development of a technique using vmat and robust

178
Development of a technique using VMAT and robust optimisation to replace the use of surface bolus during radiotherapy for patients post- mastectomy. A thesis submitted to The University of Manchester for the degree of Doctor of Clinical Science in the Faculty of Biology, Medicine and Health 2021 Helen P. Howard School of Medical Sciences Division of Cancer Sciences

Upload: others

Post on 20-Dec-2021

8 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Development of a technique using VMAT and robust

Development of a technique using VMAT and robust optimisation to replace the use of surface bolus during radiotherapy for patients post-

mastectomy.

A thesis submitted to The University of Manchester for the degree of Doctor of Clinical Science in the Faculty of Biology, Medicine and Health

2021

Helen P. Howard

School of Medical Sciences Division of Cancer Sciences

Page 2: Development of a technique using VMAT and robust

2

Contents

1 Introduction ............................................................................................................ 21

1.1 Cancer ............................................................................................................... 21

1.2 Breast Anatomy ................................................................................................ 22

1.3 Breast Cancer.................................................................................................... 22

1.3.1 Breast Cancer Diagnosis ............................................................................ 23

1.3.2 Breast Cancer Treatment .......................................................................... 25

1.4 Radiotherapy Treatment .................................................................................. 26

1.5 Radiotherapy Treatment Planning ................................................................... 27

1.6 Radiotherapy Volume Definitions .................................................................... 29

1.7 Breast Planning Volumes .................................................................................. 30

1.8 Radiotherapy Planning Techniques .................................................................. 31

1.8.1 Radiotherapy Planning Techniques for Breast Treatments ...................... 32

1.9 Radiotherapy Treatment Delivery .................................................................... 34

1.10 Radiotherapy Treatment Planning for Mastectomy Patients ...................... 34

1.11 Use of IMRT and VMAT Treatment Planning for Mastectomy Patients ....... 37

1.12 Robust Optimisation ..................................................................................... 39

1.13 Specifying Superficial Doses for Post-Mastectomy Radiotherapy ............... 41

1.14 Definition of the Skin Structure in Mastectomy Patients ............................. 42

1.15 Measurement of Superficial Doses ............................................................... 43

1.16 Scope of Project ............................................................................................ 46

2 Evaluation of current treatment method ............................................................... 48

2.1 Current technique for the treatment of post-mastectomy patients ............... 48

2.2 Dosimetric effect of bolus in post-mastectomy patients ................................. 52

2.2.1 Patient selection........................................................................................ 52

Page 3: Development of a technique using VMAT and robust

3

2.3 Method ............................................................................................................. 54

2.4 Results .............................................................................................................. 56

2.4.1 Plan comparison – Dose Distribution (single patient example) ................ 56

2.4.2 Plan comparison – PTVtoSurface (single patient example) ...................... 57

2.4.3 Plan comparison – Skin Structures (single patient example).................... 58

2.4.4 Plan comparison – Organs at Risk (single patient example) ..................... 60

2.4.5 Plan comparison – PTVtoSurface (8 patient study set) ............................ 62

2.4.6 Plan comparison – Skin Structures (8 patient study set) .......................... 64

2.4.7 Plan comparison – Organs at Risk (8 patient study set) ........................... 65

2.5 Discussion ......................................................................................................... 67

2.6 Summary ........................................................................................................... 68

3 Comparison of VMAT plans to Clinical plans .......................................................... 69

3.1 Method ............................................................................................................. 69

3.2 Results .............................................................................................................. 70

3.2.1 Plan comparison – Dose Distribution (single patient example) ................ 70

3.2.2 Plan comparison – PTVtoSurface (single patient example) ...................... 71

3.2.3 Plan comparison – Skin Structures (single patient example).................... 73

3.2.4 Plan comparison – Organs at Risk (single patient example) ..................... 75

3.2.5 Plan comparison – PTVtoSurface (8 patient study set) ............................ 77

3.2.6 Plan comparison – Skin Structures (8 patient study set) .......................... 78

3.2.7 Plan comparison – Organs at Risk (8 patient study set) ........................... 80

3.3 Discussion ......................................................................................................... 83

3.4 Summary ........................................................................................................... 84

4 Surface Dose Measurements .................................................................................. 85

4.1 Method ............................................................................................................. 86

4.2 Results .............................................................................................................. 90

4.3 Discussion ......................................................................................................... 94

Page 4: Development of a technique using VMAT and robust

4

4.4 Summary ........................................................................................................... 95

5 Effect of Perturbation ............................................................................................. 96

5.1 Method ............................................................................................................. 96

5.2 Results .............................................................................................................. 98

5.2.1 Perturbation Effect – Dose Distribution (single patient example) ........... 98

5.2.2 Perturbation Effect – PTVtoSurface (single patient example) ................ 100

5.2.3 Perturbation Effect – Skin Structures (single patient example) ............. 103

5.2.4 Perturbation Effect – Organs at Risk (single patient example) ............... 105

5.2.5 Perturbation Effect – PTVtoSurface (8 patient study set) ...................... 107

5.2.6 Perturbation Effect – Skin Structures (8 patient study set) .................... 112

5.2.7 Perturbation Effect – Organs at Risk (8 patient study set) ..................... 115

5.3 Discussion ....................................................................................................... 118

5.4 Summary ......................................................................................................... 122

6 Robust Optimisation ............................................................................................. 123

6.1 Method ........................................................................................................... 123

6.2 Results ............................................................................................................ 124

6.2.1 Plan comparison – Dose Distribution (single patient example) .............. 125

6.2.2 Plan comparison – PTVtoSurface (single patient example) .................... 125

6.2.3 Plan comparison – Organs at Risk (single patient example) ................... 125

6.2.4 Plan comparison – PTVtoSurface (8 patient study set) .......................... 128

6.2.5 Plan comparison – Organs at Risk (8 patient study set) ......................... 130

6.2.6 Plan comparison – Skin Structures (8 patient study set) ........................ 132

6.2.7 Perturbation Effect - PTVtoSurface ......................................................... 134

6.2.8 Perturbation Effect – Organs at Risk ....................................................... 139

6.2.9 Perturbation Effect – Skin Structures ..................................................... 142

6.3 Discussion ....................................................................................................... 145

6.4 Summary ......................................................................................................... 146

Page 5: Development of a technique using VMAT and robust

5

7 Summary ............................................................................................................... 148

7.1 Overview of Results ........................................................................................ 148

7.2 Limitations ...................................................................................................... 150

7.3 Further work ................................................................................................... 151

8 References ............................................................................................................. 153

Appendix 1 .................................................................................................................... 164

Appendix 2 .................................................................................................................... 167

Appendix 3 .................................................................................................................... 170

Appendix 4 .................................................................................................................... 173

Appendix 5 .................................................................................................................... 178

Word Count: 24,900

Page 6: Development of a technique using VMAT and robust

6

Figures

Figure 1.1: Cross section of the mammary gland. (Source: Cancer Research UK) ...................... 22

Figure 1.2: Diagram showing a) lobular carcinoma in situ and invasive carcinoma and b) ductal

carcinoma in situ and invasive carcinoma (Source: Cancer Research UK) .................................. 23

Figure 1.3: a) Virtual 3D representation of linear accelerator (Prosoma v4.2, MedCom,

Germany) b) MLCs defining beam shape c) Intersecting radiation beams over region of

treatment .................................................................................................................................... 28

Figure 1.4: Radiotherapy target volumes as defined in ICRU 50, ICRU 62 and ICRU 83 reports. 29

Figure 1.5: Field-based approach for PTV creation in breast treatments a) field edges defined to

cover in sup/inf and ant/post directions b) treated volume (shaded pink structure) defined by

intersection of tangential beams with breast tissue c) PTV (shaded purple structure) – treated

volume clipped from beam edges, surface and lung. ................................................................. 31

Figure 1.6: a) Beam arrangement and dose distribution for radiotherapy breast treatment b)

Beam segments for medial beam ............................................................................................... 33

Figure 1.7: Measured Depth Dose curve for a 6MV photon beam, 10x10cm field size. ............. 35

Figure 2.1: a) water equivalent bolus slab 40x40cm b) 3D rendered image from CT scan

showing bolus placement in treatment position......................................................................... 49

Figure 2.2: a) 3D rendered image from CT scan with wires defining area for treatment b)

Transverse CT image with opposing tangential beams applied ................................................. 50

Figure 2.3: Blue contour on transverse CT slice indicates position of computer- generated bolus

.................................................................................................................................................... 51

Figure 2.4: PTVtoSurface (red shaded structure) ........................................................................ 54

Figure 2.5: Skin5mm (shaded green structure) with PTVtoSurface (red contour) ...................... 56

Figure 2.6: Example 38Gy dose distribution for a) No Bolus plan b) Clinical Plan ...................... 57

Figure 2.7: Example of DVH for structure PTVtoSurface. (Dashed line = Clinical Plan, Dotted line

= No Bolus Plan) .......................................................................................................................... 58

Figure 2.8: Example DVHs for a) Skin1mm b) Skin3mm c) Skin5mm (Dashed line = Clinical Plan,

Dotted line = No Bolus Plan) ....................................................................................................... 59

Figure 2.9: Example DVHs for Heart (red line) and Ipsilateral Lung (Orange line) (Dashed line =

Clinical Plan, Dotted line = No Bolus Plan) .................................................................................. 61

Figure 2.10: Box and Whisker plots showing the a)V95% b) average dose c) V105% d) V107%

and e) D1% parameters for the PTVtoSurface structure for No Bolus Plans (Blue) and Clinical

Plans (Red) in 8 patients. ............................................................................................................ 63

Page 7: Development of a technique using VMAT and robust

7

Figure 2.11: Box and Whisker plots showing the doses for a) Skin1mm – D99% b) Skin3mm –

D99% c)skin5mm D99% d)Skin1mm – average e)Skin3mm – average f) Skin5mm – average g)

Skin1mm – D1% h) Skin3mm – D1% i) Skin5mm – D1% for No Bolus Plans (Blue) and Clinical

Plans (Red) for the 8 patients. ..................................................................................................... 65

Figure 2.12: Dosimetric parameters for organs at risk a) Box and Whisker plot for Ipsilateral

Lung V30% b) Bar Chart for Heart V5% c) Bar Chart for Heart V25% for No Bolus Plans (Blue)

and Clinical Plans (Red) for the 8 patients. ................................................................................. 66

Figure 3.1: Example Dose Distribution for a) No Bolus plan b) Clinical plan c) VMAT plan ........ 71

Figure 3.2: Example of DVH for structure PTVtoSurface. (Dashed line= Clinical Plan, Dotted line

= No Bolus Plan and Solid line = VMAT plan) .............................................................................. 72

Figure 3.3: Example DVHs for a) Skin1mm b)Skin3mm c)Skin5mm (Dashed line=Clinical plan,

Dotted line = No Bolus plan, Solid line=VMAT plan) ................................................................... 74

Figure 3.4: Example DVHS for Heart (red line), Ipsilateral Lung (Orange line) and Contralateral

Breast (Blue line). (Dashed line = Clinical plan, Dotted line = No Bolus Plan and Solid line=VMAT

plan) ............................................................................................................................................ 75

Figure 3.5: Box and Whisker plots showing the a) V95% b) average dose c) V105% d)107% and

e) D1% parameters for the PTVtoSurface structures for No Bolus Plans (Blue), Clinical Plans

(Red) and VMAT Plans (Green) in the 8 patients. ....................................................................... 77

Figure 3.6: Box and Whisker plots showing the doses for a) Skin1mm – D99% b) Skin3mm –

D99% c)skin5mm D99% d)Skin1mm – average e)Skin3mm – average f) Skin5mm – average g)

Skin1mm – D1% h) Skin3mm – D1% i) Skin5mm – D1% for No Bolus plans (Blue), Clinical plans

(Red) and VMAT plans (Green) for the 8 patients. ...................................................................... 79

Figure 3.7: Dosimetric parameters for organs at risk a) Box and Whisker plot for Ipsilateral Lung

V30% b) Bar Chart for Heart V5% (mandatory constraint shown in dashed line) c) Bar Chart for

Heart V25% d) Box and Whisker plot for Contralateral Breast for No Bolus Plans (Blue), Clinical

Plans (Red) and VMAT Plans (Green) for the 8 patients. ............................................................ 82

Figure 4.1: Dose distribution for a) No Bolus b) Bolus and c) VMAT partial arc plan on the CIRS

anthropomorphic thorax phantom ............................................................................................. 87

Figure 4.2: Points showing the TLD position on central axis of CT Scan of CIRS phantom. Each

TLD point is positioned to intersect with the body contour (green), ........................................... 88

Figure 4.3: a) anthropomorphic phantom position for treatment delivery b) packets position

with TLD either side of central axis ............................................................................................. 89

Figure 4.4: 1cm water equivalent material placed over phantom prior to delivery of bolus plan.

.................................................................................................................................................... 89

Figure 4.5: TLD results for No Bolus plan compared to treatment plan dose - 1 fraction delivery.

(Plan delivered 4 times) ............................................................................................................... 91

Page 8: Development of a technique using VMAT and robust

8

Figure 4.6: TLD results for Bolus plan compared to treatment plan dose - 1 fraction delivery

(Plan delivered 3 times) ............................................................................................................... 92

Figure 4.7: TLD results for VMAT plan compared to treatment plan dose - 1 fraction delivery.

(Plan delivered once) ................................................................................................................... 93

Figure 5.1: Example of perturbed plans a) Clinical plan -non perturbed b) Clinical plan –

perturbed 0.5cm TS direction c) Clinical plan – perturbed 0.5cm AS direction d) VMAT plan –

non perturbed e) VMAT plan – perturbed 0.5cm TS direction f) VMAT plan – perturbed 0.5cm AS

direction ...................................................................................................................................... 99

Figure 5.2: Example of DVHs for PTVtoSurface a)Clinical plan non-perturbed v Clinical plan

perturbed 0.5 in TS direction b) Clinical plan non-perturbed v Clinical plan perturbed 0.5cm in

AS direction c)VMAT plan non-perturbed v VMAT plan perturbed 0.5cm in TS direction d) VMAT

plan non-perturbed v VMAT plan perturbed 0.5cm in AS direction. (Dotted line = original plan,

dashed line = perturbed plan) ................................................................................................... 101

Figure 5.3: Example of DVHs for Skin3mm a)Clinical plan non-perturbed v Clinical plan

perturbed 0.5 in TS direction b) Clinical plan non-perturbed v Clinical plan perturbed 0.5cm in

AS direction c)VMAT plan non-perturbed v VMAT plan perturbed 0.5cm in TS direction d) VMAT

plan non-perturbed v VMAT plan perturbed 0.5cm in AS direction. (Dotted line = original plan,

dashed line = perturbed plan) ................................................................................................... 104

Figure 5.4 : Bar chart showing the perturbation effect on the parameters D99%, average dose

and D1%, for the Skin3mm. The graph shows the impact for both Clinical and VMAT plans.

(Non-perturbed = blue, perturbation 0.5cm TS = red and perturbation 0.5cm AS= green). ..... 105

Figure 5.5: DVH for Heart (Red) Ipsilateral Lung (Orange) and Contralateral Breast (Blue)

a)Clinical plan non-perturbed v Clinical plan perturbed 0.5 in TI direction b) Clinical plan non-

perturbed v Clinical plan perturbed 0.5cm in AI direction c)VMAT plan non-perturbed v VMAT

plan perturbed 0.5cm in TI direction d) VMAT plan non-perturbed v VMAT plan perturbed

0.5cm in AI direction. (Dotted line = original plan, dashed line = perturbed plan) ................... 106

Figure 5.6: Bar chart showing perturbation effect on the parameters V95% (blue), V105% (red)

and V107% (green), displayed as volume difference from non-perturbed plan, for PTVtoSurface,

averaged for 8 patient cases. The graph shows the impact for both Clinical and VMAT plans.109

Figure 5.7: Bar chart showing perturbation effect on the parameters average dose (blue) and

D1% (green), displayed as dose difference from non-perturbed plan, for PTVtoSurface,

averaged for 8 patient cases. The graph shows the impact for both Clinical and VMAT plans.109

Figure 5.8: Bar chart showing V95% for PTVtoSurface, averaged over 8 patients, under the

indicated perturbation conditions. Clinical Plans (blue) and VMAT plans (red). (Dotted line =

mandatory constraint, dashed line = optimal constraint) ........................................................ 110

Page 9: Development of a technique using VMAT and robust

9

Figure 5.9: Bar chart showing V105 values% for PTVtoSurface, averaged over 8 patients, under

the indicated perturbation conditions. Clinical Plans (blue) and VMAT plans (red). (Dotted line =

mandatory constraint, dashed line = optimal constraint) ........................................................ 110

Figure 5.10: Bar chart showing V107% values for PTVtoSurface, averaged over 8 patients,

under the indicated perturbation conditions. Clinical Plans (blue) and VMAT plans (red). (Dotted

line = mandatory constraint)..................................................................................................... 111

Figure 5.11: Bar chart showing D1% values for PTVtoSurface, averaged over 8 patients, under

the indicated perturbation conditions. Clinical Plans (blue) and VMAT plans (red). (Dotted line =

mandatory constraint) .............................................................................................................. 111

Figure 5.12: Bar chart showing perturbation effect on the parameters D99% (blue), average

dose (orange) and D1% (green) for Skin3mm, displayed as dose difference from non-perturbed

plan, averaged for 8 patient cases. The graph shows the impact for both Clinical and VMAT

plans. ......................................................................................................................................... 113

Figure 5.13: Bar chart showing D1% values for Skin3mm, averaged over 8 patients, under the

indicated perturbation conditions. Clinical Plans (blue) and VMAT plans (red). ...................... 114

Figure 5.14: Bar chart showing perturbation effect on the V25% for the heart, displayed as

volume difference from non-perturbed plan, averaged over left-sided cases (n=5). The graph

shows the impact for both Clinical and VMAT plans. ............................................................... 116

Figure 5.15: Bar chart showing V5% (2Gy) for heart for plans perturbed in the 0.5cm AI

direction. The graph shows the impact for both Clinical and VMAT plans (Dotted

line=mandatory constraint). ..................................................................................................... 116

Figure 5.16: Bar chart showing perturbation effect on V30% for the ipsilateral lung, displayed

as volume difference from non-perturbed plan, averaged over the 8 patient cases. The graph

shows the impact for both Clinical and VMAT plans (Dotted line = mandatory constraint

permitted, based on volume irradiated in non-perturbed plan) ............................................... 117

Figure 5.17: Bar chart showing perturbation effect on contralateral breast, displayed as dose

difference from non-perturbed plan, averaged over the 8 patient cases. The graph shows the

impact for both Clinical and VMAT plans. ................................................................................. 118

Figure 5.18: DRRs displaying treatment field segment for a)non-perturbed Clinical plan b)

Perturbed plan 0.5cm in TS direction showing PTVtoSurface (green contour) no longer fully

covered. ..................................................................................................................................... 119

Figure 5.19: Typical MLC segments for VMAT plan a) Non-perturbed b) Perturbed plan 0.5cm in

TS direction showing segment now outside PTVtoSurface (green contour). ............................ 120

Figure 6.1: Dose distribution a)VMATRO plan b)VMAT plan c) dose difference between

VMATRO and VMAT .................................................................................................................. 126

Page 10: Development of a technique using VMAT and robust

10

Figure 6.2: DVH for PTVtoSurface for example patient. The dashed line represents the Clinical

Plan, the dotted line the VMAT plan and the solid line is the VMATRO plan ........................... 126

Figure 6.3: Box and Whisker plots showing the a) V95% b) average dose c) V105% d)107% and

e) D1% parameters for the PTVtoSurface structures for No Bolus Plans (Blue), Clinical Plans

(Red), VMAT Plans (Green) and VMATRO Plans (Purple) in the 8 patients. .............................. 129

Figure 6.4: Dosimetric parameters for organs at risk a) Box and Whisker plot for Ipsilateral Lung

V30% b) Bar Chart for Heart V5% (mandatory constraint shown in dashed line) c) Bar Chart for

Heart V25% d) Box and Whisker plot for Contralateral Breast for No Bolus Plans (Blue), Clinical

Plans (Red),VMAT Plans (Green) and VMATRO Plans (Purple) for the 8 patients. ................... 131

Figure 6.5: Box and Whisker plot showing the doses for the D99% parameter for the Skin3mm

structure for the 8 patients for the No Bolus plans (Blue), Clinical plans (Red), VMAT plans

(Green) and VMATRO plans (Purple) ........................................................................................ 132

Figure 6.6: Box and Whisker plot showing the doses for the D1% parameter for the Skin3mm

structure for the 8 patients for the No Bolus plans (Blue), Clinical plans (Red), VMAT plans

(Green) and VMATRO plans (Purple) ........................................................................................ 133

Figure 6.7: Box and Whisker plot showing the doses for the average dose parameter for the a)

Skin1mm b) Skin3mm and C) Skin5mm structures for the 8 patients for the No Bolus plans

(Blue), Clinical plans (Red), VMAT plans (Green) and VMATRO plans (Purple). ....................... 134

Figure 6.8: Bar chart showing perturbation effect on the parameters V95% (blue), V105% (red)

and V107% (green), displayed as volume difference from non-perturbed plan, for PTVtoSurface,

averaged for 8 patient cases. The graph shows the impact for Clinical, VMAT and VMATRO

plans.) ........................................................................................................................................ 136

Figure 6.9: Bar chart showing V105 values% for PTVtoSurface, averaged over 8 patients, under

different perturbation conditions. Clinical Plans (blue), VMAT plans (red) and VMATRO (green).

(Dotted line = mandatory constraint, dashed line = optimal constraint) .................................. 137

Figure 6.10: Bar chart showing V107% values for PTVtoSurface, averaged over 8 patients,

under different perturbation conditions. Clinical Plans (blue) and VMAT plans (red (Dotted line =

mandatory constraint) .............................................................................................................. 137

Figure 6.11: Bar chart showing V95% for PTVtoSurface, averaged over 8 patients, under

different perturbation conditions. Clinical Plans (blue), VMAT plans (red) and VMATRO (green).

(Dotted line = mandatory constraint, dashed line = optimal constraint) .................................. 138

Figure 6.12: Bar chart showing perturbation effect on the parameters average dose (blue) and

D1% (green), displayed as dose difference from non-perturbed plan, for PTVtoSurface,

averaged for 8 patient cases. The graph shows the impact for Clinical, VMAT and VMATRO

plans. ......................................................................................................................................... 138

Page 11: Development of a technique using VMAT and robust

11

Figure 6.13: Bar chart showing perturbation effect on V30% for the ipsilateral lung, displayed

as volume difference from non-perturbed plan, averaged over the 8 patient cases. The graph

shows the impact for Clinical, VMAT and VMATRO plans. (Dotted line=mandatory constraint

permitted, based on volume irradiated in non- perturbed plan. For the Clinical plan the average

non-perturbed volume = 11.7%, for the VMAT plan the average non-perturbed volume = 8.4%

and for the VMATRO the average non-perturbed volume = 10.1%. The constraint for V30% <

17%, therefore permitted constraint for perturbation is 5.3%, 8.6% and 6.9% for Clinical, VMAT

and VMATRO plans, respectively). ............................................................................................ 140

Figure 6.14: Bar chart showing V5% (2Gy) for heart for plans perturbed in the 0.5cm AI

direction. The graph shows the impact for Clinical, VMAT and VMATRO plans (Blue bars= non-

perturbed, red bars=perturbed, dotted line=mandatory constraint)........................................ 141

Figure 6.15: Bar chart showing perturbation effect on contralateral breast, displayed as dose

difference from non-perturbed plan, averaged over the 8 patient cases. The graph shows the

impact for Clinical, VMAT and VMATRO plans. ........................................................................ 141

Figure 6.16: Bar chart showing perturbation effect on the parameters D99% (blue), average

dose (orange) and D1% (green) for Skin3mm, displayed as dose difference from non-perturbed

plan, averaged for 8 patient cases. The graph shows the impact for Clinical, VMAT and

VMATRO plans. ......................................................................................................................... 143

Figure 6.17: Bar chart showing D1% values for Skin3mm, averaged over 8 patients, under

different perturbation conditions. Clinical Plans (blue), VMAT plans (red) and VMATRO plans

(green). ...................................................................................................................................... 143

Figure 6.18: Bar chart showing average values for Skin3mm, for the 8 patients, under different

perturbation conditions. Clinical Plans (blue), VMAT plans (red) and VMATRO plans (green) . 144

Figure 6.19: Bar chart showing D99% values for Skin3mm, averaged over 8 patients, under

different perturbation conditions. Clinical Plans (blue), VMAT plans (red) and VMATRO plans

(green) ....................................................................................................................................... 144

Page 12: Development of a technique using VMAT and robust

12

Tables

Table 2.1: Departmental objectives and constraints used for breast planning based on those

used in FAST-Forward Trial (ISRCTN19906132) .......................................................................... 50

Table 2.2: Summary of patient information including treatment site (CW=chest wall), machine

and beam energy ........................................................................................................................ 53

Table 2.3: Summary of dosimetric parameters recorded for each patient ................................. 56

Table 2.4: Example of dosimetric parameters obtained for the structure PTVtoSurface for the

Clinical and No Bolus Plans for one patient. ............................................................................... 58

Table 2.5: Example of dosimetric parameters obtained for the Skin Structures for the Clinical

and No Bolus Plans. (Values is brackets represent the dose received as a percentage of the

prescription dose, 40Gy). ............................................................................................................ 60

Table 2.6: Example of dosimetric parameters obtained for the heart and ipsilateral lung OARS,

for the Clinical and No Bolus Plans. ............................................................................................ 61

Table 3.1: Typical starting values for planning objectives used for the VMAT plans. ................. 70

Table 3.2: Example of dosimetric parameters obtained for the structure PTVtoSurface for the

Clinical, No Bolus and VMAT plans ............................................................................................. 72

Table 3.3: Example of dosimetric parameters obtained for the skin structures for the Clinical, No

Bolus and VMAT plans (Values in brackets represent the dose received as a percentage of the

prescription dose, 40Gy) ............................................................................................................. 73

Table 3.4: Example of dosimetric parameters obtained for the heart, lung and contralateral

breast, for the Clinical, No Bolus and VMAT plans...................................................................... 76

Table 4.1: Results of TLD measurements for No Bolus plan as percentage of prescribed dose

compared to planned dose, absolute dose in brackets. .............................................................. 91

Table 4.2: Results of TLD measurement for Bolus plan as percentage of prescribed dose

compared to planned dose absolute dose in brackets. (Measurement 2 at TLD position 3

disregarded in these results) ....................................................................................................... 92

Table 4.3: Results of TLD measurements for VMAT plan as percentage of prescribed dose

compared to planned dose absolute dose in brackets. ............................................................... 93

Table 5.1: Summary of shifts for perturbed plans. Abbreviation for perturbation direction are

included. ...................................................................................................................................... 97

Table 5.2: Example of dosimetric parameters for the non-perturbed Clinical plan and with a

perturbation of 0.5cm in the TS and AS directions. Values underlined show where constraints

were not met. ............................................................................................................................ 102

Page 13: Development of a technique using VMAT and robust

13

Table 5.3: Example of dosimetric parameters for the non-perturbed VMAT plan and with a

perturbation of 0.5cm in the TS and AS directions. Values underlined show where constraints

were not met ............................................................................................................................. 102

Table 5.4: Example of dosimetric parameters for the non-perturbed Clinical plan and with a

perturbation of 0.5cm in the TI and AI directions. Values underlined show where constraints

were not met. ............................................................................................................................ 107

Table 5.5: Example of dosimetric parameters for the non-perturbed VMAT plan and with a

perturbation of 0.5cm in the TI and AI directions. Values underlined show where constraints

were not met. ............................................................................................................................ 107

Table 6.1: Dosimetric parameters achieved for PTVtoSurface single patient example. The

mandatory and optimal constraints are defined. ..................................................................... 127

Table 6.2: Dosimetric parameters achieved for organs at risk single patient example. The

mandatory and optimal constraints are defined. ..................................................................... 127

Table A1.1: Comparison of dosimetric parameters for planning structures, with VMAT plan re-

calculated using dose grids 0.2cm and 0.3cm......................................................................... 1643

Table A1.2: Comparison of dosimetric parameters for planning structures, with VMAT plan re-

calculated using dose grids 0.2cm and 0.3cm and perturbed in the 0.5cm AS direction ........ 1654

Table A1.3: Comparison of dosimetric parameters for planning structures, with VMAT plan re-

optimised using dose grids 0.2cm and 0.3cm ......................................................................... 1654

Table A1.4: Comparison of dosimetric parameters for planning structures, with VMAT plan re-

optimised using dose grids 0.2cm and 0.3cm and perturbed 0.5cm in the AS direction ........ 1665

Table A2.1 -Comparison of dose parameters for structures in VMATRO plans optimised in all

directions and limited directions, including the perturbation in AS direction…………………………168

Table A3.1: V105% parameter for PTVtoSurface for every patient. VMAT RO plans perturbed in

the AS direction to different extents……………………………………………………………………………………..170

Table A3.2: V107% parameter for PTVtoSurface for every patient. VMAT RO plans perturbed in

the AS direction to different extents……………………………………………………………………………………..170

Table A4.1: Additional information on the range of perturbation effect on the planning

parameters V95%, V105%, V107%, average dose and D1%, for the PTVtoSurface structure, for

the 8 patient cases. Data is included for the Clinical, VMAT and VMATRO plans, the data shows

the average volume difference or dose difference from the non-perturbed plan and the

min/max difference over the 8 patients. The averaged data for the planning parameters shown

in Figures 5.6, 5.7, 6.8 and 6.12. …………………………………………………………………………………………..172

Table A4.2: Additional information on the range of perturbation effect on the planning

parameters V95%, V105%, V107%, average dose and D1%, for the PTVtoSurface structure, for

the 8 patient cases. Data is included for the Clinical, VMAT and VMATRO plans, the data shows

Page 14: Development of a technique using VMAT and robust

14

the average volume or dose and the min/max values across the 8 patients. The averaged data

for the planning parameters shown in Figures 5.8-5.11 and 6.9-6.11……………………………………173

Table A4.3: Additional information on the range of perturbation effect on the planning

parameters D99%, Average and D1% for the Skin3mm structure, over the 8 patient cases. Data

is included for the Clinical, VMAT and VMATRO plans, the data shows the dose difference from

the non-perturbed plan and the min/max difference over the 8 patients. The averaged data for

the planning parameters shown in Figures 5.12, 6.16…………………………………………….…….………174

Table A4.4: Additional information on the range of perturbation effect on the planning

parameters D99%, D1% and Average dose for the Skin3mm structure, for the 8 patient cases.

Data is included for the Clinical, VMAT and VMATRO plans, the data shows the average doses

and the min/max difference over the 8 patients. The averaged data for the planning

parameters shown in Figures 5.13, 6.17-6.19……………………………………………………………………….175

Table A4.5: Additional information on the range of perturbation effect on the planning

parameters for heart and lungs structures, heart. Data is included for the Clinical, VMAT and

VMATRO plans, the data shows the average doses and the min/max difference over the 8

patients for the lung parameter and the 5 left chest wall patients for the heart parameters. The

averaged data for the planning parameters shown in Figures 5.14-5.16 and 6.13-6.14……..…176

Page 15: Development of a technique using VMAT and robust

15

Acronyms 3D-CRT Three-dimensional conformal radiotherapy

ASCO American Society of Clinical Oncology

CBCT Cone beam computed tomography

CT Computer tomography

CTV Clinical target volume

DIBH Deep inspiration breath hold

DNA Deoxyribonucleic acid

DVH Dose volume histogram

GTV Gross tumour volume

ICRP International commission on Radiological Protection

ICRU International Commission on Radiological Units &

Measurements

IMRT Intensity-modulated radiotherapy

ITV Internal target volume

MLC Multi-leaf collimator

MOSFET Metal oxide semiconductor field effect transistor

MU Monitor units

NICE National Institute for Health and Care Excellence

NTCP Normal tissue complication probability

OAR Organ at risk

OSLD Optically stimulated luminescence

PMRT Post-mastectomy radiotherapy

PRV Planning organ at risk volume

PTV Planning target volume

SMLC Segmented multileaf collimation

QEHB Queen Elizabeth Hospitals Birmingham

TCP Tumour control probability

TLD Thermoluminescent dosimeter

TNM Tumour, nodes, metastases

VMAT Volumetric modulated arc therapy

WED Water equivalent depth

Page 16: Development of a technique using VMAT and robust

16

Abstract

Radiotherapy following surgery is routine practice for patients that have had a

mastectomy and are at high risk of the cancer recurring. To ensure that the chest-wall

receives an adequate radiation dose a tissue equivalent material, or bolus, 1cm thick is

placed on the patient surface. At the Queen Elizabeth Hospitals Birmingham (QEHB)

this is applied for 7 of the 15 fractions of treatment. This technique however requires

the creation of two treatment plans, increasing planning time in the patient pathway.

In addition, the lack of flexibility of the bolus can cause air gaps between the skin and

material affecting the surface dose and since the bolus is only required for a number of

the fractions, treatment errors can occur if it is omitted by mistake. The aim of the

research was to investigate a single, no bolus planning solution for these patients to

reduce these issues.

For a sample of 8 patients it was shown that bolus increased the target volume

receiving 95% of the prescription dose by 7.7% compared to using no bolus at all. The

use of VMAT could replicate these dose distributions, including superficial doses,

without the need for bolus. Although the VMAT plans did produce a low dose bath

which in some cases increased the doses to organs at risk, the plans still met all the

required dose constraints.

However, fluence loading in the surface region (to overcome the build-up effect)

means that VMAT plans show unacceptable changes in dose distribution with changes

in patient position or contour.

Combining VMAT with robust optimisation significantly reduced dose differences

caused by perturbation. These plans however still resulted in distributions that did not

meet accepted dose constraints within the target structure, potentially causing

undesired side effects for the patients. The use of robust optimisation also

compromised the non-perturbed plans, reducing the dose enhancement effect to the

patient surface.

A single plan solution using the VMAT technique and combined with robust

optimisation produced plans that mimic the clinical plans, without the use of bolus.

Although the robust optimisation significantly reduced the variation in dose due to

perturbation, these plans are still susceptible to patient movement and can result in

tolerance doses being exceeded.

The combination of VMAT and robust optimisation shows promise in producing a

single, no bolus plan solution for chest-wall irradiation, but further work is required to

quantify patient motion if the technique is to be applied clinically.

Page 17: Development of a technique using VMAT and robust

17

Declaration

I declare that no portion of the work referred to in the thesis has been submitted in

support of an application for another degree or qualification of this or any other

university or other institute of learning.

Copyright

i. The author of this thesis (including any appendices and/or schedules to this

thesis) owns certain copyright or related rights to it (the “Copyright”) and

he has given The University of Manchester certain rights to use such

Copyright, including for administrative purposes.

ii. Copies of the thesis, either in full or in extracts and whether in hard or

electronic copy, may be made only in accordance with the Copyright,

Designs and Patents Act 1988 (as amended) and regulations issued under it

or, where appropriate, in accordance with licensing agreements which the

University has from time to time. This page must form part of any such

copies made.

iii. The ownership of certain Copyright, patents, designs, trademarks and other

intellectual property (the “Intellectual Property”) and any reproductions of

copyright works in the thesis, for example graphs and tables

(“Reproductions”), which may be described in the thesis, may not be owned

by the author and may be owned by third parties. Such Intellectual Property

and Reproductions cannot and must not be made available for use without

Page 18: Development of a technique using VMAT and robust

18

the prior written permission of the owner(s) of the relevant Intellectual

Property and/or Reproductions.

iv. Further information on the conditions under which disclosure, publication

and commercialisation of this thesis, the Copyright and any Intellectual

Property and/or Reproductions described in it may take place is available in

the University IP Policy (see

http://documents.manchester.ac.uk/DocuInfo.aspx?DocID=24420 ), in any

relevant Thesis restriction declarations deposited in the University Library,

The University Library’s regulations (see

http://www.library.manchester.ac.uk/about/regulations/ ) and in The

University’s policy on Presentation of Theses.

Page 19: Development of a technique using VMAT and robust

19

Acknowledgements

The completion of this thesis and additional aspects of this professional doctorate as

part of the Higher Specialist Scientist Training, would not have been possible without

the support and assistance I have received from many people.

I would like to thank my thesis supervisors Jason Cashmore and Helen Mayles for their

guidance and encouragement with this project. Their experience and advice has been

invaluable.

I am very grateful to my colleagues in the radiotherapy team at the Queen Elizabeth

Hospitals Birmingham for enabling me to take part in the HSST program and for their

continued support and reassurance throughout the process.

I would also like to thank the staff involved with the DClinSci for the organisation and

delivery of the course. Being one of the first cohort to go through the scheme has been

particularly challenging at times and I am grateful to have met and overcome these

issues with a wonderful group of people. In particular I would like to thank Peter

McGookin and Pedrum Kamali for their friendship and support especially in the last

few months.

To my friends and family, thank you for your endless encouragement and finally to my

partner, Jonathan Trinder, thank you for always being there.

Page 20: Development of a technique using VMAT and robust

20

Statement for Examiners

This research project forms part of the Doctor of Clinical Science (DClinSci). In addition

to the research project the DClinSci includes taught components and an innovation

project. Specialist scientific units in Medical Physics, have been delivered by the

University of Liverpool and the University of Manchester, and the teaching and

assessment for the Post Graduate Diploma in Healthcare Science Leadership has been

provided by Alliance Manchester Business School. The generic scientific units and the

innovation project were undertaken through the University of Manchester. A summary

of units and associated assessments is provided in Appendix 5. The Alliance

Manchester Business School (A units) account for 120 credits. The specialist and

generic scientific (B units) account for 150 credits, the innovation project (C1), 70

credits, and the Research Project accounts for 200 credits.

Page 21: Development of a technique using VMAT and robust

21

Chapter 1

1 Introduction

This chapter provides a brief overview of cancer biology, particularly focussing on

breast cancer, and the role of external beam radiotherapy in the treatment of patients

who have had undergone a mastectomy. The current techniques used in radiotherapy

treatment of breast cancer will be discussed and the rationale for the introduction of a

new approach to the technique will be presented.

1.1 Cancer

Cancer is a term used to describe a disease where a group of cells within the body

divide uncontrollably. Cell division, replication and multiplication are fundamental

processes within the human body to enable growth, maintenance and repair.

However, mutations to the DNA can cause disruptions to the normal cell cycle, making

cells divide more quickly or avoid cell death, resulting in abnormal growth of tissue, a

tumour. Tumours can be defined as malignant or benign. Those that divide too much,

but that do not have the potential to invade other tissues or spread around the body,

are described as benign. Tumours that can invade other tissues and spread to other

organs within the body (metastasize) are known as malignant.

Cancer Research UK report that one in two people born in the UK after 1960 will be

diagnosed with cancer at some time during their lifetime (excluding non-melanoma

skin cancer) with 367, 000 new cases reported each year in the period 2015-2017

(Cancer Research UK, 2021c). In the UK the most prevalent cancers, accounting for

more than half of all new cases, male and female combined, are breast (15%), prostate

(13%), lung (13%) and bowel (11%). In females, breast cancer accounts for 30% of all

new cases, with 1 in 7 women in the UK predicted to develop it at some point their

lifetime. The incidence rate of breast cancer in the UK has increased by 4% in the last

decade with mortality rates decreasing by 21%. The doubling of breast cancer survival

Page 22: Development of a technique using VMAT and robust

22

in the past 40 years is thought to be a result of a combination of improvements in

treatment and care, and in early detection through screening and faster diagnosis.

1.2 Breast Anatomy

The breast is defined as the tissue that overlies the pectoral muscles in the chest and

consists of glandular tissues called lobules which produce milk during lactation. The

lobules consist of a series of lobes surrounded by fat cells and connective tissue and

are supported with ligaments. Blood vessels circulate blood around the breast tissue

and nerves running through the tissue are responsible for sensation. The breast tissue

also contains lymphatic vessels which connect to lymph nodes in the axilla and behind

the sternum which maintain the fluid balance within the breast tissue and remove

unwanted toxins from the tissue. Figure 1.1 shows the key structures within the breast

(Cancer Research UK, 2021g).

Figure 1.1: Cross section of the mammary gland. (Source: Cancer Research UK)

1.3 Breast Cancer

Breast cancer most commonly occurs in the epithelial cells that line the ducts of the

breast and the breast lobules. If the tumour has started to invade other tissue within

the breast and grown out of the lining of the duct or lobule, the carcinoma is described

Page 23: Development of a technique using VMAT and robust

23

as invasive, Figure 1.2. When the cells remain contained within the basal membrane

they are referred to as, in situ. In the case of lobular carcinoma in situ, this is not

considered cancerous, though may be indicative of a higher risk of developing cancer

in the future, so treatment is usually limited to regular monitoring. Ductal carcinoma in

situ, may eventually spread to surrounding tissues if not treated and is usually graded

to establish the course of treatment required. In the UK around 70 % of invasive breast

cancers arise from the ducts, whilst around 15% start in the breast lobules (Cancer

Research UK, 2021e, 2021d). Other, rarer, types of breast cancer include inflammatory

breast cancer, which effects the lymph ducts and angiosarcomas, which effect the soft

tissue in the breast. Breast cancer can also spread via the lymphatic system to nodes in

the armpit (axillary), near the breastbone (internal mammary) or to nodes above the

collar bone (supraclavicular).

a)

b)

Figure 1.2: Diagram showing a) lobular carcinoma in situ and invasive carcinoma and b) ductal carcinoma in situ and invasive carcinoma (Source: Cancer Research UK)

1.3.1 Breast Cancer Diagnosis

The most common route for the diagnosis of breast cancer is via the ‘two-week wait’

pathway when patients who meet a particular criteria for age and symptoms, are

referred to specialists by their General Practitioner. 51% of female invasive breast

Page 24: Development of a technique using VMAT and robust

24

cancers are diagnosed via the ‘two-week wait’ referral route and 31% are detected via

mammography screening, which detects the highest proportion of cases diagnosed at

an early stage (Cancer Research UK, 2021a)

Tests for diagnosing breast cancer are often performed as part of a triple assessment,

consisting of physical examination, imaging and a biopsy. The imaging normally

consists of an ultrasound, which can distinguish whether a lump is solid or fluid. Fluid-

filled lumps may be drained, and solid lumps may be biopsied, followed by further

imaging using mammography.

To establish the prognosis of a patient with breast cancer and recommend the most

suitable treatment plan, information about the stage and grade of the tumour is

required along with the tumour type or its response to hormones.

Tumour staging takes into account the physical properties of the tumour, its size and

whether it has spread. The TNM staging system gives the complete stage of the cancer.

T describes the tumour size, N the nodal involvement and M whether the cancer has

metastasised. Similarly, breast cancer can be divided into 4 stages, with further sub-

categories for Stage 1 and 2 (A and B) and 3 for Stage 3 (A, B and C). Each category and

sub-category have a solid tumour size, and indication of nodal involvement associated

with it. Macmillan Cancer Support describe the stages as follows (Macmillan Cancer

Support, 2021). Stage 1A breast cancer is when the cancer is <2cm and has not spread

outside the breast. Stage 1B indicates small areas of breast cancer cells are found in

the lymph nodes and that no tumour is found in the breast or it is <2cm. Stage 2A

means that there is no tumour in the breast or it is <2cm, and there is 1-3 lymph nodes

involved. Stage 2B can mean that the tumour 2-5cm in size, with some cancer cells

identified in the nodes or 1-3 lymph nodes are involved. Stage 2B can also mean that

the tumour >5cm but not involving any lymph nodes. Stage 3 is referred to as locally

advanced and means that the cancer has spread to lymph nodes or to the skin of the

breast or chest wall. Stage 3A includes cases where the tumour is any size in the breast

tissues (including no tumour) but 4-9 lymph nodes are involved or is >5cm with small

clusters of cancer cells in the lymph nodes or up to 3 lymph nodes in the armpit or

near the breast bone. Stage 3B indicates the tumour has spread to the skin or into the

chest wall and Stage 3C indicates there is cancer in the skin and the breast cancer cells

have spread to 10 or more lymph nodes in the arm pits, lymph nodes above or below

Page 25: Development of a technique using VMAT and robust

25

the collar bone, or lymph nodes in the armpit and near the breastbone. The stage 3C

cancers may be considered operable or inoperable. Stage 4 breast cancer is used to

described secondary or metastatic breast cancer, when the tumour has spread to

other parts of the body. Tumour grading describes the abnormality of the cells

examined microscopically and is indicative of how aggressive the cancer is likely to be.

For invasive breast cancer there are three grades, Grade 1 to Grade 3. Grade 1 (low

grade) cancer cells look most like normal breast cells, well differentiated, usually slow-

growing and less likely to spread. Grade 2 (intermediate grade) the cells look more

abnormal and slightly faster growing that Grade 1. Grade 3 (high grade) look very

difference from normal breast tissue cells and are more likely to grow quickly and

spread to other sites.

Tumour biomarkers are also used in breast cancer to help define the prognosis and

stratify patient treatment. Tumours that are oestrogen or progesterone receptor

positive will respond well to hormone therapies that will stop the hormone stimulating

cancer cells to grow and divide. Similarly, breast cancers with high levels of the protein

HER2 are likely to respond to drugs that will attach to the protein and supress the cell

growth and division.

Survival for breast cancer is related to the stage of disease at diagnosis. Data published

by the Office for National Statistics is summarised by Cancer Research UK (2021e)

indicating that for female patients diagnosed with Stage 1 breast cancer, survival at 1

year was 100% reducing to 97.9% at 5 years. For Stage 2 breast cancer at diagnosis,

survival went from 98.9% at year one to 89.6% at year 5, and Stage 3 and 4, were

95.5% to 72% and 66% to 26.2%, respectively, at the same time points.

1.3.2 Breast Cancer Treatment

Treatment for breast cancer is primarily surgery, with 81.2% of patients diagnosed

with breast cancer in England (2013-2014) having it as part of their primary treatment

(Cancer Research UK, 2021b).The proportion of patients having surgery is dependent

on the stage at diagnosis with 92.8% of Stage 1 patients having surgery compared with

25.1% of Stage 4 patients. This may either be a breast-conserving procedure, surgery

removing part of the breast containing the cancer or mastectomy, removal of all the

Page 26: Development of a technique using VMAT and robust

26

breast tissue. Both surgeries may also involve the removal of lymph nodes, dependent

on involvement. Additional treatment such as radiotherapy, chemotherapy, endocrine

or biological therapies, will be dependent on several factors. These include the staging,

grading and presence of tumour biomarkers, the risk of local recurrence, which may be

age related, and any patient comorbidities.

For patients with invasive breast cancer the National Institute for Clinical Excellence

guidelines NG101 (NICE, 2018) recommend offering whole-breast radiotherapy

following breast-conserving surgery with clear margins, unless the patient is at very

low risk of local recurrence and are willing to take adjuvant endocrine therapy for at

least 5 years. For patients having undergone a mastectomy it is recommended that

adjuvant radiotherapy is offered to those with node-positive (macrometastases)

invasive breast cancer or those with involved margins. Radiotherapy should also be

considered for people with node-negative T3 or T4 breast cancer but not be offered to

those with invasive breast cancer at a low risk of local recurrence, where risks

associated with radiotherapy outweigh the benefits.

In 2013-2014 63.2% of all patients diagnosed with breast cancer received radiotherapy

as part of their treatment. The proportion of patients receiving radiotherapy was

dependent on the stage of cancer at diagnosis. For patients with stage 1, 2 and 3 at

diagnosis the percentage receiving radiotherapy was 70.2%, 65.1% and 80.4%

respectively. For those with Stage 4 cancer at diagnosis 39.2% received radiotherapy

(Cancer Research UK, 2021b).

1.4 Radiotherapy Treatment

Radiotherapy is the use of high energy radiation to treat cancer cells. It works by

delivering ionising radiation which damages the DNA (deoxyribonucleic acid) of the

tumour cells. As radiation can also damage normal tissue it is important that the

treatment is designed so that the tumour volume receives enough damaging radiation

to kill the tumour cells, high ‘tumour control probability’ (TCP) whilst minimising the

dose to the healthy tissues, low ‘normal tissue complication probability’ (NTCP). As

normal tissue cells are usually better at repairing from the damage from radiation than

the tumour cells, the radiation is delivered in daily fractions, allowing the healthy

Page 27: Development of a technique using VMAT and robust

27

tissue to undergo some repair, before another treatment is delivered. For breast

cancer treatments the radiotherapy is delivered using a linear accelerator with a

technique known as external beam radiotherapy. The radiation beam enters the

patient from an external source, in the head of the linear accelerator, and deposits

energy that damages the cells, when it intersects with the patient. The gantry design of

the linear accelerator allows radiation beams to enter the patient from a range of

different coplanar directions, Figure 1.3a. Overlapping the radiation beam at the site of

the tumour intensifies the dose in this area whilst minimising dose to the surrounding

area, Figure 1.3c. The beam can also be modified to conform to the shape of the

tumour. This is done in the head of the linear accelerator with the use of multi-leaf

collimators, individual tungsten leaves which can be adjusted to create the required

beam shape, Figure 1.3b.

1.5 Radiotherapy Treatment Planning

When a patient has been referred for breast radiotherapy a computer tomography

(CT) scan is performed. The scan provides detailed cross-sectional anatomical

information that is used to define the area to be treated, determine the optimal

direction that radiation beams need to enter the patient and to calculate the dose the

linear accelerator needs to deliver. To ensure that the patient is in the same position

for every fraction of treatment an immobilisation device is used. Patients receiving

radiotherapy for breast cancer are positioned on a breast board which keeps the arms

above the head. To help with daily set up the patient may also be tattooed, providing

an external reference to aid in locating the treatment area. The CT scans are

transferred to computer software where virtual simulation packages reconstruct the

data into a 3D model of the patient. The system is then used by the clinician to define

the area that needs to be treated and any organs at risk or areas of normal tissue

where dose should be kept to a minimum. Treatment planning is performed using

specialised software that is able to accurately calculate the dose deposited in the

patient. The software uses complex algorithms to model how the radiation will interact

with the patient based on electron density information acquired from the CT scan.

Page 28: Development of a technique using VMAT and robust

28

a)

b)

c)

Figure 1.3: a) Virtual 3D representation of linear accelerator (Prosoma v4.2, MedCom, Germany) b) MLCs defining beam shape c) Intersecting radiation beams over region of treatment

Page 29: Development of a technique using VMAT and robust

29

1.6 Radiotherapy Volume Definitions

To ensure that the tumour is treated adequately the international commissioning on

radiation units and measurements (ICRU) published recommendations, ICRU report 50

(International Commission on Radiation Units and Measurements, 1993), ICRU report

62 (ICRU, 1999) and ICRU report 83 (ICRU, 2010) on the use of margins to account for a

variety of uncertainties. The GTV is the Gross Tumour Volume and defined as the

extent of the malignant tumour growth. The CTV (Clinical Target Volume) is the GTV

plus a margin to encompass sub-clinical microscopic malignant disease. An ITV

(Internal Target Volume) may be created around the CTV to account for internal

movement and the PTV (Planning Target Volume) is a margin added that accounts for

geometrical variations in the CTV and inaccuracies in treatment delivery. The treated

volume is then defined as the volume receiving the prescribed dose and the irradiated

volume, the volume receiving a dose that is considered significant in relation to normal

tissue tolerance. Figure 1.4 summarises these target volumes.

Figure 1.4: Radiotherapy target volumes as defined in ICRU 50, ICRU 62 and ICRU 83 reports.

Page 30: Development of a technique using VMAT and robust

30

Other planning volumes that are defined in the reports include the Organs At Risk

(OAR), healthy tissue whose radiation sensitivity may influence treatment planning

decisions and the planning organ at risk volume (PRV). As with the PTV, the PRV

includes a margin around the OAR to account for geometric variation and inaccuracies

in delivery.

The PTV is the structure to which a treatment plan is optimised to, with the treatment

planner aiming to deliver the prescribed dose to this structure within -5% and +7%

(ICRU, 1993; ICRU, 1999), and keep the dose to the surrounding tissue as low as

possible.

1.7 Breast Planning Volumes

For breast radiotherapy treatments the approach for defining planning target volumes

can be different to that described in the previous section. For breast treatment

planning, a field- based approach is often used, although not a true PTV it is useful for

the purpose of reporting. In this situation the extent of the PTV is determined by edges

of a tangential pair of beams that have been selected to cover the treatment area. The

superior/inferior field lengths cover the breast tissue, the posterior edge, to cover the

tissue but not exposing more than 1.5cm lung, and the anterior edge to cover the

extent of the soft tissue plus a flash margin, typically 1-2cm, to ensure coverage due to

inaccuracies in treatment delivery or patient movement. A ‘treated volume’ is then

created which is the intersection of the fields with the patient tissue. The PTV is then

created by clipping 5mm from the patient surface, lung and posterior beam edge, and

10mm from the superior and inferior beam edges. Figure 1.5 outlines the field-based

approach for PTV in the case of breast treatments.

Alternatively, the CTV can be manually outlined to include soft tissues of the whole

breast from 5mm below the skin surface down to the deep fascia, excluding muscle

and underlying rib cage. A PTV is then created from the CTV by adding an appropriate

margin to take account of set-up error, breast swelling and breathing.

Page 31: Development of a technique using VMAT and robust

31

a)

b)

c)

Figure 1.5: Field-based approach for PTV creation in breast treatments a) field edges defined to cover in sup/inf and ant/post directions b) treated volume (shaded pink structure) defined by intersection of tangential beams with breast tissue c) PTV (shaded purple structure) – treated volume clipped from beam edges, surface and lung.

1.8 Radiotherapy Planning Techniques

The aim of radiotherapy is to deliver adequate dose to the tumour whilst minimising

dose to normal tissues. As mentioned previously, ICRU reports 50 and 62 (ICRU, 1993;

ICRU, 1999) recommend that the homogeneity over the PTV is between -5% to +7% of

the prescribed dose. Three key techniques in external beam radiotherapy that can be

used to achieve these objectives are three-dimensional conformal radiotherapy (3D-

CRT), intensity modulated radiotherapy (IMRT) and volumetric arc therapy (VMAT).

Page 32: Development of a technique using VMAT and robust

32

3D-CRT is a term used to describe a planning technique that uses coplanar static

beams which conform to the shape of the PTV in all 3 dimensions. To achieve the

conformity, multi-leaf collimators (MLCs) are used to shape the radiation beams which

can be delivered from multiple directions. To ensure the distribution is homogeneous

wedges, high-density material in the head of the linear accelerator, can be used to

attenuate parts of the beam. This homogeneity can also be achieved with the use of

additional beam segments to modulate the radiation, also referred to as field -in-field

treatment, this can replicate the effect of the wedge. 3D-CRT typically requires input

from the treatment planner in decisions regarding beam direction, beam weighting

and segment shaping. The field -in field method is a step towards IMRT, a technique

which uses multiple beam segments with the intensity of the beam modulated to

produce a homogeneous dose across the tumour. With the IMRT technique an inverse

approach to treatment planning is more often used. In this situation the number of

beams and beam direction may be selected by the treatment planner along with

objectives for PTV coverage and OAR sparing. The treatment planning software then

optimises the beam segments, MLC positions and segment weightings to optimise the

dose distribution to meet the planning objectives. VMAT is a type of IMRT treatment

that is delivered whilst dynamically rotating around the patient. The dose rate is

changing, and the MLCs are constantly moving to modulate the beam in order to

achieve a conformal and uniform distribution.

1.8.1 Radiotherapy Planning Techniques for Breast Treatments

For patients with breast cancer the widely used approach to treatment planning is the

use of opposing tangential fields. The fields are typically set up with a non-divergent

posterior edge to minimise lung dose. Megavoltage photon beams are used to achieve

the required depth penetration and the homogeneity of dose distribution is controlled

with wedges or a field-in-field technique. In the field-in-field technique tangents

typically consist of an open field that cover the treatment area and 4-5 segments that

control the homogeneity. A 40Gy in 15# dose schedule is the UK standard

recommended by NICE in their guidelines for the diagnosis and management of early

and locally advanced breast cancer (NICE, 2018). Figure 1.6 shows a typical beam

Page 33: Development of a technique using VMAT and robust

33

arrangement and dose distribution for a breast treatment, with the beam segments for

the medial beam also displayed.

a)

b)

Figure 1.6: a) Beam arrangement and dose distribution for radiotherapy breast treatment b) Beam segments for medial beam

Page 34: Development of a technique using VMAT and robust

34

1.9 Radiotherapy Treatment Delivery

A completed treatment plan is checked and approved by a clinician, before being

transferred electronically to the linear accelerator. When the patient comes for

treatment they will be set-up in the same position as they were imaged for their

planning CT scan and the tattoos are used to move the patient to the required position

for delivery of the radiation. Before each treatment is delivered, image verification

takes place to ensure that the patient is in the same position and assess if there has

been any anatomical changes. Typically 2D transmission (portal) or 3D cone-beam

(CBCT) images are acquired. The difference is assessed and where appropriate the

patient is shifted to ensure that the treatment plan will be accurately delivered. If the

difference exceeds pre-determined limits and the patient position cannot be adjusted

to match the planned treatment the CT scan and planning process may need to be

repeated.

More recently, as an alternative to tattoos, there has been an increased use of surface

guided radiotherapy for patient treatments. This technology uses wall mounted,

optical surface imaging to help with patient set-up and enables motion management

to be carried out during the delivery of the radiation beam, rather than just prior to

treatment. The system works by disabling the radiation beam when the surface

position exceeds a defined tolerance. When used in conjunction with Deep Inspiration

Breath Hold (DIBH) for left-sided breast treatments it can be particularly beneficial in

minimising dose to the heart.

1.10 Radiotherapy Treatment Planning for Mastectomy Patients

As previously discussed, radiotherapy is recommended by NICE (NICE, 2018) for

patients who have undergone a mastectomy and that are at risk of local recurrence.

Radiotherapy post-mastectomy can be delivered using the same technique as

previously described using opposing tangential fields. However the technique can pose

some challenges for this treatment technique which are discussed as follows.

High energy photon beams are required in radiotherapy treatments to ensure

significant penetration of the radiation beam at depth within the patient. The use of

high energy photons also results in a skin sparing at the patient surface. This is as a

Page 35: Development of a technique using VMAT and robust

35

result of the photon interactions within the tissue and is also known as the build-up

effect. When high energy photons are incident on the tissue surface secondary

electrons are produced. These electrons travel primarily in a forward direction, into

the tissue and deposit their energy at depth in the patient. Since there are minimal

interactions upstream from, or outside, the patient surface, the dose to the surface is

relatively low. The effect can be seen in Figure 1.7 which shows the variation in dose at

depth within water. The dose builds up to a maximum value at a depth of

approximately 1.4cm, and then decreases at the depth increases. This can normally be

advantageous in external beam treatments where tumours are at depth, as reduced

dose at the surface can minimise radiation induced skin reactions such as erythema.

However in the case of mastectomy patients local recurrences have been shown to

occur on the skin and in subcutaneous tissues suggesting the importance of treating

this superficial area (Andry et al., 1989). Similarly, Thoms et al. (1989) found that in a

retrospective study of 61 patients with inflammatory breast cancer, the most common

site of failure was in the chest wall and more common in patients who did not suffer

from skin toxicities such as erythema or moist desquamation.

Figure 1.7: Measured Depth Dose curve for a 6MV photon beam, 10x10cm field size.

A common method to overcome the build-up effect is the use of a tissue equivalent

material in the form of a 0.5 – 1cm thick, gel slab that covers the entire chest-wall and

is placed directly on to the skin during treatment. The bolus then provides a material

Page 36: Development of a technique using VMAT and robust

36

for photon interactions to occur outside the patient and allows dose to be deposited at

the surface.

The clinical use of bolus between centres varies widely. In 2004 an international survey

of radiation oncologists was carried out by Vu et al., (2007) which focused on the

technical details regarding the use of bolus in post-mastectomy radiotherapy (PMRT).

1035 responses were received from oncologists practising in the USA, Canada, Europe

and Australasia. The results showed that 68% of responders always used a bolus, 6%

never used one and 26% used bolus if there were specific indications (for example, skin

involvement or inflammatory disease). Respondents from the Americas were

significantly more likely to always use bolus (82%) compared to the Europeans (31%).

There was also variation in frequency of use, with 33% or respondents using bolus

throughout treatment and 46% using it on alternate days. Furthermore the thickness

of bolus varied, with 35% using <1cm and 48% ≥ 1cm. The variation in practice is

reported in other studies, Blitzblau and Horton (2013) and Mayadev et al. (2015)

similarly identifying a range in use of bolus with regards to frequency and thickness.

Mayadev et al. (2015) also recognising that a range of materials were also being

including brass bolus mesh and customised wax bolus.

As in the UK with the NICE guidelines, The American Society of Clinical Oncology

(ASCO), recommend the use of radiotherapy post-mastectomy, but similarly offer no

specific advice on how bolus should be used (Recht et al., 2016). Recent clinical trials

recruiting post-mastectomy patients SUPREMO (ISRCTN61145589) and FAST-Forward

(ISRCTN19906132) also fail to define a consistent approach to the use of bolus in their

protocols.

There is also evidence that suggests bolus is not required at all. A number of studies

have reported that retrospective data analysis showed no significant differences in

chest-wall recurrence between patients that were treated with bolus and those who

were not (Tieu et al., 2011; Turner et al., 2016; Abel et al., 2017; Nakamura et al.,

2017). The use of bolus has also been associated with increased skin toxicity (Pignol et

al., 2015) which has further been associated with treatment interruptions (Abel et al.,

2017).

Page 37: Development of a technique using VMAT and robust

37

The use of bolus also has practical challenges associated with it. Bolus materials can

increase the attenuation of dose at depth within the patient depending on the

thickness of material. When bolus is only used for a proportion of the fractions this

means that two treatment plans are required for every patient. There can also be

issues with the malleability of the bolus material, resulting in a lack of conformity of

the bolus to the chest wall during treatment delivery. It has been demonstrated that

air gaps between the bolus and chest wall can reduce the surface dose (Butson et al.,

2000; Boman et al., 2018). Since treatment planning software often assumes the bolus

is in constant contact with the skin, air gaps at treatment will result in a discrepancy

between planned and delivered superficial doses. A solution to these problems include

the use of thinner bolus material for use at all treatment fractions, being thinner the

sheets may conform better to the patient surface and if used in every treatment only

one treatment plan is required (Das et al., 2017). Alternatively, the use of a brass mesh

bolus is becoming more common, overcoming both these issues (Healy et al., 2013;

Ordonez-Sanz et al., 2014; Richmond et al., 2016). However, the use of any additional

bolus material as part of the treatment will also incur cost to purchase, requires

infection control considerations and runs the risk of potential clinical error if it there is

failure to use it during the patient’s treatment or it is incorrectly positioned.

1.11 Use of IMRT and VMAT Treatment Planning for Mastectomy Patients

As IMRT has become more common place in radiotherapy centres, the use of IMRT

(forward and inverse) and VMAT has increasingly been investigated as an alternative

treatment to 3DCRT for breast and chest-wall patients. Whilst most of this literature

focuses on radiotherapy treatment for breast patients after breast conserving surgery

(Donovan et al., 2007; Badakhshi et al., 2013; Jin et al., 2013; Haciislamoglu et al.,

2015; Virén et al., 2015; De Rose et al., 2016; Jo et al., 2017; Jensen et al., 2018)

increasingly the use of the technique is being investigated for post-mastectomy

patients too (Cavey et al., 2005; Rudat et al., 2011; Zhang et al., 2015; Xu and Hatcher,

2016). There is no consistently applied treatment approach, with the specific

techniques for IMRT and VMAT varying widely in the literature. IMRT breast plans can

range from a tangential two field beam arrangement as used by Rudat et al. (2011) to

a 7 or 9 field beam arrangement with the beams at evenly spaced angles to cover the

Page 38: Development of a technique using VMAT and robust

38

breast as used in studies by Ekambaram et al. (2015) and Popescu et al. (2010).

Similarly, VMAT plans may constitute two partial tangential arcs, or be a solid arc

around the chest-wall (Karpf et al., 2019). The optimisation methods also vary across

studies. For example Cavey et al. (2005) investigated use of the IMRT with a forward-

planned technique, requiring input from treatment planners to optimise the plans

whilst others use the inverse planning technique, allowing the treatment planning

software to optimise the distribution (Popescu et al., 2010; Virén et al., 2015).

The use of IMRT and VMAT has in general been shown to be favourable, for the

treatment of breast cancer, with conservative surgery, with more homogeneous dose

distributions, better conformity, ability to achieve similar dose to organs at risk and

increased skin sparing (Freedman et al., 2009; Johansen, Cozzi and Olsen, 2009;

Almberg, Lindmo and Frengen, 2011; Haciislamoglu et al., 2015; Jo, Kim and Son,

2017). In a randomised trial of standard 2D radiotherapy versus IMRT, it was shown

that the minimisation of in-homogeneities reduced late effects and reduced the

incidence of change in breast appearance (Donovan et al., 2007). A phase II trial of

hypo-fractionated VMAT-based treatment of early stage breast cancer also

demonstrated that the technique was well tolerated, with no pulmonary or

cardiological toxicities reported (De Rose et al., 2016). In papers that compared IMRT

and VMAT for whole breast treatment, a number reported that VMAT was less

favourable than IMRT due to higher contralateral breast doses and some organs at risk

receiving higher volumes of low dose (Badakhshi et al., 2013; Jin et al., 2013; Virén et

al., 2015).

Similarly for post-mastectomy patients the advantages of VMAT and IMRT were

reported by a number of authors comparing plans dosimetrically with standard

techniques (Ma et al., 2015; Zhang et al., 2015; Lai et al., 2016). However, the issue of

low dose bath was highlighted again, with Xu and Hatcher (2016) suggesting that this

limited the benefit of VMAT as there was not significant improvement in PTV coverage,

they did however propose that in cases where the standard technique could not

achieve the required planning constraints VMAT should be considered.

Previous studies have shown that IMRT and VMAT is a radiotherapy technique that can

be used to treat post-mastectomy patients. The definition and exact technique used

for these plans varies including the use of forward planned or inverse planned

Page 39: Development of a technique using VMAT and robust

39

approaches. However, inverse-planned IMRT can potentially pose some problems for

superficial lesions. Lee et al. (2002) reported severe skin reactions with the use of IMRT

for treatment of head and neck cancers. Similarly, a study by Higgins et al. (2007)

comparing the use of IMRT to a bilateral field arrangement for head and neck

treatments showed that IMRT increased the maximum surface dose from 69% of the

prescription dose to 82%. The effect was discussed in a study by Thomas and Hoole

(2004) which concluded that the issue was due to the PTV, created to account for

uncertainties in set-up, extending outside the patient surface. To meet the objectives

related to PTV coverage, defined in the treatment planning software, beam segments

near the patient surface require a high photon fluence. The high fluence is required to

overcome the build-up effect due to low electron density outside the patient surface,

and lack of material to induce scatter of the incident photons. If the patient position or

contour changes, the tissue intersecting the high fluence will change, potentially

resulting in a skin dose greater than originally predicted by the treatment planning

system.

1.12 Robust Optimisation

Radiotherapy treatment planning systems simulate the dose deposited within the

patient based on modelling the interactions of the radiation from the linear

accelerator as it enters the patient. For inverse planned treatments using IMRT and

VMAT, the intensity and variation of the radiation beam is determined by specifying

the ideal dose to structures defined on the patient’s CT scan. This would include the

dose needed to control the tumour and the dose to organs at risk which would

minimise any side effects. Since radiation travels through and interacts with both

tumour and normal tissue, optimising the dose distribution is a complex mathematical

problem. Within the planning system software objective functions are used on the

targets and organs at risk, which reflect the optimal clinical goals and can be expressed

as a dose to volume. The beam intensities are then modified in an iterative process to

minimise the difference between the required objective functions and the deposited

dose. Additional weightings can be put on the objective functions to increase the

probability of meeting the clinical goal.

Page 40: Development of a technique using VMAT and robust

40

The robustness of a treatment plan is its ability to maintain its clinical goals despite

differences between the patient at CT and at treatment. The differences could include

changes in patient contour or set-up position.

The typical approach to dealing with uncertainty has been described previously and is

the use of a margin around the tumour volume referred to as a PTV. Similarly margins

around organs at risk can be generated, Planning Risk Volumes (PRV). The PTV and

PRVs are then used as the structures in the treatment planning software to which

objective functions are applied.

There are however several limitations to the PTV concept, as described by Unkelbach

et al. (2018) including the assumption of a static cloud dose distribution where the

dose distribution is considered invariant. In the case of breast and chest-wall

treatments, where the target structure is in close proximity to a large density

difference (air outside the patient) this approximation is likely to be challenged. Two

alternative approaches described in literature that have been used for radiotherapy

treatment planning include probabilistic planning (Chan et al., 2006; Ramlov et al.,

2017; Tilly et al., 2019) and minimax optimisation (Fredriksson et al,. 2011; Byrne et

al.,2016; Archibald-Heeren et al., 2017; Miura et al., 2017; Wagenaar et al., 2019). In

both methods robust optimisation dose distributions are optimised in different

scenarios, for example, geometric position. The probabilistic approach optimises the

objectives on the likelihood of a particular scenario. The minimax optimisation

approach optimises the objective value in the worst-case scenario based on

positioning accuracy in different directions. The advantage of the minimax method is

that only knowledge of the different geometric scenario is needed, rather than the

probability distribution, however the minimax may over optimise in scenarios of where

likelihood of a geometric position is lower, potentially compromising treatment plan

quality.

In the context of breast radiotherapy, the combination of robust optimisation with

IMRT and VMAT has previously been established as a clinically acceptable approach

(Byrne et al., 2016, Liang et al. 2020 and Dunlop et al. 2019). As mentioned in the

previous section the challenge with treating superficial lesions with IMRT is trying to

ensure coverage with setup variations without creating high photon fluence segments,

which can result from using PTVs extending into air. Techniques such as the use of

Page 41: Development of a technique using VMAT and robust

41

virtual bolus can help overcome this, where the plan is initially optimised with a bolus

material at the surface, and then removed for the final calculation, resulting in

segments with suitable flash but not with a high fluence. However the dose differences

between the optimised and final dose, can lead to multiple re-optimisations required

to achieve an acceptable plan. Byrne et al. (2016) therefore evaluated the use of

robust optimisation with IMRT as an alternative approach to virtual bolus, showing

that it was a comparable technique to ensure coverage of breast CTV with setup

variations. Liang et al. (2020) compared the robust optimisation method to IMRT plans

where the flash margin was manually created by editing MLC positions. They

concluded that the robustly optimised plans were the only ones that met acceptable

criteria for PTV coverage under geometric error scenarios. Dunlop et al. (2019) also

demonstrated that the robust optimisation with VMAT could be used for patients

requiring breast radiotherapy that included the treatment of the internal mammary

chain. In addition to showing that compared to non-robust plans there was no

compromise in dose to organs at risk, the target coverage was improved using the

robust optimisation feature, when robustness was assessed over a typical treatment

course.

1.13 Specifying Superficial Doses for Post-Mastectomy Radiotherapy

In radiotherapy, treatment plans are usually assessed on particular, well-defined

criteria. Typically, the PTV is considered sufficiently covered if 95% of the volume is

covered by 95% of the prescription dose and 2% of the volume does not exceed 107%.

In addition to the ICRU guidelines on PTV homogeneity, clinical trials are often a good

source for PTV objectives and OAR constraints. However assessing dose distributions in

the case of post-mastectomy patients can be challenging. The variation in the use of

bolus, discussed previously, suggests that there is no consensus in the skin dose

required to ensure successful treatment and reduce recurrence. Thoms et al. (1989)

suggested that the measure of skin reaction is a good indication that sufficient dose

had been received, with reactions such as erythema and moist desquamation being

linked to an improvement in local control, though the absolute dose to skin was not

discussed. However, with it known that skin toxicity is also affected by patient habits

such as smoking (Pignol et al., 2015) the use of erythema may not be a good measure

of dose received. In the international survey on post-mastectomy radiotherapy

Page 42: Development of a technique using VMAT and robust

42

practise, Vu et al. (2007) found that responders generally defined an ‘adequate’ dose

to the skin was 85-90% of the prescription dose. This is in line with 75%-90% of the

prescribed dose that is suggested as the required dose to skin and the mastectomy

scar, in a chapter of ‘Radiation Oncology Management Decisions’ (Chao et al., 2011)

describing the management of breast cancer. Ordonez-Sanz et al. (2014) investigating

a single-plan solution for a cohort of post-mastectomy patients, measured superficial

doses which were on average 85.1% of prescription dose. Another approach to skin

dose assessment, taken by Xu and Hatcher (2016), was to create a skin volume of 3mm

below the patient and aim to ensure that 5% of the structure received 100% of the

dose and 85-95% should receive 80% of the dose.

The lack of consistency in defining the required dose and where the dose should be

delivered, means comparing techniques and patient outcomes between studies is

particularly challenging.

1.14 Definition of the Skin Structure in Mastectomy Patients

The use of skin volumes for plan assessment and patient outcome analysis could be

considered a valid approach in radiotherapy treatment, where reporting on dose to

structures is routine. The skin is composed of two distinct regions, the epidermis and

the dermis. The epidermis being the most superficial layer, made up of closely packed

epithelial cells. The dermis lies below the epidermis and is made up of connective

tissues, nerve endings, oil glands, blood vessels and lymphatic vessels. The ICRP

(International Commission on Radiological Protection and ICRU (International

Commission on Radiation Units) recommend that skin depth for practical dose

assessment is defined at 0.07mm deep, the basal layer, the deepest epidermal layer

and interface with the dermis (ICRU, 1985; ICRP, 1991). However, from a radiotherapy

treatment perspective, it is suggested by Butson et al. (2000) and Javedan et al. (2009),

that the dermal layer containing the dermal lymphatics, should be targeted as a site of

recurrence, which is thought to be >1mm deep. The actual values for epidermis and

dermis layers will also vary between patient and body region. A study by Oltulu et al.

(2018) using a histometric technique, reported the mean dermal layer in the female

breast to be 4.7mm deep. With the development of new imaging techniques, breast CT

Page 43: Development of a technique using VMAT and robust

43

studies by Huang et al. (2008) and Shi et al. (2013) found the average breast skin

thicknesses (including epidermis and dermis layers) were approximately 1.45 mm and

ranged from 0.8mm to 2.5mm.

The suggested range of breast skin thickness and ambiguity over the layers required to

be treated during radiotherapy further complicates comparisons with other planning

studies and conclusions regarding recurrences.

1.15 Measurement of Superficial Doses

In Figure 1.7 a steep dose gradient is observed when photon beams intersect an air-

tissue interface. Measurements by Devic et al. (2006) using Gafchromic film, showed

that for a 6MV photon beam and field size 10x10 cm2 the dose increases from 14% at a

depth of 4µm to 43% at 1mm. The superficial dose in this region is attributed to

contaminant electrons, from the treatment head and secondary electrons from the

irradiated material. The magnitude of these effects will depend on beam

characteristics including field size, distance from source (Bjärngard et al., 1995; Kim et

al., 1998) and beam incident angle (Gerbi et al., 1987). For correct superficial doses to

be reported by the treatment planning system it is important that the algorithm

models this correctly. However as reported by Panettieri et al. (2009), Cao et al. (2017)

and Dias et al. (2019) the measured dose difference can be up to 20% depending on

the sophistication of the algorithm. Furthermore, the accepted tolerance in this region

recommended by the American Association of Physicists in Medicine (AAPM) is ±20%

(Fraass et al., 1998). The discrepancies between planning system and measurements in

the build-up region can be attributed to type of treatment planning algorithm Cao et

al. (2017), accuracy of data measurements input into the planning system Chetty et al.

(2007) and the size of dose calculation grid (Kan et al., 2012). The size of dose grid used

may not be clinically appropriate for reporting skin dose, if larger than the defined skin

thickness.

With these limitations in mind it is appropriate to measure the skin dose directly. The

types of detectors used by research groups to assess superficial breast and chest-wall

doses include detectors thermoluminescent (TLDs), optical luminescent detectors

(OSLDs), metal-oxide semiconductor field effects transistors (MOSFETs) and

Page 44: Development of a technique using VMAT and robust

44

radiochromic film. There are no papers that directly compare all of these techniques at

once, but a few have compared a combination of the detectors. Quach et al. (2000)

measured superficial doses on a chest-wall phantom using radiochromic film, TLDs and

MOSFETs. With a single tangential field applied to a hemicylindrical phantom it was

shown that on the central axis, with an obliquity of 0⁰, the surface dose measurements

from each detector differed significantly: 28% (of Dmax) for the radiochromic film, 30%

for the TLDs (extra thin type) and 43% for the MOSFETs. This was attributed to the

effective depth of measurement of each detector. The group also identified

advantages and disadvantages of each detector from a practical view. They found that

the radiochromic film was potentially useful as a skin dosimeter due to its effective

depth of 0.17mm and ability to simultaneously measure the dose profile. However, the

sensitivity of the material meant it could only be used in the dose range of 10-100Gy.

The TLDs used (an extra thin type with an effective depth of 0.14mm) were felt to have

a more appropriate dose range for measuring a single fraction but had the

disadvantage of only being able to take point measurements. With an effective depth

of 0.5mm the MOSFETS, had the advantage of instant read outs and good spatial

resolution but also had some directional sensitivity. It was concluded that TLDs

provided good clinical skin dose measurements despite the work required in annealing

and calibrating.

Jong et al. (2016) investigated the use of the MOSkin detectors, MOSFETs designed to

provide a WED (water equivalent depth) of 0.07mm, making them suitable for

measuring skin dose. Comparing MOSkin measurements with Gafchromic (EBT2) film

and the treatment planning system (pencil beam convolution algorithm) they observed

no significant difference in skin doses when measurements were made under the

bolus. However, when measurements were made with no bolus the mean skin doses

measured using EBT2 were 11.4% higher than the MOSkin, with the planning system

reporting 14.2% higher. The dose difference between the detectors and planning

system were explained as the limitations in the planning system algorithms, grid size,

and the WED difference between detectors. Similar results using MOSkin and

Gafchromic (EBT3) film were described by another research group that used the

detectors to assess the effect of brass bolus for chest wall irradiation. EBT3 Gafchromic

film has the same composition and thickness of sensitive layer as EBT2 but has a

Page 45: Development of a technique using VMAT and robust

45

symmetric layer configuration allowing both sides of the film to be used.

Measurements were also made with an Advanced Markus ionisation chamber.

Measurements of entrance dose, attenuation and PDD were measured in a solid water

phantom, and surface beam profile on a curved phantom. In the no-bolus situation

they reported that the Markus chamber measured the surface dose for a 6MV beam as

16.5%, EBT3 22.8±3.8% and MOSkin detectors 19.2±1.0%. This difference was also

considered to be a function of the WED, with effective depth of the Markus chamber

reported as 0.023mm, EBT3 as 0.153mm and MOSkin 0.07mm.

Yusof et al. (2015) investigating dosimeters for assessing surface dose compared OSLDs

with Gafchromic EBT3 and Markus ionisation chamber measurements. The primary

objective of this study was to investigate whether OSLDs could provide accurate in-vivo

measurements for patients within a reasonable time frame and were easy to process.

They acknowledged that the advantage of Gafchromic film was with its ability to

provide a two-dimensional distribution but highlighted that waiting for 24 hours post

irradiation to ensure the film was stable could be a problem. TLDs they described as

tissue equivalent but the process for reading them was tedious and time consuming,

and MOSFETs although provided useful real-time read out had a WED of 0.8-1.8mm,

deeper than the recommended skin measurement depth. They describe OSLDs as

having good reproducibility and linearity, that reach stability 16 minutes post

irradiation and are easier and less time consuming to readout compared to TLDs and

film. The group determined the WED of the OSLDs to be 0.4mm. Surface dose

measurements in a solid water phantom with a 6MV photon beam were reported as

15.95±0.08% for the Markus chamber, 23.79±0.68% for the Gafchromic EBT3 and

37.77±2.0% for the OSLD, with the dose difference explained by the high dose gradient

and differing WEDs of the detectors.

TLDs were used to measure superficial doses by (Ordonez-Sanz et al. (2014) to assess

the use of different bolus materials for chest-wall irradiation. As part of their

investigation they found that in the build-up region TLD measurements were

consistent with diode measurements to within 3% giving them confidence that TLDs

could be calibrated and used effectively for measurements in steep dose gradients. In

comparison to the treatment planning system which used an anisotropic analytical

algorithm (AAA) it was shown that the TLDs measured slightly lower than predicted. It

Page 46: Development of a technique using VMAT and robust

46

was suggested that this was an overestimation of the dose by the treatment planning

system and due to the electron contamination source used in the AAA model.

The literature demonstrates that a range of dosimeters can be used for measuring

superficial doses. Factors to be considered in deciding which dosimeter to use include;

the requirement for discrete or 2D distributions, the need for immediate read-out of

results and the actual depth of measurement required, as discussed previously the

definition of skin depth can vary.

1.16 Scope of Project

The use of radiotherapy for treating patients following mastectomy is a well evidenced

and globally accepted technique. Evidence suggests that these patients also benefit

from increased surface dose, however there is little consensus as to the actual dose

required. A common approach to increasing surface dose is the use of bolus materials

for part of the patients’ treatment. The addition of this bolus material can present

problems including; the requirement to create two treatment plans, increasing

planning time in the patient pathway, air gaps between the patient and bolus affecting

surface dosimetry and the risk of treatment errors due to bolus placement being

missed. It has been shown that IMRT techniques can deliver higher surface doses

compared with 3DCRT by loading the build-up region with beam segments from

glancing angles. This is often unintentional but could be beneficial for patients with

lesions close to the skin surface, potentially negating the need for bolus. However,

fluence loading in the surface regions means that IMRT plans could show unacceptable

changes in dose distribution when positioning errors are examined. The use of robust

optimisation could be a means to minimise this effect as part of the treatment

planning process.

The overall aim of this thesis therefore, is to investigate whether a combination of

inverse planning with robust optimisation could provide an alternative, one plan

solution, to the current approach to radiotherapy treatments for patients requiring

post-mastectomy radiotherapy.

The thesis firstly establishes the effect of bolus on surface dose using the current

clinical technique, and then demonstrates that clinically acceptable VMAT plans can be

Page 47: Development of a technique using VMAT and robust

47

created with comparable skin doses. To confirm superficial doses predicted by the

treatment planning system, results of physical measurements with TLDs are also

presented. With limitations expected of VMAT plans due to patient positioning, the

impact of perturbation on these plans is investigated and shown to be clinically

unacceptable. Treatment plans using VMAT and robust optimisation are then

presented, demonstrating the impact that including robust optimisation has on

reducing the effect of perturbation. Finally, limitations to the results and further work

that would be required in order to use the technique clinically is discussed.

Page 48: Development of a technique using VMAT and robust

48

Chapter 2

2 Evaluation of current treatment method

The objective of this study is to assess whether a novel radiotherapy technique can be

used to plan radiotherapy treatments for post-mastectomy patients. The particular

challenge with this site of treatment is the skin surface, which is normally spared with

high energy photon beams, but that evidence suggests requires treatment to some

extent. Although the acceptable dose to this area is not well defined either in dose

required or the definition of its volume, the primary objective of this study is to see

whether similar distributions can be achieved compared to the current method using

bolus, particularly within the first 5mm of the skin surface.

As the literature does not specify in detail what radiation dose is required to reduce

local recurrence at the skin surface, dose delivered to the skin structures using the new

technique will be compared directly to the current technique used within this clinic.

However, the current level of recurrences within the department is deemed to be

acceptable, and the constraints and objectives for target coverage and organ at risk

dose used within our clinic have been implemented based on those defined in the

control arm of the breast trial, FAST-Forward (ISRCTN19906132).

2.1 Current technique for the treatment of post-mastectomy patients

At the University Hospitals Birmingham radiotherapy is prescribed following

mastectomy when there is:

Node positive disease

Incomplete or close margin (<1mm) on microscopy

Muscle invasion at deep margin and further surgery not possible

Tumours >5cm

Page 49: Development of a technique using VMAT and robust

49

Vascular/Lymphatic invasion with other risk factors e.g. high grade

High grade tumours with additional risk factors e.g. tumour >30mm

Multifocality with additional risk factors.

The prescription dose for these post-mastectomy patients is 40Gy in 15#. This is

delivered on a daily basis, over a 3-week period (not including weekends). Bolus is used

when requested by the clinicians and is typically used in cases of high-risk patients with

documented skin involvement. A large square of bolus slab of 1cm thick water

equivalent material, is used which covers the entire treatment areas and is applied for

the last 7 fractions of the patients’ treatment (Figure 2.1).

a)

b)

Figure 2.1: a) water equivalent bolus slab 40x40cm b) 3D rendered image from CT scan showing bolus placement in treatment position

As mentioned in the introduction the bolus is used to increase the surface dose by

counteracting the build-up effect when using high energy photons.

Patients are scanned on a Philips Brilliance Big Bore CT Scanner (Philips Medical

Systems, Eindhoven, Netherlands) with a slice thickness of 3mm. The patients are set-

up in a supine position with arms above their head, immobilised using a MT-350 CIVCO

(CIVCO Radiotherapy, Orange City IA, USA) breast board. Radio-opaque wires are

placed on the surface of the patient’s chest during the scan to delineate the area for

treatment and the CT images are exported to the radiotherapy simulation software

ProSoma (MedCom, Darmstadt, Germany). The patient is not routinely scanned with

Page 50: Development of a technique using VMAT and robust

50

bolus in situ, this is applied virtually by the treatment planning software during plan

optimisation. In the simulation software tangential beams, with a non-divergent back

edge are applied to the patients CT scan to cover the area of treatment, this is

reviewed and approved by the clinician (Figure 2.2).

a)

b)

Figure 2.2: a) 3D rendered image from CT scan with wires defining area for treatment b) Transverse CT image with opposing tangential beams applied

Patient CT scans are transferred to the treatment planning software, RayStation

version 6.0.0.24 (RaySearch, Stockholm, Sweden). The software uses atlas-based

segmentation to automatically generate the required OARs:- heart, left lung and right

lung, patient surface and a PTV structure. As described previously a field based PTV is

created from the intersection of the tangential fields with the patient tissue.

Treatment plans are created to meet the objectives and constraints defined in the

FAST-Forward Trial summarised in Table 2.1. The plans are calculated using a 3mm

dose grid with the RayStation collapsed cone treatment planning algorithm.

Structure Mandatory Optimal

PTV V95% ≥ 90% V95% ≥ 95% V105% ≤ 7% V105% ≤ 5% V107% ≤ 2% - D1% ≤ 110% (Dmax) -

Ipsilateral Lung V30% ≤ 17% V30% ≤ 15% Heart V25% ≤ 5% -

V5% ≤ 30% -

Table 2.1: Departmental objectives and constraints used for breast planning based on those used in FAST-Forward Trial (ISRCTN19906132)

Page 51: Development of a technique using VMAT and robust

51

For post-mastectomy patients two treatment plans are created, one with and one

without bolus. The bolus is assigned virtually to the dataset in the treatment planning

system, with the creation of a 1cm rind over the surface of the chest-wall with a

density set to 1g/cm3(water equivalence). Figure 2.3 shows an example of the bolus

structure created in the treatment planning system.

Figure 2.3: Blue contour on transverse CT slice indicates position of computer- generated bolus

The no-bolus plans are optimised to deliver 21.33Gy in 8# and the bolus plans 18.66Gy

in 7#. Plans are created using 6MV, 10MV or a mixture of both energies, dependent on

the size of the patient and the separation between the medial and lateral beams.

Treatment plans are generated using a SMLC technique for delivery on Elekta Precise

or Elekta VersaHD (Elekta, Crawley, UK) linear accelerators with 1cm and 5mm MLCs

respectively.

Page 52: Development of a technique using VMAT and robust

52

2.2 Dosimetric effect of bolus in post-mastectomy patients

2.2.1 Patient selection

To assess whether a new technique using no bolus is a suitable alternative the current

technique, the impact of bolus in the current setting was first evaluated. The dose to

the skin was of particular importance in comparison of techniques but the impact of

dose to organs at risk also needed to be considered. The treatment plans of 25

previously treated post-mastectomy patients were collated and reviewed. At

University Hospitals Birmingham the patient consent process allows the use of data for

audit and service development purposes provided these are anonymised. No further

ethics approval was sought for this planning study. All patients within the study were

receiving treatment of the chest wall only and did not include the treatment of

supraclavicular nodes, axilla or internal mammary nodes. The initial cohort of patients

were a group of consecutively planned treatments and included 12 right sided

treatments and 13 left sided treatments. Table 2.2 summarises the treatment machine

and beams energies used for these patients.

After reviewing the treatment plans of the 25 consecutively treated post-mastectomy

patients, the variation of treatment machines parameters used for each case was

analysed. It was noticed that 7 of these patients had no bolus requested at all, and a

range of energy combinations and different treatment machines had been used. The

current radiotherapy treatment technique does not aim to ensure a particular skin

dose and as discussed previously there is no consensus on what this should be and the

use of bolus is left to the discretion of the clinician. It was decided therefore, that the 7

cases without bolus originally, would be removed from the analysis as enhancement to

skin dose had not been required. As previously discussed, build-up effect, and

therefore skin dose is directly related to beam energy, this means that patients treated

with 6MV for both the bolus and no bolus plans will have different relative skin doses

to a patient treated with 10MV for both the bolus and no bolus plans. Therefore to

ensure that any conclusions drawn from this work were a result of the new technique

rather than variation in original skin dose, it was decided to rationalise the data set to

those with the same original beam energy characteristics. The largest proportions of

cases with the same characteristics were treated with 6MV for the no bolus plan and

Page 53: Development of a technique using VMAT and robust

53

10MV for the bolus plan, these 8 patients went on for further analysis. In this group of

8 patients 5 were left sides cases and 3 were right sided.

Patient Site Treatment machine

No Bolus Energy, MV

Bolus Energy, MV

1 Right CW Elekta Precise 6 NA

2 Right CW Elekta Precise 6 10

3 Left CW Elekta Precise 6 10

4 Right CW Elekta Precise 6 & 10(segments) 10

5 Right CW Elekta Precise 6 6 medial & 10 lateral

6 Left CW Elekta Precise 6 NA

7 Right CW Elekta Precise 6 NA

8 Right CW Elekta Precise 6 NA 9 Left CW Elekta Precise 6 NA

10 Left CW Elekta Precise 6 10

11 Right CW Elekta Precise 10 10

12 Right CW Elekta Precise 6 10

13 Left CW Elekta Precise 6 & 10(segments) 10

14 Left CW Elekta Precise 6 & 10(segments) 10 15 Left CW Elekta Precise 6 10

16 Right CW Elekta VersaHD

6 & 10(segments) 10

17 Left CW Elekta Precise 6 10

18 Right CW Elekta Precise 6 10

19 Left CW Elekta Precise 6 & 10(segments) 10

20 Right CW Elekta Precise 10 10

21 Left CW Elekta VersaHD

6FFF 6FFF

22 Left CW Elekta Precise 6 NA

23 Left CW Elekta Precise 6 10

24 Right CW Elekta Precise 10 & 6 (segments) NA 25 Left CW Elekta Precise 10 & 6 (segments) 10 & 6 (segments)

Table 2.2: Summary of patient information including treatment site (CW=chest wall), machine and beam energy

Page 54: Development of a technique using VMAT and robust

54

2.3 Method

To assess the impact of bolus a retrospective study was conducted in this group of 8

patients, all receiving 40Gy in 15#. The dose to clinically relevant structures were

compared for two different treatment plans. The first treatment plan being that of the

clinically delivered plan, 8# no bolus and 7# bolus, subsequently referred to as ‘Clinical’

plan. The second plan received all 15# with no bolus, subsequently referred to as ‘No

Bolus’ plan.

Structures created for the dose comparison were selected due to their clinical

relevance in the plan analysis and to assess whether the plan was clinically acceptable,

based on the dose constraints and objectives in Table 2.1.

Since dose to the skin was of particular interest in this study it was therefore decided

to include this in the PTV structure and was created using the field-based technique

described previously, without clipping 5mm from the patient surface. This structure

was named PTVtoSurface, an example is shown in Figure 2.4. It should be noted that

the chest-wall shown in Figure 2.4, and used as an example later in the text is an

outlier with regards to chest wall thickness. The average chest wall thickness for the 8

cases analysed was 2.1cm ranging from 1.1cm to 3.9cm.

Figure 2.4: PTVtoSurface (red shaded structure)

Page 55: Development of a technique using VMAT and robust

55

Due to the particular interest in the superficial area of the chest-wall tissue, 3 skin

structures were created to report on. These structures were 1mm, 3mm and 5mm skin

rinds directly below the patient’s external surface, which was defined by the treatment

planning software. The skin rinds extended laterally to the same extent as the

PTVtoSurface. These structures were named Skin1mm, Skin3mm and Skin5mm. An

example of Skin5mm is shown in Figure 2.5. The organs at risk heart and ipsilateral

lung were also used for the plan comparison. Dose volume histograms (DVH) were

generated in the treatment planning system for each patient, structure and technique.

To analyse the plan techniques quantitatively particular dosimetric parameters were

also extracted. For the PTVtoSurface structure, in addition to the parameters described

in Table 2.1 the average dose to the structure was also recorded. For the skin

structures D99% (dose covering 99% of the volume, minimum dose), D1% (near

maximum dose) and average dose were recorded. For the organs at risk, V25% and

V5% were recorded for heart and V30% for the ipsilateral lung. A summary of the

dosimetric data that was collated for each structure and is summarised in Table 2.3,

with the absolute dose values stated where appropriate based on the prescription

40Gy/15#.

Page 56: Development of a technique using VMAT and robust

56

Figure 2.5: Skin5mm (shaded green structure) with PTVtoSurface (red contour)

Structure Dosimetric Parameter (%)

Dosimetric Parameter (Gy)

PTVtoSurface V95% V105% V107% Average D1%

V38Gy V42Gy V42.8Gy Average D1%

Skin1mm, Skin3mm, Skin5mm D99% D1% Average

D99% D1% Average

Heart V25% V5%

V10Gy V2Gy

Ipsilateral Lung V30% V12Gy

Table 2.3: Summary of dosimetric parameters recorded for each patient

2.4 Results

2.4.1 Plan comparison – Dose Distribution (single patient example)

Figure 2.6 shows an example of the 38Gy coverage achieved with the No Bolus and

Clinical plans through a transverse plane for one of the patients in the group. It can be

Page 57: Development of a technique using VMAT and robust

57

observed that the dose in the Clinical Plan extends closer to the surface than for the

No Bolus plan.

a)

b)

Figure 2.6: Example 38Gy dose distribution for a) No Bolus plan b) Clinical Plan

2.4.2 Plan comparison – PTVtoSurface (single patient example)

Figure 2.7 shows an example of the DVH for the PTVtoSurface, for one of the patients.

Specific dose parameters were extracted from the DVHs for each plan technique. The

values achieved for the specified dose parameters for the PTVtoSurface, along with the

mandatory and optimal constraints, where appropriate, are shown in Table 2.4.

In the DVHs for the No Bolus and Clinical plans, for the PTVtoSurface structure (Figure

2.7), a difference in the shape at the shoulder of the curves at around 35-39Gy is

observed, with the rest of the DVH matching closely. As expected, it is the No Bolus

plan that reports lower volumes in this dose region. This effect is also seen in the

analysis of the dosimetric parameters. The volume reported for the No Bolus plan for

the V95% parameter, is 3.5% lower than the Clinical plan and the volume difference for

the V105% and V107% parameters is +0.5% and +0.1%, respectively. The maximum

dose difference to the PTVtoSurface structure between the plans is 0.5% and the

average dose is 0.2Gy lower in the case of the No Bolus plan.

Page 58: Development of a technique using VMAT and robust

58

Figure 2.7: Example of DVH for structure PTVtoSurface. (Dashed line = Clinical Plan, Dotted line = No Bolus Plan)

Structure Dosimetric Parameter

Mandatory Constraint

Optimal Constraint

Clinical Plan

No Bolus Plan

PTVtoSurface V95% V105% V107% D1 % Average

≥ 90% ≤ 7% ≤ 2% ≤ 110%

≥ 95% ≤ 5%

96.0% 1.1% 0.0% 104.9% 40.0Gy

92.4% 1.7% 0.1% 105.4% 39.8Gy

Table 2.4: Example of dosimetric parameters obtained for the structure PTVtoSurface for the Clinical and No Bolus Plans for one patient.

2.4.3 Plan comparison – Skin Structures (single patient example)

Figure 2.8 shows the dose volume histograms for the skin structures for the same

patient. The specific dose parameters as described in Table 2.3 were extracted from

the DVHs for the skin structure and each plan technique. Table 2.5 shows the doses

recorded in the case of the Clinical and No Bolus plan for the example case.

Page 59: Development of a technique using VMAT and robust

59

a)

b)

c)

Figure 2.8: Example DVHs for a) Skin1mm b) Skin3mm c) Skin5mm (Dashed line = Clinical Plan, Dotted line = No Bolus Plan)

Page 60: Development of a technique using VMAT and robust

60

Structure Dosimetric Parameter, Gy Clinical Plan No Bolus Plan

Skin1mm D99% Average D1%

34.1Gy (85.3%) 38.0Gy (95.0%) 40.7Gy (101.7%)

29.2Gy (73.0%) 36.2Gy (90.5%) 40.8Gy (102.0%)

Skin3mm D99% Average D1%

35.0Gy (87.5%) 38.7Gy (96.8%) 41.1Gy (102.8%)

30.9Gy (77.3%) 37.5Gy (93.8%) 41.5Gy (103.8%)

Skin5mm D99% Average D1%

35.4Gy (88.5%) 39.2Gy (98.0%) 41.5Gy (103.8%)

31.7Gy (79.3%) 38.3Gy (95.8%) 41.9Gy (104.8%)

Table 2.5: Example of dosimetric parameters obtained for the Skin Structures for the Clinical and No Bolus Plans. (Values is brackets represent the dose received as a percentage of the prescription dose, 40Gy).

The greatest dose difference is seen for the most superficial structure (Skin1mm),

gradually reducing for the 3mm and 5mm rinds. A difference in DVH curves is observed

for all skin structures in comparing the Clinical and No Bolus plan, with the largest

effect seen with the Skin1mm structure, the rind closest to the surface and the

smallest effect seen on the Skin5mm structure. The general trend seen in all the skin

structure DVHs is that in the No Bolus plan the volume receiving dose in the range of

30-40Gy is less than in the case of the Clinical plan. Furthermore, the maximum dose

the DVHs indicated is slightly greater using the No Bolus technique. The detail of this is

highlighted in the comparison of the dosimetric parameters in Table 2.5. The use of

bolus in the Clinical plan increases the minimum dose (D99%) to the Skin1mm,

Skin3mm and Skin5mm by 4.9Gy, 4.1Gy and 3.7Gy respectively. Similarly, the average

dose was also enhanced but to not the same extent, Skin1mm increased by 1.8Gy,

Skin3mm by 1.3Gy and Skin 5mm by 0.9Gy. In the case of the maximum dose (D1%)

the dose was lower in the Clinical cases however these differences were small, 0.1Gy

for the Skin1mm structure and 0.4Gy for the Skin3mm and Skin5mm structures.

2.4.4 Plan comparison – Organs at Risk (single patient example)

Figure 2.9 shows the DVH for the for the heart and lung OARs for the same patient

example, and Table 2.6 shows the extracted dosimetric parameters.

Page 61: Development of a technique using VMAT and robust

61

Figure 2.9: Example DVHs for Heart (red line) and Ipsilateral Lung (Orange line) (Dashed line = Clinical Plan, Dotted line = No Bolus Plan)

Structure Dosimetric Parameter

Mandatory Constraint

Optimal Constraint

Clinical Plan

No Bolus Plan

Heart

Ipsilateral Lung

V25% (10Gy) V5% (2Gy) V30%(12Gy)

≤5% ≤30% ≤17%

≤15%

0% 6.4% 13.6%

0% 6.5% 13.4%

Table 2.6: Example of dosimetric parameters obtained for the heart and ipsilateral lung OARS, for the Clinical and No Bolus Plans.

The DVH curves for heart and ipsilateral lung are almost identical in both the Clinical

and No Bolus cases. The DVH for the ipsilateral lung in the Clinical case suggests

slightly larger volumes receive dose in the range 3-15Gy but the difference is very

small. From the values reported for the dosimetric parameters shown in Table 2.6, this

is the case, with the ipsilateral lung volume that receives 12Gy being only 0.2% higher

than the Clinical plan. Table 2.6 also shows a 0.1% volume difference between the

Clinical plan and No Bolus plan for the heart parameter V5%. The shapes of the DVHs

and values reported for the dose parameters suggest that the use of bolus has very

little effect on the organs at risk.

Page 62: Development of a technique using VMAT and robust

62

2.4.5 Plan comparison – PTVtoSurface (8 patient study set)

Figure 2.10 shows each of the dosimetric parameters for the PTVtoSurface structure,

for the 8 patient cohort and for both the No Bolus and Clinical plans.

Similar trends are observed across all 8 patients. It can be seen that the volume of

PTVtoSurface receiving 95% of the prescription dose is greater in the case of the

Clinical plan compared to the No Bolus plan, on average the difference is 7.7%. The

average dose to the structure is slightly higher in the Clinical cases with the mean

difference being 0.4Gy. Over the 8 datasets the maximum dose to the PTVtoSurface

structure is on average slightly less in the Clinical case by a small value of 0.2Gy.

Similarly, the average volume of the structures receiving 105% and 107% of the

prescription dose, are slightly more in the No Bolus case, with differences of 0.6% and

0.1%, respectively. The effect observed, in using the bolus, is that it increases the

volume of the target structure receiving 95% of the prescription dose, it increases the

average dose to the target and reduces hotspots.

Page 63: Development of a technique using VMAT and robust

63

a)

b)

c)

d)

e)

Figure 2.10: Box and Whisker plots showing the a)V95% b) average dose c) V105% d) V107% and e) D1% parameters for the PTVtoSurface structure for No Bolus Plans (Blue) and Clinical Plans (Red) in 8 patients.

Page 64: Development of a technique using VMAT and robust

64

2.4.6 Plan comparison – Skin Structures (8 patient study set)

Figure 2.11 shows the results for the dosimetric parameters associated with the skin

structures for the No Bolus and Clinical plans

From the data presented in Figure 2.11a, 11b and 11c it can be seen that the minimum

dose to all three skin structures is greater in the case of the Clinical plan. The increase

on average, for the minimum dose parameter for Skin1mm, Skin3mm and Skin5mm is

5.3Gy, 4.6Gy and 4.2Gy, respectively.

A similar trend can be seen in the average dose to the skin structures (Figure

2.11d,11e, 11f). With the mean difference in the Clinical Plan being greater in all cases,

with a dose difference of 2.2Gy, 1.5Gy and 1.1Gy for Skin1mm, Skin3mm and

Skin5mm, respectively. In the case of the maximum dose to each of the skin structures

it can be seen in Figure 2.11i, 11g and 11j that there is a lot of overlap in the range of

values for each technique. This suggests that the different techniques do not have a

significant impact on this dose parameter. A similar trend for all three volumes is

observed with the addition of bolus increasing the minimum dose to the surface. The

effect is largest for the 1mm volume which is in the region of the greatest dose

gradient. Likewise, the average dose to the structures is increased with the use of

bolus, with the greatest impact observed for the 1mm skin structure. The maximum

dose in the skin structure does not appear to be affected by bolus, unlike the decrease

in hotspots that was observed for the PTVtoSurface structure. This indicates that the

hotspots lie within the main PTV rather than superficially.

Page 65: Development of a technique using VMAT and robust

65

a)

b)

c)

d)

e)

f)

g)

h)

i)

Figure 2.11: Box and Whisker plots showing the doses for a) Skin1mm – D99% b) Skin3mm – D99% c)skin5mm D99% d)Skin1mm – average e)Skin3mm – average f) Skin5mm – average g) Skin1mm – D1% h) Skin3mm – D1% i) Skin5mm – D1% for No Bolus Plans (Blue) and Clinical Plans (Red) for the 8 patients.

2.4.7 Plan comparison – Organs at Risk (8 patient study set)

The dosimetric parameters for the organs at risk are displayed in Figure 2.12 for the 8

patients. For the ipsilateral lung parameter, V30% (volume receiving 30% of the

prescription dose, 12Gy), the two techniques show a similar range in volumes and

Page 66: Development of a technique using VMAT and robust

66

median value that received that dose, suggesting that the difference between the

techniques is not significant.

a)

b)

c)

Figure 2.12: Dosimetric parameters for organs at risk a) Box and Whisker plot for Ipsilateral Lung V30% b) Bar Chart for Heart V5% c) Bar Chart for Heart V25% for No Bolus Plans (Blue) and Clinical Plans (Red) for the 8 patients.

In both techniques the optimal constraint (volume receiving 12Gy ≤15%) is achieved.

For the heart dosimetric parameters it should be noted that the patient cases cw2,

cw12 and cw18 are right sided treatments and therefore due to the position of the

Page 67: Development of a technique using VMAT and robust

67

heart the dose is significantly less. For the V5% parameter, volume of heart receiving

2Gy, the volume is on average 0.6% higher in the Clinical plan. In only 3 of the cases

the heart volume receives 10Gy (V25%) and the dose differences between the

techniques are ≤0.02%. As with the lung, the heart constraints (V5% ≤ 30% and V25% ≤

5%) are met in both treatment technique cases. The use of bolus has minimal impact

on the organs at risk.

2.5 Discussion

In this section the clinical technique which consists of a treatment plan where 8# are

delivered without any bolus and 7# are delivered with 1cm bolus, has been compared

to a treatment of 15# with no bolus used. This was done to establish the effect of

bolus in the clinical setting which we will attempt to emulate with the new treatment

technique. Dosimetric parameters for PTVtoSurface, Skin1mm, Skin3mm, Skin5mm

and the organs at risk, heart and lung, were used to analyse the effect.

For the target structure, PTVtoSurface, the results showed that the use of bolus in the

Clinical plan increased the volume that received 95% of the prescription dose (38Gy).

This was observed in the DVH of the example case, where the shoulder of the DVH

curve was shallower in the case of the No Bolus plan and the analysis of the 8 patients

which showed on average a 7.7% increase in volume receiving 38Gy when bolus was

used. With reference to the mandatory constraints for the target coverage (≥90%

volume should receive 95% prescription dose) the use of bolus meant that on average

this constraint was achieved (V95%=93.0%), where this was not the case with the No

Bolus plans (V95%=85.3%), Figure 2.10a. For the maximum dose (D1%), V105% and

V107% parameters the Clinical plans on average reported slightly lower values,

however both plans were within the required dose constraints. The difference in the

average dose to the PTVtoSurface structure was slightly higher in the Clinical cases

with a mean difference on 0.4Gy.

The use of bolus was shown to have an effect on the minimum and average dose

parameters for all the skin structures with minimal impact on the maximum dose. The

effect of bolus was seen most on the D99% and average parameters for the Skin1mm

Page 68: Development of a technique using VMAT and robust

68

structure, the most superficial, with the effect decreasing as the skin thickness

increased.

For the organs at risk, heart and lung, the comparison of the dosimetric parameters

between the Clinical and No Bolus plans showed that the two techniques were very

similar and, in both cases, met the defined constraints.

2.6 Summary

The use of bolus increases the PTVtoSurface volume receiving 95% of the

prescription dose.

The use of bolus reduces hotspots in the PTVtoSurface structure.

The use of bolus increases the minimum and average dose to the skin

structures, with most effect on 1mm. Bolus has little effect on the maximum

dose to these structures.

The use of bolus has minimal effect on dose to organs at risk.

Page 69: Development of a technique using VMAT and robust

69

Chapter 3

3 Comparison of VMAT plans to Clinical plans

This section aims to assess whether VMAT plans can be created that reproduce the

distributions of the Clinical plans without the use of bolus. The advantage of this would

be in reducing the number of treatment plans created for each patient, removing the

requirement to ensure bolus has been positioned before the treatment and reducing

any inaccuracies in dosimetry due to ill-fitting bolus.

3.1 Method

VMAT plans were created for the 8 patients analysed in the previous section. The

VMAT plans were optimised using RayStation with an Elekta VersaHD machine model

(5mm MLCs). The VMAT plans consisted of two 360⁰ arc deliveries (clockwise and anti-

clockwise), with a collimator of 10⁰ and using a beam energy of 6 MV FFF (flattening

filter free).

The plans were calculated using a 3mm dose grid with a collapsed-cone convolution

algorithm, the same conditions as for the Clinical plans (supplementary information is

provided in Appendix 1 showing impact of dose grid selection). Each plan was

normalised to 40Gy.

The same structures and dose constraints that were used to evaluate the Clinical and

No Bolus plans were used for this comparison with the addition of a structure,

Contralateral Breast. A constraint of mean contralateral breast dose <3.5Gy was used,

based on the Royal College of Radiologists recommendations for internal mammary

chain radiotherapy (RCR, 2016). This additional constraint was used due to the known,

dose bath effect when using VMAT and IMRT techniques.

The beam optimisation parameters in the treatment planning system were set as

follows; Gantry Spacing= 2⁰, Max Delivery Time = 200 seconds. Table 3.1 shows the

typical starting values used for the plan optimisation process for a left sided treatment.

Page 70: Development of a technique using VMAT and robust

70

Volume Objective/Function Starting weight

PTVtoSurface Max Dose 40.0 50 PTVtoSurface Min Dose 40.0 100

Heart Max DVH 2.0Gy to 27% volume

20

Left Lung Max Dose 38.3Gy 2 Left Lung Max DVH 5.0Gy to 15%

volume 15

Contralateral Breast Max Dose 4Gy 10 External Dose Fall-Off [H] 40.00Gy

[L] 5.0Gy, Low dose distance 3.0cm

100

External Max Dose 44.0Gy 50

Table 3.1: Typical starting values for planning objectives used for the VMAT plans.

3.2 Results

3.2.1 Plan comparison – Dose Distribution (single patient example)

Figure 3.1 shows the dose distribution for the No Bolus, Clinical and VMAT plan for the

same example patient discussed in Section 2. The difference between the No Bolus

and Clinical plans highlights visually the conclusions from the previous section, that

bolus reduces the hotspots within the treated volume, increases the dose to the

patient surface and has minimal impact on the dose to organs at risk plus normal

tissue. The main difference between the VMAT plan and both the Clinical and No Bolus

plans is the low dose distribution, where doses <20Gy, spill out of the treated area; as

typical of VMAT plans. Visually the distribution across the treated area also appears

more homogeneous for the VMAT technique. The plans are analysed quantitatively in

the following sections.

a) b)

Page 71: Development of a technique using VMAT and robust

71

c)

Figure 3.1: Example Dose Distribution for a) No Bolus plan b) Clinical plan c) VMAT plan

3.2.2 Plan comparison – PTVtoSurface (single patient example)

Figure 3.2 shows the DVH for the PTVtoSurface structure in the example patient case,

for each of the treatment techniques.

Dose values were extracted from the PTVtoSurface DVHs for the specified dose

parameters. Table 3.2 shows the values achieved for each treatment technique.

Page 72: Development of a technique using VMAT and robust

72

Figure 3.2: Example of DVH for structure PTVtoSurface. (Dashed line= Clinical Plan, Dotted line = No Bolus Plan and Solid line = VMAT plan)

Structure Dosimetric Parameter

Mandatory Constraint

Optimal Constraint

Clinical Plan

No Bolus Plan

VMAT Plan

PTVtoSurface V95% V105% V107% D1 % Average

≥ 90% ≤ 7% ≤ 2% ≤ 110%

≥ 95% ≤ 5%

96.0% 1.1% 0.0% 104.9% 40.0Gy

92.4% 1.7% 0.1% 105.4% 39.8Gy

96.6% 0.2% 0.0% 103.6% 40.0Gy

Table 3.2: Example of dosimetric parameters obtained for the structure PTVtoSurface for the Clinical, No Bolus and VMAT plans

In Figure 3.2 it is observed that the DVH for the VMAT plan displays a steeper gradient

centred on 40Gy prescription dose point than either the Clinical or No Bolus

techniques. This indicates that the VMAT technique is more homogeneous than the

other two over this target structure. The VMAT DVH falls slightly below the Clinical

plan between 35-38Gy, which suggests the minimum dose to the PTVtoSurface volume

is slightly higher in the Clinical plan. The improved homogeneity can also be observed

in the analysis of dose parameters in Table 3.2. The value for V95% is slightly higher in

Page 73: Development of a technique using VMAT and robust

73

the case of the VMAT plan compared to the Clinical plan, V105% slightly less and the

average dose remaining the same.

3.2.3 Plan comparison – Skin Structures (single patient example)

Figure 3.3 shows the DVHs for the skin structures in the same example patient. Specific

dose parameters were extracted from the DVHs for each skin structure and plan

technique. Table 3.3 reports the dose values obtained.

Structure Dosimetric

Parameter,

Gy

Clinical Plan No Bolus Plan VMAT Plan

Skin1mm D99%

Average

D1%

34.1Gy (85.3%)

38.0Gy (95.0%)

40.7Gy (101.7%)

29.2Gy (73.0%)

36.2Gy (90.5%)

40.8Gy (102.0%)

31.6Gy (79.0%)

37.9Gy (94.8%)

41.2Gy (103.0%)

Skin3mm D99%

Average

D1%

35.0Gy (87.5%)

38.7Gy (96.8%)

41.1Gy (102.8%)

30.9Gy (77.3%)

37.5Gy (93.8%)

41.5Gy (103.8%)

32.9Gy (82.3%)

39.1Gy (97.8%)

41.5Gy (103.8%)

Skin5mm D99%

Average

D1%

35.4Gy (88.5%)

39.2Gy (98.0%)

41.5Gy (103.8%)

31.7Gy (79.3%)

38.3Gy (95.8%)

41.9Gy (104.8%)

33.6Gy (84.0%)

39.6Gy (99.0%)

41.6Gy (104.0%)

Table 3.3: Example of dosimetric parameters obtained for the skin structures for the Clinical, No Bolus and VMAT plans (Values in brackets represent the dose received as a percentage of the prescription dose, 40Gy)

The DVHs in Figure 3.3 and the data in Table 3.3, show that for the example case, the

average dose for the Skin3mm and Skin5mm structures, is greater with the VMAT

technique compared to the Clinical plan and for the Skin1mm structure the average

dose is slightly lower for the VMAT technique compared to the one used clinically. For

all the skin structures the maximum dose is slightly higher in the VMAT cases than the

clinical situation, 0.5Gy, 0.4Gy and 0.1Gy for Skin1mm, Skin 3mm and Skin5mm,

respectively. The minimum dose to the skin structures is also lower with the VMAT

technique. For the Skin1mm volume, the minimum dose is 2.5Gy less with the VMAT

technique compared with the Clinical, and 2.1Gy and 1.8Gy less, for the Skin3mm and

Skin 5mm volumes.

Page 74: Development of a technique using VMAT and robust

74

Figure 3.3: Example DVHs for a) Skin1mm b)Skin3mm c)Skin5mm (Dashed line=Clinical plan, Dotted line = No Bolus plan, Solid line=VMAT plan)

a)

b)

c)

Page 75: Development of a technique using VMAT and robust

75

Comparing the VMAT technique to the Clinical plan there is a greater difference

between the minimum and maximum doses in each of the skin structures, indicating

the dose is not as homogenous, in this superficial area. Although compared to the

Clinical plans the minimum dose is not as great with the VMAT plans, the VMAT

technique still enhances the minimum dose compared to that achieved in the No Bolus

situation. Compared with the No Bolus plan the minimum dose to Skin1mm, Skin3mm

and Skin 5mm is increased by 2.4Gy, 2.0Gy and 1.9Gy, respectively. As was seen with

the comparison of the Clinical plan to the No Bolus plan, the VMAT technique has most

impact on the most superficial structure, Skin1mm.

3.2.4 Plan comparison – Organs at Risk (single patient example)

Figure 3.4 shows the DVH for the heart and lung OARs and the structure, Contralateral

Breast, for the example patient. Table 3.4 shows the extracted dosimetric parameters.

Figure 3.4: Example DVHS for Heart (red line), Ipsilateral Lung (Orange line) and Contralateral Breast (Blue line). (Dashed line = Clinical plan, Dotted line = No Bolus Plan and Solid line=VMAT plan)

Structure Dosimetric Mandatory Optimal Clinical No VMAT

Page 76: Development of a technique using VMAT and robust

76

Parameter Constraint Constraint Plan Bolus Plan

Plan

Heart

Ipsilateral Lung

Contralateral

Breast

V25% (10Gy) V5% (2Gy) V30%(12Gy) Mean Dose

≤5% ≤30% ≤17%

≤15% <3.5Gy

0% 6.4% 13.6% 0.3Gy

0% 6.5% 13.4% 0.2Gy

0% 29.9% 7.1% 2.0Gy

Table 3.4: Example of dosimetric parameters obtained for the heart, lung and contralateral breast, for the Clinical, No Bolus and VMAT plans.

It has been previously discussed that the use of bolus had minimal impact on the

organs at risk, heart and ipsilateral lung, and from Figure 3.4 and Table 3.4 we can also

observe, as expected, that it has minimal impact on the dose to the contralateral

breast. In Figure 3.4 we can observe that for the VMAT plan there is significant shift in

the DVH curve for the heart structure, towards the higher doses, and the value

reported for the V5% shows that for the VMAT plan this heart dose is just within the

mandatory constraint. For the lung structure it can be seen in the DVH, that when

using the VMAT technique more of the lung volume receives doses less than 5Gy

compared to the Clinical and No Bolus techniques however the volume of lung

receiving doses above 5Gy is greater for the Clinical or No Bolus technique. This is

supported by the V30% parameter for the lung volume, where only 7.1% of the lung

receives 12Gy using the VMAT technique compared to 13.6% for the Clinical plan. All

techniques however do produce dose distributions that meet the required constraints

for heart and lung. It can also be observed that the dose to the contralateral breast is

significantly higher using the VMAT technique, however this is still less than the

recommended constraint of <3.5Gy. The increased lung volume receiving lower doses

and the contralateral breast receiving higher dose are the effects of the low dose bath

expected with VMAT.

Page 77: Development of a technique using VMAT and robust

77

3.2.5 Plan comparison – PTVtoSurface (8 patient study set)

Figure 3.5 shows the dosimetric parameters for the PTVtoSurface structure for the 8

patient cohort, for each of the treatment techniques, with similar trends observed

across all the patients.

a)

b)

c)

d)

e)

Figure 3.5: Box and Whisker plots showing the a) V95% b) average dose c) V105% d)107% and e) D1% parameters for the PTVtoSurface structures for No Bolus Plans (Blue), Clinical Plans (Red) and VMAT Plans (Green) in the 8 patients.

Page 78: Development of a technique using VMAT and robust

78

From Figure 3.5a it can be observed that the use of the VMAT technique can increase

the volume of 38Gy (95% of the prescription dose) that the target structure receives.

The average value for the V95% objective increases from 93%, for the Clinical plan to

95.6% for the VMAT plan, resulting in the VMAT plans, on average, meeting the

optimal constraint for target coverage (V95% >95%). For this set of patients, it can also

be seen that in the VMAT plans the mandatory constraint is consistently achieved

(V95%>90%) compared with the Clinical case and the variation of this value is reduced

between patients. The average dose to the target structure is also slightly increased

with the VMAT technique compared to the Clinical case with the mean dose to the

PTVtoSurface structure increasing by 0.6Gy. For the parameters reflecting the hotter

dose distribution (areas receiving a dose greater than the prescription dose) within the

structure, D1%, V105% and V107%, it is observed that the VMAT technique results in

plans that produce slightly higher hotspots than the Clinical Plan but lower than the No

Bolus plan. However, all the volumes still meet the mandatory requirements. The

VMAT technique can produce plans with better V95% coverage than the Clinical plans,

with slightly larger hotspots, but which are clinically acceptable

3.2.6 Plan comparison – Skin Structures (8 patient study set)

In Figure 3.6 the results for the dosimetric parameters obtained for the skin structures

are summarised for the 8 patients, for each technique.

From Figure 3.6 a similar trend is seen for all the skin structures, for each of the

parameters used to compare the three techniques. It can be observed in Figure 3.6a,

Figure 3.6b and Figure 3.6c that the VMAT technique does not have the same impact

as using bolus on the minimum dose to the skin structures. Where the bolus increased

the minimum dose on average by 5.3Gy, 4.6Gy and 4.2Gy to the Skin1mm, Skin3mm

and Skin5mm, the VMAT technique increased the dose by 3.5Gy, 3.3Gy and3.3Gy,

respectively, compared with the No Bolus plan. As with the Clinical plan the VMAT

technique had the most impact on the most superficial structure, Skin1mm.

Page 79: Development of a technique using VMAT and robust

79

a)

b)

c)

d)

e)

f)

g)

h)

i)

Figure 3.6: Box and Whisker plots showing the doses for a) Skin1mm – D99% b) Skin3mm – D99% c)skin5mm D99% d)Skin1mm – average e)Skin3mm – average f) Skin5mm – average g) Skin1mm – D1% h) Skin3mm – D1% i) Skin5mm – D1% for No Bolus plans (Blue), Clinical plans (Red) and VMAT plans (Green) for the 8 patients.

With reference to the average dose parameter the VMAT technique enhances the dose

to all the skin structures more than with the use of bolus. With the VMAT technique

Page 80: Development of a technique using VMAT and robust

80

the Skin 1mm, Skin3mm and Skin 5mm receive an additional 0.7Gy, 0.9Gy and 0.8Gy,

respectively compared to the Clinical plan.

In the case of the maximum dose to the skin structures the VMAT technique results in

slightly higher hotspots than the other two techniques, particularly in the most

superficial skin layer, Skin1mm. The increase in hotpots using the VMAT technique was

also observed within the PTVtoSurface structure, however not to the same extent,

suggesting that the hotspots are within the superficial area. It should be noted that the

maximum doses are still within the accepted tolerance of <110% (44Gy).

Although the VMAT technique can increase the V95% for the PTVtoSurface structure

compared to the Clinical plan, the minimum dose to the skin structures was not

increased as much as when bolus is used. The average dose to the skin structures can

be increased with the VMAT technique, as is the maximum dose, however the

hotspots are still clinically acceptable.

3.2.7 Plan comparison – Organs at Risk (8 patient study set)

The dosimetric parameters for heart, ipsilateral lung and contralateral breast for each

of the treatment techniques are summarised for the 8 patients in Figure 3.7.

In Figure 3.7a it can be seen that on average the VMAT technique results in a smaller

volume of lung receiving 30% of the prescription dose (12Gy) than the Clinical and No

Bolus plans. On average the volume difference between the Clinical and VMAT plan is

2.8%. For this set of patients, the mean ipsilateral lung volume is 1392cm3 so with the

VMAT technique the volume receiving 12Gy is on average 38.9 cm3 less compared to

the Clinical treatment.

Page 81: Development of a technique using VMAT and robust

81

a)

b)

c)

Page 82: Development of a technique using VMAT and robust

82

d)

Figure 3.7: Dosimetric parameters for organs at risk a) Box and Whisker plot for Ipsilateral Lung V30% b) Bar Chart for Heart V5% (mandatory constraint shown in dashed line) c) Bar Chart for Heart V25% d) Box and Whisker plot for Contralateral Breast for No Bolus Plans (Blue), Clinical Plans (Red) and VMAT Plans (Green) for the 8 patients.

Figure 3.7b shows that, for the heart dosimetric parameter V5% (volume of the heart

receiving 5% of the prescription dose) all the plans are within the mandatory

constraint. However, it is clear that the VMAT technique results in larger volumes

receiving a dose of 2Gy. One reason for the higher doses in the VMAT plans is that

although the heart was used in the optimisation process for the plans, if the heart

constraint was met the plan was deemed acceptable and no further optimisation took

place. Therefore, there may be scope for further reducing these doses further. In

particular, for the three right-sided cases cw2, cw12 and cw18, the 2Gy dose received

by the heart is a likely result of the low dose bath produced by the VMAT technique

with beams entry from multiple directions, however there may be scope to reducing

this too, if additional constraints are added to the optimisation process.

Three of the eight plans showed dose in the heart over 10Gy, with the VMAT

technique resulting in a smaller volume receiving this dose. The difference in volume

compared to the Clinical technique however is relatively small, on average 0.2%, with

Page 83: Development of a technique using VMAT and robust

83

the average heart volume for these three cases being 779cm3, this is to an average

volume difference of 1.6cm3.

The effect of the VMAT dose bath is also seen for the contralateral breast structure,

with the average mean dose in the case of the VMAT technique being 1.5Gy higher.

The VMAT technique still however meets the recommended requirement than the

mean dose <3.5Gy.

Unlike the use of bolus, the use of the VMAT technique does impact the organs at risk.

However the mandatory constraints for the organs at risk can still be achieved.

3.3 Discussion

In this chapter the VMAT technique using two 360⁰ arcs was analysed to establish

whether similar dose distributions to the Clinical technique, using bolus, could be

produced. Results from the No Bolus plans were also included to compare the impact

of the two skin enhancing techniques.

For the target structure PTVtoSurface the results showed that the VMAT plan could

increase the volume that received the prescription dose more than with bolus. On

average the VMAT plans met the optimal constraint (V95%>95%) for this target

structure whereas the Clinical technique on average only met the mandatory

constraint (V95%>90%). The average dose to the PTVtoSurface structure was also

0.6Gy greater using the VMAT technique compared to the Clinical one. The maximum

dose, V105% and V107% were on average greater for the VMAT technique compared

to the Clinical plans, but still within the mandatory constraints.

The DVH for the PTVtoSurface structure in the example patient suggested that the

minimum dose was slightly lower using the VMAT technique than the Clinical case. This

was also observed in the DVH for the skin structures and seen in the average data

across the 8 patients for the minimum dose parameter. However, the results for the

average and maximum dose parameters for the skin structures showed that the VMAT

technique could enhance the dose more than the Clinical plans, compared to using no

bolus. Although the maximum dose to the skin structures was higher, using the VMAT

technique, the dose was still within the specified constraint.

Page 84: Development of a technique using VMAT and robust

84

However, unlike with the Clinical plan where the use of bolus had minimal impact on

the dose to the heart, lung and contralateral breast, the dose distribution from the

VMAT plans did affect these parameters. The VMAT technique had a positive effect on

the ipsilateral lung volume parameter V30%, where on average this volume was

smaller (2.8%), and there also a slight decrease in heart volumes receiving 10Gy (0.2%),

for the three cases where this parameter was reported. In contrast, a larger volume of

the heart received 2Gy and the mean dose to the contralateral breast on average

increased by 1.5Gy, this was expected and due to the low dose bath as consequence of

using the VMAT technique However the recommended constraints were met in all the

patients and further optimisation in the treatment planning software could have

potentially reduced these values.

3.4 Summary

VMAT plans can be produced that meet the required dose objectives for target

structure.

With the VMAT technique the volume of the target structure receiving 95% of

the prescription dose can be greater than in the Clinical case using bolus.

The mean dose to the target volume is increased using the VMAT technique

compared to the Clinical case.

The VMAT technique enhances the minimum dose to the patient surface but

not to the same extent as the Clinical plans, however average and maximum

doses to the skin structures are greater in the VMAT plans.

A consequence of using VMAT plans is a low dose bath to tissues outside the

target however dose constraints to organs at risk can still be achieved.

Page 85: Development of a technique using VMAT and robust

85

Chapter 4

4 Surface Dose Measurements

Chapter 2 quantified the effect of using bolus as part of the department’s current

technique of opposed tangential fields and in Chapter 3 it has been shown that an

alternative technique using VMAT can produce similar dose distributions, including

skin doses. The impact of dose and comparison of techniques have so far been limited

to evaluating the distributions calculated in the treatment planning system.

The limitations of treatment planning systems to accurately calculate dose at the

surface of air-tissue interfaces has been discussed in the introduction and included the

ability of the treatment planning algorithm to account for electron contamination that

originates in the treatment head and the effect of obliquity of incident beams.

In this chapter superficial doses have been measured to investigate the accuracy of the

dose calculation. The aim to determine whether the calculations for both the Clinical

and VMAT techniques are reliable and match physical measurements.

Two phases of measurements had originally been planned, measurements on an

anthropomorphic phantom followed by measurements on patients. Measurement of

superficial doses using the current technique would have taken place with the consent

of the patients, as part of a departmental audit. However, this aspect of the

investigation was unable to be completed due to the Covid-19 pandemic, which limited

unnecessary access to patients, treatment rooms and also brought about a change in

dose and fractionation for these patients. The measurements presented were carried

out on the anthropomorphic phantom as part of developing a method to be used in

patient cases. These were made just prior to Covid-19 restrictions being put into place

and therefore some repeats of experiments were unable to be completed.

Page 86: Development of a technique using VMAT and robust

86

4.1 Method

Three treatment plans were generated in RayStation using the CIRS anthropomorphic

thorax phantom (CIRS Inc., Norfolk, VA, USA). The CIRS phantom was CT scanned on a

Philips Brilliance Big Bore CT Scanner (Philips Medical Systems, Eindhoven,

Netherlands) at 3mm slice thickness. The 3 plans created were:

1. A tangential plan with no bolus

2. A tangential plan with 1cm bolus (virtually applied in the treatment planning

system)

3. A VMAT partial arc plan

The bolus and no bolus plans consisted of two opposed open fields with no segments

using a 6MV cFF (conventional flattening filter) for delivery on an Elekta VersaHD linac.

The VMAT plan was created using a 6MV FFF beam for delivery on the same machine.

All plans were prescribed to deliver 40Gy in 15#, calculated with the on a dose grid of

0.3cm. The dose distributions for the three plans are shown in Figure 4.1.

Dose measurements were performed using lithium fluoride thermoluminescent

dosimeters (TLDs), TLD-100H (ThermoFisher Scientific, Waltham, MA, USA) with a

dimension of 3mmx3mm and 0.8mm thick. The TLDs were readout using a Harshaw

3500 TLD reader (ThermoFisher Scientific, Waltham, MA, USA). Absolute dose was

obtained by irradiating a small batch of TLDs under standard calibration conditions

(6MV, 10x10 cm field, 90SSD, 10cm depth) with reference to the same measurement

using a Farmer chamber (NE2571), in accordance to our standard departmental

protocol. The TLD measurements were also corrected for dose response by irradiating

at different MU (50MU, 100MU, 200MU) and individual TLD calibration factors.

Page 87: Development of a technique using VMAT and robust

87

a)

b)

c)

Figure 4.1: Dose distribution for a) No Bolus b) Bolus and c) VMAT partial arc plan on the CIRS anthropomorphic thorax phantom

Page 88: Development of a technique using VMAT and robust

88

Five reference points, approximately 5cm apart, were created on the central slice of

the phantom, from medial to lateral, representing the positions that TLD

measurements would be made. The points were placed on the external contour, as

created by the treatment planning system (Figure 4.2).

Figure 4.2: Points showing the TLD position on central axis of CT Scan of CIRS phantom. Each TLD point is positioned to intersect with the body contour (green),

Treatment plans were delivered on the Elekta VersaHDs, using the treatment machine

laser system to align and set-up the CIRS phantom for irradiation at the planned

isocentre position. Two TLDS were placed at each measurement position, for each plan

delivered, one TLD either side of the central axis in the sup/inf direction (Figure 4.3).

For the bolus plans a slab of 1cm thick of tissue equivalent material is placed over the

whole phantom prior to plan delivery (Figure 4.4).

Page 89: Development of a technique using VMAT and robust

89

a)

b)

Figure 4.3: a) anthropomorphic phantom position for treatment delivery b) packets position with TLD either side of central axis

Figure 4.4: 1cm water equivalent material placed over phantom prior to delivery of bolus plan.

Page 90: Development of a technique using VMAT and robust

90

To establish the consistency of the TLD measurements the aim was to deliver each

treatment plan 5 times, however due to availability of treatment machines during the

Covid-19 pandemic, the no bolus treatment plan was repeated 4 times, the bolus plan

3 times and the VMAT plan was delivered once.

4.2 Results

Figure 4.5 shows the TLD dose measurements for the no bolus plan compared with the

predicted treatment planning system for one fraction. The TLD measurements at each

point, except position 3, are within 5% of each other. The variation in measurement

dose at position 3 is 6.4%. Table 4.1 summarises the average TLD measurement at

each point as a percentage of the prescribed dose compared with the treatment plan.

Figure 4.5 and Table 4.1 show that the TLD measurements are consistently lower than

the calculated doses but that the size of the difference is dependent on position. On

average the TLD measurement is 8.2% lower than planning system predicted dose.

Figure 4.6 shows the TLD measurements for the bolus plan compared with the

calculated doses in the treatment planning system. At positions 1,2, 5 and 3 (if outlier

value is ignored) the TLD measurements are within 5% of each other. At position 4 the

variation in measurement is 6.9%. From Table 4.2 it is observed that the average TLD

measurement at each position, excluding the outlier at position 3, are a higher dose

than predicted. On average the dose difference between measured doses is 2.0%

higher than predicted.

Figure 4.7 shows the measurements for the VMAT plan compared with the calculated

doses, only one set of measurements was performed. From Figure 4.7 and Table 4.3 it

can be seen that at all positions the TLDs measured a lower dose than was predicted

by the planning system. The dose difference is on average 6.4% lower with the

greatest dose difference seen at position 1, 8.0%.

Page 91: Development of a technique using VMAT and robust

91

Figure 4.5: TLD results for No Bolus plan compared to treatment plan dose - 1 fraction delivery. (Plan delivered 4 times)

Position Dose Measured, % of prescribed dose (absolute dose, Gy)

Dose Predicted, % of prescribed dose (absolute dose, Gy)

Dose Difference (%)

1 61.0 (1.63) 66.8 (1.78) -8.7 2 66.3 (1.77) 75.2 (2.0) -11.7 3 82.8 (2.21) 83.3 (2.22) -0.6 4 64.6 (1.72) 73.1 (1.95) -11.6 5 56.7 (1.51) 62.8 (1.67) -9.8

Average Difference -8.5

Table 4.1: Results of TLD measurements for No Bolus plan as percentage of prescribed dose compared to planned dose, absolute dose in brackets.

Page 92: Development of a technique using VMAT and robust

92

Figure 4.6: TLD results for Bolus plan compared to treatment plan dose - 1 fraction delivery (Plan delivered 3 times)

Position Dose Measured, % of prescribed dose (absolute dose, Gy)

Dose Predicted, % of prescribed dose (absolute dose, Gy)

Dose Difference (%)

1 104.6 (2.79) 101.9 (2.72) 2.7 2 101.9 (2.72) 100.6 (2.68) 1.3 3 101.0 (2.83) 104.1 (2.77) 2.2 4 104.0 (2.77) 102.1 (2.72) 1.9 5 101.3 (2.70) 100.0 (2.67) 1.3

Average Difference 1.9

Table 4.2: Results of TLD measurement for Bolus plan as percentage of prescribed dose compared to planned dose absolute dose in brackets. (Measurement 2 at TLD position 3 disregarded in these results)

Page 93: Development of a technique using VMAT and robust

93

Figure 4.7: TLD results for VMAT plan compared to treatment plan dose - 1 fraction delivery. (Plan delivered once)

Position Dose Measured, % of prescribed dose (absolute dose, Gy)

Dose Predicted, % of prescribed dose (absolute dose, Gy)

Dose Difference (%)

1 66.8 (1.78) 72.8 (1.94) -8.2 2 72.6 (1.94) 78.9 (2.10) -8.0 3 72.3 (1.93) 76.2 (2.03) -5.1 4 60.3 (1.61) 62.7 (1.67) -3.7 5 44.0 (1.17) 46.9 (1.25) -6.1

Average Difference -6.2

Table 4.3: Results of TLD measurements for VMAT plan as percentage of prescribed dose compared to planned dose absolute dose in brackets.

Page 94: Development of a technique using VMAT and robust

94

4.3 Discussion

The results, with the no bolus case using tangents, showed that the measured surface

dose ranged from 56.7% - 82.8% of the prescription dose. The highest dose was at the

apex of the phantom, position 3, where the incident beams are at their shallowest.

This variation in dose distribution and range of doses reflects those measured by

Quach et al. (2000) where TLDs were used on a hemicylindrical phantom and surface

doses from 49%-75% were obtained from the steepest to shallowest beam incidence.

Manger et al. (2016) also reported superficial dose distributions of the same

magnitude and profile, 40%-70% of the prescription dose, using Gafchromic EBT3 film,

with a no-bolus plan using 6MV tangential fields. Both of these groups also showed

that with the use of 1cm of tissue equivalent bolus the surface-dose profile become

shallower, with Manger et al. (2016) reporting Gafchromic film measurements

increasing to 85%-109% of the prescription dose. Although similar results are not

available to compare the VMAT plans directly the implication that beam incidence

angle influences the superficial dose provides some explanation for the dose variation

in the plan investigated in this chapter.

With the plans calculated on a dose grid of 3mm, the TLD measurements in the bolus

plan were consistent with the RayStation planning system to within 3%, however the

no bolus and VMAT plans, showed lower measurement doses compared to the

planning system of 8.5% and 6.2% respectively. This discrepancy is not unexpected as

the TLD position defined in the treatment planning system for the VMAT and No Bolus

plans lie at the intersection between tissue and air, where the treatment planning

system accuracy is most limited as discussed previously. However, the results

presented here are consistent with a study conducted by Chung et al. (2005) where it

was shown that the treatment planning systems, Pinnacle3 and Corvus overestimated

surface dose by ~15% and ~18% respectively. Similarly, Cao et al. (2017) suggested that

the collapsed cone algorithm used in RayStation performed well in the calculation of

superficial dose despite overestimation of skin dose at the reference depth of 70µm of

14.11%. Other sources of inaccuracy include the dose grid that the plans were

calculated at, 0.3cm, and calculation at a smaller dose grid could improve the dose

difference.

Page 95: Development of a technique using VMAT and robust

95

There are limitations to the accuracy of this data due to the inability to repeat some of

the measurements, particularly for the VMAT cases as access to the treatment

machines was limited due to the Covid restrictions. However, the results presented are

consistent with published data and suggest that TLDs are sensitive enough to measure

the superficial doses for comparison of the treatment techniques in addition to

comparing technique through treatment planning systems. In reporting dose to skin, it

has been previously been discussed that there is lack of consensus as to how skin is

defined. Therefore TLDs with a thickness of 0.8mm may be suitable for reporting skin

dose if this thickness is considered appropriate for skin, however for skin doses at the

ICRU recommended depth of 0.07mm, alternative dosimeters would be required.

4.4 Summary

TLD measurements on an anthropomorphic phantom were shown to be within

3% of the dose predicted by the treatment planning system when

measurement position at 1cm depth (under bolus).

Surface dose measurements with TLDs demonstrated that the treatment

planning system over-estimated the dose by 8.5% for the no bolus, tangential

plans and 6.2% for VMAT plans. This over-estimation is expected due to the

challenges in accurately modelling the build-up effect in the treatment planning

system.

Page 96: Development of a technique using VMAT and robust

96

Chapter 5

5 Effect of Perturbation

In the previous chapters it was shown that VMAT plans could be optimised that

produced similar dose distributions to the Clinical plans. The VMAT plans met the

required dose objectives for the target volume and with the exception of the D99%

value, dose parameters were comparable for the skin structures. Constraints were also

met for the heart, lung and contralateral breast.

To achieve the increase in dose to the surface region, without using bolus, IMRT

techniques need to overcome the skin-sparing effect by adding more dose into the

build-up region. This is achieved using small, narrow segments, targeting the surface

from a glancing angle. However, patient motion may result in this dose being delivered

outside the build-up region generating hotspots in the patient tissue. VMAT plans may

therefore be more susceptible to movement than the current clinical technique.

This section therefore attempts to determine the impact of changes in patient set-up

by assessing the effect on dose distribution for the Clinical and VMAT plans.

5.1 Method

To simulate the effect of setup uncertainties and patient changes the Clinical and

VMAT plans created for the cohort of 8 patients were recalculated on perturbed CT

datasets. This was performed using the ‘compute perturbed dose’ in the RayStation

software, which recalculates the original plan on the CT dataset with the isocentre

shifted by a defined value, no further optimisation of the plan is made. It is

acknowledged that this does not represent the random set-up error in the clinical

situation but instead assumes the same patient shift for all 15 fractions of treatment.

In addition, the perturbation shifts were applied to a rigid body, so non-uniform

patient changes were not taken into account. In the case of the Clinical plans, the

Page 97: Development of a technique using VMAT and robust

97

perturbation was applied to the individual no bolus and bolus parts of the plan, prior

to summation. The results in this study should be considered a worse-case scenario for

the patient treatment and should give an indication of how the dose distribution

changes with patient set-up in different directions.

Perturbed dose calculations were carried out on both the Clinical and VMAT plans for

the 8 patients. Perturbation values of 0.5cm and 0.3cm were used to establish whether

there was an acceptable limit of set-up error. The perturbation was applied in each of

the left-right, superior-inferior and anterior-posterior directions, moving the isocentre

towards and away from the patient surface, and in the superior and inferior direction.

Figure 5.1 summarises the shifts applied to the isocentres.

DVHs were obtained for each perturbed plan and the dosimetric metrics for

PTVtoSurface, skin structures and organs at risks as in previous chapters, were used for

the quantitative analysis.

Direction of Perturbation

Left Sided Treatment Shift Direction

Left Sided Perturbation Values

Right Sided Treatment Shift Direction

Right Sided Perturbation values

Towards patient surface

and superior (TS)

Left, Sup, Ant 0.5, 0.5, 0.5 0.3, 0.3, 0.3

Right, Sup, Ant -0.5, 0.5, 0.5 -0.3, 0.3, 0.3

Away from patient surface

and superior (AS)

Right, Sup, Post -0.5, 0.5, -0.5 -0.3, 0.3, -0.3

Left, Sup, Post 0.5, 0.5, -0.5 0.3, 0.3, -0.3

Towards patient surface

and inferior (TI)

Left, Inf, Ant 0.5, -0.5, 0.5 0.3, -0.3, 0.3

Right, Inf, Ant -0.5, -0.5, 0.5 -0.3, -0.3, 0.3

Away from patient surface

and inferior (AI)

Right, Inf, Post -0.5, -0.5, -0.5 -0.3, -0.3, -0.3

Left, Inf, Post 0.5, -0.5, -0.5 0.3, -0.3, -0.3

Table 5.1: Summary of shifts for perturbed plans. Abbreviation for perturbation direction are included.

Page 98: Development of a technique using VMAT and robust

98

5.2 Results

5.2.1 Perturbation Effect – Dose Distribution (single patient example)

In Figure 5.1 the impact of perturbation on the dose distribution for the Clinical and

VMAT plans perturbed by 0.5cm in the TS and AS directions (TS and AS have been

shown as they demonstrated the worse-case scenario perturbed calculations).

Figure 5.1b and Figure 5.1e show the impact of perturbation towards the patient

surface with a shift of 0.5cm. The posterior edge of the treatment volume loses

coverage of the 95% isodose line in both the plans. In the Clinical plan there is an

increased dose to the superficial region of the target volume and the opposite is

observed for the VMAT plan.

Figure 5.1c and Figure 5.1f show the impact of perturbation with a shift 0.5cm away

from the patient surface. As the isocentre is moved closer to the lung the 95% isodose

spills over the posterior edge of the target volume in both cases. In the Clinical plan a

lower dose is observed in the superficial region of the target and an increase is dose is

seen in the case of the VMAT plan.

Page 99: Development of a technique using VMAT and robust

99

a)

b)

c)

d)

e)

f)

Figure 5.1: Example of perturbed plans a) Clinical plan -non perturbed b) Clinical plan – perturbed 0.5cm TS direction c) Clinical plan – perturbed 0.5cm AS direction d) VMAT plan – non perturbed e) VMAT plan – perturbed 0.5cm TS direction f) VMAT plan – perturbed 0.5cm AS direction

Page 100: Development of a technique using VMAT and robust

100

5.2.2 Perturbation Effect – PTVtoSurface (single patient example)

The DVHs in Figure 5.2a and Figure 5.2c show the effect of a 0.5cm shift towards the

patient surface for the Clinical and VMAT plans. The coverage of the PTVtoSurface

reduces, as was observed with the dose distributions. It is clear however, from the

DVHs that the impact is greater for the VMAT case.

The DVHs in Figure 5.2b and Figure 5.2d shows the perturbation effect with a shift

away from the patient surface. For the Clinical plan the shift appears to reduce the

dose over most of the volume and slightly increase the maximum dose. For the VMAT

plan a slightly higher minimum dose is seen, however the maximum dose to the

PTVtoSurface is notably greater.

Specific dose parameters were extracted from the DVHs for each of the perturbed

plans. The values achieved, along with the mandatory and optimal constraints are

shown in Table 5.2 and Table 5.3, the values not meeting the mandatory constraints

are highlighted.

Page 101: Development of a technique using VMAT and robust

101

a)

b)

c)

d)

Figure 5.2: Example of DVHs for PTVtoSurface a)Clinical plan non-perturbed v Clinical plan perturbed 0.5 in TS direction b) Clinical plan non-perturbed v Clinical plan perturbed 0.5cm in AS direction c)VMAT plan non-perturbed v VMAT plan perturbed 0.5cm in TS direction d) VMAT plan non-perturbed v VMAT plan perturbed 0.5cm in AS direction. (Dotted line = original plan, dashed line = perturbed plan)

Page 102: Development of a technique using VMAT and robust

102

Structure Dosimetric Parameter

Mandatory Constraint

Optimal Constraint

Non – Perturbed (Clinical)

0.5cm TS Shift

0.5cm AS Shift

PTVtoSurface V95% V105% V107% D1 % Average

≥ 90% ≤ 7% ≤ 2% ≤ 110%

≥ 95% ≤ 5%

96.0% 1.1% 0.0% 104.9% 40.0Gy

81.1% 4.8% 0.2% 106.0% 38.5Gy

90.5% 3.4% 0.9% 106.7% 39.5Gy

Table 5.2: Example of dosimetric parameters for the non-perturbed Clinical plan and with a perturbation of 0.5cm in the TS and AS directions. Values underlined show where constraints were not met.

Structure Dosimetric Parameter

Mandatory Constraint

Optimal Constraint

Non – Perturbed (VMAT)

0.5cm TS Shift

0.5cm AS Shift

PTVtoSurface V95% V105% V107% D1 % Average

≥ 90% ≤ 7% ≤ 2% ≤ 110%

≥ 95% ≤ 5%

96.6% 0.2% 0.0% 103.6% 40.0Gy

74% 1.3% 0.1% 105.3% 37.7Gy

98% 22.3% 16.8% 119% 40.8Gy

Table 5.3: Example of dosimetric parameters for the non-perturbed VMAT plan and with a perturbation of 0.5cm in the TS and AS directions. Values underlined show where constraints were not met

Table 5.2 and Table 5.3 show quantitatively the impact of patient movement on the

dose distribution. A perturbation in the 0.5cm TS direction results in failure to meet

the V95% mandatory constraint in both the Clinical and VMAT cases, with a greater

reduction in volume for the VMAT case. Similarly, average dose decreases for both

plans and both plans get hotter. For shifts in the 0.5cm AS direction, the V95% for the

Clinical plan again decreases compared to its non-perturbed plan, however for the

VMAT plan the value increases. Following the same trend as V95% the average dose

decreases in the perturbed Clinical plan and increases for the VMAT plan. In both cases

a shift away from the surface increases the hotspots, and in the case of the VMAT plan

the constraints are significantly exceeded.

For perturbation towards the surface with both the Clinical and VMAT techniques a

reduction in coverage of the PTVtoSurface is seen. For shifts away from the surface a

Page 103: Development of a technique using VMAT and robust

103

slight reduction is observed for the Clinical plans and a significant increase in hotspots

is seen for the VMAT plan.

5.2.3 Perturbation Effect – Skin Structures (single patient example)

The DVHs shown in Figure 5.3a and Figure 5.3c demonstrate the impact of shifts 0.5cm

towards the patient surface for the Clinical and VMAT plans on the Skin3mm structure.

For the Clinical plan on average the dose over the structure increases though the

minimum dose decreases. In contrast for the VMAT plan a significant decrease in the

dose is observed over the whole skin structure, as was visualised in the dose

distribution (Figure 5.1e).

For perturbation in the AS direction Figure 5.3b shows that for the Clinical plan a

decrease in dose is observed for the skin structure and for the VMAT plan the opposite

is seen(Figure 5.3d), with doses increasing to more than 125% of the prescription dose.

The Skin3mm structure was chosen to demonstrate the effect, a similar trend is seen

with the Skin1mm and Skin5mm.

The bar chart in Figure 5.4 shows quantitatively the impact perturbations have on the

parameters D99%, average dose and D1% for the Skin3mm structure. In this example,

a shift towards the surface by 0.5cm (red bars), reduces the minimum dose (D99%) in

both the Clinical and VMAT cases, with more effect seen for the Clinical case, 12.6Gy

compared to 7.7Gy. The average dose is slightly increased for the Clinical plan but

reduced for the VMAT plan, as in seen for the D1% parameter. For shifts away from

the surface (green bars) the minimum dose is also reduced for both techniques,

however not to the same extent as shifts towards the surface. The impact on the

average dose for the Clinical plan is a slight reduction in dose compared to the non-

perturbed case, in contrast to the VMAT plans which see an increase in average dose.

For the D1% parameter, there is no change in dose in the perturbation of the Clinical

plan, but a significant increase in dose for the VMAT technique is observed.

For the skin structures, shifts towards the surface result in a reduction in dose for the

VMAT plan, but only reduces the minimum dose parameter for the Clinical plan. For

shifts away from the surface there is a modest decrease in minimum and average skin

dose parameters for the Clinical plan, and for the VMAT plan, as was seen in the

PTVtoSurface structure, a significant increase in maximum dose.

Page 104: Development of a technique using VMAT and robust

104

a)

b)

c)

d)

Figure 5.3: Example of DVHs for Skin3mm a)Clinical plan non-perturbed v Clinical plan perturbed 0.5 in TS direction b) Clinical plan non-perturbed v Clinical plan perturbed 0.5cm in AS direction c)VMAT plan non-perturbed v VMAT plan perturbed 0.5cm in TS direction d) VMAT plan non-perturbed v VMAT plan perturbed 0.5cm in AS direction. (Dotted line = original plan, dashed line = perturbed plan)

Page 105: Development of a technique using VMAT and robust

105

Figure 5.4 : Bar chart showing the perturbation effect on the parameters D99%, average dose and D1%, for the Skin3mm. The graph shows the impact for both Clinical and VMAT plans. (Non-perturbed = blue, perturbation 0.5cm TS = red and perturbation 0.5cm AS= green).

5.2.4 Perturbation Effect – Organs at Risk (single patient example)

The DVHs shown in Figure 5.5 and dose parameters in Table 5.4 and Table 5.5

demonstrate the impact of a 0.5cm perturbation on the heart, lung and contralateral

breast. The effect of perturbation in the TI and AI directions are shown as they

represent the worse-case scenario in the superior/inferior direction. The same impact

on dose is observed for the Clinical and VMAT plans. Perturbation towards the patient

surface (TI) reduces the dose to the heart and lung, and shifts away from the surface

(AI) increases the dose. A 0.5cm shift away from the patient surface results in both the

Clinical and VMAT plans exceeding the required dose constraint for lung, and for the

VMAT plan the heart V5% parameter is also exceeded.

There is minimal impact on the contralateral breast dose with perturbation of both the

Clinical and VMAT plans.

Page 106: Development of a technique using VMAT and robust

106

a)

b)

c)

d)

Figure 5.5: DVH for Heart (Red) Ipsilateral Lung (Orange) and Contralateral Breast (Blue) a)Clinical plan non-perturbed v Clinical plan perturbed 0.5 in TI direction b) Clinical plan non-perturbed v Clinical plan perturbed 0.5cm in AI direction c)VMAT plan non-perturbed v VMAT plan perturbed 0.5cm in TI direction d) VMAT plan non-perturbed v VMAT plan perturbed 0.5cm in AI direction. (Dotted line = original plan, dashed line = perturbed plan)

Page 107: Development of a technique using VMAT and robust

107

Structure Dosimetric Parameter

Mandatory Constraint

Optimal Constraint

Non-Perturbed (Clinical)

0.5cm TI Shift

0.5cm AI Shift

Heart

Ipsilateral Lung

Contralateral

Breast

V25% (10Gy) V5% (2Gy) V30%(12Gy) Mean Dose

≤5% ≤30% ≤17%

≤15% <3.5Gy

0% 6.4% 13.6% 0.3Gy

0% 2.9% 6.5% 0.2Gy

0.8% 13.7% 23.9% 0.3Gy

Table 5.4: Example of dosimetric parameters for the non-perturbed Clinical plan and with a perturbation of 0.5cm in the TI and AI directions. Values underlined show where constraints were not met.

Structure Dosimetric Parameter

Mandatory Constraint

Optimal Constraint

Non-perturbed (VMAT)

0.5cm TI Shift

0.5cm AI Shift

Heart

Ipsilateral Lung

Contralateral

Breast

V25% (10Gy) V5% (2Gy) V30%(12Gy) Mean Dose

≤5% ≤30% ≤17%

≤15% <3.5Gy

0% 29.9% 7.1% 2.0Gy

0% 22.3% 1.6% 2.0Gy

0% 42.3% 18.7% 2.2Gy

Table 5.5: Example of dosimetric parameters for the non-perturbed VMAT plan and with a perturbation of 0.5cm in the TI and AI directions. Values underlined show where constraints were not met.

With both the Clinical and VMAT plans, perturbation can result in dose constraints

being exceeded.

5.2.5 Perturbation Effect – PTVtoSurface (8 patient study set)

Figure 5.6 shows the difference in volume of the perturbed plans compared to the

non-perturbed plan, for the PTVtoSurface dosimetric parameters V95%, V105% and

V107%. The difference in volume is the average of the 8 patient cases, data is shown

for the Clinical and VMAT techniques and perturbations of 0.5cm and 0.3cm (Appendix

4, Table A4.1 includes the range of values for each of these parameters over the 8

patients). Similarly Figure 5.7 shows the dose difference averaged over the 8 cases, for

the mean dose to the PTVtoSurface and D1%. From these graphs it can be seen that

Page 108: Development of a technique using VMAT and robust

108

perturbation of the VMAT plan has a greater effect than perturbation of clinical plans.

As expected the magnitude of the effect is reduced with smaller shifts.

For both the Clinical and VMAT plans shifts towards the patient surface result in a

decrease in the V95% value, with the greatest impact seen for the VMAT plans. For a

0.5cmTS shift the decrease in V95%, for VMAT plans, compared to the non-perturbed

plan was on average 43.5%, ranging from 22.6%-68.1% over the 8 patients (see

Appendix 4, Table A4.1). For the Clinical plans with the same shift the decrease in

V95% was on average 23.4% (ranging from 14.9% to 29.9% over the 8 patients). A

smaller decrease in V95% is also seen for the Clinical plans with shifts away from the

patient surface. For the VMAT plans shifts away from the surface result in small

changes and in all but the 0.5cm AI direction, a slight increase in dose can be observed.

For the V105% and V107% parameters perturbation in any direction result in an

increase in PTVtoSurface volume receiving this dose. For the Clinical plans the greatest

difference between non-perturbed and perturbed, is with a shift of 0.5cm towards the

surface, which results on average with a volume difference <5%. For the VMAT plans

shifts away from the surface results in the greatest differences, with dose differences

of 52% on average for the V105% parameter for a 0.5cmAS shift, with this value

ranging from 22.1% to 73.4% across the 8 patients (Appendix 4, Table A4.1).

In Figure 5.7 it can also be observed that the impact of perturbation is in general more

signficiant for the VMAT plans. For the Clinical plans perturbation in any direction

reduces the average dose to the PTVtoSurface and increases the maximum dose, the

plans become less homogeneous. For the VMAT plans the average dose and maximum

dose, increase with shifts away from the surface. The maximum dose also increases

slightly with shifts towards the surface but average dose decreases. The average D1%

dose difference is 9.9Gy, for a shift in the 0.5cm AS direction, ranging from 6.2Gy to

15.3Gy, over the 8 patients (Appendix 4, Table A4.1).

Page 109: Development of a technique using VMAT and robust

109

Figure 5.6: Bar chart showing perturbation effect on the parameters V95% (blue), V105% (red) and V107% (green), displayed as volume difference from non-perturbed plan, for PTVtoSurface, averaged for 8 patient cases. The graph shows the impact for both Clinical and VMAT plans.

Figure 5.7: Bar chart showing perturbation effect on the parameters average dose (blue) and D1% (green), displayed as dose difference from non-perturbed plan, for PTVtoSurface, averaged for 8 patient cases. The graph shows the impact for both Clinical and VMAT plans.

Page 110: Development of a technique using VMAT and robust

110

Figure 5.8: Bar chart showing V95% for PTVtoSurface, averaged over 8 patients, under the indicated perturbation conditions. Clinical Plans (blue) and VMAT plans (red). (Dotted line = mandatory constraint, dashed line = optimal constraint)

Figure 5.9: Bar chart showing V105 values% for PTVtoSurface, averaged over 8 patients, under the indicated perturbation conditions. Clinical Plans (blue) and VMAT plans (red). (Dotted line = mandatory constraint, dashed line = optimal constraint)

Page 111: Development of a technique using VMAT and robust

111

Figure 5.10: Bar chart showing V107% values for PTVtoSurface, averaged over 8 patients, under the indicated perturbation conditions. Clinical Plans (blue) and VMAT plans (red). (Dotted line = mandatory constraint)

Figure 5.11: Bar chart showing D1% values for PTVtoSurface, averaged over 8 patients, under the indicated perturbation conditions. Clinical Plans (blue) and VMAT plans (red). (Dotted line = mandatory constraint)

Page 112: Development of a technique using VMAT and robust

112

Figure 5.8-Figure 5.11 shows the impact that the perturbation has on achieving the

required plan objectives and constraints. The data in the bar charts is the average

parameter value over the 8 patients, Appendix 4, Table A 4.2, includes additional

information on the range of these values. Figure 5.8 shows that on average the Clinical

plans only maintain the mandatory objective for V95% if the shift is <0.3cm away from

the patient surface. The range for perturbation in the 0.3cmAS perturbation direction

over the 8 patients was 84.9% to 95.2%, therefore in some cases the mandatory

objective was not achieved (Appendix 4, Table A4.2). In the other perturbation

situations the mandatory objective is, on average no longer achieved. For the VMAT

plans this objective is not achieved if perturbation is towards the patient surface.

However, for shifts up to 0.5cm away from the surface, the mandatory objective it is

maintained and this was seen across all the 8 patients (Appendix 4, Table A4.2). Figure

5.9 and Figure 5.10 show that the mandatory constraints for the parameters V105%

and V107%, are met, on average, for the Clinical plans, for perturbations of up to

0.5cm in any direction. For the VMAT plans shifts away from the surface significantly

exceed the mandatory constraints. The same is observed for the D1% parameter

(Figure 5.11).

Perturbation of the VMAT plans has a greater impact on the PTVtoSurface plan

evaluation parameters than the Clinical plans. Perturbation of both techniques can

result in mandatory objective for V95% not being achieved however the perturbation

of VMAT plans can result in dose volumes and doses that significantly exceed

mandatory planning constraints.

5.2.6 Perturbation Effect – Skin Structures (8 patient study set)

Figure 5.12 shows the difference in dose of the perturbed plans compared to the non-

perturbed plan, for the Skin3mm structure, for the dosimetric parameters D99%,

average dose and D1%. Dose differences are averaged across the 8 patients, for the

Clinical and VMAT techniques, and perturbations of 0.5cm and 0.3cm 3cm (Appendix 4,

Table A4.3 includes the range of values for each of these parameters over the 8

patients). Data for Skin3mm is shown, but the same trends were observed for the

Skin1mm and Skin5mm structures.

As was observed with the PTVtoSurface structure, in Figure 5.12 it can be seen that the

dose to the skin is impacted more by perturbation in the case of the VMAT plans

Page 113: Development of a technique using VMAT and robust

113

compared to the Clinical plans. For the Clinical plans, the maximum dose is not

significantly impacted by perturbation in any direction. For the VMAT plans,

perturbation away from the skin surface results in an increase in maximum skin dose

and shift towards the surface reduces the dose. The size of the dose difference is

proportional to the size of the perturbation. A 0.5cm TS shift on average results in a

10.5Gy dose increase, ranging from a 6.2 to 16.2Gy increase over the 8 patients

(Appendix 4, Table A4.3).

Figure 5.13 shows the impact of perturbation on the absolute dose values for the D1%

parameter for Skin3mm, it can be seen that the D1% parameter is >44Gy for the VMAT

plans with shifts away from the surface and indicates that the maximum dose

parameter is more robust to perturbations in the case of the Clinical plan (Appendix 4,

Table A4.4 shows that this is consistent over all 8 patients).

Figure 5.12: Bar chart showing perturbation effect on the parameters D99% (blue), average dose (orange) and D1% (green) for Skin3mm, displayed as dose difference from non-perturbed plan, averaged for 8 patient cases. The graph shows the impact for both Clinical and VMAT plans.

Page 114: Development of a technique using VMAT and robust

114

In Figure 5.12 as was seen for the D1% parameter, perturbation in any direction have

minimal effect on the average dose to the skin structures in the Clinical plans. The

effect on the average dose parameter in the case of the VMAT plans is consistent with

the effect observed for the PTVtoSurface structure, shifts towards the surface

reducing the average dose and shifts away increasing it. As with the maximum dose

parameter, the larger the perturbation, the larger the dose difference from non-

perturbed plan.

Figure 5.13: Bar chart showing D1% values for Skin3mm, averaged over 8 patients, under the indicated perturbation conditions. Clinical Plans (blue) and VMAT plans (red).

For the minimum dose parameter, D99%, perturbation in any direction results in a

decrease in dose for the Clinical plan. For shifts away from the surface, the average

dose differences, in the case of Clinical plans are small <0.6Gy. For Clinical plans

perturbed towards the surface, the larger the perturbation the larger the dose

difference. For Clinical plans perturbed in the 0.5cmTS direction on average this dose

difference is 11.9Gy, ranging from 8.5Gy-17.3Gy over the 8 patients (Appendix 4, Table

A4.3). Similarly, for the VMAT plans a decrease in minimum dose is also observed for

perturbations towards the surface, for the same perturbation the average dose

difference is 10.6Gy, ranging from 7.8Gy-15.4Gy. However, for the VMAT plans

Page 115: Development of a technique using VMAT and robust

115

perturbations 0.3cm away from the surface slight increase in minimum dose is

observed.

Dose to the skin structures is more robust to perturbations of the Clinical plan than the

VMAT plan.

5.2.7 Perturbation Effect – Organs at Risk (8 patient study set)

Figure 5.14 and Figure 5.15 show the average impact of perturbation on the heart

parameters for the left chest-wall cases, Appendix 4 Table 4.5 includes the range of

these values. The right sided cases have been excluded from this data as the location

of the heart results in significantly smaller doses that skew the results. As expected,

the general trend is that if the plan is perturbed away from the patient surface the

volume of the heart that receives dose increases and moving towards the surface the

volume irradiated decreases. The size of this volume correlates with the size of shift. In

addition, a perturbation in the inferior direction also impacts the volume differences

due to the location of the heart.

In Figure 5.14 it can be seen that slightly greater volume differences are observed, for

the V25% heart parameter, in the Clinical plans. The greatest difference in volume is

observed with shifts in the 0.5cmAI direction, for both techniques. However, since the

non-perturbed volumes for the Clinical and VMAT plans are 0.3% and 0.2%,

respectively, the effect of perturbation would need to increase the irradiated heart

volume by 4.7% before the dose constraint was exceeded (V25% <5%). Appendix 4,

Table 4.5 indicates that this was not exceeded in any of the patient cases.

A 0.5cm perturbation away from the patient surface also has the greatest impact on

the V5% heart parameter. In Figure 5.15 the V5% is shown for the left chest-wall

patients with the plans perturbed in this worse-case scenario. It can be seen that for

the Clinical plans the heart constraint is still met with a shift in this direction, however

in some cases the VMAT plans exceed this.

Page 116: Development of a technique using VMAT and robust

116

Figure 5.14: Bar chart showing perturbation effect on the V25% for the heart, displayed as volume difference from non-perturbed plan, averaged over left-sided cases (n=5). The graph shows the impact for both Clinical and VMAT plans.

Figure 5.15: Bar chart showing V5% (2Gy) for heart for plans perturbed in the 0.5cm AI direction. The graph shows the impact for both Clinical and VMAT plans (Dotted line=mandatory constraint).

Page 117: Development of a technique using VMAT and robust

117

Figure 5.16: Bar chart showing perturbation effect on V30% for the ipsilateral lung, displayed as volume difference from non-perturbed plan, averaged over the 8 patient cases. The graph shows the impact for both Clinical and VMAT plans (Dotted line = mandatory constraint permitted, based on volume irradiated in non-perturbed plan)

Figure 5.16 shows the impact of perturbing the plans on the ipsilateral lung volume, as

with the heart volumes a shift away from the patient surface increases the ipsilateral

lung volume irradiated and shifts towards the surface reduce this. The mandatory

constraint for the lung volume receiving 30% (12Gy) is ≤17%. The average non-

perturbed volume for the VMAT plan is 8.4% and for the Clinical plan 11.7%, therefore

for the perturbation effect to not exceed the mandatory constraint, a volume

difference of <8.6% and <5.3%, respectively is required. Figure 5.16 shows that with

the Clinical technique the 0.5cm shifts away from the surface are both > 5.3%, so

exceed the mandatory constraint (Appendix 4, Table 4.5 shows this range was from

2.2% to 12.4ki%), but the smaller 0.3cm perturbations and perturbation on the VMAT

plans are still, on average, within the required tolerance.

In Figure 5.17 the effect of plan perturbation is shown for the contralateral breast

dose. The perturbation of VMAT plans show slightly greater dose differences than the

Clinical ones, the size of the dose difference is proportional to the shift and shifts away

from the patient surface increase the dose with a decrease in dose with shifts towards

Page 118: Development of a technique using VMAT and robust

118

the surface. However, it should be noted that these dose differences are very small <

0.1Gy, so unlikely to be clinically significant. In addition, the average contralateral

breast dose for the non-perturbed Clinical plan was 0.3Gy and for the VMAT plan,

1.7Gy, therefore an increase of 0.1Gy would not exceed the tolerance of 2.5Gy.

The direction of plan perturbation had the expected effect on organ at risk dose, with

shifts towards the patient surface (away from heart and lung) reducing the volume

irradiated, and shifts away from the patient surface, increasing the volumes irradiated,

with the size of volume difference was proportional to the size of the shift.

Perturbation resulted in organ at risk dose constraints being exceeded for both Clinical

and VMAT plans.

Figure 5.17: Bar chart showing perturbation effect on contralateral breast, displayed as dose difference from non-perturbed plan, averaged over the 8 patient cases. The graph shows the impact for both Clinical and VMAT plans.

5.3 Discussion

In Chapter 3 it was concluded that VMAT plans could be created that produced dose

distributions resulting in near equivalent target and skin doses compared with Clinical

Page 119: Development of a technique using VMAT and robust

119

plans using bolus, and that met the dosimetric requirements for organs at risk, heart

and lung.

In this section the effect of set-up errors and patient contour changes on the Clinical

and VMAT plans have been investigated by applying shifts to the treatment plans and

analysing the impact on the plan evaluation parameters for target structure, skin doses

and organs at risk.

For both the Clinical and VMAT plans it was observed that shifts towards the patient

surface resulted in a decrease in PTVtoSurface volume receiving 95% of the

prescription dose, this resulted in mandatory planning objectives failing to be

achieved. The reduction in V95% is a result of the posterior edge of the target no

longer being covered by the beams, this effect is demonstrated in Figure 5.18.

a)

b)

Figure 5.18: DRRs displaying treatment field segment for a)non-perturbed Clinical plan b) Perturbed plan 0.5cm in TS direction showing PTVtoSurface (green contour) no longer fully covered.

It was seen in the DVHs, Figure 5.2a and Figure 5.2c, that the reduction in V95%, was

more significant for the VMAT plan compared to the Clinical plan, for shifts towards

the surface, suggesting that for the VMAT plans, this is not just an effect of reduced

posterior edge coverage. The reduction in skin dose also observed for VMAT plans

Page 120: Development of a technique using VMAT and robust

120

perturbed in this direction (Figure 5.3c), suggests that lack of dose coverage over the

anterior aspect of the target volume is causing this difference. In Figure 5.19 a typical

segment produced for VMAT plans is shown, non-perturbed and perturbed. It can be

seen that in the original plan the segment exposes the superficial edge of the

PTVtoSurface structure, however with the shift, most of the segment is treating air. It

is this effect that causes the V95% PTVtoSurface to be impacted more by shifts

towards the surface for VMAT plans.

The reduction in V95% for the Clinical and VMAT plans caused by under-coverage at

the posterior edge could be considered not clinically significant. This is because the

PTVtoSurface already contains a margin to take into account the effect of set-up

errors. In retrospect analysis of a CTVtoSurface structure would have been more

appropriate.

a)

b)

Figure 5.19: Typical MLC segments for VMAT plan a) Non-perturbed b) Perturbed plan 0.5cm in TS direction showing segment now outside PTVtoSurface (green contour).

Page 121: Development of a technique using VMAT and robust

121

A slight reduction in the PTVtoSurface V95% was also observed for the Clinical plans,

when the shifts were away from the patient surface. This is most likely due to the

segmented MLCS shielding a slightly larger proportion of the target area. For the VMAT

plans the V95% is largely unaffected by shifts in this direction. However unlike for the

Clinical cases, perturbation away from the surface for the VMAT plans result in the

PTVtoSurface V105%, V107% and D1% values increasing significantly. The increase

exceeds mandatory planning constraints, even with the smaller perturbation of 0.3cm.

A similar trend is observed with the maximum dose to the skin structures. This is as a

result of segments in the non-perturbed plan requiring high fluences to over-come the

build-up effect at the skin surface to deliver the required dose to the superficial

aspects of the PTVtoSurface structure. When the plan is shifted the high fluence

segments then intersect the patient where overcoming the build-up effect is not

required, resulting in significant dose to that tissue.

For the PTVtoSurface and skin structures a greater perturbation effect was seen for the

VMAT plans compared to the Clinical plans, however the same trend was not observed

for the organ at risk dose evaluation parameters with the size of the effect similar for

both techniques. The direction of perturbation effected the dose parameters as

expected, in both cases, with a shift towards the patient surface (away from heart and

lung) decreasing the organ at risk values and shifts away from the surface, increasing

them. With shifts away from the patient surface, mandatory planning constraints for

heart were shown to be exceeded with the VMAT plans, but similarly the ipsilateral

lung constraint was exceeded with the Clinical plans.

The consequences of the perturbation effect causing an underdose the target and skin

surface, is that this could affect the local control of the tumour. Perturbation resulting

in excess dose to target, skin or organs at risk could also be detrimental, resulting in

side-effects such as erythema, moist desquamation, pneumonitis or cardiomyopathy.

Although the effect of perturbation demonstrates the impact of set-up uncertainties

associated with radiotherapy deliveries, it also indicates the impact changes in the

patient contour could have, as a result of increase/decrease in swelling or weight gain

or loss.

Page 122: Development of a technique using VMAT and robust

122

5.4 Summary

VMAT plans can be created that produce similar dose distributions to the

Clinical technique using bolus however perturbation can result in under-dose or

over-dose of targets and organs at risk.

VMAT plans are not as robust as the Clinical plans with regards to target and

skin dose.

Page 123: Development of a technique using VMAT and robust

123

Chapter 6

6 Robust Optimisation

In the previous chapters it was demonstrated that VMAT plans can be created that

achieve similar target coverage and skin doses to plans produced clinically with the use

of bolus, whilst meeting organ at risk dose requirements. However, investigations

showed that VMAT plans failed to maintain target coverage when perturbations were

applied, resulting in large, high dose volumes that would unacceptable clinically. This

section investigates whether the addition of robust optimisation can resolve this issue.

The concept of plan robustness can be defined by its ability to maintain its dosimetric

qualities, which define its tumour control rate and normal tissue toxicities, despite,

changes in set-up position and variations in patient anatomy. Variations in patient

anatomy can be caused by a number of factors including change in breast shape due to

weight gain or loss, or reduction/decrease in swelling post-surgery. Limitations to the

traditional approach to maintain target coverage and organ at risk constraints, with

the use of margins around the structures, has been discussed in the introduction.

Alternative techniques to the margin approach include the optimisation of dose

distributions in different scenarios such as geometric position, using the minimax

method, as is available in RayStation. The minimax approach aims to minimise the

objective function in the worse-case scenario (e.g. the geometric position that

produces the worst result).

This chapter therefore investigates whether the addition of this optimisation feature

can produce VMAT plans that are more robust to the impact of patient movement.

6.1 Method

RayStation uses the minimax robust optimisation method whereby the objective

function is minimised in the worse-case scenario. In this study the maximum errors in

patient position were set as 0.5cm in the anterior-posterior, left-right and superior-

Page 124: Development of a technique using VMAT and robust

124

inferior directions, generating 7 scenarios to compute. For simplicity all beams were

assumed to move together (i.e. an overall isocentre shift) rather than moving

independently to each other.

As in Chapter 3, VMAT plans were created for the 8 patient cohort and consisted of

two 360⁰ arcs, collimator set to 10⁰, using a 6MV FFF beam energy. The same beam

optimisation parameters and the same starting objectives were used. The patient

position uncertainty (robustness objective) was assigned to the PTVtoSurface

minimum and maximum dose objective functions, and in cases where the heart was in

close proximity to the PTVtoSurface structure, the robustness objective was also

applied. These plans are referred to as VMATRO. As previously a dose grid of 3mm was

used (supplementary information is provided in Appendix 1 showing impact of dose

grid selection).

The robust optimisation procedure is likely to produce plans that are compromised in

some quality compared to standard optimisation as additional objectives increase the

complexity of the mathematical problem and concessions to the dose distribution will

be needed somewhere. These VMATRO were therefore assessed against the dose

evaluation metrics previously used to test for any degradation in plan quality.

The VMATRO plans were then perturbed as in the previous section by 0.5cm and

0.3cm in the TS, TS, AS and AI directions Figure 5.1 to assess to what extent the robust

optimisation function maintains plan quality when patient movement is applied.

6.2 Results

Previously it was concluded that VMAT plans could be created with dose distributions

that were equivalent to the current Clinical technique in reference to target coverage

and resulted in similar skin dose parameters. However, perturbation of the VMAT

plans in one direction resulted in greater under-dosage of the target compared with

the Clinical plan, and significant over dosage in the other.

In this section the results presented include evaluation of the VMATRO technique

against the other treatment plans using the previously described evaluation metrics,

and a comparison of the perturbation effect on VMATRO with the other treatment

Page 125: Development of a technique using VMAT and robust

125

techniques, to assess whether the use of robust optimisation on VMAT plans resolves

the issue of under and over-dosage.

6.2.1 Plan comparison – Dose Distribution (single patient example)

Figure 6.1 shows the difference in dose distribution between the VMAT and VMATRO

plans. From the dose difference map (Figure 6.1b) it is observed that the area of tissue

posterior to the target structure is receiving more dose, this is to ensure that with

perturbation away from the patient surface the target is still covered. The dose

difference map also indicates that the dose to the surface is lower.

6.2.2 Plan comparison – PTVtoSurface (single patient example)

From the DVH for the PTVtoSurface structure in Figure 6.2, the gradients of the curves

indicate that the VMATRO plan is not quite as homogeneous as the VMAT plan though

more homogeneous that the Clinical plan.

Table 6.1 shows that there is some slight degradation in the VMATRO plan compared

to VMAT and Clinical plans with regards to V95% parameter, however these

differences are small, and it still meets the optimal constraint. Hotspots are still

reduced, and homogeneity remains increased relative to the Clinical plan.

6.2.3 Plan comparison – Organs at Risk (single patient example)

Table 6.2 shows that the VMATRO meets the mandatory and optimal heart, ipsilateral

lung and contralateral breast constraints. In the case of the VMATRO plan, as with the

VMAT plan, for the low dose parameter to the heart, V5%, the volume is significantly

greater than the Clinical and No Bolus plans, this is due to the low dose bath using the

VMAT techniques. The volume however is lower for the VMATRO plan than the VMAT

plan and this will be due to the robustness objective added to the heart in the plan

optimisation. For the ipsilateral lung the V30% parameter for the VMATRO plan is 1.3%

more than the VMAT plan, this is as a result of the additional dose beyond the back

edge of target structure, observed in Figure 6.1 and the robustness objective not

added to the lung as part of the plan optimisation.

Page 126: Development of a technique using VMAT and robust

126

a)

b)

c)

Figure 6.1: Dose distribution a)VMATRO plan b)VMAT plan c) dose difference between VMATRO and VMAT

Figure 6.2: DVH for PTVtoSurface for example patient. The dashed line represents the Clinical Plan, the dotted line the VMAT plan and the solid line is the VMATRO plan

Page 127: Development of a technique using VMAT and robust

127

Structure Dosimetric Parameter

Mandatory Constraint

Optimal Constraint

Clinical Plan

No Bolus Plan

VMAT Plan

VMATRO Plan

PTVtoSurface V95% V105% V107% D1 % Average

≥ 90% ≤ 7% ≤ 2% ≤ 110%

≥ 95% ≤ 5%

96.0% 1.1% 0.0% 104.9% 40.0Gy

92.4% 1.7% 0.1% 105.4% 39.8Gy

96.6% 0.2% 0.0% 103.6% 40.0Gy

95.8% 0.3% 0% 104% 39.9Gy

Table 6.1: Dosimetric parameters achieved for PTVtoSurface single patient example. The mandatory and optimal constraints are defined.

Structure Dosimetric Parameter

Mandatory Constraint

Optimal Constraint

Clinical Plan

No Bolus Plan

VMAT Plan

VMATRO Plan

Heart

Ipsilateral Lung

Contralateral Breast

V25% (10Gy) V5% (2Gy) V30%(12Gy) Mean Dose

≤5% ≤30% ≤17%

≤15% <3.5Gy

0% 6.4% 13.6% 0.3Gy

0% 6.5% 13.4% 0.2Gy

0% 29.9% 7.1% 2.0Gy

0% 23.9% 8.4% 2.0Gy

Table 6.2: Dosimetric parameters achieved for organs at risk single patient example. The mandatory and optimal constraints are defined.

Page 128: Development of a technique using VMAT and robust

128

6.2.4 Plan comparison – PTVtoSurface (8 patient study set)

Figure 6.3 shows the grouped average results for the 8 patient cases, for each of the

dose evaluations parameters assessed for the PTVtoSurface structure. For the V95%

parameter, as was observed in the single patient example, the volumes for the

VMATRO plans suggest a slight degradation in plan quality compared to the VMAT

plans, however they are still comparable to the Clinical plans. For the average dose to

the PTVtoSurface the dose is slightly inferior compared to the VMAT plans but still

higher than the Clinical technique. For the V105% metric the VMATRO technique

creates plans with larger 105% hotspots than the VMAT and Clinical techniques, with

one patient exceeding the mandatory tolerance, V105% ≤7%, by 0.4%. A similar trend

is seen with the V107% and D1% parameters, but in these cases the VMATRO plans still

meet the plan objectives.

Page 129: Development of a technique using VMAT and robust

129

Figure 6.3: Box and Whisker plots showing the a) V95% b) average dose c) V105% d)107% and e) D1% parameters for the PTVtoSurface structures for No Bolus Plans (Blue), Clinical Plans (Red), VMAT Plans (Green) and VMATRO Plans (Purple) in the 8 patients.

a)

b)

c)

d)

e)

Page 130: Development of a technique using VMAT and robust

130

6.2.5 Plan comparison – Organs at Risk (8 patient study set)

The results in Figure 6.4 indicate that for the VMATRO plans all the organ at risk

dosimetric parameters used to define plan acceptability are met. For all the patients

the V30% lung parameter ≤17%, the heart V5%≤30%, the heart V25% ≤5% and the

contralateral breast mean dose <3.5Gy.

a)

b)

Page 131: Development of a technique using VMAT and robust

131

c)

d)

Figure 6.4: Dosimetric parameters for organs at risk a) Box and Whisker plot for Ipsilateral Lung V30% b) Bar Chart for Heart V5% (mandatory constraint shown in dashed line) c) Bar Chart for Heart V25% d) Box and Whisker plot for Contralateral Breast for No Bolus Plans (Blue), Clinical Plans (Red),VMAT Plans (Green) and VMATRO Plans (Purple) for the 8 patients.

These results show that the plans using robust optimisation remain clinically

acceptable with reference to the plan evaluation parameters for PTVtoSurface and

organs at risk. Where there is some compromise in V95% coverage compared to the

original VMAT technique, the coverage is still comparable to the Clinical technique and

Page 132: Development of a technique using VMAT and robust

132

hotter areas observed with the VMATRO plans, are still on average, within the planning

constraints.

6.2.6 Plan comparison – Skin Structures (8 patient study set)

Although the PTVtoSurface volume remains clinically acceptable it was also important

to determine the effect on skin doses and establish whether they are still equivalent to

technique using bolus. Figure 6.5 shows the D99% parameter for Skin3mm and shows

that the VMATRO plans result in a lower dose than the VMAT and Clinical plans, but

higher than without the use of bolus. The same trend was observed for the other skin

structures.

Figure 6.5: Box and Whisker plot showing the doses for the D99% parameter for the Skin3mm structure for the 8 patients for the No Bolus plans (Blue), Clinical plans (Red), VMAT plans (Green) and VMATRO plans (Purple)

For the D1% parameter, the dose variation between the techniques for all three skin

structures was <1.4Gy and all maximum dose parameters were <43.1Gy. Figure 6.6

shows the variation in maximum dose for the Skin3mm structure.

Page 133: Development of a technique using VMAT and robust

133

Figure 6.6: Box and Whisker plot showing the doses for the D1% parameter for the Skin3mm structure for the 8 patients for the No Bolus plans (Blue), Clinical plans (Red), VMAT plans (Green) and VMATRO plans (Purple)

Figure 6.7 shows a similar trend in effect for the average dose to all the skin structures

between the techniques, with the VMATRO plans resulting in a lower average dose

than the VMAT technique. For the structures, Skin3mm and Skin5mm, the VMATRO

plans demonstrate an equivalence in dose to the Clinical plans. However, in the

Skin1mm structure, the VMATRO plans result in a lower average dose than the Clinical

technique but compared to the No Bolus plans the median dose difference is 1.5Gy

greater.

As expected, the VMAT plans created using robust optimisation produce plans that

are slightly compromised compared with the original plans, but on the whole remain

clinically acceptable. The main differences are in the minimum dose to the skin

structures and average dose to the Skin1mm structure.

a) b)

Page 134: Development of a technique using VMAT and robust

134

c)

Figure 6.7: Box and Whisker plot showing the doses for the average dose parameter for the a) Skin1mm b) Skin3mm and C) Skin5mm structures for the 8 patients for the No Bolus plans (Blue), Clinical plans (Red), VMAT plans (Green) and VMATRO plans (Purple).

6.2.7 Perturbation Effect - PTVtoSurface

In the previous chapter it was concluded that whilst the VMAT plans, without bolus

were of similar quality to the Clinical plans, the impact of perturbations means these

should not be used clinically.

In this section the effect of perturbation is investigated for the VMATRO technique to

assess whether the robust optimisation removed this limitation.

Figure 6.8 shows the average change in volume seen for each of the parameters when

perturbations are applied over the 8 patients. Figure 6.8 demonstrates that the same

trend in perturbation effect is seen for the VMATRO plans as is seen in the VMAT

plans, but the magnitude of these effects is reduced.

For perturbation in the 0.5cm AS direction there is a reduction in V105% volume of

35.1% (52.1% to 17.0%), and for the V107% parameter a reduction of 38.4% (46.1% to

7.7%) between the VMAT and VMATRO techniques, with similar effects seen in the AI

Page 135: Development of a technique using VMAT and robust

135

direction. For a smaller perturbation of 0.3cm, in the AS direction the V105%

decreases from 38.7% to 8.2% and for V107%, a decrease from 30.9% to 2.7% can be

observed. However, as with the VMAT plans, on average the dose constraints for

V105% and V107% are exceeded, though not to the same extent, Figure 6.9 and Figure

6.10. It should be noted that for perturbations in the 0.3cm AS direction the volumes

for the V105% and V107% parameters, in the VMATRO plans, range from 4.9% to

20.2% and 0.4% to 9.5% respectively, so although on average across all patients the

constraint is exceeded, in some patient cases the constraints are met (Appendix 4,

Table A4.1).

For perturbations towards the surface, the volume difference for the V95% parameter

is less for the VMATRO plans compared with the VMAT plans (Figure 6.8). The average

volume difference for VMATRO plans perturbed in a 0.5cm TS direction is 16.5% less

than the VMAT plans, with a similar magnitude difference seen in the 0.5 TI direction.

A similar trend is observed for 0.3cm perturbations towards the surface resulting in

the impact of perturbation on V95% for the VMATRO plans, being equivalent to that of

the Clinical plan. Figure 6.11 shows the impact that the perturbation has on achieving

the mandatory and optimal constrains for V95% parameter. It can be seen that with

perturbations away from the surface the VMATRO plans maintain the V95% value

better than the VMAT plans, however this is well below the mandatory value. For shifts

away from the surface the V95% mandatory objectives are achieved for the VMATRO

plans for all shifts except in the 0.5cmAI direction.

Page 136: Development of a technique using VMAT and robust

136

Figure 6.8: Bar chart showing perturbation effect on the parameters V95% (blue), V105% (red) and V107% (green), displayed as volume difference from non-perturbed plan, for PTVtoSurface, averaged for 8 patient cases. The graph shows the impact for Clinical, VMAT and VMATRO plans.)

Page 137: Development of a technique using VMAT and robust

137

Figure 6.9: Bar chart showing V105% values for PTVtoSurface, averaged over 8 patients, under different perturbation conditions. Clinical Plans (blue), VMAT plans (red) and VMATRO (green). (Dotted line = mandatory constraint, dashed line = optimal constraint)

Figure 6.10: Bar chart showing V107% values for PTVtoSurface, averaged over 8 patients, under different perturbation conditions. Clinical Plans (blue) and VMAT plans (red (Dotted line = mandatory constraint)

Page 138: Development of a technique using VMAT and robust

138

Figure 6.11: Bar chart showing V95% for PTVtoSurface, averaged over 8 patients, under different perturbation conditions. Clinical Plans (blue), VMAT plans (red) and VMATRO (green). (Dotted line = mandatory constraint, dashed line = optimal constraint)

Figure 6.12: Bar chart showing perturbation effect on the parameters average dose (blue) and D1% (green), displayed as dose difference from non-perturbed plan, for PTVtoSurface, averaged for 8 patient cases. The graph shows the impact for Clinical, VMAT and VMATRO plans.

Page 139: Development of a technique using VMAT and robust

139

Figure 6.12 demonstrates that the average dose parameter for perturbed VMATRO

plans show less dose difference than perturbed VMAT plans. The same trend is

observed for VMATRO and VMAT plans, that perturbed plans shifted by a greater

distance result in a larger dose difference. VMATRO and Clinical plans perturbed

towards the surface result in similar dose differences. The dose difference for the

maximum dose parameter is also reduced with the VMATRO technique. The average

value for D1% for the non-perturbed VMATRO plan was 42.4Gy, therefore all

perturbations except 0.5cm AS (1.8Gy) meet the required dose constraint, 44Gy.

The PTVtoSurface structure is more robust for the VMATRO plans than the VMAT

plans, however still not as robust as Clinical plans. Although the maximum dose

volumes for the VMATRO are reduced compared to the VMAT plans, the dose

objectives associated with hotspots are still exceeded.

6.2.8 Perturbation Effect – Organs at Risk

Figure 6.13, Figure 6.14 and Figure 6.15 show the effect of perturbing the treatment

plans on; the ipsilateral lung receiving 30% of the prescription dose, the heart volume

receiving 5% of the prescription dose, and the mean dose to the contralateral breast.

Unlike the trend observed in the dose evaluation parameters for the PTVtoSurface,

where it was clear that the robust optimisation reduced the effect of perturbation, the

same is not true for the organs at risk. However, this is as expected as apart from in

the cases where the heart was in very close proximity to the target structure, the

robustness objective was only applied to the PTVtoSurface structure and therefore no

worse-case scenario optimisation occurred. In the case of the VMATRO plans the

effect of non-robustness and the additional dose spillage covering the posterior edge,

as shown in Figure 6.1, results in the dose constraint being exceeded for the ipsilateral

lung V30% parameter when perturbed 0.5cm in the AI direction. For the heart V5%

parameter the effect of applying the robustness objective can be seen in the cases

cw17 and cw23, where the volume receiving 5% of the prescription dose is lower for

the non-perturbed VMATRO plan than the non-perturbed VMAT plan. However, it is

Page 140: Development of a technique using VMAT and robust

140

suggested that the objective weighting was not sufficient as the heart constraint is still

exceeded in several cases.

In Figure 6.15, it can be seen that the perturbed doses to the contralateral breast are

greater for the VMATRO plans than the VMAT plans and this could be due to a trade-

off for maintaining the robustness of the PTVtoSurface structure. However, the dose

differences are very small, with the largest difference in dose =0.26Gy, for

perturbation in the 0.5cm AI direction. With a non-perturbed mean contralateral

breast dose of 1.8Gy, the perturbed VMATRO plans would still be within the tolerance

of 3.5Gy.

Figure 6.13: Bar chart showing perturbation effect on V30% for the ipsilateral lung, displayed as volume difference from non-perturbed plan, averaged over the 8 patient cases. The graph shows the impact for Clinical, VMAT and VMATRO plans. (Dotted line=mandatory constraint permitted, based on volume irradiated in non- perturbed plan. For the Clinical plan the average non-perturbed volume = 11.7%, for the VMAT plan the average non-perturbed volume = 8.4% and for the VMATRO the average non-perturbed volume = 10.1%. The constraint for V30% < 17%, therefore permitted constraint for perturbation is 5.3%, 8.6% and 6.9% for Clinical, VMAT and VMATRO plans, respectively).

Page 141: Development of a technique using VMAT and robust

141

Figure 6.14: Bar chart showing V5% (2Gy) for heart for plans perturbed in the 0.5cm AI direction. The graph shows the impact for Clinical, VMAT and VMATRO plans (Blue bars= non-perturbed, red bars=perturbed, dotted line=mandatory constraint)

Figure 6.15: Bar chart showing perturbation effect on contralateral breast, displayed as dose difference from non-perturbed plan, averaged over the 8 patient cases. The graph shows the impact for Clinical, VMAT and VMATRO plans.

Page 142: Development of a technique using VMAT and robust

142

For the organs at risk, as expected, the VMATRO plans were no more robust that the

VMAT or Clinical plans as the robustness objective was not routinely added to these

structures.

6.2.9 Perturbation Effect – Skin Structures

Figure 6.16 demonstrates for Skin3mm, the impact robust optimisation has on the

perturbation effect for the dosimetric parameters D1%, Average dose and D99%. Both

the Skin1mm and Skin5mm structures followed the same trend. In Figure 6.17 it can be

seen that for the VMATRO, the absolute doses for D1% are reduced to a value <44Gy

for the Skin3mm structure, even when perturbed, this was also observed for the 1mm

and 5mm volumes. A similar consistency in dose is observed in Figure 6.18,

demonstrating the perturbation effect for the average dose parameter.

With regard to the D99% parameter, perturbation towards the patient surface for

VMATRO plans result in a smaller dose difference compared to VMAT or Clinical plans

and perturbation away from the surface for the VMATRO plans the dose difference is

greater (the same effect was observed in the 1mm and 5mm structures). In Figure 6.19

the effect of perturbation can be seen on the D99% parameter, for the 3mm skin

structure, in direct comparison with the non-perturbed doses. It is observed that the

non-perturbed, D99% value, for the Clinical and VMAT plans, the doses are greater

than in the VMATRO cases. However, in some situations under perturbation, for

example 0.5cmTS, the VMATRO plans maintain a higher dose.

Page 143: Development of a technique using VMAT and robust

143

Figure 6.16: Bar chart showing perturbation effect on the parameters D99% (blue), average dose (orange) and D1% (green) for Skin3mm, displayed as dose difference from non-perturbed plan, averaged for 8 patient cases. The graph shows the impact for Clinical, VMAT and VMATRO plans.

Figure 6.17: Bar chart showing D1% values for Skin3mm, averaged over 8 patients, under different perturbation conditions. Clinical Plans (blue), VMAT plans (red) and VMATRO plans (green).

Page 144: Development of a technique using VMAT and robust

144

Figure 6.18: Bar chart showing average dose values for Skin3mm, averaged over the 8 patients, under different perturbation conditions. Clinical Plans (blue), VMAT plans (red) and VMATRO plans (green).

Figure 6.19: Bar chart showing D99% values for Skin3mm, averaged over 8 patients, under different perturbation conditions. Clinical Plans (blue), VMAT plans (red) and VMATRO plans (green)

Dose to the skin structures is more robust to perturbations for the VMATRO plans

compared to the VMAT plans. The maximum dose to the skin is controlled with the

VMATRO plans and the variation in minimum and average dose to the skin is reduced.

Page 145: Development of a technique using VMAT and robust

145

However, the values for the average and minimum dose parameters for the non-

perturbed VMATRO plans are slightly compromised compared to the VMAT and

Clinical plans.

6.3 Discussion

In this chapter the VMAT plans using robust optimisation were investigated as an

alternative, single plan solution to the current technique for treating post-mastectomy

patients. Unperturbed VMATRO plans were shown to be clinically acceptable when

analysing the evaluation parameters for PTVtoSurface and OAR structures. There were

slight compromises to the D99% parameter for all the skin structures, and a

compromise to the average dose parameter to the Skin1mm structure, as compared to

the Clinical technique.

Robust optimisation increased the resilience of the VMAT plans to perturbations. For

the PTVtoSurface structure, robust optimisation reduced the V105% and V107%

parameters by 35.1% and 38.7% respectively, for perturbations in the 0.5cm AS

direction. However, despite this significant reduction the volumes still exceeded the

plan objectives. The coverage of the PTVtoSurface was also shown to be better

maintained in the VMATRO plans than the VMAT plans, when perturbations were

towards the patient surface, however the V95% parameter still appeared to fall below

the mandatory required value. This may however be clinically inconsequential as the

use of robust optimisation should negate the need for a posterior margin that is

included in this PTV, and a more appropriate evaluation structure, CTVtoSurface, could

have been used.

No change in doses to the organs at risk; heart, ipsilateral lung and contralateral breast

were observed as the robust optimisation objective was not assigned to these organs

in most cases. The impact the objective does have if used on the heart constraint can

be seen in the patient example, with the V5% reduced from 29.9% to 23.9%, with the

VMATRO technique compared to the VMAT plan. For the ipsilateral lung V30%

parameter it can be seen that without using the robust optimisation objective on the

structure, the mandatory constraint is exceeded with shifts away from the patient

Page 146: Development of a technique using VMAT and robust

146

surface. This will in part be due to the spillage dose the VMATRO plans create at the

back edge of the target structure to ensure that the PTVtoSurface is robust to these

perturbations. However, including the robust optimisation function in more

optimisation objectives, will increase the calculation time. In addition, since

PTVtoSurface already contains a margin for this posterior edge, it may be more

appropriate to optimise the VMATRO plans to CTVtoSurface, which would potentially

reduce dose to heart and lung.

As was observed in the PTVtoSurface structure the robust optimisation has the desired

effect reducing maximum dose to the skin structures in the perturbed plans. The dose

reduction is to a more acceptable dose, <110% of the prescription dose, however

these are still greater than doses observed in the Clinical plans. It can also be seen that

with the VMATRO plans, perturbations towards the surface result in the D99%

parameter for the skin structures being closer to their non-perturbed values. However,

compared to the Clinical and VMAT techniques the non-perturbed value is not as

great.

Despite results showing that the VMATRO technique can produce clinically acceptable

plans, based on the dose evaluation parameters and that for skin structures of 3mm

and 5mm, the average and max dose parameters, are comparable to plans with bolus,

there are drawbacks to the technique. The most concerning issue is the high doses in

the PTVtoSurface structure. Side effects from excess dose to chest-wall tissue can

include erythema, fibrosis, chronic pain, telangiectasia and cosmetic changes, all of

which can impact quality of life for the patient.

6.4 Summary

VMAT Plans using the robust optimisation feature remain clinically acceptable

with reference to the plan evaluation parameters for PTVtoSurface and organs

at risk.

The maximum dose to the skin parameters for the VMATRO plans is

comparable to the other techniques, as is the average dose for the Skin3mm

and Skin5mm.

Page 147: Development of a technique using VMAT and robust

147

The minimum dose to the skin structures is lower for the VMATRO plans than

the Clinical or VMAT plans but higher than without the use of bolus. This also

applies to the Skin1mm structure.

The use of robust optimisation on the target structure in VMAT plans can

impact the effect of perturbation, reducing hotspots and minimising under-

dose. However perturbed plans can still exceed dose constraints.

Page 148: Development of a technique using VMAT and robust

148

Chapter 7

7 Summary

This chapter summarises the main results of the thesis, discusses the limitations to the

work undertaken and suggests further investigations that could be carried out.

7.1 Overview of Results

Radiotherapy following surgery is routine practice for patients that have had a

mastectomy and are at a high risk of recurrence. The routine approach to treating

these patients is the use of high energy, tangential photons fields which are wide

enough to encompass any change in patient contour or inaccuracies due to patient set-

up. For patients requiring radiotherapy post-mastectomy it is also common practice to

use bolus for a proportion of the treatment fractions to ensure the dose to the surface

is sufficient, though there is little consensus on the exact dose that is required and the

use of bolus can vary between clinics. The requirement for two plans adds additional

time in the patient pathway, and the use of bolus can cause inaccuracies in skin dose

due to air gaps or errors due to failure to use it. The aim of the thesis was therefore to

investigate whether a single plan solution, without the use of bolus, was achievable.

Due to the lack of consensus in the radiotherapy community in defining the required

skin dose for this group of patients, and no consistency on how skin should be defined,

the new technique was compared to the current approach used within the clinic.

The impact of bolus was initially investigated. For a sample of 8 patients it was shown

that bolus used for 7 out of 15 fractions of treatment increased the target volume

receiving 95% of the prescription dose by 7.7% compared to using no bolus at all. The

minimum dose to the 3mm thickness skin structure was enhanced by 4.6Gy, the mean

dose increased by 1.5Gy and the use of bolus had no impact on organ at risk doses.

Page 149: Development of a technique using VMAT and robust

149

Similar dose enhancements were seen for the 1mm and 5mm skin structures, with the

greatest impact observed for the 1mm structure.

It was demonstrated that VMAT plans could be created and optimised that produced

dose distributions that met required dose objectives and enhanced target coverage

compared to the clinical technique. The average dose to the skin structures could be

enhanced more than with bolus, however the minimum dose could not be increased to

the same extent but was better than no bolus treatments. The VMAT plans were also

shown to produce a low dose bath which in some cases increased the doses to organs

at risk, however still met required dose constraints.

Due to known limitations in treatment planning systems, particularly in the superficial

regions at air-tissue interfaces, physical measurements with TLDs were carried out on

an anthropomorphic phantom. The results showed measurements of dose in the build-

up region were lower than the treatment planning system predictions by, on average

8.5% for tangential treatments and 6.2% for VMAT treatments. This was consistent

with published data and sources of inaccuracy.

For VMAT plans to achieve skin doses comparable to clinical plans the optimisation

process results in high fluence segments at the surface of the patient to overcome the

build-up effect. Change in shape or movement of the patient could therefore result in

different dose deposition and was therefore investigated. For the target and skin

structures, perturbation of VMAT plans resulted in greater differences for the

evaluation parameters than with the Clinical plans, suggesting the clinical plans are

more robust. The perturbation of VMAT plans, particularly with shifts away from the

patient surface resulted in significant increases in dose which were clinically

unacceptable. Whilst the difference in dose measurements compared to the treatment

planning system suggested that the superficial dose could be 6% lower than predicted,

these increases generated by perturbation, were less superficial than where the

measurements were acquired.

Finally, the use of the mini-max robust optimisation feature in RayStation was

investigated to see whether the use of robust optimisation combined with VMAT could

overcome the limitations seen using the VMAT method alone. The robust optimisation

Page 150: Development of a technique using VMAT and robust

150

objective was applied to the target structure. The results showed a significant

reduction in dose parameter difference in the target structure, for the perturbed

VMAT plans that had been robustly optimised, compared with the original VMAT

plans. This suggested that the robust optimisation had made some impact. A reduction

in the maximum dose to skin structures was also observed when the VMATRO plans

were perturbed. The perturbed VMATRO plans however still exceeded the target

structure maximum dose constraints. Organs at risk dose constraints were also

exceeded with perturbed VMATRO plans however robust optimisation constraints

were not always applied to these structures. The use of robust optimisation also

compromised the non-perturbed VMATRO plan, reducing the enhancement to the

superficial dose. The average dose to the 3mm and 5mm skin structures remained

comparable to the clinical technique but the average dose to the 1mm structure and

minimum doses to all the skin structures were not.

Although robustly optimised VMAT plans could be created to give similar dose

distributions to the clinical technique, there was some compromise to the skin

structure doses and perturbation of these plans still resulted in distributions that did

not meet accepted dose constraints within the target structure.

7.2 Limitations

There are limitations to the data presented. The patient cohort that was used was

small, n=8, and restricted to the criteria that the Clinical plan had been created using

6MV beams for the no-bolus proportion of the plan and 10MV for the bolus part. The

small number of patients restricts confidence in the conclusions and as the choice of

energy impacts the skin dose, the findings using skin structures for comparison, are

only applicable to this group. The group of patients was also made up of both left and

right sided chest-wall sites, the position of the heart towards the left of mid-line,

meant that conclusions presented for heart doses were further impacted by the effect

of small patient numbers.

Page 151: Development of a technique using VMAT and robust

151

The data in this study represents the worst-case scenario situation, with perturbations

assumed to be in the same direction for all 15 treatments and also assumes motion in

of a rigid body. In addition, only perturbations up to 0.5cm have been examined. If the

trend is consistent with observations seen, greater perturbations will cause an increase

in the dose difference, particularly with parameters defining the hotter dose

distribution.

As previously discussed, the robust optimisation within the treatment planning

software is based on the minimax solution where the aim is to minimise the treatment

planning objective functions for the worst geometrical position. As no probability

dependence is included in this optimisation, the minimax method could over optimise

in situations which are of low probability, resulting in compromised plan quality.

Although not currently available within this treatment planning software the use of the

VMAT technique in conjunction with probabilistic optimisation functions could

produce more favourable results.

It should be noted that conclusions presented for the use of VMAT and VMATRO plans

are based on the comparison between plans within the treatment planning system and

limited TLD measurements. No further assessment or measurements were carried out

on the treatment machine’s capability of delivering the dose accurately. Similarly, the

time taken to create the VMAT plans was not compared with the creation of two plans

for the clinical technique, therefore no specific conclusions regarding reducing time in

patient pathway could be made.

7.3 Further work

Investigating the use of different methods of robust optimisation algorithms within the

planning system would require input from software manufacturer. However, there are

some aspects to the VMATRO technique where further investigation can be carried out

as an extension to this project. One aspect to this would be to evaluate the target

volume used for optimisation. In this study the structure PTVtoSurface was used, it has

been discussed that when using robust optimisation this is likely to exceed the

clinically required margin to the posterior of the treatment area, therefore

Page 152: Development of a technique using VMAT and robust

152

CTVtoSurface would be a more appropriate target optimisation structure. The target

contour is also limited at the patient surface, using a structure that extends beyond the

patient surface may enhance the surface dose but could result in more excessive

hotspots during perturbation. Alternatively, a non-uniform uncertainty of the patient

position could be used within the robustness setting, this would also reduce the

number of scenarios required to computer, reducing planning optimisation time.

Additional plans were produced to investigate whether limiting the direction of

optimisation would improve the robustness of the plans to hotspots generated within

the PTVtoSurface, by minimising the complexity of the optimisation. This however this

did not appear to have the effect required, Appendix 2.

Within this study it was observed that with a maximum error of uncertainty of 0.5cm,

used uniformly in the robustness setting, the 0.3cm perturbation dose differences for

V105% and V107%, were reduced to values that just exceeded the dose objectives.

Additional investigations into smaller perturbations, Appendix 3, showed that if the

patient was on average, perturbed in the AS direction by 0.2cm, or had an increase in

contour of 0.2cm, then in 5 out of the 8 cases, mandatory constraints would have been

met. However, it is difficult to ensure this level of accuracy.

For all the VMATRO plans used for in this planning study, two 360⁰ arcs have been

used. Other studies have shown that reduced or partial arc angles can minimise dose

to organs at risk but the compromise this may have on skin doses is not discussed in

those papers.

Page 153: Development of a technique using VMAT and robust

153

8 References

Abel, S. et al. (2017). ‘Local failure and acute radiodermatological toxicity in patients

undergoing radiation therapy with and without postmastectomy chest wall bolus: Is

bolus ever necessary?’, Practical Radiation Oncology, 7(3), pp. 167–172. doi:

10.1016/j.prro.2016.10.018.

Almberg, S. S., Lindmo, T. and Frengen, J. (2011). ‘Superficial doses in breast cancer

radiotherapy using conventional and IMRT techniques: A film-based phantom study’,

Radiotherapy and Oncology. Elsevier, 100(2), pp. 259–264. doi:

10.1016/j.radonc.2011.05.021.

Andry, G. et al. (1989). ‘Locoregional recurrences after 649 modified radical

mastectomies: incidence and significance’, European journal of surgical oncology : the

journal of the European Society of Surgical Oncology and the British Association of

Surgical Oncology. Department of Surgical Oncology, Institut Jules Bordet, Brussels,

Belgium., 15(6), pp. 476–485. Available at:

http://europepmc.org/abstract/MED/2557247.

Archibald-Heeren, B. R. et al. (2017). ‘Robust optimization of VMAT for lung cancer:

Dosimetric implications of motion compensation techniques’, Journal of Applied

Clinical Medical Physics, 18(5), pp. 104–116. doi: 10.1002/acm2.12142.

Badakhshi, H. et al. (2013). ‘Image-guided volumetric modulated arc therapy for breast

cancer: a feasibility study and plan comparison with three-dimensional conformal and

intensity-modulated radiotherapy’, The British Journal of Radiology, 86(1032). doi:

10.1259/bjr.20130515.

Bjärngard, B. E., Vadash, P. and Zhu, T. (1995). ‘Doses near the surface in high-energy

x-ray beams’, Medical Physics. John Wiley & Sons, Ltd, 22(4), pp. 465–468. doi:

10.1118/1.597477.

Blitzblau, R. C. and Horton, J. K. (2013). ‘Treatment planning technique in patients

receiving postmastectomy radiation therapy’, Practical Radiation Oncology. Elsevier,

3(4), pp. 241–248. doi: 10.1016/J.PRRO.2012.09.004.

Page 154: Development of a technique using VMAT and robust

154

Boman, E., Ojala, J., Rossi, M., & Kapanen, M. (2018). ‘Monte Carlo investigation on

the effect of air gap under bolus in post-mastectomy radiotherapy’, Physica Medica.

Elsevier, 55, pp. 82–87. doi: 10.1016/J.EJMP.2018.10.023.

Butson, M. J., Cheung, T., Yu, P., & Metcalfe, P. (2000). ‘Effects on skin dose from

unwanted air gaps under bolus in photon beam radiotherapy’, Radiation

Measurements, 32(3), pp. 201–204. doi: 10.1016/S1350-4487(99)00276-0.

Byrne, M., Hu, Y. and Archibald-Heeren, B. (2016). ‘Evaluation of RayStation robust

optimisation for superficial target coverage with setup variation in breast IMRT’,

Australasian Physical and Engineering Sciences in Medicine. Springer Netherlands,

39(3), pp. 705–716. doi: 10.1007/s13246-016-0466-6.

Cancer Research UK (2021a). Breast cancer screening and diagnosis statistics | Cancer

Research UK. Available at: https://www.cancerresearchuk.org/health-

professional/breast-cancer-screening-and-diagnosis-

statistics?_ga=2.184417830.1414124571.1609771846-

2017946323.1609771846#heading-Zero (Accessed: 4 January 2021).

Cancer Research UK (2021b). Breast cancer treatment statistics | Cancer Research UK.

Available at: https://www.cancerresearchuk.org/health-professional/cancer-

statistics/statistics-by-cancer-type/breast-cancer/diagnosis-and-treatment (Accessed:

4 January 2021).

Cancer Research UK (2021c). Cancer Statistics for the UK. Available at:

https://www.cancerresearchuk.org/health-professional/cancer-statistics-for-the-uk

(Accessed: 4 January 2021).

Cancer Research UK (2021d). Ductal carcinoma in situ (DCIS) | Breast Cancer | Cancer

Research UK. Available at: https://www.cancerresearchuk.org/about-cancer/breast-

cancer/stages-types-grades/types/ductal-carcinoma-in-situ-dcis (Accessed: 4 January

2021).

Cancer Research UK (2021e). Lobular carcinoma in situ (LCIS) | Breast conditions |

Cancer Research UK. Available at: https://www.cancerresearchuk.org/about-

Page 155: Development of a technique using VMAT and robust

155

cancer/breast-cancer/stages-types-grades/types/lobular-carcinoma-in-situ-lcis

(Accessed: 4 January 2021).

Cancer Research UK (2021f). Survival for breast cancer | Breast Cancer | Cancer

Research UK. Available at: https://www.cancerresearchuk.org/about-cancer/breast-

cancer/survival (Accessed: 4 January 2021).

Cancer Research UK (2021g). What is breast cancer? | Cancer Research UK. Available

at: https://www.cancerresearchuk.org/about-cancer/breast-cancer/about (Accessed: 4

January 2021).

Cao, Y. et al. (2017). ‘Superficial dose evaluation of four dose calculation algorithms’,

Radiation Physics and Chemistry. Pergamon, 137, pp. 23–28. doi:

10.1016/J.RADPHYSCHEM.2016.02.032.

Cavey, M. L. et al. (2005). ‘Dosimetric comparison of conventional and forward-

planned intensity-modulated techniques for comprehensive locoregional irradiation of

post-mastectomy left breast cancers’, Medical Dosimetry. Pergamon, 30(2), pp. 107–

116. doi: 10.1016/J.MEDDOS.2005.02.002.

Chan, T. C. Y., Bortfeld, T. and Tsitsiklis, J. N. (2006). ‘A robust approach to IMRT

optimization’, Physics in Medicine and Biology, 51(10), pp. 2567–2583. doi:

10.1088/0031-9155/51/10/014.

Chao, K. S. C., Perez, C. A., Wang, T. J. C., & Brady, L. W. (2011). Radiation Oncology :

Management Decisions 3rd Edition. 3rd Editio. Philadelphia: Wolters Kluwer

Health/Lippincott Williams & Wilkins. Available at:

http://ovid.com/site/catalog/books/1755.jsp (Accessed: 29 March 2019).

Chetty, I. J. et al. (2007). ‘Report of the AAPM Task Group No. 105: Issues associated

with clinical implementation of Monte Carlo-based photon and electron external beam

treatment planning’, Medical Physics, 34(12), pp. 4818–4853. doi: 10.1118/1.2795842.

Chung, H. et al. (2005). ‘Evaluation of surface and build-up region dose for intensity-

modulated radiation therapy in head and neck cancer’, Medical Physics, 32(8), pp.

2682–2689. doi: 10.1118/1.1992067.

Page 156: Development of a technique using VMAT and robust

156

Das, L. C. et al. (2017). ‘A feasibility study of 2-mm bolus for postmastectomy radiation

therapy’, Practical Radiation Oncology. Elsevier, 7(3), pp. 161–166. Available at:

https://www.sciencedirect.com/science/article/pii/S1879850016302429 (Accessed: 8

March 2018).

Devic, S. et al. (2006). ‘Accurate skin dose measurements using radiochromic film in

clinical applications; Accurate skin dose measurements using radiochromic film in

clinical applications’, Medical Physics. John Wiley & Sons, Ltd, 33(4), pp. 1116–1124.

doi: 10.1118/1.2179169.

Dias, A. G. et al. (2019). ‘Optimization of skin dose using in-vivo MOSFET dose

measurements in bolus/non-bolus fraction ratio: A VMAT and a 3DCRT study.’, Journal

of applied clinical medical physics. Wiley-Blackwell, 20(2), pp. 63–70. doi:

10.1002/acm2.12525.

Donovan, E. et al. (2007). ‘Randomised trial of standard 2D radiotherapy (RT) versus

intensity modulated radiotherapy (IMRT) in patients prescribed breast radiotherapy’,

Radiotherapy and Oncology, 82(3), pp. 254–264. doi: 10.1016/j.radonc.2006.12.008.

Dunlop, A. et al..(2019).' Evaluation of organ motion-based robust optimisation for

VMAT planning for breast and internal mammary chain radiotherapy.' Clinical and

Translational Radiation Oncology, 16, pp.60-66.

Ekambaram, V. et al. (2015). ‘Planning Aspects of Volumetric Modulated Arc Therapy

and Intensity Modulated Radio therapy in Carcinoma Left Breast - A Comparative

Study’, Asian Pacific Journal of Cancer Prevention, 16(4), pp. 1633–1636. doi:

10.7314/APJCP.2015.16.4.1633.

Fraass, B. et al. (1998). ‘American Association of Physicists in Medicine Radiation

Therapy Committee Task Group 53: Quality assurance for clinical radiotherapy

treatment planning’, Medical Physics. John Wiley and Sons Ltd, 25(10), pp. 1773–1829.

doi: 10.1118/1.598373.

Fredriksson, A., Forsgren, A. and Hårdemark, B. (2011) ‘Minimax optimization for

handling range and setup uncertainties in proton therapy’, Medical Physics, 38(3), pp.

1672–1684. doi: 10.1118/1.3556559.

Page 157: Development of a technique using VMAT and robust

157

Freedman, G. M. et al. (2009). ‘Breast intensity-modulated radiation therapy reduces

time spent with acute dermatitis for women of all breast sizes during radiation.’,

International journal of radiation oncology, biology, physics. NIH Public Access, 74(3),

pp. 689–694. doi: 10.1016/j.ijrobp.2008.08.071.

Gerbi, B. J., Meigooni, A. S. and Khan, F. M. (1987) ‘Dose buildup for obliquely incident

photon beams’, Medical Physics. John Wiley & Sons, Ltd, 14(3), pp. 393–399. doi:

10.1118/1.596055.

Haciislamoglu, E. et al. (2015). ‘Dosimetric comparison of left-sided whole-breast

irradiation with 3DCRT, forward-planned IMRT, inverse-planned IMRT, helical

tomotherapy, and volumetric arc therapy’, Physica medica : PM : an international

journal devoted to the applications of physics to medicine and biology : official journal

of the Italian Association of Biomedical Physics (AIFB). Elsevier, 31(4), pp. 360–7. doi:

10.1016/j.ejmp.2015.02.005.

Healy, E. et al. (2013). ‘Skin dose effects of postmastectomy chest wall radiation

therapy using brass mesh as an alternative to tissue equivalent bolus.’, Practical

radiation oncology. Elsevier, 3(2), pp. e45-53. doi: 10.1016/j.prro.2012.05.009.

Higgins, P. D. et al. (2007). ‘Evaluation of surface and superficial dose for head and

neck treatments using conventional or intensity-modulated techniques’, Physics in

Medicine and Biology. IOP Publishing, 52(4), pp. 1135–1146. doi: 10.1088/0031-

9155/52/4/018.

Huang, S. Y. et al. (2008). ‘The effect of skin thickness determined using breast CT on

mammographic dosimetry’, Medical Physics. John Wiley and Sons Ltd, 35(4), pp. 1199–

1206. doi: 10.1118/1.2841938.

ICRP (1991). The Biological Basis for Dose Limitation in the Skin. ICRP Skin Task Group

Report,ICRP Publication 59. Annals of the ICRP 22(2).

ICRU (1985). Determination of Dose Equivalents Resulting from External Radiation

Sources (Report 39), Journal of the International Commission on Radiation Units and

Measurements. Washington, DC: Internation Commission on Radiation Units and

Measurements. doi: 10.1093/jicru/os20.2.report39.

Page 158: Development of a technique using VMAT and robust

158

International Commission of Radiation Units and Measurements (1999). ICRU Report

62: prescribing, recording and reporting photon beam therapy (supplement to ICRU

report 50), Journal of the International Commission on Radiation Units and

Measurements. Bethesda, MD.

International Commission on Radiation Units and Measurements (1993). ICRU report

50: prescribing, recording and reporting photon beam therapy. Bethesda, MD.

International Commission on Radiation Units and Measurements (2010). ‘ICRU report

83: prescribing, recording and reporting photon beam intensity-modulated therapy

(IMRT)’.

ISRCTN Registry, ISRCTN19906132: A clinical trial testing a one week schedule of whole

breast radiotherapy against a three week schedule 2011. London. Available at:

http://www.isrctn.com/ISRCTN19906132 (Accessed: 11 April 2019).

ISRCTN Registry, ISRCTN61145589: SUPREMO, an MRC Phase III randomised trial to

assess the role of adjuvant chest wall irradiation in ‘intermediate risk’ operable breast

cancer following mastectomy 2005. London. Available at:

http://www.isrctn.com/ISRCTN61145589 (Accessed: 11 April 2019).

Javedan, K. et al. (2009). ‘Skin dose study of chest wall treatment with tomotherapy’,

Jpn J Radiol, 27, pp. 355–362. doi: 10.1007/s11604-009-0357-9.

Jensen, C. A. et al. (2018). ‘Robustness of VMAT and 3DCRT plans toward setup errors

in radiation therapy of locally advanced left-sided breast cancer with DIBH’, Physica

Medica: European Journal of Medical Physics, 45, pp. 12–18. doi:

10.1016/j.ejmp.2017.11.019.

Jin, G.-H. et al. (2013). ‘A comparative dosimetric study for treating left-sided breast

cancer for small breast size using five different radiotherapy techniques: conventional

tangential field, field-in-field, Tangential-IMRT, Multi-beam IMRT and VMAT’,

Radiation Oncology, 8(1), pp. 89–96. doi: 10.1186/1748-717X-8-89.

Jo, I. Y., Kim, S.-W. and Son, S. H. (2017). ‘Dosimetric evaluation of the skin-sparing

effects of 3-dimensional conformal radiotherapy and intensity-modulated

Page 159: Development of a technique using VMAT and robust

159

radiotherapy for left breast cancer’, Oncotarget, 8(2), pp. 3059–3063. Available at:

www.impactjournals.com/oncotarget (Accessed: 8 March 2018).

Johansen, S., Cozzi, L. and Olsen, D. R. (2009). ‘A planning comparison of dose patterns

in organs at risk and predicted risk for radiation induced malignancy in the

contralateral breast following radiation therapy of primary breast using conventional,

IMRT and Volumetric modulated arc treatment techniqu’, Acta Oncologica, 48(4), pp.

495–503. doi: 10.1080/02841860802657227.

Jong, W. L. et al. (2016). ‘In vivo skin dose measurement using MOSkin detectors in

tangential breast radiotherapy’, Physica Medica. Elsevier, 32(11), pp. 1466–1474. doi:

10.1016/J.EJMP.2016.10.022.

Kan, M. W. K., Leung, L. H. T. and Yu, P. K. N. (2012). ‘Verification and dosimetric

impact of Acuros XB algorithm on intensity modulated stereotactic radiotherapy for

locally persistent nasopharyngeal carcinoma’, Medical Physics. John Wiley and Sons

Ltd, 39(8), pp. 4705–4714. doi: 10.1118/1.4736819.

Karpf, D., Sakka, M., Metzger, M., & Grabenbauer, G. G. (2019). ‘Left breast irradiation

with tangential intensity modulated radiotherapy (t-IMRT) versus tangential volumetric

modulated arc therapy (t-VMAT): Trade-offs between secondary cancer induction risk

and optimal target coverage’, Radiation Oncology. Radiation Oncology, 14(1), pp. 1–11.

doi: 10.1186/s13014-019-1363-4.

Kim, S., Liu, C. R., Zhu, T. C., & Palta, J. R. (1998). ‘Photon beam skin dose analyses for

different clinical setups’, Medical Physics, 25(6), pp. 860–866. doi: 10.1118/1.598261.

Lai, Y. et al. (2016). ‘Modified Volumetric Modulated Arc Therapy in Left Sided Breast

Cancer After Radical Mastectomy With Flattening Filter Free Versus Flattened Beams’,

Medicine, 95(14), p. e3295. doi: 10.1097/MD.0000000000003295.

Liang, X. er al. 2020).' Using Robust Optimization for Skin Flashing in Intensity

Modulated Radiation Therapy for Breast Cancer Treatment: A Feasibility

Study.' Practical Radiation Oncology, 10(1), pp.59-69.

Page 160: Development of a technique using VMAT and robust

160

Lee, N. et al. (2002). ‘Skin toxicity due to intensity-modulated radiotherapy for head-

and-neck carcinoma’, International Journal of Radiation Oncology • Biology • Physics.

Elsevier, 53(3), pp. 630–637. doi: 10.1016/S0360-3016(02)02756-6.

Ma, C. et al. (2015). ‘Dosimetric Comparison and Evaluation of Three Radiotherapy

Techniques for Use after Modified Radical Mastectomy for Locally Advanced Left-sided

Breast Cancer’, Scientific Reports, 5(1), p. 12274. doi: 10.1038/srep12274.

Macmillan Cancer Support (2021). Staging and grading of breast cancer - Macmillan

Cancer Support. Available at: https://www.macmillan.org.uk/cancer-information-and-

support/breast-cancer/staging-and-grading-of-breast-cancer (Accessed: 4 January

2021).

Manger, R., Paxton, A. and Cerviño, L. (2016). ‘Dosimetric assessment of brass mesh

bolus for postmastectomy photon radiotherapy’, Journal of Applied Clinical Medical

Physics. Wiley-Blackwell, 17(6), pp. 86–96. doi: 10.1120/jacmp.v17i6.6221.

Mayadev, J. et al. (2015). ‘Practice Patterns in the Delivery of Radiation Therapy After

Mastectomy Among the University of California Athena Breast Health Network’,

Clinical Breast Cancer, 15(1), pp. 43–47. doi: 10.1016/j.clbc.2014.07.005.

Miura, H., Ozawa, S. and Nagata, Y. (2017). ‘Efficacy of robust optimization plan with

partial-arc VMAT for photon volumetric-modulated arc therapy: A phantom study’,

Journal of Applied Clinical Medical Physics, 18(5), pp. 97–103. doi:

10.1002/acm2.12131.

Nakamura, N. et al. (2017). ‘Post-mastectomy radiation therapy without usage of a

bolus may be a reasonable option.’, Journal of radiation research. Oxford University

Press, 58(1), pp. 66–70. doi: 10.1093/jrr/rrw055.

NICE (2018). Early and locally advanced breast cancer: Early and locally advanced

breast cancer: diagnosis and management diagnosis and management NICE guideline.

Available at: https://www.nice.org.uk/guidance/ng101/resources/early-and-locally-

advanced-breast-cancer-diagnosis-and-management-pdf-66141532913605 (Accessed:

25 March 2019).

Page 161: Development of a technique using VMAT and robust

161

Oltulu, P. et al. (2018). ‘Measurement of epidermis, dermis, and total skin thicknesses

from six different body regions with a new ethical histometric technique’, Turkish

Journal of Plastic Surgery. Medknow Publications and Media Pvt. Ltd., 26(2), p. 56. doi:

10.4103/tjps.TJPS_2_17.

Ordonez-Sanz, C., Bowles, S., Hirst, A., MacDougall, N. D. (2014). ‘A single plan solution

to chest wall radiotherapy with bolus?’, The British Journal of Radiology. British

Institute of Radiology, 87(1037), p. 20140035. doi: 10.1259/bjr.20140035.

Panettieri, V. et al. (2009). ‘AAA and PBC calculation accuracy in the surface build-up

region in tangential beam treatments. Phantom and breast case study with the Monte

Carlo code penelope’, Radiotherapy and Oncology. Elsevier, 93(1), pp. 94–101. doi:

10.1016/J.RADONC.2009.05.010.

Pignol, J.-P. et al. (2015). ‘Prospective Evaluation of Severe Skin Toxicity and Pain

During Postmastectomy Radiation Therapy’, Radiation Oncology Biology, 91(1), pp.

157–164. doi: 10.1016/j.ijrobp.2014.09.022.

Popescu, C. C. et al. (2010). ‘Volumetric Modulated Arc Therapy Improves Dosimetry

and reduces treatment time compared to conventional intensity-modulated

radiotherapy for locoregional radiotherapy of left-sided breast cancer and internal

mammary nodes.’, Radiation Oncology Biology. Elsevier, 76(1), pp. 287–295. doi:

10.1016/j.ijrobp.2009.05.038.

Quach, K. Y. et al. (2000). ‘Measurement of radiotherapy x-ray skin dose on a chest

wall phantom’, Medical Physics. John Wiley & Sons, Ltd, 27(7), pp. 1676–1680. doi:

10.1118/1.599035.

Ramlov, A. et al. (2017). ‘Clinical implementation of coverage probability planning for

nodal boosting in locally advanced cervical cancer’, Radiotherapy and Oncology.

Elsevier B.V., 123(1), pp. 158–163. doi: 10.1016/j.radonc.2017.01.015.

Recht, A. et al. (2016). ‘Postmastectomy radiotherapy: An American Society of clinical

oncology, American society for radiation oncology, and society of surgical oncology

focused guideline update’, Journal of Clinical Oncology, 34(36), pp. 4431–4442. doi:

10.1200/JCO.2016.69.1188.

Page 162: Development of a technique using VMAT and robust

162

Richmond, N. D., Daniel, J. M., Whitbourn, J. R., & Greenhalgh, A. D. (2016).

‘Dosimetric characteristics of brass mesh as bolus under megavoltage photon

irradiation’, The British Journal of Radiology. The British Institute of Radiology.,

89(1059), p. 20150796. doi: 10.1259/bjr.20150796.

De Rose, F. et al. (2016). ‘Phase II trial of hypofractionated VMAT-based treatment for

early stage breast cancer: 2-year toxicity and clinical results’, Radiation Oncology, 11,

pp. 120–128. doi: 10.1186/s13014-016-0701-z.

Rudat, V. et al. (2011). ‘Tangential beam IMRT versus tangential beam 3D-CRT of the

chest wall in postmastectomy breast cancer patients: A dosimetric comparison’,

Radiation Oncology. BioMed Central, 6(1), p. 26. doi: 10.1186/1748-717X-6-26.

Shi, L., Vedantham, S., Karellas, A., & O’Connell, A. M. (2013). ‘Technical note: Skin

thickness measurements using high-resolution flat-panel cone-beam dedicated breast

CT’, Medical Physics. John Wiley and Sons Ltd, 40(3), p. 031913. doi:

10.1118/1.4793257.

The Royal College of Radiologists (2016). Postoperative radiotherapy for breast cancer:

UK consensus statements. Available at: www.rcr.ac.uk (Accessed: 5 January 2021).

Thomas, S. J. and Hoole, A. C. F. (2004). ‘The effect of optimization on surface dose in

intensity modulated radiotherapy (IMRT).’, Physics in medicine and biology, 49(21), pp.

4919–4928. Available at: http://www.ncbi.nlm.nih.gov/pubmed/15584527 (Accessed:

5 March 2018).

Thoms, W. W. et al. (1989). ‘Multimodal Treatment For Inflammatory Breast Cancer’,

International Journal of Radiation Oncology • Biology • Physics. Elsevier, 17(4), pp.

739–745. doi: 10.1016/0360-3016(89)90060-6.

Tieu, M. T., Graham, P., Browne, L., & Chin, Y. S. (2011). ‘The Effect of Adjuvant

Postmastectomy Radiotherapy Bolus Technique on Local Recurrence’, Radiation

Oncology Biology. Elsevier, 81(3), pp. e165–e171. doi: 10.1016/j.ijrobp.2011.01.002.

Tilly, D., Holm, Å., Grusell, E., & Ahnesjö, A. (2019). ‘Probabilistic optimization of dose

coverage in radiotherapy’, Physics and Imaging in Radiation Oncology, 10(March), pp.

1–6. doi: 10.1016/j.phro.2019.03.005.

Page 163: Development of a technique using VMAT and robust

163

Turner, J. Y. et al. (2016). ‘Technique and outcome of post-mastectomy adjuvant chest

wall radiotherapy-the role of tissue-equivalent bolus in reducing risk of local

recurrence.’, The British journal of radiology. British Institute of Radiology, 89(1064), p.

20160060. doi: 10.1259/bjr.20160060.

Unkelbach, J. et al. (2018). ‘Robust radiotherapy planning’, Physics in Medicine and

Biology. IOP Publishing, 63(22). doi: 10.1088/1361-6560/aae659.

Virén, T. et al. (2015). ‘Tangential volumetric modulated arc therapy technique for left-

sided breast cancer radiotherapy’, Radiation Oncology, 10(1), pp. 79–86. doi:

10.1186/s13014-015-0392-x.

Vu, T. T. T. et al. (2007). ‘Variability in Radiation Oncologists’ Opinion on the Indication

of a Bolus in Post-mastectomy Radiotherapy: an International Survey’, Clinical

Oncology. Elsevier, 19(2), pp. 115–119. doi: 10.1016/j.clon.2006.10.004.

Wagenaar, D. et al. (2019). ‘Composite minimax robust optimization of VMAT

improves target coverage and reduces non-target dose in head and neck cancer

patients’, Radiotherapy and Oncology. Elsevier B.V., 136, pp. 71–77. doi:

10.1016/j.radonc.2019.03.019.

Xu, H. and Hatcher, G. (2016). ‘Treatment planning study of Volumetric Modulated Arc

Therapy and three dimensional field-in-field techniques for left chest-wall cancers with

regional lymph nodes’, Reports of Practical Oncology & Radiotherapy, 21(6), pp. 517–

524. doi: /10.1016/j.rpor.2016.07.005.

Yusof, F. H. et al. (2015). ‘On the Use of Optically Stimulated Luminescent Dosimeter

for Surface Dose Measurement during Radiotherapy.’, PloS one. Public Library of

Science, 10(6), p. e0128544. doi: 10.1371/journal.pone.0128544.

Zhang, Q. et al. (2015). ‘Dosimetric comparison for volumetric modulated arc therapy

and intensity-modulated radiotherapy on the left-sided chest wall and internal

mammary nodes irradiation in treating post-mastectomy breast cancer.’, Radiology

and oncology. De Gruyter Open, 49(1), pp. 91–98. doi: 10.2478/raon-2014-0033.

Page 164: Development of a technique using VMAT and robust

164

Appendix 1

Supplementary Information Dose Grids.

For all the treatment plans analysed in this work, optimisation and final dose calculation

have been carried out using a dose grid of 0.3cm. To assure the conclusions from the thesis

were not dependent on the choice of dose grid resolution, the VMAT plans, in one case,

were re-calculated with dose grids of 0.2cm and 0.1cm and a robustly optimised VMAT plan

was re-optimised with a dose grid of 0.2cm (re-optimisation at 0.1cm was deemed to take

too long). Selected structures and dosimetric parameters were then compared. Perturbation

in one direction was also analysed to check there was no significant impact on the

conclusions.

Results:

Re-calculation of VMAT plan at 0.2cm and 0.1cm:

Dosimetric Parameter

Structure Dose Grid D99% (Gy) Average (Gy) D1% (Gy)

Skin3mm 0.3cm 32.9 39.1 41.5 0.2cm 32.7 38.8 41.4 0.1cm 31.6 38.4 41.3

V95 (%) V105 (%) V107 (%)

PTVtoSurface 0.3cm 96.6 0.2 0.0 0.2cm 96.4 0.3 0.0 0.1cm 96.0 0.4 0.0

Table A1.1: Comparison of dosimetric parameters for planning structures, with VMAT plan re-calculated using dose grids 0.2cm and 0.3cm

Page 165: Development of a technique using VMAT and robust

165

Perturbation of recalculated plans in 0.5cm AS direction:

Dosimetric Parameter

Structure Dose Grid D99% (Gy) Average (Gy) D1% (Gy)

Skin 3mm 0.3cm 32.6 41.6 48.8 0.2cm 32.7 41.6 48.7 0.1cm 32.1 41.0 48.5

V95 (%) V105 (%) V107 (%)

PTVtoSurface 0.3cm 98.0 22.3 16.8 0.2cm 97.9 22.1 16.8 0.1cm 97.5 21.6 16.3

Table A1.2: Comparison of dosimetric parameters for planning structures, with VMAT plan re-calculated using dose grids 0.2cm and 0.3cm and perturbed in the 0.5cm AS direction

Re – optimisations of robustly optimised plan at 0.2cm:

Dosimetric Parameter

Structure Dose Grid D99% (Gy)

Average (Gy) D1% (Gy)

Skin3mm 0.3cm 31.6 34.0 41.3 0.2cm 30.9 33.3 42.0

V95 (%) V105 (%) V107 (%)

PTVtoSurface 0.3cm 95.8 0.3 0.0 0.2cm 95.6 3.2 0.5

Table A1.3: Comparison of dosimetric parameters for planning structures, with VMAT plan re-optimised using dose grids 0.2cm and 0.3cm

Page 166: Development of a technique using VMAT and robust

166

Perturbation of re-optimised plan in 0.5cm AS direction:

Dosimetric Parameter

Structure Dose Grid D99% (Gy) Average (Gy) D1 %(Gy)

Skin 3mm 0.3cm 28.0 37.8 43.9 0.2cm 29.5 37.4 44.4

V95 (%) V105 (%) V107 (%)

PTVtoSurface 0.3cm 93.7 14.1 5.6 0.2cm 93.6 13.2 6.4

Table A1.4: Comparison of dosimetric parameters for planning structures, with VMAT plan re-optimised using dose grids 0.2cm and 0.3cm and perturbed 0.5cm in the AS direction

Summary:

In the case of re-calculation, small differences in absolute doses are observed, D99% for the

Skin3mm structure ranging the most by 1.3Gy (31.6Gy to 32.9Gy calculated at 0.1mm and

0.3mm, respectively). Similarly a 0.6% difference is noted in the V95% parameter for

PTVtoSurface. However, the magnitude of the effect, due to perturbation, for the D1% Skin

3mm, and V105%/V107% PTVtoSurface, indicate that using recalculated plans at a smaller

dose grid would have minimal impact on the conclusions.

This is also true in the cases of the re-optimised plans with a dose grid of 0.2cm. Although

small differences in dose and volumes are observed, Skin 3mm D99% differing by 0.7Gy,in

the non-perturbed plan and PTVtoSurface V105% differing by 2.9%, the magnitude of the

parameters when plans are perturbed suggest that re-optimisation with a smaller dose grid

does not significantly change or improve the problem.

Page 167: Development of a technique using VMAT and robust

167

Appendix 2

Supplementary information: Robust Optimisation Direction.

Within this study the robust optimisation was applied to the structure PTVtoSurface for the

VMATRO plans. In the discussion it was noted that robust optimisation may not have been

necessary in all directions as the PTV defining the back edge of the treatment volume

already accounted for set-up errors in this direction. Since robust optimisation is a

challenging process for the treatment planning system, re-optimisation in limited directions,

to account for changes in position of the patient surface only, was carried out to assess

whether the effect of perturbation could be controlled better.

Two left sided patients had the VMATRO plans re-optimised, with robust optimisation

applied in the anterior and left directions only. As with the previous plans a 0.5cm

uncertainty was applied. Selected structures and dosimetric parameters were then

compared, and the effect of perturbation in one direction analysed (AS- away from surface)

at 0.5cm and 0.3cm,

Results:

Table A2.1 shows the dosimetric parameters for Skin3mm and PTVtoSurface for robustly

optimised plans 0.5cm in all directions (VMATRO all dir) and VMAT plans optimised in the

anterior and left lateral directions only (VMAT RO limited). The dosimetric parameters are

included for the non – perturbed, perturbed 0.5cm AS and 0.3cm AS plans.

Summary:

In both cases it can be seen there are slight differences in the non-perturbed plans robustly

optimised in all directions and those robustly optimised in limited directions. For the

Skin3mm structure for the D99% parameter, in Case 2 a dose difference of 0.9Gy is seen,

and for the PTVtoSurface structure volume differences up to 2.3% are observed for the

Page 168: Development of a technique using VMAT and robust

168

V105% parameter. For the perturbed plans it can be seen that the dosimetric parameters

are of a similar magnitude for both the robustly optimised plans in limited directions and

robustly optimised plans in all directions. For PTVtoSurface the V107% parameter in case 1,

is 13.7% for the plan optimised in all directions compared to 15.6% for that optimised in

limited directions, when perturbed by 0.5cm. A similar trend can be observed with the D1%

parameter for Skin 3mm, and for perturbations at 0.3cm. In case 2, it appears that the use

of limited direction robust optimisation results in plans that are worse when perturbed than

those robustly optimised in all directions.

The conclusions from the re-optimisation of only a small number of the patient plans are

limited. However, it is clear from these two cases that, despite minimising the complexity of

the calculation, the robustly optimised plans in limited directions still result in perturbed

plans with dose parameters of similar magnitudes to the plans optimised in all directions.

The conclusions from the main body of work suggested that whilst using robust optimisation

can reduce the hotpots observed in perturbed plans, these hotspots can still be of a size that

would be clinically unacceptable. These additional results show that even if optimisation

complexity is reduced, the hotspots seen on perturbed plans are of a similar magnitude and

therefore would also be clinically unacceptable.

Page 169: Development of a technique using VMAT and robust

169

Structure Plan Non-perturbed 0.5cm AS Perturbation 0.3cm AS Perturbation

Case 1

D99% (Gy)

Average (Gy)

D1% (Gy)

D99% (Gy)

Average (Gy)

D1% (Gy)

D99% (Gy)

Average (Gy)

D1% (Gy)

Skin 3mm VMATRO all dir 31.6 38.2 41.3

29.8 38.3 43.7

31.4 38.7 42.8 VMAT RO limited 32.3 38.6 40.6

32 40.1 44.7

32.9 39.7 43.2

V95 (%) V105 (%)

V107 (%)

V95 (%) V105 (%) V107 (%)

V95 (%) V105 (%) V107 (%)

PTVtoSurface VMAT RO all dir 95.8 0.3 0.0

94.5 5.6 13.7

96.1 7 1.3 VMAT RO limited 95.2 0.1 0.0 97.5 8.8 15.6 97.8 5.2 1

Case 2

D99% (Gy)

Average (Gy)

D1% (Gy)

D99% (Gy)

Average (Gy)

D1% (Gy)

D99% (Gy)

Average (Gy)

D1% (Gy)

Skin 3mm VMATRO all dir 32.1 38.8 42.1

29.1 39 45.6

31.9 39.7 44.5

VMAT RO limited 33.0 39.1 42

32.6 41.3 46.6

34.0 40.9 44.9

V95 (%) V105 (%)

V107 (%)

V95 (%) V105 (%) V107 (%)

V95 (%) V105 (%) V107 (%)

PTVtoSurface VMAT RO all dir 95.4 2.0 0.2

93.5 29.3 17.5

95.8 20.2 9.5

VMAT RO limited 95.3 4.3 1.5 97.5 38.2 26.5 98.0 23.3 11.6

Table A2.1 -Comparison of dose parameters for structures in VMATRO plans optimised in all directions and limited directions, including the perturbation in AS direction.

Page 170: Development of a technique using VMAT and robust

170

Appendix 3

Supplementary Information – Perturbation Effect

The VMATRO plans used in this study were optimised to a maximum uncertainty of 0.5cm in

all directions on the PTVtoSurface structure. The results of perturbing these plans by

displacing the isocentre by 0.5cm and 0.3cm showed that one of the most significant issues

was that the dose volume parameters V105%, V107% were exceeded for PTV if the shifts

were away from the surface, with the mandatory constraint for V105%<7% (less than 7%

volume should receive 42Gy) and V107% (less than 2% volume should receive 42.8Gy).

Alternatively, this could be a considered as a change in patient contour, as an increase in

surface by 0.3cm or 0.5cm. The conclusions assumed that the perturbation or change in

contour was the same for the entire treatment, so if on average the patient position

resulted in a perturbation by 0.5cm in the AS direction across the whole treatment or the

surface increased by 0.5cm on average throughout the whole treatment, the PTVtoSurface

volume receiving 105% of the dose, averaged over the 8 patients, was 19.6% ranging from

9.3% to 29.3%, with none of the plans meeting the mandatory constraint. With a shift of

0.3cm in the AS direction, on average PTVtoSurface V105% was 10.8% ranging from 4.9% to

20.2%, therefore some patients would meet the mandatory constraint, but some still

received a significant volume. To try to establish whether there was an acceptable average

perturbation with a shift away from the surface or average increase in contour that would

meet the mandatory criteria for V105% and V107%, 3 of the patient plans were further

investigated with perturbations calculated on their VMATRO plans which included 0.1cm,

0.2cm and 0.4cm.

Page 171: Development of a technique using VMAT and robust

171

Results:

V105% values for PTVtoSurface for perturbed VMATRO plans (red boxes indicate value fails

mandatory constraint, green boxes indicate value passes mandatory constraint):

Perturbation (cm)

Patient 0.5 0.4 0.3 0.2 0.1 0.0

cw 17 13.7 10.7 7.0 3.4 1.0 0.3 cw 10 29.3 25.6 20.2 13.6 7.4 2.0 cw3 19.2 10.5 4.5 2.8 2.1 1.1 cw2 25.8 17.5 10.5 5.3 2.9 1.2 cw12 28.7 20.9 13.3 7.8 4.0 2.1 cw15 14.1 11.5 8.4 5.6 4.0 3.1 cw18 10.3 9.7 7.8 5.9 3.6 1.4 cw23 15.9 15.7 14.7 12.2 9.3 7.4

Average 19.6 15.3 10.8 7.1 4.3 2.3

Table A3.1: V105% parameter for PTVtoSurface for every patient. VMAT RO plans perturbed in the AS direction to different extents.

V107% values for PTVtoSurface for perturbed VMATRO plans (red boxes indicate value fails

mandatory constraint, green boxes indicate value passes mandatory constraint):

Perturbation (cm)

Patient 0.5 0.4 0.3 0.2 0.1 0.0

cw 17 5.6 3.3 1.3 0.3 0.0 0.0 cw 10 17.5 14.0 9.5 4.8 2.1 0.2 cw3 4.5 1.0 0.4 0.3 0.2 0.0 cw2 10.6 5.3 1.8 0.8 0.3 0.0 cw12 12.0 6.8 3.2 1.3 0.6 0.2 cw15 5.6 4.0 2.4 1.3 0.8 0.6 cw18 2.9 2.6 2.0 1.1 0.4 0.1 cw23 7.0 6.0 4.8 3.6 2.4 1.5

Average 8.2 5.4 3.2 1.7 0.8 0.3

Table A3.2: V107% parameter for PTVtoSurface for every patient. VMAT RO plans perturbed in the AS direction to different extents.

Page 172: Development of a technique using VMAT and robust

172

Summary:

For two of the key dosimetric parameters, V105% and V107%, which resulted in the

conclusion that the VMATRO was not a viable technique, were investigated further to see if

there was an ‘average’ perturbation over a whole treatment, that would be acceptable.

From the results we can see that for this set of patients, the V105% objective is met in 5/8

cases, when perturbed by 0.2cm in the AS direction and for the same perturbation the

V107% objective is met in 6/8 cases. The results indicate that the impact of perturbation is

patient specific however, in most cases if the patient set-up and contour difference over the

entire patient treatment is within 0.2cm of the planned treatment, these parameters could

be achievable.

Page 173: Development of a technique using VMAT and robust

173

Appendix 4

Perturbation V95% (%) V105% (%) V107% (%) D1% (Gy) Ave. (Gy)

Direction Average Min Max Average Min Max Average Min Max Average Min Max Average Min Max

Clinical 0.5cm TS -23.4 -14.9 -29.9

4.9 8.8 0.6

0.7 2.7 0.0

0.8 1.6 0.2

-2.4 -2.4 -3.0

0.5cm TI -18.3 -10.2 -24.4

4.3 9.6 0.7

0.7 4.3 0.0

0.8 1.4 0.3

-1.5 -1.5 -2.8

0.5cm AS -7.7 -1.0 -19.9

2.2 4.9 1.0

0.4 0.9 0.0

0.7 0.9 0.2

-0.5 -0.5 -0.8

0.5cm AI -9.6 -1.5 -18.3

3.4 6.2 2.0

0.8 1.6 0.0

0.9 1.2 0.5

-0.5 -0.5 -0.8

0.3cm TS -11.2 -8.0 -14.2

0.8 3.8 -0.2

0.0 0.2 0.0

0.3 0.9 -0.1

-0.6 -0.6 -1.1

0.3cm TI -8.3 -5.0 -12.3

0.7 2.3 -0.1

0.0 0.4 0.0

0.2 0.6 -0.1

-0.4 -0.4 -1.0

0.3cm AS -2.2 0.5 -6.9

1.3 2.9 0.4

0.1 0.5 0.0

0.4 0.7 0.1

-0.2 -0.2 -0.5

0.3cm AI -2.7 -0.2 -6.5

2.0 3.1 1.0

0.3 0.9 0.0

0.6 0.8 0.3

-0.2 -0.2 -0.5

VMAT 0.5cm TS -43.5 -22.6 -68.1

2.1 4.8 -0.5

0.6 1.7 0.0

0.6 1.2 -0.1

-3.9 -3.9 -6.2

0.5cm TI -39.6 -18.6 -64.9

2.1 6.2 -0.2

0.6 1.9 -0.1

0.5 1.2 -0.4

-3.1 -3.1 -5.4

0.5cm AS 1.4 4.4 -1.8

52.1 72.4 22.1

46.1 66.7 16.8

9.9 15.3 6.2

3.0 3.0 0.8

0.5cm AI -1.4 0.7 -5.6

51.8 72.8 23.1

46.1 67.4 18.5

9.6 14.9 6.5

2.8 2.8 0.8

0.3cm TS -28.9 -13.8 -49.9

0.3 1.9 -1.0

0.1 0.2 -0.1

0.1 0.5 -0.4

-1.8 -1.8 -3.1

0.3cm TI -26.2 -10.8 -47.4

0.2 1.8 -0.9

0.0 0.2 -0.1

0.0 0.5 -0.5

-1.6 -1.6 -2.8

0.3cm AS 2.5 5.2 0.8

38.8 58.2 13.8

30.9 48.5 8.3

6.2 9.6 3.8

1.8 1.8 0.5

0.3cm AI 1.3 2.8 -0.4

39.5 59.7 15.4

31.6 49.8 10.0

6.2 9.4 4.1

1.7 1.7 0.5

VMATRO 0.5cm TS -27.0 -16.4 -40.7

-0.5 2.7 -3.6

-0.1 0.6 -1.2

-0.3 0.4 -1.3

-2.2 -2.2 -2.8

0.5cm TI -22.6 -12.2 -33.9

-0.3 3.2 -2.7

0.0 0.7 -0.8

-0.2 0.5 -1.0

-1.7 -1.7 -2.2

0.5cm AS -2.1 2.8 -9.2

17.0 26.6 8.5

7.7 15.9 2.9

1.9 2.6 1.1

0.4 0.4 -0.4

0.5cm AI -3.8 1.0 -10.8

17.5 36.1 7.9

7.6 17.0 2.4

1.6 2.5 1.0

0.3 0.3 -0.5

0.3cm TS -10.7 -5.9 -13.6

-1.2 0.0 -3.5

-0.3 0.0 -0.9

-0.5 0.0 -1.1

-0.8 -0.8 -1.0

0.3cm TI -8.7 -4.4 -10.9

-1.0 0.2 -2.8

-0.2 0.0 -0.6

-0.3 0.0 -0.8

-0.7 -0.7 -0.9

0.3cm AS 0.3 2.5 -2.6

8.2 15.9 3.8

2.7 8.0 0.4

1.0 1.2 0.5

0.3 0.3 0.0

0.3cm AI -0.2 1.7 -3.6 8.4 15.5 4.2 2.2 4.5 0.6 0.8 1.3 0.5 0.3 0.3 0.0

Table A4.1: Additional information on the range of perturbation effect on the planning parameters V95%, V105%, V107%, average dose and D1%, for the PTVtoSurface structure, for the 8 patient cases. Data is included for the Clinical, VMAT and VMATRO plans, the data shows the average volume difference or dose difference from the non-perturbed plan and the min/max difference over the 8 patients. The averaged data for the planning parameters shown in Figures 5.6, 5.7, 6.8 and 6.12.

Page 174: Development of a technique using VMAT and robust

174

Perturbation V95% (%) V105% (%) V107% (%) D1% (Gy) Ave. (Gy) Direction Average Min Max Average Min Max Average Min Max Average Min Max Average Min Max

Clinical non-perturbed 93.0 88.4 96.1

0.6 0.0 1.8

0.0 0.0 0.0

39.8 39.5 40.0

41.8 41.3 42.2 0.5cm TS 69.6 62.3 81.1

5.5 0.8 8.8

0.7 0.0 2.7

37.4 36.5 38.5

42.6 41.9 43.1

0.5cm TI 74.7 64.1 85.8

4.8 0.9 11.4

0.7 0.0 4.3

38.3 36.7 39.4

42.5 42.0 43.6 0.5cm AS 85.3 72.9 93.7

2.8 1.7 5.0

0.4 0.0 0.9

39.3 38.9 39.6

42.4 42.2 42.7

0.5cm AI 83.4 71.5 93.1

3.9 2.4 6.4

0.8 0.0 1.6

39.3 38.7 39.7

42.7 42.3 43.1 0.3cm TS 81.8 75.6 88.0

1.4 0.2 3.9

0.0 0.0 0.2

39.1 38.4 39.6

42.0 41.6 42.5

0.3cm TI 84.7 76.1 91.0

1.2 0.1 4.1

0.0 0.0 0.4

39.4 38.5 39.9

42.0 41.6 42.5 0.3cm AS 90.8 84.9 95.2

1.8 0.7 3.0

0.1 0.0 0.5

39.6 39.2 39.8

42.2 41.9 42.6

0.3cm AI 90.3 82.5 95.1

2.5 1.4 3.7

0.3 0.0 0.9

39.5 39.1 39.8

42.4 42.1 42.8

0.0 0.0 0.0

0.0 0.0 0.0

VMAT non-perturbed 95.6 92.9 97.3

0.7 0.2 1.5

0.1 0.0 0.1

40.0 39.8 40.1

41.9 41.4 42.1 0.5cm TS 52.1 24.9 74.0

2.9 0.3 5.5

0.6 0.0 1.7

36.1 33.6 37.7

42.4 41.7 43.0

0.5cm TI 56.0 28.1 78.0

2.9 0.0 6.9

0.7 0.0 1.9

36.9 34.4 38.5

42.3 41.2 43.1 0.5cm AS 97.0 94.5 98.6

52.8 22.3 72.6

46.2 16.8 66.7

43.0 40.8 44.5

51.7 47.6 57.2

0.5cm AI 94.3 90.8 97.9

52.5 23.3 72.9

46.2 18.5 67.4

42.8 40.7 45.1

51.5 48.4 56.8 0.3cm TS 66.8 43.0 82.8

1.0 0.0 2.6

0.1 0.0 0.3

38.2 36.7 38.9

41.9 41.3 42.4

0.3cm TI 69.4 45.5 85.9

1.0 0.0 2.5

0.1 0.0 0.2

38.4 37.0 39.3

41.9 41.0 42.4 0.3cm AS 98.2 97.2 99.1

39.5 14.0 58.4

31.0 8.3 48.5

41.8 40.5 42.6

48.0 45.3 51.5

0.3cm AI 96.9 95.7 99.0

40.2 15.6 59.9

31.7 10.0 49.8

41.7 40.5 42.8

48.0 45.9 51.3

0.0 0.0 0.0

0.0 0.0 0.0

VMATRO non-perturbed 92.8 90.9 95.8

2.6 0.3 7.4

0.5 0.0 1.5

40.0 39.8 40.3

42.4 41.6 43.4 0.5cm TS 65.8 50.5 79.4

2.1 0.0 4.8

0.4 0.0 1.1

37.8 37.1 38.8

42.1 41.2 42.8

0.5cm TI 70.2 57.3 83.6

2.3 0.0 5.3

0.5 0.0 1.2

38.3 37.7 39.3

42.2 41.0 42.9 0.5cm AS 90.8 82.9 95.9

19.6 10.3 29.3

8.2 2.9 17.5

40.4 39.7 41.2

44.2 43.3 46.0

0.5cm AI 89.0 81.5 95.4

20.1 9.3 38.2

8.1 2.5 17.1

40.3 39.6 41.4

44.0 43.3 45.0 0.3cm TS 82.1 77.6 88.5

1.5 0.0 3.9

0.2 0.0 0.7

39.2 38.9 39.8

41.9 40.9 42.6

0.3cm TI 84.1 80.0 90.1

1.7 0.0 4.6

0.3 0.0 1.1

39.3 38.9 39.9

42.0 41.2 42.8 0.3cm AS 93.1 89.1 96.1

10.8 4.9 20.2

3.2 0.4 9.5

40.3 40.0 40.8

43.3 42.5 44.5

0.3cm AI 92.6 88.8 95.9 11.0 6.2 17.6 2.7 0.7 6.0 40.3 40.0 40.9 43.2 42.7 44.0

Table A4.2: Additional information on the range of perturbation effect on the planning parameters V95%, V105%, V107%, average dose and D1%, for the PTVtoSurface structure, for the 8 patient cases. Data is included for the Clinical, VMAT and VMATRO plans, the data shows the average volume or dose and the min/max values across the 8 patients. The averaged data for the planning parameters shown in Figures 5.8-5.11 and 6.9-6.11

Page 175: Development of a technique using VMAT and robust

175

Perturbation D99% (Gy) Average (Gy) D1% (Gy) Direction Average Min Max Average Min Max Average Min Max

Clinical 0.5cm TS -11.9 -8.5 -17.3

-0.4 0.2 -1.3

0.9 1.7 0.3 0.5cm TI -6.8 -3.2 -15.6

-0.1 0.5 -1.1

0.8 1.4 0.1

0.5cm AS -0.6 -0.1 -1.3

-0.5 -0.2 -0.9

0.0 0.3 -0.3 0.5cm AI -0.5 -0.1 -0.9

-0.5 -0.1 -0.9

0.1 0.4 -0.2

0.3cm TS -2.0 -0.9 -4.7

0.1 0.3 -0.3

0.4 0.9 0.0 0.3cm TI -1.1 -0.3 -4.1

0.2 0.4 -0.3

0.3 0.7 -0.1

0.3cm AS -0.3 0.1 -0.7

-0.3 -0.1 -0.6

0.0 0.3 -0.3 0.3cm AI -0.2 0.1 -0.5

-0.3 -0.1 -0.6

0.1 0.3 -0.1

VMAT 0.5cm TS -10.6 -7.8 -15.4

-4.1 -3.3 -5.9

-1.3 0.2 -1.9 0.5cm TI -8.1 -4.9 -11.5

-3.8 -3.0 -5.3

-1.4 0.0 -2.1

0.5cm AS -1.0 1.9 -4.7

4.2 7.2 1.8

10.5 16.2 6.2 0.5cm AI -3.4 -0.9 -5.6

3.9 7.5 1.3

10.0 15.9 6.0

0.3cm TS -4.5 -3.2 -6.8

-3.0 -2.4 -3.7

-1.7 -1.1 -2.1 0.3cm TI -3.8 -2.3 -5.2

-2.9 -2.1 -3.5

-1.7 -1.1 -2.1

0.3cm AS 1.0 2.5 -0.1

3.4 4.9 2.1

7.1 11.3 4.6 0.3cm AI 0.4 2.3 -0.9

3.3 5.4 2.2

7.0 10.9 4.5

VMATRO 0.5cm TS -4.3 -2.6 -6.6

-1.6 -1.0 -2.1

-0.6 0.4 -1.5 0.5cm TI -2.6 -1.9 -3.7

-1.4 -0.8 -1.9

-0.7 0.4 -1.4

0.5cm AS -2.0 0.3 -5.2

-0.1 1.1 -2.3

2.0 3.5 0.2 0.5cm AI -2.8 -0.1 -5.9

-0.3 1.2 -2.6

1.8 2.8 0.1

0.3cm TS -1.1 -0.8 -1.5

-0.8 -0.6 -1.1

-0.6 -0.1 -1.0 0.3cm TI -0.9 -0.5 -1.4

-0.8 -0.5 -1.2

-0.7 0.0 -1.1

0.3cm AS -0.2 0.5 -1.3

0.3 0.9 -0.7

1.1 2.4 0.3 0.3cm AI -0.4 0.4 -1.5 0.2 0.9 -0.7 1.0 1.7 0.3

Table A4.3: Additional information on the range of perturbation effect on the planning parameters D99%, Average and D1% for the Skin3mm structure, over the 8 patient cases. Data is included for the Clinical, VMAT and VMATRO plans, the data shows the dose difference from the non-perturbed plan and the min/max difference over the 8 patients. The averaged data for the planning parameters shown in Figures 5.12, 6.16.

Page 176: Development of a technique using VMAT and robust

176

Perturbation D1% (Gy) D99% (Gy) Ave, (Gy) Direction Average min max Average Min Max Average Min Max

Clinical non-perturbed 41.3 40.8 41.8 34.6 33.9 35.0 38.6 38.4 38.9 0.5cm TS 42.2 41.9 42.7 22.7 17.2 26.4 38.2 37.1 38.9 0.5cm TI 42.0 41.7 42.6 27.9 18.9 31.5 38.6 37.4 39.2 0.5cm AS 41.3 40.9 41.8 34.0 33.7 34.4 38.1 37.7 38.7 0.5cm AI 41.4 40.9 41.8 34.2 33.8 34.6 38.2 37.7 38.7 0.3cm TS 41.7 41.3 41.9 32.6 29.8 34.0 38.7 38.1 39.0 0.3cm TI 41.6 41.4 41.8 33.5 30.4 34.6 38.8 38.2 39.1 0.3cm AS 41.3 40.8 41.8 34.4 34.0 34.7 38.3 38.0 38.8 0.3cm AI 41.3 40.9 41.8 34.4 34.0 34.7 38.3 38.0 38.8

VMAT non-perturbed 42.0 41.5 42.4 33.4 32.2 35.1 39.5 39.1 40.0 0.5cm TS 40.8 39.8 42.3 22.8 17.8 25.1 35.4 33.5 36.4 0.5cm TI 40.6 39.6 42.1 25.2 21.6 28.0 35.7 34.1 36.5 0.5cm AS 52.5 48.5 58.3 32.3 28.2 36.5 43.7 41.4 47.2 0.5cm AI 52.0 48.2 58.0 29.9 27.3 34.2 43.4 40.9 47.5 0.3cm TS 40.3 39.8 41.0 28.8 26.3 30.3 36.5 35.7 37.1 0.3cm TI 40.3 39.7 41.0 29.6 28.3 30.6 36.7 35.9 37.0 0.3cm AS 49.1 46.5 53.4 34.4 32.8 37.6 42.9 41.1 44.9 0.3cm AI 49.0 46.9 53.0 33.8 32.0 37.4 42.9 41.3 45.4

VMATRO non-perturbed 41.7 41.3 42.4 31.6 29.8 33.0 38.5 37.9 39.0 0.5cm TS 41.1 40.1 42.1 27.2 25.7 30.4 36.9 36.5 38.0 0.5cm TI 41.0 40.1 42.2 28.9 27.6 31.1 37.0 36.7 38.2 0.5cm AS 43.7 41.9 45.6 29.5 24.6 33.3 38.3 35.8 40.1 0.5cm AI 43.5 42.0 44.9 28.8 23.9 32.9 38.1 35.5 40.3 0.3cm TS 41.2 40.4 42.2 30.4 28.9 32.0 37.6 37.1 38.4 0.3cm TI 41.1 40.2 41.9 30.6 29.2 32.2 37.7 37.2 38.5 0.3cm AS 42.9 42.1 44.5 31.3 28.5 33.5 38.8 37.4 39.7 0.3cm AI 42.7 42.1 43.8 31.2 28.2 33.4 38.7 37.4 39.9

Table A4.4: Additional information on the range of perturbation effect on the planning parameters D99%, D1% and Average dose for the Skin3mm structure, for the 8 patient cases. Data is included for the Clinical, VMAT and VMATRO plans, the data shows the average doses and the min/max difference over the 8 patients. The averaged data for the planning parameters shown in Figures 5.13, 6.17-6.19.

Page 177: Development of a technique using VMAT and robust

177

Heart (Left CWs) Heart (Left CWs) Ipsilateral Lung (all patients)

V5 (%)

V25 (%)

V30 (%)

Perturbation Direction Average Min Max Average Min Max Average Min Max

Clinical 0.5cm TS -5.1 -6.5 -2.6 -0.3 -0.7 0.0 -5.6 -9.8 0.7 0.5cm TI -4.1 -5.1 -2.4

-0.3 -0.7 0.0

-3.3 -7.1 11.7

0.5cm AS 5.5 3.7 6.9

1.4 0.2 2.9

6.7 3.4 12.4 0.5cm AI 7.7 5.4 9.2

2.1 0.8 3.7

6.2 2.2 10.4

0.3cm TS -3.4 -4.3 -1.9

-0.2 -0.6 0.0

-3.2 -6.2 2.7 0.3cm TI -2.7 -3.3 -1.6

-0.2 -0.6 0.0

-1.8 -4.4 9.2

0.3cm AS 3.2 2.1 3.9

0.6 0.1 1.5

4.3 2.0 9.7 0.3cm AI 4.3 2.9 5.2

0.9 0.2 1.9

3.1 -3.5 6.2

VMAT 0.5cm TS -9.7 -11.1 -7.8

-0.2 -0.5 0.0

-5.9 -6.7 -4.0 0.5cm TI -7.2 -9.2 -3.9

-0.2 -0.5 0.0

-5.2 -5.9 -3.5

0.5cm AS 9.1 4.7 12.4

1.1 0.0 2.5

6.3 3.9 7.8 0.5cm AI 12.9 11.7 14.2

1.5 0.0 2.8

7.4 4.4 11.6

0.3cm TS -6.2 -6.9 -5.2

-0.2 -0.5 0.0

-3.9 -5.1 -2.6 0.3cm TI -4.6 -5.7 -2.4

-0.2 -0.5 0.0

-3.3 -3.8 -2.2

0.3cm AS 5.4 2.9 7.2

0.5 0.0 1.3

3.7 2.4 4.7 0.3cm AI 7.5 6.7 8.1

0.7 0.0 1.5

4.4 2.7 6.9

VMAT RO 0.5cm TS -10.0 -10.9 -8.5

-0.4 -1.1 0.0

-6.7 -7.9 -4.6 0.5cm TI -8.0 -10.8 -5.8

-0.4 -1.2 0.0

-5.8 -6.7 -3.9

0.5cm AS 9.2 6.3 12.5

1.5 0.0 3.1

6.5 4.0 7.8 0.5cm AI 12.7 11.1 14.4

1.8 0.1 3.4

7.7 4.5 11.7

0.3cm TS -6.4 -7.0 -5.4

-0.3 -1.0 0.0

-4.3 -5.9 -2.9 0.3cm TI -5.0 -6.9 -3.5

-0.3 -1.0 0.0

-3.6 -4.3 -2.4

0.3cm AS 5.4 3.8 7.4

0.7 0.0 1.7

3.9 2.5 4.7 0.3cm AI 7.3 6.4 8.1 0.9 0.0 1.8 4.6 2.8 7.0

Table A4.5: Additional information on the range of perturbation effect on the planning parameters for heart and lungs structures, heart. Data is included for the Clinical, VMAT and VMATRO plans, the data shows the average doses and the min/max difference over the 8 patients for the lung parameter and the 5 left chest wall patients for the heart parameters. The averaged data for the planning parameters shown in Figures 5.14-5.16 and 6.13-6.14.

Page 178: Development of a technique using VMAT and robust

178

Appendix 5

List of Alliance Manchester Business School - A units, and Specialist Medical Physics -B

units, together with credit value and assignment word count.

Unit title Credits Assignment wordcount

AMBS – A Units

A1: Professionalism and professional development in the healthcare environment

30 Practice paper – 2000 words A1 – assignment 1 – 1500 words A1 – assignment 2 – 4000 words

A2: Theoretical foundations of leadership 20 A2 – assignment 1 – 3000 words A2 – assignment 2 – 3000 words

A3: Personal and professional development to enhance performance

30 A3 – assignment 1 – 1500 words A3 – assignment 2 – 4000 words

A4: Leadership and quality improvement in the clinical and scientific environment

20 A4 – assignment 1 – 3000 words A4 – assignment 2 – 3000 words

A5: Research and innovation in health and social care

20 A5 – assignment 1 – 3000 words A5 – assignment 2 – 3000 words

Medical Physics – B Units

B1: Medical Equipment Management 10 2000 word assignment

B2: Clinical and Scientific Computing 10 2000 word assignment

B3: Dosimetry 10 Group presentation + 1500 word assignment

B4: Optimisation in Radiotherapy and Imaging

10 Group presentation + 1500 word assignment

B6: Medical statistics in medical physics 10 2000 word assignment (study design) 2000 word assignment (statistical analysis)

B8: Health technology assessment 10 3000 word assignment

B9: Clinical applications of medical imaging technologies in radiotherapy physics

20 Group presentation 2000 word assignment

B10a: Advanced Radiobiology 10 Virtual experiment + 1500 word report

B10c: Novel and specialised external beam radiotherapy

10 2000 word assignment

B10f: Radiation Protection Advice 10 1500 word report/pieces of evidence for portfolio

Generic - B Units

B5: Contemporary issues in healthcare science

20 1500 word assignment + creative project

B7: Teaching Learning Assessment 20 20 minute group presentation

Section C

C1: Innovation Project 70 4000-5000 word Literature Review Lay Presentation