18
Applied Spectroscopy Determination of the experimental error of high spectral resolution infrared observations from spectral residuals: application to IASI C. Serio 1 , G. Masiello 1 , C. Camy-Peyret 2 , E. Jacquette 3 , O. Vandermarcq 3 , F. Bermudo 3 , D. Coppens 4 , D. Tobin 5 1 School of Engineering, University of Basilicata, Potenza, Italy 2 IPSL (UPMC/UVSQ) Paris, France 3 CNES, Toulouse, France 4 EUMETSAT, Darmstadt, Germany 5 CIMMS/SSEC, University of Wisconsin, Madison, USA 29 November - 5 December 2017 ITSC-XXI Darmstadt, Germany, 1

Upload: others

Post on 02-Aug-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Determination of the experimental error of high spectral ... · The FM and PCA/BIC approaches complete each other, the PCA/BIC is much faster. There are a lot of lessons learned by

Applied Spectroscopy

Determination of the experimental error of high spectral resolution infrared observations

from spectral residuals: application to IASI

C. Serio1, G. Masiello1, C. Camy-Peyret2, E. Jacquette3, O. Vandermarcq3, F. Bermudo3, D. Coppens4, D. Tobin5

1School of Engineering, University of Basilicata, Potenza, Italy2IPSL (UPMC/UVSQ) Paris, France3CNES, Toulouse, France4EUMETSAT, Darmstadt, Germany5CIMMS/SSEC, University of Wisconsin, Madison, USA

29 November - 5 December 2017ITSC-XXI Darmstadt, Germany, 1

Page 2: Determination of the experimental error of high spectral ... · The FM and PCA/BIC approaches complete each other, the PCA/BIC is much faster. There are a lot of lessons learned by

Background

1. The problem of how to characterize instrument noise directly fromobserved radiances has received a deal of attention in the past years.Most studies address the problem with heuristic or empirical approaches.

2. In contrast, our analysis is based on a firm statistical approach, whichleaves no room to ad hoc assumptions.

3. Our methodology does not address the problem of forward model noise(that is the noise due to instrument and forward model uncertainty andbias), which is the topics of many concurrent studies.

4. We are focused on the instrument noise alone.

29 November - 5 December 2017ITSC-XXI Darmstadt, Germany, 2

Page 3: Determination of the experimental error of high spectral ... · The FM and PCA/BIC approaches complete each other, the PCA/BIC is much faster. There are a lot of lessons learned by

We are focused on the instrument noise alone and have developed two tools, which will be exemplified through application to IASI

A forward model based approach A PCA/BIC based approach

ITSC-XXI Darmstadt, Germany, 29 November - 5 December 2017 3

Page 4: Determination of the experimental error of high spectral ... · The FM and PCA/BIC approaches complete each other, the PCA/BIC is much faster. There are a lot of lessons learned by

Both Approaches rely on the concept of spectral residuals

FM based approach 𝛿𝛿𝑹𝑹 = 𝑹𝑹𝑜𝑜𝑜𝑜𝑜𝑜 − 𝐹𝐹(𝒗𝒗), 𝐹𝐹, the forward model 𝒗𝒗, the state vector estimated based on a

suitable retrieval procedure

After a proper localization, the instrument noise covariance matrix is estimated by

�𝑺𝑺𝜀𝜀 =< (𝛿𝛿𝑹𝑹 − 𝛿𝛿𝑹𝑹)(𝛿𝛿𝑹𝑹 − 𝛿𝛿𝑹𝑹)𝒕𝒕 >

PCA approach After appropriate centering of the radiance

vector, 𝛿𝛿𝑹𝑹 = 𝑹𝑹𝑜𝑜𝑜𝑜𝑜𝑜 − 𝑼𝑼𝜏𝜏𝒄𝒄 𝑼𝑼 orthogonal basis of PCA 𝒄𝒄 vector,

truncated at 𝜏𝜏 For a given 𝜏𝜏 and a-priori normalizing

covariance, �𝑺𝑺𝜀𝜀, the instrument noise covariance is estimated by

�𝑺𝑺𝜀𝜀 = �𝑺𝑺𝜀𝜀1/2𝑼𝑼−𝜏𝜏𝜦𝜦𝑼𝑼−𝜏𝜏

𝑡𝑡/2�𝑺𝑺𝜀𝜀𝑡𝑡/2

with 𝜦𝜦 the diagonal matrix of eigenvalues, �𝑺𝑺𝜀𝜀 is the full IASI covariance matrix

according to the EUMETSAT/CNES release.

ITSC-XXI Darmstadt, Germany, 29 November - 5 December 2017 4

Page 5: Determination of the experimental error of high spectral ... · The FM and PCA/BIC approaches complete each other, the PCA/BIC is much faster. There are a lot of lessons learned by

For both approaches, we use the whole IASI spectral coverage (8461 channels).The FM approach is combined with Random Projections and we simultaneouslyretrieve parameters and gas species. (Poster 8p.02)

Surface Temperature Emissivity

(spectrum)

29 November - 5 December 2017 ITSC-XXI Darmstadt, Germany, 5

Atmospheric Profiles of Temperature H2O O3

HDO CO2

N2O CO CH4

SO2

HNO3

NH3

OCS CF4

Retrieved parameters and species include

Page 6: Determination of the experimental error of high spectral ... · The FM and PCA/BIC approaches complete each other, the PCA/BIC is much faster. There are a lot of lessons learned by

The PCA approach estimator is fully analytical, we do not need for each truncation point 𝜏𝜏 to re-compute the spectral residuals. However it critically depends on the given 𝜏𝜏 and the estimator is biased

ITSC-XXI Darmstadt, Germany, 29 November - 5 December 2017 6

PCA approach; application to IASI A: simulation

Page 7: Determination of the experimental error of high spectral ... · The FM and PCA/BIC approaches complete each other, the PCA/BIC is much faster. There are a lot of lessons learned by

Develop an approach, which simultaneously and optimally estimates noise and truncation point

ITSC-XXI Darmstadt, Germany, 29 November - 5 December 2017 7

𝒙𝒙 = 𝑯𝑯𝑯𝑯 + 𝜼𝜼

ℒ =𝑵𝑵𝟐𝟐 𝑑𝑑log 𝟐𝟐𝝅𝝅 + log 𝑺𝑺𝒙𝒙 + 𝐭𝐭𝐭𝐭(𝑺𝑺𝒙𝒙−𝟏𝟏𝑺𝑺)

𝑩𝑩𝑩𝑩𝑩𝑩 = −𝟐𝟐 log𝒑𝒑 𝒙𝒙 𝝉𝝉

Step 1: Write the PCA problem in a probabilistic fashion using the formalism of latent variables

Step 2: Define a log-likelihood cost function

Step 3: Compute the Bayesian Information Criterion, BIC and get the optimal solution by seeking for BIC minimum Gideon Schwarz, Estimating the

Dimension of a Model, Ann. Statist. Volume 6, Number 2 (1978), 461-464.

Page 8: Determination of the experimental error of high spectral ... · The FM and PCA/BIC approaches complete each other, the PCA/BIC is much faster. There are a lot of lessons learned by

PCA/BIC approach, application to IASI in simulation

Seek for BIC minimumEstimated radiometric noise at the BIC minimum, 𝝉𝝉 = 𝟐𝟐𝟐𝟐

ITSC-XXI Darmstadt, Germany, 29 November - 5 December 2017 8

Page 9: Determination of the experimental error of high spectral ... · The FM and PCA/BIC approaches complete each other, the PCA/BIC is much faster. There are a lot of lessons learned by

Application to IASI: IASI data in external calibration (ExtCal) mode for the year 2016

ITSC-XXI Darmstadt, Germany, 29 November - 5 December 2017 9

Page 10: Determination of the experimental error of high spectral ... · The FM and PCA/BIC approaches complete each other, the PCA/BIC is much faster. There are a lot of lessons learned by

IASI-A, PCA/BIC perfectly matches BB in flight calibration. Data averaged over the whole year 2016

ITSC-XXI Darmstadt, Germany, 29 November - 5 December 2017 10

Page 11: Determination of the experimental error of high spectral ... · The FM and PCA/BIC approaches complete each other, the PCA/BIC is much faster. There are a lot of lessons learned by

IASI-B, PCA/BIC perfectly matches BB in flight calibration. Data averaged over the whole year 2016

ITSC-XXI Darmstadt, Germany, 29 November - 5 December 2017 11

Page 12: Determination of the experimental error of high spectral ... · The FM and PCA/BIC approaches complete each other, the PCA/BIC is much faster. There are a lot of lessons learned by

IASI-B, PCA/BIC perfectly matches BB in flight calibration. Data averaged over the whole year 2016. NEDT Analysis

ITSC-XXI Darmstadt, Germany, 29 November - 5 December 2017 12

Page 13: Determination of the experimental error of high spectral ... · The FM and PCA/BIC approaches complete each other, the PCA/BIC is much faster. There are a lot of lessons learned by

Radiometric noise as a function of month. The case of IASI-A

ITSC-XXI Darmstadt, Germany, 29 November - 5 December 2017 13

Page 14: Determination of the experimental error of high spectral ... · The FM and PCA/BIC approaches complete each other, the PCA/BIC is much faster. There are a lot of lessons learned by

Radiometric noise as a function of month. The case of IASI-A. PCA/BIC can track differences as small as that due to icing

ITSC-XXI Darmstadt, Germany, 29 November - 5 December 2017 14

Page 15: Determination of the experimental error of high spectral ... · The FM and PCA/BIC approaches complete each other, the PCA/BIC is much faster. There are a lot of lessons learned by

COVARIANCE MATRIX, IASI-A

ITSC-XXI Darmstadt, Germany, 29 November - 5 December 2017 15

Page 16: Determination of the experimental error of high spectral ... · The FM and PCA/BIC approaches complete each other, the PCA/BIC is much faster. There are a lot of lessons learned by

IASI-A, comparison CD ON vs CD OFF

CD ON CD OFF

ITSC-XXI Darmstadt, Germany, 29 November - 5 December 2017 16

Page 17: Determination of the experimental error of high spectral ... · The FM and PCA/BIC approaches complete each other, the PCA/BIC is much faster. There are a lot of lessons learned by

IASI-B, comparison CD ON vs CD OFF

CD ON CD OFF

ITSC-XXI Darmstadt, Germany, 29 November - 5 December 2017 17

Page 18: Determination of the experimental error of high spectral ... · The FM and PCA/BIC approaches complete each other, the PCA/BIC is much faster. There are a lot of lessons learned by

Conclusions IASI is spectrally and radiometrically stable. The spectral quality has improved after

switching off the CD mechanism. The FM and PCA/BIC approaches complete each other, the PCA/BIC is much faster. There are a lot of lessons learned by applying the PCA/BIC, which, e.g., are

important for PCA compression, In case of IASI, the normalization to the diagonal of the covariance matrix yields

the worst results, the dimensionality is artificially increased! Normalization tothe diagonal is even worse than not considering normalization at all. The bestway to normalize is to use the correct covariance matrix we have estimated withthe PCA/BIC approach.

In addition, the IASI covariance provided by us should be used when IASI is used forspectroscopy.

We are planning to apply PCA/BIC methodology to AIRS, CrIs, possibly FY-A4 andhopefully ESA-Forum Mission Explorer in the far infrared.

ITSC-XXI Darmstadt, Germany, 29 November - 5 December 2017 18