digital light meter
DESCRIPTION
DIGITAL LIGHT METER. 1994. LIGHT METER. CT Lutron LX-105. Dr. Lim Jac Fang Occupational Health Doctor Department of Health, Sabah. FEATURES. Microprocessor circuit ensure high accuracy an also provide special function and features - PowerPoint PPT PresentationTRANSCRIPT
DIGITAL LIGHT METER
Dr. Lim Jac Fang
Occupational Health Doctor
Department of Health, Sabah
1994
CT Lutron LX-105
LIGHT METER
FEATURES
Microprocessor circuit ensure high accuracy an also provide special function and features
Super large LCD display with contrast adjustment for viewing angel.
Dual function display
Heavy duty & compact case
Record Minimum, Maximum & Average readings
Data Hold
Auto power off saves battery life
Operates from 9V battery
FEATURES (cont..)
RS 232 PC serial interface
Spectrum of photo sensor meets C.I.E
Wide range measurement both LUX & foot candle units
Relative % light measurement
User selectable lighting type (Tungsten, Fluorescent, Daylight or Mercury)
Zero adjustment by push button
GENERAL SPECIFICATIONS
Circuit – Custom one-chip microprocessor LSA circuit
Display – 13mm with contrast adjustment
Lighting type – Daylight, Tungsten, Fluorescent, Mercury
Measurement range – LUX 0- 50,000 LUX , 3 ranges
- Foot candle 0 – 5,000 ft-cd, 3 ranges
- Relativity – 0 – 1999%
Sensor – Photo diode & color correction filter
Memory – Max, Min and Average readings
Sample Time – Approx 0.4 sec
Zero Adj. – By push button
Power off – manual off or auto shut off (after 10 min)
GENERAL SPECIFICATIONS (cont…)
Data output – RS 232 PC serial
Operating temperature - 0°C - 50°C (32°F - 122°F)
Operating Humidity – Max 80% RH
Power supply - 06P DC 9V battery (heavy duty)
Weight – 335g/0.77IB (Included batteries)
Size – 180 x 72 x 32 mm (7.1 x 2.8 x 1.3 inch)
Accessories - instruction manual
- Sensor cover
DIGITAL LIGHT METER
2001
CT Lutron LX-105
LIGHT METER
1
2
34
567
89 10
16
11
12
13
14
151- Display
2- Power Off/On
3- Data Hold Button
4- LUX/FC Button
5- Memory Record Button
6- Memory Call Button
7- Light Source Button
8- Zero button
9- % Button
10- Range Switch
11- LCD contrast adjust
12- Light Sensor
13- Sensor cover
14- Light sensor plug
15- Light sensor input socket
16- RS-232 Output
17- Battery compartment cover
17
MEASURING PROCEDURE
Push Power Off/On Button
Zero Adjust Procedures1. Cover the light sensor-using the Sensor Cover
2. Slide the Range Switch to the 2000 LUX Position
3. Push the Zero Button-display will show zero values
4. Upon completion, remove the sensor cover
Select the desired measuring unit by pressing the LUX/FC Button
Determine the lighting type by pressing the Light Source Select Button
L= Tungsten, F= Fluorescent, S= Day Light, C= Mercury
Select the max. Range using the Range Switch
If display shows “- - - -”, it indicates an overload condition, select the next higher range
If display shows “- - - -”, it indicates an out of range, select the next lower range
Position the Light Sensor directly under light source
-On the 20000 LUX range, the last digit will be shown on the lower line of LCD display
-On 50000 LUX range, the last two digit will be shown on the lower line of LCD display
For example:-
On the 20000 LUX range, if display show
1562 LUX
0
Data Hold
During measurement, pushing the “Data Hold Button will hold the display values and the LCD will show “D.H” symbol
To cancel the Data Hold function, press the Data Hold Button, once more
Relative % light measurement
During measurement, press the “%” Button. The current measured value will indicated as 100%
All the subsequent measurement will be indicated as percentage, relative to the value when the button was pressed
The formula used is as shown below:The new light valuesThe light values when the % button was pressed X 100
To de-activated- press “%” Button
Data Record (Max., min & Average reading
The DATA RECORD function displays the max, min & Average readings.
To start the Data Record function, press the “Record Button” once
With the “REC” symbol indicated on display
C. Push the “CALL Button” once more – average (AVG) values
A. Push the “CALL Button” once - maximum values (max)
B. Push the “CALL Button” once again – minimum (Min) values
To de-activate the Data Record function, Press the “Record Button” once again
BATTERY REPLACEMENT
When the left corner LCD display show “LBT”
-spec measurement may still be made for several hours after low battery indicator appears before the instrument become inaccurate
THANK YOU FOR YOUR ATTENTION