detection in hplc -...
TRANSCRIPT
Detection in HPLC
Dr. Shulamit Levin, 1
Selecting the Right Detector:Selecting the Right Detector:Types of Detectors in HPLCTypes of Detectors in HPLC
Shulamit LevinShulamit Levin
•UV/VIS•Refractive index •Fluorescence•Electrochemical •Conductivity •Mass-spectrometric (LC/MS) •Evaporative light scattering
•Appendix:•Cutoff of solvents UV•Troubleshooting of RI detector as an example
The Detector is the The Detector is the ““EyeEye”” of the HPLC Systemof the HPLC System
HPLC Columnin Oven
Auto Sampler
Pumpflows 50-5000µL/min)
Fraction Collector
Waste
DetectorControl &Data Processing
1. Fucose2. Galactosamine3. Glucosamine4. Galactose5. Glucose6. Mannose
20.00Minutes5.00
mV
0.00
300
12
34
5
6
a bc d
DetectorsDetectors
UV/VISUV/VISRefractive index Refractive index FluorescenceFluorescenceElectrochemical Electrochemical Conductivity Conductivity MassMass--spectrometric (LC/MS) spectrometric (LC/MS) Evaporative light scatteringEvaporative light scattering
Optical Bench of UVOptical Bench of UV--VIS DetectorVIS Detector
DeuteriumArc Lamp
RotatingDiffraction
Grating190 to 600nm
Taper-CellFlowCell
Beam SplitterMirrors
DualPhotodiode
ApertureSlit
Illumination Lens
Beam-DefiningApparatus
Optical Light Path
Sample side
Reference side
Detection in HPLC
Dr. Shulamit Levin2
Beer's LawBeer's Law
Reduce Pathlength Reduce Concentration
Absorbance = Extinction Coefficient x Pathlength x Concentration
UV ChromophoresUV Chromophores
UV ChromophoresUV Chromophores UVUV--Vis chromophoresVis chromophores
λmax Em x 10-3 @ λmax
Adenine 260.5 E = 13.4Guanine 275 E = 8.1Cytosine 267 E = 6.1Thymine 264.5 E = 7.9Uracil 259.5 E = 8.2NADH 340 E = 6.23NAD 260 E = 18
Detection in HPLC
Dr. Shulamit Levin3
UV spectrum of 10 UV spectrum of 10 nMnM mobile phasemobile phase
AU
AU
0.00
0.10
0.20
0.30
0.40
0.50
0.60
0.70
0.80
0.90
1.00
nm210.00 220.00 230.00 240.00 250.00 260.00 270.00 280.00 290.00
pH 0.94
TFA
AU
0.00
0.10
0.20
0.30
0.40
0.50
0.60
0.70
0.80
0.90
1.00
nm210.00 220.00 230.00 240.00 250.00 260.00 270.00 280.00 290.00
265.1 295.8
pH 2.0
AU
0.00
0.10
0.20
0.30
0.40
0.50
0.60
0.70
0.80
0.90
1.00
nm210.00 220.00 230.00 240.00 250.00 260.00 270.00 280.00 290.00
205.2
pH 2.26
AU
0.00
0.10
0.20
0.30
0.40
0.50
0.60
0.70
0.80
0.90
1.00
nm210.00 220.00 230.00 240.00 250.00 260.00 270.00 280.00 290.00
pH 2.78
Sodium PhosphateFormic Acid
Acetic Acid
U.V. CutU.V. Cut--offs for some Common Solventsoffs for some Common Solvents
SolventSolvent UV CutoffUV Cutoff SolventSolvent UV CutoffUV CutoffWaterWater 180180 N-HeptaneN-Heptane 197197MethanolMethanol 205205 CyclohexaneCyclohexane 200200N-PropanolN-Propanol 205205 Carbon tetrachlorideCarbon tetrachloride 265265AcetonitrileAcetonitrile 190190 ChloroformChloroform 245245THFTHF 225225 BenzeneBenzene 280280AcetoneAcetone 330330 TolueneToluene 285285Methyl acetateMethyl acetate 260260 Methylene chlorideMethylene chloride 232232Ethyl AcetateEthyl Acetate 260260 TetrachloroethyleneTetrachloroethylene 280280NitromethaneNitromethane 380380 1,2-Dichloroethane1,2-Dichloroethane 225225
All wavelengths reported in nm.All wavelengths reported in nm.
Remember Solvents chosen can affect detection!!Remember Solvents chosen can affect detection!!
UV Detection of UV Detection of AccQAccQ--Tag Amino Acid DerivativesTag Amino Acid Derivatives
0.000
0.002
0.004
0.006
0.008
0.010
0.012
0.014
0.016
0.018
0.020
0.022
0.024
15.00 20.00 25.00 30.00 35.00 40.00 45.00 50.00 55.00
AU
Minutes
Aspa
rtic
Acid
Glu
tam
ic A
cid
Hyd
roxy
prol
ine
Ser
ine
Asp
arag
ine
AMQ
Gly
cine
Glu
tam
ine H
istid
ine
NH
3Th
reon
ine
Arg
inin
e
Alan
ine
Pro
line
Alph
a-am
inob
utyr
ic a
cid
Tyro
sine
Cys
teic
Aci
dVa
ine
Mte
hion
ine
Orn
ithin
eLe
ucin
e
Lysi
ne
Phe
nyla
lani
ne
Tryp
toph
anIsol
euci
ne
Diode Array DetectorDiode Array Detector
Detection in HPLC
Dr. Shulamit Levin4
Extraction of 3D DataExtraction of 3D Data
Time
Abso
rban
ce
Chromatogram
λ1
λ2
Wavelength
Abso
rban
ce Spectrum
PDA Spectrum Index PlotPDA Spectrum Index PlotDNPH Derivatives 0.25 ng Each PeakDNPH Derivatives 0.25 ng Each Peak
260.00 260.00280.00 280.00300.00 300.00320.00 320.00340.00 340.00360.00 360.00380.00 380.00400.00 400.00420.00 420.00440.00 440.00
nm
0.00000.0000
0.00020.0002
0.00040.0004
0.00060.0006
AUAU
Millennium PDA Spectrum Index Plot - SampleWeight 0.25 ng - PDA 360.0 nm
CoelutionCoelution of 2 Peaksof 2 Peaks
AU
A
B
Coelution detection at a single wavelength
Coelution is the sum of absorbance of 2 peaks A and B
Peak Purity MeasurementPeak Purity Measurement
2.20 2.40 2.60 2.80 3.00
200.00
250.00
300.00
0.00
0.20
0.40
MaximumImpurity
Spectra at apex and inflection points are displayed
Spectrum at maximum impurity is different
AU
nm
Detection in HPLC
Dr. Shulamit Levin5
Maximum Impurity DetectionMaximum Impurity Detection
260.00 260.00280.00 280.00300.00 300.00320.00 320.00340.00 340.00360.00 360.00380.00 380.00400.00 400.00420.00 420.00440.00 440.00
nm
Minutes
nm
-0.00010 0.00010
0.00000 0.00000
0.00010 0.00010
0.00020 0.00020
0.00030 0.00030
18.40 18.60 18.80 19.00 19.20
AU AU
Millennium PDA Spectrum Index Plot - SampleWeight 0.25 ng 360nm 996PDA 360.0 nm
Hexaldehyde 2,5-Dimethylbenzaldehyde
Coelution of DNPH Hexaldehyde and
2,5-Dimethylbenzaldehyde
Determination of Peak PurityDetermination of Peak Purity
Abs
orba
nce
Time
S t a n d a r d
T im e
U n k n o w n
Peak Purity analyzes all spectra (minimum 15) within a peak against the apex spectrum of the peak itself.
Peak PurityPeak Purity Spectral MatchingSpectral Matching
Spectral match of apex spectrum of the unknown against the apex spectrum of a standard, stored in a user’s library.
Different Spectra Different Spectra –– 53 deg53 deg
200.00 240.00 280.00 320.00nm
EthylPabaEthylparaben
Abs
orba
nce
10 deg of Spectral Contrast10 deg of Spectral Contrast
Similar spectra for structurally related compounds
230.00 250.00 270.00 290.00 310.00nm
TheophyllineDyphylline
Abs
orba
nce
Detection in HPLC
Dr. Shulamit Levin6
Spectral Contrast 0.5 DegreesSpectral Contrast 0.5 Degrees
200.00 240.00 280.00 320.00nm
MethylparabenEthylparaben
Abs
orba
nce
Very similar spectra, CH2 difference
Spectral Contrast can differentiate these spectra
Spectra of nonSpectra of non--UV Active CompoundsUV Active Compounds
210.00 230.00 250.00 270.00 290.00nm
Abs
orba
nce
Analyte and 2 Impurities
DetectorsDetectors
UV/VISUV/VISRefractive index Refractive index FluorescenceFluorescenceElectrochemical Electrochemical Conductivity Conductivity MassMass--spectrometric (LC/MS) spectrometric (LC/MS) Evaporative light scatteringEvaporative light scattering
Refractive Index DetectorRefractive Index Detector
Detection in HPLC
Dr. Shulamit Levin7
Refractive Index DetectorRefractive Index DetectorDifferential Refractive Index DetectorDifferential Refractive Index Detector
LAMP
SLED
R
RS
To Amplifier
No sample = n
With sample = n+∆ν
∆X = Const x ∆n
Sugar Analysis Sugar Analysis
-160.00
-140.00
-120.00
-100.00
-80.00
-60.00
-40.00
-20.00
0.00
20.00
40.00
60.00
80.00
100.00
120.00
5.00 6.00 7.00 8.00 9.00 10.00 11.00
mV
Minutes
Fructose
Dextrose Sucrose
Maltose Lactose
SampleName: Sugars D Vial: 1 Inj: 1 Ch: SATIN Type: Standard
Bagel Extract
Polymer AnalysisPolymer Analysis
0.0050.00
100.00150.00200.00250.00300.00350.00400.00450.00500.00550.00600.00650.00700.00750.00800.00
18.00 20.00 22.00 24.00 26.00 28.00 30.00
MV
Minutes
1260
000
9640
0
5570
SampleName: GPC STDS
Dow 1683
2890
000
1900
00
1030
0
192300
Detection in HPLC
Dr. Shulamit Levin8
LipidsLipids
Del
RIU
5.0 6.0 7.0 8.0Minutes
1 2
250 ng on column 1=Tristearin2=Myristic acid
Styragel HR 0.5, 4.6 x 300 mm, 35°C, 0.35 mL/min
dRI sensitivity = 32X, 32°C
DetectorsDetectors
UV/VISUV/VISRefractive index Refractive index FluorescenceFluorescenceElectrochemical Electrochemical Conductivity Conductivity MassMass--spectrometric (LC/MS) spectrometric (LC/MS) Evaporative light scatteringEvaporative light scattering
Fluorescence ProcessFluorescence Process
Excitation
Emission
Energy Levels
Ground State
Excited States
ExcitationExcitation--Emission SpectraEmission Spectra
λ Maximum of Excitation Spectrum
λ Maximum of Emission Spectrum
Stoke’s λ shift
Lifetime= 10-9 – 10-15 sec
Detection in HPLC
Dr. Shulamit Levin9
Fluorescence DetectorsFluorescence DetectorsExcitation filter
LAMPCell
Emission filter
Photomultiplier
Fluorescence Detector Optical BenchFluorescence Detector Optical Bench
Photom ultiplier tube
Em issionG rating
ExcitationGrating
M irror
Torroidal M irror
Beam Splittter
FlowCell
M irror
Torroidal M irror
Photodiode
ExcitationSlit
Em ission Slit
UV UV vsvs Fluorescence SensitivityFluorescence Sensitivity
Minutes
AMQ
20.00 40.00 60.00
Res
pons
e
AccQ-Tag amino acid analysis
FluorescenceExcitation=250 nmEmission=395 nm
UV 254 nm
DetectorsDetectors
UV/VISUV/VISRefractive index Refractive index FluorescenceFluorescenceElectrochemical Electrochemical Conductivity Conductivity MassMass--spectrometric (LC/MS) spectrometric (LC/MS) Evaporative light scatteringEvaporative light scattering
Detection in HPLC
Dr. Shulamit Levin10
Electrochemical DetectorElectrochemical Detector
Reference Electrode
Working Electrode
Electrolyte (mobile phase)
Auxiliary Electrode
Analyte is oxidized or reduced
+ -
As compounds are oxidized or reduced, a current proportional to concentration is produced.
Electrochemical Detection of Electrochemical Detection of CatecholaminesCatecholamines & Related Compounds & Related Compounds
1. Norepinepherine 150 ppb2. Epinepherine 200 ppb3. Normetanepherine 50 ppb4. Dopamine 200 ppb5. Metanepherine 200 ppb6. 3-Methoxytyramine 75 ppb7. 4-Methoxytyramine 500 ppb
2.00 4.00 6.00 8.00 10.00 12.00
Minutes
0.00
nAmps
1 2
3
45
6
7
Pulsed Pulsed AmperometricAmperometric Detection of Detection of MonosaccharidesMonosaccharides
1. Fucose2. Galactosamine3. Glucosamine4. Galactose5. Glucose6. Mannose
20.00Minutes5.00
mV
0.00
300
12
34
5
6
DetectorsDetectors
UV/VISUV/VISRefractive index Refractive index FluorescenceFluorescenceElectrochemical Electrochemical Conductivity Conductivity MassMass--spectrometric (LC/MS) spectrometric (LC/MS) Evaporative light scatteringEvaporative light scattering
Detection in HPLC
Dr. Shulamit Levin11
Conductivity DetectorConductivity Detector
M o bile ph ase
M o bile p hase p lu s sam ple
++ --
Conductivity DetectorConductivity Detector
Mobile phase
Mobile phase plus sample
Anion Analysis by ICAnion Analysis by IC
0.70
1.05
1.40
µS
0.00 5.00 10.00 15.00 20.00 25.00Minutes
1
2
3
45
6 7
1. Fluoride2. Chloride3. Nitrite4. Bromide5. Nitrate6. Phosphate7. Sulfate
1 ppm2 ppm4 ppm4 ppm4 ppm6 ppm4 ppm
Column:Eluent:Flow rate:Injection vol.:Detection:
Waters IC-Pak Anion HCBorate/Gluconate2.0 mL/min100 µLDirect Conductivity
Anion analysis by ICAnion analysis by IC
µS
0.00
0.40
0.80
1.20
1.60 1 23
4 5
67
0.00 4.00 8.00 12.00 16.00 20.00 24.00Minutes
0.00
0.01
0.02
0.03
0.04
0.05
AU
3
4
5 Column:Eluent:
Flow rate:Injection vol.:
Waters IC-Pak Anion HR1.2 mM Sodium Carbonate/1.2 mM Sodium Bicarbonate1.0 mL/min50 µL
Detection: Direct Conductivity after Suppression
Detection: UV (PDA) at 214 nm
1. Fluoride2. Chloride3. Nitrite4. Bromide5. Nitrate6. Phosphate7. Sulfate
1 ppm2 ppm4 ppm4 ppm4 ppm6 ppm4 ppm
Detection in HPLC
Dr. Shulamit Levin12
ApplicationsApplications
Sensitivities for compounds such as phenol, catecholamines, nitrosamines, and organic acids are in the picomole (nanogram) range.
The mobile phase must be made electrically conductive, usually by the addition of a suitable salt:
Ion Exchange
Reversed Phase and Ion-Pair RP
No normal phase separations
DetectorsDetectors
UV/VISUV/VISRefractive index Refractive index FluorescenceFluorescenceElectrochemical Electrochemical Conductivity Conductivity MassMass--spectrometric (LC/MS) spectrometric (LC/MS) Evaporative light scatteringEvaporative light scattering
DATA SYSTEM
DETECTION OF IONS
+
+
SAMPLE DESOLVATIONAND IONIZATION
LC/MSINTERFACE
SOURCE
ANALYZERION DETECTOR
HPLCMASS SPECTRUM
SORTING OF IONS
+
++
++ ++ ++
-+
+
+ +
++
-
+
+
+
CHROMATOGRAM
Typical LC/MS System ProgressionTypical LC/MS System Progression
DATA SYSTEM
DETECTION OF IONS
+
+
SAMPLE DESOLVATIONAND IONIZATION
LC/MSINTERFACE
SOURCE
ANALYZERION DETECTOR
HPLCMASS SPECTRUM
SORTING OF IONS
+
++
++ ++ ++
-+
+
+ +
++
-
+
+
+
CHROMATOGRAM
Typical LC/MS System ProgressionTypical LC/MS System Progression
Detection in HPLC
Dr. Shulamit Levin13
Transition from LC to MSTransition from LC to MS
• State of Matter: LiquidLiquid to GasGas
• Charge State: “NeutralNeutral” to IonIon
• Pressure: 760 760 torrtorr to 1010--55 to 10to 10--88 torrtorr
APCI MechanismAPCI Mechanism
Ionization produces solvent ions
The solvent ions reactwith analyte moleculesforming clusters
Corona Needle
X = Solvent Molecules e.g.H2O, MeCNM = Sample Molecule
Heated Nebulizer
xx
xH+ M
xx
xH+
Mx
xM
MH+
xx
xx
XH+x
xH+
MH+
ElectrosprayElectrospray IonizationIonization Positive or Negative?Positive or Negative?
Basic Compounds (-NH2) (M+H)+
Acidic Compounds (-CO2H, -OH) (M-H)-
Detection in HPLC
Dr. Shulamit Levin14
Recognizing Multiply Charged Ions
Mass spectrometers operate on the basis of mass-to-charge ratio (m/z). Mass assignments are normally made assuming a single charge per ion (i.e. m/z = m)
Single charge Mass = (M+H)
Double charge Mass = 1/2 (M+2H)
n charge Mass = 1/n (M+nH)
Isotopes of doubly charged ions are separated by 0.5 Da
n = 20
n = 22
n = 18
n = 16, m/z = 1060
n = 23, m/z = 738
n = 21
n = 19
n = 17
Horse Heart Myoglobin
Mass RangeMass RangeMultiply Charged MoleculesMultiply Charged Molecules
Calculated MassAcquired Mass range
Hemoglobin SpectrumHemoglobin SpectrumPresence of More Than One Charged EnvelopePresence of More Than One Charged Envelope
1000 1050 1100 1150 1200 1250 1300 1350 1400m/z0
100
%
1081.60
1009.36
1058.83
1164.52
1133.92
1261.64
1221.181376.08
1323.14
DeconvolutionDeconvolution by by MaxEntMaxEntHemoglobinHemoglobin
15000 15100 15200 15300 15400 15500 15600 15700 15800 15900 16000 16100mass0
100
%
15125.0
15857.0
15149.0
15866.0
Detection in HPLC
Dr. Shulamit Levin15
Multiply Charged Ions – How Many Charges?
(M+H)+
(M+2H)2+
1 Da
0.5 Da
DATA SYSTEM
DETECTION OF IONS
+
+
SAMPLE DESOLVATIONAND IONIZATION
LC/MSINTERFACE
SOURCE
ANALYZERION DETECTOR
HPLCMASS SPECTRUM
SORTING OF IONS
+
++
++ ++ ++
-+
+
+ +
++
-
+
+
+
CHROMATOGRAM
Typical LC/MS System ProgressionTypical LC/MS System Progression
Mass Mass SpectrometerSpectrometer’’ss
AnalyzersAnalyzers
T im e O f F lig h t M a s s A n a ly ze rsT im e O f F lig h t M a s s A n a ly ze rs
S O U R C E
D E T E C T O RR E F L E C T R O N M O D E
R E F L E C T R O N O ND E T E C T O RL IN E A R M O D E
D R IF T T U B E
S O U R C E
D E T E C T O RR E F L E C T R O N M O D E
R E F L E C T R O N O F F D E T E C T O RL IN E A R M O D ED R IF T T U B E
T im e O f F lig h t M a s s A n a ly ze rsT im e O f F lig h t M a s s A n a ly ze rs
S O U R C E
D E T E C T O RR E F L E C T R O N M O D E
R E F L E C T R O N O ND E T E C T O RL IN E A R M O D E
D R IF T T U B E
S O U R C E
D E T E C T O RR E F L E C T R O N M O D E
R E F L E C T R O N O F F D E T E C T O RL IN E A R M O D ED R IF T T U B E
199
FTFT--ICRICR--SpectrometerSpectrometer
DCDC
DCSource
Filament
Transferoptic
Trapping Plates
Transmitter Plates
Receiver Plates
Sender
Elektroden Electrodes
Y
ZX
Magnetic Fielt B
Starting with the Starting with the quadrupolequadrupole
Source
DetectorNonresonant Ion
Resonant Ion
dc and Rf voltages
V(t) = - Vdc - Vrf cosΩt
V(t) = Vdc + Vrf cosΩt
190
Electron Multiplier
Inlet
Ring Electrode, Rf
End Cap Electrode
Axial Modulation
IonIon TrapsTraps
+ + ++++ ++
IonSource
Slit
Magnetic sectorElectrostatic Sector
(ESA)
Detector
SlitNier-Johnson-Geometry (EB) DATA SYSTEM
DETECTION OF IONS
+
+
SAMPLE DESOLVATIONAND IONIZATION
LC/MSINTERFACE
SOURCE
ANALYZERION DETECTOR
HPLCMASS SPECTRUM
SORTING OF IONS
+
++
++ ++ ++
-+
+
+ +
++
-
+
+
+
CHROMATOGRAM
Typical LC/MS System ProgressionTypical LC/MS System Progression
Detection in HPLC
Dr. Shulamit Levin16
MS DetectorsMS Detectors
++ ++ -
-
-Electron Multiplier(voltage setting lower than Dynode)
Conversion Dynode (Voltage 1- 20 kV)
Current is measured
+ +Mass Analyser
dynodedynode
phosphorphosphorphotomultiplierphotomultiplier
Photomultiplier
Electron Multiplier
Mass Spectrometer 3D RunMass Spectrometer 3D Run
Mixture
2.00 4.00 6.00 8.00 10.00Time0
Int.1: Mass Chromatogram
5.054.65
3.820.74
8.62
5.65
8.02
10.62
Total-Ion-Current MS Chromatogramwith poor resolution
Mix
60 80 100 120 140 160 180 200 220 240 260 280 300 320 340m/z0
100
%
(10.696) 1: Scan ES+ 4.34e5262.87
59.99
213.90
195.98120.8068.92
98.8576.87 128.82
170.92
222.87
235.87
240.88
263.87
264.85
267.91 287.01 309.02333.84
Mixture
2.00 4.00 6.00 8.00 10.00Time0
Abs1: Diode Array
5.054.65
3.820.74
8.62
5.65
8.02
10.62
200210220230240250260270280290300310nm0
Int. 210.00
246.00
294.00
UV-Diode Array Chromatogramwith poor resolution
Mix
2.00 4.00 6.00 8.00 10.00 12.00 14.00Time0
100
%
1: Scan ES+ 262.874.59e5
10.60
Mixture
2.00 4.00 6.00 8.00 10.00 Time0
Int. 5.054.65
3.820.74
8.625.65
8.0210.62
Selectivity of Mass Spectrometer DetectorSelectivity of Mass Spectrometer Detector
Extracted Ion Chromatogramof a single Component from
a mixture of Components
Mix
60 80 100 120 140 160 180 200 220 240 260 280 300 320 340m/z0
100
%
(10.696) 1: Scan ES+ 4.34e5262.87
59.99
213.90
195.98120.8068.92 98.85
76.87 128.82 170.92
222.87
235.87
240.88
263.87
264.85
267.91 287.01 309.02333.84
LC-MS Analysis
XTerra™ MS C18, 2.1 x 50 mm ( 5 µm)
10 µL injection of 200 ng/mL sample (in 40% MeOH),1=Propranolol, 2=Doxepin, 3=Nortriptyline, 4=Trimipramine, 65/35 0.1 % Formic Acid / MeCN0.2 mL/min
Time (min)
1.00 2.00 3.00 4.00 5.000
100
0
100
0
100
0
100
0
100 2958.11e4
2.56
2802.21e5
1.57
2641.26e5
2.16
2601.53e5
1.29
Total Ion Chromatogram3.50e5
1.571.29 2.16
2.56
100 125 150 175 200 225 250 275 300Mass/Charge (m/z)
0
100
0
100
100
0
1004.59e4295
296
1.13e5280
281
7.67e4264
233 265
9.25e4260
261
0O
N
N
N
NH
O NHOH
(1)
(2)
(3)
(4)
Ding
Detection in HPLC
Dr. Shulamit Levin17
Triple Triple QuadrupolesQuadrupoles: MS: MS--MS ModesMS Modes
Multiple Reaction Monitoring
Typically used in Quantitative Work of
Triple Quadrupoles
Multiple Reaction Monitoring
Typically used in Quantitative Work of
Triple Quadrupoles
MS1 Collision
Cell
MS2
StaticStatic
Multiple Reaction Monitoring
Typically used in Quantitative Work of
Triple Quadrupoles
Multiple Reaction Monitoring
Typically used in Quantitative Work of
Triple Quadrupoles
MS1 Collision
Cell
MS2
StaticStatic
Daughter (Product) Ion SpectraDaughter (Product) Ion Spectra
MS1 CollisionCell
MS2
ScanningStatic
Daughter (Product) Ion SpectraDaughter (Product) Ion Spectra
MS1 CollisionCell
MS2
ScanningStatic
Parent (Precursor) Ion SpectraParent (Precursor) Ion Spectra
MS1 CollisionCell
MS2
StaticScanning
Parent (Precursor) Ion SpectraParent (Precursor) Ion Spectra
MS1 CollisionCell
MS2
StaticScanning
Constant Neutral Loss SpectraConstant Neutral Loss Spectra
MS1 CollisionCell
MS2
ScanningScanning
Constant Neutral Loss SpectraConstant Neutral Loss Spectra
MS1 CollisionCell
MS2
ScanningScanning
0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00 7.50 8.00
Time
0
100
%
0
100
%
0
100
%
0
100
%
0
100
%
0
100
%
Std_017 MRM of 12 Channels ES+ 208 > 162.9
1.73e75.10
Std_017 MRM of 12 Channels ES+ 149.95 > 91
1.49e73.80
Std_017 MRM of 12 Channels ES+ 135.8 > 91
4.29e62.95
Std_017 MRM of 12 Channels ES+ 165.85 > 148
2.05e72.25
Std_017 MRM of 12 Channels ES+ 179.8 > 163
4.13e63.20
Std_017 MRM of 12 Channels ES+ 193.65 > 162.7
6.48e63.95
MDEA
Ephedrine
MDMA
MDA
Amphetamine
Methamphetamine
Typical Quantitative Analysis Using Triple Quadrupoles:Simultaneous MRM analysis of 6 amphetamines
Highly specific and sensitive chromatograms
DetectorsDetectors
UV/VISUV/VISRefractive index Refractive index FluorescenceFluorescenceElectrochemical Electrochemical Conductivity Conductivity MassMass--spectrometric (LC/MS) spectrometric (LC/MS) Evaporative light scatteringEvaporative light scattering
Evaporative Light Scattering Evaporative Light Scattering -- ELSELS
Nebulization Desolvation
DetectionTo detector cell
Nebulizer
Lamp
Detection in HPLC
Dr. Shulamit Levin18
Rayleigh Scattering Rayleigh Scattering –– Why the Sky is blueWhy the Sky is blue
• Scattering is independent of the particle’s chemical properties, where:– N = # of particles– α = Polarizability i.e. the sum of the dipoles of all the molecules in the particle.
For a homogeneous particle this is proportional to the particle volume. – R = Distance of observer from scatterer– Dependence on wavelength of incident light, shorter wavelengths produce
greater scattering
λ4R2
I = I0 8 π4 Nα2 (1 +cos2θ)
Scattering ModelsScattering Models
If D/λ< 0.1then I = f (D6)
0.1<D/λ< 10then I = f (D4)
D/λ>10then I = f (D2)
I = Intensity of the scattered lightλ = wavelength of the light
Scattering is dependent on particle size “D” Increasing particle size
D α C1/3
(Often see solute density)I α (Cb)
With 2>b>2/32 is the limiting value for Rayleigh
symmetrical scattering
Depends on which type of scattering is predominant
Non-linear mass detectoruse chromatography data software
quadratic curve or log/log curve to fit calibration curve
ELSD ELSD vsvs UVUV ELSD ELSD vsvs RIRI
Detection in HPLC
Dr. Shulamit Levin19
ELSD Used with Other DetectorsELSD Used with Other Detectors
ELSD to monitor all compounds and determine purity levels
Diode Array TIC
Min
ES+TIC
ELSD
24.5 7
Diode array TIC
PDA to monitor UV/Vis friendly compounds
Mass Spec to verify that compound has been synthesized
Not a Universal DetectorNot a Universal DetectorTypically Used with Other DetectorsTypically Used with Other Detectors
UV TIC
MS ES+ TIC
ELSD
Erythromycins Separation
Min1 3 5
ELSD
Diode Array TIC
See NonSee Non--UV Absorbing CompoundsUV Absorbing Compounds See Your Peaks FasterSee Your Peaks FasterUse of Gradients Versus IsocraticUse of Gradients Versus Isocratic
RI Detection
ELSD Detection
Detection in HPLC
20
Estrogen analogues and saltm
V
0.00
10.00
20.00
30.00
40.00
50.00
60.00
70.00
80.00
90.00
100.00
110.00
Minutes0.10 0.20 0.30 0.40 0.50 0.60 0.70 0.80 0.90 1.00 1.10 1.20 1.30 1.40 1.50 1.60 1.70 1.80 1.90 2.00
NaC
l-0.
207
Estr
adio
l-0.
849
Estr
iol-
1.21
0
Evaporative LightEvaporative Light--scattering Detectorscattering Detector