digital panel meter · asahi keiki co.,ltd. digital panel meter a1000 model code a1 1 - range power...
TRANSCRIPT
ASAHI KEIKI CO.,LTD.
DIGITAL PANEL METER
DC Voltage, current
Features
· Low cost
· Bright LED 14.2mm (red)
· AC power and DC power availableNEW!
A1000
Input Specifications
• DC voltage measurement (Range 12 to 14)Measurement
RangeRange InputImpedance
Accuracy(23°C±5°C, 35 to 85%)
InputProtection
±1.999V±19.99V±199.9V
1MΩ121314
±100V±250V±500V
± 0.3% of F.S
• DC current measurement (Range 23 to 25)Measurement
RangeRange InputResistance
Accuracy(23°C±5°C, 35 to 85%)
InputProtection
±19.99mA±199.9mA±1.999A
10Ω1Ω
0.1Ω
232425
±150mA±500mA
±3A ± 0.5% of F.S
AC Voltage, current
• AC voltage measurement (Range 12 to 14 Average)
MeasurementRangeRange Input
ImpedanceAccuracy
(23°C±5°C, 35 to 85%)Input
Protection1.999V19.99V199.9V
1MΩ121314
50V50V
500V ± 0.5% of F.S
MeasurementRangeRange Input
ResistanceAccuracy
(23°C±5°C, 35 to 85%)Input
Protection19.999mA199.9mA1.999A
5A
10Ω1Ω
0.1Ω1Ω
23242526
150mA500mA
3A8A
± 1.0% of F.S
Input Frequency: 40 Hz to 100HzRectifier Circuit: Reprsents effective value of sine wave by
demodulation of mean valueResponse time: approx 1 sec (10 to 90%)
Input Frequency: 40 Hz to 100HzRectifier Circuit: Reprsents effective value of sine wave by
demodulation of mean valueResponse time: approx 1 sec (10 to 90%)
• AC current measurement (Range 23 to 26 Average)
Thermocouple (Range KC, JC)Measurement
RangeRange SensorType
Accuracy(23°C±5°C, 35 to 85%)
MaxResolution
0 to +1000°C0 to +400°C
KJ
KCJC
1°C1°C ± 0.8% of F.S
Sensor lead Resistance: less than 50ΩBurn out: Flash -1999Over input voltage: ±5VDCCold Junction Compensator Accuracy:±2°C
RTD (Pt100Ω) (Range PA, PJ)Measurement
RangeRange SensorType
Accuracy(23°C±5°C, 35 to 85%)
MaxResolution
-100.0 to +199.9°C-200 to +600°C
Pt100ΩPt100Ω
PAPJ
0.1°C1°C
± 0.2% of F.S± 0.4% of F.S
Current for resistance: 1mA (TYP)External lead resistance: less than 1 ohm per lead
Process (Range 1V, 2A)Measurement
RangeRange InputImpedance
Accuracy(23°C±5°C, 35 to 85%)
InputProtection
1 to 5V4 to 20mA
1MΩ51Ω
1V2A
±100V50mA
± 0.3% of F.S± 0.4% of F.S
Display
offset ±1000fullscale ±100 to 1999
Common SpecificationInput bias current: 50 pA (TYP)Sampling speed: 2.5/secOver range indication: When input exceeds the maximum display
flash 1999Max. display: 1999Display: Red LED 14.2mm heightDimensions: 96mm (H) X 48mm (W) X 75mm (D) DIN sizePower supply: 90 to 132 VAC
180 to 264 VAC5 to 24 VDC
Power Consumption: 1.5 VA at AC power100mA at DC power
Operating Temp.: 0 to 50°C, 35 to 85%RHStorage Temp.: -10 to 70°C less than 60% RHDielectric Strength: Power supply and input, 1500 VAC/1min at
AC powerPower supply and input, 500 VDC/1min at DCpower
Insulattion Resistance : 500 VDC 100MΩ at above terminals
ASAHI KEIKI CO.,LTD.
DIGITAL PANEL METERA1000
Model Code
A11 -
Range
Power Supply 1. 90 to 132 VAC2. 180 to 264 VAC3. 5 to 24 VDC ±10%
Input: 1. DC voltage 2. DC current 4. AC voltage 5. AC current 6. AC current 5A B. Process C. Thermocouple D. RTD
Dimensions
Front View
Side View
96
48
538.9
7.8
8.7
Connection Diagram
12 r
ange
HI
13-1
4 ra
nge
HI
NC
LO
HO
LD
CO
M
PO
W (-
)
PO
W (+
)
23 r
ange
HI
24 r
ange
HI
25 r
ange
HI
LO
HO
LD
CO
M
PO
W (-
)
PO
W (+
)
+ - S
NC
HO
LD
CO
M
PO
W (-
)
PO
W (+
) A B C
NC
HO
LD
CO
M
PO
W (-
)
PO
W (+
)
1V r
ange
HI
2A r
ange
HI
NC
LO
HO
LD
CO
M
PO
W (-
)
PO
W (+
)
DC voltage (12 to 14 range)AC voltage (12 to 14 range)
DC current (23 to 25 range)AC current (23 to 25 range)
Thermocouple (KC, JC) RTD (PA, PJ)
Process
ASAHI KEIKI CO.,LTD.
DIGITAL PANEL METER
DC Voltage Measurement
DC Current Measurement
MeasurementRangeRange Input Impedance AccuracyInput Protection
±199.9mV±1.999V±19.99V±199.9V
100MΩ
1MΩ
11121314
±50V
±250V ± (0.1% of F.S)
MeasurementRangeRange Input Impedance AccuracyInput Protection
±1.999mA±19.99mA±199.9mA±1.999A
100Ω10Ω1Ω
0.1Ω
22232425
±50mA±150mA±500mA
±3A
± (0.2% of F.S)
Power supply Non-isolation typeVoltage: 5VDC±5%Consumption: 60mA
Isolation typeVoltage: 5 to 24VDCConsumption: 100mA
SpecificationInput configuration: Single EndedConversion rate: 2.5/secMax display: 1999Overrange indication : When input exceeds the maximum display,
0000 or -0000 flashesDisplay : Red LED, 10mmPolarity : A "-" is displayed automaticallyDecimal point : Settable to any digit position by front SWExternal control : HoldOperating Temp. : 0 to 50˚C 35 to 85% RHDimensions : 48mm(W)24mm(H)39.7mm(D) DIN sizeWeight : 40gDielectric strength : Power supply and input terminal at 500VDC/
1minCase and each terminals at 1500VAC/1min
Insulation resistance : 500VDC 100M at the above terminals
* Replacement for AP-200, AP-201A, AP-202A(no option)
Features
· LED hight 10mm
· Power supply 5VDC or 5 to 24VDC
· 3 1/2 digitNEW!
A2100
Ordering Code
A21 0 -
Range
Power Supply 1. non-isolation type (5VDC)2. isolation type (5 to 24VDC)
11. ±199.9mV12. ±1.999V13. ±19.99V14. ±199.9V22. ±1.999mA23. ±19.99mA24. ±199.9mA25. ±1.999A
Dimensions
22.2
24
Min65
48 40
4.5
Min35
45
Panel Cutout
ASAHI KEIKI CO.,LTD.
DIGITAL PANEL METER
DC Voltage MeasurementMeasurement
RangeRange Input Impedance AccuracyInput Protection
±1.999V±19.99V±199.9V
100MΩ
1MΩ
121314
±50V
±250V ± (0.1% of F.S)
Features
· LED hight 10mm
· Power supply 5 to 24VDC (Isolation)
· 4 1/2 digitNEW!
A2200
Ordering Code
A222 0 -
Range12. ±1.9999V13. ±19.999V14. ±199.99V
Dimensions
22.2
24
Min65
48 40
4.5
Min35
45
Panel Cutout
Power supply Isolation typeVoltage: 5 to 24VDCConsumption: 100mA
SpecificationInput configuration: Single EndedConversion rate: 2.5/secMax display: 19999Overrange indication : When input exceeds the maximum display,
0000 or -0000 flashesDisplay : Red LED, 10mmPolarity : A "-" is displayed automaticallyDecimal point : Settable to any digit position by front SWExternal control : HoldOperating Temp. : 0 to 50˚C 35 to 85% RHDimensions : 48mm(W)24mm(H)39.7mm(D) DIN sizeWeight : 40gDielectric strength : Power supply and input terminal at 500VDC/
1minCase and each terminals at 1500VAC/1min
Insulation resistance : 500VDC 100M at the above terminals
* Replacement for AP-241A, AP-242A, AP-244A(no option)
Input Specification (A5XXX-)• DC voltage, current
Range
11
Displayoffset ±9999full scale 0 to ±9999
Maximum Resolution
10 µV
Input Impedance
100 MΩ
Input Protection
±100 V
Input configuration : Single endedMeasuring method : ∆∑typeNormal mode rejection : More than NMR 50dB (50/60 Hz)
• AC voltage, current (Average)
• AC voltage, current (TRUE-RMS)
Range
111213
Measurementrange99.99 mV999.9 mV9.999 V
Input Frequency
40 Hz to 1 kHz
Display
offset ±9999full scale 0 to ±9999
MaximumResolution10 µV100 µV1 mV
InputImpedance
more than 1 MΩ
InputProtection100 V100 V250 V
Accuracy
±(0.2% of rdg + 20 digit)±(0.2% of rdg + 20 digit)±(0.2% of rdg + 20 digit)
Input configuration : Single endedResponse time : Approx. 1 sec.Crest factor : 4:1 at fullscale (only for TRUE-RMS)Dead zone : 0 to 99 digit
• Resistance
Range
11121314
Measurementrange99.99 Ω999.9 Ω9.999 kΩ99.99 kΩ
Display
offset ±9999full scale 0 to ±9999
Accuracy
±(0.1% of FS)
Range12131415
Measurement range±999.9 mV±9.999 V±99.99 V±600 V
Display
offset ±9999full scale 0 to ±9999
Maximum Resolution10 µV1 mV10 mV100 mV
Input Impedance100 MΩ1 MΩ10 MΩ10 MΩ
Input Protection±100 V±250 V±250 V±600 V
Accuracy±(0.1% of FS)±(0.1% of FS)±(0.1% of FS)±(0.15% of FS)
Range232425
Measurement range±9.999 mA±99.99 mA±999.9 mA
Display
offset ±9999full scale 0 to ±9999
Maximum Resolution1 µA10 µA100 µA
Input Impedance10 Ω1 Ω0.1 Ω
Input Protection±100 mA±500 mA±3 A
Accuracy
±(0.2% of FS)±(0.2% of FS)±(0.3% of FS)
Range
111213
Measurementrange99.99 mV999.9 mV9.999 V
Input Frequency
40 Hz to 1 kHz
Display
offset ±9999full scale 0 to ±9999
MaximumResolution10 µV100 µV1 mV
InputImpedance
more than 1 MΩ
InputProtection100 V100 V250 V
Accuracy
±(0.2% of rdg + 10 digit)±(0.2% of rdg + 10 digit)±(0.2% of rdg + 10 digit)
Range
1415
Measurementrange99.99 V600 V
Input Frequency
40 Hz to 1 kHz
Display
offset ±9999full scale 0 to ±9999
MaximumResolution10 mV100 mV
InputImpedance
more than 1 MΩ
Accuracy
±(0.2% of rdg + 10 digit)±(0.3% of rdg + 10 digit)
InputProtection250 V600 V
Range
232425
Measurementrange9.999 mA99.99 mA999.9 mA
Input Frequency
40 Hz to 1 kHz
Display
offset ±9999full scale 0 to ±9999
MaximumResolution1 µA10 µA100 µA
InputImpedance10 Ω1 Ω0.1 Ω
InputProtection100 mA500 mA3 A
Accuracy
±(0.5% of rdg + 10 digit)±(0.5% of rdg + 10 digit)±(0.5% of rdg + 10 digit)
Range
26
Measurementrange
5 A
Input Frequency
50 Hz or 60 Hz
Display
offset ±9999full scale 0 to ±9999
MaximumResolution
1 mA
InputImpedance
CT
InputProtection
8 A
Accuracy
±(0.5% of rdg + 10 digit)
Range
1415
Measurementrange99.99 V600 V
Input Frequency
40 Hz to 1 kHz
Display
offset ±9999full scale 0 to ±9999
MaximumResolution10 mV100 mV
InputImpedance
more than 1 MΩ
InputProtection250 V600 V
Accuracy
±(0.2% of rdg + 20 digit)±(0.3% of rdg + 20 digit)
Range
232425
Measurementrange9.999 mA99.99 mA999.9 mA
Input Frequency
40 Hz to 1 kHz
Display
offset ±9999full scale 0 to ±9999
MaximumResolution1 µA10 µA100 µA
InputImpedance10 Ω1 Ω0.1 Ω
InputProtection100 mA500 mA3 A
Accuracy
±(0.5% of rdg + 20 digit)±(0.5% of rdg + 20 digit)±(0.5% of rdg + 20 digit)
Range
26
Input Frequency
50 Hz or 60 Hz
Display
offset ±9999full scale 0 to ±9999
MaximumResolution
1 mA
InputImpedance
CT
InputProtection
8 A
Accuracy
±(0.5% of rdg + 20 digit)
Measurementrange
5 A
available Fahrenheit displayCold junction compensator accuracy : ±1°C (10 to 40°C)Sensor lead resistance : less than 50ΩLinearizing method : Digital linearizingBurn out alarm : -----
Range
KAKBJTSRB
SensortypeKKJTSRB
Measurementrange–50.0 to 199.9°C–50 to 1200°C–50 to 1000°C–50 to 400°C0 to 1700°C–10 to 1700°C100 to 1800°C
MaximumResolution0.1°C1°C1°C1°C1°C1°C1°C
Accuracy
±(0.5% of FS)±(0.2% of FS)±(0.2% of FS)±(0.6% of FS)±(0.4% of FS)±(0.4% of FS)±(0.4% of FS) over 500°C
• Thermocouple
Current
5 mA0.5 mA50 µA5 µA
Accuracy
±(0.2% of FS)±(0.1% of FS)±(0.1% of FS)±(0.1% of FS)
Maximum Resolution
10 mΩ100 mΩ1 Ω10 Ω
Measurement range
±99.99 mV
02
03
04
05
08
09
06
07
10
11
12
13
01
Output Specification
• RS-485 (Conforming to EIA RS-485) (A5XX4-XX)
Communication method : Full duplexTransmission speed : 2400/4800/9600/19200/38400 bpsStart bit : 1 bitData length : 7 bit/8 bitParity : Even parity/odd parityError detection : BCCStop bit : 1 bit/2 bitCharacter code : ASCII codeTransmission control process : Ignored processSignal name : + non reversal output
– reversal outputMaximum No. of meter connected : 31Line length : Up to 500 m in total
• RS-232C (Conforming to EIA RS-232C) (A5XX3-XX)
Communication method : Full duplexTransmission speed : 2400/4800/9600/19200/38400 bpsStart bit : 1 bitData length : 7 bit/8 bitParity : Even parity/odd parityStop bit : 1 bit/2 bitCharacter code : ASCII codeTransmission control process : Ignored process
Output0 to 1 V0 to 10 V1 to 5 V4 to 20 mA
Resistive loadmore than 10 kΩmore than 10 kΩmore than 10 kΩless than 550 Ω
Output method : PWM methodScaling : Digital scaling
Accuracy
±(0.5% of FS)
• Analog output (A5XX2-XX)
• HI & LO setpoint output (A5XX1-XX)
Setting range : –9999 to 9999Hysteresis : 1 to 999 digit for each setpointsRelay contact capacity : AC240V 8A resistive load DC30V 8A resistive load
Indication > High setpointHigh setpoint Indication Lo setpointIndication < Lo setpoint
HIGOLO
• RTD
available Fahrenheit displayCurrent for resistance : Approx. 1 mAExternal lead resistance : Less than 10 Ω/leadLinearizing method : Digital linearizingBurn out alarm : -----
14
• FrequencyRange
11121314
Measurement range0.1 to 200 Hz1 to 2000 Hz0.01 to 20 kHz0.1 to 200 kHz
Display
Prescale: 0.001 to 5 1 to 100
Maximum Resolution0.1 Hz1 Hz10 Hz100 Hz
Display Renewal time1 to 10 s1 s100 ms100 ms
Accuracy±(0.2% of FS)±(0.2% of FS)±(0.2% of FS)±(0.2% of FS)
Input typeOpen collectorLogicMagnet
Input voltage levelL: less than 1 V (5 V, 2.2 kΩ) pullupL: less than 1 V, HI: 2.5 to 15 V0.3 to 30 V P-P
Input typeVoltage
Input voltage level
50 to 500 V rms
Input Protection
500 V
Input Protection30 V15 V15 V
15
16
• Strain gaugePower supplyfor sensor
5 V10 V
Zero adjustment range
–0.3 to +2 mV/V
Span adjustment range
1 to 3 mV/V
Maximum Resolution
0.5 µV/digit1 µV/digit
Accuracy
±(0.1% of FS) +2 digit
Sensor : 350 ΩPower supply for sensor : 5 V±5% (less than 15 mA) 10 V±5% (less than 30 mA)
17
• Process with sensor power supplyRange
1 V2 A
Measurement range
1 to 5 V4 to 20 mA
Display
offset ±9999full scale 0 to ±9999
Input Impedance
1 MΩ10 Ω
Input protection
±100 V±100 mA
Accuracy±(0.2% of FS)±(0.2% of FS)
19
Power supply for sensor : 12 V DC ±10% 50 mA 24 V DC ±10% 25 mA
• ProcessRange
1 V2 A
Measurement range
1 to 5 V4 to 20 mA
Display
offset ±9999full scale 0 to ±9999
Input Impedance
1 MΩ10 Ω
Input protection
±100 V±100 mA
Accuracy±(0.2% of FS)±(0.2% of FS)
18
Sensor typePT100ΩPT100Ω
Measurement range–100.0 to 199.9°C–100 to 600°C
Maximum Resolution0.1°C1°C
Accuracy±(0.15% of FS)±(0.3% of FS)
RangePAPB
We reserve the right to change specifications without notice.
33-6, YAGUCHI 2-CHOME, OHTA-KU,TOKYO 146-8505, JAPANFAX : (03)3757-2989PHONE: (03)3579-3893http:// www.asahikeiki.co.jpE-mail: [email protected]
Printed in Japan, Aug. 2004 (KP)
45+
0.6
-0
92 +0.8- 0
MIN
.70
MIN.120
92
96
487
121
45
147.
5
Common SpecificationDisplay : Main display Red LED 14.2 mm height
Sub display Green LED 8 mm heightConversion rate : 12.5 times/sec.Maximum display : 9999Overrange indication : When input exceeds the maximum display,
display OL or –OLZero display : Leading zero suppressionDecimal point : Settable to any digit positionExternal control : Start/Hold, Peak Hold, Digital ZeroOperating temp. : 0 to 50°C 35 to 85% RHStorage temp. : –10 to 70°C less than 60% RHPower supply : AC 100 to 240V±10%(AC main unit)
DC 9 to 60 V (DC main unit)Power consumption : approx 4 VA (at 100 V)Dimensions : 96 mm (W) X 48 mm (H) X 147.5 mm (D) DIN sizeWeight : approx. 450 gDielectric strength (AC) : Power supply/input terminal/output terminal AC2000 V/1 min.
Input terminal/output terminal DC500 V/1 min.Case/power supply/input terminal output terminal AC2000 V/1 min.
Dielectric strength (DC) : Power supply/input terminal/output terminal DC500 V/1 min.Input terminal/output terminal DC500 V/1 min.Case/power supply/input terminal output terminal AC2000 V/1 min.
Insulation resistance : DC500 V more than 100 MΩ at the above terminals
Dimensions Ordering CodeA 5
Input
Output
Display board
Main board
01. DC voltage (±99.99 mV)02. DC voltage (±999.9 mV to ±600V)03. DC current (±9.999 mA to ±999.9 mA)04. AC voltage AVG (99.99 mV to 9.999 V)05. AC voltage AVG (99.99 V to 600 V)06. AC voltage RMS (99.99 mV to 9.999 V)07. AC voltage RMS (99.99 V to 600 V)08. AC current AVG (9.999 mA to 999.9 mA)09. AC current AVG (5 A)10. AC current RMS (9.999 mA to 999.9 mA)11. AC current RMS (5 A)12. Resistance (99.99 Ω to 99.99 kΩ)13. Temperature (Thermocouple)14. Temperature (RTD)15. Frequency (Open collector, Logic, Magnet)16. Frequency (50 to 500 Vrms)17. Strain gauge18. 1 to 5 V, 4 to 20 mA19. 1 to 5 V, 4 to 20 mA with sensor power supply
∗ Impossible to use only No.19 input for DC power supply
0. None1. HI & LO setpoint2. Analog output3. RS-232C4. RS-4855. HI & LO setpoint + analog output6. HI & LO setpoint + analog output + RS-232C7. HI & LO setpoint + analog output + RS-485
1. Single2. Multiple
1. AC100 - 240 V (±10%)2. DC9 - 60 V
Display single type (A6X2X-XX)Display single type (A6X2X-XX)
Display multi type (A6X1X-XX)Display multi type (A6X1X-XX)
Input Specifications DC voltage,current• A6XX1-XX
Minimum display 50 digits
Range Measurement Range Display Input Protection Accuracy (23˚C ± 5˚C, 35 ~ 85% RH)1112131415
±99.99 mV±999.9 mV±9.999 V±99.99 V±700 V
±50 V±50 V±250 V±250 V±700 V
offset ±9999full scale ±9999
±(0.03% of rdg + 1digit)
±(0.1% of rdg + 2digit)
• A6XX2-XXRange Measurement Range Display Input Protection Accuracy (23˚C ± 5˚C, 35 ~ 85% RH)
21222324
±99.99 uA±999.9 uA±9.999 mA±99.99 mA
±10 mA±10 mA±50 mA±500 mA
±(0.1% of rdg + 2 digit)offset ±9999full scale ±9999
• A6XX5-XXRange Measurement Range Display Input Protection Accuracy (23˚C ±5˚C, 35 ~ 85% RH)
21222324
99.99 uA999.9 uA9.999 mA99.99 mA
10 mA10 mA50 mA500 mA
±(0.5% of rdg + 20 digit)offset ±9999full scale ±9999
AC voltage, current (TRUE-RMS)• A6XX4-XX
Range Measurement Range Display Input Protection Accuracy (23˚C ± 5˚C, 35 ~ 85% RH)1112131415
99.99 mV999.9 mV9.999 V99.99 V700.0 V
50 V50 V250 V250 V700 V
offset ±9999 full scale ±9999
±(0.2% of rdg + 20 digit)
±(0.3% of rdg + 20 digit)
• A6XX3-XXRange Measurement Range Display Input Protection Accuracy (23˚C ± 5˚C, 35 ~ 85% RH)
2526
±999.9 mA±2.000 A
±3 Aoffset ± 9999full scale ±9999
±(0.1% of rdg + 2 digit)
• A6XX6-XXRange Measurement Range Display Input Protection Accuracy (23˚C ±5˚C, 35 ~ 85% RH)
2526
999.9 mA5 A
3 A8 A
offset ±9999full scale ±9999
±(0.7% of rdg + 20 digit)
Minimum display 50 digits
Measurement method: Two or Four wire (internal socket changable)
Resistance• A6XX7-XX
Range Measurement Range Display Current Accuracy (23°C ±5˚C, 35 ~ 85% RH)11121314
99.99 Ω999.9 Ω9.999 KΩ99.99 kΩ
5 mA0.5 mA50 uA5 uA
offset ±9999full scale ±9999
±(0.1% of rdg + 4digit)
Frequency• A6XX9-XX
Range Measurement Range Accuracy (23˚C ±5, 35 ~ 85% RH)111213
0.1 to 999.9 Hz1 to 9.999 kHz10 to 99.99 kHz
±(0.2% of FS)
Thermocouple • RTD• A6XX8-XX
Range Sensor type Measurement Range Maximum Resolution Accuracy (23˚C ±5˚C, 35 ~ 85% RH)KAKBJTSRB
PAPB
KKJTSRBPT100 ΩPT100 Ω
0.1˚C1˚C1˚C1˚C1˚C1˚C1˚C0.1˚C1˚C
-50.0 to 199.9˚C-50 to 1200˚C-50 to 1000˚C-50 to 400˚C0 to 1700˚C-10 to 1700˚C100 to 1800˚C-100.0 to 199.9˚C-100 to 600˚C
±(0.5% of FS)±(0.2% of FS)±(0.2% of FS)±(0.6% of FS)±(0.4% of FS)±(0.4% of FS)±(0.4% of FS)±(0.15% of FS)±(0.3% of FS)
Minimum display 50 digits
available Fahrenheit displayCold junction compensator accuracy : ±2˚C (10 to 40˚C)Sensor lead resistance : less than 50 ΩLinearzing method : Digital linearizingBurn out alarm : - - - - -Conversion time : 2.5/sec
• A6XX9-XXInput type Input voltage level Input ProtectionOpen collectorLogicMagnetVoltage
L: less than 1.5 V (5V, pullup)L: less than 1 V, HI: 2.5 to 15 V0.3 to 30 V P-P30 V rms to 500 V rms
30 V 15 V 30 V 500 V
Prescale : 0.01 ~ 10.00PPR : 1~100Power supply for sensor : 12 VDC ±10%
Strain gauge• A6XXA-XX
sensor Zero adjustment range Span adjustment range Measurement range Accuracy 5 V10 V
-1 to + 1 mV/V -4 ~ + 4 mV/V1 to 3mV/V ±(0.1% of FS) + 2 digit
Sesnor : 350 ΩPower supply for sensor : 5 V ± 5% (less than 15 mA)
10 V ± 5% (less than 30 mA)Conversion speed : 1000/sec
Process• A6XXB-XX
sensor Measurement Range Display Accuracy 1 V2 V2 A3 A
1 to 5 V±5 V4 to 20 mA±20 mA
offset: ±9999full scale 0 to ±9999
±(0.03% of rdg + 2digit)±(0.1% of rdg + 3digit)
Excitation power supply : 12 VDC ±5%24 VDC ±5% 25 mA
Output Specification
Start/Hold : Shorted between COM terminal and start/hold terminal or same level / holdOpen COM terminal and start/hold terminal or 5 V / start
D. ZERO : Shorted between COM terminal and D.ZERO terminal / ONP.HOLD : Shorted between COM terminal and P. HOLD terminal / ONR.RE (relay reset) : Shorted between COM terminal and R.RE terminal / ON P.SEL1 to 3 : as below
pattern 1OFFOFFOFF
P.SEL1 P.SEL2 P.SEL3
pattern 2ONOFFOFF
pattern 3OFFONOFF
pattern 4ONONOFF
pattern 5OFFOFFON
pattern 6ONOFFON
pattern 7OFFONON
pattern 8ONONON
ON : Shorted between COM terminal and P.SEL terminal or same levelOFF: Open between COM terminal and P.SEL terminal or 5 V
• HI & LO setpoints output (A6XXX-1X Relay output) (A6XXX-2X Photocoupler)
Comparative condition:Indication > High setpointHIgh setpoint Indication Lo setpointIndication < Lo setpoint
HIGOLO
Option• External control (A6XXX-X1)
Communication method :Full duplexTransmission speed :4800/9600/19200/38400 bpsStart bit :1 bitData length :7 bit/8 bitParity :Even parity/ odd parity/ noneStop bit :1 bit/2 bitCharacter code :ASCII codeTransmission control process :Ignord process
• RS232C (Conforming to EIA RS-232C)(A6XXX-X5)
Communication method : Full duplexTransmission speed : 4800/9600/19200/38400 bpsStart bit : 1 bitData length : 7 bit/ 8 bitParity : Even parity/ odd marity/ noneError detection : BCCStop bit : 1 bit/2 bitCharacter code : ASCII codeTransmssion control process : Ignord processSignal name : +non reversal output
-reversal outputMaximum no of meter connected : 31Line length : up to 500m in total
• RS-485 (conforming to EIA RS-485)(A6XXX-X6)
Type of output : Open collector or TTLLogic : ChangeableOutput rate : DC30 V 10 mA (open collector)
Funout 2 (TTL)
• BCD output (A6XXX-X2 TTL)(A6XXX-X3 Open collector)
Setting range : -9999 to 9999Hysteresis : 1 to 999 digit for each setpointsRelay contact capacity : AC250 V 2 A resistive load
DC30 V 2 A resistive loadPhotocoupler capacity : DC30 V 20 mA
• Analog output (A6XXX-X4)output Load resistance Accuracy Ripple
0 to 1 V 0 to 10 V 1 to 5 V
4 to 20 mA
more than 10 kΩ
less than 550 Ω
±(0.5% of F.S)50 mV p-p
25 mV p-p
Output method : PWM methodScaling : Digital scalingResolution : 13 bitReponse time : approx 0.5 sec
Common SpecificationDisplay : Main display Red or Green LED 20 mm
Sub display Red LED 6 mmConversion rate : 1000 times/sec.Maximum display : 9999Overrange indication : When input exceed the maximum display
display OVER or –OVERZero display : Leading zero suppressionDecimal point : Settable to any digit positionOperating temp. : 0 to 50°C 35 to 85% RHStorage temp. : –10 to 70°C less than 60% RHPower supply : AC 100 to 240 V ±10%
DC 12 to 48 V ±10%Power consumption : 8 VA (max. load)Dimensions : 96 mm (W) X 48 mm (H) X 97.5 mm (D) DIN sizeWeight : approx. 450 gDielectric strength (AC) : Power supply/input terminal/output terminal AC1500 V/min.
(DC) : Power supply/input terminal/output terminal DC500 V/min. (Common) : Case/each terminal AC1500 V/min.
Insulation resistance : DC500 V more than 100 MΩ at the above terminals
A 6
Option 0. none1. External control2. BCD output (TTL)3. BCD output (open collector)4. Analog output5. RS-232C6. RS-4857. RS-232C, analog output8. RS-485, analog output
Comparator output 0. none1. Relay2. Photocoupler
Input 1. DC voltage2. DC current3. DC current4. AC voltage5. AC current6. AC current7. Resistance8. Temperature (T/C, RTD)9. FrequencyA. Strain gaugeB. Process
Displayb board 1. Multiple2. Single
Power supply 1. AC 100 to 240 V2. DC12 to 48 V
We reserve the right to change specifications without notice.
33-6, YAGUCHI 2-CHOME, OHTA-KU,TOKYO 146-8505, JAPANFAX : (03)3757-2989PHONE: (03)3579-3893http:// www.asahikeiki.co.jpE-mail: [email protected]
Printed in Japan, Dec. 2003 (YG)
Muliple
Input Option
Comparator Power
Single
48
(97.
5)
10.9
77
0.5
(86.
5)
(21.
5)
44.8
91.5
102.5
45+
0.6
-0
92 +0.8- 0
MIN
.70
MIN.120
Dimensions Ordering Code
Input Specifications DC voltage,current• A7X11-X
Minimum display 50 digitsFrequency range: 40Hz to 1kHzResponse time: Approx 1 sec(10% to 90%)Conversion rate: 1000times/sec
Range Measurement Range Display Input Protection Accuracy (23˚C ± 5˚C, 35 ~ 85% RH)1112131415
±99.99 mV±999.9 mV±9.999 V±99.99 V±700 V
±50 V±50 V±250 V±250 V±700 V
offset ±9999full scale ±9999
±(0.03% of rdg + 1digit)
±(0.1% of rdg + 2digit)
• A7X12-XRange Measurement Range Display Input Protection Accuracy (23˚C ± 5˚C, 35 ~ 85% RH)
21222324
±99.99 uA±999.9 uA±9.999 mA±99.99 mA
±10 mA±50 mA±50 mA±700 mA
±(0.1% of rdg + 2 digit)offset ±9999full scale ±9999
• A7X15-XRange Measurement Range Display Input Protection Accuracy (23˚C ±5˚C, 35 ~ 85% RH)
21222324
99.99 uA999.9 uA9.999 mA99.99 mA
10 mA10 mA50 mA500 mA
±(0.5% of rdg + 20 digit)offset ±9999full scale ±9999
AC voltage, current (TRUE-RMS)• A7X14-X
Range Measurement Range Display Input Protection Accuracy (23˚C ± 5˚C, 35 ~ 85% RH)1112131415
99.99 mV999.9 mV9.999 V99.99 V700.0 V
50 V50 V250 V250 V700 V
offset ±9999 full scale ±9999
±(0.2% of rdg + 20 digit)
±(0.3% of rdg + 20 digit)
• A7X13-XRange Measurement Range Display Input Protection Accuracy (23˚C ± 5˚C, 35 ~ 85% RH)
2526
±999.9 mA±2.000 A
±3 Aoffset ± 9999full scale ±9999
±(0.1% of rdg + 2 digit)
• A7X16-XRange Measurement Range Display Input Protection Accuracy (23˚C ±5˚C, 35 ~ 85% RH)
2526
999.9 mA5 A
3 A8 A
offset ±9999full scale ±9999
±(0.7% of rdg + 20 digit)
Minimum display 50 digitsFrequency range: 40Hz to 1kHzResponse time: Approx 1 sec(10% to 90%)Conversion rate: 1000times/sec
Conversion rate: 1000times/sec
Conversion rate: 1000times/sec
Conversion rate: 1000times/sec
Conversion rate: 100times/secMeasurement method: Two or Four wire (internal socket changable)
Resistance• A7X17-X
Range Measurement Range Display Current Accuracy (23°C ±5˚C, 35 ~ 85% RH)11121314
99.99 Ω999.9 Ω9.999 KΩ99.99 kΩ
5 mA0.5 mA50 uA5 uA
offset ±9999full scale ±9999
±(0.1% of rdg + 4digit)
Thermocouple • RTD• A7X18-X
Range Sensor type Measurement Range Maximum Resolution Accuracy (23˚C ±5˚C, 35 ~ 85% RH)KAKBJTSRB
PAPB
KKJTSRBPT100 ΩPT100 Ω
0.1˚C1˚C1˚C1˚C1˚C1˚C1˚C0.1˚C1˚C
-50.0 to 199.9˚C-50 to 1200˚C-50 to 1000˚C-50 to 400˚C0 to 1700˚C-10 to 1700˚C100 to 1800˚C-100.0 to 199.9˚C-100 to 600˚C
±(0.5% of FS)±(0.2% of FS)±(0.2% of FS)±(0.6% of FS)±(0.4% of FS)±(0.4% of FS)±(0.4% of FS)±(0.15% of FS)±(0.3% of FS)
Minimum display 50 digitsFrequency range: 40Hz to 1kHzResponse time: Approx 1 sec(10% to 90%)Conversion rate: 1000times/sec
available Fahrenheit displayCold junction compensator accuracy : ±2˚C (10 to 40˚C)Sensor lead resistance : less than 50 ΩLinearzing method : Digital linearizingBurn out alarm : - - - - -Conversion time : 2.5times/sec
Strain gauge• A7X1A-X
sensor Zero adjustment range Span adjustment range Measurement range Accuracy 5 V10 V
-1 to + 1 mV/V -4 ~ + 4 mV/V1 to 3mV/V ±(0.1% of FS) + 2 digit
Frequency• A7X19-X
Range Measurement Range Accuracy (23˚C ±5, 35 ~ 85% RH)111213
0.1 to 999.9 Hz1 to 9.999 kHz10 to 99.99 kHz
±(0.2% of FS)
Input type Input voltage level Input ProtectionOpen collector Logic Magnet Voltage
L: less than 1.5 V (5V, pullup)L: less than 1 V, HI: 2.5 to 15 V0.3 to 30 V P-P30 V rms to 500 V rms
30 V 15 V 30 V 500 V
Prescale : 0.01 ~ 10.00PPR : 1~100Power supply for sensor : 12 VDC ±10%
Sesnor : 350 ΩPower supply for sensor : 5 V ± 5% (less than 15 mA)
10 V ± 5% (less than 30 mA)Conversion speed : 1000/sec
Process• A7X1B-X
sensor Measurement Range Display Accuracy 1 V2 V2 A3 A
1 to 5 V±5 V4 to 20 mA±20 mA
offset: ±9999full scale 0 to ±9999
±(0.03% of rdg + 2digit)±(0.1% of rdg + 3digit)
Conversion rate : 1000times/secExcitation power supply : 12 VDC ±5%
24 VDC ±5% 25 mA
Output Specification• HH, HI & LO, LLsetpoints outputComparative condition:
Indication > High High setpointHigh High setpoint Indication > High setpointHigh setpoint Indication Lo setpointLow setpoint > Indication Low Low setpointLow Low setpoint > Indication
HHHI
GOLOLL
Communication method :Full duplexTransmission speed :4800/9600/19200/38400 bpsStart bit :1 bitData length :7 bit/8 bitParity :Even parity/ odd parityStop bit :1 bit/2 bitCharacter code :ASCII codeTransmission control process :Ignord process
• RS232C (Conforming to EIA RS-232C)
Communication method : Full duplexTransmission speed : 4800/9600/19200/38400 bpsStart bit : 1 bitData length : 7 bit/ 8 bitParity : Even parity/ odd marityError detection : BCCStop bit : 1 bit/2 bitCharacter code : ASCII codeTransmssion control process : Ignord processSignal name : +non reversal output
-reversal outputMaximum no of meter connected : 31Line length : up to 500m in total
• RS-485 (conforming to EIA RS-485)
Type of output : Open collector or TTLLogic : ChangeableOutput rate : DC30 V 10 mA (open collector)
Funout 2 (TTL)
• BCD output
Setting range : -9999 to 9999Hysteresis : 1 to 999 digit for each setpointsRelay contact capacity : AC125 V 0.3 A resistive load
DC30 V 1 A resistive loadPhotocoupler capacity : DC30 V 50 mA
• Analog output (PWM)output Load resistance Accuracy Ripple
0 to 1 V 0 to 10 V 1 to 5 V
4 to 20 mA
more than 10 kΩ
less than 550 Ω
±(0.5% of F.S)±50 mV p-p
±25 mV p-p
Output method : PWM methodScaling : Digital scalingResolution : 13 bitReponse time : approx 0.5 sec
Common SpecificationMaximum display : 9999Overrange indication : When input exceed the maximum display
display OVER or -OVERZero display : Leading zero suppressionDecimal point : Settable to any digit positionOperating temp : 0 to 50°C (35 to 85% RH)Storage temp. : -10 to 70°C less than 60%RHPower supply : 100 to 240VAC ± 10%
12 to 48VDCPower consumption : 8VA (At AC) 7W(At DC)Dimensions : 72mm (W) × 36mm (H) × 118mm (D)Weight:approx : 160gDielectric strength (AC) : Power supply/input terminal/output terminal AC1500 V/min
(DC) : Power supply/input terminal/output terminal DC500 V/min: Input terminal/output terminal DC500V/min
Insulation resistance : DC500V more than 100MΩ at the above terminals
We reserve the right to change specifications without notice.
3rd floor, Ehara bldg., 5-7, Kaji-cho 1-chome, Chiyoda-ku,TOKYO 101-0044, JAPANFAX : (03)3251-5576PHONE: (03)3251-5567http:// www.asahikeiki.co.jpE-mail: [email protected]
Printed in Japan, Feb. 2006 (KS)
Dimensions Ordering CodeA 7 1 –
Output
Input
0. none1. Relay output2. Photocoupler output3. BCD output (TTL)4. BCD output (Open collector)5. BCD output (TTL), Photocoupler output6. BCD output (Open collector), Photocoupler
output7. Analog output8. Analog output, Photocoupler output9. RS-232CA. RS-485B. RS-232C, Photocoupler outputC. RS-485, Photocoupler output
1. DC voltage2. DC current3. DC current4. AC Voltage5. AC current6. AC current7. Resistance8. Temperature (T/C, RTD)9. FrequencyA. Strain gaugeB. Process
Power supply 1. AC100 to 240V2. DC12 to 48V
68
80
87
72
36
7
118
33
HHHIGOLOLL
REPHDZMEP1P2P3
ME
Min.96
68+0.7-0
33-0 +0
.6
Input Specifications
DC voltage , current
A83 1-0 (DC voltage)
A83 2-0 (DC current)
AC voltage, current (TRUE-RMS)
A83 4-0 (AC voltage)
A83 5-0 (AC current)
Process
A83 B-0
ASAHI KEIKI CO.,LTD.
UNIVERSAL TYPE DIGITAL PANEL METERA8000
Featuers
· Power Supply : 5V to 12VDC, 12V to 24VDC
· BCD Output (option)
· 4 1/2 digit
NEW!
Accuracy(23°C ± 5°C, 35 to 85%RH)
± (0.1% of rdg + 2 digit)
range
11121314
MeasurementRange
± 199.99mV± 1.9999V± 19.999V± 199.99V
Display
offset ± 19999full scale ± 19999
InputProtection
± 50V± 50V
± 250V± 250V
Accuracy(23°C ± 5°C, 35 to 85%RH)
± (0.2% of rdg + 2 digit)
range
22232425
MeasurementRange
± 1.9999mA± 19.999mA± 199.99mA± 1999.9mA
Display
offset ± 19999full scale ± 19999
InputProtection
± 50mA± 50mA
± 3A± 3A
Accuracy(23°C ± 5°C, 35 to 85%RH)
± (0.2% of rdg + 20 digit)
range
11121314
MeasurementRange
199.99mV1.9999V19.999V199.99V
Display
offset ± 19999full scale ± 19999
InputProtection
50V50V250V250V
Frequency range : 40Hz to 1kHzResponse time : Approx 1 sec (10% to 90%)Dead band : 100 digit
Accuracy(23°C ± 5°C, 35 to 85%RH)
± (0.5%of rdg + 20 digit)
range
232425
MeasurementRange
19.999mA199.99mA1999.9mA
Display
offset ± 19999full scale ± 19999
InputProtection
50mA3A3A
Frequency range : 40Hz to 1kHzResponse time : Approx 1 sec (10% to 90%)Dead band : 100 digit
Accuracy(23°C ± 5°C, 35 to 85%RH)
± (0.1% of rdg + 3digit)± (0.2% of rdg + 3digit)
range
1V2V
MeasurementRange
1 to 5V4 to 20mA
Display
Offset : ± 19999Fullscale ± 9999
InputProtection
± 50V± 50mA
Excitation power supply : 24VDC ± 10% 30mA
Accuracy(23°C ± 5°C, 35 to 85%RH)
± (0.5% of FS)± (0.2% of FS)± (0.2% of FS)± (0.6% of FS)± (0.4% of FS)± (0.4% of FS)
± (0.4% of FS) applied over 500°C
range
KAKBJTSRB
Sensortype
KKJTSRB
MeasurementRange
-50.0 to 199.9°C-50 to 1200°C-50 to 1000°C-50 to 400°C0 to 1700°C
-10 to 1700°C100 to 1800°C
MaximumResolution
0.1°C1°C1°C1°C1°C1°C1°C
Cold junction compensator accuracy : ± 2°C (10 to 40°C)Sensor lead resistance : less than 50ΩLinearzing method : Digital linearizingBurn out alarm : - - - - -
Thermocouple • RTD
A83 C-0 (Thermocouple)
A83 D-0 (RTD)
Output Specification
BCD outputType of output: Open collector or TTLLogic: ChangeableOutput rate: DC30V 10mA (open collector)
Funout 2 (TTL)
External ControlHold: Shorted HOLD terminal and COM terminal or
equipotential, Hold ONDigital Zero: Shorted DZ terminal and COM terminal or equipotential,
Digital zero ONPeak Hold: Shorted PH terminal and COM terminal or equipotential,
Peak hold ON
Common Specification
Display: Red LED 10mm heightMaximum display: -19999 to 19999Conversion rate: 20times/secOverrange indication: When input exceed the maximum
display display OL or -OLZero display: Leading zero suppressionDecimal point: Settable to any digit positionOperating temp: 0 to 50°C (35 to 85% RH)Storage temp: -10 to 70°Cless than 60%RHPower supply: 5V to 12VDC, 12V to 24VDC Power consumption: Appro x 1.5WDimensions: 48mm (W) x 24mm (H) x 96mm (D)
DIN sizeWeight (unit only) : Appro x 100gDielectric strength: Power supply/input terminal/output
terminal/external control DC500V/minInput terminal/output terminal/external control DC500V/minCase/each terminal AC1500V/min
Insulation resistance: DC500V more than 100MΩ at the above terminals
Accuracy(23°C ± 5°C, 35 to 85%RH)
± (0.15% of FS)± (0.3% of FS)
range
PAPB
Sensor type
PT100ΩPT100Ω
MeasurementRange
-100.0 to + 199.9°C-100 to + 600°C
MaximumResolution
0.1°C1°C
available Fahrenheit display
Ordering Code
Dimensions
ASAHI KEIKI CO.,LTD.
With sheet switch48.0
24.0
Without sheet switch48.0
24.0
89.1
3.5
Front View
SideView
A8 – 0Option 1. none
2. BCD (TTL), external control3. BCD (Open collector), external control
Input
Front sheet
Power supply
1. DC voltage2. DC current4. AC voltage5. AC currentB. ProcessC. ThermocoupleD. RTD
1. with sheet switch2. without print sheet swtich
3. 5V to 12VDC4. 12V to 24VDC
Printed in Japan, Jun. 2005 (KP)
We reserve the right to change specifications without notice.
3rd Fl., Ehara Bldg., 5-7, Kaji-cho 1-chome, Chiyoda-ku, Tokyo 101-0044, JapanTEL:+81-3-3251-5576FAX:+81-3-3251-5567http:// www.asahikeiki.co.jpE-mail: [email protected]
UNIVERSAL TYPE DIGITAL PANEL METERA8000
ASAHI KEIKI CO.,LTD.
Proc
ess
Mon
itor
Ordering Code
Dimensions
Connection Diagram
LOHIOV +5V
D.P
COM
Connect input Lo and Ov.
Power
INPUT
HO
LD
1 2 3 4 5 6 7 8 9 10
101 102 103
48
73
514.8
24
Fullscale Adjust
Offset Adjust
Min.65
Min
.3522
.2+
0.3
–0
45+0.6–0
Front
Panel thickness 0.8 to 3.2mm
Side
Panel Cutout
AS-201 -
1V2V3VRange
DC Voltage Measurement
Accuracy: ±0.1% rdg ±2 digit (23°C ±5°C)
SpecificationsInput Configuration: Single EndedInput Bias Current: 2nA(Typ.)Conversion Rate: 2.5/secNormal Mode Rejection: 40dB Typ. (50/60Hz)Temperature Coefficient: ±0.1 digit (Typ.)/°C
±0.3 digit (Max.)/°CDisplay: LED, 10.2mm 3 1/2 digitPolarity: 1V Range:When input less than 1V, a "-" is
displayed automatically.2V, 3V Range:When input less than 0V, a "-"is displayed automatically.
Decimal Point: Settable to any digit positionOverrange Indication: When input exceeds the maximum display,
1999 flashesFull-scale Adjustment: 100~1999External Hold: Negative 0V or ShortPower Supply: DC 5V±5% 70mA (Typ.)Operating Temperature: 0~50°C, 35 to 85%RHPower Consumption: 350mWDimensions: 24(H) 48(W) 73(D)mm DIN SizeWeight: 40gDielectric Strength: Between Input(Lo) and Mounted panel,
AC1500V/1 min.Insulation Resistance: DC500V 100MΩ or more at above terminals
1~5V0~5V
0~10V
Model Code RangeInput
ImpedanceInput
Protection
AS-201-1VAS-201-2VAS-201-3V
1MΩ1MΩ1MΩ
±250V±250V±250V
0-100~19990-100~19990-100~1999
DisplayAdjustable
PROCESS MONITORAS-201
Features• Adjustable full scale
• Bright LED 10.2mm
• Din size 24(H) 48(W)
• 1-5V, 0-5V, 0-10V measurement
• Power supply DC5V (NO-ISOLATED)
Fullscale Adjustable
ASAHI KEIKI CO.,LTD.
PROCESS MONITORAS-243
Standard 4 1/2 Digit
Dimensions
Connection Diagram
Front
Side
Panel Cutout
Panel thickness 0.8 to 3.5mm
24
48
87
655
45
Min.65
Min
.35
+0.6-0
22.2+
0.3
-0
1 2
LOHI NC NC NCNCCOMS/H - +
DC POWERINPUT
3 4 5 6 7 8 9 10
Features
•Easily adjustable offset-fullscale (Span 10000 counts)
•Bright LED, 8 mm (Red)
•DIN size 24 (H) 48 (W) mm
•Power supply DC 24 V (isolated)
•1-5 V, 0-5 V, 0-10 V, 4-20 mA measurement
DC Voltage Measurement
General SpecificationsInput configuration : Single endedInput bias current : 2 mA (TYP.)Conversion rate : 2.5/secNormal mode rejection : 50dB (TYP.) (50/60 Hz)Temp. coefficient : Offset display ±0.5 digit/˚C
Fullscale display ±0.01% of span/˚COverrange indication : When input exceeds the maximum display,
display flashesFullscale adjustment : 100 to 9999Offset a djustment : ±5000Span : 10000 countsDisplay : LED, 8 mm (Red)Polarity : A "-" is displayed automaticallyScaling : By switch (rear) and fine volume (front)Hold : COM and S/H terminal shortedDecimal point : Settable to any digit positionOperating temperature : 0 to 50˚C 35 to 85%RHPower supply : DC24V±20%Power consumption : 30 mA at DC24VDimensions : 24(H) 48(W) 87(D)mm DIN sizeWeight (unit) : Approx. 50gDielectric strength : Between input (Lo) and mounted panel, AC
1500V/l min.Between input (Lo) and power supply (0V), DC500 V/l min.
Insulation resistance : DC 500 V 100 MΩ or more at the aboveterminals
Accessory : Connector, Instruction manual
Model Range Display Input Impedance Input ProtectionAS-243-1V 1-5 V Offset ±5000 1MΩ ±250VAS-243-2V 0-5V fullscale 1MΩ ±250VAS-243-3V 0-10 V 100 to 9999 1MΩ ±250V
Accuracy: ±(0.1%of rdg+2 digit) (23˚C±5˚C 45~75%RH)
DC Current MeasurementModel Range Display Internal Resistance Input Protection
AS-243-2A 4-20mA Offset ±5000 50Ω ±50mAfullscale100 to 9999
Accuracy: ±(0.1% of rdg+2 digit) (23˚C±5˚C 45~75%RH)
Ordering Code
AS-243 4-
Power Supply
-
Range
4. DC24V
1V2V3V2A