diagnostics instrument target product profile...(home, mobile, and community clinics), postpartum...
TRANSCRIPT
Diagnostics Instrument–Target
Product Profile
Diagnostic Instrument: Hemoglobinometer
Version 1.1–June 2018
Mailing Address
PO Box 900922
Seattle, WA 98109 USA
Street Address
2201 Westlake Avenue
Suite 200
Seattle, WA 98121 USA
www.path.org
i
Contents
EXECUTIVE SUMMARY .............................................................................................................................. 1
PRODUCT USE SUMMARY / MEDICAL NEED / DIFFERENTIATION STRATEGY ................................. 4
TARGET POPULATIONS AND USE CASES .............................................................................................. 5
TARGET PERFORMANCE ........................................................................................................................ 10
PRODUCT CONFIGURATION AND FORMAT .......................................................................................... 13
PRODUCT COSTS AND CHANNELS TO MARKET................................................................................. 17
REFERENCES ............................................................................................................................................ 19
CHANGE MANAGEMENT ......................................................................................................................... 22
1
Executive summary
Variable Minimum Optimistic
Product use
Intended use
The diagnostic device is intended to measure blood hemoglobin (Hb) concentration. The test will provide results expressed in grams Hb per deciliter and must clinically differentiate mild, moderate, and severe anemia.
Same.
Target population
Pregnant women receiving care at antenatal care (ANC) clinics.
Pregnant women receiving ANC in all settings, postpartum women (0–6 months after delivery), and prepregnancy girls and women.
End users Trained facility-based health care workers: nurses, medical officers, and midwives.
Trained facility- and community-based health care workers.
Infrastructure level
ANC clinics in community health facilities, district health facilities and hospitals, and regional hospitals. Tier 4 health care settings and above (WHO heath facility classifications).
All heath care settings where women seek care during the pre- and postnatal period, including through mobile clinics and home visits.
Target countries All countries categorized by WHO as having severe (>40% prevalence) or moderate (20%–39% prevalence) anemia public health problems.
All countries where antenatal services are provided in settings without access to full blood-count testing.
Product performance
2
Variable Minimum Optimistic
Clinical sensitivity and specificity
Sensitivity >85% and specificity >80% in each trimester of pregnancy at the moderate anemia threshold of 11.0g/dL and the severe anemia threshold point of 7.0g/dL.
Sensitivity >95% and specificity >90% in each trimester of pregnancy at the moderate anemia threshold of 11.0g/dL and the severe anemia threshold point of 7.0g/dL.
Precision Coefficient of variation (CV) should not exceed 6%. To be determined.
Analytical accuracy relative to reference method
In comparison to the standard reference, 80% of the results must fall within +1.0g Hb/dL of the reference across the range of Hb concentrations.
In comparison to the standard reference, 85% of the results must fall within +1.0g Hb/dL of the reference across the range of Hb concentrations.
Operating temperature The device should be able to operate at 15oC–35oC, 30%–85% humidity.
The device should be able to operate at 15oC–40oC, 20%–90% humidity.
Reference methods
Standard reference method for capillary or venous blood: 1. HemoCue®* (capillary blood samples). 2. Automated hematology analyzer (venous blood sample).
Both standard reference methods for capillary or venous blood.
Product design
Complexity No more than 5 steps. Requires one health worker to execute.
No more than 3 steps. Requires one health worker to execute.
Instrumentation Handheld, portable reader device and disposable/ consumable component (less than 500 grams).
Handheld portable device with no consumables.
3
Variable Minimum Optimistic
Sample type Capillary blood from fingerstick and venous whole blood collected in anticoagulant.
Noninvasive device: none.
Sample collection and processing Finger prick using a standardized lancet or venipuncture with no sample processing.
Noninvasive sensor probe: none.
Quality control Manufacturer provides controls specific for test in intended use
Compatible with commercial hemoglobin controls. Clear instructions for quality controls appropriate for intended use.
Costs and channels to market
Target cost per result Less than US$0.75 cents. Includes the consumable cost plus the reader cost pre result.
To be determined.
Channels to market
Through UNICEF Supply Division, Unitaid, and the Global Financing Facility. Additionally, local distributors to procure the consumable components locally.
Same.
*HemoCue is a registered trademark of HemoCue AB, Ängelholm, Sweden.
Note: UNICEF, United Nations Children’s Fund; WHO, World Health Organization.
4
Product use summary / medical need / differentiation strategy
Clinical and/or surveillance need:
Improving the nutritional status of women before and during pregnancy can reduce the risk of adverse birth outcomes. Anemia is a significant
driver of poor nutritional status and the global prevalence of anemia among pregnant women is about 38%.1 Clinical assessment is a common
but inaccurate method for detecting anemia. Laboratory-based methods, such as an automated hematology analyzer, are unsuitable for use in
antenatal care (ANC) settings. Currently available point-of-care tests are considered expensive, and the high cost can prohibit routine use.
There is a need for a point-of-care (POC) hemoglobinometer for routine screening of pregnant women for anemia in all health settings where
women seek ANC. More accurate and reliable anemia-detection tools will support better clinical decision-making, help stratify pregnant women
according to their risk of anemia-related pregnancy complications, and potentially improve provider and patient compliance with recommended
anemia treatment options, such as iron-folic acid supplementation.
Intended use case scenario:
A low-cost POC hemoglobinometer will be used for routine screening in ANC clinics. The 2016 World Health Organization (WHO)
recommendations on ANC for a positive pregnancy experience recommends routine anemia testing at weeks 12, 26, and 36 in countries where
the prevalence of anemia is at least 40% among pregnant women.1 The test will evaluate hemoglobin (Hb) concentration in capillary blood as
part of routine screening at ANC and at delivery. The test will identify women with all severity levels of anemia (mild, moderate, severe) and
provide accurate results to inform treatment options and monitor hemoglobin concentration throughout pregnancy. Trained clinicians and
facility-based health workers, including nurses and midwives as well was trained community health workers, will use the test in all health
settings where women seek ANC, primarily in ANC clinics.
5
Target populations and use cases
Target Annotation
3.1 Intended Use
The diagnostic device is intended to measure blood hemoglobin
(Hb) concentration. The test will provide results expressed in
grams Hb per deciliter and must clinically differentiate mild,
moderate, and severe anemia.
A robust, low-cost, accessible, and accurate point-of-care (POC) Hb
measurement tool would enable routine anemia screening within the
antenatal care (ANC) programs in low- and middle-income countries
(LMIC).
This diagnostic must be able to discriminate between healthy women and
women with mild anemia, moderate anemia, and severe anemia with
sufficient accuracy to support clinical decision-making. While this diagnostic
must identify all cases of anemia (<11.0g/dL) to ensure adequate treatment
and follow-up is received, the priority is to identify all cases of severe
anemia (<7.0g/dL). The results will be used to inform treatment options,
including an iron and folic acid (IFA) supplementation regimen. Depending
on the severity of anemia and the time of testing, women may receive IFA
supplements, counseling, and treatment regarding malaria or deworming
and iron intravenous or intramuscular injections. The result may also be
used to facilitate referral for treatment at a higher-tier health facility.
ANC is delivered in both the public and private health sectors.2 ANC is
commonly provided in primary health facilities. As such, the test will need to
provide a result at the POC and be no more complex than current
components of ANC care, such as blood pressure measurement or malaria
rapid tests.
Screening may be achieved through invasive tests that require a blood
sample or noninvasive tests that require no sample. Some literature and
clinical experts report that noninvasive devices have significant
performance limitations that limit their utility in clinical decision-making.3–5
6
3.2 Target Population
Minimum: Pregnant women receiving care at ANC clinics.
Optimistic: Pregnant women receive care in all ANC settings
(home, mobile, and community clinics), postpartum women (0–6
months after delivery), and prepregnancy girls and women.
The primary target is pregnant women undergoing routine ANC visits at
primary clinics in LMIC. The 2016 WHO recommendations on ANC for a
positive pregnancy experience recommends routine anemia testing at
weeks 12, 26, and 36 in countries where the prevalence of anemia is at
least 40% among pregnant women.1 If there is no Hb result on record, Hb
should be checked when pregnant women arrive at a birthing facility for
delivery in order to understand risk of postpartum haemorrhage.6
In addition to routine ANC, the test may also be used outside of clinic
settings and in populations before and after pregnancy. Stakeholders
interviewed reported an interest in expanding anemia screening to the
prepregnancy and postnatal period and through screening in school
settings and during postpartum care.6,7
3.3 End Users
Minimum: Facility-based health care workers: nurses, medical
officers, and midwives.
Optimistic: Facility and community-based health care workers.
End users of the test will vary across countries and facility types.6 All users
will receive training in the use of the test along with other components of
routine screening in ANC. ANC services at facilities are generally delivered
by nurses, medical officers, clinical officers, and midwives. A formative
usability assessment in a representative ANC setting in Ghana found that
end users ranged from community health nurses to laboratory technicians
with between 2 and 5 years of medical training.
Optimistically, a POC Hb measurement tool would be sufficiently robust
and simple to use outside of the ANC clinic setting. This would expand
access to anemia testing in the prepregnancy and postnatal period through
home visits and community- and school-based screening and entail use by
lower-skilled community health workers.
3.4 Infrastructure Level This diagnostic will be used at primary care clinics and other health facilities
where ANC is provided, such as district hospitals or facilities with limited or
7
Minimum: ANC clinics in community health facilities, district
health facilities and hospitals, and regional hospitals. Tier 4 health
care settings and above (WHO heath facility classifications).
Optimistic: All heath care settings where women seek care
during the pre- and postnatal period, including through mobile
clinics and home visits.
no laboratory capacity. Full blood-count testing is recommended in facilities
where this level of laboratory capacity is present.1
Optimistically, a POC Hb measurement tool would be sufficiently robust
and simple to use outside of the ANC clinic setting. This would expand
access to anemia testing in the prepregnancy and postnatal period through
home visits and community- and school- based screening and entail use by
lower-skilled community health workers.
3.5 Target Countries
Minimum: All countries categorized by WHO as having severe
(>40% prevalence) or moderate (20%–39% prevalence) anemia
public health problems.
Optimistic: All countries where antenatal services are provided
in settings without access to full blood-count testing.
All countries with severe or moderate prevalence of anemia in pregnant
women would benefit from a POC Hb measurement tool to expand access
to screening in ANC and reduce the burden of maternal anemia. Globally,
in countries with a lower prevalence of anemia, a POC hemoglobin
measurement tool will expand access to anemia screening in settings
without access to full blood-count testing.
As a target launch country, India is well suited in terms of health need and
policy alignment. The prevalence of anemia during pregnancy in India is
one of the highest globally, estimated at 59%.8 Operational guidelines to
reduce the prevalence of anemia in India among all at-risk groups include
the use of digital hemoglobinometers, indicating an opportunity to introduce
and scale hemoglobin measurement tools.9
8
Geographic summary of anemia prevalence in pregnant women
globally.(10)
3.6 Target Customer
Public and private clinics that provide ANC services. In order for
diagnostics to be procured and used in public ANC clinics, they must be
registered in-country and be included on lists of products that can be
covered by the national health insurance program. A hemoglobinometer is
included in the World Health Organization Model List of
Essential In Vitro Diagnostics.
Additional comments:
Fit with clinical workflow / linkage to care.
A low-cost POC hemoglobinometer would enable routine screenings of Hb
concentration among pregnant women at 12, 26, and 36 weeks, as per
WHO recommendations, and be offered for free as part of ANC.1 There are
9
some reports of women being asked to pay for the cost of screening in a
laboratory.11 Anemia screening and use of the hemoglobinometer will be
done in parallel with other recommended ANC screening practices included
in global and national guidelines.1 Health workers using the diagnostic need
to be able to read the result of the test and interpret it in the context of the
relevant anemia thresholds and other clinical indicators and measurements
taken during ANC.
Providers will conduct the test at the POC. The result must be available
immediately and used to communicate the hemoglobin measurement value
and anemia status. The result will be used to inform treatment options,
including iron and folic acid (IFA) supplementation regimen. Depending on
the severity of anemia and the time of testing, women may receive IFA
supplements, counseling, and treatment regarding malaria or deworming
and iron intravenous or intramuscular injections. The result may also be
used to facilitate referral for treatment at a higher-tier health facility.
10
Target performance
Target Annotation
4.1 Clinical Sensitivity and Specificity
Minimum: Sensitivity >85% and specificity >80% in each
trimester of pregnancy at the moderate anemia threshold of
11.0g/dL and the severe anemia threshold point of 7.0g/dL.
Optimistic: Sensitivity >95% and specificity >90% in each
trimester of pregnancy at the moderate anemia threshold of
11.0g/dL and the severe anemia threshold point of 7.0g/dL.
This diagnostic must discriminate between healthy individuals, mild anemia,
moderate anemia, and severe anemia with sufficient accuracy to support
clinical decision-making, as in routine ANC there will be limited or no access
to confirmatory testing.
Evidence generated in multiple studies and reviewed by WHO found that
some POC hemoglobin measurement techniques have sufficient performance
to improve detection of anemia in pregnancy.12–14 Expert interviews indicated
that sensitivity is prioritized over specificity to identify cases of anemia and, in
particular, all cases of severe anemia.6
Some evidence suggests that Hb measurement tools, particularly noninvasive
tools, may have insufficient performance—specifically, decreased accuracy in
the second or third trimester due to changes in circulatory status.3 There may
be a need for further evidence to inform the risk and the cost of both over-
and underestimating anemia in pregnant women.
4.2 Precision
Minimum: Coefficient of variation (CV) should not exceed 6%.
Optimistic: To be determined.
Hb concentration in capillary blood likely varies more than in venous blood
due to the variable presence of interstitial fluids. This variation may be
compounded by varying finger-prick techniques that may impact precision
measurements.14 The standard reference for capillary measurement—the
HemoCue®*—reports CV ranging from 0.51-1.3%.15
This variation may be addressed through repeat measurements, which may
be facilitated by multiple ANC visits over the course of the pregnancy.
11
Minimal and optimistic precision criteria may warrant further study. The
criteria reported here take into account the precision estimates of the
standard reference for capillary measurement. The optimal precision criteria
should be a judgement of the corresponding clinical risk associated with
missing someone with severe anemia or the cost to the health system of
overestimating anemia prevalence or severity.
4.3 Analytical Accuracy Relative to Reference Method
Minimum: In comparison to the standard reference, 80% of
the results must fall within +1.0g Hb/dL of the reference across
the range of hemoglobin concentrations.
Optimistic: In comparison to the standard reference, 85% of
the results must fall within +1.0g Hb/dL of the reference across
the range of hemoglobin concentrations.
The clinically acceptable bias is within 1.0g/dL of the reference across the
dynamic range of Hb concentrations (4 – 17g/dL).4,16
The POC results should be strongly correlated with reference standard with a
correlation coefficient >0.80.16–19
Mean difference (bias) should be less than or equal to +0.5g/dL. Absolute
mean difference should be less than 1.0g/dL in each trimester.3,4,16,17,20,21
4.4 Limit of Detection
Minimum: Clinically relevent threshold of 4g Hb/ dL or lower.
Optimistic: Same.
This diagnostic must be able to discriminate between healthy women and
women with mild anemia, moderate anemia, and severe anemia to inform
linkage to care. Global and national clinical guidelines indicate that
measurements below 4 grams Hb/ dL do not indicate different clinical
management strategies. The hemoglobin measurements below 4 g/ dL
require blood transfusion.22–24
4.5 Time to Result
Minimum: 5 minutes or less.
Optimistic: 2 minutes or less.
The time to result should be comparable with other diagnostics used in ANC.
This may include blood pressure measurements and rapid diagnostic tests for
HIV or malaria, which entail a 10–15 min time to result. The HemoCue®
requires 15–60 seconds to a result.15
4.6 Throughput
Minimum: 1 sample per test run.
In an ANC setting, a provider will see one patient at a time. The volume of
patients attending ANC will vary across facilities. One study in a
12
Optimistic: Same.
representative ANC setting found that a provider may see up to 8 women per
day and thus run up to 8 tests per day.25
4.7 Operating Temperature
Minimum: The device should be able to operate at 15oC–
35oC, 30%–85% humidity.
Optimistic: The device should be able to operate at 15oC–
40oC, 20%–90% humidity.
In a study monitoring the temperature and humidity at surveillance sites in
four African countries, the mean and median operating temperatures were
26.6°C and 26.4°C, respectively, with a range of 17.6°C–37.0°C.26 For the
same sites the mean and median relative humidity values were 70.4% and
72.1%, with a range of 40.2% to 119.5%.26 These are similar to expected
conditions in ANC clinics in intended-use settings.
Recent field testing in ANC clinics suggest that technologies used in ANC
need to be robust to extreme environmental conditions, including
temperatures of about 35.0°C. Certain models of the HemoCue® for POC
capillary testing are more robust to extreme temperatures than others.27
*4.8 Reference Methods
Minimum: Standard reference method for capillary or venous
blood:
1. HemoCue (capillary blood samples).
2. Automated hematology analyzer (venous blood sample).
Optimistic: Both standard reference methods for capillary or
venous blood.
The test should be comparable to reference standing using both capillary
blood (HemoCue®) and venous blood (automated analyzer).28 There is
extensive data demonstrating the accuracy of the HemoCue® compared to
the reference automated hematology analyzer.29 However, sufficient
differences are also evident and, if possible, both methods should be used as
reference methods to triangulate.
*HemoCue is a registered trademark of HemoCue AB, Ängelholm, Sweden.
13
Product configuration and format
Target Annotation
5.1 Complexity
Minimum: No more than 5 steps. Requires only one health worker to operate. Optimistic: No more than 3 steps. Requires only one health worker to operate.
If an invasive test is used, the trained health care worker will use an alcohol
wipe to clean the finger. A lancet will be used to prick the finger to obtain a
capillary blood sample, applied to the test device. The hemoglobin diagnostic
will then measure the blood sample. A noninvasive test will require fewer
steps but may require navigating digital interface. The device must be
appropriate for the setting with respect to cost, resources required, human
capacity, and usability/acceptability.12 Ease of use is a critical feature for
widespread adoption of diagnostic technologies in low-resource settings.
5.2 Instrumentation Minimum: Handheld, portable reader device and disposable/ consumable component for individual measurement. Optimistic: Handheld portable device with no consumables.
Instrumentation for an invasive test includes a reader plus consumables, a
strip or a cuvette. Instrumentation for noninvasive test includes a digital
device either with or without a probe with a sensor. If a noninvasive test is
used, the instrument might be a mobile platform that includes other
diagnostics, clinical support tools, or decision-making algorithms. The
complexity of clinical algorithms in antenatal care (ANC) settings are well
suited to translation to a digital platform and could improve ANC service
delivery and health outcomes.30 Instruments must be robust devices, resilient
to high heat, humidity, and dusty environments.
5.3 Power Requirements Minimum: AC adapter for running on power with battery backup. Optimistic: Same.
Some ANC facilities do not have electricity and will require battery-operated
devices that can operate during all ANC hours (sometime more than 8 hours
per day).25
5.4 Quality Control Minimum: Manufacturer provides controls configured appropriately for intended use.
Some commercially available controls may require cold chain or other
infrastructure. Controls need to be able to be used in the same setting by the
same target end user of the test.
14
Optimistic: Compatible with commercial hemoglobin controls and clear instructions for use, appropriate for use in low- and middle-income countries (LMIC) ANC settings.
5.5 Time to Result
Minimum: 5 minutes or less.
Optimistic: 2 minutes or less.
The time to result should be comparable with other diagnostics used in ANC
settings. This may include blood pressure measurements and rapid
diagnostic tests for HIV or malaria. The standard reference for capillary
measurement requires 15–60 seconds to a result.15
5.6 Throughput
Minimum: 1 sample per test run.
Optimistic: Same.
In an ANC setting, a provider will see one patient at a time. The volume of
patients attending ANC will vary across facilities. One study in a
representative ANC setting found that a provider may see up to 8 women per
day and thus run up to 8 tests per day, though others may see significantly
more.25
5.7 Sample Type Minimum: Capillary blood from fingerstick and venous whole blood collected in anticoagulant. Optimistic: Noninvasive device: none.
In order to demonstrate performance and validate tests for the ANC intended
use, invasive devices must be run using capillary blood collected through
fingerstick. Noninvasive devices will require no specimen and can more easily
obtain repeat measurements, suggested in some testing protocols.14
However, some literature and clinical experts report that noninvasive devices
have significant performance limitations that limit their utility in clinical
decision-making. More performance data on noninvasive devices is needed.3–
5
5.8 Sample Collection and Processing Minimum: Finger prick or venipuncture with no sample processing. Optimistic: Noninvasive sensor probe: none.
Test should be resilient to varying finger-prick and blood-application
techniques between users.14
5.9 Sample Volume Minimum: One blood drop (10uL). Optimistic: Noninvasive: none.
The size of the blood drop must not influence the hemoglobin concentration
result. Alternatively, the device must have a mechanism to standardize the
amount of blood read by the device. PATH experiments with 2, 5, 10, and
20 uL indicated that the size of the blood drop may influence the hemoglobin
measurement. If blood-drop size does influence the result, then the device
15
should include an error message to alert the user that the volume was
insufficient.
5.10 Target Analyte Minimum: Hemoglobin concentration Optimistic: Same
Expert interviews indicated that clinicians would greatly benefit from a
diagnostic that, in addition to Hb measurement, provided additional
information regarding the root cause of anemia among ANC patients. This
might include infection, malaria, and/or hemoglobinopathies or dietary
deficiencies of iron, folate, and vitamin A.
5.11 Calibration Minimum: Manufacturer provides all required calibration and specifics when and how device must be calibrated. Optimistic: No calibration needed.
Readers with disposable strips may need to be recalibrated when a new lot is
used or at other regular intervals. A digital device may require no calibration.
5.12 Data Handling Minimum: None. Optimistic: Up to 500 previous test results would be stored on device and downloaded via USB or other connection.
Data-management method will need to be compatible with local record
system. Many settings use an ANC booklet kept by patient that tracks
pregnancy health record during and between pregnancies. The device output,
including any digital version of the results, should be compatible with local
systems.
5.13 Other Supplies Needed Minimum: Lancets and alcohol swabs. Optimistic: None.
Apart from supplies to conduct blood sample collection, the kit should contain
all necessary supplies for running the test.
5.14 Waste Disposal Minimum: Does not include material that cannot be disposed of in a normal ANC setting. Optimistic: No biohazard waste for disposal.
Disposable strips and other testing supplies could be disposed of using routine methods at primary care clinics.
16
5.15 Target Shelf Life Minimum: For disposable component: 3 months upon opening container at test operating range. No cold chain required. Optimistic: None.
The minimum and optimistic criteria for shelf life will vary depending on whether the test is invasive and requires a reader plus consumables or noninvasive and requires a digital device and possible a sensor/probe. More information is needed to inform these requirements across varying contexts. The shelf life of the HemoCue® (HemoCue AB, Ängelholm, Sweden) is 3 months upon opening container at test operating range with no cold chain requirements. Stakeholder interviews indicate that this is acceptable and aligned with patient volumes in most target settings. A noninvasive test would not have any shelf life restrictions.
17
Product costs and channels to market
Target
Annotations
6.1 Target Cost per Result
Minimum: Less than US$0.75 per test. Includes the consumable cost plus the reader cost per result. Optimistic: To be determined.
Given the test use as part of routine screening at multiple time points during
pregnancy, cost is a critical factor. Expert interviews and global guidance
acknowledges that proposed use may entail high recurrent costs, and
alternatives to a hemoglobinometer may be used if high cost would prohibit
routine and regular screening.1
The estimated cost per result using a digital hemoglobinometer is $0.75
(HemoCue®; HemoCue AB, Ängelholm, Sweden), which, given the recurrent
cost of screening at multiple timepoints, may be too costly for routine use in
ANC.31 Stakeholders interviewed indicated that the cost per result should be in
the $0.02–$0.07 range.32,33 The willingness of manufacturers to meet such low
prices points in a relatively undefined market must also be carefully considered.
The total cost of the test would include the cost of the reader plus any
consumable component. For invasive tests, this would include any disposable
strip or cuvette, and for noninvasive tests, this would include any sensor or
probes needed. More costing and cost-effectiveness data is needed to
understand the minimal and optimistic criteria for test cost. In particular, the
different cost structures between an invasive and noninvasive method should be
better understood.
6.2 Stability Requirements, Storage Conditions, Cold
Chain Requirements
See 5.15 Target Shelf Life.
18
Minimum: See 5.15 Target Shelf Life. Optimistic: See 5.15 Target Shelf Life.
6.3 Channels to Market Minimum: Approved by relevant global bodies (i.e., UNICEF to enable distribution by UNICEF Supply Division). Any consumables must be available through local distributors in each target country to ensure a consistent supply. Optimistic: Same as well as ability to procure and distribute through the Global Financing Facility (GFF).
Once approved, certified, and accepted by UNICEF, it would be the preferred
method, distributed by UNICEF Supply Division. Procurement through GFF.
Consumables must be available locally in order to ensure a consistent supply.
Two core attributes identified as necessary for hemoglobinometer uptake
included user training on test use and maintenance and local availability of
disposable components.11
6.4 Supply, Service, and Support Mechanisms
Minimum: Service and support supplied via distribution channels, either directly from NGOs procuring tests or from their agents/distributors. Optimistic: No service or support required.
Delivery of service and support may come through the product distribution channels. Ideally, the device would not require external service or support for the duration of use. Clear training materials should be made widely available.
19
References
1. World Health Organization. WHO Recommendations on Antenatal Care for a Positive Pregnancy
Experience. Geneva: World Health Organization; 2016. 152 p.
2. Campbell OMR, Benova L, MacLeod D, Baggaley RF, Rodrigues LC, Hanson K, et al. Family
planning, antenatal and delivery care: cross-sectional survey evidence on levels of coverage and
inequalities by public and private sector in 57 low- and middle-income countries. Tropical Medicine &
International Health. Apr. 2016;21(4):486–503.
3. Yoshida A, Saito K, Ishii K, Azuma I, Sasa H, Furuya K. Assessment of noninvasive, percutaneous
hemoglobin measurement in pregnant and early postpartum women. Medical Devices (Auckl).
2014;7:11–16.
4. Skelton VA, Wijayasinghe N, Sharafudeen S, Sange A, Parry NS, Junghans C. Evaluation of point-of-
care haemoglobin measuring devices: a comparison of Radical-7TM pulse co-oximetry, HemoCue(®)
and laboratory haemoglobin measurements in obstetric patients.* Anaesthesia. Jan. 2013;68(1):40–
45.
5. Butwick A, Hilton G, Carvalho B. Non-invasive haemoglobin measurement in patients undergoing
elective Caesarean section. British Journal of Anaesthesia. Feb. 2012;108(2):271–277.
6. Avery L. Hemoglobinometer Target Product Profile - Stakeholder Interviews. 2018.
7. Smith J. Hemoglobinometer Target Product Profile - Stakeholder Interviews. 2018.
8. Rahman MM, Abe SK, Rahman MS, Kanda M, Narita S, Bilano V, et al. Maternal anemia and risk of
adverse birth and health outcomes in low- and middle-income countries: systematic review and meta-
analysis. American Journal of Clinical Nutrition. Feb. 2016;103(2):495–504.
9. Ministry of Health and Family Welfare. Intensified National Iron Plus Initiative (I-NIPI): Operational
Guidelines for Programme Managers. New Delhi: Ministry of Health and Family Welfare, Government
of India; Apr. 2018: 76. (Anemia Mukt Bharat).
10. WHO. The Global Prevalence of Anaemia in 2011. Geneva: World Health Organization; 2015:43
[cited 2018 Jun 19]. Available at
http://www.who.int/entity/nutrition/publications/micronutrients/global_prevalence_anaemia_2011/en/in
dex.html.
11. PATH. Assessing the TrueHb Hemometer for Maternal Care: A Usability Evaluation in Kintampo,
Ghana [DRAFT]. Seattle: PATH; June 2018.
12. Sobhy S, Rogozinska E, Khan KS. Accuracy of on-site tests to detect anemia during prenatal care.
International Journal of Gynecology & Obstetrics. Nov. 2017;139(2):130–136.
13. van den Broek NR, Ntonya C, Mhango E, White SA. Diagnosing anaemia in pregnancy in rural
clinics: assessing the potential of the Haemoglobin Colour Scale. Bulletin of the World Health
Organization. 1999;77(1):15–21.
14. Morris SS, Ruel MT, Cohen RJ, Dewey KG, de la Brière B, Hassan MN. Precision, accuracy, and
reliability of hemoglobin assessment with use of capillary blood. American Journal of Clinical
Nutrition. Jun. 1999;69(6):1243–1248.
20
15. HemoCue. Package Insert for Hb201+ Cuvettes. Ängelholm, Sweden: HemoCue AB; 2016. Available
at https://hemocuepacific.com/blogs/news/103937286-package-insert-for-hb-201-cuvettes.
16. Parker M, Han Z, Abu-Haydar E, Matsiko E, Iyakaremye D, Tuyisenge L, et al. An evaluation of
hemoglobin measurement tools and their accuracy and reliability when screening for child anemia in
Rwanda: A randomized study. PLoS ONE. Jan. 4, 2018;13(1):e0187663.
https://doi.org/10.1371/journal.pone.0187663.
17. Neufeld L, García-Guerra A, Sánchez-Francia D, Newton-Sánchez O, Ramírez-Villalobos MD,
Rivera-Dommarco J. Hemoglobin measured by Hemocue and a reference method in venous and
capillary blood: a validation study. Salud Pública de México. Jun. 2002;44(3):219–227.
18. Hudson-Thomas M, Bingham KC, Simmons WK. An evaluation of the HemoCue for measuring
haemoglobin in field studies in Jamaica. Bulletin of the World Health Organization. 1994;72(3):423–
426.
19. Boghani S, Mei Z, Perry GS, Brittenham GM, Cogswell ME. Accuracy of Capillary Hemoglobin
Measurements for the Detection of Anemia among U.S. Low-Income Toddlers and Pregnant Women.
Nutrients. Mar. 9, 2017;9(3).
20. Nkrumah B, Nguah SB, Sarpong N, Dekker D, Idriss A, May J, et al. Hemoglobin estimation by the
HemoCue® portable hemoglobin photometer in a resource poor setting. BMC Clinical Pathology. Apr.
21, 2011;11:5.
21. Hiscock R, Simmons S, Carstensen B, Gurrin L. Comparison of Massimo Pronto-7 and HemoCue Hb
201+ with laboratory haemoglobin estimation: a clinical study. Anaesthesia and Intensive Care. Sep.
2014;42(5):608–613.
22. Carson JL, Kleinman S. Indications and hemoglobin thresholds for red blood cell transfusion in the
adult. UpToDate website. Available at https://www.uptodate.com/contents/indications-and-
hemoglobin-thresholds-for-red-blood-cell-transfusion-in-the-adult. Accessed Jun 20, 2018.
23. National Blood Service, Ghana. National Guidelines for the Clinical Use of Blood and Blood Products
in Ghana. Accra, Ghana: NBS, Ghana; Jun. 2013: 57. Available at http://nbsghana.org/wp-
content/uploads/2015/07/National-Guidelines-for-the-Clinical-Use-of-Blood-and-Blood-Products-e-
version.pdf.
24. Muñoz M, Peña‐Rosas JP, Robinson S, Milman N, Holzgreve W, Breymann C, et al. Patient blood
management in obstetrics: management of anaemia and haematinic deficiencies in pregnancy and in
the post-partum period: NATA consensus statement. Transfusion Medicine. 28(1):22–39.
25. Magoma M, Requejo J, Merialdi M, Campbell OM, Cousens S, Filippi V. How much time is available
for antenatal care consultations? Assessment of the quality of care in rural Tanzania. BMC
Pregnancy Childbirth. Sep. 24, 2011;11:64.
26. Golden A, Stevens EJ, Yokobe L, Faulx D, Kalnoky M, Peck R, et al. A Recombinant Positive Control
for Serology Diagnostic Tests Supporting Elimination of Onchocerca volvulus. PLoS Neglected
Tropical Diseases. Jan. 8, 2016;10(1):e0004292.
27. Mandava U, Drammeh BS, Zhang M, Schleicher RL, Pfeiffer CM. Three models of HemoCue
photometers: comparison of venous blood hemoglobin concentrations and influence of heat on
cuvettes. FASEB Journal. Apr. 1, 2010;24(1_supplement):538.8-538.8.
21
28. Pullum T, Collison DK, Namaste S, Garrett D. Hemoglobin Data in DHS Surveys: Intrinsic Variation
and Measurement Error (DHS Methodological Reports 18). Rockville, Maryland: Demographic and
Health Surveys Program; Aug. 2017. Available at https://dhsprogram.com/pubs/pdf/MR18/MR18.pdf.
29. Briggs C, Kimber S, Green L. Where are we at with point- of- care testing in haematology? British
Journal of Haematology. 2012;158(6):679–690. Available at https://doi.org/10.1111/j.1365-
2141.2012.09207.x.
30. Feroz A, Perveen S, Aftab W. Role of mHealth applications for improving antenatal and postnatal
care in low and middle income countries: a systematic review. BMC Health Services Research. Nov
7, 2017;17:704. doi:10.1186/s12913-017-2664-7.
31. Medina Lara A, Mundy C, Kandulu J, Chisuwo L, Bates I. Evaluation and costs of different
haemoglobin methods for use in district hospitals in Malawi. Journal of Clinical Pathology. Jan.
2005;58(1):56–60.
32. Saxena R. Hemoglobinometer Target Product Profile - Stakeholder Interviews. 2018.
33. Liljestrand J. Hemoglobinometer Target Profile – Stakeholder Interview. 2018.
22
Change management
Version Key changes from previous version
1.0
First draft of the TPP for point-of-care hemoglobinometers.
1.1
June 29,
2018
PATH edits based on literature review and stakeholder interviews.