do not open the tf analyzer 2000 system under … do not open the tf analyzer 2000 system under any...
TRANSCRIPT
-
i
Do not open the TF Analyzer 2000 system under any circum-stances!!! Neither the Basic Unit nor the Probe Head. It existsdanger to life because of the supply voltage. No adjustments arepossible by the user.
If you find out malfunctioning of the system, please contact:
aixACCT Systems GmbHDennewartstr. 25 - 27
D-52068 AachenGermany
phone ++49 241 963 1410fax ++49 241 963 1411e-mail [email protected]
Information furnished in this manual by aixACCT is believed to be accurate andreliable. aixACCT assumes no liability for the user or misuse of the informationprovided herein.
aixACCT is not liable for any damages or injuries incurred as the result of open-ing the Basic Unit or the Probe Head of the TF Analyzer 2000 system by anyperson or the operation of this system in a manner inconsistent with the proce-dures and recommendations in the operation manual of the TF Analyzer 2000system. Any changes or modifications not expressly approved by aixACCT willvoid the users authority to operate the equipment.
Note: This equipment has been tested and found to comply with the limits for a Class A digital device,pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection againstharmful interference when the equipment is operated in a commercial environment. This equipment gen-erates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with theinstruction manual, may cause harmful interference to radio communications. Operation of this equipmentin a residential area is likely to cause harmful interference in which case the user will be required to correctthe interference at his own expense.
-
ii
Licenses, Warranty, Export Restrictions
Hardware and software licensesaixACCT will repair or replace, at its location of business, within the period of one year fromdelivery at no expense to the purchaser faulty or damaged hardware or software where suchfaults or damage are the direct result of design or manufacture of the TF Analyzer 2000 sys-tem and / or its measurement software.
Limitations on warrantyThe warranty set forth above does not extend to and shall not apply to:
1. TF Analyzer 2000 systems which have been repaired or altered by other than aixACCTpersonnel, unless the user have properly altered or repaired the TF Analyzer 2000 sys-temin accordance with procedures previously approved in writing by the aixACCT Sys-tems GmbH.
2. TF Analyzer 2000 system which have been subject to misuse, neglect, accident, or im-proper installation.
Export limitationsThe TF Analyzer 2000 system contains subsystems whose export are governed by the Bundes-ausfuhramt of the Federal Republic of Germany concerning commodities of this type. Neitherthe TF Analyzer 2000 system nor its subcomponents may be exported without first contactingaixACCT and the Bundesausfuhramt.
Trademarks used in this manualProducts and company names listed are trademarks or trade names of their respective compa-nies: Pentium, IBM, WinNT.
-
iii
Contents
Licenses, Warranty, Export Restrictions
1 General
1.1 Concept of the TF Analyzer 2000 system . . . . . . . . . . . . . . . . . . . 1
1.2 Basic unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.3 Probe head . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2 System installation
2.1 Package contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.2 Hardware installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.3 TF Analyzer in combination with high voltage amplifiers (optional) . . . . . 9
2.3.1 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.3.2 Testing the high voltage option . . . . . . . . . . . . . . . . . . . . 9
2.4 TF Analyzer with displacement sensors for piezo measurements (optional) . . 13
2.5 TF Analyzer with oscilloscopes for high-speed measurements (optional) . . . 17
2.6 TF Analyzer in combination with a Racal switch box (optional) . . . . . . . . 21
2.6.1 Switch box installation . . . . . . . . . . . . . . . . . . . . . . . . . 21
2.7 TF Analyzer in combination with an Agilent switch box E5250A (optional) . 25
2.7.1 Switch box installation . . . . . . . . . . . . . . . . . . . . . . . . . 25
3 Measuring with the TF Analyzer
3.1 General program overview . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
3.1.1 Measurement dialog . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.1.2 Setting of the measurement temperature (optional) . . . . . . . . . . 31
3.1.3 Manual waveform dialog . . . . . . . . . . . . . . . . . . . . . . . . 32
3.1.4 Settings dialog box . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.1.5 Compensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
3.1.6 Description of the graph windows . . . . . . . . . . . . . . . . . . . 39
3.1.7 Data Formats . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
3.1.8 Multi-sample measurements with a switch box (optional) . . . . . . . 43
3.2 Dynamic Hysteresis Measurement - DHM . . . . . . . . . . . . . . . . . . . 47
-
iv
3.2.1 Measurement procedure and typical measurement . . . . . . . . . . 47
3.2.2 The characteristic parameters of the hysteresis loop . . . . . . . . . . 48
3.2.3 Execution of a hysteresis measurement . . . . . . . . . . . . . . . . 49
3.2.4 ASCII-File of the Hysteresis Program . . . . . . . . . . . . . . . . . 50
3.3 Pulse Measurement - PM . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
3.3.1 Measurement procedure and typical measurement . . . . . . . . . . 53
3.3.2 Execution of a PM measurement . . . . . . . . . . . . . . . . . . . . 54
3.3.3 ASCII-format of the Pund measurement . . . . . . . . . . . . . . . . 56
3.4 Static Hysteresis Measurement - SHM . . . . . . . . . . . . . . . . . . . . . 59
3.4.1 Measurement procedure and typical measurement . . . . . . . . . . 59
3.4.2 Execution of a static hysteresis measurement . . . . . . . . . . . . . 60
3.4.3 ASCII-File of the Static Hysteresis Program . . . . . . . . . . . . . 61
3.5 Leakage Measurement - LM . . . . . . . . . . . . . . . . . . . . . . . . . . 63
3.5.1 Measurement procedure and typical measurement . . . . . . . . . . 63
3.5.2 Description of the leakage measurement method . . . . . . . . . . . 63
3.5.3 ASCII file format of the Leakage measurement . . . . . . . . . . . . 65
3.6 Piezo Measurement - PZM (optional) . . . . . . . . . . . . . . . . . . . . . 67
3.6.1 Measurement procedure and typical measurement . . . . . . . . . . 67
3.6.2 Execution of a Piezo measurement . . . . . . . . . . . . . . . . . . . 68
3.6.3 ASCII-format of the piezo program . . . . . . . . . . . . . . . . . . 70
3.7 CV and piezo coefficient Measurement - CVM (optional) . . . . . . . . . . . 73
3.7.1 Measurement procedure and typical measurement . . . . . . . . . . 73
3.7.2 Execution of a CV measurement . . . . . . . . . . . . . . . . . . . . 73
3.7.3 ASCII-format of the CVM program . . . . . . . . . . . . . . . . . . 76
3.8 Thermo Measurement - THM (optional) . . . . . . . . . . . . . . . . . . . . 77
3.8.1 Measurement procedure and typical measurement . . . . . . . . . . 77
3.8.2 Execution of a Thermo measurement . . . . . . . . . . . . . . . . . 77
3.8.3 ASCII-File of the Thermo Program . . . . . . . . . . . . . . . . . . 79
3.9 Fatigue Measurement - FM . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
3.9.1 Measurement procedure and typical measurement . . . . . . . . . . 83
3.9.2 Execution of a fatigue measurement . . . . . . . . . . . . . . . . . . 84
-
v
3.9.3 ASCII-File of the Fatigue Measurement . . . . . . . . . . . . . . . . 86
3.10 Imprint Measurement - IM . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
3.10.1 Measurement procedure and typical measurement . . . . . . . . . . 89
3.10.2 Executing an Imprint measurement . . . . . . . . . . . . . . . . . . 91
3.10.3 ASCII-File of the Imprint Program . . . . . . . . . . . . . . . . . . 93
3.11 Retention Measurement - RM . . . . . . . . . . . . . . . . . . . . . . . . . 95
3.11.1 Measurement procedure and typical measurement . . . . . . . . . . 95
3.11.2 Executing a retention measurement . . . . . . . . . . . . . . . . . . 96
3.11.3 ASCII-File of the Retention program . . . . . . . . . . . . . . . . . 98
3.12 Access Time Measurement - ATM (optional) . . . . . . . . . . . . . . . . . 101
3.12.1 Measurement procedure and typical measurement . . . . . . . . . . 101
3.12.2 Execution of an Access Time measurement . . . . . . . . . . . . . . 102
3.12.3 ASCII-format of the Access Time program . . . . . . . . . . . . . . 104
4 GPIB control of the TF Analyzer (optional)
4.1 Setup of the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
4.2 GPIB commands for the TF Analyzer . . . . . . . . . . . . . . . . . . . . . 108
4.2.1 General commands . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
4.2.2 File operation commands . . . . . . . . . . . . . . . . . . . . . . . 109
4.2.3 Option commands . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
4.2.4 Sampledata commands . . . . . . . . . . . . . . . . . . . . . . . . . 110
4.2.5 Hysteresis module commands . . . . . . . . . . . . . . . . . . . . . 111
4.2.6 PUND module commands . . . . . . . . . . . . . . . . . . . . . . . 112
4.2.7 Static hysteresis module commands . . . . . . . . . . . . . . . . . . 112
4.2.8 Leakage module commands . . . . . . . . . . . . . . . . . . . . . . 113
4.2.9 Piezo module commands . . . . . . . . . . . . . . . . . . . . . . . . 113
4.2.10 CV module commands . . . . . . . . . . . . . . . . . . . . . . . . . 114
4.2.11 Thermo module commands . . . . . . . . . . . . . . . . . . . . . . 114
4.2.12 Fatigue module commands . . . . . . . . . . . . . . . . . . . . . . . 115
4.2.13 Imprint module commands . . . . . . . . . . . . . . . . . . . . . . . 115
4.2.14 Retention module commands . . . . . . . . . . . . . . . . . . . . . 116
4.2.15 Accesstime module . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
4.2.16 Graph commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
-
vi
5 Troubleshooting
5.1 Measurement problems and possible solutions . . . . . . . . . . . . . . . . . 119
-
1
1 General
1.1 Concept of the TF Analyzer 2000 system
The TF Analyzer 2000 system is the first modular designed electrical characterization system
for electroceramic thin films and bulk ceramic samples. Comprehensive material characteriza-
tion by small expense is the intention of this setup. aixACCT is able to offer a system both with
excellent electrical performance and unique test methods.
Within the TF Analyzer 2000 system four modules are offered:
FE-module to investigate the ferroelectric hysteresis loop
PS-module to investigate the switching kinetics of ferroelectric materials
RX-module to investigate the relaxation behavior of electroceramic materials
DR-module to investigate the self discharge behaviour
The different modules are optimized for special measurement requirements and complement
each other. For example completes the pulse switching module the investigation of the fre-
quency dependence of the hysteresis loop very well. So, the investigation of this effect can
be observed from the mHz range, using the static hysteresis and the dynamic hysteresis mea-
surement of the FE module to the MHz range, using the PS module. Newest characterization
methods are implemented like static hysteresis measurement. This method helps to separate re-
versible and irreversible parts of the polarization. In correlation with the data from the imprint
measurement the user is able to perform a life time extrapolation for his material in view of
memory application.
The specific amplifiers are located in a so called Probe Head. Since it can be placed very close
to the sample the loop between drive and return is much reduced and so the signal to noise ration
is increased significantly. The computer and the high precision power supplies are located in
the Basic Unit. Changing the Probe Head means to change the characterization method. To do
this you simply have to solve and reconnect four cables.
1.2 Basic unit
The TF Analyzer 2000 System consists of the Basic Unit and the Probe Head. The Basic Unit
contains the whole of the computer and the power supplies.
The computer is build of an IBM compatible computer, containing a Pentium class processor, at
least 32 Mbyte memory, a hard disk with more than 2 GByte capacity, and a VGA color graphic
-
2 1 General
card. The TF Analyzer 2000 system offers an efficient personal computer system. Several
interface cards which control the interaction with the Probe Head will complete the PC unit. A
digital I/O-card provides 48 bit to control the amplification of the current to voltage converter
and the electrometer. The analog input card is a high speed, high resolution data acquisition
board. The maximum sample rate amounts to 1 MHz and the resolution is 12 bit at a voltage
range of 10 V, equivalent to 4.88 mV resolution. The arbitrary waveform generator applies the
selected pulse train to the device under test. The vertical resolution amounts to 12 bit at 10 Vpp,
(2.44 mV = LSB) and the sample frequency is up to 80 MHz.
The second part of the Basic Unit contains the power supplies for the Probe Head and additional
circuit parts to handle the different Probe Heads. Electromagnetic shielding is realized and
separates the two parts of the Basic Unit. This guarantees the performance of the high precision
analog power supplies and fits in the overall concept of the TF Analyzer 2000 system to fulfill
the requirements of the EMC.
1.3 Probe head
The concept of the TF Analyzer 2000 system includes the minimization of the incoupling loop
between drive and return, because the Probe Head of the system contains the necessary ampli-
fiers to drive a capacitive load and to record the data of the measuring cycle. So, this setup can
be used with any professional probe station without the need of long cables, because the Probe
Head can be placed close to the device under test. This concept guarantees an optimum signal
to noise ratio. Three amplifiers are located in the FE-module. To drive the capacitive load two
different amplifiers are used depending on the selected program, e. g. Dynamic Hysteresis or
Fatigue. The hysteresis program allows cycletimes between one second and 1 ms, respectively
1 microsecond depending on the configuration. In this case the TF Analyzer 2000 system offers
an amplifier which provides voltage amplitudes up to 58 Vpp. In the case of a fatigue measure-
ment the amplifier is changed to offer the user frequencies up to 20 MHz driving a 100 pF linear
Styroflex-capacitor by an amplitude of 10 Vpp. The third amplifier detects the current response
of the device under test. The method used to collect the data is the feedback method, which is
realized by a current to voltage converter. This method reduces the influence of parasitic capac-
itance and back voltage known from the Sawyer Tower measurement drastically and enable for
the first time direct hysteresis measurements on nanosized samples with submicron square area.
Fig. 1.3.1 shows the circuit design in principle.
Fig. 1.3.1: Principle of the feed-back method.
-
2.2 Hardware installation 3
2 System installation
2.1 Package contents
After receiving the TF Analyzer 2000 system please unpack the system and inspect it for any
damage.
The delivery consists of the following items:
1. TF Analyzer 2000 Basic Unit
2. TF Analyzer 2000 FE-Probe Head
3. PS 2 mouse
4. Keyboard
5. Power cable
6. 2 BNC cables male to male
7. Lemo cable 18 pins male to male
8. Lemo cable 16 pins male to male
9. Windows NT 4.0 software
10. Hysteresis software version 1.44
11. TREK high voltage amplifier (optional)
12. Manual of the TREK amplifier (optional)
13. High voltage cable (optional)
14. 2 BNC cables male to male (optional)
15. Certificate with the license key
16. This manual
If any item is damaged or missing, please contact the aixACCT Systems GmbH immediately at
phone ++49 241 963 1410fax ++49 241 963 1411e-mail [email protected]
2.2 Hardware installation
Please proceed with the following steps for installation:
1. Connect the monitor cable to the VGA plug on the rear plane.
2. Connect the PS 2 mouse to plug socket on the rear panel.
3. Connect the keyboard to plug socket on the rear panel.
-
4 2 System installation
Make sure that the power switch of the Basic Unit is turned off.
4. Connect the power cable to the Basic Unit.
5. Connect the power cable of the Basic Unit to mains.
It is very important for personal safety and orderly operation of the system that the
ground connection is established!
6. Connect the power cable of the monitor to mains.
7. Connect the Probe Head to the Basic Unit:
(a) Connect lemo cables from the Basic Unit to the Probe Head. These cables cannot
be mixed up with each other.
(b) Connect BNC cables from Basic Unit to the Probe Head. To avoid wrong connection
both Basic Unit and Probe Head have one BNC plug which is marked by a red
border. Connect the red BNC cable to these two plugs. Connect the black BNC
cable to the other two plugs.
(c) Connect the BNC connectors at the narrow side of the Probe Head to the device
under test.
Please notice that no cabling longer than 3 m is allowed between Probe Head and Basic Unit
to operate this system pursuant to Part 15 of the FCC Rules or the CE standard. Furthermore
the user is responsible for a proper shielding of the sample or probe station according to these
standards.
For additional information to the LEDs and to the other connectors refer to Fig. 2.2.1 to Fig. 2.2.3.
On the right hand of the front panel there are the usual switches of a personal computer. The
Power LED indicates the supply power of the PC unit. On the left of the power LED there is
the ON/OFF switch of the system in addition to the main switch at the rear. The hard disk LED
lights when the hard disk is accessed by the PC. On the right of the hard disk LED is the RESET
switch of the PC located. Above the BNC connectors on the front panel four LEDs indicate the
measurement method correlated with the connected Probe Head.
On the rear panel of the Basic Unit the power plug and the connectors of the interfaces of a
personal computer are placed:
serial ports (two)
parallel port and
monitor interface
mouse connector
keyboard connector.
-
2.2 Hardware installation 5
Fig. 2.2.1: FE Probe Head of the TF Analyzer.
Fig. 2.2.2: Front panel of the TF Analyzer 2000.
-
6 2 System installation
Fig. 2.2.3: Rear panel of the TF Analyzer 2000.
Now you can start to get familiar with the software of the TF Analyzer 2000. The software has
already been installed by aixACCT Systems.
To check, if the system works perfectly right, please follow the given instructions:
1. Turn on the power switch of the Basic Unit. First the main switch at the rear and after-
wards the ON/OFF switch at the front (e.g. with the tip of a pencil).
2. Turn on the power switch of the monitor.
3. The system starts Windows NT automatically.
4. To login press CTRL+Alt+Del when required.
5. Use TFA as username in the login dialog box and press enter. There is no password
required by delivery, so this line can be left empty.
6. Now the operating system continues booting and finally the main graphical user interface
of Windows NT comes up.
7. Move the cursor on the hysteresis icon on the desktop and double-click on the left mouse
button.
8. The window with the main menu of the FE-software opens up.
9. Open the measurement dialog window (entry in the Options menu or button in the tool
bar).
10. Click on Hysteresis measurement (default).
11. Interconnect the In and Out BNC connectors on the front panel of the Basic Unit with a
BNC cable and press Start.
12. When the measurement is finished an ellipsoid is displayed in the appearing graph win-
dow.
13. Remove the short circuit between plug In and Out and connect the BNC and Lemo cables
-
2.2 Hardware installation 7
to the FE-Probe Head as mentioned before.
14. Connect a linear capacitor to the BNC plugs at the narrow side of the Probe Head.
15. Choose 5 V amplitude and 100 Hz frequency in the Input Parameters of the hysteresis
program.
16. Press AutoRange and wait until the Start button is enabled again.
17. Press Start.
18. Now nearly a line should be showed in the graph window. The reason for the slight
opening in the line is the output resistance of the amplifier driving the capacitive load.
These tests ensure that Basic Unit and FE-Probe Head are working perfectly right.
In case the system has been reinstalled or the license file is corrupted, a window appears at
program start which allows the restoration of the license file. Please type in the license key
which can be found on the hard disc (c:\Program Files\TFAnalyzer\Licensekey.txt) or printedon the certificate delivered with the system. In any other case please contact aixACCT Systems
GmbH immediately at
phone ++49 241 963 1410fax ++49 241 963 1411e-mail [email protected]
-
8 2 System installation
FE
ampl
ifier
in
put
sign
al
outp
ut
prob
e st
atio
n
BN
C c
able
, shi
eld
grou
nded
BN
C c
able
, shi
eld
grou
nded
red
mar
ked
BN
C c
able
botto
m
elec
trode
top
elec
trode
conn
ect c
huck
gr
ound
to F
E-
mod
ule
grou
nd
for
best
res
ults
Fig. 2.2.4: Wiring scheme for FE-module.
-
2.3 TF Analyzer in combination with high voltage amplifiers (optional) 9
2.3 TF Analyzer in combination with high voltage amplifiers (optional)
2.3.1 Installation
The TF Analyzer 2000 system can be used in combination with miscellaneous high voltage
amplifiers. Standard HV amplifier ranges provided with the system are 200 V, 500 V, 1 kV,
2 kV, 4 kV, or 10 kV. The resolution of the high voltage system will be: Maximum Output
Voltage 2 4096. The connection between the power supply and the high voltage amplifier
is described in the wiring scheme in Fig. 2.3.2.
Connect the earth terminal of the Basic Unit and the High Voltage Amplifier.
Connect the enable BNC connector on the rear panel of the TF Analyzer to the digital
enable input on the front panel of the high voltage amplifier. This connection ensures that
voltage is applied only during measurements controlled by the software. An interlock
switch can be optionally wired in series.
In order to select between the external high voltage amplifier and the built-in amplifier of the
TF Analyzer 2000, choose the Options - Settings menu item of the main menu (see Fig. 2.3.1).
The external high voltage amplifier is selected if the enable checkbox is checked. In case the
user has more than one high voltage amplifier installed, the connected model can be selected in
the selection box beside the enable button. A message box is used to confirm the selection of
the high voltage option after the Settings dialog is closed with the OK button.
Notice that in case of use of the high voltage amplifier, the output of the high voltage amplifier
is directly connected to the sample holder which contacts the top electrode of the device under
test.
Please ensure that the cable connection corresponds to the wiring diagram in Fig. 2.3.2.
Notice to turn off the high voltage in any case before you change the cabling and ensure that the
device under test is discharged.
There is danger to life if you touch the high voltage line or a charged sample !
Dont press the Start button before the connection corresponding to the wiring diagram is en-
sured. At any time the sample should be protected from touching during operation.
2.3.2 Testing the high voltage option
To test the high voltage option, please follow the steps below. Refer to the wiring diagram
in Fig. 2.3.2 to connect the Basic Unit and the Probe Head, the high voltage amplifier (e. g.
TREK) and the positioner unit.
-
10 2 System installation
Fig. 2.3.1: How to select the external high voltage amplifier.
The sample must be protected from touching in any case, when high voltage is applied to
the sample !
Connect a linear capacitor which maximum voltage parameter exceeds the maximum applied
voltage of 500 V to the positioner unit. Start the Hysteresis main program and select Options
- Settings out of the main menu and enable the high voltage amplifier. Select Hysteresis in the
Measurement dialog, choose appropriate parameters and a sufficient current range and press the
Start button. After the measurement is done, a rising line for the linear capacitor can be seen in
the P (V ) diagram of the graph window.
This test ensures that Basic Unit and Probe Head are working perfectly right together with the
high voltage amplifier.
In any other case contact aixACCT Systems GmbH immediately at
phone ++49 241 963 1410fax ++49 241 963 1411e-mail [email protected]
-
2.3 TF Analyzer in combination with high voltage amplifiers (optional) 11
TFA
NA
LYZ
ER
20
00
Pow
erI/
OIn
Out
Hig
h
Vo
lta
ge
Out
pu
t
)(
ampl
ifier
in
put
prob
esH
igh
Vol
tage
cabl
e
BN
C c
able
, shi
eld
grou
nded
,ho
t wire
= v
irtua
l gro
und
inpu
t
+9
$P
SOL
ILH
U
Inpu
t Sig
nal
rear
Gro
und
Con
nect
ion
Volta
ge
Mon
itor
conn
ect C
H2
to
volta
ge m
onito
r
Touc
h Pr
otec
tion
Dig
ital E
nabl
e(s
witc
hH
V o
ff)
outp
ut
not u
sed
DA
NG
ER
!H
igh
volta
ge is
app
lied
durin
g m
easu
rem
ents
!O
bser
ve s
afet
y pr
ecau
tions
!
Fig. 2.3.2: Wiring scheme for FE-module together with a high voltage amplifier.
-
12 2 System installation
-
2.4 TF Analyzer with displacement sensors for piezo measurements (optional) 13
2.4 TF Analyzer with displacement sensors for piezo measurements (op-tional)
The TF ANALYZER 2000 system can be combined with various displacement sensors for large
and small signal characterization of piezoelectric and electrostrictive materials. An analog sig-
nal output, which is proportional to the displacement, can directly be connected to the TF An-
alyzer as an additional input channel. The input has a resolution of 12 bit which equals with
Vpp = 10 V to 4.88 mV. For measurements including a high voltage amplifier please refer to
chapter 2.3. Simultaneous measurements of polarization and displacement and of capacitance
and piezoelectric coefficient versus an applied bias voltage range can be performed.
Notice that in case of use of the high voltage unit, the output of the high voltage amplifier is
directly connected to the sample holder which contacts the top electrode of the device under
test. To select between the external high voltage amplifier and the build in amplifier of the
TF Analyzer 2000, access the Options - Settings menu item of the main menu. Select Output
amplifier and choose HV Amplifier. You will be requested to ensure the cable connection
corresponding to the wiring scheme in this chapter. Notice to turn off the high voltage in any
case and ensure that the device under test is discharged before you change the cabling. In
the same menu (see Fig. 2.4.1) the Piezo Factor can be specified. It correlates the output voltage
of the displacement sensor to the change in length of the sample. The Piezo Factor should be in
the unit nm/V.
There is danger to life if you touch the high voltage line or a charged sample!
Dont press the Start button before the connection corresponding to the wiring diagram is en-
sured.
For displacement measurements with an additional high voltage amplifier, please refer to the
wiring scheme in Fig. 2.4.3 to connect the Basic Unit and the Probe Head, the TREK High
Voltage Amplifier and the positioner unit.
The sample must be protected from touching in any case, when high voltage is applied to
the sample.
In case fo any malfunction please contact aixACCT Systems GmbH immediately at
phone ++49 241 963 1410fax ++49 241 963 1411e-mail [email protected]
-
14 2 System installation
Fig. 2.4.1: Setting of the Piezo Factor to connect the displacement sensor to the TF Analyzer2000 system.
-
2.4 TF Analyzer with displacement sensors for piezo measurements (optional) 15
TFA
NA
LYZ
ER
20
00
Pow
erI/
OIn
Out
)(
ampl
ifier
in
put
prob
esdr
ive
BN
C c
able
(bot
tom
)
BN
C c
able
, shi
eld
grou
nded
,ho
t wire
= v
irtua
l gro
und
inpu
t
grou
nded
shi
eld
outp
ut
conn
ect C
H1
to
disp
lace
men
t sig
nal
volta
ge o
utpu
t fo
r di
spla
cem
ent
'LV
SOD
FH
PH
QW
0H
DV
XUH
PH
QW
XQ
LW/
DV
HU
,QWH
UIH
URP
HWH
UR
UP
HF
KD
QLF
DO
Fig. 2.4.2: Wiring scheme for piezoelectric measurement setup with the FE-module.
-
16 2 System installation
! "!
FE
ampl
ifier
in
put
prob
esH
igh
Vol
tage
cabl
e
BN
C c
able
, shi
eld
grou
nded
,ho
t wire
= v
irtua
l gro
und
inpu
t
#$% &
'( )* ) +
,
- ./01
23 4.56
7857
9 7:0.
; ? @A
B C
DD
EFG
H? @A
I
C
J KLM
N OPK
Q RMQS KT
Dig
ital E
nabl
e(s
witc
hH
V o
ff)
outp
ut
not u
sed
DD
EF U
V C WX? @
DI
WC A@
?
H? @A
X
Y V C WX? @
DI
D
AN
GE
R !
Hig
h vo
ltage
is a
pplie
d du
ring
mea
sure
men
ts !
Obs
erve
saf
ety
prec
autio
ns!
Z[ \]^ _`abacd
e a_\fgabacdfc[
d
h i jk+, lm n
o +,* +,p&+
o +,p,&+qr jn)qj
(s
t uvw x
yz{| x}
~w x
v
Fig. 2.4.3: Wiring scheme for piezoelectric measurement setup with the FE-module and anadditional high voltage amplifier.
-
2.5 TF Analyzer with oscilloscopes for high-speed measurements (optional) 17
2.5 TF Analyzer with oscilloscopes for high-speed measurements (optional)
The TF Analyzer 2000 system can optionally be combined with various oscilloscopes for per-
forming high-speed characterization of materials. The oscilloscope is used to perform the data
acquisition at high speeds to increase the frequency range of hysteresis or pulse measurements
in combination with the enhanced version of the FE module. For the connection of oscillo-
scopes refer to the corresponding drawings in this chapter. For the oscilloscope model types,
several types have been tested to work correctly, e.g. Tektronix TDS684, TDS220, TDS1000
or TDS2000 series. Since the number of models that have been tested to work is steadily in-
creasing, please contact aixACCT for further manufacture and model types. Make sure that the
oscilloscope is set-up to be controlled remotely by the TF Analyzer via GPIB and the oscillo-
scope is set to GPIB address 25 by default. All necessary settings of the oscilloscope are then
controlled by the TF Analyzer. By default the current is monitored by channel 1, the voltage is
monitored by channel 2. Use of active probes is recommended to reduce the cabling influence.
The external trigger output on the rear of the TF Analyzer Basic Unit has to be connected to the
auxiliary trigger input of the oscilloscope by a short (
-
18 2 System installation
FE
virtu
al g
roun
d in
put
volta
ge o
utpu
t
prob
e st
atio
n
BN
C c
able
keep
as
shor
t as
poss
ible
BN
C c
able
keep
as
shor
t as
poss
ible
red
mar
ked
BN
C c
able
( bot
tom
el
ectro
de )
( top
ele
ctro
de )
Infi
niiu
mH
P5
48
x5
AG
PIB
con
trol c
able
BN
C tr
igge
rou
tto
AU
X tr
igge
r in
put
(rea
r)
!"#$ !
%& ''
(#$ )"*!+
2
use
Ch.
2 fo
r vo
ltage
mon
itorin
g
NC
conn
ectc
urre
ntou
tput
to
Cha
nnel
1 (a
nd to
bas
ic u
nit v
iaT-
conn
ecto
r, op
tiona
lly)
conn
ectt
o C
hann
el2
here
vi
aB
NC
(T) f
or
volta
ge
mon
itorin
g
orco
nnec
tbot
tom
el
ectro
de(d
rive)
to C
h.2
here
by
activ
e pr
obe
setG
PIB
add
ress
to 2
5
,
-
Fig. 2.5.1: Wiring scheme for FE-module together with an HP oscilloscope
-
2.5 TF Analyzer with oscilloscopes for high-speed measurements (optional) 19
Fig. 2.5.2: Wiring scheme for the high speed FE-module together with a Tektronix TDS 220oscilloscope.
-
20 2 System installation
FE
virtu
al g
roun
d in
put
volta
ge o
utpu
t
prob
e st
atio
n
BN
C c
able
keep
as
shor
t as
poss
ible
BN
C c
able
keep
as
shor
t as
poss
ible
red
mar
ked
BN
C c
able
( bot
tom
el
ectro
de )
( top
el
ectro
de )
Te
ktr
on
ix T
DS
68
4C
GP
IB c
ontro
l cab
le
BN
C tr
igge
rout
to a
ux
trigg
er in
put
! """ #
$% &
'() *+,-
.
2
use
Ch.
2 fo
r vo
ltage
mon
itorin
g
NC
conn
ectc
urre
ntou
tput
to
Ch.
1(an
d to
bas
ic u
nit v
ia
T-c
onne
ctor
, opt
iona
lly)
conn
ectt
o C
h.2
here
via
BN
C (T
) fo
r sim
plifi
ed
volta
ge m
onito
ring
orco
nnec
tbot
tom
el
ectro
de(d
rive)
to C
h.2
here
by
activ
e pr
obe
Fig. 2.5.3: Wiring scheme for the high speed FE-module together with a Tektronix TDS 684Coscilloscope.
-
2.6 TF Analyzer in combination with a Racal switch box (optional) 21
2.6 TF Analyzer in combination with a Racal switch box (optional)
For industrial application the TF ANALYZER 2000 standard system can be extended by a
switch box. This enables the user to successively select and measure several samples to the
Probe Head. The switchbox is automatically controlled via GPIB (general purpose interface
bus) by the measurement software of the TF Analyzer.
To perform multi-sample measurements with the switch box choose enable Switchbox in the
Options - Settings - Measurement menu. Additional settings have to be specified to adjust the
hardware of the TF Analyzer, the switch box, and the probe card to contact the samples. This
is done in the mapping file switches.ini to map each switch of the switch matrix to a contact pin
on the probe card (see below).
The selection of the contacts and their measuring sequence is located in the Options - Sam-
pledata menu. If the switch box is enabled, this menu is extended with input lines for top and
bottom contact and buttons for editing a contact list. This is described in detail in the multisam-
ple measureing section 3.1.8.
2.6.1 Switch box installation
Editing the mapping file Instead of directly choosing the switch number of the switch matrix
a mapping file can be used to correlate the physical switch number with a mnemonic name.
This can be the number of the probe needle of a probe card or a text on the measured wafer
itself. The mapping file is named switches.ini and must be located in the directory C:\ProgramFiles\TFAnalyzer. It is a plain ASCII file which can be edited with any text editor (e. g.notepad).
The left column is a mnemonic name which is used in the sampledata menu to specify the top
and bottom contact of a sample to be measured. This can be for example a name of the probe
card tip or a text on the wafer. The rigth column contains the names respectively numbers to
address the switches of the switch box (for details refer to the manual of the switchbox). A
clipping of a mapping file example is listed below.
[Contacts]ModuleAddr1 11Bottom = 311Top = 152Bottom = 302Top = 143Bottom = 293Top = 134Bottom = 284Top = 12
-
22 2 System installation
Fig. 2.6.1: View of the Racal switch box rear panel .
Connection of switch box and TF Analyzer 2000 system The switch matrix is controlled
by the TF Analyzer via IEEE. So, an IEEE connection between switch matrix and TF Analyzer
has to be build up. Fig. 2.6.1 shows the rear panel of the switch matrix. In case of a 64 channel
switch box one of the four frames is equipped with a switching matrix.
A detailed drawing of the frame itself and its pin configuration is shown in Fig. 2.6.2. Fig. 2.6.3
illustrates where the cables from the Probe Head and the 64 cables to the probe needles are to
be connected to.
The upper plugs are used to connect the four rows. In this configuration the bottom and top
contact are connected to row 2 and 3 of the matrix. Both cables in the lower plugs connect 64
switches each of them with separate shielding for excellent signal to noise ratio.
Each 64 pins plug offers 32 contacts. Each signal line is used with a separate shielding line.
Row A contains the signal line, row B of the plug represents the shielding lines. As shown in
Fig. 2.6.3 the plug on the left contains switch number 1 - 32, the plug on the right contains
switch 33 - 64. Each cable is signed with the logical switch number which is related to the list
in the mapping file switches.ini.
-
2.6 TF Analyzer in combination with a Racal switch box (optional) 23
Fig. 2.6.2: Pin configuration of a Racal 64 channel switching frame.
-
24 2 System installation
Fig. 2.6.3: Scheme how to connect one switching frame.
-
2.7 TF Analyzer in combination with an Agilent switch box E5250A (optional) 25
2.7 TF Analyzer in combination with an Agilent switch box E5250A (op-tional)
For industrial application the TF ANALYZER 2000 standard system can be extended by a
switch box. This enables the user to successively select and measure several samples to the
Probe Head. The switchbox is automatically controlled via GPIB (general purpose interface
bus) by the measurement software of the TF Analyzer.
To perform multi-sample measurements with the switch box choose enable Switchbox in the
Options - Settings - Measurement menu. Additional settings have to be specified to adjust the
hardware of the TF Analyzer, the switch box, and the probe card to contact the samples. This
is done in the mapping file switches.ini to map each switch of the switch matrix to a contact pin
on the probe card (see below).
The selection of the contacts and their measuring sequence is located in the Options - Sam-
pledata menu. If the switch box is enabled, this menu is extended with input lines for top and
bottom contact and buttons for editing a contact list. This is described in detail in the multisam-
ple measureing section 3.1.8.
2.7.1 Switch box installation
Editing the mapping file Instead of directly choosing the switch number of the switch matrix
a mapping file can be used to correlate the physical switch number with a mnemonic name.
This can be the number of the probe needle of a probe card or a text on the measured wafer
itself. The mapping file is named switches.ini and must be located in the directory C:\ProgramFiles\TFAnalyzer. It is a plain ASCII file which can be edited with any text editor (e. g.notepad).
The left column is a mnemonic name which is used in the sampledata menu to specify the top
and bottom contact of a sample to be measured. This can be for example a name of the probe
card tip or a text on the wafer. The rigth column contains the names respectively numbers to
address the switches of the switch box (for details refer to the manual of the switchbox). A
clipping of a mapping file example is listed below.
[Contacts]TopRow = 7BottomRow = 81Bottom = 1011Top = 1022Bottom = 1032Top = 1043Bottom = 105
-
26 2 System installation
Port 7 and 8
Fig. 2.7.1: View of the Agilent E5250A switch box rear panel.
3Top = 1064Bottom = 1074Top = 1085Bottom = 1095Top = 1106Bottom = 1116Top = 112
Connection of switch box and TF Analyzer The switch matrix is controlled by the TF An-
alyzer via IEEE commands. So, an IEEE connection between switch matrix and TF Analyzer
has to be build up. Fig. 2.7.1 shows the rear panel of the switch matrix. Up to four modules
can be plugged into the mainframe. For the connection of the probe card to the module please
refer to the switch box manual. The Probe Head is connected to the input ports 7 and 8 of the
mainframe.
-
27
3 Measuring with the TF Analyzer
3.1 General program overview
Within the FE-module of the TF Analyzer 2000 system several types of measurement proce-
dures are implemented to characterize ferroelectric materials. This includes dynamic and quasi
static hysteresis, PUND and leakage measurements. Furthermore, reliability investigations like
fatigue, imprint, and retention measurements can be performed. Some special type measure-
ments, like Piezo, CV, or Access Time measurements, can be obtained on request optionally.
All these measurement procedures can be accessed within the hysteresis software.
In this section general functions and menu entries of the hysteresis software are described. To
start the hysteresis software either double click on the icon Hysteresis on the desktop or launch
it from the Task bar with Start Programs Hysteresis (see Fig. 3.1.1).
Fig. 3.1.1: Graphical user interface (GUI) of WinNT with the icon of the Hysteresis software.
The main program window in Fig. 3.1.2 contains the title bar containing the product name
TF Analyzer 2000 and the module name FE-MODULE, the menu bar and optional a tool bar
below the menu bar and a status bar at the bottom of the window. When placing the cursor
on a specific icon in the tool bar without pressing a button, a short description of the function
appears. The mentioned tool and status bar are only visible if the entries in the View pull down
-
28 3 Measuring with the TF Analyzer
menu are selected (default setting). All other program functions can be reached by pull down
menus in the menu bar. A detailed description of the contents of the File and Options pull down
menus is given below. Edit contains operations like copy, paste, and cut on open measurement
data. Using Open Compare Graph, which is available for some measurement types, the user is
able to display a second graph in an already open measurement. So, for example it is possible to
compare the dynamic hysteresis data to the static hysteresis data of the same sample in the graph
window. Or, this function can be used for examinations such as the analysis of the influence
of the sample preparation on the electrical properties of a material. To delete this second graph
click on Delete Compare Graph in the menu. The Window menu allows the arrangement of
multiple measurement graphs.
Fig. 3.1.2: Main program window of the hysteresis software with two open measurements andthe Measurement dialog.
The File menu In the File menu all kinds of file operations are summarized. Some entries ap-
pear only in case of an open measurement graph. Also, menu entries for printing and the printer
setup can be found in the File menu. Additionally, a list of recently used files (if available) is
displayed and the entry to exit the program can be found here (see Fig. 3.1.3).
Open is used to reload saved measurements of the FE-Module. In case of an open or loaded
measurement additional entries in the file menu exist. These are operations like Save and Save
-
3.1 General program overview 29
Fig. 3.1.3: Title bar, menu bar with opened File menu and tool bar of the hysteresis software.
As to save measurement data in the TF Analyzer file format and an Export function to save
data in ASCII file format. A description of the ASCII data file format can be found at the end
of each measurement procedure description. Furthermore, Close exits the actual measurement
procedure and Print Preview opens a window with the graph, which will be printed with the
Print menu.
Exit terminates the hysteresis program. The software checks whether measured data has been
saved or not. In the case of unsaved data a dialog box appears and asks the user whether he
wants to save the data or not. This will be done for all open unsaved graph windows.
The Options menu In the pull down menu Options (Fig. 3.1.4) the choices for program set-
tings are summarized. First of all, with the Sampledata entry the user can change parameters
like name, area and thickness according to the sample he wishes to measure. The other Options
menu entries are described in the following sections.
Fig. 3.1.4: Menu choice Options.
-
30 3 Measuring with the TF Analyzer
3.1.1 Measurement dialog
The Open Measure Dialog opens a dialog window which allows the selection of all available
measurement procedures. One out of twelve measurement procedures can be selected from the
Measurement dialog (see Fig. 3.1.5). Some of the measurement procedures (like Piezo, CV,
Cell Test, AccessTime) are only optional available in the program. Please contact the aixACCT
Systems GmbH for further information. When choosing the card index with the name of the
measurement to be performed the necessary input parameters and control buttons for this mea-
surement will be displayed below. To step through the different edit lines and push buttons one
can use the Tab key to step forward, or SHIFT + Tab for backward stepping. Detailed descrip-
tions on the measurements and their specific parameters are given in the following chapters.
Fig. 3.1.5: Measurement dialog to select the various measurement procedures, to apply therequired parameters, and to control the measurement. In this figure the dialog for the hysteresismeasurement is displayed. This box can also be accessed directly from the tool bar icon.
-
3.1 General program overview 31
The Current Amplifier field is located below the Input Parameter field. It is used to select the
current range with which the Probe Head measures. For best measurement results the current
range should be close to the current response of the sample during the measurement. It depends
on input parameters like frequency and amplitude and the measured sample material and geom-
etry. A range between 10 pA and 1 A can be selected for the FE-Probe Head. An automatic
determination of the correct current range can be started with the AutoRange button. The pro-
gram starts a test measurement with the specified input parameters on the connected sample and
detects and selects the best fitting range. It is displayed in the edit line.
The measurement control buttons Start, Stop and Close are located at the bottom of the dialog
boxes. A mouse click on Start runs the selected measurement procedure with the specified
parameters. If the checkbox Append to file (if active) is not selected, a new graph window will
be opened for each measurement. Otherwise the measurement data will be added to an already
open measurement of this type, if this window was active before the Start button was pressed.
This function is not available for all measurement procedures. A running measurement can be
stopped with the Stop button. After clicking on this button it can take some seconds until the
program stops. This depends on the task management of the software and is necessary to ensure
that recorded data does not get lost. All data recorded so far is still accessible. Close is used to
exit the measurement dialog window.
As long as a measurement is running the text Measuring ... is displayed on the left side of the
status bar. In the middle of the status bar a progress bar and the estimated remaining measure-
ment time is displayed. Total is the time to the end of the whole measurement. Next gives the
time to the end of the next hysteresis or PUND measurement, which is of special interest for
long fatigue or retention measurements.
3.1.2 Setting of the measurement temperature (optional)
The toolbar has been extended with a selection box to set the measurement temperature (see
Fig. 3.1.6). This field can be used to constitute a temperatrue which is set by a connected tem-
perature controller before any of the measurements of the measurement dialog is performed.
Except for Thermo measurements in which this field is not evaluated. Temperature values can
be user defined by clicking on and editing the selection box. By default the input tempera-
tures are assumed to be in the unit C. However, if the inserted number is followed by an F
the temperature is accepted in degree Fahrenheit and will be automatically converted into the
C value. Selecting none resets the temperature controller to a default temperature (20C) and
skips the temperature setting for the following measurements. The sample temperature is saved
as additional characteristic value of the measurement and can be viewed in the Graph View.
This feature is only available in systems with an optional temperature controller unit and an
-
32 3 Measuring with the TF Analyzer
Fig. 3.1.6: Selection box in the tool bar to choose a temperature for the next measurement(optional).
appropriate sample holder.
3.1.3 Manual waveform dialog
For advanced users it is possible to replace the typical triangular, sine, or PUND waveforms by
user specific waveforms with the Manual Waveform dialog.
Fig. 3.1.7: Dialog window to create, edit, and save user defined waveforms to be applied to thesample for measurement.
This dialog provides all necessary functions to define, edit, and save new waveforms. Add
allows the easy creation of standard waveforms like sine, triangle or rectangle with user defined
-
3.1 General program overview 33
amplitude, frequency, and number of repetitions. Alternative to the manually edited waveform it
is possible to import an ASCII data file with the Import button. The selected file should contain
the name of the waveform as first line and subsequently lines containing two numeric values
for time and voltage separated by spaces or Tab characters. The edited waveform is displayed
as red curve in the graph of the manual waveform dialog window. Start and duration of the data
recording can be set with the Trigger and Recording Time when the check box is enabled. This
is visualized as green curve in the graph. If both values are set to zero data will be recorded
during the whole duration of the excitation signal. A check function is implemented to control
if the edited waveform matches the hardware specifications of the Basic Unit and FE-Probe
Head. A user defined waveform file (file extension tfw) can be loaded in the Hysteresis and
Piezo measurement procedure. Press Load beside the Waveform selection menu and select a
saved waveform file. The defined waveforms will appear as additional entries in the selection
menu. Please note that the recorded data depend on the polarization state of the sample which
maybe changes if a waveform is applied for a second time or if the Autorange function has been
carried out.
3.1.4 Settings dialog box
Program specific adjustments can be reached via the menu entry Settings or the button with
the hammer symbol in the tool bar. They are grouped under the topics Measurement, Compen-
sation, Default, and View. Some of these fields maybe grayed and cannot be selected. This
depends on the user specific configuration of the TF Analyzer 2000 system, e.g. if it includes
the high speed or high voltage option with an external amplifier or not.
In the Measurement dialog box (see Fig. 3.1.8) the choice Data Aquisition allows to switch be-
tween internal data aquisition with the FE Probe Head and external aquisition with a connected
oscilloscope for high speed measurements. Please note that leakage, leakage compensation,
static, CV, and piezo measurements cannot be performed in external mode. Please refer to
Chapter 2.5 how to connect and set up the system for measurements with an oscilloscope.
The following three fields are used to enable and select external hardware components which
can be connected to the TF Analyzer 2000 system. This include for example high voltage
amplifiers, switchboxes, or temperature controllers. Please refer to Chapter 2.3 how to connect
and set up your system for high voltage measurements properly.
The Piezo Factor in the field Piezo Settings is used to adapt the analog output voltage of a
connected displacement sensor to the change in length of the sample. Its unit has to be in
nm / V . For supplementary information on piezo measurements please refer to Chapter 2.4
and Chapter 3.6.
-
34 3 Measuring with the TF Analyzer
Fig. 3.1.8: Dialog box to apply measurement specific settings.
Be aware that the parameter ranges of all measurement types described in the following chapters
may be influenced by the software settings and hardware set-up of the TF Analyzer. x
For a general overview refer to the following Table 3.1.1 for some important dependent param-
eter ranges.
-
3.1 General program overview 35
Table 3.1.1: Examples for dependent input parameters. These range limits may further dependon system specific settings, software changes or hardware set-up.
(1) Frequency 1 Hz to 2 kHz standard frequency range of the excitationsignal
1 Hz to 200 kHz high-speed (enhanced FE) option frequencyrange (external mode)
.001 Hz to 2 kHz high-voltage frequency range, dependent onhigh-voltage amplifier
1 Hz to 20 MHz frequency range of the fatigue signal (loadand amplitude dependent)
(2) Pulse Widthor Rise Time
125 s to 1 s standard pulse width range
1 s to 1 s high-speed (enhanced FE) option pulsewidth range (external mode)
125 s to 1 s high-voltage pulse width range, dependenton high-voltage amplifier
25 ns to 1000 s pulse width range of the write pulse (load de-pendent)
(3) Amplitude 0.05 V to 25 V
standard voltage range for frequencies below100 kHz
0.02 V to 12 V
voltage range for frequencies above 100 kHz
10 V to 4000 V
high-voltage (here 4 kV option) range of theexcitation signal, dependent on high-voltageamplifier
The Compensation dialog box is described in the following chapter 3.1.5. In the Default dialog
box shown in Fig. 3.1.9 the user has the possibility to select the input parameter file in which his
specific measurement settings will be saved. Beside the aixACCT default settings several more
settings e. g. for high speed or high voltage can be defined. The selected set will be activated
after the Settings dialog box is closed with the OK button. A new parameter set can be defined
with the New Set dialog box. With save active parameter the currently choosen parameter from
the Measurement dialog are saved into this new set.
A file can be specified with multi sample data information for the use of the TF Analyzer 2000
systemtogether with a switchbox. Use the Browse button to select a *.msd file. Please refer to
Chapter 2.6 for more information on the use of a switchbox.
-
36 3 Measuring with the TF Analyzer
Fig. 3.1.9: Setting of user defined input parameter sets and the selection path and file name formulti sample data information.
The last dialog box called View allows the user to make changes regarding the graphical ap-
pearance of the measurement data (see Fig. 3.1.10). This includes the definition of symbols to
display the points of measurements (e.g. in fatigue or retention measurements). The size of
these symbols and the thickness of the data curve can be influenced with the dot size parameter.
Also, there are two more checkboxes in this dialog. The first checkbox classic graph style tog-
gles between the new graph display style with axes at the left, right and bottom of the diagram
and the one used in previous versions of the hysteresis software. The second one changes the
default for the checkbox Append to file. A change of this checkbox will take place after closing
and reopening the measurement dialog.
-
3.1 General program overview 37
Fig. 3.1.10: Adjustments for the graphical view and data representation.
3.1.5 Compensation
The Compensation menu entry in the Options menu is included to give a quick access to the
Settings - Compensation menu (Fig. 3.1.11) described hereafter.
There are three different types of compensation methods implemented in the hysteresis soft-
ware which can be enabled in the compensation dialog box (see Fig. 3.1.11). The first one is
the compensation of the parasitic capacitance of the measurement setup. Please perform an
open measurement with open contacts by clicking on the do open measurement button. Option-
ally, the open measurement can be exported to make the compensation data available for data
processing in other programs.
The second compensation method is a frequency compensation, which is especially of interest
for high frequency measurements. With this method the decrease of amplitude and phase of the
current amplifier at high frequencies will be compensated.
Third is the leakage compensation which subtracts the leakage current from the current response
of the sample. This is only necessary with samples having high leakage. Frequency and leakage
compensation will be automaticly applied to the measured data, if the check-boxes are enabled.
An optional compensation method is the in-situ parasitic capacitance compensation (lower part
-
38 3 Measuring with the TF Analyzer
of Fig. 3.1.11). It is similar to the capacitance compensation but performed by hardware. The
compensation can be adjusted automatically according to the open capacitance of the set-up by
hitting the automatic compensation button. Therefore the set-up should be contacted to an open-
contact structure or should be set-up close to the real device open capacitance. The adjustment is
done for each current range separately, so it takes a few seconds. Afterwards, the compensation
can be adjusted manually according to the actual measurements as shown below in Fig. 3.1.12
to Fig. 3.1.14. During hysteresis measurement, a measurement can be adjusted and taken by
hitting the test-button, this will start another hysteresis acquisition and show the change. For
an undercompensated loop (Fig. 3.1.12) the compensation should be increased (move coarse or
fine slider to the right), for an overcompensated loop (Fig. 3.1.13) the compensation should be
decreased (move slider left) until the loop is ok (Fig. 3.1.14). A very strong overcompensation
may even lead to oscillation of the amplifier and no loop will be recorded. The fine slider equals
to approx. 0.05 fF per step. The adjustment has to be repeated, if the current range is changed
in hysteresis measurement. The settings are stored and saved between program starts. Once
adjusted, the compensation is automatically performed on all measurements.
Fig. 3.1.11: Dialog box to fix the compensation methods which are applied in the measurements.This dialog box can directly be accessed from the Options menu, too.
-
3.1 General program overview 39
Fig. 3.1.12: under compen-sated hysteresis loop
Fig. 3.1.13: over compen-sated hysteresis loop
Fig. 3.1.14: well compen-sated hysteresis loop
3.1.6 Description of the graph windows
Now a description of the graph windows is given, which are used to display the measurement
data. It is possible to open several graph windows even of different measurement types. So,
various measurements of different type performed on one sample or measurements on differ-
ent samples can easily be compared. A typical graph view window of a Dynamic Hysteresis
Measurement is displayed in Fig. 3.1.15.
On the right hand side of the window below the aixACCT logo informations can be found about
the measurement type and conditions. The first is the acronym of the type of measurement dis-
played (see Table 3.1.2 below). Further data are the measurement frequency, the amplitude, and
the used current range. Data for an optional opened compare graph are displayed in brackets.
If more than one curve is displayed a legend specifying the colors of the curves can be found
below the diagram.
-
40 3 Measuring with the TF Analyzer
Fig. 3.1.15: Open graph view window of a hysteresis measurement.
Table 3.1.2: Abbreviations for the measurement procedures and their data file extensions
Measurement Procedure Acronym File exten-sion
Dynamic Hysteresis Measurement DHM tfhPUND Measurement PM tfpStatic Hysteresis Measurement SHM tfsLeakage Measurement LM tflPiezo Measurement PZT tfzCV Measurement CVM tfcThermo Measurement THM tftFatigue Measurement FM tffRetention Measurement RM tfrImprint Measurement IM tfiAccess Time Measurement ATM tfq
Directly below the graph are three fields for sample data and characteristic measurement values.
The Sample Data field on the left contains data like the sample name, area, thickness. A double
click with the left mouse button with the mouse-pointer located in this field opens the Sample
data dialog box, which allows the user to change name, thickness and area of this particular dis-
-
3.1 General program overview 41
played measurement. A change of the sample area or thickness is only possible in Hysteresis,
Pund, and Piezo measurements. For these the polarization and field values will be recalculated
according to the changed area or thickness values. This functionality is not available for mea-
surements which have been performed with a Hysteresis software version lower than 1.43. By
default the third field at the right is empty. A double click with the left mouse button with the
mouse-pointer located in this field opens the Index Dialog box, which allows the selection of
additional measurement information. One possible selection is the entry named Setting. The
following abbreviations are used to represent special measurement conditions:
Table 3.1.3: Abbreviations used for special measurement settings
Acronym DescriptionNP no prepolarisation pulseSL single loop measurementExt Mon measurement with oscilloscope and voltage monitoringHV Mon high voltage measurement with voltage monitoringUP unipolar voltage amplitudeFC measurement with frequency compensation enabledLC measurement with leakage compensation enabledCC measurement with capacitance compensation enabled
At the top of the graph window a digit is displayed (behind the text Graph). This number
identifies the number of the shown graph out of a set of consecutively performed measurements
of the same type. All the data of one measurement cycle is stored in the same file.
The scroll bar below the preview window allows the user to switch the displayed graph if there
are multiple graphs stored in the measurement. If a compare graph has been opened an ad-
ditional scroll bar appears to select the required data separately. With the X-Axis and Y-Axis
selection boxes at the bottom left it is possible to switch between the displayed data of one
measurement, e.g. the hysteresis of the polarization and the corresponding current response.
This function is also available by pressing the right mouse button with the mouse pointer in the
preview window. There are additional functions to copy, delete, and export the displayed mea-
surement. Copy and Copy graph only are used to add the graph to the clipboard. From there
the graph can be added with Paste to another measurement or included as graphic into other
programs. For the function Copy graph only the clipboard graphic contains only the diagram
without Measurement Data, Sample Data, and Characteristic Values.
The Contact Select selection box at the bottom right of the graph view window is only available
if a switchbox is enabled. Then one sample of a multisample measurement can be selected by
choosing its contact pair.
-
42 3 Measuring with the TF Analyzer
A special feature in Fatigue, Retention, Imprint, and Access Time graph views is the possi-
bility to view the single measurement graphs by a double click on the dots representing the
measurements. To visualize the measurement dots please refer to Fig. 3.1.10. Furthermore, an
additional entry Properties exists for these measurements in the menu which can be accessed by
a right mouse click with the mouse pointer on the graph. With this selection the values which
are displayed as measurement results can be choosen.
3.1.7 Data Formats
Measurement data can be saved in two different formats. Data files of the TF Analyzer 2000
system program use a specific TFA data file format which allows quick access to the data.
The file extensions of the TF Analyzer 2000 system files consist of three characters tf?. The
questionmark in the file extension is a wildcard for a character which represents the different
measurement procedures. (see Table 3.1.2). The acronyms in the table are used in the graph
windows to quickly identify the displayed data.
Hysteresis files (extension tfh) and PUND files (tfp) can be loaded into an already open mea-
surement with the compare graph function mentioned above. This is also possible for Hysteresis
or PUND files in Fatigue measurements.
The second format is used to export measurement data as an user readable ASCII-file (menu
choice File - Export). The file extension in this case is dat. ASCII data files can be imported into
any data presentation program. A detailed description of the contents of the ASCII file format
can be found at the end of each measurement procedure description. Ahead of each data section
text lines are included which describe the type of data, sample data and measurement conditions.
Data values are separated by tab characters. In Hysteresis or PUND measurements characteristic
result data (like the relaxed remanent polarization values in a Hysteresis measurement) which
are extracted from the measured data can be saved, too.
Beside other measurement data and characteristic values the current range of the measurement
is also included as a number after the tag name Current Range. The following table Table 3.1.4
shows the corresponding ranges.
-
3.1 General program overview 43
Table 3.1.4: Values corresponding to current ranges
Value Current Range0 1 A1 100 mA2 not used3 not used4 10 mA5 1 mA6 100 A7 10 A8 1 A9 100 nA
10 10 nA11 1 nA12 100 pA13 10 pA
3.1.8 Multi-sample measurements with a switch box (optional)
The multi sample measurement feature of the TF Analyzer 2000 system offers automatic testing
of many samples in parallel using a switch box and probe card system. The differences to the
standard software are quite few.
The first major difference is the declaration of the sample data for each contact pair of the switch
box, which is identical to top and bottom contact of the capacitor under test. The information
for each sample has to be filled into the Multiple Sampledata window (see Fig. 3.1.16). This
is opened via the Sampledata entry in the Options menu when a switch box is enabled in the
Options - Settings dialog box.
Equivalent to the single measurements described before sample name, area, and thickness have
to be filled into the edit lines. This declaration is necessary to calculate the polarization values
of each sample in consideration of their electrode area. E.g. on a wafer test capacitors with
different top electrode areas may be arranged. So, the sample data has to fit to the device
under test to calculate the true polarization value for each sample. In addition for multisample
measurements top and bottom electrode have to be selected, which are connected to the physical
switches of the switch box. The relation between the mnemonic name of the top and bottom
contact to the physical switches of the switch matrix is established by the wafer mapping file
(switches.ini, see Chapter 2.6). In case of names for top and bottom contact which are not
defined in the wafer mapping file, Invalid appears next to the edit line. In case there is no
information in the wafer mapping file the user has to fill in the physical switch numbers of the
-
44 3 Measuring with the TF Analyzer
Fig. 3.1.16: Input window for the declaration of a multiple sampledata contact list.
switch box in the edit line of the top contact and bottom contact.
Usually, for a specific layout configuration and probe card, the wafer mapping file and the
multisample data is stored, as it is fix for a certain layout on the wafer. To edit the contact
pair configuration the user can double click on any line of the text field. The configuration of
this contact pair will be displayed in the edit lines above. Now, any modification of the contact
configuration can be established by the help of the push buttons explained below:
Add: If a new contact configuration is generated, it will be added to the list by pressing the
Add button. The new contact configuration is added at the bottom of the list.
Insert: If you want to insert a new contact configuration at a certain position of the list, use
the Insert button. It will be included above the highlighted contact configuration in the
list.
Remove: The selected contact configuration in the list will be deleted.
Replace: In this case the generated contact configuration will replace the selected contact
configuration in the list.
Once the configuration is defined, it can be saved and reloaded at any time with Save and Load.
-
3.1 General program overview 45
For multisample measurements current ranges need to be specified for every sample beside the
usual measurement parameter. This can be done automatically with the Autorange button in
the measurement dialog or manually by applying a current range using Edit. The current range
selection field in the measurement dialog is disabled in multisample measurements.
The appearance of the graph windows with an enabled switch box is almost the same as in
the case of single sample measurements. The difference is, that the selction box Contact Se-
lect beside the Graph Select box is enabled for Fatigue, Retention, Imprint, and Access Time
measurements. This enables the user to select between the measured samples.
For the four measurements types mentioned above the following data management is used.
After one of these measurement procedures the data of each contact configuration will be saved
in separated files. The text field in the sampledata menu mentions the file names under which
the recorded data is stored. The filename consists of the given filename which is equal for
all contact configurations and is followed by the sample name and the top and bottom contact
notation. Each name element is separated by underscores respectively a minus between top and
bottom. The filename extension (see Table 3.1.2) is added to indicate which data are stored in
the file. So, a typical filename for fatigue data will look like:
filename_samplename_top-bottom.tff
All the data of a multisample measurement will be saved in an automatic created subfolder to
the actual folder. This subfolder is named identical to the given filename. Additionally, a file
named filename.tfm is created, which contains the names of all files of this measurement. This
file can be used to reload the data of the whole measurement. But, single measurements can be
loaded out of the subfolder too.
For the other measurement procedures the measurement data of the multiple samples will be
safed consecutive in one file. The different measurements can be viewed with the scroll bar
below the graph.
-
46 3 Measuring with the TF Analyzer
-
3.2 Dynamic Hysteresis Measurement - DHM 47
3.2 Dynamic Hysteresis Measurement - DHM
The dynamic hysteresis program allows a comfortable and comprehensive investigation of the
characteristic values of a hysteresis loop in a frequency range from 1 Hz to 1 kHz. The cor-
relation of several process parameters to the shape of the hysteresis loop can be examined. In
addition, measurement parameters such as excitation signal, frequency, and amplitude of the
excitation signal can be varied.
3.2.1 Measurement procedure and typical measurement
The voltage excitation signal is shown in Fig. 3.2.1 in the case of triangular voltage excitation.
Fig. 3.2.1: Excitation signal for hysteresis measurement.
The prepolarization pulse establishes a defined polarization state, the negative state of relaxed
remanent polarization. The prepolarization pulse is followed by three consecutive bipolar exci-
tation signals, each signal separated by 1 second relaxation time. The corresponding hysteresis
loops to the bipolar excitation signal are shown in Fig. 3.2.2. The hysteresis loop correspond-
ing to pulse no. 1 starts in the negative relaxed remanent polarization state (Prrel) and turns
into the positive saturation (Pmax+). When the voltage is equal to zero the polarization reaches
the positive remanent polarization state (Pr+), afterwards it turns into the negative saturation
(Pmax) and back to the remanent polarization state (Pr). This point is normally not equal to
the starting point (Prrel), because of the polarization loss over time.
The second loop establishes the sample into the positive remanent polarization state without
sampling data. The third loop starts in the positive relaxed remanent polarization state (Prrel+),
turns into the negative saturation (Pmax) crosses the polarization axis at zero volts excitation
signal in the negative remanent polarization state (Pr). Afterwards the sample is driven into the
positive saturation (Pmax+) and ends in the positive remanent polarization state (Pr+) when the
voltage is zero. Afterwards the hysteresis loop is balanced respectively to the values P (+Vmax)
and P (Vmax). From the data of the first loop the parameters Vc, Pr, Prrel are determinedand from the data of the third loop the parameters Vc+ , Pr+ , Prrel+. The closed hysteresis loop
-
48 3 Measuring with the TF Analyzer
Fig. 3.2.2: Typical hysteresis measurement graph.
(red curve) is calculated from the second half of the first (green curve) and the second half of
the third loop (blue curve).
3.2.2 The characteristic parameters of the hysteresis loop
In the following the nomenclature of characteristic values for the evaluation of the measured
data is introduced (see table below and Fig. 3.2.3).
Pr+ positive state of remanent polarization of the dynamicallymeasured hysteresis loop
Prrel+ positive state of relaxed remanent polarization, relaxedfor one second in the Pr+ state. Equal to the positive stateof remanent polarization of the quasi statically measuredloop (see Chapter 3.4)
Pmax+ state of polarization when the stimulating signal reaches itsmaximum value - positive saturation
Vc+ positive coercive voltage, voltage where the polarizationcrosses the x-axis by increasing voltage values
Pr, Prrel, Pmax, Vc are the corresponding values on the negativefield and polarization direction
Psw (Pmax Prrel) change of polarization when the sampleis switched from the negative state of the relaxed remanentpolarization into the positive saturation - switching case
Pnsw (Pmax Prrel+) change of polarization when the sample isdriven into the positive saturation from the positive state ofthe relaxed remanent polarization - non-switching case
dPsw (Psw Pnsw) detectable polarization difference betweenswitching and non-switching case
-
3.2 Dynamic Hysteresis Measurement - DHM 49
Fig. 3.2.3: Nomenclature used within the TF Analyzer 2000 system.
3.2.3 Execution of a hysteresis measurement
In the following table the input parameters of a dynamic hysteresis measurement are listed
including their posssible ranges and values and a short description:
Input Parameters
(1) Frequency edit line 1 Hz to 2 kHz frequency of the excitation sig-nal, (refer also to Chap. 3.1.4)
(2) Waveform selection box triangle, sineor manual
waveform of the excitation sig-nal
(3) Amplitude edit line 0.1 V to 25 V
amplitude of the excitation sig-nal (refer also to Chap. 3.1.4)
(4) No. of points edit line 20 to 1000 number of data points measuredfor one hysteresis loop
(5) No. of aver-ages
edit line 1 to 100 number of measurement cyclesto build the averaged loop
(6) Single loop check box performs a measurement withonly one pulse (instead of three).With No. of averages > 1 therewill be only one prepol pulse.
(7) No prepolpulse
check box performs a measurement withoutprepolarisation pulse (no averag-ing is possible with this optionactivated)
-
50 3 Measuring with the TF Analyzer
For a general program overview and how to start the hysteresis measurement procedure please
refer to chapter 3.1. To start a hysteresis measurement with values of your choice act as follows:
Check the sample data which you find in the menu choice Options - SampleData. The
window coming up offers three edit lines. Here the user can fill in the sample name,
the area (pad size) and the thickness of the sample. Move the cursor on an edit line you
want to change, click the left mouse button and enter the value of your choice. Press OK
to close this window by saving the changes. The sample data will be shown in a graph
window together with the characteristic values after a measurement.
Now, open the Measurement Dialog and select the card indexed Hysteresis. Check the
values in the box Input Parameters. Move the cursor to an edit line and click on the
left mouse button if you want to change one of these values. The Waveform selection
box enables to select the excitation signal. Click on the right corner of the box and the
selectable values are displayed. As standard excitation triangular and sine signals are
provided.
In the Current Amplifier field a current range for the measurement can be selected. Press
the button AutoRange to start a software routine to automatically find the correct current
range. During this process the other input fields and buttons are disabled.
When the Start button is enabled again, press it to start the measurement. An empty graph
window opens up. In the status bar of the main window (line at the bottom) the status of
the program is displayed, in this case Measure...
Directly, after the measurement is completed the new recorded data are displayed in the
graph window.
If the hysteresis graph is closed, the software checks whether the measured data has been
saved or not. In the case of loosing unsaved data, a dialog box appears and asks the user
whether he wants to save the data are not.
3.2.4 ASCII-File of the Hysteresis Program
A hysteresis measurement can be saved as an ASCII data file with the File - Export function. In
case that the characteristic values have been saved too, the ASCII format of the DHM contains
two parts: The first part starting with the line HysteresisResult contains the characteristic values
which have been extracted from the measurement data. The second part starts with the line
DynamicHysteresis. It contains the measurement data. After the first line (HysteresisResult or
DynamicHysteresis) four more text lines follow which describe the program version, the type
-
3.2 Dynamic Hysteresis Measurement - DHM 51
of data which follow (result or data), the TFA-File version, and a time stamp at which time the
file has been created.
After this header section follows a blank line and than the table data. Before the numeric data
there are a variable number of text lines describing the measurement conditions under which
the data where recorded. The first text line contains always the tag Table and the number of the
table, separated with a tabulator. The last text line before the numeric data describe the data
columns contents and their units. The denotation and meaning is listed in Table 3.2.1.
Table 3.2.1: Denotation of the columns of the DHM program
Column Denotation1 t Time2 V+ V+ Voltage (starting
positive)3 V- V- Voltage (starting
negative)4 I1 Current from closed
loop5 P1 Polarization from
closed loop6 I2 Current from pulse
No. 17 P2 Polarization from pulse
No. 18 I3 Current from pulse
No. 39 P3 Polarization from pulse
No. 3
The hysteresis module of the TF Analyzer 2000 system offers a P (V )- and a I(t)- representation
of the data. The I(t) curves display the measured current response of the sample versus time.
Polarization curves are combined from a total of three measurements. The polarization data
determined from the closed loop, the data determined when starting from the negative remanent
polarization state and the data acquired after starting from the positive remanent polarization
state. They all can be displayed versus the excitation voltage.
To display the curve of the current response vs. time column 1 is assigned to the x-axis and col.
4, 6, or 8 is assigned to the y-axis. To display the polarization of the measurement you have
to assign V+ to the x-axis and column 5 or 7 to the y-axis. To display the P (V ) of the curve
starting in the positive relaxed remanent polarization state assign V to the x-axis and col. 9 to
the y-axis.
-
52 3 Measuring with the TF Analyzer
A clipping of the ASCII-file of the DHM is shown below.
HysteresisResultProgram: Hyster 1.44BasicUnit: BU 313TfaFileType: resultTfaVersion: 3.0TimeStamp: 10/29/2002 17:41:33Table 1Volt [V] Freq[Hz] Pr+[uC/cm2] Pr-[uC/cm2] Prrel+[uC/cm2] Prrel-[uC/cm2] Pmax[uC/cm2] Vc+[V]Vc-[V] Area [mm] Current Range[RESULT DATA, 11 columns]
DynamicHysteresisProgram: Hyster 1.44BasicUnit: BU 313TfaFileType: dataTfaVersion: 3.0TimeStamp: 10/29/2002 17:41:33
Table 1Timestamp: 10/29/2002 17:41:33Averages: 1SampleName: PZTThickness [nm]: 160Area [mm2]: 0.125Current Range: 5Hysteresis Frequency [Hz]: 100Hysteresis Amplitude [V]: 3Vc+ [V]: 1.56671Vc- [V]: -1.20417Pr+ [uC/cm2]: 36.3141Prrel+ [uC/cm2]: 35.5749Prrel- [uC/cm2]: -36.7047Pr- [uC/cm2]: -36.7978Pmax [uC/cm2]: 43.7104Wloss [uJ/cm2]: 0Psw [uC/cm2]: 80.4151Pnsw [uC/cm2]: 8.1355dPsw [uC/cm2]: 72.2796CompensateFrequency: 1CompensateLeakage: 1Waveform: 0CompensateCapacitance: 1Settings: FC LC CCTime[s] V+[V] V-[V] I1[A] P1[uC/cm2] I2[A] P2[uC/cm2] I3[A] P3[uC/cm2][DATA, 9 columns]
-
3.3 Pulse Measurement - PM 53
3.3 Pulse Measurement - PM
3.3.1 Measur