I nte rna tional J ourn a l of Mode rn Engine e ring Res e a rch (IJ MER) www.ijmer.com Vol.3, I ss ue .2, Ma rch-April . 20 13 pp -888-890 I SSN: 22 49 -6645  Th omas Le k sc h a, 1 Xua n Vinh Pha m , 2 Stepha n Ba rtelm e i 3  1 Faculty of Engine ering S cie nce , De partm e nt of Med ical Techn ology , J ade Unive rsity, Germ any , Mail: [email protected] 2 Ins titute of I nn ov ations -Transfe r, Fie ld Mech atronics, J ade University, Germ any 3 Faculty of Engine ering S cience , De partm e nt of Med ical Techn ology , J ade Unive rsity, Germ any Abstract:  Th e a im o f t h is d ev e lo p ment w as t o p ro d uce an in no v at iv e exam in a t io n an d measu rin g d e v ic e f or t h e ana ly s is an d evaluation of the phys ical statics of pati en ts. The new meas uring sys tem should be able to de term ine and e val uate any dysb alance of the leg leng th, pelvi c obliquity and possibl e misalignm ent of the fe et. In future, the syste m should be implemented as a semi-automatic measuring device within the scope of medical diagnosis. Keywords: Device sy ste m , Vi de o-F ootprint, Leg -P e lvis defe ctive position, Measu re ment, Med ical te chno logy I.  I NTRODUCTION A wide va ri ety of processe s a nd m ea suring m ethods are currently use d to dete rmi ne leg or pelvic misalignm ent of  patients. These include purel y ma nual palpating or tapping the body, m ea surem ent with tape mea sures and angl e m ea sures or even image processing such as computer tomography and magne tic resonance tom ography [1]. A ll these m ethods atte mpt to dete rmi ne m isali gnm ent of the body by an alyzi ng leng th or angul ar m ea surem ents . T he ne wly developed se m i-autom atic measuring system (Fig. 1) could be used to compile the following parameters:  Body size  Shoulder height  L e g l e ng th differe nce  Pelvic obliquity  Footp rint a na lysis  Gravimetric analysis. Fig. 1 Mea suring system to dete rmi ne leg a nd pe lvic m isa lignm ent wi th m ea suring pl acean d da ta stora ge II.  PROCEDURAL METHOD In orde r to compile the a bove m entioned param ete rs, a se m i-automatic m ea suring system was de veloped which can produce and compensate for dysbalances using motorically adjustable footsteps, determine body sizes and differences using laser-assisted length and difference measurements and compile, archive and analyse all the data by implementing a specially develop software. The following sections describe the individual devices and/or measuring modules. III.  MEASUREMENT OF L ENGTHS  Th e leng t h measuring d evice c o nce r ns a lase r -assisted, magnetic absolute length m e asu r ing sys t em. Wit h t his t y p e of m ea surem ent, a se nsor is gui ded over a m agn etically coded m eta llic band . Due to the differe nt len gths of three m agn etic tracks, phase shifts are produced which, together with the signals from the sensor, produce absolute length values [2]. The m ea suring syste m us ed can realistical ly achieve a ccuracies u p to 1 mm . The m ea sured va lues establishe d are trans ferred via an RS232 interface to a PC where th ey are process ed or e val uate d by proj ect-speci fic sof tware. Pri or to each m ea surem en t, m arker poi nts are a dhe red to patien ts in relevant body area s. The las er beam is then f ocusse d on the se rel eva nt me asu ri ng points an d the distance be twee n them m ea sured. The l as er use d is as signe d to L as er Clas s 2M [3] with a wave len gth of  650 nm and output powe r of 5 mW. The len gth m ea surem ents ca n be take n with the pa tien t standing and in three di ffere nt positions (Fig. 2). Each measurement involves measuring the right and left halves of the patient as well as a rear view, following which the measurements are analysed. Pivotable measuring place Mea surem en t, ana lysis anddatastorage Developm e nt of a ne w de vice sys tem f or the de termination of  L e g - de f e ctive or/and Pe lvis - de f e ctive pos ition  www.ijmer.com 888 | Page  

Upload: tania-johnson

Post on 03-Apr-2018

213 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Development of a new device system for the determination of   Leg-defective or/and Pelvis-defective position

 

International Journal of Modern Engineering Research (IJMER)www.ijmer.com Vol.3, Issue.2, March-April. 2013 pp-888-890 ISSN: 2249-6645

 Thomas Lekscha,1 Xuan Vinh Pham,2 Stephan Bartelmei3 1Faculty of Engineering Science, Department of Medical Technology, J ade University, Germany,

Mail: [email protected] of Innovations-Transfer, Field Mechatronics, J ade University, Germany

3Faculty of Engineering Science, Department of Medical Technology, J ade University, Germany

Abstract: The aim of this development was to produce an innovative examination and measuring device for the analysis andevaluation of the physical statics of patients. The new measuring system should be able to determine and evaluate anydysbalance of the leg length, pelvic obliquity and possible misalignment of the feet. In future, the system should beimplemented as a semi-automatic measuring device within the scope of medical diagnosis.

Keywords: Device system, Video-Footprint, Leg-Pelvis defective position, Measurement, Medical technology

I .  INTRODUCTION A wide variety of processes and measuring methods are currently used to determine leg or pelvic misalignment of 

patients. These include purely manual palpating or tapping the body, measurement with tape measures and angle measures oreven image processing such as computer tomography and magnetic resonance tomography [1]. All these methods attempt todetermine misalignment of the body by analyzing length or angular measurements. The newly developed semi-automaticmeasuring system (Fig. 1) could be used to compile the following parameters:

  Body size

  Shoulder height

  Leg length difference

  Pelvic obliquity

  Footprint analysis

  Gravimetric analysis.

Fig. 1 Measuring system to determine leg and pelvic misalignment with measuring place and data storage

II .  PROCEDURAL METHOD In order to compile the above mentioned parameters, a semi-automatic measuring system was developed which can

produce and compensate for dysbalances using motorically adjustable footsteps, determine body sizes and differences usinglaser-assisted length and difference measurements and compile, archive and analyse all the data by implementing a speciallydevelop software. The following sections describe the individual devices and/or measuring modules.

I I I .  MEASUREMENT OF LENGTHS  The length measuring device concerns a laser-assisted, magnetic absolute length measuring system. With this type

of measurement, a sensor is guided over a magnetically coded metallic band. Due to the different lengths of three magnetictracks, phase shifts are produced which, together with the signals from the sensor, produce absolute length values [2]. Themeasuring system used can realistically achieve accuracies up to 1 mm. The measured values established are transferred viaan RS232 interface to a PC where they are processed or evaluated by project-specific software. Prior to each measurement,marker points are adhered to patients in relevant body areas. The laser beam is then focussed on these relevant measuringpoints and the distance between them measured. The laser used is assigned to Laser Class 2M [3] with a wavelength of 650 nm and output power of 5 mW. The length measurements can be taken with the patient standing and in three differentpositions (Fig. 2). Each measurement involves measuring the right and left halves of the patient as well as a rear view,following which the measurements are analysed.

Pivotable

measuring place

Measurement, analysisand datastorage

Development of a new device system for the determination of Leg-defective or/and Pelvis-defective position 

www.ijmer.com 888 | Page 

Page 2: Development of a new device system for the determination of   Leg-defective or/and Pelvis-defective position

 

  International J ournal of Modern Engineering Research (IJMER)www.ijmer.com Vol.3, Issue.2, March-April. 2013 pp-888-890 ISSN: 2249-6645

Fig. 2 Illustration of the various measuring positions (Positions L, B, R) as a plan view

IV.  FOOT POSITIONING K INEMATICS In order to be able to compensate or simulate the differences in leg length or pelvic misalignment, foot positioning

devices have been developed which enable each leg (foot) to be raised, lowered and/or tilted independent of the opposite sideof the body. These devices (Fig. 3) can compensate for dysbalances in the human body. While the patient is standing on thetwo (right and left) motorically adjustable moving units, an external operating panel can be used to move the right and/or leftleg (foot) up and down or tilt them to the side. Following compensation of any misalignment, the length measurement of thebody size can be completed again.

Fig. 3 Foot positioning devices, right and left, with illustration of the extreme positions in lateral, dorsal and up/down

V.  FOOTPRINT ANALYTICS A further special feature of our newly developed system is the capability to take photographs of the soles of the feed

when a load is applied and when the load applied is changed. To do this, green light is derived laterally from the cold lightcathodesin two glass plates (left and right foot positions). The light is reflected to the top surfaces of the glass through totalreflection and, therefore, cannot escape from the glass plates. However, as a result of the pressure applied by a foot, the lightcan escape from the glass plates at the pressure points exerted by the feet [4], and these point are particularly wellhighlighted visually (Fig. 4). A visual foot sole analysis or pressure diagnosis can be completed. The pressure images of thefeet are recorded in real time by a video camera and stored as individual frames. The images of the feet can be producedbefore and after compensation of a dysbalance. A comparison of the footprint images can also be used as evidence of successful compensation of a dysbalance.

Fig. 4 Illustration of two video-footprints with highlighted pressure points through decoupling light

Light extractionat pressure marks

www.ijmer.com 889 | Page 

Page 3: Development of a new device system for the determination of   Leg-defective or/and Pelvis-defective position

 

  International J ournal of Modern Engineering Research (IJMER)www.ijmer.com Vol.3, Issue.2, March-April. 2013 pp-888-890 ISSN: 2249-6645

VI.  CONCLUSION During the development of this innovative system of devices for measuring leg and pelvic misalignment, a

prototype was produced which is currently in the practical trial phase. The prototype is being tested with regard to itseveryday practicality in a large, North German orthopaedic practice by medical students. The results of the practical testsshould serve as a basis for optimising the system. An application for a patent has been registered at the Deutschen Patent-und Markenamt (German Patent and Trademark Office) for the above measuring system. The patent is registered underPatent Number: DE 10 2010 047 460 A1 [5]. Following the practical tests and optimisation phase in 2013 and 2014, it isplanned to put the new measuring system into series production.

VII.  ACKNOWLEDGEMENTS For the support by realization of this project, we would like to thank the Faculty of Engineering Science and the

Institute of Innovations-Transfer, both at the Jade University in Wilhelmshaven, Germany. We would also like to thankBrigitte Bohlen from ESBO Schuhtechnik, domiciled in Oldenburg, Germany. Some further thanks go at the Lower SaxonyCountry, Germany, and onto the European Union, for the partial financing of the development.

REFERENCES [1] Zerres, S.: Vermessungen in der Orthopädie. Ortho Press, Edition 1/2013, Page 34-35, Version 19 North, Deutscher

Patientenverlag, D-51065 Köln, Germany[2] ELGO Electric: EMAX und FEMAX, Magnetische Absolut-Längenmesssysteme.EMAX_FEMAX_D_06-03, ELGO

ElectricGmbH2003, D-78239 Rielasingen, Germany[3] German Institute forStandardization: Sicherheit von Lasereinrichtungen, Klassifizierung von Anlagen und Anforderungen.

DIN EN 60825-1, VDE 0837-1 (2012-11-00), Beuth Verlag, D-10787 Berlin, Germany[4] Ott, P.: Lichttechnik. Chapter: Lichtleiter für Beleuchtungsanwendungen. E-Learning University Heilbronn, http://ilias.hs-

heilbronn.de/goto.php?target=lm_151982&client_id=hshn, current access 28.01.2013[5] German Patent and Trade Mark Office: Verfahren und Vorrichtung zur Erfassung der Körperstatik. DE 10 2010 047 460

A1 2012.04.12, D-80331 München, Germany

AUTORS

Dr. Thomas Lekscha received his Ing. From the Faculty of Engineering Science, Jade University in Wilhelmshaven,Germany. He received his M.Sc. and also his Ph.D. from the Faculty of Electrical Engineering and Information Technology,STU University of Technology in Pressburg. His research activities are in the field of medical technologies and electronics. Today, he is a member of a research staff for medical technologies and Assistant Professor at the Department of Medical Technology, Jade University, Germany.

XuanVinh Pham (Final Year Student) received his Bc.Eng. From the Faculty of Engineering Science, J ade University inWilhelmshaven, Germany. He is still working at his M.Sc. at the RWTH Aachen University, Germany.

Prof. Dr. Stephan Bartelmei received his Ing. From the TUHH Technical University Hamburg Harburg, Germany. Hereceived his Dr. from the University Hannover, Germany. His research activities are in the field of hydraulic systems andmanufacturing systems engineering. Today, he is Professor for Mechanical Engineering at the Faculty of EngineeringScience and also the Head of the Institute of Innovations-Transfer at the Jade University in Wilhelmshaven, Germany. 

www.ijmer.com 890 | Page