DIVA-DDIVA-DDIVA-DDIVA-D 0123012301230123Digital Imaging Video Acquisition
with full DICOM compatibility
User’s manualUser’s manualUser’s manualUser’s manual
-
8/15/2019 DIVA-D (User's Manual)
2/58
USER’S MANUAL Contents
(Rev. 0) DIVA-Di
Contents1 INTRODUCTION 1
1.1 Icon appearing in the manual................................................................. 1
2 SAFETY ASPECT 2
2.1 Warning.................................................................................................. 3
2.2 Environmental risks and displacement................................................... 5
2.3 Symbols used ......................................................................................... 6
3 DESCRIPTION 73.1 Identification labels ................................................................................ 7
3.2 Description ............................................................................................. 8
3.2.1 DIVA-D composition ................................................................... 9
3.2.2 Monitors..................................................................................... 9
3.2.3 Accessories................................................................................. 9
4 TECHNICAL DATA 10
4.1 DIVA-D AS 25....................................................................................... 10
4.2 CRT Monitor ......................................................................................... 15
4.3 LCD Monitor ......................................................................................... 17
4.4 Environmental conditions..................................................................... 18
4.5 Standards............................................................................................. 18
5 KEYBOARD AND FUNCTIONS 19
5.1 Functions description ........................................................................... 20
6 UNIT SWITCHING ON 22
6.1 CCD camera not connected .................................................................. 24
6.2 System fault ......................................................................................... 25
7 PATIENT DATA 26
8 FLUOROSCOPY 28
8.1 Fluoro special function ......................................................................... 30
9 MODALITY SELECTION 31
10 GRAPHY 33
-
8/15/2019 DIVA-D (User's Manual)
3/58
USER’S MANUAL Contents
DIVA-D (Rev. 0)ii
11 POST PROCESSING 35
11.1 Special function.................................................................................... 39
11.2 Multi image........................................................................................... 41
11.3 Image print ........................................................................................... 42
12 DICOM FUNCTIONS 43
12.1 Waiting list ........................................................................................... 44
12.2 Queue view and flush ........................................................................... 45
12.3 Print ..................................................................................................... 47
12.4 Store..................................................................................................... 49
12.5 CD - ROM ............................................................................................. 51
13 MAINTENANCE AND CLEANING 54
13.1 Frequency of maintenance and cleaning actions ................................... 54
This publication can only be reproduced, transmitted, transcribed, or translatedinto any human or computer language with the written consent of VILLA SISTEMIMEDICALI S.p.a.
This manual in English is the original version.
-
8/15/2019 DIVA-D (User's Manual)
4/58
USER’S MANUAL Introduction
(Rev. 0) DIVA-D1
1111
INTRODUCTIONINTRODUCTIONINTRODUCTIONINTRODUCTION
NOTE: The present manual is updated for the product it is sold with in order togrant an adequate reference to use properly and safely the product.
The manual may not reflect changes to the product not impactingoperating modes or safety.
DIVA-D is an acquisition, storing and processing digital system studiedto process digitalized images.
The purpose of this manual is to give the User instructions for a safe andefficient use of the DIVA-D.
The device must be used in compliance with the procedures described inthe user manual and never used for purposes different than the onesdescribed in the user manual.
Qualified personnel only, having necessary knowledge on X-ray protection can use the device.
1.11.11.11.1 Icon appearing in the manualIcon appearing in the manualIcon appearing in the manualIcon appearing in the manual
This icon indicates a NOTE: please read thoroughly the items marked by this picture.
This icon indicates a WARNING message; the items marked by this iconrefer to the safety aspects of the patient and/or of the operator.
-
8/15/2019 DIVA-D (User's Manual)
5/58
USER’S MANUAL Safety aspect
DIVA-D (Rev. 0)2
2222
SAFETY ASPECTSAFETY ASPECTSAFETY ASPECTSAFETY ASPECT
WARNING: Please read this chapter thoroughly.
VILLA SISTEMI MEDICALI devices comply with the related safety requirements; furthermore it supplies all information necessary for a correct use and the warnings related to danger associated with X-raysgenerating units.
VILLA SISTEMI MEDICALI, has not to be held responsible for:
• use of DIVA-D different from the intended one as described in thismanual;
• damages to the unit, to the operator, to the patient, caused either by wrong installation or maintenance procedures, different from thosedescribed in the Servcie manual and in the user manual supplied
with the unit, and by wrong operations;
• mechanical and/or electrical modifications performed during and
after the installation, different from those described in the officialdocumentation of the product.
Any technical intervention must only be performed by VILLA SISTEMI MEDICALI qualified technicians.
Only the authorized personnel can remove the covers and/or haveaccess to the components under voltage.
-
8/15/2019 DIVA-D (User's Manual)
6/58
USER’S MANUAL Safety aspect
(Rev. 0) DIVA-D3
2.12.12.12.1 WarningWarningWarningWarning
This device has not been designed to be used in environments where vapors, anaestethic mixes flammable with air, or oxygen and nitrousoxide can be detected.
Avoid pouring water, even accidentally, or other liquids into the device,as this could cause short-circuits.
Before cleaning the device, please disconnect it from the line voltage.
Wherever necessary, use the fit accessories, such as the leaded aprons,to protect the patient from radiation exposure.
While performing the radiography, no one, apart from the operator andthe patient, must remain in the room.
Though this unit has been designed with a quite acceptable protectionlevel from electromagnetic interference, it is advisable to install it at a certain distance from electrical energy transformation rooms, from staticcontinuity units, from portable receiving-transmitting units for amateurial use and for cellular use. Cellular telephones are only admitted at a distance of more than 1.5 m from any component of thedevice.Other medical instruments and devices that must be used in the sameinstallation area of the unit must comply with the ElectromagneticCompatibility rules in force. Non-complying instruments, of which thepoor immunity from electromagnetic fields is well known, must beinstalled at least 3 m away from the DIVA-D and supplied by a different electrical line.
DIVA-D cannot be installed in operating theatres.DIVA-D must be off while using devices such as electrical scapels or similar instruments.
The equipment automatically performs some self-test functions andtherefore is able to detect some anomalous conditions. However,
whenever there are doubts on the performance of the equipment (e.g.images too clear, too dark or with artifacts), the activity must be
suspended and a Villa Sistemi Medicali qualified and authorized engineer must be contacted.
WARNING:If during the self-test phase the equipment shows a failure messagecaused by the electronic circuits or the Ram disks, the images of a givenpatient retrieved for post processing may not belong to the requestedpatient.In this case, immediately contact an authorized Villa Sistemi Medicaliengineer and do not make any diagnosis or evaluation on the storedimages.
In addition, as the failure may have occurred during the last workingsession, diagnosis carried out after the last exam cannot beconsidered valid.
-
8/15/2019 DIVA-D (User's Manual)
7/58
USER’S MANUAL Safety aspect
DIVA-D (Rev. 0)4
During the use of the equipment, the operator must comply with the
local work safety regulation, in particular with the ones referring to theuse of video terminals.
The cathode X-ray tube (CRT) of the monitors, when present, ismanufactured in order to minimize the risk of implosion. However, if heavily hit it may break thus projecting out pieces of glass, powder andtoxic gas. Avoid hitting the CRT with any object and make sure that it does not hit any object.
These risks are not present when LCD monitors are used.
-
8/15/2019 DIVA-D (User's Manual)
8/58
USER’S MANUAL Safety aspect
(Rev. 0) DIVA-D5
2.22.22.22.2 Environmental risks and displacementEnvironmental risks and displacementEnvironmental risks and displacementEnvironmental risks and displacement
The DIVA-D contains in some of its parts, materials that at the end of theunits life, must be recycled at an approved location.Particularly the device contains the following materials and/or components:
• Central unit and control console: iron, copper, aluminum, non- biodegradable plastic material, glass-resin for printed circuit board(PCB)
• CRT Monitor, TV camera: glass vacuum tubes, iron, copper,
aluminum, glass-resin for printed board, non-biodegradable plasticmaterial.
WARNINGGlass vacuum tube may implode with projection of glass pieces, powder and toxic gas.
-
8/15/2019 DIVA-D (User's Manual)
9/58
USER’S MANUAL Safety aspect
DIVA-D (Rev. 0)6
2.32.32.32.3 Symbols usedSymbols usedSymbols usedSymbols used
Symbol Description
Warning, read the relevant documentation
The equipment contains in some of its parts,materials that at the end of the units life, must berecycled at an approved location
∼ A.C.
0123Complies with Medical Equipment Directive 93/42
Device with type B applied parts
Dangerous voltage
-
8/15/2019 DIVA-D (User's Manual)
10/58
USER’S MANUAL Description
(Rev. 0) DIVA-D7
3333
DESCRIPTIONDESCRIPTIONDESCRIPTIONDESCRIPTION
3.13.13.13.1 Identification labelsIdentification labelsIdentification labelsIdentification labels
1
Main label (rack)
2
Control keyboard label
3
CRT Monitor label
4
LCD Monitor label
1
2
4
3
-
8/15/2019 DIVA-D (User's Manual)
11/58
USER’S MANUAL Description
DIVA-D (Rev. 0)8
3.23.23.23.2 DescriptionDescriptionDescriptionDescription
DIVA-D is a digital acquisition, storing and processing system for radiological images. The large storing capacity allows performing thefollowing functions:
• Image acquisition and storing in FLUOROSCOPY
• Image acquisition and storing in EXPOSURE with the following working programs: SPOT, BONE, ANGIO, PERIF, STEP, TOMO
• Post-processing of the acquired images, thanks to a dedicatedprogram
• Printing of the acquired images through a laser printer.
• Sending via Dicom network acquired images to Pacs system.
The images coming from the image intensifier are acquired through a CCD camera, then fed to the digitalization stage (A to D Converter) after that they can be stored in the permanent memory (Ram disk).From this moment on the images can be processed thanks to thededicated software programs.
The system has a large accessible database, which allows for the entry and retrieval of patient information along with exam description, remark and notes.
DIVA-D is available in 2 versions:
DIVA-D RF 12
Ram Disk capacity 36 Gbytes
Operator monitor 1 or 2 on request
"In room" monitor 0, 1 or 2 on request
Imaging acquisition (max speed andmatrix)
12 imag/sec matrix 1024x1024
12 imag/sec matrix 1024x1024
Functions • standard acquisition: YES
• imaging subtraction: NO• road mapping: NO
• fluoro store: YES
DIVA-D AS 25
Ram Disk capacity 72 Gbytes
Operator monitor 1 or 2 on request
"In room" monitor 0, 1 or 2 on request
Imaging acquisition (max speed andmatrix)
25 imag/sec matrix 1024x1024
25 imag/sec matrix 1024x1024
Functions • standard acquisition: YES
• imaging subtraction: YES
• road mapping: YES
• fluoro store: YES
-
8/15/2019 DIVA-D (User's Manual)
12/58
USER’S MANUAL Description
(Rev. 0) DIVA-D9
3.2.13.2.13.2.13.2.1 DIVA-D compositionDIVA-D compositionDIVA-D compositionDIVA-D composition
DIVA-D contains the following components:
• Rack containing all electronics (computer, PCB's, Ram disk, power supply, etc).
• Control keyboard.
• CCD TV camera.
• POD (Printed Ouput/input Device) pcb interface for table andgenerator.
• Laser interface for standard digital protocol (P831) when Dicommodule is not present only.
3.2.23.2.23.2.23.2.2 MonitorsMonitorsMonitorsMonitors
All DIVA-D models are available with CRT or LCD monitors for operatorsand "in room".
The size of these monitors depends on customer request.
3.2.33.2.33.2.33.2.3 AccessoriesAccessoriesAccessoriesAccessories
Depending on the request, DIVA-D can be supplied with accessories:
• DICOM module equipped with SW package for:
− print
− store
− worklist
− CD write
• Remote control function.
• Ceiling suspension for "in room" monitor.
-
8/15/2019 DIVA-D (User's Manual)
13/58
USER’S MANUAL Technical data
DIVA-D (Rev. 0)10
4444
TECHNICAL DATATECHNICAL DATATECHNICAL DATATECHNICAL DATA
NOTE: Technical data, performance and instructions provided in this manualare referred to the DIVA-D AS 25 model. Therefore, some of thesecharacteristics are not applicable for DIVA-D RF 12 systems.
4.14.14.14.1 DIVA-D AS 25DIVA-D AS 25DIVA-D AS 25DIVA-D AS 25
Identification
Type DIVA-D AS 25
Model NDR 632 E00 + NDR transf 02
Components
Central unit Rack composed by:Power supply; Computer; Pcb's; Ramdisk
Control system Flat control keyboard
Acquisition by CCD Camera 1024 x 1024
Interface system Generator and Remote Control Tableinterface PCB
Connection cables Prepared interconnection cables
Monitor Operator: 1 or 2 on request "In room": 1 or 2 on request
Dimension and weight Central Unit (Rack)
Weight 15 Kg
Height 254 mmDepth 345 mm
Width 440 mm
Dimension and weight Control keyboard
Weight 6 Kg
Height 50 mm
Depth 235 mm
Width 460 mm
-
8/15/2019 DIVA-D (User's Manual)
14/58
-
8/15/2019 DIVA-D (User's Manual)
15/58
USER’S MANUAL Technical data
DIVA-D (Rev. 0)12
DICOM module (optional) Technical characteristics
Processor management Intel ™
Working frequency 1 GHz
Ram memory 512 Mbytes
Transit memory Ram disk 40 Gbytes
Dicom images backing Ram disk 40 Gbytes
Network access speed 100 Mbit
SW package available -Print (1024x1024 16bit)-Store (1024x1024 16bit @ 3 Mbyte)-CD write (1024x1024 8bit)
(512x512 8 bit)-Worklist
Imaging acquisition (speed and matrix)
Mode 12 bits progressive scanning acquisitionand visualization
Fluoroscopy Matrix 1024x1024 Acquisition max rate25 fr/sec
Pulsed fluoroscopy Matrix 1024x1024 Acquisition selectable rate:
1-3-6-12-25 fr/sec
Exposure and angiography Matrix 1024x1024 Acquisition max rate25 fr/sec
Imaging memorization
Storage Automatic and permanent.
Maximum access time to disk for image write and read (with matrix 1024x102416 bit)
30 msec.
Memory capacity 72 Gigabytes
Maximum storage capacity of images
with matrix 1024x1024
36000 images
Electrical characteristics
Standard power supply (transf 02) 230 V ac ±15%
Standard power supply (DIVA-D rack) 24 V ac ±15%
Frequency 50-60 Hz
Current absorbed 650 mA @ 230 V ac
Power absorbed 150 W @ 230 V ac
-
8/15/2019 DIVA-D (User's Manual)
16/58
USER’S MANUAL Technical data
(Rev. 0) DIVA-D13
Working configuration
Remote R\ F control table - Apollo- Mercury 332
R\ F tilting table - Vision
Integration with generator Genius 65/80 HF-A with serial mode(RS 232) interface
Clinical application
Fluoroscopy Continuous and pulsed HCF
Standard exams and with contrast agent Gastroenterological, bone exams, ERCP,Urology, Mielography, etc.
Tomography In low dose modality
Vascular exams Thoracic angiography, carotid,peripheral, phlebography.Exams with on-line subtraction andpost processing.
Peripheral steps Angiography Optional. Possible application with Apollo remote control table only.
Working characteristics
User interface Windows representation
Fluoroscopy mode
Storage Manual start on Ram disk
Regulation of the parameter Auto video level regulation. Auto radiological parameter regulation.Manual selection of six field video levels
Edge enhancement Two selectable levels
Positive/Negative function (grey scaleinversion)
In line/off line
Noise reducer Two selectable level
Road mapping Direct maskingImaging comparison Automatic or manual transfer of the
image
Last image hold (LIH) With integrated image
Manual regulation of the video level Three selectable levels
Motion detection filter Automatic exclusion of the recursivefilter during the movement of the image(no blurring effect)
Dynamic recursive filter DRF Automatic noise/blurring optimizationeffect offers zero persistence
-
8/15/2019 DIVA-D (User's Manual)
17/58
-
8/15/2019 DIVA-D (User's Manual)
18/58
USER’S MANUAL Technical data
(Rev. 0) DIVA-D15
4.24.24.24.2 CRT Monitor CRT Monitor CRT Monitor CRT Monitor
NOTE: The characteristics of the monitors are the same for both DIVA-D'smodel.
Technical data
Type Monitor CRT B&W
Manufacturer Nical
Model Refer to optional
Dimensions and weight Monitor 17” Monitor 21”
Weight 17,5 Kg 21,8 Kg
Height 335 (359) mm 394 (418) mm
Depth 350 mm 380 mm
Width 405 mm 489 mm
Screen dimensions 17" (414 mm) 21" (511 mm)
Electrical characteristicsStandard power supply 230 V ac ±15%
Frequency 50-60 Hz
Current absorbed 350 mA
Power absorbed 80 W
Scanning
Type Interlaced
Line number 1049
"Refresh" frequency 60 Hz
Standards EIA
Video input signal
Type Video input connected ac
Connection type BNC
Impedance 75 Ω
Input video signal 0,5 to 1,5 V pp (negativesynchronism)
-
8/15/2019 DIVA-D (User's Manual)
19/58
USER’S MANUAL Technical data
DIVA-D (Rev. 0)16
Display characteristics
Deflection angle 90°
Aspect ratio 4:3
Band width 40 MHz ±3 dB
Raster 0.25 mm
Phosphor White P45
Geometric error
-
8/15/2019 DIVA-D (User's Manual)
20/58
USER’S MANUAL Technical data
(Rev. 0) DIVA-D17
4.34.34.34.3 LCD Monitor LCD Monitor LCD Monitor LCD Monitor
NOTE: The characteristics of the monitors are the same for both DIVA-D'smodel.
Technical data
Type Monitor LCD B&W
Manufacturer Nical
Model Refer to optionals
Dimensions and weight Monitor 18” Monitor 20”
Weight 5,5 Kg 6,0 Kg
Height 340 mm 355 mm
Depth 60 mm 60 mm
Width 430 mm 457 mm
Screen dimensions 18" 20"
Electrical characteristicsStandard power supply 230 V ac ±15%
Frequency 50-60 Hz
Current absorbed 250 mA
Power absorbed 50 W
Display characteristic
Visibility angle 180°
Contrast 600:1
Resolution 1280x1024
Gray scale 768 levels
Max luminance 700 Cd/m2
Calibration automatic
Optional
Contrast and brightness control with digitalpush-button on front panel and rear remotecontrol connection
Available on request on all models
Contrast and brightness control only withrear remote control connection
Available on request on all models
-
8/15/2019 DIVA-D (User's Manual)
21/58
USER’S MANUAL Technical data
DIVA-D (Rev. 0)18
4.44.44.44.4 Environmental conditionsEnvironmental conditionsEnvironmental conditionsEnvironmental conditions
Operative Temperature: from -10° to +45° CelsiusHumidity: up to 95 % not condensing
Transport and storage Temperature: from -35° to +65° CelsiusHumidity: up to 95 % not condensing
4.54.54.54.5 StandardsStandardsStandardsStandards
DIVA-D system with accessories and optional complying with the relevant standards:
EN 60601-1Electro medical equipment. General safety requirement (included amendment A1+A2+A11+A12+A13)
EN 60601-1-2
Electro medical equipment. General safety requirement.2 – Collateral standards: Electromagneticcompatibility.
EN 60950 Information device safety
0123 The symbol CE grants that DIVA-D complies withEuropean directive 93/42
Further information regarding standards complying can be requested to:
VILLA SISTEMI MEDICALI S.p.A. Via delle Azalee, 320090 - Buccinasco - (MI) - Italia
Tel: +39-02-48.859.1Fax: +39-02-48.818.44
The manufacturer of the equipment DIVA-D is (according to the93/42/CEE European directive):
NICAL S.p.A. Via Soffredini, 4320126 - Milano - Italia
Tel.: +39-02-25.711.10Fax: +39-02-25.722.07
-
8/15/2019 DIVA-D (User's Manual)
22/58
USER’S MANUAL Keyboard and functions
(Rev. 0) DIVA-D19
5555
KEYBOARD AND FUNCTIONSKEYBOARD AND FUNCTIONSKEYBOARD AND FUNCTIONSKEYBOARD AND FUNCTIONS All functions of DIVA-D are activated through buttons located on a dedicated keyboard.
The following picture and relevant buttons functions descriptions permitsto locate easily all reference of the present manual.
-
8/15/2019 DIVA-D (User's Manual)
23/58
USER’S MANUAL Keyboard and functions
DIVA-D (Rev. 0)20
5.15.15.15.1 Functions descriptionFunctions descriptionFunctions descriptionFunctions description
To adjust the kV value when automatic mode isactive.It is possible to increase and decrease the imagelevel with XX steps.
R.O.I. selection. 6 different areas can be selected for the auto kV correction in function of dimension andposition of the image area to be optimized.
To store fluoroscopy images.
To enable the road mapping function. It is used assupport during catheter insertion.
To select the fluoro modality between continuousand pulsed.
To select the frame rate exposure acquisition.
When a new patient must be recorded the key permits to access the patient data window.Mode is used to enter into modality and anatomicexam selection.
To enter into patient data base to perform post processing mode.
In post processing mode, when a patient is recalledfrom data base, to select run into the patient andthe frame into run.
To play a run that contains more then one frameand to stop the sequence visualization.
-
8/15/2019 DIVA-D (User's Manual)
24/58
USER’S MANUAL Keyboard and functions
(Rev. 0) DIVA-D21
These keys allows to enter into different post
processing mode:− symmetric and asymmetric image limitation
− negative-positive
− zoom x4 with scroll
− cine loop review
− multi image runs selection.
These keys allows to enter into different post processing angiographic mode:
− mask shift with 1/8 pixel step
− mask present or removed
− manual mask selection
− remasking.
To modify images quality in post processing:
− manual windowing
− edge enhancement
− auto windowing
− reset of the modifications.
To access into fluoroscopy or post processingspecial function.
To enter into DICOM (optional) modality, to savedata into DICOM module and to print or store thecaptured images.
This group of keys are used as auxiliary functionsduring the different modality.
Alphanumeric keys used to write characters indifferent working modalities.
This manual is divided in different chapters, each of one is dedicated to a working modality which DIVA-D supports.In any chapter it is explained how to use the buttons located on thekeyboard to optimize the DIVA-D use.
-
8/15/2019 DIVA-D (User's Manual)
25/58
USER’S MANUAL Unit switching ON
DIVA-D (Rev. 0)22
6666
UNIT SWITCHING ONUNIT SWITCHING ONUNIT SWITCHING ONUNIT SWITCHING ON Any time that DIVA-D is switched on, it performs an auto test to verify if the CCD camera is connected and correctly working and if the electronicsof the main rack, including RAM disks are normally operating.
At start up, self test starts and the window is displayed:
where software and hardware information is displayed.In the first area the software versions present in the unit are indicated.
This information allows the user and service engineer to identify the unit configuration.In the second area the hardware characteristics and the “license level”are displayed. This field permits to identify which functions are availableon the DIVA-D.
At the end of start up, three conditions are possible:
- No error code The Unit automatically goes to full operation and the display shows.
SYSTEM READY
Device is ready to operate.
-
8/15/2019 DIVA-D (User's Manual)
26/58
USER’S MANUAL Unit switching ON
(Rev. 0) DIVA-D23
If a DIVA-D is equipped with DICOM module after about 2 minutes
from this condition, the monitors shows:
SYSTEM READY
DICOM READY
This delay is determined by the DICOM module initialization timethat is longer than the DIVA-D one.
- Camera not connected (see paragraph 6.1)
It is necessary to press ENTER key to escape from this condition.Unit goes to off line operation only.
- System fault (see paragraph 6.2)
Unit stops. A service engineer is required.
-
8/15/2019 DIVA-D (User's Manual)
27/58
USER’S MANUAL Unit switching ON
DIVA-D (Rev. 0)24
6.16.16.16.1 CCD camera not connectedCCD camera not connectedCCD camera not connectedCCD camera not connected
During auto test, the unit could recognize a fault in the CCD camera connection.In this condition the window shows:
It is necessary to press ENTER key to escape from this message.In this condition the device goes into post processing mode in which it
will be possible to perform review only of exams already stored.No acquisition of new images is possible.
-
8/15/2019 DIVA-D (User's Manual)
28/58
USER’S MANUAL Unit switching ON
(Rev. 0) DIVA-D25
6.26.26.26.2 System faultSystem faultSystem faultSystem fault
If auto test recognizes an analogic, digital or disks fault the window shows:
The unit is out of work and a service engineer must be contacted.
-
8/15/2019 DIVA-D (User's Manual)
29/58
USER’S MANUAL Patient data
DIVA-D (Rev. 0)26
7777
PATIENT DATAPATIENT DATAPATIENT DATAPATIENT DATABefore starting with a new exam is necessary to insert into DIVA-D thedata of the new patient.If this is not done, as soon the exposure images acquisition is required,the message:
is displayed.It is necessary to press NEW PATNT key to open the patient data entering window.
NOTE:If fluoroscopy mode is activated, without storing any LIH or any fluorosequence, the images visualization is present also if no patient data isinserted. This because a patient could perform a fluoroscopicinvestigation only. In all other cases, because all radiographic images are
automatically stored on RAM disks, the patient data must be entered intothe data base.
When the key NEW PATNT is activated the new patient window isdisplayed:
-
8/15/2019 DIVA-D (User's Manual)
30/58
USER’S MANUAL Patient data
(Rev. 0) DIVA-D27
If DICOM module is present the “worklist” selection is allowed, otherwise
no.For this function, please refer to DICOM functions chapter.
By using UP or DOWN arrow keys activate the selection required between"DATA ENTRY" or "LAST PATIENT".Press ENTER key to confirm.
If “data entry” is selected, the window appears:
By using UP or DOWN arrow keys select the required fields. The surname field is mandatory, the others are optional.
NOTE:If the birth date is not entered, the patient age in the data base will berecorded as 0.
If sex is not inserted, O (other) will be present in the data base).
Confirming by ENTER, the data are stored into the data base and themodality window is displayed to permit the acquisition mode selection.
If "last patient" is selected, the PATIENT DATA window appears with thedata relevant to the last patient stored.
This selection is used when during exam on a patient, the unit has beenswitched off.
This feature is very useful to avoid 2 files for the same patient.
-
8/15/2019 DIVA-D (User's Manual)
31/58
USER’S MANUAL Fluoroscopy
DIVA-D (Rev. 0)28
8888
FLUOROSCOPYFLUOROSCOPYFLUOROSCOPYFLUOROSCOPYIf DIVA-D is equipped with 2 monitors in the control desk area, one of them is reserved to fluoroscopy images only, otherwise the fluoro window is present any time the fluoro command is activated.
The following functions and information are available in this modality:
• Status
The visualization could be X-ray during acquisition or L.I.H. (Last Image Hold) in off line.
• Reg. of int
The visualization is modified according to selection.ROI (Region Of Interest) must be modified in function of position anddimension of the organ. 6 different ROI are available troughcorrespondent keys. Maintaining pressed the key relevant to the ROIselected for 3 sec, the ROI dimension and position is displayed on theimage.
• Modality
This field could be STD or PULSE xx fr/sec.
By means of the key the fluoro mode can be selected. When led is off, STD is displayed in the window, meaning that thefluoro mode selected is continuous.
When led is on PULSE with the frame rate is displayed in the window. This rate can be modified using keys FR/s + and FR/s -.
-
8/15/2019 DIVA-D (User's Manual)
32/58
USER’S MANUAL Fluoroscopy
(Rev. 0) DIVA-D29
• Store
In normal use the condition is OFF. To activate the fluoro storingmodality it is necessary to press STORE. The relevant led turns onand the images number stored are indicated at the bottom of thedisplay.
When fluoro acquisition is interrupted also the storage is stopped. At new fluoro acquisition it will be necessary to press the STORE key again to perform a new storage.In LIH condition, pressing the STORE key, the LIH will be stored intothe hard disk.
• Edge
By default the value is OFF. It means the edge level used is the
default one. Using E key, the visualization is ON and the edge levelselected is the one configured at the installation.
• Noise
By default the value is OFF. It means the noise suppression levelused is the default one. Using N key, the visualization is ON and thenoise level selected is the one configured at the installation.
• Road map
Road mapping function is available on DIVA-D AS25 only. The function is activated by using the ROAD MAP key. Pressing thekey, the relevant led turns on and the fluoro images will be in
subtractions mode.Releasing fluoro, on the display the LIH will be in sub mode andfollowing fluoro images will have the mask overlapped.
To reset the road mapping mode and the mask, press the key withfluoro off.
The Led will be turned off and next fluoro will be performed instandard mode.
• Kv – mA – Dose
In these fields the actual kV and mA are displayed.Dose field is not yet activated.
• Run – FrameIn these fields the fluoro run number stored for the present patient and the images contained in the run are displayed.
The dose level can be modified by using the keys kV+ and kV-. The effect of this action is the image contrast level becoming higher or lower depending by the key activated. 3 decreasing steps and 3increasing steps are available starting from the standard level.
The radiological parameters are automatically adjusted by the system,during fluoro acquisition to optimize the image video level.
This adjustment is calculated based on ROI selection.
-
8/15/2019 DIVA-D (User's Manual)
33/58
USER’S MANUAL Fluoroscopy
DIVA-D (Rev. 0)30
8.18.18.18.1 Fluoro special functionFluoro special functionFluoro special functionFluoro special function
In fluoro off line mode, by pressing the SPEC FUNCT key, the relevant window is displayed:
By using UP or DOWN arrow keys select the field required.Press RESET key to activate the relevant selection.
• Store XX Fr/s
The storage rate in fluoro can be selected to preserve disk space. This selection modifies the storage speed only without modifying the visualization rate.
• Standard / Pulsed
This selection can also be done in on line mode by using the pulsedkey as described above.
• Ref image This selection appears when DIVA-D is equipped with 2 monitors incontrol desk area only.
When this selection is active the graphy monitor is used to display the LIH while on fluoro monitor the live images are displayed.
• Pattern 1 / 2
Allows to display 2 different patterns in order to adjust correctly themonitor brightness and contrast.
-
8/15/2019 DIVA-D (User's Manual)
34/58
USER’S MANUAL Modality selection
(Rev. 0) DIVA-D31
9999
MODALITY SELECTIONMODALITY SELECTIONMODALITY SELECTIONMODALITY SELECTION When patient data are inserted, the modality and relevant anatomicprograms must be chosen.
The access to this function is automatic when exiting from patient data window with ENTER key or any time the MODE key is activated. This because when a new patient is inserted it is necessary to select themodality for the first acquisition, and to change this modality if a patient requires different exams type.
The acquisition modality can be:
• SPOT
• BONE• ANGIO
• PERIF
• STEP
• TOMO.
SPOT, BONE and TOMO are allowed for both DIVA-D models RF and AS, ANGIO and PERIF on AS only, STEP could be present on a AS model only as supplementary option.
The step modality is allowed only if the system is based on APOLLO tableonly.
The modality window looks as follows:
It is divided in 3 areas:
• modality selection
• anatomic selection
• parameters modification.By default the left area is active when modality is selected. To informabout the area selected, the background is light gray.
-
8/15/2019 DIVA-D (User's Manual)
35/58
USER’S MANUAL Modality selection
DIVA-D (Rev. 0)32
By using UP and DOWN arrow keys, select the required modality, by
using RIGHT arrow key select the anatomic area, this background becomes light grey and select now the desired anatomic program.
NOTE: The not available modality (depending by the DIVA-D license level) will bedisplayed in gray and not reachable by selection.
On the right area the parameters associated to the selected anatomicprogram are displayed.
To modify permanently one or more parameters, press RIGHT arrow key,
the background will be light gray (as in picture) and by using relevant keys, the associated parameter can be changed.Using alphanumeric keys the anatomic name will be modified and themessage “ENTER to save” will appear on window bottom side.By ENTER key the new anatomic configuration will be stored.
If a temporary modification, on an existing anatomic program, need to bedone it will be sufficient to modify the relevant parameter by using theassociated key.Confirming by ENTER key, the acquisition mode will be selected with themodified parameters.
In modality function, the parameters associated will be different assummarized below:
SPOT BONE ANGIO PERIF STEP TOMO
ROI ROI ROI ROI ROI ROI
Frame rate Edge Frame rate Frame rate Frame rate Tomo frames
Edge Low dose Edge Edge Edge Edge
Low dose Neg – Pos Low dose Low dose Low dose Low dose
Neg -Pos Neg – Pos Neg – Pos Neg – Pos Neg – Pos
Auto mask Auto mask Auto mask Pulsed fluoro
Sub cont Sub cont Sub cont
Kv reduction
-
8/15/2019 DIVA-D (User's Manual)
36/58
USER’S MANUAL Graphy
(Rev. 0) DIVA-D33
10101010
GRAPHYGRAPHYGRAPHYGRAPHY When the modality selection is done and confirmed, DIVA-D is ready for graphy acquisition.
The window displayed is:
on which the patient data, modality and anatomic parameters aredisplayed.Now the acquisition can start or anatomic parameters modification can
be done by using the associated keys.If a key for modification is pressed (i.e. ROI, Frame/sec ecc) the modality
window will be displayed.
Before to start with graphy acquisition, a fluoro acquisition must bedone.
This action allows DIVA-D to calculate the correct exposure parameters
(technique factors). As matter of fact at the exposure request, DIVA-D informs the generator about the technique factors to be used for the graphy acquisition.
The area “Run” is reserved to the number of exams stored for the present patient.During acquisition the areas “Run”, “Frame”, “mSec” and “kV” areupdated.
-
8/15/2019 DIVA-D (User's Manual)
37/58
USER’S MANUAL Graphy
DIVA-D (Rev. 0)34
At the end of acquisition the following window will be displayed:
-
8/15/2019 DIVA-D (User's Manual)
38/58
USER’S MANUAL Post processing
(Rev. 0) DIVA-D35
11111111
POST PROCESSINGPOST PROCESSINGPOST PROCESSINGPOST PROCESSING To review the exams contained for the current “patient”, 2 options arepossible:
• by MENU key
• by STOP key.
Using STOP the Post Processing window is immediately displayed,selecting MENU the patient selection window will appears.
Select the required patient using UP and DOWN keys.Press ENTER to confirm.
-
8/15/2019 DIVA-D (User's Manual)
39/58
USER’S MANUAL Post processing
DIVA-D (Rev. 0)36
The window reporting the data relevant to the different runs of the
patient is displayed.Select the desired run using UP and DOWN keys.Press ENTER to confirm.
The post processing window appears:
On this window, in addition to the data relevant to patient, acquisitionmodality, anatomic program etc, the following information is present:
• ROI used to acquire the present run
• Frame rate used to acquire the present run
• I.I. field used to acquire the present run
• mSec exposure time.
In addition to these information the following parameters are present:
• total runs number and run showed
•
total frame number present on the active run and the frame showed• in angio or perif mode only, the mask number selected
• in angio or perif mode only, the sub contrast value
• the windowing level
• the edge level.
-
8/15/2019 DIVA-D (User's Manual)
40/58
USER’S MANUAL Post processing
(Rev. 0) DIVA-D37
All of them and other settings can be modified using the dedicated keys:
To select run in a patient and frame in a run. The frame keys allow to review a run frame by frameindependently by frame rate acquisition.
To play a run containing more then one frame andto stop the visualization sequence. The frame rate in play mode is the same usedduring acquisition and cannot be modified.
To introduce electronic diaphragm.
When this function is activated (led ON) by usingUP, DOWN, LEFT and RIGHT keys, upper and right shutters will be closed and opened.Using the four arrows contemporarily with SHIFT key the lower and left shutters will be activated. To reset the function press the key.
To reverse the gray scale of image from negative topositive and reverse.
To zoom the present image 4 times. When this function is activated (led ON) by usingUP, DOWN, LEFT and RIGHT keys, it is possible toscroll the enlarged image.
To reset the function press the key. When this function is activated the frames present in the run are displayed with the frame rate used inacquisition. The number of sequence depends by the technical setting. At the end of the last sequence the led turns OFF. Any time the cine loop is required, press the key toactivate the function.
This function is available for exams acquired inangio mode only. When the subtracted image isdisplayed the pixel shift function can be activated by relevant key. The led become ON and the mask position can be modified by 1/8 pixel step. Thismovement is obtained by UP, DOWN, LEFT andRIGHT keys.
This function is available for exams acquired inangio mode only. When the subtracted image isdisplayed the mask can be removed or replaced by means of the relevant key.
This function is available for exams acquired inangio mode only. When a run is scrolling thecontained frames, using this key a new mask can beselected. The mask will be captured at the exact moment when the key is pressed.
This function is available for exams acquired in
angio mode only. When subtracted image isdisplayed it is possible to increase or decrease theimage mask number.
-
8/15/2019 DIVA-D (User's Manual)
41/58
USER’S MANUAL Post processing
DIVA-D (Rev. 0)38
To modify images characteristics by means of the
relevant keys it is possible to change:− windowing (changing the black and white levels)
− edge enhancement. With AUTO WNDW key the windowing will beautomatically optimized at default level. Pressingreset key any modification described above is reset.
When subtracted image is displayed it is possible toincrease or decrease the image background level by means of the keys on the left side.
In post processing special functions are available inaddition to the functions described. This functionsare described in the next paragraph.
-
8/15/2019 DIVA-D (User's Manual)
42/58
USER’S MANUAL Post processing
(Rev. 0) DIVA-D39
11.111.111.111.1 Special functionSpecial functionSpecial functionSpecial function
When the post processing modality is activated by means of the SPECIAL FUNCTION key the lower part of the window:
is represented as in the above figure.By means of the UP and DOWN key it is selectable one of the four specialfunctions available:
• overwrite
• angle measure
• length measure
• test pattern display.
• Overwrite
Allows to report comments on the image. To position the cursor onthe required place use UP, DOWN, LEFT and RIGHT arrow keys.Press ENTER key to confirm when the text insertion is completed.Changing the image or using STOP key, the text inserted is cancelled.
• Angle measure
Allows to take a measure of the angle of a part of image.Using UP and DOWN keys select the approx level from the table topon which the part to be measured is located between the 3 levelsproposed 0, 10 or 20 cm. Press ENTER to confirm.
To position the cursor on the required place use UP, DOWN, LEFT and RIGHT arrow keys.Press ENTER key to confirm the starting measuring point. Move the
cursor on the angle vertex and confirm whit ENTER. Move the cursor
-
8/15/2019 DIVA-D (User's Manual)
43/58
USER’S MANUAL Post processing
DIVA-D (Rev. 0)40
at the end of second segment of the angle, the angle value is
displayed.Changing the image or using STOP key, the text inserted is cancelled.
WARNING:Because of different reasons ( II deformation, error in the level organposition and others) the angle value obtained are only indicative. Do not use this value for diagnostic purpose.
• Length measure
Allows to take a measure of a part of the image.Using UP and DOWN keys select the approx level from the table topon which the part to be measured is located between the 3 levelsproposed 0, 10 or 20 cm. Press ENTER to confirm.
To position the cursor on the required place use UP, DOWN, LEFT and RIGHT arrow keys.Press ENTER key to confirm the starting measuring point. Move thecursor on the final measuring point and confirm with ENTER, thelength value is displayed.Changing the image or using STOP key, the text inserted is cancelled.More than one measure can be done on the same image. To do that repeat the above described steps.
Ratio between 2 measure can be done selecting the relevant window.
WARNING:Because of different reasons (II deformation, error in the level organposition and others) the measures obtained are only indicative. Do not use this value for diagnostic purpose.
• Test pattern
Allow to display a test pattern to adjust correctly the monitor
brightness and contrast.Starting with any modality the test pattern is reset from monitor.
-
8/15/2019 DIVA-D (User's Manual)
44/58
USER’S MANUAL Post processing
(Rev. 0) DIVA-D41
11.211.211.211.2 Multi imageMulti imageMulti imageMulti image
When the patient menu window is displayed, multi image function can beused.It is a short and easy way how to select the run to be review.
Using the key, on monitor the first image of any run present intothe selected patient is displayed.
By means of the UP, DOWN, LEFT and RIGHT arrow keys, select thedesired run and press ENTER to confirm.
The run selected will be displayed in full screen mode.
-
8/15/2019 DIVA-D (User's Manual)
45/58
USER’S MANUAL Post processing
DIVA-D (Rev. 0)42
11.311.311.311.3 Image printImage printImage printImage print
If DIVA-D is not equipped with DICOM module, it is possible to send theimage to a laser printer directly.
The image present on display is also present on the printer output channel.
To capture this image it is necessary to confirm from the printer keyboard supplied by the printer manufacturer.
NOTE:Do not change the image displayed until the laser printer acquisition is
completed.
-
8/15/2019 DIVA-D (User's Manual)
46/58
USER’S MANUAL DICOM functions
(Rev. 0) DIVA-D43
12121212
DICOM FUNCTIONSDICOM FUNCTIONSDICOM FUNCTIONSDICOM FUNCTIONSIf the DIVA-D is equipped with DICOM module, the DICOM key is active.By means of this key the following window is displayed:
Depending from the license present the following functions can beactivated:
• print (to send images to a DICOM laser printer)
• store (to send images to a Pac’s via network)
• CD-ROM (to burn CD-ROM with images).
In addition to these functions another software package can be present:
• waiting list.
-
8/15/2019 DIVA-D (User's Manual)
47/58
USER’S MANUAL DICOM functions
DIVA-D (Rev. 0)44
12.112.112.112.1 Waiting listWaiting listWaiting listWaiting list
This function is allowed if the network is equipped with HIS/RIS. When new patient window is displayed (chapter 7) the waiting list selection is allowed.Confirm by ENTER, the window
is displayed.On this window the patients present on the worklist are displayed.By means of the UP and DOWN arrow key select the desired one, pressENTER to confirm.
The patient data window will be displayed (chapter 7) reporting all data inserted in HIS/RIS.By this function any manual patient data entering is avoided.
-
8/15/2019 DIVA-D (User's Manual)
48/58
USER’S MANUAL DICOM functions
(Rev. 0) DIVA-D45
12.212.212.212.2 Queue view and flushQueue view and flushQueue view and flushQueue view and flush
To enter into DICOM functions the dedicated key must be used. The window
will be displayed.
If no queue error or messages are present, the last used function will beproposed, anyway the queue view will be reachable by means of ENTER key directly.In any case UP and DOWN arrow keys can be used to reach the functiondesired or queue view.If errors or messages are present in the queue, the relevant field is with
white background; pressing ENTER key the window
will appears.
-
8/15/2019 DIVA-D (User's Manual)
49/58
USER’S MANUAL DICOM functions
DIVA-D (Rev. 0)46
The queue list will be showed and erasable.
To cancel the list, using UP and DOWN arrow keys, activate FLUSHfunction and confirm with ENTER key.
-
8/15/2019 DIVA-D (User's Manual)
50/58
USER’S MANUAL DICOM functions
(Rev. 0) DIVA-D47
12.312.312.312.3 PrintPrintPrintPrint
When DICOM window is selected, the last function used in DICOM modeis proposed.By means of UP and DOWN arrow keys select the required function andconfirm with ENTER key.If print function is selected the printer window selection is showed.
With UP and DOWN arrow keys, select the required printer and confirm. With LEFT and RIGHT arrow key select the film subdivision and confirm. The DICOM print window
is displayed.Select the image required and modify, if necessary the characteristic asin standard post processing mode.In the lower section of the window, the frame number and position aredisplayed.
-
8/15/2019 DIVA-D (User's Manual)
51/58
USER’S MANUAL DICOM functions
DIVA-D (Rev. 0)48
To add the image, the key must be pressed 2 times. After the first time the window
is displayed.Pressing once more the addition is confirmed and the image stored intoDICOM print temporary archive.
The previous window will be showed allowing a new image addition. When the above window is present, 3 other functions are available (useUP and DOWN arrow keys to select and press ENTER to confirm):
• Delete & exit
To reset the temporary DICOM print archive and exit from DICOMfunctions.
• Study print
To print the image stored.
To print the study it is also possible to press 2 times the key starting from the main DICOM print window.
• Change frame
To store the present image instead of an image already present intothe DICOM print archive.
To do that use LEFT and RIGHT arrow keys, the white portion of thefilm will be moved on the film representation showing the place on
which the present image will be located.Press ENTER to confirm, the previous image will be erased and thenew one stored.
-
8/15/2019 DIVA-D (User's Manual)
52/58
USER’S MANUAL DICOM functions
(Rev. 0) DIVA-D49
12.412.412.412.4 StoreStoreStoreStore
By means of UP and DOWN arrow keys select the store function andconfirm with ENTER key.
When store function is selected the window is showed:
On the lower part of the window the following information are listed:
• Total Mbytes
Mbytes used by the images stored in the present DICOM store study. This is referred to the temporary DICOM archive.
• Total Patient
This value can be 0 or 1. As matter of fact a DICOM study to be stored cannot contain morethan 1 patient.
•
Total runsIs the number of runs stored to be transferred in the network.
• Total frames
Is the number of single image to be transferred in the network.
• Total sub-runs
Is the number of sub-runs stored to be transferred in the network.
-
8/15/2019 DIVA-D (User's Manual)
53/58
-
8/15/2019 DIVA-D (User's Manual)
54/58
USER’S MANUAL DICOM functions
(Rev. 0) DIVA-D51
12.512.512.512.5 CD - ROMCD - ROMCD - ROMCD - ROM
By means of UP and DOWN arrow keys select the CD-ROM function andconfirm with ENTER key.
When CD-ROM function is selected the media type window is showed:
allowing to select the CD-ROM format between High resolution1024x1024 or High speed 512x512.High resolution saves images with 1024x1024 matrix but for staticimages only.High speed has matrix 512x512 but is capable to store images up to25 images/sec.By means of UP and DOWN arrow keys select the media type requiredand confirm with ENTER.
-
8/15/2019 DIVA-D (User's Manual)
55/58
USER’S MANUAL DICOM functions
DIVA-D (Rev. 0)52
The following window will be displayed:
On the lower part of the window the following information are listed:
• CD free space
The disk percentage space still available.
• Total Patient
This value can be 0 or 1. As matter of fact a DICOM study to be stored cannot contain morethan 1 patient.
• Total runs
Is the number of runs stored to be transferred in the network.
• Total frames
Is the number of single image to be transferred in the network.
• Total sub-runs
Is the number of sub-runs stored to be transferred in the network.
-
8/15/2019 DIVA-D (User's Manual)
56/58
USER’S MANUAL DICOM functions
(Rev. 0) DIVA-D53
To store into temporary DICOM archive, select the patient, the run or the
image required and press ENTER. The window:
is displayed.By means of UP and DOWN arrow keys select if the memory addition will
be of the showed frame only, of the run or of the total patient.
Press ENTER to confirm.
The previous window will be showed allowing a new addition. When the above window is present, 2 other functions are available (useUP and DOWN arrow keys to select and press ENTER to confirm):
• Delete & exit
To reset the temporary DICOM CD-ROM archive and exit fromDICOM functions.
• Burn disk
To send the study stored to CD-ROM.
Before to select "burn disk", put a new CD-ROM into driver and press
key. The dicomization procedure will start and at the end of this operation theCD-ROM burning start.
This sequence is displayed on the window with a waiting notice.
NOTE:If no CD-ROM or not valid one is present into the driver, an error will be
present in the queue.DIVA-D will try again to access the CD after 30 sec.Insert a valid CD-ROM into the driver before a new writing attempt.
-
8/15/2019 DIVA-D (User's Manual)
57/58
USER’S MANUAL Maintenance and cleaning
DIVA-D (Rev. 0)54
13131313
MAINTENANCE AND CLEANINGMAINTENANCE AND CLEANINGMAINTENANCE AND CLEANINGMAINTENANCE AND CLEANINGLike all electrical device, this equipment requires periodical checks.
This will insure an efficient and correct functioning for years to come.
This equipment is made up of electrical and electronic parts that need to be checked out in order to guarantee high quality radiological results asdescribed in previous chapter.
The maintenance and cleaning procedures described in this chapter must be done by the user.In addition to the above, the authorized service technician will have toperform the maintenance actions described in the Service Manual.
WARNING:
• Before proceeding with the cleaning of the equipment, disconnect it from the mains.
• Make sure that water and any other liquids are not spilled inside theequipment as this may cause corrosion and short circuits.
• Clean the painted surfaces by using a moist cloth and neutraldetergent, taking care to wipe them by using a dry cloth. Do not use
solvents (alcohol, petrol, and trichloroethilene) or corrosive or abrasive products.
13.113.113.113.1 Frequency of maintenance and cleaning actionsFrequency of maintenance and cleaning actionsFrequency of maintenance and cleaning actionsFrequency of maintenance and cleaning actions
Frequency Description
1 month Check the presence of labels and their condition
1 months Check of the image quality.
6 months Clean the monitors and the external parts.
6 year Visual checking of the status of the interconnection cablesand relevant connection.
-
8/15/2019 DIVA-D (User's Manual)
58/58