determinationofchromiuminnaturalwaterbyadsorptive...
TRANSCRIPT
-
Research ArticleDetermination of Chromium in Natural Water by AdsorptiveStripping Voltammetry Using In Situ Bismuth Film Electrode
Nguyen Thị Hue ,1 Nguyen Van Hop,1 Hoang Thai Long ,1 Nguyen Hai Phong ,1Tran Ha Uyen,1 Le Quoc Hung,2 and Nguyen Nhi Phuong1,3
1University of Sciences, Hue University, Hue 530000, Vietnam2Institute of Chemistry, Vietnam Academy of Science and Technology, Ha Noi 100000, Vietnam3Pham Van Dong University, Quảng Ngãi 570000, Vietnam
Correspondence should be addressed to Nguyen �ị Hue; [email protected] and Hoang �ai Long; [email protected]
Received 2 August 2019; Revised 7 January 2020; Accepted 10 April 2020; Published 14 May 2020
Academic Editor: Evelyn O. Talbott
Copyright © 2020 Nguyen �ị Hue et al. �is is an open access article distributed under the Creative Commons AttributionLicense, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properlycited.
Development of adsorptive stripping voltammetry (AdSV) combined with in situ prepared bismuth film electrode (in situ BiFE)on glassy carbon disk surface using diethylenetriamine pentaacetic acid (DTPA) as a complexing agent and NO−3 as a catalyst todetermine the trace amount of chromium (VI) is demonstrated. According to this method, in the preconcentration step atEdep � −800mV, the bismuth film is coated on the surface of glassy carbon electrodes simultaneously with the adsorption ofcomplexes Cr(III)-DTPA. In addition to the influencing factors, the stripping voltammetry performance factors such as de-position potential, deposition time, equilibration time, cleaning potential, cleaning time, and technical parameters of differentialpulse and square wave voltammetries have been investigated, and the influence of Cr(III), Co(II), Ni(II), Ca(II), Fe(III), SO2−4 , Cl
−,and Triton X has also been investigated. �is method gained good repeatability with RSD
-
chromium traces in different objects, and achieves very lowLOD, size < 10−9–10−12M [10–20].
Another problem raised was the working electrode usedin the AdSV method. Currently, most studies on the AdSVmethod use hanging mercury drop electrodes (HMDEs)[21–34], or static mercury drop electrodes (SMDE) [19, 35]which are expensive and very difficult to fabricate. Researchon the use of mercury film electrode (MFE) and bismuth filmelectrode on glassy carbon disk surface (BiFE), which are lessexpensive and easier to fabricate, and environmentallyfriendly BiFE has been improved.
AdSV has been applied in combination with differenttypes of electrodes and stripping voltammetry signalingtechniques, and many authors have been successful indetermining the amount of traces and super chromiumtraces in complex objects. Most studies use HMDEelectrodes, some use MFE electrodes, and there are alsomany ex situ BiFE electrodes [10, 11, 14, 16, 17, 36] andgold film electrodes (AuFE) [13, 15, 37–39], and the latermodified electrodes [20, 36, 40–44] are applied more forchromium analysis in complex objects. �e complex li-gands used are DTPA [1, 21, 22, 25, 26, 28, 32, 35, 45],triethylenetetramine hexaacetic acid (TTHA)[19, 29, 31, 37], diphenylcarbazide (DPCB) [23, 29],pyrocatechol violet [24, 30], pyrogallol [34], rubeanicacid [33], neo TT [40], and quercetin [20], and thecommon base ingredients are CH3COONa, acetate buffer(CH3COOH/CH3COONa), and CH3COONa/NaNO3; allstudies analyzed chromium in river water, seawater,groundwater, tea water, and wastewater, and almost noworks have analyzed chromium in sediments by theadsorptive stripping voltammetry method and no authorhas used in situ BiFE to analyze chromium in environ-mental objects. Overall, studies have achieved very lowdetection limits from 10-9 to 10-10M (or from 0.05 ppb to0.005 ppb). Table 1 shows the summary of publishedstudies on chromium determination by the AdSVmethod.
When analyzing Cr(VI) by the AdSVmethod, Yong et al.[10] and Lin et al. [11] proposed the adsorption mechanismas follows.
�e first stage of the adsorptive stripping voltammetrymethod is the movement of Cr(VI) in the solution to theelectrode surface and is reduced by the following reaction.
Cr(VI) + 3e⟶Cr(III); E1/2 � −0.2V in comparisonwith 3M Ag/AgCl/KCl electrode; in the presence of DTPAions on the interface between the electrode membrane andthe solution, Cr(III) ions form complexes quickly andadsorbed onto the electrode surface. �e main nature of thiscomplex is unknown, and some authors [21, 46] said thatthis complex exists mainly as Cr(III)-DTPA2− and a little inCr(III)-HDTPA−, or Khan [47] suggested that the complexof Cr(III) with DTPA exists as [Cr(III) (H2O)HY]−, [Cr(III)-DTPA]2−, and so on.
After complex adsorption, potential scan was conductedfrom −0.9V to −1.35V [10], or from −0.8V to −1.4V [11]; asa result of this process, the Cr(III)-DTPA complex is re-duced to the Cr(II)-DTPA complex [10, 11]:
Cr(III) − DTPA + e⟶ Cr(II) − DTPA (1)
�e signal recorded with this reduction is at E1/2� −1.15V[10, 11] and at E1/2� −1.1V in comparison with 3MAg/AgCl/KCl electrode.
�e role of NO−3 ions is to oxidize Cr(II)-DTPA com-plexes into Cr(III)-DTPA complexes.
�anks to the oxidation of NO−3 , the peak is higher andthe sensitivity of the method is increased.
Cr(II) − DTPA⟶N O−3 Cr(III) − DTPA (2)
In order to contribute to the development of the AdSVmethod, we have conducted several studies to determinechromium by the AdSV method using mercury film elec-trodes, ex situ bismuth film electrode, and in situ bismuthfilm electrode. In this paper, we present the results ofchromium determination by the AdSV method using in situbismuth film electrode (in situ BiFE) in the presence ofDTPA as a complexing agent and ion NO−3 as a catalyst.Literature survey revealed that in situ BiFE has never beenused for chromium determination by the adsorptive strip-ping voltammetry.
2. Experimental
2.1. Apparatus and Reagents. Square wave stripping vol-tammetric measurements were conducted on Electro-chemical Analyzer 797 VA Computrace (Metrohm,Switzerland) accompanied with three electrodes. �eseelectrodes were inserted into the 80ml capacity electro-chemical cell. �e working electrode was a glassy carbonrotating disk electrode with d� 2.8± 0.1mm; the referenceelectrode was a 3M Ag/AgCl/ KCl electrode and the aux-iliary electrode was a platinum wire. All measurements werecarried out at 25± 1°C.
DTPA (diethylenetriamine pentaacetic acid) is used as acomplexing agent for chromium. 50.10−3M DTPA solutionwas prepared by dissolving 4.916 g of DTPA in double-distilled water and then adding 25% aqueous ammoniasolution until the pH reaches 6.0.
0.2M Cr(VI) stock solution was prepared by weighing7.35 g of K2Cr2O7 (Merck, purity of 95–98%), dissolving,and making it up to 250ml with double-distilled water.Cr(VI) working solutions such as Cr(VI) 10−6M and Cr(VI)10−7M are diluted daily from this stock solution.
0.48.10−3M Bi(III) working solution was prepared from4.8 .10−3M Bi(III) (the type used for analyzing atomic ab-sorption spectrometry from Merck).
Acetate buffer (pH� 6) was prepared from NaCH3COO(Merck, purity of 95–98%) and CH3COOH (Merck, purity97%). 2.5M NaNO3 solution was prepared from NaNO3(Merck, purity of 95–98%).
Othermetal ionic solutions such as Co(II), Ni(II), Zn(II),Cr(III), Fe(III), and Ca(II) are from the corresponding1000mg/l stock solution (the type used for analyzing atomicabsorption spectrometry from Merck). Triton X-100working solution was prepared from Triton X-100 (Merck).
2 Journal of Environmental and Public Health
-
Tabl
e1:
Summaryof
publish
edworks
onstripp
ingvoltammetry
metho
dforchromium
determ
ination.
No.
Ligand
sBa
ckgrou
ndsolutio
nWorking
electrod
eMeasurement
techniqu
esLO
D(pbb
)Analytical
object
Determined
form
Timeof
publication
References
1DTP
ACH
3COONH
4,NaN
O3
(pH
�5.2)
HMDE
DP-AdS
V1.2
Fake
template
Total
chromium
1987
[21]
2Dipheny
lcarbazide
H2SO40.1M
GE
LSV
0.05
Fake
template
Cr(VI)
1988
[29]
3TT
HA
CH
3COONa,NaN
O3
(pH
�5.5)
HMDE
DP-CSV
1.04
Fake
template
Cr(VI)
1988
[29]
4TT
HA
CH
3COONa,NaN
O3
(pH
�5.5)
HMDE
SqW-A
V0.02
Fake
template
Cr(VI)
1988
[29]
5DTP
AKH
2PO
4,Na 2HPO
4(pH
�6–
7)HMDE
DP-CSV
0.01
Fake
template
Cr(VI)
1990
[32]
6Dipheny
lcarbazide
H2SO40.3M
HMDE
DP-AdS
V0.02
Groun
dwater
Cr(VI)
1992
[23]
7DTP
A2.5mM
Acetate
buffer(pH
5.2)
HMDE
SqW-A
dSV
0.05
Seaw
ater
Cr(VI),total
chromium
1992
[45]
8DTP
A0.01
MNaN
O30.5M,M
ES(m
orph
olinoethanesulfonic
acid)(pH
6.1)
HMDE
DP-AdS
V0.62
Riverwater,tap
water
Total
chromium
2000
[28]
9DTP
ANaN
O3,acetatebu
ffer(pH
5.7)
HMDE
DP-AdS
V
Cr(VI):
0.005
activ
ated
Cr(III):
0.52
Riverwater,
lake
water,
sewer
water
Cr(VI),
activ
ated
Cr(III),total
chromium
2001
[1]
10TT
HA
0.1M
(triethylenetetram
inhexaacetic
acid)
NaC
H3C
OO
1M,N
aNO35M
(pH
6.2–
6.5)
SMDE
SqW-A
dSV
0.52
Leaf
Cr(VI)
2003
[19]
11DTP
AAcetate
buffer(pH
�6),K
NO3
0.25
MHMDE
DP-AdS
V0.004
Wastewater
Cr(VI),total
chromium
2004
[22]
12Ru
beanic
acid
(dith
iooxam
ide)
Acetate
buffer(pH
6),K
NO3
0.20
MHMDE
DP-AdS
V1976
Riverwater,
seaw
ater,
sewage,vinegar
Cr(VI)
2012
[33]
13DTP
AAcetate
buffer(pH
6)NaN
O3
HMDE
AdS
V18.2
Hum
anurine
Cr(VI)
2005
[26]
14TT
HA
0.2M
NaC
H3C
OO0.1M,N
aNO35M
(pH
6.2)
HMDE
CV-A
dSV
0.3
Cr(VI)
1997
[31]
15DTP
ANaC
H3C
OO
0.01
M,N
aNO3
0.5M
(pH
8.5)
SMDE
156.10
−5
Riverwater,tap
water
Cr(VI)
1999
[35]
16DTP
A5mM
NaC
H3C
OO
0.15
M,N
aNO3
0.7M
(pH
6)HMDE
SqW-A
dSV
0.05
Cem
ent
Cr(VI)
2011
[25]
17Py
rogallo
lred
0.4M
acetatebu
ffer(pH
4.5)
HMDE
SqW-A
dSV
0.05
Seaw
ater
Cr(VI),
Cr(III),total
chromium
2012
[34]
18DTP
A0.05
MNaC
H3C
OO
0.2M
(pH
6.2)
HMDE
DP-AdS
V1.04
Electrop
latin
gwaste
water
Cr(VI)
2004
[27]
19Py
rocatechol
violet
Acetate
buffer
HMDE
DP-AdS
VCr(VI)
1997
[24]
20
HED
TA(N
-2-hydroxyethyl
ethylenediam
ine-N,N′,N″-triacetic
acid)andPC
V(pyrocatecho
lviolet)
0.1M
acetatebu
ffer(pH
6)KNO
32M
HMDE
DP-AdS
VCr(III),
Cr(VI)
2002
[30]
Journal of Environmental and Public Health 3
-
Tabl
e1:
Con
tinued.
No.
Ligand
sBa
ckgrou
ndsolutio
nWorking
electrod
eMeasurement
techniqu
esLO
D(pbb
)Analytical
object
Determined
form
Timeof
publication
References
21Cup
ferron
0.01
MPIPE
S0.2M
(pH
�7)
BiFE
exsitu
SqW-A
dSV
0.1
Labo
ratory
water,
cigarette
s,soil
sample
Total
chromium
2004
[14]
22DTP
A5mM
Acetate
0.1M
(pH
�6.0)
BiFE
exsitu
SqW-A
dSV
0.015
Riverwater
Total
chromium
2005
[11]
23DTP
A5mM
NaO
Ac0.1M
(pH
�6.0)
BiFE
exsitu
SqW-A
dSV
0.015
Bloo
dsample
Total
chromium
2006
[10]
24DTP
A5mM
Acetate
0.1M
(pH
�6.0)
BiFE
exsitu
SqW-A
dSV
0.017
(CrV
I );0.022(C
rtotal)
Riverwater
Cr(VI);
Cr(III)
2010
[16]
25PA
R(4-(2-pyridylazo)resorcinol)
CH
3COOH–C
H3C
OONa,
triso
dium
citrate
BiFE
exsitu
SqW-A
dSV
0.01
Tapwater,lake
water,soil
sample
Total
chromium
2013
[17]
26DTP
ACH
3COONH
4,NaN
O3
(pH
�5.2)
Bifilm
wrapp
edsin
glewalled
carbon
nano
tubes
DP-AdS
V0.12.10−
3Fake
template
Total
chromium
2013
[18]
27DTP
A0.01
M0.1M
acetatebu
ffer(pH
�6.0)
Hg(A
g)FE
DP-AdS
V0.004
Natural
water,
drinking
water
Cr(VI)
2006
[12]
28TT
HA
CH
3COONa,NaN
O3
(pH
�5.5)
Goldfilm
mod
ified
carbon
compo
site
electrod
eDP-CSV
4.0
Fake
template
Cr(VI)
2014
[37]
29PE
T(4pyridine-ethanethiol)
NaF
0.15
M(pH
�4.5)
PET/nano
-Au/Pt-
RDelectrod
eDP-AdS
V0.001
Seaw
ater
Cr(VI)
2015
[13]
30DTP
AKH
2PO
4,Na 2HPO
4(pH
�6–
7)
(Flower-like
self-
assemblyof
gold
nano
particles)
AuN
Ps/G
CE
DP–
CSV
0.001
Fake
template
Cr(VI)
2012
[38]
31HCl(pH
�2)
AuN
Ps/nano-TiC/
GCE
DP-AdS
V2.08
Seaw
ater
Cr(VI)
2015
[15]
32HClO
40.06
MAuN
Ps/SPC
EASV
0.002
Tapwater,
seaw
ater
Cr(VI)
2015
[39]
33QH2(quercetin)
Acetate
buffer(pH
�6),K
NO3
0.7M
QH2/MWCNT-
SPCE(quercetin/
multiw
alledcarbon
nano
tubesscreen-
printedcarbon
electrod
e)
DP-AdS
V15.9
Drink
ingwater
Cr(VI)
2013
[20]
34-
Acetate
buffer(pH
�5)
μNPs/G
CE
DP-SV
0.01
Electrop
latin
gwaste
water
Cr(III)
2015
[44]
35DTP
A0.1M
Acetate
buffer(pH
6),K
NO3
0.25
MμNPs/BiFE
DP-AdS
V0.12.10−
3Fake
template
Cr(VI)
2011
[36]
4 Journal of Environmental and Public Health
-
Tabl
e1:
Con
tinued.
No.
Ligand
sBa
ckgrou
ndsolutio
nWorking
electrod
eMeasurement
techniqu
esLO
D(pbb
)Analytical
object
Determined
form
Timeof
publication
References
36Po
lyviny
lbutyral/SPE
s+4.7%
DTP
AH
2SO4(pH
1)SP
Es(screen-
printedelectrod
e)CV-A
dSV
52.0
Fake
template
Cr(VI)
2014
[42]
372,5,8,11,14-Pentaaza-15,16,29-
phenanthrolin
ophane
(NeoTT
)1,6-Dichloro-hexane
(DCH),
LiCl1
0mM,H
Cl1
mM
Liqu
id/liqu
idinterface
SqW-A
dSV
250.0
Fake
template
Cr(VI)
2005
[40]
38
Silver:A
gClO
40.1mM,b
riton
robinson
(pH
2)gold:
HAuC
lO40.1mM,H
2SO4
0.5mM.
Carbo
nscreen-
printedelectrod
e(C
SPEs)
DPV
Silver:4
4.2
gold:2
0.8
Fake
template
Cr(VI)
2008
[43]
39Septon
ex10
−6 M
(1-
pentadecyltrim
ethylamon
ium
brom
ide)
HCl0
.25M,N
aCl0
.1M
(pH<2)
Carbo
npaste
DP-CSV
2.6
Tea
CrO
2− 42004
[41]
Journal of Environmental and Public Health 5
-
2.2. -e Working Electrode and Adsorptive Stripping Vol-tammetric Procedure. In this method, the working electrodeis bismuth film electrode created in in situ on glassy carbonrotating disk (in situ BiFE) and it is formed during thedeposition process in the following way: �e glassy carbondisk electrode was inserted into the electrochemical cellcontaining the reference electrode, platinum auxiliaryelectrode, and analysis solution (0.4.10−3M DTPA, 28.8.10−5M Bi(III), 5.0.10−6M KBr, 0.4M NaNO3, 0.4M acetatebuffer solution, and Cr(VI)).�e glassy carbon electrode wasrotated with constant speed and deposition at −800mV wasobserved (deposition voltage, Edep) at a definite time (de-position time, tdep); in the process, Bi(III) is reduced to Bi0which adheres to the surface of glassy carbon plate formingin situ BiFE; at the same time, Cr(VI) is reduced to Cr(III),and then new Cr(III) forms complexes with DTPA in thesolution layer close to the electrode surface and Cr(III)-DTPA complex adsorbed onto the surface of in situ BiFE, sochromium is enriched on the surface of the in situ BiFE [10].At the end of this period, the electrode stops rotating for30–60 seconds (equilibration time, tequal). Subsequently, thepotential scan was carried out in a negative potential di-rection from −800mV to −1450mV, and at the same time,the stripping voltammogram was recorded using a certainstripping voltammetry technique, differential pulse ad-sorptive stripping voltammetry (DP-AdSV), or square waveadsorptive stripping voltammetry (SqW-AdSV). During thisperiod, Cr(III) in the Cr(III)-DTPA complexes is reduced toCr(II) forming Cr(II)-DTPA complexes and generating thestripping peak current of chromium (Ip) [10]. If NO−3 is notpresent in solution, Ip will be very small, NO−3 present in thesolution will oxidize Cr(II)-DTPA to Cr(III)-DTPA andthen Cr(III)-DTPA is electrochemically reduced to Cr(II)-DTPA, and the repeated cycle increases the height of Ip [10].In other words, the NO−3 ion acts as a catalyst. After dis-solving, the electrode was cleaned by electrolysis at +400mVfor 30 seconds to dissolve Bio and other metals that may bepresent into the solution. Ip is proportional to the con-centration of Cr(VI) in the solution.
In all experiments, for Cr(VI) with trace, the firstmeasurement result must be discarded because it is unstable.�e stripping voltammogram was recorded 3 times (n� 3),and the peak current (Ip) and peak potential (Ep) values areaveraged from three repetitions.
�e glassy carbon electrodes were cleaned by polishingthe surface with fine Al2O3 powder (particle size 0.6 μm) andthen washed with distilled water and then with 1MNaOH toremove all Al2O3 particles on the glassy carbon surface, andthen the electrodes were dipped into 1M HCl solution andfinally washed with distilled water and the electrodes weredried with soft filter paper.
3. Results and Discussion
3.1. Differential Pulse Adsorptive Stripping Voltammetry (DP-AdSV) Using In Situ BiFE. In order to select the appropriateconditions for themethod, the experimental conditions werefixed as shown in Table S1. A univariate method is applied toexamine the effect of factors. �e magnitude of the stripping
peak current (Ip) and the relative standard deviation of the Ip(RSD) are used for selecting the appropriate test conditions.
3.1.1. Effect of Acetate Buffer Concentration. Acetate bufferis chosen to stabilize the pH of the solution. Acetate bufferis one of the factors that strongly influence the complex ofCr(III) and DTPA [10]. �e complex formation betweenCr(III) and DTPA usually occurs at pH� 6 [10, 11]. Atchromium concentration CCr(VI) � 3.8.10−8M, ligand con-centration CDTPA � 0.4.10−3M, and bismuth concentrationCBi(III) � 24.10−5M, the survey results of the effect of acetatebuffer concentration (CAc) in the range of 0.1M–0.6M(pH � 6) showed that CAc � 0.4M was appropriate. Withthis condition, the peak current is 31.36 μA and the re-peatability is relatively good (RSD� 1.6% with n� 3)(Figure 1(a)).
3.1.2. Effect of Bi(III) Concentration and DTPAConcentration. Previous studies with ex situ BiFE electrodeshave suggested that the presence of KBr in the solutionincreases the bismuth’s sustainability on the glassy disksurface and at the same time improves the conductivity ofthe solution [48]. In the acetate buffer with CAc � 0.4M andthe presence of KBr at a concentration of 5.0.10−6M,CDTPA � 0.4.10−3M, and CCr(VI) � 3.8.10−8M, a Bi(III) con-centration of 28.8.10−5M is appropriate. At those concen-trations, the peak current is 37.6 μA and the repeatability isgood (RSD� 1.7% with n� 3) (Figure 1(b)). �e surveyresults of the effect of DTPA concentrations in the range of0.1 to 1.0.10−3M to Cr peak current show that DTPAconcentration of 0.4.10−3M was appropriate (Figure 1(f)).
3.1.3. Effect of NaNO3 Concentration. In the presence ofNO−3 , peak current of chromium (Ip) was enhanced sig-nificantly. Many authors argue that NO−3 ions act as oxi-dizing agents that convert Cr(II)-DTPA complexes toCr(III)-DTPA complexes and thus increase the concentra-tion of Cr(III) on the electrode surface, and this leads to anincrease in Ip [10, 11, 21]. At CAc � 0.4M, CKBr � 5.0.10−6M,CDTPA � 0.4.10−3M, CBi(III) � 28.8 .10−5M, andCCr(VI) � 3.8.10−8M, Ip increased when the NaNO3 con-centration increased from 0.1M to 0.4M (Figures 1(e) and2(c)). However, when the NaNO3 concentration is greaterthan 0.4M, it increases the baseline and may contaminatethe analysis solution. When the NaNO3 concentration isequal to 0.4M, the peak current is 17.6 µA and the re-peatability is quite good (RSD� 3.7% with n� 2). NaNO3concentration value of 0.4M was selected for furtherinvestigation.
3.1.4. Effect of Deposition Potential (Edep) and Deposition Time(tdep). When CAc � 0.4M, CKBr � 5.0.10−6M, CCr(VI) � 2 ppb,CDTPA � 0.4.10−3M, and CBi(III) � 28.8.10−5M, the surveyresults of the effect of the deposition potential in the rangefrom −700mV to −1000mV are shown in Figure 1(c) andthe stripping voltammetry is shown in Figure 2(a).
6 Journal of Environmental and Public Health
-
From this result, it shows that when Edep is equal to−800mV, the peak current is 42.1 µA and the repeatability isgood (RSD� 1.3% with n� 3). Edep of −800mV was selectedfor further studies.
With the above conditions and when Edep is equal to−800mV, Ip is almost unchanged when the deposition time(tdep) is greater than 80 s, which means that it tends to reach
saturation (Figure 1(d)). tdep of 50 s is selected for the nextexperiment (the stripping voltammetry at tdep � 50 s isshown in Figure 2(b)).
3.1.5. Effect of Rotating Rate of Electrode (ω) and Equili-bration Time (tequal). By increasing the rotation speed of the
IpRSD
0
5
10
15
20
25
30
35I p
(μA
)
0.25 0.5 0.750CAc (M)
0
5
10
15
20
25
30
RSD
(%)
(a)
0
10
20
30
40
I p (μ
A)
148 248 34848CBi(III) (10–4M)
0
2
4
6
8
RSD
(%)
IpRSD
(b)
IpRSD
34
36
38
40
42
44
I p (μ
A)
–1.0 –0.9 –0.8 –0.7 –0.6–1.1Edep (V)
0
1
2
3
4
5
RSD
(%)
(c)
IpRSD
0
10
20
30
40
I p (μ
A)
40 50 60 70 80 90 100 11030tdep (s)
0
2
4
6
8
10
12
RSD
(%)
(d)
IpRSD
0
5
10
15
20
I p (μ
A)
0.1 0.2 0.3 0.4 0.50CNaNO3 (M)
0
2
4
6
8
10
12
RSD
(%)
(e)
IpRSD
0
2
4
6
8
I p (μ
A)
0.2 0.4 0.6 0.8 1 1.20CDTPA (10–3M)
0
1
2
3
4
5
RSD
(%)
(f )
Figure 1: Effect of acetate buffer concentration (a), Bi(III) concentration (b), deposition potential (c), deposition time (d), NaNO3concentration (e), and DTPA concentration (f) on the Ip of chromium in DP-AdSV using in situ BiFE. Experimental conditions are asmentioned in Sections 3.1.1 to 3.1.4, and other conditions are as shown in Table S1.
Journal of Environmental and Public Health 7
-
electrode to a specified value, it will increase the masstransfer and the efficiency of the enrichment will be better.�e survey results of the rotating rate of the electrode in therange from 800 rpm to 2400 rpm showed that ω of 2000 rpmwas appropriate. At the end of the enrichment phase, theelectrode should not be rotated for a specified period of timeto keep the solution quiet and the electrode surface is sta-bilized (this time is also called equilibration time, symbol-ized as tequal). Ip survey results according to tequal showed thattequal of 50 s was appropriate.
3.1.6. Effect of Cleaning Potential (Eclean) and Cleaning Time(tclean). �e cleaning of the surface of the carbon glassy discelectrode at the end of each stripping voltammetry is es-sential, as it will create repeating electrode surfaces for
subsequent measurements. In terms of experimental con-ditions as in Section 3.1.3, the survey results of the influenceof Eclean in the range of 200mV to 500mV and tclean in therange of 60 s to 120 s showed that Eclean is equal to 300mVand tclean is equal to 110 s which are appropriate.
3.2. Square Wave Adsorptive Stripping Voltammetry (SqW-AdSV) Using In Situ BiFE. Some authors argue that, inaddition to differential pulse stripping voltammetry tech-niques, square wave stripping voltammetry can be used torecord the signal (at this time, the method is called squarewave adsorptive stripping voltammetry (SqW-AdSV)) andalso allow the determination of very sensitive chromium.Based on the experimental conditions initially fixed asshown in Table S1, the effects of the factors were investigated
Cr
–70.0u
–80.0u
–90.0u
–100u
–110u
–120uI (
A)
–1100 –1300 –1400–1200–900 –1000–800E (mV)
(a)
Cr
–60.0u
–70.0u
–80.0u
–90.0u
I (A
)
–900 –1000 –1100 –1200 –1300 –1400–800E (mV)
(b)
Cr
0.4 M
0.3 M
0.2 M
0.1 M
Baseline–30.0u
–40.0u
–50.0u
–60.0u
–70.0u
I (A
)
–900 –1000 –1100 –1200 –1300 –1400–800E (mV)
(c)
Figure 2: �e stripping voltammetry DP-AdSV using in situ BiFE of chromium is recorded for (a) Edep � −800mV and (b) tdep � 50 s.(c) NaNO3 concentrations are changed: the bottom line is the baseline, followed by concentration of NaNO3 increasing from 0.1 to 0.4M.Other experimental conditions are as shown in Figure 1.
8 Journal of Environmental and Public Health
-
in a similar way to the DP-AdSV method, and we obtainedthe appropriate conditions for SqW-AdSV (using in situBiFE) to determine Cr(VI) as shown in Table S2.
3.3. Interferences. Interferences for the determination ofCr(VI) consist of metallic cations that have the strippingpeak current near the stripping peak current of chromiumand anions that can form complexes or make conjugateswith the forms of chromium and Bi(III) which can beadsorbed onto the surface of the in situ BiFE, and surfactantscan be adsorbed onto the working surface of the electrode.
�e influence of interferences can be estimated by rel-ative error values of stripping peak current (RE). Considerthat RE for Ip was equal to RE for C (because Ip � kC). RE forIp (or C) was accepted when it was equal to ½ Horwitzfunction RSD (RE Ip(Cr)≤½ RSDHorwitz �½.2(1−0.5lgC) � 32%with C� 0.2 ppb). RE was calculated as follows:
RE Ip(Cr)(%) �Ip(Cr) − Ip(Cr)
0
Ip(Cr)0∗100, (3)
where RE is the relative error values of stripping peakcurrent, Ip (Cr)0 is the stripping peak current without addinginterferences and Ip(Cr) is the stripping peak current withadding interferences.
3.3.1. Interference Studies. When the chromium(III) con-centration is about 100 times higher than the chromium(VI)concentration, the Ip does not change significantly(RE< 18%). In fact, rarely encountered CCr(III) case is 300times higher than CCr(VI), so it can be assumed that Cr(III)does not affect the Cr(VI) determination. �is investigationagain confirms that Cr(III) does not affect the determinationof Cr(VI) (Table S3 and Figure S3).
In the acetate buffer (pH� 5–6), Zn(II), Co(II),and Ni(II) can affect the determination of Cr(VI) because ithas a stripping peak current close to the stripping peakcurrent of Cr(VI). Ep (Zn)≈−1040÷−1050mV, Ep (Co)≈−1290÷−1300mV, Ep (Ni)≈−1080÷−1100mV, and Ep(Cr)≈−1180÷−1240mV.
From the experimental results at approximately 3.810−9M (0.2 ppb) chromium concentration, 120 s depositiontime, and the suitable conditions as in Table 2, we have seenthat Zn has not influenced the determination of chromiumwhen Zn(II) concentration is 800 times larger than Cr(VI)concentration (REIp(Cr)≤ 16%) (Table 2), and in conse-quence, we can determine chromium in the natural waterwith attendance of Zn(II) because ordinarily Zn(II) con-centration is 500 times smaller than Cr(VI) concentration innatural water (Table 2). Co and Ni have not influenced thedetermination of chromium when Co(II) and Ni(II) con-centrations are 90 times larger than Cr(VI) concentration(RE Ip(Cr) � 1.7–8.0% for Co(II) and 3.9–19.4% for Ni(II)(Table 2).
In natural water, Fe(III) and Ca(II) usually exist at highconcentrations of mM; in seawater, Ca(II) exists at quitehigh concentration, about 10−2÷10−3M, so it is necessary toexamine the effect of Fe(III) and Ca(II) on the stripping peak
current of Cr(VI).�e results in Table S4 show that when theconcentration of Fe(III) and Ca(II) increases to 36.10−6Mand 50.10−6M, respectively, meaning that the concentrationsof Fe(III) and Ca(II) are about 10,000 times greater than theconcentration of Cr(VI), the determination of Cr(VI) is notaffected. RE is less than 17%.
In natural water, Cl− and SO2−4 ions have significantconcentrations (mM and larger), and they can form com-plexes with metals present in the study solution and maytherefore affect the determination of Cr(VI). To investigatethe effects of Cl− and SO2−4 , a series of experiments wereconducted with Cl− concentrations ranging from 0 to281.7.10−3M and SO2−4 concentrations ranging from 0 to10.4.10−3M. �e results in Table S5 show that Cl− does notaffect the determination of Cr(VI) when CCl− is in the rangeof 0 to 14.09. 10−3M.When CCl− is greater than 28.17.10−3M(nearly equivalent to Cl− concentration in brackish water),Cl− affects the determination of Cr(VI) with RE> 32%.�erefore, when analyzing Cr(VI) in samples with highCl−concentration, it is necessary to take Cl− removal methodfrom the sample. SO2−4 did not affect the determination ofCr(VI) by SqW-AdSV/in situ BiFE method with RE< 27%.
In the adsorption stripping voltammetry method, thesurfactant can be adsorptive on the surface of the workingelectrode, and this can affect the adsorption process of themetallic complexes on the working electrode. Triton X-100(polyethylene glycol) is a typical nonionic surfactant andusually is used in order to observe the influence of thesurfactant on the adsorption stripping voltammetry method.�e effects of Triton X-100 are investigated at concentrationsbetween 0 and 93. 10−9M, and the results in Table S6 showthat, when increasing the concentration of Triton X-100 to25 times higher than the concentration of Cr(VI), it still didnot affect the determination of Cr(VI) with RE< 8%.
In fact, the concentration of natural surfactants is rarelygreater than 77.10−9M, and therefore, it can be assumed thatthey do not affect the Cr(VI) determination. �us, whendetermining Cr(VI) by the adsorption stripping voltam-metry method, it is not necessary to remove the surfactantfrom the analytical solution.
In some cases, natural water and wastewater contain manyorganic substances including surfactants. It is necessary totreat the sample to exclude organic substances before analysisusing UV irradiation and decomposition in acid mixture.
3.4. Evaluation of Reliability of DP-AdSV and SqW-AdSVMethods
3.4.1. Repeatability. Repeat recording of 7 stripping vol-tammetry lines (n� 7) on the same in situ BiFE according tothe DP-AdSV or SqW-AdSV method in Figure 3 shows thatIp in both approaches has good repeatability with RSD< 4%(n� 9) and RSD< 3% (n� 7), respectively, for DP-AdSV andSqW-AdSV. �e stripping peak of chromium (Ep) is neg-ligible, only about 20mV toward the positive side.
3.4.2. Linear Range and Detection Limits. �e linear rangeand LOD of the two methods SqW-AdSV and DP-AdSV
Journal of Environmental and Public Health 9
-
were investigated with the appropriate experimental con-ditions as shown in Table S2 and the stripping voltammetryspecifications as shown in Table S1, and the following resultswere obtained:
(1) -e Linear Range. For the SqW-AdSV method, Ip andCCr(VI) have a good linear correlation in the rangeCCr(VI) � 0.3÷1.8 ppb with a correlation coefficient (R) of 0.9994 (linear regression equation is shown in Figure 4(a), andthe stripping voltammetry is shown in Figure 4(b)).
For the DP-AdSV method, there is a good linear cor-relation in the range CCr(VI) � 2÷12 ppb with R� 0.9989(linear regression equation is shown in Figure 4(d)).
(2) Sensitivity.�e SqW-AdSVmethod achieved a sensitivity(23 µA/ppb) of about 34 times higher than the DP-AdSV(0.682 µA/ppb) method.
(3) Detection Limits and Quantitative Limits. For SqW-AdSV (when Edep � −800mV and tdep � 160 s):
LOD� 0.1 ppb; LOQ� 0.3 ppb. For DP-AdSV (whenEdep � −800mV and tdep � 50 s): LOD� 0.6 ppb;LOQ� 2 ppb.
�us, the SqW-AdSVmethod achieves a narrower linearrange than the DP-AdSV method, but it achieves highersensitivity than the DP-AdSVmethod (due to its lower LODand greater slope linearity). It can be said that with LOD asabove, DP-AdSV and SqW-AdSV methods can be used within situ BiFE to determine the trace amount of Cr(VI).
3.5. Determination of Chromium in Natural Water by theSqW-AdSV Using In Situ BiFE. In natural water samples,chromium usually exists in both Cr(VI) and Cr(III)forms. As investigated, the SqW-AdSV/in situ BiFEmethod identifies Cr(VI) and also determines the totalCr(VI) + Cr(III) if during the decomposition of thesample, an additional oxidizer is added to oxidize Cr(III)to Cr(VI). �us, we can determine chromium in indi-vidual forms by determining Cr(VI) (∗) and total
Table 2: Influence of Zn(II), Co(II), and Ni(II) concentrations on peak current.
Cation Zn(II) Co (II) Ni (II)No. CZn(II)(nM) Ip(Cr)(μA) RE Ip(Cr)(%) CCo(II)(nM) Ip(Cr)(μA) REIp(Cr)(%) CNi(II)(nM) Ip(Cr)(μA) RE Ip(Cr)(%)1 0 96.4 0 0 76.5 0 0 49.7 02 770 96.3 0.1 84 82.4 7.6 84 53.2 3.93 1540 94.6 1.9 168 82.5 7.9 168 58.3 10.44 2310 87.9 8.7 252 81.7 6.8 252 61.5 14.65 3080 80.9 16.0 336 77.8 1.7 336 62.4 19.4Conditions: CCr(VI) � 3.8.10−9M� 0.2 ppb; CBi(III) � 28.8.10−5M; tad � 120 s; Eclean � 400mV; tclean � 100 s; Ustep � 6mV; v � 210mV/s; ∆E� 30mV; f� 35Hz.;ω � 2000 rpm; CDTPA � 0.4mM; CAc � 0.4M; CNaNO3 � 0.4M; Ead � −800mV.
Cr
–80.0u
–100u
–120u
–140u
–160u
–180u
I (A
)
–900 –1000 –1100 –1200 –1300 –1400–800E (mV)
(a)
Cr
–40.0u
–50.0u
–60.0u
I (A
)
–900 –1000 –1100 –1200 –1300 –1400–800E (mV)
(b)
Figure 3:�e stripping voltammetry lines were repeated: (a) according to the SqW-AdSVmethod (n� 7) with CCr(VI) � 1 ppb; (b) accordingto the DP-AdSV method (n� 9) with CCr(VI) � 2 ppb. Other experimental conditions are as shown in Table S2.
10 Journal of Environmental and Public Health
-
Cr(VI) + (III) (∗∗). It follows that the Cr(III) content isthe difference of (∗∗) and (∗).
Based on the above results, it is possible to apply theSqW-AdSV/in situ BiFE to determine the trace of Cr(VI)
with LOD≈ 0.1 ppb. With that LOD, the SqW-AdSV/in situBiFE can directly determine the amount of Cr(VI) in naturalwater, without the stage of getting rich, and this is a greatadvantage of the SqW-AdSV/BiFE method.
Ip = 44,50 + 23,19 [Cr(VI)]R = 0,9994
30
40
50
60
70
80
90
100I p
(μA
)
0.4 0.8 1.2 1.6 20CCr(VI) (ppb)
(a)
1.8 ppb1.6 ppb1.4 ppb1.2 ppb1.0 ppb0.8 ppb0.6 ppb0.4 ppb0.2 ppb
nen
Cr
–50.0u
–75.0u
–100u
–125u
–150u
–175u
I (A
)
–900 –1000 –1100 –1200 –1300 –1400–800E (mv)
(b)
0
2
4
6
8
10
12
I P (μ
A)
2 4 6 8 10 12 14 16 18 20 220CCr(VI) (ppb)
(c)
Ip = 0,89 + 0,68 [Cr(VI)]R = 0,9989
0
2
4
6
8
10
12
I p (μ
A)
2 4 6 8 10 12 140CCr(VI) (ppb)
(d)
Figure 4: (a) Linear regression line for the SqW-AdSV method; (b) stripping voltammetry of SqW-AdSV method: the bottom line is thebaseline, followed by nine additional standard lines, each adding 0.2 ppb; (c) relation between Ip andCCr(VI) when examining the linear rangeof the DP-AdSV method; (d) linear regression line for the DP-AdSV method. Experimental conditions are as shown in Table S2.
Table 3: Appropriate experimental conditions for the SqW-AdSV/in situ BiFE for the determination of Cr(VI).
No. Parameter (unit of measure) Symbol SqW-AdSV/in situ BiE1 DTPA concentration (M) CDTPA 0.4. 10−3
2 Concentration of acetate buffer (pH� 6) (M) CAc 0.403 NaNO3 concentration (M) CNaNO3 0.404 KBr concentration (M) CKBr 5.10−6
5 Bi(III) concentration (M) CBi(III) 28.8. 10−5
6 Cleaning potential (mV) Eclean 3007 Cleaning time (s) tclean 1008 Rotating speed of working electrode (rpm) (ω) 2000 20009 Deposition potential (mV) Edep -80010 Deposition time (s) tdep 20011 Equilibration time (s) tequal 5012 Potential sweep range (mV) Erange −800÷−1450
13
Technical parameters• Amplitude (mV) ∆E 30
• Voltage step (mV) Ustep 6• Sweep rate (mV/s) v 210
• Frequency (Hz) f 35
Journal of Environmental and Public Health 11
-
In order to answer the question of whether the analyticalmethod used to analyze the amount of chromium in naturalwater samples can be applied, we have conducted experi-ments to verify the correctness (through the CertifiedReference Material (CRM)) and an analysis of some naturalwater samples. On the basis of the experiments mentionedabove, the analysis process of Cr(VI) and total Cr(III, VI) innatural water by the SqW-AdSV method was proposed.
3.5.1. Quality Control of Analytical Methods through Stan-dard Sample Analysis. In order to confirm the practicalapplicability of the SqW-AdSV method to analyze chro-mium traces using BiFE electrodes, it is necessary to controlthe analytical method quality by evaluating the accuracy andrepeatability when analyzing standard samples.
(1) For Surface Water Samples. Surface water CertifiedMaterial Reference (SPS-SW1 Batch 122) was selected toevaluate the accuracy of the method. �e actual value of thechromium content of the sample is 2.00± 0.02 ppb (95%confidence boundary e�± 0.02 ppb). Analysis of standardSPS-SW1 surface water (CRM) samples by the SqW-AdSVusing in situ BiFE with the appropriate experimental con-ditions is shown in Table 3. �e analysis was repeated 3times. �e volume of the solution to be charged to theelectrolyser is 2mL, and the volume of solution in theelectrolyser is 10mL.
�e results in Table 4 show that the SqW-AdSV/in situBiFE has good repeatability (the standard deviation is 4% for
the repetition of 3, and it is less than half the standarddeviation based on the Horwitz function (RSDH � 2(1−0,5lgC);when the chromium concentration is 2 ppb, the standarddeviation of the Horwitz equation is 41%) [49] and themethod has good accuracy because the chromium content iswithin the 95% confidence interval of the CRM sample.
(2) For Seawater Samples. Analysis of the standard seawaterCRM coded NASS 6 by the SqW-AdSV/in situ BiFE with theappropriate experimental conditions as shown in Table 3.
Because the concentration of chromium in NASS 6seawater was too small to be directly analyzed, only NASS 6standard sample was used as the matrix for analysis andvalidity. �e actual value of the chromium content in theNASS 6 sample is 0.116± 0.008 ppb (the 95% confidencebound e�± 0.008 ppb). NAAS 6 standard sample was addedwith standard Cr(VI) solution to attain 3 levels of 2 ppb,6 ppb, and 10 ppb and then analyzed with the standardadded samples to determine recovery.
�e results showed that the SqW-AdSV/in situ BiFE forchromium analysis in seawater samples has a good accuracy(recoverability from 94 to 109%). According to the AOAC(American Association of Analytical Chemists) when ana-lyzing the levels of 1.0 to 10 ppb, achieving a recovery rate of80 to 110% is acceptable [50]. �erefore, it is possible to usethis method to analyze chromium in seawater samples(Table 5).
�e results of the linear range, sensitivity, limit ofdetection, and accuracy showed that it is possible to use
Table 4: Accuracy of the SqW-AdSV/in situ BiFE for the determination of chromium in surface water.
Information [Cr(VI)](ppb) CCr (ppb)
Experiment1 0.38 1.902 0.40 2.003 0.40 2.00
Average± S (ppb) 1.97± 0.08Cr content in the CRM sample (ppb) 2.00± 0.02 (CCr � 1.98÷ 2.02 ppb)RSD (%), n� 3 4(a)[Cr(VI)] is the concentration of Cr(VI) in the electrolyte minus the blank. White sample has [Cr(VI)]� 0.034 ppb; CCr is the Cr content in the sample(calculated by the formula: CCr � [Cr(VI)] .V2/V1). V1: volume of solution taken into the electrolyser (V1 � 2ml), V2: volume of solution in the electrolyser(V2 �10mL), S is the standard deviation. Experimental conditions are as shown in Table 3.
Table 5: Determination of the accuracy of the SqW-AdSV/in situ BiFE on the NASS 6a.
[NAAS6]
�e content of chromium in thesample (ppb) ×1
Chromium standard added(ppb)×o
�e content of chromium in standard addedsamples (ppb) ×2
Recovery(%)
2 ppb 0.116 1.8841.915 962.043 1021.941 97
Average± S 1.966± 0.054
6 ppb 0.116 5.8845.958 996.258 1045,655 94
Average± S 5.957± 0.213
10 ppb 0.116 9.88410.687 10710.125 10110.887 109
Average± S 10.566± 0.279aRecovery� (x2−x1).·100/x0; S is the standard deviation; experimental conditions are as shown in Table 3.
12 Journal of Environmental and Public Health
-
SqW-AdSV/in situ BiFE to determine chromium in surfacewater and seawater.
3.5.2. Real Sample Analysis. For the purpose of testing thepossibility of applying the SqW-AdSV/in situ BiFE method forthe analysis of chromium in water environment, well water, tapwater, lagoon water, and seawater in some different areas in�ua �ien Hue province were taken for analysis.
Water samples were taken in clean PET bottles and acidifiedwith concentrated HCl (500μl HCl/500ml of sample). Sampleswere filtered through 0.45μm porous fiberglass filter paper andanalyzed immediately after filtration.
Collected and stored samples were analyzed directly(after filtration through a 0.45 μm porous fiberglass filterpaper) to determine the Cr(VI) content by the SqW-AdSV/in situ BiFE; the total chromium content of the sample wasdetermined after the decomposition of the sample by themethod of (a) ((a): add 50 μl of concentrated HCl, 25 μl ofH2O2 35% to 50ml of sample in the Teflon cup, boil for 90minutes, let it cool, and adjust up to 25ml) [51].
Samples were analyzed by SqW-AdSV/in situ BiFEaccording to the process shown in Figure S2. �e results ofactual sample analysis are presented in Table 6.
4. Conclusion
Using in situ BiFE electrodes with DTPA complexing ligands inacetate buffer solution pH 6 with the presence of KBr and NO−3ion, DP-AdSV and SqW-AdSV methods can determinechromium(VI) concentrations of 0.3 ppb and 2.0 ppb, respec-tively. �e proposed method has been successfully applied forchromium analysis in some natural water samples such aslagoon water, well water, tap water, and saltwater in some areas
of �ua�ien Hue province, Vietnam. �is CCr(VI+III). analysisprocedure in water sample by SqW-AdSV/in situ BiFE methodwas satisfactorily applied for the determination of chromium inreal water such as tap water, river water, and well water samplesin all countries of the world.�e determination of chromium inthe above real water sample could be carried out within 60min.
Data Availability
�e data used to support the findings of this study areavailable from the corresponding author upon request.
Conflicts of Interest
�e authors declare that they have no conflicts of interest.
Supplementary Materials
Table S1: for the DP-AdSV and SqW-AdSV using in situBiFE. Table S2: suitable experimental conditions for DP-AdSV and SqW-AdSV methods using in situ BiFE. Table S3:influence of chromium(III). Figure S1: SqW-AdSV/BiFE insitu stripping voltammograms of chromium(VI) when ex-amining the effects of chromium(III). Table S4: influence ofFe(III) and Ca(II). Table S5: influence of Cl− and SO42−.Table S6: influence of Triton X-100. Figure S2: diagram ofCCr(VI + III) analysis procedure in water sample by SqW-AdSV/in situ BiFE method. . (Supplementary Materials)
References
[1] Y. Li and H. Xue, “Determination of Cr (III) and Cr (VI)species in natural waters by catalytic cathodic strippingvoltammetry,” Analytica Chimica Acta, vol. 448, no. 1-2,pp. 121–134, 2001.
Table 6: Chromium content in lagoon water samples, tap water, well water, and saltwater.
No Sample type Sample symbolChromium concentration
(Cmean± ε) ppb, n� 3, P� 0.95CCr(VI + III) CCr(VI)
1
Water sample of Cau Hai Lagoon
M1 13.8± 0.2 1.0± 0.12 M2 19.0± 1.0 1.0± 0.23 M3 7.3± 0.4 1.5± 0.24 M4 26.1± 5.8 1.6± 0.35 M5 14.1± 0.8 1.3± 0.76 M6 1.0± 0.1 0.7± 0.27 M7 11.1± 4.4 0.8± 0.38
Tap waterPTN 20.0± 2.3
9 GÐ 19.2± 3.310 GÐ1 18.1± 0.811
Well water
G1 28.6± 1.012 G1′ 24.6± 4.013 G2 22.3± 4.214 G2′ 12.5± 4.815 G3 6.4± 0.616 G3′ 14.4± 0.517 G4 13.6± 3.318 G4′ 21.2± 4.419 Saline water B1 1.3± 0.3 1.0± 0.220 B2 16.1± 1.3 12.1± 1.4
Journal of Environmental and Public Health 13
http://downloads.hindawi.com/journals/jeph/2020/1347836.f1.pdf
-
[2] A. Bobrowski, A. Królicka, and J. Zarębski, “Characteristics ofvoltammetric determination and speciation of chromium—areview,” Electroanalysis, vol. 21, no. 13, pp. 1449–1458, 2009.
[3] B. R. James, J. C. Petura, R. J. Vitale, and G. R. Mussoline,“Hexavalent chromium extraction from soils: a comparison offive methods,” Environmental Science & Technology, vol. 29,no. 9, pp. 2377–2381, 1995.
[4] R. J. Vitale, G. R. Mussoline, J. C. Petura, and B. R. James,“Hexavalent chromium extraction from soils: evaluation of analkaline digestion method,” Journal of Environmental Quality,vol. 23, no. 6, pp. 1249–1256, 1994.
[5] U. Divrikli, M. Soylak, and L. Elci, “Determination of totalchromium by flame atomic absorption spectrometry aftercoprecipitation by cerium (IV) hydroxide,” EnvironmentalMonitoring and Assessment, vol. 138, no. 1-3, pp. 167–172,2008.
[6] L. Yao, Y. Zhu, W. Xu et al., “Combination of dispersive solidphase extraction with dispersive liquid-liquid microextractionfor the sequential speciation and preconcentration of Cr (III)and Cr (VI) in water samples prior to graphite furnace atomicabsorption spectrometry determination,” Journal of Industrialand Engineering Chemistry, vol. 72, pp. 189–195, 2019.
[7] H. Sereshti, V. Khojeh, and S. Samadi, “Optimization ofdispersive liquid-liquid microextraction coupled with in-ductively coupled plasma-optical emission spectrometry withthe aid of experimental design for simultaneous determina-tion of heavy metals in natural waters,” Talanta, vol. 83, no. 3,pp. 885–890, 2011.
[8] Q.-y. Zhu, L.-y. Zhao, D. Sheng et al., “Speciation analysis ofchromium by carboxylic group functionalized mesoporoussilica with inductively coupled plasma mass spectrometry,”Talanta, vol. 195, pp. 173–180, 2019.
[9] A. J. Laghari, M. Y. Khuhawar, and L. A. Zardari, “Capillarygas chromatographic analysis of pyrrolidinedithiocarbamatemetal chelates,” Journal of Separation Science, vol. 30, no. 16,pp. 2727–2732, 2007.
[10] L. Yong, K. C. Armstrong, R. N. Dansby-Sparks,N. A. Carrington, J. Q. Chambers, and Z.-L. Xue, “Quanti-tative Analysis of trace Chromium in blood samples. Com-bination of the advanced oxidation Process with catalyticadsorptive stripping voltammetry,” Analytical Chemistry,vol. 78, no. 21, pp. 7582–7587, 2006.
[11] L. Lin, N. S. Lawrence, S. �ongngamdee, J. Wang, Y. Wang,and Y. Lin, “Catalytic adsorptive stripping determination oftrace chromium (VI) at the bismusth film electrode,” Talanta,vol. 65, pp. 144–148, 2005.
[12] B. Bas’, “Refreshable mercury film silver based electrode fordetermination of chromium (VI) using catalytic adsorptivestripping voltammetry,” Analytica Chimica Acta, vol. 570,pp. 195–201, 2006.
[13] N. H. Du, N. T. T. Luyen, N. K. Manh et al., “Determination ofppt level chromium (VI) using the gold nano-flakes elec-trodeposited on platinum rotating disk electrode and mod-ified with 4-thiopyridinium,” American Journal of AnalyticalChemistry, vol. 06, no. 05, pp. 457–467, 2015.
[14] E. Chatzitheodorou, A. Economou, and A. Voulgaropoulos,“Trace determination of chromium by square-wave adsorp-tive stripping voltammetry on bismuth film electrodes,”Electroanalysis, vol. 16, no. 21, pp. 1745–1754, 2004.
[15] H. Han, D. Pan, D. Liu, and X. Hu, “Cathodic strippingvoltammetric determination of chromium in coastal waterson cubic Nano-titanium carbide loaded gold nanoparticlesmodified electrode,” Frontiers inMarine Science, vol. 75, no. 2,pp. 1–7, 2015.
[16] E. O. Jorge, M.M. Rocha, I. T. E. Fonseca, andM.M. M. Neto,“Studies on the stripping voltammetric determination andspeciation of chromium at a rotating–disc bismuth filmelectrode,” Talanta, vol. 81, pp. 556–564, 2010.
[17] Q. Zhang, S.-wei Zhong, J.-li Su, X.-jun Li, and H. Zou,“Determination of trace chromium by square-wave adsorp-tive cathodic stripping voltammetry at an improved bismuthFilm Electrode,” -e Electrochemical Society, vol. 160, no. 4,pp. 237–242, 2013.
[18] R. Ouyang, W. Zhang, S. Zhou et al., “Improved Bi filmwrapped single walled carbon nanotubes for ultrasensitiveelectrochemical detection of trace Cr (VI),” ElectroChimicaActa, vol. 113, pp. 686–693, 2013.
[19] A. S. Misiego, R. M. G. Carra, P. A. Carracedo, andM. E. M. Torre, “Determination of Cr (in small quantities) byadsorptive stripping voltammetry: a comparative study ofsquare wave versus differential pulse,” Analytical and Bio-analytical Chemistry, vol. 376, no. 7, pp. 1119–1125, 2003.
[20] S. Sadeghi and A. Garmroodi, “A highly sensitive and selectiveelectrochemical sensor for determination of Cr (VI) in thepresence of Cr (III) using modified multi-walled carbonnanotubes/quercetin screen-printed electrode,” MaterialsScience and Engineering: C, vol. 33, no. 8, pp. 4972–4977, 2013.
[21] K. Torrance and C. Gatford, “Determination of solublechromium in simulated P.W. R coolant by differential pulseadsorptive stripping voltammetry,” Talanta, vol. 34,pp. 934–944, 1987.
[22] A. Bobrowski, B. Ba´s, J. Dominik, and E. Niewiara,“Chromium speciation study in polluted waters using cata-lytic adsorptive stripping voltammetry and tangential flowfiltration,” Talanta, vol. 63, no. 4, pp. 1003–1012, 2004.
[23] C. Elleouet, F. Quentel, and C. Madec, “Determination oftrace amounts of chromium (VI) in water by electrochemicalmethods,” Analytica Chimica Acta, vol. 257, no. 2,pp. 301–308, 1992.
[24] D. V. Vukomanovic, G. W. vanLoon, K. Nakatsu, andD. E. Zoutman, “Determination of chromium (VI) and (III)by adsorptive stripping voltammetry with pyrocatechol vio-let,” Microchemical Journal, vol. 57, no. 1, pp. 86–95, 1997.
[25] E. Panaščikaitė, I. Latvėnaitė, and S. Armalis, “Determinationof chromium in cement by catalytic adsorptive strippingvoltammetry,” Chemija, vol. 22, no. 4, pp. 210–215, 2011.
[26] L. Husakova, A. Bobrowski, J. Sramkova, A. Krolicka, andK. Vytras, “Catalytic adsorptive stripping voltammetry versuselectrothermal atomic absorption spectrometry in the de-termination of trace cobalt and chromium in human urine,”Talanta, vol. 66, no. 4, pp. 999–1004, 2005.
[27] J. Kiptoo, J. C. Ngila, and G. M. Sawula, “Speciation studies ofnickel and chromium in wastewater from an electroplatingplant,” Talanta, vol. 64, no. 1, pp. 54–59, 2004.
[28] M. Korolczuk, “How faster and cheaper to determine chro-mium by adsorptive cathodic stripping voltammetry in thepresence of DTPA and nitrate,” Fresenius Journal of Ana-lytical Chemistry, vol. 367, no. 8, pp. 761-762, 2000.
[29] M. Goldoni, A. Caglieri, D. Poli et al., “Determination ofhexavalent chromium in exhaled breath condensate andenvironmental air among chrome plating workers,” AnalyticaChimica Acta, vol. 562, no. 2, pp. 229–235, 2006.
[30] O. Dom´ınguez and M. Julia Arcos, “Simultaneous deter-mination of chromium (VI) and chromium (III) at trace levelsby adsorptive stripping voltammetry,” Analytica ChimicaActa, vol. 470, pp. 241–252, 2002.
14 Journal of Environmental and Public Health
-
[31] M. Palrecha and P. K. Mathur, “Adsorptive stripping vol-tammetric determination of chromium in gallium,” Talanta,vol. 45, no. 2, pp. 433–436, 1997.
[32] F. Scholz, B. Lange, M. Draheim, and J. Pelzer, “�e catalyticadsorptive stripping voltammetric determination of chro-mium with DTPA and nitrate,” Fresenius’ Journal of Ana-lytical Chemistry, vol. 338, no. 5, pp. 627–629, 1990.
[33] S. Abbasi and A. Bahiraei, “Ultra trace quantification ofchromium (VI) in food and water samples by highly sensitivecatalytic adsorptive stripping voltammetry with rubeanicacid,” Food Chemistry, vol. 133, no. 3, pp. 1075–1080, 2012.
[34] V. Arancibia, E. Nagles, M. Gómez, and C. Rojas, “Speciationof Cr (VI) and Cr (III) in water samples by adsorptivestripping voltammetry in the presence of pyrogallol red ap-plying a selective accumulation potential,” InternationalJournal of Electrochemical Science, vol. 7, pp. 11444–11455,2012.
[35] M. Korolczuk and M. Grabarczyk, “Voltammetric determi-nation of Cr (VI) in a flow system in the presence ofdiethylenetriaminepentaacetic acid (DTPA) following itsdeposition in the metallic state,” Analytica Chimica Acta,vol. 387, no. 1, pp. 97–102, 1999.
[36] J. Saturno, D. Valera, H. Carrero, and L. Fernández, “Elec-troanalytical detection of Pb, Cd and traces of Cr at micro/nano-structured bismuth film electrodes,” Sensors and Ac-tuators B: Chemical, vol. 159, no. 1, pp. 92–96, 2011.
[37] R. Torabi Kachoosangi and R. G. Compton, “Voltammetricdetermination of Chromium (VI) using a gold film modifiedcarbon composite electrode,” Sensors and Actuators B,vol. 178, pp. 555–562, 2014.
[38] R. Ouyang, S. A. Bragg, J. Q. Chambers, Z.-L. Xue, andZi-L. Xue, “Flower-like self-assembly of gold nanoparticles forhighly sensitive electrochemical detection of chromium (VI),”Analytica Chimica Acta, vol. 722, pp. 1–7, 2012.
[39] S. A. Tukur, N. A. Yusof, R. Hajian, and N. R. Hajian, “Linearsweep anodic stripping voltammetry: determination ofChromium (VI) using synthesized gold nanoparticles mod-ified screen-printed electrode,” Journal of Chemical Sciences,vol. 127, no. 6, pp. 1075–1081, 2015.
[40] A. M. OMahony, M.´l D. Scanlon, A. Berduque et al.,“Voltammetry of chromium (VI) at the liquid|liquid inter-face,” Electrochemistry Communications, vol. 7, pp. 976–982,2005.
[41] I. Svancara, P. Foret, and K. Vytras, “A study on the deter-mination of chromium as chromate at a carbon paste elec-trode modified with surfactants,” Talanta, vol. 64, no. 4,pp. 844–852, 2004.
[42] S. A. Miscoria, C. Jacq, T. Maeder, and R. Martin Negri,“Screen-printed electrodes for electroanalytical sensing, ofchromium (VI) in strong acid media,” Sensors and ActuatorsB: Chemical, vol. 195, pp. 294–302, 2012.
[43] O. Dom´ınguez-Renedo, L. Ruiz-Espelt, N. Garc´ıa-Astor-gano, and M. Julia Arcos-Mart´ınez, “Electrochemical de-termination of chromium (VI) using metallic nanoparticle-modified carbon screen-printed electrodes,” Talanta, vol. 76,pp. 854–858, 2008.
[44] S. Wyantuti, Y. W. Hartati, M. L. Firdaus, C. Panatarani, andR. Tjokronegoro, “Fabrication of gold nanoparticles -modifiedglassy carbon electrode and its application for voltammetricdetection of Cr (III),” International Journal of Scientific andTechnology Research, vol. 4, no. 1, pp. 135–139, 2015.
[45] M. Boussemart, C. M. G. van den Berg, and M. Ghaddaf, “�edetermination of the chromium speciation in sea water using
catalytic cathodic stripping voltammetry,” Analytica ChimicaActa, vol. 262, no. 1, pp. 103–115, 1992.
[46] L. Narayana, A. Suvarapu, R. Somala, P. Bobbala, I. Hwang,and V. Reddy Ammireddy, “Simultaneous spectrophoto-metric determination of chromium (VI) and vanadium (V) byusing 3,4-dihydroxybenzaldehyde isonicotinoyl hydrazone(3,4-DHBINH),” E–Journal of Chemistry, vol. 6, pp. 459–465,2009.
[47] M. R. Khan and S. B. Khoo, “Optimization of the simulta-neous batch determinations of Bi (iii), Hg (ii) and Cu (ii) at anepoxy-graphite electrode bulk modified with 2-mercapto-benzothiazole,” -e Analyst, vol. 126, no. 12, pp. 2172–2177,2001.
[48] E. A. Hutton, S. B. Hočevar, and B. Ogorevc, “Ex situpreparation of bismuth film microelectrode for use in elec-trochemical stripping microanalysis,” Analytica ChimicaActa, vol. 537, no. 1-2, pp. 285–292, 2005.
[49] N. G. Naseri, S. J. Baldock, A. Economou, J. Nicholas, andP. R. F. Goddard, “Disposable electrochemical flow cells forcatalytic adsorptive stripping voltammetry CAdSV at a bis-muth film electrode,” Analytical Bioanalytical Chemistry,vol. 391, pp. 1283–1292, 2008.
[50] I. Taverniers, M. De Loose, and E. Van Bockstaele, “Trends inquality in the analytical laboratory. II. Analytical methodvalidation and quality assurance,” Trends in AnalyticalChemistry, vol. 23, no. No.8, pp. 535–555, 2004.
[51] J. R. Dean, Method for Environment Trace Analysis, pp. 99–105, John Wiley, Hoboken, NJ, USA, 2003.
Journal of Environmental and Public Health 15