00 No.7-15 P.42
8-21 H42:CALVALUE Calorie -INF*2to+INF*2
Numberofdecimalplaces:0to5-999999to+999999Numberofdecimalplaces:0 to 5
R - No.3-18/No.3-19
P.34
8-22 H50:SETSIL Set SIL NoYes
NoYes
RW No - P.90
*1:Determinedbasedontheinformationaboutspecifieditemsatthetimeoforderingandthesensortobecombinedwith.*2:Therangeofsingleprecisionfloat(IEEE754).
<5. Parameter Lists> 175
IM 01E22C02-01EN
(9) AlarmconfigurationparametersTheseparametersareisrelatedtothealarmoutput,burnoutandhistory.
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
9-1 I10:HIGHALARM Highalarm -INF*2to+INF*2
Numberofdecimalplaces:0 to 5
-999999to+999999Numberofdecimalplaces:0to5
RW 300.000 Flowvelocity:No.3-15Volumetricflowrate: No.3-16/No.3-19Massflowrate:No.3-17/No.3-19Flownoisecm/s
P.126
9-2 I11:LOWALARM Lowalarm -INF*2to+INF*2
Numberofdecimalplaces:0 to 5
-999999to+999999Numberofdecimalplaces:0to5
RW -300.000 Flowvelocity:No.3-15Volumetricflowrate: No.3-16/No.3-19Massflowrate:No.3-17/No.3-19Flownoisecm/s
P.126
9-3 I12:HIHIALARM HHalarm -INF*2to+INF*2
Numberofdecimalplaces:0 to 5
-999999to+999999Numberofdecimalplaces:0to5
RW 300.000 Flowvelocity:No.3-15Volumetricflowrate: No.3-16/No.3-19Massflowrate:No.3-17/No.3-19Flownoisecm/s
P.126
9-4 I13:LOLOALARM LLalarm -INF*2to+INF*2
Numberofdecimalplaces:0 to 5
-999999to+999999Numberofdecimalplaces:0to5
RW -300.000 Flowvelocity:No.3-15Volumetricflowrate: No.3-16/No.3-19Massflowrate:No.3-17/No.3-19Flownoisecm/s
P.126
9-5 I14:H/LALMHYS H/Lalarmhyst 0 to 10Numberofdecimalplaces:0
0 to 10Numberofdecimalplaces:0
RW 5 % P.126
9-6 I15:ALMOUTACT Activemode OnactiveOffactive
OnactiveOffactive
RW Onactive - P.109
9-7 I20:4-20BURNOUT 4-20burnout HighLow
HighLow
R - - P.109
*2:Therangeofsingleprecisionfloat(IEEE754).
<5. Parameter Lists> 176
IM 01E22C02-01EN
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
9-7 I30:OUTMSKSET1 Mask1-1 Maskalloff 013:MainEEPdflt RW AllOff - P.103Mask1-2 Maskallon
M/BEEPdfltoffM/BEEPdfltonPWM1stopoffPWM1stoponPWM2stopoffPWM2stoponOptmismatchoffOptmismatchonOptEEPFAILoffOptEEPFAILonOptA/DFAILoffOptA/DFAILonOptSPIFAILoffOptSPIFAILonIndbdFAILoffIndbdFAILonIndEEPFAILoffIndEEPFAILonLCDdrvFAILoffLCDdrvFAILonIndmismatchoffIndmismatchonIndcommERRoffIndcommERRonSDFAILoffSDFAILon
021:PWM1stop022:PWM2stop023:Optbdmismatch024:OptbdEEPFAIL025:OptbdA/DFAIL026:OptbdSPIFAIL028:IndbdFAIL029:IndbdEEPFAIL030:LCDdrvFAIL031:Indbdmismatch032:IndcommERR033:microSDFAIL
RW 032:IndcommERR - P.103
I31:OUTMSKSTS1 - MaskoffMaskonMainEEPdfltonOptSPIFAILonOptA/DFAILonOptEEPFAILonOptmismatchonPWM2stoponPWM1stoponSDFAILonIndcommERRonIndmismatchonLCDdrvFAILonIndEEPFAILonIndbdFAILon
- R - P.103
*2:Therangeofsingleprecisionfloat(IEEE754).
<5. Parameter Lists> 177
IM 01E22C02-01EN
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
9-7 I32:OUTMSKSET2 Mask2-1 MaskalloffMaskallonSigoverflowoffSigoverflowonEmptydetectoffEmptydetectonH/LalmoffH/LalmonAdhoverlv4offAdhoverlv4onSpancfgERRoffSpancfgERRonPVFcfgERRoffPVFcfgERRonAO14-20lmtoffAO14-20lmtonAO24-20lmtoffAO24-20lmtonAO1mltrngoffAO1mltrngonAOH/LcfgoffAOH/LcfgonDenscfgERRoffDenscfgERRon
050:Signaloverflow051:Emptydetect052:H/LHH/LLalm053:Adhoverlv4060:SpancfgERR061:PVFcfgERR062:AO14-20lmt063:AO24-20lmt064:AO1mltrng065:H/LcfgERR066:DensitycfgERR
RW 052:H/LHH/LLalm053:Adhoverlv4
- P.103
Mask2-2 Pls1cfgERRoffPls1cfgERRonPls2cfgERRoffPls2cfgERRonNomsizecfgoffNomsizecfgonAdhcfgERRoffAdhcfgERRonFLNcfgERRoffFLNcfgERRonLognotstrtoffLognotstartonAO1saturateoffAO1saturateonAO2saturateoffAO2saturateonPls1saturatoffPls1saturatonPls2saturatoffPls2saturatonAIsaturateoffAIsaturateonCablemisconoffCablemiscononCoilinsuoffCoilinsuonTrnmismatchoffTrnmismatchon
067:Pls1cfgERR068:Pls2cfgERR069:Nomisizecfg070:AdhcfgERR071:FLNcfgERR072:Lognotstart080:AO1saturate081:AO2saturate082:Pls1saturate083:Pls2saturate084:AIsaturate085:Cablemiscon086:Coilinsulation131:Transmismatch
RW 081:AO2saturate082:Pls1saturate083:Pls2saturate084:AIsaturate086:Coilinsulation
- P.103
I33:OUTMSKSTS2 - MaskoffMaskonAdhoverlv4onH/LHH/LLalmonEmptydetectonSigoverflowonDenscfgERRonAOH/LcfgonAO1mltrngonAO24-20lmtonAO14-20lmtonPVFcfgERRonSpancfgERRonLognotstartonFLNcfgERRonAdhcfgERRonNomisizecfgonPls2cfgERRonPls1cfgERRonTrnmismatchonCoilinsuonCablemiscononAIsaturateonPls2saturteonPls1saturteonAO2saturateonAO1saturateon
- R - P.103
*2:Therangeofsingleprecisionfloat(IEEE754).
<5. Parameter Lists> 178
IM 01E22C02-01EN
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
9-7 I34:OUTMSKSET3 Mask3-1 MaskalloffMaskallonAdhoverlv3offAdhoverlv3onLCwarnoffLCwarnonInsudetectoffInsudetectonFLNoverlv3offFLNoverlv3onFLNoverlv4offFLNoverlv4onAZwarnoffAZwarnonVerifwarnoffVerifwarnonFCNwarnoffFCNwarnonSimactiveoffSimactiveonAO1fixoffAO1fixonAO2fixoffAO2fixonPls1fixoffPls1fixonPls2fixoffPls2fixonAIfixoffAIfixon
087:Adhesionlv3088:LCwarn089:Insudetect090:FLNoverlv3091:FLNoverlv4092:AZwarn093:Verifwarn094:Factnoisewarn095:Simulateactive096:AO1fix097:AO2fix098:Pls1fix099:Pls2fix100:AIfix
RW 087:Adhesionlv3088:LCwarn089:Insudetect090:FLNoverlv3091:FLNoverlv4092:AZwarn093:Verifwarn094:Factnoisewarn
- P.103
Mask3-2 RestorerunoffRestorerunonDispoveroffDisp over onSDsizewngoffSDsizewngonBkupincmpltoffBkupincmpltonSDmismatchoffSDmismatchonSDremovERRoffSDremovERRonWatchdogoffWatchdogonPoweroffoffPoweroffonInstPWFailoffInstPWFailonPrmbkuprunoffPrmbkuprunonDatalogrunoffDatalogrunon
101:Paramrestorerun102:Disp over103:SDsizewarn104:Bkupincmplt105:SDmismatch106:SDremovalERR120:Watchdog121:Poweroff122:InstpowerFAIL123:Parambkuprun124:Datalogrun
101:Paramrestorerun102:Disp over103:SDsizewarn105:SDmismatch106:SDremovalERR120:Watchdog121:Poweroff122:InstpowerFAIL123:Parambkuprun124:Datalogrun
P.103
I35:OUTMSKSTS3 - MaskoffMaskonFCnoisewarnonVerifwarnonAZwarnonFLNoverlv4onFLNoverlv3onInsudetectonLCwarnonAdhoverlv3onAIfixonPls2fixonPls1fixonAO2fixonAO1fixonSimactiveonSDremovERRonSDmismatchonBkupincmpltonSDsizewarnonDisp over onRestorerunonDatalogrunonPrmbkuprunonInstPWFailonPoweroffonWatchdogon
- R - P.103
*2:Therangeofsingleprecisionfloat(IEEE754).
<5. Parameter Lists> 179
IM 01E22C02-01EN
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
9-8 I40:RECMSKSET1 Mask1-1 MaskalloffMaskallonM/BEEPdfltoffM/BEEPdfltonExciterFAILoffExciterFAILon
013:MainEEPdflt RW AllSpace - P.103
Mask1-2 PWM1stopoffPWM1stoponPWM2stopoffPWM2stoponOptmismatchoffOptmismatchonOptEEPFAILoffOptEEPFAILonOptA/DFAILoffOptA/DFAILonOptSPIFAILoffOptSPIFAILonIndbdFAILoffIndbdFAILonIndEEPFAILoffIndEEPFAILonLCDdrvFAILoffLCDdrvFAILonIndmismatchoffIndmismatchonIndcommERRoffIndcommERRonSDFAILoffSDFAILon
021:PWM1stop022:PWM2stop023:Optbdmismatch024:OptbdEEPFAIL025:OptbdA/DFAIL026:OptbdSPIFAIL028:IndbdFAIL029:IndbdEEPFAIL030:LCDdrvFAIL031:Indbdmismatch032:IndcommERR033:microSDFAIL
AllSpace P.103
I41:RECMSKSTS1 - MaskoffMaskonMainEEPdfltonExciterFAILonOptSPIFAILonOptA/DFAILonOptEEPFAILonOptmismatchonPWM2stoponPWM1stoponSDFAILonIndcommERRonIndmismatchonLCDdrvFAILonIndEEPFAILonIndbdFAILon
- R - P.103
I42:RECMSKSET2 Mask2-1 MaskalloffMaskallonSigoverflowoffSigoverflowonEmptydetectoffEmptydetectonH/LalmoffH/LalmonAdhoverlv4offAdhoverlv4on
050:Signaloverflow051:Emptydetect052:H/LHH/LLalm053:Adhoverlv4
RW AllSpace - P.103
Mask2-2 CablemisconoffCablemisconon
085:Cablemiscon AllSpace P.103
I43:RECMSKSTS2 - MaskoffMaskonAdhoverlv4onH/LHH/LLalmonEmptydetectonSigoverflowonCablemisconon
- R - P.103
*2:Therangeofsingleprecisionfloat(IEEE754).
<5. Parameter Lists> 180
IM 01E22C02-01EN
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
9-9 I50:ALMRECORD1 Recordalarm1 AllSpace10:MainboardCPUfailure11:Reversecalculationfailure12:MainboardEEPROMfailure13:MainboardEEPROMdefault14:Sensorboardfailure15:Sensorcommunicationerror16:A/D1failure[Signal]17:A/D2failure[Exciter]18: Coil open19:Coilshort20:Exciterfailure21:PWM1stop22:PWM2stop23:Optionboardmismatch24:OptionboardEEPROMfailure25:OptionboardA/Dfailure26:OptionboardSPIfailure27:Parameterrestoreincomplete28:Indicatorboardfailure29:IndicatorboardEEPROMfailure30:LCDdriverfailure31:Indicatorboardmismatch32:Indicatorcommunicationerror33:microSDfailure50:Signaloverflow51:Emptypipedetection52:H/LorHH/LLalarm53:Adhesionoverlevel485:Cablemisconnect120:Watchdogon121:Poweroffon122:Instantpowerfailureon
AllSpace010:MainCPUFAIL011:RevcalcFAIL012:MainEEPFAIL013:MainEEPdflt014:SnsrbdFAIL015:SnsrcommERR016:AD1FAIL[Sig]017:AD2FAIL[Excit]018:Coil open019:Coilshort020:ExciterFAIL021:PWM1stop022:PWM2stop023:Optbdmismatch024:OptbdEEPFAIL025:OptbdA/DFAIL026:OptbdSPIFAIL027:RestoreFAIL028:IndbdFAIL029:IndbdEEPFAIL030:LCDdrvFAIL031:Indbdmismatch032:IndcommERR033:microSDFAIL050:Signaloverflow051:Emptydetect052:H/LHH/LLalm053:Adhoverlv4085:Cablemiscon120:Watchdog121:Poweroff122:InstpowerFAIL
R AllSpace - P.102
9-10 I51:ALMTIME1 Recordtime1 00000D 00: 00 to 99999D 23: 59
00000D 00: 00 to 99999D 23: 59 R 00000D 00: 00 - P.102
9-11 I52:ALMRECORD2 Recordalarm2 AllSpace10:MainboardCPUfailure11:Reversecalculationfailure12:MainboardEEPROMfailure13:MainboardEEPROMdefault14:Sensorboardfailure15:Sensorcommunicationerror16:A/D1failure[Signal]17:A/D2failure[Exciter]18: Coil open19:Coilshort20:Exciterfailure21:PWM1stop22:PWM2stop23:Optionboardmismatch24:OptionboardEEPROMfailure25:OptionboardA/Dfailure26:OptionboardSPIfailure27:Parameterrestoreincomplete28:Indicatorboardfailure29:IndicatorboardEEPROMfailure30:LCDdriverfailure31:Indicatorboardmismatch32:Indicatorcommunicationerror33:microSDfailure50:Signaloverflow51:Emptypipedetection52:H/LorHH/LLalarm53:Adhesionoverlevel485:Cablemisconnect120:Watchdogon121:Poweroffon122:Instantpowerfailureon
AllSpace010:MainCPUFAIL011:RevcalcFAIL012:MainEEPFAIL013:MainEEPdflt014:SnsrbdFAIL015:SnsrcommERR016:AD1FAIL[Sig]017:AD2FAIL[Excit]018:Coil open019:Coilshort020:ExciterFAIL021:PWM1stop022:PWM2stop023:Optbdmismatch024:OptbdEEPFAIL025:OptbdA/DFAIL026:OptbdSPIFAIL027:RestoreFAIL028:IndbdFAIL029:IndbdEEPFAIL030:LCDdrvFAIL031:Indbdmismatch032:IndcommERR033:microSDFAIL050:Signaloverflow051:Emptydetect052:H/LHH/LLalm053:Adhoverlv4085:Cablemiscon120:Watchdog121:Poweroff122:InstpowerFAIL
R AllSpace - P.102
*2:Therangeofsingleprecisionfloat(IEEE754).
<5. Parameter Lists> 181
IM 01E22C02-01EN
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
9-12 I53:ALMTIME2 Recordtime2 00000D 00: 00 to 99999D 23: 59
00000D 00: 00 to 99999D 23: 59 R 00000D 00: 00 - P.102
9-13 I54:ALMRECORD3 Recordalarm3 AllSpace10:MainboardCPUfailure11:Reversecalculationfailure12:MainboardEEPROMfailure13:MainboardEEPROMdefault14:Sensorboardfailure15:Sensorcommunicationerror16:A/D1failure[Signal]17:A/D2failure[Exciter]18: Coil open19:Coilshort20:Exciterfailure21:PWM1stop22:PWM2stop23:Optionboardmismatch24:OptionboardEEPROMfailure25:OptionboardA/Dfailure26:OptionboardSPIfailure27:Parameterrestoreincomplete28:Indicatorboardfailure29:IndicatorboardEEPROMfailure30:LCDdriverfailure31:Indicatorboardmismatch32:Indicatorcommunicationerror33:microSDfailure50:Signaloverflow51:Emptypipedetection52:H/LorHH/LLalarm53:Adhesionoverlevel485:Cablemisconnect120:Watchdogon121:Poweroffon122:Instantpowerfailureon
AllSpace010:MainCPUFAIL011:RevcalcFAIL012:MainEEPFAIL013:MainEEPdflt014:SnsrbdFAIL015:SnsrcommERR016:AD1FAIL[Sig]017:AD2FAIL[Excit]018:Coil open019:Coilshort020:ExciterFAIL021:PWM1stop022:PWM2stop023:Optbdmismatch024:OptbdEEPFAIL025:OptbdA/DFAIL026:OptbdSPIFAIL027:RestoreFAIL028:IndbdFAIL029:IndbdEEPFAIL030:LCDdrvFAIL031:Indbdmismatch032:IndcommERR033:microSDFAIL050:Signaloverflow051:Emptydetect052:H/LHH/LLalm053:Adhoverlv4085:Cablemiscon120:Watchdog121:Poweroff122:InstpowerFAIL
R AllSpace - P.102
9-14 I55:ALMRECORD3 Recordtime3 00000D 00: 00 to 99999D 23: 59
00000D 00: 00 to 99999D 23: 59 R 00000D 00: 00 - P.102
*2:Therangeofsingleprecisionfloat(IEEE754).
<5. Parameter Lists> 182
IM 01E22C02-01EN
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
9-15 I56:ALMRECORD4 Recordalarm4 AllSpace10:MainboardCPUfailure11:Reversecalculationfailure12:MainboardEEPROMfailure13:MainboardEEPROMdefault14:Sensorboardfailure15:Sensorcommunicationerror16:A/D1failure[Signal]17:A/D2failure[Exciter]18: Coil open19:Coilshort20:Exciterfailure21:PWM1stop22:PWM2stop23:Optionboardmismatch24:OptionboardEEPROMfailure25:OptionboardA/Dfailure26:OptionboardSPIfailure27:Parameterrestoreincomplete28:Indicatorboardfailure29:IndicatorboardEEPROMfailure30:LCDdriverfailure31:Indicatorboardmismatch32:Indicatorcommunicationerror33:microSDfailure50:Signaloverflow51:Emptypipedetection52:H/LorHH/LLalarm53:Adhesionoverlevel485:Cablemisconnect120:Watchdogon121:Poweroffon122:Instantpowerfailureon
AllSpace010:MainCPUFAIL011:RevcalcFAIL012:MainEEPFAIL013:MainEEPdflt014:SnsrbdFAIL015:SnsrcommERR016:AD1FAIL[Sig]017:AD2FAIL[Excit]018:Coil open019:Coilshort020:ExciterFAIL021:PWM1stop022:PWM2stop023:Optbdmismatch024:OptbdEEPFAIL025:OptbdA/DFAIL026:OptbdSPIFAIL027:RestoreFAIL028:IndbdFAIL029:IndbdEEPFAIL030:LCDdrvFAIL031:Indbdmismatch032:IndcommERR033:microSDFAIL050:Signaloverflow051:Emptydetect052:H/LHH/LLalm053:Adhoverlv4085:Cablemiscon120:Watchdog121:Poweroff122:InstpowerFAIL
R AllSpace - P.102
9-16 I57:ALMTIME4 Recordtime4 00000D 00: 00 to 99999D 23: 59
00000D 00: 00 to 99999D 23: 59 R 00000D 00: 00 - P.102
*2:Therangeofsingleprecisionfloat(IEEE754).
<5. Parameter Lists> 183
IM 01E22C02-01EN
(10) DisplayconfigurationparametersTheseparametersarerelatedtodisplaysettings.
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
10-1 J10:LINE1SEL Line 1 Flowrate(%)PVVelocityVolumeflowMassflowFlowrate(%Bar)CalorieTotalizer1Totalizer2Totalizer3TagnumberCommunprotocolAdhesionAnalogout1Analogout2FlownoiselevelTotalizer1countTotalizer2countTotalizer3count
Flowrate(%)PVVelocityVolumeflowMassflowFlowrate(%Bar)CalorieTotalizer1Totalizer2Totalizer3TagnumberCommunprotocolAdhesionAnalogout1Analogout2FlownoiselevelTotalizer1countTotalizer2countTotalizer3count
RW PV - P.111
10-2 J11:LINE2SEL Line 2 NoneFlowrate(%)PVVelocityVolumeflowMassflowFlowrate(%Bar)CalorieTotalizer1Totalizer2Totalizer3TagnumberCommunprotocolAdhesionAnalogout1Analogout2FlownoiselevelTotalizer1countTotalizer2countTotalizer3count
NoneFlowrate(%)PVVelocityVolumeflowMassflowFlowrate(%Bar)CalorieTotalizer1Totalizer2Totalizer3TagnumberCommunprotocolAdhesionAnalogout1Analogout2FlownoiselevelTotalizer1countTotalizer2countTotalizer3count
RW Flowrate(%) - P.11110-3 J12:LINE3SEL Line 3 RW Analogout1 - P.11110-4 J13:LINE4SEL Line 4 RW None - P.11110-5 J14:LINE5SEL Line 5 RW None - P.11110-6 J15:LINE6SEL Line 6 RW None - P.11110-7 J16:LINE7SEL Line 7 RW None - P.11110-8 J17:LINE8SEL Line 8 RW None - P.111
10-9 J20:FORMATFR FormatPV Auto0digit1digit2digit3digit4digit5digitAuto2
Auto0digit1digit2digit3digit4digit5digitAuto2
RW Auto2 - P.112
10-10 J21:FORMATTOT1 Formattotal1 Auto0digit1digit2digit3digit4digit5digit6digit7digit
Auto0digit1digit2digit3digit4digit5digit6digit7digit
RW Auto - P.11210-11 J22:FORMATTOT2 Formattotal2 RW10-12 J23:FORMATTOT3 Formattotal3 RW
10-13 J24:DISPCONTR Contrast -5-4-3-2-10+1+2+3+4+5
-5-4-3-2-1012345
RW 0 - P.118
10-14 J25:DISPLINE Linemode 1Line(Big)1 Line2 Line3 Line4 Line
1line(big)1 line2 line3 line4 line
RW 3 line - P.114
*1:Determinedbasedontheinformationaboutspecifieditemsatthetimeoforderingandthesensortobecombinedwith.*2:Therangeofsingleprecisionfloat(IEEE754).
<5. Parameter Lists> 184
IM 01E22C02-01EN
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
10-15 J26:DISPPERIOD Period 0.2 s0.4 s1.0 s2.0 s4.0 s8.0 s
0.2 s0.4 s1.0 s2.0 s4.0 s8.0 s
RW 0.4 s - P.117
10-16 J27:DISPNE107 NE107 display NormalNE107
NormalNE107
RW Normal - P.101
10-17 J28:DISPALARM Alarmdisplay NormalDetail
NormalDetail
RW Normal - P.101
10-18 J29:DISPSCROLL Scrollmode OffManualAuto(2s)Auto(4s)Auto(8s)
OffManualAuto(2s)Auto(4s)Auto(8s)
RW Off*3 - P.114
10-19 J30:DISPDAMPING Damp 0.0 to 200.0Numberofdecimalplaces:1
0.0 to 200.0Numberofdecimalplaces:1
RW 0.0 s P.118
10-20 J31:FORMATDATE Formatdate MM/DD/YYYYDD/MM/YYYYYYYY/MM/DD
MM/DD/YYYYDD/MM/YYYYYYYY/MM/DD
RW MM/DD/YYYY - P.118
10-21 J32:LANGUAGE Language EnglishFrenchGermanItalianSpanishPortugueseRussianChineseJapanese
EnglishFrenchGermanItalianSpanishPortugueseRussianChineseJapanese
RW English - P.110
10-22 J40:DISPMODE Displaymode NormalTrend
NormalTrend
RW Normal - P.116
10-23 J41:OFFTRENDLRV TrendofflnLRV -INF*2to+INF*2
Numberofdecimalplaces:0 to 5
-999999to+999999Numberofdecimalplaces:0to5
RW 0.0 - P.117
10-24 J42:OFFTRENDURV TrendofflnURV -INF*2to+INF*2
Numberofdecimalplaces:0 to 5
-999999to+999999Numberofdecimalplaces:0to5
RW 10.0 - P.117
10-25 J43:TREND1SEL Trend1 Flowrate(%)PVVelocityVolumeflowMassflowCalorieAnalogout1Analogout2Totalizer1Totalizer2Totalizer3
Flowrate(%)PVVelocityVolumeflowMassflowCalorieAnalogout1Analogout2Totalizer1Totalizer2Totalizer3
RW PV - P.116
10-26 J44:TREND2SEL Trend2 NoneFlowrate(%)PVVelocityVolumeflowMassflowCalorieAnalogout1Analogout2Totalizer1Totalizer2Totalizer3
NoneFlowrate(%)PVVelocityVolumeflowMassflowCalorieAnalogout1Analogout2Totalizer1Totalizer2Totalizer3
RW None - P.11610-27 J45:TREND3SEL Trend3 RW None - P.11610-28 J46:TREND4SEL Trend4 RW None - P.116
10-29 J47:DISPINVERSE Inversion NormalInvert
NormalInvert
RW Normal - P.118
10-30 J50:LCDTEST LCD test NotexecuteExecuteShowPattern1ShowPattern2ShowPattern3ShowPattern4
NotexecuteExecuteShowpattern1Showpattern2Showpattern3Showpattern4
RW Notexecute - -
10-31 J51:DISPSQUAWK: Squawk OffOnSquawkOnce
OffOnSquawkonce
RW Off - P.119
10-32 J52:LANGPACKAGE Languagepackage Package1Package2
Pack1Pack2
R - *1 - P.110
10-33 J53:DISPINSTALL Disp install No dispWithdisp
No dispWithdisp
RW Withdisp*1 - -
10-34 J54:IRSWOPERATE - DisableEnable
- RW Enable - P.119
*1:Determinedbasedontheinformationaboutspecifieditemsatthetimeoforderingandthesensortobecombinedwith.*2:Therangeofsingleprecisionfloat(IEEE754).*3:WhenMainsoftrev(K50:MAINBREV/Mainsoftrev)isR2.01.02orearlier,thedefaultvalueisManual.
<5. Parameter Lists> 185
IM 01E22C02-01EN
(11) DeviceinformationconfigurationparametersTheseparametersarerelatedtothedeviceinformation.
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
11-1 K01:TAGNO TagNo ASCII16characters ASCII16characters RW AllSpace*1 - P.12211-2 K02:OPERATETIME Operationtime 00000D 00: 00 to 99999D 23: 59 00000D 00: 00 to 99999D 23: 59 R 00000D 00: 00 - P.12411-3 K03:CURDATE Currentdate 01/01/1900 to 12/31/2155 01/01/1900 to 12/31/2155 R 2019/01/01 - P.12411-4 K04:CURTIME Currenttime 00: 00: 00 to 23: 59: 59 00: 00: 00 to 23: 59: 59 R 00: 00: 00 - P.12411-5 - Set date - 01/01/1900 to 12/31/2155 RW 01/01/2019 - P.124
- Settime - 00: 00: 00 to 23: 59: 59 RW 00: 00: 00 - P.124K05:SETCURDAY - 1 to 31 - RW 1 - P.124K06:SETCURMONTH
- 1 to 12 - RW 1 - P.124
K07:SETCURYEAR - 1900 to 2155 - RW 2019 - P.124K08:SETCURHR - 0 to 23 - RW 0 - P.124K09:SETCURMIN - 0 to 59 - RW 0 - P.124K10:SETCURSEC - 0 to 59 - RW 0 - P.124
11-6 K11:TRNSTYPE Transmittertype Non1AType
None1Atype
R - - P.123
11-7 K12:OPTBOARDID OptionboardID NonMulti
NoneMulti
R - - P.123
11-8 K15:EL SIZE Electrodesize 1mm3mm8mm10mm6mm
1mm3mm8mm10mm6mm
RW 3mm - P.137
11-9 K17:EXPROTECT Explosionprotection No/Yes No/Yes RW No - P.12511-10 K20:MODEL CODE Modelcode ASCII16characters ASCII16characters RW AllSpace*1 - P.12211-11 K21:SUFFIXCONF1 Suffixconfig1 ASCII16characters ASCII16characters RW AllSpace*1 - P.122
K22:SUFFIXCONF2 Suffixconfig2 ASCII16characters ASCII16characters RW AllSpace*1 - P.12211-12 K23:OPTION1 Option 1 ASCII16characters ASCII16characters RW AllSpace*1 - P.122
K24:OPTION2 Option 2 ASCII16characters ASCII16characters RW AllSpace*1 - P.122K25:OPTION3 Option 3 ASCII16characters ASCII16characters RW AllSpace*1 - P.122K26:OPTION4 Option 4 ASCII16characters ASCII16characters RW AllSpace*1 - P.122
11-13 K27:RSMDLCD Modelcode ASCII16characters ASCII16characters RW AllSpace*1 - P.12211-14 K28:RSSUFCONF1 Suffixconfig1 ASCII16characters ASCII16characters RW AllSpace*1 - P.122
K29:RSSUFCONF2 Suffixconfig2 ASCII16characters ASCII16characters RW AllSpace*1 - P.12211-15 K30:RSOPT1 Option 1 ASCII16characters ASCII16characters RW AllSpace*1 - P.122
K31:RSOPT2 Option 2 ASCII16characters ASCII16characters RW AllSpace*1 - P.122K32:RSOPT3 Option 3 ASCII16characters ASCII16characters RW AllSpace*1 - P.122K33:RSOPT4 Option 4 ASCII16characters ASCII16characters RW AllSpace*1 - P.122
11-16 K34:TRNSSRNO TransserialNo ASCII16characters ASCII16characters RW AllSpace - P.12211-17 K35:FTSRNO Sensor serial No ASCII16characters ASCII16characters RW AllSpace - P.12211-18 K40:MEMO1 Memo1 ASCII16characters ASCII16characters RW AllSpace - P.123
K41:MEMO2 Memo2 ASCII16characters ASCII16characters RW AllSpace - P.123K42:MEMO3 Memo3 ASCII16characters ASCII16characters RW AllSpace - P.123
11-19 K50:MAINBREV Mainsoftrev R2.01.01*4 R2.01.01*4 R - - P.12311-20 K51:SENSORBREV Snsrsoftrev R1.01.01*4 R1.01.01*4 R - - P.12311-21 K52:INDBREV Indsoftrev R2.01.01*4 R2.01.01*4 R - - P.12311-22 K57:MIO MultiI/O NAN
AO2ActAIActPLS/SO2PasPLS/SO2Act1PLS/SO2Act2/3
NoneAO2ActAIActPLS/SO2PasPLS/SO2Act1PLS/SO2Act2/3
R None - -
*1:Determinedbasedontheinformationaboutspecifieditemsatthetimeoforderingandthesensortobecombinedwith.*4:Setatthefactorybeforeshipment.
<5. Parameter Lists> 186
IM 01E22C02-01EN
(12) DiagnosisfunctionconfigurationparametersTheseparametersarerelatedtotheelectrodeadhesiondetection,flownoisediagnosis,andverificationfunctionsettings.
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
12-1 L10:ADHESIONCHK Function DisableEnable
DisableEnable
RW Enable - P.128
12-2 L11:ADHLEVEL1 Thresholdlevel1 0.00to+100.00Numberofdecimalplaces:2
0.00to+100.00Numberofdecimalplaces:2
RW 0.10 Mohm P.128
12-3 L12:ADHLEVEL2 Thresholdlevel2 0.00to+100.00Numberofdecimalplaces:2
0.00to+100.00Numberofdecimalplaces:2
RW 0.50 Mohm P.128
12-4 L13:ADHLEVEL3 Thresholdlevel3 0.00to+100.00Numberofdecimalplaces:2
0.00to+100.00Numberofdecimalplaces:2
RW 4.00 Mohm P.128
12-5 L14:ADHLEVEL4 Thresholdlevel4 0.00to+100.00Numberofdecimalplaces:2
0.00to+100.00Numberofdecimalplaces:2
RW 12.00 Mohm P.128
12-6 L16:ADHMEASVAL Value 0.00000to+1000.0Numberofdecimalplaces:0to5
0.00000to+1000.0Numberofdecimalplaces:0to5
R - Mohm P.128
12-7 L17:ADHSTATUS Status Level0Level1Level2Level3Level4
Level 0Level 1Level 2Level 3Level 4
R - - P.128
12-8 L18:ADHCHKCYC Checkcycle 0.5min1min2min10min
0.5min1min2min10min
RW 2min - P.128
12-9 L20:FLNOISECHK Function DisableEnable
DisableEnable
RW Disable - P.136
12-10 L21:FLNOISELV1 Thresholdlevel1 0.0to+999.9Numberofdecimalplaces:1
0.0to+999.9Numberofdecimalplaces:1
RW 5.0 cm/s P.136
12-11 L22:FLNOISELV2 Thresholdlevel2 0.0to+999.9Numberofdecimalplaces:1
0.0to+999.9Numberofdecimalplaces:1
RW 10.0 cm/s P.136
12-12 L23:FLNOISELV3 Thresholdlevel3 0.0to+999.9Numberofdecimalplaces:1
0.0to+999.9Numberofdecimalplaces:1
RW 30.0 cm/s P.136
12-13 L24:FLNOISELV4 Thresholdlevel4 0.0to+999.9Numberofdecimalplaces:1
0.0to+999.9Numberofdecimalplaces:1
RW 400.0 cm/s P.136
12-14 L25:FLNOISEVAL Value 0.00000to+100.00Numberofdecimalplaces:0to5
0.00000to+100.00Numberofdecimalplaces:0to5
R - cm/s P.136
12-15 L26:FLNOISESTS Status Level0Level1Level2Level3Level4
Level 0Level 1Level 2Level 3Level 4
R - - P.136
12-16 L27:FLNOISEDMP Damp 0.1 to 500.0Numberofdecimalplaces:1
0.1 to 500.0Numberofdecimalplaces:1
RW 3.0 s P.37
12-17 L28:FLNOISESPN Span 0.1 to 999.9Numberofdecimalplaces:1
0.1 to 999.9Numberofdecimalplaces:1
RW 150.0 cm/s P.35
12-18 L30:LOCONDFUNC Function DisableEnable
DisableEnable
RW Disable - P.137
12-19 L31:CONDUCTIVITY Value 0.00000to+INF*2
Numberofdecimalplaces:0to50.00000 to 999999Numberofdecimalplaces:0to5
R - mS/cm P.137
12-20 L32:CONDUCTLMT Lowlimit 0.0000 to 99.9999Numberofdecimalplaces:4
0.0000 to 99.9999Numberofdecimalplaces:4
RW 0.0010 mS/cm P.137
12-21 L33:DIAGEXE Diagnosticexecute NotexecuteElecinsexeConnchkexe
NotexecuteElectrodeinsulexeConnectcheckexe
RW Notexecute - P.130P.135
12-22 L37:COILINSTH Coilinsulthreshold 0.0 to 100.0Numberofdecimalplaces:1
0.0 to 100.0Numberofdecimalplaces:1
RW 25.0 % P.138
12-23 L38:IEXCOMP IEXcompare 0.0 to 999.9Numberofdecimalplaces:1
0.0 to 999.9Numberofdecimalplaces:1
R 260.0 mA P.138
12-24 L40:VPEAKHOLD Vpeakhold 0.000 to 3.000Numberofdecimalplaces:3
0.000 to 3.000Numberofdecimalplaces:3
R - V P.138
12-25 L41:IEXCOILR IEXresistance 0to+9999.9Numberofdecimalplaces:1
0to+9999.9Numberofdecimalplaces:1
R - ohm P.138
12-26 L42:ELECVOLA ElectrodevoltageA -3.00 to 3.00Numberofdecimalplaces:2
-3.00 to 3.00Numberofdecimalplaces:2
R - V P.129
12-27 L43:ELECVOLB ElectrodevoltageB -3.00to+3.00Numberofdecimalplaces:2
-3.00to+3.00Numberofdecimalplaces:2
R - V P.129
12-28 L44:EMPTYSTS Emptystatus FullEmpty
FullEmpty
R - - P.129
*2:Therangeofsingleprecisionfloat(IEEE754).
<5. Parameter Lists> 187
IM 01E22C02-01EN
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
12-29 L47:VFTGTSET VFtargetselect AllonAlloffMagneticonMagneticoffExcitationonExcitationoffCalculationonCalculationoffDevstatusonDevstatusoffConnstatusonConnstatusoff
MagneticcircuitExcitecircuitCalccircuitDevicestatusConnectstatus
RW Allon - P.131
12-30 L48:VFTGTSTS - MagneticExcitationCalculationDevicestatusConnstatus
- R - P.131
12-31 L49:DIAGOUTPUT Diagnosticoutput ZeroMeasuredvalueHold
ZeroMeasuredvalueHold
RW Zero - P.130P.131P.135
12-32 L50:VFMODE Mode NoflowFlow
NoflowFlow
RW Noflow - P.131
12-33 L51:VFEXE Execute NotexecuteExecute
NotexecuteExecute
RW Notexecute - P.131
12-34 L52:VFNO VFNo FactoryPreviousPresent
FactoryPreviousPresent
RW Factory - P.131
12-35 L53:VFCHKRES Failed/Passed PassedFailedCanceledNo DataUnknown
PassedFailedCanceledNo dataUnknown
R No data - P.131
12-36 L54:VFOPETIME VFoperatetime 00000D 00: 00 to 99999D 23: 59 00000D 00: 00 to 99999D 23: 59 R 00000D 00: 00 - P.13112-37 L55:VFMAGRES Magneticcircuit Passed
FailedCanceledNo DataUnknownSkip
PassedFailedCanceledNo dataUnknownSkip
R No data - P.131
12-38 L56:VFEXCITRES Excitecircuit PassedFailedCanceledNo DataUnknownSkip
PassedFailedCanceledNo dataUnknownSkip
R No data - P.131
12-39 L57:VFCALCRES Calccircuit PassedFailedCanceledNo DataUnknownSkip
PassedFailedCanceledNo dataUnknownSkip
R No data - P.131
12-40 L58:VFDEVRES Devicestatus PassedFailedCanceledNo DataUnknownSkip
PassedFailedCanceledNo dataUnknownSkip
R No data - P.131
12-41 L59:VFCONNRES Connectstatus PassedFailedCanceledNo DataUnknownSkip
PassedFailedCanceledNo dataUnknownSkip
R No data - P.131
*2:Therangeofsingleprecisionfloat(IEEE754).
<5. Parameter Lists> 188
IM 01E22C02-01EN
(13) TestmodeconfigurationparametersTheseparametersarerelatedtotestmodesettings.
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
13-1 M01:TESTRELTIM Releasetime 10min30min1h3h12h
10min30min1h3h12h
RW 30min - P.141
13-2 M02:TESTMODE - TestalloffVelotestonVolFtestonMassFtestonCalorie test onAO1testonP1testonSO1 test onAO2testonP2testonSO2 test onAItestonSO11 test onSI11 test onSO12 test onSI12 test onTestallonAlmteston
- RW Testalloff - P.139
M03:TESTSTATUS - TestoffTestonCalorie test onMassFtestonVolFtestonVelotestonAItestonSO2 test onP2testonAO2testonSO1 test onP1testonAO1testonAlmtestonSI12 test onSO12 test onSI11 test onSO11 test on
- R - - P.139
- Inputtest►Testmode - VelocitytestVolumetestMass testCalorie testAItestSI11 testSI12 test
RW - - P.139
- Outputtest►Testmode
- AO1testPO1testSO1 testAO2testPO2testSO2 testSO11 testSO12 testAlarmtest
RW - - P.139
13-3 M10:VELOTVAL Velocity -INF*2to+INF*2
Numberofdecimalplaces:0 to 5
-999999to+999999Numberofdecimalplaces:0 to 5
RW 0.0 No.3-15 P.140
13-3 M11:VOLUMTVAL Volume -INF*2to+INF*2
Numberofdecimalplaces:0 to 5
-999999to+999999Numberofdecimalplaces:0 to 5
RW 0.0 No.3-16/ No.3-19
P.140
13-4 M12:MASSTVAL Mass -INF*2to+INF*2
Numberofdecimalplaces:0 to 5
-999999to+999999Numberofdecimalplaces:0 to 5
RW 0.0 No.3-17/ No.3-19
P.140
13-5 M13:CALOTVAL Calorie -INF*2to+INF*2
Numberofdecimalplaces:0 to 5
-999999to+999999Numberofdecimalplaces:0 to 5
RW 0.0 No.3-18/ No.3-19
P.140
13-6 M41:SI11TVAL SI11 OpenClose
OpenClose
RW Open - P.140
13-7 M43:SI12TVAL SI12 OpenClose
OpenClose
RW Open - P.140
13-8 M14:AO1TESTVAL AO1 2.400 to 21.600Numberofdecimalplaces:3
2.400 to 21.600Numberofdecimalplaces:3
RW 4.000 mA P.140
*2:Therangeofsingleprecisionfloat(IEEE754).
<5. Parameter Lists> 189
IM 01E22C02-01EN
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
13-9 M15:P1TESTVAL PO1 0 to 12500Numberofdecimalplaces:0
0 to 12500Numberofdecimalplaces:0
RW 0.0 pps(pulse/s) P.140
13-10 M16:SO1TESTVAL SO1 OpenClose
OpenClose
RW Open - P.140
13-11 M20:AO2TESTVAL AO2 2.400 to 21.600Numberofdecimalplaces:3
2.400 to 21.600Numberofdecimalplaces:3
RW 4.000 mA P.140
13-12 M21:P2TESTVAL PO2 0 to 12500Numberofdecimalplaces:0
0 to 12500Numberofdecimalplaces:0
RW 0 pps(pulse/s) P.140
13-13 M22:SO2TESTVAL SO2 OpenClose
OpenClose
RW Open - P.140
13-14 M40:SO11TVAL SO11 OpenClose
OpenClose
RW Open - P.140
13-15 M42:SO12TVAL SO12 OpenClose
OpenClose
RW Open - P.140
13-16 M30:AITESTVAL AI 0.000 to 21.600Numberofdecimalplaces:3
0.000 to 21.600Numberofdecimalplaces:3
RW 0.000 mA P.140
13-17 M44:ALMTESTVAL Alarm OpenClose
OpenClose
RW Open - P.140
13-18 M50:TEST2MODE Test2mode NormalTest
NormalTest
RW Normal - P.141
13-19 M51:TEST2OUT Test2value -10.0 to 110.0 -10.0 to 110.0 RW 0.0 % P.141
*2:Therangeofsingleprecisionfloat(IEEE754).
(14) Backup/restore/dataloggingconfigurationparametersTheseparametersarerelatedtothebackupfunction,restorefunctionanddataloggingfunctionofparameters.
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
14-1 N10:FBCKUPNAME Fbackupname ASCII16characters ASCII16characters R FactoryDelivery
- P.142
14-2 N11:FBCKUPDATE Fbackupdate ASCII16characters ASCII16characters R 01/01/2019*4
- P.142
14-3 N12:SDBCKNAME SDbackupname ASCII8characters ASCII8characters RW SD_FILE - P.14214-4 N13:BCKNAME1 Backupname1 ASCII16characters ASCII16characters RW Backup1 - P.14214-5 N14:BCKDATE1 Backupdate1 ASCII16characters ASCII16characters RW 01/01/2019 - P.14214-6 N15:BCKNAME2 Backupname2 ASCII16characters ASCII16characters RW Backup2 - P.14214-7 N16:BCKDATE2 Backupdate2 ASCII16characters ASCII16characters RW 01/01/2019 - P.14214-8 N17:BCKNAME3 Backupname3 ASCII16characters ASCII16characters RW Backup3 - P.14214-9 N18:BCKDATE3 Backupdate3 ASCII16characters ASCII16characters RW 01/01/2019 - P.14214-10 N19:BACKUPEXEC Backupexecute Notexecute
Store Main to 1Store Main to 2Store Main to 3Store Main to SDStoreEEP1toSDStoreEEP2toSDStoreEEP3toSD
NotexecuteStoremainto1Storemainto2Storemainto3StoremaintoSDStoreEEP1toSDStoreEEP2toSDStoreEEP3toSD
RW Notexecute - P.142
14-11 N20:RESTOREEXEC Restoreexecute NotexecuteDuplicateData1DuplicateData2DuplicateData3DuplicateSDRestoreData1RestoreData2RestoreData3RestoreSDCompulsionData1CompulsionData2CompulsionData3CompulsionSDRestoreFactory
NotexecuteDuplicatedata1Duplicatedata2Duplicatedata3DuplicateSDRestoredata1Restoredata2Restoredata3RestoreSDCompulsiondata1Compulsiondata2Compulsiondata3CompulsionSDRestorefactory
RW Notexecute - P.145
14-12 N21:RESTORERSLT Restoreresult UnexecutedSuccessFailureRunning
UnexecutedSuccessFailureRunning
R Unexecuted - P.145
14-13 N22:BACKUPRSLT Backupresult UnexecutedSuccessFailureRunning
UnexecutedSuccessFailureRunning
R Unexecuted - P.142
14-14 N30:LOGGINGFILE Filename ASCII8characters ASCII8characters RW LOG_FILE - P.153
*4:Setatthefactorybeforeshipment.
<5. Parameter Lists> 190
IM 01E22C02-01EN
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
14-15 N31:LOGINTRTIM Intervaltime 1 s10 s30 s1min5min30min1h
1 s10 s30 s1min5min30min1h
RW 1min - P.153
14-16 N32:LSTARTDATE Start date 01/01/1900 to 31/12/2155 01/01/1900 to 31/12/2155 R 01/01/2019 - P.15314-17 N33:LSTARTTIME Starttime 00: 00: 00 to 23: 59: 59 00: 00: 00 to 23: 59: 59 R 00:00:00 - P.15314-18 N34:LOGENDTIME Endtime 10min
30min1h3h12h24h72h240h720h1440h
10min30min1h3h12h24h72h240h720h1440h
RW 12h - P.153
14-19 N35:LOG1SELECT Log1 VelocityVolumeflowMassflowCaloriePVFlownoiseAdhesionElectrodeAElectrodeBVpeak
VelocityVolumeflowMassflowCaloriePVFlownoiseAdhesionElectrodeAElectrodeBVpeak
RW PV - P.15314-20 N36:LOG2SELECT Log2 RW Velocity - P.15314-21 N37:LOG3SELECT Log3 RW Volumeflow - P.15314-22 N38:LOG4SELECT Log4 RW Massflow - P.153
14-23 N39:LOGGINGEXEC Execute NotexecuteExecute
NotexecuteExecute
RW Notexecute - P.153
*4:Setatthefactorybeforeshipment.
(15) Parameter protection parametersTheseparametersarerelatedtothewriteprotectfunction.
No.Parameter Name Data Range
R/W Default Value Unit ReferenceBRAIN Display BRAIN Display
15-1 P10:KEYCODE Keycode 0000 to 9999Numberofdecimalplaces:0
0000 to 9999Numberofdecimalplaces:0
RW 0000 - P.156
15-2 P20:WRTPROTECT Writeprotectsts NoYes
NoYes
R No - P.155
15-3 P21:ENABLEWRITE Enablewrite ASCII8characters ASCII8characters RW AllSpace - P.15515-4 P22:NEW
PASSWORDNewpassword ASCII8characters ASCII8characters RW AllSpace - P.155
15-5 P23:SOFTSEAL Softsealstatus KeepBreak
KeepBreak
R Keep - P.155
<6. Parameter Menu Tree> 191
IM 01E22C02-01EN
6. Parameter Menu TreeNOTETheavailablemenusandparametersvarydependingontheconnectionterminaltypeandtheoptionalcodesspecifiedatthetimeofordering.
6.1 Display Menu TreeTheoutlineofthedisplaymenustructureisshownbelow.
Operation levelExit R ReadonlyOperator R/W1 ReadandwritepossibleMaintenance R/W2 Readalwayspossible,
Writinglimitedtothefollowingoperationlevels (Operationlevel:Maintenance,Specialist)Specialist
R/W3 Readalwayspossible, Writinglimitedtothefollowingoperationlevels (Operationlevel:Specialist)
Device setupCurrentalarm →Page192 Easysetupwizard →Page199
VolumeLanguage R/W1 Pulse/Statusout
Display setProcessvariables →Page193 Autozeroexe
Flowrate(%)Flowrate Detailedsetup →Page200Velocity ProvarVolume SensorMass UserSpanCalorie TotalizerAnalogout1 Pulse/StatusoutAnalogout2 StatusinTotalizer Analogout/in
MultirangeDiag/Service →Page194 AUXcalculation
Sts/Selftest Display setTimestamp AccesscfgDiagnosis DeviceinfoVerification ProtectionAutozeroAO/AItrim microSD →Page207H/Lalarmcfg ContentsTest UnmountParambkup/restore FormatDatalog PropertyDispindicator
<6. Parameter Menu Tree> 192
IM 01E22C02-01EN
�� Current alarmCurrent alarm
SettinguploadStatus0 RStatus1 RStatus2 RStatus3 RStatus4 RStatus5 RStatus14 RStatus15 RStatus16 RStatus17 RStatus18 RStatus19 RStatus20 RSettingdownload
Status0 Status1 Status2 Status3010:MainCPUFAIL 014:SnsrbdFAIL 021:PWM1stop 027:RestoreFAIL011:RevcalcFAIL 015:SnsrcommERR 022:PWM2stop 028:IndbdFAIL012:MainEEPFAIL 016:AD1FAIL[Sig] 023:Optbdmismatch 029:IndbdEEPFAIL013:MainEEPdflt 017:AD2FAIL[Excit] 024:OptbdEEPFAIL 030:LCDdrvFAIL
018:Coil open 025:OptbdA/DFAIL 031:Indbdmismatch019:Coilshort 026:OptbdSPIFAIL 032:IndcommERR020:ExciterFAIL 033:microSDFAIL
Status4 Status5 Status14 Status15050:Signaloverflow 060:SpancfgERR 067:Pls1cfgERR 080:AO1saturate051:Emptydetect 061:PVFcfgERR 068:Pls2cfgERR 081:AO2saturate052:H/LHH/LLalm 062:AO14-20lmt 069:Nomisizecfg 082:Pls1saturate053:Adhoverlv4 063:AO24-20lmt 070:AdhcfgERR 083:Pls2saturate
064:AO1mltrng 071:FLNcfgERR 084:AIsaturate065:H/LcfgERR 072:Lognotstart 085:Cablemiscon066:DensitycfgERR 086:Coilinsulation
131:Transmismatch
Status16 Status17 Status18 Status19087:Adhesionlv3 095:Simulateactive 101:Paramrestorerun 120:Watchdog088:LCwarn 096:AO1fix 102:Disp over 121:Poweroff089:Insudetect 097:AO2fix 103:SDsizewarn 122:InstpowerFAIL090:FLNoverlv3 098:Pls1fix 104:Bkupincmplt 123:Parambkuprun091:FLNoverlv4 099:Pls2fix 105:SDmismatch 124:Datalogrun092:AZwarn 100:AIfix 106:SDremovalERR093:Verifwarn094:Factnoisewarn
NOTEOnlythestatusitemsthatoccurintheBRAINcommunicationtypearedescribedabove.Otherthantheabovestatusitemsarealsodisplayedonthedisplayunit.
<6. Parameter Menu Tree> 193
IM 01E22C02-01EN
�� Process variablesProcess variables
Flowrate(%) RFlowrate RVelocity RVolume RMass RCalorie RAnalogout1 RAnalogout2 RTotalizer
Totalizer1 RTotalizer2 RTotalizer3 RTotalizer1count RTotalizer2count RTotalizer3count R
<6. Parameter Menu Tree> 194
IM 01E22C02-01EN
�� Diagnosis/ServiceDiag/Service
Sts/Selftest →Page195
TimestampDate RTime ROptime R
Diagnosis →Page196
Verification →Page197
AutozeroExecute R/W2Result
Zerovalue R/W3
AO/AItrim →Page197
H/LalarmcfgHighalarm R/W3Lowalarm R/W3HHalarm R/W3LLalarm R/W3H/Lalarmhyst R/W34-20burnout R
Test →Page198
Parambkup/restoreFbackupname RFbackupdate RSDbackupname R/W3Backupexecute R/W3Backupresult RRestoreexecute R/W3Restoreresult RBackupname1 R/W3Backupdate1 R/W3Backupname2 R/W3Backupdate2 R/W3Backupname3 R/W3Backupdate3 R/W3
DatalogFilename R/W3Intervaltime R/W3Start date RStarttime REndtime R/W3Execute R/W3Log1 R/W3Log2 R/W3Log3 R/W3Log4 R/W3
DispindicatorLCD test R/W1Squawk R/W1
<6. Parameter Menu Tree> 195
IM 01E22C02-01EN
�z Status/Self testSts/Self test
CurrentalarmStatus0 RStatus1 RStatus2 RStatus3 RStatus4 RStatus5 RStatus14 R
Status15 RStatus16 RStatus17 RStatus18 RStatus19 RStatus20 R
AlarmAlarmrecordmask
Mask1-1 R/W3Mask1-2 R/W3Mask2-1 R/W3Mask2-2 R/W3
AlarmrecordRecordalarm1 RRecordtime1 RRecordalarm2 RRecordtime2 RRecordalarm3 RRecordtime3 RRecordalarm4 RRecordtime4 R
AlarmoutmaskMask1-1 R/W3Mask1-2 R/W3Mask2-1 R/W3Mask2-2 R/W3Mask3-1 R/W3Mask3-2 R/W3
AlarmoutputActivemode R/W3
<6. Parameter Menu Tree> 196
IM 01E22C02-01EN
�z DiagnosisDiagnosis
AdhesionFunction R/W3Thresholdlevel1 R/W3Thresholdlevel2 R/W3Thresholdlevel3 R/W3Thresholdlevel4 R/W3Result
Value RStatus R
Checkcycle R/W3
FlownoiseFunction R/W3Thresholdlevel1 R/W3Thresholdlevel2 R/W3Thresholdlevel3 R/W3Thresholdlevel4 R/W3Result
Value RStatus R
Damp R/W3Span R/W3
ConductivityFunction R/W3Lowlimit R/W3Result
Value R
Diagnosticexecute R/W3Coilinsulthreshold R/W3IEXcompare RDiagnosticoutput R/W3Vpeakhold RIEXresistance REmptycheck
ElectrodevoltageA RElectrodevoltageB REmptystatus R
<6. Parameter Menu Tree> 197
IM 01E22C02-01EN
�z VerificationVerification
Mode R/W3Execute R/W3VFNo R/W3VFtargetselect R/W3Result
Failed/Passed RVFoperatetime RMagneticcircuit RExcitecircuit RCalccircuit RDevicestatus RConnectstatus R
�z Analog output/input adjustmentAO/AI trim
AOtrimAO1trimclear R/W3AO1trim4mA R/W3AO1trim20mA R/W3AO2trimclear R/W3AO2trim4mA R/W3AO2trim20mA R/W3
AItrimTrimclear R/W3Trim4mA R/W3Trim20mA R/W3
<6. Parameter Menu Tree> 198
IM 01E22C02-01EN
�z TestTest
Releasetime R/W3Inputtest
Testmode R/W3Velocity R/W3Volume R/W3Mass R/W3Calorie R/W3AI R/W3SI11 R/W3SI12 R/W3
OutputtestTestmode R/W3AO1 R/W3PO1 R/W3SO1 R/W3AO2 R/W3PO2 R/W3SO2 R/W3SO11 R/W3SO12 R/W3Alarm R/W3
Test2mode R/W3Test2value R/W3
<6. Parameter Menu Tree> 199
IM 01E22C02-01EN
�� Easy setup wizardEasy setup wizard
VolumeSettinguploadDampAO/F R/W2Damppls/ttl R/W2Unit R/W2Timeunit R/W2Span R/W2Settingdownload
Pulse/StatusoutSettinguploadP1unit R/W2P1val R/W2F1at0% R/W2F1at100% R/W2Settingdownload
Display setSettinguploadLine 1 R/W1Line 2 R/W1Line 3 R/W1Settingdownload
Autozeroexe R/W2
<6. Parameter Menu Tree> 200
IM 01E22C02-01EN
�� Detailed setupDetailed setup
Provar →Page201
SensorLowMF R/W3HighMF R/W3Flowsensorsel R/W3Nominalsizeunit R/W3Nominalsize R/W3
UserSpan →Page201
Totalizer →Page202
Pulse/Statusout →Page203
Statusin →Page204
Analogout/in →Page204
MultirangeForwardspan2 R/W3Forwardspan3 R/W3Forwardspan4 R/W3Reversespan1 R/W3Reversespan2 R/W3Reversespan3 R/W3Reversespan4 R/W3Autorangehyst R/W3Bidirectionhyst R/W3
AUXcalculationFlowdirect R/W3Ratelimit R/W3Deadtime R/W3Noisefilter R/W3Pulsingflow R/W3Powersync R/W3Setpowerfreq R/W3Iexpowerfrequency RMeaspowerfreq RSet SIL R/W3
Display set →Page205
AccesscfgUserrole RChgmainte R/W2Chgspecial R/W3
Deviceinfo →Page206
ProtectionKeycode R/W3Writeprotectsts REnablewrite R/W3Newpassword R/W3Softsealstatus R
<6. Parameter Menu Tree> 201
IM 01E22C02-01EN
�z Process variablesPro var
PVflowselect R/W3Velocity Calorie
DampAO/F R/W3 DampAO/F R/W3Damppls/ttl R/W3 Damppls/ttl R/W3Unit R/W3 Unit R/W3Span R/W3 Timeunit R/W2
Span R/W3Volume Specificheat R/W3
DampAO/F R/W2Damppls/ttl R/W2 DensityUnit R/W2 Valueselect R/W3Timeunit R/W2 Unit R/W3Span R/W2 Fixeddensity R/W3
Std density R/W3Mass Correctdensity R
DampAO/F R/W3Damppls/ttl R/W3 TemperatureUnit R/W3 Stdtemperature R/W3Timeunit R/W2 Meastemperature RSpan R/W3 Fixedtemperature R/W3
CoefA1 R/W3CoefA2 R/W3
Velocitycheck R
�z User spanUser Span
UserspanAO1Select R/W3Span R/W3Unit R/W3
UserspanAO2Select R/W3Span R/W3Unit R/W3
<6. Parameter Menu Tree> 202
IM 01E22C02-01EN
�z TotalizerTotalizer
Totalizer1Unit RConvfactor R/W3Lowcut R/W3Failureopts R/W3Options R/W3Start/Stop R/W3Reset/Preset R/W3Presetvalue R/W3Set point R/W3
Totalizer2Unit R/W3Convfactor R/W3Lowcut R/W3Failureopts R/W3Options R/W3Start/Stop R/W3Reset/Preset R/W3Presetvalue R/W3Set point R/W3
Totalizer3Unit R/W3Convfactor R/W3Lowcut R/W3Failureopts R/W3Options R/W3Start/Stop R/W3Reset/Preset R/W3Presetvalue R/W3Set point R/W3
<6. Parameter Menu Tree> 203
IM 01E22C02-01EN
�z Pulse/Status outputPulse/Status out
PO1/SO1Outputmode R/W3Activemode R/W3Fixwidth R/W3Rateunit R/W2Ratevalue R/W2Lowcut R/W3Alarmout R/W3Frequencyat0% R/W2Frequencyat100% R/W2SO1function R/W3Options R/W3
PO2/SO2Activepulse R/W3Outputmode R/W3Pulseselect R/W3Activemode R/W3Fixwidth R/W3Rateunit R/W3Ratevalue R/W3Lowcut R/W3Alarmout R/W3Frequencyat0% R/W3Frequencyat100% R/W3SO2function R/W3Options R/W3
SO11Activemode R/W3Function R/W3
SO12Activemode R/W3Function R/W3
<6. Parameter Menu Tree> 204
IM 01E22C02-01EN
�z Status inputStatus in
SI11Activemode R/W3Function R/W3State R
SI12Activemode R/W3Function R/W3State R
�z Analog output/inputAnalog out/in
AO1Lowcut R/W3Highlimit R/W3Lowlimit R/W3Alarmout R/W3Rangemode R/W3
AO2AO2select R/W3Lowcut R/W3Highlimit R/W3Lowlimit R/W3Alarmout R/W3
AIFunction R/W3Highlimit R/W3Lowlimit R/W3Value RUnit R/W3URV R/W3LRV R/W3
<6. Parameter Menu Tree> 205
IM 01E22C02-01EN
�z Display setDisplay set
LineselectLine 1 R/W1Line 2 R/W1Line 3 R/W1Line 4 R/W1Line 5 R/W1Line 6 R/W1Line 7 R/W1Line 8 R/W1
TrendselectTrend1 R/W1Trend2 R/W1Trend3 R/W1Trend4 R/W1
DispformatFormatPV R/W1Formattotal1 R/W1Formattotal2 R/W1Formattotal3 R/W1
OptionalconfigContrast R/W1Linemode R/W1Period R/W1NE107 display R/W1Alarmdisplay R/W1Scrollmode R/W1Damp R/W1Formatdate R/W1Inversion R/W1Languagepackage R/W1Displaymode R/W1TrendofflnLRV R/W1TrendofflnURV R/W1
<6. Parameter Menu Tree> 206
IM 01E22C02-01EN
�z Device informationDevice info
Date/Time Ver/NuminfoCurrentdate R Transmittertype RCurrenttime R OptionboardID ROperationtime R TransserialNo R/W3Set date R/W3 Sensor serial No R/W3Settime R/W3 Mainsoftrev R
Snsrsoftrev ROrderinfo Indsoftrev R
TagNo R/W3Longtag R/W3 IOstatusElectrodesize R/W3 MultiI/O RExplosionprotection R/W3MScode Memo
Modelcode R/W3 Memo1 R/W3Suffixconfig1 R/W3 Memo2 R/W3Suffixconfig2 R/W3 Memo3 R/W3Option 1 R/W3Option 2 R/W3Option 3 R/W3Option 4 R/W3
RSMScodeModelcode R/W3Suffixconfig1 R/W3Suffixconfig2 R/W3Option 1 R/W3Option 2 R/W3Option 3 R/W3Option 4 R/W3
Disp install R/W3
<6. Parameter Menu Tree> 207
IM 01E22C02-01EN
�� microSDmicroSD
Contents RUnmount R/W1Format R/W1Property
Totalspace RAvailablespace RFilesystem R
208
IM 01E22C02-01EN
Revision InformationzTitle : ADMAGTISeriesAXG1AMagneticFlowmeterRemoteTransmitterBRAIN
CommunicationTypezManualNo. : IM 01E22C02-01EN
Edition Date Page Revised Item
1st Aug.2019 — Newpublication
2nd Mar. 2020 —1332102114128
Incorporatethemanualchange19-0020-E.Correctionoferrors.2.2.1 AddNOTE.4.1.2 Addthesettingexample.4.8.5 Addthesettingexample.4.9.4 AddNOTE.4.11.4 AddIMPORTANT
3rd Mar. 2020 14-16 2.3 Changetheoperationprocedurebythedisplayunit.