Integra ® Jarit ® Endoscope & Instrument Holder DEVICE DESCRIPTION All Integra ® Jarit ® instruments are carefully inspected before shipment. Because damage may occur during transit, the instruments should be thoroughly inspected upon receipt and prior to use. Before first use, all instruments must be thoroughly cleaned prior to sterilization or high level decontamination. If you have any questions or require additional information regarding your Jarit ® Instrument Holder, please contact your JARIT sales representative or Jarit ® Technical Services at 1-800-431-1123. INDICATIONS FOR USE The Integra ® Jarit ® Endoscope and Instrument Holder consists of a table-mounted endoscope and instrument holding system intended for use by surgeons to hold instruments, endoscopes and arthroscopes with a diameter of 4mm to 10mm during general diagnostic and therapeutic procedures. The device is also intended for use by qualified surgeons for holding endoscopes during diagnostic and therapeutic neurologic procedures. The product is supplied non-sterile. Prior to the initial and following each use, the product must be cleaned, disinfected, and sterilized, as well as checked for visible irregularities and malfunctioning according to the indications given in these instructions for use. WARNING This product is not to be used for “non-grounded” patient positioning. Special care should be taken when these products are used together with high-frequency applications. Contact between them must be avoided. PRODUCT DESCRIPTION The Endoscope and Instrument Holder consists of the rail clamping base and the articulating arm. 620-800 JARIT ® Endoscope & Instrument Holder - complete set includes the following components: (Total length is 85cm) • Articulated Arm • Scope Holder - 4/10mm • Column - length 30cm • Rail Clamping Base, double clamp (620-804) CAUTION The Articulating arm, scope holder and column are one piece and cannot be taken apart. 620-804 RAIL CLAMPING BASE - is included in 620-800 and 620-805, but may be purchased separately. 620-805 JARIT ® BARIATRIC ENDOSCOPE & INSTRUMENT HOLDER - set includes the following components: (Total length is 100cm) • Articulated Arm • Scope Holder - 4/10mm • Column - length 45cm • Rail Clamping Base, double clamp (620-804) Articulating Arm Scope Holder Rails Clamping Base Wing Handle (Pos.1) Wing Handle (Pos.2) Column Central Handle CAUTION The Articulating arm, scope holder and column are one piece and cannot be taken apart. ARTICULATING ARM The articulating arm can be adjusted in any desired position. With the central handle, all joints can be fixed and quickly released. The central handle allows for a quick and easy positioning of endoscopes and instruments. The articulated arm is made of stainless steel and the central handle is made of plastic. SCOPE HOLDER The scope holder allows form-fitting clamping of endoscopes and instruments. The scope holder and wing handle are made of plastic. RAIL CLAMPING BASE The rail clamping base has a rail clamp opening of 35mm and provides a secure and strong base for the endoscope and instrument holder. The hole diameter for the articulating arm column is 16mm. The rail clamping base is made of aluminum and the wing handles are made of plastic. The rail clamping base is not electrically insulated. DIRECTIONS FOR DISASSEMBLY 1. Remove the endoscope or instrument from the scope holder by turning the wing handle in a counter clockwise direction. 2. Hold the articulating arm with one hand and loosen the wing handle on the rail clamping base to remove the articulating arm (Pos. 2). 3. Loosen the handle on the rail clamping base and lift the rail clamping base from the rail of the surgery table (Pos. 1). DIRECTIONS FOR ASSEMBLY 1. Place the rail clamping base onto the rail of the surgery table and fix by turning the handle in the clockwise direction (Pos. 1). 2. Place the column of the articulating arm into the hole of the base and tighten the wing handle by turning the handle in a clockwise direction (Pos. 2). 3. The desired height and orientation of the column, including the articulated arm, can be adjusted at any time by loosening the handle (Pos. 2). To do so, hold the column with one hand and loosen the handle with the other hand. When the desired position is reached, tighten the handle. WARNING If the rail clamping base is not tightened correctly, it may become loose which may cause injuries. WARNING This product is not to be used for “non-grounded” patient positioning. Special care should be taken when these products are used together with high- frequency applications. Contact between them must be avoided. CAUTION The fixation of the articulated arm is based on the principle of friction. Changing the position without loosening the clamping mechanism can cause damage and will shorten the lifespan of the Jarit ® Endoscope & Instrument Holder. DIRECTIONS FOR ASSEMBLY As with any decontamination procedure, personnel should follow accepted guidelines for hand washing, the use of protective attire, etc., as recommended by A.A.M.I. Standards and Recommended Practice, “Comprehensive Guide to Steam Sterilization and Sterility Assurance in Health Care Facilities,” ANSI/AAMI ST79:2006. Disinfection or sterilization cannot be accomplished without first thoroughly cleaning medical devices/ equipment. DECONTAMINATION is a two step process. 1. Thorough cleaning and rinsing. 2. Sterilization or disinfection. A. MANUAL DECONTAMINATION: PRECLEANING: Remove gross debris from Jarit ® Endoscope & Instrument Holder with a lap sponge and sterile water after the procedure to prevent drying on of blood and body fluids, etc. It is important to rinse the JARIT Endoscope & Instrument Holder that has been exposed to blood and saline solution before these substances dry. Blood and body fluids as well as saline solutions are highly corrosive. In addition, blood can produce a stain that is difficult to remove. PREPARING THE JARIT ® ENDOSCOPE & INSTRUMENT HOLDER FOR TRANSPORT TO DECONTAMINATION: Immediately after the surgical procedure, place the Jarit ® Endoscope & Instrument Holder into an instrument tray and cover with a towel. Transport tray in an impervious plastic bag or container to the decontamination environment (keep the outside of the containment clean). Pos.1 Pos.2

Upload: others

Post on 04-May-2022

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: DIRECTIONS FOR ASSEMBLY - Integra Life

Integra® Jarit® Endoscope & Instrument HolderDEVICE DESCRIPTIONAll Integra® Jarit® instruments are carefully inspected before shipment. Because damage may occur during transit, the instruments should be thoroughly inspected upon receipt and prior to use. Before first use, all instruments must be thoroughly cleaned prior to sterilization or high level decontamination. If you have any questions or require additional information regarding your Jarit® Instrument Holder, please contact your JARIT sales representative or Jarit® Technical Services at 1-800-431-1123.

INDICATIONS FOR USEThe Integra® Jarit® Endoscope and Instrument Holder consists of a table-mounted endoscope and instrument holding system intended for use by surgeons to hold instruments, endoscopes and arthroscopes with a diameter of 4mm to 10mm during general diagnostic and therapeutic procedures. The device is also intended for use by qualified surgeons for holding endoscopes during diagnostic and therapeutic neurologic procedures.

The product is supplied non-sterile. Prior to the initial and following each use, the product must be cleaned, disinfected, and sterilized, as well as checked for visible irregularities and malfunctioning according to the indications given in these instructions for use.

WARNINGThis product is not to be used for “non-grounded” patient positioning. Special care should be taken when these products are used together with high-frequency applications. Contact between them must be avoided.

PRODUCT DESCRIPTIONThe Endoscope and Instrument Holder consists of the rail clamping base and the articulating arm.620-800 JARIT® Endoscope & Instrument Holder - complete setincludes the following components: (Total length is 85cm)• Articulated Arm• Scope Holder - 4/10mm• Column - length 30cm• Rail Clamping Base, double clamp (620-804)CAUTION The Articulating arm, scope holder and column are one piece andcannot be taken apart.

620-804 RAIL CLAMPING BASE - is included in 620-800 and 620-805, but may be purchased separately.

620-805 JARIT® BARIATRIC ENDOSCOPE & INSTRUMENT HOLDER - set includes the following components: (Total length is 100cm)• Articulated Arm • Scope Holder - 4/10mm• Column - length 45cm• Rail Clamping Base, double clamp (620-804)

Articulating Arm

Scope Holder

Rails Clamping Base

Wing Handle (Pos.1)Wing Handle (Pos.2)

Column

Central Handle

CAUTION The Articulating arm, scope holder and column are one piece and cannot be taken apart.

ARTICULATING ARMThe articulating arm can be adjusted in any desired position. With the central handle, all joints can be fixed and quickly released. The central handle allows for a quick and easy positioning of endoscopes and instruments. The articulated arm is made of stainless steel and the central handle is made of plastic.

SCOPE HOLDERThe scope holder allows form-fitting clamping of endoscopes and instruments. The scope holder and wing handle are made of plastic.

RAIL CLAMPING BASEThe rail clamping base has a rail clamp opening of 35mm and provides a secure and strong base for the endoscope and instrument holder. The hole diameter for the articulating arm column is 16mm. The rail clamping base is made of aluminum and the wing handles are made of plastic. The rail clamping base is not electrically insulated.

DIRECTIONS FOR DISASSEMBLY1. Remove the endoscope or instrument from the scope holder by turning the wing handle in a counter clockwise direction.

2. Hold the articulating arm with one hand and loosen the wing handle on the rail clamping base to remove the articulating arm (Pos. 2).

3. Loosen the handle on the rail clamping base and lift the rail clamping base from the rail of the surgery table (Pos. 1).

DIRECTIONS FOR ASSEMBLY1. Place the rail clamping base onto the rail of the surgery table and fix by turning the handle in the clockwise direction (Pos. 1). 2. Place the column of the articulating arm into the hole of the base and tighten the wing handle by turning the handle in a clockwise direction (Pos. 2).3. The desired height and orientation of the column, including the articulated arm, can be adjusted at any time by loosening the handle (Pos. 2). To do so, hold the column with one hand and loosen the handle with the other hand. When the desired position is reached, tighten the handle. WARNINGIf the rail clamping base is not tightened correctly, it may become loose which may cause injuries.WARNINGThis product is not to be used for “non-grounded” patient positioning. Special care should be taken when these products are used together with high-frequency applications. Contact between them must be avoided.CAUTIONThe fixation of the articulated arm is based on the principle of friction. Changing the position without loosening the clamping mechanism can cause damage and will shorten the lifespan of the Jarit® Endoscope & Instrument Holder.

DIRECTIONS FOR ASSEMBLYAs with any decontamination procedure, personnel should follow accepted guidelines for hand washing, the use of protective attire, etc., as recommended by A.A.M.I. Standards and Recommended Practice, “Comprehensive Guide to Steam Sterilization and Sterility Assurance in Health Care Facilities,” ANSI/AAMI ST79:2006. Disinfection or sterilization cannot be accomplished without first thoroughly cleaning medical devices/equipment.

DECONTAMINATION is a two step process.

1. Thorough cleaning and rinsing.

2. Sterilization or disinfection.

A. MANUAL DECONTAMINATION:

PRECLEANING: Remove gross debris from Jarit® Endoscope & Instrument Holder with a lap sponge and sterile water after the procedure to prevent drying on of blood and body fluids, etc. It is important to rinse the JARIT Endoscope & Instrument Holder that has been exposed to blood and saline solution before these substances dry. Blood and body fluids as well as saline solutions are highly corrosive. In addition, blood can produce a stain that is difficult to remove.

PREPARING THE JARIT® ENDOSCOPE & INSTRUMENT HOLDER FOR TRANSPORT TO DECONTAMINATION:Immediately after the surgical procedure, place the Jarit® Endoscope & Instrument Holder into an instrument tray and cover with a towel. Transport tray in an impervious plastic bag or container to the decontamination environment (keep the outside of the containment clean).

Pos.1 Pos.2

Page 2: DIRECTIONS FOR ASSEMBLY - Integra Life

CLEANING: To prevent the formation of biofilm, cleaning shouldoccur as soon as possible after the Jarit® Endoscope & Instrument Holder is used. Biofilm is an accumulation of a biomass of bacteria and extracellular material that tightly adheres itself to the surface of the Jarit® Endoscope & Instrument Holder. It cannot be easily removed, and protects microorganisms from being easily removed by ordinary cleaning/decontamination methods used in hospitals.

Step 1. Pre-Rinse, as necessary with tap water.

WARNING The central handle of the articulated arm must be tightened during the cleaning, rinsing and disinfection process.

Step 2. Cleaning the Jarit® Endoscope & Instrument Holder: The use of a neutral pH detergents (≥ 4pH and ≤9pH) is vital to the maintenance of the Jarit® Endoscope & Instrument Holder. Contact with acidic or alkaline solution will remove the Jarit® Endoscope & Instrument Holders’ protective barrier of chromium oxide, often leading to corrosion, pitting, and breakage. The ideal cleaning agent is nonabrasive, low-foaming and free-rinsing. Using a small clean hand-held brush, remove soil from all surfaces of the Jarit® Endoscope & Instrument Holder. Never use steel wool, wire brushes, scalpel blades or highly abrasive detergent or cleansers to remove soil from the Jarit® Endoscope & Instrument Holder. These will damage the Jarit® Endoscope & Instrument Holders’ protective surface and lead to corrosion. Use a clean soft bristled brush to clean the Jarit® Endoscope & Instrument Holder. A build-up of soil, debris, lubricants, etc. will make it difficult to use the Jarit® Endoscope & Instrument Holder and eventually irreparably damage it. Manual cleaning should remove all visible residue.

WARNING Instruments made of aluminum alloys will be damaged by basic(pH>9) cleaning agents and solvents.Step 3. Rinse: Thoroughly rinse the Jarit® Endoscope & Instrument Holder by immersing it in tap water and wiping with a clean, soft cloth.

Step 4. Final Rinse: Should be with “treated water”. Softened or deionized water should be used for the final rinse to better remove detergents etc.

• Softening water removes calcium and magnesium ions that cause waterto be hard. Iron ions may also be removed by this treatment.

• Deionization removes ionized salts and particles from the water. Excessively hard water can spot or stain the Jarit® Endoscope & InstrumentHolder and excessive chlorine in water can cause pitting of the Jarit®

Endoscope & Instrument Holder.

• Deionized water is preferred for the final rinse.

Step 5. Decontaminate Clean Jarit® Endoscope & Instrument Holder: Once the Jarit® Endoscope & Instrument Holder has been cleaned, it must be rendered safe for handling, inspection and assembly.

Step 6. Visual Inspection of the Jarit® Endoscope & Instrument Holder: Visually inspect the Jarit® Endoscope & Instrument Holder for cleanliness and to ensure all parts are in proper working order. Inspection is a vital part of proper care and maintenance. The Jarit® Endoscope & Instrument Holder in need of repair will not perform accurately in surgery and breakage is likely to occur. DO NOT USE a damaged Jarit® Endoscope & Instrument Holder.

B. MECHANICAL DECONTAMINATIONThe Jarit® Endoscope & Instrument Holder may be processed in an automatic washer decontaminator/disinfector. Follow the manufacturer’s specifications when using automatic washer decontaminators/disinfectors. They usually require the use of a low foaming, free rinsing detergent with a neutral pH (7.0). A high-foaming detergent may clean effectively but will often leave residual deposits on the Jarit® Endoscope & Instrument Holder and do harm to mechanical washers. Automated washer decontaminator/disinfectors usually have adjustable wash and rinse times. Some washers enable the user to customize extra cycles to process heavily soiled Jarit® Endoscope & Instrument Holders more effectively. Water quality for rinsing cycles should be sterile or germ-poor (maximum 10 germs/ml) and endotoxin-free (maximum 0.25 endotoxin units/ml) water (Aqua purificata). Air used for drying should be filtered. The cleaning and disinfecting apparatus should be serviced and checked regularly.

When choosing cleaning and disinfecting agents, please make sure that they do not contain the following:

• No organic/inorganic, respectively oxidizing acids (pH should not be lessthan 4, the use of neutral/enzymatic agents is recommended)

• No base (pH should not be higher than 9, the use of neutral/enzymatic agents is recommended).

• Solvents (alcohol, benzene, acetone, ...)

• Phenol

• Chlorine, bromine, iodine

• Chlorine salts (in particular ammonium chloride compounds), chlorinated/halogenated hydrocarbons

• Oxidizing agents, peroxide, hypochloride

Do not clean any of the instruments with metal brushes or steel wool.

All the instruments should not be exposed at temperatures higher than 137ºC (279ºF).

Check with a Jarit® Technical Service representative at 1-800-431-1123 for questions regarding processingdelicate, complex and/or multipart instruments by this method.

C. PREPARING THE JARIT® ENDOSCOPE & INSTRUMENTHOLDER FOR STERILIZATIONDrying: Before the Jarit® Endoscope & Instrument Holder is wrapped for sterilization or storage, it must be thoroughly dry. If the Jarit® Endoscope & Instrument Holder is wet when wrapped for sterilization, it is likely to come out of the sterilizer wet. “Wet Packs” are not suitable for use after sterilization because they may be easily contaminated when handled. In addition, remaining moisture, particularly in the hinges, may result in corrosion that will weaken the Jarit® Endoscope & Instrument Holder and lead to breakage during use. Place the Jarit® Endoscope & Instrument Holder into a surgical instrument tray. Prepare the Jarit® Endoscope & Instrument Holder for sterilization using a wrapper or rigid sterilization container that is appropriate for the method of steam sterilization.

WARNINGThe central handle of the articulated arm must be open during the sterilization process.

D. TERMINAL STEAM STERILIZATIONAfter following the decontamination and preparation recommendations, the Jarit® Endoscope & Instrument Holder is ready for sterilization. See also, AAMI Standards and Recommended Practice, “Comprehensive Guide to Steam Sterilization and Sterility Assurance in Health Care Facilities,” ANSI/AAMI ST79:2006. AAMI standards recommend that the sterilizer manufacturer’s written instructions for minimum cycle parameters should also be followed.

Medical device manufacturer’s exposure times to sterilization temperature may need to be longer than the minimum indicated by the sterilizer manufacturer but must never be shorter.

Below are the recommended sterilization parameters:

Sterilizer Exposure

Temperature

Exposure

Time

Minimum Dry

Time

Pre-vacuum

(wrapped)

121ºC (250ºF)

132ºC (270ºF)

134ºC (273ºF)

20 minutes

5 minutes

5 minutes

20-30 minutes

20-30 minutes

20-30 minutes

Never use the flash-sterilization.

Do not use hot air-sterilization, irradiation sterilization, formaldehyde or ethylene oxide sterilization, or plasma sterilization.

WARNING For additional information and inservicing assistance regarding instrument care and handling, contact your local Jarit® representative, or call Jarit® Technical Service Department at 1-800-431-1123.

WARNINGConsult individual national infection control/prevention protocols for specific guidance regarding processing medical devices with suspected exposure to Creutzfeldt-Jakob Disease (CJD).

Page 3: DIRECTIONS FOR ASSEMBLY - Integra Life

JL-00050 Rev C 02/20 1488266-1

WARRANTY Jarit® hereby guarantees the Jarit® Endoscope & Instrument Holders, 620-800, 620-804 and 620-805 for the period of two (2) years free from defects in material and workmanship when used normally for its intended purpose. This Warranty shall apply only to the original purchaser. This Warranty shall not apply to any conditions(s) or damage resulting from negligence, misuse, improper handling, improper cleaning, improper maintenance, improper opening techniques, unauthorized repair work or accident. Please direct your questions concerning this Warranty to: Integra LifeSciences Production Corporation11 Cabot BoulevardMansfield, MA 02048 USARETURNED GOODS POLICYProducts must be returned in unopened packages with manufacturer’s seals intact to be accepted for replacement or credit unless returned due to a complaint of product defect. Determination of a product defect will be madeby Integra. Products will not be accepted for replacement if they have beenin the possession of the customer for more than 90 days.REPAIRS AND MAINTENANCEShould your instruments require repair or maintenance, contact Jarit®

for return authorization and address. Instruments returned to Jarit® for repair must have a statement testifying that each instrument has been thoroughly cleaned and sterilized. Failure to supply evidence of cleaning anddisinfection will result in a cleaning charge and delayed processing of yourinstrument repair.PRODUCT INFORMATION DISCLOSUREINTEGRA AND ITS SUBSIDIARIES (“INTEGRA”) AND MANUFACTURER EXCLUDEALL WARRANTIES, EXCEPT INTEGRA’S APPLICABLE STANDARD WARRANTYWHETHER EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO, ANYIMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR APARTICULAR PURPOSE AS WELL AS ANY EXPRESS OR IMPLIED WARRANTY TOPATIENTS. NEITHER INTEGRA NOR MANUFACTURER SHALL BE LIABLE FORANY INCIDENTAL OR CONSEQUENTIAL LOSS, DAMAGE, OR EXPENSE,DIRECTLY OR INDIRECTLY ARISING FROM USE OF THIS PRODUCT. NEITHERINTEGRA NOR MANUFACTURER ASSUME NOR AUTHORIZE ANY PERSON TOASSUME FOR THEM ANY OTHER OR ADDITIONAL LIABILITY OR RESPONSIBILITY IN CONNECTION WITH THESE PRODUCTS.

Jarit, Integra and the Integra logo are registered trademarks of Integra LifeSciences Corporation or its subsidiaries in the United States and/or other countries. ©2020 Integra LifeSciences Corporation. All Rights Reserved.

SYMBOLS USED ON LABELINGManufacturer

Catalog number

Lot number

See instructions for use

Non-sterile - Sterilize prior to use

Consult instructions for use

Integra LifeSciences Production Corporation

11 Cabot Boulevard

Mansfield, MA 02048 USA

integralife.com

US Federal Law restricts this device to sale by or on the order of a physician only.

NONSTERILE